Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Important Precautions
THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.
• IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER
THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE
TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE
MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE
PROVIDER, OPERATOR OR PATIENT FROM ELECTRIC SHOCK, MECHANICAL
OR OTHER HAZARDS.
CE MANUEL DE MAINTENANCE N’EST DISPONIBLE QU’EN ANGLAIS.
• SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE
LANGUE QUE L’ANGLAIS, C’EST AU CLIENT QU’IL INCOMBE DE LE FAIRE
TRADUIRE.
• NE PAS TENTER D’INTERVENTION SUR LES éQUIPEMENTS TANT QUE LE
MANUEL SERVICE N’A PAS éTé CONSULTé ET COMPRIS.
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAîNER CHEZ LE
TECHNICIEN, L’OPéRATEUR OU LE PATIENT DES BLESSURES DUES à DES
DANGERS éLECTRIQUES, MéCANIQUES OU AUTRES.
DIESES KUNDENDIENST-HANDBUCH EXISTIERT NUR IN ENGLISCHER
SPRACHE.
• FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENö TIGT,
IST ES AUFGABE DES KUNDEN FüR EINE ENTSPRECHENDE ÜBERSETZUNG
ZU SORGEN.
• VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES
KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN
WURDE.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN
DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN
DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE
GEFAHREN KOMMEN.
- i
Page 4
GE
AVISO
(ES)
ATENÇÃO
(PT-Br)
(PT-pt)
AVISO
AVVERTENZA
(IT)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLéS.
• SI ALGúN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA
QUE NO SEA EL INGLéS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN
SERVICIO DE TRADUCCIó N.
• NO SE DEBERá DAR SERVICIO TéCNICO AL EQUIPO, SIN HABER
CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL
PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN
LESIONES PROVOCADAS POR CAUSAS ELéCTRICAS, MECá NICAS O DE
OTRA NATURALEZA.
ESTE MANUAL DE ASSISTêNCIA TéCNICA Só SE ENCONTRA DISPONíVEL EM
INGLêS.
• SE QUALQUER OUTRO SERVIç O DE ASSISTê NCIA Té CNICA, QUE NãO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, é DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIç OS DE TRADUçã O.
• Nã O TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E
COMPREENDIDO ESTE MANUAL DE ASSISTêNCIA TéCNICA.
• O Nã O CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANçA
DO TéCNICO, OPERADOR OU PACIENTE DEVIDO A‘ CHOQUES ELé TRICOS,
MECâ NICOS OU OUTROS.
ESTE MANUAL DE ASSISTÊNCIA ESTÁ DISPONÍVEL APENAS EM INGLÊS.
• SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A
GEHC, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA
RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE EFECTUAR REPARAÇÕES NO EQUIPAMENTO SEM TER
CONSULTADO E COMPREENDIDO PREVIAMENTE ESTE MANUAL.
• A INOBSERVÂNCIA DESTE AVISO PODE RESULTAR EM FERIMENTOS NO
TÉCNICO DE ASSISTÊNCIA, OPERADOR OU PACIENTE EM CONSEQUÊNCIA
DE CHOQUE ELÉCTRICO, PERIGOS DE ORIGEM MECÂNICA, BEM COMO DE
OUTROS TIPOS.
IL PRESENTE MANUALE DI MANUTENZIONE è DISPONIBILE SOLTANTO IN
INGLESE.
• SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEHC RICHIEDE IL
MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE è TENUTO A PROVVEDERE
DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL’APPARECCHIATURA SOLO DOPO
AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL
CONTENUTO.
• NON TENERE CONTO DELLA PRESENTE AVVERTENZA POTREBBE FAR
COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL’ADDETTO ALLA
MANUTENZIONE, ALL’UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.
ii -
Page 5
GE
HOIATUS
(ET)
VAROITUS
(FI)
ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ
(EL)
FIGYELMEZTETÉS
(HU)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.
• KUI KLIENDITEENINDUSE OSUTAJA Nõ UAB JUHENDIT INGLISE KEELEST
ERINEVAS KEELES, VASTUTAB KLIENT Tõ LKETEENUSE OSUTAMISE EEST.
• ä RGE ü RITAGE SEADMEID TEENINDADA ENNE EELNEVALT KäESOLEVA
TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• Kä ESOLEVA HOIATUSE EIRAMINE Võ IB PõHJUSTADA TEENUSEOSUTAJA,
OPERAATORI Võ I PATSIENDI VIGASTAMIST ELEKTRILöö GI, MEHAANILISE
Võ I MUU OHU TAGAJäRJEL.
TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.
• JOS ASIAKKAAN PALVELUNTARJOAJA VAATII MUUTA KUIN
ENGLANNINKIELISTä MATERIAALIA, TARVITTAVAN Kää NNö KSEN
HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ä Lä YRITä KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT
JA YMMä RTä NYT TäMä N HUOLTO-OHJEEN.
• MIKä LI Tä Tä VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA
PALVELUNTARJOAJAN, LAITTEISTON Kä YTTä Jä N TAI POTILAAN
VAHINGOITTUMINEN Sä HKöISKUN, MEKAANISEN VIAN TAI MUUN
VAARATILANTEEN VUOKSI.
ΤΟΠΑΡΟΝΕΓΧΕΙΡΙ∆ΙΟΣΕΡΒΙΣ∆ΙΑΤΙΘΕΤΑΙΣΤΑΑΓΓΛΙΚΑΜΟΝΟ.
• ΕΑΝΤΟΑΤΟΜΟΠΑΡΟΧΗΣΣΕΡΒΙΣΕΝΟΣΠΕΛΑΤΗΑΠΑΙΤΕΙΤΟΠΑΡΟΝ
ΕΓΧΕΙΡΙ∆ΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ
ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ
ΕΚΤΟΣ ΕΑΝ
ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ
ΕΓΧΕΙΡΙ∆ΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ∆Ε ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝ∆ΕΧΕΤΑΙ ΝΑ
ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ.
EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL.
• HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT,
AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A
KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ,
MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB
VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT EREDMÉNYEZHETI.
- iii
Page 6
GE
VIÐVÖRUN
(IS)
VÝSTRAHA
(CS)
ADVARSEL
(DA)
WAARSCHUWING
(NL)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.
• EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN
ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU.
• REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG
SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
• EF EKKI ER FARIÐ AÐ ÞESSARI VIÐVÖRUN GETUR ÞAÐ VALDIÐ MEIÐSLUM
ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS VEGNA RAFLOSTS,
VÉLRÆNNAR EÐA ANNARRAR HÆTTU.
TENTO SERVISNÍ NÁVOD EXISTUJE POUZE V ANGLICKé M JAZYCE.
•VPř íPADě, ŽE POSKYTOVATEL SLUŽEB ZÁKAZNÍKŮM POTř EBUJE Ná VOD
V JINéM JAZYCE, JE ZAJIšTěNí Př EKLADU DO ODPOVíDAJíCíHO JAZYKA
úKOLEM Zá KAZNíKA.
• NEPROVÁDĚJTE úDRŽBU TOHOTO ZAř íZENí, ANIŽ BYSTE SI Př Eč ETLI
TENTO SERVISNÍ NÁVOD A POCHOPILI JEHO OBSAH.
•VPř íPADě NEDODRŽOVá Ní TéTO VýSTRAHY MůŽE DOJíT ÚRAZU
ELEKTRICKÁM PROUDEM PRACOVNíKA POSKYTOVATELE SLUŽEB,
OBSLUŽNéHO PERSONá LU NEBO PACIENTů VLIVEM ELEKTRICKéHOP
PROUDU, RESPEKTIVE VLIVEM K RIZIKU MECHANICKÉHO POŠKOZENÍ NEBO
JINÉMU RIZIKU.
DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.
• HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END
ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE
DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFŘRE SKADE
PĹ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR
TEKNIKEREN, OPERATŘREN ELLER PATIENTEN.
DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.
• ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE
KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE
ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET
ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND
KUNNEN RAKEN ALS GEVOLG VAN EEN ELEKTRISCHE SCHOK,
MECHANISCHE OF ANDERE GEVAREN.
iv -
Page 7
GE
BRĪDINĀJUMS
(LV)
ĮSPĖJIMAS
(LT)
ADVARSEL
(NO)
OSTRZEŻENIE
(PL)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
šĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.
• JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ
VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS
IZLASĪŠANAS UN SAPRAŠANAS.
•ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS
TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES
SNIEDZĒJAM, OPERATORAM VAI PACIENTAM.
ŠIS EKSPLOATAVIMO VADOVAS YRA IŠLEISTAS TIK ANGLŲ KALBA.
• JEI KLIENTO PASLAUGŲ TEIKĖJUI REIKIA VADOVO KITA KALBA – NE ANGLŲ,
VERTIMU PASIRŪPINTI TURI KLIENTAS.
•NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS DARBŲ, NEBENT
VADOVAUTUMĖTĖS ŠIUO EKSPLOATAVIMO VADOVU IR JĮ SUPRASTUMĖTE
• NEPAISANT ŠIO PERSPĖJIMO, PASLAUGŲ TEIKĖJAS, OPERATORIUS AR
PACIENTAS GALI BŪTI SUŽEISTAS DĖL ELEKTROS SM
ŪGIO, MECHANINIŲ AR
KITŲ PAVOJŲ.
DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.
• HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET
KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE
SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT
SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ
GRUNN AV ELEKTRISK STØT, MEKANISKE ELLER ANDRE FARER.
NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU
ANGIELSKIM.
•JEśLI FIRMA śWIADCZĄCA KLIENTOWI USłUGI SERWISOWE WYMAGA
UDOSTęPNIENIA PODRęCZNIKA W JęZYKU INNYM NIŻ ANGIELSKI,
OBOWIĄZEK ZAPEWNIENIA STOSOWNEGO TłUMACZENIA SPOCZYWA NA
KLIENCIE.
• NIE PRó BOWAć SERWISOWAć NINIEJSZEGO SPRZęTU BEZ UPRZEDNIEGO
ZAPOZNANIA SIę Z PODRęCZNIKIEM SERWISOWYM.
• NIEZASTOSOWANIE SIę DO TEGO OSTRZEŻENIA MOżE GROZIć
OBRAŻENIAMI CIAłA SERWISANTA, OPERATORA LUB PACJENTA W WYNIKU
PORAŻENIA PRĄDEM, URAZU MECHANICZNEGO LUB INNEGO RODZAJU
ZAGROŻEń.
- v
Page 8
GE
ATE NŢIE
(RO)
ОСТОРОЖНО!
(RU)
(BG)
ПРЕДУПРЕЖДЕНИЕ
UPOZORENJE
(SR)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.
• DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ
DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O
TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR
CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA
DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA
PERICOLELOR DE ELECTROCUTARE, MECANICE SAU DE ALTĂ NATURĂ.
Данное рук о во дс т во по обслуживанию ПРЕДОСТАВЛЯЕТСЯ только
на английском Яз ы ке.
• Ес ли сервисноМУ ПЕРСОНАЛУ клиента необходимо рук о во дс тво
не на английском ЯЗЫКЕ, клиенту следует самостоЯтельно
ОБЕСПЕЧИТЬ перевод.
ЧТО СПЕЦИАЛИСТ ПО ТЕХОБСЛУЖИВАНИЮ, ОПЕРАТОР ИЛИ ПАЦИЕНТ
ПОЛУЧАТ УДАР ЗЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ
ДРУГОЕ ПОВРЕЖДЕНИЕ.
ТОВА СЕРВИЗНО РЪКОВОДСТВО Е НАЛИЧНО САМО НА АНГЛИЙСКИ ЕЗИК.
