GE Sensing O2X1 Operating Manual

GE Infrastructure
Sensing
Model O2X1
Oxygen Transmitter
User’s Manual
GE Infrastructure
Sensing
Model O2X1
Oxygen Transmitter
916-048G September 2004
September 2004
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Mounting the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Wiring the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Making a Custom Transmitter Cable . . . . . . . . . . . . . . . . . . . . 6
Installing an Oxygen Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
The O2X1 Keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
The O2X1 Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Entering and Exiting the User Program. . . . . . . . . . . . . . . . . . . . 13
Programming and Calibrating the O2X1 . . . . . . . . . . . . . . . . . . 14
Selecting the Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Trimming the Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Span Gas Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Air Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Model Number. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Electrical Classification/Certification . . . . . . . . . . . . . . . . . . . . . . 23
Intrinsically Safe Installation. . . . . . . . . . . . . . . . . . . . . . . . . . 24
Non-Hazardous Installation . . . . . . . . . . . . . . . . . . . . . . . . . . 25
I
September 2004
Warranty
Each instrument manufactured by GE Infrastructure Sensing, Inc. is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE Infrastructure Sensing, Inc. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE Infrastructure Sensing, Inc. determines that the equipment was defective, the warranty period is:
one year from delivery for electronic or mechanical failures
one year from delivery for sensor shelf life
If GE Infrastructure Sensing, Inc. determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE Infrastructure Sensing, Inc. , the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu
of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course
of dealing or usage or trade).
III
September 2004
Return Policy
If a GE Infrastructure Sensing, Inc. instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE Infrastructure Sensing, Inc., giving full details of
the problem, and provide the model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE Infrastructure Sensing, Inc. will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE Infrastructure Sensing, Inc. instructs you to send your
instrument to a service center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Infrastructure Sensing, Inc. will evaluate
the instrument to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty,
the instrument will be repaired at no cost to the owner and returned.
If GE Infrastructure Sensing, Inc. determines that the
damage is if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
IV
not covered under the terms of the warranty, or
September 2004

Introduction

The GE Infrastructure Sensing O2X1 oxygen trans mitt er (F igu re 1 below) accurately measures oxygen content in a gas mixture at 0 to 250,000 ppmv (25%), in six ranges. It is loop-powered from a 9-28 VDC power source and provides data readings via a 4-20 mA analog output that is updated once per second. Programmed information is stored in a nonvolatile memory.

Figure 1: The O2X1 Oxygen Transmitter

O2X1 Oxygen Transmitter 1
September 2004
Introduction (cont.)
The O2X1 is available as either an ambient air monitor or as a process unit. The process units are easily installed by completing the following steps:
insert the oxygen sensor into the O2X1 transmitter
connect the power to the transmitter
calibrate the oxygen sensor
attach the transmitter body to the sensor manifold with the
knurled slip nut
A variety of oxygen sensors can be used with the O2X1, but a quick and easy calibration with ambient air must be performed upon startup whenever a new sensor is installed.
This guide shows how to properly mount, wire, calibrate, and adjust the O2X1 oxygen transmitter. Proceed to the appropriate section for information on the following topics:
Note: The installation instructions in this manual apply only to
O2X1 process
units.
Mounting the Transmitter
Wiring the Transmitter
Installing an Oxygen Sensor
Entering and Exiting the User Program
Programming and Calibrating the O2X1
Specifications
Electrical Classification/Certification
2 O2X1 Oxygen Transmitter
September 2004

Mounting the Transmitter

T o install the O2X1 oxygen transmitter into the process or sample system, refer to Figure 2 below and proceed to the next page.
Connector Key
Black: Return (-)
2.75 [70]
Red: Signal (+)
Transmitter Module
LEDs
7.26 [184]
Knurled Nut
Oxygen Sensor
Manifold
1/8 NPT Thread
2.24 [57]
0.50 [13] 8-32 UNC-2B
.30 Deep, 2 Plcs.
1.00 [25]
Figure 2: Outline and Installation Drawing
(316 SS and Delrin Process Units)
O2X1 Oxygen Transmitter 3
September 2004
Mounting the Transmitter (cont.)
Note: To avoid collecting condensate that may damage the
oxygen sensor, mount the O2X1 in an upright position, with the sensor manifold below the electronics module.
Install the O2X1 transmitter by completing the following steps:
1. Remove the O2X1 transmitter and the separately-packaged oxygen sensor (see Figure 3 below) from the shipping container. Keep the shipping container and packaging material for possible future use.

