The FGA 311 In Situ Flue Gas Analyzer is a GE Panametrics product. GE Panametrics has joined other GE
high-technology sensing businesses under a new name—GE Industrial, Sensing.
March 2006
WarrantyEach instrument manufactured by GE Infrastructure Sensing, Inc. is
warranted to be free from defects in material and workmanship.
Liability under this warranty is limited to restoring the instrument to
normal operation or replacing the instrument, at the sole discretion of
GE Infrastructure Sensing, Inc. Fuses and batteries are specifically
excluded from any liability. This warranty is effective from the date of
delivery to the original purchaser. If GE Infrastructure Sensing, Inc.
determines that the equipment was defective, the warranty period is:
• one year for general electronic failures of the instrument
• one year for mechanical failures of the sensor
If GE Infrastructure Sensing, Inc. determines that the equipment was
damaged by misuse, improper installation, the use of unauthorized
replacement parts, or operating conditions outside the guidelines
specified by GE Infrastructure Sensing, Inc., the repairs are not
covered under this warranty.
The warranties set forth herein are exclusive and are in lieu of
all other warranties whether statutory, express or implied
(including warranties of merchantability and fitness for a
particular purpose, and warranties arising from course of
dealing or usage or trade).
Return PolicyIf a GE Infrastructure Sensing, Inc. instrument malfunctions within the
warranty period, the following procedure must be completed:
1. Notify GE Infrastructure Sensing, Inc., giving full details of the
problem, and provide the model number and serial number of the
instrument. If the nature of the problem indicates the need for
factory service, GE Infrastructure Sensing, Inc. will issue a RETURN
AUTHORIZATION number (RA), and shipping instructions for the
return of the instrument to a service center will be provided.
2. If GE Infrastructure Sensing, Inc. instructs you to send your
instrument to a service center, it must be shipped prepaid to the
authorized repair station indicated in the shipping instructions.
3. Upon receipt, GE Infrastructure Sensing, Inc. will evaluate the
instrument to determine the cause of the malfunction.
Then, one of the following courses of action will then be taken:
• If the damage is covered under the terms of the warranty, the
instrument will be repaired at no cost to the owner and returned.
• If GE Infrastructure Sensing, Inc. determines that the damage is not
covered under the terms of the warranty, or if the warranty has
expired, an estimate for the cost of the repairs at standard rates
will be provided. Upon receipt of the owner’s approval to proceed,
the instrument will be repaired and returned.
IntroductionThe FGA 311 In Situ Flue Gas Oxygen Transmitter monitors the
combustion efficiency of a boiler or furnace by measuring the
percentage of oxygen in the flue gases. This self-contained instrument
is installed so that its zirconium oxide oxygen sensor is located
directly in the stream of the flue gases. The measurements may be
recorded via the built-in 4-20 mA output or transmitted to a computer
terminal via the built-in RS232 interface.
The FGA 311 is available with either the standard weatherproof
enclosure or the optional explosion-proof enclosure, as shown in
Figure 1-1 on page 1-2. However, the operation and the internal
components are the same for both configurations.
The FGA 311 consists of the following sub-assemblies:
• An aluminum enclosure, which includes a breather port, the unit’s
mounting threads, a calibration gas connection and a reference air
connection. In addition, the enclosure houses the printed circuit
board and provides two ports for electrical conduit or cable glands.
• All of the electrical components, including the replaceable fuse,
are mounted on the printed circuit board (PCB). All electrical
connections for the line power, 4-20 mA recorder output and
RS232 interface are made to the PCB.
• A probe assembly that extends into the flue gas stream.
• A heater sub-assembly that maintains the oxygen sensor at the
proper operating temperature for efficient operation.
• A zirconium oxide oxygen sensor to measure the percentage of
oxygen in the flue gases.
• A filter assembly that protects the oxygen sensor from particulates
in the flue gas stream.
The FGA 311 has been designed to permit troubleshooting,
maintenance and adjustment of the instrument without removing the
unit from the flue.