• АКО ДОСТАВЧИКЪТ НА СЕРВИЗНИ УСЛУГИ НА КЛИЕНТ
СЕ НУЖДАЕ ОТ
ЕЗИК, РАЗЛИЧЕН ОТ АНГЛИЙСКИ, ЗАДЪЛЖЕНИЕ НА КЛИЕНТА Е ДА
ПРЕДОСТАВИ ПРЕВОДАЧЕСКА УСЛУГА.
• НЕ СЕ ОПИТВАЙТЕ ДА ИЗВЪРШВАТЕ СЕРВИЗНО ОБСЛУЖВАНЕ НА ТОВА
ОБОРУДВАНЕ, ОСВЕН ВСЛУЧАЙ, ЧЕ СЕРВИЗНОТО РЪКОВОДСТВО Е
ПРОЧЕТЕНО И СЕ РАЗБИРА.
• НЕСПАЗВАНЕТО НА ТОВА ПРЕДУПРЕЖДЕНИЕ МОЖЕ ДА ДОВЕДЕ ДО
НАРАНЯВАНЕ НА ДОСТАВЧИКА НА СЕРВИЗНИ УСЛУГИ, НА
ОПЕРАТОРА
ИЛИ ПАЦИЕНТА ВСЛЕДСТВИЕНА ТОКОВ УДАР, МЕХАНИЧНИ ИЛИ ДРУГИ
РИСКОВЕ.
OVAJ PRIRUČNIK ZA SERVISIRANJE DOSTUPAN JE SAMO NA ENGLESKOM
JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST JE NA KLIJENTU DA PRUŽI USLUGE PREVOĐENJA.
• NEMOJTE POKUŠAVATI DA SERVISIRATE OPREMU AKO NISTE PROČITALI I
RAZUMELI PRIRUČNIK ZA SERVISIRANJE.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO POVREĐIVANJA
SERVISERA, OPERATERA ILI PACIJENTA UZROKOVANOG ELEKTRIČNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
vi -
Page 9
GE
OPOZORILO
(SL)
UPOZORENJE
(HR)
UPOZORNENIE
(SK)
VARNING
(SV)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI.
• ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V
DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA.
• NE POSKUŠAJTE SERVISIRATI OPREME, NE DA BI PREJ PREBRALI IN
RAZUMELI SERVISNI PRIROČNIK.
• ČE TEGA OPOZORILA NE UPOŠTEVATE, OBSTAJA NEVARNOST
ELEKTRIČNEGA UDARA, MEHANSKIH ALI DRUGIH NEVARNOSTI IN
POSLEDIČNIH POŠKODB PONUDNIKA SERVISNIH STORITEV, UPORABNIKA
OPREME ALI PACIENTA.
OVAJ SERVISNI PRIRUČNIK DOSTUPAN JE SAMO NA ENGLESKOM JEZIKU.
• AKO KLIJENTOV SERVISER ZAHTIJEVA JEZIK KOJI NIJE ENGLESKI,
ODGOVORNOST KLIJENTA JE PRUŽITI USLUGE PREVOĐENJA.
• NEMOJTE POKUŠAVATI SERVISIRATI OPREMU AKO NISTE PROČITALI I
RAZUMJELI SERVISNI PRIRUČNIK.
• AKO NE POŠTUJETE OVO UPOZORENJE, MOŽE DOĆI DO OZLJEDE
SERVISERA, OPERATERA ILI PACIJENTA PROUZROČENE STRUJNIM
UDAROM, MEHANIČKIM I DRUGIM OPASNOSTIMA.
TÁTO SERVISNÁ PRÍRUČKA JE K DISPOZíCII LEN V ANGLIČTINE.
• AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO
ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE
ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA VYKONÁVAŤ SERVIS ZARIADENIA SKÔR, AKO SI
NEPREČÍTATE SERVISNÚ PRÍRUČKU A NEPOROZUMIETE JEJ.
• ZANEDBANIE TOHTO UPOZORNENIA Mô ŽE VYÚSTIŤ DO ZRANENIA
POSKYTOVATEĽA SLUŽIEB, OBSLUHUJúCEJ OSOBY ALEBO PACIENTA
ELEKTRICKýM PRúDOM, PRÍPADNE DO MECHANICKéHO ALEBO INéHO
NEBEZPEč ENSTVA.
DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.
• OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN
ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA
ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST
OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I
SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL
FÖLJD AV ELEKTRISKA STÖTAR, MEKANISKA FAROR ELLER ANDRA FAROR.
- vii
Page 10
GE
DİKKAT
(TR)
(JA)
Traditional
Chinese
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR.
• BU UYARININ GÖZ ARDI EDİLMESİ, ELEKTRİK ÇARPMASI YA DA MEKANİK
VEYA DİĞER TÜRDEN KAZALAR SONUCUNDA TEKNİSYENİN, OPERATÖRÜN
YA DA HASTANIN YARALANMASINA YOL AÇABİLİR.
viii -
Page 11
GE
(ZH-CN)
(KO)
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
- ix
Page 12
GE
DANGER
WARNINGWARNING
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In
Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by
a GE representative or hospital receiving agent. Whether noted or concealed, damage MUST be
reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the
contents and containers held for inspection by the carrier. A transportation company will not pay a claim
for damage if an inspection is not requested within this 14 day period.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT - FOR USA ONLY
All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the
equipment shall be performed by licensed electrical contractors. Other connections between pieces of
electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel.
In performing all electrical work on these products, GE will use its own specially trained field engineers.
All of GE’s electrical work on these products will comply with the requirements of the applicable
electrical codes.
The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers,
personnel of third-party service companies with equivalent training, or licensed electricians) to perform
electrical servicing on the equipment.
OMISSIONS & ERRORS
If there are any omissions, errors or suggestions for improving this documentation, please contact the
GE Healthcare Global Documentation Group with specific information listing the system type, manual
title, part number, revision number, page number and suggestion details.
Mail the information to:
Service Documentation,
GE Medical Systems (China) Co., Ltd.
No.19 Changjiang Road
WuXi National Hi-Tech Development Zone
Jiangsu, P.R China 214028
TEL: +86 510 85225888; FAX: +86 510 85226688
GE Healthcare employees should use TrackWise to report service documentation issues. These issues
will then be in the internal problem reporting tool and communicated to the writer.
SERVICE SAFETY CONSIDERATIONS
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN
THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND
ADJUSTING.
Use all Personal Protection Equipment (PPE) such as gloves, safety shoes, safety
glasses, and kneeling pad, to reduce the risk of injury.
For a complete review of all safety requirements, see the Chapter 1, Safety Considerations section in
the Service Manual.
x -
Page 13
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
LEGAL NOTES
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without
prior written permission of GE Healthcare.
GE Healthcare may revise this publication from time to time without written notice.
TRADEMARKS
All products and their name brands are trademarks of their respective holders.
•Contractors (Some parts of Chapter 2 - Site Preparations)
1-2 -
Page 27
GE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-1-5 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Models Covered by this
Manual
Table 1-2 LOGIQ e R4.x.x Model Designations
Part NumberDescription
5151219LOGIQ e R4.x.x Console for USA
5151220LOGIQ e R4.x.x Console for Argentina
5151243 LOGIQ e R4.x.x Console for Europe
5151247 LOGIQ e R4.x.x Console for China
5151252LOGIQ e R4.x.x Console for Japanese
5151253 LOGIQ e R4.x.x Console for Australia
5151251LOGIQ e R4.x.x Console for India
Table 1-3 LOGIQ e R5.0.x / LOGIQ e Vet Model Designations
Part NumberDescription
5199260LOGIQ e / LOGIQ e Vet R5.0.x Console for USA
5198329LOGIQ e / LOGIQ e Vet R5.0.x Console for Argentina
5199670LOGIQ e / LOGIQ e Vet R5.0.x Console for Europe
5198793LOGIQ e / LOGIQ e Vet R5.0.x Console for China
5198161LOGIQ e / LOGIQ e Vet R5.0.x Console for Japanese
5199693LOGIQ e / LOGIQ e Vet R5.0.x Console for Australia
5212841LOGIQ e / LOGIQ e Vet R5.0.x Console for India
Table 1-4 LOGIQ e R5.2.x Model Designations
Part NumberDescription
5323370LOGIQ e R5.2.x Console for USA
5323375LOGIQ e R5.2.xConsole for Argentina
5323371LOGIQ e R5.2.xConsole for Europe
5323380LOGIQ e R5.2.xConsole for Japanese
5323378LOGIQ e R5.2.xConsole for Australia
5323377LOGIQ e R5.2.x Console for India
5198793-3LOGIQ e R5.2.x Console for China
5357467LOGIQ e R5.2.x Console for India
5357230LOGIQ e R5.2.x Console
5389811LOGIQ e R5.2.x Console for China
Table 1-5 LOGIQ e R6.x.x Model Designations
Part NumberDescription
5389028LOGIQ e R6.x.x Console for USA
5389808LOGIQ e R6.x.x Console for Argentina
5389810LOGIQ e R6.x.x Console for Europe
5389811LOGIQ e R6.x.x Console for Chinese
Chapter 1 - Introduction1-3
Page 28
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-5 LOGIQ e R6.x.x Model Designations
Part NumberDescription
5389812LOGIQ e R6.x.x Console for Japanese
5389813LOGIQ e R6.x.x Console for Australia
5389814LOGIQ e R6.x.x Console for India
5410642LOGIQ e R6.x.x Console for CKD USA
5410642LOGIQ e R6.x.x Console for CKD Taiwan
5427106LOGIQ e R6.x.x Console for Korean
5405339LOGIQ e R6.x.x Console for EMEA
Table 1-6 LOGIQ e R7.x.x Model Designations
Part NumberDescription
5432745LOGIQ e R7.x.x Console for USA
5432746LOGIQ e R7.x.x Console for Europe
5432747LOGIQ e R7.x.x Console for Japan
5432748LOGIQ e R7.x.x Console for Argentina
5432750LOGIQ e R7.x.x Console for Australia
5432753LOGIQ e R7.x.x Console for India
5432754LOGIQ e R7.x.x Console for Korea
5439035LOGIQ e R7.x.x Console for CKD USA
5447891LOGIQ e R7.x.x Console for China
Table 1-7 LOGIQ i R4.x.x Model Designations
Part NumberDescription
5176830LOGIQ i R4.1.x Console for USA
5179969LOGIQ i R4.1.x Console for Argentina
5179685LOGIQ i R4.1.x Console for Europe
5179209LOGIQ i R4.1.x Console for China
5179748LOGIQ i R4.1.x Console for Japanese
5179191LOGIQ i R4.1.x Console for Australia
5179176LOGIQ i R4.1.x Console for India
Table 1-8 LOGIQ i R5.x.x Model Designations
Part NumberDescription
5245475LOGIQ i R5.x.x Console for USA
5245476LOGIQ i R5.x.x Console for Argentina
5245477LOGIQ i R5.x.x Console for Europe
5245478LOGIQ i R5.x.x Console for China
5245480LOGIQ i R5.x.x Console for Australia
5245481LOGIQ i R5.x.x Console for India
1-4 -
Page 29
GE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-9 Vivid e R4.x.x Model Designations
Part NumberDescription
5171613Vivid e R4.x.x Console for USA
5172528Vivid e R4.x.x Console for Argentina
5171633 Vivid e R4.x.x Console for Europe
5171603 Vivid e R4.x.xConsole for China
5171755Vivid e R4.x.x Console for Japanese
5172272Vivid e R4.x.x Console for Australia
5172286Vivid e R4.x.x Console for India/South Africa
5183318Vivid e R4.x.x Console with SKD ECG
Table 1-10 Vivid e R5.x.x Model Designations
Part NumberDescription
5198601Vivid e R5.x.x Console for USA
5212203Vivid e R5.x.x Console for Argentina
5212894 Vivid e R5.x.x Console for Asia
5199815Vivid e R5.x.x Console for China
5213329Vivid e R5.x.x Console for Japan
5212851 Vivid e R5.x.x Console for Australia
5199644Vivid e R5.x.x Console for India/South Africa
5183318Vivid e R5.x.x Console with SKD ECG for Europe
5435024Vivid e R5.x.x Console for USA
5435025Vivid e R5.x.x Console for Europe
5435026Vivid e R5.x.x Console for Argentina
5435027Vivid e R5.x.x Console for India
5435028Vivid e R5.x.x Console for Australia
5435030Vivid e R5.x.x Console for Japan
5435031Vivid e R5.x.x Console for China
5435032Vivid e R5.x.x Console for SKD ECG
Table 1-11 Vivid e R6.x.x Model Designations
Part NumberDescription
5411312Vivid e R6.x.x Console for EU
5411381Vivid e R6.x.x Console for USA
5411382 Vivid e R6.x.x Console for China
5411383Vivid e R6.x.x Console for Argentina
5411384 Vivid e R6.x.x Console for Japan
5411385Vivid e R6.x.x Console for Australia
5411386Vivid e R6.x.x Console for India
5411388Vivid e R6.x.x Console for SKD ECG
5410641Vivid e R6.x.x Console for CKD Taiwan
Chapter 1 - Introduction1-5
Page 30
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-11 Vivid e R6.x.x Model Designations
Part NumberDescription
5435033Vivid e R6.x.x Console for USA
5435034Vivid e R6.x.x Console for EU
5435035Vivid e R6.x.x Console for Argentina
5435036Vivid e R6.x.x Console for India
5435037Vivid e R6.x.x Console for Australia
5435038Vivid e R6.x.x Console for Japan
5435040Vivid e R6.x.x Console for SKD ECG
Table 1-12 LOGIQ e Vet R6.x.x Model Designations
Part NumberDescription
5198415-2LOGIQ e Vet R6.x.x Console for EU
5199518-2LOGIQ e Vet R6.x.x Console for USA
Table 1-13 LOGIQ e Vet R7.x.x Model Designations
Part NumberDescription
5198415-3LOGIQ e Vet R7.x.x Console for EU
5199518-3LOGIQ e Vet R7.x.x Console for USA
1-1-6Purpose of Operator Manual(s)
The Operator Manual(s) should be fully read and understood before operating the LOGIQ
e/LOGIQ e Vet/LOGIQ i/Vivid eand also kept near the unit for quick reference.
