ibia nailing in horizontal or angled traction position. Use Eschmann
Converting T30 to
o
rthopaedic table or just adding orthopaed
ic traction using auto
Easy patient setting and full
T30a/m
with
orthopaedic
attachment
self guiding docking trolley preventing risk of injury to operating room staff.
Offers option of single or double click-on traction beams with offset choice,
and leg support for trouble free patient setting and ultimate patient comfort.
Patient weight load of 200 kg on traction mode without support pole.
Optional accessory trolley
flexibility of X-ray imaging for
DHS procedures
T30 leg support or DP leg holder.
Large wheels option with single locking
with built
This literature should be read in
T30a/m 5/6 section Operation
table convertible to orthopaedic
and trauma table
New generation, new concept,
completely versatile, all purpose
operation table
5 or 6 sections general and
specialised operation table
Fully X-ray translucent table top,
-in X-ray cassette tunnel
through the table top
Detachable and interchangeable
sections
Convertible to fully functional
orthopaedic and trauma table
Special designed shoulder surgery attachment for Eschmann “T” range
conjunction with EschmannT20
catalogue publicity No. PS-217c
pedal (T30m)
Small wheels with double pedal (T30a)
Eschmann Equipment, Peter Road, Lancing, West Sussex BN158TJ, England Tel: +44 1903 753322 Fax: +44 1903 875789 www.eschmann.co.ukAsia Pacific: Eschmann Holdings Pte Ltd., 300 Beach Road, #32-07 The Concourse, Singapore 199555
Tel: +65 62983344 Fax: +65 62928918 Email: Eschmann_spore@pacific.net.sg
Details given in this leaflet are correct at the time of going to press. The Company however, reserves the right to improve or modify the equipment shown. The Eschmann name and
logo are trademarks Eschmann Holdings Ltd. Eschmann Equipment is a trading name of Eschmann Holdings Limited.
Optional accessories
Full X-ray imaging at lowest height for spinal surgery
T30a/mwith orthopaedic attachment
New generation, new concept, completely versatile, all
purpose operation table
5 or 6 sections general and specialised operation table
Fully X-ray translucent table top, with built-in X-ray cassette
tunnel through the table top
Detachable and interchangeable sections
Convertible to fully functional orthopaedic and trauma table
Detachable, and click-on traction beams,
centre positioning with double joint
knuckles.
Two options of double pedal locking
mechanism, with standard wheels(T30a)
Tibia nailing downward traction
position
Large wheels option with single
locking pedal (T30m)
This literature should be read in
conjunction with EschmannT20
catalogue publicity No. PS-217c
Tibia Nailing in horizontal traction
position. Use Eschmann T30 leg
support or DP leg holder
Easy click-on docking cart with autolock mechanism
T30
Perennial
Post
Carbon Fibre
with
, Range Configuration and Specification
T30a and T30m
5 or 6 (with divided leg section) sections table top.
Features and specification as T20a/m, but without
transverse top
T30m, (single pedal
and large wheels)
Interchangeable
sections
5 Section T30a, (double pedals
& small wheels)
Orthopaedic Attachment for
T30a and T30m
Sacral Support
Docking Cart (optional)
Traction beam & T. assembly
Orthopaedic boot
Docking Cart
T30a/m-Ortho
An Orthopaedic and Trauma Operating Table complete
detachable double traction beams assembly, leg
supports, swivel joint knuckles, pair of adjustable booties,
carbon fibre perennial post,
convertible to 5 or 6 sections fully
X-ray translucent top general and
specialised operation table
Perennial Post
Carbon Fibre
Gynae & Urology with
perennial extension
Sacral assembly
Gynae & Urology
Traction Beam &
Traction Assembly
Optional accessories and attachments for Orthopaedic Surgery
Accessory Trolley Hip Fixation Accessories Hand Surgery Table Shoulder Chair
Optional accessories for General & Specialised Surgery
Width Extender
Eschmann Equipment, Peter Road, Lancing, West Sussex BN158TJ, England Tel: +44 1903 753322 Fax: +44 1903 875789 www.eschmann.co.ukAsia Pacific: Eschmann Holdings Pte Ltd., 300 Beach Road, #32-07 The Concourse, Singapore 199555
Tel: +65 62983344 Fax: +65 62928918 Email: Eschmann_spore@pacific.net.sg
Details given in this leaflet are correct at the time of going to press. The Company however, reserves the right to improve or modify the equipment shown. The Eschmann name and
logo are trademarks Eschmann Holdings Ltd. Eschmann Equipment is a trading name of Eschmann Holdings Limited.
Hand Traction
Tibia Nailing Set
Accessory Trolley Foot Support Light weight Leg S Divided Leg S Ophthalmic Head S DP Leg Holder
T30
OPERATING TABLES
Series
Instructions for use
113364
T-IM122b
Read these Instructions before use
Keep these ‘Instructions for use’ in a safe convenient place for future reference.
Eschmann After Sales Service Department
The Eschmann After Sales Service Department is staffed and equipped to provide advice and
assistance during normal office hours. To avoid delays when making enquiries, please quote the
Model and Serial Number of your Operating Table which is shown on the Serial Number Plate
located on the table base (or the trunk section for the T30-m). Please ensure you include all
alpha and numeric digits of the Serial Number.
For further information visit www.eschmann.co.uk
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 875711.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Patents and Trade marks
The ESCHMANN name and logo; “T30-a” and “T30-m” are trade marks of Eschmann Holdings Limited.
“Eschmann Equipment” is a trading name of Eschmann Holdings Limited.
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.
The CE marking affixed to the product certifies that it complies with the European
Medical Devices Directive 93/42/EEC as amended and related legislation.
T-IM122b March 2007
Instructions for use
T30
OPERATING TABLES
Series
The T30 Series Operating Tables
The T30 powered operating table is a modular system suitable for
a full range of general, specialist and Orthopaedic surgical
procedures. A range of 4, 5 or 6 section tops are
available along with a superior range of additional
accessories to make this a truly versatile
general and specialist powered
operating table. With the simple
removal of the general surgical
lower trunk section, via a docking
cart, and the addition of a range of
orthopaedic accessories, the T30 can
easily be converted for a full range of
orthopaedic procedures.
TheT30 operating table has an X-ray translucent tabletop
with a built-in X-ray cassette tunnel. The slimline column is
offset relative to top and base for ease of C-arm access.
The T30 Table features a corded handset controlling Trendelenburg,
reverse Trendelenburg, lateral tilt, flexion/extension (including
90° chair position) and height. The handset also provides
a battery level indication.
The batteries in the table base are
mains rechargeable with a standby
battery in case of emergency.
Covers to the top and base
are purple and made
of a special scratch
resistant, hard-wearing
and easy to clean
seamless acrylic
capped ABS.
