Jenway 73 User Manual

73 Series Spectrophotometer
Accessory fitting and operation manual
061 672
Safety
Please read this information carefully prior to installing or using this equipment.
1. The unit described in this manual is designed be operated only by trained personnel. Any adjustments,
maintenance and repair must be carried out as defined in this manual, by a person qualified to be aware of the hazards involved.
2. It is essential that both operating and service personnel employ a safe system of work, in addition to the detailed instructions specified in this manual.
3. Other than for those items defined in the maintenance procedures herein there are no user serviceable items in this instrument. Removal of covers and attempted adjustment or service by unqualified personnel will invalidate the warranty and may incur additional charges for repair.
4. References should always be made to the Health and Safety data supplied with any chemicals used. Generally accepted laboratory procedures for safe handling of chemicals should be employed.
5. If it is suspected that safety protection has been impaired in any way, the unit must be made inoperative and secured against any intended operation. The fault condition should immediately be reported to the appropriate servicing authority.
2
SECTION 1 – Accessories and Spare Parts
1.1 OPTIONAL ACCESSORIES
Part Code Description of Accessory
660 101 Internal printer
735 401 Automatic 8 cell turret
735 201 Sipper pump
735 301 Peltier
735 701 Combined sipper peltier pump
735 801 10x10mm path length cuvette holder
735 901 16/24mm test tube holder
736 001 10 x 100mm path length cuvette holder
736 101 10x10 (70µl cell holder)
736 201 Water heated 10x10 single cell holder
735 601 Boiling tube holder
736 301 Film holder
035 088 Visible calibration set
035 091 UV/Visible calibration set
060 422 Moulded cuvette rack for 16 10x10mm cuvettes
735 001 Dust cover
019 146 2GB USB memory sticks for external memory
037 551 RS232 to USB converter for use with computer without a serial port
1.2 CONNECTING THE ACCESSORIES
There are two types of accessories which can be fitted in the sample chamber – passive or active accessories. The range of passive accessories includes 10 x 10mm single cuvette holders, single water heated cuvette holders, adjustable path length (10 to 100 mm) cuvette holders, test tube holders, boiling tube holders, film holders and micro-cuvette holders. The range of active accessories includes an automated 8 cell changer, sipper pump, peltier and combined peltier sipper pump. The instrument must be turned off before any accessories are fitted.
1.2.1 INTERNAL PRINTER
Clips
Use a small screw driver to lift the blanking panel on the top of the instrument. Squeeze the two clips in order to remove the blanking panel. Disconnect the printer wires which are secured to the underside of the blanking plate.
Unpack the printer from the packaging. Turn the printer upside down and connect the printer wires by clipping into the connector on the printer.
3
1.2.2 PASSIVE ACCESSORIES
Thumb screw
Squeeze the grey plastic clips together so that the printer top opens. Slot the printer into the top of the instrument and push down until it fits flush to all four sides.
Insert the paper roll into the printer – ensuring that there is some paper sticking out of the printer before clicking the grey plastic back into place. Switch the instrument on. The power and error lights on the printer will flash. Once the instrument power on tests are complete press the feed button to check that the paper is fed correctly.
Unscrew the thumb screw to undo the passive accessory. Lift out the passive accessory. To fit a different passive accessory simply place the accessory in the correct orientation, align the thumb screw and tighten to fix in place.
To replace the passive accessory with an active accessory refer to section 1.2.3.
1.2.3 ACTIVE ACCESSORIES
1.
Thumb screw
3.
Unscrew the thumb screw to undo the passive accessory. Lift out the passive accessory. To fit an active accessory unscrew screws 1 to 4 and lift out the metal base plate.
2.
4.
This will expose the bottom of the sample chamber with the power supply connection needed to operate the active accessories.
4
1.2.3.1 Automatic 8 cell turret
Take the 8 cell turret base plate. Connect the power supply in the bottom of the sample chamber to the connector on the underside of the baseplate. Place the base plate in the sample chamber. Replace screws 1 to 4.