1-6 -
Page 31
GE
DANGER
WARNINGWARNING
CAUTION
NOTICE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 1-2
Important Conventions
1-2-1Conventions Used in Book
Icons
Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and
conventions used on the product and in the service information are described in this chapter.
Safety Precaution Messages
Various levels of safety precaution messages may be found on the equipment and in the service
information. The different levels of concern are identified by a flag word that precedes the precautionary
message. Known or potential hazards are labeled in one of following ways:
DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL
CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE
IGNORED.
WARNING IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT CAN CAUSE
SEVERE PERSONAL INJURY AND PROPERTY DAMAGE IF INSTRUCTIONS ARE
IGNORED.
Caution is used to indicate the presence of a hazard that will or can cause minor personal injury
and property damage if instructions are ignored.
Equipment Damage Possible
Notice is used when a hazard is present that can cause property damage but has absolutely no
personal injury risk.
Example: Disk drive will crash.
NOTE:Notes provide important information about an item or a procedure.
Information contained in a NOTE can often save you time or effort.
Chapter 1 - Introduction1-7
Page 32
GE
LASER
LIGHT
Signed
Date
TAG
&
LOCKOUT
EYE
PROTECTION
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-2-2Standard Hazard Icons
Important information will always be preceded by the exclamation point contained within a triangle, as
seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be
used to make you aware of specific types of hazards that could cause harm.
Table 1-14 Standard Hazard Icons
ELECTRICALMECHANICALRADIATION
LASERHEATPINCH
Other hazard icons make you aware of specific procedures that should be followed.
Table 1-15 Standard Icons Indicating a Special Procedure Be Used
AVOID STATIC ELECTRICITYTAG AND LOCK OUTWEAR EYE PROTECTION
1-8Section 1-2 - Important Conventions
Page 33
GE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-2-3Product Icons
The following table describes the purpose and location of safety labels and other important information
provided on the equipment.
Table 1-16 Warnings
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
Identification and Rating Plate• Manufacture’s name and address
• Date of manufacture
• Model and serial numbers
• Electrical ratings (Volts, Amps, phase,
and frequency)
Bottom panel of the console
Type/Class LabelUsed to indicate the degree of safety or
protection.
IP Code (IPX1 or IPX8)
IPX1: FSU-2001
IPX8: MKF 2-MED GP26, FSU-1000
Indicates the degree of protection
provided by the enclosure per IEC60
529. IPX1 cannot be used in operating
room environment;
IPX8 can be used in operating room
environment.
Authorized European Representative
address
United States only Prescription
Requirement label
Equipment Type BF (man in the box
symbol) IEC 878-02-03 indicates B
Type equipment having a floating
applied part.
Bottom panel of the adapter.
Bottom of Footswitch
Bottom panel
Bottom panel
Probe connectors
General Warning.Various
“CAUTION - Dangerous voltage” (the
lightning flash with arrowhead in
equilateral triangle) is used to indicate
electric shock hazards.
Various
Chapter 1 - Introduction1-9
Page 34
GE
REFREF
SNSN
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-16 Warnings
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
“Protective Earth” indicates the
protective earth (grounding) terminal.
“ON” indicates the power on position of
the power switch.
CAUTION
This Power Switch DOES NOT
ISOLATE Mains Supply
“TUV” Listing and Certification Mark is
used to designate conformance to
nationally recognized product safety
standards. The Mark bears the name
and /or logo of the testing laboratory,
product category, safety standard is
assessed and a control number.
Date of manufacture.
The date could be a year, year and
month, or year, month and day, as
appropriate. See ISO 8601 for date
formates.
Catalog or model number.Rating Plate
Inside of AC adapter
Stick to Power Switch
Bottom panel of the console
Rating Plate
Serial numberRating Plate
Direct Current.
For products to be powered from a DC
supply.
InputRating Plate
For use with adapter model TWADP
100
DescriptionRating Plate
Type CF Defib-Proof Applied Part (heart
in the box with paddle) symbolis in
accordance with IEC 60878-02-06.
Rating Plate
Rating Plate
ECG Module
1-10Section 1-2 - Important Conventions
Page 35
GE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-16 Warnings
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
“Consult accompanying documents” is
intended to alert the user to refer to the
operator manual or other instructions
when complete information cannot be
provided on the label.
Various
Do not push the system.
This symbol indicates that the waste of
electrical and electronic equipment
must not be disposed as unsorted
municipal waste and must be collected
separately. Please contact an
authorized representative of the
manufacturer for information concerning
the decommissioning of your
equipment.
When closing the LCD cover, use
caution to avoid injuring hands or
fingers as there is a closing mechanism
which allows the LCD cover to
automatically close.
Indicates the product contains
hazardous materials in excess of the
limits established by Chinese standard
SJ/T11363-2006 Requirements for
Concentration Limits for Certain
Hazardous Substances in Electronic
Information Products. The number in
the symbol is the Environment-friendly
Use Period (EFUP), which indicates the
period during which the toxic or
hazardous substances or elements
contained in electronic information
products will not leak or mutate under
normal operating conditions so that the
use of such electronic information
products will not result in any severe
environmental pollution, any bodily
injury or damage to any assets.
Rear of Docking Cart and rear of
Isolation Cart.
Rating Plate
Rating Plate
Rear panel, rating plate
This product consists of devices that
may contain mercury, which must be
recycled or disposed of in accordance
with local, state, or country laws. (Within
this system, the backlight lamps in the
monitor display contain mercury.)
Bottom panel of the console
Chapter 1 - Introduction1-11
Page 36
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 1-16 Warnings
LABEL/SYMBOLPURPOSE/MEANINGLOCATION
The separate collection symbol is
affixed to a battery, or its packaging, to
advise you that the battery must be
recycled or disposed of in accordance
with local or country laws. The letters
below the separate collection symbol
indicate whether certain elements
(Pb=Lead, Cd=Cadmium, Hg=Mercury)
are contained in the battery. To minimize
potential effects on the environment and
human health, it is important that all
marked batteries that you remove from
the product are properly recycled or
disposed. For information on how the
battery may be safely removed from the
device, please consult the service
manual or equipment instructions.
Information on the potential effects on
the environment and human health of
the substances used in batteries is
available at this url: http://
www.gehealthcare.com/euen/weeerecycling/index.html
Battery Pack
No hazardous substance, above the
maximum concentration value, is
present. Maximum concentration values
for electronic information products, as
set by the People’s Republic of China
Electronic Industry Standard SJ/
T11364-2006, include the hazardous
substances of lead, mercury, hexavalent
chromium, cadmium, polybrominated
biphenyl (PBB), and polybrominated
diphenyl ether (PBDE).
Do not connect the DVD-RW to the
system while scanning.
GOST Symbol. Russia Regulatory
Country Clearance.
DVD-RW
Bottom
1-12Section 1-2 - Important Conventions
Page 37
GE
WARNINGWARNING
WARNINGWARNING
CAUTION
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 1-3
Safety Considerations
1-3-1Introduction
The following safety precautions must be observed during all phases of operation, service and repair of
this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this
manual, violates safety standards of design, manufacture and intended use of the equipment.
1-3-2Human Safety
Operating personnel must not remove the system covers.
Servicing should be performed by authorized personnel only.
Only personnel who have participated in a LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid etraining
are authorized to service the equipment.
1-3-3Mechanical Safety
Ultrasound probes are highly sensitive medical instruments that can easily be damaged
by improper handling. Use care when handling and protect from damage when not in
use. Do not use a damaged or defective probe. Failure to follow these precautions can
result in serious injury and equipment damage.
Never use a probe that has fallen to the floor. Even if it looks ok, it may be damaged.
The LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e weights 4.6kg or more, depending on
installed peripherals, when ready for use. To avoid possible injury and equipment damage:
ALWAYS:
•Use the handle to move the system.
•Do not let the system strike walls or door frame.
•Limit movement to a slow careful walk.
NOTE:Special care should be taken when transporting the unit in a vehicle:
•Before transporting, place the system in its special storage case.
•Ensure that the system is firmly secured while inside the vehicle.
•Secure system with straps or as directed otherwise to prevent motion during transport.
•Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and
erratic stops or starts.
1-3-4Electrical Safety
To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The
system is equipped with a three-conductor AC power cable. This must be plugged into an approved
electrical outlet with protective ground.
The power outlet used for this equipment should not be shared with other types of equipment.
Both the system power cable and the power connector meet international electrical standards.
1-3-5Label Location
Please refer to Basic User Manual for label location information.
Chapter 1 - Introduction1-13
Page 38
GE
WARNINGWARNING
CAUTION
NOTICE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-3-6Battery Safety
To avoid the risk of injury, follow the warning and cautions to make sure that the battery does not burst,
ignite, or generate heat of fumes.
•The battery has a safety device. Do not disassemble or alter the battery.
