Eschmann LS5, SES2555, QuickVac Service manual

LS5/SES2555
& QuickVac
VACUUM & NON-VACUUM AUTOCLAVES
Service Manual
606023
ST-SM33l
Introduction
Read these Instructions before use
Keep thIs ‘Service Manual’ in a safe convenient place for future reference.
Operating controls
& display
Autoclave
components
Operation cycle
Maintenance
Part removal
& replacement
Technical data
Eschmann After Sales Service Department
The Eschmann After Sales Service Department is staffed and equipped to provide advice and assistance during normal office hours. To avoid delays when making enquires, please quote the Model and Serial Number of your Autoclave which is shown on the Serial Number plate, the location of which is shown below. Please ensure you include all alpha and numeric digits of the Serial Number.
The Serial Number
F 4
L
0
S
0
4
m
VA
A
C
6
U
9
U
6
M
plate is located here
1 8
1
1
S
3
N
A
L
2
S
x
4
3
5
8
B
1
0
5
9
0
0
R
9
0
3
9
E
2
4
-
F
0
5
6 8 3
­4 5
6
­7
6 2 3 0
V
4
.
4
A
E s c
5
h
0
m
-
a
6
n n
0
P
E
H
e
q
t
u
e
z
i
r
p
R
m
o
e
a
n
W
d
t ,
,
e
L
s
a
t
n
S
c
u
i
s
n
s
g
e
, x , E n
B
g
N
la
1
n
5
M
d 8
A
T
D
J .
E I N E
N G L A
N
D
789789-1
For further information visit www.eschmann.co.uk
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England. T el: +44 (0) 1903 765040. Fax: +44 (0) 1903 762006.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Patents and Trade marks
The ESCHMANN name and logo are registered trade marks of Eschmann Holdings Limited . “Eschmann Equipment” is a trading name of Eschmann Holdings Limit ed. “LS5”, “Little Sister”, “QuickV ac” and “SES2555” are t rade marks of Eschmann Holdings Limit ed.
Patents : Patents Pending plus - Pat. US5090033 and Pat. GB2238407
Copyright © 2003
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying, recording or otherwise without written permission from Eschmann Holdings Limited. The information in this publication was correct at the time of going to print. The Company, however, reserves the right to modify or improve the equipment referred to.
If the CE mark is affixed to the product, it indicates compliance with Council Directive 93/42/EEC of 14 June 1993 concerning medical devices.
Service Manual
ST-SM33l September 2003
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
CONTENTS
1 Introduction
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Description
2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Operating controls & display
3 Operating controls & display . . . . . . . . . . . . . . . . . 4
Push buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Digital displays . . . . . . . . . . . . . . . . . . . . . . . . . 5
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 Autoclave components
4 Autoclave components . . . . . . . . . . . . . . . . . . . . . . 5
5 Operation cycle
5 Operation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
6 Cycle details
6 Cycle Details . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
7 Maintenance
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
7.1 Special functions . . . . . . . . . . . . . . . . . . . . 12
7.2 Protected functions . . . . . . . . . . . . . . . . . . . 12
7.3 Errors and faults . . . . . . . . . . . . . . . . . . . . . 12
Power failure . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Aborted cycles and errors . . . . . . . . . . . . . . . . 12
Reset thermostat (overheat) . . . . . . . . . . . . . . 12
7.4 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 13
7.5 LED Function test . . . . . . . . . . . . . . . . . . . . 13
7.6 Connections, air, water and electrical . . . . . 13
7.7 Safety checks (Weekly & Annual) . . . . . . . . 13
8.17 Gauze filter . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.18 Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.19 Safety valve . . . . . . . . . . . . . . . . . . . . . . . . 33
8.20 Band heater temp. sensor . . . . . . . . . . . . . 33
8.21 Mains switch . . . . . . . . . . . . . . . . . . . . . . . . 33
8.22 Band heater . . . . . . . . . . . . . . . . . . . . . . . . 33
8.23 Chamber water level sensors . . . . . . . . . . . 33
8.24 Reset thermostat (overheat) . . . . . . . . . . . . 34
8.25 Cycling thermostat . . . . . . . . . . . . . . . . . . . 34
8.26 Thermal fuse . . . . . . . . . . . . . . . . . . . . . . . 34
8.27 Display board . . . . . . . . . . . . . . . . . . . . . . . 34
8.28 Mains fuse holders . . . . . . . . . . . . . . . . . . . 34
8.29 Membrane key pad . . . . . . . . . . . . . . . . . . . 34
8.30 Reservoir level sensor . . . . . . . . . . . . . . . . 34
8.31 Water pump . . . . . . . . . . . . . . . . . . . . . . . . 35
8.32 Test probe port and pressure gauge . . . . . . 35
8.33 Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.34 Band heater resettable thermostat . . . . . . . 35
8.35 Vacuum pump maintenance . . . . . . . . . . . . 35
8.37 Pressure door lock . . . . . . . . . . . . . . . . . . . 36
9 Technical data
9 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Parts lists
Parts list 1
Electrical . . . . . . . . . . . . . . . . . . . . . . . . . 39 - 41
Parts list 2
Pipes and plumbing . . . . . . . . . . . . . . . . . . . . . 43
Parts list 3
General assembly . . . . . . . . . . . . . . . . . . 45 - 47
Tables
Table 1 Key to display A2 . . . . . . . . . . . . . . . . . . . . 10
Table 2 Special functions . . . . . . . . . . . . . . . . . . . . 15
Table 3 Protected functions . . . . . . . . . . . . . . . . . . . 17
Table 4 Fault diagnosis & error number details . . . . 22
Table 5 Protected function P7 quick reference . . . . 29
8 Part removal & replacement
8 Part removal & replacement . . . . . . . . . . . . . . . . . 30
8.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
8.2 Fuse renewal . . . . . . . . . . . . . . . . . . . . . . . 30
8.3 Cover removal & replacement . . . . . . . . . . 30
8.4 Chamber solenoid valves . . . . . . . . . . . . . . 31
8.5 Reservoir . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.6 Mains transformer . . . . . . . . . . . . . . . . . . . 32
8.7 Power board . . . . . . . . . . . . . . . . . . . . . . . . 32
8.8 Door microswitches . . . . . . . . . . . . . . . . . . 32
8.9 Door lock/open solenoids . . . . . . . . . . . . . . 32
8.10 Door lock magnetic catch . . . . . . . . . . . . . . 32
8.11 Chamber temperature sensors . . . . . . . . . . 32
8.12 Heating element . . . . . . . . . . . . . . . . . . . . . 32
8.13 Door seal . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.14 Vacuum pump . . . . . . . . . . . . . . . . . . . . . . 33
8.15 Cooling fan . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.16 Bacterial air filter . . . . . . . . . . . . . . . . . . . . . 33
Figures
Fig.1a Control panel (SES2555/LS5) . . . . . . . . . 5 + 51
Fig.1b Control panel (SES2555/QuickVac) . . . . . 5 + 51
Fig.2 Main components (covers on) . . . . . . . . . . . . . 6
Fig.3 Main components (covers off) . . . . . . . . . . . . . 7
Fig.4 Main components (covers off, non-vacuum) . . 8
Fig.5 Pipe connections and identification . . . . . . . . 14
Fig.6 Vacuum pump part identification . . . . . . . . . . 35
Fig.7 Parts list 1 - Electrical . . . . . . . . . . . . 38 and 40
Fig.8 Parts list 2 - Pipes and plumbing . . . . . . . . . . 42
Fig.9 Sheet 1 Parts list 3 - General assembly . . . 44 Fig.9 Sheet 2 Parts list 3 - General assembly . . . 46
Fig.10 Schematic electrical diag. (non-vacuum) . . . . 49
Fig.11 Schematic electrical diag. (vacuum) . . . . . . . 50
Appendix 1
Appendix 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
ST-SM33l P3/51
1 Introduction
WARNING
Read this service manual before servicing the autoclave and keep it in a safe convenient place for future reference. Always have at hand the ‘Instructions for Use’ as supplied with the autoclave and be familiar with the cautions and warnings contained within it. The servicing of this autoclave should be under the control of a competent person with sterilization training. The autoclave should only be serviced as specified by the manufacturer and detailed within these service instructions.
1.1 This service manual applies to the following autoclaves, fitted with software version 1.51 or later (software version is displayed briefly when the autoclave is switched ‘on’ or B1 is pressed):-
SES Little Sister 5 Autoclave (from Serial Number L5NA3C1001) REF 87-060-01 SES Little Sister 5 Vacuum Autoclave (from Serial Number L5SA3C1001) REF 87-060-25 SES QuickVac V acuum Autoclave (from Serial Number LQSA3D1001) REF 87-061-47 SES 2555 Autoclave (from Serial Number S5NA3C1001) REF 87-060-61 SES 2555 Vacuum Autoclave (from Serial Number S5SA3C1001) REF 87-060-73 SES 2555 Vacuum Autoclave (from Serial Number SQSA3D1001) REF 87-060-73 SES 2555 Autoclave - Australian (from Serial Number S5NA3C1001) REF 87-061-09 SES 2555 Vacuum Autoclave - Australian (from Serial Number S5SA3C1001) REF 87-061-21 SES 2555 Autoclave - Singapore and New Zealand (from Serial Number S5NA3C1001) REF 87-060-85
SES 2555 Vacuum Autoclave - Singapore and New Zealand (from Serial Number S5SA3C1001) R EF 87- 060-97
1.2 This Service Manual contains a technical description, complete maintenance procedures and an illustrated list of spare parts that are available from the Eschmann After Sales Service Department.
Note: When sterilizing lubricated dental handpieces, the reservoir water should be changed every week to prevent contamination of the door seal, and other elastomer components, used in the pressure system.
1.5 It is recommended that the reservoir is drained, allowed to dry and refilled with water* every week. At every service interval the reservoir must be removed, thoroughly cleaned and dried (see section 8.5), and then refilled with water*. * DO NOT USE T A P WATER Eschmann recommend the use of ‘Sterile Water for Irrigation’ alternatively clean distilled or deionized water, or water treated by reverse osmosis may be used.
2 Description
2.1 The autoclave is a portable steam unit heated by an electric element (or elements depending on the model). For sterilization of wrapped loads the vacuum unit creates a vacuum in the chamber. The unit is supplied to suit the mains electrical supply shown in the technical data section.
2.2 The autoclave is electronically controlled and has various sterilizing programmes as detailed in the ‘Instructions for Use’. For typical sterilization cycle times, refer to the technical data section.
2.3 The required sterilizing cycle is selected by pressing various buttons on the control panel. LEDs illuminate to show which temperature and cycle have been selected. The cycle is started by pressing the start button. The sterilizing/drying cycle proceeds automatically until complete and the printer (if fitted) starts automatically when the start button is pressed.
2.4 Indication of cycle status or error codes during a cycle are provided by the digital display and printer (if fitted).
3 Operating controls & display
3.1 The following operating controls and displays are all positioned on the front of the unit. Their use is explained in greater detail within the ‘Instruction for Use’.
Mains switch The symbol ‘I‘ indicates ‘on’ and the symbol
0’ indicates ‘off’ (see 9, Fig. 2).
1.3 Ensure that routine servicing is carried out at regular intervals by either Eschmann trained personnel or suitably trained engineers only, otherwise the warranty could be infringed.
1.4 Keep the ‘Instructions for Use’ and this ‘Service Manual’ readily accessible for reference purposes prior to and during operation, cleaning and servicing of the autoclave.
CAUTION
In common with other systems containing static water reservoirs, water used in this unit can become contaminated over a period of time, or following an aborted cycle, and should be treated as a potential risk of infection.
Push buttons (see Fig. 1a and 1b)
B1 Display change button. Press B1 to display the
cycle counter in display window D1.
B2 Cycle temperature button. Each time B2 is
pressed the cycle temperature alternates between 134°C and 121°C.
B3 Unwrapped load (with drying) cycle button. B4 Unwrapped load (without drying) cycle button. B5 Programme ‘P’ button. For details see Section 7.2. B6 Cycle start button. (Always press and hold) B7 Open door button. (Always press and hold) B8 Single wrapped load (with drying) cycle button.
(V acuum autoclaves only)
P4/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
B9 Hollow instruments (without drying) cycle
button. (QuickV ac and later SES2555 models only)
Note: During some maintenance procedures you can
press and hold buttons to scroll display data quickly .
Digital displays (see Fig. 1a and 1b)
D1 Time, cycle counter, error number display or
minutes left during sterilization or drying phase.
For more details on the error codes see Table 4.
D2 Temperature display. D3 Pressure display.
NOTE: Displays may show other information in special
functions mode, see the ‘Special functions table’.
