GE Sensing Druck PC6-IDOS User Manual

Druck PC6-IDOS
Pressure calibrator/indicator
User manual - K0355
GE
Sensing
A-8010 Graz, Riesstraße 146
Sensorik
Messtechnik
Handelsgesellschaft m.b.H.
Publication
K0355, Issue 1
Copyright
© 2006 General Electric Company. All rights reserved.
Trademarks
All product names are trademarks of their respective companies.
CONTENTS
Welcome ( Indicator Note ) Safety Warning Quick Reference
Specifications
1.0
1.1
1.2
1.3
1.4
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
PC6-IDOS Arrangement Functions (menu 1) Functions (menu 2)
1.5
1.6
1.6.1
1.6.3
1.6.5
1.6.6
1.6.7
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6
1.7.7
1.7.8
1.7.9
1.7.10
1.7.11
1.7.12
1.7.13
1.7.14
1.7.15
1.7.16
1.7.17
1.7.18
1.7.19
1.8
1.9
1.10
1.11
1.12
1.13
2.0
2.1
2.2
2.3
2.4
2.5
2.6
2.10
2.12
3.0
3.1
3.2
3.3
Menu 1, Menu 2 & Menu 3 Menu 1
Menu 2
Menu 3
...................... ii
........................................... iii
iv
............................................. v
Introduction ...................................... 2
General ............................................. 2
Keypad ............................................. 2
LCD .................................................. 3
Connection Panel ............................. 3
2 mm Sockets ............................ . 3
Quick connect coupling ........... .. 3
RS232 ..................................... .... 3
Charger ...................................... . 3
EPM ........................................... . 3
............................... 4
....................................
.................................... 6
Introduction ......................................... 10
Connections
Pressure Input .................................... 10
Voltage/Current Input ...................... 10
RS232 ................................................ 11
Battery charger ................................... 11
Voltage Output .................................. 11
Messages & Parameters
Retained parameters .... .. 12
Battery low state .............................. . 12
Tare Value .......................................... 12
Maximum/Minimum ......................... 12
Void leak test ..................................... 12
Switch Off .......................................... 12
Auto Switch Off .................................. 12
Pressure Overload ............................ 12
Low Battery Detect ............................ 12
Pressure Switch State Change .......... 12
Printer Busy ........................................ 12
Outside User Set Alarm Value ........... 12
Display Hold ...................................... 12
Connection Fault ............................... 13
Short Detected .................................. 13
Voltage / Current Overscale .............. 13
RS232 Output Format ....................... 13
Alarm Flag ......................................... 13
Symbols ............................................. 13
Units ................................................... 15
Max,Min, Switch, Temp.,
Tare .................................................... 15
V / mA ................................................. 15
Leak Test ............................................ 17
File (logged files) ................................ 18
Alarm Settings - High / Low ............... 21
V / mA (Measure mode)............................ 22
V-Out, Display light, Port ................... 26
Resolution ........................................... 26
Event ................................................... 26
Log ..................................................... 27
Zero .................................................... 30
Port
Power_Off ........................................... 31
Span ................................................... 33
History (overload, zero/span, ...)........... 35
Language .................................... 35
Units ................................................... 35
..........................
Unit, Leak ....................................... 38
V / mA ............................................ 39
File ................................................. 40
High & Low .................................... 41
V / mA (measure) ........................... 42
V / mA (source) .............................. 43
V-out, Display light ....... 48
Resolution, Event, RS232 ............. 48
Log ................................................ 49
Date&Time, Language, Zero, Port,
Power_Off ...................................... 50
Span .............................................. 51
History ............................................ 52
Language ........................................ 53
Units ................................................ 54
(Keys, Activating functions, Exit,
Menu selection, Data Entry, Tare) ............................
Maintenance
Functions (menu 3)
1.6.2
1.6.4
2.7
2.8
2.9
................................................ vi
........
..........
........
........
........
.................................... 8
......................................... 10
.
EPM ..................................................... 10
Switch Test........................................... 11
.
.
.................... 12
.................. ....
.
Date& Time,% 15
.
V / mA (Source mode)...............................23
..
.
RS232 ................................................. 29
Date & Time ....................................... 30
Language ............................................. 30
....................................................... 31
.
