GE Sensing OxyTrak 390 Operating Manual

GE
Sensing
OxyTrak 390
Panametrics Flue Gas Analyzer
User’s Manual
910-264B3 September 2007
The OxyTrak™ 390 Flue Gas Analyzer is a Panametrics product. Panametrics has joined other GE high­technology sensing businesses under a new name—GE Sensing.
September 2007
Warranty Each instrument manufactured by GE Sensing, Inc. is warranted to be
free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE determines that the equipment was defective, the warranty period is:
one year for general electronic failures of the instrument
one year for mechanical failures of the sensor
If GE determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of all other warranties whether statutory, express or implied (including warranties of merchantability and fitness for a particular purpose, and warranties arising from course of dealing or usage or trade).
Return Policy If a GE Sensing, Inc. instrument malfunctions within the warranty
period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the model
number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE will issue a RETURN AUTHORIZATION number (RA), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE instructs you to send your instrument to a service center, it
must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine the
cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
If GE determines that the damage is not covered under the terms
of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iii
September 2007
Table of Contents
Chapter 1: General Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Sample System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Display Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Zirconium Oxide Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Platinum-Catalyst Combustibles Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Heater Control Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Chapter 2: Installation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Installation Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Selecting the Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Preparing the Site . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
CE Mark Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Wiring the Analog Outputs (A-C). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Wiring the Alarm Relays (A-D) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Wiring the Calibration Relays (E-H) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Wiring the RS232 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Wiring the RS485 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Remote Display Option . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Wiring the Line Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Chapter 3: Operation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Preventing Common Problems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Cleaning the Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Powering Up the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
The Display and Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Entering Programming Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Exiting Programming Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Temporary Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Locking Programming Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Powering Down the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Taking Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
v
September 2007
Table of Contents (cont.)
Chapter 4: Setting Up the Display
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Selecting the Number of Views . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Adjusting the Display Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Selecting the Measurement Mode and Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Measurement Units Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Chapter 5: General Programming
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Setup Menu (Real Time Clock). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Calibrate Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Manual, One-Gas Oxygen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Manual, Two-Gas Combustibles Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Automatic, One-Gas Oxygen Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Output Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Trim . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Cal Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
Relays Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Relays A-D . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
Relays E-H . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
Communications Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
RS232/RS485 Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
Ethernet Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Chapter 6: Advanced Programming
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Display, Relays, and Communications Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Output Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Calibrate Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
O2 Tolerance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Combustibles Tolerance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Setup Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
O2 Sensor Temp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Heater Block Temp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Auto Cal Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Blow Back . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Factory Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Default Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Additional Menu Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
vi
September 2007
Table of Contents (cont.)
Chapter 7: Specifications
Performance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Appendix A: The Nernst Equation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Equilibrium Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
The OxyTrak™ 390 Equation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Appendix B: Menu Maps
Display and Setup Menu Map (2719 passcode). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Outputs and Relays Menu Map (2719 passcode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
Calibrate and Communications Menu Map (2719 passcode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3
Calibrate, Output and Factory Menu Map (2719 passcode) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
Setup Menu Map (7378 passcode). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
Appendix C: Calibration, Standard Convection
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Method #1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Method #2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Method #3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Method #4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-3
Method #5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Method #6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-4
Method #7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
Calibration Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-5
Appendix D: Calibration, Aspirated Flow
Response Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .D-1
Calibration Gas Flow Rate Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1
Oxygen Sensor Field Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
Manual Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2
vii
September 2007
Table of Contents (cont.)
Appendix E: Default Settings
Calibration Sheet Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
Display Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-2
Calibrate Defaults. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-3
Output Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
Output A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4
Output B. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-5
Output C. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-6
Relays Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-7
Communications Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-13
Setup Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-14
Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-15
Appendix F: Blow Back Sample System
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1
viii

Chapter 1

General Information

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
September 2007

Introduction Process plant managers are usually looking for ways to reduce

expense and increase profitability . When combustibles are burned as part of the operation, and that combustion is incomplete (allowing unburned fuel to escape), costs go up and profits go down.
A reliable system for analyzing flue gas can provide the necessary information to:
adjust the flow of oxygen
increase the efficiency of the combustion
gain significant cost savings for the overall operation
To meet these specific needs, GE Sensing provides the OxyTrak™ 390 Flue Gas Analyzer which monitors the efficiency of a furnace or
boiler by measuring excess oxygen and/or ppm combustibles in the flue gases.
To measure these two parameters, the OxyTrak™ 390 uses:
unburned
v
a zirconium oxide oxygen sensor
a platinum-catalyst combustibles sensor (optional)
The oxygen sensor measures excess oxygen or, in a fuel rich environment, equivalent combustibles. The combustibles sensor monitors partially combusted fuel, only in the presence of excess oxygen (i.e. there must be enough oxygen present to burn the fuel). Each OxyTrak™ 390 may be equipped with an oxygen sensor, a combustibles sensor, or both.

