Technical content subject to change without notice.
[no content intended for this page
ii
Preface
Information Paragraphs
• Note paragraphs provide information that provides a deeper understanding of
the situation, but is not essential to the proper completion of the instructions.
• Important paragraphs provide information that emphasizes instructions that
are essential to proper setup of the equipment. Failure to follow these
instructions carefully may cause unreliable performance.
• Caution! paragraphs provide information that alerts the operator to a hazardous
situation that can cause damage to property or equipment.
• Warning! paragraphs provide information that alerts the operator to a
hazardous situation that can cause injury to personnel. Cautionary information
is also included, when applicable.
Safety Issues
WARNING! It is the responsibility of the user to make sure all local, county,
state and national codes, regulations, rules and laws related to
safety and safe operating conditions are met for each
installation.
Auxiliary Equipment
Local Safety Standards
The user must make sure that he operates all auxiliary equipment in accordance
with local codes, standards, regulations, or laws applicable to safety.
Working Area
WARNING! Auxiliary equipment may have both manual and automatic
modes of operation. As equipment can move suddenly and
without warning, do not enter the work cell of this equipment
during automatic operation, and do not enter the work
envelope of this equipment during manual operation. If you do,
serious injury can result.
WARNING! Make sure that power to the auxiliary equipment is turned OFF
and locked out before you perform maintenance procedures on
the equipment.
Model PM880 Portable Hygrometer User’s Manualiii
Preface
Qualification of Personnel
Make sure that all personnel have manufacturer-approved training applicable to the
auxiliary equipment.
Personal Safety Equipment
Make sure that operators and maintenance personnel have all safety equipment
applicable to the auxiliary equipment. Examples include safety glasses, protective
headgear, safety shoes, etc.
Unauthorized Operation
Make sure that unauthorized personnel cannot gain access to the operation of the
equipment.
Environmental Compliance
Waste Electrical and Electronic Equipment (WEEE) Directive
GE Measurement & Control Solutions is an active participant in Europe’s Waste
Electrical and Electronic Equipment (WEEE) take-back initiative, directive
2002/96/EC.
The equipment that you bought has required the extraction and use of natural
resources for its production. It may contain hazardous substances that could impact
health and the environment.
In order to avoid the dissemination of those substances in our environment and to
diminish the pressure on the natural resources, we encourage you to use the
appropriate take-back systems. Those systems will reuse or recycle most of the
materials of your end life equipment in a sound way.
The crossed-out wheeled bin symbol invites you to use those systems.
If you need more information on the collection, reuse and recycling systems, please
contact your local or regional waste administration.
Visit http://www.ge-mcs.com/en/about-us/environmental-health-and-
safety/1741-weee-req.html for take-back instructions and more information
about this initiative.
The GE PM880 is a versatile, battery-powered portable hygrometer that is
designed for spot sampling moisture measurements. Although the PM880
offers many options, it requires only battery charging and minimal setup to
begin taking measurements.
This chapter is designed to provide a step-by-step guide to getting the meter
up and running as quickly as possible. Additional programming options are
discussed in later chapters.
Follow the sections in this chapter to setup your meter. This chapter
includes the following information:
• Charging the Battery Pack
• Powering On and Off
• Entering Data Using the Keypad
• Entering Setup Data
• Displaying Measurements
• Saving Data in Site Files
Model PM880 Portable Hygrometer User’s Manual1
Chapter 1. Getting Started
1.1 Charging the Battery Pack
The PM880 is powered by a rechargeable NiMH battery pack. When you
receive the PM880, you will need to fully charge the pack. The pack must
be charged for 4-5 hours, until the amber LED on the charger goes out, to
receive the maximum charge. When fully charged, the pack provides 16 to
24 hours of continuous operation, depending on the probe type and the
PM880 usage. An internal battery gauge indicates the remaining power in
the pack.
Before charging the battery pack, make sure the charger is properly
configured. Use the instructions below:
1.1.1Removing the Battery Pack
1. Make sure you are in a general-purpose location where you can safely
charge the pack.
