GE Sensing MIS 1 Operating Manual

Moisture Image Series 1 Hygrometer
Programming Manual
April 2004
Process Control Instruments
Moisture Image Series 1 Hygrometer
Programming Manual 910-108PB2
This manual contains inst ructions for Serie s 1 unit s
equipped with controller card p/n 703-1250. This
controller card supports the PanaCom/PanaView
user interface software.
June 2003
Warranty Each instrument manufactured by GE Panametrics is warranted to be
free from defects in material and workmanship. Liability under this warranty is limited to restoring the instr ument to normal operation or replacing the instrument, at the sole discretion of GE Panametrics. Fuses and batteries are specif ically exclu ded from any liability. This warranty is effective from the date of de livery to the original purch as er. If GE Panametrics determines that the equipment was defective, the warranty period is:
one year for general electronic failures of the instrument
one year f or mechanical failures o f the transducers
If GE Panametrics determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE Panametrics, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course of dealing or usage or trade).
Return Policy If a GE Panametrics instrument malfunctions within the warranty period,
the following procedure must be completed:
1. Notify GE Panametrics, giving full details of the problem, and
provide the model numbe r and s eri al numbe r of t he instr umen t. I f the nature of the problem indicates the need for factory service, GE Panametrics will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping in str uc tions for the return of the i nst rume nt to a service center will be p ro vided.
2. If GE Panametrics instructs you to send your instrument to a service
center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Panametrics will evaluate the instrument to
determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
If GE Panametrics determines that the damag e is not covered under
the terms of the warranty, or if the warra nty has expired , an es timate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iii
April 2004
Table of Contents
Chapter 1: Features and Capabilities
Introduction to the Moisture Image
Series 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Electronics Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
BASEEFA Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
CSA Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1- 3
Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
M Series and TF Series Moisture Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Moisture Image Series Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Delta F Oxygen Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Cabling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
M Series and TF Series Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Moisture Image Series Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Delta F Oxygen Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Pressure Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
RS232 Communications Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
User Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
Chapter 2: Basic Programming
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 -1
Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Powering Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Enter i n g Da t a in t o the U s er P ro g ram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Screen and Key Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Entering the Passcode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Getting On-Line Help . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Verifying and Entering Setup Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Activating and Changing Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Entering Calibration Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Entering High and Low Reference Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21
Displaying Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-28
Setting Up the Matrix Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-32
Setting Up the Line Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-35
Tagging the Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-37
Adjusting the Contrast of the Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-40
Setting the Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-41
v
April 2004
Table of Contents (cont.)
Chapter 3: A d vanced Prog r amming
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Setting Up the Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Selecting Measurement Mode and Units for Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Setting the Range for Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Setting Up the Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Selec t ing th e Measur ement Mo d e , Un i t s an d H ig h / L o w Va lu es fo r A la rm s . . . . . . . . . . . . 3-5
Entering the Setpoint and Deadband Values for Alarms. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Using Automatic Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Entering Constants and User Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Entering User Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Specifying User Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Entering User-Defined Tables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Entering Saturation Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Using Computer Enhanced Response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Setting Up the Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 7
Viewing the Data Logger Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Viewing Logged Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Removing Logged Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Entering Data Logger Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Setting Up the RS232 Communications Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Configuring the Serial Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Description of a Sample Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Setting Error Processing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Procedure for Setting Error Processing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Loading New Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3- 3 0
Chapter 4: Programming with PanaView
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Changing Measurement Modes and Units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Open the Meter Display Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Change Meter Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Changing a Matrix Display to a Graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Changing Alarm Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Changing Recorder Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4- 9
Setting Recorder Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Testing Recorder Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Programming Error Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Programming User Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Entering User Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 -1 4
Entering Saturation Constants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
vi

Chapter 1

Features and Capabilities

Introduction to the Moisture Image Series 1 . . . . . . . . . . . . . . . . .1-1
Electronics Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Probes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
User Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
June 2003