Physical Description The standard GE Sensing OxyTrak™ 390 Flue Gas Analyzer is

provided in a general-purpose weatherproof (IP52, NEMA 2) enclosure. The analyzer consists of a convection loop/analyzer package and a display controller, which may be mounted either locally or remotely . Figure 1-1 on page 1-2 shows the OxyTrak™ 390 with local and remote display controllers.
General Information 1-1
September 2007
Local Controller
ESC ENT
Remote Controller
Junction Box
Figure 1-1: Standard OxyTrak™ 390 Configurations
ESC ENT
1-2 General Information
September 2007
Sample System The convection loop/analyzer package houses the sample system,
which consists of the components shown in Figure 1-2 on page 1-4 and Figure 1-3 on page 1-5. The functions of the sample system components are as follows:
a manifold with removable thermocouple and cartridge heaters to
prevent acid components of the flue gas from condensing in the sample system and causing corrosion
a zirconium oxide oxygen sensor
an optional platinum-catalyst combustibles sensor to monitor
incomplete combustion of the fuel by burning it in the presence of excess oxygen
a temperature-controlled sensor furnace to maintain the oxygen
sensor at a stable operating temperature and to act as the engine for convective sampling
a convection loop to circulate the sample gases through the sample
system
an aspirator port to connect to an aspirated probe.
General Information 1-3
September 2007
Sample System (cont.)
Furnace T/C
Sensor
Furnace
Convection
Loop
Aspirator
Port
Manifold
Oxygen Sensor
Combustibles
Sensor
Aspirated
Probe
Connection
Figure 1-2: The Sample System
1-4 General Information
Sample System (cont.)
Sensor
Furnace
September 2007
Convection Loop
Combustibles Sensor
Oxygen Sensor Washer
Zirconium Oxide Oxygen Sensor
Combustibles Sensor Washers
Figure 1-3: Sensor Locations (ref. dwg #705-1088, sht 3)
General Information 1-5
September 2007
Sample System (cont.)
Figure 1-4: Combustibles Sensor
SENSOR POSITIVE
Figure 1-5: Zirconium Oxide Oxygen Sensor
1-6 General Information
September 2007
Display Controller The display controller (see Figure 1-6 below) includes the terminal
blocks for making all electrical connections and the furnace temperature control (FTC) circuit board. The FTC board maintains a
constant sensor furnace temperature to improve the accuracy of the oxygen analysis and to extend the life of the oxygen sensor.
Figure 1-6: Display Controller Interior
The display controller performs the following functions:
amplifies the oxygen and combustibles sensor outputs
linearizes the oxygen signal
controls the sensor temperature
outputs the reading on a 64 x 128 pixel graphic display
enables programming using an integral keypad
provides a linear 4-20 mA analog output
provides four alarm relays
provides four auto-calibration relays
provides RS232/RS485 communications outputs
General Information 1-7
September 2007

Principles of Operation Ideally, every furnace/burner should mix a precise ratio of air to fuel,