The mattress is moulded,
latex free and antistatic.
A choice of table
bases offers a
range of ergonomically
designed foot pedals to suit all
operating theatre environments. The T30-a pedals are located at the head and leg end of the
table base to provide braked, wheel (move in a straight line) and 360° mobility. The T30-a is
capable of supporting a maximum patient weight of 300kg or 135kg when moved around the
theatre.
The T30-m single foot pedal is located at the head end of the table and has three positions to
enable the table to be static (braked), rotated 360° or moved in a straight line. The T30-m is
capable of supporting a maximum patient weight of 300kg or 200kg when moved around the
theatre.
APPENDIX 5 - The traction accessories ..................................................................................................65
APPENDIX 6 - The Tibial Nailing Accessory............................................................................................68
APPENDIX 7 - The Lateral Femoral Nailing Accessory ...........................................................................70
APPENDIX 8 - The orthopaedic accessory trolley ...................................................................................72
APPENDIX 9 - Orthopaedic accessory and spares list ............................................................................74
APPENDIX 10 - Orthopaedic positioning and imaging procedures.......................................................... 75
T-IM122bP5/77
1.0 PRELIMINARY INFORMATION
WARNING
Read this preliminary information carefully and
note ALL of the warnings, cautions and safety
notes contained within these ‘Instructions for
use’ before using this Operating Table. Keep
these ‘Instructions for use’ close-to-hand at all
times for reference.
1.1General
1.1.1The T30 Tables are classified as battery
powered, mobile, general purpose and orthopaedic
operating tables. Their intended function is to support and
position a patient, in conjunction with their associated
accessories, during general surgical and orthopaedic
operations and procedures in an operating theatre. Their
intended application is for use by medically qualified
personnel, trained in the use of the T30 Tables, during
surgical operations and procedures in accordance with
these instructions.
1.1.2These ‘Instructions for Use’ should be referred to
for details of the following T30 Tables, see the table’s Serial
Number Plate for the actual table Serial Number and REF
No.:
T30-a Catalogue (REF) number prefixed TOR2
Serial Numbers (SN) prefixed TAOA* or above.
T30-m Catalogue (REF) number prefixed TOR2
Serial Numbers (SN) prefixed TMOA* or above.
* The last letter of the SN prefix is the design modification
state, this may increase during the build life of the tables.
NOTE: Some models have ‘Denyer’, ‘USA’, or ‘Euro (Kifa)’
accessory sidebars. Standard UK Eschmann accessories
that locate onto these may not fit, please check with
Eschmann Equipment before purchasing accessories.
1.1.3T30 Tables and accessories must only be used
as detailed within these ‘Instructions for use’. Failure to do
so could result in injury to patients or users, or damage to
the operating table and accessories. Always ensure that
all warnings and cautions detailed within these ‘Instructions
for use’, are strictly complied with. Appendix 1 is provided
for Hospitals that wish to keep a log of those people trained
in the safe use of this table. Eschmann Equipment offers
full training in the safe use of these tables, please contact
our After Sales Service Department, Tel. +44 (0) 1903
765040 for details.
1.1.4‘Instructions for use’ and ‘Service manuals’ (see
section 3.3.9) as applicable should be readily accessible
for reference prior to and when operating, cleaning and
servicing the operating table. All manuals are available from
Eschmann Equipment, see inside front cover for address.
1.1.5Ensure the table has been correctly installed
before starting to use it, see section 3.0.
1.1.6Ensure that these operating tables and
accessories are regularly serviced and checked for safety
at least every six months. All servicing and maintenance
procedures should be carried out by engineers who have
been trained by Eschmann. Training, a Service, or a Service
Contract can be arranged through our After Sales Service
Department [Tel. +44 (0) 1903 765040]. Do not remove
the table covers at any time (danger of accessible mains
voltage), this should only be carried out during maintenance
procedures.
1.2 About this manual
1.2.1Within the text of these ‘Instructions for use’ the
following terminology is used:
i)
Normal patient orientation.
head section in a supine position, with the table in its
normal configuration.
ii)
Normal table configuration.
trunk section fitted, infill section fitted to long trunk,
head section fitted to infill section, leg section fitted
to general surgical short trunk (see Fig. 2.1).
iii)
Orthopaedic configuration.
section fitted with hip and pre-operative (split) leg
sections attached (see Fig. 2.1).
iv)
Left and right
the side of the table when viewed from the head
end with the table in its normal configuration.
v)
Auto-level
the automatic sequence of movements to return the
table to a preprogrammed position by pressing and
holding a single handset button. This does not affect
the head or leg sections, or any other manual or
powered accessory, fitted to the table.
vi)
LEDs.
This abbreviation is used when referring to
the various indicator lights on the table or handset
which are light emitting diodes (LEDs).
vii) The term T30 Table is used to cover both the T30-a
and the T30-m version of these operating tables.
1.2.2These ‘Instructions for use’ have been split into
specific sections for ease of finding information (see the
main headings in ‘Contents’). Where applicable and within
each section, adequate cross references to other sections
are made to eliminate the need to duplicate information.
Orthopaedic procedures and table set up are covered in
Appendix 5 onwards
1.2.3Section 5 (Operation) details how to use the T30
Tables and operate their controls correctly. It is strongly
recommended that the user has read and is familiar with
sections 1, 2 and 4 before passing onto section 5 and
starting to use this operating table.
1.2.4Within this manual the sections in bold type
headed ‘WARNING’ give guidance on possible actions that
could lead to injury of the patient, or theatre staff, and
potential damage to the operating table. Sections headed
‘CAUTION’ give guidance on possible actions that could
lead to damage of this operating table which could then
lead to injury of the patient, or theatre staff.
. The terms ‘left’ and ‘right’ refer to
. The term ‘auto-level’ is used to describe
.
Patient’s head on the
General surgical short
Orthopaedic short trunk
P6/77T-IM122b
T30
OPERATING TABLES
Series
1.3Table description
1.3.1T30 Tables have been designed to provide
facilities for General Surgery* and Orthopaedic Surgery**
with traction, they allow for intra-operative radiography
using a C-arm image intensifier.
*including Minimal Access procedures, Urology and
Gynaecology, Thoracic, Ophthalmic and ENT,
Neurosurgery, Plastic and Maxillo-Facial surgery.
**including dynamic hip screw (DHS) in the supine
position, femoral nailing in the supine position,
femoral nailing in the lateral position, femoral
reconstructive nailing in the supine position,
retrograde femoral nailing, tibial nailing, humeral
nailing and shoulder arthroscopy.
