GE Sensing HygroTrace Operating Manual

GE
Sensing
HygroTrace
Moisture Transmitter
User’s Guide
GE
Sensing
HygroTrace
Moisture Transmitter
916-102A3 September 2007
The HygroTrace Moisture Transmitter is a GE Panametrics product. GE Panametrics has joined other GE high-technology sensing businesses under a new name—GE Sensing.
September 2007
Warranty
Each instrument manufactured by GE Infrastructure Sensing, Inc. is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE determines that the equipment was defective, the warranty period is:
one year from delivery for electronic or mechanical failures
one year from delivery for sensor shelf life
If GE determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu
of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course
of dealing or usage or trade).
iii
September 2007
Return Policy
If a GE Infrastructure Sensing, Inc. instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the
model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE instructs you to send your instrument to a service
center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine
the cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty,
the instrument will be repaired at no cost to the owner and returned.
If GE determines that the damage is not covered under the
terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iv
September 2007
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Sample System Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Insertion into the Sample System/Process. . . . . . . . . . . . . . . 3
Standard Wiring Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Digital Communications Wiring Connections . . . . . . . . . . . . 7
Powering Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Using the HygroTrace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Keypad Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
The Default Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Unlocking the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Accessing the Menus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Entering Numeric Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting Up the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4-20 Analog Output and Error Reporting . . . . . . . . . . . . . . 14
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Communications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
LOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Menu Item Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Moisture Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
v
September 2007

Introduction

The GE Sensing HygroTrace is a compact moisture transmitter designed for measuring ultra-low water content in the ultra high purity and semiconductor markets and related applications. The HygroTrace measures water content in nitrogen or argon gas in the range of 0 to 100 ppb
calibrated range. The unit features an integrated display and a six­button keypad, and is housed in an aluminum enclosure. The HygroTrace operates on 24 VDC and provides 4-20 mA analog and RS485 digital outputs.
The HygroTrace uses an aluminum oxide sensor manufactured with semiconductor techniques. The patented measurement technique enables the HygroTrace to respond very quickly to both dry-to-wet and wet-to-dry situations. The technique applies a temperature pulse to ‘dry’ the sensor. Then, the re-adsorption rate is measured while holding a constant sensor temperature. This measurement is proportional to the moisture concentration in the sample gas. As a result, the HygroTrace offers the sensitivity and response time needed to effectively measure parts per billion levels of moisture.
with trending analysis beyond the
v

Installation

Sample System Guidelines

The HygroTrace transmitter is typically installed in a sample system or on the bypass stream of an ultra-high purity gas distribution system. The factory recommends that the unit be installed in a sample system to protect the sensor from coming in contact with damaging elements in the process, and to allow the transmitter to be removed for service without interrupting the main process gas flow.
Before constructing a sample system, consult a GE Sensing applications engineer and adhere to the guidelines below. See Figure 1 on page 2 for an example of a sample system.
HygroTrace Moisture Transmitter 1
September 2007
Sample System Guidelines (cont.)
A sample system should be kept very simple. It should
contain as few components as possible and all or most of those components should be located downstream of the measurement location.
Sample system components should not be made of material
that will affect measurement. Most common filters and pressure regulators are not suitable for sample systems because they have wetted parts that may adsorb or release moisture, etc. into the sample system. They may also allow ambient contamination to enter the sample system. If possible, use stainless steel material for all wetted parts.
The transmitter should be installed perpendicular to the
sample inlet. For dimensions and other requirements see Specifications on page 26.
Note: A typical sample system for the hygrometer will use VCR
components.
Note: At least 5 feet of 1/4” tubing, if vented to the atmosphere,
should be installed on the outlet to prevent back diffusion of ambient moisture into the sample tee cell of the HygroTrace.
Sample systems should be leak-tested prior to operation to
verify the integrity of the connections, components and fittings. Tighten any loose fittings.
IMPORTANT: Caution must be taken when pressurizing or
depressurizing the sample system to prevent damage to the moisture sensor.