Take the 8 cell carousel and place on top of the motor, taking care to align the three ball bearings with the grooves on the motor shaft. Gently push the carousel down onto the motor shaft until it is located into place. Gently rotate the carousel until there is some resistance. The carousel is now in the correct position.
If the fitting is too tight use a small screw driver to loosen the ball bearings before pushing the carousel down onto the shaft.
1.2.3.2 Peltier
5.
6.
For this accessory as well as removing the passive accessory base plate, the front panel of the instrument must also be removed. Loosen screws 5 and 6 until the front panel can be lifted out in the forwards direction.
Take the peltier base plate. Connect the power supply in the bottom of the sample chamber to the connector on the underside of the base plate. Place the base plate in the sample chamber. Replace screws 1 to 4. Take the peltier front panel and slot into place before retightening screws 5 and 6.
When the accessory is fitted the instrument will look like this.
5
1.2.3.3 Sipper pump
Bi directional flow A (sipping)
6.
5.
For this accessory as well as removing the passive accessory base plate, the front panel of the instrument must also be removed. Loosen screws 5 and 6 until the front panel can be lifted out in a forward direction.
Take the sipper base plate. Connect the power supply in the bottom of the sample chamber to the connector on the underside of the base plate. Place the base plate in the sample chamber. Replace screws 1 to 4. Take the sipper front panel and slot into place before re-tightening screws 5 and 6.
Bi directional flow B (pumping)
Continuous flow
For sipping:
6.
9.
5.
3.
2.
8.
7.
4.
1.
The tubing should be connected depending on the function that the sipper pump is going to perform. All tubing must be kept as short as possible and the tubing must not be allowed to obstruct the ligth path.
1. Connect the sipper pump tubing to the outlet port on the flow-through cuvette.
2. Secure the tubing using the clip on the righthand side of the pump head.
3. Ease the tubing round the rollers by carefully rotating them clockwise, by hand. Clamp the tubing into the clip on the left hand side of the motor.
4. Once secured, ensure the tubing is routed into the two retaining clips located on the base plate at the side of the pump head.
5. Cut the tubing at the point where it fits comfortably onto the left hand tube located on the inside of the front bulk head.
6. Connect a suitable length of this tubing to the external waste pipe.
6
For pumping:
7. Cut a small length of the sipper pump tube and push
this over one end of the capillary tube. Connect this to the inlet port of the flow-through cuvette.
8. Route the tube into the two retaining clips located on the base plate at the side of the pump head.
9. Fit the sipper probe and secure using the thumbscrew. Feed the capillary tubing through the tube and up through the sipper probe, allowing sufficient length for it to pass into a suitable receptacle.
10.
7.
1.
9.
6. 5. 8. 2.
3.
4.
1. Cut two pieces of sipper pump tubing approximately 300mm in length. Take one length of tubing and fit this to the pump head, as shown, securing the tubing using the clip on the right hand side of the pump head.
2. Ease the tubing round the rollers carefully rotating them clockwise, by hand. Clamp the tubing into the clip on the left hand side of the motor.
3. Fit the other end onto the inlet port on the flow­through cuvette.
4. Fit the second 300mm length of tubing to the outlet port of the flow-through cuvette. Once secured, ensure the tubing is routed into the two retaining clips located on the base plate at the side of the pump head.
5. Fit the other end of the tubing onto the outlet port, located on the inside of the front bulkhead.
6. Connect a suitable length of sipper pump tubing to the external outlet port.
7. Insert one end of the capillary tube into the sipper pump tubing, as shown.
8. Feed the other end through the inlet port located on the inside of the bulkhead.
9. Fit the sipper probe and secure using the thumbscrew.
10. Carefully feed the tubing through the sipper probe,
allowing sufficient length for it to pass into a suitable receptacle.
When the sipper accessory has been fitted and the tubing has been connected the instrument will look like this.