Introduction to the Moisture Image Series 1

The GE Panametrics Moisture Image Seri es 1 is a microproc essor based multi-cha nnel ana lyze r designe d to mea sure dissolve d moistur e concentration in gas es and non-aqueous liquids, and dissolved oxygen concentration in gases. It is a highly advanced industrial analyzer whi ch is ve rs ati le and ad ap tab l e to ma ny ap pl ic ation s.
The Series 1 combines the use of hardware and software to make various measur ements. The user conne cts the need ed input s (moi sture probes, oxygen cells, pressure transducers, etc.) to the back panel of the electronics unit using several types of cables. Typically , the user installs moisture probes and oxygen cells into the process using a sample system that is specifically designed for the application. The sample system delivers a sa mple of the process to the probes. The probes then se nd a signal to the Series 1 electronics unit, which interprets the signa l and converts it into a measurement.
The complete analyzer, including electronics, cables, and measurement probes, forms a “syste m” which is designed to monitor the process accurately and reliably. The following sections discuss each part of this “system.”

Electronics Unit The center of the “sys tem” is the el ectroni cs unit. The elect ronics unit

has many components, including terminal blocks for all connections, an RS232 port, a Liquid Crystal Display (LCD), a keypad, and menu keys. See Figure 1-1.
The back panel of the e lectro nics unit cont ai ns the ter minal bl ocks for connecting moisture probes, oxygen cells, recorders, and optional alarms (standard or hermetically sealed) for up to six channels. The Series 1 also provides connections for two isolated outputs per channel. Input connections for the M Series, TF Series, Moisture Image Series and Delta F oxygen sensors are intrinsically safe.
As an optional feature, the Series 1 provides connections for two auxiliary inputs ( not intr insi cally safe ) on each channe l. The auxil iary inputs can accept signals f rom any 0/4 to 20-mA or 0 to 2-VDC device, including other GE Panametrics analyzers.
The RS232 port enables the user to connect a remote terminal, printer, or computer to the S eries 1. Using an optional Personal Compute r (PC ) int erfa ce s oft ware p ack a g e, the user ca n als o set up and operate one or more analyzers from a remote location, transfer data for further analysis, and create data logs.
The front panel of the electronics unit has a Liquid Crystal Display (LCD) to show da ta.The LCD is capable of displaying data in a line graph, or a text matrix format.
Features and Capabilities 1-1
April 2004
123
YES
12
2
3
Electronics Unit (cont.) The front panel a lso has a 16-key keypa d th at is used t o en ter data a nd
operate the S eries 1. There are four menu keys below the LCD, as shown in Figure 1-1 below. The function of a menu key is indicated directly above it on the LCD. The menu key functions chan ge as the user makes menu selections.
The electron ics unit wi th al l thes e featu re s is avai l abl e in a rack mount, bench mount, panel mount, weatherproof, or explosion-proof versions.
LCD Screen
POWER
Menu Keys
Terminal Blocks
Keypad
4 7
5 8 0-
6NO 9
.
50
RS232 Port
Figure 1-1: The Serie s 1 Electronics Unit Features
1-2 Features and Capabilities
June 2003

BASEEFA Certification Rack, bench, panel Series 1s s/n 2000 and above and WPF Series 1s

s/n 2300 and a bove are BASEEFA certified to supply intrinsically safe levels. Installation requires that the Series 1 be mounted in a general purpose area only. M Series, TF Series, and Moisture Image Series probes and Delta F oxygen cells a re intrinsically safe when connected to the Series 1, allowing f or sensor installation in IE C/ CENELEC zone 0 areas.

CSA Certification Newer versions of rack, bench, panel and WPF Series 1’s are

CSA-NR TL certified intrinsically safe. Install ation requires the Series 1 be mounted in either a NEC CL.1, Div. 2, Group B, C or D hazardous area or a general purpose area. M Series, TF Series, and Moisture Image Series probes and Delta F oxygen cells are intrinsically sa fe when connected to the Series 1, allowing for sensor installati on in NEC CL.1, Div. 1, Groups A, B, C and D hazardous areas.
Note: Refer to certification labels on your instrument and
installation drawing #752-138 to determine the applicable level of certificat i on that yo ur meter ca r rie s.
Features and Capabilities 1-3
June 2003

