QUICK START GUIDE
How to Set your IRwin Methane Leak Detector for 1 Point Calibration
2.2% (2.5%) - Universal Mode
(for a factory set unit)
1
Make sure your IRwin is set in Universal Mode (see the How to set your IRwin for Universal
Mode quick start guide)
2
Press followed by
3
Use to navigate to General and press
4
Use to navigate to Custom and press
5
Use to navigate to AbG Cal Setup and press
1 of 5
mimc66en1-02 (1905)

6
If prompted, enter the Supervisor (Advanced) password code
If not prompted, go ahead to next step
7
Use to select Generic and press
followed by
How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.2% (2.5%)
8
Use and to navigate to Calibration
and press
9
Use to navigate to Settings and press
10
If prompted, enter the (min) Intermediate password code
If not prompted, go ahead to next step
2 of 5
mimc66en1-02 (1905)