1.3.2T30 Tables are stable, rigid in use and the robust
construction provides protection from patient trolley or
C-arm knocks whilst still providing easy access for
servicing. Careful design has minimised traps for potential
contamination and stops fluid entering the table during
normal use, cleaning and disinfection procedures.
1.3.3Tabletop movements (i.e. Trendelenburg, height,
tilt and break) are electrically powered and incorporate ‘soft
start and stop’ motions for patient comfort and safety. The
pedestal base is foot operated and the standard head and
leg sections are hand operated (see sections 5.1 and 5.2
respectively). All orthopaedic accessories are manually
adjusted and operated.
1.3.4T30 Tables are easy to operate, theatre staff can
quickly learn how to use them correctly and safely. Tabletop
control is provided by a corded handset, or an optional
footswitch (for Trendelenburg and height control only).
Handset control overrides footswitch control at all times.
1.3.5Power for all powered table movements is
provided by internal rechargeable batteries. There are two
battery sets, main batteries and standby batteries. These
are recharged by an internal battery charger which requires
connection to mains voltage using the mains cord supplied.
Note: If the main table batteries fail or become critically
low (i.e. warning indicators were ignored and batteries have
not been recharged) the table cannot be powered directly
from the mains. However by depressing and holding the
standby battery switch, powered table movements can
again be achieved (using power from standby batteries).
1.3.6In emergencies (e.g. handset failure or a critically
low battery) a standby control panel on the column can be
used to control the table (however this MUST be limited to
emergency use ONLY, certain safety features are overridden when this panel is used, see section 5.3.4). It may
be required to press the standby battery switch if the main
battery charge level is too low.
1.3.7The base on the T30-a Table is fitted with
enclosed multidirectional castors and has two foot pedals.
The T30-m Table is provided with four large castors that
can be adjusted by a single foot pedal. The castors on all
models can provide either castor, wheel or braked
orientations (see section 5.1).
1.3.8Visual indicators have been restricted to essential
functions and information only (e.g. table switched ‘on’ see
section 5.3.2.2, battery level indication and battery charging
state see section 5.3.1). Audible signals are only used to
signal that the table has been switched ‘on’ for a long time
without operation and that it should be switched ‘off’, see
section 8.1 (this inactivity ‘beep’ can be configured ‘off’
during a service if required), a ‘beep’ is also provided when
switching ‘on’.
1.3.9Tabletops have a lightweight X-ray translucent
surface (designed to reduce shadows on images) and an
X-ray cassette tunnel with the facility for an X-ray cassette
(430mm x 340mm) to be placed at any point beneath the
full length of the patient’s body, see section 6.3. The
standard sidebars allow placement of clamps and most
standard accessories. Simple buttons or catches release
the head, leg, hip, pre-operative leg, infill and short trunk
sections.
1.3.10The tabletop can be adjusted into the following
patient positions (Note: Orthopaedic positions are detailed
in Appendix 5 onwards):
♦Supine with C-arm access to patient from nipple
region to feet
♦Supine with C-arm access to patient from groin to
head
♦Supine Extension with C-arm access to break area
(Cholecystectomy-type procedure)
♦Supine Flexion (‘Lawn chair’ position)
♦Lateral
♦Lateral Extension with C-arm access to break
(Nephrectomy position)
♦Supine Lithotomy with C-arm access to whole of
the Urinary tract
♦Supine Lithotomy with or without Trendelenburg
♦Prone with or without extension at waist or hips (with
C-arm access)
♦90° Chair position with patient’s knees at the same
level as the heart, offset to the head end.
1.3.11For additional information see section 6.0 for
patient positioning notes, section 6.2 for obese patients,
6.3 for radiographic procedures and 6.4 for illustrations
and details of safe loading. Tabletop configurations are
shown pictorily in section 6.5.
T-IM122bP7/77
2.0 TABLE PARTS AND SYMBOLS
REFThis symbol indicates the catalogue number is
as indicated adjacent to the symbol.
2.1Part identification
2.1.1Fig. 2.1 shows the T30-a and T30-m Table tops
in their normal configuration and identifies the major parts
of the tabletop.
2.1.2Fig. 2.2 identifies the various parts of the T30-a
and the T30-m Table base and column.
2.1.3Orthopaedic accessories and components are
detailed in Appendix 5 onwards.
2.2Symbols and graphics
To enable an easy reference to all the symbols and graphics
used on the T30 Tables (and within these ‘Instructions for
use’) the following grouped sections show all the symbols
and graphics used.
2.2.1 Symbols general
The following symbols are shown on various parts of the
table, handset or Serial Number Plate.
IPX 4 This symbol (splash proof) denotes that the
equipment (the table) meets the requirements of IEC529
for protection from splashing water.
IPX 6 This symbol (protection against heavy seas) denotes
that the equipment (the handset) meets the requirements
of IEC529 in that water from heavy seas or water projected
from powerful jets shall not enter in harmful quantities.
This symbol indicates that the equipment is for
use on alternating current.
This symbol indicates that fuses adjacent to the
symbol have a rating and type as detailed.
This symbol warns the user to read the accompanying documents, these ‘Instructions for use’.
This symbol indicates that the manufacturer is
as indicated adjacent to the symbol.
This symbol indicates that the date of manufacture is as indicated adjacent to the symbol.
This symbol indicates the connection point for
a footswitch.
This symbol indicates the connection point for
the corded handset.
This symbol indicates that the table section
to which it is applied (e.g. head section, under
the mattress) should not be used as a seat.
This symbol indicates the ‘Safe working load’
of the section to which it is applied can safely
support an evenly distributed load to the value
indicated, in this example 25kg.
This symbol indicates the ‘Minimum breaking
load’ of the section to which it is applied. An
evenly distributed load (in this example 100kg
or greater) may break the section.
This symbol on the table base indicates the
maximum weight that can be placed on the
table when static and braked, or being moved
around the theatre (see section 6.4 for more
information).
This symbol is used to identify the standby
battery switch.
113150-01
This graphic (T30-m Table only) adjacent to the
mains socket, identifies the relationship between
the colour of the mains ‘on’ LED and the battery
charge state, see section 5.3.1.
Symbols and near the main table ‘on/off’
switch, indicate ‘OFF’ and ‘ON’ respectively.
With the mains cord attached the equipment has
‘Class II’ protection against electric shock.
The patient leakage current, with mains voltage
on the applied parts, meets the requirement for
type BF medical electrical equipment and are
defibrillator proof.
This symbol is used to indicate the table’s duty
cycle which is the ratio of the operating time to
the sum of the operating time and the ensuing
interval.
SNThis symbol indicates the unit serial number is
as indicated adjacent to the symbol.
This symbol indicates that this product was
placed on the market after 13th August 2005. At
the end of its working life it should be deposited
at an appropriate facility to enable recycling. For
information on how to recycle this product
responsibly contact Eschmann.
