Page 1
Technical
Publications
Direction 5264895-100
Rev. 1
GE Healthcare
LOGIQ C5/C5 PRO
Advanced Reference Manual
Copyright© 2008 By General Electric Co.
Operating Documentation
Page 2
Regulatory Requirement
This product complies with regulatory requirements of the following European
Directive 93/42/EEC concerning medical devices
This manual is a reference for the LOGIQ C5/C5 PRO. It applies to R1.X .X and
backward software version for the LOGIQ C5/C5 PRO ultrasound systems.
GE Healthcare
GE Medical Systems: Telex 3797371
P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.
(Asia, Pacific, Latin America, North America)
GE Ultraschall TEL: 49 212.28.02.208
Deutschland GmbH & Co. KG FAX: 49 212.28.02.431
Beethovenstraße 239
Postfach 11 05 60
D-42655 Solingen GERMANY
Page 3
Revision History
Table i-1: Reason for Change
REV DATE REASON FOR CHANGE
Rev. 1 July ,2008 Initial Release
List of Effective Pages
Table i-2: List of Effective Pages
REVISION
PAGE NUMBER
Title Page Rev. 1 2-1 thru 2-10 Rev. 1
Revision History i-1 and i-2 Rev. 1 3-1 thru 3-74 Rev. 1
Table of Contents ii-1 thru ii-2 Rev. 1 Index 1 Rev. 1
1-1 thru 1-46 Rev. 1
NUMBER PAGE NUMBER
REVISION
NUMBER
Please verify that you are using the latest revision of this document.
Information pertaining to this document is maintained on GPC (GE
Medical Systems Global Product Configuration). If you need to know the
latest revision, contact your distributor, local GE Sales Representative or
in the USA call the GE Ultrasound Clinical Answer Center at
1 800 682 5327 or 1 262 524 5698.
LOGIQ C5/C5 PRO Advanced Reference Manual i-1
Direction 5264895-100 Rev. 1
Page 4
This page intentionally left blank.
i- 2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 5
Table of Contents
Table of Contents
Chapter 1 — Acoustic Output
Bioeffects - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-2
Concerns Surrounding the Use of Diagnostic Ultrasound - - - - - - - - - - - 1-2
Measurement Precision and Uncertainty - - - - - - - - - - - - - - - - - - - - - - - 1-3
FDA Acoustic Output Data - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4
Maximum output summary- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-4
IEC Acoustic Output Tables - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-60
Chapter 2 — Measurement Formulas
Measurement Formulas - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -2-2
Formulas–Generic - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-2
Formulas–OB - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-4
Formulas–GYN - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-8
Formulas–Vascular - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2-10
Chapter 3 — OB Tables
OB Tables- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
ASUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-2
Berkowitz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-3
Campbell - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
Eriksen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-4
French College of Fetal Ultrasound (CFEF) - - - - - - - - - - - - - - - - - - - 3-5
Goldstein - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-10
Hadlock - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-11
Hansmann - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-17
Hellman - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-25
Hill - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-25
Hohler - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-26
Jeanty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-26
JSUM - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-38
Kurtz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-42
Mayden - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-43
Mercer - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-44
Merz - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-45
Moore - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-55
Nelson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-55
Osaka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-56
Paris - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-60
Rempen - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-63
Robinson - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-68
Tokyo - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-68
LOGIQ C5/C5 PRO Advanced Reference Manual ii-1
Direction 5264895-100 Rev. 1
Page 6
Tokyo Shinozuka - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-72
Yarkoni - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -3-78
Index - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - i-1
ii- 2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 7
Chapter 1
Acoustic Output
Provides Acoustic Output information and tables, as
well as possible bioeffects and prudent use.
LOGIQ C5/C5 PRO Advanced Reference Manual 1-1
Direction 5264895-100 Rev. 1
Page 8
Acoustic Output
Bioeffects
Concerns Surrounding the Use of Diagnostic Ultrasound
During a diagnostic ultrasound examination, high frequency
sound penetrates and interacts with tissue in and around the
area of anatomy to be imaged. Only a small portion of this sound
energy is reflected back to the probe for use in constructing the
image while the remainder is dissipated within the tissue. The
interaction of sound energy with tissue at sufficiently high levels
can produce biological effects (aka bioeffects) of either a
mechanical or thermal nature. Although the generation of
bioeffect is intentional with therapeutic ultrasound, it is generally
undesired in diagnostic applications and may be harmful in
some conditions.
NOTE: The American Institute of Ultrasound in Medicine has published
a document entitled "Medical Ultrasound Safety". This three
part document covers Bioeffects and Biophysics, Prudent Use
and Implementing ALARA.
Ultrasound users should read the AIUM to become more familiar
with Ultrasound safety. A copy of this document is included as
part of the documentation CD (5268173-200).
To contact the AIUM concerning their publications:
• In the USA, by telephone at 1-800-638-5352.
• To write them, use the following address:
AIUM
14750 Sweitzer Lane
Suite 100
Laurel, MD, USA 20707-5906
1- 2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 9
Measurement Precision and Uncertainty
Precision and measurement uncertainty of the system used to
measure the acoustic output is provided in the following table.
These values are determined in accordance with Section 6.4 of
the Standard for Real-Time Display of Thermal and Mechanical
Acoustic Output Indices on Diagnostic Ultrasound Equipment,
rev 2 AIUM/NEMA 2004, and the Acoustic Output Measurement
Standard for Diagnostic Ultrasound Equipment, rev 3 NEMA
2004.
Table 1-1: Measurement Precision and Uncertainty
Center
Frequency
Acoustic
Power
Peak
Rarefactional
Pressure
Bioeffects
Acoustic
Intensity
Measurement
Uncertainty
+ 0.25% + 5.6% + 9.5% + 18.9%
LOGIQ C5/C5 PRO Advanced Reference Manual 1-3
Direction 5264895-100 Rev. 1
Page 10
Acoustic Output
FDA Acoustic Output Data
Maximum output summary
The following tables list the typical maimum acoustic output
levels achievalbe with LOGIQ C5/C5 PRO for all probes and
operational modes. It is intended that this is information be
useful in making ALARA decisions and selecting the most
appropriate probe for the application. In accordance with US
FDA Guidelines, the overall maximum acoustic SPTA intensity
for LOGIQ C5/C5 PRO is limited to 720 mW/cm
limited to 1.9. Modes for which TI does not exceed 0.1 are
indicated by <0.1. TI is limited to 6.0.
Table 1-2 shows all of the probes’ operating modes. The
following pages contain probe acoustic output tables for those
noted in Table 1-2.
2
and MI is
Table 1-2: Summary Table
Transducer Model:
Operatin
g Mode
B-Mode X X X X X X X X
M-Mode
(inc. BMode)
Color flow X X X X X X X X
Power
Doppler
Pulsed
Doppler
Other
(Specify)
E7C
E6C
-RC
-RC
XXX XX XXX
XXX XX XXX
XXX XX XXX
6.5C-RC4C-RC3.8C-RC3S-RC7.5L
-RC
7LRC
1- 4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 11
Maximum Probe Temperature
In accordance with IEC 60601-2-37 requirement, the maximum
lens temperature is limited to 43 degC with thermal phantom,
while 50 degC in air
Table Release Notes-2: Maximum Probe Temperature (Degrees C)
Probe Max Temp
FDA Acoustic Output Data
.
With TMM Phan-
tom*
E7C-RC 41.4 Dop 37.7 Dop
E6C-RC 42.2 CFM 40.5 CFM
6.5C-RC 37.7 Dop 37.9 Dop
4C-RC 41.2 CFM 43.1 CFM
3.8C-RC 38.9 CFM 41.3 CFM
7.5L-RC 38.8 Dop 37.4 Dop
7L-RC 38 CFM 35.5 CFM
3S-RC 40.3 B 42.1 B
Notes:
Lens temperature measured under following conditions per IEC 60601-2-37:
Thermocouple was placed at the geometric center of the lens.
Thermal phantom* placed in 33 degC water bath for non-endo probes (all probes except (E6C-RC/E7CRC).
Thermal phantom* placed in 37 degC water bath for endo probes (E6C-RC/E7C-RC)
Probe placed upright in contact with above thermal phantom.
Autofreeze capability disabled.
Lens temperature monitored for 30 min.
Measurement uncertainty: 1.56 degC.
*Thermal phantom made with tissue-mimicking material as referenced in IEC60601-2-37: Appendix J.
Mode In Air Mode
LOGIQ C5/C5 PRO Advanced Reference Manual 1-5
Direction 5264895-100 Rev. 1
Page 12
Acoustic Output
Table Key
Table 1-1: Table Key
FDA IEC Meaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3
a α Acoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz)
A
aprt
D
eq
d
-6
d
eq
f
c
I
pa
I
pa.3
PII I
PII
.3
I
TA
(Z) I
I
TA. 3
(Z) I
I
SPTA
I
(Z) I
SPTA.3
A
arpt
C
MI
D
eq
d
-6
d
eq
ƒ
awf
I
pa
I
pa,α
pi
I
pi,α
Ita(z) Temporal-Average Intensity
(z ) Attenuated Temporal-Average Intensity / (at depth z)
ta,α
(z ) Spatial-Peak Temporal-Average Intensity
zpta
(z ) Attenuated Spatial-Peak Temporal-Average Intensity
zpta,α
MI MI Mechanical Index
W
o
(Z) P
W
.3
W
o1
PII p
p
r
p
r.3
P Output Power / Time average acoustic power at the source
α
P
1
i
p
r
p
rα
PRF prr Pulse Repetition Rate / Pulse repetition frequency
TI TI Thermal Index / (same)
TIB TIB Bone Thermal Index / (same)
TIC TIC Cranial-Bone Thermal Index / (same)
TIS TIS Soft-Tissue Thermal Index / (same)
PD t
x
-12,y-12
d
X, Y -12 dB Output Beam Dimensions / (same)
Z Z Distance from the Source to a Specified Point / (same)
Z
sp
Z
bp
Z
sp
Z
b
Z
bp
Z
s
-12db Output Beam Area / Active aperture area
Normalizing Coefficient
Equivalent Aperture Diameter / (same)
Pulse Beam Width / Beam diameter at –6 dB
Equivalent Beam Diameter
Acoustic Working Frequency / Center frequency
Pulse-Average Intensity
Attenuated Pulse-Average Intensity
Pulse-Intensity Integral
Attenuated Pulse-Intensity Integral
Attenuated Output Power / Time average acoustic power derated to depth z
Bounded Output Power / Power emitted from the central 1cm of aperture
Pulse Pressure Squared Integral / Pulse intensity integral
Peak-Rarefactional Acoustic Pressure / (same)
Attenuated Peak-Rarefactional Acoustic Pressure / (same)
Pulse Duration / (same)
Depth for TIB / Depth at which the relevant index is maximum
Break-Point Depth / (same)
Depth for TIS / Depth at which the relevant index is maximum
1- 6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 13
FDA Acoustic Output Data
Table 1-2: Table Key (in French 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-7
Direction 5264895-100 Rev. 1
Page 14
Acoustic Output
Table 1-3: Table Key (in French 2)
1- 8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 15
FDA Acoustic Output Data
Table 1-4: Table Key (in German 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-9
Direction 5264895-100 Rev. 1
Page 16
Acoustic Output
Table 1-5: Table Key (in German 2)
1- 10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 17
FDA Acoustic Output Data
Table 1-6: Table Key (in Italian 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-11
Direction 5264895-100 Rev. 1
Page 18
Acoustic Output
Table 1-7: Table Key (in Italian 2)
1- 12 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 19
FDA Acoustic Output Data
Table 1-8: Table Key (in Brazilian Portuguese 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-13
Direction 5264895-100 Rev. 1
Page 20
Acoustic Output
Table 1-9: Table Key (in Brazilian Portuguese 2)
1- 14 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 21
FDA Acoustic Output Data
Table 1-10: Table Key (in Danish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-15
Direction 5264895-100 Rev. 1
Page 22
Acoustic Output
Table 1-11: Table Key (in Danish 2)
1- 16 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 23
FDA Acoustic Output Data
Table 1-12: Table Key (in Latin American Spanish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-17
Direction 5264895-100 Rev. 1
Page 24
Acoustic Output
Table 1-13: Table Key (in Latin American Spanish 2)
1- 18 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 25
FDA Acoustic Output Data
Table 1-14: Table Key (in Norwegian 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-19
Direction 5264895-100 Rev. 1
Page 26
Acoustic Output
Table 1-15: Table Key (in Norwegian 2)
1- 20 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 27
FDA Acoustic Output Data
Table 1-16: Table Key (in Swedish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-21
Direction 5264895-100 Rev. 1
Page 28
Acoustic Output
Table 1-17: Table Key (in Swedish 2)
1- 22 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 29
FDA Acoustic Output Data
Table 1-18: Table Key (in Finnish 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-23
Direction 5264895-100 Rev. 1
Page 30
Acoustic Output
Table 1-19: Table Key (in Finnish 2)
1- 24 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 31
FDA Acoustic Output Data
Table 1-20: Table Key (in Japanese 1)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-25
Direction 5264895-100 Rev. 1
Page 32
Acoustic Output
Table 1-21: Table Key (in Japanese 2)
1- 26 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 33
FDA Acoustic Output Data
Table 1-22: Table Key (in Chinese)
LOGIQ C5/C5 PRO Advanced Reference Manual 1-27
Direction 5264895-100 Rev. 1
Page 34
Acoustic Output
4C-RC
Table 1-23: Transducer Mode 4C-RC Operating Mode: B Mode
Index Label MI
scan
TIS TIB
non-scan
nonscan Aaprt< Aaprt>1
TIC
Global Maximum: Index Value 0.98
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.56
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 48.8 48.8
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 5.2
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.70 2.87 2.87
X( c m ) 1.2 1.2
Y( c m )
t
d
PD (µsec) 0.47
prr PRF (Hz) 17
pr at max. Ipi pr@PII
(MPa) 2.16
max
0.67
1.3 1.3
0.37
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
@MI
max
FL
ma
(cm)
(cm) 2.75 2.75
X
(cm) 5.24 5.24
Y
(W/
cm
130.