Probes Probe s are the part of the system that come into contact with the

process flow. The Series 1 uses various types of probes to fit the needs of a wide range of applications.
The term “probe” includes devices such as moisture se nsors, pressure transducers, and oxygen cells. Although you can supply your own probes for certain measurements, GE Panametrics provides the following standar d probes to make moisture, temperature, pressur e and oxygen measurements:
Moisture (M) Series - moisture and temperature (in gases and
liquids)
Three Function (TF) Series and Moisture Image Series Probes -
moisture, temperature, and pressure (in gases and liqui ds)
Delta-F Cells - oxygen measurement (in gases)
The most commonly used probes are discussed in more detail below.
M Series and TF Series Moisture Probes
The M Series and the TF Series probes are very similar. Both probes use GE Panametrics aluminum oxide sensor s to measure moisture, and thermistor beads to measure temperature (the thermistor is optional). The TF Series, howe ver, also has an optional built-in strain-gag e-ty p e pres su re tr ans d ucer. The sensor ass em b lies are secured to a probe mount and are usually protected with a sintered stainless-ste el shield (TF-9 probes do not have a shield). Othe r types of shields are availabl e (refer to Figure 1-2 below and Figure 1-3 on page 1-5).
When using an M Series probe, a separate pressure transmitter is required if in-li ne pressure measurement is needed. The Series 1 can use any pressure transmitte r with a 0/4 to 20 mA or 0 to 2-V output. Alternativel y, the user can enter fixed values for temperature and pressure into the Series 1 memory if the process conditions are constant.
Figure 1-2: The M Serie s Probe
1-4 Features and Capabilities
M Series and TF Series Moisture Probes (cont.)
June 2003
Figure 1-3: The TF Series Probe
Although the M and the TF S eries probes may be installed directly in the process line, more often these probes are inserted into a sample system. The sample system protects the probes and enables the user to easily remove the probes for calibrating, servicing and cle aning. Sample systems also control the flow of the process stream so that it is within the specifications of the measurement probe or cell.
The M and TF Series probes, when used with a BASEEFA-approved Series 1, are intrinsically safe (see Chapter 3, Specifications, of the Start up Guide for BASEEFA certification numbers) and designed to meet the requirements of IEC/CENELEC zone 0 areas.
When operated with a CSA-NRTL approved Series 1, the M and TF Series probes are intrinsically safe (see Chapter 3, Specifications, of the Series 1 Startup Guide for CSA-NRTL certification numbers) and designed to meet the requirements of NEC C1.1, Di v. 1, Groups A, B, C and D hazardous areas.
The M and TF Series probes measure moisture conten t in a standa rd range from –110°C to 20°C (–166°F to 68°F) dew/frost point temperature and option ally from –110°C to 60°C (–166° F to 140°F). Both probes optionally measure temperature in a range of –30°C to 70°C (–22°F to 158°F). The TF probe optionally measures pressure from 30 to 300, 50 to 500, 100 to 1000, 300 to 3000, or 500 to 5000 PSIG.
Features and Capabilities 1-5
June 2003
Moisture Image Series Probe
The Moisture Image Series Probe measures moisture, temperature (optional), and pressure (optional). The Moistu re Image Ser ies Probe has its own electronics module that consists of a built-in microcontroller with 16-bit resolution (see Figure 1-4 below).
The electronics module stores the probe cali bration and refere nce data in non-volatile memory; the refore, the user never needs to enter the data by hand. The Moisture Image Series Probe also features continuous compensation of its own analog circuitry to assure long­term electronic stability.
The Moisture Image Series Probe, when operated with a BASEEFA­approved Series 1, is intrinsically safe (see Cha pter 3, Specifications, of the Startup Guide for BASEEFA certification numbers) and designed to meet the require ments of IEC/CENELEC zone 0 areas.
The Moisture Image Series Probe, when operated with a CSA-NRTL certified Series 1, is intrinsically safe (see Chapter 3, Specifications, of the Series 1 St artup Guide for CSA-NRTL certification numbers) and designed to meet the requirements of NEC C1.1, Div. 1, Groups A, B, C and D hazardous areas.
The Moisture Image Series Probe measures moisture content in a standard range from –110°C to 20°C (–166°F to 68°F) dew/frost point temperature and optionally from –110°C to 60°C (–166°F to 140°F). Additionally, the Moisture Image Series Probe can be used to measure temperature in a range of –30°C to 70°C (–22°F to 158°F) and pressure from 30 to 300, 50 to 500, 100 to 1000, 300 to 3000, or 500-5000 PSIG.
Electronics Module
Figure 1-4: The Moisture Image Series Probe
1-6 Features and Capabilities
June 2003
Delta F Oxygen Cell The Series 1 stand ard oxygen measurement probe is the Delta F Cell.
Overall oxygen content range is from 0 to 25%. You may choose from seven standard oxygen r anges.The lowest standard ra nge is 0 to 1/10/100 ppm, and the highest st andar d range is 0 to 25% oxygen. An ultra-low range sensor that covers 0-500 ppb
than 5 ppb is also available. The Series 1 also accepts oxygen inputs from other GE Panametrics oxygen sens ors.
If your application has a high conc entration of acid-forming components, GE P anametrics can supply a STAB-EL option for the Delta F Cell. The STAB-EL option is specially designed to effective ly neutralize these components from the sample gas stream.
The Delta F Cell is available in a general purpose model with both a top and bottom drain or only a top drain (see Figure 1-5 below) . The Delta F Cell is usually inst alled in its own sample system.The Delta F Cell is available with VCR® fittings, mounted in a NEMA-4X enclosure for weatherproof/cor rosion resistant applications, or mounted in a NEMA-7 enclosure for hazardous areas.
and is sensitive to le ss
v
The Delta F Cel l, when ope rat ed w i th a BASE E FA-approved Series 1, is intrinsically safe (see Chapter 3, Specifications, of the Start up Guide for BASEEFA certification numbers) and designed to meet the requirements of IEC/CENELEC zone 0 areas.
The Delta F Cell, when used with a CSA-NRTL certified Series 1, is intrinsically sa fe (see Chapter 3, Specifications, of the Series 1 Start up Guide for CSA-NRTL certification numbers) a nd designed to meet the requirements of NEC C1.1, Div. 1, Groups A, B, C and D hazardous areas.
Bottom Drain
Top Drain
Figure 1-5: Delta F Oxygen Cells - Bottom and Top Drain
Features and Capabilities 1-7
April 2004

