Jenway 6285 User Manual

Fluorimeters
Models 6270, 6280 & 6285
Operating Manual
628 020/REV A/04- 06
Safety
This is important information; please read carefully before installing or using this instrument.
1. The 6
2. The fluorimeter is a sensitive electronic and optical instrument designed for use in a
3. In addition to observing the instructions detailed in the Operating Manual and Service
4. Voltage levels hazardous to life are present in this instrument, for personal safety only
5. This instrument is designed for minimal maintenance, which must be carried out carefully
6. Other than for those items defined in the maintenance procedures herein there are no
7. Reference should always be made to the Health and Safety Data for any chemicals or
8. It is important that good laboratory practice is observed when handling samples,
9. If it is suspected that safety protection has been impaired in any way, the fluorimeter must
2 Series Fluorimeters are designed for operation by trained personnel that are
aware of the principles and applications involved. For further help and advice please contact your local distributor, e-mail sales@jenway.com
or visit www.jenway.com
laboratory environment. Careful adherence to the installation instructions must be
observed. If in doubt contact a relevant and competent authority for advice before
proceeding.
Manual for this instrument all installation, operating and service personnel must be aware
of, and employ, a safe system of work.
trained engineers aware of the risk and avoidance of electric shock should remove
protective covers from the instrument.
following the procedures detailed in this manual. All safety instructions in these procedures, as well as those defined locally for the area or environment where the work
is being carried out must be observed.
user serviceable items in this instrument. Removal of covers and attempted adjustment
or service by unqualified personnel will invalidate any warranty and incur additional charges for repair.
reagents used. All available information, advice and warnings on the handling, storage, use and disposal of such must be carefully observed. When not available this data must be requested from the supplier before proceeding in any way.
chemicals, reagents and ancillary equipment in order to carry out measurement and
analysis with this instrument. Suitable safety and personal protective equipment must
be used at all times.
be made inoperative and secured against any intended operation. The fault condition must be reported to the appropriate servicing authority. In all such reports the model
number and serial number of the fluorimeter must be quoted.
Contents:
Section 1 Introduction
Instrument description Good practice guidelines
Section 2 Getting Started
Unpacking Installation
Section 3 Fluorescence Theory Section 4 Instrument Menu Options Section 5 Instrument Settings
Instrument Lock
Date, Time and Language Security
Section 6 Setting up a method Section 7 Using a stored method Section 8 Calibration Section 9 Performing Measurements
Section 10 Timed Readings Section 11 Saving Results Section 12 Printing Results Section 13 Maintenance & Troubleshooting
General and routine maintenance Error codes and troubleshooting
Section 14 Accessories and Spares
Optional accessories
Section 15 Specification & Data
Technical specification RS232 serial interface
EC Declaration of Conformity
Operator ID
Glossary of Terms
Section 1
Introduction
1.1 Instrument Description
Three models are available to cover a wide range of applications. The model 6280 covers the most sensitive determinations with emission wavelengths up to 650nm. Where higher emission wavelengths will be utilised the model 6285 with its red­enhanced detector is applicable. For less sensitive applications with a broader wavelength range the model 6270 will meet the necessary requirements.
All three models offer intuitive operation with a user interface based on logical menus that can be navigated from the simple keypad. Up to 20 methods can be created without restriction and be saved for future use, they can also be locked against accidental change by password entry, ensuring data integrity.
The permanent time and date tag on every stored reading further enhances Good Laboratory Practice, while calibration reminders and operator identity can also be entered to support conformance and traceability of operation.
With press-to-read operation and Total energy transfer (TET) technology the output of the high-energy xenon lamp is maximised and its expected life extended so that it should never need replacing in normal use.
The high quality optics is complemented by the Intelligent Filter Modules (IFM) that can be identified by the system, this enables error messages to be generated and displayed should the wrong filters be fitted or their positions be incorrect.
All models are powered from an external, universal mains adapter suitable for use from 90Vac to 264Vac, while the 12V dc input enables use of the fluorimeters in vehicles or from suitable battery packs.
Where temperature controlled fluorescence studies are necessary an optional electrically heated sample holder is easily fitted while for continuous flow analysis an external sipper pump and a wide range of flow through cuvettes are also available.
(1)
1.2 Good Practice Guidelines
1. The selection of the optimum excitation and emission wavelengths (filters) is
critical to achieving the best performance from the analysis.
2. All fluorimeters should be sited in a clean, dry, dust free environment. When in use ambient temperature and light levels should remain as constant as possible.
3. Adherance to Standard Operating Procedures (SOP) and Good Laboratory Practice (GLP) should be maintained, with regular calibration checks and a suitable Quality Control (QC) programme.
