GE Logiq C5, LOGIQ C5 PRO Reference Manual

Page 1
Technical Publications
Direction 5264895-100 Rev. 1
LOGIQ C5/C5 PRO Advanced Reference Manual
Copyright© 2008 By General Electric Co.
Operating Documentation
Page 2
Regulatory Requirement
This product complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices
This manual is a reference for the LOGIQ C5/C5 PRO. It applies to R1.X.X and backward software version for the LOGIQ C5/C5 PRO ultrasound systems.
GE Healthcare
GE Medical Systems: Telex 3797371 P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A. (Asia, Pacific, Latin America, North America)
GE Ultraschall TEL: 49 212.28.02.208 Deutschland GmbH & Co. KG FAX: 49 212.28.02.431 Beethovenstraße 239 Postfach 11 05 60 D-42655 Solingen GERMANY
Page 3
Revision History
Table i-1: Reason for Change
REV DATE REASON FOR CHANGE
Rev. 1 July ,2008 Initial Release

List of Effective Pages

Table i-2: List of Effective Pages
REVISION
PAGE NUMBER
Title Page Rev. 1 2-1 thru 2-10 Rev. 1
Revision History i-1 and i-2 Rev. 1 3-1 thru 3-74 Rev. 1
Table of Contents ii-1 thru ii-2 Rev. 1 Index 1 Rev. 1
1-1 thru 1-46 Rev. 1
NUMBER PAGE NUMBER
REVISION
NUMBER
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on GPC (GE Medical Systems Global Product Configuration). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
LOGIQ C5/C5 PRO Advanced Reference Manual i-1 Direction 5264895-100 Rev. 1
Page 4
This page intentionally left blank.
i-2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 5

Table of Contents

Table of Contents
Chapter 1 — Acoustic Output
Bioeffects - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2
Concerns Surrounding the Use of Diagnostic Ultrasound - - - - - - - - - - - 1-2 Measurement Precision and Uncertainty - - - - - - - - - - - - - - - - - - - - - - - 1-3
FDA Acoustic Output Data - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4
Maximum output summary- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4 IEC Acoustic Output Tables - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-60
Chapter 2 — Measurement Formulas
Measurement Formulas - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -2-2
Formulas–Generic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2 Formulas–OB - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4 Formulas–GYN - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8 Formulas–Vascular - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-10
Chapter 3 — OB Tables
OB Tables- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
ASUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2 Berkowitz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3 Campbell - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4 Eriksen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4 French College of Fetal Ultrasound (CFEF) - - - - - - - - - - - - - - - - - - - 3-5 Goldstein - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-10 Hadlock - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-11 Hansmann - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-17 Hellman - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-25 Hill - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-25 Hohler - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-26 Jeanty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-26 JSUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-38 Kurtz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-42 Mayden - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-43 Mercer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-44 Merz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-45 Moore - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-55 Nelson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-55 Osaka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-56 Paris - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-60 Rempen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-63 Robinson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-68 Tokyo - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-68
LOGIQ C5/C5 PRO Advanced Reference Manual ii-1 Direction 5264895-100 Rev. 1
Page 6
Tokyo Shinozuka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-72 Yarkoni - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-78
Index - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-1
ii-2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 7

Chapter 1

Acoustic Output

Provides Acoustic Output information and tables, as well as possible bioeffects and prudent use.
LOGIQ C5/C5 PRO Advanced Reference Manual 1-1 Direction 5264895-100 Rev. 1
Page 8
Acoustic Output

Bioeffects

Concerns Surrounding the Use of Diagnostic Ultrasound

During a diagnostic ultrasound examination, high frequency sound penetrates and interacts with tissue in and around the area of anatomy to be imaged. Only a small portion of this sound energy is reflected back to the probe for use in constructing the image while the remainder is dissipated within the tissue. The interaction of sound energy with tissue at sufficiently high levels can produce biological effects (aka bioeffects) of either a mechanical or thermal nature. Although the generation of bioeffect is intentional with therapeutic ultrasound, it is generally undesired in diagnostic applications and may be harmful in some conditions.
NOTE: The American Institute of Ultrasound in Medicine has published
a document entitled "Medical Ultrasound Safety". This three part document covers Bioeffects and Biophysics, Prudent Use and Implementing ALARA.
Ultrasound users should read the AIUM to become more familiar with Ultrasound safety. A copy of this document is included as part of the documentation CD (5268173-200).
To contact the AIUM concerning their publications:
In the USA, by telephone at 1-800-638-5352.
To write them, use the following address:
AIUM
14750 Sweitzer Lane
Suite 100
Laurel, MD, USA 20707-5906
1-2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 9

Measurement Precision and Uncertainty

Precision and measurement uncertainty of the system used to measure the acoustic output is provided in the following table. These values are determined in accordance with Section 6.4 of the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, rev 2 AIUM/NEMA 2004, and the Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment, rev 3 NEMA
2004.
Table 1-1: Measurement Precision and Uncertainty
Center
Frequency
Acoustic
Power
Peak
Rarefactional
Pressure
Bioeffects
Acoustic Intensity
Measurement Uncertainty
+ 0.25% + 5.6% + 9.5% + 18.9%
LOGIQ C5/C5 PRO Advanced Reference Manual 1-3 Direction 5264895-100 Rev. 1
Page 10
Acoustic Output

