GE Sensing HygroTrace Operating Manual

GE
Sensing
HygroTrace
Moisture Transmitter
User’s Guide
GE
Sensing
HygroTrace
Moisture Transmitter
916-102A3 September 2007
The HygroTrace Moisture Transmitter is a GE Panametrics product. GE Panametrics has joined other GE high-technology sensing businesses under a new name—GE Sensing.
September 2007
Warranty
Each instrument manufactured by GE Infrastructure Sensing, Inc. is warranted to be free from defects in material and workmanship. Liability under this warranty is limited to restoring the instrument to normal operation or replacing the instrument, at the sole discretion of GE. Fuses and batteries are specifically excluded from any liability. This warranty is effective from the date of delivery to the original purchaser. If GE determines that the equipment was defective, the warranty period is:
one year from delivery for electronic or mechanical failures
one year from delivery for sensor shelf life
If GE determines that the equipment was damaged by misuse, improper installation, the use of unauthorized replacement parts, or operating conditions outside the guidelines specified by GE, the repairs are not covered under this warranty.
The warranties set forth herein are exclusive and are in lieu
of all other warranties whether statutory, express or implied (including warranties or merchantability and fitness for a particular purpose, and warranties arising from course
of dealing or usage or trade).
iii
September 2007
Return Policy
If a GE Infrastructure Sensing, Inc. instrument malfunctions within the warranty period, the following procedure must be completed:
1. Notify GE, giving full details of the problem, and provide the
model number and serial number of the instrument. If the nature of the problem indicates the need for factory service, GE will issue a RETURN AUTHORIZATION NUMBER (RAN), and shipping instructions for the return of the instrument to a service center will be provided.
2. If GE instructs you to send your instrument to a service
center, it must be shipped prepaid to the authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE will evaluate the instrument to determine
the cause of the malfunction.
Then, one of the following courses of action will then be taken:
If the damage is covered under the terms of the warranty,
the instrument will be repaired at no cost to the owner and returned.
If GE determines that the damage is not covered under the
terms of the warranty, or if the warranty has expired, an estimate for the cost of the repairs at standard rates will be provided. Upon receipt of the owner’s approval to proceed, the instrument will be repaired and returned.
iv
September 2007
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Sample System Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Insertion into the Sample System/Process. . . . . . . . . . . . . . . 3
Standard Wiring Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Digital Communications Wiring Connections . . . . . . . . . . . . 7
Powering Up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Using the HygroTrace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Keypad Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
The Default Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Unlocking the Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Accessing the Menus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Entering Numeric Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting Up the Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4-20 Analog Output and Error Reporting . . . . . . . . . . . . . . 14
Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
User. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Communications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
About . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
LOCK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Menu Item Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Moisture Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
v
September 2007

Introduction

The GE Sensing HygroTrace is a compact moisture transmitter designed for measuring ultra-low water content in the ultra high purity and semiconductor markets and related applications. The HygroTrace measures water content in nitrogen or argon gas in the range of 0 to 100 ppb
calibrated range. The unit features an integrated display and a six­button keypad, and is housed in an aluminum enclosure. The HygroTrace operates on 24 VDC and provides 4-20 mA analog and RS485 digital outputs.
The HygroTrace uses an aluminum oxide sensor manufactured with semiconductor techniques. The patented measurement technique enables the HygroTrace to respond very quickly to both dry-to-wet and wet-to-dry situations. The technique applies a temperature pulse to ‘dry’ the sensor. Then, the re-adsorption rate is measured while holding a constant sensor temperature. This measurement is proportional to the moisture concentration in the sample gas. As a result, the HygroTrace offers the sensitivity and response time needed to effectively measure parts per billion levels of moisture.
with trending analysis beyond the
v

Installation

Sample System Guidelines

The HygroTrace transmitter is typically installed in a sample system or on the bypass stream of an ultra-high purity gas distribution system. The factory recommends that the unit be installed in a sample system to protect the sensor from coming in contact with damaging elements in the process, and to allow the transmitter to be removed for service without interrupting the main process gas flow.
Before constructing a sample system, consult a GE Sensing applications engineer and adhere to the guidelines below. See Figure 1 on page 2 for an example of a sample system.
HygroTrace Moisture Transmitter 1
September 2007
Sample System Guidelines (cont.)
A sample system should be kept very simple. It should
contain as few components as possible and all or most of those components should be located downstream of the measurement location.
Sample system components should not be made of material
that will affect measurement. Most common filters and pressure regulators are not suitable for sample systems because they have wetted parts that may adsorb or release moisture, etc. into the sample system. They may also allow ambient contamination to enter the sample system. If possible, use stainless steel material for all wetted parts.
The transmitter should be installed perpendicular to the
sample inlet. For dimensions and other requirements see Specifications on page 26.
Note: A typical sample system for the hygrometer will use VCR
components.
Note: At least 5 feet of 1/4” tubing, if vented to the atmosphere,
should be installed on the outlet to prevent back diffusion of ambient moisture into the sample tee cell of the HygroTrace.
Sample systems should be leak-tested prior to operation to
verify the integrity of the connections, components and fittings. Tighten any loose fittings.
IMPORTANT: Caution must be taken when pressurizing or
depressurizing the sample system to prevent damage to the moisture sensor.