Indicators (see Fig. 1a & 1b)
A1 Low water indicator. A2 Process display indicator. (See Table 1 in the
instruction for Use).
A3 Door unlocked indicator. A4 Door locked indicator.
NOTE: FOR EASY REFERENCE THESE FIGURES (1a & 1b) ARE REPEATED
ON A FOLD OUT AT THE BACK OF THIS MANUAL
C Selection indicators. These indicators illuminate
to show which function has been selected (cycle or temperature).
4 Autoclave components
4.1 The following components and equipment are incorporated in the autoclave. They are illustrated in Fig.2, 3 and 4. Fig.2 shows the autoclave with covers on Fig. 3 and 4 with covers off. In Fig. 2 and 3 the vacuum unit is shown, some components are particular to the vacuum unit only (vacuum and QuickV ac models are similar apart from the control panel). Fig.4 shows the position of components peculiar to the non-vacuum unit and some common components.
Door close solenoid, Fig.3 item 8. This is operated when the
door closed microswitch (Fig.3 item 19) is actuated by closing the door. The lock (Fig.3 item 12) holds the door closed until the sterilizing cycle is complete. It will also keep the door closed under all fault conditions. As absence of power is also a ‘fault’ the unit power switch must be set to ‘on’ in order to open the door. (See errors and error clearing section).
Fig. 1a Control panel for SES2555 and LS5 Fig. 1b Control panel for QuickVac
and later SES2555 Models
ST-SM33l P5/51
Fig. 2 Main component identification (covers on)
Door open solenoid, Fig.3 item 6. This is operated when the
door open button B7 is pressed to release the primary door lock, or the controller signals it to open the door onto the secondary catch during drying on non­vacuum autoclaves.
Secondary catch, Fig.2 item 12. To ensure that the door
does not fly open should there be residual pressure in the chamber when the open door button is operated.
Primary door lock, Fig.3 item 12. Used to lock the door shut
during all cycles, cannot be released until Pressure door lock (Fig.3 item 34) opens when pressure falls below 0.2Bar(Gauge).
Door closed microswitch, Fig.3, item 19. This is used to
signal the controller that the door is properly closed. It should operate just as the door is fully closed.
Door locked microswitch, Fig. 3, item 18. This is used to
signal the controller that the door is properly locked. It should operate just as the door is fully locked.
Chamber pressure gauge, Fig.3 item 21. Fitted behind the
door cover, it indicates the approximate chamber pressure (without electrical power).
Water reservoir, Fig.3 item 5. This is used to hold water
which is admitted into the chamber via the water fill valve (Fig.3, item 11). The water reservoir also receives hot water and steam vapour discharged from the chamber towards the end of the cycle, via the water discharge valve (Fig.3, item 16). The vacuum pump (Fig.3, item 9) on vacuum units also discharges into the water reservoir.
Reservoir water level sensor (fitted in base of reservoir
position indicated in Fig. 4 item 1). The water level sensor in the reservoir will stop the cycle being started if there is insufficient water in the reservoir to complete a chamber fill. Indicator A1 will be illuminated in a low water condition.
Gauze filter, Fig. 2 item 6. This filters the water leaving the
chamber (easily removed for cleaning).
Mains switch, Fig. 2 item 9. This switch controls mains
power to the autoclave.
Heating element, Fig.3 item 13, and Fig.4 item 4. The
heating element consists of a single immersion element inside the chamber. The heating element is controlled by a solid state relay on the Power PCA (Fig.3, item 26) and protected from overheating by a manual reset thermostat (Fig.3, item 25). Refer to TECHNICAL DATA for heater element loading.
Power board, Fig.3 item 26. There are a number of key
functions provided by the power board: Solid state relays (SSR) control the mains supply to
the water heater element (10A), band heater (10A) and vacuum pump (2A).
Mechanical relay provides additional safety for heaters and pump circuits.
Connections for display board (40-way ribbon cable, J1), 24 volts from transformer (J2), temperature sensors (J3), solenoid valve drives, door microswitches and water level sensors (J4), mains to heaters and vacuum pump (J5) and mains in (J6).
P6/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
Note: Item 34 the ‘Pressure door lock’ is only fitted on later units.
Note: Items 15 and 20 are shown with ‘twinlocks’ fitted, the autoclave is supplied with a blank plug only
Fig. 3 Main component identification (covers off) vacuum and QuickVac autoclave
cables and tubing removed for clarity.
ST-SM33l P7/51
4
1
2
3
Fig. 4 Main component identification (covers off) non-vacuum autoclave with cables,
pressure door lock and tubing removed for clarity (some parts common to vacuum autoclave).
Fuse for protecting the 24V d.c. supply (T3.15A). Two voltage regulators. Autoclave bleeper. Pressure transducer with connection via pipe to
chamber neck ring.
Display board (fitted behind membrane key pad, Fig.2
item 3). The display board interfaces with the power board to control every aspect of autoclave management. The main features are:
7 segment displays (D1, D2 and D3 of Fig. 1) to provide time, temperature and pressure indication and other information for special and protected functions and error detail.
LEDs to indicate low water, cycle status and door status (A1, A2, A3 and A4 of Fig. 1)
LEDs to provide visual indication of cycle selection (C of Fig. 1).
Connections to Power PCA (40 way ribbon cable, J4), key pad (11 way ribbon cable, J3), printer (16 way ribbon cable, J5) and serial port (10 way ribbon cable, J2).
Battery to maintain real time clock. Microprocessor to control the autoclave. This receives
information from the front panel and all the sensors (temperature, pressure, door microswitches, chamber water level, reservoir level etc.). If any errors are detected they are shown on the display, and printed (if a printer is fitted) as error codes.
Reset thermostat (overheat). Fig.3 item 25. The reset
thermostat is fitted underneath the unit and is connected in series with the chamber heater element. The reset thermostat is operated by a fluid-filled capsule clamped to the heating element, providing protection if the temperature of the heater surface exceeds 250°C. It will remake electrically if the reset
button underneath the cabinet is pressed, after giving the heater element time to cool. (Position is indicated by a label on the left-hand side of the unit).
Cycling thermostat, Fig.4 item 3. (Non-vacuum only) This
is used to control the temperature of the chamber heating element, in conjunction with the microprocessor, during drying.
Band heater overheat cutout, Fig.3 item 29. (Vacuum units
only). The band heater cutout is fitted on the band heater and is connected in series with the band heater. It contains a bimetallic disc thermostat which operates if the temperature of the heater surface exceeds 250°C. The cutout will remake electrically if the reset button is pressed when the heater has cooled.
Mains fuses, Fig. 2 item 8. Two fuses are on the front panel
of the unit rated as shown in TECHNICAL DATA, connected in the ‘mains supply’ to the unit.
Transformer, Fig.3 item 27. The transformer converts the
mains voltage to 20V a.c. It is rated at 75VA.
Water fill solenoid valve, Fig.3 item 11. The water fill solenoid
valve controls the entry of water into the chamber. It is electrically operated from the 24V d.c. supply generated and signalled from the power board.
Water discharge solenoid valve, Fig.3 item 16. The water
discharge solenoid valve is used at the end of the sterilizing cycle to allow water and steam vapour from the chamber to pass back into the reservoir. The valve is electrically operated from a 24V d.c. supply generated and signalled from the power board.
Steam bleed solenoid valve, Fig.3 item 2. The steam bleed
solenoid valve is used at the beginning of the heating phase to allow air and steam vapour to pass back into the reservoir, and at the start of discharge to remove
P8/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
pressure before the discharge valve opens. The valve is electrically operated from a 24V d.c. supply generated and signalled from the power board.
Safety valve, Fig.3 item 24. The safety valve is factory set
[pressure rating 2.85Bar (Gauge)] to release excess elevated pressure from within the chamber. It is a primary safety device and must not be readjusted. It should operate between 2.72 and 3.13Bar (Gauge).
Air admit solenoid valve, Fig.3 item 1. (Vacuum units only)
The air admit solenoid valve controls the admission of air (filtered by the Bacterial Filter, Fig.2 item 1) into the chamber during the drying phase.
Vacuum solenoid valve, Fig.3 item 3. (Vacuum units only)
When open, this valve allows the vacuum pump to suck air and steam from the chamber. The valve is electrically operated from a 24V d.c. supply generated and signalled from the power board.
Vacuum pump, Fig.3 item 9. (Vacuum units only) The
vacuum pump is a two stage diaphragm pump and used to suck air and steam from the chamber. The pump is electrically operated from the mains supply and signalled from the power board.
Bacterial Filter, Fig.2 item 1. (Vacuum units only) The
bacterial filter filters the air entering the chamber during the drying phase.
Chamber temperature sensors, Fig.3 item 23. These are
used to sense the chamber temperature. Two are provided within the single sensor unit to provide controller accuracy.
Band heater temperature sensor, Fig.3 item 28. (Vacuum
units only) The band heater temperature sensor is used to control the band temperature during the drying phase.
Thermocouple entry port, Fig.3 item 22. Marked TT on the
chamber door. This is used to insert a thermocouple into the chamber to allow the operating temperature to be measured and, if necessary, adjusted.
Pressure test port, Fig.3 item 15#. Marked PT on the
chamber door. The pressure test port is for inserting a pressure probe to monitor chamber pressure.
Spare test port, Fig.3 item 20#. Spare temperature or
pressure test port for use as required during test procedures.
Membrane key pad, Fig.2 item 3 & Fig.1a & 1b. Incorporates
programme selector buttons, digital display windows and programme indicator LEDs. It is connected to the display board via an 11-way ribbon cable.
Pressure transducer (on the power board). The pressure
transducer monitors the pressure in the chamber and generates chamber pressure signals for cycle monitoring, control, and display.
Printer, Fig.2 item 5. The printer, if fitted, starts automatically
when a cycle is selected and will print out hardcopy of the sterilization cycle. Details of printer output is given in the ‘Instruction for Use’.
Water pump, Fig.4 item 2. (Non-vacuum units only) Used
to pump water into the chamber from the reservoir. The pump is operated from a 24V d.c. supply generated and signalled from the power board.
Fan, Fig.3 item 10. The fan draws air over the chamber for
rapid cooling between cycles. The fan is electrically operated from the 24V supply and signalled from the power board.
Band heater, Fig.3 item 4. (Vacuum units only) The band
heater heats the chamber during the drying cycle. The heating element is controlled by a solid state relay on the Power PCA and protected from overheating by a manual reset thermostat (Fig.3, item 25). Refer to TECHNICAL DATA for heater element loading.
Chamber water level sensors, Fig. 2 items 10 and 11 and
Fig.3 items 14 and 17. These signal the power board when the minimum and maximum water levels have been achieved in the chamber during filling. They are connected to the power board.
Thermal fuse, not illustrated, strapped on the outside of the
chamber behind neck ring (Non vacuum units only). The thermal fuse is connected in series with the chamber heater element, providing protection if the temperature of the thermal fuse exceeds 184°C (chamber will be above 250°C). It cannot be reset and must be replaced if it trips.
RS232 serial connection port (9-way D type), Fig.2 item 4.
This connection port can be used to link the autoclave to a P.C. for calibration, customization, software upgrades, fault diagnosis etc. providing the correct equipment is available (e.g. software).
Magnetic door latch, Fig.3 item 7. (Vacuum units only)
Used to hold the primary door lock (Fig.3 item 12) in place during the vacuum cycle.
Pressure door lock, Fig. 3 item 34. This lock stops the door
opening when the pressure inside the chamber is above 0.2Bar (Gauge). Note: Not fitted on early units.
5 Operation cycle
5.1 A detailed knowledge of the operation of the autoclave is not necessary to be able to repair it effectively; however, a basic understanding of the various processes of autoclave operation which occur during a cycle is given in the following paragraphs.
CAUTION
Ensure that the Autoclave is switched off
before filling the reservoir.
# Note: Items 15 and 20 are shown with ‘twinlocks’ fitted, the autoclave is supplied with a blanking plug only
ST-SM33l P9/51
TABLE 1 - KEY TO DISPLAY A2
Phase 1 LED 1 Cycle started.
LED 2 Chamber minimum fill achieved. LED 3 Chamber filled with water.
Phase 2 LED 1 Heater on.
LED 2 Approximately 90°C reached. LED 3 Final heating phase started.
Phase 3 LED 1 Sterilization phase started.
LED 2-7 One sixth of the sterilization time
completed for each LED illuminated.
Phase 4 LED 1 Condensing stage started
LED 2 Water removed from chamber. LED 3 Pressure approaching ambient.
Note: Eschmann recommend the use of ‘Sterile Water for Irrigation’ alternatively clean distilled or deionized water , or water treated by reverse osmosis may be used.