Figure 3: Packaged Oxygen Sensor

4 O2X1 Oxygen Transmitter
September 2004
Mounting the Transmitter (cont.)
2. Using Teflon tape as a sealant, connect the sample gas inlet
and outlet to the 1/8” NPT ports on the sensor manifold (see Figure 4 below). Either port may be used as the inlet or the outlet; the direction of flow does not matter.
Sensor Manifold
Sample Inlet Sample Outlet
Teflon Tape

Figure 4: Sensor Manifold Installation

O2X1 Oxygen Transmitter 5
September 2004

Wiring the Transmitter

T o wire the O2X1 oxygen transmitter refer to F igure 5 on the next page and Figure 6 on page 8. Then, proceed as follows:
!WARNING!
For FM/CSA/BASEEFA certified applications, the O2X1
must be installed with a zener barrier (see Figure 5 on
the next page). Also, for installations in a hazardous
location, the blue OCI (5) T5 cable must be used.
1. A standard 5 ft long, two-wire cable is supplied with the
O2X1. Screw the cable’s five-pin connector to the mating connector on the O2X1 unit.
Note: Longer cables are available for quotation.
2. Connect the other end of the cable according to one of the
following steps: a. No Zener Barrier: Attach the red wire to the positive
terminal of a 9-28 VDC power supply. Attach the black lead to the negative terminal of the power supply.
b. With Zener Barrier: Attach the red and black wires to the
zener barrier, as shown in Figure 6 on page 8.

Making a Custom Transmitter Cable

T o construct a transmitter cable longer than the one supplied with the unit (up to a maximum of 1,100 ft), refer to Note 1 in Figure 5 on the next page and the specifications on pages 24 and 25. Solder the two leads on one end of the cable to the connector provided by GE Infrastructure Sensing. For the pin connections on the transmitter, see Figure 2 on page 3.
Note: The orientation of the connector depends on how the
transmitter is installed. Use the connector key shown in Figure 2 on page 3 as a reference when making lead connections.
6 O2X1 Oxygen Transmitter
Wiring the Transmitter (cont.)
September 2004
HAZARDOUS LOCATION
(CLASS I, II, III, DIV 1, GROUP A, B, C, D, E, F, G)
See Note 1
O2X1
NON-HAZARDOUS LOCATION
MTL706
BARRIER
(SEE NOTE 2)
NOTE 4
SEE
1. Cable Requirements
A.FM Approved Installations
inductance due to the cable cannot exceed the following values:
Group A, B
0.068 µF 0.30 µF0.90 µF
4.0 mH 18 mH 35 mH
B.CSA Installations
- Refer to the Canadian Electrical Code for
allowable cable parameters and lengths.
2. Barriers must be installed in accordance with the barrier manufacturer’s specifications
3. Installations should be in accordance with ANSI RP12.6, “Installations of Intrinsically Safe Systems for Hazardous (Classified) Locations” and the National Electrical Code (ANSI/NFPA 70).
4. Control room equipment connected to the associated apparatus (barrier inputs) must not use or generate more than 250 V.
5. CSA installations must be installed in accordance with the Canadian Electrical Code, Part 1.
6. WARNING - SUBSTITUTION OF COMPONENTS MAY IMPAIR SAFETY. AVERTI SSEMENT - LA S UBSTITUTI ON DE COMPOSANTS PEUT COMPROMETTRELA SECURITE INTRINSIQUE
(per Drawing #752-099)
- Total cable capacitance and
Group C, E Group D, F, G
Figure 5: Typical Certification Requirements for
316 SS Process Unit Installations
O2X1 Oxygen Transmitter 7
September 2004
Wiring the Transmitter (cont.)
With Zener Barrier
Without Zener Barrier