4. The correct selection of cuvettes is imperative for accurate and reproducible results:
a) Ensure all cuvettes used are compatible with fluorimetric measurements
where the emission beam is at 90° to the excitation source. Typical examples have 4 clear sides.
b) Ensure any native fluorescence from the cuvette material is minimal at the
analysis wavelengths. c) Plastic cuvettes should be used once only. d) Glass and quartz cuvettes should be thoroughly cleaned after use. Discard
when scratches or marks are evident in their optical surfaces. e) Ensure any cleaning agents used do not fluoresce at the analytical
wavelengths and are thoroughly rinsed away before drying. f) Ensure the cuvettes used are compatible with the constituents of both the
samples and standards they are to hold. Plastic cuvettes are not compatible
with some organic solvents. g) Cuvettes must be handled with care; by the top and non-optical surfaces only.
Any finger marks must be removed by using a suitable cleaning process. h) Flow through cuvettes must be selected with additional consideration for the
sample type, aspirated volume, pumping system and rinse cycle, as well as
the waste handling to be used.
5. The high sensitivity of fluorimetric analysis means that all glassware used in the preparation of samples and standards must be totally free from contamination.
6. Chemicals and reagents used in sample and standard preparation should be of the highest grade of purity (AR Grade) and all should be checked for excessive background fluorescence at the analytical wavelengths.
7. Samples and working standards should not be stored due to the effects of evaporation, as well as chemical and photo-degradation. Only prepare samples and working standards when they are required for analysis.
8. Fluorescence is inversely proportional to temperature. Ensure that all samples and standards have equilibrated to ambient temperature before analysis. If in doubt, use a temperature controlled cuvette holder.
9. Refrigerated or cold samples will form micro-bubbles on the cuvette wall as they warm up. These are a common cause of drift in readings. Ensure all samples and standards have equilibrated to ambient temperature before analysis.
(2)
10. Check the linear range for each method and, where necessary, use a multi-point calibration or calibration curve.
11. Be aware of the effects of quenching and, where necessary, use sample dilution or extraction methods to eliminate this.
12. Monitor the blank during and between batches to identify any increase due to contamination.
13. Sources of contamination to be considered include cleaning agents, micro­organisms, particles in suspension, stop-cock grease, filter paper residues and plasticisers leached from containers, caps or sealing materials.
14. Many fluorescent assays are pH dependent. Ensure the pH of all samples and standards is within specified limits before carrying out the analysis.
(3)
Section 2
Getting Started
2.1 Unpacking
Remove the universal 12V power supply (with UK, US and EU leads) and the pack of 100 disposable cuvettes from the packaging.
Remove the fluorimeter from the carton by lifting it in the centre between the two support cheeks; do not lift it by the support cheeks.
Place all items on a clean workbench then remove the support cheeks and the polythene bag from the fluorimeter.
Any shortages or damage must be reported to your local distributor or the manufacturer as soon as possible.
Keep all packing materials in case the unit has to be re-shipped at a later date. It is important that when re-packing the instrument it is first sealed in a strong, clean polythene bag to protect it from the dust and particles that are present in all packing materials.
2.2 Installation
2.21 Location
In ideal circumstances the installation environment will be clean, dry and dust free with the instrument protected from extreme variations in ambient lighting and temperature change. For field use it is recommended that the instrument is used in the optional storage case for additional protection. Where conditions are less than ideal, maintenance and cleaning must be carried out regularly and additional protection offered where possible. The optional dust cover should always be fitted when the unit is not being used or is stored for short periods.
2.22 Supply Voltage
The fluorimeter is powered by a low voltage dc power supply that operates from a 90­264Vac mains supply. The universal power supply is supplied with 3 mains leads for UK, EU and US sockets. The correct lead for your supply should be selected.
2.23 Mains Connections
The leads supplied have a moulded on plug. However, if this is removed for any reason the wires in the mains lead are colour coded to conform to the internationally recognised standard such that:
UK CONNECTIONS BROWN LIVE BLACK LIVE BLUE NEUTRAL WHITE NEUTRAL GREEN/YELLOW EARTH GREEN EARTH
Safety When disposing of any removed plug the connectors must be removed or made incapable of insertion into a mains socket.
US CONNECTIONS
(4)
2.24 Keypad Functions
1
425673135
246
1. UP ARROW key - used to navigate through menus, to increase values
and for paging up in stored results.
2. DOWN ARROW key - used to navigate through menus, to decrease values
and for paging down in stored results.
3. LEFT ARROW key - used to navigate through menus and to highlight the
selected digit when setting values.
4. RIGHT ARROW key - used to navigate through menus and to highlight the
selected digit when setting values.
5. ENTER key - used to accept the highlighted menu option.
6. CAL key - initiates a calibration sequence from within the
measurement mode.
7. PRINT key - initiates a print from the measurement display or stored
results. Sends data to RS232 serial port.
2.25 Instrument Display
1. Method Name - this will appear on all measurement screens, with the exception of
Raw Fluorescence.