5.2 Power is switched on by selecting the power switch (O-I) to I (Fig.2 item 9).
NOTE: If the door was locked closed when switching ‘on’ indicator A4 will flash and button B7 should be pressed to unlock the door, the secondary catch can then be released to open the door fully.
5.3 After loading the chamber close the door and hold it firmly shut until the automatic door lock operates. This is shown by indicator A3 switching off and indicator A4 switching on.
NOTE: A cycle cannot be selected (as detailed in the ‘Instruction for Use’) with the door open.
5.4 When the door is closed, with the power switched on, this is sensed by the power board via the door closed microswitch. Indicator A4 will not illuminate until the automatic lock has operated and the door locked microswitch has been actuated by the primary lock. After cycle selection (as detailed in the ‘Instruction for Use’) the cycle is started by pressing B6. If the autoclave has a printer it will start printing automatically .
5.5 As the cycle progresses indicator A2 will show the cycle stage reached (see Table 1) and the digital displays D2 and D3 will display the actual chamber temperature and pressure respectively .
5.6 Control of the cycle is fully automatic with temperature information being monitored by temperature sensors and displayed on the front panel. Timing is controlled by the power board and cycle times cannot be adjusted. By comparing measured values with known time/ temperature relationships, the power board is able to detect faults such as lack of water at the fill stage, or loss of water and steam during the process, and it will indicate such problems by displaying errors codes (e.g. 'Err03') which will be accompanied by an audible warning signal (see Table 4 for details on clearing errors and error codes).
Phase 5 LED 1 Start of drying phase.
LED 2 Drying phase 50% completed. LED 3 Cycle completed (audible bleep).
5.7 The autoclave operates at temperatures slightly above the usual recommended minimums. The operating temperature for the 121°C cycle is set to 122.5°C, and the 134°C cycle is set for 135.5°C.
5.8 Operating information relating to the printer is given in the ‘Instructions for Use’.
5.9 The overall time for the cycle is not fixed and depends on many factors such as the supply voltage, the load, and the ambient temperature. However, the power board will ensure a satisfactory sterilization cycle even when these factors vary over wide ranges. T ypical cycles details are explained in section 6.
6 Cycle details
6.1 A typical vacuum cycle (e.g. 134° Wrapped extended cycle) progresses through several stages. The following information provides details of the controls actuated and the resultant functions performed by the autoclave during selection and running of a typical vacuum sterilization cycle.
Note: The water reservoir must be above minimum and the door must be closed.
Select temperature 134° with B2 button. (121° light
extinguished, 134° light illuminated or vice versa for each press of button B2).
Push B8 button. (Wrapped LED illuminated, all other
cycle LED’s extinguished).
Push B5 button. (Extend LED illuminates, gives
sterilization time of 18 minutes at 134°C).
Push B6 button. (First LED on progress indicator
illuminates. If fitted, the printer records time and cycle data. Mains relay is on. V acuum pump is on. V acuum inlet valve open. All other valves closed).
Wait for pressure to drop to 32kPa. (Vacuum inlet valve
closed. V acuum pump off. If fitted, the printer outputs ‘TPV’ followed by the time and temperature and pressure values).
P10/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
Programmed pause. (If chamber pressure does not rise
by more than 0.9kPa in the last minute, water fill valve opens).
Wait for water to reach minimum level. (Second LED on
progress indicator illuminates).
Wait for water to reach maximum level. (W ater fill valve
closes. Third LED on progress indicator illuminates).
Immersion heater on. (Fourth LED on progress indicator
illuminates. T emperature rises).
Wait for the pressure to rise to 50kPa above atmospheric.
(Fifth LED on progress indicator illuminates. Steam bleed valve opens. If fitted, the printer outputs ‘TPP’ followed by the time and temperature and pressure values).
Programmed pause. (Steam bleed valve closes.
Sixth LED on progress indicator illuminates. Heater off. If pressure above 115kPa, fan on until 115kPa achieved. Fan off, vacuum inlet valve opens and pump starts).
Wait for pressure to fall 20kPa below atmospheric. (Pump
off, vacuum valve closes, heater comes on).
Wait for temperature to reach 135.0°. (Seventh LED on
progress indicator illuminates. If fitted, the printer outputs ‘TSS’ followed by current time, temperature and pressure values, D1 indicates remaining sterilization time in minutes, initially 18 mins).
Eighth to thirteenth LED’s on progress indicator illuminate
at regular intervals. (T emperature should not deviate from 135.5° by more than ±1.5° during sterilisation plateau. If fitted, the printer outputs the temperature and pressure values every 20 seconds. Immersion heater duty drops to approximately 20%).
Programmed pause. (Immersion heater disabled. Beep
sounds. If fitted, the printer outputs ‘TSE’ followed by the temperature and pressure values. Fan on. Fourteenth LED on progress indicator illuminates. Steam bleed valve pulses to exhaust steam).
Wait for 20 seconds. (Air admit valve opens). (Note: Above two steps repeat until end of drying time) Wait for half of specified drying time. (Eighteenth LED
on progress indicator illuminates).
Wait until five minutes before the end of the specified
drying time. (Band heater disabled. Fan on).
Wait until the end of specified drying time. (Nineteenth
LED on progress indicator illuminates. Mains relay off. Three beeps sound. If fitted, the printer outputs ‘TDE’ followed by the temperature and pressure values, ‘Cycle Complete’ followed by the time. D1 displays time).
6.2 A typical non-vacuum cycle (e.g. 134° unwrapped with drying extended cycle) progresses through several stages. The following information provides details of the controls actuated and the resultant functions performed by the autoclave during selection and running of a typical non-vacuum sterilization cycle. Note: The water reservoir must be above minimum and the door must be closed.
Select temperature 134° with B2 button. (121° light
extinguished, 134° light illuminated or vice versa for each press of button B2).
Push B3 button. (With drying LED illuminated, all other
cycle LED’s extinguished).
Push the button B5. (Extend LED illuminated, gives
sterilization time of 18 minutes at 134°C).
Push the button B6. (First LED on progress indicator
illuminates. If fitted, the printer records time and cycle data. Water pump is on. Water fill valve opens. Steam bleed valve opens. All other valves are closed).
Wait for water to reach minimum level. (Second LED on
progress indicator illuminates).
Wait for water to reach maximum level. (W ater fill valve
closes. Water pump is off. Third LED on progress indicator illuminates).
Wait for temperature to drop to 120°C. (W ater discharge
valve opens. Band heater enabled.
Wait for the water in the chamber to drain. (Fifteenth
LED on progress indicator illuminates).
Wait for rate of pressure drop below 40kPa/minute.
(Sixteenth LED on progress indicator illuminates. Seventeenth LED on progress indicator illuminates. If fitted, the printer outputs ‘TDS’ followed by the temperature and pressure values. Vacuum pump on. Vacuum inlet valve opens at 115kPa. Beep sounds. D1 displays drying time left).
Wait for chamber pressure to fall 5kPa below
atmospheric. (Air admit valve opens).
Wait for pressure to return to atmospheric. (Air admit
valve closes).
Mains relay on. (Immersion heater on. Fourth LED on
progress indicator illuminates. Temperature rises).
Wait for the temperature to be above 90° and to rise at
over 1.8°/minute. (Fifth LED on progress indicator illuminates).
Programmed pause and condition check. (Steam bleed
valve closed. Sixth LED on progress indicator illuminates).
Wait for temperature to reach 135.0°. (Seventh LED on
progress indicator illuminates. If fitted, the printer outputs ‘TSS’ followed by the current time and the temperature and pressure values. Immersion heater duty drops. Temperature should not deviate from
135.5° by more than ±1.5° during sterilisation plateau. Eighth to thirteenth LED’s on progress
ST-SM33l P11/51
indicator illuminate at regular intervals. If fitted, the printer outputs the temperature and pressure values every 20 seconds. D1 indicates remaining sterilization time, initially 18 mins).
Programmed pause. (Immersion heater disabled. Beep
sounds. If fitted, the printer outputs ‘TSE’ followed by the temperature and pressure values, and maximum and minimum sterilization temperatures).
Fan on. (Fourteenth LED on progress indicator
illuminates. Steam bleed valve pulses to exhaust steam).
Wait for temperature to drop to 120°C. (W ater discharge
valve opens).
Wait for the water in the chamber to drain. (Chamber
minimum indicator off. Fifteenth LED on progress indicator illuminates).
Wait for rate of pressure drop to go below 60kPa/minute.
(Sixteenth LED on progress indicator illuminates. Seventeenth LED on progress indicator illuminates. Fan off at 1 15kPa. Door catch releases and door is opened by residual pressure. Beep sounds. D1 displays remaining drying time).
display D1 will show F0, each time the ‘P’ button B5 is pressed, this will increase from F0 up to F8* (F4, F5, F6 vacuum autoclave only), and then back to F0 again, until either the start button B6 is pressed to select that function, or, the autoclave is switched ‘off’ or B7 is pressed (with the door open) to cancel special function selection. To access F0 the autoclave PIN number (password) is required. If this is known the protected functions are available and are detailed in T able 3 - protected functions.
* Door must be shut before access to F4, F5, F6 and F7 is permitted.
7.2 Protected functions.
Protected functions are accessed through special function F0 (see above) using the autoclave PIN number (password). If this is known the protected functions are available as detailed in Table 3 - protected functions. When access to the protected functions has been made all displays and indicators are blank except D1 which shows P0. For each press of B5 the “P” number will increase by 1 and then back to P0. Cycle to the number required and press button B6 to perform that function as detailed in Table 3.
7.3 Errors and faults.
Immersion heater duty to 4%. (Steam bleed valve opens.
If fitted, the printer outputs ‘TDS’ followed by the temperature and pressure values. D1 displays drying time left).
Wait for half of specified drying time. (Eighteenth LED
on progress indicator illuminates).
Wait for second half of specified drying time. (Nineteenth
LED on progress indicator illuminates. Immersion heater disabled. Mains relay off. Fan on. Three beeps sound. If fitted, the printer outputs ‘TDE’ followed by the temperature and pressure values, ‘Cycle Complete’ followed by the time. D1 displays time).
7 Maintenance
For easy reference to the buttons, display and indicator numbers (B1, D1, A1, etc.) used within the following sections note that Figs. 1a and 1b are repeated on a fold out page at the back of this manual. Open out this page to provide a quick reference guide.
7.1 Special functions.
Switch ‘off’ the mains switch (item 9, Fig.2). Press and hold in the ‘P’ button B5 whilst switching the mains switch back ‘on’ (display D1 will show F0) to make available several special functions as detailed in Table 2 - Special functions. When the autoclave is switched on in this way display D1 only will be illuminated (with other displays as required depending on which function is selected). Initially
7.3.1 Power failure
If, after switching ‘on’ power there is no visual display , first check power supply connections, both fuses at the front of the unit (see Fuse Renewal) and the fuse in the mains plug (if it is a fused plug). If power fails during a cycle, check supply conditions and fuses at the front of the unit and in the mains plug if fused, also check the mains lead is correctly connected at both ends. Once power is restored the display D1 will show the error number (D2 and D3 will be blank) until cleared by switching the autoclave ‘off’ and then ‘on’ whilst depressing the start button B6.
7.3.2 Aborted cycles and errors
If an error occurs during a cycle (i.e. any time after pressing the start button B6), the cycle will abort and provided power supply to the unit is maintained, the error number will be indicated by the visual display D1 (displays D2 and D3 will continue to show actual temperature and pressure). To display the last error number in display D1 press B1 and B5 together at any time. Errors are detailed in Table 4 ­Fault diagnosis with error number details.
7.3.3 Reset thermostat (overheat)
Some solutions to an error number will indicate that the manual reset thermostat may have tripped and caused the error. Should the heating element manual reset thermostat operate this can be reset by pressing the ‘reset button’ on the underside of the autoclave (approximately 11cm in on the left hand side underneath the unit as indicated on the left hand side of the unit).
P12/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
7.4 Calibration.
The autoclave can be calibrated by running the calibration cycles (Protected functions P3, P4 and P5 consecutively) This will require the positioning of probes (from calibrated instruments or chart recorders) into the autoclave chamber through the test ports. The three available test ports are positioned in the autoclave door, one marked TT (a twinlock) for the chamber test probe, one marked PT (a blank plug) for the chamber pressure probe and a spare blank plug for probes that could be positioned in the load should this be required. To insert a probe remove the port cover and carefully insert ROUNDED tipped probes into the chamber through the seal built into the port, take care not to damage the port seals and replace the cover after use. During calibration of the band heater a temperature probe should be placed under the spring clip provided on the heater, this will require the top cover to be removed to gain access. It is important that the top cover is replaced for accurate calibration. Running protected function P5 will enable chamber temperature and pressure calibration, band heater calibration and a pressure relief valve test. The latter can be avoided if this test is not required during calibration.