Figure 1: Sample System Example

2 HygroTrace Moisture Transmitter
September 2007

Insertion into the Sample System/Process

!CAUTION!
If the HygroTrace is being installed directly into
the process line, consult the factory for proper
installation instructions and precautions before
beginning the following procedure.
!CAUTION! No maintenance of the sensor can be performed in the field. Accessing the sensor chamber will break the VCR
seal and should be done only at the factory.
To install the transmitter, refer to Figure 2 below and use a wrench to thread the female 1/4” fittings of the sample system or process line onto the 1/4” male VCR fittings of the tee sample cell configuration.
1/4” VCR (Male)
2 places
Tee Sample Cell

Figure 2: HygroTrace Installation

HygroTrace Moisture Transmitter 3
September 2007

Standard Wiring Connections

Note: The HygroTrace is not certified for use in Hazardous
(Classified) locations.
This procedure is for those units not connected to a computer.
The transmitter must be wired using the factory-supplied cable (two meters in length). If an alternate length is required, please contact the factory for assistance.
Note: If cables need to be lengthened, r efer to Table 1 on page 5
to splice an extension onto the existing cable. Connect positive to positive and negative to negative.
Use the following steps to wire the transmitter to the system.
1. Align the red dot with the red mark (see Figure 3 below) and push the connector on the transmitter cable into the mating connector on the transmitter module. Make sure the pins are properly aligned. Once it is inserted, push the cable connector until it locks into place.

Figure 3: Transmitter Cable Connection

4 HygroTrace Moisture Transmitter
September 2007
Standard Wiring Connections (cont.)
2. Using the flying leads at the other end of the transmitt er cable,
connect the transmitter to the power supply and data acquisition system (DAS) as shown in Figure 4 on page 6. Refer to Table 1 below for a description of the leads in the factory-supplied cable.
Table 1: Cable Lead Connections - No PC
Lead Connection Description
Red (+) 24VDC 1 AMP
Black (–) 24VDC 1 AMP
Orange (+) 4-20mA Output
Blue (–) 4-20mA Output
Cable Braid System Ground
3. Trim any unused leads back to the outer cable jacket in order to remove the bare tinned wire and prevent accidental short circuits.
The HygroTrace is now ready for operation.
Note: To remove the probe cable, pull back the spring-loaded
sleeve and pull the cable out of the connector.
HygroTrace Moisture Transmitter 5
September 2007
+24V 1 AMP
+
RS232
TO PC
RS232
RS485 TO
CONVERTER
POWER
SUPPLY
+
4-20mA
METER
COMMON
+ RS485
– RS485
SYSTEM
GROUND
RED
BLACK
ORANGE
BLUE
WHITE
GREEN
BRAID
CABLE

Figure 4: Standard Wiring Connections (ref. dwg #702-684)

6 HygroTrace Moisture Transmitter
September 2007

Digital Communications Wiring Connections

If the unit is to be operated using PanaView™ instrumentation software installed on a computer, an RS232/RS485 converter must be used and the wiring must be set up as follows.
The transmitter must be wired using the factory-supplied cable (two meters in length).
Note: To lengthen cables, refer to Table 2 below to splice an
extension onto the existing cable. Connect positive to positive, negative to negative, and ground to ground.
Note: EMI ferrite shielding may be required for communication
through an ungrounded serial port. Please consult the factory for guidance.
Use the following steps to wire the transmitter to the system.
1. Align the red dot with the red mark (see Figure 3 on page 4) and push the connector on the transmitter cable into the mating connector on the transmitter module. Make sure the pins are properly aligned. Secure the connectors by pushing the cable connector until it locks in place.
2. Using the flying leads at the other end of the transmitt er cable, connect the transmitter to the power supply, the data acquisition system (DAS) and the computer as shown in Figure 4 on page 6. Refer to Table 2 below for a description of the leads in the factory-supplied cable.
Table 2: Cable Lead Connections - with PC
Lead Connection Description
Red (+) 24VDC 1 AMP
Black (–) 24VDC 1 AMP
Orange (+) 4-20mA Output
Blue (–) 4-20mA Output White (+) RS485 Green (–) RS485
Cable Braid System Ground
HygroTrace Moisture Transmitter 7
September 2007
Digital Communications Wiring Connections (cont.)
3. Trim any unused leads back to the outer cable jacket in order
to remove the bare tinned wire and prevent accidental short circuits.
The HygroTrace is now ready for operation.
Note: To remove the probe cable, pull back the spring-loaded
sleeve and pull the cable out of the connector.