.........
..
.................
........
5
37
1.7
2.11
Description 1
Operating Instructions 9
How to Select Functions (Menu 1) 14
Set-menu (menu 2) 20
Cal-menu (menu 3) 32
Flowcharts 36
Appendices 55
Index 60
1 External Pressure Module (EPM) 55
2 Calibration Procedures......................... 58
Page i
WELCOME
This detailed operating manual will help you to become familiar
with the many features of the PC6-IDOS Calibrator. The simple step by step
instructions will quickly guide you through the procedures for using the
calibrator and other items needed for accurately testing and calibrating
numerous types of instruments. Please t
efore you begin to use the PC6-IDOS.
ake time to carefully read
the whole manual b
Two versions of the PC6-IDOS are available:
PC6-IDOS Calibrator:
PC6-IDOS Indicator:
All of the functions shown in this manual apply to this unit. The Calibrator can be supplied with a PV211 pneumatic test system for pressures up to 40 bar (600 psi) or PV212 hydraulic test system for pressures up to 700 bar (10 psi). Refer to the instructions supplied with the applicable pressure system.
This is a simplified version of the Calibrator and does not include the Voltage/Current input/output, functions.
PC6-IDOS
PV210 Low pressure pneumatic test system for pressures up to 3 bar (45 psi),
,000
PC6-IDOS
or EPM
Page ii
SAFETY WARNING!
HIGH PRESSURE
RECHARGEABLE Ni-MH BATTERIES
:
Uncontrolled release of high pressure is hazardous to personnel and may cause damage to equipment. Before connection of any pressure component to the PC6­IDOS ensure that the component(s) is/are isolated from the pressure supply and any internal pressure is released slowly.
Rechargeable Ni-MH batteries must be recycled or disposed of properly.
CAUTION: Use charger supplied by GE Sensing only.
May explode if damaged or disposed of in fire. DO NOT short-circuit.
Ni-MH
Page iii
QUICK REFERENCE
Keys
Press to turn PC6-IDOS on and off
Press to access the menus
Press to accept functions or settings
Press to move cursor up/left or increase values
Press to move cursor down/right or decrease values
Press to print/log/start event depending on previous setting
Select menu position cursor press the Enter key
under function
To exit out of any menu select 'Exit' option, or press key.
Activating Functions
Exit
Menu Selection
Pressure Display
1st Press for Menu 1
2nd Press for Menu 2
3rd Press back to Pressure Display /
Calibration Menu (see page 33)
Press or to position cursor and press
to select required alphanumeric. Press
key to release cursor.
For next data entry, repeat the above procedures then press once all data is entered.
Note: To quickly scroll through alphanumeric lists hold down the or key.
Data Entry
Tare
This is a temporary function and will be cancelled when the unit is switched off.
Hold down the key for 4 seconds while in the pressure display mode. The unit will bleep after two seconds when the Max. and Min. values reset. After another two seconds a second bleep will sound as the displayed pressure now resets to zero.
MENU
MENU
MENU
MENU
MENU
ON/OFF
MENU
MENU
Page iv
SPECIFICATIONS
* Temperature coefficient ± 0.005% of reading / °C w.r.t. 20°C (68°F)
15% FS 6 Digits External Pressure Modules. See Appendix 1 for EPM details. bar, mbar, MPa, kPa, psi, kg/cm², atm, inH 0, mH 0, mmH 0, inHg, mmHg, UNIT1, UNIT2 (user defined)
emperature effects are included in the accuracy.
426 records into a maximum of 20 files, 32k memory additional 585 records, ( 32k memory option available ) Backlit LCD, 16 character x 2 line alpha numeric
Overload Pressure Resolution EPM
Pressure Units
Data Capacity
Display
2
2 2
Overload Warning Temperature Display
Pressure Switch Test
Operating Temperature Calibration Temperature Calibrated Temp. range Storage Temperature Temperature Effects
RS232 Parameters
Zero Reset
Humidity Resolution Select Power Supply
Voltage Output
Flashing display and audible tone at 115% of FS Temperature of pressure sensor (Accuracy: ±1 °C (1.8 °F) ) Status OPEN/CLOSED
0 to 50 °C (32 to 122 °F) 20 ±2 °C (68 ±3.6 °F) 5 to 45 °C (41 to 113 °F)
-20 to 70°C (-4 to 158 °F) T
baud rate 1200, 2400, 4800, 9600, 19200 and
38400. Stop bits 1 or 2, status ON/OFF Manual keyboard operation
5 to 95% Relative Humidity non condensing Increase or decrease by a factor of 10 6V Ni-MH rechargeable battery pack (see warning on page iii)/ mains operation via charger (supplied).