2. Results display - provides direct readout of standards and sample results.
3. Units of measure - shows selected measurement units: ppm, RFU, U/ml, mU/l,
U/l, µM/l, mM/l, M/l, µg/ml, mg/ml, µg/dl, mg/dl, g/dl, ng/l, µg/l, mg/l, g/l, none, %, ppb
4. Status message - shows the current instrument status, such as reading … or
printing … and provides a reminder when calibration is due.
5. Gain - shows the current photomultiplier tube gain setting (0-100%). Gain should
be optimised either manually or by using auto set gain for each method.
Note: This option is not available on the Model 6270.
6. Menu Options - EXIT - allows the user to return to the main menu, READ - press read to read the sample or standard, SAVE - saves the currently
displayed result to instrument memory
(5)
2.26 Rear Panel Layout
1. Output Socket – 9 way output socket for RS232
2. Switch – Power On/Off switch
3. Connection Sockets – 4 x pin sockets for heated cell block controller.
4. Power In Socket – Connection socket for 12V DC universal mains adapter
2.27 Sample Chamber Filter and Cuvette Positioning
(6)
2.28 Power on and Self-Tests
Connect the mains supply cable to the rear panel mains input socket and plug the other end into a suitable mains supply socket.
Lift the sample chamber lid on the instrument and ensure that there is no sample or other item present in the sample holder, close the lid.
Switch on the supply socket, then the instrument, using the Power switch located on the rear panel.
The instrument will then perform a power on self-test protocol. The following screen will be shown for approx. 3 seconds:
Please ensure sample chamber door remains closed during power on tests
Followed by:
On successful completion of these tests the Main Menu screen will be displayed. For optimum performance a 10-minute warm-up period is required if the ambient temperature is below 10°C. The unit must be re-calibrated and sample measurement repeated if this calibration check shows excessive drift.
Should a problem occur during the self-tests an information box or error support message will be displayed. For assistance please refer to the troubleshooting section in this manual.
(7)
Example Menu Operation
A common operating system is used throughout this and similar Jenway products, a brief overview of navigating through the menu system with the cursor keys follows;
Use the up and down arrow keys to highlight a menu option, press the enter key to confirm.
Use the right and left arrow keys to select a digit for adjustment with the up and down arrow
Left arrow
Up arrow key
Right arrow
Down arrow
Enter key
Left arrow
Up arrow key
Right arrow
Down arrow
keys, press the enter key to confirm the setting when ALL digits have been correctly set. Use the up or down arrow keys to browse through pre-set options, when the correct selection
is displayed press the enter key to confirm your choice.
(8)
Enter key
Left arrow
Up arrow key
Right arrow
Down arrow
Enter key
Section 3
Fluorescence Theory
The interaction between electromagnetic radiation and matter provides a useful, qualitative and quantitative analytical tool, known as spectroscopy. The region of the electromagnetic spectrum, to which matter under investigation is subjected to, defines the type of transitions that occur within the molecules. Fluorimetry uses radiation from the UV-Vis region of the electromagnetic spectrum to study transitions between electronic levels in a molecule or atom. The absorption of energy from light radiation (photons) by a molecule or atom, promotes electrons from a low energy ground state to a higher energy excited level. This is known as excitation and the amount of energy transferred to the molecule or atom will depend on two main factors. The composition of the matter under investigation and the energy and wavelength of the radiation, have a significant effect on the transitions of electrons. The molecule or atom converts the excitation energy to vibrational or light energy and the electron returns to its ground state. Vibrational energy is transferred through movement and collision with other molecules, but energy not lost in this way is released as light radiation. The light emission is known as fluorescence and if some energy has been removed through vibration, it will be of a lower energy and longer wavelength than the excitation energy. The wavelength and intensity of the emitted radiation is dependent on the structure and composition of the molecule and the excitation wavelength used.
Relationship between concentration and fluorescence
The fluorescence signal F and concentration C of the matter under investigation, are related by:
ε
-
F = KQP
(1-10
0
bC
)
Where
K = A constant characteristic of the instrument (Including instrument
electronics, pH and Temperature) Q = Quantum efficiency (= Photons emitted/Photons absorbed) P
= Power of incident radiation
0
ε = Molar absorptivity of the species (matter)
b = Absorption path length
If the concentration of the matter in question is low (dilute), relationship is then linear and the equation can be written as
F = 2.3KQP
εbC
0
The accuracy of fluorescent measurements is very high because the radiant energy being formed is measured directly. There are also only a few, easily controlled limits on the sensitivity of fluorescence measurements. From the equation above it can be seen that adjustments made to instrument electrical noise and competing radiations, along with physical limitations such as radiation energy, sample volume and cell size affect the measurement sensitivity.
(9)
εbC is small. The
Loading...
+ 26 hidden pages