Figure 6: O2X1 Wiring Diagrams

8 O2X1 Oxygen Transmitter
September 2004

Installing an Oxygen Sensor

T o install a new or replacement oxygen sensor in the O2X1, refer to Figure 7 below and complete the following steps:
Sensor Base
Knurled Nut
Oxygen Sensor Ring
Sensor Manifold

Figure 7: Oxygen Sensor Installation

1. Disconnect the power from the O2X1
2. Loosen the knurled slip nut and remove the O2X1 body from
the sensor manifold. If a previous oxygen sensor is already in place, remove it now.
3. Apply power to the unit. The O2X1 initializes in about 15 seconds and then all three LEDs light momentarily.
O2X1 Oxygen Transmitter 9
September 2004
Installing an Oxygen Sensor (cont.)
Note: Before continuing with the installation, become familiar
with the procedures for programming and calibrating the O2X1 oxygen transmitter discussed later in this manual.
4. Trim the 4-20 mA analog output (see page 15) and set the range to 0-25% oxygen (see page 14).
5. Open the airtight package (see Figure 4 on page 5) and remove the oxygen sensor from the package. To ma intain the oxygen sensor’s energy level, remove the red grounding tab and immediately proceed to the next step.
6. Orient the sensor so that its gold-plated electrodes are facing the spring-loaded contact pins in the sensor base (see Figure 7 on the previous page). Firmly press the oxygen sensor into the sensor base on the O2X1 transmitter.
7. It is best to perform an air calibration on the new oxygen sensor at this time (see page 18). On the 0-25% oxygen scale, a properly calibrated oxygen sensor generates a current of
17.4 mA at the 4-20 mA analog output terminals.
8. Using the knurled slip nut, attach the transmitter with the calibrated oxygen sensor to the sensor manifold. Rotate the transmitter as desired and hand-tighten the knurled slip nut.
9. Begin the flow of either the process or the calibration gas. The analog output reading should begin to drop as the oxygen sensor adjusts to the reduced oxygen level. During this time, reset the range as required.
10. For improved accuracy in the ppm oxygen ranges, a span gas
calibration should now be performed (see page 16).
IMPORTANT: Sensor life is dependent on the application. High
oxygen concentrations and/or contaminants such as acid gases shorten sensor life.
10 O2X1 Oxygen Transmitter
September 2004

The O2X1 Keypad

The O2X1 keypad consists of three keys and three light-emitting diodes (LEDs), as shown in see Figure 1 o n page 1. Table 1 below lists all of the keypad components and shows the symbols used to represent them in this document.
Note: During programming, all keystrokes must be completed
within 60 seconds of the previous keystroke.
Table 1: The O2X1 Keys and LEDs
Symbol Component Name
Mode Set Key
Up Arrow Key
Down Arrow Key
5 1 5 2 5 3
LED #1 LED #2 LED #3

The O2X1 Menu Map

As an aid in navigating through the User Program, a complete Menu Map of the program is presented in Figure 8 on the next
page. Please refer to this figure as needed, while programming the O2X1 transmitter.
O2X1 Oxygen Transmitter 11
September 2004
(Press and Hold
for 3 seconds)
U N E
M N
I A
M
MEASUREMENT
23
1
23
1
RANGE SELECT
132321
Current Range
(see table at right)
OUTPUT TRIM
1 32
SPAN GAS CALIBRATION
1 32
Expose
Sensor to
Span Gas*
* before entering Main Menu
123
MODE
20 mA Trim
(read with ammeter)
4 mA Trim
(read with ammeter)
All Other Cals
1st & 2nd Cals
* current <80% of stored value
New Range
(see table at right)
132
Output = 20 mA
132
Output = 4 mA
Stores Value as Reference
or <50 microamps.
= Press Mode Set Key = Instructions for User
= Press Up Arrow Key
= Press Down Arrow Key
321 321 321
= 0-10 ppm
321 321 321
= 0-100 ppm = 0-10,000 ppm
RANGE DISPLAYS
321
NOTE: All LEDs turn OFF during output trim.
132
Read Output
with Ammeter
TEXT
123
= LED Status = "Blinking" LED
= "ON" LED = "OFF" LED
= 0-1,000 ppm
Use ammeter to read output value.
Passed
Weak
1132
Sensor*
= 0-10%
= 0-25%
32
AIR CALIBRATION
1 2 3
Expose
Sensor to
Exit Without Saving Changes
EXIT
Save Changes and Exit
All Other Cals
Air*
1st & 2nd Cals
Wait 5 Seconds
Stores Value as Reference
Stop Pressing Keys
for 60 Seconds
(or cut power for 5 sec.)
Passed
132
1 2 3
Weak
Sensor*
MEASUREMENT
132
132
123
MODE