Regulated DC supply 10V ± 0.1% max current 10mA Unregulated DC supply 24V ± 10% max current 50mA
Pressure Range
Bar psi
-0.07 to 0.07 -1.0 1.0
-1 to 1 15
-1 to 2 30
-1 to 3.5 50
-1 to 7 100
-1 to 10 150
-1 to 20 300
-1 to 35 500
0 to 70 1,
0 to 100 1,500
0 to 200 3,
0 to 350 5,
0 to 700
to
-0.2 to 0.2 -3.0 to 3.0
-0.35 to 0.35 -5.0 to 5.0
-0.7 to 0.7 -10 to 10
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
-15 to
0 to 000
0 to
0 to 000
0 to 000
0 to 10,000
Accuracy (FS)
up to
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.05%
0.05%
0.05%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Model
-2-C
-10-C
-20-C
PC6-IDOS-350-C SG
PC6-IDOS-700-C SG
PC6-IDOS-0.07-C G
PC6-IDOS-0.2-C G
PC6-IDOS-0.35-C G
PC6-IDOS-0.7-C G
PC6-IDOS-1-C G
PC6-IDOS G or A
PC6-IDOS-3.5-C G
PC6-IDOS-7-C G or A
PC6-IDOS G
PC6-IDOS G or A
PC6-IDOS-35-C G
PC6-IDOS-70-C G
PC6-IDOS-100-C G
PC6-IDOS-200-C G
Resolution
1 V
0.1mV 1mV
1 Aµµ
Measurement
Voltage *
Current *
Auto Range
(Auto Range)
Accuracy
0.025% FS
0.025% FS
0.025% FS
0.025% FS
0.025% FS
Range
0 to 100mV
0 to 5V
0 to 50V 0 to 50mA 0 to 25mA
Input resistance
1 M
1 M
1 M
6
Resolution
0.2mV
1 Aµ
Sourcing Voltage * Current *
Accuracy
0.025% FS
0.025% FS
Range
0 to 10V
0 to 50mA
Max. Load
20mA
330 Ohm
Other Pressure ranges are available to special order. SG = Sealed Gauge,G = Vented Gauge, A = Absolute
° (32 to 122 °F)† 0.075% FS accuracy, 0 to 50 C
Page v
SPECIFICATIONS / MAINTENANCE
Battery Life Recharge Time Low Battery
Dimensions (l: w: h)
Weight
Microprocessor Recommended Recalibration Period
Electrical Connections
Pressure Media
Pressure Connection
Software (optional)
System requirements:Fully charged: up to 8 hours
1
14-16 hours for full charge Continuous check, audio and visual warning
92 x 110 x 59 mm (3.6 x 4.3 x 2.3 in)
850 g (30 oz)
H8S/2138
1 Year (Max)
Ground Standard 2 mm socket Voltage/Current Input/Output Standard 2 mm sockets Pressure Switch Input Standard 2 mm sockets Battery 2 pole miniature round connector RS232 8 pole miniature round connector EPM (P2) 5 pole miniature round connector
Water, Mineral Oil, Non-corrosive gases (for other media, contact GE). 3/8"BSP male ‘quick-fit’ / 1/4”BSP or 1/4”NPT male (Pneumatic). 3/8"BSP female ‘quick-fit’ / 3/8”BSP or 3/8”NPT male (Hydraulic).
The ‘ ’ software will present downloaded data in various graphical forms such as gauges or a scrolling bar & remotely control the PC6-IDOS via the RS232 cable linked to a PC. The downloaded data can be stored in a variety of file types compatible with most analysis database and word processing programs. Calibration procedures can be uploaded to the PC6-IDOS.
Calibrator head only:
Calibrator head only: Maximum
Minimum Intel Pentium with Windows 95 or
higher.
