Series 35
Specifications
Electronics
European Compliance
Complies with EMC Directive 89/336/EEC and
73/23/EEC LVD (Installation Category II, Pollution
Degree 2)
Functions
• Dew point
• Ppmv in gases at constant pressure (pressure by
programmable constant)
Inputs
• Moisture: Rack, bench, panel and weatherproof models
have a single input for M Series probe connected to a
terminal strip. Probe may be remotely located up to
600 m (2000 ft) from electronic console
• Auxiliary: 4 to 20 mA input
Recorder Output
0 to 20 mA, 4 to 20 mA or 0 to 2 V isolated analog, linear
in parameter chosen
Current Output Load
• 500 S maximum for AC units
• 250 S maximum for portable units
Voltage Output Load
10 kS minimum
Computer-Enhanced Response Software
Optional: Software provides fast measurements in abrupt
dry-down applications (3- to 5-minute speed of
response)
Alarm Relays
• Fault alarm
• Optional Standard Form C, 2 A at 120/250 VAC
• Optional hermetically sealed relay 0.3 A at 120 VAC
Display Units
• Dew/frost point temperature: °F (°C)
• Concentration: ppmv
Power Requirements
• AC power supply: Specify as 100, 120, 220 or 240 VAC,
50/60 Hz
• DC power supply: 24 VDC
Temperature
• Operating: 32°F to 140°F (0°C to 60°C)
• Storage: –4°F to 158°F (–20°C to 70°C)
Configuration Dimensions (h x w x d)
• Rack mount: 3.47 in x 19.00 in x 12.00 in
(8.81 cm x 48.26 cm x 30.48 cm)
• Bench mount: 3.62 in x 10.98 in x 10.78 in
(9.19 cm x 27.88 cm x 27.38 cm)
• Panel mount: 4.25 in x 11.00 in x 12.00 in
(10.80 cm x 27.94 cm x 30.48 cm)
• Weatherproof: 13.56 in x 11.44 in x 5.22 in
(34.44 cm x 29.06 cm x 13.26 cm)
Moisture Measurement
Sensor Type
GE M Series thin-film aluminum oxide
Traceability
Each sensor is individually computer-calibrated against
known moisture concentrations. Calibrations are
traceable to the NIST.
Operating and Storage Temperature
–166°F to 158°F (–110°C to 70°C)
Operating Pressure
5 µ of Hg to 5000 psig (345 bar)
Dew/Frost Point Temperature
Overall calibration range capability: 140°F to -166°F
(60°C to –110°C)
Calibration Ranges
• Standard: 68°F to -112°F (20°C to –80°C) with data to
-166°F (–110°C)
• Extended high: 140°F to -112°F (60°C to –80°C) with
data to -166°F (–110°C)
Accuracy
• ±1.1°F (±2°C) from 140°F to -85°F (60°C to –65°C)
• ±1.6°F (±3°C) from -86°F to -166°F (–66°C to –110°C)