Inficon IRwin Methan Leak Detector Quick Start Manual

QUICK START GUIDE
How to Perform GC Function Test on IRwin Methane Leak Detector (for a factory set unit)
1
Make sure your IRwin is set in Universal Mode (see
the How to set your IRwin for Universal Mode quick start guide)
2
Press followed by
3
4
Press . Use to select GC in the drop-down menu, press
5
Press to perform a function test with GC Mix
1.0% CH4 50ppm C2H6 1000ppm C3H
1 of 3
8
mimd66en1-03 (1910)
6
Press to start the Function Test for GC
7
Expose the instrument to GC gas mix (GC mix 1.0% CH4 50ppm C2H6 1000ppm C3H8)
How to Perform GC Function Test on IRwin Methane Leak Detector
8
The Function Test is in progress
9
Expose both probe and instrument to fresh air
10
The GC Analysis is in progress
2 of 3
mimd66en1-03 (1910)