2
6
)
Frequency (MHz) 3.0 3.0 3.0
Image Depth (cm) 7.0 7.0
Focus (cm) 6.4 5.12
7.0
5.12
1- 28 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 35
FDA Acoustic Output Data
Table 1-24: Transducer Mode 4C-R
C Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.80 0.51 - 0.30 0.78 0.28
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
zs z1(cm) 3.1
zbp zbp(cm) 3.1
zb zsp(cm) 5.2
z at max. Ipi,a z
deq(zb) deq(zsp) (cm) 3.15
ƒ
awf
(MPa) 1.31
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
sp
f
c
(mW) 37.4 - 37.4 37.4
(z1)]) 40.45
(cm) 4.2
(MHz) 2.70 2.87 - 2.87 2.87 2.87
non-scan
Aaprt>
TIC
non-
1
scan Aaprt<
Dim of A
t
d
aprt
PD (µsec) 0.86
X( c m ) 1.2 - 6 6 1.2
Y( c m )
1.3 - 1.3 1.3 1.3
prr PRF (Hz) 513.9
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,a
x
PA. 3
(MPa) 2.13
max
(cm) 3
max
(cm) 2.75 - 2.75 2.75
X
(cm) 5.24 - 5.24 5.24
FL
Y
@MI
ma
(W/
cm
93.04
2
)
Frequency (MHz) THI5.24.0 - 4.0 4.0
Image Depth (cm) 17.0 17.0 - 17.0
Focus (cm) 6.4 4.1 - 4.1
17.0
4.1
LOGIQ C5/C5 PRO Advanced Reference Manual 1-29
Direction 5264895-100 Rev. 1
Page 36
Acoustic Output
Table 1-25: Transducer Mode 4C-R
C Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 0.83 1.26 0.8
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.52
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 106 106
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.0
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 3.3 2.5 2.5
X( c m ) 6.0 6.0
Y( c m )
1.3 1.3
TIC
t
d
PD (µsec) 2.24
prr PRF (Hz) 7
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.13
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 11 .7 11. 7
(cm) 11. 9 11. 9
100.
2
9
)
Frequency (MHz) 2.5 2.5 2.5
Image Depth (cm) 8 10
Vel Scale (kHz) 2.3 1.6
ROI Max Max
PS 10 10
Focus (cm) 44.8 51.2
10
1.6
Max
10
51.2
1- 30 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 37
FDA Acoustic Output Data
Table 1-26: Transducer Mode: 4C-R
C Operating Mode: Pulsed Doppler (Inc.B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 0.85 0.68 - 0.68 2.63 0.88
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.33
r.3
(mW) 57.6 - 57.6 57.6
.3(Z1),ITA. 3
ta,a(z s
o
)] [(W
(z1)]) 137.1
zs z1(cm) 5.4
zbp zbp(cm) 5.4
zb zsp(cm) 3.8 5.4
z at max. Ipi,a z
sp
(cm) 5.4
deq(zb) deq(zsp) (cm) 3.2
ƒ
awf
f
c
(MHz) 2.49 2.49 2.49 2.49 2.49 2.49
TIC
Dim of A
t
d
aprt
PD (µsec) 2.06
X( c m ) 1.43 - 6.0 6.0 1.43
Y( c m )
1.3 - 1.3 1.3 1.3
prr PRF (Hz) 2604.1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 1.99
max
(cm) 3.2
max
(cm) -2 . 5 2.5
X
FL
(cm) -3 . 7 3.7
Y
@MI
max
(W/
cm
2
98.25
)
Frequency (MHz) 2.5 2.5 2.5 2.5 2.5 2.5
SVD (cm) 8.0 8.0 8.0 8.0 8.0 8.0
Vel Scale (kHz) 2.6 3.9 3.9 3.9 3.9 3.9
SV 4 4 - 4 4 4
Image Depth(cm) 10 10 10 10 10 10
LOGIQ C5/C5 PRO Advanced Reference Manual 1-31
Direction 5264895-100 Rev. 1
Page 38
Acoustic Output
E6C-RC
Table 1-27: Transducer Mode E6C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.83 0.58 0.50
IEC FDA Units
non-scan
nonscan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 1.78
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.9
z at max. Ipi,a z
sp
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
X( c m ) 0.81 0.81
Y( c m )
t
d
PD (µsec) 0.16
prr PRF (Hz) 35
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 2.41
max
(cm)
max
X
FL
Y
(mW) 27.6 27.6
(z1)])
(cm)
(MHz) 4.63 4.46 4.46
0.5 0.5
(cm) 0.81 0.81
(cm) 3.0 3.0
I
at max. MI I
pa,a
PA. 3
x
@MI
ma
(W/
cm
2
226
)
Frequency (MHz) 6.5 6.5 6.5
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
1- 32 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.6
Page 39
FDA Acoustic Output Data
Table 1-28: Transducer Mode E
6C-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.81 0.53 0.53 0.63 0.45
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.73
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 24.9 24.9 24.9 24.9
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.9
z at max. Ipi,a z
sp
(cm) 1.4
deq(zb) deq(zsp) (cm) 1.13
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.62 4.49 4.49 - 4.49 4.49
X( c m ) 0.81 2.5 - 2.5 0.81
Y( c m )
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.31
prr PRF (Hz) 522.4
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz)
Image Depth (cm) 4.0 4.0 4.0 4.0
Focus (cm) 2.0 1.6 1.6 1.6
(MPa) 3.51
max
(cm) 1.13
max
(cm) 0.81 2.5 2.5 0.81
X
FL
(cm) 3.0 3.0 3.0 3.0
Y
@MI
max
(W/
cm
226.6
2
)
6.5 6.5 6.5 6.5
6.5
4.0
1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-33
Direction 5264895-100 Rev. 1
Page 40
Acoustic Output
Table 1-29: Transducer Mode E
6C-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.76 0.88 0.68
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.68
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 37.6 37.6
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.7
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.86 4.91 4.91
X( c m ) 0.81 0.81
Y( c m )
0.5 0.5
TIC
t
d
PD (µsec) 0.70
prr PRF (Hz) 7
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 2.23
max
(cm)
max
Focal Length FLX(cm) 0.81 0.81
FL
(cm) 3.0 3.0
Y
at max. MI I
I
pa,a
PA. 3
@MI
max
(W/
cm
103.7
2
)
Frequency (MHz) 5.0 5.0 5.0
Image Depth (cm) 9.0 9.0
Vel Scale (kHz) 1.3 1.3
ROI Max Max
PS 10 10
Focus (cm) 2.0 1.8
9.0
1.3
Max
10
1.8
1- 34 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 41
FDA Acoustic Output Data
Table 1-30: Transducer Mode: E
6C-RC Operating Mode: Pulsed Doppler (Inc.B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.66 0.86 0.86 - 3.13 1.07
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.46
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 36.2 36.2 36.2 36.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.9 1.9
z at max. Ipi,a z
sp
(cm) 1.9
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.83 4.98 4.98 - 4.98 4.98
X( c m ) 0.81 2.5 - 2.5 0.81
Y( c m )
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.47
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.01
max
(cm) 0.63
max
(cm) 0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
ma
(W/
cm
79.6
2
)
3
Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0
SVD (cm) 2.8 2.8 2.8 - 2.8 2.8
Vel Scale (kHz) 1.3 3.9 3.9 - 3.9 3.9
SV 1 3 3 - 3 3
Image Depth(cm) 9 9 9 9 9
LOGIQ C5/C5 PRO Advanced Reference Manual 1-35
Direction 5264895-100 Rev. 1
Page 42
Acoustic Output
E7C-RC
Table 1-31: Transducer Mode E7C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.32 0.56 0.41
IEC FDA Units
non-scan
nonscan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 2.83
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 23.5 23.5
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.63 4.96 4.96
X( c m ) 0.81 0.81
Y( c m )
t
d
PD (µsec) 0.40
prr PRF (Hz) 25
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.48
max
(cm)
max
(cm) 0.81 0.81
X
0.5 0.5
(cm) 3.0 3.0
FL
Y
I
at max. MI I
pa,a
PA. 3
x
@MI
ma
(W/
cm
244.1
2
)
Frequency (MHz) THI10 8 8
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
4.0
1.6
1- 36 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 43
FDA Acoustic Output Data
Table 1-32: Transducer Mode E7C
-RC Operating Mode: M Mode (Inc. B-Mode)
Index Label MI
scan
Aaprt< Aaprt>1
Global Maximum: Index Value
IEC FDA Units
pra p
P W
min of [ P
α( zs
, I
(MPa)
r.3
o
)] [(W
ta,α (z s
.3(Z1),ITA. 3
(mW)
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm)
z at max. Ipi, α z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz)
X( c m )
0.99 0. 57 0 .57
2.15
23.9 23.9 23.9 23.9
1.4
4.66 4.96 4.96
0.81 2.5
TIS TIB
non-scan
nonscan
-
-
-
0.69 0.42
1. 4
1.13
4. 96 4. 96
2.5 0.81
TIC
Y( c m )
t
d
prr PRF (Hz)
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,α
Frequency (MHz)
Image Depth (cm)
PD (µsec)
(MPa)
max
(cm)
max
(cm)
X
FL
(cm)
Y
x
PA. 3
@MI
ma
(W/
cm
2
)
0.39
524.1
2.64
132.2
THI10 8 8
4.0 4.0 4.0
Focus (cm) 2.0
0.5 0.5
0.81 2.5
3.0 3.0
1.6 1.6
-
0.5 0.5
1.13
-
-
-
-
-
0.81
3.0
88
4.0 4.0
1.6 1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-37
Direction 5264895-100 Rev. 1
Page 44
Acoustic Output
Table 1-33: Transducer Mode E7C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.00 0.94 0.71
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.21
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 40.1 40.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.90 4.93 4.93
X( c m ) 0.81 0.81
Y( c m )
0.5 0.5
TIC
t
d
PD (µsec) 0.62
prr PRF (Hz) 2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.81
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 0.81 0.81
(cm) 3.0 3.0
205.