General Information1-1
March 2006
Standard
(Weatherproof)
Explosion-Proof
(Flameproof)
Side Views
Flame Arrestor
R
I
S
R
O
V
U
S
T
T
E
N
S
I
O
N
N
D
E
S
I
G
R
E
N
E
U
O
S
A
P
E
N
D
O
N
O
T
O
P
E
N
E
L
W
I
H
End Views
N
O
D
N
E
P
H
I
W
L
E
N
E
P
O
T
O
A
S
O
U
E
N
E
R
G
I
S
E
D
N
O
I
S
N
E
T
S
U
V
O
R
S
I
R
Flame Arrestors
Figure 1-1: The FGA 311 Enclosures
1-2General Information
March 2006
Principles of OperationIn an ideal combustion process, a precise ratio of air to fuel is burned
efficiently to yield only heat, water vapor, and carbon dioxide.
However, because of burner aging, imperfect air to fuel mixtures,
variable firing rates and/or inaccurate ignition timing, this situation
rarely happens.
A sure sign of a less than ideal combustion process is the presence of
excess oxygen in the flue gases. The level of this excess oxygen is
easily monitored with the FGA 311 In Situ Flue Gas Oxygen
Transmitter, and the information can then be used to make the
necessary adjustments to improve the efficiency of the combustion
process. The following two major components are included in the
FGA 311 analyzer:
• a zirconium oxide oxygen sensor
• a loop-controlled heater circuit
These components are described in the sections that follow.
The Oxygen SensorThe inside and outside of the zirconium oxide oxygen sensor are
coated with a porous platinum catalyst, forming two electrodes. Flue
gases flow past the outside of the sensor, while atmospheric air
circulates freely on the inside of the sensor. The atmospheric air is
used as the reference gas for the oxygen measurements. See the
oxygen sensor sub-assembly shown in Figure 1-2 below.
At the normal operating temperature (650° to 1100°C) of the sensor,
the oxygen molecules in the atmospheric reference air (20.93%
oxygen) are electrochemically reduced at the inner electrode. The
resulting oxygen ions seek an equilibrium with the lower oxygen
concentration on the sample gas side of the sensor, by migrating
through the porous ceramic toward the outer electrode.
Packing ScrewSleeveO-Ring
O-RingPackingOxygen Cell
Figure 1-2: The FGA 311 Oxygen Sensor
General Information1-3
March 2006
The Oxygen Sensor
(cont.)
At the outer electrode, the oxygen ions give up their extra electrons
and revert to oxygen molecules, before being swept away by the flue
gas stream. This exchange of electrons at the electrodes generates a
voltage gradient across the sensor. See Figure 1-3 below.
The lower the concentration of oxygen in the flue gases, the greater
the rate of ion migration through the ceramic and the higher the
resulting voltage gradient across the sensor. In fact, the sensor’s
voltage output rises logarithmically as the percentage of oxygen in
the flue gases decreases. This enables the FGA 311 to accurately
measure very small levels of oxygen in the flue gases.
Oxygen ions migrate through the Zirconium Oxide
from Inside to Outside
O
2
O
O
2
2
O
2
O
2
From Outside
Electrode
When O concentration in sample gas falls,
cell voltage rises with increased O migration rate.
Volts
2
O
2
Zirconium Oxide Ceramic
with Lattice Imperfections
From Inside
Electrode
2
Figure 1-3: Oxygen Migration Through the Sensor
1-4General Information
March 2006
The Heater Control
Circuit
The oxygen sensor temperature in the FGA 311 is maintained by a
heater, which is part of a complex temperature control loop. This
circuit constantly monitors the oxygen sensor temperature, compares
it to the set point temperature (700°C), and turns the heater ON or
OFF accordingly. The specific type of control circuit used is called a
Proportional Integral Derivative (PID) loop, because of the three
adjustable parameters involved:
• Proportional Band: Because the system cannot respond
instantaneously to temperature changes, the actual temperature of
the oxygen sensor oscillates about the set point. In general,
increasing the proportional band reduces the magnitude of these
temperature oscillations.
