Multi-Parameter
Water Quality
TROLL
Information Sheet
Catalog No. 33250
QUICK CAL
SOLUTION
APPLICATION
A calibration standard of an appropriate chemical composition to perform a one-
point calibration of all four MP TROLL Basic sensors (pH, ORP, Conductivity,
Polarographic Dissolved Oxygen) at the same time with a single solution. Four
250 mL containers, each sufficient for at least 2 Quick Cals.
Shelf Life, unopened 6 months
After opening 7-21 days
®
pH, ORP, & Conductivity values
at various temperatures
pH mV µS/cm
Temp. (± 0.02) (± 5) (± 40)
5oC 7.10 255 4910
10oC 7.04 247 5635
15oC 7.03 239 6390
20oC 7.02 231 7180
25oC 7.00 224 8000
30oC 6.98 217 8850
35oC 6.97 209 9735
Specific Conductivity Coefficient:
1.95% / degree C (0-25oC)
2.06% / degree C (15-35oC)
2.23% / degree C (25-50oC)
EXPECTED ACCURACIES
The following accuriacies may be expected when calibrating the MP TROLL
using Quick Cal solution and following instructions in the Win-Situ Quick Cal
Wizard and/or in the MP TROLL Operator's Manual:
Quick Cal Meets USGS
Parameter accuracy guidelines*
pH: at pH 7 ± 0.09 pH units
over range 0-14 pH ± 0.2 pH units
ORP ± 25 mV NA
DO (with stirrer) ± 0.25 mg/L
Conductivity:
Low Range (3 – 50,000 µS/cm) ± 3%
High Range (70 – 200,000 µS/cm) ± 3%
* Calibration criteria for continuous water-quality monitors, from USGS Water-Resources
Investigations Report 00-4252, p. 13
GUIDELINES AND PRECAUTIONS
• Store the solution cool and dark. Shake well before use. For best results,
calibrate at the temperature expected during field use
• Do not store sensors in Quick Cal solution
• For best results, use fresh solution for each Quick Calibration
• Dispose of used solution by flushing down the drain with plenty of water
• Solution is best if used before date on bottle, but may be usable beyond that
date if stored cool, dark, and tightly capped
✓✓
✓
✓✓
✓✓
✓
✓✓
✓✓
✓
✓✓
✓✓
✓
✓✓
✓✓
✓
✓✓