2
)
5
Frequency (MHz) 5.0 5.0 5.0
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 1.3 1.4
ROI Max Max
PS 10 10
Focus (cm) 1.8 1.6
4.0
1.4
Max
10
1.6
1- 38 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 45
FDA Acoustic Output Data
Table 1-34: Transducer Mode: E7C
-RC Operating Mode: Pulsed Doppler (inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.92 0.64 0.64 - 2.68 0.95
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.04
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 26.9 26.9 26.9 26.9
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.5 1.8
z at max. Ipi,a z
sp
(cm) 1.8
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.95 4.98 4.98 - 4.98 4.98
X( c m ) 0.81 2.5 - 2.5 0.81
Y( c m )
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 1.07
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 2.64
max
(cm) 0.63
max
(cm) 0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
max
(W/
cm
155.6
2
)
Frequency (MHz) 5.0 5.0 5.0 - 5.0 5.0
SVD (cm) 2.0 2.0 2.0 - 2.0 2.0
Vel Scale (kHz) 2.0 2.0 2.0 - 2.0 2.0
SV 2 3 3 - 3 3
Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-39
Direction 5264895-100 Rev. 1
Page 46
Acoustic Output
7L-RC
Table 1-35: Transducer Mode 7L-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.83 0.71 0.40
IEC FDA Units
non-scan
non-
scan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 2.11
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 22.4 22.4
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 6.47 6.64 6.64
X( c m ) 3.84 3.84
Y( c m )
t
d
PD (µsec) 0.19
prr PRF (Hz) 92
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 2.69
max
(cm)
max
(cm) 3.84 3.84
X
0.4 0.4
(cm) 2.3 2.3
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
133.6
2
)
Frequency (MHz) THI12 11 11
Image Depth (cm) 4.0 4.0
Focus (cm) 1.6 1.2
1- 40 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.2
Page 47
FDA Acoustic Output Data
Table 1-36: Transducer Mode 7L-RC
Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 0.83 0.52 0.52 - 0.33 0.29
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.21
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 16.2 16.2 16.2 16.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm) 1.2
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 6.48 6.78 6.78 - 6.78 6.78
X( c m ) 3.84 0.83 - 0.83 3.84
Y( c m )
0.4 0.4 - 0.4 0.4
TIC
t
d
PD (µsec) 0.19
prr PRF (Hz) 516.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz) THI1
(MPa) 2.71
max
(cm) 0.63
max
(cm) 3.84 3.84 - 3.84 3.84
X
FL
(cm) 2.3 2.3 - 2.3 2.76
Y
@MI
max
(W/
cm
132.
2
)
5
11 11 - 11 11
2
Image Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
Focus (cm) 1.6 1.2 2.2 - 1.2 1.2
LOGIQ C5/C5 PRO Advanced Reference Manual 1-41
Direction 5264895-100 Rev. 1
Page 48
Acoustic Output
Table 1-37: Transducer Mode 7L-RC
Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.09 0.78 0.59
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.44
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 32.8 32.8
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.98 5.00 5.00
X( c m ) 3.84 3.84
Y( c m )
0.4 0.4
TIC
t
d
PD (µsec) 0.64
prr PRF (Hz) 11
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.94
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 3.84 3.84
(cm) 2.3 2.3
280
2
)
Frequency (MHz) 5 5 5
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 3.0 3.0
ROI Max Min
PS 12 12
Focus (cm) 1.8 1.8
4.0
3.0
Min
12
1.8
1- 42 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 49
FDA Acoustic Output Data
Table 1-38: Transducer Mode: 7L
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
n
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 0.73 0.73 0.73 - 3.07 1.09
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.62
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 30.5 30.5 30.5 30.5
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 2.1 2.3
z at max. Ipi,a z
sp
(cm) 2.3
deq(zb) deq(zsp) (cm) 0.35
ƒ
awf
f
c
(MHz) 4.97 4.99 4.99 - 4.99 4.99
TIC sca
Dim of A
t
d
aprt
PD (µsec) 1.05
X( c m ) 3.84 0.83 - 0.83 3.84
Y( c m )
0.4 0.4 - 0.4 0.4
prr PRF (Hz) 130
2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.32
max
(cm) 0.58
max
(cm) 3.84 0.83 - 3.84
X
FL
(cm) 2.3 2.3 - 2.3
Y
@MI
ma
(W/
cm
136.
2
6
)
Frequency (MHz) 5 5 5 - 5 5
SVD (cm) 3.0 3.0 3.0 - 3.0 3.0
Vel Scale (kHz) 1.3 2.6 2.6 - 2.6 2.6
SV 2 3 3 - 3 3
Depth(CM0 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-43
Direction 5264895-100 Rev. 1
Page 50
Acoustic Output
7.5L-RC
Table 1-39: Transducer Mode7.5L-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.28 0.65 0.51
IEC FDA Units
non-scan
nonscan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 2.78
r.3
(mW) 28.5 28.5
.3(Z1),ITA. 3
ta,a(z s
o
)] [(W
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.3
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.75 4.80 4.80
X( c m ) 3.84 2.5
Y( c m )
t
d
PD (µsec) 0.22
prr PRF (Hz) 77
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.45
max
(cm)
max
(cm) 3.84 3.84
X
0.4 0.4
(cm) 2.3 2.3
FL
Y
I
at max. MI I
pa,a
x
PA. 3
@MI
ma
(W/
cm
2
263.2
)
Frequency (MHz) THI12 6 6
Image Depth (cm) 4.0 4.0
Focus (cm) 2.0 1.6
1- 44 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
4.0
1.6
Page 51
FDA Acoustic Output Data
Table 1-40: Transducer Mode 7.5
L-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 1.27 0.61 0.61 - 0.34 0.45
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.75
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 25.3 25.3 25.3 25.3
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.3
z at max. Ipi,a z
sp
(cm) 1.4
deq(zb) deq(zsp) (cm) 0.63
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.72 5.04 5.04 - 5.04 5.04
X( c m ) 3.84 0.83 - 0.83 3.84
Y( c m )
0.4 0.4 - 0.4 0.4
TIC
t
d
PD (µsec) 0.21
prr PRF (Hz) 516.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
Frequency (MHz) THI1
(MPa) 3.40
max
(cm) 0.71
max
(cm) 3.84 3.84 - 3.84 3.84
X
FL
(cm) 2.3 2.3 - 2.3 2.3
Y
@MI
max
(W/
cm
271.
2
2
)
66 - 66
2
Image Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
Focus (cm) 2.0 1.6 1.6 - 1.6 1.6
LOGIQ C5/C5 PRO Advanced Reference Manual 1-45
Direction 5264895-100 Rev. 1
Page 52
Acoustic Output
Table 1-41: Transducer Mode 7.5
L-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.24 0.91 0.69
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.75
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 38.5 38.5
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.1
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.93 4.95 4.95
X( c m ) 3.84 3.84
Y( c m )
0.4 0.4
TIC
t
d
PD (µsec) 0.64
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 3.32
max
(cm)
max
X
FL
Y
@MI
ma
(W/
cm
(cm) 3.84 3.84
(cm) 2.3 2.3
392.
2
)
4
Frequency (MHz) 5.0 4 4
Image Depth (cm) 4.0 4.0
Vel Scale (kHz) 3.3 2.6
ROI Max Max
Packet Size 10 8
4.0
2.6
Max
8
Focus (cm) 1.8 1.8 1.8
1- 46 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 53
FDA Acoustic Output Data
Table 1-42: Transducer Mode: 7.5
L-RC Operating Mode: Pulsed Doppler (inc. B-Mode)
Index Label MI
scan
Aaprt< Aaprt>1
Global Maximum: Index Value
IEC FDA Units
pra p
P W
min of [ P
α( zs
, I
(MPa)
r.3
o
)] [(W
ta,α (z s
.3(Z1),ITA. 3
(mW)
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm)
z at max. Ipi, α z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz)
X( c m )
0.98 0.78 0.78
2.18
32.8 32.8 32.8 32.8
1.7
4.97 4.99 4.99
3.84 0.8 3
TIS TIB
non-scan
nonscan
-
-
-
3.54 1.25
2.0
2.0
0.35
4.99 4.99
0.8 33 .8 4
TIC
Y( c m )
t
d
PD (µsec)
1.04
0.4 0.4
-
0.4 0.4
prr PRF (Hz) 1302
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
I
at max. MI I
pa,α
PA. 3
Frequency (MHz) 5.0
SVD (cm)
Vel Scale (kHz) 1.3
SV
max
max
@MI
(MPa)
(cm)
(cm)
X
FL
(cm)
Y
(W/
max
cm
2.93
245.
2
)
3
2.6 2.6 2.6
2
0.58
3.84 0.83
2.3 2.3
5.0 5.0
2.6 2.6
-
-
-
-
-
3.84
2.3
5.0 5.0
2.6 2.6
2.6 2.6
33 - 33
Depth (cm) 4.0 4.0 4.0 - 4.0 4.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-47
Direction 5264895-100 Rev. 1
Page 54
Acoustic Output
3S-RC
Table 1-43: Transducer Mode 3S-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.17 0.69 1.24
IEC FDA Units
non-scan
non-
scan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 1.5
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 82.1 82.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.3
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 1.66 1.76 1.76
X( c m ) 0.9 0.6
Y( c m )
t
d
PD (µsec) 1.16
prr PRF (Hz) 27
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 1.81
max
(cm)
max
(cm) 0.9 0.9
X
1.2 1.2
(cm) 4.8 4.8
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
Frequency (MHz) THI
315.
2
15
)
3.2 3.2
3.2
Image Depth (cm) 30.0 30.0
Focus (cm) 3.2 3.2
30.0
3.2
1- 48 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 55
FDA Acoustic Output Data
Table 1-44: Transducer Mode 3S
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 1.17 0.69 - 0.69 1.28 1.39
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.5
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 82.1 - 82.1 82.1
(z1)]) 2.2
zs z1(cm) 2.2
zbp zbp(cm) 2.2
zb zsp(cm) 3.3
z at max. Ipi,a z
sp
(cm) 2.2
deq(zb) deq(zsp) (cm) 1.48
ƒ
awf
Dim of A
aprt
f
c
(MHz) 1.66 1.76 - 1.76 1.76 1.76
X( c m ) 0.9 - 1.9 1.9 1.9
Y( c m )
4.8 - 4.8 4.8
TIC
t
d
PD (µsec) 1.16
prr PRF (Hz) 113
4.3
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 1.81
max
(cm) 1.48
max
Focal Length FLX(cm) 0.9 - 1.9 0.9
FL
(cm) 4.8 - 4.8 4.8
Y
at max. MI I
I
pa,a
PA. 3
x
@MI
ma
(W/
cm
Frequency (MHz) THI
315.
2
15
)
3.2 - 3.2 3.2
3.2
Image Depth (cm) 30.0 30.0 - 30.0
Focus (cm) 3.2 3.2 - 3.2
30.0
3.2
LOGIQ C5/C5 PRO Advanced Reference Manual 1-49
Direction 5264895-100 Rev. 1
Page 56
Acoustic Output
Table 1-45: Transducer Mode 3S
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.72 1.23 2.23
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 0.86
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 146.8 146.
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 3.5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
f
c
(MHz) 1.96 1.76 1.76
TIC
8
Dim of A
t
d
aprt
PD (µsec) 1.31
X( c m ) 0.9 0.9
Y( c m )
1.2 1.2
prr PRF (Hz) 17
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
(MPa) 1.09
max
(cm)
max
Focal Length FLX(cm) 0.9 0.9
(cm) 4.8 4.8
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
416.
2
93
)
Frequency (MHz) 2.0 2.0 2.0
Image Depth (cm) 15.0 15.0
Vel Scale (kHz) 4.1 4.1
Penet Mid Mid
ROI 8 8
PS 3.52 3.52
15.0
4.1
Mid
8
3.52
1- 50 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 57
FDA Acoustic Output Data
Table 1-46: Transducer Mode: 3S
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.89 0.53 0.53 - 2.27 0.77
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.26
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 44.2 44.2 44.2 44.2
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.8 4.2
z at max. Ipi,a z
sp
(cm) 4.2
deq(zb) deq(zsp) (cm) 1.44
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.0 2.51 2.51 - 2.51 2.51
X( c m ) 0.9 1.9 - 1.9 0.9
Y( c m )
1.2 1.2 - 1.2 1.2
TIC
t
d
PD (µsec) 0.78
prr PRF (Hz) 113
2.2
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 1.42
max
(cm) 1.44
max
(cm) 0.9 - 0.9
X
FL
(cm) 4.8 - 4.8
Y
@MI
ma
(W/
cm
273.
2
48
)
Frequency (MHz) 2.0 2.0 2.0 - 2.0 2.0
SVD (cm) 2.4 2.4 2.4 - 2.4 2.4
Vel Scale (kHz) 1.1 1.1 1.1 - 1.1 1.1
SV 1 1 1 - 11
LOGIQ C5/C5 PRO Advanced Reference Manual 1-51
Direction 5264895-100 Rev. 1
Page 58
Acoustic Output
3.8C-RC
Table 1-47: Transducer Mode 3.8C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 0.76 0.56 0.28
IEC FDA Units
non-scan
nonscan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 1.24
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 36.4 36.4
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.68 3.1 3.1
X( c m ) 1.2 1.2
Y( c m )
t
d
PD (µsec) 0.46
prr PRF (Hz) 4 8
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 1.97
max
(cm)
max
(cm) 12.