1
The battery usage time specified here is heavily dependent upon the functions selected. For instance, using the back-light function will reduce the battery usage time, as will source mode functions.
Clean the case with a moist, lint-free cloth and a weak detergent. Do not use solvents or abrasive materials.
Do not do local repairs. Return the equipment to the manufacturer or an approved service agent.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment.
For more information, contact one of these:
• our customer service department
(Contact us at www.gesensing.com)
• your local government office
SiCal PRO
® ®
Ordering Codes:
PC6 - IDOS- XXXX - C - 1
Maintenance
Range (Add ‘A’ for absolute unit)
C Ior (Calibrator or Indicator)
1 2 3 4=UK / =EUR / =USA / =AUS
Page vi
DESCRIPTION
1Page
1.0 Introduction
1.1 General
1.2 Keypad
The Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal 6V rechargeable battery pack. It is portable and capable of accurate pressure measurement, voltage or current measurements and sourcing when calibrating transducers, pressure switches, etc. This section describes the equipment from an external viewpoint, allowing the user to become familiar with the various controls and connections provided.
The PC6-IDOS is housed in a three part anodised aluminium case with a liquid crystal display unit at the front and a six button membrane keyboard. The main body contains the circuitry, connections, internal transducer and battery pack.
The Keypad is a non-tactile membrane keyboard with six buttons. The keyboard layout is shown below.
An audible tone accompanies
every key operation.
PC6-IDOS
Note:
PRINT
LOG
ON/OFF
MENU
ENTER
ZERO
2Page
1.3 Liquid Crystal Display (LCD)
1.4 Connection Panel
The LCD is a 16 character x 2 line alphanumeric LED backlit display which is capable of displaying special characters.
Pressure reading Unit of measurement
Function Function Value
(See Page 4, Fig 2)
The connection panel is located at the top of the case and provides input / output connections as detailed in Fig.2
Six standard 2 mm electrical sockets mounted along the top of the case provide input or output current and voltage and switch status.
Voltage outputs of 10 and 24V are available for use with sensors, transducers, transmitters, etc.
Pressure connections are made to either a 3/8"BSP (1/4”BSP / 1/4”NPT adaptors) male for low pressure systems (below 40 bar (600 psi)) or 3/8”BSP female (3/8”BSP/ 3/8”NPT adaptors) for systems greater than 40 bar
.
An 8 way circular miniature connector is used for communication with RS232 standard communication protocols/equipment. The user software ‘ ’ (optional) can be used to remotely monitor or download/upload data from the PC6-IDOS via this connector.
(Optional RS232 cable - Part No. 01-0625)
The 5 way circular miniature connector is used to connect to the EPM (External Pressure Module). These optional modules cover pressure ranges from 0.07 to 700 bar (1.0 to 10,000 psi).
1.4.1 2 mm sockets.
1.4.2 Quick connect coupling.
1.4.3 RS232.
SiCal PRO
1.4.4 Charger.
1.4.5 EPM.
(600 psi)
The 2 way circular miniature connector is used for the 9V charger/power supply (provided).
RS232
COUPLING
CHARGER
Special Symbols
3Page
EPM
PRINT
LOG
ON/OFF
MENU
ENTER
ZERO
Druck PC6-IDOS
Front View: LCD & keypad
Quick Connect Coupling.
54-0101 3/8”BSP 54-0364
- Above
40 bar
¼”NPT
PC6-IDOS
(600 psi)
General Purpose Adaptors
54-0158 ¼”BSP 54-0363
- 40 bar
(600 psi) or below
¼”NPT
PC6-IDOS
Top View: connection panel
(See Page 10
for V/mA
connections)
EPM
+24 V Volts output. +10 V Volts output. GND Common Ground Socket
mA Milliamp Input (controls input current in
source mode). V Volt Input (output in source mode). SWITCH Pressure Switch Input.
Display.
Keypad
See Page 2
Fig 1
Fig 2
+
-
ARRANGEMENT OF PC6-IDOS
Battery Charger Input Socket
External Pressure Module Socket
RS232 Communication Socket
See Page 11
See Page 10
See Page 11
BATTERY
CHARGER
4Page
GND
V mA SWI TCH
+10V +24V
CHARGER
RS232
IN/ OU T
OU TPU TS
EPM
PRESSURE DISPLAY -
Selection of pressure unit.
inches of mercury (at 0°C) millimetres of mercury (at 0°C) user defined unit second user defined unit Returns to menu 1
Displays the maximum pressure measured since the last reset (60 samples per second).