Pushing prohibited. This symbol, on the trolley,
indicates that it should not be pushed from the
side on which the symbol is attached.
This graphic (T30-a Table only)
adjacent to the mains socket, identifies
the relationship between the colour of
the mains ‘on’ LED and the battery
charge state for the main and standby
batteries, see section 5.3.1.
P8/77T-IM122b
T30
OPERATING TABLES
Series
2.2.2 Handset button symbols
The following symbols are shown on the handset buttons
to indicate their function. Use of the handset is fully detailed
in section 5.3.3 of this manual.
Trendelenburg - Press to adjust tabletop in the
Trendelenburg (head down) direction.
Reverse Trendelenburg - Press to adjust
tabletop in the Reverse Trendelenburg (head
up) direction.
Height down - Press to move tabletop down.
Height up - Press to move tabletop up.
Tilt - Press to tilt tabletop down on the left (when
viewed from the long trunk end)
Tilt - Press to tilt tabletop down on the right
(when viewed from the long trunk end)
2.2.3 Handset graphics
This graphic is shown on the corded
handset to indicate the battery charge
level for both the main and standby
batteries, see section 5.3.1.
2.2.4 Standby control panel button symbols
The following symbols are shown on the standby control
panel buttons, indicating the function they select. The
arrows (upper or lower) indicate the direction the selected
function will move, if the corresponding direction button
(i.e. upper or lower) is pressed, see section 5.3.4.
Button selects Trendelenburg function.
Button selects Break function.
Button selects Height function.
Break down - Press to move the break down
(i.e. short trunk moves down w.r.t. long trunk)
Break up - Press to move the break up
(i.e. short trunk moves up w.r.t. long trunk)
Flexion - Press to move tabletop into Flexion.
Extension - Press to move table into Extension.
Return to Level - Press to return tabletop
to a preset level position.
Button selects Tilt function.
Non functional Button (selects Traverse function
on other tables in the ‘T’ series).
Direction button - Press button to obtain movement
indicated by upper arrow of function button.
Direction button - Press button to obtain movement
indicated by lower arrow of function button.
T-IM122bP9/77
TYPICAL GENERAL SURGICAL
TABLETOP ARRANGEMENT
Note: The infill section can be removed, or be
positioned between the leg and short trunk
section. The leg section shown can be replaced
by a divided (split) leg section.
3
1
5
2
11
16
17
8
9
9
1
9
2
10
4
6
12
6
9
18
19
7
TYPICAL ORTHOPAEDIC
13
12
7
14
15
9
TABLETOP ARRANGEMENT
17
1Head section
2Long trunk section
3General surgical short trunk section
4Orthopaedic short trunk section
5Infill section
6Hip sections
7Pre-operative leg sections
8Leg section
9Section release buttons
10Leg section gas spring
11Head section end block (use to stow UK/EURO handset)
12Accessory sidebar (can also be used to stow handset*)
13End block, also on head section (use to stow UK/EURO handset)
14Head section release bar
15Head section gas spring
16Leg section release bar
17Traction beam
18Fixed knuckle
19Intermediate knuckle
* Clip on handset must match sidebar type, see accessory section 4.8
Fig. 2.1 Part identification of the T30 Table tops.
P10/77T-IM122b
T30
OPERATING TABLES
Series
10
5
3
1
4
2
6
87
9
13
Detail A Detail B Detail C
3
12
11
B
A
C
12
10
1
15
9
1Standby battery switch
2T30-a Table's battery fuses
3Connection socket for Eschmann handset (ONLY)
4Table ‘On’ LED (green). Bright or Dim see section 5.3.1)
5Table ‘On/Off’ switch (
6Connection socket for Eschmann footswitch (ONLY)
7T30-a Table, mains ‘on’ and Main Battery charging state LED. Red, Amber or Green.
8T30-a Table, mains ‘on’ and Standby Battery charging state LED. Red, Amber or Green.
9Connection socket for mains (ONLY use Eschmann mains cord supplied)
10T30-a Table, castor foot pedal
11T30-a Table, wheel foot pedal
12Standby control panel
13T30-a Table, mains fuses
14T30-m Table, mains ‘on’ and Battery charging state LED. Red, Amber or Green.
15T30-m Table, foot pedal
16T30-m Table, mains fuses
= Off, = On)
14
16
Fig. 2.2 Part identification of the T30 Table base and column.
T-IM122bP11/77
3.0 INSTALLATION
3.1General
3.1.1In the U.K. the table is delivered un-crated with
the head and leg section fitted. The mattress set is boxed
individually and placed on the trunk sections. The handset,
mains cord and literature are supplied loose. Any
accessories ordered with the table will be packed
individually.
3.1.2 For overseas markets the table is usually packed in
a container with the head and leg section fitted, together
with a boxed mattress set a mains cord and the literature.
The handset is packed within the container in an antistatic
bag. Other accessories are usually packed separately, but
some may be included in the main case and should be
unpacked and stored separately during table installation.
3.1.2(Export only) Carefully remove the T30 Table
(having first removed any accessories and packing
restraints from within the container) from the packing case
as follows:-
iRemove the walls of the case leaving the table on
the base still in the braked position as packed.
Remove any chocks from the pallet base to enable
later table movement (iv below).
iiPosition the ramp provided (in the packing case)
adjacent to the base of the case.
iiiFollow the instruction provided in section 5.1 of this
manual and place the table base into the ‘Wheel’
orientation or mode.
ivWith at least two people to support the table’s weight,
push it (do not pull it) off of the pallet base and down
the ramp.
vNote that the table should not be wheeled over rough
ground, always use a trolley until a smooth floor area
has been reached.
viShould it be necessary to lift the operating table refer
to section 3.2 where suitable lifting points and
methods are detailed.
3.1.3Unpack the mattress set and fit a mattress to
each tabletop section as detailed in section 5.2.10.
3.1.4Any packaging materials should be recycled or
disposed of in accordance with current legislation.
supplied fitted to the cord is not suitable it should be
replaced with a suitable plug wired as below. If the plug is
a fused type, a 10A fuse must be fitted. The mains supply
cord must always be wired as follows:
Brown internal cord to LIVE
Blue internal cord to NEUTRAL
Green internal cord to Earth
(Note: The T30-a and T30-m Tables are Class II,
Type BF, there is no EARTH connection through to
the table)
3.1.7T30 Tables are powered by internal rechargeable
batteries which are connected and charged before delivery.
An internal mains powered battery charger is incorporated
in the table’s base. The table batteries should be recharged
(see section 5.3.1) and the table operated through the cycle
of movements detailed below to check and ensure correct
function, before the tables are first used.
