GE Sensing PT878 Quick Start Guide

Page 1
PT878
Quick Start Guide
Page 2
September 2003
Process Control Instruments
PT878 Portable Flowmeter
Quick Start Guide 916-068A1
Page 3
July 2003
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Return Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Keypad Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Powering ON and OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Programming Transducer, Pipe and Fluid Parameters . . . . . . . 10
Transducer Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Pipe Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Fluid Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Saving Site Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Changing the Number of Displayed Measurements . . . . . . . . . 16
Changing the Measured Parameters . . . . . . . . . . . . . . . . . . . . . 17
PT878 Quick Start Guide 1
Page 4
July 2003

Warranty

Each instrument manufactured by GE Panametrics is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Panametrics. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE Panametrics determines that the equipment was defective, the warranty period is:
one year from delivery for electronic or mechanical failures
one year from delivery for sensor shelf life
If GE Panametrics determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE Panametrics, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in
lieu of all other warranties whether statutory, express
or implied (including warranties or merchantability and
fitness for a particular purpose, and warranties arising
from course of dealing or usage or trade).
PT878 Quick Start Guide 3
Page 5
July 2003

Return Policy

If a GE Panametrics instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE Panametrics, giving full details of the problem, and
provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE Panametrics will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE Panametrics instructs you to send your instrument to a
service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Panametrics will evaluate the instrument to
determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
If GE Panametrics determines that the damage is not covered
under the terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s app roval to proceed, the instrument will be repaired and returned.
4 PT878 Quick Start Guide
Page 6
July 2003

Introduction

This booklet provides directions for making basic flow measurements with the PT878, including the following:
Keypad Description
Powering Up
Programming
Displaying Measurements
This document is designed solely to get you up and running as quickly as possible. It does not provide a detailed description of the procedures, but simply the steps needed to make flow measurements.
PT878 Quick Start Guide 5
Page 7
July 2003

Keypad Description

Since you need to use the keypad in the procedures that follow, you should familiarize yourself with the keys and their functions.
The P T878 keypad has 25 keys. The functions for each key are as follows (see Figure 1 on the next page):
3 function keys ([F1], [F2], [F3]) — enable you to select the
special functions which appear at the bottom of the screen.
12 numeric keys (including - and .) — enable you to enter
numeric data.
4 arrow keys ([W ], [X], [S], [T]) — enable you to move
through the menu options.
[?] Help key— enables you to access on-line help.
[MENU] Menu key — enables you to access the Main Menu.
[ENTER] — enables you to enter a particular menu, and
enters selected values into the PT878 memory.
[SEL] — enables you to move between data measurements on
the screen.
[ESC] — enables you to exit menus or menu options at any
time; cancels a numeric entry.
Red key [ ] — turns the power ON or OFF, and toggles the
backlight ON or OFF.
6 PT878 Quick Start Guide
Page 8
Keypad Description (cont.)
July 2003

Figure 1: TransPort PT878 Keypad

PT878 Quick Start Guide 7
Page 9
July 2003

Powering ON and OFF

To operate the PT878, the power cord must be plugged into line voltage or the battery must be fully charged as described in Chapter 2, Initial Setup, of the User’s Manual. You must also install your transducers as discussed in the Liquid Transducer Installation Guide (document #916-055), and make transducer cable connections to the top of the P T878 as described in Chapter 2 of the User’s Manual.
To turn the PT878 ON, press the red button in the upper-right­hand corner of the keypad. Immediately upon power up the PT878 emits a short beep and displays a “PCI Loader” message. It then validates the instrument programming, displays the GE Panametrics logo and the software version, and emits a long beep. If the meter fails any of these tests, contact GE Panametrics.
!WARNING!
If the meter fails the backup battery test, you
must send the unit back to the factory for a
battery replacement. Make sure you keep the
NiCad batteries charged until you are ready to
ship the unit back to the factory. Before
shipping, print out all the log and site data, or
transfer it to your PC.
After the meter conducts all the self checks, the screen then appears similar to the one shown in Figure 2 on the next page.
To turn the PT878 OFF, press the red key for 3 seconds. The screen now appears similar to Figure 3 on the next page. Three options are available:
Press [F1] to turn the PT878 completely OFF.
Press [F2] to send the PT878 into sleep mode. In this mode,
some of the power supplies shut down, but the PT878 remains in a standby mode. Users can resume taking measurements immediately by pressing the power button.
Press [F3] to cancel the command and return the PT878 to
normal operation.
8 PT878 Quick Start Guide
Page 10
Powering On and Off (cont.)
July 2003
ABC.SIT
2000/11/30 09:53 AM
Velocity, ft/s
0.00
Delta-T, ns
0.10
E0: No Errors