•Charge and discharge the batteries only when the ambient temperature is between 10 and
40 C (50 F and 104 F).
•Do not short-circuit the battery by directly connecting the negative terminals with metal
objects.
•Do not heat the battery or discard it in a fire.
•Do not expose the battery to temperature over 50° C (122° F). Keep it away from fire and other
heat sources.
•Do not charge the battery near a heat source, such as a fire or heater.
•Do not leave the battery in direct sunlight.
•Do not drop packs from height to prevent them from possible malfunction damage.
•Do not pierce the battery with a sharp object, hit it, or step on it.
•Do not use a damaged battery.
•Do not solder a battery.
•Do not connect the battery to an electrical power outlet.
•Do not contact PCM (Power Control and Monitor, it’s a small board in the battery) directly to
prevent packs from ESD damage.
•In case of longer non-use of the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e, please
make sure the battery is removed.
To avoid the battery bursting, igniting, or fumes from the battery causing equipment damage,
observe the following precautions:
•Do not immerse the battery in water or allow it to get wet.
•Do not put the battery into a microwave oven or pressurized container.
•If the battery leaks or emits an odor, remove it from all possible flammable sources.
•If the battery emits an odor or heat, is deformed or discolored, or in a way appears abnormal
during use, recharging or storage, immediately remove it and stop using it. If you have any
questions about the battery, consult GE or your local representative.
•Short term (less than one month) storage of battery pack:
•Store the battery in a temperature range between -5° C (23° F) and 50° C (122°F).
•Use only GE recognized batteries.
•In case of the long term (3 months or more) storage:
•Store the battery in a temperature range of -5° C (-23° F) and 50° C (122°F).
•When charging for the first time after long-term storage. Recover such packs to
original performance through repeating several cycles of full charging and
discharging.
•When store packs for more than 6 months, charge at lease once charging require
per 6 months to prevent leakage and deterioration in performance due to selfdischarging.
•When the system isn't powered on continuously more than 6 months, in order to prevent
leakage and deterioration in performance of CMOS battery, power on the system at least
once per 6 months for more than 10 hours to have CMOS battery fully charged. Time and
date need to be re-setup.
The battery shall be shipped in about 30% charged state. Those packs have to be fully charged and
discharged up to 3 times to utilize Li-lon smart packs before use.
1-14Section 1-3 - Safety Considerations
Page 39
GE
DANGER
WARNING
WARNING
WARNING
WARNING
WARNING
WARNING
NOTICE
Signed
Date
TAG
&
LOCKOUT
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-3-7Dangerous Procedure Warnings
Warnings, such as the examples below, precede potentially dangerous procedures throughout this
manual. Instructions contained in the warnings must be followed.
DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT
IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING
AND ADJUSTING.
EXPLOSION WARNING
DO NOT OPERATE THE EQUIPMENT IN AN EXPLOSIVE ATMOSPHERE.
OPERATION OF ANY ELECTRICAL EQUIPMENT IN SUCH AN ENVIRONMENT
CONSTITUTES A DEFINITE SAFETY HAZARD.
DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT
BECAUSE OF THE DANGER OF INTRODUCING ADDITIONAL HAZARDS, DO NOT
INSTALL SUBSTITUTE PARTS OR PERFORM ANY UNAUTHORIZED MODIFICATION
OF THE EQUIPMENT.
SHUT DOWN FORCEDLY OR PLUG IN/OUT ACDC INVALID MAY CAUSE THE
DAMAGE OF SYSTEM FILES.
1-3-8Lockout/Tagout (LOTO) requirements
Follow OSHA Lockout/Tagout requirements (USA) or local Lockout/Tagout requirements by ensuring
you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Remove battery.
6.) Control all stored and residual energy.
7.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE
GREATER THAN 30 VOLTS:
1. TURN OFF THE SCANNER.
2. UNPLUG THE SYSTEM.
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON
WILL TURN OFF.
5. REMOVE THE SYSTEM BATTERY.
Chapter 1 - Introduction1-15
Page 40
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
1-3-9Returning/Shipping Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEMS policy states that body fluids must be properly removed from any part or equipment prior to
shipment. GEMS employees, as well as customers, are responsible for ensuring that parts/equipment
have been properly decontaminated prior to shipment. Under no circumstance should a part or
equipment with visible body fluids be taken or shipped from a clinic or site (for example, body coils or
an ultrasound probe).
The purpose of the regulation is to protect employees in the transportation industry, as well as the
people who will receive or open this package.
NOTE:The US Department of Transportation (DOT) has ruled that “items that were saturated and/or
dripping with human blood that are now caked with dried blood; or which were used or intended
for use in patient care” are “regulated medical waste” for transportation purposes and must be
transported as a hazardous material.
NOTE:The USER/SERVICE staff should dispose all the waste properly as per federal, state, and local
waste disposal regulation.
The ultrasound system is not meant to be long term storage of patient data or images. The user is
responsible for the date on the system and a regular backup is highly recommended.
If the system is sent for repair, please ensure that any patient information is backup and erased from
the system before shipping. It is always possible during system failure and repair to lose patient data.
GE is not responsible for the loss of this data.
If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes,
GE will ascertain agreement from the customer. The patient information shall only be transferred by
approved service processes, tools and devices restricting access, protecting or encrypting data where
required, and providing traceability in the form of paper or electronic documents at each stage of the
procedure while maintaining compliance with cross-border restrictions of patient information transfers.
1-16Section 1-3 - Safety Considerations
Page 41
GE
WARNING
WARNING
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 1-4
EMC, EMI, and ESD
1-4-1Electromagnetic Compatibility (EMC)
Electromagnetic compatibility describes a level of performance of a device within its electromagnetic
environment. This environment consists of the device itself and its surroundings including other
equipment, power sources and persons with which the device must interface. Inadequate compatibility
results when a susceptible device fails to perform as intended due interference from its environment or
when the device produces unacceptable levels of emission to its environment. This interference is often
referred to as radio–frequency or electromagnetic interference (RFI/EMI) and can be radiated through
space or conducted over interconnecting power of signal cables. In addition to electromagnetic energy,
EMC also includes possible effects from electrical fields, magnetic fields, electrostatic discharge and
disturbances in the electrical power supply.
1-4-2CE Compliance
The LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e unit conforms to all applicable conducted
and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF
fields, magnetic fields and power line transient requirements.
For applicable standards refer to the Safety Chapter in the Basic User Manual.
NOTE:For CE Compliance, it is critical that all covers, screws, shielding, gaskets, mesh, clamps, are in
good condition, installed tightly without skew or stress. Proper installation following all
comments noted in this service manual is required in order to achieve full EMC performance.
1-4-3Electrostatic Discharge (ESD) Prevention
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING
THE NECESSARY ESD PRECAUTIONS:
1.FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC
SENSITIVE EQUIPMENT.
Chapter 1 - Introduction1-17
Page 42
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 1-5
Customer Assistance
1-5-1Contact Information
If this equipment does not work as indicated in this service manual or in the User Manual, or if you
require additional assistance, please contact the local distributor or appropriate support resource, as
listed below.
Prepare the following information before you call:
-System ID serial number.
-Software version.
Table 1-17 Phone Numbers for Customer Assistance
LocationPhone Number
USA
GE Medical Systems
Ultrasound Service Engineering
9900 Innovation Drive
Wauwatosa, WI 53226
14470 Bergen Blvd
Suite 400
Noblesville IN 46060
Canada1-800-668-0732
Latin America
Europe
GE Ultraschall Deutschland GmbH& Co. KG
BeethovenstraBe 239
Postfach 11 05 60, D-42665 Solingen
Germany
Asia (Singapore)
GE Ultrasound Asia
Service Department - Ultrasound
298 Tiong Bahru Road #15-01/06
Central Plaza
Singapore 169730
Service: On-site
Service: Parts
Applications support
Inquires/Loaner Return
Depot PTL
Service
Applications support
Phone: +33 (0)130-831-300 (General Imaging and Cardiac)
Fax: +49 (0)212-2802-431
Tel: +65 291-8528
Fax: +65 6291-7006
1-800–437–1171
1-800-558-2040
1-800-682-5327 or 1-262-524-5698
317-814-8421
317-814-8409
1-800-321-7937
1-262-524-5698
Japan Support Center
Phone: 81-42-648-2944
Fax: 81-42-648-2905
1-18Section 1-5 - Customer Assistance
Page 43
GE
Direction 5370626-100, Revision18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
65-1, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do,
Korea 462-120
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC
DBA GE Healthcare
9900 Innovation Drive
Wauwatosa, WI 53226
U.S.A
TEL: +86 510-85225888
FAX: +86 510-85226688
ManufacturerPhone Number
TEL: +86 510-85225888
FAX: +86 510-85226688
TEL: +82-31-740-6112
FAX: +82-31-740-6435
TEL: (1) 800-437-1171
FAX: (1) 414-721-3865
Chapter 1 - Introduction1-19
Page 44
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
This page was intentionally left blank.
1-20Section 1-5 - Customer Assistance
Page 45
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 2
Site preparations
Section 2-1
Overview
2-1-1Purpose of this chapter 2
This chapter provides the information required to plan and prepare for the installation of a LOGIQ e/
LOGIQ e Vet/LOGIQ i/Vivid e . Included are descriptions of the facility and electrical needs to be met
by the purchaser of the unit.
2-1-2Chapter Contents
Table 2-1 Contents in Chapter 2
SectionDescriptionPage Number
2-1
2-2
2-3
Overview
General Console Requirements
Facility Needs
2-1
2-2
2-6
Chapter 2 Site preparations2-1
Page 46
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 2-2
General Console Requirements
2-2-1Console Environmental Requirements
Table 2-2 Environmental Requirements for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Scanners
OperationalStorageTransport
Temperature
Humidity
Pressure
Temperatures in degree C, conversion to degree F =(degree C*(9/5) + 32)
10 - 40 degree C-5 - 50 degree C-5 - 50 degree C
30 - 75%
non-condensing
700 - 1060hPa700 - 1060hPa700 - 1060hPa
10 - 90%
non-condensing
10 - 90%
non-condensing
2-2-1-1Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and diameters can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interface.
2-2-2Electrical Requirements
NOTE:GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Power Requirements
Table 2-3 Electrical Specifications for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
AdapterVoltagePowerCurrentFrequency
TWADP100100-240 VAC130VA max.1.3 A (max.)50/ 60HZ
2-2Section 2-2 - General Console Requirements
Page 47
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-2-2-2Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
Table 2-4 Inrush Current
Inrush Current
Voltage
100V0.38A0.41A
240V0.20A0.21A
2-2-2-3Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
2-2-2-4Site Power Outlets
A dedicated AC power outlet must be within reach of the unit without extension cords. Other adequate
outlets for the external peripherals, medical and test equipment needed to support this unit must also
be present within 1 m (3.2 ft.) of the unit. Electrical installation must meet all current local, state, and
national electrical codes.
2-2-2-5Unit Power Plug
If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
2-2-2-6Power Stability Requirements
Voltage drop-out
Max 10 ms.
Power Transients
Console OnlyConsole with all peripherals
(All applications)
Less than 25% of nominal peak voltage for less than 1 millisecond for any type of transient, including
line frequency, synchronous, asynchronous, or aperiodic transients.
Chapter 2 Site preparations2-3
Page 48
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-2-3EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ e/LOGIQ e Vet/
LOGIQ i/Vivid e complies with limits as stated on the EMC label. However there is no guarantee that
interface will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of defect.
These sources include:
•medical lasers,
•scanners,
•cauterizing guns,
•computers,
•monitors,
•fans,
•gel warmers,
•microwave ovens,
•light dimmers,
•portable phones.