Displays the pressure as a percentage between user pre-set high and low alarm values. Displays the pressure at the time of tare. Volts / milliamps functions
Displays the following input measurements.
Displays actual input measurements. Displays converted input measurements. Returns to Pressure Display.
Indicates the status of the pressure switch under test (OPEN/CLOSED). Allows the PC6-IDOS to measure pressure changes over time.
Starts the leak test. Sets the leak test time.
Returns to Pressure Display. Allows the user to monitor the temperature around the selected pressure transducer in pressure display mode. Allows the user to monitor the date & time of the on board real time clock in pressure display mode. Allows the user to view or manipulate a file previously stored in the filing system.
Allows the user to see logged file.
Outputs a logged file
Imports a file (refer to ‘ instructions for exporting files to PC6-IDOS)
Deletes a single file or all logged files.
Returns to Pressure Display. Returns to Pressure Display mode.
Allows the user to operate the PC6-IDOS for pressure measurements. (Display Functions)
SiCal PRO’
(SHORTCUT: To abort any menu level, simply press the MENU key).
Menu 1
Unit
inHg mmHg UNIT1 UNIT2 Exit
Max
% Tare V / mA
Measure
Actual Converted Exit
Switch Leak
Start Set-Time
Exit Temp C Date & Time File
View
Export
Import
Delete
Exit Exit
bar millibar megapascals kilopascals pounds force per square inch kilograms force per square centimetre atmospheres i inches of water (at 4°C) metres of water (at 4°C) millimetres of water (at 4°C)
Displays the minimum pressure measured since the last reset (60 samples per second).
Displays the following outputs (status must be ON - see ‘Setup’ in Menu 2)
Displays actual output measurements. Displays converted output measurements. Displays Pressure to Volts / milliamps conversion. Returns to Pressure Display.
Returns to pressure display
bar mbar
MPa kPa
psi kg/cm²
atm nH O
mH O mmH O
Min
Source
Actual Converted P to V / P to mA Exit
Exit
2
2 2
FUNCTIONS (MENU 1)
5Page
6Page
Continued on
Page 7
}
FUNCTIONS (MENU 2)
SETTING MENU Menu 2
- To set user specific functions. (Set Functions)
High
Low
Measure
Zero FS Text Root
Source
Setup
Status Root Step Ramp
Status Rate Dwell_time Exit
Exit
Exit
Allow the user to set a High or Low Pressure Alarm Limit. (by using the increase/decrease facility above which an alarm sounds.)
Voltage / Current input settings for Converted Measurement.
Set Zero. Set Full scale value. User defined measurement text. Switch measurement root function ‘on’ or ‘off’.
Voltage / Current output settings.
Sets the parameters for voltage/current sourcing.
Switches output on or off. Switch sourcing root function ‘on’ or ‘off’. Set step value (%FS) Settings for Ramp mode
Switches ramp mode on or off. Sets the ramp rate (V or mA/Sec). Sets the time for which the output remains at the same level. Returns to Source sub-menu.
Returns to Source sub-menu.
Returns to Pressure Display.
V / mA
Exit
Range
Zero FS Exit
Convert
Zero FS Text Exit
P to V / mA
Zero FS Exit
Exit
Reverts back to Pressure Display.
Sets the following values:-
Sets Zero value Sets Full Scale value Returns to Source sub-menu.
Sets the following converted values:-
Sets converted Zero values. Sets converted Full Scale values. User defined converted text. Returns to Source sub-menu.
Sets Pressure to V / mA conversion factors.
Set zero values. Set Full Scale values. Returns to Source sub-menu.
FUNCTIONS (MENU 2) Continued.
Page 7
Switches the Power Supply outputs +10V & +24V ON or OFF
(separate pc software utility needed -contact supplier).
Vout
Switches ON or OFF the LED back light. Selects low, normal or high resolution. Sets up the event status for RS232 output and for logging functions.
