A-100 Series
Peltier/Sipper System
User’s Manual
All right reserved © 2008 Version:1.0.1
Contents
Preface…………………………………………………..…………………………………………1
1. Safety………………………………………….………………………………………1
2. About the Instrument…………………………..……………………………………1
3. Symbol………………………..…………………………………………………………1
Chapter II Summary…………………………………………………………………………2
1. Working principle and structure………………………………………………………2
2. Main Specifications…………………………………………………………………2
Chapter III Install Instrument………………………………………………………………3
1. Environment Required………………………………………………………………3
2. Voltage Confirmation…………………………………………………………………3
3. Installation……………………………………………………………………………3
Chapter IV
1. Introduction………………………………………..……………………………………5
2. Operation Panel……………………………..…………………………………………5
3. Keypad Description…………………….………………………………………………5
Chapter V
1. Software Structure………………….…….……………………………………………7
2. Basic Operation……………….…….………………………………………………7
Chapter VI
1. Daily Maintenance……………………………………………………………………9
Appendix :Ⅰ Consumable parts……………………….………………………………………10
Appendix :Ⅱ Packing List………………….…..………………….……………………………10
Introduction of the Instrument…
Instrument Operation
1) Temperature Control………………………………………………………………7
2) Automatic Sipper………………...…………………………………………………8
Instrument Maintenance
1) Surface Clean………………………………………………………………………9
2) Clean the Silica Gel Pipe…………..………………………………………………9
………………………………………………………7
……………………………..………………………9
……………………………………………5
I
- 1 -
PREFACE
1.
. Safety
..
This instruments are designed according to the following safety documents:《IEC
61010-1:2001》(Which is :《Measurement, Control and Laboratory Electric Equipments
Safety Requirement》Part 1) and 《IEC 60601 — 1:1988》(Which is:《Electric Apparatus
of Medical Treatment》 Part 1).
2.
. About the Instruments
..
A-100 Series Peltier/Sipper System is an electrical test instruments which is widely
used in the laboratories.
Use Frequency: Intermittence
Excessive Voltage(Current): No
Pollution Class: Class 1
3.
. Symbols
..
Caution, Danger!
Caution, Danger!
Caution, Hot!
Ground
Fuse
Recall, this instrument will be called back by the
appointed Electrical Treatment Department or by the
original Manufacturer when wasted.
- 2 -
Series A-100 Peltier/Sipper System
Chapter I Summary
A-100 Series Peltier/Sipper System (A-100: Peltier/Sipper System; A-101: Sipper Unit;
A-102: Ambient Peltier Unit), as a spectrophotometer matching accessories, with high
stability, flexible and accurate. It can be widely used in chemical, pharmaceutical,
biological and chemical, metallurgy, light industry, textiles, materials, environmental
protection, medical laboratories and education sectors. It is a conventional laboratory
preparation of apparatus.
1.
. Working principle and structure
..
Peltier temperature controller using the device as temperature control element. By
controlling the current size to raise or lower the temperature. With a fast response and
accurate temperature control characteristics. Autosampler selected stepper motor driven
vacuum pump drawing samples. With a flow rate of a balanced and accurate
characteristics of injection.
2.
. Main Specifications
..
Specifications for A-100 Series Peltier/Sipper System:
Temp. Range: 15—40℃(Environment Temp.:22℃)
Temp. Accuracy: ±0.2℃
Temp. Display: 0.1℃
Sipper Sample Volume: 1、2、3、4、5、7.5、10、12.5、15ml/min
Sipper Sample Accuracy: ±5%
Display: 128×64 dot Matrix LCD
Dimensions: 270×170×115mm
Weight: 3kg