• Intrinsically safe option for Class 1 Div 1 and
Div 2 locations
• Ex ameproof option
• Proven galvanic fuel cell O2 sensor technology
• User selectable ranges for ppm and percent
oxygen
• User-friendly and intuitive user interface with
diagnostics
• Microprocessor-based, all-digital technology for
reliable operation
• Low maintenance, economical and compact
• Sensor failure output error
bhge.com
• Sensor lifetime indication
• NAMUR error indication
Panametrics Oxygen Transmitter
Control at the Tip of a Finger
The oxy.IQ is a highly reliable and cost-eective two-wire,
loop-powered transmitter with a linearized
4 to 20 mA output. It measures oxygen in ten ppm ranges
and seven percentage ranges. All ranges are userselectable. This compact transmitter uses proven sensor
technology to accurately measure O2 in a variety of gases,
even in hazardous environments.
Proven Sensor Technology
The oxy.IQ’s oxygen sensor is an advanced galvanic
fuel cell that provides superior performance, accuracy,
stability and long life. The cell’s innovative design
eliminates the potential for negative signal output,
reduces sources of contamination.
The cell is unaected by other background gases or
hydrocarbons and is compatible with acid gases
(OX-2 and OX-4 cells). Recovery from air at low ppm levels
takes just a few minutes. Because the cell is
self-contained, minimal maintenance is required. There is
no electrolyte to change or electrodes to clean.
Intrinsically Safe
When equipped with optional zener barriers or a galvanic
isolator, the oxy.IQ can be mounted in a hazardous
(classied) location.
The rugged oxy.IQ is controlled by a microprocessor that
enables the user to select the range, trim outputs and
perform calibration. Improved user interface includes
user-selectable and programmable ranges, sensor
diagnostics including a sensor lifetime feature and
sensor raw signal, low reading/sensor failure signal and
temperature compensation. Programming is easily done
using the keypad and display. These keypad buttons allow
complete functionality in hazardous areas when equipped
with a zener barrier or a galvanic isolator per drawing
752-347.
Installation Flexibility
The compact oxy.IQ, with its built-in microprocessor,
is designed to t easily into any installation site. The
oxy.IQ can be installed right at the sampling point,
whereas other transmitters must be rack or panel
mounted.
Sample Systems
In addition to standard features and options, GE oers
a full line of sample handling systems for a variety
of applications. If needed, GE can design and build a
sample conditioning system to meet unique application
requirements. Please contact GE for details.
1:2 ratio, ppm levels only. For example, 100 ppm NO2 looks like 200 ppm O
(3) Minimaleectonsensorperformance
OX-2,
ppm
Cont.
OX-3, %
Cont. Int. (1)
OX-4%
Cont. Int.
2
oxy.IQ Specications
Intrinsically Safe (IS) Installation
Intrinsically safe installations require a zener barrier, one
IS cable and one non-IS cable.
Power Requirements
24 to 28 VDC at 50 mA
Cable
• PN 704-1318: blue jacket, twisted pair with connector;
26 AWG; 2 or 10 m length
Output
Total load must equal 250 W ±5% when using zener
barrier
General Purpose and Non-Incendive
(Div 2) Installations
Standard option package or IS package,
no Zener barrier or galvanic isolator is required.