X
1.3 1.3
0.9
75
FL
(cm) 5.24 4.8
Y
I
at max. MI I
pa,a
Frequency (MHz) THI
PA. 3
@MI
max
(W/
cm
67.5
2
5
)
4.0 4.0
5.2
Image Depth (cm) 17 17
Focus (cm) 6.4 4.1
17
4.1
1- 52 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 59
FDA Acoustic Output Data
Table 1-48: Transducer Mode 3.8C
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value .076 0.64 - 0.26 0.72 0.32
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.25
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 41.8 - 41.8 41.8
(z1)]) 82.83
zs z1(cm) 3.1
zbp zbp(cm) 3.1
zb zsp(cm) 5
z at max. Ipi,a z
sp
(cm) 3.1
deq(zb) deq(zsp) (cm) 2.8
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.67 3.23 - 3.23 3.23 3.23
X( c m ) 1.2 - 6 6 1.2
Y( c m )
1.3 - 1.3 1.3 1.3
TIC
t
d
PD (µsec) 0.46
prr PRF (Hz) 513.
9
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
Frequency (MHz) THI
(MPa) 1.98
max
(cm) 2.8
max
(cm) 2.75 - 2.75 2.75
X
FL
(cm) 5.24 - 5.24 5.24
Y
@MI
ma
(W/
cm
315.
2
15
)
4.0 - 4.0 4.0
5.2
Image Depth (cm) 17 17 - 17
Focus (cm) 6.4 4.1 - 4.1
17
4.1
LOGIQ C5/C5 PRO Advanced Reference Manual 1-53
Direction 5264895-100 Rev. 1
Page 60
Acoustic Output
Table 1-49: Transducer Mode 3.8C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.87 1.41 0.9
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.38
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 117 11 7
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 4.4
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.51 2.52 2.52
X( c m ) 6.0 6.0
Y( c m )
1.3 1.3
TIC
t
d
PD (µsec) 1.36
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 2.03
max
(cm)
max
X
FL
Y
@MI
(W/
max
cm
(cm) 11 .7 11.7
(cm) 11.9 11 .9
112.
2
7
)
Frequency (MHz) 2.5 2.5 2.5
Image Depth (cm) 10 10
Vel Scale (kHz) 2.0 2.0
Penet Max Max
ROI 10 12
PS 60.8 60.8
10
2.0
Max
12
60.8
1- 54 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 61
FDA Acoustic Output Data
Table 1-50: Transducer Mode:3.8C
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
nonscan
Global Maximum: Index Value 0.74 0.93 - 0.93 3.95 1.26
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.17
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 77.9 - 77.9 77.9
(z1)]) 116.3
zs z1(cm) 4.8
zbp zbp(cm) 4.8
zb zsp(cm) 4.1 4.8
z at max. Ipi,a z
sp
(cm) 4.8
deq(zb) deq(zsp) (cm) 2.53
ƒ
awf
Dim of A
aprt
f
c
(MHz) 2.49 2.51 - 2.51 2.51 2.51
X( c m ) 1.43 - 6.0 6.0 6.0
Y( c m )
1.3 - 1.3 1.3 1.3
TIC
t
d
PD (µsec) 2.05
prr PRF (Hz) 320
5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 1.66
max
(cm) 2.53
max
(cm) -2 . 5 2.5
X
FL
(cm) -2 . 7 3.7
Y
@MI
ma
(W/
cm
70.1
2
4
)
Frequency (MHz) 2.5 2.5 - 2.5 2.5 2.5
SVD (cm) 5.6 5.6 - 5.6 5.6 5.6
Vel Scale (kHz) 3.2 3.2 - 3.2 3.2 3.2
SV 4 3 - 3 3 3
Image Depth(cm) 10 10 10 10 10
LOGIQ C5/C5 PRO Advanced Reference Manual 1-55
Direction 5264895-100 Rev. 1
Page 62
Acoustic Output
6.5C-RC
Table 1-51: Transducer Mode 6.5C-RC Operating Mode: B Mode
TIS TIB
Index Label MI
scan
Global Maximum: Index Value 1.11 1.50 0.45
IEC FDA Units
non-scan
nonscan Aaprt< Aaprt>1
TIC
pra p
P W
min of [P
a(z s
, I
(MPa) 2.40
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 25.1 25.1
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.5
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.65 4.2 4.2
X( c m ) 0.62 0.62
Y( c m )
t
d
PD (µsec) 0.38
prr PRF (Hz) 58
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
(MPa) 3.06
max
(cm)
max
(cm) 0.62 0.62
X
0.5 0.5
(cm) 2.76 2.76
FL
Y
I
at max. MI I
pa,a
PA. 3
@MI
max
(W/
cm
67.5
2
5
)
Frequency (MHz) THI106.5 6.5
Image Depth (cm) 4.0 4.0
Focus (cm) 1.6 42.4
4.0
2.4
1- 56 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 63
FDA Acoustic Output Data
Table 1-52: Transducer Mode 6.5C
-RC Operating Mode: M Mode (Inc. B-Mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 1.02 0.44 0.44 - 0.45 0.39
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 2.20
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 21.6 21.6 21.6 21.6
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.5
z at max. Ipi,a z
sp
(cm) 1.6
deq(zb) deq(zsp) (cm) 1.6
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.66 4.25 4.25 - 4.25 4.25
X( c m ) 0.62 2.5 - 2.5 0.62
Y( c m )
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 0.39
prr PRF (Hz) 524.
1
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
x
PA. 3
(MPa) 2.80
max
(cm) 1.6
max
(cm) 0.62 2.5 - 0.62
X
FL
(cm) 2.76 2.76 - 2.76
Y
@MI
ma
(W/
cm
188.
2
9
)
Frequency (MHz) THI106.5 6.5 - 6.5
Image Depth (cm) 4.0 4.0 4.0 -
Focus (cm) 2.4 2.4 2.4 -
4.0
2.4
LOGIQ C5/C5 PRO Advanced Reference Manual 1-57
Direction 5264895-100 Rev. 1
Page 64
Acoustic Output
Table 1-53: Transducer Mode 6.5C
-RC Operating Mode: Color Flow Mode
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.59 0.46 0.35
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.30
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 19.7 19.7
(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 2
z at max. Ipi,a z
sp
(cm)
deq(zb) deq(zsp) (cm)
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.90 4.92 4.92
X( c m ) 0.81 0.81
Y( c m )
0.5 0.5
TIC
t
d
PD (µsec) 0.63
prr PRF (Hz) 5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
(MPa) 1.82
max
(cm)
max
X
FL
Y
@MI
(W/
max
cm
(cm) 0.81 0.81
(cm) 3.0 3.0
79.3
2
3
)
Frequency (MHz) 5 5 5
Image Depth (cm) 6.0 6.0
Vel Scale (kHz) 2.3 1.9
ROI Max Max
Packet Size 14 14
Focus(cm) 2.4 2.24
6.0
1.9
Max
14
2.24
1- 58 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 65
FDA Acoustic Output Data
Table 1-54: Transducer Mode:6.5C
-RC Operating Mode: Pulsed Doppler (Inc. B-mode)
TIS TIB
Index Label MI
scan
non-scan
Aaprt< Aaprt>1
non-
scan
Global Maximum: Index Value 0.90 0.67 0.67 - 3.17 0.95
IEC FDA Units
pra p
P W
min of [P
a(z s
, I
(MPa) 1.99
r.3
o
)] [(W
ta,a(z s
.3(Z1),ITA. 3
(mW) 28.5 28.5 28.5 28.5
(z1)]) -
zs z1(cm) -
zbp zbp(cm) -
zb zsp(cm) 1.8 2.2
z at max. Ipi,a z
sp
(cm) 2.2
deq(zb) deq(zsp) (cm) 0.75
ƒ
awf
Dim of A
aprt
f
c
(MHz) 4.95 4.96 4.96 - 4.96 4.96
X( c m ) 0.81 2.5 - 2.5 0.81
Y( c m )
0.5 0.5 - 0.5 0.5
TIC
t
d
PD (µsec) 1.06
prr PRF (Hz) 196
5
pr at max. Ipi pr@PII
deq at max. Ipi deq@PII
Focal Length FL
at max. MI I
I
pa,a
PA. 3
x
(MPa) 2.71
max
(cm) 0.75
max
(cm) -0.81 - 0.81
X
FL
(cm) 3.0 - 3.0
Y
@MI
ma
(W/
cm
185.
2
5
)
Frequency (MHz) 5 5 5 - 5 5
SVD (cm) 2.2 2.2 2.2 - 5.6 5.6
Vel Scale (kHz) 2.0 5.2 5.2 - 5.2 5.2
SV 2 2 2 - 2 2
Image Depth(cm) 6.0 6.0 6.0 - 6.0 6.0
LOGIQ C5/C5 PRO Advanced Reference Manual 1-59
Direction 5264895-100 Rev. 1
Page 66
Acoustic Output
IEC Acoustic Output Tables
4C-RC
Table 1-55: 4C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.16 2.13 2.93 1.99 1.99
(mW/cm2)
I
spta
(mm) 52 52 30 47 47
I
p
w
pb6
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
prr (kHz) - 513.9 - 2604.1 2604.1
srr (Hz) 17 513.9 7 - -
Output Beam Dim.
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
(MHz) 2.87 2.87 2.5 2.49 2.49
F
awf
APF (%)
(Acoustic Power up
Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
106.4 112.5 271.1 1174 1174
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode B B B B B
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
48.8 37.4 106 57.6 57.6
31.28 23.97 13.59 30.98 30.98
yes yes yes yes yes
1- 60 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 67
FDA Acoustic Output Data
3.8C-RC
Table 1-56: 3.8C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 1.97 1.98 2.03 1.66 1.66
(mW/cm2)
I
spta
(mm) 50 50 44 41 41
I
p
w
pb6
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
prr (kHz) - 513.9 - 3205 3205
srr (Hz) 18 513.9 5 - -
Output Beam
Dim.
• ( || ) (mm) 12 12 60 14.3 14.3
• ( ) (mm) 13 13 13 13 13
(MHz) 3.1 3.23 2.3 2.51 2.51
F
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
87.4 92.7 96.37 1232 1232
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization
Mode
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) Contact Contact Contact Contact Contact
I
ts
24.9 41.8 117 77.9 77.9
23.33 26.79 15 41.9 41.9
BB B BB
Yes Yes Yes Yes Yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-61
Direction 5264895-100 Rev. 1
Page 68
Acoustic Output
6.5C-RC
Table 1-57: 6.5C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.06 2.80 1.82 2.71 2.71
(mW/cm2)
I
spta
(mm) 15 15 20 18 18
I
p
w
pb6
• ( || ) (mm) 6.2 6.2 8.1 8.1 8.1
• ( ) ( m m ) 5555 5
prr (kHz) - 524.1 - 5204 5204
srr (Hz) 58 524.1 5 - -
Output Beam
Dim.
• ( || ) (mm) 6.2 6.2 8.1 8.1 8.1
• ( ) ( m m ) 5555 5
(MHz) 4.2 4.25 4.92 4.96 4.96
F
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
102.1 106.5 106.7 1273 1273
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
P o w e r - u p M o d e BBBB B
Initialization
Mode
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) contact contact contact contact contact
I
ts
25.1 21.6 19.7 28.5 28.5
80.97 69.68 48.64 70.37 70.37
BBBB B
y e sy e sy e sy e s y e s
1- 62 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 69
FDA Acoustic Output Data
E6C-RC
Table 1-58: E6C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.41 2.34 2.23 2.01 2.01
I
(mW/cm2)
spta
I
(mm) 19 19 17 18 18
p
w
pb6
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) ( m m )55555
prr (kHz) - 501.1 - 3930 3930
srr (Hz) 35 524.1 7 - -
Output Beam Dim.
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) ( m m )55555
(MHz) 4.46 4.49 4.91 4.98 4.98
F
awf
APF (%)
(Acoustic Power up
Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
93.6 95.8 98.2 398.7 398.7
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization Mode B B B B B
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
27.6 24.9 37.6 36.2 36.2
68.15 61.48 92.84 89.38 89.38
yes yes yes yes yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-63
Direction 5264895-100 Rev. 1
Page 70
Acoustic Output
E7C-RC
Table 1-59: E7C-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.48 2.64 2.81 2.64 2.64
I
(mW/cm2)
spta
I
(mm) 14 14 14 15 15
p
w
pb6
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) (mm) 5 5 5 5 5
prr (kHz) - 524.1 - 2040 2040
srr (Hz) 25 524.1 5 - -
Output Beam
Dim.