WARNING! THE BATTERY PACK CAN BE INSTALLED AND REMOVED IN
HAZARDOUS AREAS, HOWEVER, IT MUST BE CHARGED IN
GENERAL-PURPOSE AREAS ONLY.
2. Remove the four thumb screws on the back panel of the PM880 that
secure the pack.
3. Lift the pack out of the unit.
2Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.1.2Configuring the Battery Charger
The PM880 battery charger has a switchable voltage input of 115 to
230 VAC. Make sure the voltage selector is in the correct position before
plugging it in.
WARNING! The battery charger is intended for indoor use only.
See Specifications for power cord type.
Risk of electrical shock; use in dry locations only.
Charge on NiMH type rechargeable batteries.
Other types of batteries may burst,
causing injury and/or damage.
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
!DANGER!
TO REDUCE THE RISK OF FIRE OR ELECTRICAL SHOCK, CAREFULLY
FOLLOW THE GUIDELINES BELOW.
BE SURE THE VOLTAGE SELECTOR IS IN THE CORRECT VOLTAGE
POSITION BEFORE PLUGGING IT IN.
(SEE THE BOTTOM OF THE CHARGER.)
FOR USE IN THE UNITED STATES, THE VOLTAGE SELECTOR SWITCH
MUST BE PLACED IN THE 115 VOLT POSITION.
FOR USE IN COUNTIRES OUTSIDE THE UNITED STATES, CONFIRM THE
VOLTAGE AVAILABLE AT YOUR LOCATION
AND PLACE THE VOLTAGE SELECTOR IN THE CORRECT POSITION
BEFORE USING THE PRODUCT.
FOR CONNECTION TO A SUPPLY NOT IN THE UNITED STATES, USE AN
ATTACHMENT PLUG OR ADAPTOR OF THE PROPER CONFIGURATION
FOR THE POWER OUTLET.
Model PM880 Portable Hygrometer User’s Manual3
Chapter 1. Getting Started
1.1.3Charging the Battery Pack
1. Plug the pack into the battery charger.
2. Connect the charger to AC power. When power is applied, the amber
LED glows and the charger automatically begins charging the pack.
Note: The pack is fully charged when the amber LED goes out
(approximately 4-5 hours).
IMPORTANT: When you receive your unit, you should completely charge the
battery pack before using it for the first time.
3. Remove the pack from the charger and reinstall it in the PM880. Refer to
Checking the Battery Status on page 133 for details on monitoring the
battery.
4Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.2 Powering On and Off
IMPORTANT: For CE compliance, the PM880 is classified as a
battery-powered device.
1.2.1Powering On
T o turn the PM880 on, press the red button in the upper-right-hand corner of
the keypad. Immediately upon powerup the PM880 displays a series of
messages and performs various checks. The screen then appears similar to
the one shown in Figure 1.
Note: If the meter displays a Probe Reminder message, press [ENTER].
Figure 1: Screen After Powering On
Note: If the meter fails any of these tests, contact the factory.
Model PM880 Portable Hygrometer User’s Manual5
Chapter 1. Getting Started
1.2.2Powering Off
1. To turn the PM880 off, press the red key for 3 seconds. The screen now
appears similar to Figure 2.
2. Press [F1] to shut down the PM880.
Figure 2: Shutdown Menu
6Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.3 Entering Data Using the Keypad
Use the information below to familiarize yourself with how to enter data
using the PM880 keypad.
Use the arrow keys to scroll to a menu topic and then press
the menu.
When entering data into a menu window, press:
[ENTER] to open
• The [] key to scroll forward through the menu options.
• The [] key to scroll back through menu options.
• The [F2] key (Cancel) or the [ESC] key to exit a menu at any time and
return to Operate Mode without changing data.
Note: If you enter an incorrect numeric value, press the [] key to erase
the last digit entered.
1.4 Entering Setup Data
For immediate operation, the PM880 requires only a moisture probe and its
corresponding calibration data. You should enter data for each site where
you plan to make a measurement. Once entered, you can save it into a file
that can be recalled at the measurement site. Use the sections that follow to
program your meter.