• Integral Action: A consequence of increasing the proportional
band is the introduction of an offset between the set point and the
control point. The integral portion of the control loop acts to move
the control point back toward the set point within a specified
period of time. Thus, decreasing this integration time reduces the
offset more quickly.
• Derivative Action: The derivative portion of the control loop
applies a corrective signal based on the rate at which the actual
temperature is approaching the set point. In effect, the derivative
action reduces overshoot by counteracting the control signal
produced by the proportional and integral parameters.
The heater control circuit is configured at the factory for optimum
performance. Because of the strong interaction between the three
parameters involved, properly setting up the PID loop is a very
complex matter. As a result, randomly changing the P, I and/or O
parameters can seriously degrade the performance of the FGA 311.
IMPORTANT:Always contact the factory before attempting to
IntroductionThis chapter gives directions for the proper installation and wiring of
the FGA 311. The following specific topics are included:
• unpacking the unit
• selecting the site
• mounting the analyzer
• wiring the analyzer
!WARNING!
To ensure safe operation of the FGA 311, it must be
installed and operated as described in this manual. In
addition, be sure to follow all applicable local safety codes
and regulations for installing electrical equipment.
Unpacking the UnitRemove the analyzer from its shipping container and make sure that
all items on the packing slip have been received. If anything is
missing, contact the factory immediately. The analyzer, as shown in
Figure 2-1 below, is shipped fully assembled and ready to install.
3/4" NPT
6"
Enclosure
6
6
5
5
4
4123
123
3/4" NPT
Caution!
When unpacking the analyzer, be careful not to damage
the probe. It is covered with a porous ceramic filter that
will easily crack if it is knocked against a hard surface.
PC Board
Calibration Gas Inlet
1-1/2" NPT
26-1/4"
Heater
19"
Probe
Sensor
Figure 2-1: The FGA 311 In Situ Flue Gas Analyzer
1-1/4"
Installation2-1
March 2006
Line Power RequirementsEach FGA 311 analyzer is factory-configured for the proper line
voltage, as specified at the time of purchase. The available options
include the following:
• Japan = 100 VAC
• U.S.A. = 110/120 VAC
• Europe = 220 VAC
• Australia = 240 VAC.
Caution!
To change the line voltage to the unit, contact the factory
for instructions. DO NOT make such a change without first
obtaining proper instructions.
Selecting the SiteEnvironmental and installation factors should already have been
discussed with a GE Sensing applications engineer or field sales
person before the FGA 311 arrives.The analyzer must be installed
either in a furnace or boiler wall or in a flue duct. Ideally, the end of
the probe assembly should extend approximately 1 ft (30.5 cm) into
the flue gas stream. Also, the analyzer should be positioned so that
the probe holes are on the downstream side of the probe (see
Figure 2-2 on page 2-4).
• For furnaces, locate the analyzer close to the combustion zone,
typically within the radiant section and always before the
convection section. Make sure that the probe’s maximum operating
temperature is not exceeded and that the probe is not situated in a
non-homogeneous flue gas mixture.
IMPORTANT:If the ambient temperature in the vicinity of the probe
can exceed 650°C (1202°F), a high temperature
probe assembly is required.
• For boilers, locate the analyzer downstream of the heat exchanger
and just before the economizer air heater, if one is installed. The
analyzer should not be placed downstream of any air heater,
because of possible air leaks that can cause inaccurate readings.
In general, the sample point should be an area of high turbulence,
which will ensure a good homogeneous mixture of the flue gases.
Conditions to be avoided would include air leaks upstream of the
sample point and dead spaces in the vicinity of the sample point.
2-2Installation
March 2006
Mounting the AnalyzerThe FGA 311 has integral male 1-1/2” NPT mounting threads. This
permits a flange to be threaded onto the analyzer, and the resulting
assembly is then bolted to a mating flange on the furnace/boiler wall
or flue duct.