and the mixture should burn efficiently to yield only heat, water vapor and carbon dioxide. However, because of burner aging, imperfect air to fuel mixtures and changing firing rates, this rarely happens. Monitoring the actual efficiency of the combustion process is easily accomplished with the OxyTrak™ 390.
A flue gas sample is drawn into the probe by gaseous diffusion and a gentle convective flow. The sample passes through the probe and into the sample system, where it is maintained at a temperature above 200°C (392°F) by the heater block. In the presence of oxygen, this sample temperature is high enough to burn any partial combustion products that reach the active (platinum-coated) element of the combustibles sensor. The resulting temperature differential between the two combustibles sensor elements is related to the concentration of partial combustion products in the test sample.
Note: The sampled gas is maintained above 200°C (392°F) to
prevent flue gas acids from condensing in the analyzer and causing corrosion.
The sample then passes into the sensor furnace, which heats the sample gas and the oxygen sensor to 700°C (1,292°F) (a temperature above 650°C (1,202°F) is required for proper operation of the oxygen sensor). The oxygen sensor is covered with a platinum catalyst that causes the burning of all remaining combustibles, enabling the sensor to measure the excess oxygen (or fuel) in the flue gas.
The sensor furnace also generates the convective flow that circulates the sample gas through the sample system. The hot sample gas in the sensor furnace rises out of the furnace and cools, as it is pushed from behind by the hot gases still in the furnace. The cooled sample gases then drop down the other branch of the convection loop and into the annular space between the probe and probe sleeve, where they are carried away by the gas flow in the flue.
1-8 General Information
September 2007
Zirconium Oxide Oxygen Sensor
The inside and outside of the zirconium oxide oxygen sensor are coated with porous platinum, forming two electrodes. The sample gas flows past the outside of the sensor, while atmospheric air circulates freely inside the sensor. This atmospheric air is used as the reference gas for making oxygen measurements. See Figure 1-7 below.
Oxygen ions migrate through the zirconium oxide along the concentration gradient.
O
O
2
O
2
O
O
2
Atmospheric O
Inside Cell
20.9%
2
Sample O
2
Outside Cell
2
O
2
2
Zirconium Oxide Ceramic with Lattice Imperfec tio n s
From O u ts ide
Electrode
When O concentration in sample gas falls, the cell voltage rises
2
Volts
From Ins id e Electrode
with increased oxygen migration through the zirconium oxide.
Figure 1-7: Oxygen Migration in the Zirconium Oxide Sensor
At the operating temperature of the oxygen sensor, the atmospheric reference oxygen is electrochemically reduced at the inner electrode, and the resulting oxygen ions seek to equalize with the lower oxygen concentration on the sample side of the cell by migrating through the porous ceramic toward the outer electrode. At the outer electrode they give up electrons to become oxygen molecules again, and are swept away by the sample gas flow.
The lower the concentration of oxygen in the flue gas sample, the greater the rate of ion migration through the ceramic, and the higher the cell voltage due to electron exchange at the electrodes. The cell voltage rises logarithmically as the amount of oxygen in the flue gas falls, allowing the accurate measurement of very low levels of excess oxygen in the flue gas.
General Information 1-9
September 2007
Platinum-Catalyst Combustibles Sensor
The combustibles sensor consists of two platinum thermistors mounted side by side in the sample stream. One thermistor, the active element, is used to detect/react partial combustion products, while the other thermistor, the reference element, provides a baseline. The active element is coated with a black platinum catalyst and the reference element has a white inert surface. As the sample gas passes over the active element, the platinum catalyst causes any combustibles to burn (in the presence of excess oxygen), thereby raising the temperature of the active element above that of the reference element (see Figure 1-8 below).
Flue Gas Flow
Reference
Element
(Inert Coating)
Active
Element
(Platinum C ata ly st)
Combustibles Sensor
Figure 1-8: Combustibles Sensor Elements
The resulting temperature differential between the active and reference elements is proportional to the concentration of combustibles in the sample, and a corresponding resistance change is then converted into a reading of parts per million by volume (ppm
of combustibles.
)
V
1-10 General Information
September 2007
Heater Control Circuit The oxygen sensor temperature in the OxyTrak™ 390 is maintained
by a heater, which is part of a complex temperature control loop. This circuit constantly monitors the oxygen sensor temperature, compares it to the set point temperature (700°C), and turns the heater ON or OFF accordingly. The specific type of control circuit used is called a Proportional Integral Derivative (PID) loop, because of the three adjustable parameters involved:
Proportional Band: Because the system cannot respond
instantaneously to temperature changes, the actual temperature of the oxygen sensor oscillates about the set point. In general, increasing the proportional band reduces the magnitude of these temperature oscillations.
Integral Action: A consequence of increasing the proportional
band is the introduction of an offset between the set point and the control point. The integral portion of the control loop acts to move the control point back toward the set point within a specified period of time. Thus, decreasing this integration time reduces the offset more quickly.
Derivative Action: The derivative portion of the control loop
applies a corrective signal based on the rate at which the actual temperature is approaching the set point. In effect, the derivative action reduces overshoot by counteracting the control signal produced by the proportional and integral parameters.
The heater control circuit is configured at the factory for optimum performance. Because of the strong interaction between the three parameters involved, properly setting up the PID loop is a very complex matter. As a result, randomly changing the P, I and/or D parameters can seriously degrade the performance of the OxyTrak™
390.
IMPORTANT: Always contact the factory before attempting to
change the default P, I and/or D values.
General Information 1-11