Cabling Each probe requires an interc onne cting cable. Cables differ

depending on the probe. The various cables are discussed below under the appropr iate pr obe type . All of these cable assembl ies c an be ordered from GE Panametrics.
M Series and TF Series Probes
Moisture Image Series Probe
The M and the TF Series probes are connected to the analyzer with a special GE Panametrics shielded cable. Probes can be located up to 600 meters (2000 feet) from t he Series 1 (consult GE Panametrics for distances up to 1200 meters). To measure pressure with a TF probe, the maximum cable length is approximately 152 meters (500 feet). Small electronic offsets, especially with longer cables, are possibl e and should be corrected as described in Performing a MH C alibrati on Test/Adjustment on page 1-20 of the Service Manual.
The Moisture Image Series Probe can be connected to the Series 1 with an unshie lded, twisted-pair cable. The probe may be located up to 915 meters (3000 feet) from the analyz er using 22 AWG twisted pair cable. (Consult GE Panametrics for longer cable lengths.)
Delta F Oxygen Cells The standard Delta F Cell is connected to the analyzer with a four-
wire shielded cable (22 AW G). With 22-AWG cable, oxygen cells with a range from 0 to 1/10/100 ppm
located up to 90 meters (300 feet) away from the analyz er. All other oxygen cells can be located up to 15 meters (50 feet) away. For greater cable lengths, 20, 18 or 16 AWG wir e is required. Consult a factory engineer when a longer length is required.
or 0 to 0.5/5/50 ppmv can be
v
Pressure Sensors The Series 1 accep ts either p res su re t rans du cer s or pr es sur e
transmitters. Most pressure transducers can be connect ed with a standard four-wire shielded cable. Most pressure tr ansmitters are connected using either a two or four-wire non-shielded cable and can be either loop or self-power ed.
RS232 Communications Port
A personal computer or serial printer can be connected to the Series 1 RS232 communications port using a standard serial cable. A special switch enabl es the u ser to co n fig u re the un it as Da ta Terminal Equipment (DTE) or Data Communications Equipment (DCE). (See Connecting a PC or Printer in Chapte r 1 of the Service Manual for details on DTE/DCE configuration.)
1-8 Features and Capabilities
June 2003