FDA Acoustic Output Data

Maximum output summary

The following tables list the typical maimum acoustic output levels achievalbe with LOGIQ C5/C5 PRO for all probes and operational modes. It is intended that this is information be useful in making ALARA decisions and selecting the most appropriate probe for the application. In accordance with US FDA Guidelines, the overall maximum acoustic SPTA intensity for LOGIQ C5/C5 PRO is limited to 720 mW/cm limited to 1.9. Modes for which TI does not exceed 0.1 are indicated by <0.1. TI is limited to 6.0.
Table 1-2 shows all of the probes’ operating modes. The following pages contain probe acoustic output tables for those noted in Table 1-2.
2
and MI is
Table 1-2: Summary Table
Transducer Model:
Operatin
g Mode
B-Mode X X X X X X X X
M-Mode (inc. B­Mode)
Color flow X X X X X X X X
Power Doppler
Pulsed Doppler
Other (Specify)
E7C
E6C
-RC
-RC
XXX XX XXX
XXX XX XXX
XXX XX XXX
6.5C-RC4C-RC3.8C-RC3S-RC7.5L
-RC
7L­RC
1-4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 11
Maximum Probe Temperature
In accordance with IEC 60601-2-37 requirement, the maximum lens temperature is limited to 43 degC with thermal phantom, while 50 degC in air
Table Release Notes-2: Maximum Probe Temperature (Degrees C)
Probe Max Temp
FDA Acoustic Output Data
.
With TMM Phan-
tom*
E7C-RC 41.4 Dop 37.7 Dop
E6C-RC 42.2 CFM 40.5 CFM
6.5C-RC 37.7 Dop 37.9 Dop
4C-RC 41.2 CFM 43.1 CFM
3.8C-RC 38.9 CFM 41.3 CFM
7.5L-RC 38.8 Dop 37.4 Dop
7L-RC 38 CFM 35.5 CFM
3S-RC 40.3 B 42.1 B
Notes: Lens temperature measured under following conditions per IEC 60601-2-37: Thermocouple was placed at the geometric center of the lens. Thermal phantom* placed in 33 degC water bath for non-endo probes (all probes except (E6C-RC/E7C­RC). Thermal phantom* placed in 37 degC water bath for endo probes (E6C-RC/E7C-RC) Probe placed upright in contact with above thermal phantom. Autofreeze capability disabled. Lens temperature monitored for 30 min. Measurement uncertainty: 1.56 degC.
*Thermal phantom made with tissue-mimicking material as referenced in IEC60601-2-37: Appendix J.
Mode In Air Mode
LOGIQ C5/C5 PRO Advanced Reference Manual 1-5 Direction 5264895-100 Rev. 1
Page 12
Acoustic Output
Table Key
Table 1-1: Table Key
FDA IEC Meaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3
a α Acoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz)
A
aprt
D
eq
d
-6
d
eq
f
c
I
pa
I
pa.3
PII I
PII
.3
I
TA
(Z) I
I
TA. 3
(Z) I
I
SPTA
I
(Z) I
SPTA.3
A
arpt
C
MI
D
eq
d
-6
d
eq
ƒ
awf
I
pa
I
pa,α
pi
I
pi,α
Ita(z) Temporal-Average Intensity
(z) Attenuated Temporal-Average Intensity / (at depth z)
ta,α
(z) Spatial-Peak Temporal-Average Intensity
zpta
(z) Attenuated Spatial-Peak Temporal-Average Intensity
zpta,α
MI MI Mechanical Index
W
o
(Z) P
W
.3
W
o1
PII p
p
r
p
r.3
P Output Power / Time average acoustic power at the source
α
P
1
i
p
r
p
rα
PRF prr Pulse Repetition Rate / Pulse repetition frequency
TI TI Thermal Index / (same)
TIB TIB Bone Thermal Index / (same)
TIC TIC Cranial-Bone Thermal Index / (same)
TIS TIS Soft-Tissue Thermal Index / (same)
PD t
x
-12,y-12
d
X, Y -12 dB Output Beam Dimensions / (same)
Z Z Distance from the Source to a Specified Point / (same)
Z
sp
Z
bp
Z
sp
Z
b
Z
bp
Z
s
-12db Output Beam Area / Active aperture area
Normalizing Coefficient
Equivalent Aperture Diameter / (same)
Pulse Beam Width / Beam diameter at –6 dB
Equivalent Beam Diameter
Acoustic Working Frequency / Center frequency
Pulse-Average Intensity
Attenuated Pulse-Average Intensity
Pulse-Intensity Integral
Attenuated Pulse-Intensity Integral
Attenuated Output Power / Time average acoustic power derated to depth z
Bounded Output Power / Power emitted from the central 1cm of aperture
Pulse Pressure Squared Integral / Pulse intensity integral
Peak-Rarefactional Acoustic Pressure / (same)
Attenuated Peak-Rarefactional Acoustic Pressure / (same)
Pulse Duration / (same)
Depth for TIB / Depth at which the relevant index is maximum
Break-Point Depth / (same)
Depth for TIS / Depth at which the relevant index is maximum
1-6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 13
FDA Acoustic Output Data
Table 1-2: Table Key (in French 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-7 Direction 5264895-100 Rev. 1
Page 14
Acoustic Output
Table 1-3: Table Key (in French 2)
1-8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 15
FDA Acoustic Output Data
Table 1-4: Table Key (in German 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-9 Direction 5264895-100 Rev. 1
Page 16
Acoustic Output
Table 1-5: Table Key (in German 2)
1-10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 17
FDA Acoustic Output Data
Table 1-6: Table Key (in Italian 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-11 Direction 5264895-100 Rev. 1
Page 18
Acoustic Output
Table 1-7: Table Key (in Italian 2)
1-12 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 19
FDA Acoustic Output Data
Table 1-8: Table Key (in Brazilian Portuguese 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-13 Direction 5264895-100 Rev. 1
Page 20
Acoustic Output
Table 1-9: Table Key (in Brazilian Portuguese 2)
1-14 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 21
FDA Acoustic Output Data
Table 1-10: Table Key (in Danish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-15 Direction 5264895-100 Rev. 1
Page 22
Acoustic Output
Table 1-11: Table Key (in Danish 2)
1-16 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 23
FDA Acoustic Output Data
Table 1-12: Table Key (in Latin American Spanish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-17 Direction 5264895-100 Rev. 1
Page 24
Acoustic Output
Table 1-13: Table Key (in Latin American Spanish 2)
1-18 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 25
FDA Acoustic Output Data
Table 1-14: Table Key (in Norwegian 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-19 Direction 5264895-100 Rev. 1
Page 26
Acoustic Output
Table 1-15: Table Key (in Norwegian 2)
1-20 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 27
FDA Acoustic Output Data
Table 1-16: Table Key (in Swedish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-21 Direction 5264895-100 Rev. 1
Page 28
Acoustic Output
Table 1-17: Table Key (in Swedish 2)
1-22 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 29
FDA Acoustic Output Data
Table 1-18: Table Key (in Finnish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-23 Direction 5264895-100 Rev. 1
Page 30
Acoustic Output
Table 1-19: Table Key (in Finnish 2)
1-24 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 31
FDA Acoustic Output Data
Table 1-20: Table Key (in Japanese 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-25 Direction 5264895-100 Rev. 1
Page 32
Acoustic Output
Table 1-21: Table Key (in Japanese 2)
1-26 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 33
FDA Acoustic Output Data
Table 1-22: Table Key (in Chinese)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-27 Direction 5264895-100 Rev. 1
Page 34
Acoustic Output
4C-RC
Table 1-23: Transducer Mode 4C-RC Operating Mode: B Mode
Index Label MI
scan
TIS TIB
non-scan
non­scanAaprt< Aaprt>1
TIC
Global Maximum: Index Value 0.98
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.56
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 48.8 48.8
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 5.2
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.70 2.87 2.87
X(cm) 1.2 1.2
Y(cm)
t
d
PD (µsec) 0.47
prr PRF (Hz) 17
pr at max. Ipi pr@PII
(MPa) 2.16
max
0.67
1.3 1.3
0.37
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
@MI
max
FL
ma
(cm)
(cm) 2.75 2.75
X
(cm) 5.24 5.24
Y
(W/
cm
130.
2
6
)
Frequency (MHz) 3.0 3.0 3.0
Image Depth (cm) 7.0 7.0
Focus (cm) 6.4 5.12
7.0
5.12
1-28 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 35
FDA Acoustic Output Data
Table 1-24: Transducer Mode 4C-R
C Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.80 0.51 - 0.30 0.78 0.28
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
zs z1(cm) 3.1
zbp zbp(cm) 3.1
zb zsp(cm) 5.2
z at max. Ipi,a z
deq(zb) deq(zsp) (cm) 3.15
ƒ
awf
(MPa) 1.31
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
sp
f
c
(mW) 37.4 - 37.4 37.4
(z1)]) 40.45
(cm) 4.2
(MHz) 2.70 2.87 - 2.87 2.87 2.87
non-scan
Aaprt>
TIC
non-
1
scanAaprt<
Dim of A
t
d
aprt
PD (µsec) 0.86
X(cm) 1.2 - 6 6 1.2
Y(cm)
1.3 - 1.3 1.3 1.3
prr PRF (Hz) 513.9
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,a
x
PA. 3
(MPa) 2.13
max
(cm) 3
max
(cm) 2.75 - 2.75 2.75
X
(cm) 5.24 - 5.24 5.24
FL
Y
@MI
ma
(W/
cm
93.04
2
)
Frequency (MHz) THI5.24.0 - 4.0 4.0
Image Depth (cm) 17.0 17.0 - 17.0
Focus (cm) 6.4 4.1 - 4.1
17.0
4.1
LOGIQ C5/C5 PRO Advanced Reference Manual 1-29 Direction 5264895-100 Rev. 1
Page 36
Acoustic Output
Table 1-25: Transducer Mode 4C-R
C Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 0.83 1.26 0.8
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.52
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 106 106
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.0
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 3.3 2.5 2.5
X(cm) 6.0 6.0
Y(cm)
1.3 1.3
TIC
t
d
PD (µsec) 2.24
prr PRF (Hz) 7
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.13
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 11 .7 11. 7
(cm) 11. 9 11. 9
100.
2
9
)
Frequency (MHz) 2.5 2.5 2.5
Image Depth (cm) 8 10
Vel Scale (kHz) 2.3 1.6
ROI Max Max
PS 10 10
Focus (cm) 44.8 51.2
10
1.6
Max
10
51.2
1-30 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 37
FDA Acoustic Output Data
Table 1-26: Transducer Mode: 4C-R
C Operating Mode: Pulsed Doppler (Inc.B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 0.85 0.68 - 0.68 2.63 0.88
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.33
r.3
(mW) 57.6 - 57.6 57.6
.3(Z1),ITA. 3
ta,a(zs
o
)] [(W
(z1)]) 137.1
zs z1(cm) 5.4
zbp zbp(cm) 5.4
zb zsp(cm) 3.8 5.4
z at max. Ipi,a z
sp
(cm) 5.4
deq(zb) deq(zsp) (cm) 3.2
ƒ
awf
f
c
(MHz) 2.49 2.49 2.49 2.49 2.49 2.49
TIC
Dim of A
t
d
aprt
PD (µsec) 2.06
X(cm) 1.43 - 6.0 6.0 1.43
Y(cm)
1.3 - 1.3 1.3 1.3
prr PRF (Hz) 2604.1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 1.99
max
(cm) 3.2
max
(cm) -2.5 2.5
X
FL
(cm) -3.7 3.7
Y
@MI
max
(W/
cm
2
98.25
)
Frequency (MHz) 2.5 2.5 2.5 2.5 2.5 2.5
SVD (cm) 8.0 8.0 8.0 8.0 8.0 8.0
Vel Scale (kHz) 2.6 3.9 3.9 3.9 3.9 3.9
SV 4 4 - 4 4 4
Image Depth(cm) 10 10 10 10 10 10
LOGIQ C5/C5 PRO Advanced Reference Manual 1-31 Direction 5264895-100 Rev. 1
Page 38
Acoustic Output
E6C-RC
Table 1-27: Transducer Mode E6C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.83 0.58 0.50
IEC FDA Units
non-scan
non­scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 1.78
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.9
z at max. Ipi,a z
sp
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
X(cm) 0.81 0.81
Y(cm)
t
d
PD (µsec) 0.16
prr PRF (Hz) 35
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 2.41
max
(cm)
max
X
FL
Y
(mW) 27.6 27.6
(z1)])
(cm)
(MHz) 4.63 4.46 4.46
0.5 0.5
(cm) 0.81 0.81
(cm) 3.0 3.0
I
at max. MI I
pa,a
PA. 3
x
@MI
ma
(W/
cm
2
226
)
Frequency (MHz) 6.5 6.5 6.5
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
1-32 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.6
Page 39
FDA Acoustic Output Data
Table 1-28: Transducer Mode E
6C-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.81 0.53 0.53 0.63 0.45
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.73
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 24.9 24.9 24.9 24.9
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.9
z at max. Ipi,a z
sp
(cm) 1.4
deq(zb) deq(zsp) (cm) 1.13
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.62 4.49 4.49 - 4.49 4.49
X(cm) 0.81 2.5 - 2.5 0.81
Y(cm)
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.31
prr PRF (Hz) 522.4
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz)
Image Depth (cm) 4.0 4.0 4.0 4.0
Focus (cm) 2.0 1.6 1.6 1.6
(MPa) 3.51
max
(cm) 1.13
max
(cm) 0.81 2.5 2.5 0.81
X
FL
(cm) 3.0 3.0 3.0 3.0
Y
@MI
max
(W/
cm
226.6
2
)
6.5 6.5 6.5 6.5
6.5
4.0
1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-33 Direction 5264895-100 Rev. 1
Page 40
Acoustic Output
Table 1-29: Transducer Mode E
6C-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.76 0.88 0.68
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.68
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 37.6 37.6
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.7
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.86 4.91 4.91
X(cm) 0.81 0.81
Y(cm)
0.5 0.5
TIC
t
d
PD (µsec) 0.70
prr PRF (Hz) 7
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 2.23
max
(cm)
max
Focal Length FLX(cm) 0.81 0.81
FL
(cm) 3.0 3.0
Y
at max. MI I
I
pa,a
PA. 3
@MI
max
(W/
cm
103.7
2
)
Frequency (MHz) 5.0 5.0 5.0
Image Depth (cm) 9.0 9.0
Vel Scale (kHz) 1.3 1.3
ROI Max Max
PS 10 10
Focus (cm) 2.0 1.8
9.0
1.3
Max
10
1.8
1-34 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 41
FDA Acoustic Output Data
Table 1-30: Transducer Mode: E
6C-RC Operating Mode: Pulsed Doppler (Inc.B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.66 0.86 0.86 - 3.13 1.07
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.46
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 36.2 36.2 36.2 36.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.9 1.9
z at max. Ipi,a z
sp
(cm) 1.9
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.83 4.98 4.98 - 4.98 4.98
X(cm) 0.81 2.5 - 2.5 0.81
Y(cm)
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.47
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.01
max
(cm) 0.63
max
(cm) 0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
ma
(W/
cm
79.6
2
)
3
Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0
SVD (cm) 2.8 2.8 2.8 - 2.8 2.8
Vel Scale (kHz) 1.3 3.9 3.9 - 3.9 3.9
SV 1 3 3 - 3 3
Image Depth(cm) 9 9 9 9 9
LOGIQ C5/C5 PRO Advanced Reference Manual 1-35 Direction 5264895-100 Rev. 1
Page 42
Acoustic Output
E7C-RC
Table 1-31: Transducer Mode E7C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.32 0.56 0.41
IEC FDA Units
non-scan
non­scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 2.83
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 23.5 23.5
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.63 4.96 4.96
X(cm) 0.81 0.81
Y(cm)
t
d
PD (µsec) 0.40
prr PRF (Hz) 25
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.48
max
(cm)
max
(cm) 0.81 0.81
X
0.5 0.5
(cm) 3.0 3.0
FL
Y
I
at max. MI I
pa,a
PA. 3
x
@MI
ma
(W/
cm
244.1
2
)
Frequency (MHz) THI10 8 8
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
4.0
1.6
1-36 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 43
FDA Acoustic Output Data
Table 1-32: Transducer Mode E7C
-RC Operating Mode: M Mode (Inc. B-Mode)
Index Label MI
scan
Aaprt< Aaprt>1
Global Maximum: Index Value
IEC FDA Units
pra p
P W
min of [P
α(zs
, I
(MPa)
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA. 3
(mW)
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm)
z at max. Ipi,α z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz)
X(cm)
0.99 0. 57 0 .57
2.15
23.9 23.9 23.9 23.9
1.4
4.66 4.96 4.96
0.81 2.5
TIS TIB
non-scan
non­scan
-
-
-
0.69 0.42
1.4
1.13
4.96 4.96
2.5 0.81
TIC
Y(cm)
t
d
prr PRF (Hz)
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,α
Frequency (MHz)
Image Depth (cm)
PD (µsec)
(MPa)
max
(cm)
max
(cm)
X
FL
(cm)
Y
x
PA. 3
@MI
ma
(W/
cm
2
)
0.39
524.1
2.64
132.2
THI10 8 8
4.0 4.0 4.0
Focus (cm) 2.0
0.5 0.5
0.81 2.5
3.0 3.0
1.6 1.6
-
0.5 0.5
1.13
-
-
-
-
-
0.81
3.0
88
4.0 4.0
1.6 1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-37 Direction 5264895-100 Rev. 1
Page 44
Acoustic Output
Table 1-33: Transducer Mode E7C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.00 0.94 0.71
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.21
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 40.1 40.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.90 4.93 4.93
X(cm) 0.81 0.81
Y(cm)
0.5 0.5
TIC
t
d
PD (µsec) 0.62
prr PRF (Hz) 2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.81
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 0.81 0.81
(cm) 3.0 3.0
205.
2
)
5
Frequency (MHz) 5.0 5.0 5.0
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 1.3 1.4
ROI Max Max
PS 10 10
Focus (cm) 1.8 1.6
4.0
1.4
Max
10
1.6
1-38 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 45
FDA Acoustic Output Data
Table 1-34: Transducer Mode: E7C
-RC Operating Mode: Pulsed Doppler (inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.92 0.64 0.64 - 2.68 0.95
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.04
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 26.9 26.9 26.9 26.9
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.5 1.8
z at max. Ipi,a z
sp
(cm) 1.8
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.95 4.98 4.98 - 4.98 4.98
X(cm) 0.81 2.5 - 2.5 0.81
Y(cm)
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 1.07
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 2.64
max
(cm) 0.63
max