Chapter 2

Installation

Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Installation Site. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
September 2007

Introduction This chapter provides instructions on how to properly install and wire

the OxyTrak™ 390. Be sure to observe all installation limits and precautions described in this chapter. Pay particular attention to the specified ambient temperature range of –30 to +70°C (-22 to +158°F) for the analyzer and –30 to +60°C (-22 to +140°F) for the controller.
!WARNING!
To ensure safe operation, the OxyTrak™ 390 must be
installed and operated as described in this manual. Also, be
sure to follow all applicable local safety codes and
regulations for installing electrical equipment. All
procedures should be performed by trained service
personnel only.

Unpacking Remove the analyzer (see Figure 2-1 below) from its shipping

container, and make sure that all items on the packing slip have been received. If anything is missing, contact the factory immediately.
Note: See Figure 2-6 on page 2-17 (local controller) or Figure 2-7
on page 2-18 (remote controller) for a complete outline and installation drawing of the OxyTrak™ 390.
Figure 2-1: Typical OxyTrak™ 390 with Local Controller
Installation 2-1
September 2007

Installation Site Environmental and installation factors should already have been

discussed with a GE Sensing applications engineer or field sales person before the OxyTrak™ 390 arrives.
Selecting the Site The tip of the probe is typically inserted into the stack to a distance of
1/3 of the stack diameter. Also, the flue gas flow direction should be either perpendicular to the probe or angled away from the open end of the probe (see Figure 2-2 below).
IMPORTANT: Never allow the flue gas flow to be angled directly into
the end of the probe.
For furnaces, locate the analyzer close to the combustion zone,
typically within the radiant section and always before the convection section. Make sure that the probe’s maximum operating temperature is not exceeded and that the probe is not situated in a non-homogeneous flue gas mixture.
For boilers, locate the analyzer downstream of the heat exchanger
and just before the economizer air heater, if one is installed. The analyzer should not be placed downstream of any air heater, because of possible air leaks that can cause inaccurate readings.
In general, the sample point should be an area of high turbulence, which will ensure a good homogeneous mixture of the flue gases. Conditions to be avoided would include air leaks upstream of the sample point and dead spaces in the vicinity of the sample point.
Mounting Flange
Shroud
90°
90°
Probe
Probe Sleeve
Figure 2-2: Permitted Flue Gas Flow Angles
2-2 Installation
September 2007
Selecting the Site (cont.) Finally, the following installation requirements should be observed:
Install the OxyTrak™ 390 in a location that provides ready access
for programming, testing, and servicing the unit.
Protect all cables from excessive physical strain (bending, pulling,
twisting, etc.).
Be sure that the input voltage at the planned installation site is
within the limits specified for the OxyTrak™ 390.
Preparing the Site Preparation of the installation site should include the following steps
(see Figure 2-8 on page 2-19 and Figure 2-3 below):
Note: Although a horizontal installation is shown in this manual,
other mounting angles are permissible.
FRONT VIEW
GAS FLOW
SIDE VIEW
3" min
Welds
2" Sch 80 Pipe
.
Mounting Plate
1 7/8"
min.
90°
Masonry
Wall
Mating Flange
Figure 2-3: A Typical Installation Setup
Installation 2-3
September 2007
Preparing the Site (cont.) 1. At the chosen analyzer location on the furnace or boiler wall or on
the side of a horizontal or vertical flue duct, drill a hole of the proper diameter to accommodate a short length of pipe having at least a 1 7/8 in. 48 mm) inside diameter. A length of 2” Schedule 80 pipe is suitable for this purpose.
2. Weld the short pipe into a mounting plate, with welds on both sides of the plate. The pipe length must be sufficient to meet the following requirements:
One end of the pipe should extend through the rear of the
mounting plate sufficiently to enter the wall. For installation in a masonry wall, the pipe should extend entirely through the wall to prevent the probe from becoming trapped, if the wall should crumble.
To provide clearance for installing the flange bolts, the pipe
must be long enough to provide 4 in. (100 mm) of clearance between the front surface of the mounting plate and the back surface of the mating flange.
3. Weld the mating flange onto the end of the short pipe so that the raised face of the flange faces away from the mounting plate. Be sure that the following requirements are met:
One end of the short pipe should be flush with the raised face of
the flange.
The mating flange should be oriented so that its bolt holes
straddle the vertical and horizontal center lines of the mounting plate.
Note: The OxyTrak™ 390 can be supplied with an optional flange. If
a flange is desired, it must be specified (e.g. 3”-150# flange) at the time of purchase
4. Attach the mounting plate to the wall with the pipe extending into the drilled hole.
For probe lengths greater than 2 meters (6 feet), a support sleeve is recommended. Refer to Figure E-3 on page E-3.
2-4 Installation
September 2007