Powering Up

After the HygroTrace is wired as described in the previous sections, power may be applied to the unit. The transmitter takes up to 60 seconds to initialize and begin normal operation. If the sensor has been exposed to ambient conditions of less than 10% RH for less than 72 hours, the HygroTrace will meet its specified accuracy within 24 hours following startup.
8 HygroTrace Moisture Transmitter

Using the HygroTrace

Keypad Features

September 2007
Cancel
Left

Figure 5: HygroTrace Keypad

The HygroTrace has only six keys: four arrow keys, a Cancel key and an Enter key.
Use the arrow keys to navigate among menu choices, and to increment/decrement numeric entries.
Use the Cancel key to cancel a numeric entry change, or exit a menu.
Use the Enter key to accept a numeric entry or select a menu option.
Up
Down
Enter
Right
HygroTrace Moisture Transmitter 9
September 2007

The Default Display

Figure 6 on page 11 shows the normal, default display of the HygroTrace/TFM.
Menu Prompt Area - Titles, prompts, status displayed here.
Measurement - 3 ½ digits (0 - 1999). Automatic decimal
placement based on range.
Unit Symbol - ppb = Parts Per Billion water by volume.
Keypad Lock Indicator - Unlock sequence required to modify
settings.
Measurement Progress Bar - HygroTrace takes
~2 minutes to acquire data and complete the measurement. The progress bar gives a visual indication of the measurement status.
Warning/Error Indicator - In the absence of an error, this
region is blank. Errors that effect the quality of the measurement (e.g. heater failure) are indicated in ALL CAPITAL letters. Warnings that indicate problems detected with the measurement are indicated in Mixed Case letters. Refer to Table 3 below.
Table 3: Warning/Error Messages
No. Label Type Description
0 Status OK ---- No Errors detected.
The calculated PPBv value is greater
Sns Ovr Rng
3
12 Acquiring... WARN
101 HTR FAIL LO ERROR
102 HTR FAIL HI ERROR The heater exceeded 200°C.
Sns Too Wet
204
10 HygroTrace Moisture Transmitter
WARN
WARN
than the highest calibration point, and is an extrapolation.
Displayed at startup while the HygroTrace collects data for the first measurement.
The heater failed to reach the desired temperature in the time allotted.
The sensor is exposed to excessive moisture conditions.
September 2007
Menu Prompt Area
Measurement
Units Symbol
Warning/Error Indicator
Measurement Progress Bar
Keypad Lock Indicator

Figure 6: Default Display

Unlocking the Keypad

After power-on, the HygroTrace keypad is locked. It is necessary to enter the keypad unlock sequence to make any changes to the HygroTrace.
Similar to a mobile phone, the HygroTrace will prompt the operator to unlock if any key is pressed. A passcode is required to use certain factory service features only.
To unlock the keypad, press Cancel , Enter , Cancel in sequence.

Accessing the Menus

After successfully unlocking the keypad, the HygroTrace will display the Main Menu (see Figure 7 on page 12). Use the arrow keys to highlight the menu item desired. Refer to the Menu Map, Figure 9 on page 23.
Press Enter to select the highlighted item. Many menu items will display another menu. Use Cancel to return to the previous menu page. Pressing Cancel from the Main Menu
will return the screen to the Measurement Display.
HygroTrace Moisture Transmitter 11
September 2007
Accessing the Menus (cont.)
Note: Menu items displayed with an ellipsis (…) will bring up
more choices, while those without take immediate action.