• ( || ) (mm) 8.1 8.1 8.1 8.1 8.1
• ( ) (mm) 5 5 5 5 5
(MHz) 4.96 4.96 4.93 4.98 4.98
F
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
87.5 91.3 1115.7 864.3 864.3
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization
Mode
Acoustic Output
Freeze
I
(mm) 0.0 0.0 0.0 0.0 0.0
tt
I
(mm) contact contact contact contact contact
ts
23.5 23.9 40.1 26.9 26.9
58.02 59.01 99.01 66.42 66.42
BBBBB
yes yes yes yes yes
1- 64 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 71
FDA Acoustic Output Data
3S-RC
Table 1-60: 3S-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 1.81 1.81 1.09 1.42 1.42
(mW/cm2)
I
spta
(mm) 33 33 35 18 18
I
p
w
pb6
• ( | | ) ( m m ) 99999
• ( ) (mm) 12 12 12 12 12
prr (kHz) - 1134.3 - 1132.2 1132.2
srr (Hz) 27 1134.3 17 - -
Output Beam Dim.
• ( | | ) ( m m ) 99999
• ( ) (mm) 12 12 12 12 12
(MHz) 1.76 1.76 1.76 2.51 2.51
F
awf
APF (%)
(Acoustic Power up
Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
281.12 259.2 184.5 457.3 457.3
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
P o w e r - u p M o d e BBBBB
I n i t i a l i z a t i o n M o d e BBBBB
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
82.1 82.1 146.8 44.2 44.2
76.02 76.02 135.9 40.93 40.93
yes yes yes yes yes
LOGIQ C5/C5 PRO Advanced Reference Manual 1-65
Direction 5264895-100 Rev. 1
Page 72
Acoustic Output
7L-RC
Table 1-61: 7L-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 2.96 2.71 2.94 2.32 2.32
(mW/cm2)
I
spta
(m m ) 11 11 11 21 21
I
p
w
pb6
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
• ( ) (mm) 4 4 4 4 4
prr (kHz) - 516.1 - 2604 2604
srr (Hz) 92 516.1 11 - -
Output Beam
Dim.
• ( || ) (mm) 38.4 38.4 38.4 23.84 3.84
• ( ) (mm) 4 4 4 4 4
(MHz) 6.64 6.78 5.0 4.99 4.99
F
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
79.4 82.6 185.2 963.4 963.4
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
Power-up Mode B B B B B
Initialization
Mode
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
(mm) contact contact contact contact contact
I
ts
22.4 16.2 32.8 30.5 30.5
14.58 10.55 21.35 19.86 19.86
BBB BB
yes yes yes yes yes
1- 66 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 73
FDA Acoustic Output Data
7.5L-RC
Table 1-62: 7.5L-RC IEC Acoustic Output Information
Parameter B B+M B+CFM B+PWD B+CFM+PWD
Pr_(MPa) 3.45 3.49 3.32 3.93 2.93
(mW/cm2)
I
spta
(mm) 13 13 11 17 17
I
p
w
pb6
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
• ( ) ( m m )4444 4
prr (kHz) - 516.1 - 2604 2604
srr (Hz) 77 516.1 5 - -
Output Beam
Dim.
• ( || ) (mm) 38.4 38.4 38.4 3.84 3.84
• ( ) ( m m )4444 4
(MHz) 4.80 5.04 5.0 4.99 4.99
F
awf
APF (%)
(Acoustic Power
up Fraction)
AIF (%)
(Acoustic
Initialization
Fraction)
92.3 96.8 125.2 1426 1426
100 100 100 100 100
100 100 100 100 100
Maximum Power
(mW)
(mW/cm2)
I
ob
P o w e r - u p M o d e BBBB B
Initialization
Mode
Acoustic Output
Freeze
(mm) 0.0 0.0 0.0 0.0 0.0
I
tt
I
(mm) contact contact contact contact contact
ts
28.5 25.3 38.5 32.8 32.8
18.55 16.47 25.07 21.35 21.35
BBBB B
y e sy e sy e sy e s y e s
LOGIQ C5/C5 PRO Advanced Reference Manual 1-67
Direction 5264895-100 Rev. 1
Page 74
Acoustic Output
1- 68 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 75
Chapter 2
Measurement Formulas
Reference for measurement formula tables for the
various applications.
LOGIQ C5/C5 PRO Advanced Reference Manual 2-1
Direction 5264895-100 Rev. 1
Page 76
Measurement Formulas
Formulas–Generic
Table 2-1: Generic Calculation Formulas
Calc
Mnemonic
Calc Name Input Measurements Formula
Measurement Formulas
MaxPG Maximum Pressure
Gradient
MeanPG Mean Pressure Gradient flow velocities from one
% Stenosis Stenosis Ratio two areas (by ellipse,
PI Pulsatility Index two Doppler blood flow
RI Resistivity Index two Doppler blood flow
HR Heart Rate (beats/minute) one 2 beat time interval HR[BPM]=120[sec]/
A/B Ratio Velocities Ratio two Doppler blood flow
TAMAX Time Averaged Maximum
Velocity (Trace Method is
Peak or manual)
TAMIN Time Averaged Minimum
Velocity (Trace method is
Floor)
two Doppler blood flow
peak velocities
time marker to another
time marker in a Doppler
display
trace, circle or distance)
peak velocities and
TAM AX
peak velocities
peak velocities
two time marks in a
Doppler display
two time marks in a
Doppler display
MaxPG[mmHg]=4x(v1^2v2^2)
MeanPG[mmHg]=
n
4x (Vi^2)/n
i=1
% Stenosis= [1-(A
A
)]x100
lumen
PI=(V
max-Vdiastole
a
TAM AX
RI=(V
max-Vdiastole
2beat time [sec]
A/B=V
1/V2
TAMAX=sum{Vt} from t1
to t2/(t2-t1) [cm/s or m/s]
TAMIN=sum{V
to t2/(t2-t1) [cm/s or m/s]
)/
)/V
} from t1
t
residual
max
/
a
TAMEAN Time Averaged Mean Velocity
(Trace method is Mean)
a.Vdiastole = Vmin or Vend-diastole (depends on preset selection)
two time marks in a
Doppler display
TAMEAN=sum{V
to t2/(t2-t1) [cm/s or m/s]
} from t1
t
2- 2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 77
Measurement Formulas
Table 2-2: Volume Calculation Formulas
Calc Name Input Measurements Formula
Volume (spherical) one distance
Volume (prolate spheroidal) two distances, d1>d2
Volume (prolate spheroidal) one ellipse, d1 major axis, d2
minor axis
Volume (spheroidal) three distances
Volume (spheroidal) one distance d1, one ellipse, d2
ma jor axis, d3 minor axis
Vol [ml ]=(
Vol[ml]= (
Vol[ml]= (
Vol[ml]= (
Vol[ml]= (
π/6)xd^3
π/6)xd1xd2^2
π/6)xd1xd2^2
π/6)xd1xd2xd3
π/6)xd1xd2xd3
LOGIQ C5/C5 PRO Advanced Reference Manual 2-3
Direction 5264895-100 Rev. 1
Page 78
Measurement Formulas
Formulas–OB
Table 2-3: OB Calculation Formulas
Calc
Mnemonic
AC Abdominal
BPD Biparietal
CRL Crown Rump
FL Femur Length one distance FL = -3.91 + 0.427 (GA) -
GS Gestational Sac three distances GS [wk] = 1.42450142 *
HC Head
HC Head
Calc Name
Circumference
Diameter
Length
Circumference
Circumference
Input
Measurements
circumference by
trace, ellipse, circle
or two distances
one distance BPD = -3.08+0.41 (GA) -
one distance CRL=1.684969+
circumference by
trace, ellipse, circle
or two distances
one ellipse HC [mm] = 2.325 * (BPD
AC=13.3+1.61 (GA) -
0.00998 (GA)
0.000061 (GA)
0.315646xd1+
0.049306xd1^2+
0.004057xd1^3+
0.000120456xd1^4
0.0034 (GA)
(d1+d2+d3)/3+ 3.6225
HC=-11.48 + 1.56 (GA) -
0.0002548 (GA)
[mm]^2 + OFD [mm]^2)^0.5
Formula
2
3
2
3
Author
Reference
Hadlock et al,
Radiology,
152:497-501,
1984
Hansmann,
Ultrasound
Diagnosis in
Obstetrics and
Gynecology,
438-9, 1985
EF Ejection
Fraction
two distances on MMode (End-diastolic
dimension and Endsystolic dimension
on M-Mode)
EF = (1 - Ds^3 / Dd^3) n/a
2- 4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 79
Table 2-4: CUA Hadlock Formulas
Measurement Formulas
Calc
Mnemonic
a
CUA
Calc
Name
Composite
Ultrasound
Age
Formula
1. CUA (BPD) = 9.54 + 1.482 * BPD + 0.1676 * BPD
2. CUA (HC) = 8.96 + 0.540 * HC + 0.0003 * HC
3. CUA (AC) = 8.14 + 0.753 * AC + 0.0036 * AC
4. CUA (FL) = 10.35 + 2.460 * FL + 0.170 * FL
5. CUA (BPD, HC) = 10.32 + 0.009 * HC2 + 1.3200 *
BPD + 0.00012 * HC
6. CUA (BPD, AC) = 9.57 + 0.524 * AC + 0.1220 * BPD
7. CUA (BPD, FL) = 10.50 + 0.197 * BPD * FL + 0.9500
* FL + 0.7300 * BPD
8. CUA (HC, AC) = 10.31 + 0.012 * HC
9. CUA (HC, FL) = 11.19 + 0.070 * HC * FL + 0.2630 *
HC
10. CUA (AC, FL) = 10.47 + 0.442 *AC + 0.3140 * FL
0.0121 * FL
11. CUA (BPD, HC, AC) = 10.58 + 0.005 * HC2 + 0.3635
* AC + 0.02864 * BPD * AC
12. CUA (BPD, HC, FL) = 11.38+0.070 * HC * FL +
0.9800 * BPD
13. CUA (BPD, AC, FL) = 10.61 + 0.175 * BPD * FL +
0.2970 * AC + 0.7100 * FL
14. CUA (HC, AC, FL) = 10.33 + 0.031 * HC * FL +
0.3610 * HC + 0.0298 * AC * FL
15. CUA (BPD, HC, AC, FL) = 10.85 + 0.060 * HC * FL +
0.6700 * BPD + 0.1680 * AC
3
3
2
+ 0.3850 * AC
2
3
2
2
2
-
Author
Reference
Hadlock,
Radiology,
1984 152:497501
2
a.Formulas are used only if Hadlock HC, FL, AC and BPD are used and CUA is selected as the preset
in the CUA/AUA for Hadlock preset in the System M&A Preset Menu. If other authors are used, CUA
automatically changes to AUA and an average value is displayed.