(cm) 0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
max
(W/
cm
155.6
2
)
Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0
SVD (cm) 2.0 2.0 2.0 - 2.0 2.0
Vel Scale (kHz) 2.0 2.0 2.0 - 2.0 2.0
SV 2 3 3 - 3 3
Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-39 Direction 5264895-100 Rev. 1
Page 46
Acoustic Output
7L-RC
Table 1-35: Transducer Mode 7L-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.83 0.71 0.40
IEC FDA Units
non-scan
non-
scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 2.11
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 22.4 22.4
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 6.47 6.64 6.64
X(cm) 3.84 3.84
Y(cm)
t
d
PD (µsec) 0.19
prr PRF (Hz) 92
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 2.69
max
(cm)
max
(cm) 3.84 3.84
X
0.4 0.4
(cm) 2.3 2.3
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
133.6
2
)
Frequency (MHz) THI12 11 11
Image Depth (cm) 4.0 4.0
Focus (cm) 1.6 1.2
1-40 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.2
Page 47
FDA Acoustic Output Data
Table 1-36: Transducer Mode 7L-RC
Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 0.83 0.52 0.52 - 0.33 0.29
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.21
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 16.2 16.2 16.2 16.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm) 1.2
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 6.48 6.78 6.78 - 6.78 6.78
X(cm) 3.84 0.83 - 0.83 3.84
Y(cm)
0.4 0.4 - 0.4 0.4
TIC
t
d
PD (µsec) 0.19
prr PRF (Hz) 516.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz) THI1
(MPa) 2.71
max
(cm) 0.63
max
(cm) 3.84 3.84 - 3.84 3.84
X
FL
(cm) 2.3 2.3 - 2.3 2.76
Y
@MI
max
(W/
cm
132.
2
)
5
11 11 - 11 11
2
Image Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
Focus (cm) 1.6 1.2 2.2 - 1.2 1.2
LOGIQ C5/C5 PRO Advanced Reference Manual 1-41 Direction 5264895-100 Rev. 1
Page 48
Acoustic Output
Table 1-37: Transducer Mode 7L-RC
Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.09 0.78 0.59
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.44
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 32.8 32.8
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.98 5.00 5.00
X(cm) 3.84 3.84
Y(cm)
0.4 0.4
TIC
t
d
PD (µsec) 0.64
prr PRF (Hz) 11
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.94
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 3.84 3.84
(cm) 2.3 2.3
280
2
)
Frequency (MHz) 5 5 5
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 3.0 3.0
ROI Max Min
PS 12 12
Focus (cm) 1.8 1.8
4.0
3.0
Min
12
1.8
1-42 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 49
FDA Acoustic Output Data
Table 1-38: Transducer Mode: 7L
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
n
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 0.73 0.73 0.73 - 3.07 1.09
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.62
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 30.5 30.5 30.5 30.5
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 2.1 2.3
z at max. Ipi,a z
sp
(cm) 2.3
deq(zb) deq(zsp) (cm) 0.35
ƒ
awf
f
c
(MHz) 4.97 4.99 4.99 - 4.99 4.99
TICsca
Dim of A
t
d
aprt
PD (µsec) 1.05
X(cm) 3.84 0.83 - 0.83 3.84
Y(cm)
0.4 0.4 - 0.4 0.4
prr PRF (Hz) 130
2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.32
max
(cm) 0.58
max
(cm) 3.84 0.83 - 3.84
X
FL
(cm) 2.3 2.3 - 2.3
Y
@MI
ma
(W/
cm
136.
2
6
)
Frequency (MHz) 5 5 5 - 5 5
SVD (cm) 3.0 3.0 3.0 - 3.0 3.0
Vel Scale (kHz) 1.3 2.6 2.6 - 2.6 2.6
SV 2 3 3 - 3 3
Depth(CM0 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-43 Direction 5264895-100 Rev. 1
Page 50
Acoustic Output
7.5L-RC
Table 1-39: Transducer Mode7.5L-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.28 0.65 0.51
IEC FDA Units
non-scan
non­scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 2.78
r.3
(mW) 28.5 28.5
.3(Z1),ITA. 3
ta,a(zs
o
)] [(W
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.3
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.75 4.80 4.80
X(cm) 3.84 2.5
Y(cm)
t
d
PD (µsec) 0.22
prr PRF (Hz) 77
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.45
max
(cm)
max
(cm) 3.84 3.84
X
0.4 0.4
(cm) 2.3 2.3
FL
Y
I
at max. MI I
pa,a
x
PA. 3
@MI
ma
(W/
cm
2
263.2
)
Frequency (MHz) THI12 6 6
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
1-44 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.6
Page 51
FDA Acoustic Output Data
Table 1-40: Transducer Mode 7.5
L-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 1.27 0.61 0.61 - 0.34 0.45
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.75
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 25.3 25.3 25.3 25.3
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.3
z at max. Ipi,a z
sp
(cm) 1.4
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.72 5.04 5.04 - 5.04 5.04
X(cm) 3.84 0.83 - 0.83 3.84
Y(cm)
0.4 0.4 - 0.4 0.4
TIC
t
d
PD (µsec) 0.21
prr PRF (Hz) 516.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz) THI1
(MPa) 3.40
max
(cm) 0.71
max
(cm) 3.84 3.84 - 3.84 3.84
X
FL
(cm) 2.3 2.3 - 2.3 2.3
Y
@MI
max
(W/
cm
271.
2
2
)
66 - 66
2
Image Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
Focus (cm) 2.0 1.6 1.6 - 1.6 1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-45 Direction 5264895-100 Rev. 1
Page 52
Acoustic Output
Table 1-41: Transducer Mode 7.5
L-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.24 0.91 0.69
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.75
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 38.5 38.5
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.93 4.95 4.95
X(cm) 3.84 3.84
Y(cm)
0.4 0.4
TIC
t
d
PD (µsec) 0.64
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 3.32
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 3.84 3.84
(cm) 2.3 2.3
392.
2
)
4
Frequency (MHz) 5.0 4 4
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 3.3 2.6
ROI Max Max
Packet Size 10 8
4.0
2.6
Max
8
Focus (cm) 1.8 1.8 1.8
1-46 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 53
FDA Acoustic Output Data
Table 1-42: Transducer Mode: 7.5
L-RC Operating Mode: Pulsed Doppler (inc. B-Mode)
Index Label MI
scan
Aaprt< Aaprt>1
Global Maximum: Index Value
IEC FDA Units
pra p
P W
min of [P
α(zs
, I
(MPa)
r.3
o
)] [(W
ta,α(zs
.3(Z1),ITA. 3
(mW)
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm)
z at max. Ipi,α z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz)
X(cm)
0.98 0.78 0.78
2.18
32.8 32.8 32.8 32.8
1.7
4.97 4.99 4.99
3.84 0.83
TIS TIB
non-scan
non­scan
-
-
-
3.54 1.25
2.0
2.0
0.35
4.99 4.99
0.833.84
TIC
Y(cm)
t
d
PD (µsec)
1.04
0.4 0.4
-
0.4 0.4
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,α
PA. 3
Frequency (MHz) 5.0
SVD (cm)
Vel Scale (kHz) 1.3
SV
max
max
@MI
(MPa)
(cm)
(cm)
X
FL
(cm)
Y
(W/
max
cm
2.93
245.
2
)
3
2.6 2.6 2.6
2
0.58
3.84 0.83
2.3 2.3
5.0 5.0
2.6 2.6
-
-
-
-
-
3.84
2.3
5.0 5.0
2.6 2.6
2.6 2.6
33 - 33
Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-47 Direction 5264895-100 Rev. 1
Page 54
Acoustic Output
3S-RC
Table 1-43: Transducer Mode 3S-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.17 0.69 1.24
IEC FDA Units
non-scan
non-
scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 1.5
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 82.1 82.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.3
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 1.66 1.76 1.76
X(cm) 0.9 0.6
Y(cm)
t
d
PD (µsec) 1.16
prr PRF (Hz) 27
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 1.81
max
(cm)
max
(cm) 0.9 0.9
X
1.2 1.2
(cm) 4.8 4.8
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
Frequency (MHz) THI
315.
2
15
)
3.2 3.2
3.2
Image Depth (cm) 30.0 30.0
Focus (cm) 3.2 3.2
30.0
3.2
1-48 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 55
FDA Acoustic Output Data
Table 1-44: Transducer Mode 3S
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 1.17 0.69 - 0.69 1.28 1.39
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.5
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 82.1 - 82.1 82.1
(z1)]) 2.2
zs z1(cm) 2.2
zbp zbp(cm) 2.2
zb zsp(cm) 3.3
z at max. Ipi,a z
sp
(cm) 2.2
deq(zb) deq(zsp) (cm) 1.48
ƒ
awf
Dim of A
aprt
f
c
(MHz) 1.66 1.76 - 1.76 1.76 1.76
X(cm) 0.9 - 1.9 1.9 1.9
Y(cm)
4.8 - 4.8 4.8
TIC
t
d
PD (µsec) 1.16
prr PRF (Hz) 113
4.3
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 1.81
max
(cm) 1.48
max
Focal Length FLX(cm) 0.9 - 1.9 0.9
FL
(cm) 4.8 - 4.8 4.8
Y
at max. MI I
I
pa,a
PA. 3
x
@MI
ma
(W/
cm
Frequency (MHz) THI
315.
2
15
)
3.2 - 3.2 3.2
3.2
Image Depth (cm) 30.0 30.0 - 30.0
Focus (cm) 3.2 3.2 - 3.2
30.0
3.2
LOGIQ C5/C5 PRO Advanced Reference Manual 1-49 Direction 5264895-100 Rev. 1
Page 56
Acoustic Output
Table 1-45: Transducer Mode 3S
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.72 1.23 2.23
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 0.86
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 146.8 146.
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
f
c
(MHz) 1.96 1.76 1.76
TIC
8
Dim of A
t
d
aprt
PD (µsec) 1.31
X(cm) 0.9 0.9
Y(cm)
1.2 1.2
prr PRF (Hz) 17
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 1.09
max
(cm)
max
Focal Length FLX(cm) 0.9 0.9
(cm) 4.8 4.8
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
416.
2
93
)
Frequency (MHz) 2.0 2.0 2.0
Image Depth (cm) 15.0 15.0
Vel Scale (kHz) 4.1 4.1
Penet Mid Mid
ROI 8 8
PS 3.52 3.52
15.0
4.1
Mid
8
3.52
1-50 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 57
FDA Acoustic Output Data
Table 1-46: Transducer Mode: 3S
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.89 0.53 0.53 - 2.27 0.77
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.26
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 44.2 44.2 44.2 44.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.8 4.2
z at max. Ipi,a z
sp
(cm) 4.2
deq(zb) deq(zsp) (cm) 1.44
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.0 2.51 2.51 - 2.51 2.51
X(cm) 0.9 1.9 - 1.9 0.9
Y(cm)
1.2 1.2 - 1.2 1.2
TIC
t
d
PD (µsec) 0.78
prr PRF (Hz) 113
2.2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 1.42
max
(cm) 1.44
max
(cm) 0.9 - 0.9
X
FL
(cm) 4.8 - 4.8
Y
@MI
ma
(W/
cm
273.
2
48
)
Frequency (MHz) 2.0 2.0 2.0 - 2.0 2.0
SVD (cm) 2.4 2.4 2.4 - 2.4 2.4
Vel Scale (kHz) 1.1 1.1 1.1 - 1.1 1.1
SV 1 1 1 - 11
LOGIQ C5/C5 PRO Advanced Reference Manual 1-51 Direction 5264895-100 Rev. 1
Page 58
Acoustic Output
3.8C-RC
Table 1-47: Transducer Mode 3.8C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.76 0.56 0.28
IEC FDA Units
non-scan
non­scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 1.24
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 36.4 36.4
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.68 3.1 3.1
X(cm) 1.2 1.2
Y(cm)
t
d
PD (µsec) 0.46
prr PRF (Hz) 4 8
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 1.97
max
(cm)
max
(cm) 12.
X
1.3 1.3
0.9
75
FL
(cm) 5.24 4.8
Y
I
at max. MI I
pa,a
Frequency (MHz) THI
PA. 3
@MI
max
(W/
cm
67.5
2
5
)
4.0 4.0
5.2
Image Depth (cm) 17 17
Focus (cm) 6.4 4.1
17
4.1
1-52 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 59
FDA Acoustic Output Data
Table 1-48: Transducer Mode 3.8C
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value .076 0.64 - 0.26 0.72 0.32
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.25
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 41.8 - 41.8 41.8
(z1)]) 82.83
zs z1(cm) 3.1
zbp zbp(cm) 3.1
zb zsp(cm) 5
z at max. Ipi,a z
sp
(cm) 3.1
deq(zb) deq(zsp) (cm) 2.8
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.67 3.23 - 3.23 3.23 3.23
X(cm) 1.2 - 6 6 1.2
Y(cm)
1.3 - 1.3 1.3 1.3
TIC
t
d
PD (µsec) 0.46
prr PRF (Hz) 513.
9
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
Frequency (MHz) THI
(MPa) 1.98
max
(cm) 2.8
max
(cm) 2.75 - 2.75 2.75
X
FL
(cm) 5.24 - 5.24 5.24
Y
@MI
ma
(W/
cm
315.
2
15
)
4.0 - 4.0 4.0
5.2
Image Depth (cm) 17 17 - 17
Focus (cm) 6.4 4.1 - 4.1
17
4.1
LOGIQ C5/C5 PRO Advanced Reference Manual 1-53 Direction 5264895-100 Rev. 1
Page 60
Acoustic Output
Table 1-49: Transducer Mode 3.8C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.87 1.41 0.9
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.38
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 117 11 7
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 4.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.51 2.52 2.52
X(cm) 6.0 6.0
Y(cm)
1.3 1.3
TIC
t
d
PD (µsec) 1.36
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 2.03
max
(cm)
max
X
FL
Y
@MI
(W/
max
cm
(cm) 11 .7 11.7
(cm) 11.9 11 .9
112.
2
7
)
Frequency (MHz) 2.5 2.5 2.5
Image Depth (cm) 10 10
Vel Scale (kHz) 2.0 2.0
Penet Max Max
ROI 10 12
PS 60.8 60.8
10
2.0
Max
12
60.8
1-54 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 61
FDA Acoustic Output Data
Table 1-50: Transducer Mode:3.8C
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non­scan
Global Maximum: Index Value 0.74 0.93 - 0.93 3.95 1.26
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.17
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 77.9 - 77.9 77.9
(z1)]) 116.3
zs z1(cm) 4.8
zbp zbp(cm) 4.8
zb zsp(cm) 4.1 4.8
z at max. Ipi,a z
sp
(cm) 4.8
deq(zb) deq(zsp) (cm) 2.53
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.49 2.51 - 2.51 2.51 2.51
X(cm) 1.43 - 6.0 6.0 6.0
Y(cm)
1.3 - 1.3 1.3 1.3
TIC
t
d
PD (µsec) 2.05
prr PRF (Hz) 320
5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 1.66
max
(cm) 2.53
max
(cm) -2.5 2.5
X
FL
(cm) -2.7 3.7
Y
@MI
ma
(W/
cm
70.1
2
4
)
Frequency (MHz) 2.5 2.5 - 2.5 2.5 2.5
SVD (cm) 5.6 5.6 - 5.6 5.6 5.6
Vel Scale (kHz) 3.2 3.2 - 3.2 3.2 3.2
SV 4 3 - 3 3 3
Image Depth(cm) 10 10 10 10 10
LOGIQ C5/C5 PRO Advanced Reference Manual 1-55 Direction 5264895-100 Rev. 1
Page 62
Acoustic Output
6.5C-RC
Table 1-51: Transducer Mode 6.5C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.11 1.50 0.45
IEC FDA Units
non-scan
non­scanAaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(zs
, I
(MPa) 2.40
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 25.1 25.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.65 4.2 4.2
X(cm) 0.62 0.62
Y(cm)
t
d
PD (µsec) 0.38
prr PRF (Hz) 58
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.06
max
(cm)
max
(cm) 0.62 0.62
X
0.5 0.5
(cm) 2.76 2.76
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
67.5
2
5
)
Frequency (MHz) THI106.5 6.5
Image Depth (cm) 4.0 4.0
Focus (cm) 1.6 42.4
4.0
2.4
1-56 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 63
FDA Acoustic Output Data
Table 1-52: Transducer Mode 6.5C
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.02 0.44 0.44 - 0.45 0.39
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 2.20
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 21.6 21.6 21.6 21.6
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.5
z at max. Ipi,a z
sp
(cm) 1.6
deq(zb) deq(zsp) (cm) 1.6
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.66 4.25 4.25 - 4.25 4.25
X(cm) 0.62 2.5 - 2.5 0.62
Y(cm)
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.39
prr PRF (Hz) 524.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.80
max
(cm) 1.6
max
(cm) 0.62 2.5 - 0.62
X
FL
(cm) 2.76 2.76 - 2.76
Y
@MI
ma
(W/
cm
188.
2
9
)
Frequency (MHz) THI106.5 6.5 - 6.5
Image Depth (cm) 4.0 4.0 4.0 -
Focus (cm) 2.4 2.4 2.4 -
4.0
2.4
LOGIQ C5/C5 PRO Advanced Reference Manual 1-57 Direction 5264895-100 Rev. 1
Page 64
Acoustic Output
Table 1-53: Transducer Mode 6.5C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.59 0.46 0.35
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.30
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 19.7 19.7
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 2
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.90 4.92 4.92
X(cm) 0.81 0.81
Y(cm)
0.5 0.5
TIC
t
d
PD (µsec) 0.63
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 1.82
max
(cm)
max
X
FL
Y
@MI
(W/
max
cm
(cm) 0.81 0.81
(cm) 3.0 3.0
79.3
2
3
)
Frequency (MHz) 5 5 5
Image Depth (cm) 6.0 6.0
Vel Scale (kHz) 2.3 1.9
ROI Max Max
Packet Size 14 14
Focus(cm) 2.4 2.24
6.0
1.9
Max
14
2.24
1-58 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 65
FDA Acoustic Output Data
Table 1-54: Transducer Mode:6.5C
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.90 0.67 0.67 - 3.17 0.95
IEC FDA Units
pra p
P W
min of [P
a(zs
, I
(MPa) 1.99
r.3
o
)] [(W
ta,a(zs
.3(Z1),ITA. 3
(mW) 28.5 28.5 28.5 28.5
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.8 2.2
z at max. Ipi,a z
sp
(cm) 2.2
deq(zb) deq(zsp) (cm) 0.75
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.95 4.96 4.96 - 4.96 4.96
X(cm) 0.81 2.5 - 2.5 0.81
Y(cm)
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 1.06
prr PRF (Hz) 196
5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.71
max
(cm) 0.75
max
(cm) -0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
ma
(W/
cm
185.
2
5
)
Frequency (MHz) 5 5 5 - 5 5
SVD (cm) 2.2 2.2 2.2 - 5.6 5.6
Vel Scale (kHz) 2.0 5.2 5.2 - 5.2 5.2
SV 2 2 2 - 2 2
Image Depth(cm) 6.0 6.0 6.0 - 6.0 6.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-59 Direction 5264895-100 Rev. 1
Page 66
Acoustic Output