Figure 7: Main Menu

Entering Numeric Values

Since the HygroTrace has no numeric keypad, numeric values are entered using a “combination lock” style of entry:
Use the left arrow key and right arrow key to select the digit to change. The digit selected will be indicated with a .
Use the up arrow key and down arrow key to increment or decrement the digit.
Note: If incrementing or decrementing a digit would cause the
numeric value to exceed its allowable range (maximum/ minimum value), the digit will not change.
Press Enter to save the new value and return, or Cancel to return, leaving the original value intact.

Figure 8: Numeric Entry

12 HygroTrace Moisture Transmitter
September 2007

Setting Up the Transmitter

After proper installation, the HygroTrace Transmitter can be set up to accommodate the user’s requirements. Typically, the user may need to configure the analog outputs, trim the analog outputs, and program the digital outputs. Refer to the Menu Map, Figure 9 on page 23, and complete the following steps. Upon startup, the HygroTrace proceeds through several displays until a screen similar to the following appears:
The symbol in the lower right corner indicates that the screen is locked. To unlock the screen, press
Cancel, Enter, Cancel.
Note: In most instances; use the Enter key to save an entry and/
or move ahead to the following screen; use the Cancel key to reject an entry and/or return to the previous screen.
9

Display

When the screen is unlocked, the Main Menu appears with several options. To set up the display, select Display... and press Enter. The following screen appears:
To show the primary ppb select Measurement and press
Enter. A screen similar to the following appears:
Press Cancel to return to the Main Menu.
display,
v
Diagnostics - has limited access. The service passcode is
required. For a description, see Diagnostics on page 29.
HygroTrace Moisture Transmitter 13
September 2007

4-20 Analog Output and Error Reporting

The HygroTrace has a single 4-20 mA analog (recorder) output. Normally, the output reports the measurement in ppb
user-specified range. The default range is 0-100 ppb that is, 4 mA represents the zero (0 ppb the span (100 ppb
The recorder output can exceed this range, within limits. A measurement greater than the span will be reported up to
20.5 mA, and a reading below the zero will be reported down to
3.8 mA. Faults are indicated by readings outside this range.
) value.
v
), and 20 mA represents
v
using a
v
;
v
These faults will be indicated by a 3.5 mA signal
:
No Data - transmitter has not completed a measurement cycle,
so it cannot report.
No Calibration - transmitter was not programmed with
calibration data.
Out of Range - measurement cannot be determined with the
given calibration.
These faults will be indicated by a 21.1 mA signal
:
Heater Fail Low - heater failed to reach the operating setpoint.
Heater Fail High - heater exceeded its operating limit.
Limits are taken from the NAMUR Recommendation NE 043. Measurement is valid from 3.8 mA to 20.5 mA. Fault is indicated by 3.6 mA or 21.0 mA.
The HygroTrace has a special operating mode used to dry the sensor (see Sensor Dry Down on page 19). When in Dry Down mode, the unit will report a fixed output of 20.5 mA, the maximum valid measurement. When returned to normal operation, the output returns to reporting the actual measurement.
In the case of a measurement exceeding the maximum calibration value, the output will be limited to the maximum valid signal,
20.5 mA. The HygroTrace will never produce a negative reading (less than 0 ppb).
14 HygroTrace Moisture Transmitter