Table 2-5: EFW Calculation Formulas
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
Calc Name
Weight
Weight
Input
Measurements
AC and HC EFW [kg] = 10^
(1.182+(0.0273*HC
[cm])+(0.07057*AC [cm] (0.00063*AC [cm]^2)(0.0002184*AC [cm] * HC
[cm]))
BPD, AC, and FL EFW [g]=10^(1.335-
(0.0034*AC [cm] * FL [cm]) +
(0.0316 * BPD [cm]) +
(0.0457 * AC [cm]) + (0.1623
* FL [cm]
Formula
Author
Reference
Hadlock,
Radiology,
150:535, 1984
Hadlock,
AJOG,
151:333, 1985
LOGIQ C5/C5 PRO Advanced Reference Manual 2-5
Direction 5264895-100 Rev. 1
Page 80
Measurement Formulas
Table 2-5: EFW Calculation Formulas (Continued)
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
EFW Estimated Fetal
Calc Name
Weight
Weight
Weight
Weight
Weight
Input
Measurements
AC, HC, and FL EFW [g]=10^(1.326-
(0.00326 * AC [cm] * FL
[cm]) + (0.0107 * HC [cm]) +
(0.0438 * AC [cm]) + (0.158
* FL [cm]))
AC, HC, BPD, FL EFW [g] = 10^(1.3596-
(0.00386 * AC [cm] * FL
[cm]) + (0.0064 * HC [cm]) +
(0.00061 * BPD [cm] * AC
[cm]) + (0.0424 * AC [cm]) +
(0.174 * FL [cm]))
AC and FL EFW [g] = 10^(1.304 +
0.05281 * AC [cm] + 0.1938
* FL [cm] - 0.004 * AC [cm] *
FL [cm])
EFW [g] = -3200.40479 +
157.07186 * AC [cm] +
15.90391 * BPD [cm]^2
EFW [g] = 0.515263 -
0.105775 * BPD [mm] +
(0.000930707 * BPD [mm]^2
+ 0.0649145 * TAD [mm] -
0.00020562 * TAD [mm]^2
Formula
Author
Reference
Hadlock,
AJOG,
151:333, 1985
Hadlock,
AJOG,
151:333, 1985
Hadlock,
Radiology,
150:535, 1984
Merz
German
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
EFW Estimated Fetal
Weight
AC and BPD EFW [kg] = 10^(-1.7492 +
BPD [cm] and
AC [cm]
BPD [cm] and
AC [cm]
BPD, AxT, FL [cm] EFW1 [g] =1.07 * BPD
0.166 * BPD [cm] + 0.046 *
AC [cm] - 2.646 * AC [cm] *
BPD [cm]/1000)
EFW [g] =
10^(1.7288+0.09184 * BPD
[cm] + 0.02581 * AC [cm] +
0.00011 * BPD [cm] * AC
[cm])
EFW [g] =10^(3-1.7492 +
(0.166 * BPD [cm]) + (0.04 *
A [cm]) - (0.002646 * AC
[cm] * BPD [cm]))
EFW [g] = 1.07 * BPD
[cm]^3 + 3.42 * APTD [cm] *
TTD [cm] * FL [cm]
[cm]^3 + 3.42 * AxT [cm2] *
FL [cm]
Shepard,
AJOG, 142:47,
1982
Shepard/
Warsoff
Richards/
Berkowitz
Tokyo
University
Tokyo
Shinozuka
2- 6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 81
Measurement Formulas
Table 2-5: EFW Calculation Formulas (Continued)
Calc
Mnemonic
EFW Estimated Fetal
EFW Estimated Fetal
Amniotic Fluid
Index (AFI)
Calc Name
Weight
Weight
The normal values are considered to be:
36-40 weeks
0-5 cm = very low
5.1-8.0 cm = low
8.1 - 18.0 cm = normal
>18.0 = high
Dr. Rutherford/Dr. Phelan, Obstetrics and Gynecology, Volume
70, No. 3, Part 1, p.353-6, Sept. 1987.
28-40 weeks
15.0 cm = average
Input
Measurements
BPD, AC, FL [cm] EFW2 [g] = 1.07 * BPD
[cm]^3 + 0.30 * AC [cm]^2 *
FL [cm]
BPD, AxT, SL [cm] EFW3 [g] = 1.07 * BPD [cm]
^3 + 2.91 * AxT [cm2] *
SL[cm]
Formula
Sagittal
Author
Reference
Tokyo
Shinozuka
Tokyo
Shinozuka
>20.0 - 24.0 = hydramnios
<5.0-6.0 = Oligohydramnios
Transverse
Dr. C.C. Smith, The Female Patient, Volume 15, p.85-97, March
1990.
LOGIQ C5/C5 PRO Advanced Reference Manual 2-7
Direction 5264895-100 Rev. 1
Page 82
Measurement Formulas
Formulas–GYN
Table 2-6: GYN Calculation Formulas
Calc
Mnemonic
UT-L Uterine Length one distance Ut-L[cm or mm]=d1
UT-H Uterine Height one distance Ut-H[cm or mm]=d1
UT-W Uterine Width one distance Ut-W[cm or mm]=d1
UT-Volume Uterine Volume
UtPFD Uterus Portio-Fundus
UtAP Anterior-Posterior Uterus
UtQ Transverse Uterus
Endo Endometrium Thickness one distance Endo[cm or mm]=d1
Lt. Ov-L Left Ovarian Length one distance Lt. Ov-L[cm or mm]=d1
Lt. Ov-H Left Ovarian Height one distance Lt. Ov-H[cm or mm]=d1
Lt. Ov-W Left Ovarian Width one distance Lt. Ov-W[cm or mm]=d1
Lt. Ov-Volume Left Ovarian Volume
Rt. Ov-L Right Ovarian Length one distance Rt. Ov-L[cm or mm]=d1
Calc Name
Distance
Diameter
Diameter
Input
Measurements
Formula
Rt. Ov-H Right Ovarian Height one distance Rt. Ov-H[cm or mm]=d1
Rt. Ov-W Right Ovarian Width one distance Rt. Ov-W[cm or mm]=d1
Rt. Ov-Volume Right Ovarian Volume
Lt. Ov-RI Left Ovarian Vessel
Resistive Index
Ut-RI Uterine Vessel Resistive
Index
Rt. Ov-RI Right Ovarian Vessel
Resistive Index
LtOvFo[ml] Left Ovary Follicles
two Doppler blood
flow peak velocities
two Doppler blood
flow peak velocities
two Doppler blood
flow peak velocities
One distance
Two distances
Three distances
b
b
Lt. Ov-RI= (V
Ut-RI= (V
Rt. Ov-RI= (V
V
max
D1[cm]3 x π /6
D1[cm]2 x D2x[cm] π /6
: (D1 <D2)
D1[cm] x D2[cm]
: (D2 <D1)
D1[cm] x D2[cm]x D3[cm] x
max-Vdiastole
a
max-Vdiastole
)/V
max-Vdiastole
2
x π /6
)/V
max
)/
max
a
π/6
a
2- 8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 83
Measurement Formulas
Table 2-6: GYN Calculation Formulas
Calc
Mnemonic
RtOvFo[ml] Right Ovary Follicles
Lt. Ov-PI Left Ovarian Vessel
Rt. Ov-PI Right Ovarian Vessel
diastole
= V
min
a.V
b.To calculate LtOvFo or RtOvFo with one (or two distances), press the Clear key after the first (or
second distance) measurement(s).
Calc Name
Pulsatility Index
Pulsatility Index
or V
end-diastole
(depends on preset selection)
Input
Measurements
One distance
Two distances
Three distances
b
Formula
D1[cm]3 x π /6
b
D1[cm]2 x D2[cm] x π /6
: (D1 <D2)
D1[cm] x D2[cm]
: (D2 <D1)
D1[cm] x D2[cm] x D3[cm] x
2
x π /6
π/6
LOGIQ C5/C5 PRO Advanced Reference Manual 2-9
Direction 5264895-100 Rev. 1
Page 84
Measurement Formulas
Formulas–Vascular
Table 2-7: Vascular Calculation Formulas
Calc
Mnemonic
Calc Name Input Measurements Formula
RT ECA Right External Carotid Artery
Vel oci ty
RT CCA Right Common Carotid Artery
Vel oci ty
RT BIFURC Right Carotid Bifurcation
Vel oci ty
RT ICA Right Internal Carotid Artery
Vel oci ty
RT ICA/CCA Right Internal Carotid Artery
Velocity/Common Carotid Artery
one Doppler blood flow
peak velocity
one Doppler blood flow
peak velocity
one Doppler blood flow
peak velocity
one Doppler blood flow
peak velocity
two Doppler blood flow
peak velocities
Velocity Ratio
LT E C A ,
LT C C A ,
Same as above, for Left Carotid
Artery
Same as above Same as above
LT BIFURC,
LT I C A ,
LT ICA/CCA
A/B Ratio Velocities Ratio two Doppler blood flow
peak velocities
% Stenosis Stenosis Ratio two areas (by ellipse,
trace,
circle or distance)
S/D Ratio Systolic Velocity/Diastolic
Velocities Ratio
two Doppler blood flow
peak velocities
RT ECA=v1
[cm/s or m/s]
RT CCA=v1
[cm/s or m/s]
RT BIFURC=v1
[cm/s or m/s]
RT ICA=v1
[cm/s or m/s]
RT ICA/CCA=V
V
CCA
A/B=V
1/V2
% Stenosis=[1(A
S/D=V
/ A
residual
systole/Vdiastole
lumen
/
ICA
)]x100
a
PI Pulsatility Index two Doppler blood flow
peak velocities and
TAM AX
RI Resistivity Index two Doppler blood flow
peak velocities
HR Heart Rate (beats/minute) one 2 beat time interval
(measured manually or
PI=(V
max-Vdiastole
a
TAM AX
RI=(V
max-Vdiastole
a
V
max
HR[BPM]=120[sec]/
2 beat time[sec]
)/
)/
automatically)
a.V
diastole
= V
min
or V
end-diastole
(depends on preset selection)
2- 10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 85
Chapter 3
OB Tables
Reference for Obstetric Measurement Tables.
LOGIQ C5/C5 PRO Advanced Reference Manual 3-1
Direction 5264895-100 Rev. 1
Page 86
OB Tables
OB Tables
ASUM
Table 3-1: AC: ASUM, Deler (Fetal Age)
Unit: AC (mm); Age (Days); 2SD (Days)
AC Age 2SD AC Age 2SD AC Age 2SD AC Age 2SD
<35
35
46
57
69
80
92
103
114
n/a
70
77
84
91
98
105
112
119
——
8
8
8
8
9
9
9
9
126
137
149
160
171
183
194
206
217
126
133
140
147
154
161
168
175
182
10
10
10
10
10
10
12
12
12
228
240
251
263
274
285
297
308
320
189
196
203
210
217
224
231
238
245
14
14
14
14
14
16
16
18
18
331
342
354
365
377
>377
252
259
266
273
280
n/a
18
18
20
20
20
——
Table 3-2: BPD: ASUM, Aust NZ, Obstet Gynaecol 1989: 29:26 (Fetal Age)
Unit: BPD (mm); Age (Week); 2SD (Week - * signifies No Data)
BPD Age 2SD BPD Age 2SD BPD Age 2SD BPD Age 2SD
<20
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
n/a
84
86
88
90
92
94
95
97
99
101
103
105
107
109
111
113
115
117
119
121
——
4
4
4
4
5
5
5
5
5
6
6
6
6
7
7
7
7
8
8
8
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
123
126
128
130
132
134
136
139
141
143
145
147
149
152
154
157
159
161
164
166
169
8
9
9
9
9
9
10
10
10
10
11
11
11
11
12
12
12
12
13
13
13
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
171
173
176
178
181
183
186
188
191
193
196
199
201
204
206
209
212
214
217
220
222
13
13
14
14
14
14
15
15
15
15
16
16
16
16
17
17
17
17
17
18
18
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
>98
225
228
231
234
237
240
244
247
251
255
259
264
270
276
284
292
301
n/a
18
18
19
0
0
0
0
0
0
0
0
0
0
0
0
0
0
——
3- 2 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 87
OB Tables
Table 3-3: CRL: ASUM, Silva et al 1991.6 (Fetal Age)
Unit: CRL (mm); Age (Days); 2SD (* No Data available)
CRL Age 2SD CRL Age 2SD CRL Age 2SD CRL Age 2SD
<2
2
3
4
5
6
7
8
9
10
11
12
13
14
15
n/a
42
43
44
45
46
47
48
49
50
51
52
53
54
55
——
*
*
*
*
*
*
*
*
*
*
*
*
*
*
16
17
18
19
20
22
23
24
25
26
27
29
30
31
33
56
57
58
59
60
61
62
63
64
65
66
67
68
69
70
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
34
36
37
38
40
41
43
45
46
48
50
51
53
55
57
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
*
*
*
*
*
*
*
*
*
*
*
*
*
*
*
58
60
62
64
66
68
70
72
74
76
78
80
82
>82
86
87
88
89
90
91
92
93
94
95
96
97
98
n/a
Berkowitz
Table 3-4: BD: Berkowitz (Fetal Age)
Unit: BD (mm); Age (Day); SD (mm)
BD Age SD BD Age SD BD Age SD BD Age SD
<13
13
14
15
16
17
18
19
20
21
22
23
24
n/a
81
82
84
86
88
91
95
98
102
105
109
110
——
0
0
0
0
0
0
0
0
0
0
0
0
25
26
27
28
29
30
31
32
33
34
35
36
37
112
116
120
124
128
130