1.2.3.4 Combined sipper peltier pump Refer to section 1.2.3.3 for more details.
7
1.3 USING THE ACCESSORIES
1.3.1 Automatic 8 cell turret
0.000
0.000
500
ppm
ABS
nm
When the automatic 8 cell turret is in use the 8 cell turret icon is displayed in the bottom right hand corner of the screen. The current cell position is displayed adjacent to the 8 cell turret icon. The 0 position should always be used for the zero calibration sample.
09 : 02
To perform measurements using the automatic 8 cell turret, insert the cuvettes containing the samples into turret positions 1 to 7. Insert the cuvette containing the blank solution into turret position 0. Enter the required measurement mode and set up the required measurement parameters. Press the key below the calibrate to zero icon. The instrument will automatically move the turret around to position zero to perform the measurement. Once the calibration is complete the measure sample icon will appear and the turret will return to its original starting position.
ppm
0.000
500
ABS
nm
0.000
09 : 02
To measure the next sample select the next turret position and press the key below the measure sample icon. Repeat this process until all the samples have been measured. To adjust the wavelength press the key below the 8 cell turret icon and use the arrow icons to adjust the wavelength.
0
Press the key below the 8 cell turret icon to highlight the icon and the two arrow icons above. Press the keys adjacent to the arrow icons to increase or decrease the current cell position of the turret, until the required sample position has been selected. Press the key below
0
the measure sample icon. The instrument will perform a reading and display the result on the screen.
1.3.1.1 Automatic 8 cell turret - supporting creation of a standard curve in quantitation
The 8 cell turret can be used to support creation of a new standard curve in the quantitation measurement mode.
When the standard measurement screen is open the 8 cell turret icon will be displayed in the bottom left hand corner of the screen. The current cell position is displayed adjacent to the 8 cell turret icon. The 0 position should always be used for the zero calibration sample.
0.000 ABS
To measure the standards using the automatic 8 cell
0
turret, insert the cuvettes containing the standards into turret, insert the cuvettes containing the standards into turret positions 1 to 6 (depending on how many standards needed). Insert the cuvette containing the blank solution into turret position 0. Press the key adjacent to the tick icon to perform an initial calibration to zero absorbance.
Use the keys adjacent to the arrow icons to increase the turret position, until the required standard position has been selected. Press the key adjacent to the tick icon to measure the standard. The standard concentration and photometric value will then be displayed. The standard can be re-measured by pressing the key adjacent to the back icon.
To measure the next standard select the next turret position and press the key adjacent to tick icon. Repeat this process until all the standards have been measured.
8
1.3.2 Peltier
09 :0 2
ºC
ºF
0.000
100.0
400
02 0
ABS
%T
nm
.
19.9
-> 120.5
When the peltier is in use the peltier icon is displayed in the bottom right hand corner of the screen. The current temperature is displayed above the set point temperature adjacent to the peltier icon. Below the peltier icon there is an arrow icon to indicate if the current temperature is below or above the set temperature. To adjust the set point temperature hold the key below the peltier icon for 2 seconds.
This opens the peltier settings screen. Use the keys at the bottom of the screen to select the digit to be changed and use the keys adjacent to the arrow icons to increase or decrease the number. The temperature can be set in °C or °F by pressing the key adjacent to the °C icon. Repeat presses will cycle between °C and °F.
Once the required temperature has been selected press the key adjacent to the tick icon to save and return to the expanded operating menu. The peltier will begin to heat or cool depending on the current temperature.
.
80 06
1.3.3 Sipper pump
ABS
0.000
400
%T
nm
100.0
09 : 02
1.3.3.1 MANUAL SIPPER PUMP SETTINGS
When the sipper is in use the sipper pump icon is displayed in the bottom right hand corner of the screen. The pump direction is displayed by an arrow icon below the sipper pump icon. The sipper pump can operate in manual or timed mode, depending on the option selected in sipper pump settings. To open the sipper pump settings hold the key below the sipper pump icon for 2 seconds.