Outputs

Recorder Zero
September 2007
To set up outputs, from the Main Menu select Output... and press
Enter. The following screen appears:
To specify the ppb reading equivalent to 4 mA, select Rcd Zero and press Enter. A screen similar to the following appears:
Use the arrow keys to change the output value. Press Enter to save (or Cancel to keep the previous value), and return to the Output Menu.
Recorder Span
To specify the ppb reading equivalent to 20 mA, select Rcd Span and press Enter. A screen similar to the following appears:
Use the arrow keys to change the output value. Press Enter to save (or Cancel to keep the previous value), and return to the Output Menu.
HygroTrace Moisture Transmitter 15
September 2007
Recorder Trim
To adjust/calibrate the 4-20 mA output, select Rcd Trim... and press Enter. The following screen appears:
To return an adjusted 4-20mA calibration to the factory defaults, press Enter. The following screen appears.
Note: Resetting the Trim erases any previous values before
calibrating the output.
Select YES or NO and press Enter, or press Cancel. If you select NO or press Cancel, the unit returns to the previous screen and no change is made to the 4-20mA calibration.
If you select YES, the following screen appears to adjust and calibrate the 4-20mA signal.
With Trim Zero highlighted, to adjust the 4 mA output from the factory default, press Enter and a screen similar to the following appears.
Note: The HygroTrace will attempt to output exactly 4.00 mA.
The operator should read the output, in mA, at the DAS or other indicator, and enter that value as follows.
Use the arrow keys to change the output value. Press Enter to save (or Cancel to keep the previous value), and return to the Recorder Trim menu.
16 HygroTrace Moisture Transmitter
September 2007
Recorder Trim (cont.)
If a new Trim Zero was saved, the Trim Span is now highlighted. To
adjust the 20 mA output from the factory default, press Enter and see a screen similar to the following:
Note: The HygroTrace will attempt to output exactly
20.00 mA. The operator should read the output, in mA, at the DAS or other indicator, and enter that value in the following screen.
Use the arrow keys to change the output value. Press Enter to save (or Cancel to keep the previous value), and return to the Recorder Trim menu. Press Cancel twice to return to the Main Menu.
Test
To force the 4-20 mA output to a specified percent of scale, select Test on the Output Menu and press Enter. A screen similar to the following appears.
Use the arrow keys to change the test value. Press Enter to save (or Cancel to keep the previous value), and return to the Output Menu. Press Cancel to return to the Main Menu.
HygroTrace Moisture Transmitter 17
September 2007

User

PPB Offset
To specify an offset value in PPB and add to the measured PPB value, or to dry down the sensor, select User... on the Main Menu, press Enter and see the following screen:
Use the arrow keys to enter the user passcode (2719), press Enter and the following screen appears. The User Mode indicator U appears in the lower right corner of the screen.
T o specify a PPB of fset value, select PPB Offset, press Enter and the following screen appears:
U
Use the arrow keys to enter the PPB offset desired, press Enter and the screen returns to the User Settings Menu.
U
Clear PPB Offset
To clear the PPB offset entered above, select Clear PPB Offset and press Enter. The following screen appears. Press Cancel to return to the Main Menu or proceed to the
U
next option. Select YES or NO and press Enter, or
press Cancel. If you select NO or press Cancel, the unit returns to the previous screen and the PPD offset is not cleared.
U
If you select YES, the PPB Offset is cleared and the unit returns to the previous screen.
18 HygroTrace Moisture Transmitter
September 2007
Sensor Dry Down
Following factory calibration, the HygroTrace is vacuum packaged to minimize sensor uptake of ambient moisture during shipment and storage. Once the unit is installed in the process or sample system, it is recommended that the user establish a nominal sample flow of 1 SLM and initiate a manual dry down of the sensor using the Sensor Dry Down option in the User Settings Menu.
The initiation of this dry down cycle depends on the exposure of the unit to ambient moisture conditions. The criteria for exposure is detailed below . It is recommended that the heater control be set to 190°C for 24 hours following initial installation. The dry down cycle must be manually started and stopped by the user.
General rule of thumb: >72 hours @ <10% RH or >24 hours @ >40% RH
This sensor-drying procedure can be repeated following prolonged exposure to ambient moisture conditions. This will ensure the highest accuracy for moisture measurement.
Note: When in Dry Down mode, the unit will report a fixed
output of 20.5 mA, the maximum valid measurement. When returned to normal operation, the output will return to reporting the actual measurement.
T o initiate a manual dry down of the sensor, select Sensor Dry Down in the User Settings Menu. Press Enter and the following screen appears:
U
Press Enter to start the Sensor Dry Down (or press Cancel to not
begin a sensor dry down cycle) and return to the previous screen. If you pressed Enter, the Sensor Dry Down
U
has begun.
Note: The sensor will be heated to and r emain at 190°C until the
user stops this cycle.
HygroTrace Moisture Transmitter 19
September 2007
Sensor Dry Down (cont.)
To stop the Sensor Dry Down, select it again on the User Settings Menu, press Enter, and this screen appears. Press Enter to stop the Sensor Dry Down (or press Cancel to continue)
U
and return to the previous screen.
Note: If you pressed Enter, the Sensor Dry Down has ended.