Note: For installations where the FGA 311 may be exposed to water
or other fluids, install the unit with the breather (see
Figure 2-2 on page 2-4) facing downward.
Do not use any thread sealant during the installation. Upon
heating, Teflon tape will melt and other sealants may emit
gases that interfere with the oxygen readings.
Carefully follow the instructions on page 2-5 to mount the FGA 311
In Situ Flue Gas Oxygen Transmitter. The unit may be mounted in
either a horizontal or vertical orientation.
IMPORTANT:Direct mounting of the FGA 311 into a threaded hole
using its mounting threads is not recommended.
Always use a mounting flange.
To prepare the installation site for mounting the analyzer, complete
the following preliminary steps:
1. Fasten a short section of pipe having at least a 2” inside diameter
into the process wall.
IMPORTANT:Make sure that the pipe is long enough to permit
installation of the flange mounting hardware and that
the flange is oriented with its bolt holes straddling
the vertical and horizontal center lines.
2. Weld a mating flange onto the end of the pipe, as shown in
Figure 2-2 on page 2-4.
Note: Standard available factory options include 3”-300 lb ANSI,
4”-150 lb ANSI and DN80 PN16 mounting flanges. A
separate mating flange can also be supplied with the unit.
Installation2-3
March 2006
GAS
FLOW
Input Scoop
GAS
Holes
Holes
FLOW
Standard Probe
Mounting Wall
Cable Gland
Calibration
1m, 1.5m & 2m Probes
90°
2" Min.
Hardware
Flanges
Gas Line
Gasket
FG311
Cable Gland
Control Cable
Power Cable
Reference
Air Line
Breather
Figure 2-2: Flange Mounting the FGA 311
2-4Installation
March 2006
Mounting the Analyzer
(cont.)
Before mounting the FGA 311 analyzer assembly onto the mating
flange, mark the enclosure in line with the probe holes. Then,
complete the following steps:
IMPORTANT:The probe holes must be positioned on the
downstream side of the probe. See Figure 2-2 on
page 2-4 for the proper orientation.
1. Slide the gasket over the probe assembly and up against the
mounting flange on the analyzer.
Note: Be sure to use a suitable high temperature gasket for this
application.
2. Orient the analyzer so the probe holes are properly positioned on
the downstream side of the probe, and slide the probe through the
hole in the mounting wall until the two flanges meet.
3. Secure the analyzer in place by fitting bolts into the matching
flange mounting holes and fastening the bolts with nuts and
washers. Make sure that the gasket is properly positioned between
the two flanges.
This completes the mounting of the FGA 311. The required external
connections are discussed in the following sections.
Installation2-5
March 2006
Wiring the AnalyzerConnect the power and control signal wiring as described below and
as shown in Figure 2-3 on page 2-7, which shows the printed circuit
board mounted within the open FGA 311 enclosure. The necessary
connectors are supplied with the unit, and they are plugged into the
mating connectors on the printed circuit board prior to shipment. DO NOT power up the unit until instructed to do so!
ATTENTION EUROPEAN CUSTOMERS!
In order to meet CE Mark requirements, all electrical cables
must be installed as described in Appendix C, CE Mark
Compliance.
Wiring the Line PowerUse one of the 3/4” NPT connections on the enclosure for installation
of the power cable gland or conduit. The FGA 311 has been preset at
the factory for the line voltage specified at the time of purchase.
Never connect a different line voltage to the unit without first
obtaining instructions from the factory.
!WARNING!
Improper connection of the power line or connection to the
wrong voltage may result in an electrical hazard.
Note: For compliance with the European Union’s Low Voltage
Directive (73/23/EEC), this unit requires an external power
disconnect device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly visible,
directly accessible, and located within 1.8 m (6 ft) of the
FGA 311.