Full Trendelenburg / reverse Trendelenburg
Maximum to minimum height
Maximum tilt, left and right
Maximum to minimum break
CAUTION
To complete and maintain the antistatic
pathway the table must be used on an
electrically conductive or antistatic floor and
with mattresses supplied by Eschmann
Equipment.
3.1.8The table has an antistatic pathway from the
tabletop, through an internal resistor, to the castors. To
complete the antistatic pathway, the table must be used
on an electrically conductive, or on an antistatic floor (also
see the warning in section 6.1).
3.1.9As with all medical electrical equipment care
should be taken with regard to electromagnetic
compatibility (EMC) during installation. These instructions
are written in line with the latest international standards
(EN 60601-1-2:2001) and are designed to minimise the
risk of electromagnetic compatibility issues. T30 Tables
should be installed and put into service in accordance with
the EMC information provided in the Technical Data section
of these ‘Instructions for Use’ (Section 9.11).
CAUTION
If the mains plug is changed it is most important
that a fuse of the correct type, size and rating
is used (see Technical Data, section 9.5.4).
3.1.6T30 Tables require a mains electrical supply
corresponding to the voltage shown on the Serial Number
Plate located on the table base. Only use the Eschmann
mains supply cord provided with the table. If the plug
3.1.10The table should be cleaned and disinfected prior
to its first use as detailed in section 8.2 and 8.3 and then
commissioned in accordance with any local procedures
applicable to new equipment, this should include staff
training. Eschmann supply a range of wall charts with the
table, additional training aids and on-site training can be
arranged, contact Eschmann [Tel. +44 (0) 1903 765040]
for more information.
P12/77T-IM122b
T30
OPERATING TABLES
Series
3.2Lifting the operating table
WARNING
The table is heavy and at least four strong
people are required to lift it. Ensure that
adequate precautions are taken (e.g. wear
protective shoes, use the correct straps).
3.2.1The T30 Table should only be lifted as a last
resort. Ideally it should be placed on a trolley directly from
the delivery vehicle, or moved on the base of the delivery
packing case (overseas only) and then rolled down the
ramp provided.
3.2.2If required the T30 Table should only be lifted by
placing suitable webbing straps underneath the table base
in the positions indicated in Fig. 3.1 (having placed the
T30-a Table into its ‘castor’ orientation to increase ground
clearance) and observing the notes that follow.
3.2.3Extreme care should be taken to pad the straps
where they pass the base covers and the tabletop sections
to avoid damage. Take special care not to cause damage
to the lower edge of the base covers.
3.2.4Before lifting remove all tabletop sections,
accessories and mattresses, to minimise the weight to
242.7kg* (T30-a Table) or 190.7kg* (T30-m Table). Place
the tabletop into a level plane in both directions
(i.e. tilt and Trendelenburg) as shown in Fig. 3.1. Lower
the short trunk and tabletop to their maximum limits.
*Nominal weight for table with a General Surgical short
trunk, add 14.7kg if an Orthopaedic short trunk is fitted.
3.2.5The table should only be lifted the minimum
amount required and not carried. Lift the table sufficiently
high to allow a fully decked pallet to be slid underneath.
3.2.6When lowering the table after the lift take care
not trap feet under the table’s base. The table should be
placed into its ‘braked’ orientation whilst on the pallet.
3.2.7 Transport the table on the pallet using a forklift
truck or similar equipment ensuring the table is strapped
securely to the pallet.
3.2.8Inspect the table for any signs of damage and
check all functions prior to placing the table into service.
3.3Technical
3.3.1The following sections are provided for the user
to note prior to using a T30 Table.
3.3.2The T30 Table meets the requirements of international standards (see section 9.8) and conform
dimensionally to meet most requirements, for the full table
technical specification details refer to the Technical Data,
section 9.0.
3.3.3The T30 Table should only be used on an antistatic floor and is classified as type ‘BF’ (i.e. the table has
isolation from earth equivalent to that of type ‘BF’ equipment
when the mains cord is attached).
3.3.4The antistatic properties of the table depend upon
the use of the recommended mattresses (i.e. Eschmann
antistatic mattresses ONLY) also see section 3.1.8.
Use webbing straps of minimum width 100mm
and a minimum breaking load of 1000kg.
Position adequate padding in the areas indicated
(on both sides) to avoid damaging the table covers.
T30-a Position straps inside, but as
close as possible to the castors.
T30-m Position straps inside
castors as indicated.
Fig. 3.1 Lifting the T30 Table
T-IM122bP13/77
3.3.5Only Eschmann accessories listed in this manual
should be used on the table and in accordance with the
‘User Handbook’ supplied with the accessory. Accessories
available from Eschmann are listed in section 7.0 and
Appendix 9 (for Orthopaedic accessories). Other
accessories, especially those that could compromise table
stability, must not be used. Use of other equipment with
T30 Tables should only be considered after evaluating the
safety of the patient and personnel. For accessories that
fit onto the sidebar ensure that they are compatible with
the sidebar fitted to the table. These tables can be supplied
with standard UK, Euro USA or Denyer style sidebars, see
section 1.1.2
3.3.6Provision of a diagnostic port within the table
enables access for reprogramming the table’s software,
down-loading of fault information and service functions.
Use of this port MUST be limited to trained service
personnel only, and should only be used in accordance
with the correct Eschmann manuals (see 3.3.9 below).
3.3.7The table has four fuses (two on the T30-m) that
the user has access to in the event of failure. The position
of these are shown in Fig. 2.2 and are replaced as detailed
in section 5.3.5.
3.3.8Eschmann can provide customers with manuals
(see 3.3.9 below), for use by them in maintaining their own
equipment. These manuals contain schematic diagrams,
component part lists, descriptions and calibration
instructions which will assist the customer’s Eschmann
trained personnel to service the equipment or replace parts
(which should only be obtained from Eschmann).
3.3.9The following manuals are applicable to the T30
Tables and their accessories (the part number is in brackets
following the manual reference) they are available to order,
see inside front cover for contact details:-
T-SM54 (113389) - Service manual
T-IPL41 (113390) - Illustrated parts list
T-IM94 (111012) - Application software manual
T-IM56 (698907) - General accessory leaflet
Note: Some accessories are provided with their
own ‘User/Service Handbooks’.
3.3.10Appendix 1 provides a log that can be used to
record those people trained in the safe use of this operating
table. It is suggested that this is used to ensure that ALL
personnel using this table, are aware of all the warnings
and cautions contained within these ‘Instructions for use’.
3.3.11The T30 Tables and their accessories, as listed
in these ‘Instructions for use’, do not contain ‘Latex’.
