Step 1: Tape adapter threads.
Level
TROLL®
Accessories
Information Sheet
Catalog No. 0051470
NPT ADAPTER
APPLICATION
1/4"–18 NPT thread adapter permits installation of a Level TROLL in threaded
process piping.
The kit includes:
• 015 Viton
• Teflon® thread tape
• Vacuum grease
• NPT adapter (titanium grade 5)
Tools required:
• 11/16" wrench
®
o-ring
Step 3: Remove instrument nose
cone—be very careful not to touch
the exposed pressure sensor!
Pressure sensor
opening
Lubricated o-ring
Step 4. Lubricate o-ring and install
on probe.
1/4" NPT threaded
pipe or fitting
INSTALLATION
1. Tape threads, clockwise (when looking down threaded end), 2 turns.
2. Thread taped adapter into pipe or fitting using 11/16" wrench on wrench flats.
3. Remove the instrument nose cone.
Do not touch the pressure sensor face with any object when the
nose cone is removed. Damage caused by digging or scraping
in the pressure sensor opening is not covered by the warranty.
4. Lubricate the supplied o-ring with vacuum grease and install on threads at front
end of instrument as shown.
5. Thread probe into adapter and hand-tighten.
GUIDELINES AND PRECAUTIONS
• Pressure is limited by the pressure range of the Level TROLL.
Taped adapter
Level TROLL
1 800 446 7488
(toll-free, US and Canada) or 970 498 1500 www.in-situ.com