User Program The Series 1 has a user program that is accessed and controlled by the

keypad (to the right of the LCD screen) and the four menu keys (located under the LCD screen).The user program enables the user to enter the necessary pr obe data, set up the LCD screen, and control the other operational features. The meter stores this data in bat tery­backed RAM for up to five years, even if power is off.
The user program consists of the following four main menus:
HELP - provides on-li ne he lp for t he various m enus wit hin the use r
program.
SELECT - lets the user selec t the type of measur ement and units to
display.
SETTINGS - enables the user to enter probe data and set up basic
functions suc h as recor de rs an d alarm s .
OPTIONS - enables the user to set up the Series 1 screen to display
measurements in matrix or graphic format and perform advanced functions.
The Settings Menu pro mpts the user to e nter a passcod e (see Entering the Passcode in Chapter 2) .
All the functions and f eature s of the S eries 1 progr am are dis cussed in Chapters 2 and 3, Basic Programming and Advanced Programming.
Features and Capabilities 1-9

Chapter 2

Basic Programming

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Ver ifying and Entering Setup Data . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Displaying Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-28
Adjusting the Contrast of the Screen. . . . . . . . . . . . . . . . . . . . . .2-40
Setting the Clock and Calendar . . . . . . . . . . . . . . . . . . . . . . . . . .2-41
June 2003

Introduction The Moisture Image Series 1 is a highly advanced and versatile

instrument. Because of this, operation ranges from basic to advanced functions. This section is designed to provide a step-by-step guide to basic operations. For information on other functions, refer to Chapter 3, Advanced Programming.
IMPORTAN T: You must install a switch or circuit breaker on the
input power line. For greatest safety, locate the circ uit breaker or power switch near the electronics console. To comply with the Low Voltage Directive, you must install an external current protection device.
The following sections cover basic programming functions:
Getting Started - describes how to power up, gives a description of
the user program, and expla ins how to obtain on-line help.
Verifying Setup Data - describes how to verify and, if neces sary,
enter setup data f or moistur e, p ressure, oxygen, and an y other input devices.
Displaying Measur e ments - expla ins how to displ ay measur ements
using differ ent screen formats.
Adjusting the Contrast of the Screen - explains how to adjust the
screen for optimal viewing.
Setting the Clock and Calendar - explains how to set the internal
clock and calendar.
Note: Because they are so essential, the instructions covered in the
Getting Star ted, Verifying and Entering Setup Data and Displaying Measurement s sections are also inc lude d in the
Start up Guide. If you have read this material, you may skip those sections of this chapter.
Basic Programming 2-1
June 2003

Getting Started The purpose of this section is to give a brief description of the Series