To operate the sipper pump in manual mode press the key adjacent to the manual sipper icon. Select the preferred pump direction by pressing the key below the forwards or backwards arrow icon. Press the key adjacent to the tick icon to save and return to the expanded operating menu.
To perform a measurement place the sipper tubing into
ABS
0.000
%T
the sample and press the key below the sipper pump icon.
100.0
nm
400
09 : 02
9
ABS
0.000
400
%T
nm
ABS
100.0
09 : 02
0.000
400
%T
nm
100.0
09 : 02
1.3.3.2 TIMED SIPPER PUMP SETTINGS
Confirmation will be needed to start the sipper pump. Press the key adjacent to the tick icon to confirm and start the sipper pump. Press the key adjacent to the cross icon to cancel and return to the expanded operating menu.
To stop the sipper pump press the key adjacent to the stop icon. Ensure that the flow through cuvette contains enough sample before pressing the key below the measure sample icon.
To operate the sipper pump in timed mode press the key adjacent to the timed sipper pump icon.
Press the key below the calibrate timed sipper icon. Select the required pump direction by pressing the key below the forwards or backwards arrow icon. Press the key adjacent to the tick icon to continue to the next stage of the calibration sequence.
Insert the inlet tubing into the sample container and press the key adjacent to the single greater than icon. The sipper pump will start and the sample will be pumped through the tubing to the flow through cuvette. It is possible to skip this setup stage by pressing the key adjacent to the double greater than icon.
Once the cuvette is full press the key adjacent to the stop icon to stop the sipper pump. The time taken for sample uptake is recorded.
10
To fine tune the amount of sample uptake press the key below the plus or minus icon to increase or decrease the amount of sample taken up. The recorded time will be adjusted accordingly. Once the fine tuning is complete, or if none is required, press the key adjacent to the tick icon to move to the next stage of the calibration sequence.
This stage allows an air gap to be added to the calibration sequence. If an air gap is not required press the key below the 000 icon to set the air gap to zero. If a previously programmed air gap is to be used press the key adjacent to the double greater than icon to skip this stage and
000
retain the current air gap time.
To program an air gap remove the inlet tubing from the sample container and press the key adjacent to the single greater than icon. The sipper pump will start and air will be pumped through the tubing to the flow through cell.
Once the required amount of air has been taken up press the key adjacent to the stop icon. The time taken for air uptake is recorded.
To fine tune the amount of air uptake press the keys below the plus or minus icons to increase or decrease the amount of air taken in. The recorded time will be adjusted accordingly. Once the fine tuning is complete, or if none is required, press the key adjacent to the tick icon to move to the next stage of the calibration sequence.
Once the sample uptake and air gap have been programmed the preferred disposal of the sample can be set. There are two options, the sample can either be sent back to the sample container or it can be sent to the waste pipe. Press the key below the forward or backward arrows to select what happens to the sample after measurement.
If the original pump direction selected was forwards, selecting the forwards direction at this stage will send the sample to waste and selecting the backwards direction will send the sample back to the sample container. Once the required direction has been selected press the key adjacent to the tick icon to save the calibration sequence and return to the expanded operating menu. To exit the sipper calibration sequence without saving any changes press the back key at any point during the calibration sequence.
11
0.000
ABS
To perform a measurement place the sipper tubing into the sample and press the key below the sipper pump icon.
09 : 02
09 : 02
09 : 02
100.0
400
0.000
100.0
400
0.000
100.0
400
%T
nm
ABS
%T
nm
ABS
%T
nm
Confirmation will be needed to start the sipper pump. Press the key adjacent to the cross icon to cancel and return to the expanded operating menu. Press the key adjacent to the tick icon to confirm and start the sipper pump. The pump will run for the previously recorded sample take up time. Ensure that the flow through cuvette contains enough sample before pressing the key below the measure sample icon.