Power Requirements
9 to 28 VDC loop powered, 0.7 W maximum
Cable
• PN 704-1317: black jacket, twisted pair with connector;
26 AWG; 2 or 10 m length
All Installations
Process Wetted Materials
• SS process unit: 316 stainless steel, Viton® O-ring,
gold plated sensor electrical contacts and glass
User-Selectable Measurement Ranges
• PPM sensors:
– 0 to 10 ppmv O2 (OX-1 or OX-2 only)
– 0 to 20 ppmv O2 (OX-1 or OX-2 only)
– 0 to 50 ppmv O2 (OX-1 or OX-2 only)
– 0 to 100 ppmv O
– 0 to 200 ppmv O
– 0 to 500 ppmv O
– 0 to 1000 ppmv O
– 0 to 2000 ppmv O
– 0 to 5000 ppmv O
– 0 to 10,000 ppmv O
• Percent sensors:
– 0% to 1% O
– 0% to 2% O
– 0% to 5% O
– 0% to 10% O
– 0% to 25% O
– 0% to 50% O
2
2
2
2
2
2
2
2
2
2
2
2
2
oxy.IQ Specications
All Installations
Accuracy
• ±1% of range at calibration point
• ±2% of range at the calibration point for the
0 to 10 ppmv range (OX-1 or OX-2 only)
Repeatability
• ±1% of range
• ±2% of Range for the 0 to 10 ppmv range
(OX-1, 2 only)
Resolution
±0.1% of range
Linearity
±2% of range (OX-1, 2, 3, 5)
±5% of range (OX-4)
O2 Sensor Operating Temperature
32 to 113°F (0 to 45°C)
Sample Pressure
Vented to atmosphere during operation and calibration
Dimensions
4.10 x 2.75 x 2.05 in. (104.1 x 69.9 x 52.1 mm)
Weight
1.35 lb (612 grams)
Sample Flow Rate
1.0 SCFH (500 cc/min) recommended for process units
Electrical Classication/Certication
Intrinsically Safe
USA/Canada
IS for Class I, Div 1, Groups ABCD, T4
AEx ia IIC T4
ATEX and IECEx
Ex ia IIC Ga T4 Tamb -20 to 60°C
IS package, non-incendive without use of Zener
barrier or galvanic isolator:
• USA/Canada: Class 1, Div 2, Groups A, B, C, D; T4
• ATEX/IECEx: Ex na IIC T4
European Compliance
Complies with EMC Directive 2004/108/EC
Atmospheric Pressure Eect
±0.13% of reading per mmHg (directly proportional to
absolute pressure). During calibration, pressure and ow
must be kept constant.
Process Connection
• 1/8 in. NPT inlet and outlet
Dimensions
1/8-27NPT-2B
2 X
1.58 [40.1]
4.10 [104.1]
8-32 UNF-2B
2 X
2.75
[69.9]
2.05 [52.1]
1.33 [33.8]
1.03 [26.2]
1.00 [25.4]
.30
[7.7]
Ø1.75 [44.5]
Note: all dimensions are inches [mm]
Installation Options
Hazardous LocationNon-Hazardous Location
ZBB BUS BAR
POWER SUPPLY
oxy.IQ Transmitter
Class 1 Div 2 Enclosure
oxy.IQ Transmitter
BROWN
BLUE
3
42
GREEN
1
RED
GREEN
BLACK
MTL7706
IS Cable
704-1318-xx
(Blue Jacket)
1. Equipment connected to barrier
inputs must not use or generate
more than 250V.
2. Total load of R1 + R2 must equal
250 ohms ±5%.
IS GROUND
Non-IS Cable
Hazardous LocationNon-Hazardous Location
BLUE
BROWN
704-1317-xx
(Black Jacket)
BROWN
+24V
24V RETURN
4-20MA ANALOG
INPUT DEVICE
R1
BLACK
R2
POWER SUPPLY
+24V
24V RETURN
4-20MA ANALOG
INPUT DEVICE
1. For Class 1 Div 2 installation, the oxy.IQ shall be installed in
an appropriately rated final enclosure accepting a Division 2
wiring method per NEC/CEC.
oxy.IQ Transmitter
704-1317-xx
(Black Jacket)
Maximum load of analog input device is
dependent on power supply voltage
and cable resistance.
BLUE
BROWN
Maximum load of analog input device is
dependent on power supply voltage
and cable resistance.
BROWN
POWER SUPPLYNon-Hazardous Location
+24V
24V RETURN
4-20MA ANALOG
INPUT DEVICE
Ordering Information
Record selected option in blank spaces indicated at bottom of form.
OXY.IQ-BCD-E Option Code
A - Model Only
oxy.IQ Oxygen Transmitter; 4 to 20 mA output
B - Sensor
0 No sensor
1 Standard ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
2 Acid ppm, 0 to 10, 20, 50, 100, 200, 500, 1000 ppm
3 Standard percent sensor
4 Acid percent sensor
5 Standard ppm, 0 to 100, 200, 500 and 1000 ppm
C - Package 1 Standard
3 Intrinsically safe (US/CAN Class 1 Div 1) or Non-incendive (US/CAN Class 1 Div 2)
4 Ex ameproof
D - Cable Length
0 No cable
1 2 meter cable
2 10 meter cable
E - Zener Barrier0 None
1 Zener barrier
2 Galvanic isolator
Note: For Class 1 Div 1, either zener barrier or galvanic isolator must be
selected. For Class 1 Div 2, no barriers needed. Please refer to dwg 752-347
for installation guidelines.
__ – __ ___ ___ ___ [Use this number to order product]
Baker Hughes reserves the right to make changes in specications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your BHGE
representative for the most current information. The Baker Hughes logo is a trade mark of Baker Hughes, a GE company. The GE Monogram is a trademark of the General Electric Company.
920-603D
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.