IEC Acoustic Output Tables

4C-RC
Table 1-55: 4C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.16 2.13 2.93 1.99 1.99
(mW/cm2)
I
spta
(mm) 52 52 30 47 47
I
p
w
pb6
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
prr (kHz) - 513.9 - 2604.1 2604.1
srr (Hz) 17 513.9 7 - -
Output Beam Dim.
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
(MHz) 2.87 2.87 2.5 2.49 2.49
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
106.4 112.5 271.1 1174 1174
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode B B B B B
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
48.8 37.4 106 57.6 57.6
31.28 23.97 13.59 30.98 30.98
yes yes yes yes yes
1-60 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 67
FDA Acoustic Output Data
3.8C-RC
Table 1-56: 3.8C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 1.97 1.98 2.03 1.66 1.66
(mW/cm2)
I
spta
(mm) 50 50 44 41 41
I
p
w
pb6
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
prr (kHz) - 513.9 - 3205 3205
srr (Hz) 18 513.9 5 - -
Output Beam Dim.
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
(MHz) 3.1 3.23 2.3 2.51 2.51
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
87.4 92.7 96.37 1232 1232
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) Contact Contact Contact Contact Contact
I
ts
24.9 41.8 117 77.9 77.9
23.33 26.79 15 41.9 41.9
BB B BB
Yes Yes Yes Yes Yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-61 Direction 5264895-100 Rev. 1
Page 68
Acoustic Output
6.5C-RC
Table 1-57: 6.5C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.06 2.80 1.82 2.71 2.71
(mW/cm2)
I
spta
(mm) 15 15 20 18 18
I
p
w
pb6
• ( || ) (mm) 6.2 6.2 8.1 8.1 8.1
( ) (mm)5555 5
prr (kHz) - 524.1 - 5204 5204
srr (Hz) 58 524.1 5 - -
Output Beam Dim.
• ( || ) (mm) 6.2 6.2 8.1 8.1 8.1
( ) (mm)5555 5
(MHz) 4.2 4.25 4.92 4.96 4.96
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
102.1 106.5 106.7 1273 1273
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up ModeBBBB B
Initialization Mode
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) contact contact contact contact contact
I
ts
25.1 21.6 19.7 28.5 28.5
80.97 69.68 48.64 70.37 70.37
BBBB B
yesyesyesyes yes
1-62 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 69
FDA Acoustic Output Data
E6C-RC
Table 1-58: E6C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.41 2.34 2.23 2.01 2.01
I
(mW/cm2)
spta
I
(mm) 19 19 17 18 18
p
w
pb6
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
( ) (mm)55555
prr (kHz) - 501.1 - 3930 3930
srr (Hz) 35 524.1 7 - -
Output Beam Dim.
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
( ) (mm)55555
(MHz) 4.46 4.49 4.91 4.98 4.98
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
93.6 95.8 98.2 398.7 398.7
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode B B B B B
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
27.6 24.9 37.6 36.2 36.2
68.15 61.48 92.84 89.38 89.38
yes yes yes yes yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-63 Direction 5264895-100 Rev. 1
Page 70
Acoustic Output
E7C-RC
Table 1-59: E7C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.48 2.64 2.81 2.64 2.64
I
(mW/cm2)
spta
I
(mm) 14 14 14 15 15
p
w
pb6
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) (mm) 5 5 5 5 5
prr (kHz) - 524.1 - 2040 2040
srr (Hz) 25 524.1 5 - -
Output Beam Dim.
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) (mm) 5 5 5 5 5
(MHz) 4.96 4.96 4.93 4.98 4.98
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
87.5 91.3 1115.7 864.3 864.3
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode
Acoustic Output Freeze
I
(mm) 0.0 0.0 0.0 0.0 0.0
tt
I
(mm) contact contact contact contact contact
ts
23.5 23.9 40.1 26.9 26.9
58.02 59.01 99.01 66.42 66.42
BBBBB
yes yes yes yes yes
1-64 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 71
FDA Acoustic Output Data
3S-RC
Table 1-60: 3S-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 1.81 1.81 1.09 1.42 1.42
(mW/cm2)
I
spta
(mm) 33 33 35 18 18
I
p
w
pb6
( || ) (mm) 99999
• ( ) (mm) 12 12 12 12 12
prr (kHz) - 1134.3 - 1132.2 1132.2
srr (Hz) 27 1134.3 17 - -
Output Beam Dim.
( || ) (mm) 99999
• ( ) (mm) 12 12 12 12 12
(MHz) 1.76 1.76 1.76 2.51 2.51
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
281.12 259.2 184.5 457.3 457.3
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up ModeBBBBB
Initialization ModeBBBBB
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
82.1 82.1 146.8 44.2 44.2
76.02 76.02 135.9 40.93 40.93
yes yes yes yes yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-65 Direction 5264895-100 Rev. 1
Page 72
Acoustic Output
7L-RC
Table 1-61: 7L-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.96 2.71 2.94 2.32 2.32
(mW/cm2)
I
spta
(m m ) 11 11 11 21 21
I
p
w
pb6
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
• ( ) (mm) 4 4 4 4 4
prr (kHz) - 516.1 - 2604 2604
srr (Hz) 92 516.1 11 - -
Output Beam Dim.
• ( || ) (mm) 38.4 38.4 38.4 23.84 3.84
• ( ) (mm) 4 4 4 4 4
(MHz) 6.64 6.78 5.0 4.99 4.99
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
79.4 82.6 185.2 963.4 963.4
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) contact contact contact contact contact
I
ts
22.4 16.2 32.8 30.5 30.5
14.58 10.55 21.35 19.86 19.86
BBB BB
yes yes yes yes yes
1-66 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 73
FDA Acoustic Output Data
7.5L-RC
Table 1-62: 7.5L-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.45 3.49 3.32 3.93 2.93
(mW/cm2)
I
spta
(mm) 13 13 11 17 17
I
p
w
pb6
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
( ) (mm)4444 4
prr (kHz) - 516.1 - 2604 2604
srr (Hz) 77 516.1 5 - -
Output Beam Dim.
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
( ) (mm)4444 4
(MHz) 4.80 5.04 5.0 4.99 4.99
F
awf
APF (%) (Acoustic Power up Fraction)
AIF (%) (Acoustic Initialization Fraction)
92.3 96.8 125.2 1426 1426
100 100 100 100 100
100 100 100 100 100
Maximum Power (mW)
(mW/cm2)
I
ob
Power-up ModeBBBB B
Initialization Mode
Acoustic Output Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
28.5 25.3 38.5 32.8 32.8
18.55 16.47 25.07 21.35 21.35
BBBB B
yesyesyesyes yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-67 Direction 5264895-100 Rev. 1
Page 74
Acoustic Output
1-68 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 75