7.5 LED Function test.
Annually
i Check that the pressure relief valve operates
correctly at the set pressure. ii Inspect the pressure system for integrity. iii Check door microswitches and interlocks. iv Check door locking mechanism for integrity v Check pressure indicators for correct operation.
NOTE:
Annual inspections should only be undertaken by a ‘Competent Person’. Eschmann can provide comprehensive service contracts which cover preventive maintenance to ensure trouble free operation of your autoclave as well as six monthly and annual inspections of the pressure system to satisfy the requirements of the Pressure Systems Safety Regulations 2000.
To test the function of all LEDs (i.e. to check that they are capable of illumination) press and hold buttons B6 and B7 together with the door open. All LEDs will be illuminated while the buttons are held depressed together.
7.6 Connections, air, water and electrical.
During maintenance refer to Figs. 5, 10 and 1 1 for details on all autoclave connections. Fig. 5 shows all the air and water connections whilst Figs. 10 and 1 1 show the electrical connections for the non-vacuum and vacuum units respectively. (Note: Figs. 10 and 11 open out at the back of this manual). When refitting tie wraps to pipes, use ‘Tie wrap tool 5401’ set at tension 2.
7.7 Safety Checks
The following safety checks, detailed in the ‘Instructions for Use’, are repeated here as they are important.
Operators should ensure that the following safety checks are carried out.
Weekly
i Check that the door opens and closes easily . ii Check the door seal for any signs of damage. iii Check the secondary door catch latches effectively. iv Check for any obvious escape of steam or water
during a cycle (apart from the normal escape via reservoir vents).
Any of the above defects should be attended to by a ‘Competent Person’ immediately and the autoclave should not be used until repair has been effected (see note below).
ST-SM33l P13/51
Fig. 5 Pipe connections and identification
P14/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 2 - SPECIAL FUNCTIONS (See section 7.1 which provides instructions for
accessing these special functions and Fig. 1a & 1b foldout, at the back of this manual)
DISPLAY FUNCTION METHOD OF USE
F0 Protected functions To use protected P functions requires knowledge of the PIN code (code is in HEX
format). With display D1 showing F0 press the start button B6, display D1 shows 0. Use B2 and B3 to adjust D1 to show the first digit of the PIN number, then press B4 to confirm it. D1 will show a 0 in the next position of the PIN number, use B2, B3 and B4 to adjust and confirm the second digit. Continue until all digits have been entered, D1 will show P0. Press button B5 to cycle through the protected functions
P0-P9. Use the information in T able 3 - Protected Functions to carry out the selected P function and return the autoclave to normal operation mode.
F1 Printer test (Note: The printer should be loaded with paper before entering this special
functions mode). With display D1 showing F1 press the start button B6 to start the printer test, a character set printout, display D1 shows F11. To stop the test press button B7, the printer will stop at the end of the test cycle, display D1 shows F0. T o resume normal use switch ‘off’ the mains switch (item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open.
F2 Cycle history With display D1 showing F2 press the start button B6 to start printing a summary
of the last five cycles completed by the unit (cycle temperature, cycle type, cycle count, date and time at start of cycle, date and time at end of cycle and the cycle result). Display D1 returns to F0. To resume normal use switch ‘off’ the mains switch (item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open.
F3 Set delayed start With display D1 showing F3 press the start button B6 to enter the delayed start
time setting mode and delayed start initiation, display D1 will show . If the start hour displayed in D2 is correct (as set last time) and the minutes can be assumed correct (as set last time) press the start button B6 to programme the autoclave to start automatically at this set time. The display D1 will change to F0 when B6 is pressed. Press B7 with the door open to exit the setting mode. If the start time is not correct when entering function F3 (display D1 showing
, display D2 showing the hour set in 24 hour clock mode) press B2 to increase the hour (or B3 to reduce it). Press B4 to confirm the hour when it is correct. Display D1 will show , to increase the minutes (shown in D2) press B2, or B3 to reduce them. Press B6 to confirm the minutes. The display D1 will show F0. Press B7 with the door open to exit the setting mode. The autoclave should be left switched ‘on’ with the required load in place, the door shut and the required cycle selected as follows. When the door is shut a ‘ ’ flashes ‘on and off’ after the time in display D1. This continues while the program selection and any adjustments to it are made (see ‘Instructions for Use’ if ‘ ’ appears in display D2). When cycle selection has been completed and B6 is pressed to confirm them, the ‘ ’ stops flashing. The cycle selected will be run automatically , starting at the time set, a ‘ ’ will be displayed in D1 after the time to show that a delayed start time has been set. On completion of the cycle the autoclave is ready for the load to be removed and normal use resumed without the need to exit from the special functions mode. Note: If the same delayed start time is required each day the setting can be entered quickly once display D1 shows F3, by pressing (B6 B6 B7).
F4 Leak Test With display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONLY) Note. To conduct this test the
chamber temperature must be below 50°C to avoid an Error 38, and the door closed before selecting F4.
display D1 shows F4, press B6. Indicator A4 will flash, press B7 to open the door, indicator A3 will illuminate. Close the door, A4 will illuminate, press B6 this starts an automatic leak test which checks to see if the autoclave system is still pressure tight. Display D3 will show the chamber pressure, after a few minutes display D2 will show the pressure rise to the nearest 0.1kPa. After a further 10 minutes if the pressure has not risen by more than 1.3kPa display D2 will show indicating the autoclave has passed the test (or to indicate the autoclave has failed the test) and an audible sound will be heard. (continues on page 16)
ST-SM33l P15/51
TABLE 2 (continued) - SPECIAL FUNCTIONS
DISPLAY FUNCTION METHOD OF USE
F4 (continued from page 15) Press button B5 and wait until display D2 and D3 go blank. To resume normal
use switch ‘off’ the mains switch (item 10, Fig. 2) and then switch back ‘on’, or press B7 with the door open.
F5 Air Detection Test With display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONL Y) Note. To conduct this test the
chamber temperature must be below 50°C to avoid an Error 38, and the door closed before selecting F5. Connect a needle valve (calibrated to provide a leak of between 1.0 and 1.3kPa per minute) via a short length of 6mm hose to the antibacterial filter before pressing B6.
F6 Steam penetration test. With display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONLY) Note. To conduct this test the
door must be closed before selecting F6.
display D1 shows F5, press B6. Indicator A4 will flash, press B7 to open the door , indicator A3 will illuminate. Close the door , A4 will illuminate, press start button B6. This starts an automatic air detection test which checks if the autoclave system detects a leak if present (it performs a leak test similar to F4 first, to ensure no leaks are present, then admits air through the needle valve). Display D3 shows chamber pressure, after a few minutes display D2 will show pressure rise to the nearest
0.1kPa. After a further 10 minutes if pressure has not risen by more than 1.3kPa display . Display D2 will then show or according to result (test aborts on
). If first test is the air valve will open to conduct detection test. After one minute if pressure <34kPa the leak rate is measured for one minute (if =>34kPa test aborts), display D3 will show pressure. If the rate of pressure rise is more than
0.9kPa display D2 will show indicating the autoclave has passed the test, or
to indicate the autoclave has failed the test (i.e. pressure rise equal or less than 0.9kPa). To resume normal use switch ‘off’ the mains switch (item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open.
display D1 shows F6, press B6. Indicator A4 will flash, press B7 to open the door, indicator A3 will illuminate. Place the Eschmann LS5 Vacuum T ype-S and QuickV ac Steam Penetration T est Device in it’ s pouch (see test device instructions for use) and into the chamber. Close the door, A4 will illuminate, press the start button B6, this will start a special B8 cycle (no drying), D2 and D3 will show temperature and pressure as normal during the cycle. At the end of the cycle the normal audible sound will be heard. Press B7 to open the door, display D1 will be showing F0. To resume normal use switch ‘off’ the mains switch (item 9, Fig.
2) and then switch back ‘on’, or press B7 with the door open. (NOTE: This function can also be used to carry out HTM2010 Performance Qualification (PQ) Tests in conjunction with the ‘Instructions for Testing’ manual ST -IM55).
F7 Disinfect function With display D1 showing F7 load the autoclave with the load to be disinfected
Note. To conduct this test the door must be closed before selecting F7.
F8 Adjust date and time With display D1 showing F8 press the start button B6 to enter the adjust date and
and close the door. Press the start button B6 to start the disinfect function (a 2.5 minute disinfect time at 1 10°C), D1 will continue to show F7 during the cycle and D2 and D3 will show temperature and pressure. At the end of the cycle the normal audible sound will be heard. Press B7 to open the door and remove the load, display D1 will be showing F0 . T o resume normal use switch ‘off’ the mains switch (item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open. (NOTE: The ex-factory condition is for this function to be disabled.)
time function. Display D1 will show and D2 will show the year, to increase the year press B2 (or B3 to reduce it) when correct press button B4 to confirm the year. The display D1 will show and D2 will show the month in numerical format, to increase the month press B2 (or B3 to reduce it) when correct press button B4 to confirm. The display D1 will show and D2 will show the day in numerical format, to increase the day press B2 (or B3 to reduce it) when correct press button B4 to confirm. The display D1 will show and D2 will show the hour , to increase the hour press B2 (or B3 to reduce it) when correct press button B4 to confirm. The display D1 will show and D2 will show the minute, to increase the minute press B2 (or B3 to reduce it) when correct press button B6, display D1 shows F0 display D2 and D3 will be blank. The time and date are now set (seconds were set to zero when B6 was pressed). Press B7 with the door open to continue normal use.
P16/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 3 - PROTECTED FUNCTIONS (See section 7.2 which provides instructions for
accessing these special functions and Fig. 1a & 1b foldout, at the back of this manual)
DISPLAY AND FUNCTION METHOD OF USE
P0 - Print last 5 failed cycles This protected function prints a summary of the last five failed cycles. Display
D1 shows P0. Press button B6. The printer prints a summary of the last five failed
cycles containing the following information for each cycle; Cycle temperature, Cycle type, the word “Extended” if the cycle was extended, Cycle count, Date and time at start of cycle, Date and time at end of cycle and the word “Error” followed by the error number. Display shows F0. With the door open press B7 to exit back to normal mode, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed).
P1 - Print last failed cycle This protected function prints a summary of the last failed cycle. Display D1
shows P1. Press the button B6. The printer prints a summary of the last failed cycle, containing the following information; Cycle temperature, Cycle type, the word “Extended” if the cycle was extended, Cycle count, Date and time at start of cycle, Date and time at end of cycle, the word “Error” followed by the error number and up to nineteen lines containing the time, temperature and pressure. Display shows P1. With the door open press B7 to exit back to special function F0, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed).
P2 - Recall factory calibration This protected function recalls and copies the factory calibration to the field
calibration area. Display D1 shows P2. Press button B6, display D1 shows P21. Press and hold B3 and B5 buttons together until a beep is heard. The factory calibration information overwrites the current field calibration. Display D1 shows P2. With the door open press B7 to exit back to special function F0, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed). Always recalibrate after using this function, see section 7.4.
P3 - Room T emperature Calibration This protected function performs a room temperature calibration. Ensure
autoclave is at room temperature before starting to eliminate calibration errors. Insert the required probes from the calibrated instruments and/or chart recorders (e.g. chamber temperature probe, chamber pressure probe) as detailed in section 7.4. Replace the autoclave cover during calibration. The door must be open and on the secondary catch for this protected function. Display D1 shows P3, press button B6, display D1 shows “t” followed by the current temperature. Display D2 shows the current temperature, display D3 shows the current pressure in kPa. Use buttons B2 and B3 to set the value in display D1 to the current temperature indicated on the calibrated instrument. Press button B6 until a beep is heard. Press button B1, display D1 shows “P” followed by the current pressure in mBars. Use buttons B2 and B3 to set the value in display D1 to the current pressure indicated on the calibrated instrument. (Note: The value in display D1 changes in steps of 2 mBars). Press button B6 until a beep is heard and the temperature and pressure displays change to the values entered in the previous steps. Press button
B7, display D1 shows P3. With the door open press B7 to exit back to special function F0, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed).