Figure 2: Screen After Power Up

Velocity, ft/s
SHUTDOWN: Meter OFF SLEEP: Meter Idle CANCEL: Resume Operations
Signal, dB
32
Volume, l/s
0.0
Signal, dB
Delta-T, ns
Shutdown Sleep

Figure 3: Screen After Power Down

PT878 Quick Start Guide 9
Volume, l/s
Cancel
Page 11
July 2003

Programming Transducer, Pipe and Fluid Parameters

In order for the PT878 to begin taking measurements, you must enter the transducer, pipe and fluid parameters. Once these parameters are entered, you may save them so that you can edit them at a later time.
To enter the Program Menu, press the right of the PT878 keypad. Press the [
[MENU] key at the lower
X] arrow key once to
scroll from the Site Menu to the Program Menu. At the Program Menu, press
[ENTER]. The screen appears similar to Figure 4
below.
Site Program Meter Logging Service
Transducer
Vel
ocity, ft/s
Pipe Lining
Fluid Path
Energy
Delta
Inputs Analog Output Digital Output User Functions
Co
Correction Factors
E0: No Errors
0.00
-T, ns
0.10
Signal, dB
32
Volume, l/s
0.0

Figure 4: The PROGRAM Menu

10 PT878 Quick Start Guide
Page 12
July 2003

Transducer Parameters

This section describes how to enter parameters for all standard transducer types. To step through each parameter within the option, press the [
Note: If you are using a special transducer, refer to Chapter 3 in
your User’s Manual.
1. T o enter the T ransducer option from the Program Menu, press
[ENTER].
2. The first prompt asks you to select whether you are using a
wetted or a clamp-on transducer.
T] key. Begin by using the following steps:
a. Use the [ b. Press
Note: If the PT878 does not scroll to a particular parameter, it
is not necessary for that transducer or pipe type.
3. The next prompt asks you to enter the transducer number. a. From the Type prompt, press the [
Transducer prompt, and press
b. A drop-down list of transducer numbers opens. (The list
varies, depending on whether you have selected wetted or clamp-on in the previous prompt.) Press the [ keys to scroll to the appropriate number. To speed scrolling, you can press the [ page, or the [
c. Press
4. The PT878 supplies the needed parameters for the selected
transducer. Press the [ at the top and proceed to the Pipe tab.
W] and [X] keys to scroll between the two types.
[ENTER] to confirm the choice.
T] key to reach the
[ENTER].
T] or [S]
X] key to scroll down by a
W] key to scroll up by a page.
[ENTER] to confirm your selection.
T] key to return to the Transducer tab
PT878 Quick Start Guide 11
Page 13
July 2003