Once the measurement has been performed remove the tubing from the sample and press the key below the sipper pump icon to perform the next stage of the calibration sequence.
Confirmation will be needed to start the sipper pump. Press the key adjacent to the cross icon to cancel and return to the expanded operating menu. Press the key adjacent to the tick icon to confirm and start the sipper pump. The pump will run for the previously recorded air gap take up time.
ABS
0.000
400
%T
nm
ABS
100.0
09 : 02
1.3.4 Combined sipper peltier pump
0.000
400
%T
nm
100.0
09 : 02
20.0
-20.0
If an air gap of zero was previously selected this screen will not appear and the calibration sequence will continue to sample disposal.
Once this stage of the calibration sequence is complete press the key below the sipper pump icon to dispose of the sample. Confirmation will be needed to start the sipper pump. Press the key adjacent to the cross icon to cancel and return to the expanded operating menu. Press the key adjacent to the tick icon to confirm and start the sipper pump. Depending on the disposal route previously selected the sample will either go to drain or back to the sample container.
When the combined sipper peltier is in use the sipper peltier icon is displayed in the bottom right hand corner of the screen. The current temperature is displayed above the set point temperature adjacent to the sipper peltier icon. Adjacent to the peltier icon is an arrow to indicate if the current temperature is below or above the set temperature.
13
The pump direction is displayed by an arrow icon below the sipper peltier icon. The combined sipper peltier pump combines the functionality of the peltier and sipper pump. To open the sipper peltier settings hold the key below the sipper peltier icon for 2 seconds.
1.4 SPARES
Part Code Description of Spare Part
012 075 Tungsten halogen lamp
730 545 Xenon lamp module
735 801 10x10mm path length cuvette holder
060 084 Pack of 100 disposable plastic visible wavelength 10x10 cuvettes
060 229 Pack of 500 disposable plastic visible wavelength 10x10 cuvettes
060 230 Pack of 100 disposable plastic UV wavelength 10x10 cuvettes
037 702 Paper roll for printer
021 060 24V 65W power supply unit with various plug attachments
The settings menu is the same as the sipper pump settings except for the peltier icon in the top left hand corner. Pressing the key adjacent to the peltier icon will open the peltier settings enabling the temperature to be set. Refer to section 1.3.2 for more details. The sipper pump can operate in a manual or timed mode. Refer to section 1.3 for more details.
14
Bibby Scientific Ltd
Beacon Road Stone Staffordshire ST15 0SA United Kingdom Tel: +44 (0)1785 812121 Fax: +44 (0)1785 813748 e-mail: info@bibby-scientific.com www.bibby-scientific.com
Bibby Scientific France SAS
ZI du Rocher Vert - BP 79 77793 Nemours Cedex France Tel: +33 1 64 45 13 13 Fax: +33 1 64 45 13 00 e-mail: bsf@bibby-scientific.fr www.bibby-scientific.com
Bibby Scientific Italia Srl
Via Alcide de Gasperi 56 20077 Riozzo di Cerro al Lambro Milano Italia Tel: +39 (0)2 98230679 Fax: +39 (0)2 98230211 e-mail: marketing@bibby-scientific.it www.bibby-scientific.it
Bibby Scientific Middle East Ltd
BPO Box 27842 Engomi 2433 Nicosia Cyprus Tel: +357 22 660 423 Fax: +357 22 660 424 e-mail: sales@bibby-scientificme.com www.bibby-scientific.com
Bibby Scientific US Ltd
3 Terri Lane Suite 10 Burlington NJ 08016 USA Tel: 800-225-9243 Fax: 609-589-2571 www.bibby-scientific.com
Bibby Scientific (Asia) Ltd
Room 607 Yen Sheng Centre 64 Hoi Yuen Road Kwun Tong Kowloon Hong Kong Tel: +852 3583 1581 Fax: +852 3583 1580 e-mail bibby@bibby-scientificasia.com www.bibby-scientificasia.com
Loading...