Jenway A-100, A-101, A-102 Instruction Manual

Page 1
A-100 Series
Peltier/Sipper System
User’s Manual
Page 2
Page 3
Contents
Preface…………………………………………………..…………………………………………1
1. Safety………………………………………….………………………………………1
2. About the Instrument…………………………..……………………………………1
3. Symbol………………………..…………………………………………………………1
Chapter II Summary…………………………………………………………………………2
1. Working principle and structure………………………………………………………2
2. Main Specifications…………………………………………………………………2
Chapter III Install Instrument………………………………………………………………3
1. Environment Required………………………………………………………………3
2. Voltage Confirmation…………………………………………………………………3
3. Installation……………………………………………………………………………3
Chapter IV
1. Introduction………………………………………..……………………………………5
2. Operation Panel……………………………..…………………………………………5
3. Keypad Description…………………….………………………………………………5
Chapter V
1. Software Structure………………….…….……………………………………………7
2. Basic Operation……………….…….………………………………………………7
Chapter VI
1. Daily Maintenance……………………………………………………………………9
Appendix :Ⅰ Consumable parts……………………….………………………………………10 Appendix :Ⅱ Packing List………………….…..………………….……………………………10
Introduction of the Instrument…
Instrument Operation
1) Temperature Control………………………………………………………………7
2) Automatic Sipper………………...…………………………………………………8
Instrument Maintenance
1) Surface Clean………………………………………………………………………9
2) Clean the Silica Gel Pipe…………..………………………………………………9
………………………………………………………7
……………………………..………………………9
……………………………………………5
I
Page 4
Page 5
- 1 -
PREFACE
1
Safety
This instruments are designed according to the following safety documents:IEC 61010-12001》(Which is :Measurement, Control and Laboratory Electric Equipments Safety Requirement》Part 1) and 《IEC 60601 — 1:1988》(Which is:《Electric Apparatus of Medical Treatment Part 1).
2
About the Instruments
A-100 Series Peltier/Sipper System is an electrical test instruments which is widely used in the laboratories.
Use Frequency: Intermittence
Excessive Voltage(Current): No
Pollution Class: Class 1
3
Symbols
Caution, Danger!
Caution, Danger!
Caution, Hot!
Ground
Fuse
Recall, this instrument will be called back by the appointed Electrical Treatment Department or by the original Manufacturer when wasted.
Page 6
- 2 -
Series A-100 Peltier/Sipper System
Chapter I Summary
A-100 Series Peltier/Sipper System (A-100: Peltier/Sipper System; A-101: Sipper Unit; A-102: Ambient Peltier Unit), as a spectrophotometer matching accessories, with high stability, flexible and accurate. It can be widely used in chemical, pharmaceutical, biological and chemical, metallurgy, light industry, textiles, materials, environmental protection, medical laboratories and education sectors. It is a conventional laboratory preparation of apparatus.
1
Working principle and structure
Peltier temperature controller using the device as temperature control element. By controlling the current size to raise or lower the temperature. With a fast response and accurate temperature control characteristics. Autosampler selected stepper motor driven vacuum pump drawing samples. With a flow rate of a balanced and accurate characteristics of injection.
2
Main Specifications
Specifications for A-100 Series Peltier/Sipper System
 Temp. Range: 15—40(Environment Temp.:22)  Temp. Accuracy: ±0.2  Temp. Display: 0.1  Sipper Sample Volume 123457.51012.515ml/min  Sipper Sample Accuracy: ±5%  Display 128×64 dot Matrix LCD  Dimensions 270×170×115mm  Weight 3kg
Page 7
- 3 -
Chapter II Install Instrument
1
Environment Required
To ensure the best performance, A-100 Series Automatic Peltier/Sipper System has its special design to prevent the negative influence from environment. Before installation, please confirm the following conditions:
The best working temperature range is 16-35 and the humidity is 45-80%. Keep it as far as possible away from the strong magnetic or electrical fields or
any electrical device that may generate high-frequency fields.
Set the unit up in an area that is free of dust, corrosive gases and strong
vibrations.
Remove any obstructions or materials that could hinder the flow of air under and
around the instrument.
The power requirement is 220±22V@50±1Hz or 110±11V@60±1 Hz. Use the appropriate power cord and plug into a grounded outlet. Be away from direct sunlight Be away from dust.
2
Voltage Confirmation
Be sure to set the instrument’s voltage switch at your local power
supply, or severe damage may occur!
3
Installation
Step 1:: Check the packing list
Unpack the contents; check the materials with the Packing List. Any damage or
Lost found, please contact us or the local dealer.
Step 2:: Position
Place the instrument on the stable table carefully.
