The HygroPro Moisture Transmitter is a compact, intrinsically-safe,
loop-powered, 4-20 mA transmitter that provides accurate dew/frost
point measurements over a range of –110 to 20°C (-166 to 68°F). It
features an integrated display and a six-button keypad, and it is
housed in an IP67/Type 4X enclosure.
The HygroPro transmitter uses an aluminum oxide moisture sensor,
and it includes a temperature thermistor and a pressure transducer on
a common mount for calculation of parameters such as:
• ppm
• ppm
in gases
v
in liquids
w
• pounds per million standard cubic feet in natural gas
• process relative humidity.
IMPORTANT: To install the HygroPro in a hazardous (classified) area,
see Hazardous Area Wiring Connections on page 15.
HygroPro User’s ManualPage 1
Chapter 1: Installation
Flow
Probe
Valve 1
Valve 2
Flow Cell
See note below
Intake near the center of the
pipe is recommended.
HygroPro
Needle
Needle
1.2 Sample System Guidelines
The HygroPro transmitter can be installed in a sample system or
directly in the process line. However, GE recommends that the unit be
installed in a sample system to protect the probe from potentially
damaging components in the process stream. Figure 1 below shows a
typical sample system.
Figure 1: A Typical Sample System
Note: At least 5 ft (1.5 m) of 1/4” (6 mm) tubing vented to
atmosphere will ensure an accurate process sample and avoid
diffusion of ambient air moisture back into the process.
Page 2HygroPro User’s Manual
Chapter 1: Installation
1.2 Sample System Guidelines (cont.)
In the sample system shown in Figure 1 on the previous page, fully
open Valve 1 and use Val v e 2 to regulate the sample flow for
measurements at process system pressure. For measurements at
atmospheric pressure, fully open Valve 2 and use Valve 1 to regulate
the sample flow.
Before constructing a sample system, consult a GE application
engineer and observe the following guidelines:
• The sample system should be very simple and should contain as
few components as possible. All or most of those components
should be located downstream of the measurement point.
• Sample system components must not affect moisture readings.
Most common filters and pressure regulators are not suitable for
sample systems because the wetted parts adsorb moisture or
release moisture into the sample system. They may also allow
ambient contamination to enter the sample system. If possible, use
stainless steel for all wetted parts.
• The HygroPro probe should be oriented perpendicular to the
sample system inlet. For dimensions and other sample system
requirements see Mechanical Specifications on page 46.
• Sample systems should be tested for leaks prior to operation, using
a Snoop leak detector, to verify the integrity of connections,
components and fittings.
IMPORTANT: When pressurizing or depressurizing the sample system,
be careful to avoid shock damage to the moisture sensor.
HygroPro User’s ManualPage 3
Chapter 1: Installation
Probe Hex Nut
3/4-16 UNF-2A
Shield
1.3 Mounting the Transmitter
!CAUTION!
If the HygroPro will be installed directly into the process
line, consult GE for proper installation instructions and
precautions before proceeding.
Refer to Figure 2 below and complete the steps on the next page to
install the HygroPro transmitter.
Figure 2: HygroPro Installation
Page 4HygroPro User’s Manual
Chapter 1: Installation
1.3 Mounting the Transmitter (cont.)
1. Make sure the sintered or stainless-steel shield is in place over the
sensor. This shield protects the aluminum oxide sensor from
damage during operation.
2. Using the integral 3/4-16 straight male thread, screw the probe end
of the transmitter into the process or sample system fitting. Make
sure not to damage the threads.
Note: A 3/4-16 to G 1/2 thread adapter is available from GE.
3. Using a 1-1/8” wrench on the probe hex nut, tighten the probe
securely into the process or sample system fitting.
!CAUTION!
Do not apply torque to the transmitter electronic module
to tighten the unit into its fitting.
Note: If there is insufficient space to rotate the HygroPro during
installation, remove the Replaceable Transducer Element
from the transmitter, install it in the fitting, then re-install the
the transmitter onto the RTE.
(RTE)
1.4 Wiring the Transmitter
Proceed to the appropriate section for instructions on:
• Standard Wiring Connections (proceed to next page)
• Hazardous Area Wiring Connections (proceed page 15)
HygroPro User’s ManualPage 5
Chapter 1: Installation
1.4.1 Standard Wiring Connections
The HygroPro is a 4-20 mA loop-powered device that can use the
same two wires for its measurement signal lines and its power supply
lines. Follow the instructions in this section only for units not
requiring hazardous (classified) area protection.
IMPORTANT: To install the HygroPro in a hazardous (classified) area,
see Hazardous Area Wiring Connections on page 15.
The transmitter must be wired using the factory-supplied, 2-meter
cable. If a different cable length is required, please contact the factory
for assistance.
Note: If a longer cable is required, you may splice an extension onto
the standard cable. Be sure to connect leads of the same color
together. For example, connect blue lead to blue lead, brown
lead to brown lead, etc.
The HygroPro may be connected to either of the following types of
external systems:
• an external device that can provide the loop power to the
HygroPro and can also receive and display the 4-20 mA
measurement output from the HygroPro (proceed to next page)
• an external power supply to provide power to the HygroPro plus a
personal computer (PC) running GE’s PanaView™ interface
software (proceed to page 13)
Page 6HygroPro User’s Manual
Chapter 1: Installation
1.4.1a Standard Connections - Without a PC
IMPORTANT: To install the HygroPro in a hazardous (classified) area,
see Hazardous Area Wiring Connections on page 15.