Note: The PM880 has additional menus that enable you to tailor
measurements as specifically as possible to your particular
application. For more information on menu options refer to
Chapter 4, Using Special Features.
Model PM880 Portable Hygrometer User’s Manual7
Chapter 1. Getting Started
1.4.1Selecting Probe Type
Use the steps below to select the probe type for the measurement site. Refer
to Figure 9 on page 19 for a menu map.
Note: You may also enter a constant value (rather than a live input) for
moisture, temperature or pressure measurement. For more
information refer to Entering Constants and User Functions on
page 54.
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to Probe and press [ENTER]. The screen
appears similar to Figure 3.
Figure 3: Probe Window
8Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.4.1 Selecting Probe Type (cont.)
4. At Type, press [ENTER] to open the drop-down list of probe types.
Note: If you do not know the probe type, refer to the Calibration Data
Sheet.
5. Use the arrow keys to scroll to the desired selection and press [ENTER].
6. Press [F3] (OK) to exit.
Next, do one of the following:
• If you are using an M or TF Series probe, proceed to Entering
Calibration Data on the next page.
• If you are using an MIS or an MISP2 Probe, proceed to Displaying
Measurements on page 15.
Note: If you have an MIS probe that was sent back to the factory for
calibration without the electronics module, you need to enter
calibration data as described in Entering Calibration Data on
page 10.
Model PM880 Portable Hygrometer User’s Manual9
Chapter 1. Getting Started
1.4.2Entering Calibration Data
The PM880 needs moisture and/or pressure calibration data for probes. You
must enter this data when setting up:
• a new unit
• a probe(s) sent back to the factory for calibration
Be sure you have the Calibration Data Sheets that are supplied with each
GE probe. Each Calibration Data Sheet consists of a list of data points that
you will need to enter or verify, and lists its corresponding probe serial
number. Calibration Data Sheets are usually packed inside the probe cases.
Since the PM880 enables you to use more than one probe, the meter
provides a place to enter the probe serial number to help you match the
probes to the corresponding site files. The probe serial number is stored as
part of the site file.
Use the sections which follow to enter the serial number, and the moisture
and pressure calibration data for the probes.
10Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.4.2aEntering Probe ID
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to Calibrate and press [ENTER].
4. Use the arrow keys to scroll to the Probe ID tab and press [ENTER]. A
screen similar to Figure 4 appears.
Figure 4: Probe ID Window
5. Use the arrow keys to scroll to S/N and press [ENTER].
6. Use the numeric keys to enter the serial number from the Calibration
Data Sheet and press
[ENTER]. The serial number also appears on the
hex nut of the moisture probe.
7. Use the arrow keys to scroll to the suffix box and press [ENTER] to open
the drop-down list.
8. Use the arrow keys to scroll to the desired suffix and press [ENTER].
9. Press [F3] (OK) to exit.
Model PM880 Portable Hygrometer User’s Manual11
Chapter 1. Getting Started
1.4.2bEntering Moisture Calibration Data
You need to enter calibration data for M and TF Series probes only. The
Moisture Image
electronics module, and upload it into the PM880 memory when needed.
Note: It is not necessary to enter calibration data for the Moisture Image
Series (MIS) probe unless you send it back to the factory for
calibration without its electronics module. If this is the case, you
must manually enter the calibration data as described below.
Note: It is not necessary to enter calibration data for the MISP2 probe.
Enter moisture calibration data (MH or FH) and dewpoint readings (up to
20 data points) for the moisture probe as described below. Refer to Figure 9
on page 19 for a menu map.
Note: If you ar e using an MIS probe and do not need to enter data, pr oceed
to Displaying Measurements on page 15.
Series (MIS) probes store all calibration data in their
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to Calibrate and press [ENTER].
4. If necessary, use the arrow key to scroll to Hygro and press [ENTER]. A
screen appears similar to Figure 5 on page 13.