Chapter 2

Measurement Formulas

Reference for measurement formula tables for the various applications.
LOGIQ C5/C5 PRO Advanced Reference Manual 2-1 Direction 5264895-100 Rev. 1
Page 76

Measurement Formulas

Formulas–Generic
Table 2-1: Generic Calculation Formulas
Calc
Mnemonic
Calc Name Input Measurements Formula
Measurement Formulas
MaxPG Maximum Pressure
Gradient
MeanPG Mean Pressure Gradient flow velocities from one
% Stenosis Stenosis Ratio two areas (by ellipse,
PI Pulsatility Index two Doppler blood flow
RI Resistivity Index two Doppler blood flow
HR Heart Rate (beats/minute) one 2 beat time interval HR[BPM]=120[sec]/
A/B Ratio Velocities Ratio two Doppler blood flow
TAMAX Time Averaged Maximum
Velocity (Trace Method is Peak or manual)
TAMIN Time Averaged Minimum
Velocity (Trace method is Floor)
two Doppler blood flow peak velocities
time marker to another time marker in a Doppler display
trace, circle or distance)
peak velocities and TAM AX
peak velocities
peak velocities
two time marks in a Doppler display
two time marks in a Doppler display
MaxPG[mmHg]=4x(v1^2­v2^2)
MeanPG[mmHg]= n 4x (Vi^2)/n i=1
% Stenosis= [1-(A A
)]x100
lumen
PI=(V
max-Vdiastole
a
TAM AX
RI=(V
max-Vdiastole
2beat time [sec]
A/B=V
1/V2
TAMAX=sum{Vt} from t1 to t2/(t2-t1) [cm/s or m/s]
TAMIN=sum{V to t2/(t2-t1) [cm/s or m/s]
)/
)/V
} from t1
t
residual
max
/
a
TAMEAN Time Averaged Mean Velocity
(Trace method is Mean)
a.Vdiastole = Vmin or Vend-diastole (depends on preset selection)
two time marks in a Doppler display
TAMEAN=sum{V to t2/(t2-t1) [cm/s or m/s]
} from t1
t
2-2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 77
Measurement Formulas
Table 2-2: Volume Calculation Formulas
Calc Name Input Measurements Formula
Volume (spherical) one distance
Volume (prolate spheroidal) two distances, d1>d2
Volume (prolate spheroidal) one ellipse, d1 major axis, d2
minor axis
Volume (spheroidal) three distances
Volume (spheroidal) one distance d1, one ellipse, d2
ma jor axis, d3 minor axis
Vol [ml ]=(
Vol[ml]= (
Vol[ml]= (
Vol[ml]= (
Vol[ml]= (
π/6)xd^3
π/6)xd1xd2^2
π/6)xd1xd2^2
π/6)xd1xd2xd3
π/6)xd1xd2xd3
LOGIQ C5/C5 PRO Advanced Reference Manual 2-3 Direction 5264895-100 Rev. 1
Page 78
Measurement Formulas
Formulas–OB
Table 2-3: OB Calculation Formulas
Calc
Mnemonic
AC Abdominal
BPD Biparietal
CRL Crown Rump
FL Femur Length one distance FL = -3.91 + 0.427 (GA) -
GS Gestational Sac three distances GS [wk] = 1.42450142 *
HC Head
HC Head
Calc Name
Circumference
Diameter
Length
Circumference
Circumference
Input
Measurements
circumference by trace, ellipse, circle or two distances
one distance BPD = -3.08+0.41 (GA) -
one distance CRL=1.684969+
circumference by trace, ellipse, circle or two distances
one ellipse HC [mm] = 2.325 * (BPD
AC=13.3+1.61 (GA) -
0.00998 (GA)
0.000061 (GA)
0.315646xd1+
0.049306xd1^2+
0.004057xd1^3+
0.000120456xd1^4
0.0034 (GA)
(d1+d2+d3)/3+ 3.6225
HC=-11.48 + 1.56 (GA) -
0.0002548 (GA)
[mm]^2 + OFD [mm]^2)^0.5
Formula
2
3
2
3
Author
Reference
Hadlock et al, Radiology, 152:497-501, 1984
Hansmann, Ultrasound Diagnosis in Obstetrics and Gynecology, 438-9, 1985
EF Ejection
Fraction
two distances on M­Mode (End-diastolic dimension and End­systolic dimension on M-Mode)
EF = (1 - Ds^3 / Dd^3) n/a
2-4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 79
Table 2-4: CUA Hadlock Formulas
Measurement Formulas
Calc
Mnemonic
a
CUA
Calc
Name
Composite Ultrasound Age
Formula
1. CUA (BPD) = 9.54 + 1.482 * BPD + 0.1676 * BPD
2. CUA (HC) = 8.96 + 0.540 * HC + 0.0003 * HC
3. CUA (AC) = 8.14 + 0.753 * AC + 0.0036 * AC
4. CUA (FL) = 10.35 + 2.460 * FL + 0.170 * FL
5. CUA (BPD, HC) = 10.32 + 0.009 * HC2 + 1.3200 *
BPD + 0.00012 * HC
6. CUA (BPD, AC) = 9.57 + 0.524 * AC + 0.1220 * BPD
7. CUA (BPD, FL) = 10.50 + 0.197 * BPD * FL + 0.9500 * FL + 0.7300 * BPD
8. CUA (HC, AC) = 10.31 + 0.012 * HC
9. CUA (HC, FL) = 11.19 + 0.070 * HC * FL + 0.2630 * HC
10. CUA (AC, FL) = 10.47 + 0.442 *AC + 0.3140 * FL
0.0121 * FL
11. CUA (BPD, HC, AC) = 10.58 + 0.005 * HC2 + 0.3635 * AC + 0.02864 * BPD * AC
12. CUA (BPD, HC, FL) = 11.38+0.070 * HC * FL +
0.9800 * BPD
13. CUA (BPD, AC, FL) = 10.61 + 0.175 * BPD * FL +
0.2970 * AC + 0.7100 * FL
14. CUA (HC, AC, FL) = 10.33 + 0.031 * HC * FL +
0.3610 * HC + 0.0298 * AC * FL
15. CUA (BPD, HC, AC, FL) = 10.85 + 0.060 * HC * FL +
0.6700 * BPD + 0.1680 * AC
3
3
2
+ 0.3850 * AC
2
3
2
2
2
-
Author
Reference
Hadlock, Radiology, 1984 152:497­501
2
a.Formulas are used only if Hadlock HC, FL, AC and BPD are used and CUA is selected as the preset in the CUA/AUA for Hadlock preset in the System M&A Preset Menu. If other authors are used, CUA automatically changes to AUA and an average value is displayed.
Table 2-5: EFW Calculation Formulas
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
Calc Name
Weight
Weight
Input
Measurements
AC and HC EFW [kg] = 10^
(1.182+(0.0273*HC [cm])+(0.07057*AC [cm] ­(0.00063*AC [cm]^2)­(0.0002184*AC [cm] * HC [cm]))
BPD, AC, and FL EFW [g]=10^(1.335-
(0.0034*AC [cm] * FL [cm]) + (0.0316 * BPD [cm]) + (0.0457 * AC [cm]) + (0.1623 * FL [cm]
Formula
Author
Reference
Hadlock, Radiology, 150:535, 1984
Hadlock, AJOG, 151:333, 1985
LOGIQ C5/C5 PRO Advanced Reference Manual 2-5 Direction 5264895-100 Rev. 1
Page 80
Measurement Formulas
Table 2-5: EFW Calculation Formulas (Continued)
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
Calc Name
Weight
Weight
Weight
Weight
Weight
Input
Measurements
AC, HC, and FL EFW [g]=10^(1.326-
(0.00326 * AC [cm] * FL [cm]) + (0.0107 * HC [cm]) + (0.0438 * AC [cm]) + (0.158 * FL [cm]))
AC, HC, BPD, FL EFW [g] = 10^(1.3596-
(0.00386 * AC [cm] * FL [cm]) + (0.0064 * HC [cm]) + (0.00061 * BPD [cm] * AC [cm]) + (0.0424 * AC [cm]) + (0.174 * FL [cm]))
AC and FL EFW [g] = 10^(1.304 +
0.05281 * AC [cm] + 0.1938 * FL [cm] - 0.004 * AC [cm] * FL [cm])
EFW [g] = -3200.40479 +
157.07186 * AC [cm] +
15.90391 * BPD [cm]^2
EFW [g] = 0.515263 -
0.105775 * BPD [mm] + (0.000930707 * BPD [mm]^2 + 0.0649145 * TAD [mm] -
0.00020562 * TAD [mm]^2
Formula
Author
Reference
Hadlock, AJOG, 151:333, 1985
Hadlock, AJOG, 151:333, 1985
Hadlock, Radiology, 150:535, 1984
Merz
German
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
AC and BPD EFW [kg] = 10^(-1.7492 +
BPD [cm] and AC [cm]
BPD [cm] and AC [cm]
BPD, AxT, FL [cm] EFW1 [g] =1.07 * BPD
0.166 * BPD [cm] + 0.046 * AC [cm] - 2.646 * AC [cm] * BPD [cm]/1000)
EFW [g] = 10^(1.7288+0.09184 * BPD [cm] + 0.02581 * AC [cm] +
0.00011 * BPD [cm] * AC [cm])
EFW [g] =10^(3-1.7492 + (0.166 * BPD [cm]) + (0.04 * A [cm]) - (0.002646 * AC [cm] * BPD [cm]))
EFW [g] = 1.07 * BPD [cm]^3 + 3.42 * APTD [cm] * TTD [cm] * FL [cm]
[cm]^3 + 3.42 * AxT [cm2] * FL [cm]
Shepard, AJOG, 142:47, 1982
Shepard/ Warsoff
Richards/ Berkowitz
Tokyo University
Tokyo Shinozuka
2-6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 81
Measurement Formulas
Table 2-5: EFW Calculation Formulas (Continued)
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
Amniotic Fluid
Index (AFI)
Calc Name
Weight
Weight
The normal values are considered to be:
36-40 weeks
0-5 cm = very low
5.1-8.0 cm = low
8.1 - 18.0 cm = normal
>18.0 = high
Dr. Rutherford/Dr. Phelan, Obstetrics and Gynecology, Volume 70, No. 3, Part 1, p.353-6, Sept. 1987.
28-40 weeks
15.0 cm = average
Input
Measurements
BPD, AC, FL [cm] EFW2 [g] = 1.07 * BPD
[cm]^3 + 0.30 * AC [cm]^2 * FL [cm]
BPD, AxT, SL [cm] EFW3 [g] = 1.07 * BPD [cm]
^3 + 2.91 * AxT [cm2] * SL[cm]
Formula
Sagittal
Author
Reference
Tokyo Shinozuka
Tokyo Shinozuka
>20.0 - 24.0 = hydramnios
<5.0-6.0 = Oligohydramnios
Transverse
Dr. C.C. Smith, The Female Patient, Volume 15, p.85-97, March
1990.
LOGIQ C5/C5 PRO Advanced Reference Manual 2-7 Direction 5264895-100 Rev. 1
Page 82
Measurement Formulas
Formulas–GYN
Table 2-6: GYN Calculation Formulas
Calc
Mnemonic
UT-L Uterine Length one distance Ut-L[cm or mm]=d1
UT-H Uterine Height one distance Ut-H[cm or mm]=d1
UT-W Uterine Width one distance Ut-W[cm or mm]=d1
UT-Volume Uterine Volume
UtPFD Uterus Portio-Fundus
UtAP Anterior-Posterior Uterus
UtQ Transverse Uterus
Endo Endometrium Thickness one distance Endo[cm or mm]=d1
Lt. Ov-L Left Ovarian Length one distance Lt. Ov-L[cm or mm]=d1
Lt. Ov-H Left Ovarian Height one distance Lt. Ov-H[cm or mm]=d1
Lt. Ov-W Left Ovarian Width one distance Lt. Ov-W[cm or mm]=d1
Lt. Ov-Volume Left Ovarian Volume
Rt. Ov-L Right Ovarian Length one distance Rt. Ov-L[cm or mm]=d1
Calc Name
Distance
Diameter
Diameter
Input
Measurements
Formula
Rt. Ov-H Right Ovarian Height one distance Rt. Ov-H[cm or mm]=d1
Rt. Ov-W Right Ovarian Width one distance Rt. Ov-W[cm or mm]=d1
Rt. Ov-Volume Right Ovarian Volume
Lt. Ov-RI Left Ovarian Vessel
Resistive Index
Ut-RI Uterine Vessel Resistive
Index
Rt. Ov-RI Right Ovarian Vessel
Resistive Index
LtOvFo[ml] Left Ovary Follicles
two Doppler blood flow peak velocities
two Doppler blood flow peak velocities
two Doppler blood flow peak velocities
One distance
Two distances
Three distances
b
b
Lt. Ov-RI= (V
Ut-RI= (V
Rt. Ov-RI= (V
V
max
D1[cm]3 x π/6
D1[cm]2 x D2x[cm] π/6 : (D1 <D2)
D1[cm] x D2[cm] : (D2 <D1)
D1[cm] x D2[cm]x D3[cm] x
max-Vdiastole
a
max-Vdiastole
)/V
max-Vdiastole
2
x π/6
)/V
max
)/
max
a
π/6
a
2-8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 83
Measurement Formulas
Table 2-6: GYN Calculation Formulas
Calc
Mnemonic
RtOvFo[ml] Right Ovary Follicles
Lt. Ov-PI Left Ovarian Vessel
Rt. Ov-PI Right Ovarian Vessel
diastole
= V
min
a.V b.To calculate LtOvFo or RtOvFo with one (or two distances), press the Clear key after the first (or second distance) measurement(s).
Calc Name
Pulsatility Index
Pulsatility Index
or V
end-diastole
(depends on preset selection)
Input
Measurements
One distance
Two distances
Three distances
b
Formula
D1[cm]3 x π/6
b
D1[cm]2 x D2[cm] x π/6 : (D1 <D2)
D1[cm] x D2[cm] : (D2 <D1)
D1[cm] x D2[cm] x D3[cm] x
2
x π/6
π/6
LOGIQ C5/C5 PRO Advanced Reference Manual 2-9 Direction 5264895-100 Rev. 1
Page 84
Measurement Formulas
Formulas–Vascular
Table 2-7: Vascular Calculation Formulas
Calc
Mnemonic
Calc Name Input Measurements Formula
RT ECA Right External Carotid Artery
Vel oci ty
RT CCA Right Common Carotid Artery
Vel oci ty
RT BIFURC Right Carotid Bifurcation
Vel oci ty
RT ICA Right Internal Carotid Artery
Vel oci ty
RT ICA/CCA Right Internal Carotid Artery
Velocity/Common Carotid Artery
one Doppler blood flow peak velocity
one Doppler blood flow peak velocity
one Doppler blood flow peak velocity
one Doppler blood flow peak velocity
two Doppler blood flow peak velocities
Velocity Ratio
LT E C A , LT C C A ,
Same as above, for Left Carotid Artery
Same as above Same as above
LT BIFURC, LT I C A , LT ICA/CCA
A/B Ratio Velocities Ratio two Doppler blood flow
peak velocities
% Stenosis Stenosis Ratio two areas (by ellipse,
trace, circle or distance)
S/D Ratio Systolic Velocity/Diastolic
Velocities Ratio
two Doppler blood flow peak velocities
RT ECA=v1 [cm/s or m/s]
RT CCA=v1 [cm/s or m/s]
RT BIFURC=v1 [cm/s or m/s]
RT ICA=v1 [cm/s or m/s]
RT ICA/CCA=V V
CCA
A/B=V
1/V2
% Stenosis=[1­(A
S/D=V
/ A
residual
systole/Vdiastole
lumen
/
ICA
)]x100
a
PI Pulsatility Index two Doppler blood flow
peak velocities and TAM AX
RI Resistivity Index two Doppler blood flow
peak velocities
HR Heart Rate (beats/minute) one 2 beat time interval
(measured manually or
PI=(V
max-Vdiastole
a
TAM AX
RI=(V
max-Vdiastole
a
V
max
HR[BPM]=120[sec]/ 2 beat time[sec]
)/
)/
automatically)
a.V
diastole
= V
min
or V
end-diastole
(depends on preset selection)
2-10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 85