P14/77T-IM122b
4.0 SAFETY NOTES & CAUTIONS
4.1Warnings
Ensure you are familiar with all the warnings and cautions
provided within these ‘Instructions for use’ before using
the table.
T30
OPERATING TABLES
Series
WARNINGS
T30 Tables have been designed to minimise the
possibility of accidental electrosurgery burns.
Contact with any metal surfaces (e.g. table
sidebar, or other equipment etc.) can cause burns
during electrosurgery and must be avoided.
T30 Tables are not rated as AP or APG and should
not therefore be used in the presence of explosive
gases.
T30 Tables have been designed for patients
weighing up to 300kg (47 stone) with their centre
of gravity (normally the umbilicus) positioned
close to the column on the trunk sections.
However patient positioning and additional loads
from accessories can compromise table stability
and strength. Refer to the graphs in section 6.4
for safe loading.
To comply with BS EN 60601-1:1990 some
accessories have been designed for a maximum
evenly distributed load, see the ‘User Handbook’
supplied with each accessory.
With the table in (or during transition into) the
‘castor’ or ‘wheel’ orientation, the centre of
gravity of the patient (normally the umbilicus)
should lie no more than 200mm away from the
centre of the column. Whenever this is not
practical the table should be adequately
supported (e.g. by at least two able people).
The head and leg sections are designed to
support and position the corresponding part of
the patient’s weight only. Damage leading to
failure of the section may be caused if excessive
weight is applied. Take care when handling these
sections to avoid strain and ensure no body parts
or objects are trapped when replacing or
adjusting them.
Only use Eschmann accessories and sections
that are compatible with this table. When parts
are replaced during maintenance procedures,
ensure that ONLY parts supplied by, or from,
Eschmann Equipment are used. Alternatives,
although similar, may affect the safety of the
table. Eschmann cannot be held responsible for
service, modification or adjustments to the
equipment, when performed by other than
Eschmann accredited personnel.
Illustrations and descriptions of patient
positioning are for guidance only. It is the
responsibility of the operating surgeon to make
sure the patient is positioned correctly for each
procedure. (See section 6.0).
During any table positioning procedure care
should be taken to ensure the patient’s safety.
In particular during Trendelenburg and tilt
movements the patient should be supported to
ensure they remain secure on the tabletop. The
patient’s weight should be supported whenever
the sections are adjusted or removed from the
table during repositioning.
During ANY movement of the table or tabletop,
ensure that no part of either patient or hospital staff,
or object (e.g. drapes, infusion tubing, diathermy
connections, ECG cords etc.) can become trapped
between any moving and/or stationary equipment,
or in a pinch point, causing injury or damage to
equipment. Particular table movements that should
be operated with care are reducing the height and
Trendelenburg, which can cause trapping
situations. Always ensure adequate slack is
available in drapes and tubing for the movement
required (e.g. maximum Trendelenburg).
When pushing the table with a patient (maximum
weight 135kg for T30-a or 200kg for T30-m , see
section 5.1) always ensure that the patient’s limbs
are secure on the tabletop to prevent crushing
or trapping them against another object, always
use cot sides (available as an accessory).
Always keep the patient under observation (e.g.
check respiratory and circulatory system and for
the possibility of pressure sores etc.) and
correctly positioned whilst on the table.
Ensure that electrical equipment connected to the
communication port (available during
maintenance procedures only) complies with
appropriate electrical safety standards. Note that
standards compliance of this product may be
affected if noncompliant equipment is attached
to the communication port.
Do not use any table or accessory if there are
visible signs of damage or wear and tear that
could compromise safety.
T-IM122bP15/77
CAUTIONS
Do not place either heavy accessories, or, long
accessories that could impose high torques,
to the sidebars, as this may lead to damage of
the rails.
Do not exceed the duty cycle for any table
motor drive as detailed in the technical data
section 9.5.5.
Care should be taken moving the table over soft
floors (e.g. carpet or ‘cushion’ flooring) as these
will increase resistance to movement
compared to normal hospital ‘hard’ flooring.
4.2Do’s and Don’ts
Attention to the following points will prolong the life and
efficiency of the T30 Table and will help to avoid the risk of
accidents, or damage. Other safety notes and warnings
are also given within the text of this manual and these
should be noted during use of the table.
DO:
♦Keep these ‘Instructions for use’ close-to-hand.
♦Read these ‘Instructions for use’ carefully before
adjusting, moving or using the table.
♦Use the table on an antistatic floor to prevent
inadvertent static buildup.
♦Use only the correct Eschmann mattresses and
accessories that are compatible with the T30 Table as
detailed in these instructions.
♦Check that handset cables and standby controls are
not damaged before use.
♦Check that the table and its accessories are not worn
or damaged, or are in any way not suitable for the
intended purpose, before use.
♦Check that all the sections (e.g. head and leg) and
accessories are secure, and put the table base in the
‘braked’ position before use.
♦Ensure that all cables are not stretched leading to
disconnection or damage during movement or
readjustment of the operating table or patient.
♦Remove table accessories and their clamps (in
particular rotary clamps) from sidebars, when they
are not being used.
♦Read and follow the instructions for cleaning, and for
the care of the table and mattresses.
♦Switch ‘off’ and disconnect from the mains electrical
supply prior to cleaning and/or disinfecting the table
and when it is not being recharged.
♦Place the table batteries on charge at the end of every
day or shift (see section 5.3.1)
♦Ensure that the table and accessories are serviced at
regular intervals (every six months is the
recommended frequency) only by Eschmann trained
personnel.
Arrange service contracts through Eschmann
Equipment [Tel. +44 (0) 1903 765040]
♦Ensure that only the Eschmann mains cord supplied
with the table is used to connect the table to the mains
supply.
♦Ensure that only Eschmann supplied parts are used
during part replacement.
DO NOT:
♦Do not lift the table by its tabletop.
♦Do not move the table with a patient without cot sides
in place on either side of the tabletop.
♦Do not push the table over rough surfaces, use a
trolley.
♦Do not drop the table (or individual sections).
♦Do not put heavy weights on the table sections,
observe the maximum advised loading.
♦Do not put sharp objects on, or against, mattresses,
pads, or the radiographic tabletop.
♦Do not place any objects on the base covers
♦Do not drop heavy objects onto the radiographic
tabletop or base covers.
♦Do not spill oil, ether, or other fluids onto the mattresses
or the pads.
♦Do not pull the table by any of the tabletop sections,
or accessories, always push it.
♦Do not service this equipment unless you have been
trained by Eschmann.
♦Do not attempt to fit an infill section to another infill
section or a T20 series table.
4.3Daily ‘Before use’ test
It is recommended that the following ‘Daily test’ is carried
out before using the table:
iCheck batteries and charge if required.
ii Check table responds to the all
handset commands.
iii Check table responds to all Standby
control panel commands.