132
135
138
142
145
149
152
0
0
0
0
0
0
0
0
0
0
0
0
0
38
39
40
41
42
43
44
45
46
47
48
49
50
155
159
162
166
169
173
179
185
191
197
202
207
212
0
0
0
0
0
0
0
0
0
0
0
0
0
51
52
53
54
55
56
57
58
59
>59
217
223
230
237
244
251
258
266
275
n/a
*
*
*
*
*
*
*
*
*
*
*
*
*
——
0
0
0
0
0
0
0
0
0
——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-3
Direction 5264895-100 Rev. 1
Page 88
OB Tables
Campbell
Table 3-5: HC/AC Ratio: Campbell, Br J Obstet Gynaecol 1977, 84:165-174
(Fetal Growth)
Unit: GA (Weeks); Min/Max (Index)
GA Min Max GA Min Max GA Min Max
<13
13
15
17
19
21
n/a
1.14
1.05
1.07
1.09
1.06
——
1.31
1.39
1.29
1.26
1.25
23
25
27
29
31
33
1.05
1.04
1.05
0.99
0.96
0.96
1.21
1.22
1.22
1.21
1.17
1.11
35
37
39
41
>42
0.93
0.92
0.87
0.93
n/a
1.11
1.05
1.06
1.00
n/a
Eriksen
Table 3-6: TAD: Eriksen (Fetal Age)
Unit: TAD (mm); Age (Day); SD (mm)
TAD Age SD TA D Age SD TAD Age SD TAD Age SD
<23
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
n/a
91
93
95
97
99
101
103
105
107
109
111
113
115
117
119
120
122
124
126
128
130
132
——
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
65
66
67
134
136
138
140
142
144
146
148
150
152
154
156
158
161
163
165
167
169
171
173
175
177
179
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
182
184
186
188
190
192
195
197
199
201
203
206
208
210
212
214
217
219
221
223
225
228
230
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
91
92
93
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
>112
232
234
236
239
241
243
245
247
250
252
254
256
259
261
263
266
268
270
273
275
277
280
n/a
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
——
3- 4 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 89
French College of Fetal Ultrasound (CFEF)
Table 3-7: BPD
GA
(weeks)
11 12.08 13.12‘ 15.36 17.60 18.63
12 15.81 16.96 19.40 21.81 22.92
13 19.47 20.71 23.30 25.92 27.12
14 23.05 24.36 27.14 29.92 31.23
15 26.56 27.93 30.89 33.82 35.23
16 29.97 31.41 34.53 37.62 39.08
17 33.32 34.85 38.12 41.35 42.87
18 36.55 38.15 41.58 44.97 46.56
19 39.76 41.46 45.00 48.52 50.18
20 42.85 44.56 48.22 51.90 53.64
21 45.86 47.66 51.43 55.23 57.00
22 48.79 50.61 54.53 58.44 60.30
23 51.63 53.48 57.51 61.54 63.45
24 54.38 56.31 60.42 64.57 66.50
25 57.04 59.00 63.25 67.48 69.42
26 59.62 61.64 65.94 70.24 72.27
27 62.12 64.15 68.55 72.62 75.00
28 64.50 66.61 71.03 75.52 77.60
29 66.84 68.98 73.50 77.97 80.09
30 69.07 71.21 75.80 80.37 82.52
31 71.22 73.39 78.00 82.63 84.80
32 73.30 75.49 80.16 84.80 87.00
33 75.24 77.46 82.14 86.84 89.04
34 77.14 79.36 84.07 88.8 91.00
35 78.94 81.14 85.90 90.61 92.83
36 80.64 82.88 87.61 92.35 94.56
37 82.27 84.50 89.24 93.97 96.19
38 83.78 86.00 90.70 95.42 97.66
39 85.22 87.43 92.10 96.86 99.05
40 86.47 88.78 93.45 98.13 100.31
41 87.00 89.00 94.00 99.00 101.00
3% 10% 50% 90% 97%
OB Tables
LOGIQ C5/C5 PRO Advanced Reference Manual 3-5
Direction 5264895-100 Rev. 1
Page 90
OB Tables
Table 3-8: FL
GA
(weeks)
12 2.76 3.89 6.33 8.79 10.00
13 6.09 7.29 9.88 12.42 13.65
14 9.40 10.65 13.33 16.00 17.27
15 12.56 13.87 16.66 19.44 20.77
16 15.70 17.00 19.95 22.80 24.18
17 18.74 20.12 23.12 26.13 27.53
18 21.69 23.14 26.23 29.30 30.80
19 24.59 26.06 29.25 32.44 33.91
20 27.42 28.94 32.23 35.48 37.03
21 30.12 31.72 35.05 38.41 40.00
22 32.83 34.39 37.87 41.30 42.91
23 35.34 37.00 40.50 44.03 45.71
24 37.89 39.58 43.16 46.75 48.42
25 40.33 42.04 45.69 49.36 51.08
26 42.66 44.40 48.17 51.88 53.62
27 44.95 46.72 50.53 54.32 56.09
28 47.13 48.94 52.80 56.64 58.45
29 49.22 51.06 54.94 58.91 60.72
30 51.30 53.14 57.13 61.08 62.92
31 53.26 55.13 59.15 63.14 65.04
32 55.12 57.04 61.11 65.19 67.07
33 56.96 58.87 63.00 67.10 69.03
34 58.69 60.62 64.76 68.88 70.84
35 60.33 62.29 66.47 70.65 72.63
36 61.90 63.89 68.13 72.34 74.30
37 63.40 65.36 69.63 73.91 75.89
38 64.81 66.79 71.11 75.38 77.41
39 66.16 68.19 72.48 76.81 78.84
40 67.42 69.47 73.79 78.14 80.17
41 68.00 70.00 74.00 79.00 81.00
3% 10% 50% 90% 97%
3- 6 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 91
OB Tables
Table 3-9: HC
GA
(weeks)
16 105.80 110.58 120.86 131.25 136.11
17 118.67 123.78 134.49 145.38 150.39
18 131.08 136.36 147.55 158.92 164.11
19 143.00 148.53 160.29 172.14 177.48
20 154.53 160.21 172.47 184.86 190.54
21 165.41 171.49 184.21 197.12 203.09
22 176.12 182.35 ‘195.74 208.91 215.15
23 186.32 192.31 206.64 220.26 226.76
24 196.19 203.00 217.18 231.39 238.00
25 205.50 212.40 227.32 241.91 248.81
26 214.44 221.57 236.72 252.00 259.23
27 222.87 230.33 246.00 261.75 269.13
28 231.00 238.56 254.77 271.00 278.57
29 238.40 246.35 263.00 279.71 287.56
30 245.86 253.74 270.84 288.13 296.00
31 252.54 260.81 278.33 296.00 304.27
32 258.86 267.22 285.29 303.54 312.00
33 264.62 278.38 292.00 310.40 319.10
34 270.14 279.00 298.10 317.00 325.91
35 275.33 284.23 303.62 323.00 332.16
36 279.79 289.00 308.81 328.75 338.00
37 283.90 293.32 313.52 334.00 343.34
38 287.63 297.29 317.88 338.64 348.29
39 290.88 300.76 321.86 343.00 352.67
40 293.00 303.00 324.00 346.00 356.00
3% 10% 50% 90% 97%
LOGIQ C5/C5 PRO Advanced Reference Manual 3-7
Direction 5264895-100 Rev. 1
Page 92
OB Tables
Table 3-10: AC
GA
(weeks)
15 80.70 85.30 95.00 104.40 108.80
16 91.30 96.10 106.40 116.80 121.60
17 101.70 106.80 118.00 129.00 134.00
18 111.80 117.40 129.20 141.00 146.60
19 122.00 128.00 140.40 153.00 158.80
20 132.00 138.00 151.40 164.70 171.00
21 141.60 148.20 162.30 176.30 183.00
22 151.40 158.20 173.00 187.80 194.70
23 160.90 168.20 183.60 199.00 206.30
24 170.20 177.80 194.00 210.30 218.00
25 179.30 187.30 204.40 221.30 229.30
26 188.40 196.70 214.50 232.30 240.60
27 197.30 206.00 224.50 243.00 251.60
28 206.20 215.10 234.40 253.60 262.60
29 214.70 224.00 244.00 264.00 273.30
30 223.20 232.80 253.60 274.20 283.70
31 231.60 241.60 263.00 284.20 294.40
32 239.70 250.00 272.20 294.30 304.60
33 247.80 258.40 281.20 304.00 314.80
34 255.60 266.70 290.20 313.80 324.80
35 263.20 274.70 298.80 323.30 334.50
36 271.00 282.60 307.40 332.50 344.30
37 278.30 290.30 316.00 341.70 353.80
38 285.60 298.00 324.70 350.70 363.00
39 292.70 305.30 332.40 359.60 372.20
40 298.00 311.00‘ 339.00 367.00 380.00
3% 10% 50% 90% 97%
3- 8 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 93
OB Tables
Table 3-11: TAD
GA
(weeks)
11 9.68 11.00 13.50 16.00 17.25
12 12.68 14.00 17.00 20.00 21.46
13 15.60 17.25 20.56 24.00 25.51
14 18.69 20.41 24.00 27.84 29.56
15 21.76 23.64 27.69 31.74 33.61
16 25.00 27.00 31.21 35.53 37.48
17 28.23 30.34 34.70 39.21 41.39
18 31.54 33.64 38.31 42.89 45.14
19 34.78 37.00 41.69 46.42 48.58
20 38.16 40.26 45.21 50.00 52.20
21 41.14 43.46 48.34 53.22 55.63
22 44.21 46.61 51.57 56.75 59.08
23 47.00 49.47 54.72 60.00 62.46
24 49.77 52.39 57.88 63.43 66.00
25 52.54 55.18 61.00 66.74 69.44
26 55.17 58.00 64.00 70.12 72.89
27 57.72 60.73 67.11 73.42 76.42
28 60.43 63.58 70.27 76.80 79.87
29 63.13 66.36 73.27 80.17 83.33
30 65.80 69.17 76.17 83.45 86.75
31 68.35 71.88 79.25 86.68 90.13
32 70.90 74.43 82.10 89.76 93.36
33 73.08 76.75 84.78 92.89 96.64
34 75.25 79.08 87.55 95.89 99.86
35 77.00 81.10 90.00 99.00 103.00
36 78.48 82.90 92.36 102.00 106.31
37 79.79 84.60 94.81 105.00 109.67
38 80.92 86.10 97.00 108.19 113.29
39 81.85 87.41 99.33 111.34 117.00
40 82.58 88.59 101.64 114.52 120.70
41 82.80 89.20 103.00 117.00 123.00
3% 10% 50% 90% 97%
LOGIQ C5/C5 PRO Advanced Reference Manual 3-9
Direction 5264895-100 Rev. 1
Page 94
OB Tables
Goldstein
Table 3-12: TCD: Goldstein et a, Am J OB/GYN, May 1987 (Fetal Growth)
Unit: TCD (Weeks); Age/Quat1/Mean/Quat3/Max (mm)
Age Min Quat1 Mean Quat3 Max
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
10
14
16
17
18
18
19
21
22
22
23
25
26
27
29
31
32
33
32
33
31
36
37
40
52
12
16
17
18
18
19
20
23
23
24
21.5
28
28.5
30
32
32
35
36
36
38
37
29
37
40
52
14
16
17
18
19
20
22
23
24
25
28
29
30
31
34
35
38
38
40
40
40.5
43
45
48.5
52
15
16
18
19
19
20
23
24
25
27
28
30
31
32
36
37
39
40
43
41
43
52
52
52
55
16
17
18
19
22
22
24
24
26
28
29
32
32
34
38
40
43
42
44
44
47
55
55
55
55
3- 10 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 95
OB Tables
Hadlock
Table 3-13: AC: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: AC (mm); Age (Week); SD (Week)
AC Age SD AC Age SD AC Age SD AC Age SD
<50
50
55
60
65
70
75
80
85
90
95
100
105
110
115
120
123
124
125
130
n/a
12.0
12.4
12.8
13.2
13.6
14.0
14.4
14.8
15.2
15.6
16.0
16.4
16.9
17.3
17.7
17.9
18.0
18.1
18.5
——
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 2.1
± 2.1
± 2.1
135
140
145
150
155
160
165
170
175
180
185
190
192
193
195
200
205
210
215
220
19.0
19.4
19.8
20.2
20.7
21.1
21.5
22.0
22.4
22.9
23.3
23.7
23.9
24.0
24.2
24.6
25.1
25.5
26.0
26.4
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
225
230
235
240
245
250
255
258
259
260
265
270
275
280
285
290
295
300
305
310
26.9
27.4
27.8
28.3
28.7
29.2
29.7
30.0
30.1
30.2
30.6
31.1
31.6
32.0
32.5
33.0
33.5
34.0
34.5
34.9
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
315
320
321
325
330
335
340
345
350
355
360
365
370
375
380
385
>385
35.4
35.9
36.0
36.4
36.9
37.4
37.9
38.4
38.9
39.4
39.9
40.4
40.9
41.4
42.0
42.5
n/a
± 3.0
± 3.0
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
——
Table 3-14: AC: Hadlock, AJR; 139: 367-370; 1982 (Fetal Age)
Unit: AC (mm); Age (Days); SD (Days)
AC Age SD AC Age SD AC Age SD AC Age SD
<47
47
53
60
67
74
80
87
93
100
106
113
119
126
132
n/a
84
87
91
94
98
101
105
106
112
115
119
122
126
129
——
13
13
13
13
13
13
13
13
13
13
13
13
13
14
138
144
151
157
163
174
180
186
192