Communications

Baud Rate
To change the baud rate, select Comms... on the Main Menu, press Enter and the following screen appears:
Select Baud Rate... and press Enter. A screen similar to the following appears.
Use the arrow keys to select a new baud rate and press Enter. The number at the top right side of the screen changes. Press Cancel to return to the Comms Menu.
20 HygroTrace Moisture Transmitter
September 2007
Node ID
Note: If more than one HygroTrace is on the RS485 network,
each unit must have a unique ID.
To change the node ID, select Node ID on the Comms Menu, then press
Enter and the following screen appears:
Use the arrow keys to change the node ID value. Press Enter to save (or Cancel to keep the previous value), and return to the Comms
Menu.
Test
To test the serial connection over RS485, select Test and press Enter. The HygroTrace will transmit a short message, “HygroTrace Comm Test - Node 16”, over RS485 for testing the serial connection.
To return to the Main Menu, press Cancel.
Service - has limited access. The service passcode is required.
For a description of the service menu, see Service on page 30.
HygroTrace Moisture Transmitter 21
September 2007

About

To access HygroTrace copyright and software version information, and instrument data such as unit and sensor serial numbers, the number of sensor hours and the system uptime:
Select About on the Main Menu screen and press Enter. The following screen appears:
+
To access HygroTrace copyright and serial number information, select ID on the About HygroTrace screen and press Enter. A screen
+
similar to the following appears: To return to the About HygroTrace
menu, press Cancel. To access up time information, select System Status and press Enter. A screen similar to the following appears.
+
To return to the About HygroTrace menu, press Cancel. T o access software version information specific to the unit, select Software
+
Versions and press Enter. A screen similar to the following appears.
Press Cancel twice to return to the Main Menu.
+

LOCK

To lock the unit from further changes, select LOCK on the Main Menu and press Enter. To restore access to the unit menus, press Cancel, Enter, Cancel.
+
22 HygroTrace Moisture Transmitter
September 2007
Coeff A
Coeff B
About LOCK
Dump
Update
Params
Service**
Main
User*
57600
Baud Rate
PPB Offset ID
38400
19200
100°C
9600
60°C
50°C
Erase
OFF
150°CHeater Test..
190°C
Resume
Reboot
Cal Data
1200
2400
Reset Trim
Trim Zero
Node ID
Trim Span
Test
Test
Rcd Span
Rcd Trim
Notes:
*User passcode required.
**Service passcode required . Contact factory for assistance.
Display Output Comms
Rcd Zero
Diagnostic**

Figure 9: Menu Map

HygroTrace Moisture Transmitter 23
September 2007

Menu Item Descriptions

Display
1. Measurement - Show the primary ppb
display
v
2. Diagnostics - (Service passcode required
)
Output
1. Rcd Zero - Specify the ppb reading equivalent to 4 mA
2. Rcd Span - Specify the ppb reading equivalent to 20 mA
3. Rcd Trim - Adjust/calibrate the 4-20 mA output a. Reset Trim - Clear trim values prior to recalibrating output b. Trim Zero - Adjust the 4 mA output c. Trim Span - Adjust the 20 mA output
4. Test - Force the 4-20 output to a specified percent of scale
User - (User passcode required
)
1. PPB Offset - Enter PPB Offset
2. Clear PPB Offset - Clear PPB Offset
3. Sensor Dry Down - Dry the Sensor
Comms
1. Baud Rate - Specify the data rate for RS485 communications
2. Node ID - Specify the PanaView Network Node ID
3. Test - Force meter to transmit an ID message for test purposes
Service - (Service passcode required
)
1. Update - Update the unit software.
2. Params - Access the parameter storage. a. Dump - Dump the parameter storage. b. Erase - Erase the EPROM.
3. Heater Test - Test the heater. a. 50°C to 190°C - Select the temperature for testing. b. OFF - Turn the heater off. c. Resume - Resume normal measurement operation.
24 HygroTrace Moisture Transmitter
September 2007
Menu Item Descriptions (cont.)
4. Cal Data - Change the calibration data. a. Bulk Limit - Set the bulk limit. b. Cal Curve - Set a calibration point and edit PPB and ZH
values.
c. Extrapolation - Set coefficient A and/or B.
5. Reboot - Reboot and return to the initial screen.
About - Display Product, Copyright, and Version information. LOCK - Lock the keypad.
HygroTrace Moisture Transmitter 25
September 2007