The presence of a broadcast station or broadcast van may also cause interference.
See Ta bl e 2 -5 for EMI Prevention tips.
Table 2-5 EMI Prevention/abatement
EMI RuleDetails
Be aware of RF sources
Ground the unit
Replace all screws, RF
gaskets, covers, cores
Replace broken RF gaskets
Do not place labels where RF
gaskets touch metal
Use GE specified harnesses
and peripherals
Take care with cellular phones Cellular phones may transmit a 5 V/m signal; that could cause image artifacts.
Keep the unit at least 5 meters or 15 feet away from other EMI sources. Special shielding
may be required to eliminate interference problems caused by high frequency, high
powered radio or video broadcast signals.
Poor grounding is the most likely reason a unit will have noisy images. Check grounding of
the power cord and power outlet.
After you finish repairing or updating the system, replace all covers and tighten all screws.
Any cable with an external connection requires a magnet wrap at each end. Install the
shield over the front of card cage. Loose or missing covers or RF gaskets allow radio
frequencies to interfere with the ultrasound signals.
If more than 20% or a pair of the fingers on an RF gasket are broken, replace the gasket.
Do not turn on the unit until any loose metallic part is removed.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit
RF leakage. Or, if a label has been found in such a position, move the label.
The interconnect cables are grounded and require ferrite beads and other shielding. Also,
cable length, material, and routing are all important; do not change from what is specified.
Properly dress peripheral
cables
Do not allow cables to lie across the top of the card cage or hang out of the peripheral bays.
Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor
cables to the frame.
2-4Section 2-2 - General Console Requirements
Page 49
GE
NOTICE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-2-4Scan Probe Environmental Requirements
Operation:10° to 40° C
Storage:-10° to 50° C
NOTE:The recommended storage temperature range of 6Tc-RS is between 0°C and 45°C.
NOTE:Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF
-10 TO + 60 degrees C. WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS, THE
PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE.
Chapter 2 Site preparations2-5
Page 50
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 2-3
Facility Needs
2-3-1Recommended Ultrasound Room Layout
2-3-1-1Purchaser Responsibilities
The work and materials needed to prepare the site is the responsibility of the purchaser. Delay,
confusion, and waste of manpower can be avoided by completing pre installation work before delivery.
User the Pre Installation checklist to verify that all needed steps have been taken,
Purchaser reasonability includes:
•Procuring the materials required.
•Completing the preparations before delivery of the ultrasound system.
•Paying the costs for any alternations and modifications not specifically provided in the sales
contract.
NOTE:All electrical installation that are preliminary to the positioning of the equipment at the site
prepared for the equipment must be performed by licensed electrical contractors. Other
connections between pieces of electrical equipment, products involved (and the accompanying
electrical installations) are highly sophisticated and special engineering competence is
required. All electrical work on these product must comply with the requirements of applicable
electrical codes. The purchaser of GE equipment must only utilize qualified personnel to
perform electrical servicing on the equipment.
The desire to use a non-listed or customer provided product or to place an approved product further
from the system than the interface kit allows presents challenges to the installation team. To avoid
delays during installation, such variances should be made known to the individuals or group performing
the installation at the earliest possible date (preferable prior to purchase).
The ultrasound suite must be clean prior to delivery of the machine. Carpet is not recommended
because it collects dust and creates static. Potential sources of EMI (electromagnetic interference)
should also be investigated before delivery. Dirt, static, and EMI can negatively impact system.
2-6Section 2-3 - Facility Needs
Page 51
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-3-2Required Features
NOTE:GE Medical Systems requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Please note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
•Dedicated single branch power outlet of adequate amperage meeting all local and national codes
which is located less than 2.5 m (8 ft.) from the unit’s proposed location
•Door opening is at least 76 cm (30 in) wide
•Proposed location for unit is at least 0.2m (0.67 ft.) from the wall for cooling
•Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m of the unit to connect cables.
•Power outlets for other medical equipment and gel warmer
•Power outlets for test equipment and modem within 1 m (3.2 ft.) of unit
•Clean and protected space to store transducers (in their cases or on a rack)
•Material to safely clean probes (done with a plastic container, never metal)
2-3-3Desirable Features
•Door is at least 92 cm (3 ft.) wide
•Circuit breaker for dedicated power outlet is easily accessible
•Sink with hot and cold water
•Receptacle for bio–hazardous waste, like used probe sheaths
•Emergency oxygen supply
•Storage for linens and equipment
•Nearby waiting room, lavatory, and dressing room
•Dual level lighting (bright and dim)
•Lockable cabinet ordered by GE for its software and proprietary manuals.
Chapter 2 Site preparations2-7
Page 52
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
2-3-3-1Recommended and Alternate Ultrasound Room Layout
Recommended standard floor plan and a minimal floor plan for ultrasound equipment:
FILM
FILE
C AB INE T
SE CR ETA R YS OR
DO C TO R 'S DES K
DEDIC ATE D
,
R OOM
PR O C ESSIN G
F IL M VIE WE R
C OUNTER TO P
POWER
RECEP TACLE
C OUN T E R
TO P
SIN K
SU C T IO N LINE
EMERGECY
OX Y GEN
OV E RH EA D
LI G HT S D IMME R
DOOR
42 IN .
(107 C M)
A 14 by 17 foot Recommended Floor P lan
S cale : E ach square equals one square foot
18 IN .
(46 CM)
C ONS O LE
F OOT
SW
S TOO L
76 IN .
(193 C M)
LINEN SUPPLY
E XAM IN AT ION
TAB LE
24 IN .
(61 CM)
PAT IEN T
TO ILE T
F AC ILI T Y
F IL M VIE WE R
SIN K
F IL M SUPPLIES
LINEN SUPPLY
PR OB ES /SUPPLIES
EXT ERNAL
PE R IP HE R ALS
DEDIC ATED POWE R
OUTLETS
DOOR
30 IN .
(76 CM)
24 IN .
(61 CM)
F OOT
SW
S TOO L
E XAM IN AR ION
TAB LE
LOGIQ Book XP
C ONS O LE
76 IN .
(193 C M)
G E C AB INE T
FOT S OFTWARE
A ND MANUALS
A n 8 by 10 foot Minimal Floor P lan
Figure 2-1 RECOMMENDED ULTRASOUND ROOM LAYOUT
2-8Section 2-3 - Facility Needs
Page 53
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
DICOM services provide the operator with clinically useful features for moving images and patient
information over a hospital network. Examples of DICOM services include the transfer of images to
workstations for viewing or transferring images to remote printers. As an added benefit, transferring
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at
a lower cost.
2-3-4-4DICOM Option Pre-installation Requirements
To configure the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e to work with other network connections, the
site’s network administrator must provide some necessary information.
Information must include:
•A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ e/LOGIQ e Vet/
LOGIQ i/Vivid e .
•The IP addresses for the default gateway and other routers at the site for ROUTING
INFORMATION.
•The host name, IP address, port and AE Title for each device the site wants connected to the
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e for DICOM APPLICATION INFORMATION. A field for
the make (manufacturer) and the revision of the device, is also included. This information may be
useful for solving errors.
Chapter 2 Site preparations2-9
Page 54
GE
LOGIQ e/
Host Name
AE Title
Local Port
IP Address
Net Mask
...
...
ROUTING INFORMATION
ROUTER2
ROUTER3
ROUTER1
...
...
...
...
...
...
...
Destination
IP Addresses
GATEWAY IP Addresses
Default
DICOM APPLICATION INFORMATION
NAME
Store 2
Store 3
Store 1
MAKE/REVISIONIP ADDRESSESPORTAE TITLE
Store 5
Store 6
Store 4
Storage
Commit
MPPS
Worklist
...
...
...
...
...
...
...
...
...
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Figure 2-2 Worksheet for DICOM Network Information
2-10Section 2-3 - Facility Needs
Page 55
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 3
System Setup
Section 3-1
Overview
3-1-1Purpose of Chapter 3
This chapter contains information needed to install the unit. Included are references to a procedure that
describes how to receive and unpack the equipment and how to file a damage or loss claim.
How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes,
and external peripherals for electrical safety are included in this procedure. Also included in this section
are guidelines for transporting the unit to a new site.
Table 3-1 Contents in Chapter 3
SectionDescriptionPage Number
3-1
3-2
3-3
3-4
3-5
3-6
3-7
3-8
3-9
3-10
3-11
Overview
Receiving and Unpacking the Equipment
Packing the Equipment
Preparing for Installation
Completing the Installation
System Configuration
Software/Option Configuration
Connectivity Installation Worksheet
Loading Base Image Software
Software Version check out
Paperwork
3-1
3-4
3-7
3-8
3-9
3-12
3-30
3-31
3-32
3-33
3-34
Chapter 3 System Setup3-1
Page 56
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-1-2Average Installation Time
Table 3-2 Average Installation Time
DescriptionAverage Installation TimeComments
Unpacking the scanner
Scanner wo/options
DICOM Option
20 minutes
30 minutes
30 minutes
Dependent on the configuration that is required
Dependent on the amount of configuration
The LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e installation and functional checkout will take approximately
one hour. LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e consoles with optional equipment may take slightly
longer.
3-2Section 3-1 - Overview
Page 57
GE
DANGER
DANGER
CAUTION
DANGER
DANGER
DANGER
DANGER
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-1-3Installation Warnings
1.) There are no operator serviceable components. To prevent shock, do not remove any covers or
panels. Should problems or malfunctions occur, unplug the power cord. Only qualified service
personnel should carry out servicing.
NOTE:For information regarding packing labels, refer to LABELS ON PACKAGE.
2.) After being transported, the unit may be very cold or hot. If this is the case, allow the unit to
acclimate before you turn it on. It requires one hour for each 2.5×C increment it's temperature is
below 10×C or above 40×C.
Equipment damage possibility. Turning the system on without acclimation after arriving at site
may cause the system to be damaged.
WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE
AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH
THE UNIT!
If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to
its operating environment.
To prevent electrical shock, connect the unit to a properly grounded power outlet. Do not use a
three to two prong adapter. This defeats safety grounding.
Do not operate this unit unless all board covers are securely in place.
OPERATOR MANUAL(S)
The User Manual(s) should be fully read and understood before operating the LOGIQ e/LOGIQ e
Vet/LOGIQ i/Vivid e and kept near the unit for quick reference.
ACOUSTIC OUTPUT HAZARD
Although the ultrasound energy transmitted from the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
probe is within FDA limits, avoid unnecessary exposure. Ultrasound energy can produce heat
and mechanical damage
Chapter 3 System Setup3-3
Page 58
GE
Adhesive tape
Plastic Bands
1
2
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-2
Receiving and Unpacking the Equipment
When a new system arrives, check that any components are not damaged and are not in short supply.
If shipping damage or shortage occurs, contact the address shown in Chapter 1.
1.) Cut the four PLASTIC BANDs.
2.) Cut the adhesive tape and open top covers of paper carton.
Figure 3-1 Open top covers of paper carton.
3-4Section 3-2 - Receiving and Unpacking the Equipment
Page 59
GE
CAUTION
Paper Pad
Console and interleavers
Console Package
Accessories Package
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-2 Receiving and Unpacking the Equipment (cont’d)
3.) Take out the Paper pad.
4.) Take out console together with 2 interleavers from console package.
5.) Take out the interleavers beside Accessories Package.
6.) Take out Accessories Package.
Figure 3-2 Unpacking the equipment
Do not lift the unit by the rubber band. Equipment damage may result.