Sets key driven output or logging. Sets timed output defined by the user. Reverts back to Pressure Display.
Allows the user to setup, amend and view files for logging.
To select a log file and start logging. To look at a file. Deletes a file. To enable or disable logging. To export logged files. To import files from a PC. (see ‘ ’ software instructions for exporting files to the PC6-IDOS) Reverts back to Pressure Display.
Settings for RS232 communications.
To enable RS232 communication. To select the baud rate. To select the number of stop bits. Reverts back to Pressure Display.
Sets up Date & Time parameters
Sets the date & time. To look at date & time.
Reverts back to Pressure Display. Allows user to change language Will reset pressure reading to Zero. ( This will remove Zero offset permanently / NOT available for absolute units ). Selects current pressure module (P1, P2) Allows the user to set Power_Off time or to switch the function on or off.
Switches the auto-power function off.
Allows the user to set the auto-power off time in minutes.
Reverts back to Pressure Display. Returns to Pressure Display
Display Light Resolution Event
Key Timed Exit
Log
Log View Delete Status Export Import Exit
RS232
Status Baud-rate Stop-bits Exit
Date/Time
Set View
Exit Language Zero
Port Power_Off
Key
Timed
Exit Exit
SiCal PRO
FUNCTIONS (MENU 3)
8Page
CAL MENU
Menu 3
Span
Continue Exit
History
Overload Zero-Span
Exit Language Units
View
Alter
Exit Exit
Allows user to set span.
Starts span calibration procedure. Returns to Cal Menu.
Allows user to see overload and zero-span history.
Displays period of time over which overload occurred. Displays date last calibrated, last span set, range and overload value.
Returns to Cal Menu.
Allows user to download different languages.
Displays the two current user-defined units. Allows user to modify the conversion factor and/or text for the two user-defined units. Returns to Cal Menu
Returns to Pressure Display.
PC6S/No Displays the unit’s serial number.
Displays the serial number of the currently selected pressure module.PM S/No
OPERATING
INSTRUCTIONS
9Page
Fig.3a
10Page
+V
Input
0V
-V
Output
GND
V mA SWITCH
+10V +24V
CHARG ER RS232
IN /O UT
OU TPU TS
GND
V m A SW ITCH
Two Wire
Transmitter
4 to 20 mA
Three Wire
Transmitter
0 to 5V
+10V +24V
IN /O UT
OU TPU TS
EPM
Fig.3
EPM
1.5 Introduction
The PC6-IDOS enables the user to carry out checks on pressure transducers, transmitters, pressure switches, pressure gauges and process indicators quickly and accurately. It is intended for use as a portable, on-site calibrator, used in close proximity with the equipment under test and without the need to remove the unit from its parent system. This section describes the connections to be made when checking switches, transmitters, etc. It also lists menu options, selection procedures and purpose of options available.
1.6 Connections
1.6.1 Pressure Input The PC6-IDOS measures pressure
directly via the connector found underneath the case (i.e. the connection between the PC6-IDOS and a hand held pressure test system). Alternatively, connect the item under test to the PC6-IDOS with the adaptor and seals provided.
1.6.2 EPM (External Pressure Module)
Different pressure ranges can be read by attaching an External Pressure Module to the 5-way socket.
(See Fig.3 opposite)
Warning: The PC6-IDOS must be switched off prior to
connecting or disconnecting the EPM .
1.6.3 Voltage / Current Input / Output Voltage and Current
readings can be taken by connecting the outputs from the item under test to the 'V' or 'mA' and 'GND' input sockets on the connection panel using the connection leads provided.
Warning: When measuring voltage or current, make sure 'GND' is connected (where applicable, see fig. 3a)
1.6.4 Switch Test Activate ‘Switch’ function in menu 1. The switching pressure of a pressure switch can now be monitored by connecting the contacts of the pressure switch between the INPUT 'GND' and 'SWITCH' sockets on the connection panel and pressurizing the pressure switch.
1.6.5 RS232
If the RS232 facility is to be used then connections to the 8­way socket must be made to the specifications given in the table below. Alternatively an RS232 communication cable can be purchased from GE suitable for most RS232 applications as shown below.
To RS232 Standard 9-way socket on PC6- 'D'type female IDOS connector
Pin 6 and pin 8 should not be used.