Figure 8: O2X1 Menu Map

12 O2X1 Oxygen Transmitter
September 2004

Entering and Exiting the User Program

Table 2 below shows how to use the O2X1 keypad to na vigate through the User Program.
Table 2: O2X1 Menu Navigation
Operation Key/LED Sequence
Enter the Main
Menu
1. Press and hold the key until
ON (about 3 seconds).
turns
5 3
Move to next
item in Main
Menu
Save changes
and exit
Exit without
saving changes
2. Press the key, and
3. Press the key, and
5 2 turns ON.
5 1 starts
blinking. This indicates that you have entered the Main Menu.
Note: The O2X1 operates on a 4-20 mA
loop current, which varies as the LEDs turn
ON and OFF. Therefore,
do not take any measurements while in the User Program.
Press the key until the appropriate LED status is obtained (see Figure 8 on page 12).
While in the Main Menu, press and hold down both the and keys until all
three LEDs begin blinking. Then release the keys. After your changes have been saved, all three LEDs will turn
OFF.
Note: If any of the LEDs are still blinking,
you are still in the Main Menu.
Stop pressing keys and wait until all three LEDs turn
OFF (about 60 seconds), or
disconnect the power for at least 5 seconds.
The O2X1 User Program consists of a Main Menu and four submenus. A complete Menu Map of the program is shown in Figure 8 on page 12.
O2X1 Oxygen Transmitter 13
September 2004

Programming and Calibrating the O2X1

Upon startup, a five-step adjustment and calibration procedure must be performed on the O2X1:
1. Select the desired measurement range.
2. Trim the low (4 mA) and high (20 mA) analog outputs.
3. Upon installation of a new oxygen sensor, calibrate the unit
with air for either a ppm or % sensor.
4. For ppm sensors only, purg e the sensor with a low ppm oxygen gas.
5. For all subsequent calibrations, use a span gas that is appropriate for the sensor and range selected.
To begin the above tasks, proceed to the next section.

Selecting the Range

1. Scroll through the Main Menu until 5 1 is blinking, and
press the key to display the current range. See Table 3 below to determine the current range from the status of the LEDs.
Table 3: LED Range Codes
5 1 5 2 5 3
OFF ON ON 0-10 ppm
ON OFF OFF 0-100 ppm OFF ON OFF 0-1,000 ppm OFF OFF ON 0-10,000 ppm
ON ON OFF 0-10%
ON ON ON 0-25%
2. To select the desired range, press the or key to scroll through the available options until the desired range is displayed via the LED codes.
3. When done, press the key. The blinking of that you have returned to the Main Menu.
14 O2X1 Oxygen Transmitter
Range
5 1 indicates
September 2004

Trimming the Output

To trim the analog output, calibrate the low (4 mA) end of the output then the high (20 mA) end of the output.
IMPORTANT: The low and high adjustments interact with each
other. Therefore, r echeck the calibration after the procedure has been completed.
Preparing to Trim the Output:
Prepare to trim the analog output as follows:
1. Connect an ammeter in series with the positive O2X1 power supply lead, to monitor the analog output.
2. Enter the Main Menu as described in Table 2 on page 13.
3. Press the key repeatedly until
5 2 blinks. This is the
beginning of the Output Trim submenu.
Trimming the Output Low (4 mA) End:
1. Press the key. All LEDs turn
OFF and the analog output is
driven to approximately 4 mA.
2. Press the or key to adjust the output up or down, until it equals 4.00 ± 0.01 mA.
3. When done, press the key.
5 2 blinks, indicating that
you are back at the Main Menu.
Trimming the Output High (20 mA) End:
1. Press the key. All LEDs turn
OFF and the analog output is
driven to approximately 20 mA.
2. Press the or key to adjust the output up or down, until it equals 20.00 ± 0.01 mA.
3. When done, press the key.
5 2 blinks, indicating that
you are back at the Main Menu.
O2X1 Oxygen Transmitter 15
September 2004
Trimming the Output (cont.)
Completing the Trim Procedure:
1. Repeat both the low end and high end analog output trimming steps until no further trimming adjustments are required.
2. Exit the Main Menu and Save your changes, as described in Table 2 on page 13.