Mounting This section explains how to mount OxyTrak™ 390 analyzer at the

site that was prepared in the previous section. The OxyTrak™ 390 has integral male 1-1/2” NPT mounting threads. This permits a flange to be threaded onto the analyzer, and the resulting assembly is then bolted to the mating flange on the furnace/boiler wall or flue duct.
Caution!
Flue gas condensate is extremely corrosive. The OxyTrak™
390 must be wired and powered up immediately after
mounting to prevent damage to the unit. If a blowback
(purge) system is to be used, install this system and turn it on
right away also.
IMPORTANT: Direct mounting of the OxyTrak™ 390 into a threaded
hole using its mounting thr e ads is not Always use a mounting flange.
Note: Rather than the use of a thread sealant, a high temperature
lubricant such as Molykote 1000 is recommended.
recommended.
Refer to Figure 2-8 on page 2-19, and complete the following steps to mount the OxyTrak™ 390 convection loop/analyzer package:
1. Slide a suitable flange gasket over the probe and up against the mounting flange on the analyzer.
Note: Be sure to use a suitable high temperature gasket for this
application.
2. Orient the analyzer so that the convection loop/ana lyzer package is vertical, and slide the probe through the hole in the mounting wall until the two flanges meet.
3. Using suitable hardware, make sure the gasket is properly positioned between them, and bolt the two flanges together.
4. Continue as follows: a. If you have a local display controller, the physical installation
is complete. Proceed to the wiring section on the next page.
b. If you have a remo te display controller, proceed to Step 5.
5. Refer to Figure 2-7 on page 2-18 and mount the remote display
controller in a convenient location. Be sure to allow sufficient clearance for programming and operation of the unit.
6. Install suitable cable glands and conduit for the environment, to connect the junction box on the bottom of the convection loop/ analyzer package to the display controller (2 places).
Installation 2-5
September 2007

Wiring !Attention European Customers!