Chapter 3

OB Tables

Reference for Obstetric Measurement Tables.
LOGIQ C5/C5 PRO Advanced Reference Manual 3-1 Direction 5264895-100 Rev. 1
Page 86

OB Tables

OB Tables

ASUM

Table 3-1: AC: ASUM, Deler (Fetal Age)
Unit: AC (mm); Age (Days); 2SD (Days)
AC Age 2SD AC Age 2SD AC Age 2SD AC Age 2SD
<35 35 46 57 69 80 92 103 114
n/a 70 77 84 91 98 105 112 119
—— 8 8 8 8 9 9 9 9
126 137 149 160 171 183 194 206 217
126 133 140 147 154 161 168 175 182
10 10 10 10 10 10 12 12 12
228 240 251 263 274 285 297 308 320
189 196 203 210 217 224 231 238 245
14 14 14 14 14 16 16 18 18
331 342 354 365 377 >377
252 259 266 273 280 n/a
18 18 20 20 20 ——
Table 3-2: BPD: ASUM, Aust NZ, Obstet Gynaecol 1989: 29:26 (Fetal Age)
Unit: BPD (mm); Age (Week); 2SD (Week - * signifies No Data)
BPD Age 2SD BPD Age 2SD BPD Age 2SD BPD Age 2SD
<20 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
n/a 84 86 88 90 92 94 95 97 99 101 103 105 107 109 111 113 115 117 119 121
—— 4 4 4 4 5 5 5 5 5 6 6 6 6 7 7 7 7 8 8 8
40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
123 126 128 130 132 134 136 139 141 143 145 147 149 152 154 157 159 161 164 166 169
8 9 9 9 9 9 10 10 10 10 11 11 11 11 12 12 12 12 13 13 13
61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81
171 173 176 178 181 183 186 188 191 193 196 199 201 204 206 209 212 214 217 220 222
13 13 14 14 14 14 15 15 15 15 16 16 16 16 17 17 17 17 17 18 18
82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 >98
225 228 231 234 237 240 244 247 251 255 259 264 270 276 284 292 301 n/a
18 18 19 0 0 0 0 0 0 0 0 0 0 0 0 0 0 ——
3-2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 87
OB Tables
Table 3-3: CRL: ASUM, Silva et al 1991.6 (Fetal Age)
Unit: CRL (mm); Age (Days); 2SD (* No Data available)
CRL Age 2SD CRL Age 2SD CRL Age 2SD CRL Age 2SD
<2 2 3 4 5 6 7 8 9 10 11 12 13 14 15
n/a 42 43 44 45 46 47 48 49 50 51 52 53 54 55
—— * * * * * * * * * * * * * *
16 17 18 19 20 22 23 24 25 26 27 29 30 31 33
56 57 58 59 60 61 62 63 64 65 66 67 68 69 70
* * * * * * * * * * * * * * *
34 36 37 38 40 41 43 45 46 48 50 51 53 55 57
71 72 73 74 75 76 77 78 79 80 81 82 83 84 85
* * * * * * * * * * * * * * *
58 60 62 64 66 68 70 72 74 76 78 80 82 >82
86 87 88 89 90 91 92 93 94 95 96 97 98 n/a

Berkowitz

Table 3-4: BD: Berkowitz (Fetal Age)
Unit: BD (mm); Age (Day); SD (mm)
BD Age SD BD Age SD BD Age SD BD Age SD
<13 13 14 15 16 17 18 19 20 21 22 23 24
n/a 81 82 84 86 88 91 95 98 102 105 109 110
—— 0 0 0 0 0 0 0 0 0 0 0 0
25 26 27 28 29 30 31 32 33 34 35 36 37
112 116 120 124 128 130 132 135 138 142 145 149 152
0 0 0 0 0 0 0 0 0 0 0 0 0
38 39 40 41 42 43 44 45 46 47 48 49 50
155 159 162 166 169 173 179 185 191 197 202 207 212
0 0 0 0 0 0 0 0 0 0 0 0 0
51 52 53 54 55 56 57 58 59 >59
217 223 230 237 244 251 258 266 275 n/a
* * * * * * * * * * * * * ——
0 0 0 0 0 0 0 0 0 ——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-3 Direction 5264895-100 Rev. 1
Page 88
OB Tables

Campbell

Table 3-5: HC/AC Ratio: Campbell, Br J Obstet Gynaecol 1977, 84:165-174
(Fetal Growth)
Unit: GA (Weeks); Min/Max (Index)
GA Min Max GA Min Max GA Min Max
<13 13 15 17 19 21
n/a
1.14
1.05
1.07
1.09
1.06
——
1.31
1.39
1.29
1.26
1.25
23 25 27 29 31 33
1.05
1.04
1.05
0.99
0.96
0.96
1.21
1.22
1.22
1.21
1.17
1.11
35 37 39 41 >42
0.93
0.92
0.87
0.93 n/a
1.11
1.05
1.06
1.00 n/a