Chapter 3 System Setup3-5
Page 60
GE
Console
Interleaver
Interleaver
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-2 Receiving and Unpacking the Equipment (cont’d)
7.) Remove 2 interleavers.
8.) Remove plastic bag.
Figure 3-3 Removing interleavers and plastic bag
3-6Section 3-2 - Receiving and Unpacking the Equipment
Page 61
GE
CAUTION
CAUTION
CAUTION
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-2 Receiving and Unpacking the Equipment (cont’d)
NOTE:Check the shipping container for special instructions. Verify that the container is intact. In some
cases a secondary container may be used. If so, ask the carrier for unpacking instructions.
Please carefully unpack the system, and do not dispose the package of LOGIQ e/LOGIQ e Vet/
LOGIQ i/Vivid e , so that it can be reused for service.
3-2-1Moving into Position
Do not lift the unit by the rubber band. Use handle to move system.
Equipment Damage Possibility. Lifting the console by holding covers may damage the covers.
Do not lift the console by holding any covers.
In general, a single adult can move the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e . Before moving, store
all loose parts in original accessory box or in back pack. Return probes to original box.
Section 3-3
Packing the Equipment
Please pack LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e in the reverse order of unpacking.
Figure 3-4 Labels on Package
Chapter 3 System Setup3-7
Page 62
GE
WARNINGWARNING
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-4
Preparing for Installation
3-4-1Verify Customer Order
Compare items received by the customer to that which is listed on the delivery order. Report any items
that are missing, back ordered or damaged.
3-4-2Physical Inspection
3-4-2-1System Voltage Settings
•Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ e/LOGIQ e
Vet/LOGIQ i/Vivid e Scanner is found on a label located on the AC adapter.
•220-240VAC(China); 100-120VAC(USA/Japan); 220-240VAC(Europe, Latin America).
Connecting a LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e scanner to the wrong voltage level
will most likely destroy the scanner.
3-4-3EMI Protection
This Unit has been designed to minimize the effects of Electro Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the system from image artifacts caused by
this interference. For this reason, it is imperative that all covers and hardware are installed and secured
before the unit is put into operation.
3-8Section 3-4 - Preparing for Installation
Page 63
GE
UPUP
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-5
Completing the Installation
3-5-1Power On / Boot Up
NOTE:After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-5-1-1Scanner Power On
Lower the handle. Plug the AC adapter output connector into the system DC input port (located on the
system’s rear panel) with the arrow side upward. Plug the AC adapter power cord into a grounded,
protective earth outlet.
Figure 3-5 Connect AC adapter
When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD,
Peripherals and the Back-end Processor.
Chapter 3 System Setup3-9
Page 64
GE
Power On/Off switch
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-5-1-2Turn on the system
Press the Power On/Off switch at the front of the system once.
Figure 3-6 Power On/Off Switch
When the Power On/Off switch on the Control Panel is pressed once, the Back-end Processor starts
and the software code is distributed to initiate the scanner.
No status messages are displayed during this process.
3-5-2Power Off/ Shutdown
NOTE:After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-5-2-1Back-end Processor Power Down
To power down the system:
1.) Press the Power On/Off switch at the front of the system once.
2.) The System-Exit window is displayed.
Figure 3-7 System Exit Window
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and is complete when the power status LED is turned
off.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
6.) Close LCD cover.
3-10Section 3-5 - Completing the Installation
Page 65
GE
Probe Latch
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-5-2-2Scanner Shutdown
Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
3-5-3Transducer Connection
1.) Carefully open the system LCD display, plug the probe connector into the probe port, then lock the
probe latch upward.
NOTE:Please ensure that the probe latch is in an unlocked position before you connect the probe to the
system.
Figure 3-8 Connect the probe
NOTE:It is not necessary to turn OFF power to connect or disconnect a probe.
Chapter 3 System Setup3-11
Page 66
GE
WEIGHT: 4.6 KG
NOTE: Length is in mm
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-6
System Configuration
3-6-1System Specifications
3-6-1-1Physical Dimensions and weight with old LCD
The physical dimensions of the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console with old LCD are
summarized in Figure 3-9 on page 3-12 .
Table 3-4 Physical Dimensions of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e with old LCD
HeightWidthDepthUnit
Console
Only
61100340375287337mm
2.43.913.414.811.313.3inches
Console
with handle
Console
Only
Console
with handle Console Only
Console with
handleUnit
Figure 3-9 Overall Dimensions and weight with old LCD
3-12Section 3-6 - System Configuration
Page 67
GE
WEIGHT: 5.2 KG
NOTE: Length is in mm
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-1-2Physical Dimensions and weight with new LCD
The physical dimensions of the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console with new LCD are
summarized in Figure 3-10 on page 3-13 .
Table 3-5 Physical Dimensions of LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e with new LCD
HeightWidthDepthUnit
Console
Only
70100346372295340mm
2.83.913.614.611.613.4inches
Console
with handle
Console
Only
Console
with handle
Console Only
Console with
handle
Unit
Figure 3-10 Overall Dimensions and weight with new LCD
Chapter 3 System Setup3-13
Page 68
GE
new LCD
old LCD
new LCD
old LCD
new LCD
old LCD
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-1-3How to identify old LCD and new LCD
•Identify LCD type from the appearance of Ultrasound System.
Figure 3-11 Difference between Old LCD and New LCD
3-6-2Electrical Specifications
Table 3-6 Electrical Specifications for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
AdapterVoltageTolerancesCurrentFrequency
TWADP100100-240 VAC+/-10%1.3A (max)50/60Hz
3-14Section 3-6 - System Configuration
Page 69
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-3Approved peripherals
Table 3-7 Approved peripherals
DeviceManufacturerModelInterfaceRemark
SONYUP-D897USB
B/W Printer
SONYUP-D898MDUSB
MITSUBISHI P95DUSB
Digital Color Printer
HP Color Printer
DVD-RW
Wireless Lan Adapter
3-pedal footswitchSteuteMKF 2-MED GP26USB
SONYUP-D23MDUSB
SONYUP-D25MDUSB
HPHP Officejet Pro K5400USB
HPHP Officejet Pro K8600USB
HPHP Deskjet 470USB
HPHP Officejet 100USB
LITEONLITEON DX-20A4PUSB
LITEONLITEON Model eHAU 120USB
LITEONLITEON Model eHAU 324USB
PLEXTORPX-L890UEUSB
LITEONLITEON eSAU108 DVDRWUSB
LITEONLITEON eUAU108 DVDRWUSB
LITEONLITEON eBAU108 DVDRWUSB
NetGearWNA3100USB
NetGear WN111V2USB
NetGearWG111V3USB
LinksysLinksys WUSB54GUSB
1-pedal footswitch
USB Memory
USB HUB
ECG
USB HDD
WhanamFSU2001USB
WhanamFSU-1000USB
SanDiskCRUZER 2GUSB
SanDiskCRUZER 4GUSB
SanDiskCRUZER 4G with U3 ApplicationUSB
Shenzhen GaojianHE 702AUSB
Shenzhen GaojianHE420GEUSB
NORAVECG-USB1USB
GEWUXIECG-USB1USB
SeagateUSB HDD 80GUSB
SeagateUSB HDD 250GUSB
SeagateUSB HDD 500GUSB
Chapter 3 System Setup3-15
Only supported by LOGIQ e R6.x.x/R7.x.x
and LOGIQ e Vet R6.x.x/R7.x.x
Page 70
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Table 3-7 Approved peripherals
DeviceManufacturerModelInterfaceRemark
DVD-RecorderPanasonic
Video Transfer Pinnacle
HandheldHandheld 3800GUSBOnly supported by LOGIQ e R5.2.x,
BarCode Reader*
HandheldHandheld 4600GUSB
MD-800E
MD-800U
Pinnacle Video Transfer - 8230-
10022-11
USB
USB
Not supported by LOGIQ e R4.x.x, Vivid e
R4.x.x, LOGIQ i R4.x.x
Not supported by LOGIQ e R4.x.x/5.0.x,
LOGIQ e Vet R5.0.x, Vivid e R4.x.x/5.0.x,
LOGIQ i R4.x.x/5.0.x
LOGIQ e R6.x.x, LOGIQ e R7.x.x, LOGIQ
e Vet R6.x.x, LOGIQ e Vet R7.x.x
NOTE:For detailed installation information and connection procedures, please refer to Peripheral
Installation manual.
3-16Section 3-6 - System Configuration
Page 71
GE
WARNINGWARNING
4
3
2
1
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-4Connecting Cables
Equipment damage possibility. Be sure to use the following recommended connecting cables to
connect recording devices and a network with LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e console.
Table 3-8 List of Connecting Cables
NamePart No.FigureNOTE
USB Cable5122305
3-6-5Peripherals/Accessories Connector Panel
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e peripherals and accessories can be properly connected using
the side connector panel.
3-6-5-1Rear Panel Connector
Located on the rear panel are video input and output connectors, power connector and ethernet port.
For USB Printer & USB DVD-RW
Figure 3-12 Rear Connector Panel
1.) Port for DC input (AC Adapter)
2.) Docking Connector
3.) SVGA Output
4.) Ethernet port
Chapter 3 System Setup3-17
Page 72
GE
2
1
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-2Left Panel Connector
Located on the left side panel are two USB Ports for control connections for printer, service tools and
earphone port.
Figure 3-13 Left Side Panel
1.) Two interchangeable USB Port (Digital Printers, DVD-RW and/or FootSwitch, etc)
2.) Earphone Port.
NOTE:Each outer (case) ground line of peripheral/accessory connectors are protectively grounded.
Signal ground lines are not isolated.
3-18Section 3-6 - System Configuration
Page 73
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-3This section indicates the pin assignment for each connector.
1. Pin Assignment of DC input
Connector: 4 Pin, Female
Table 3-9 Pin Assignments of DC input
Pin No.SignalPin No.Signal
1+20V3GND
2+20V4GND
2. Pin Assignment of USB
Table 3-10 Pin assignment of USB1
Pin No.SignalPin No.Signal
1+5VDC3DATA+
2DATA-4GND
Table 3-11 Pin assignment of USB2
Pin No.SignalPin No.Signal
1+5VDC3DATA+
2DATA-4GND
3. Pin assignment of RS232C for external VGA
Connector: D-SUB, 15Pin, Female
Table 3-12 Pin Assignments of RS232C for External VGA
Pin No.SignalPin No.Signal
1RED9 N/A
2GREEN10SGND
3BLUE11 N/A
4N/A12N/A
5GND13HSYNC
6RGND14VSYNC
7GGND15N/A
8BGND16
Chapter 3 System Setup3-19
Page 74
GE
NOTICE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals
A.) Connect B/W printer to the system.
B/W Printer can be properly connected using USB Port1 or USB Port 2 (Figure 3-14 on page 3-20 ).
Figure 3-14 Connect B/W printer to the system
B.) Connect UP-D23MD color printer to the system.
UP-D23MD Color Printer can be properly connected using USB Port1 or USB Port 2 (Figure 3-15
on page 3-20 ).
Figure 3-15 Connect digital color printer to the system
HP Deskjet 470 /HP Officejet Pro K5400 Color Printer must connect with PIT (Printer Isolation
Transformer) when being used.
3-20Section 3-6 - System Configuration
Page 75
GE
CAUTION
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
C.) Connect DVD-RW to the system.
DVD-RW can be properly connected using USB Port1 or USB Port2 (Figure 3-16 on page 3-21 ).
Figure 3-16 Connect DVD-RW to the system
DO NOT connect DVD-RW to the system while scanning.