P4 - Low pressure calibration This protected function performs a low-pressure calibration. This function
can only be run on a vacuum autoclave. The door must be closed for this protected
function. Insert a suitable pressure probe from a calibrated instrument and/or chart recorder as detailed in section 7.4. Display D1 shows P4, press button B6. Display D1 shows P 200, display D3 shows the current pressure in kPa. (Note: D3 pressure is always in kPa). Wait until the pressure drops to about 200 mBars on the calibrated instrument then press button B5 (this closes the vacuum solenoid valve, but leaves the pump running). Use buttons B2 and B3 to set the value in display D1 to the current
(Continued on page 18)
ST-SM33l P17/51
TABLE 3 (continued) - PROTECTED FUNCTIONS
DISPLAY AND FUNCTION METHOD OF USE
pressure indicated on the calibrated instrument. (Note: The pump is running during this calibration and the vacuum solenoid valve can be toggled open and closed by pressing B5, this enables chamber pressure to be adjusted if required). Press button B6 until a beep is heard to save calibration data . Press button B7, wait for the pressure to return to within 5 mBars of atmospheric. Display D1 shows P4. With the door open press B7 to exit back to special function F0, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed).
P5 - Low and high temperature calibration Insert the required probes from the calibrated instruments and/
or chart recorders (e.g. chamber temperature probe, chamber pressure probe and band heater probe) as detailed in section 7.4. Replace the autoclave cover during
calibration. The door must be closed for this protected function. Display D1 shows P5, press button B6.
For Non-Vacuum - A 121°C Naked without Drying cycle is started. For Vacuum - A 121°C Naked with Drying cycle is started. (Note: For both the above cycles the sterilisation time is extended to give adequate
time to achieve an accurate calibration). Button B5 LED is illuminated and (if fitted) the printer records time and cycle data with
the words “Calibration Cycle” after the cycle type. Wait for the sterilisation plateau to be reached, then press button B1. Display D1 shows “t” followed by the current temperature. Use buttons B2 and B3 to set the value in display D1 to the current temperature indicated on the calibrated instrument. Continue to adjust the temperature in D1 until a stable condition is achieved. (Note: The sterilisation time can be extended if required by pressing B5). Press button B4 until a beep is heard. Display D1 shows the new temperature value. Press button B1, display D1 shows “P” followed by an estimated pressure value based on the current temperature. Use buttons B2 and B3 to set the value in display D1 to the current pressure indicated on the calibrated instrument (the value in display D1 changes in steps of 2). Press button B4 until beep is heard. Display D3 shows the new pressure value. Press and hold buttons B2 and B6 until beep is heard. The 121° LED is extinguished and the 134° LED is illuminated, the temperature rises towards 135.5°C, process display A2 returns to the seventh LED illuminated, display D1 shows P5. W ait for the temperature to approach 135.5°C, then press button B1. Display D1 shows t followed by the current temperature. Use buttons B2 and B3 buttons to set the value in display D1 to the current temperature indicated on the calibrated instrument. Continue to adjust the temperature in D1 until a stable condition is achieved. (Note: The sterilisation time can be extended if required by pressing B5). Press button B4 until beep is heard. Display D1 shows the new temperature value. Press button B1, display D1 shows “P” followed by an estimated pressure value based on the current temperature. Use buttons B2 and B3 to set the value in display D1 to the current pressure indicated on the calibrated instrument. Press button B4 until beep is heard. Display D3 shows new pressure value.
If it is required to test the safety valve press and hold buttons B2 and B7 until beep is heard, the 121° LED illuminates, the temperature rises towards 148°C (approx.), the process display A2 returns to the seventh LED illuminated, display D1 shows P5. Wait for the pressure to reach about 385kPa, the safety relief valve starts to release the pressure from the chamber, note valve limits are 2.72 - 3.13Bar (Gauge.)
Press and hold buttons B5 and B7, the water discharge valve opens, process display
A2 advances to the fourteenth LED. Non-Vacuum - The steriliser completes the cycle normally, display D1 shows P5 all
other displays are blank.
(Continued on page 19)
P18/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 3 (CONTINUED) - PROTECTED FUNCTIONS
DISPLAY AND FUNCTION METHOD OF USE
Vacuum - The steriliser continues onto the drying phase, wait for band heater temperature
to reach about 158°C on the calibrated instrument, then press button B1. Display D1 shows “t” followed by current band heater temperature. Use buttons B2 and B3 to set the value in display D1 to the current temperature indicated on the calibrated instrument. Continue to adjust the temperature in D1 until a stable condition is achieved. Press button B4, a beep is heard. Press and hold button B5 and B7, the autoclave completes the cycle immediately . Display D1 shows P5, all other displays are blank. With the door open press B7 to exit back to special function F0, or B5 to cycle through and gain access to other protected functions (by pressing B6 when the correct P number is displayed).
P6 - Cycle Enable/Disable This protected function enables or disables the sterilization cycles. Display D1 shows
P6, press button B6. The programme and temperature LEDs display the default
programme. All LEDs on the progress indicator A2 are illuminated to indicate this programme is enabled. Button B4 can now be used to toggle all LEDs on the progress indicator A2 “on” or “off” showing that the indicated programme is enabled or disabled respectively.
Press button B2, the programme and temperature LEDs display the next programme. Use B4 to toggle all LEDs in A2 “on” or “off” to enable or disable the indicated programme.
Repeat pressing B2 and enable or disable each programme as required (when indicated) until all programmes have been set as required. (Note: When B2 is pressed and no indicator lights are “on” display D2 will show either , to indicate the disinfect cycle, or
, to indicate the ‘Hollow Naked’ cycle. These can also be enabled or disabled by pressing button B4. Autoclaves without button B9 will still display the option to enable or disable the ‘Hollow Naked’ cycle even though the cycle cannot be run).
Note: Pressing button B3 will enable all programmes, it is not possible to disable all programmes. When button B5 indicator illuminates this indicates an extended cycle.
Press button B6 until a beep is heard, all the programme enabling or disabling settings made above are stored in memory , display D1 shows P6 all other displays or indicators are “off”. To exit press B7 (with the door open) or B5 to cycle through and gain access to another protected functions. When the correct P number is displayed by pressing B6.
P7 - I/O T est This test verifies all the inputs and outputs to the Microcontroller. When familiar with the
use of this protected function, see T able 5 for a quick reference chart which details how to use P7 to check individual components, rather than progressing through each one, as detailed below. Before using Table 5 ensure you know how to activate and leave this protected function. The water reservoir should be empty (if not, the 8th LED of the progress indicator A2 will be “on” throughout this test and it will not be possible to test the reservoir sensor). Display D1 shows P7, press button B6. Display D1 shows the raw ADC value for the first pressure transducer, display D2 shows P1, display D3 shows 1, the fourth and sixth LEDs on the progress indicator A2 are illuminated if the door is open. Press button B1, display D1 shows the raw ADC value for the second pressure transducer , display D2 shows P2. Press button B1, display D1 shows the raw ADC value for the first chamber temperature transducer, display D2 shows t1. Press button B1, display D1 shows the raw ADC value for the second chamber temperature transducer, display D2 shows t2. Press button B1, display D1 shows the raw ADC value for the Band Heater temperature transducer, display D2 shows t3. Press button B1, display D1 shows the raw ADC value for the first pressure transducer, display D2 shows P1, display D3 shows 1.
Press button B4, B4 LED illuminates, and the water discharge valve opens. Press button B4, B4 LED goes “off” and the water discharge valve closes.
(Continued on page 20)
ST-SM33l P19/51
TABLE 3 (CONTINUED) - PROTECTED FUNCTIONS
DISPLAY AND FUNCTION METHOD OF USE
P7 - I/O T est (cont’d) Press button B2, display D3 shows 2. Press button B4, B4 LED illuminates, and the
water fill valve opens. Press button B4, B4 LED goes “off” and the water fill valve closes. Press button B2, display D3 shows 3. Press button B4, B4 LED illuminates, and the air
admit valve opens. Press button B4, B4 LED goes “off” and the air admit valve closes. (Only on vacuum autoclaves).
Press button B2, display D3 shows 4. Press button B4, B4 LED illuminates, and the fan runs. Press button B4, B4 LED goes “off” and the fan stops.
Press button B2, display D3 shows 5. Press button B4, B4 LED illuminates, and the steam bleed valve opens. Press button B4, B4 LED goes “off” and the steam bleed valve closes.
Press button B2, display D3 shows 6. Press button B4, B4 LED illuminates, and the vacuum inlet valve opens. Press button B4, B4 LED goes “off” and the vacuum inlet valve closes. (Only on vacuum autoclaves).
Press button B2, display D3 shows 7. Press button B4, B4 LED illuminates, and the door open solenoid operates. Press button B4, B4 LED goes “off” and the door open solenoid operates.
Press button B2, display D3 shows 8. Press button B4, B4 LED illuminates, and the door close solenoid operates. Press button B4, B4 LED goes “off” and the door close solenoid operates.
Press button B2, display D3 shows 9. Press button B4, B4 LED illuminates, and the mains relay is “on”. Press button B4, B4 LED goes “off” and the mains relay is “off”. Press button B4, to turn the mains relay “on” and leave it “on” to enable other tests (e.g. heaters) that follow.
Press button B2, display D3 shows 10. Press button B4, B4 LED illuminates, and the water pump operates. Press button B4, B4 LED goes “off” and the water pump stops. (Only on non-vacuum autoclaves).
Press button B2, display D3 shows 11*. Press button B4, B4 LED illuminates, and the vacuum pump operates. Press button B4, B4 LED goes “off” and the vacuum pump stops. (Only on vacuum autoclaves).
Press button B2, display D3 shows 12*. Press button B4, B4 LED illuminates, and the band heater is “on”. Press button B4, B4 LED goes “off” and the band heater is “off”. (Only on vacuum autoclaves).
Press button B2, display D3 shows 13*. Press button B4, B4 LED illuminates, and the immersion heater is “on”. Press button B4, B4 LED goes “off” and the immersion heater is “off”.
Press button B2, display D3 shows 14. Press button B4, B4 LED illuminates, and the beeper is “on”. Press button B4, B4 LED goes “off” and the beeper is “off”.
Press button B2, display D3 shows 1. Press button B4, B4 LED illuminates, and the water discharge valve opens. Press button B3, B4 LED goes “off” and the water discharge valve closes.
Short the chamber minimum water level sensor to the door, the second LED on progress indicator A2 illuminates.
(Continued on page 21)
* Ensure mains relay was left “on” (when D3 shows 9 above) to perform this test.
P20/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 3 (CONTINUED) - PROTECTED FUNCTIONS (See section 7.2)
DISPLAY AND FUNCTION METHOD OF USE
P7 - I/O T est (cont’d) Short the chamber maximum water level sensor to the door, the third LED on progress
indicator A2 illuminates. Operate the door closed microswitch, the fourth LED on progress indicator A2 is
extinguished and the fifth LED on progress indicator A2 is illuminated. Operate the door locked microswitch, the sixth LED on progress indicator A2 is
extinguished and the seventh LED on progress indicator A2 is illuminated. Short the two reservoir water level sensors together, the eighth LED on progress indicator
A2 illuminates. Press button B7, display D1 shows P7, all other displays are blank. T o exit press B7 (with the door open) or B5 to cycle through and gain access to another
protected functions. When the correct P number is displayed by pressing B6.
P8 - Service completion When service is carried out in response to a service request from the autoclave, it is
necessary to reset the service request flag in the EEPROM. This occurs automatically if the unit is re-calibrated. It can also be done through the protected function P8. Display D1 shows P8. Press button B6. The autoclave beeps. The service request is set for 1 1 months later and will print ‘Service and Certification due’. Display D1 shows P8. T o exit press B7 (with the door open) or B5 to cycle through and gain access to another protected functions. When the correct P number is displayed by pressing B6.
P9 - Enable/disable This function changes the configuration file to disable, or to enable, the facility which
adjustable drying and sterilization times
allows the user to alter the preset sterilization and drying times when selecting a cycle. Display D1 shows P9. Press button B6. D2 displays either if the function is disabled or if it is enabled. Press button B2 to enable it, or press button B3 to disable it. Press
button B6 to confirm the setting. The autoclave beeps to acknowledge. Display D1 shows P9. T o exit press B7 (with the door open) or B5 to cycle through and gain access to another protected functions. When the correct P number is displayed by pressing B6.
ST-SM33l P21/51
TABLE 4 - FAULT DIAGNOSIS AND ERROR NUMBER DETAILS
CLEARING ERROR DISPLAY AND
REST ARTING A CYCLE
WARNING
If an error display appears during a cycle, do not attempt to open the chamber door until the unit has cooled (for safety below 50°C) and internal pressure has fallen sufficiently to release the door, see ‘3 Check pressure’. Loads whose cycles are aborted by one of these error displays should be treated as non­sterile, they should be re-wrapped if wrapped and then sterilized by running the cycle again. Also note the Caution after section 16.1 in the ‘Instructions’ concerning the contamination of water in the reservoir.
