GE Sensing Moisture Probe User Manual

GE Oil & Gas
An Automated High Precision Saturation/Dilution Calibration System for Trace Moisture Sensors
The paper discusses the physical construction and operating principles of this saturation/dilution type calibration system for water vapor in a nitrogen carrier gas used for the purpose of trace moisture sensor calibration. The system generates known moisture standards, allows for settling time, takes readings from sensors, records sensor readings and creates data tables. The data tables are either loaded into “smart” probes with on-board non-volatile memory and/or printed as data sheets provided to customers so that calibration data may be inputted into analyzer electronics. The system is capable of calibrating 128 aluminum oxide moisture sensor probes per week, over a range of –80°C to +10°C dew/frost point temperature. The system is traceable to the National Institute of Standard and Technology (NIST). Each aluminum oxide moisture probe has a dedicated channel for reading the signal from the probe speeding data acquisition time. A chilled mirror hygrometer is used as a traceable transfer standard.
The heritage design, in operation since the early 1980’s, is upgraded with a modern data acquisition, computer controls, data historian, human-machine­interface, and report generation based around National Instruments LabView measurement and automation software. The newly automated system provides faster system throughput and more diagnostic information about calibration performance than the heritage systems. In addition, the system can provide local and remote operator alarms should errors in calibration system operation be detected during the calibration run.
Various aluminum oxide moisture probes that can
be calibrated in the described system
Introduction
One of the major challenges, which every manufacturer of measurement equipment must face, is the calibration of the equipment against a secondary standard on a production line basis while maintaining strict quality control and quality assurance. As the number of instruments or devices, which must be calibrated increases, so
does the challenge. Every eort must be made to
assure consistent accuracy and repeatability of calibration as throughput increases.
GE manufactures process-oriented hygrometers, which use aluminum oxide moisture sensors. The aluminum oxide moisture sensor’s basic construction is shown in the graphic. The aluminum oxide moisture sensor is a transducer, which responds to the vapor pressure of water in the medium to which it is exposed, either gas phase
or liquid phase. Water molecules diuse through
the porous gold layer of the sensor and into the aluminum oxide dielectric layer of the sensor.
Because the water molecule is a polar molecule, it adsorbs to the pore walls of the AlO2 layer, thereby changing the dielectric of the sensor.
The popularity of the aluminum oxide-based hygrometer among process industries has increased by several orders of magnitude since its introduction in the 1960’s. Thus the development of a sophisticated calibration system capable of calibrating a large number of moisture probes in a single run is required to meet customer demand for new moisture probes and provide capacity for customer return probes for periodic recalibration. This transducer must be calibrated against a secondary standard. The latest system design upgrades the automation of the system to provide faster thru-put, more data, better control, and better data manipulation while keeping the basis for the moisture generation calibration gas on the proven double-dilution system. The system is also scalable as a stand-alone system that can be replicated easily and provided to global production and service locations, thus providing customers with regionally located turn-around for their service needs.
impedance. The volumetric mixing ratio of the gas/water vapor mixture is then altered and the process is repeated. The process of calibration consists of two main sections.
1. Production of the carrier gas/water vapor mixture
2. Data acquisition, storage and reduction
Shown below is a block diagram of the calibration system. Gaseous nitrogen is withdrawn from a large tank of liquid nitrogen and transported via a single section of electro-polished, orbital-welded 316SS tubing into the calibration laboratory. Although the theoretical frost point of the nitrogen at this point is equivalent to the boiling point of nitrogen (-198°C), due to trace moisture ingress from various sources
(transportation from the gas supplier, tank relling,
etc.), its frost point temperature is actually -85°C or less at 85 to 95 psig (5.9 to 6.6 bar) pressure. The pressure of the nitrogen is reduced to ambient pressure, which further reduces the frost point to approximately -95°C. The dry nitrogen then travels through a chamber, which contains several “standard” aluminum oxide moisture sensor probes. These standard probes have been previously calibrated in a laboratory, which is independent of the facility being described here. They are used to assure that the nitrogen entering the calibration system does in fact contain a moisture
concentration, which is insignicant compared to the nal mixture that is being generated. The dry
gas then enters the moisture generation system.