To meet CE Mark requirements, install all cables as
described on the next page.
IMPORTANT: For compliance with the European Union’s Low
Voltage Directive (73/23/EEC), the OxyTrak™ 390 requir es an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
!WARNING!
To ensure safe operation, the OxyTrak™ 390 must be
installed and operated as described in this manual. Be sure
to follow all applicable local safety codes and regulations
for installing electrical equipment. All procedures should
be performed by trained service personnel only.
To wire the OxyTrak™ 390, see Figure 2-9 on page 2-20 for a local assembly or Figure 2-10 on page 2-21 for a remote assembly, and connect the following items to the display controller (do not
line power through the same conduit as the other connections):
run the
alarm relays A-D
calibration relays E-H
4-20 mA analog output
RS232 or RS485 output
line power (connect through the right-hand port)
If you have a remote display controller, you must also make the following connections between the controller and the junction box:
oxygen and combustibles sensors
furnace and manifold thermocouples
thermocouple cold junction compensation
furnace and manifold heaters
IMPORTANT: Do not alter any of the factory-installed wiring.
To access the terminal blocks for wiring, unthread the four screws on the front of the display controller and swing the cover open. If you have a system with a remote display controller, you must also unthread the three screws on the junction box and swing the cover open.
2-6 Installation
September 2007
CE Mark Compliance For CE Mark compliance, the OxyTrak™ 390 must meet both the
EMC and LVD directives.
IMPORTANT: CE Mark compliance is required for all units used in
EEC countries.
EMC Compliance For EMC compliance, the electrical connections must be shielded and
grounded as shown in Table 2-1 below. After all the necessary electrical connections have been made, seal any unused cable entry holes with standard conduit plugs or equivalent.
Note: If the instructions in this section are followed, the unit will
comply with the EMC Directive 89/336/EEC.
Table 2-1: Wiring Modifications for EMC Compliance
Connection Wiring Modification
Power 1. When connecting the power, select the cable
entry closest to the chassis ground.
2. Use shielded cable* to connect the power to the OxyTrak™ 390 enclosure. Connect the shield to the nearest chassis ground terminal.
3. Connect the power line ground wire to the nearest chassis ground terminal.
Input/Output 1. Use shielded cable* to interconnect the
OxyTrak™ 390 enclosure with any external I/O devices.
2. Connect the shields to the nearest chassis ground terminal.
*Wires enclosed in a properly-grounded metal conduit do not
require additional shielding.
LVD Compl iance For compliance with the European Union’ s Low Voltag e Directive
(73/23/EEC), the analyzer requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marked as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
Note: If the instructions in this section are followed, the unit will
comply with the Low Voltage Directive (73/23/EEC).
Installation 2-7
September 2007
Wiring the Analog Outputs (A-C)
Wiring the Alarm Relays (A-D)
To wire an analog output device to the OxyTrak™ 390, refer to Figure 2-4 on page 2-12 and Figure 2-9 on page 2-20 or Figure 2-10 on page 2-21, and make the following connections to terminal block
J11 in the display controller:
1. Connect the positive pin to the input of the analog output device: a. Output A - J11–5 (+) b. Output B - J11–3 (+) c. Output C - J11–1 (+)
2. Connect the negative pin to the return of the analog output
device:
a. Output A - J11–6 (–) b. Output B - J11–4 (–) c. Output C - J11–2 (–)
To wire a warning device to any of the OxyTrak™ 390 ala rm relay s (A-D), refer to Figure 2-4 on page 2-12 and Figure 2-9 on page 2-20 or Figure 2-10 on page 2-21, and make the following connections to terminal blocks J7 and J8 in the display controller:
1. Connect the NC pin to the alarm device input for failsafe operation, or leave this pin unused for non-failsafe operation:
a. Relay A - J7–4 (NC) b. Relay B - J7–1 (NC) c. Relay C - J8–4 (NC) d. Relay D - J8–1 (NC)
2. Connect COM pin to the alarm device return: a. Relay A - J7–6 (COM) b. Relay B - J7–3 (COM) c. Relay C - J8–6 (COM) d. Relay D - J8–3 (COM)
3. Connect the NO pin to the alarm device input for non-failsafe
operation, or leave this pin unused for failsafe operation:
a. Relay A - J7–5 (NO) b. Relay B - J7–2 (NO) c. Relay C - J8–5 (NO) d. Relay D - J8–2 (NO)
2-8 Installation
September 2007
Wiring the Calibration Relays (E-H)
To wire a warning device to any of the OxyTrak™ 390 calibrati on relays (E-H), refer to Figure 2-4 on page 2-12 and Figure 2-9 on p age 2-20 or Figure 2-10 on page 2-21, and make the following connections to terminal blocks J9 and J10 in the display controller:
1. Connect the NC pin to the alarm device input for failsafe
operation, or leave this pin unused for non-failsafe operation:
a. Relay E - J9–3 (NC) b. Relay F - J9–6 (NC) c. Relay G - J10–3 (NC) d. Relay H - J10–6 (NC)
2. Connect the COM pin to the alarm device return:
a. Relay E - J9–1 (COM) b. Relay F - J9–4 (COM) c. Relay G - J10–1 (COM) d. Relay H - J10–4 (COM)
3. Connect the NO pin to the alarm device input for non-failsafe
operation, or leave this pin unused for failsafe operation:
a. Relay E - J9–2 (NO) b. Relay F - J9–5 (NO) c. Relay G - J10–2 (NO) d. Relay H - J10–5 (NO)
Note: The OxyTrak™ 390 relays do not provide power. To use the
Blow Back process, connect a power supply in series with Relay H and the Blow Back solenoid valve.
Installation 2-9
Loading...
+ 132 hidden pages