1 user program and explain how to use its front pane l for viewing and entering data. This section includes the following:
powering up the meter
entering data into the user program
screen and key functions
entering the passcode
getting on-line help
Read the following sections c are fully to properly set up and operate your instrument.
!WARNING! TO ENSURE THE SAFE OPERATION OF THIS UNIT, YOU MUST I N STALL A ND OPERAT E T H E SERIES 1
AS DESCRIBED IN THIS MANUAL. IN ADDITION, BE
SURE TO FOLLOW ALL APPLICABLE SAFETY CODES AND REGULATIONS FOR INSTALLING
ELECTRICAL EQUIPMENT IN YOUR AREA.
Powering Up The Series 1 has a universal power supply that a utomatically adjusts
to line voltages from 90 to 260 VAC. After making electrical connections (including grounding) as described in Chapter 1, Installation, of the Start up Guide, pre ss the powe r switch to powe r up the meter.
IMPORTAN T: .The AC power cord is the main disconnect device.
For compliance with the EU’s Low Voltage Directive (IEC 1010), this unit requires an external power disconnect device such as a switch or circuit breaker. The disconnect device must be marke d as such, clearly visible, directly accessible, and located within 1.8 m (6 ft) of the unit.
2-2 Basic Programming
Powering Up (cont.) Caution!
Do not op erate the De lta F oxy gen cell for ex t ended
peri ods of time at oxyg en concentrations th at are over
range. Trace and low percent range sensors may be
damaged if expose d t o high levels of ox y gen, such as
air, for l ong periods (>1 hour) while the Se ries 1 is on . If
exposure is unavoidable, either disconnect the oxygen
cell from the Ser ies 1 or equi p the sample system with a
valve that allows the cell t o be switched to purge gas .
If the meter passes its self-test, the screen begins displaying measurements in a matrix format similar to the one shown in Figure 2-1 below.
June 2003
1 Dew Point
-18.0 ° C
2 Temperature
+35.9 °C
3 H
O
2
+0.535 ppm
v
4 Oxygen
48.3 ppb
v
5 Pressure
27.10 PSIg
X Unassigned
Main Menu 11:35 01/22/2000
pg: 1
HELP SELECT SETTINGS OPTIONS
Figure 2-1: Screen Matrix Format After Power Up
Basic Programming 2-3
June 2003
123
YES
Entering Data into the User Program
The Series 1 has a user program that enables you to enter data to set up alarms, recorders, a nd probes, as well as its other features.
The user program c onsists of the following four main menus (refer to Figures 3-17 and 3-18, page s 3-31 and 3-32, for menu maps of the user program):
HELP - provides on-line help for the vari ous menus of the
user program.
SELECT - lets you select the type of measurement a nd units to
display.
SETTINGS - enables you to s et up r ecorde rs and al arms, as we ll as
enter probe data for basic operation. (To enter this menu you must enter the passcode. See page 2-6.)
OPTI ONS - enables you to set up the screen to display
measurements in text or graphical form. It is also used to perform advanced functions .
You can use the screen, four menu keys, and the keypad to enter data into the unit (see Figure 2-2 below).
Screen
POWER
Keypad
4 7
5 8 0-
6NO 9
.
Menu Keys
Figure 2-2: The Moisture Image Series 1 Front P anel
2-4 Basic Programming
June 2003
Screen and Key Functions
The screen displays measur ements, up to four menu options, system messages , and a pointer. The pointer acts as a measurement indicator, as well as a guide during data entry.
While the unit displays measure ments in the matrix format, the pointer moves sequenti ally from box to box to indicate measurement updates. However, while in the user program, you can use the arrow keys to move the pointer to the desired location for data entry.
The menu options appear at the bottom of the screen. Below the screen are the corresponding menu keys (refer to Figure 2-3). Each menu key corr esponds to each of the menu options in the user program. Up to four n ew menu options appea r each tim e you en t er a menu.
The line above the menu options is the message line (see Figure 2-3 below). The message line displays the menu title, the time, and the date. The message line also displays a list of selections de pending on the menu option you enter.
The page indicator is located in between the message line and the menu keys. Since the Series 1 is capable of displayi ng up to six pages of data, the page indicator displays the currently selected pa ge.
Message Line
Main Menu 11:35 01/22/2000
HELP SELECT SETTINGS OPTIONS
Menu Options
Page Indicator
pg: 1
Menu Keys
Figure 2-3: Screen Components
Basic Programming 2-5
June 2003
Screen and Key Functions (cont.)
The keypad to the side of the screen consi sts of 16 keys, incl uding t he [.] and [-] symbol s , two arrow k eys , and two respon s e key s: [YE S ] and [NO]. The numeric keys are for numeric entry only; however , the arrow and response keys have more than one function.
The arrow keys perform two functions:
Use either arrow key to move the screen pointer to the desired
location for data entry.
Use the left arrow key as a backspace during number entry.
The response keys perform thre e funct ions:
Use either the [YES] or [NO] key to respond to questions.
Use the [NO] key to erase data.
Press [YES] to confirm a num ber entry or after e rasing data to
retrieve the previ ous number.
Entering the Passcode The Settings Menu is the only menu that requir es a passcode. The
passcode is a four-digit number that enables only author ized users to enter setup data. The Series 1 pr ompts you to ente r the pa sscode whe n you enter the Settings Menu. See the end of this chapter for your default passcode.
2-6 Basic Programming
June 2003
Getting On-Line Help The Series 1 offers on- line help screens that contain descriptions and
instructions for various topics. The first menu option in most of the menus is HELP.
When you press the HELP menu ke y, the screen display gives a list of help topics as shown in Figure 2-4. This list enables you to get help for most subjects, regardless of the current menu. To select a topic, use the arrow keys to move the pointer to the topic you want to view , then press [YES]. Use the NEXT and PREVIOUS menu options to scroll from one page to another. The NEXT menu option brings you to the following page and the PREV menu options brings you to the previous page. Press DONE to exit and return to the current menu.
Select Topic, then press YES.
Alarms Entering Text
AutoCal Error Messages
Calibration Error Processing Clock/Calendar Graphs
Cs Constant
Configuration Legends,Entering
Data Logging Logger,Capacity
Delta-F Tables Logger,Setting Enhanced Response Logger,Viewing
Entering Numbers Main Menu
Help Topic Index Menu
ABOUT NE X T PREV DONE
Henry’s Law
Figure 2-4: HELP Screen
By pressing the ABOUT key, users can access the history of their particular Series 1: its serial number, P CI number, software revision number and date of manufa ctur e. Thi s informati on r emains part of the HELP file and cannot be changed by the user.
Basic Programming 2-7
April 2004