1 Record error number : For future reference
record the error number before clearing the display. In the event of mains power failure, power must be restored to the autoclave to achieve this.
2 Clear error and reset autoclave : To clear the error display and reset the autoclave wait until any audible discharge of water and steam has stopped and switch OFF the autoclave with the mains switch (item 9, Fig. 2). Wait 10 seconds, press-and-hold-in the ‘Start button’ ( B6 Fig.1) whilst switching back ON with the mains switch (item 9, Fig. 2), immediately the normal display appears release the ‘Start button’ (B6 Fig.1).
3 Check pressure : Before attempting to open the door note the pressure reading, this should be below 104kPa before the door is opened.
4 Open the door : Note A4 will be flashing, open the door by pressing B7. If the door fails to open see NOTE below, there may still be water in the autoclave.
NOTE: Door opening problems. There may be up to a litre of water (which could be HOT) still in the chamber, be prepared to collect this in a suitable container. A built-in safety feature inhibits door opening unless the following procedure is adopted. Switch OFF at mains switch (item 9, Fig. 2). Press-and-hold-in the ‘Door open button’ (B7 Fig.1) whilst switching back ON with the mains switch (item 9, Fig. 2) wait two seconds and then before releasing button B7 (which will open the door) hold a container in place under the door to collect any (HOT) water.
5 Check and prepare the aborted load : Loads must be considered non-sterile if an error occurs and should be resterilized. All aborted wrapped loads should be dried, placed in new pouches and reprocessed.
6 Check chamber before starting another cycle : The chamber should be allowed to dry and cool before
starting another cycle. To start another cycle select the cycle and start the autoclave in the normal manner.
7 If the error persists or an error number occurs that is listed at the end of this table then call the Eschmann After Sales Service Department (see details inside front cover) or contact your approved service engineer. Do not attempt to solve an error using tools or tampering with the autoclave.
FAULT DIAGNOSIS
Note: Consult T able 3, the protected functions table, section P7. This function will enable each of the main components within the autoclave to be tested by pressing the specified buttons on the control panel thus verifying the inputs and outputs to the microcontroller. This procedure will help eliminate working components from those listed below as the possible cause of unit faults, during error correction and fault diagnosis.
1. No display or LEDs illuminated when power switched on.
a) Mains supply failure. a) Check mains supply is available and switched on. Check mains plug
and plug fuse. Check supply cable for loose connections or breaks.
b) No 24V on power board. b) Either of the mains fuses (item 8, Fig.2) failed, see 8.2 to replace fuses.
Faulty power switch (item 9, Fig.2), see 8.21 to replace switch. Transformer (item 27, Fig.3) failed, see 8.6 to replace transformer . Wiring fault (mains to fuses, fuses to switch, switch to transformer, transformer to power board).
c) No power on display board. c) Faulty power board (item 26, Fig.3) or faulty wiring or faulty connection
to display board. See 8.7 to replace power board.
d) Faulty display board. d) Replace display board see 8.27.
(continued on page 23)
P22/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
FAULT DIAGNOSIS (Continued)
2. A1 illuminated but reservoir full.
a) Reservoir level sensor fault. a) Reservoir level sensor (item 1, Fig.4) faulty or loose (see section 8.30)
or bad connection to power board.
b) Reservoir contaminated. b) Drain, clean and dry reservoir, then refill.
3. A1 Flashing at cycle start - Cycle start requested with low water .
a) Button B6 pressed with A1 'on'. a) Low reservoir water level, fill as detailed in the 'Instructions for Use'. If
indicator A1 is still illuminated when reservoir is full check function of reservoir water level sensor (item 1 Fig.4) for broken wire etc. If indicator A1 was not illuminated when B6 pressed and reservoir empty check indicator A1 using the LED test detailed in 7.5.
4. Cycle selection not possible.
a) Water level in reservoir low. a) Check A1 Fig.1, if 'on' fill reservoir. b) Door not closed. b) Close door, check A4 Fig.1 illuminated (if not see 'c', 'd' and 'e' below). c) Door microswitch fault. c) Faulty door microswitch (items 18 and 19, Fig.3) or faulty wiring to
them. See 8.8 to replace microswitches.
d) Door close solenoid fault. d) Door close solenoid (item 8 Fig.3) faulty, see 8.9 to replace solenoid.
24V power supply to solenoid failed or connections faulty .
e) Door adjustment faulty. e) Door adjustment stopping primary lock engaging.
5. Safety valve leaking.
a) Dirt on valve seat. a) With low pressure in chamber carefully operate safety valve (item 24
Fig.3) by hand (Warning: Beware of risk of scalds from escaping
steam). If leakage persists, fit new safety valve, see 8.19.
6. Safety valve operating at low pressure during cycle.
a) Safety valve fault. a) Check safety valve and replace if required see 8.19.
7. Safety valve operating at high pressure during cycle.
a) Calibration fault. a) Recalibrate see 7.4. b) Chamber temperature sensor fault. b) Fit new temperature sensor see 8.11. c) Power board fault. c) Fit new power board see 8.7.
8. Unusual display when switching on power.
a) Power board failed to reset a) Switch-off power, wait for 10 seconds and switch back on again. b) Power board fault. b) Fit new power board (see 8.7) or reprogramme via P.C.
9. Error 01 displayed - Power failure during cycle.
a) Mains supply failure. a) For display to show error 01 power has already been restored
successfully. If error 01 continues to occur check for loose mains connections and wiring or over tightened fuse cover item 8 Fig.2.
10. Error 02 displayed - Door opened during cycle.
a) Door closed microswitch fault. a) Faulty door microswitch (item 19, Fig.3) or wiring to it. See 8.8 to replace
microswitch.
11. Error 03 displayed - T ime for chamber to fill with water too long (time out).
a) Water fill solenoid valve fault. a) Water fill solenoid valve (item 11, Fig.3) faulty or connections or supply
to it faulty or failed. See 8.4 to replace valve.
(continued on page 24)
ST-SM33l P23/51
FAULT DIAGNOSIS (Continued)
b) Water pipe connection fault. b) Pipes to water fill valve from reservoir disconnected, blocked or kinked
(pipes pass via water pump on non-vacuum unit).
c) Water pump fault (non-vacuum only) c) Water pump (item 2, Fig. 4) faulty, or pipe connections to it faulty, or
supply to it faulty or failed. See 8.31 to replace pump. Also check on older units that pump is in the new position close to reservoir, if not, or
in doubt, contact Eschmann Equipment. d) Low water in reservoir. d) Reservoir water level sensor error see fault 2 above. e) Chamber water level sensor error. e) Water level sensor failure (items 14 and 17, Fig.3) or wiring to them
faulty (see 8.23).
12. Error 04 displayed - Low water level during cycle (water / steam loss).
a) Insufficient 'overfill'. a) Chamber fill below that required due to slow fill rate, check faults and
solutions in 10 above. b) Chamber integrity fault. b) Chamber has a slow leak (below that to trigger leak test) from any of
the following areas. Door seal (check and or replace see 8.13), solenoid
valves (check and or replace see 8.4), safety valve (see faults 4 and 5
above), test probe entry ports (check and or replace see 8.32), door
face damaged (check and or replace see 8.33), water level sensors in
door (check and or replace see 8.23). c) Large or obscure load c) Try a cycle with a smaller load and ensure load can drain, or empty. d) Chamber too hot at start. d) Allow unit to cool and try another cycle. e) Bench top not level e) Ensure autoclave is level to enable correct water fill levels.
13. Error 06 displayed - Chamber temperature too low during sterilization.
a) Power to heater fault. a) Low mains voltage check supply. Heater (item 4 Fig.4) gone open-
circuit check and or fit new heater see 8.12. Solid state relay or
mechanical relay on power board (item 26 Fig.3) failed, fit new power
board see 8.7. Thermal fuse open-circuit, replace fuse see 8.26. b) Chamber temperature sensor fault. b) Chamber temperature sensor (item 23 Fig.3) failed or connections to it
faulty (see 8.11 to fit new sensor) or in need of calibration see 7.4.
Control signal path corruption, check wiring, connections, etc.
14. Error 07 displayed - Chamber temperature too high during sterilization.
a) Chamber temperature sensor fault. a) Chamber temperature sensor (item 23 Fig.3) failed or connections to it
faulty (see 8.1 1 to fit new sensor). Control signal path corruption, check
wiring, connections, etc. b) Power board fault. b) Solid state relay or mechanical relay on power board (item 26 Fig.3)
failed, fit new power board see 8.7.
15. Error 10 displayed - Time to create chamber pressure drop too long (time out).
a) Chamber integrity fault. a) See 7b. b) Vacuum pump fault. b) Vacuum pump (item 9 Fig.3) failed or electrical power or pipe
connections to it faulty. See section 8.14 to replace pump. If pump
running check tubing to vacuum solenoid valve not leaking and pumps
diaphragm and valves, see 8.35. Check pipe from chamber to pressure
transducer on power board connected and not leaking. Pressure
transducer fault change power board see 8.7. c) Vacuum inlet valve fault. c) Vacuum inlet solenoid valve (item 3, Fig.3) faulty or connections or
supply to it faulty or failed. See 8.4 to replace valve. d) Wet, warm chamber d) Chamber conditions at start of cycle too wet and or warm. Restart cycle
with dryer colder conditions.
(continued on page 25)
P24/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
FAULT DIAGNOSIS (Continued)
16.Error 13 displayed - Steam quality error, temperature pressure relationship outside set limits.
a) Air in chamber a) Check chamber integrity see 12b. Obscure load (air entrapped) adjust
load and try cycle again. Temperature and or pressure sensors faulty or in need of re-calibration, see 8.1 1 and 8.7 to change power board.
b) Steam bleed valve blocked b) Check and clean / replace as required see section 8.4.
17.Error 14 displayed - Time to achieve vacuum (before drying) outside set limits (time out).
a) Chamber integrity fault. a) See 12b. b) Vacuum pump fault. b) See 15b. c) V acuum inlet valve fault. c) See 15c.
18.Error 15 displayed - Chamber temperature sensors differing by more than preset limit.
a) Chamber temperature sensor fault. a) See 13b.
19.Error 16 displayed - Real-time clock error
a) Real-time clock error/stopped. a) Restart clock by resetting time using special function F8. If this is not
successful replace display board see 8.27.
20.Error 17 displayed - Band heater not working
a) Band heater fault. a) Check band heater not open circuit and power is available from SSR
etc. check all connections. Check mains voltage . If required replace band heater (see 8.22) or power board (see 8.7).
b) Reset thermostat fault. b) Reset thermostat (item 29 Fig.3) and repeat cycle, if thermostat trips
again the band heater temperature sensor (item 28 Fig.3) could be at fault (ensure sensor is not dirty or corroded) or in need of calibration (see 7.4). If fault continues replace band heater thermostat (item 29 Fig.3) see 8.34.
c) Band heater temp. sensor fault. c) Ensure sensor is not dirty or corroded, check calibration (see 7.4) fit
new sensor if required see 8.20.
d) Heat loss (covers not fitted). d) Refit covers and run cycle, covers should always be in place.
21.Error 18 displayed - System configuration invalid.
a) EEPROM corrupted. a) Reconfigure via P .C. or replace Display PCA see 8.27.
22.Error 19 displayed - Calibration invalid.
a) EEPROM corrupted. a) Recalibrate see section 7.4, or transfer factory calibration using
‘Protected function’ P2 then recalibrate as section 7.4.
23.Error 23 displayed - Wrapped load leak test, pressure rise above set limit.
a) Chamber integrity fault. a) See 12b.
24.Error 24 displayed - Wrapped load leak test, pressure fall outside set limit.
a) Invalid test. a) Repeat test. b) Pressure transducer fault b) Replace Power PCA c) Power PCA fault c) Replace Power PCA
25.Error 25 displayed - Time to create initial fill vacuum above set limit.
a) Chamber integrity fault. a) See 12b. b) Vacuum pump fault. b) See 15b. c) V acuum inlet valve fault. c) See 15c. d) Wet, warm chamber d) See 15d.
(continued on page 26)
ST-SM33l P25/51
FAULT DIAGNOSIS (Continued)
26.Error 26 displayed - Chamber water level sensors, 'maximum' before 'minimum'.
a) Sensor wiring fault. a) Sensors (item 14 and 17, Fig.3) wired wrong way round. Minimum fill
sensor (item 14 Fig.3) wire dislodged or broken. Maximum fill sensor
(item 17 Fig.3) short after cycle started.
27.Error 27 displayed - Wrapped load purge time outside set limit.
a) Power to heater fault. a) See 13a. b) Chamber integrity fault. b) See 12b. c) Pressure sensor fault. c) Check pipe from chamber to pressure transducer on power board
connected and not leaking. Pressure transducer fault change power
board see 8.7. Check pressure calibration see 7.4.
