Safety
Please read this information carefully prior to installing or using this equipment.
1. The unit described in this manual is designed to be operated only by trained personnel. Any
adjustments, maintenance and repair must be carried out as defined in this manual, by a person
qualified to be aware of the hazards involved.
2. It is essential that both operating and service personnel employ a safe system of work, in addition to
the detailed instructions specified in this manual.
3. References should always be made to the Health and Safety data supplied with any chemicals
used. Generally accepted laboratory procedures for safe handling of chemicals should be employed.
4. If it is suspected that safety protection has been impaired in any way, the unit must be made
inoperative and secured against any intended operation. The fault condition should immediately be
reported to the appropriate servicing authority.
Good Practice Guidelines3.1
Set Up Parameters3.2
Preparation of Conductivity Standards3.3
Calibration with Known Cell Constant3.4
Calibration with Standard Solution3.5
Sample Measurement3.6
A fully specified laboratory Conductivity/Resistivity/TDS/Salinity/Temperature meter that includes full
support for good laboratory practices (GLP). The meter supports 1, 2 or 3 point conductivity
calibration. Powerful data logging capabilities are included with the ability to store up to 500
readings either manually, at timed intervals or alarmed events.
1.2 Instrument Specifications
Conductivity
Ranges0 to 19.99µS / 0 to 199.9µS / 0 to 1999µS /
0 to 19.99mS / 0 to 199.9mS / 0 to 1999mS / 0 to 19.99S*
Remove the Model 4520 from the packaging and ensure the following items are included:
1. Model 4520 Conductivity/TDSMeter
2. Glass bodied conductivity cell with ATC K=1 (027 013)
3. Electrode holder
4. Power Supply (as specified at time of ordering the product)
5. Condensed operating instructions (452 051)
6. Operating Manual (452 050)
The electrode stand requires minimal assembly (refer to the diagram below).
Any shortages or damage should be reported immediately to the manufacturer or your local
distributor.
2.2 Installation
The Model 4520 is supplied ready to use. Connect the conductivity cell to the rear panel DIN socket.
The electrode stand requires minimal assembly (refer to the diagram below).
Fig. 2.2.1 Electrode Holder Assembly
4
1
2
3
2
452 050/REV A/03-03
2.3 Displays
1.Primary display – 4½ digit. Provides direct readout in Conductivity and TDS of samples and
standards.
2.Mode annunciators – shows selected measurement mode; Conductivity or TDS.
3.Secondary display – 3½ digit display. Provides direct readout of automatic or manual
temperature.
4.Mode annunciators – indicates temperature in °C or °F and whether the measurements are
manually temperature compensated (MAN symbol).
5.Endpoint symbol – this symbol is displayed when an endpoint has been detected.
6.Mode tags – Each mode tag is highlighted when selected; SETUP, MODE,
STATUS or RESULTS. If a double headed arrow symbol is present this indicates that the
mode can be changed to an alternative option (pH/mV).
7.Real time clock - will display either date or time.
8.The following symbols will appear along the display:
Padlock - Set up parameters security locked
Notepad symbol - data logging to internal memory
IrDA status
Alarm indication - an Up arrow refers to Hi alarm / a Down arrow refers to Low alarm
3
452 050/REV A/03-03
2.4 Keypad
1. ESCused to switch the instrument on and to place into standby mode (only if power
supply lead remains connected to the instrument). Also used to escape/exit a
mode.
2. CAL / CLRused to select and perform a calibration sequence. This key is also used to clear
readings from Memory.
3. Print keyused to initiate a print.
4. Up Arrowused for adjustment during set up, to scroll results and to toggle between
Conductivity (µS or mS) and TDS (mg/l or g/l) modes.
5. Down Arrowused for adjustment during set up, to scroll results and to toggle between
Conductivity µS or mS) and TDS (mg/l or g/l) modes.
6. Left Arrowused for adjustment during set up and to move between mode tags.
7. Right Arrowused for adjustment during set up and to move between mode tags.
8. STOused to accept an entered value in set-up mode and to instigate a stored reading.
This key can also be used as a CAL key during calibration.
4
452 050/REV A/03-03
2.5 Inputs/Outputs
2.5.1 Rear panel layout
1. Conductivity Socket7 pin DIN socket which allows the conductivity cell to be connected.
2. Alarm Out2 x 4mm sockets. Open collector alarm output.
3. Output Socket9 way connection for RS232.
4. Power InAC 9V I/P socket. 2.1 x 5.5mm socket allowing the power supply to be
connected to the instrument.
5. Analog Out2 x 4mm sockets. Analogue output (buffered electrode potential).
5
452 050/REV A/03-03
Section 3
Operation
3.1Good Practice Guidelines
1.For greatest accuracy ensure no particulate matter is suspended in the solution under test. If
necessary, filter or allow the particles to settle prior to use. Do not allow the cell to come into
contact with any sediment which may be present.
2.Ensure no air bubbles are trapped in the cell area between the plates.
3.Ensure the cell plates are completely immersed in the solution under test.
4.Thoroughly rinse the cell plates in deionised water after use, and for short term storage
immerse the cell in deionised water. Although it is not essential to store the conductivity cell
with the plates in a wetted condition, if they are allowed to dry out completely, initial stability on
re-use may be impaired until the cell plates become re-wetted.
5.Do not attempt to clean the cell plates as this will remove the black platinization and alter the
calibration and accuracy of the probe. If the cell plates become worn or damaged the plates
should be re-platinised and the cell constant recalculated.
6.Ensure no salt deposits or particulate matter are allowed to build up around the cell plates or
on the probe body as this may produce a conductivity path lower than that through the
solution. It is recommended that such deposits be removed by soaking the cell in deionised
water. No attempt should be made to wipe off these deposits as this may cause damage to
the cell plates.
7.Ensure the correct reference temperature is selected for the operating procedures being
used.
6
452 050/REV A/03-03
Loading...
+ 19 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.