Fisher INNERVATOR252 User manual

INNERVATOR252

Constant Current Peripheral Nerve Stimulator

OPERATING INSTRUCTIONS

English
The NS252 provides neuromuscular block monitoring for general anesthesia required for surgical procedures in the operating theater.

OPERATOR CONTROL BUTTONS

Switches the INNERVATOR252 on & off.
Sets the amount of current to be delivered by 10mA steps. If CURRENT UP button is depressed the set current advances to 80mA. If CURRENT DOWN button is depressed the set current decreases to 0mA. Maximum set current is 160mA, (To deliver 80 - 160 mA see Programming Features)
Mutes audible pulse delivery sound. To cancel press again.
Cannot be used to mute the audio alarm sound.
B. DEMAND BUTTONS
TW
Delivers a single pulse.
TOF
Delivers a train of four pulses where each pulse is 0.5 seconds apart
TET
Delivers a tetanus pulse train for as long as the button is depressed.
DBS
Delivers a Double Burst pulse train. DBS = three pulses 20 ms, a 750 ms lull then two features).
OR
three 20ms pulses, (see programming
PTC
Post Tetantic Count. Consists of five sec TET pulse, three sec pause, twenty twitch pulses at 1 Hz. For safety, PTC can only be activated once every five minutes.
C. REPEAT BUTTONS
Activates a preset pulse pattern. Press again to terminate the cycle.
1 SEC TW
Single pulse delivered at one second intervals.
10 SEC TW
Single pulse delivered at ten second intervals
12 SEC TOF
Train of four pulses repeated every 12 seconds.
, (1 Second Twitch (1Hz))
, (10 Second Twitch (0.1Hz))
.
, (12 Second Train-of Four)
D. PROGRAM, PTC BUTTONS
PROG
Activates programming capabilities. Press once to change settings, twice to store program.
RUN
Starts stored program. Press again to stop. Default program = Double Burst at 12 sec intervals.
VISUAL INDICATORS
CURRENT
Current delivered to the patient, shown in milliamps
PULSE DELIVERY
Shown as a solid dot in the centre of a circle. It flashes every time a pulse is delivered.
MUTE
Indicates that audible pulse signal is muted.
.
mA
!
PROGRAM
Indicates INNERVATOR252 is in Programming mode
RUN
Indicates that the stored program is running.
PTC
Flashes to indicates that PTC, Post Tetanic Count pattern is
OR
running
REPEAT
Indicates that one of the three repeat pulse patterns, 1SEC TW, 10 SEC TW or 12 SEC TOF is running.
ERROR
Indicates that the INNERVATOR252 is not able to deliver the specified current. Indicates excessive impedance, or disconnected leads or dry patient electrodes.
continuous display if safety lockout is still in effect.
RUN
PTC
REPEAT
ERROR
P
LOW BATTERY
Indicates that the batteries require replacing as soon as possible.

AUDIO INDICATORS

PULSE DELIVERY
Monotone, single beep every pulse.
AUDIBLE ALARM
Half second, monotone beep each time a pulse is delivered incorrectly AND, at the completion of the pulse train there is a 2 second continuous beep.
Indicates pulse does not fall within the tolerance limits each side of the set current. May be caused by, excessive impedance, disconnected leads or dry patient electrodes.
If the alarm sound is occurring repeatedly:
Do not continue running a repeat function.
Do not deliver any further pulses until the fault causing
the condition is repaired.
USE & CONNECTION TO A PATIENT
Only shrouded lead sets (200NS033 (supplied), 200NS032 or 200NS031) may be used with the INNERVATOR252.
1. Ensure that the INNERVATOR252 is switched OFF.
2. Connect external lead to the INNERVATOR252, proximal (red, positive) to red socket and distal (black, negative), to black socket.
3. Connect the other end of the lead to the patient
electrodes, usually attached along the ulnar nerve. Proximal (red) connector to proximal (red) electrode, distal (black) connector to distal (black) electrode.
Red Black
Following the induction of anesthesia
4. INNERVATOR252 on and increase the current until desired level of supramaximal stimulation has been attained.
NOTE: A two second continuous tone indicates that alarm conditions exist. This may be caused by excessive impedance, dried or faulty electrode(s), or disconnected or faulty leads.
5. After use, always ensure that the INNERVATOR252 is switched OFF.
Auto power shut-off after 30 minutes of non-use.
INNERVATOR252 will emit three beeps, if no buttons are pressed within next minute, INNERVATOR252 will automatically switch off. Press any button to cancel the automatic shut down for another 30 minutes. Will not occur if running a repeat function.
switch the