28.Error 28 displayed - Unwrapped purge time 1, outside set limit.
a) Power to heater fault. a) See 13a. b) Chamber temperature sensor fault. b) See 13b.
29.Error 29 displayed - Unwrapped purge time 2, outside set limit.
a) Power to heater fault. a) See 13a. b) Chamber temperature sensor fault. b) See 13b.
30.Error 30 displayed - Time to heat to set point outside set limit.
a) Power to heater fault. a) See 13a. b) Chamber temperature sensor fault. b) See 13b. c) Steam bleed valve fault. c) Check steam bleed valve (item 2 Fig.3) wiring and that it opens and
closes correctly , replace if required see 8.4.
31.Error 31 displayed - Chamber maximum temperature outside set limit.
a) Power board fault. a) See 14b. b) Chamber temperature sensor fault. b) See 13b.
32.Error 32 displayed - Chamber maximum pressure outside set limit.
a) Chamber temperature sensor fault. a) See 13b. b) Pressure sensor fault. b) See 27c. c) Power board fault. c) See 14b.
33.Error 33 displayed - Water discharge time outside set limit.
a) Steam bleed valve fault. a) Stuck open see 30c. b) Water discharge valve fault. b) Water discharge valve (item 2, Fig.3) not opening, check wiring and
that it opens and closes correctly , replace if required see 8.4. c) Chamber water level sensor fault. c) Check sensors (item 14 and 17, Fig.3) not shorted to earth. Also in
extreme cases condensation on sensor can give a false reading., repeat
cycle and check for repeated error. d) Gauze filter blocked. d) Gauze filter (item 6, Fig.2) blocked by debris, clean around filter then
remove it and clean filter or replace with a new one.
34.Error 34 displayed - Steam discharge time outside set limit.
a) Steam bleed valve fault. a) Valve failed shut, see 30c. b) Pressure sensor fault. b) See 27c.
(continued on page 27)
P26/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
FAULT DIAGNOSIS (Continued)
35.Error 35 displayed - Non vacuum door open (for drying) time outside set limit.
a) Door open solenoid fault. a) Check door open solenoid (item 6, Fig.3) wiring and function. b) Door held closed by vacuum. b) Check door seal clean, check all valve functions. c) Door closed microswitch fault. c) Check door closed microswitch (item 19, Fig.3) function and wiring,
replace if required see 8.8.
36.Error 36 displayed - Door unlocked during cycle.
a) Door locked microswitch fault. a) Check door locked microswitch (item 18, Fig.3) function and wiring,
replace if required see 8.8. b) Door solenoid adjustment fault b) Check door solenoid adjustment/fitting is correct (see section 8.9). c) Door lock magnet fault c) Check door lock magnet adjustment/fitting is correct (see section 8.10).
37.Error 37 displayed - Band heater above set limit.
a) Band heater temp. sensor fault. a) See 20c. b) Power board fault. b) See 14b.
38.Error 38 displayed - Chamber temperature above set limit at cycle start.
a) Chamber temperature sensor fault. a) See 13b. b) Ambient temperature extreme. b) Only run within design parameters.
39.Error 39 displayed - Chamber temperature below set limit at cycle start.
a) Chamber temperature sensor fault. a) See 13b. b) Ambient temperature extreme. b) Only run within design parameters.
40.Error 40 displayed - Chamber pressure above set limit at cycle start.
a) Chamber pressure sensor fault. a) See 27c. b) Ambient pressure extreme. b) Only run within design parameters.
41.Error 41 displayed - Chamber pressure below set limit at cycle start.
a) Chamber pressure sensor fault. a) See 27c. b) Ambient pressure extreme. b) Only run within design parameters.
42.Error 42 displayed - Band heater temperature above set limit at cycle start.
a) Band heater temperature sensor fault. a) See 20c. b) Ambient pressure extreme. b) Only run within design parameters.
43.Error 43 displayed - Band heater temperature below set limit at cycle start.
a) Band heater temp. sensor fault. a) See 20c. b) Ambient pressure extreme. b) Only run within design parameters. c) Calibration error c) Recalibrate, see section 7.4.
44.Error 44 displayed - Chamber water level high at cycle start.
a) Maximum water level sensor fault. a) Water level sensor failure (item 17, Fig.3) or wiring to it.
45.Error 45 displayed - Chamber water level low at cycle start.
a) Minimum water level sensor fault. a) Water level sensor failure (item 14, Fig.3) or wiring to it.
(continued on page 28)
ST-SM33l P27/51
FAULT DIAGNOSIS (Continued)
46.Error 46 displayed - Door open button pressed 10 times and door failed to open.
a) Chamber pressure too high. a) Wait for pressure to drop and try again. (Door open solenoid
programmed not operate with chamber pressure above 115kPa and 1 15°C).
b) Chamber pressure too low. b) V acuum in chamber holding door closed, check all valve functions. Also
check door seal and face clean. c) Calibration fault. c) Recalibrate unit see 7.4. d) Drain tube tap trapped d) Release drain tube tap from between door and panel.
47.Error 47 displayed - Steam bleed valve failure.
a) Steam bleed valve fault. a) Valve failed to open during purge, see 30c. b) Chamber temperature sensor fault. b) See 13b.
48. Error 49 displayed - Fan failed to cool chamber during vacuum pulse
a) Fan not operating. a) Check fan connections and function etc. b) Grill blocked limiting air flow. b) Check air flow into chamber via grill etc.
49. Error 50 displayed - Band heater sensor reading too high
a) Calibration error a) Recalibrate autoclave see section 7.4. b) Band heater sensor fault b) Replace sensor, see section 8.20 and recalibrate.
50. Error A0 displayed - Test ‘Abort Cycle’ function selected from PC via ‘Support Application’.
This function aborts the cycle running (error 100 recorded for cycle, the part of the cycle under investigation has been passed allowing the remaining part of the cycle to be aborted to save time. (Note: Loads used during test should be considered non-sterile).
ErrAO
on autoclave display). This is useful if
P28/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 5 - PROTECTED FUNCTION P7, QUICK REFERENCE TABLE
NOTES:
1. Use button B4 to switch the activated component ‘on’ or ‘off’. State is indicated (‘on’ or ‘off’) by LED adjacent to button B4 (on = on / off = of f).
2. If testing 230V items the mains relay should be ‘on’, or left ‘on’.
DISPLAYED ADC VALUES (scroll with button B1)
Display D2 ADC (analogue to digital) value displayed in D1
P1 First pressure transducer P2 Second pressure transducer (NB. Not used) t1 First chamber temperature transducer t2 Second chamber temperature transducer t3 Band heater temperature transducer
COMPONENT ACTIVATION (scroll up with button B2, down with button B3)
Display D3 Component selected ‘on’ or ‘off’ by button B4
1 Water discharge valve 2 Water fill valve 3 Air admit valve 4 Fan 5 Steam bleed valve 6 V acuum inlet valve 7 Door open solenoid 8 Door close solenoid 9 Mains relay 10 Water pump 11 V acuum pump (see note 2 above) 12 Band heater (see note 2 above) 13 Chamber immersion heater (see note 2 above) 14 Audible beeper
SENSOR & MICROSWITCH TEST
Action to activate component Progress display LED Illuminated
Short chamber minimum water level sensor to door Second Short chamber maximum water level sensor to door Third Operate door closed microswitch Fifth (Fourth extinguished) Operate door locked microswitch Seventh (Sixth extinguished) Short reservoir water level sensors together Eighth
ST-SM33l P29/51
8 Part removal and replacement
8.3 Cover removal and replacement
WARNING
Switch-off and disconnect mains power supply before removing the autoclave cover , or doing maintenance procedures. During certain procedures mains voltage may have to be present with the cover removed and extreme care should be taken to avoid contact with mains voltage.
Check that chamber is at atmospheric pressure before opening the door (see 8.1.3). When the door is opened beware of possible very hot water or steam escaping from the chamber.
8.1 General
8.1.1 All the major parts of the autoclave can be easily removed and replaced following removal of the autoclave cover, as detailed in section 8.3. It is important that the autoclave is recalibrated after certain part replacement, where this is required it is called for at the end of the relevant part replacement procedure. DO NOT FORGET TO REPLACE COVERS BEFORE CALIBRATION. After all maintenance procedures always run at least one complete cycle to check all functions are working correctly .
8.1.2 The vacuum pump can be refurbished if performance deteriorates as an alternative to complete replacement, see section 8.35.
8.1.3 Before any maintenance procedure especially in the case of an electrical fault, always check the chamber pressure is close to, or at ambient, before attempting to open the door*. In the case of no electrical power the chamber pressure can be checked on the safety pressure gauge (see 21, Fig.3) behind the door cover. Note that the door cover can be released when the door is locked shut as detailed in section 8.3.1. * Note: The door cannot be opened on later models fitted with the pressure door lock until the pressure has fallen below 0.2Bar (Gauge).
8.1.4 After part replacement always ensure that all cable ties removed are replaced correctly after reassembly .
8.2 Fuse renewal
Fuses are fitted at the front of the autoclave (item 8, Fig. 2). For fuse ratings refer to ‘TECHNICAL DAT A’. To extract a fuse switch ‘off’ the unit and disconnect from the mains power supply , insert a screwdriver or small coin in the slot of the fuse holder and twist it counter-clockwise. After inspecting or renewing a fuse, reverse the above procedure to re-secure fuse holder (DO NOT OVER TIGHTEN). See section 8.28 if the fuse holder assembly is to be replaced.
8.3.1 Front cover removal (see warning above section 8.1). To remove the front cover place the end of a flat bladed
screwdriver into the slot behind the plate marked ‘B’ in Fig.2 and depress the sprung loaded door interlock pin (see 8.3.4 for orientation of this pin when re-engaging it). Plate ‘B’ can only be removed using a TX20 Torx pin tamperproof screwdriver. In the event that the door is locked shut, access to plate ‘B’ in Fig.2 can be gained by opening the printer door (item 15, Fig.2). When the interlock pin is depressed (down) the door cover can be separated from the main autoclave door and door beam. Release the drain pipe and tap (item 14, Fig. 2) from the base of the door and then remove the four front cover retaining screws marked ‘A’ in Fig.2. Carefully release the door cover and front cover together from the unit (pulling the right hand side out first) until access to the inside of the autoclave has been gained. Release the ribbon cable attached to the display board and the air inlet pipe from the rear of the air filter (vacuum units only). Continue to remove the front covers feeding the drain pipe and tap through the hole marked ‘C’ in Fig.1. If it is required to operate the autoclave with the front cover removed the ribbon cable can be extended by releasing the ribbon cable clamp.
8.3.2 Top cover removal (see warning above section 8.1). On early models (no side fixing screws) Remove the front
cover as detailed in 8.3.1. Slightly raise the front of the top cover and pull it forwards to release the cover from the two retaining spigots on the top of the reservoir (item 30, Fig.3). Lift cover clear of reservoir or continue to slide cover forwards until clear. On later models (with side fixing screws) On later models the front cover does not need to be removed. Using a TX25 Torx pin tamperproof screwdriver remove the screw on each side. The top cover can then be slide backwards and then up until clear.
8.3.3 Top cover replacement.
On early models (no side fixing screws) Slide the top cover into place tilted slightly down at the back until slightly forward of the reservoir. Push the cover down and back to enable the clips inside the cover to engage on the reservoir spigots (item 30, Fig.3). Finally push the cover down at the front to engage the lower retaining clips onto the base of the autoclave. Ensure all four clips are correctly located before replacing the front covers. On later models (with side fixing screws) Slide the top cover back into place and replace the two side fixing screws (TX25 Torx pin tamperproof screwdriver) to secure it.
8.3.4 Front cover replacement.
Ensure the safety valve shield (item 27a, Fig. 9 sheet 2) is positioned correctly as shown in Fig. 9 sheet 2. Position the front cover assembly over the open door in front of the
P30/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
autoclave and pass the reservoir drain pipe and tap through hole ‘C’ Fig.2. Reconnect the ribbon cable to the display board and on vacuum units, the air inlet pipe to the rear of the inlet filter. Replace the ribbon cable clamp if removed at 8.3.1 above. Position the cover assembly back into place (left hand side first), replace the cover retaining screws (with the correct shouldered washers in place where required, see Parts List 3) and return the drain pipe and tap into its retaining clips. Check the door spacer (item 34, Fig. 9) is in place before finally depressing the sprung loaded door interlock pin and locating it into the slot at the bottom of the door beam. Note that the top of the door interlock pin (item 2, Fig. 9) has a machined flat that should be aligned to face the door cover (item 1, Fig. 9). Some early units did not have this flat and orientation is not critical. Replace plate ‘B’ in Fig.2 using the correct tamper proof screws.