Refer to Figure 3 on the next page and Table 1 below, and complete
the following steps to wire the transmitter.
Table 1: Cable Leads - Without a PC
LeadConnection Description
BluePower Supply (+) [12-28 VDC]
BrownPower Supply (–) [Return]
ShieldEarth Ground [recommended]
1. Push the female connector end of the factory-supplied cable into
the mating male connector on the transmitter module. Make sure
the pins are properly aligned. Then, secure the connectors together
by sliding the metal sleeve on the cable over the connectors and
turning it clockwise until it is tight.
2. Using the flying leads at the other end of the factory-supplied
cable, connect the transmitter to the external system.
Note: The blue and brown leads also carry the measurement signal
current output of 4-20 mA.
3. Trim any unused leads even with the outer cable jacket to remove
any bare wire and prevent accidental short circuits.
HygroPro User’s ManualPage 7
Chapter 1: Installation
12-28 VDC
Return
Power and
Measurement
Circuits
+–
Shield
Earth
Ground
brownblue
1.4.1a Standard Connections - Without a PC (cont.)
Figure 3: Standard Connections - Without a PC
Page 8HygroPro User’s Manual
Chapter 1: Installation
1.4.1b Standard Connections - With a PC
IMPORTANT: To install the HygroPro in a hazardous (classified) area,
see Hazardous Area Wiring Connections on page 15.
Refer to Figure 4 on the next page and Table 2 below, and complete
the following steps to wire the transmitter.
Table 2: Cable Leads - With a PC
LeadConnection Description
BluePower Supply (+) [12-28 VDC]
BrownPower Supply (–) [Return]
WhiteRS485-RS232 Converter (+) [positive]
BlackRS485-RS232 Converter (–) [negative]
GroundEarth Ground
1. Connect an RS485-RS232 converter (customer-supplied) to an
available serial port on the PC.
2. Push the female connector end of the factory-supplied cable into
the mating male connector on the transmitter module. Make sure
the pins are properly aligned. Then, secure the connectors together
by sliding the metal sleeve on the cable over the connectors and
turning it clockwise until it is tight.
3. Using the flying leads at the other end of the factory-supplied
cable, connect the transmitter to the external system.
4. Trim any unused leads even with the outer cable jacket to remove
any bare wire and prevent accidental short circuits.
HygroPro User’s ManualPage 9
Chapter 1: Installation
+ White
Earth
Power
?Black
Ground
PanaView
RS232
RS485 ? RS232
Converter Box
24 VDC
Supply
Blue
Brown
+–
1.4.1b Standard Connections - With a PC (cont.)
Figure 4: Standard Connections - With a PC
Page 10HygroPro User’s Manual
Chapter 1: Installation
1.4.2 Hazardous Area Wiring Connections
Before installing and using the HygroPro in a hazardous (classified)
area, be sure to read and understand all applicable reference materials.
This includes:
• all EU or North American Standards and Directives (see Tab le 3
and Tab l e 4 on page 17)
• all local safety procedures and practices
• the FM Schematic drawing (see Figure 5 on the next page)
• this user’s manual
Note: It is the installer’s responsibility to follow all applicable
standards and procedures.
!WARNING!
The procedures in this section must be performed only
by trained technicians who have the necessary skills
and qualifications.
HygroPro User’s ManualPage 11
Chapter 1: Installation
Hazardous or Non-Hazardous Location
Class I, Division 1, Group ABCD
Class II, Division 1, Group EFG
Class III, Division 1
Non-Hazardous Location
HygroPro
Loop Terminal
Safety Entity Parameters:
Vmax = 28 V
Imax = 93.3 mA
Pi = 0.653 W
Ci = 0
Li = 62 uH
RS485 Comm Terminal
Safety Entity Parameters:
Vmax = 3.72 V
Imax = 228 mA
Pi = 212 mW
Ci = 0
Li = 62 uH
Notes:
1. Installation shall be in accordance with the National Electrical Code
(ANSI/NFPA 70) or Part One of the Canadian Electrical Code (C221),
as applicable.
2. The subject equipment is FM Approved, and FM Approved for Canada for the
following Hazardous (Classified) Locations and Protection Methods:
Intrinsically Safe, Class I, II, III, Division 1, Group ABCDEFG, T4
T Ambient: -40 C to 60 C
Enclosure: IP 67
3. Galvanically Isolated FM Approved (US) or Canadian Certified (for Canada)
associated apparatus installed in accordance with the manufactruer's instructions.
blue
brown
black
white
Associated Apparatus
Note 3
Loop-Power Device
Note 4
RS485 Comm
Note 4
Associated Apparatus
Note 3
4. Equipment to which associated apparatus is connected shall not
use or generate in excess of 250V.
5. Maintain the following relationship between Intrinsic Safety (Entity)
Parameters of HygroPro and associated apparatus:
HygroPro Vmax > associated apparatus Voc
HygroPro Imax > associated apparatus Isc
HygroPro Pi > associated apparatus Po
HygroPro Ci + Ccable < associated apparatus Ca
HygroPro Li + Lcable < associated apparatus La
Use cable P/N 230-049-01
1.4.2 Hazardous Area Wiring Connections (cont.)
Figure 5: FM Schematic (ref. dwg #752-262)
Page 12HygroPro User’s Manual
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.