DEFINITIONS

Type BF ATTENTION 9 3/42/EEC Applied Part Consult Clas s IIb ac companying
documents
‘PERIPHERAL NERVE ST IMULATOR’
WITH RESPECT TO ELEC TRIC SHOCK, FIRE,
MECHANICAL HAZARDS ONLY IN ACCORDANCE
WITH UL 2601-1, C AN/CSA C22.2 NO. 601.1
WARNING:
the possibility of injury to the patient or user exists if a procedure is not followed correctly.
CAUTION:
damage to the equipment exists if a procedure is not followed correctly.
NOTE:
to clarify points, procedures or instructions.
A WARNING statement refers to the conditions when
A Caution statement indicates the possibility of
A Note statement provides additional information intended
84DJ
Page 1 of 2
WARNING
CAUTION
Interference from electrosurgical units (diathermy), may be experienced. Ensure that the Nerve Stimulator is positioned away from electrosurgical unit. Lead-set should not run parallel to the diathermy lead and should not cross the diathermy lead at any stage.
Remove batteries if equipment is not likely to be used for some time. Clean with a damp cloth. All maintenance must be carried out by qualified personnel. For further information and technical data contact your Fisher & Paykel Healthcare dealer.
:
This device should not be used in a manner that allows trans-thoracic stimulation.
Not to be used for Internal Nerve Location. Use recommended stimulator lead sets only. Do not use with needle electrodes.
Dangerous voltages may exist within this device for up to ten minutes after removal of the batteries.
This device i s not suitable f or use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxid e.
This product is intended for use by qualified medical practitioners and such per sons should e nsure that they are totally familiar with the use of the nerve stimulator before using the device.
:
HOW TO PROGRAM
A. PROGRAMMING A PULSE PATTERN
1. Push PROG button to enter programming mode, (display will read P). Press again to store mode, (display will read P00ma).
2. Use DEMAND BUTTONS* ONLY to deliver desired pulse pattern. Any number may be pressed in any order.
3. Allow desired time interval between pulses to elapse between button pushes, (the display will count upwards in seconds). Each time a demand button is pressed the display returns to zero. Maximum elapsed time between button pushes is 99 seconds.
4. Push PROG again to store the program.
5. Reset the INNERVATOR252 to the desired current delivery level, (current is automatically zeroed during programming).
6. Push RUN to execute the stored program at set output current. Program will run until RUN is pressed again or unit is switched off. The RUN indicator will be displayed while program is running.
*
PTC button cannot be used whe n storing a program.
B. PROGRAMMING TETANUS PULSE TO 50HZ OR
100HZ.
1. Push PROG button ONCE, (display will read P).
2. Push TET button to select Tetanus frequency. The frequency will be displayed as “50” or “100”.
NOTE: There is no visual indicator to display TET frequency when in use. However the 100Hz audio indicator has a faster pulse delivery sound than the 50Hz.
C. PROGRAMMING THE MAXIMUM CURRENT
SETTING TO 80mA OR 160mA.
1. Push PROG button ONCE, (display will read P).
2. Push CURRENT UP button to select maximum current at 80mA. “Lo” will be displayed to indicate the maximum current is 80mA. Pressing the CURRENT UP button again will select the maximum current at 160mA and “Hi” will be displayed.
If 80mA is programmed, depressing the CURRENT UP button will increase the current in 10mA steps to 80mA.
If 160mA is programmed, depressing the CURRENT UP button will increase the current in 10mA steps to 80mA. To set beyond 80mA, the CURRENT UP button
pressed and released
10mA is desired.
each time a further increment of
must be
D. PROGRAMMING THE DOUBLE BURST
STIMULATION PATTERN.
1. Push PROG button ONCE, (display will read P).
2. Push the DBS button to select either the 3:2 or 3:3 pattern. The display will indicate either “3:2” or “3:3”.
E. IMPORTANT NOTES:
All programs are retained when unit is switched off.
Programs are retained for 5 minutes when batteries are
removed to allow sufficient time for battery replacement.
If batteries removed for longer than 5 minutes, program
features return to default, (see specifications).

