GE Sensing oxy.IQ Data sheet

oxy.IQ
Panametrics Oxygen Transmitter
Applications
An oxygen transmitter for use in:
Glove box purge and leak detection
Natural gas
Semiconductor wafer machines
Membrane air separators
Inert welding gases
Pure gaseous hydrocarbon streams
Process monitoring of gaseous monomers
Heat treating and bright annealing
Features
Two-wire, loop-powered 4 to 20 mA transmitter
Display with keypad
Intrinsically safe option for Class 1 Div 1 and Div 2 locations
Ex ameproof option
Proven galvanic fuel cell O2 sensor technology
User selectable ranges for ppm and percent oxygen
User-friendly and intuitive user interface with diagnostics
Microprocessor-based, all-digital technology for reliable operation
Low maintenance, economical and compact
Sensor failure output error
bhge.com
Sensor lifetime indication
NAMUR error indication
Panametrics Oxygen Transmitter
Control at the Tip of a Finger
The oxy.IQ is a highly reliable and cost-eective two-wire,
loop-powered transmitter with a linearized 4 to 20 mA output. It measures oxygen in ten ppm ranges and seven percentage ranges. All ranges are user­selectable. This compact transmitter uses proven sensor technology to accurately measure O2 in a variety of gases, even in hazardous environments.
Proven Sensor Technology
The oxy.IQ’s oxygen sensor is an advanced galvanic fuel cell that provides superior performance, accuracy, stability and long life. The cell’s innovative design eliminates the potential for negative signal output, reduces sources of contamination.
The cell is unaected by other background gases or
hydrocarbons and is compatible with acid gases (OX-2 and OX-4 cells). Recovery from air at low ppm levels takes just a few minutes. Because the cell is self-contained, minimal maintenance is required. There is no electrolyte to change or electrodes to clean.
Intrinsically Safe
When equipped with optional zener barriers or a galvanic isolator, the oxy.IQ can be mounted in a hazardous
(classied) location.
The rugged oxy.IQ is controlled by a microprocessor that enables the user to select the range, trim outputs and perform calibration. Improved user interface includes user-selectable and programmable ranges, sensor diagnostics including a sensor lifetime feature and sensor raw signal, low reading/sensor failure signal and temperature compensation. Programming is easily done using the keypad and display. These keypad buttons allow complete functionality in hazardous areas when equipped with a zener barrier or a galvanic isolator per drawing 752-347.
Installation Flexibility
The compact oxy.IQ, with its built-in microprocessor, is designed to t easily into any installation site. The oxy.IQ can be installed right at the sampling point, whereas other transmitters must be rack or panel mounted.
Sample Systems
In addition to standard features and options, GE oers
a full line of sample handling systems for a variety of applications. If needed, GE can design and build a sample conditioning system to meet unique application requirements. Please contact GE for details.
Advanced Galvanic Fuel Cell Sensor
Oxygen Sensor Interference Gases
Gas OX-1 and OX-
5, ppm Cont.
H2S <5 ppm <10 ppm 0.0005% 0.01% 0.001% 0.1%
SO3<10 ppm <10 ppm 0.01% 0.1% 0.01% 0.1%
SO2<10 ppm (3) 0.01% 0.1% (3) (3)
HCI <1000 ppm (3) 0.1% 1.0% (3) (3)
HCN <1000 ppm (3) 0.1% 1.0% (3) (3)
CO2<1000 ppm (3) 0.1% 20% (3) (3)
NO2(2) (2) (2) (2) (2) (2)
CL2(2) (2) (2) (2) (2) (2)
Cont. = Continuous, Int. = Intermittent
(1) Recommendedmaximumexposure30minutesfollowedbyushingwith
ambient air for equal period
(2) Minimaleectonsensorperformance,butproducessignalinterferenceof
1:2 ratio, ppm levels only. For example, 100 ppm NO2 looks like 200 ppm O
(3) Minimaleectonsensorperformance
OX-2, ppm Cont.
OX-3, %
Cont. Int. (1)
OX-4%
Cont. Int.
2
oxy.IQ Specications
Intrinsically Safe (IS) Installation
Intrinsically safe installations require a zener barrier, one IS cable and one non-IS cable.
Power Requirements
24 to 28 VDC at 50 mA
Cable
PN 704-1318: blue jacket, twisted pair with connector; 26 AWG; 2 or 10 m length
Output
Total load must equal 250 W ±5% when using zener barrier
General Purpose and Non-Incendive (Div 2) Installations
Standard option package or IS package, no Zener barrier or galvanic isolator is required.
Power Requirements
9 to 28 VDC loop powered, 0.7 W maximum
Cable
PN 704-1317: black jacket, twisted pair with connector; 26 AWG; 2 or 10 m length
All Installations
Process Wetted Materials
• SS process unit: 316 stainless steel, Viton® O-ring, gold plated sensor electrical contacts and glass
User-Selectable Measurement Ranges
• PPM sensors:
– 0 to 10 ppmv O2 (OX-1 or OX-2 only) – 0 to 20 ppmv O2 (OX-1 or OX-2 only) – 0 to 50 ppmv O2 (OX-1 or OX-2 only) – 0 to 100 ppmv O – 0 to 200 ppmv O – 0 to 500 ppmv O – 0 to 1000 ppmv O – 0 to 2000 ppmv O – 0 to 5000 ppmv O – 0 to 10,000 ppmv O
• Percent sensors:
– 0% to 1% O – 0% to 2% O – 0% to 5% O – 0% to 10% O – 0% to 25% O – 0% to 50% O
2
2
2
2
2
2
2
2
2
2
2
2
2
Loading...