Verifying and Entering Setup Data

Before the Seri es 1 can make measur em en ts, it must ha ve the p ro per setup data e ntered into its memory. GE Panametrics has entere d all or most of the setup data for each probe before delivery; however, you should verify that all the data is corr ect and entered into your unit.
Note: If you want to switch a probe from one channel to another,
you must re-e nter some or all of the setup data as descri bed in this section.
Verifying setup data consists of three steps:
Verify that probes are properly activated in the Probe
Configuration Menu as descri bed on page 2-9.
Verify that calibration data for all necessary probes is properly
entered in th e System Calibration Menu as described on page 2-12.
Verify tha t app lic ab le hig h and low ref e re nc e value s for the Series
1 measurement circuitry are properly entered into the Reference Menu as described on page 2-21.
Caution!
All high and low reference v alues are fac t ory s et and
normally do not need adjustmen t . H ow ever, t he factory
may instru c t you to adjust th e high and low moistur e
reference values. If necessary, you should do so as
described in Performing an MH Calibration/Adjustment
Test on page 1-20 of the Service Manual.
Use the following sections to pr operly verify and enter setup data.
IMPORTAN T: While you are reviewing data, you should take the
time to record it on the Program Information List supplied in Appendix A of the Startup Guide, or on a separate sheet of paper.
2-8 Basic Programming
April 2004
Activating and Changing Probes
Verifying Probe Configuration Data
Although the probes are physica lly connected to the back of the electronics unit, you must “tell” the electronics what type of measurements the probe is capable of makin g. I n addition, if you want to use a constant va lue rather than a live input, you must “tell” the electronics that you are using a constant value.
If you do not activate probes, or if you activate them incorrectly, the unit will display
No Probe o r other error messages.
Once you verify and/or change confi gur ation data, refer to Entering Calibration Data on page 2-12.
IMPORTAN T: You should record this data on the Program
Information List supplied in Appendix A of the Start up Guide, or on a separate shee t of paper, and keep it in a safe place.
1. Enter the Probe Configurat ion Menu (for a menu map, refer to
Figure 3-19 on page 3-32). A screen similar to Figure 2-5 below appears.
Table 2-1: Entering the Probe Configura tion Menu
Press th e Main Men u key: To enter the:
SETTINGS* Settings Menu
SYSTEM System Settings Menu
CONFIG Probe Configuration Menu
* The Settings Menu will prompt for a passcode (see page 2-6).
SYSTEM CONFIGURATION
Ch O2 HYG T P Aux1 Aux2 1 2 3 4 5 6
Probe Configuration Menu
HELP TAG ERRORS DONE
Figure 2-5: Probe Configuration Menu
Basic Programming 2-9
April 2004
Activating and Changing Probes (cont.)
Verify that the correct probe types have been a ctivated for each channel. To do this, make sure a probe is activated for each measurement it is capable of taking. For example, if a Moisture Image Series Probe with the temperature and pressure functions is connected to channel 1, “MIS” must be select ed in the hygrometry, temperature, and pressure columns as shown in Figure 2-6 below.
If the optiona l auxiliary inputs are installed and active, you only need to select whe ther the input is c urrent or voltage. If you do not select current or voltage, the Series 1 defaults to current.
