The Service Information provides the following information:
Precautions and safety information
Specifications
Basic maintenance (cleaning, replacing the battery and fuses)
Performance test procedures
Calibration and calibration adjustment procedures
To avoid shock or injury, do not perform the verification tests or calibration procedures
described in the manual unless you are qualified to do so.
The information provided in this document is for the use of qualified personnel only.
Caution
The 61-320 serials contain parts that can be damaged by static discharge.
Follow the standard practices for handling static sensitive devices.
For additional information about IDEAL INDUSTRIES, INC. and its products, and services, visit
IDEAL INDUSTRIES, INC. web site at:
www.idealindustries.com
Precautions and Safety Information
Use the Meter only as described in the Service Manual. If you do not do so, the protection provided by the
Meter may be impaired. Read the “Safety Information” page before servicing this product.
In this manual, a Warning identifies conditions and actions that pose hazard (s) to the user; a Caution
identifies conditions and actions that may damage the Meter or the test instruments.
The Symbols
The symbols used on the Meter and in this manual are explained in Table A.
Table A. The Symbols
Risk of electric shock
See instruction card
DC measurement
Equipment protected by double or reinforced
insulation
Battery
Earth
AC measurement
Conforms to EU directives
1
Form Number TM61320-2-4 Rev 3 April 2004
SAFETY
Review the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To avoid potential hazards, use the product only as specified.
CAUTION: These statements identify conditions or practices that could result in damage to the
equipment or other property.
WARNING: These statements identify conditions or practices that could result in personal injury or
loss of life.
Specific precautions
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the
product without covers in place.
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified
for that connector.
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads
while they are connected to a voltage source.
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or
damp conditions.
2
Form Number TM61320-2-4 Rev 3 April 2004
SPECIFICATIONS
All specifications are warranted unless noted typical and apply to the 61-320 & 61-322 & 61-324
Stated accuracies are at 23°C±5°C at less than 80% relative humidity and without the battery indicator
displayed.
General specifications
Characteristics Description
Display count 3 3/4
Numeric update rate 1.5 times / sec
Polarity display Automatic
Overrange display “OL” is display
Low voltage indicator is indicated
Automatic power-off time Automatic backslit off = 30 minutes
Power source
9V×1 battery for 61-322 & 61-324;
1.5V×2 batteries for 61-320
Maximum input voltage 1000V CAT III between V and COM
Maximum floating voltage
1000V CAT
III
between any terminal and earth ground
Maximum input current 400mA between mA and COM
Maximum open circuit Voltage (current inputs)600V between mA and COM
Overload protection mA connector 1A (600V) fast blow fuse.
V connector V, V, Ω, , , , Hz, µA
Temperature Coefficient 0.2×(Spec. Accuracy) / °C, <18°C or >28°C
Battery Life
Alkaline 9V 200 hours for 61-322 & 61-324
Alkaline 1.5V×2 AA size 600 hours for 61-320
3
Form Number TM61320-2-4 Rev 3 April 2004
Measurement Characteristics
Accuracy is ±(% reading + number of digits) at 23°C ± 5°C, less than 80% R.H.
(1) DC Volts
Range Resolution Accuracy
Over voltage
protection
400.0mV 100µV
4.000V 1mV
40.00V 10mV
±(0.5% reading + 2 digit) 1000V rms
400.0V 100mV
1000V 1V
Input Impedance: 10M (over 1000M in 400mV range).
(2) AC Volts
Range Resolution Accuracy
400.0mV 100µV
4.000V 1mV
±(1.3% reading + 5 digits) *
Unspecified
40.00V 10mV
400.0V 100mV
750V
1V
±(1.2% reading + 5 digits) *
1
2,3
Over voltage
protection
1000V rms
Input Impedance: 10M // less than 100pF.
1
*
Frequency Response: 50Hz ~ 300Hz
2
Frequency Response: 50Hz ~ 500Hz
*
CMRR / NMRR: (Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
V
: CMRR > 60dB at DC, 50Hz / 60Hz
AC
V
: CMRR > 100dB at DC, 50Hz / 60Hz
DC
NMRR > 50dB at DC, 50Hz / 60Hz
AC Conversion Type: 61-320: Average sensing rms indication calibrated to the sine wave input.
61-322 / 61-324: AC conversions are ac-coupled, true rms responding, calibrated to
the sine wave input.
3
*
61-322 / 61-324 The specified accuracy is for sine wave at full scale and non-sine wave at half scale
with crest factor up to 2.
Crest Factor: C.F. = Peak/RMS
+1.5% addition error for C.F. from 1.4 to 3
+3.0% addition error for C.F. from 3 to 4.
Form Number TM61320-2-4 Rev 3 April 2004
4
(3) DC Current
Range Resolution Accuracy Voltage Burden
400.0µA 0.1µA 5mV
4000µA 1µA 2V max.
±(1.0% reading + 2 digits)
40.00mA* 10µA 5mV
400.0mA* 0.1mA
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse. (61-322 / 61-324)
µA Input : 600V rms.
* For 61-322 / 61-324 only.
(4) AC Current (61-322 / 61-324)
Range Resolution Accuracy Voltage Burden
40.00mA 10µA 5mV
400.0mA 100µA
±(1.5% reading + 5 digits) *
50Hz ~ 500Hz
1
2V max.
Overload Protection: mA Input : 1A (600V) fast blow fuse.
AC Conversion Type: AC conversions are ac-coupled, true rms responding, calibrated to the sine wave
input.
1
*
The specified accuracy is for sine wave at full scale and non-sine wave at half scale with crest factor
up to 2.
(5) Resistance
Range Resolution Accuracy
400.0 *
1
0.1±(1.0% reading + 5 digits)
Over voltage
protection
4.000K 1
40.00K 10
±(0.7% reading + 2 digits)
600V rms
400.0K 100
4.000M 1K
40.00M *
2
10K±(1.5% reading + 5 digits)
Open circuit Voltage: -1.3V approx.
1
< 5 digit of reading rolling.
*
2
*
< 2% of reading rolling.
(6) Diode Check and Continuity
Range Resolution Accuracy
Max. Test
Current
Max. Open
Circuit Voltage
1mV ±(1.5% reading + 5 digits) * 1.5mA 3V
* For 0.4V ~ 0.8V
Overload Protection: 600V rms.
Continuity: Built-in buzzer sounds when resistance is less than approximately 450 Response time
is approximately 100 msec.
5
Form Number TM61320-2-4 Rev 3 April 2004