Span Gas Calibration

Before entering the Main Menu, use a low oxygen content purge gas to prepare the O2X1. Then, start the flow of the span gas to the sensor and wait until the output reading is stable. After initial exposure to the calibration gas, obtaining a stable reading takes a few seconds in the higher ranges (0-1,000 ppm and above) and a minute or more in the lower ranges (0-10 and 0-100 ppm). For accurate calibration, the span gas should have an oxygen content of 70-90% of the range being calibrated.
To perform the calibration, complete the following steps:
1. Use the equation below to calculate the expected mA output that corresponds to the known oxygen content of the span gas:
Span Gas ppm
--------------------------------------
4.0 16.0
×+ mA Output=
Full Range ppm
For example, if the span gas contains 80 ppm oxygen and the 0-100 ppm range is being calibrated, the analog output should equal 4 + 16 x (80/100) = 16.8 mA.
2. Start the flow of span gas to the sensor, and allow the 4-20 mA output reading to stabilize.
3. After the reading has stabilized, enter the Main Menu as described on page 13.
4. Press the key until
16 O2X1 Oxygen Transmitter
5 1 is ON and 5 3 is blinking.
September 2004
Span Gas Calibration (cont.)
5. For the first two calibrations after installing the transmitter,
use the key to store the calibration value in memory as a reference to detect a weak sensor. For all subsequent
calibrations, press the key to reference this value against the one stored in memory. All LEDs turn pressed.
6. While using an ammeter to monitor the 4-20 mA output, press the or key to adjust the analog output until the reading
equals the value calculated in Step 1 on the previous page.
7. When done, press the key. If the calibration was successful,
5 1 should be ON and 5 3 should be blinking.
At this point, the mA output is equal to the calibration setting of Step 5 above and must be saved. Save the calibration by
holding down both the and keys until all three LEDs flash and then turn
OFF.
OFF after the key is
Note: The second calibration after installation should be
performed within 1-2 days of the first calibration.
Note: If
5
1 and 5 3 blink alternately with 5 2, the sensor is
weak and sensor failure may be imminent.
IMPORTANT: To detect a weak sensor, the same span gas that
was used for the first two calibrations must be used for all subsequent calibrations.
If the sensor is weak and a replacement sensor is not immediately available, repeat the calibration using the key in Step 5 above.
This stores a weaker sensor calibration value into memory for temporary use. However, be aware that sensor drift may be increasing, and accuracy cannot be guaranteed. The sensor should be replaced as soon as possible.
O2X1 Oxygen Transmitter 17
September 2004

Air Calibration

An air calibration, as described below, is always recommended upon installation of a new oxygen sensor. However, because of the non-linearity of the oxygen sensor, a span gas calibration (see page 16) should also be performed to ensure more accurate readings in the ppm ranges.
Caution!
The useful life of ppm sensors will be extended by
minimizing exposure of the sensor to air.
1. Expose the oxygen sensor to ambient air for about two minutes.
2. In the Main Menu, press the key until
5 2 is ON and 5
3 is blinking.
3. For the first two calibrations after installing the transmitter, use the key to store the sensor current value in memory as a reference. For all subsequent calibrations, press the key
to reference this new value against the one stored in memory. After pressing the key,
5 2 and 5 3 will come on for about
five seconds, until the calibration is complete.
4. Reconnect the O2X1 transmitter to the process and switch back to a calibration gas of low oxygen content.
5. When
5 2 is ON and 5 3 is blinking, you are back at the
Main Menu. The new calibration must be saved by holding down both the and keys until all three LEDs flash and
then turn
OFF.
18 O2X1 Oxygen Transmitter
September 2004
Air Calibration (cont.)
Note: If 5 1 and 5 3 blink alternately with 5 2, the calibration
was rejected because the sensor current was less than 50 µA. The sensor is weak and sensor failure may be imminent - replace the sensor as soon as possible.
Note: The second calibration after installation should be
performed within 1-2 days of the first calibration.
If the sensor is weak and a replacement sensor is not immediately available, repeat the calibration using the key in Step 3 on the
previous page. If this reading is still less than 50 µA, the reading cannot be stored and the sensor must be replaced immediately.
O2X1 Oxygen Transmitter 19
September 2004