Note: If you are using an auxil iary input to measur e pressure,
activate Au1 (or Au2 depe nding on which input you ar e using ) in the pres sure column for the desired channel; then activate curre nt or voltage in the Aux 1(or Aux 2) column.
In addition, if you connect a pressure transducer to the PRESS/ OTHER terminal bloc k, you must activate the TF probe in the pressure col umn fo r that ch ann el .
If you need to change any data in the Probe Configuration Menu, see the following sectio n, Ente ring or Changing Probe Configuration Data. If the data is correct, refer to Tagging the Inputs on page 2-37.
SYSTEM CONFIGURATION
Ch O2 HYG T P Aux1 Aux2 1 2 3 4 5 6
Probe Configuration Menu
MIS MIS MIS
HELP TAG ERRORS DONE
Figure 2-6: MIS Probe Configuration on Channel 1
2-10 Basic Programming
April 2004
Entering or Changing Probe Configuration Data
Figure 3-19 on page 3-32 for a menu map.
1. From the Probe Configuration Menu (Table 2-1 on page 2-9),
move the pointer to the channel and probe y ou want to select. The arrow will move only to installed channels.
2. Press [YES].
3. The possible probe types appear on the message line at the bottom
of the screen. Table 2-2 on page 2-12 shows the available probe types.
If you choose to enter a constant for moisture (Kh), temperature (Kt), or pressure ( Kp), you must ente r a c onstant value in the User Const ant Menu. See Entering Constants and User Functions on page 3-9 for instructions.
4. Move the brac kets t o t he corre ct pr obe t ype, then pre ss [YES]. The
selected probe type displays.
5. Repeat Steps 1 through 4 for each channel.
6. T o exit, press the DONE menu key until Main Menu appears on
the message line.
Note: You can only sele ct probes for an installed channel.
If you are only using Moisture Image Series probes, pr oceed to Displaying Measurements on page 2-28. If you are using any other type of input device, pr oceed to Entering Calibration Data on page 2-12.
Basic Programming 2-11
April 2004
Entering or Changing Probe Configuration Data (cont.)
Table 2-2: Probe Types
Measurement Mode Probe Type
– (None)
OXY (Oxygen)
HYG (Hygrometry)
T (Temperature)
P (Pressure)
Aux1 (A u x iliar y 1)
Aux2 (A u x iliar y 2)
% (Percent) ppM (Parts per Million) ppB (Parts per Billion)
– (None) MIS (MIS Probe) TF (Three-Function Probe) Mxx (M Series Probe) Kh (Constant Dew Point)
– (None) MIS (MIS probe) TF (Three-Function probe) Mxx (M-Series probe) Kt (Constant Temperature)
– (None) MIS (MIS probe) TF (Three-Function probe) Au1 (Auxiliary 1) Au2 (Auxiliary 2) Kp (Constant Pressure)
– (None) I (Current) V (Voltage)
– (None) I (Current) V (Voltage)
Entering Calibration Data
The Series 1 needs calibration data for moisture, oxyge n, and pressure probes, as well as auxiliary inputs. GE Panametrics has already entered cali bration data for probes; however, you should verify this data befor e star tup. You will need to enter this data if you:
sent the probes back to the factory for calibration
use a different pr obe
use a non-GE Panametrics input
T o enter, verify, or change calibration data for any probe type, enter the System Calibration Menu as described in Table 2-3 below. (see Figure 3-19 on page 3-32 for a menu map.)
.
Table 2-3: M o vi n g to th e System Ca l ibr ation Men u
Press th e M ai n M en u key: To enter the:
SETTINGS* Settings Menu
SYSTEM System Settings Menu
CALIB System Calibration Menu
*The Settings Menu will prompt for a passcod e (see page 2-6).
2-12 Basic Programming
Loading...
+ 93 hidden pages