Step 3:: Install Holder
Open the sample room lid of Spectrophotometer, unscrew the four screws in conjunction 4-cell holder and the rod, and then remove them, assemble the Constant Temperature Micro Cell Holder in sample room at the same location using the 2 screws as Fig.2-1. Then connect the plug with the power supply.
Note: A101, just use Micro Cell Holder
Page 8
- 4 -
Series A-100 Peltier/Sipper System
Fig. 2-1
Step 4:: Link the power cord(except A-101)
Make sure the instrument’s power switch is in the Off condition, link the power
cord to the instrument as Fig.2-2
Fig. 2-2
Step 5:: Link the sample silica gel pipe to spectrophotometer(Except A-102))
Check again; make sure that the instrument power switch is in the Off condition,
Page 9
- 5 -
link the sample silica gel pipe as Fig.2-3, open the pump holder, insert the silica gel pipe like No.1 in the Fig.2-3, then lock the holder like No.2. Connect the Right gel pipe with the Flow Through Cell through the hole of No.3, and connect the Left gel pipe with the Flow Through Cell through the hole of No.4. Screw the shaft accordingly like Fig.2-4.
2-3
Fig.2-4
Step 1:: Install Power Cord
Check instrument to make sure the power switch is in Off condition. Connect the two sides of power line to instrument and the power outlet.
Step 2:: Connected to power
Check again and then switch on, the instrument can be used after self-checking.
Page 10
- 6 -
Series A-100 Peltier/Sipper System
Page 11
- 7 -
Chapter III Introduction of the Instrument
1
Introduction
Appearance of the instrument (Fig.3-1):
1 — Operation Panel 2 — Pump 3 — Temperature Control Port 4 — Power Socket 5 — Power Switch
2
Operation Panel
Operation panel of Series A-100( Fig.3-2)
Fig.3-1
Page 12
- 8 -
Series A-100 Peltier/Sipper System
1 — LCD 2 — Keypad
3
Keypad Description
Fig.3-2
Start/Stop
Set Parameter
Cancel operation
Scroll the menu or data
Page 13
- 9 -
Chapter IV Instrument Operation
1
Software Structure
Software structure of Series (Fig.4-1)
Main Menu
2
Basic Operation
1)
1) Temperature Control
1)1)
Setting Interface
Fig.4-1
Temperature Switch
Temperature Setting
Sipper Setting
In the main menu, press to enter interfaceFig.4-2), press
to choose ”Set Heater”, then press to enter interfaceFig.4-3),
press to choose “On” to enter temperature control function, then
press to confirm and return
Fig. 4-2
Fig.4-3
Page 14
- 10 -
Series A-100 Peltier/Sipper System
Press to choose “Set Temp”. Then press to enter “Set
Temp” (Fig.4-4),press to add/reduce temperature,press
to confirm and return.
Fig.4-4
Press to return Main Menu, the instrument will start testing after transferred to the set temperature.
2)
2) Automatic Sipper
2)2)
In the main menu, press to enter; press to choose“Set
Sipper” then press to set sipper speed(Fig.4-5), press
to choose relative speed, press to confirm and return.
Fig.4-5
Press to return main menu, press to start sipper, press
again to stop.
Page 15
- 11 -
Chapter V Instrument Maintenance
To keep the instrument work in good condition, constant maintain is needed.
1
Daily Maintain
1)
1) Surface Clean
1)1)
The cover of the instrument is with paint. Please use wet towel to wipe off the drips on the surface immediately. Organic solution is forbidden to be used to clean the cover. Please wipe off the dirt on the cover timely.
2)
2) Clean the Silica Gel Pipe
2)2)
Clean the sipper silica gel pipe after testing or changing the sample solution to avoid being corroded or polluted.
Page 16
- 12 -
Series A-100 Peltier/Sipper System
Appendix ⅠⅠ ::Consumable parts
Name & Specifications Order No.
Sipper silica gel pipe 900111 10mm Masking Con. Flowthrough G.Cell 916132 10mm Masking Con. Flowthrough Q.Cell 916142
Page 17
Appendix
Packing List
1) A-100: Peltier/Sipper System
Constant Temperature and Automatic Sipper System 1set Constant Temperature Sample Holder 1pc Silica Gel Pipe 1pc Connection Line 1pc Power Cord 1pc User’s Manual 1set
2) A-101: Sipper Unit;
Constant Temperature and Automatic Sipper System 1set Silica Gel Pipe 1pc Power Cord 1pc User’s Manual 1set
3) A-102: Ambient Peltier Unit)
Constant Temperature System 1set Constant Temperature Sample Holder 1pc Connection Line 1pc Power Cord 1pc User’s Manual 1set