197
203
208
214
219
225
133
136
140
143
147
154
157
161
164
168
171
175
178
182
185
14
14
14
14
14
14
14
14
14
14
15
15
15
15
15
230
235
241
246
251
256
261
266
271
276
281
286
291
296
300
189
192
196
199
203
206
210
213
217
220
224
227
231
234
238
15
15
15
15
15
15
15
21
21
21
21
21
21
21
21
305
310
314
319
323
328
332
337
341
344
349
353
>353
241
245
248
252
255
259
262
266
269
273
276
280
n/a
21
21
21
21
18
18
18
18
18
18
18
18
——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-11
Direction 5264895-100 Rev. 1
Page 96
OB Tables
a
Table 3-15: BPD: Hadlock, Radiology 1984, Vol. 152:497
Unit: BPD (mm); Age (Week); SD (Week)
BPD Age SD BPD Age SD BPD Age SD BPD Age SD
(Fetal Age)
<14
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
n/a
11. 9
12.1
12.3
12.5
12.8
13.0
13.2
13.4
13.6
13.8
14.1
14.3
14.5
14.8
15.0
15.2
15.5
15.7
16.0
16.3
16.5
16.8
a.Variability of GA estimate by BPD at term is ± 2 SD (6 weeks)
——
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
17.0
17.3
17.6
17.9
18.1
18.4
18.7
19.0
19.3
19.6
19.9
20.2
20.5
20.8
21.1
21.5
21.8
22.1
22.4
22.8
23.1
23.4
23.8
± 1.2
± 1.2
± 1.2
± 1.2
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
± 1.7
59
60
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
24.1
24.5
24.8
25.2
25.5
25.9
26.3
26.6
27.0
27.4
27.7
28.1
28.5
28.9
29.3
29.7
30.1
30.5
30.9
31.3
31.7
32.1
32.5
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 2.2
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
101
102
103
>103
Table 3-16: BPD: Hadlock, J Ultrasound Med 1:97-104, April 1982 (Fetal Age)
Unit: BPD (mm); Age (Days); SD (Days)
33.0
33.4
33.8
34.2
34.7
35.1
35.6
36.0
36.5
36.9
37.4
37.8
38.3
38.7
39.2
39.7
40.2
40.6
41.1
41.6
42.1
42.6
n/a
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
± 3.2
——
BPD Age SD BPD Age SD BPD Age SD BPD Age SD
<20
20
21
22
23
24
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
n/a
85
88
90
92
93
95
97
99
102
103
105
107
109
111
113
116
118
120
122
124
——
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
126
128
130
132
134
137
139
141
144
146
148
151
153
155
158
160
162
163
167
169
172
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
9
9
61
62
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
175
177
180
183
185
188
190
193
196
198
201
204
207
209
213
216
218
221
224
228
230
9
9
9
9
9
9
9
9
9
9
9
9
9
9
14
14
14
14
14
14
14
82
83
84
85
86
87
88
89
90
91
92
93
94
95
96
97
98
99
100
>100
233
237
239
243
246
249
253
256
259
263
266
270
272
276
279
284
287
291
294
n/a
14
14
14
14
14
14
25
25
25
25
25
25
25
25
25
25
25
25
25
——
3- 12 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 97
OB Tables
Table 3-17: CI: Hadlock, AJR, 137: 83, 1981 (Fetal Growth)
Min (Index) Max (Index)
70 86
Table 3-18: CRL: Hadlock, Radiology 1992, Vol. 182:501 (Fetal Age)
Unit: CRL (mm); Age (Week); SD (Week)
CRL Age SD CRL Age SD CRL Age SD CRL Age SD
<2
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
n/a
5.7
5.9
6.1
6.2
6.4
6.6
6.7
6.9
7.1
7.2
7.4
7.5
7.7
7.9
8.0
8.1
8.3
8.4
8.6
8.7
8.9
9.0
9.1
9.2
9.4
9.5
9.6
9.7
9.9
10.0
——
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.3
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
± 0.4
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
10.1
10.2
10.3
10.4
10.5
10.6
10.7
10.8
10.9
11. 0
11. 1
11. 2
11. 2
11. 3
11. 4
11. 5
11. 6
11. 7
11. 7
11. 8
11. 9
12.0
12.0
12.1
12.2
12.3
12.3
12.4
12.5
12.6
12.6
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.5
± 0.6
± 0.6
± 0.6
± 0.6
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
84
85
86
87
88
89
90
91
92
93
12.7
12.8
12.8
12.9
13.0
13.1
13.1
13.2
13.3
13.4
13.4
13.5
13.6
13.7
13.8
13.8
13.9
14.0
14.1
14.2
14.2
14.3
14.4
14.5
14.6
14.7
14.8
14.9
15.0
15.1
15.2
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.6
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
94
95
96
97
98
99
100
101
102
103
104
105
106
107
108
109
110
111
112
113
114
115
116
117
118
119
120
121
>121
15.3
15.3
15.4
15.5
15.6
15.7
15.9
16.0
16.1
16.2
16.3
16.4
16.5
16.6
16.7
16.8
16.9
17.0
17.1
17.2
17.3
17.4
17.5
17.6
17.7
17.8
17.9
18.0
n/a
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.7
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
± 0.8
——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-13
Direction 5264895-100 Rev. 1
Page 98
OB Tables
Table 3-19: EFW: Hadlock (Fetal Age)
Unit: EFW (grams); Mean (Weeks); SD (grams)
EFW Mean SD EFW Mean SD EFW Mean SD
<35
35
45
58
73
93
117
146
181
223
273
331
n/a
10
11
12
13
14
15
16
17
18
19
20
——
4
6
7
9
12
15
19
23
28
35
42
399
478
568
670
785
913
1055
1210
1379
1559
1751
1953
21
22
23
24
25
26
27
28
29
30
31
32
51
61
72
85
101
116
134
154
175
198
222
248
2162
2377
2595
2813
3028
3236
3435
3619
3787
>3787
33
34
35
36
37
38
39
40
41
n/a
275
302
330
357
385
411
436
460
481
——
Table 3-20: FL: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: FL (mm); Age (Week); SD (Week)
FL Age SD FL Age SD FL Age SD FL Age SD
<6
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
n/a
11. 9
12.2
12.4
12.7
13.0
13.3
13.5
13.8
14.1
14.4
14.7
15.0
15.3
15.6
16.0
16.3
16.6
16.9
17.2
——
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
± 1.4
25
26
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
17.6
17.9
18.2
18.6
18.9
19.3
19.6
20.0
20.3
20.7
21.0
21.4
21.8
22.2
22.5
22.9
23.3
23.7
24.1
24.5
± 1.4
± 1.4
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 1.8
± 2.1
± 2.1
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
63
64
24.9
25.3
25.7
26.1
26.5
26.9
27.3
27.7
28.2
28.6
29.0
29.5
29.9
30.3
30.8
31.2
31.7
32.1
32.6
33.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
80
81
82
83
>83
33.5
34.0
34.5
34.9
35.4
35.9
36.4
36.9
37.4
37.9
38.4
38.9
39.4
39.9
40.4
40.9
41.4
42.0
42.5
n/a
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
± 3.1
——
3- 14 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1
Page 99
OB Tables
Table 3-21: FL: Hadlock, AJR 138: 875-878, May 1982 (Fetal Age)
Unit: FL (mm); Age (Days); SD (Days)
FL Age SD FL Age SD FL Age SD FL Age SD
<10
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
n/a
90
92
94
95
97
99
101
104
106
108
110
112
114
116
118
120
123
——
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
27
28
29
30
31
32
33
34
35
36
37
38
39
40
41
42
43
44
125
127
130
132
134
137
139
142
145
147
150
153
155
157
160
163
166
169
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
10
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
60
61
62
171
174
177
180
183
185
189
192
195
197
201
204
207
210
213
216
220
223
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
10
63
64
65
66
67
68
69
70
71
72
73
74
75
76
77
78
79
>79
226
230
233
237
239
243
246
250
253
257
260
264
268
272
275
279
283
n/a
10
10
10
10
10
10
10
10
11
11
11
11
11
11
11
11
11
——
Table 3-22: HC: Hadlock, Radiology 1984, Vol. 152:497 (Fetal Age)
Unit: HC (mm); Age (Week); SD (Week)
HC Age SD HC Age SD HC Age SD HC Age SD
<55
55
60
65
70
75
80
85
90
95
100
105
110
115
120
125
130
n/a
12.0
12.3
12.6
12.8
13.1
13.4
13.7
14.0
14.3
14.7
15.0
15.3
15.6
16.0
16.3
16.6
——
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
± 1.2
135
140
145
149
150
155
160
165
170
175
180
185
190
195
200
205
210
17.0
17.3
17.7
18.0
18.1
18.4
18.8
19.2
19.6
20.0
20.4
20.8
21.3
21.7
22.2
22.6
23.1
± 1.2
± 1.2
± 1.2
± 1.2
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
± 1.5
215
219
220
225
230
235
240
245
250
255
260
265
270
274
275
280
285
23.6
23.9
24.0
24.5
25.0
25.5
26.1
26.6
27.1
27.7
28.3
28.9
29.4
29.9
30.0
30.7
31.3
± 1.5
± 1.5
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 2.1
± 3.0
± 3.0
± 3.0
290
295
300
305
310
315
319
320
325
330
335
340
345
350
355
360
>360
31.9
32.6
33.3
33.9
34.6
35.3
35.9
36.1
36.8
37.6
38.3
39.1
39.9
40.7
41.6
42.4
n/a
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 3.0
± 2.7
± 2.7
± 2.7
± 2.7
± 2.7
± 2.7
± 2.7
± 2.7
± 2.7
——
LOGIQ C5/C5 PRO Advanced Reference Manual 3-15
Direction 5264895-100 Rev. 1
Page 100
OB Tables
Table 3-23: HC: Hadlock, AJR 138: 649-653, 1982 (Fetal Age)
Unit: HC (mm); Age (Days); SD (Days)
HC Age SD HC Age SD HC Age SD HC Age SD
<69
69
75
81
88
96
103
110
117
124
131
137
144
150
157
163
n/a
84
87
91
94
98
101
105
108
112
115
119
122
126
129
133
——
9
9
9
9
9
9
9
9
9
9
9
9
9
11
11
169
175
181
187
193
198
204
209
215
220
225
230
240
245
250
255
136
140
143
147
150
154
157
161
164
168
171
175
182
185
189
192
11
11
11
11
11
11
11
11
11
11
16
16
16
16
16
16
260
264
269
273
278
282
286
291
294
298
302
306
309
312
316
319
196
199
203
206
210
213
217
220
224
227
231
234
238
241
245
248
16
16
16
16
16
19
19
19
19
19
19
19
19
19
19
19
322
325
328
331
334
337
340
343
345
348
351
353
354
>354
252
255
259
262
266
269
273
276
280
286
287
290
294
n/a
19
24
24
24
24
24
24
24
24
24
24
24
24
——
Table 3-24: FL/HC Ratio: Hadlock, J Ultrasound Med 1984, 3: 439-442 (Fetal Growth)
Unit: GA (Weeks)
GA Min Max GA Min Max GA Min Max
<15
15
16
17
18
19
20
21
22
23
n/a
15.3
13.3
14.6
15.8
16.1
16.8
15.9
18.4
19.2
——
17.1
16.5
17.6
18.0
18.3
19.8
20.3
20.2
20.8
24
25
26
27
28
29
30
31
32
33
18.7
18.7
18.6
18.6
18.8
19.6
19.2
19.3
19.1
19.9
20.9
20.3
20.4
20.4
20.6
20.8
21.4
21.3
21.3
21.5
34
35
36
37
38
39
40
41
42
>42
19.4
20.1
20.1
20.8
20.9
20.6
20.7
21.6
20.1
n/a
21.8
22.3
22.1
22.6
22.7
23.4
22.5
23.2
23.9
n/a
Table 3-25: FL/AC Ratio: Hadlock (Fetal Growth)
Unit: Age (Weeks)
Age Min (Index) Max (Index)
21
42
20
20
24
24
3- 16 LOGIQ C5/C5 PRO Advanced Reference Manual
Direction 5264895-100 Rev. 1