Eriksen

Table 3-6: TAD: Eriksen (Fetal Age)
Unit: TAD (mm); Age (Day); SD (mm)
TAD Age SD TA D Age SD TAD Age SD TAD Age SD
<23 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44
n/a 91 93 95 97 99 101 103 105 107 109 111 113 115 117 119 120 122 124 126 128 130 132
—— 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67
134 136 138 140 142 144 146 148 150 152 154 156 158 161 163 165 167 169 171 173 175 177 179
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90
182 184 186 188 190 192 195 197 199 201 203 206 208 210 212 214 217 219 221 223 225 228 230
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
91 92 93 94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 >112
232 234 236 239 241 243 245 247 250 252 254 256 259 261 263 266 268 270 273 275 277 280 n/a
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 ——
3-4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 89

French College of Fetal Ultrasound (CFEF)

Table 3-7: BPD
GA
(weeks)
11 12.08 13.12‘ 15.36 17.60 18.63
12 15.81 16.96 19.40 21.81 22.92
13 19.47 20.71 23.30 25.92 27.12
14 23.05 24.36 27.14 29.92 31.23
15 26.56 27.93 30.89 33.82 35.23
16 29.97 31.41 34.53 37.62 39.08
17 33.32 34.85 38.12 41.35 42.87
18 36.55 38.15 41.58 44.97 46.56
19 39.76 41.46 45.00 48.52 50.18
20 42.85 44.56 48.22 51.90 53.64
21 45.86 47.66 51.43 55.23 57.00
22 48.79 50.61 54.53 58.44 60.30
23 51.63 53.48 57.51 61.54 63.45
24 54.38 56.31 60.42 64.57 66.50
25 57.04 59.00 63.25 67.48 69.42
26 59.62 61.64 65.94 70.24 72.27
27 62.12 64.15 68.55 72.62 75.00
28 64.50 66.61 71.03 75.52 77.60
29 66.84 68.98 73.50 77.97 80.09
30 69.07 71.21 75.80 80.37 82.52
31 71.22 73.39 78.00 82.63 84.80
32 73.30 75.49 80.16 84.80 87.00
33 75.24 77.46 82.14 86.84 89.04
34 77.14 79.36 84.07 88.8 91.00
35 78.94 81.14 85.90 90.61 92.83
36 80.64 82.88 87.61 92.35 94.56
37 82.27 84.50 89.24 93.97 96.19
38 83.78 86.00 90.70 95.42 97.66
39 85.22 87.43 92.10 96.86 99.05
40 86.47 88.78 93.45 98.13 100.31
41 87.00 89.00 94.00 99.00 101.00
3% 10% 50% 90% 97%
OB Tables
LOGIQ C5/C5 PRO Advanced Reference Manual 3-5 Direction 5264895-100 Rev. 1
Page 90
OB Tables
Table 3-8: FL
GA
(weeks)
12 2.76 3.89 6.33 8.79 10.00
13 6.09 7.29 9.88 12.42 13.65
14 9.40 10.65 13.33 16.00 17.27
15 12.56 13.87 16.66 19.44 20.77
16 15.70 17.00 19.95 22.80 24.18
17 18.74 20.12 23.12 26.13 27.53
18 21.69 23.14 26.23 29.30 30.80
19 24.59 26.06 29.25 32.44 33.91
20 27.42 28.94 32.23 35.48 37.03
21 30.12 31.72 35.05 38.41 40.00
22 32.83 34.39 37.87 41.30 42.91
23 35.34 37.00 40.50 44.03 45.71
24 37.89 39.58 43.16 46.75 48.42
25 40.33 42.04 45.69 49.36 51.08
26 42.66 44.40 48.17 51.88 53.62
27 44.95 46.72 50.53 54.32 56.09
28 47.13 48.94 52.80 56.64 58.45
29 49.22 51.06 54.94 58.91 60.72
30 51.30 53.14 57.13 61.08 62.92
31 53.26 55.13 59.15 63.14 65.04
32 55.12 57.04 61.11 65.19 67.07
33 56.96 58.87 63.00 67.10 69.03
34 58.69 60.62 64.76 68.88 70.84
35 60.33 62.29 66.47 70.65 72.63
36 61.90 63.89 68.13 72.34 74.30
37 63.40 65.36 69.63 73.91 75.89
38 64.81 66.79 71.11 75.38 77.41
39 66.16 68.19 72.48 76.81 78.84
40 67.42 69.47 73.79 78.14 80.17
41 68.00 70.00 74.00 79.00 81.00
3% 10% 50% 90% 97%
3-6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 91
OB Tables
Table 3-9: HC
GA
(weeks)
16 105.80 110.58 120.86 131.25 136.11
17 118.67 123.78 134.49 145.38 150.39
18 131.08 136.36 147.55 158.92 164.11
19 143.00 148.53 160.29 172.14 177.48
20 154.53 160.21 172.47 184.86 190.54
21 165.41 171.49 184.21 197.12 203.09
22 176.12 182.35 ‘195.74 208.91 215.15
23 186.32 192.31 206.64 220.26 226.76
24 196.19 203.00 217.18 231.39 238.00
25 205.50 212.40 227.32 241.91 248.81
26 214.44 221.57 236.72 252.00 259.23
27 222.87 230.33 246.00 261.75 269.13
28 231.00 238.56 254.77 271.00 278.57
29 238.40 246.35 263.00 279.71 287.56
30 245.86 253.74 270.84 288.13 296.00
31 252.54 260.81 278.33 296.00 304.27
32 258.86 267.22 285.29 303.54 312.00
33 264.62 278.38 292.00 310.40 319.10
34 270.14 279.00 298.10 317.00 325.91
35 275.33 284.23 303.62 323.00 332.16
36 279.79 289.00 308.81 328.75 338.00
37 283.90 293.32 313.52 334.00 343.34
38 287.63 297.29 317.88 338.64 348.29
39 290.88 300.76 321.86 343.00 352.67
40 293.00 303.00 324.00 346.00 356.00
3% 10% 50% 90% 97%
LOGIQ C5/C5 PRO Advanced Reference Manual 3-7 Direction 5264895-100 Rev. 1
Page 92
OB Tables
Table 3-10: AC
GA
(weeks)
15 80.70 85.30 95.00 104.40 108.80
16 91.30 96.10 106.40 116.80 121.60
17 101.70 106.80 118.00 129.00 134.00
18 111.80 117.40 129.20 141.00 146.60
19 122.00 128.00 140.40 153.00 158.80
20 132.00 138.00 151.40 164.70 171.00
21 141.60 148.20 162.30 176.30 183.00
22 151.40 158.20 173.00 187.80 194.70
23 160.90 168.20 183.60 199.00 206.30
24 170.20 177.80 194.00 210.30 218.00
25 179.30 187.30 204.40 221.30 229.30
26 188.40 196.70 214.50 232.30 240.60
27 197.30 206.00 224.50 243.00 251.60
28 206.20 215.10 234.40 253.60 262.60
29 214.70 224.00 244.00 264.00 273.30
30 223.20 232.80 253.60 274.20 283.70
31 231.60 241.60 263.00 284.20 294.40
32 239.70 250.00 272.20 294.30 304.60
33 247.80 258.40 281.20 304.00 314.80
34 255.60 266.70 290.20 313.80 324.80
35 263.20 274.70 298.80 323.30 334.50
36 271.00 282.60 307.40 332.50 344.30
37 278.30 290.30 316.00 341.70 353.80
38 285.60 298.00 324.70 350.70 363.00
39 292.70 305.30 332.40 359.60 372.20
40 298.00 311.00‘ 339.00 367.00 380.00
3% 10% 50% 90% 97%
3-8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 93
OB Tables
Table 3-11: TAD
GA
(weeks)
11 9.68 11.00 13.50 16.00 17.25
12 12.68 14.00 17.00 20.00 21.46
13 15.60 17.25 20.56 24.00 25.51
14 18.69 20.41 24.00 27.84 29.56
15 21.76 23.64 27.69 31.74 33.61
16 25.00 27.00 31.21 35.53 37.48
17 28.23 30.34 34.70 39.21 41.39
18 31.54 33.64 38.31 42.89 45.14
19 34.78 37.00 41.69 46.42 48.58
20 38.16 40.26 45.21 50.00 52.20
21 41.14 43.46 48.34 53.22 55.63
22 44.21 46.61 51.57 56.75 59.08
23 47.00 49.47 54.72 60.00 62.46
24 49.77 52.39 57.88 63.43 66.00
25 52.54 55.18 61.00 66.74 69.44
26 55.17 58.00 64.00 70.12 72.89
27 57.72 60.73 67.11 73.42 76.42
28 60.43 63.58 70.27 76.80 79.87
29 63.13 66.36 73.27 80.17 83.33
30 65.80 69.17 76.17 83.45 86.75
31 68.35 71.88 79.25 86.68 90.13
32 70.90 74.43 82.10 89.76 93.36
33 73.08 76.75 84.78 92.89 96.64
34 75.25 79.08 87.55 95.89 99.86
35 77.00 81.10 90.00 99.00 103.00
36 78.48 82.90 92.36 102.00 106.31
37 79.79 84.60 94.81 105.00 109.67
38 80.92 86.10 97.00 108.19 113.29
39 81.85 87.41 99.33 111.34 117.00
40 82.58 88.59 101.64 114.52 120.70
41 82.80 89.20 103.00 117.00 123.00
3% 10% 50% 90% 97%
LOGIQ C5/C5 PRO Advanced Reference Manual 3-9 Direction 5264895-100 Rev. 1
Page 94
OB Tables

Goldstein

Table 3-12: TCD: Goldstein et a, Am J OB/GYN, May 1987 (Fetal Growth)
Unit: TCD (Weeks); Age/Quat1/Mean/Quat3/Max (mm)
Age Min Quat1 Mean Quat3 Max
15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
10 14 16 17 18 18 19 21 22 22 23 25 26 27 29 31 32 33 32 33 31 36 37 40 52
12 16 17 18 18 19 20 23 23 24
21.5 28
28.5 30 32 32 35 36 36 38 37 29 37 40 52
14 16 17 18 19 20 22 23 24 25 28 29 30 31 34 35 38 38 40 40
40.5 43 45
48.5 52
15 16 18 19 19 20 23 24 25 27 28 30 31 32 36 37 39 40 43 41 43 52 52 52 55
16 17 18 19 22 22 24 24 26 28 29 32 32 34 38 40 43 42 44 44 47 55 55 55 55
3-10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 95
OB Tables