RS232 Connections Table
Note:
1.6.5.1 RS232 Output
Warning:
1.6.6 Battery Charger
CAUTION:
1.6.7 Voltage Output
Connect the PC6-IDOS to a printer or computer using a suitable cable (not supplied) as detailed above and set the parameters as detailed in Para. 2.5 & 2.7. The output can now be obtained as ‘single shot' or ‘timed' as follows:
- This output is available in Display mode when RS232 status is ON & Log status is OFF. The output cycle is initiated by pressing the key. Pressing the key cancels the output. Once initiated, the output is repeated at the selected time interval (see ‘Event' function, page 26 ) until the key is pressed.
The internal battery pack may be recharged (when indicated by the battery low message) by plugging the charger (supplied) into the 'CHARGER' socket.
When recharging internal batteries, only the charging
unit supplied with the PC6-IDOS should be used.
A voltage of +10V or +24V may be sourced from the PC6-IDOS by connecting to the 2 mm sockets. A typical application for this voltage output would be to power a pressure transmitter. ( see Fig. 3. page 10 ) To switch on voltage output function:­a) Press the key twice to activate 'Set' menu. b) Select 'V-out' and press the key. c) Select 'ON' and press the key.
1.6.5.2 Single Shot - This output is always available in Display mode when RS232 status & Log status are OFF. The output is obtained each time the key is pressed (provided any previous output is complete).
The PC6-IDOS must be switched off prior to
connecting or disconnecting the RS232 cable.
1.6.5.3 Timed
SIGNAL
Data Terminal Ready (DTR)
Transmit Data (TxD) Receive Data (RxD)
Request To Send (RTS)
Clear To Send (CTS)
Do not use
Ground
Do not use
STATE
Held High
Active High
INPUT/OUTPUT
Output Output
Input
Output
Input
Output
Common
Input
PIN No.
1 2 3 4 5 6 7 8
PRINT
LOG
PRINT
LOG
MENU
MENU
MENU
Cable Part No: 01-0625
11Page
ENTER
ZERO
ENTER
ZERO
1.7 Messages & Parameters
The PC6-IDOS registers pressure input at switch on after the initial start­up message. The pressure reading is displayed on the top line of the LCD (see Page 3). Pressure values are monitored, alarm values are compared, alarm warnings and other information messages displayed until the unit is switched off.
Any read / write errors to memory detected at switch on will be reported as: Memory Error ‘Error No.’
When switched on, the unit registers the same state as when last switched off. The display mode (Operating/Set), units of measurement, function and parameter settings, alarm limits and RS232 output control, Port (P1 or P2) selection, logged parameters are all retained at switch off, with the exception of the following conditions:
A battery low state occurring prior to the switch off. When switched on, the unit registers the latest retained set of parameters.
Any TARE value set during use is not retained and
therefore should be set, if required, each time the unit is switched on.
Maximum and Minimum values are not retained. Current values reflect pressure monitored from the start of the pressure input.
If the unit is switched off during a LEAK test then the test will be void. When switched on again the unit will start up in LEAK test mode.
When the key is pressed the 'switch off' message is displayed for two seconds accompanied by an audible bleep. Normal operation may be resumed by pressing any key during switch off.
The PC6-IDOS will automatically turn itself of
after a set period (See Page 31 for instructions)
(Bleep rate 4 Hz) ‘OVERLOAD' occurs when
input pressure exceeds overload pressure level.
(Bleep rate 1 Hz). ‘Low Battery' occurs when internal battery pack drops below 5.5V. Message flashes for 4 seconds, after a further 30 seconds, low battery detection is re-enabled.
(Bleep rate 1 Hz)
If the receiving device is not ready for input (e.g. device not connected or handshake LOW), 10 seconds after starting an output cycle (single shot or timed output) the message 'Printer Busy' is flashed on the display, together with an audible warning. If the device does not accept input within the next 5 seconds then the output cycle is cancelled and output is stopped until the key is pressed again.
(Bleep rate 2 Hz)
Occurs when measured pressure is lower than the user set Low alarm value or greater than the user set High alarm value and the
message is flashed. Indication ceases when pressure is equal to
either alarm value or between the values.