The FGA 311 is designed to comply with the LVD Directive
per the requirements of EN 61010 with the following
exception: The 230 VAC unit passes with a test voltage of
1,800 VAC. (Refer to Table D.10 of Annex D in EN 61010.)
Connect the line power to terminal block TB1, as shown in Figure 2-3
on page 2-7, by completing the following steps:
1. Connect the
2. Connect the
LINE (black) lead to pin #1 on TB1.
NEUTRAL (white) lead to pin #2 on TB1.
3. Connect the
GROUND (green) lead to the earth ground screw inside
the enclosure.
Proceed to the next section to wire the control signals.
2-6Installation
March 2006
Heater Connections
J3 Terminal Block
Pin
1
2
3
Description
Heater
No Connection
Heater
NOTE: These connections are made
at the factory.
U15
+
C1
1
E4
TB2
MODE
CAL
S3
ADJ
S1
S2
TB1 Power*
Terminal Block
Description
Pin
1
Line
Neutral
2
Power Voltage Options
Factory Installed
PowerJumpersFuse
115VW1,W3,W4,W51/2A
W2,W6230V1/4A
*Ground wire goes to
earth gnd connection.
6
7
8
2
3
C1
+
C2
U5
1
3
9
2
1
R
U
V
0
1
E
4
3
R
1
FAULT
10E85
2
1
S
S
D
D
W4
W6W5
E6
J2
4
E11 E12
S
D
E9
25
3
S
D
J1
W2
1
703-1216
T1
TB1
E3
1
W3
J3
C16
F1
3
L1
4
C15
U18
5
S
D
E7
E2
E1
E5
3
1
5
4
W1
RT1
C14
1
2
Fuse
RS232 Analog Output
TB2 Terminal Block
PinDescription
1Return
2Receive
3Send
44-20 mA Out–
54-20 mA Out+
1. PC Board may be oriented differently,
depending on how the unit is installed.
Future Use (CJC–)6
NOTE: These connections are made
at the factory.
2. J1 is reserved for future use.
(Shown with sensor/heater assembly and cable removed for clarity.)
Figure 2-3: Connections to the FGA 311 PC Board
Installation2-7
March 2006
Wiring the Control SignalsUse the remaining 3/4” NPT port on the FGA 311 enclosure for
connecting the control cable gland or conduit. For cable runs of less
than 5 m (16.4 ft), standard twisted pair cable may used for these
connections. However, shielded cable should be used for longer cable
lengths and/or for CE Mark compliance.
IMPORTANT:This symbol indicates Caution - risk of electric
shock:
The control signal connections are made to the printed circuit board
on terminal block TB2, as shown in Figure 2-3 on page 2-7. To make
the necessary connections, complete the following steps:
1. Connect the RS232 serial port leads to TB2 as follows:
Note: The serial port connection should be made with a GE Sensing
#704-668 cable assembly or its equivalent. To assemble an
equivalent cable, refer to Figure B-6 in Appendix B of this
manual.
a. Connect the
Ground lead (the green wire from pin #5 of the
DB9 connector on the PC) to pin #1.
b. Connect the
Receive lead (the red wire from pin #2 of the DB9
connector on the PC) to pin #2.
c. Connect the
Transmit lead (the white wire from pin #3 of the
DB9 connector on the PC) to pin #3.
2. Connect the 4-20 mA analog output leads to TB2 as follows:
a. Connect the 4-20 mA
b. Connect the 4-20 mA
Return (-) lead to pin #4.
Signal (+) lead to pin #5.
Proceed to the next section for a description of the internal, factoryinstalled wiring.
2-8Installation
March 2006
Factory ConnectionsIn addition to the user connections described above, the following
connections are made at the factory to terminal blocks J2 and J3.
• Thermocouple connections to J2 on pins #1 (+) and #2 (-).
• Oxygen sensor connections to J2 on pins #3 (+) and #4 (-).
• Heater connections to J3 on pins #1 and #3.
For reference purposes only, these factory connections are shown in
Figure 2-3 on page 2-7.