BATTERY REPLACEMENT

To replace batteries use three 1.5 volt “AA” batteries

SPECIFICATIONS

OUTPUT CURRENT RANGE:

0-160 mA selectable in 10mA steps

TETANUS FREQUENCY:

Programmable 50Hz or 100Hz

OUTPUT PULSE FORM:

Square wave, constant current, monophasic, unidirectional

OUTPUT PULSE WIDTH:

195 µs ± 5µs

OUTPUT PULSE RISE/FALL TIME:

< 10
µs
maximum, (resistive load)

OUTPUT VOLTAGE:

350 volts, maximum

MAXIMUM LOAD IMPEDANCE

TW, TOF, DBS - 4K ohms at 80mA, 3.5K ohms at 160mA
TET, PTC – 1.8K ohms at 80mA, 1.3K ohms at 160mA

SET CURRENT OFFSET

at 0-80mA ± 2.5mA at 85-160mA ± 5.0mA

ALARM TOLERANCE LIMITS

± 2.5mA of the set current

DEFAULT PROGRAM

10sec twitch
Temperature 5 °C to 30 °C -10 °C to 60 °C -10 °C to 60 °C

DEFAULT PROGRAM SETTINGS

TET = 50Hz, Current Max = 80mA, Current steps 10mA,
DBS = 3:3

POWER REQUIREMENT

Internally powered. Three 1.5 volt “AA” batteries.
Continuous operation.

BATTERY LIFE

More than 300 hours continuous use with alkaline batteries

DIMENSIONS

Height - 170mm
Width - 73mm
Depth - 43mm
Weight - 0.18kg, (without batteries)
LEAD SETS (shrouded):
200NS033 (supplied): dome connectors.
200NS032: pin connectors.
200NS031: alligator clips.
ENVIRONMENTAL
Operation Transport Storage
Humidity 0 % to 90 % 0 % to 90 % 0 % to 90 % ‘Ordinary equipment’

CONTACTS

INTERNATIONAL
AUSTRALIA
FRANCE/BENELUX
GERMANY/AUSTRIA
UK/IRELAND
USA
NOTE: US Federal law restricts this device to sale by, or
REF 185042350 Rev B January 2002 Page 2 of 2
Tel: +61-(0)3-9879 5022
on the order of, a physician.
P.O. Box 14-348, Panmure, Auckland 6, New Zealand Email: info@fphcare.com
Web Site: www.fphcare.com Tel: +64-(0)9-574 0100 Fax:+64-(0)9-574 0158
Fax: +61-(0)3-9879 5232
Tel: +33-(0)1-64 46 52 01 Fax: +33-(0)1-64 46 52 21
Tel: +49-(0)7182-93777-0 Fax: +49-(0)7182-93777-99
Tel: +44-(0)1628-626 136 Fax: +44-(0)1628-626 146 (EU Authorised Representative)
Tel: 1800 446 3908 Fax: +1 949 470 3933 Tel: +1 949 470 3900
Loading...