Pipe Parameters

This section describes how to enter parameters for most standard pipe materials and diameters. To step through each parameter within the option, press the [ enter the pipe parameters.
T] key. Use the following steps to
1. From the Transducer tab, press the [ the Pipe tab, and press
2. The first prompt asks you to select the pipe material. a. Press b. A drop-down list of materials opens. Press the [
c. Press
d. (This prompt only appears if you have selected “Other” in
3. At the Diameter prompt, the meter asks for the pipe outside
diameter and thickness. a. You must select from two choices, outside diameter or
[ENTER] to enter the material prompt.
keys to scroll to the appropriate material, or scroll to “Other” for a material not on the list. You can press the
X] key to scroll down by a page, or the [W] key to scroll
[ up by a page.
[ENTER] to confirm the choice.
step b.) The meter prompts you to enter the sound speed. Use the numeric keys to type the desired sound speed in the text box, and press
circumference.
[ENTER].
[ENTER] to confirm the choice.
X] arrow key to reach
T] or [S]
The prompt moves to the OD (outside diameter) text box.
Type the thickness (in mm or in.) into the text box, and
[ENTER] to confirm the choice, or
press
Move the prompt to the OD X PI (circumference) text box.
Type the OD (in mm or in.) into the box, and press
[ENTER] to confirm your choice.
12 PT878 Quick Start Guide
Page 14
July 2003
Pipe Parameters (cont.)
Note: If you do not know the outside diameter, refer to Sound
Speeds and Pipe Size Data (914-004) or use the On-Line Help function (refer to the User’s Manual).
b. In either case, the next prompt asks for the wall thickness.
Type the value (in mm or in.) into the box, and press
[ENTER] to confirm your choice.
Note: If you have selected certain pipe materials (carbon or
stainless steel, cast iron, PVC and CPVC), you have the option of entering the pipe dimensions by a standardized schedule. Refer to Chapter 3 of the User’s Manual to learn more about this option.
4. Press the [
S] key twice to return to the Pipe tab.
You have completed entering data in the Pipe tab. Proceed to the Fluid tab.

Fluid Parameters

This section describes how to enter parameters for most common fluids. To step through each parameter, press the [
1. From the Pipe tab, press the [X] arrow key twice to reach the Fluid tab, and press
[ENTER].
2. The first prompt asks you to select whether or not you want Tracking Windows. This feature is used when the process fluids soundspeed fluctuates considerably. (Default operation is “No.”)
a. Use the [
W ] and [X] keys to scroll to the appropriate
radio button.
b. Press
[ENTER] to confirm your selection.
T] key.
PT878 Quick Start Guide 13
Page 15
July 2003
Fluid Parameters (cont.)
3. To select the fluid type: a. Press
b. Scroll to the appropriate fluid. If you do not see your fluid
c. Press
[ENTER] to open the drop-down menu. (The choices
vary, depending on whether or not you have selected Tracking Windows.)
on the list, select “Other.”
[ENTER] to confirm your selection.
If you have selected a preprogrammed fluid, the PT878
supplies the sound speed. Pressing the [T] key returns you to the Tracking Windows prompt.
If you have selected “Other,” the prompts dif fer, depending on
whether or not you have enabled the Tracking Windows.
If you have not enabled the Tracking Windows, the PT878
prompts you to enter the sound speed. Use the numeric keys to enter the desired sound speed, and press [ENTER].
If you have enabled the Tracking Windows, the PT878
prompts you for minimum and maximum sound speeds.
a. Use the numeric keys to enter the desired minimum speed.
[ENTER].
Press
b. Press the [T ] key to move to the maximum speed text
box. Use the numeric keys to enter the desired maximum speed. Press
At the end of either sequence, pressing the [T] key returns you to the Tracking Windows prompt. You have completed entering data in the Fluid tab. To confirm the entries and return to Operate Mode, press
It is a good idea to read Chapter 3, Programming Site Data, in the User’s Manual for more details on programming site parameters.
14 PT878 Quick Start Guide
[ENTER].
[F3] (OK).The PT878 returns to Operate Mode.
Page 16
July 2003

Saving Site Data

Now that you have entered all the necessary data, the PT878 will give you the opportunity to save it into a site file using the following steps:
1. To open the Site Menu, press the right of the P T878 keypad. Then press appears similar to Figure 5 below.
Site Program
Manager
Velocity, ft/s
Save Now
1 View 2 Views 3 Views 4 Views
Delta-T, ns
Drive Manager FKeys
0.00
Meter Logging Service
4
0.10
E0: No Errors
[MENU] key at the lower
[ENTER]. The screen
Signal, dB
32
Volume, l/s
0.0