Specifications

ppm Sensor Measurement Ranges (Field-Programmable):
0–10 ppmv O2 (316 SS package only) 0–100 ppmv O2 0–1,000 ppmv O2 0–10,000 ppmv O2
% Sensor Measurement Ranges (Field-Programmable):
0–1% O2 0–10% O2 0–25% O2
Accuracy:
±1% of span at calibration point for all ranges, except ±2% of span at the calibration point for the 0–10 ppmv range (316 SS package only)
Repeatability:
±1% of span or ±2% of span for 0–10 ppmv range (316 SS package only)
Resolution:
±0.1% of span
Linearity:
±2% of span
Operating Temperature:
o
C (32–113°F)
0–45
Ambient Temperature Effect:
±3% of reading over operating temperature range
20 O2X1 Oxygen Transmitter
September 2004
Specifications (cont.)
Sample Pressure:
atmospheric (during operation and calibration)
Atmospheric Pressure Effect:
±0.13% of reading per mm Hg (directly proportional to absolute pressure). During calibration, pressure and flow rate must be kept constant.
Process Connection:
316 SS & Delrin process units: 1/8” NPT inlet and outlet ambient air monitoring units: none
Sample Flow Rate:
1.0 SCFH (500 cc/min) recommended
Response Time:
90% step change with standard sensors:
10 seconds from 1–12 ppm O2 10 seconds from 12–1 ppm O2 15 seconds from 5–10,000 ppm O2 10 seconds from 10,000–5 ppm O2 22 seconds from 0.1–21% O2 16 seconds from 21–0.1% O2
Process Wetted Materials
316 SS Process Unit: 316 SS, Viton, Teflon Delrin Process Unit: Delrin, Viton, Teflon Ambient Air Monitoring Unit: Delrin, Viton, Teflon
O2X1 Oxygen Transmitter 21
September 2004

Model Number

The model number for the O2X1 oxygen transmitter is constructed as follows:
O2X1 - B C
Where, B (Sensor) = 0 - no sensor
1 - standard ppm sensor 2 - acid ppm sensor 3 - standard percent sensor 4 - acid percent sensor
Where, C (Package) = 0 - standard stainless steel
1 - Delrin plastic 2 - ambient air monitor S - special
For example, an O2X1 with a standard percent sensor and a Delrin plastic package would have the following model number:
O2X1-31
22 O2X1 Oxygen Transmitter
September 2004

Electrical Classification/Certification

Weatherproof (Process Units Only):
NEMA 4X, IP66
Intrinsically Safe (316 SS Units Only):
Class I, II, III, Div. 1, Groups A, B, C, D, E, F, G; FM J.I.2D6A6.AX (3610); CSA LR 44204
o
II 1 G EEx ia IIC T4 (-20° to +60 BAS01ATEX1094 X
CE European Union Compliance:
Note: See the Declaration of Conformity and ATEX Compliance
certificates at the back of this manual for additional information.
All O2X1 package options comply with EMC Directive 89/336/EEC and Low-Voltage Directive 73/23/EEC (Installation Category II, Pollution Degree 2).
C)
For EN 61000-4-3, the O2X1 meets performance criteria A and, in a limited number of frequencies, performance criteria B per EN 61326.
IMPORTANT: For CE compliance, the O2X1 cable must be
shielded and the shield must be terminated at the supply end.
O2X1 Oxygen Transmitter 23
September 2004

Intrinsically Safe Installation

Intrinsically safe installations require the use of an MTL706 Zener barrier.
Power Requirements with an MTL706:
24–28 VDC, at 50 mA
Cable:
OCI(*)T5 in hazardous area: blue, 2-conductor twisted-pair with connector, 22 AWG, 1,100 ft maximum length (* = length in ft)
OCB(*) in nonhazardous (safe) area:
3-conductor, 22 AWG, 0.04 ohms/ft
Note: Cable requirements for certified installations must also
comply with Note 1 in Figure 5 on page 7.
Output:
Total load must equal 250 ±5%
24 O2X1 Oxygen Transmitter
September 2004