(Bleep rate 2 Hz)
Occurs in pressure mode. If the key is pressed once (less than 1 second) the pressure value (upper line) is frozen and the lower line is flashed with the ‘Display Hold’ message at 1 Hz. Normal display update resumes when any top row key is pressed.
Memory Errors:
Retained parameters
Battery low state
Tare value
Maximum/Minimum
Void leak test
Error Numbers:-
0 8 9- Onboard eeprom; - RTC; - 2nd Onboard eeprom. Errors with Pressure Modules will be indicated by the message: "No P1 (or P2) Module" preceeded by "PM EEPROM ERROR" at power on.
1.7.1
1.7.2
1.7.3
1.7.4
1.7.5
1.7.6 Switch off
1.7.7 Auto Switch off
1.7.8 Pressure Overload
1.7.9 Low Battery Detect
1.7.11 Printer Busy
1.7.12 Outside User Set Alarm Values
1.7.13 Display Hold
1.7.10 Pressure Switch State Change (Bleep rate 2 Hz)
Occurs in pressure mode with the pressure switch state display being selected. When the connected pressure switch changes state, the pressure value (upper line) is frozen and the pressure switch state (lower line) display is flashed at 1 Hz unless a log file is open. Normal display update resumes when the key is pressed.
’outside
limits’
ON/OFF
12Page
ENTER
ZERO
ENTER
ZERO
PRINT
LOG
Note: n pressing the key.
The display contents can be printed out duri g display hold by
1.7.14 Connection Fault:
1.7.15 Short Detected:
1.7.16 Voltage / Current Overscale:
1.7.17 RS232 Output Format
1.7.18 Alarm Flags
In voltage / current source mode, if the PC6-IDOS is unable to control the voltage/current output (e.g. because there is no loop back in the current mode), the message “Connection Fault” will flash on the lower line accompanied by a bleep at 1 Hz. The message will continue until the user presses the or key ( after fault is rectified ) or the user exits from source mode.
If the 24V output is ON and it is shorted to 0V, the message “*Short detected*” will be displayed on the lower line accompanied by a bleep at 1 Hz. This message will be cleared after approximately 5 seconds. In this condition the 24V supply will be switched off for safety.
In Voltage measure mode, if a voltage in excess of 55V is connected to the PC6-IDOS the message “*OVERSCALE*” will be displayed on the lower line, until the voltage falls within the input limits. In Current measure mode, if a current in excess of 55mA enters the mA terminal, the PC6-IDOS will display “*OVERSCALE*” on the upper line and the lower line will scroll the message “Check connections: Press MENU to continue”. Pressing the key will remove the message provided the current measured is within range. This message may also appear in the source mode under connection fault conditions. Source mode will change to measure mode for safety. If the current overscale message occurs when the PC6-IDOS is in timed RS232 output mode, the unit will exit from the timed output mode. If there is an excess voltage at the mA input when the PC6-IDOS is placed into current measurement mode, the message “Connection Fault” will be scrolled on the lower line. Once the fault has been rectified, press the key to return to measuring mode.
The output format is as follows: LF Time & date ( 1100) +LF display top line (16 characters ) + LF display lower line (16 characters ) + LF alarm flag (1 character ) + LF
A single character is output to indicate the various
pressure alarm conditions as given below:
* overload pressure level exceeded
(pressure warnings, 1.7.8 & 1.7.12) > pressure exceeds user set high limit < pressure below user set low limit “ ” a space if no alarms apply. c connection fault s short detected
23:13:30 08
Note: If two (or more) conditions occur which require a message display, the highest priority condition in the above list will be displayed. The sequence of messages given above set the priority except 'Switch Off' which comes last.
If more than one symbol is applicable, they are displayed one after the other.
1.7.19 Display Symbols
Top line symbols
Bottom line symbols
13Page
or
or
Log Status ON
Pressure Port 1 selected
Pressure Port 2 (EPM) selected
Rs232 Output Status ON
Timed Log / output
Keyed Log / output
Source Step Mode
Measure Mode & P-V / P-mA
Root Function
Source Mode Ramp
Nudge Mode (UP)
Nudge Mode (DOWN)
Log Status ON File Open
MENU
MENU
PRINT
LOG