P16/77T-IM122b
4.4Accessories
The accessories available from Eschmann for the T30
Tables are listed in section 7.0 and Appendix 9. Use of
other equipment with T30 Tables should only be considered
after evaluating the safety of the patient and personnel.
Inadvertent use of incorrect accessories could damage the
table and lead to injury. Always ensure that the information
in the ‘Instruction’ or ‘User Handbook’ supplied with the
accessory are complied with and follow all the safety notes
contained within them during use. For accessories that fit
onto the sidebars ensure that they are compatible with the
sidebars fitted to these tables. These tables can be supplied
with standard UK, USA or Denyer style sidebars.
4.5Manual handling
WARNING
Ensure care is taken when moving adjusting
or lifting any part of the table or patient. Note
the guidelines provided.
T30
OPERATING TABLES
Series
4.5.1During adjustment or changing the configuration
of the T30 Tables, there are occasions when the user
should be aware of the safe practises to be employed
during manual handling or adjustment of parts of the table.
Appendix 4 provides manual handling advice and the
weights of the heaviest sections and accessories
commonly used. When lifting, carrying or fitting heavy
components it is recommended that care is taken and two
or more people are employed when required.
4.5.2These ‘Instructions for use’ advise supporting the
weight of the patient during adjustment of any section, this
requires the intervention of several personnel, some
supporting the patient’s limbs and others adjusting the table
sections.
4.5.3When moving the table note that extra effort is
required to start the table moving, take care not to strain
limbs or back. The table should not be moved when heavily
loaded. Note the warnings and cautions provided. It is good
practice to use two or more people when moving a table.
4.5.4When changing the orthopaedic short trunk
section the special ‘cart’ is recommended and should
always be used, this is fully detailed in section 5.2.5.
4.5.5To ensure all the orthopaedic accessories are
stored correctly and are easily accessible for use, the
orthopaedic trolley should be used. This is fully detailed
and illustrated in Appendix 8.
T-IM122bP17/77
5.0 OPERATING THE TABLE
This section has been split into sections as follows:
5.1Moving/Operating the table base.
5.2Using the removable sections.
5.3Using the table’s powered/electrical functions.
Note: Orthopaedic procedures are covered in Appendix 5
onwards.
5.1Moving/Operating the table base
WARNING
Always push the table (do not pull it) at a
suitable height ensuring that it is stable at all
times (maximum stability will be at minimum
height).
Do not move the table around the theatre with
a patient weighing more than 135kg (T30-a
Table) or 200kg (T30-m Table). Take care not to
collide with personnel or equipment.
Ensure that the patient is adequately supported
and restrained (especially limbs) using cot
sides as appropriate.
Ensure all connections via cord or tube, to the
patient or table, have either been disconnected,
or are only attached to equipment that will move
with the table to avoid inadvertent
disconnection.
T30-a Table only When changing from ‘castor’
or ‘wheel’ orientation, to ‘braked’, ensure no
objects (e.g. cords, tubing etc.) can become
trapped beneath the table base and floor.
If the table is difficult to move, check for objects
under castors and that castors are maintained
and kept clean and free from foreign objects.
orientations does not require battery power, they are
manual operations achieved by using one or both of the
foot pedals on the table base.
The table should always be left in the ‘braked’ orientation
with both pedals in their raised positions. Do not leave the
table with the wheel pedal up and the castor pedal down.
When familiar with this section the label on the table base
serves as a quick reference guide to pedal operation.
5.1.1.1‘Wheel’ orientation (from ‘braked’)
T30-a wheel pedal operation is easily achieved if the pedal
is pressed down with the right foot whilst steadying yourself
with hands on the tabletop. Stand on the side of the table
on which the footpad is located. Use a steady ‘press’ rather
than a ‘rapid depression’ of the pedal.
To place the table into the ‘Wheel’ orientation from the
‘braked’ orientation, press the wheel pedal (shown in
Fig. 5.1) down steadily until you hear an audible ‘click’ (action
indicated in Fig. 5.1). This ‘click’ indicates that the pedal
has locked in the down position. Do not continue to press
the pedal after the ‘click’ as this will release the internal
catch and the pedal will not lock down. If this does happen
the pedal must be allowed to rise fully (this resets the internal
catch) before pressing it down again.
When placed into the ‘Wheel’ orientation the table base is
supported on two wheels at the short trunk end and two
castors at the long trunk end. To move the table in ‘wheel’
orientation always push it (do not pull it) from the long trunk
end, moving the end nearest to you, left or right, to steer the
table in the required direction. The table is in ‘wheel’
orientation when the wheel pedal is in the lowered position
and castor pedal is in the raised position.
5.1.1.2‘Castor’ orientation (from ‘wheel’)
T30-a Table's castor pedal operation is easily achieved if
the pedal is pressed with the left foot whilst steadying yourself
with hands on the tabletop. Stand on the side of the table
on which the footpad is located. Use a steady ‘press’ rather
than a ‘rapid depression’ of the pedal.
The T30-a Table has two foot pedals, see section 5.1.1,
the T30-m Table has a single foot pedal, see section 5.1.2.
5.1.1 T30-a Table base
The T30-a Table can be moved easily on built-in castors
and wheels. Normally the table rests on brake pads at the
long trunk end and wheels at the short trunk end, these
provide a secure and static location on the operating theatre
floor. Lower the table to a suitable height to achieve a stable
position before moving the table.
To move the T30-a Table it can be placed onto its wheels
and castors in two ways, providing both a ‘castor’
orientation and a ‘wheel’ orientation. The latter enables
easy movement of the table in a straight line (down a
corridor for example). Moving the table into either of these
To place the table into the ‘castor’ orientation from the ‘wheel’
orientation, press the castor pedal (shown in Fig. 5.2) down
steadily until you hear an audible ‘click’ (action indicated in
Fig. 5.2). This ‘click’ indicates that the pedal has locked in
the down position. Do not continue to press the pedal after
the ‘click’ as this will release the internal catch and the pedal
will not lock down. If this does happen the pedal must be
allowed to rise fully (this resets the internal catch) before
pressing it down again.
When placed into the ‘castor’ orientation the table base is
supported on four castors, two at each end of the table.
This orientation enables the table to be moved in any
direction including sideways and swivelling within its own
length. To move the table always push it in the required
direction, never pull it. The table is in ‘castor’ orientation when
both pedals are in their lowered positions.
P18/77T-IM122b
T30
OPERATING TABLES
Series
5.1.1.3‘Castor’ orientation (from ‘braked’)
To place the table into ‘castor’ orientation from the ‘braked’
orientation follow 5.1.1.1 to place the table into ‘wheel’
orientation and then follow 5.1.1.2 to complete the move
into the ‘castor’ orientation.