Reference Air and
Calibration Gas Lines
During normal operation, the FGA 311 requires a constant supply of a
reference air for the zirconium oxide oxygen sensor. The
recommended gas for this purpose is instrument air (containing
20.93% oxygen) at a flow rate of 20-50 cc/min. Connect this gas
supply, with 1/4” tubing, to the 1/4” compression fitting provided.
Refer to Figure 2-4 on page 2-10 for the location of this connector.
Also, a separate gas supply is required for calibration of the
instrument. This line should be connected, with 1/4” tubing, to the
remaining 1/4” compression fitting on the FGA 311. See Figure 2-4
on page 2-10 for the location of this connector.
Note: The calibration gas port must remain capped, if no permanent
plumbing is attached. See Chapter 4, Calibration, for a
discussion of the recommended calibration gases and
procedures.
In addition to the basic reference air and calibration gas lines, a
variety of ancillary equipment is recommended. A typical system that
will ensure efficient and reliable operation of the FGA 311 is
illustrated in Figure 2-4 on page 2-10.
IntroductionSince the FGA 311 In Situ Flue Gas Oxygen Transmitter is a
monitoring device, operation of the installed analyzer is simple. Once
it has been properly installed and set up, it will simply begin taking
readings. However, the analyzer should be allowed to warm up for at
least one hour (three hours if possible) prior to use. See Chapter 2,
Installation, if all of the required installation requirements have not
yet been completed.
Calibration of the unit should be checked once or twice a week for the
first month of operation and once every 2-3 months thereafter. See
Chapter 4, Calibration, for the correct procedures.
!WARNING!
To ensure safe operation of the FGA 311, it must be
installed and operated as described in this manual. In
addition, be sure to follow all applicable local safety codes
and regulations for installing electrical equipment
Preventing Common
Problems
Because of the extreme conditions in monitoring flue gases and the
complexity of the FGA 311’s measurement techniques, some simple
precautions should be taken with the instrument. Failure to observe
these basic procedures can lead to operational difficulties.
Compliance with the following instructions will help to eliminate
such common problems:
• Do not use pipe thread compounds on any part of the FGA 311.
Many pipe thread compounds emit combustible vapors that may
cause inaccurate readings.
• Do not handle the sensor assembly any more than is absolutely
necessary. Although some scratches on the platinum electrode can
be tolerated, rubbing the coating should be avoided. Also, the
transfer of skin oils to the electrode can cause erroneous readings.
• Scrubbing the sensor while washing it or washing a hot sensor can
damage or destroy it. Clean the sensor only by rinsing it with clean
water, after the sensor has cooled.
• Installing a cold probe assembly into a hot flue gas stream can
cause damage to the filter and/or sensor. Always allow the probe
assembly to gradually heat up to normal operating temperature,
before subjecting it to hot flue gases.
If any problems not covered in this manual are encountered, contact a
GE Sensing representative for assistance.
Operation and Programming3-1
March 2006
Powering UpBefore powering up the unit, start the flow of reference air. Be sure
that the reference air source is connected to the correct port, as shown
in Figure 2-4 on page 2-9. As stated in Chapter 2, Installation, the
recommended reference air is instrument air (20.93% oxygen) at a
flow rate of 20-50 cc/min.
Note: The zirconium oxide oxygen sensor can not provide accurate
readings without a known oxygen percentage on the reference
side of the cell. Allow the reference air to flow for at least five
minutes prior to operation.
Power may now be applied to the unit. Since the FGA 311 does not
have its own power switch, the main disconnect must be used to
power the analyzer on. Simply place this switch in the ON position,
and allow the analyzer to warm up for at least one hour (three hours if
possible) before taking any readings.
Note: For compliance with the European Union’s Low Voltage
Directive (73/23/EEC), this unit requires an external power
disconnect device such as a switch or circuit breaker. The
disconnect device must be marked as such, clearly visible,
directly accessible, and located within 1.8 m (6 ft) of the
FGA 311.