LITEON DVD-RW drive recommended media list
Media issues are common throughout the DVD-RW drive industry. Because DVD media vendors often
change disc suppliers, quality levels may change due to manufacturing differences. This means that
you may encounter DVD/CD creation problems with media that may have worked successfully before.
Overall system configuration and other factors may also affect the success of creating a DVD/CD.
The following media types have been tested. LITEON highly recommends that you use the media types
on this list when creating your DVDs/CDs.
Not all brands of media have been tested and, therefore, you may encounter success with other brands
not listed. This list will be updated as other media is tested and approved.
Table 3-13 LITEON DVD-RW drive recommended media list
Brand NameType
SONYDVD-R
SONYCD-R
SONYDVD-RW
MitsubishiCD-R
MitsubishiCD-RW
MitsubishiDVD-R
MitsubishiDVD-RW
JVCDVD-R
MBICD-R
MBIDVD-R
Taiyo YudenCD-R
Taiyo YudenDVD-R
Chapter 3 System Setup3-21
Page 76
GE
3-pedal footswitch
1-pedal footswitch
Wireless Lan
Adapter
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
D.) Connect FootSwitch to the system.
FootSwitch can be properly connected using USB Port1 or USB Port2.
Figure 3-17 Connect Foot Switch to the system
E.) Connect Wireless LAN Adapter to the system.
Wireless LAN Adapter can be properly connected using USB port.
Figure 3-18 Connect Wireless LAN Adapter to the system
F.) Connect the CRT to the system.
CRT can be properly connected using the SVGA output.
3-22Section 3-6 - System Configuration
Page 77
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
G.) Connect the USB Memory to the system. The USB Memory can be properly connected using USB
port 1 or 2.
Figure 3-19 USB Memory Connection
H.) Connect the ECG to the system. ECG can be properly connected using USB port 1 or 2.
Figure 3-20 ECG
Chapter 3 System Setup3-23
Page 78
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
I.) Connect the USB HDD to the system. The USB Harddisk can be properly connected using USB
port 1 and 2.
Figure 3-21 Emergency Disk Connection
J.) Connect the BarCode Reader to the system, the BarCode Reader can be properly connected via
USB port 1 or 2.
Figure 3-22 BarCode Reader Connection
3-24Section 3-6 - System Configuration
Page 79
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
K.) Connect the Sandisk U3 USB Flash Drive to the system. The Sandisk U3 USB Flash Drive can be
properly connected using USB port 1 and 2.
Figure 3-23 Sandisk U3 USB Flash Drive Connection
NOTE:The Password of the Sandisk U3 USB Flash Drive is Predefined.
NOTE:The Sandisk U3 USB Flash Drive is only available on LOGIQ e R6.x.x/R7.x.x and LOGIQ e Vet
R6.x.x/R7.x.x.
Connect the security enabled Sandisk U3 USB Flash Drive into the system, input the password in the
pop-up dialog. Click OK to login, click Cancel to cancel.
Figure 3-24 Input the password
The Sandisk U3 USB Flash Drive can be used if the right password is entered. An error message will
display if the wrong password is entered.
Chapter 3 System Setup3-25
Page 80
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
Figure 3-25 Wrong password error
Click OK, input the password again.
The Sandisk U3 USB Flash Drive will be locked if reaching the maximum number of failed password
attempts.
Figure 3-26 Sandisk U3 USB Flash Drive locked
NOTE:Please refer to the operation manual of each peripheral for information needed by the user to
operate the system safely.
For detailed installation information, please refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Peripheral Installation Instruction manual.
3-26Section 3-6 - System Configuration
Page 81
GE
For LOGIQ e R5.x.x, R6.x.x, R7.x.x, LOGIQ i R5.x.x,
For LOGIQ e R4.x.x, LOGIQ i R4.x.x, Vivid e R4.x.x
Vivid e R5.x.x/R6.x.x, LOGIQ e Vet R6.x.x, R7.x.x
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-5-4Connect peripherals (cont’d)
L.) Connect the external monitor to the system. The monitor can be properly connected using the VGA
port.
Figure 3-27 VGA Connection
Press Ctrl+Alt+V
Select Intel(R) Dual Display Clone
and select OK
on the keyboard, a dialog box appears.
Figure 3-28 Graphics Controller Properties
.
; check the Same display configuration driver on both display box
NOTE:Please refer to the operation manual of each peripheral for information needed by the user to
operate the system safely.
NOTE:For LOGIQ e R5.2.x, R6.x.x, R7.x.x, LOGIQ e Vet R6.x.x, R7.x.x, the default set is dual, it will
not pop up Figure 3-28.
For detailed installation information, please refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Peripheral Installation manual.
Chapter 3 System Setup3-27
Page 82
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-6Available Probes
See in specification in the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e User Reference Manual for Probes
and intended use.
For LOGIQ e R4.0.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS
Probes.
For LOGIQ e R5.0.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS,
i/t739-RS, 9L-RS Probes.
For LOGIQ e R5.2.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS,
i/t739-RS, 9L-RS, 16L-RS Probes.
For LOGIQ e R6.x.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS,
i/t739-RS, 9L-RS, 16L-RS, 6Tc-RS, 6S-RS Probes.
For LOGIQ e R7.x.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS,
i/t739-RS, 9L-RS, 6Tc-RS, 6S-RS, L8-18i-RS,P2D Probes.
For LOGIQ i R4.x.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 12L-RS
Probes.
For LOGIQ i R5.x.x, the system supports 4C-RS, E8C-RS, 8C-RS, i12L-RS, 8L-RS, 3S-RS, 6S-RS,
12L-RS, i/t739-RS, P2D, 9L-RS Probes.
For Vivid e R4.x.x, the system supports 4C-RS, i12L-RS, 8L-RS, 3S-RS Probes.
For Vivid e R5.0.x, the system supports 4C-RS, i12L-RS, 8L-RS, 3S-RS, 8C-RS, 9L-RS, P2D Probes.
For Vivid e R5.2.x, the system supports 4C-RS, 8C-RS, E8C-RS, 8L-RS, 9L-RS, 12L-RS, i12L-RS, 3SRS, 6S-RS, P2D Probes.
For Vivid e R6.x.x, the system supports 4C-RS, 8C-RS, E8C-RS, 8L-RS, 9L-RS, 12L-RS, i12L-RS, 3SRS, 6S-RS, P2D, 6Tc-RS Probes.
For LOGIQ e Vet R5.0.x, the system supports 4C-RS Vet, E8C-RS Vet, 8C-RS Vet, i12L-RS Vet, 8LRS Vet, 3S-RS Vet, 12L-RS Vet, 9L-RS Vet, i739-RS LC, 6S Vet-RS.
For LOGIQ e Vet R5.2.x, the system supports 4C-RS Vet, E8C-RS Vet, 8C-RS Vet, i12L-RS Vet, 8LRS Vet, 3S-RS Vet, 12L-RS Vet, 9L-RS Vet.
For LOGIQ e Vet R6.x.x, the system supports 4C-RS Vet, E8C-RS Vet, 8C-RS Vet, i12L-RS Vet, 8LRS Vet, 3S-RS Vet, 12L-RS Vet, 9L-RS Vet, 6S Vet-RS, i739-RS-LC, 6Tc-RS Vet.
For LOGIQ e Vet R7.x.x, the system supports 4C-RS Vet, E8C-RS Vet, 8C-RS Vet, i12L-RS Vet, 8LRS Vet, 3S-RS Vet, 12L-RS Vet, 9L-RS Vet, 6S Vet-RS, i739-RS-LC, 6Tc-RS Vet, t739-RS Vet, L8-18iRS Vet, P2D probe Vet.
NOTE:16L-RS is not available in China.
3-28Section 3-6 - System Configuration
Page 83
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
3-6-6Available Probes (cont’d)
Table 3-14 List of Probes for LOGIQ e/LOGIQ i/Vivid e
Probe Name Material of HeadshellArea of UsingTYPECatalog NumberPart Number
NOTE:i739-RS-LC, 6SVet-RS and 6Tc-RS Vet probes are only available for LOGIQ e Vet R6.x.x and R7.x.x.
t739-RS Vet, L8-18i-RS Vet and P2D probe Vet probes are only available for LOGIO e Vet R7.x.x.
Section 3-7
Software/Option Configuration
Refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic User Manual, Chapter 16, Customizing Your
System for information on configuring items like Hospital, Department, Language, Units (of measure),
Date, Time and Date Format.
For information on configuring Software Options, Refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
Basic User Manual, Chapter 16, Customizing Your System.
For information on configuring DICOM Connectivity, Refer to the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid
e Basic User Manual, Chapter 16, Customizing Your System.
3-30Section 3-7 - Software/Option Configuration
Page 85
GE
Site System Information
Services (Destination Devices)
TCP/IP Settings
Device Type
Manufacturer
Name
IP Address
Port
AE Title
1
2
3
4
5
6
7
8
9
10
11
12
Name - AE Title:
Dept:
LOGIQ SN:
Type:
Floor:
Room:
REV:
Comments:
CONTACT INFORMATION
Name
Title
Phone
E-Mail Address
Site:
IP Settings
IP Address:
Subnet Mask:
Default Gateway:
Remote Archive Setup
Remote Archive IP:
Remote Archive Name:
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-8
Connectivity Installation Worksheet
Chapter 3 System Setup3-31
Page 86
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-9
Loading Base Image Software
Please refer to:
Section 8-3 "Loading the System Software" on page 8-4
3-32Section 3-9 - Loading Base Image Software
Page 87
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-10Software Version check out
3-10-1Functional Check-out
1.) Power on LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e scanner and wait until system booting to main
screen.
2.) Press Utility/Config
3.) Choose the About
key on control panel.
button on the right.
Figure 3-29 About
4.) Check whether "Software version" is the right version for use.
Figure 3-30 Software version
Chapter 3 System Setup3-33
Page 88
GE
Direction 5370626-100, Revision 16 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 3-11
Paperwork
NOTE:During and after installation, the documentation (i.e. User Manuals, Installation Manuals...) for the
peripheral units must be kept as part of the original system documentation. This will ensure that all
relevant safety and user information is available during the operation and service of the complete
system.
3-11-1Product Locator Installation
NOTE:The Product Locator Installation Card shown may not be same as the provided Product Locator
card.
Figure 3-31 Product Locator Installation Card
3-11-2User Manual(s)
User Check that the correct User Manual(s) for the system and software revision, is included with the
installation. Specific language versions of the User Manual may also be available. Check with your GE
Sales Representative for availability.
3-34Section 3-11 - Paperwork
Page 89
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Chapter 4
Functional Checks
Section 4-1
Overview
4-1-1Purpose for Chapter 4
This chapter provides procedures for quickly checking major functions of the LOGIQ e/LOGIQ e Vet/
LOGIQ i/Vivid e console, diagnostics by using the built-in service software, and power supply
adjustments.
Table 4-1 Contents in chapter 4
SectionDescriptionPage Number
4-1
4-2
4-3
4-4
4-5
Section 4-2
Required Equipment
To perform these tests, you'll need any of the sector, linear, or convex transducers.
(normally you should check all the transducers used on the system)
Overview
Required Equipment
General Procedure
Software Configuration Checks
Peripheral Checks
4-1
4-1
4-2
4-37
4-37
Chapter 4 Functional Checks4-1
Page 90
GE
CAUTION
NOTICE
Signed
Date
TAG
&
LOCKOUT
UPUP
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
Section 4-3
General Procedure
SYSTEM REQUIRES ALL COVERS
Operate this unit only when all board covers and frame panels are securely in place. The covers
are required for safe operation, good system performance and cooling purposes.