Specifications

General

Measurement Range:
0 to 100 ppb
Compatible Gases:
with trend indication beyond the calibrated range
v
Nitrogen and Argon
Process Gas Temperature Range:
14°F to 95°F (–10°C to 35°C)
Storage Temperature:
–40°F to 158°F (–40°C to 70°C)
Operating Temperature:
14°F to 95°F (–10°C to 35°C)
Warm-Up Time:
Meets specified accuracy within 24 hours, after sensor
exposure <72 hours @ 25°C and 60% RH
Calibrated Uncertainty @ 77°F (25°C):
±20% of reading or ±5 ppb
Response Time:
, whichever is greater
v
Less than 20 min. for 95% of 25 ppb
26 HygroTrace Moisture Transmitter
step change
v

Electrical

Power:
20 to 28VDC, 20 Watts
Analog Output:
4 to 20 mA, 400Ω load max
Digital Output:
RS485
Output Resolution:
14 bits
Display:
128 X 64 LED backlit LCD
Display of primary measurements and diagnostics

Mechanical

September 2007
Sample Connection:
In-line flow, ¼" male VCR process connection
Sample Flow Rate:
1 to 4.3 SCFH (0.5 to 2 SLM)
Operating Pressure:
0-10 psig (0 to 0.69 bar)
Proof Pressure:
3000 psig (207 bar)
Enclosure:
Aluminum construction; black color; powder coated finish
Overall Dimensions (H x W x D): 7.3 in x 4.6 in x 2.5 in
(185 mm x 117 mm x 63.5 mm)
Weight: 2.5 lbs (1.13 kg)
HygroTrace Moisture Transmitter 27
September 2007

Moisture Sensor

Sensor Type:
Thin-film aluminum oxide moisture sensor
Calibration:
Each sensor is individually computer-calibrated against known
moisture concentrations.
Calibration Interval:
Sensor recalibration by GE is recommended every 6 to 12
months depending on the application.

Certification

Complies with EMC Directive 89/336/EEC and
PED 97/23/EC for DN < 25
UL 508

Figure 10: Certification Label (for information only)

28 HygroTrace Moisture Transmitter
September 2007

Appendix

The Diagnostics and Service menus are intended for factory­trained personnel only, and access is limited by requiring a service passcode. These menus allow the factory default data, including calibration data and sensor parameters, to be adjusted. Please contact an applications or service engineer at GE Sensing if access to these menus is required. Normal operation of the HygroTrace does not require access to the information contained in these menus.

Diagnostics

A Diagnostics Display page can be selected from the Display Menu (see Figure 11 below), but can be accessed only by using
the service passcode.
= Heating = Cooling
... = Holding at Temp.
+ = Acquiring Data
= = Calculating Result
Hygrometry
Voltage
Number of Data
Samples
Heater
Temperature
Menu: x H:+1.234567 60.2°C
100
N: 85 R :0.855 M:+1.195886e–05
2
Control Status
ppb
Slope
of Fitted Line
Correlation Coefficient
of Fitted Line
Calculated
Result
Service Mode
Indicator

Figure 11: Diagnostics Display

HygroTrace Moisture Transmitter 29
September 2007
!CAUTION! No maintenance of the sensor can be performed in the field. Accessing the sensor chamber will break the VCR
seal and should be done only at the factory.