Hadlock

Table 3-13: AC: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: AC (mm); Age (Week); SD (Week)
AC Age SD AC Age SD AC Age SD AC Age SD
<50 50 55 60 65 70 75 80 85 90 95 100 105 110 115 120 123 124 125 130
n/a
12.0
12.4
12.8
13.2
13.6
14.0
14.4
14.8
15.2
15.6
16.0
16.4
16.9
17.3
17.7
17.9
18.0
18.1
18.5
—— ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 2.1 ± 2.1 ± 2.1
135 140 145 150 155 160 165 170 175 180 185 190 192 193 195 200 205 210 215 220
19.0
19.4
19.8
20.2
20.7
21.1
21.5
22.0
22.4
22.9
23.3
23.7
23.9
24.0
24.2
24.6
25.1
25.5
26.0
26.4
± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2
225 230 235 240 245 250 255 258 259 260 265 270 275 280 285 290 295 300 305 310
26.9
27.4
27.8
28.3
28.7
29.2
29.7
30.0
30.1
30.2
30.6
31.1
31.6
32.0
32.5
33.0
33.5
34.0
34.5
34.9
± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0
315 320 321 325 330 335 340 345 350 355 360 365 370 375 380 385 >385
35.4
35.9
36.0
36.4
36.9
37.4
37.9
38.4
38.9
39.4
39.9
40.4
40.9
41.4
42.0
42.5 n/a
± 3.0 ± 3.0 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ——
Table 3-14: AC: Hadlock, AJR; 139: 367-370; 1982 (Fetal Age)
Unit: AC (mm); Age (Days); SD (Days)
AC Age SD AC Age SD AC Age SD AC Age SD
<47 47 53 60 67 74 80 87 93 100 106 113 119 126 132
n/a 84 87 91 94 98 101 105 106 112 115 119 122 126 129
—— 13 13 13 13 13 13 13 13 13 13 13 13 13 14
138 144 151 157 163 174 180 186 192 197 203 208 214 219 225
133 136 140 143 147 154 157 161 164 168 171 175 178 182 185
14 14 14 14 14 14 14 14 14 14 15 15 15 15 15
230 235 241 246 251 256 261 266 271 276 281 286 291 296 300
189 192 196 199 203 206 210 213 217 220 224 227 231 234 238
15 15 15 15 15 15 15 21 21 21 21 21 21 21 21
305 310 314 319 323 328 332 337 341 344 349 353 >353
241 245 248 252 255 259 262 266 269 273 276 280 n/a
21 21 21 21 18 18 18 18 18 18 18 18 ——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-11 Direction 5264895-100 Rev. 1
Page 96
OB Tables
a
Table 3-15: BPD: Hadlock, Radiology 1984, Vol. 152:497
Unit: BPD (mm); Age (Week); SD (Week)
BPD Age SD BPD Age SD BPD Age SD BPD Age SD
(Fetal Age)
<14 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35
n/a
11. 9
12.1
12.3
12.5
12.8
13.0
13.2
13.4
13.6
13.8
14.1
14.3
14.5
14.8
15.0
15.2
15.5
15.7
16.0
16.3
16.5
16.8
a.Variability of GA estimate by BPD at term is ± 2 SD (6 weeks)
—— ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2
36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58
17.0
17.3
17.6
17.9
18.1
18.4
18.7
19.0
19.3
19.6
19.9
20.2
20.5
20.8
21.1
21.5
21.8
22.1
22.4
22.8
23.1
23.4
23.8
± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7 ± 1.7
59 60 61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81
24.1
24.5
24.8
25.2
25.5
25.9
26.3
26.6
27.0
27.4
27.7
28.1
28.5
28.9
29.3
29.7
30.1
30.5
30.9
31.3
31.7
32.1
32.5
± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 2.2 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1
82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 101 102 103 >103
Table 3-16: BPD: Hadlock, J Ultrasound Med 1:97-104, April 1982 (Fetal Age)
Unit: BPD (mm); Age (Days); SD (Days)
33.0
33.4
33.8
34.2
34.7
35.1
35.6
36.0
36.5
36.9
37.4
37.8
38.3
38.7
39.2
39.7
40.2
40.6
41.1
41.6
42.1
42.6 n/a
± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ± 3.2 ——
BPD Age SD BPD Age SD BPD Age SD BPD Age SD
<20 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39
n/a 85 88 90 92 93 95 97 99 102 103 105 107 109 111 113 116 118 120 122 124
—— 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
126 128 130 132 134 137 139 141 144 146 148 151 153 155 158 160 162 163 167 169 172
10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 9 9
61 62 63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81
175 177 180 183 185 188 190 193 196 198 201 204 207 209 213 216 218 221 224 228 230
9 9 9 9 9 9 9 9 9 9 9 9 9 9 14 14 14 14 14 14 14
82 83 84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 100 >100
233 237 239 243 246 249 253 256 259 263 266 270 272 276 279 284 287 291 294 n/a
14 14 14 14 14 14 25 25 25 25 25 25 25 25 25 25 25 25 25 ——
3-12 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 97
OB Tables
Table 3-17: CI: Hadlock, AJR, 137: 83, 1981 (Fetal Growth)
Min (Index) Max (Index)
70 86
Table 3-18: CRL: Hadlock, Radiology 1992, Vol. 182:501 (Fetal Age)
Unit: CRL (mm); Age (Week); SD (Week)
CRL Age SD CRL Age SD CRL Age SD CRL Age SD
<2 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
n/a
5.7
5.9
6.1
6.2
6.4
6.6
6.7
6.9
7.1
7.2
7.4
7.5
7.7
7.9
8.0
8.1
8.3
8.4
8.6
8.7
8.9
9.0
9.1
9.2
9.4
9.5
9.6
9.7
9.9
10.0
—— ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.3 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4 ± 0.4
32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62
10.1
10.2
10.3
10.4
10.5
10.6
10.7
10.8
10.9
11. 0
11. 1
11. 2
11. 2
11. 3
11. 4
11. 5
11. 6
11. 7
11. 7
11. 8
11. 9
12.0
12.0
12.1
12.2
12.3
12.3
12.4
12.5
12.6
12.6
± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.5 ± 0.6 ± 0.6 ± 0.6 ± 0.6
63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 84 85 86 87 88 89 90 91 92 93
12.7
12.8
12.8
12.9
13.0
13.1
13.1
13.2
13.3
13.4
13.4
13.5
13.6
13.7
13.8
13.8
13.9
14.0
14.1
14.2
14.2
14.3
14.4
14.5
14.6
14.7
14.8
14.9
15.0
15.1
15.2
± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.6 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7
94 95 96 97 98 99 100 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 >121
15.3
15.3
15.4
15.5
15.6
15.7
15.9
16.0
16.1
16.2
16.3
16.4
16.5
16.6
16.7
16.8
16.9
17.0
17.1
17.2
17.3
17.4
17.5
17.6
17.7
17.8
17.9
18.0 n/a
± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.7 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ± 0.8 ——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-13 Direction 5264895-100 Rev. 1
Page 98
OB Tables
Table 3-19: EFW: Hadlock (Fetal Age)
Unit: EFW (grams); Mean (Weeks); SD (grams)
EFW Mean SD EFW Mean SD EFW Mean SD
<35 35 45 58 73 93 117 146 181 223 273 331
n/a 10 11 12 13 14 15 16 17 18 19 20
—— 4 6 7 9 12 15 19 23 28 35 42
399 478 568 670 785 913 1055 1210 1379 1559 1751 1953
21 22 23 24 25 26 27 28 29 30 31 32
51 61 72 85 101 116 134 154 175 198 222 248
2162 2377 2595 2813 3028 3236 3435 3619 3787 >3787
33 34 35 36 37 38 39 40 41 n/a
275 302 330 357 385 411 436 460 481 ——
Table 3-20: FL: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: FL (mm); Age (Week); SD (Week)
FL Age SD FL Age SD FL Age SD FL Age SD
<6 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
n/a
11. 9
12.2
12.4
12.7
13.0
13.3
13.5
13.8
14.1
14.4
14.7
15.0
15.3
15.6
16.0
16.3
16.6
16.9
17.2
—— ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4 ± 1.4
25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44
17.6
17.9
18.2
18.6
18.9
19.3
19.6
20.0
20.3
20.7
21.0
21.4
21.8
22.2
22.5
22.9
23.3
23.7
24.1
24.5
± 1.4 ± 1.4 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 1.8 ± 2.1 ± 2.1
45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64
24.9
25.3
25.7
26.1
26.5
26.9
27.3
27.7
28.2
28.6
29.0
29.5
29.9
30.3
30.8
31.2
31.7
32.1
32.6
33.1
± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0
65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 80 81 82 83 >83
33.5
34.0
34.5
34.9
35.4
35.9
36.4
36.9
37.4
37.9
38.4
38.9
39.4
39.9
40.4
40.9
41.4
42.0
42.5 n/a
± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ± 3.1 ——
3-14 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 99
OB Tables
Table 3-21: FL: Hadlock, AJR 138: 875-878, May 1982 (Fetal Age)
Unit: FL (mm); Age (Days); SD (Days)
FL Age SD FL Age SD FL Age SD FL Age SD
<10 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
n/a 90 92 94 95 97 99 101 104 106 108 110 112 114 116 118 120 123
—— 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6
27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44
125 127 130 132 134 137 139 142 145 147 150 153 155 157 160 163 166 169
6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 6 10
45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62
171 174 177 180 183 185 189 192 195 197 201 204 207 210 213 216 220 223
10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10
63 64 65 66 67 68 69 70 71 72 73 74 75 76 77 78 79 >79
226 230 233 237 239 243 246 250 253 257 260 264 268 272 275 279 283 n/a
10 10 10 10 10 10 10 10 11 11 11 11 11 11 11 11 11 ——
Table 3-22: HC: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: HC (mm); Age (Week); SD (Week)
HC Age SD HC Age SD HC Age SD HC Age SD
<55 55 60 65 70 75 80 85 90 95 100 105 110 115 120 125 130
n/a
12.0
12.3
12.6
12.8
13.1
13.4
13.7
14.0
14.3
14.7
15.0
15.3
15.6
16.0
16.3
16.6
—— ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.2
135 140 145 149 150 155 160 165 170 175 180 185 190 195 200 205 210
17.0
17.3
17.7
18.0
18.1
18.4
18.8
19.2
19.6
20.0
20.4
20.8
21.3
21.7
22.2
22.6
23.1
± 1.2 ± 1.2 ± 1.2 ± 1.2 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5 ± 1.5
215 219 220 225 230 235 240 245 250 255 260 265 270 274 275 280 285
23.6
23.9
24.0
24.5
25.0
25.5
26.1
26.6
27.1
27.7
28.3
28.9
29.4
29.9
30.0
30.7
31.3
± 1.5 ± 1.5 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 2.1 ± 3.0 ± 3.0 ± 3.0
290 295 300 305 310 315 319 320 325 330 335 340 345 350 355 360 >360
31.9
32.6
33.3
33.9
34.6
35.3
35.9
36.1
36.8
37.6
38.3
39.1
39.9
40.7
41.6
42.4 n/a
± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 3.0 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ± 2.7 ——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-15 Direction 5264895-100 Rev. 1
Page 100
OB Tables
Table 3-23: HC: Hadlock, AJR 138: 649-653, 1982 (Fetal Age)
Unit: HC (mm); Age (Days); SD (Days)
HC Age SD HC Age SD HC Age SD HC Age SD
<69 69 75 81 88 96 103 110 117 124 131 137 144 150 157 163
n/a 84 87 91 94 98 101 105 108 112 115 119 122 126 129 133
—— 9 9 9 9 9 9 9 9 9 9 9 9 9 11 11
169 175 181 187 193 198 204 209 215 220 225 230 240 245 250 255
136 140 143 147 150 154 157 161 164 168 171 175 182 185 189 192
11 11 11 11 11 11 11 11 11 11 16 16 16 16 16 16
260 264 269 273 278 282 286 291 294 298 302 306 309 312 316 319
196 199 203 206 210 213 217 220 224 227 231 234 238 241 245 248
16 16 16 16 16 19 19 19 19 19 19 19 19 19 19 19
322 325 328 331 334 337 340 343 345 348 351 353 354 >354
252 255 259 262 266 269 273 276 280 286 287 290 294 n/a
19 24 24 24 24 24 24 24 24 24 24 24 24 ——
Table 3-24: FL/HC Ratio: Hadlock, J Ultrasound Med 1984, 3: 439-442 (Fetal Growth)
Unit: GA (Weeks)
GA Min Max GA Min Max GA Min Max
<15 15 16 17 18 19 20 21 22 23
n/a
15.3
13.3
14.6
15.8
16.1
16.8
15.9
18.4
19.2
——
17.1
16.5
17.6
18.0
18.3
19.8
20.3
20.2
20.8
24 25 26 27 28 29 30 31 32 33
18.7
18.7
18.6
18.6
18.8
19.6
19.2
19.3
19.1
19.9
20.9
20.3
20.4
20.4
20.6
20.8
21.4
21.3
21.3
21.5
34 35 36 37 38 39 40 41 42 >42
19.4
20.1
20.1
20.8
20.9
20.6
20.7
21.6
20.1 n/a
21.8
22.3
22.1
22.6
22.7
23.4
22.5
23.2
23.9 n/a
Table 3-25: FL/AC Ratio: Hadlock (Fetal Growth)
Unit: Age (Weeks)
Age Min (Index) Max (Index)
21 42
20 20
24 24
3-16 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1