Figure 5: The Site Menu

1. Press the [T] key to reach the “Save Now” option, and press
[ENTER].
2. A window appears, asking if you want to save the current site (DEFAULT. SIT).
Press [F2] (No) to cancel saving the site, or
Press [F3] (Yes) to save the site.
The PT878 returns to displaying measurements. If you want to save the site under another name, or create a new site, refer to Chapter 4, Creating and Managing Sites, in the User’s Manual.
PT878 Quick Start Guide 15
Page 17
July 2003

Changing the Number of Displayed Measurements

The P T 87 8 can display one to four different measurement parameters simultaneously. However, sometimes you might wish to display only one or two parameters. To change the number of open display windows from within the Site Menu (shown in Figure 5 on the previous page), use the [ move to the desired number of views (1 View, 2 Views, etc.)
[ENTER]. The screen displays the desired number of
Press windows.
To display data in other formats (such as bar or line graphs) or modify the softkeys, refer to Chapter 5, Displaying and Configuring Data, in the User’s Manual.
T]or [S] arrow keys to
16 PT878 Quick Start Guide
Page 18
July 2003

Changing the Measured Parameters

Once the PT878 is programmed, you can display the following measurements on the screen:
Velocity
Volumetric
Forward Totals
Reverse Totals
Power
Energy
Diagnostics
To enter the Measurement option:
1. Press
2. Press
[SEL] from the Operate Mode window until you have
reached the desired measurement.
[ENTER] to open the Display Menu. The screen appears
similar to Figure 6 below.
ABC.SIT
4
Velocity, ft/s
Format View Measurement
2000/1 1 /30 09:53 AM
0.00
Delta-T, ns
0.10
E0: No Errors

Figure 6: The Display Menu

Signal, dB
32
Volume, l/s
0.0
PT878 Quick Start Guide 17
Page 19
July 2003
Changing the Measured Parameters (cont.)
3. Press the [X] twice to highlight the Measurement option, and
[ENTER].
press
4. A drop-down menu entry, Change, appears. Press
[ENTER]
again.
Another window opens, shown in Figure 7 below; the left column displays the measurement data sources, while the right column displays English and metric measurement units (or, with the Diagnostics source shown, a list of diagnostics parameters).
Select Measurement
Diagnostics
Velocity
Meters/sec
Feet/sec Volume Fwd Totalizer
Rev Totalizer
Power
Fwd Energy
Rev Energy
Temperature
No Unit
Cancell OK