Service

To access the Service Menu, select Service... on the Main Menu and
press Enter. The following screen appears:
Use the arrow keys to enter the Service Passcode. Press Enter and the following screen appears:
Update
Due to product improvements or software enhancements, it may be necessary to install new software provided by GE Sensing. This is done over the RS485 digital communications link, using a PC application capable of performing an XMODEM file transfer.
Note: For updates, the baud rate setting on the COMMS menu is
ignored. Updates will use the following parameters:
Baud Rate: 38400 Word Size: 8 bits Stop Bits: 1 Parity: None Flow Control: None/disabled
To update the instrument software, select Update on the Service Menu, press Enter and the following screen appears:
+
30 HygroTrace Moisture Transmitter
September 2007
Update (cont.)
Select YES or NO and press Enter, or press Cancel. If you select NO or press Cancel, the unit returns to the previous screen. If you select YES, the following screen appears.
+
The HygroTrace software is updated and the unit reboots (returns to power up, proceeds through several initiation displays and returns to normal operation).
Parameter Storage
Dump
This function transmits the contents of the EEPROM used for calibration and parameter storage via the RS485 digital link. It is provided to assist GE Sensing service personnel in maintaining the product.
To dump the parameter storage, select Dump, press Enter and the following message appears at the top of the screen for a few seconds.
+
When the dump is complete, Param Storage reappears.
HygroTrace Moisture Transmitter 31
September 2007
Erase
Note: Erase will delete ALL calibration and user settings. The
HygroTrace will be incapable of performing accurate
measurements until this data is restored.
Note: This function is provided to assist GE Sensing service
personnel in maintaining the product.
To erase the EPROM, select Erase, press Enter and the following message appears at the top of the screen for a few seconds.
+
Select YES or NO and press Enter, or press Cancel. If you select NO or press Cancel, the unit returns to the previous screen. If you select YES, the following message appears.
+
When the erasure is complete, Param Storage reappears. To return to the Service Menu, press Cancel.
Heater Test
T o perform a test of the heater, select Heater Test..., press Enter and a screen similar to the following appears:
+
Use the arrow keys to select the appropriate temperature to test, and press Enter. The temperature in the upper right corner of the screen should approximate the one selected.
+
To turn off the test, select OFF and press Enter. The temperature in the upper right corner of the screen returns to its original value. To resume normal operation, select RESUME and press Enter. Press Cancel to return to the Service Menu.
32 HygroTrace Moisture Transmitter
September 2007
Calibration Data
T o change the calibration data, select Cal Data... on the Service Menu, press Enter and the following screen appears:
+
Bulk Limit
To set the bulk limit, select it on the screen and press Enter. The following screen appears:
+
Use the arrow keys to change to the desired value and press Enter (or Cancel to keep the previous value). The screen returns to the Cal Data menu.
+
Calibration Curve
To edit the calibration curve, select Cal Curve... on the Cal Data menu and press Enter. The following screen appears:
To set a calibration point, highlight Select Cal Point, press Enter and the following screen appears.
+
Use the arrow keys to enter the desired calibration point to change and press Enter (or Cancel to keep the previous value). The screen
+
returns to Edit Cal Curve.
Proceed in the same way to Edit PPB Value and/or ZH Value to change the PPB and/or ZH value for a particular calibration point. Then select the next calibration point and repeat the process. When Edit Cal Curve is complete, press Cancel to return to the Cal Data menu.
HygroTrace Moisture Transmitter 33
September 2007
Extrapolation
Reboot
To set coefficients A and/or B, select Extrapolation... and press Enter.
Select the coefficient to change, press Enter and a screen similar to the following appears:
+
Use the arrow keys to change to the desired value and press Enter (or Cancel to keep the previous value). The screen returns to the
+
Extrapolation menu. Press Cancel twice to return to the Service Menu.
T o reboot the meter and return to the initial screen, select Reboot and press Enter. The following screen appears.
+
Select YES or NO and press Enter, or press Cancel. If you select NO or press Cancel, the unit returns to the previous screen. If you select YES, a reboot occurs.
+
Note: The HygroTrace returns to power up, proceeds through
several initiation displays and returns to normal operation.
34 HygroTrace Moisture Transmitter
USA
1100 Technology Park Drive Billerica, MA 01821-4111 Web: www.gesensing.com
Ireland
Sensing House Shannon Free Zone East Shannon, Co. Clare Ireland
Loading...