Lockout/Tagout Requirements (For USA only)
Follow OSHA Lockout/Tagout requirements by ensuring you are in total control of the Power Cable on
the system.
4-3-1Power On/Boot Up
After AC/DC is connected correctly to the scanner, the power is applied to the scanner. When the
Control panel Power On/Off key is pressed once, the System starts.
4-3-1-1Scanner Power On
Lower the handle. Plug the AC adapter output connector into the system DC input port (located on the
system’s rear panel) with the arrow side upward. Plug the AC adapter power cord into a grounded,
protective earth outlet.
Figure 4-32 connect AC adapter
4-2Section 4-3 - General Procedure
Page 91
GE
Power On/Off switch
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-1-1Scanner Power On (cont’d)
When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD,
Peripherals and the Back-end Processor.
4-3-1-2Turn on the system
Press the Power On/Off switch at the front of the system once.
Figure 4-33 Power On/Off Switch
When the Power On/Off switch on the Control Panel is pressed once, the Back-end Processor starts
and the software code is distributed to initiate the scanner.
No status messages are displayed during this process.
4-3-2Power Off/ Shutdown
NOTE:After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
4-3-2-1Back-end Processor Power Down
To power down the system:
1.) Press the Power On/Off switch at the front of the system once.
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
Figure 4-1 System Exit Window
Chapter 4 Functional Checks4-3
Page 92
GE
CAUTION
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-2-1Back-end Processor Power Down (cont’d)
4.) The shutdown process takes 15 seconds and the power off sequence is complete when the power
status LED is turned off.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
6.) Close LCD cover.
4-3-2-2Scanner Shutdown
Disconnect the Mains Power Cable is necessary. For example: Relocating the scanner.
DO NOT unplug and/or transport the unit until after the power off sequence has been completed.
Failure to do so may result in corrupted patient files.
4-3-2-3Full Maintenance Reboot
Full Maintenance Reboot will fully restart the whole system for performance improvement.
NOTE:It’s recommended to do Full Maintenance Reboot at least daily or when system starts to slow
down or lag in its reaction.
1.) Press the Power On/Off switch at the front of the system once.
2.) Click “Full Maintenance Reboot”.
3.) Full Maintenance Reboot dialog is displayed.
Figure 4-2 “Full Maintenance Reboot” dialog
4-3-2-4Exit (only activated for LOGIQ e R6.0.2)
Exit will reboot application.
1.) Press the Power On/Off switch at the front of the system once.
2.) Click “Exit”.
3.) The application will reboot.
4-4Section 4-3 - General Procedure
Page 93
GE
NOTICE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-3Archiving and Loading Presets
NOTE:Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an DVDR disk (or USB memory device) for reloading on the system.
Presets should NOT be saved on the same DVD-R disk (or USB memory device) as images. The
Archive Menu lists the images but does NOT list the presets stored on a DVD-R disk (or USB memory
device).
4-3-3-1Archiving Presets to an DVD-R Disk (or USB memory device)
1.) Insert an empty (blank) DVD-R disk into the DVD-RW.
2.) Access to the Utility/Config Menu, and select System. The Backup sheet will be shown on the LCD
display.
Figure 4-3 Backup Sheet
3.) Select the item to Backup/Restore.
4.) Enter backup destination or browse through the disk to locate the destination.
5.) Select Backup now. The backup status for each item is displayed on the Result column.
Chapter 4 Functional Checks4-5
Page 94
GE
NOTICE
Signed
Date
TAG
&
LOCKOUT
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-3-2Loading Presets from an DVD-R disk (or USB memory device)
1.) Insert the DVD-R disk with the archived Presets into the DVD-RW.
2.) Access to Utility->System->Backup/Restore. The Restore sheet will be shown on the LCD display.
(see Figure 4-3 on page 4-5 )
3.) Select the items needed to be stored.
4.) Select Restore. The system performs the restore and restarts.
4-3-4Adjusting the Display Monitor
Please refer to Section 6-2 "Monitor Adjustments" on page 6-2.
4-3-5Lockout/Tagout (LOTO) requirements
Follow OSHA Lockout/Tagout requirements (USA) or local Lockout/Tagout requirements by ensuring
you are in total control of the AC power plug at all times during the service process.
To apply Lockout/Tagout:
1.) Plan and prepare for shutdown.
2.) Shutdown the equipment.
3.) Isolate the equipment.
4.) Apply Lockout/Tagout Devices.
5.) Remove battery.
6.) Control all stored and residual energy.
7.) Verify isolation.
All potentially hazardous stored or residual energy is relieved.
Energy Control and Power Lockout for LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e
WHEN SERVICING PARTS OF THE SYSTEM WHERE THERE IS EXPOSURE TO VOLTAGE
GREATER THAN 30 VOLTS:
1. TURN OFF THE SCANNER.
2. UNPLUG THE SYSTEM.
3. MAINTAIN CONTROL OF THE SYSTEM POWER PLUG.
4. WAIT FOR AT LEAST 20 SECONDS FOR CAPACITORS TO DISCHARGE AS THERE ARE NO
TEST POINTS TO VERIFY ISOLATION. THE AMBER LIGHT ON THE OP PANEL ON/OFF BUTTON
WILL TURN OFF.
5. REMOVE THE SYSTEM BATTERY.
4-6Section 4-3 - General Procedure
Page 95
GE
1
2
3
4
5
6
7
8
9
10
11
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-6System Features
4-3-6-1Control Panel
Figure 4-4 Control Panel Tour
1.) TGC
2.) New Patient
3.) End Exam
4.) Mode/Gain/Auto Keys
5.) Preset Key
6.) Imaging/Measurement Keys
7.) Depth/Zoom/Ellipse
8.) Image Keys
9.) Print Keys
10.)Freeze
11.)Keyboard
Chapter 4 Functional Checks4-7
Page 96
GE
2
1
3
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-6-2LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e SoftMenu Key Tour
In general, there are two types of softMenu keys: Paddle Switch and adjustable knobs.
Figure 4-5 SoftMenu Key Tour
1.) The Paddle Switch is used to access and adjust the Sub SoftMenu.
2.) Press the adjustable knobs to toggle option menu between line one and line two.
3.) Rotate the adjustable knobs to adjust the corresponding parameters.
4-8Section 4-3 - General Procedure
Page 97
GE
4
5
6
7
8
9
10
11
19
20
2122
18
12
16
17
15
14
13
23
123
24
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-6-3Monitor Display
Figure 4-6 Monitor Display Tour
Table 4-2 Monitor Display Features
1. Institution/Hospital Name, Date, Time, Operator
Identification, system status (real-time of frozen).
12. Probe Identifier. Exam Study.24. Active key for Depth/Zoom/Ellipse
13. Imaging Parameters by Mode (current mode highlighted).
16. Body Pattern.
and Analysis), Position, Size, Scan Area Width and Tilt.
Chapter 4 Functional Checks4-9
Page 98
GE
SoftMenu Key
Primary Menu
Scan Area
B Mode
Zoom/Depth
Freeze
Secondary Menu
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-7B Mode Checks
4-3-7-1Preparations
1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-28, in Chapter 3 Installation
to the System probe connector.
2.) Turn ON the scanner (if it isn’t turned on already)
NOTE:The keyboard layout varies from each Product, the following figures show LOGIQ e R5.x.x as
an example.
Figure 4-7 Controls available in B Mode
Figure 4-8 B Mode Screen Picture Example
4-10Section 4-3 - General Procedure
Page 99
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-7-2B Mode OP Panel Controls
Table 4-3 B Mode Control Panel Controls
StepTaskExpected Result(s)Comments
1
Press B Mode keyB Mode Starts
Adjust the field of view. Increasing the depth
may view larger/deeper structures rates, and
2
3
4
Adjust Depth
Adjust Gain
Adjust Focus
decreasing the depth may view near the skin
line.Press Up/Down Button to increase/
decrease. Depth displays on the monitor in
cm.
Controls the amount of echo information
displayed in an image. Turn B Mode dial to
the left/right to increase/decrease Gain. Gain
displays on the monitor in Gn (dB).
Increases the number of focal zones or
moves the focal zone(s) to tighten up the
beam for specific area. Press the control to
toggle between Focus Position and Focus
Number. Press Up/Down Button to move or
adjust the focal numbers.
5
Activate Auto Optimize
7
Adjust Time Gain Compensation (TGC)
8
Adjust Scan Area
9
Adjust Zoom
10
Reverse
Optimize the image based upon a specified
region of interest or anatomy. Press the
Center Button in the Gain Dial to toggle the
ATO/ACE On and Off.
Amplifies the returning signals to correct for
the attenuation caused by tissues at
increasing depth. TGC slide pots spaced
proportional to the depth. Move the slide pots
to the left/right to decrease/increase TGC. A
TGC curve appears on the display.
Widen or narrow the size of the sector angle
to maximize the image’s region of interest
(ROI). Press Scan Area and move the
Trackball to narrow/widen the angle.
Changes the location of the focal point(s). A
triangular focus marker indicates the depth of
the focal point.
Toggles the left/right orientation of the scan
image.
Chapter 4 Functional Checks4-11
Page 100
GE
Direction 5370626-100, Revision 18 LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e Basic Service Manual
4-3-7-3B Mode Softmenu Key
Table 4-4 B Mode Softmenu Key
StepTaskExpected Result(s)comments
1
Adjust Rejection
2
Activate Colorize
Adjust Edge Enhance
3
Adjust Edge Enhance/Contour
4
Activate Gray Map
5
Adjust Frequency
Adjust Frame Average
6
Adjust DDP
Selects a level below which echoes will not be
amplified (an echo must have a certain
minimum amplitude before it will be processed).
Enables gray scale image colorization. To
deactivate, reselect a Gray Map.
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
Edge Enhance brings out subtle tissue
differences and boundaries by enhancing the
gray scale differences corresponding to the
edges of structures. Adjustments to M Mode's
edge enhancement affects the M Mode only.
Determines how the echo intensity levels
received are presented as shades of gray.
Multi Frequency mode lets you downshift to the
probe's next lower frequency or shift up to a
higher frequency.
Temporal filter that averages frames together.
This has the effect of presenting a smoother,
softer image.
Temporal filter that averages frames together.
This has the effect of presenting a smoother,
softer image.
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
For Vivid e
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
For Vivid e
7
Adjust Rotation
Adjust Line Density
8
Adjust Frame Rate
9
Power output
10
Dynamic Range
11
Focus Number and Position
12
Virtual ConvexVirtual Convex for linear probe
13
Virtual ApexVirtual Convex for Sector probe
Rotates the image by selecting the value from
the pop-up menu.
Optimizes B Mode frame rate or spatial
resolution for the best possible image.
Optimizes B Mode frame rate or spatial
resolution for the best possible image.
Optimizes image quality and allows user to
reduce beam intensity. 2% increments between
0-100%. Values greater than 0.1 are displayed.
Dynamic Range controls how echo intensities
are converted to shades of gray, thereby
increasing the adjustable range of contrast.
Increases the number of transmit focal zones or
moves the focal zone(s) so that you can tighten
up the beam for a specific area. A graphic caret
corresponding to the focal zone position(s)
appears on the right edge of the image.
4-12Section 4-3 - General Procedure
For LOGIQ e, LOGIQ e Vet and
LOGIQ i
For Vivid e
Only for LOGIQ e R5.x.x/R6.x.x/
R7.x.x, LOGIQ e Vet, LOGIQ i R5.x.x,
Vivid e R5.x.x, Vivid e R6.x.x