Figure 7: Measurement Selection Window

5. Use the [
source, and press
6. Then use the [
measurement unit (or diagnostic parameter). Press
T] and [S] arrow keys to reach the desired data
[SEL].
T] or [S] arrow keys to select the desired
[F3], OK,
to confirm the entry.
The screen returns to Operate Mode, and displays any changes.
18 PT878 Quick Start Guide
Page 20
July 2003
Index
A
Arrow Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
C
Changing Screen Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
D
Diameter, Pipe Outside . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
E
ESC Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Exit Key. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
F
Feature Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Fluid Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Fluid Type, Selecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Function Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
H
Help Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
K
Key
ESC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6, 7
Keys
Arrow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
ESC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Help. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Numeric. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
SEL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
M
Measurement Types, Changing. . . . . . . . . . . . . . . . . . . . . . . . . . . 17, 18
Measurements, Changing Number of . . . . . . . . . . . . . . . . . . . . . . . . . 16
Menu Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
PT878 Quick Start Guide 19
Page 21
July 2003
Index (cont.)
N
Numeric Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
P
Pipe Diameter, Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Pipe Material. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Pipe Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12, 13
Power Key. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Powering Off. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Powering On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8, 9
Program Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Programming
Fluid Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Pipe Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12, 13
Programming Transducer Parameters. . . . . . . . . . . . . . . . . . . . . . . . . .11
R
Return Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
S
Schedule for Pipe Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
SEL Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Site Data, Saving. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Site Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Sleep Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Sound Speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12, 14
Sound Speeds, Minimum and Maximum . . . . . . . . . . . . . . . . . . . . . . 14
T
Tracking Windows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Transducer Number, Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Transducer Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Transducer Type, Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Transducers
Wetted or clamp-on. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
W
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
20 PT878 Quick Start Guide
Page 22
WORLDWIDE
OFFICES
MAIN OFFICES: GE PANAMETRICS INTERNATIONAL OFFICES:
USA
GE Panametrics 221 Crescent St., Suite 1 Waltham, MA 02453-3497 USA Telephone: 781-899-2719 Toll-Free: 800-833-9438 Fax: 781-894-8582 E-mail: panametrics@ps.ge.com Web: www.gepower.com/panametrics
ISO 9001 Certified
Ireland
GE Panametrics Shannon Industrial Estate Shannon, Co. Clare Ireland Telephone 353-61-470200 Fax 353-61-471359 E-mail info@panametrics.ie
ISO 9002 Certified
Australia
P.O. Box 234 Gymea N.S.W. 2227 Australia Telephone 61 (02) 9525 4055 Fax 61 (02) 9526 2776 E-mail panametrics@panametrics.com.au
Austria
Waldgasse 39 A-1100 Wien Austria Telephone +43-1-602 25 34 Fax +43-1-602 25 34 11 E-mail panametrics@netway.at
Benelux
Postbus 111 3870 CC Hoevelaken The Netherlands Telephone +31 (0) 33 253 64 44 Fax +31 (0) 33 253 72 69 E-mail info@panametrics.nl
France
BP 106 11 Rue du Renard 92253 La Garenne Colombes Cedex France Telephone 33 (0) 1 47-82-42-81 Fax 33 (0) 1 47-86-74-90 E-mail panametrics@panametrics.fr
Germany
Mess-und Pruftechnik Robert-Bosch-Straße 20a 65719 Hofheim Germany Telephone +49-6122-8090 Fax +49-6122-8147 E-mail panametrics@t-online.de
Italy
Via Feltre, 19/A 20132 Milano Italy Telephone 02-2642131 Fax 02-26414454 E-mail info@panametrics.it
Japan
2F, Sumitomo Bldg. 5-41-10, Koishikawa, Bunkyo-Ku Tokyo 112-0002 Japan Telephone 81 (03) 5802-8701 Fax 81 (03) 5802-8706 E-mail pci@panametrics.co.jp
Korea
Kwanghee Bldg., 201, 644-2 Ilwon-dong, Kangnam-Ku Seoul 135-945 Korea Telephone 82-2-445-9512 Fax 82-2-445-9540 E-mail jkpark@panaeng.co.kr
Spain
Diamante 42 28224 Pozuelo de Alarcon Madrid Spain Telephone 34 (91) 351.82.60 Fax 34 (91) 351.13.70 E-mail info@panametrics.infonegocio.com
Sweden
Box 160 S147 23 Tumba Sweden Telephone +46-(0)8-530 685 00 Fax +46-(0)8-530 357 57 E-mail pana@panametrics.se
Taiwan
7th Fl 52, Sec 3 Nan-Kang Road Taipei, Taiwan ROC Telephone 02-2788-3656 Fax 02-2782-7369 E-mail rogerlin@lumax.com.tw
United Kingdom
Unit 2, Villiers Court 40 Upper Mulgrave Road Cheam Surrey SM2 7AJ England Telephone 020-8643-5150 Fax 020-8643-4225 E-mail uksales@panametrics.ie
July 2003
Page 23
USA
GE Panametrics 221 Crescent Street, Suite 1 Waltham, MA 02453-3497 Telephone: (781) 899-2719 Toll-free: (800) 833-9438 Fax: (781) 894-8582 E-Mail: panametrics@ps.ge.com Web: www.gepower.com/panametrics
Ireland
GE Panametrics Shannon Industrial Estate Shannon, County Clare Ireland Telephone: 353-61-470200 Fax: 353-61-471359 E-Mail: info@panametrics.ie
Loading...