Aluminum Oxide Moisture
Sensor Construction
The Calibration System
The calibration system, against which GE aluminum oxide moisture sensor probes are calibrated, operates on the principal of saturation of a gas with water vapor, and then dilution of that saturated gas with
dry gas to achieve a nal mixture containing a
known concentration of water vapor. Calibration of the moisture sensor is accomplished by exposing the probe to a generated carrier gas/water vapor mixture and recording the probe’s electrical
2
Aluminum Oxide Moisture Probe
Calibration System
Dry gas stream is here split into two parts. One part remains dry while the other part is saturated with water vapor by passing it through a large packed tower containing distilled water. The tower is maintained at a constant temperature of 15.2°C and a constant pressure of 20.7 psia (244Kpa). Temperature of saturation is monitored by a laboratory thermometer, which has a NIST traceable calibration and is accurate to within ±0.1°C. Pressure of saturation is monitored by a pressure transducer regularly calibrated against a pressure gauge with an NIST traceable accuracy of ±0.05 psi. This gauge is also used to calibrate transducers, which monitor the pressure in the calibration tanks. The saturated gas mixture will produce a gas mixture with a dew point of 10°C ±0.1°C when the pressure is reduced to one atmosphere.
The 10°C dew point gas exits the saturator section of the calibration system and enters the dilution section. To produce dew/frost point temperatures above
-50°C, only one dilution stage is necessary. Frost point generation below -50°C requires two-stage dilution.
Each of the ve ow control valves used in the dilution
stage of the calibration system is of the thermal mass type and the calibration of each is traceable to NIST.
Their rated accuracy is ± 2% of reading. These ow control valves consist of two units, a ow sensor and
an automatic controller with control valve. Output
of the ow sensor goes to a comparator, which
automatically adjusts the control valve to deliver the quantity of dry or wet gas required to produce the
desired nal mixture.
The diagram shows the ow mixing system principal
for the various combinations of “wet” and “dry” gas
owmeters used to generate carrier gas/water
vapor mixtures equivalent to various dew/frost point
temperatures. The table below shows the owrates for
the respective valves to achieve dew/frost set points.
Flow Block Diagram
3
Flowrates & Accuracies for Generation of Dew/Frost Point Temperatures
Dew/Frost Point Flowmeter(s) in Use Flowrate (cc/min)
None
(System Dry-Down)
10°C FCV-2 4854 +/- 0.1
0°C
-10°C
-20°C
-30°C
-40°C
-50°C
-60°C
-70°C
-80°C
FCV-3 FCV-5
FCV-2 FCV-3
FCV-2 FCV-5
FCV-2 FCV-5
FCV-1 FCV-5
FCV-1 FCV-5
FCV-1 FCV-5
FCV-1 FCV-3 FCV-4 FCV-5
FCV-1 FCV-3 FCV-4 FCV-5
FCV-1 FCV-3 FCV-4 FCV-5
4900
15000
4854 4900
4029
15000
1380
15000
480
15000
159
15000
48
15000
159
4900
433
15000
48
4900
337
15000
48
4900
69
15000
Calculated Dew/Frost
Point Accuracy (°C)
-
+/- 0.2
+/- 0.3
+/- 0.3
+/- 0.3
+/- 0.3
+/- 0.4
+/- 0.3
+/- 0.4
+/- 0.4
The accuracy of the calibration system varies with the generated dew point, as shown in the table. These accuracy estimates were obtained from an error analysis of the moisture mixing system, given in the Appendix of this paper. The overall accuracy of the calibration system is better than ±0.4°C dew/frost point.
The outlet of the dilution stage of the calibration system is routed to the calibration tank capable of holding 128 probes. Each system can support up to 3 calibration tanks providing an overall capacity of 384 probes. Equilibrium conditions in each tank during calibration are assured by random sampling of the impedance of probes from each of the three tanks. These probes are continuously monitored at each generated dew/frost point temperature until the change in impedance becomes within the error of the A-D converter.
4
Chilled Mirror Hygrometer
Loading...
+ 7 hidden pages