5. Press [] to move the pointer to the 00/MH box in the calibration table.
6. Press [ENTER] to change the value. Use the numeric keys to enter the
desired MH or FH value, and press
12Model PM880 Portable Hygrometer User’s Manual
[ENTER].
Chapter 1. Getting Started
1.4.2b Entering Moisture Calibration Data (cont.)
Figure 5: Hygro Window in Calibration
7. Press [] to move to the next text box, and press [ENTER]. Use the
numeric keys to enter the corresponding dewpoint reading and press
[ENTER].
8. Use the arrow keys to scroll to additional data points and repeat steps 5,
6 and 7 until you have entered the values for each data point.
Note: To insert an additional data point, press [F1] (Insert).
To delete a data point, press
9. When you have completed entering values, press [F3] (EXIT).
[F2] (Delete).
Next, do one of the following:
• If the probe has a pressure transducer, proceed to step 3 in Entering
Pressure Calibration Data on page 14.
• If the probe does not have a pressure transducer, press [F3] (OK).
Proceed to Displaying Measurements on page 15.
Model PM880 Portable Hygrometer User’s Manual13
Chapter 1. Getting Started
1.4.2cEntering Pressure Calibration Data
To enter calibration data for the pressure transducer, you must list the zero
and span range in mV (or FP) and psig. Refer to Figure 9 on page 19 for a
menu map.
1. If the menu is not active, press [MENU].
2. Use the arrow key to scroll to Program and press [ENTER].
3. Use the arrow key to scroll to Calibrate and press [ENTER].
4. Use the arrow key to scroll to Pressure and press [ENTER]. The screen
appears similar to Figure 6.
Note: The corresponding mV value can be in psig (English) or kPa
(metric). The supplied Calibration Data Sheets list these values in
psig. If you want to change the system units, refer to Selecting English or Metric Units on page 46.
Figure 6: Calibration Pressure Window
14Model PM880 Portable Hygrometer User’s Manual
Chapter 1. Getting Started
1.4.2c Entering Pressure Calibration Data (cont.)
5. Press [] to move the pointer to the Zero/mV (Zero/FP) box in the
table.
6. Press [ENTER] to change the value. Use the numeric keys to enter the
desired value and press
7. The corresponding pressure value box is highlighted. Use the numeric
keys to enter the corresponding pressure value and press
8. The span mV box is highlighted. Repeat steps 5, 6 and 7 to enter the
[ENTER].
[ENTER].
Span values.
9. When you have completed entering values, press [F3] (Exit).
10. Press [F3] (OK) and proceed to Displaying Measurements.
1.5 Displaying Measurements
The PM880 can display one to four measurement parameters
simultaneously . There are two parts for displaying measurements: selecting
the number of measurements and then selecting the type of measurements.
See Figure 9 on page 19 for a menu map.
1.5.1Selecting the Number of Measurements
1. If the menu is not active, press [MENU].
2. Use the arrow keys to scroll to Site and press [ENTER].
3. Use the arrow keys to move to the desired number of views or
measurement windows (1 View, 2 Views, etc.).
4. Press [ENTER] at your selection. The screen displays the designated
number of views (measurement windows).
Model PM880 Portable Hygrometer User’s Manual15
Chapter 1. Getting Started
1.5.2Selecting the Types of Measurements
1. Press [SEL] or arrow keys to move the pointer to the windo w you want to
change and press
2. Use the arrow keys to scroll to Measurement and press [ENTER]. The
screen appears similar to Figure 7. The left column displays the five
measurement types and the right column displays the measurement
units.
[ENTER].
Figure 7: Select Measurement Window
3. Use the arrow keys to scroll to the desired measurement type and press
[SEL].
4. Use the arrow keys to select the desired measurement unit (or diagnostic
parameter).
Note: Press [F3] (No Unit) to display a measurement with no units.
5. Press [F3] (OK).
6. Repeat this section for displaying other measurements.
16Model PM880 Portable Hygrometer User’s Manual
Loading...
+ 188 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.