Figure 1: Sample System Example

2 HygroTrace Moisture Transmitter
September 2007

Insertion into the Sample System/Process

!CAUTION!
If the HygroTrace is being installed directly into
the process line, consult the factory for proper
installation instructions and precautions before
beginning the following procedure.
!CAUTION! No maintenance of the sensor can be performed in the field. Accessing the sensor chamber will break the VCR
seal and should be done only at the factory.
To install the transmitter, refer to Figure 2 below and use a wrench to thread the female 1/4” fittings of the sample system or process line onto the 1/4” male VCR fittings of the tee sample cell configuration.
1/4” VCR (Male)
2 places
Tee Sample Cell

Figure 2: HygroTrace Installation

HygroTrace Moisture Transmitter 3
September 2007

Standard Wiring Connections

Note: The HygroTrace is not certified for use in Hazardous
(Classified) locations.
This procedure is for those units not connected to a computer.
The transmitter must be wired using the factory-supplied cable (two meters in length). If an alternate length is required, please contact the factory for assistance.
Note: If cables need to be lengthened, r efer to Table 1 on page 5
to splice an extension onto the existing cable. Connect positive to positive and negative to negative.
Use the following steps to wire the transmitter to the system.
1. Align the red dot with the red mark (see Figure 3 below) and push the connector on the transmitter cable into the mating connector on the transmitter module. Make sure the pins are properly aligned. Once it is inserted, push the cable connector until it locks into place.

Figure 3: Transmitter Cable Connection

4 HygroTrace Moisture Transmitter
September 2007
Standard Wiring Connections (cont.)
2. Using the flying leads at the other end of the transmitt er cable,
connect the transmitter to the power supply and data acquisition system (DAS) as shown in Figure 4 on page 6. Refer to Table 1 below for a description of the leads in the factory-supplied cable.
Table 1: Cable Lead Connections - No PC
Lead Connection Description
Red (+) 24VDC 1 AMP
Black (–) 24VDC 1 AMP
Orange (+) 4-20mA Output
Blue (–) 4-20mA Output
Cable Braid System Ground
3. Trim any unused leads back to the outer cable jacket in order to remove the bare tinned wire and prevent accidental short circuits.
The HygroTrace is now ready for operation.
Note: To remove the probe cable, pull back the spring-loaded
sleeve and pull the cable out of the connector.
HygroTrace Moisture Transmitter 5
September 2007
+24V 1 AMP
+
RS232
TO PC
RS232
RS485 TO
CONVERTER
POWER
SUPPLY
+
4-20mA
METER
COMMON
+ RS485
– RS485
SYSTEM
GROUND
RED
BLACK
ORANGE
BLUE
WHITE
GREEN
BRAID
CABLE

Figure 4: Standard Wiring Connections (ref. dwg #702-684)

6 HygroTrace Moisture Transmitter
September 2007

Digital Communications Wiring Connections

If the unit is to be operated using PanaView™ instrumentation software installed on a computer, an RS232/RS485 converter must be used and the wiring must be set up as follows.
The transmitter must be wired using the factory-supplied cable (two meters in length).
Note: To lengthen cables, refer to Table 2 below to splice an
extension onto the existing cable. Connect positive to positive, negative to negative, and ground to ground.
Note: EMI ferrite shielding may be required for communication
through an ungrounded serial port. Please consult the factory for guidance.
Use the following steps to wire the transmitter to the system.
1. Align the red dot with the red mark (see Figure 3 on page 4) and push the connector on the transmitter cable into the mating connector on the transmitter module. Make sure the pins are properly aligned. Secure the connectors by pushing the cable connector until it locks in place.
2. Using the flying leads at the other end of the transmitt er cable, connect the transmitter to the power supply, the data acquisition system (DAS) and the computer as shown in Figure 4 on page 6. Refer to Table 2 below for a description of the leads in the factory-supplied cable.
Table 2: Cable Lead Connections - with PC
Lead Connection Description
Red (+) 24VDC 1 AMP
Black (–) 24VDC 1 AMP
Orange (+) 4-20mA Output
Blue (–) 4-20mA Output White (+) RS485 Green (–) RS485
Cable Braid System Ground
HygroTrace Moisture Transmitter 7
Loading...
+ 28 hidden pages