Non-Hazardous Installation

Non-Hazardous installations do not require the use of an MTL706 Zener barrier.
Power Requirement:
9–28 VDC, loop-powered, 0.6 W max.
Cable:
OC(*)T5: gray, 2-conductor twisted pair with connector, 22 AWG, 0.04 ohms/ft (* = length in ft)
Output:
Max. Load (ohms) = [40 x (PSV - 8)] - RC PSV = Power Supply Voltage in VDC RC = Cable Resistance (22 AWG cable has 0.04 ohms/ft)
Example:
Given a 24 VDC power supply and a 1,000 ft, 22 AWG cable
RC = 1,000 ft. x 0.04 ohms/ft = 40 ohms Max. Load = [40 x (24 - 8)] - 40 = 600 ohms
O2X1 Oxygen Transmitter 25
September 2004
Index
A
Air Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
C
Cable, Custom. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Calibration
Ambient Air. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
Five-Step Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
Span Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-16
CE Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
Certification
Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
E
Electrical Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-23
I
Installation Specifications
Intrinsically Safe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-24
Non-Hazardous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
K
Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
L
LED’s . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
M
Menu Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-12
Model Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-22
O
Outline and Installation Drawing. . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Output, Trimming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-15
Oxygen Sensor, Installing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
R
Range, Selecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
1
September 2004
Index (cont.)
S
Sensor Manifold, Installing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Span Gas Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-16
Specifications
Hazardous Locations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-24
Non-Hazardous Locations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-25
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-20
U
Unit Types. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
User Program
Entering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Exiting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Navigating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
W
Wiring Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Wiring the Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
2
GE Infrastructure
Sensing
DECLARATION
OF
CONFORMITY
We, Panametrics Limited
declare under our sole responsibility that the
to which this declaration relates, are in conformity with the following standards:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4; BAS01ATEX1094 X
Baseefa (2001) Ltd/EECS, Buxton, SK17 9JN, UK
• EN 50104:1998
Deviation from Standard: unit not tested from -10°C to 0°C
• EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation
(For EN 61000-4-3, the O2X1 meets performance Criteria A, and in a limited number of frequencies, performance Criteria B per EN 61326.)
• EN 60529:1991+A1:2000
IP66
following the provisions of the 89/336/EEC EMC Directive and the 94/9/EC ATEX Directive.
The units listed above and any ancillary sample handling systems supplied with them do not bear CE marking for the Pressure Equipment Directive, as they are supplied in accordance with Article 3, Section 3 (sound engineering practices and codes of good workmanship) of the Pressure Equipment Directive 97/23/EC for DN<25.
Shannon Industrial Estate
Shannon, County Clare
Ireland
O2X1 Oxygen Transmitter
Shannon - July 1, 2003
Mr. James Gibson
GENERAL MANAGER
TÜV
TÜV ESSEN
ISO 9001
CERT-DOC-H1 August 2004)
U.S.
GE Infrastructure
Sensing
DECLARATION
DE
CONFORMITE
Nous, Panametrics Limited
déclarons sous notre propre responsabilité que les
rélatif á cette déclaration, sont en conformité avec les documents suivants:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4; BAS01ATEX1094 X
Baseefa (2001) Ltd/EECS, Buxton, SK17 9JN, UK
• EN 50104:1998
Deviation from Standard: unit not tested from -10°C to 0°C
• EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation
