QUICK START GUIDE
How to Set your IRwin Methane Leak Detector for 1 Point Calibration
2.2% (2.5%) - Universal Mode
(for a factory set unit)
1
Make sure your IRwin is set in Universal Mode (see the How to set your IRwin for Universal
Mode quick start guide)
2
Press followed by
3
Use to navigate to General and press
4
Use to navigate to Custom and press
5
Use to navigate to AbG Cal Setup and press
1 of 5
mimc66en1-02 (1905)

6
If prompted, enter the Supervisor (Advanced) password code
If not prompted, go ahead to next step
7
Use to select Generic and press
followed by
How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.2% (2.5%)
8
Use and to navigate to Calibration
and press
9
Use to navigate to Settings and press
10
If prompted, enter the (min) Intermediate password code
If not prompted, go ahead to next step
2 of 5
mimc66en1-02 (1905)

11
Press to get to the drop-down menu
12
Press on the rst choice of the drop-down
menu
How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.2% (2.5%)
13
Navigate to far right drop-down menu and press
14
Use to select Disabled and press
15
Press again to view drop-down menu
3 of 5
mimc66en1-02 (1905)

16
In Universal Low Range, use to select the
second choice in the drop-down menu, press
17
Navigate to far right drop-down menu and press
How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.2% (2.5%)
18
Use to select Disabled and press
19
Press again to view the drop-down menu
20
In Universal Low Range, use to select the third
choice in the drop-down menu, press
4 of 5
mimc66en1-02 (1905)

21
Use to navigate to drop-down menus. Set the
actual value of the calibration gas you have, nominally
2.2% or 2.5%, by entering:
2 and press
2 (5) and press
22
Navigate to far right drop-down menu and press
How to Set your IRwin Methane Leak Detector
for 1 Point Calibration 2.2% (2.5%)
23
Select Set by using and and press
24
Press followed by and
www.inficon.com reach.sweden@inficon.com
Due to our continuing program of product improvements, specifications are subject to change without notice.
All trademarks are the property of the ir respective owners.
©2019 INFICON mimc66en1-02 (1905)