The FGA 311 is designed to comply with the LVD Directive
per the requirements of EN 61010 with the following
exception: The 230 VAC unit passes with a test voltage of
1,800 VAC. (Refer to Table D.10 of Annex D in EN 61010.)
The red fault light (DS4) and one of the green oxygen range
indicators (DS1, DS2 or DS3) on the printed circuit board will blink
until the FGA 311 has reached its normal operating temperature of
700°C. Then, the fault light will go out and the range indicator will
glow steadily.
3-2Operation and Programming
March 2006
Taking MeasurementsAfter the FGA 311 has warmed up, the voltage output of the
zirconium oxide oxygen sensor will vary logarithmically with the
oxygen concentration in the flue gases, according to the Nernst
equation (see Appendix A, The Nernst Equation, for details):
20.93
EmV()48.274
In Equation 3-1 above, “E” is the voltage in millivolts generated by
the sensor at an operating temperature of 700°C.
The built-in thermocouple temperature sensor in the FGA 311
automatically adjusts the constant used in the Nernst equation to
reflect the precise actual temperature of the oxygen sensor. In
addition, the non-linear output voltage signal generated by the oxygen
sensor is internally converted into a linear 4-20 mA output current
signal, which is sent to pins 4 and 5 of terminal block TB2. The 4-20
mA current range corresponds to a flue gas oxygen range of 0% to the
programmed O
ammeter or a recording device to these terminals, the oxygen content
of the flue gases may be continuously monitored.
range (5, 10 or 25%). By connecting a digital
2
-------------
log•=
%O
2
(3-1)
Data RecordsAppendix D, Data Records, at the back of this manual provides
several tables for entering all of the relevant data pertaining to the
installation and programming of the FGA 311. Be sure to maintain
the accuracy of this data on a regular basis. In the event of problems
with the unit, the data records may provide valuable information to
assist in the troubleshooting procedure.
Menu MapA complete menu map of the FGA 311’s built-in software is shown in
Figure 3-3 on page 3-22. Refer to this figure as needed to supplement
the step-by-step programming instructions that follow.
Programming OptionsThe following two methods for programming the FGA 311 In Situ
Flue Gas Oxygen Transmitter may be used to navigate through the
User Program:
• manual switches, which are located inside the enclosure, on the
analyzer’s printed circuit board
• a terminal or computer, utilizing the unit’s built-in RS232 serial
interface
Note: The entire
programming switches.
EXTRA menu is not accessible via the manual
Operation and Programming3-3
March 2006
Programming with
Manual Switches
CAL
S3
MODE
9
8
7
6
5
4
Manual programming of the FGA 311 is accomplished with a series
of switches and LEDs located on the printed circuit board. These
components, which are shown in Figure 3-1 below, may be accessed
by removing the cover from the FGA 311.
Although the menu map shown in Figure 3-3 on page 3-22 does apply
to both the manual programming mode and the RS232 programming
mode, some of the programming options are not available via the
manual switches. Specifically, the entire EXTRA menu can not
accessed via the PC board switches.
6
78
W2
W1
T1
TB1
E3
1
W3
C14
J3
C16
F1
3
L1
4
C15
U18
E7
E2
E1
E5
3
1
5
RT1
1
2
DS5
0
1
2
3
TB2
S1
510
E8
1
FAULT
4
S
D
25
E9
2
3
1
C
U15
+
C2
1
U5
+
C1
1
CAL
S3
TB2
MODE
ADJ
9
2
1
R
U
V
E4
0
1
E
4
3
R
1
S1
FAULT
10E85
25
3
2
1
S
S
S
D
D
D
S2
34
W4
W6 W5
E6
J2
4
E11 E12
S
D
E9
1
J1
703-1216
5
S
D
1
2
3
S
S
S
D
D
S2
D
ADJ
Figure 3-1: Circuit Board Switches and LEDs
3-4Operation and Programming
Loading...
+ 82 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.