(For EN 61000-4-3, the O2X1 meets performance Criteria A, and in a limited number of frequencies, performance Criteria B per EN 61326.)
• EN 60529:1991+A1:2000
IP66
suivant les régles de la Directive de Compatibilité Electromagnétique 89/336/EEC et d’ATEX 94/9/EC.
Les matériels listés ci-dessus ainsi que les systèmes d'échantillonnages pouvant être livrés avec, ne portent pas le marquage CE de la directive des équipements sous pression, car ils sont fournis en accord avec la directive 97/23/EC des équipements sous pression pour les DN<25, Article 3, section 3 qui concerne les pratiques et les codes de bonne fabrication pour l'ingénierie du son.
Shannon Industrial Estate
Shannon, County Clare
Ireland
O2X1 Oxygen Transmitter
Shannon - July 1, 2003
Mr. James Gibson
DIRECTEUR GÉNÉRAL
TÜV
TÜV ESSEN
ISO 9001
CERT-DOC-H1 August 2004)
U.S.
GE Infrastructure
Sensing
KONFORMITÄTS-
ERKLÄRUNG
Wir, Panametrics Limited
erklären, in alleiniger Verantwortung, daß die Produkte
folgende Normen erfüllen:
• EN 50014:1997+A1+A2:1999
• EN 50020:1994
• EN50284:1999
• II 1 G EEx ia IIC T4; BAS01ATEX1094 X
Baseefa (2001) Ltd/EECS, Buxton, SK17 9JN, UK
• EN 50104:1998
Deviation from Standard: unit not tested from -10°C to 0°C
• EN 61326:1998, Class A, Annex A, Continuous Unmonitored Operation
(For EN 61000-4-3, the O2X1 meets performance Criteria A, and in a limited number of frequencies, performance Criteria B per EN 61326.)
• EN 60529:1991+A1:2000
IP66
gemäß den Europäischen Richtlinien, Niederspannungsrichtlinie EMV-Richtlinie Nr.: 89/336/EG und ATEX Richtlinie Nr. 94/9/EG.
Die oben aufgeführten Geräte und zugehörige, mitgelieferte Handhabungssysteme tragen keine CE-Kennzeichnung gemäß der Druckgeräte-Richtlinie, da sie in Übereinstimmung mit Artikel 3, Absatz 3 (gute Ingenieurpraxis) der Druckgeräte-Richtlinie 97/23/EG für DN<25 geliefert werden.
Shannon Industrial Estate
Shannon, County Clare
Ireland
O2X1 Oxygen Transmitter
Shannon - July 1, 2003
Mr. James Gibson
GENERALDIREKTOR
TÜV
TÜV ESSEN
ISO 9001
CERT-DOC-H1 August 2004)
U.S.
GE Infrastructure
Sensing
ATEX COMPLIANCE
We, GE Infrastructure Sensing, Inc.
as the manufacturer, declare under our sole responsibility that the product
to which this document relates, in accordance with the provisions of ATEX Directive 94/9/EC Annex II, meets the following specifications:
Furthermore, these additional requirements and specifications apply to the product:
• Having been designed in accordance with EN 50014, EN 50020, and EN 50284, the
product meets the fault tolerance requirements of electrical apparatus for category “ia”.
• The product is an electrical apparatus and must be installed in the hazardous area in
accordance with the requirements of the EC Type Examination Certificate. The installation must be carried out in accordance with all appropriate international, national and local standard codes and practices and site regulations for flameproof apparatus and in accordance with the instructions contained in the manual. Access to the circuitry must not be made during operation.
• Only trained, competent personnel may install, operate and maintain the equipment.
• The product has been designed so that the protection afforded will not be reduced due
to the effects of corrosion of materials, electrical conductivity, impact strength, aging resistance or the effects of temperature variations.
• The product cannot be repaired by the user; it must be replaced by an equivalent
certified product. Repairs should only be carried out by the manufacturer or by an approved repairer.
• The product must not be subjected to mechanical or thermal stresses in excess of those
permitted in the certification documentation and the instruction manual.
• The product contains no exposed parts which produce surface temperature infrared,
electromagnetic ionizing, or non-electrical dangers.
• As a special condition for safe use, the product must not be used to measure oxygen
concentrations greater than 21%.
1100 Technology Park Drive
Billerica, MA 01821-4111
1180
U.S.A.
O2X1 Oxygen Transmitter
II 1 G EEx ia IIC T4 (-20°C to +60°C) BAS01ATEX1094X
CERT-ATEX-D (Rev. August 2004)
USA
1100 Technology Park Drive Billerica, MA 01821-4111 Web: www.gesensing.com
Ireland
Shannon Industrial Estate Shannon, County Clare Ireland
Loading...