Note: It is not critical that sections 5.1.1.1 and 5.1.1.2 are
carried out in this sequence the reverse is equally suitable
and correct.
The table is in ‘castor’ orientation when both pedals are in
their lowered positions.
5.1.1.4‘Braked’ orientation
To place the table into the ‘braked’ orientation move both the
pedals into their raised position as follows.
Stand on the side of the table on which the footpad is located
and use the foot advised in 5.1.1.1 or 5.1.1.2 as appropriate.
Steady yourself with your hands on the tabletop. Press the
pedal ‘firmly down’ (action indicated in Fig. 5.2), this
disengages the internal locking catch. Release pressure
on the pedal and allow it to rise. The table will gently lower
into its ‘braked’ orientation, the motion is softened by an
internal damper.
The table is in ‘braked’ orientation when both pedals are in
their raised positions.
The T30-m Table base has been provided with four large
castors. The table rests on these castors at all times
providing either a secure and static location on the
operating theatre floor (‘braked’), or two modes for easy
movement (‘castor’ and ‘wheel’ modes).
Adjust the table to a suitable height to achieve a stable
position before moving the table. The T30-m Table should
always be left ‘braked’ with the foot pedal in its lowest
position. When familiar with this section the label on the
table base serves as a quick reference guide to pedal
operation.
Place the table into ‘castor’ mode or ‘wheel’ mode to move
it. The latter enables easy movement of the table in a
straight line (down a corridor for example). The ‘castor’
mode provides full free wheeling mobility with 360° rotation
and sideways movement.
Adjusting the table into either mode does not require battery
power, they are manual operations achieved using the
single foot pedal on the table base (see Fig. 5.4).
5.1.2.1Foot pedal operation
Foot pedal operation is easily achieved when the pedal is
operated with either foot whilst steadying yourself with your
hands on the tabletop. Stand on either side of the table or
the pedal end, which ever is the most suitable and easy.
Do not operate the pedal from the end of the table when a
long table section (e.g. a leg section) has been fitted to the
long trunk end, this may require unnecessary stretching
by the operator to reach the pedal. Operate the pedal from
the side of the table.
Use a steady motion rather than a ‘rapid’ movement of the
foot pedal, this provides easy identification of the ‘snap’
into any of its three positions.
5.1.2.2‘Braked’
To place the table into the ‘braked’ orientation press the
foot pedal (see Fig. 5.4) down to its lowest position. Operate
the pedal as detailed in section 5.1.2.1. and press the foot
pedal down until it snaps into its lowest position.
The pedal can be moved from its highest ‘wheel’ position
through its central ‘castor’ position and into the ‘braked’
position in one easy movement.
The table is ‘braked’ when the foot pedal is in its lowest
position.
The pedal is moved up from its lowest ‘braked’ position by
lifting the pedal up with the top of the foot, or down from its
raised ‘wheel’ position by pressing the pedal down with the
ball of the foot, until the pedal snaps into the central ‘castor’
position.
The table is in ‘castor’ mode when the foot pedal is in its
central position.
Note: When the table has been moved to the required
location always leave the table ‘braked’.
5.1.2.4‘Wheel’ mode
WARNING
Do not push the table in the ‘wheel’ mode until
you are sure the castors have moved into their
correct orientation for the ‘wheel’ mode, as
detailed in the second stage below and Fig. 5.4.
If the table is pushed with the castors locked out
of position this will cause undue wear leading to
failure of the short trunk end castors.
Placing the table into the ‘wheel’ mode is a four stage
procedure which will ensure that the short trunk end
castors are locked in position correctly.
First, if the table is ‘braked’ move the foot pedal into its
‘castor’ position as detailed in section 5.1.2.3.
Second, push the table forwards from the long trunk end
until both the short trunk end castors have swivelled into
the position ‘B’ shown in Fig. 5.4. They do not need to be
exactly in line, but they should not be as shown in ‘A’ Fig. 5.4
(i.e. leading their mounts).
Third, lift the pedal up with the top of the foot, as detailed
in section 5.1.2.1, until the pedal snaps into its highest
‘wheel’ position.
Fourth and finally, continue to push the table in a straight
line, the short trunk end castors will lock automatically in
line with the table base. This enables the table to be moved
easily down a corridor, steer from the long trunk end.
The table is in ‘wheel’ mode when the foot pedal is in its
highest position and the short trunk end castors have locked
in-line with the table base.
Note: When the table has been moved to the required
location adjust the table into ‘castor’ mode to enable full
mobility and positioning and then leave the table ‘braked’
(press the pedal fully down).
5.1.2.3‘Castor’ mode
To place the table into the ‘castor’ mode, move the foot
pedal (see Fig. 5.4) into its central position. Operate the
pedal as detailed in section 5.1.2.1 and move the foot pedal
until it snaps into its central position.
P20/77T-IM122b
T30
OPERATING TABLES
Illustration of the table’s base ‘quick
reference guide’ label. Note the
booklet symbol that indicates
reference to these instructions
should be observed.
113136-01
Detail of the required short trunk end castor postion, which should be obtained, before
lifting the foot pedal into its top position, placing the table base into 'wheel' mode.
Note:
In position A the castor leads its mount, in position B the castor trails its mount.
Series
A
See A above
See B above
B
WHEEL
Select 'castor' mode' (pedal in the central position).
Push the table until the short trunk end castors are
correctly aligned, as indicated. Lift the pedal up fully
into the 'wheel' position (to stop the short trunk end
castors swivelling). This will enable the table to be
'wheeled' in a straight line. Push and steer the table
from the pedal (head) end (see WARNING in section
5.1.2.4).
CASTOR
Lift pedal up, or press pedal down,
into the central position to enable the
table to be 'castored'. All four castors
will swivel allowing full table mobility.
BRAKE
Press pedal down fully with the castors
in any position to brake all four castors.
Fig. 5.4 Operating the T30-m Table base
T-IM122bP21/77
5.1.3 Moving the table with a patient
The normal table position when moving it with a patient is
level in both directions (i.e. tilt and Trendelenburg) and
with the patient in the supine or lateral recovery position.
Fig. 5.5 shows an alternative position that can be used,
the notes in Fig. 5.5 also apply when moving the table with
the patient in the supine or lateral recovery position.
Notes:
Always use cot sides.
Lower table to a suitable height.
Ensure patient limbs are secure.
Check patient weight and position complies with Fig. 6.3.
Always push table from the head end (do not pull).
T20-m table illustrated but position also applies to T30-a and T30-m
tables with or without an infill section fitted to the long or short trunk.
Fig. 5.5 Alternative table position for moving table with a patient
P22/77T-IM122b
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.