8.4 Chamber solenoid valves
8.4.1 If the valve assemblies are to be dismantled for cleaning and inspection do not misplace the small internal spring around the sealing plunger or the seal. If only the valve coil has failed (e.g. short-circuiting) it can be replaced.
8.4.2 To clean, replace or repair valves first remove the electrical connection from the valve by removing the crimps from the coil. If replacing the valve, remove the tubing connected to the valve body (do not forget to clamp the
pipe or drain the reservoir when working on the water fill valve) by cutting through the cable tie (take care not to
cut into the pipe).
8.4.3 To replace the coil release the nut and washer on the top of the coil and remove the coil from the valve body . Replace the coil ensuring that the replacement has the same electrical rating (24V, 14W), replace the washers and nut (do not overtighten as this can damage the coil) and reattach the electrical connection. Test the valve function using protected function P7.
8.4.4 To clean the valve remove the coil as detailed above and then release the exposed nut on the top of the valve body. Remove the valve stem and internal parts, clean them and the inside of the valve seating. Reassemble taking care to return parts in the correct sequence and do not overtighten the coil retaining nut. Finally reattach the electrical connection (release the clamp on fill pipe or fill reservoir if carried out at 8.4.2) and test the valve function using protected function P7.
8.4.5 To replace the valve completely release the two retaining screws and shakeproof washers from the base of the valve body and discard. Reattach the new valve (ensure coil has the correct electrical rating 24V , 14W and the small ‘O’ ring part number 391075 is in place) reconnect the electrical connection and pipe to the valve body, securing it with a cable tie (using Hellerine oil part number 670170 sparingly). T est the valve function using protected function P7.
8.5 Reservoir
8.5.1 Removal. To remove the reservoir assembly
(item 5, Fig.3) for cleaning or replacement drain the reservoir as detailed in the 'Instructions for use'. Cut off the cable ties (taking care not to damage the pipe) from the silicone discharge and vacuum exhaust pipes where they connect to the top of the condensing coil (item 31, Fig.3). Loosen the jubilee clip from the reservoir filling pipe (item 32, Fig.3) and remove both the discharge and exhaust pipe from the reservoir. Remove the reservoir retaining screws and washers from the left and right hand side of the reservoir. Partly withdraw the reservoir with the cap and condensing coil in place. Cut off the cable tie (taking care not to damage the pipe) from the silicone chamber fill pipe where it connects to the fitting at the base of the reservoir and remove the chamber fill pipe. Disconnect the reservoir level sensors electrical connections and then remove the reservoir completely from the autoclave.
8.5.2 Cleaning. To clean the reservoir first release the cap and withdraw the condensing coil carefully from the reservoir. W ash the condensing coil and reservoir using a detergent solution (‘Teepol’ L, is recommended, used as detailed by the manufacturer) using a suitable bottle brush to clean inside the reservoir. Rinse all components thoroughly with tap water to remove any residual detergent. Then rinse all components thoroughly with distilled or deionized water (or water treated by reverse osmosis) to remove any residual tap water. Allow components to dry and reassemble the condensing coil back into the reservoir ensuring that the silicone pipe on the end of the condensing coil lies along the base of the reservoir.
8.5.3 Replacement. If replacing the reservoir completely remove the old condensing coil and clean it before reassembly as detailed above. Also remove the two cover retaining spigots (screw and shouldered washer, item 30, Fig.3) the water level sensor and the fill pipe spigot and large washer from the old reservoir. Reassemble the reservoir, condensing coil, fill pipe spigot with large washer (use PTFE tape part number 301600 on spigot thread) the water level sensor and cover retaining spigots. When replacing the top electrical connection for the water level sensor note the following order for the parts - PTFE washer against the reservoir with a steel washer on top, then the electrical connection a steel washer and finally the screw with PTFE tape around thread.
Reconnect both the electrical connections and then the fill pipe (using Hellerine oil part number 670170 sparingly and a cable tie) at the base of the reservoir. Replace the assembly into the autoclave and secure in place. The jubilee clip around the reservoir filling pipe should be tightened to a maximum torque setting of 1.7N m, do not exceed this as excess torque could damage the reservoir and filling pipe. Replace the two silicone pipes at the top of the reservoir (vacuum exhaust and discharge) using new cable ties.
ST-SM33l P31/51
8.6 Mains transformer
Detach the mains transformer (item 27, Fig.3) connections, noting the position of each one. Remove the two retaining screws and shakeproof washers and withdraw transformer. The replacement transformer should be an identical unit, rated at 75VA. To replace the transformer, reverse the removal procedure ensuring the connections are remade in the correct manner.
8.7 Power board
Note: See Appendix 1 before proceeding.
8.7.1 T o remove the power board (item 26, Fig.3) release the two thumb nuts at the top of the board taking care not to loose the spacers between the board and supports. Carefully tilt the board away from the reservoir and pull it clear from the base clip with the insulation sheet. When free remove all the connecting cable clips from their sockets on the power board. Cut off (taking care not to damage the pipe) the cable tie around the silicone pipe where it connects to the top of the pressure transducer.
open and push the right-hand solenoid right before tightening the right-hand solenoid screws. Ensure the door lock mechanism is working freely .
8.10 Door lock magnetic catch (vac. only)
To replace the magnetic catch (item 7, Fig.3) release its retaining screws and remove. When replacing adjust the close solenoid (Fig. 3, item 8) until the primary lock is fully closed and secure.
8.11 Chamber temperature sensors
Disconnect the temperature sensors from the wiring loom by releasing their plug from its socket on the loom. Remove the temperature sensors (item 23, Fig.3) up and clear of the chamber. Replace the sensor by the reverse of the removal procedure ensuring a new sealing washer is used and is in place. Finally for vacuum units perform a leak test (special function F4)
Note: When a new temperature sensor is fitted, the autoclave must be recalibrated (see Calibration 7.4).
8.7.2 Replacement is simply the reverse of the removal procedure but ensure the correct procedures are used when handling the PCA to protect it from static discharge. Ensure the spacers and the insulation sheet are returned to their original positions and the silicone pipe is reconnected to the pressure transducer (using Hellerine oil part number 670170 sparingly) with a new cable tie.
Note: When a new power board is fitted, it will be necessary to recalibrate the autoclave (see Calibration section 7.4).
8.8 Door microswitches
To remove and replace either of the door microswitches (items 18 and 19, Fig.3) remove the two 'truarc' retaining washers from the microswitch location pillars. Expose the soldered connections to unsolder the join and release the microswitch. Replacement is the reverse of removal but ensure the correct microswitch is used (they are not the same), the insulation sleeve is replaced after soldering the connection and new retaining washers are used. There may be some adjustment required after replacement, ensure they operate and change state correctly .
8.9 Door lock/open solenoids
To remove a door lock (item 8, Fig.3) or open (item 6, Fig.3) solenoid release the two securing screws and shakeproof washers from the top of the solenoid assembly . To remove the solenoid slide it back (open solenoid) or forwards (lock solenoid) to clear the armature. Remove the two electrical connections and if removing both solenoids note the electrical connections. Replacement is the reverse of the removal procedure but ensure to move the lock closed and push the left-hand solenoid right before tightening the left-hand solenoid screws and move the lock
8.12 Heating element CAUTION
Avoid kinking the capillary tube/s.
8.12.1 To remove the heating element (item 4, Fig.4)
disconnect the electrical spade connections from the element terminals at the back of the chamber. Remove the sterilizing trays, carriers, and chamber furniture from inside chamber and release the thermostat sensor from its clip. (Note: On the non-vacuum unit there are two sensors). Unscrew and remove the large nut from the heater mounting boss at the rear of the chamber. Withdraw the heating element through the chamber door.
CAUTION
Ensure that the sensor(s) are
correctly positioned, see details
in sections 8.24 and 8.25.
8.12.2 Replacement of the heating element is the
reverse of the removal procedure. Ensure that a new sealing washer is used and the element is horizontal when fitted and does not touch the chamber furniture when replaced. Also ensure that the nut is fully tightened, to avoid leaks. Finally for vacuum units perform a leak test (special function F4).
8.13 Door seal
Open the chamber door and remove the door seal (item 34, Fig.3) from its retaining groove. Ensure the front face and groove in the neck ring and the door sealing face are cleaned with a dry lint free wipe. When fitting a new door seal wipe it with diluted washing up liquid first. Start at the top and work down both sides to the bottom pushing the
P32/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
seal into the groove at approximately 45°. Ensure that it is correctly seated into its groove and that it is not kinked. (NOTE: DO NOT USE GREASE). Finally for vacuum units perform a leak test (special function F4).
8.14 Vacuum pump
Note: For pump refurbishment see section 8.35.
8.14.1 To remove the vacuum pump (item 9, Fig.3) disconnect the electrical connection from the wiring loom. Cut off the cable ties (taking care not to damage the pipes) from the silicone pipes where they connect to the pump and release the pipes from the vacuum pump, noting the orientation to ensure correct reconnection. Release the large 'O' ring holding the pump in place and remove the vacuum pump.
8.14.2 Replace the pump by the reverse of the removal procedure ensuring the pipes are connected to the correct pump ports (using Hellerine oil part number 670170 sparingly) with new cable ties.
8.15 Cooling fan
8.15.1 T o remove the cooling fan (item 10, Fig.3) it is necessary to remove the power board (see section 8.7) to gain access to the fixings. Disconnect the electrical connection from the wiring loom and remove the four screws and shakeproof washers and remove the fan.
thread. Rotate the spout cup back into position, replace the top screw but do not tighten screws. Ensure the safety valve shield (item 27a, Fig. 9 sheet 2) is positioned correctly as shown in Fig. 9 sheet 2. Fit the front cover but not the top cover and then tighten the spout cup screws with the spout cup central to the front panel. Finally for vacuum units perform a leak test (special function F4).
8.20 Band heater temperature sensor
To remove the band heater temperature sensor (item 28, Fig.3) release the sensor wire from it’s connection to the wiring loom and remove the securing nut and shakeproof washer. Refit by reversing the above and ensure that the sensor is in tight contact with the band heater.
Note: When a new temperature sensor is fitted, the band heater must be recalibrated (see section 7.4).
8.21 Mains switch
T o remove and replace the mains switch remove the cable tie securing the cover, lift the cover and remove the four spade connectors from the switch noting the terminals each was connected to. Release the switch from the base tray by pressing the retaining lugs in on each side at the back of the switch. Replace switch by reversing the above ensuring correct orientation and the cover and cable tie are replaced.
8.15.2 Installation is the reverse of the removal procedure. However, when installing the fan ensure that the label is to the top and the cable exits at the bottom to the back left (looking from the front) This ensures that the flow is in the correct direction. Check finger guard is in place.
8.16 Bacterial air filter
Remove and replace the bacterial air filter (item 1, Fig.2) as detailed in the 'Instructions for Use'.
8.17 Gauze filter
Remove clean and/or replace the gauze filter (item 6, Fig.2) as detailed in the 'Instructions for Use'.
8.18 Printer
Remove the printer (item 5, Fig.2) by reversing the fitting instructions provided in the 'Instructions for Use'. T o replace the printer follow the notes provided in the 'Instructions for Use'.
8.19 Safety valve
To remove the safety valve (item 24, Fig.3) first remove the top screw securing the reservoir spout cup, loosen the bottom screw and rotate the cup out of the way . T o remove the safety valve, unscrew it from the chamber neck ring. When replacing the safety valve clean the female thread in the neck ring and apply PTFE tape to the new valves
8.22 Band heater
8.22.1 To remove the band heater (item 4, Fig.3) remove the band heater temperature sensor (see section 8.20) and the power board (see section 8.7). Remove the heater spade connections and release the heater clamp bolts. Carefully slide heater off the chamber vertically by allowing the heater to open up into a ‘U’ shape.
8.22.2 Refit a new band heater (use threadlock part number 670650 on retaining screws) by reversing the above, referring to sections 8.20 and 8.7 for details on how to refit the temperature sensor and power board.
8.23 Chamber water level sensors
To remove, clean and replace parts of the chamber water level sensors (items 14 and 17, Fig.3) release the door cover to gain access to the fixings and wiring. Remove the outer nut and washers retaining the wire terminal and the nut holding the sensor assembly in the chamber door. Remove the outer bush and the sensor (an M4 x 35mm cap head screw) and bush from inside the chamber door. Clean the internal chamber seating and refit new components as follows:
Remove any burr on the lead-in edge on both sides of the bore and ensure bore is clean. Fit one bush onto the screw, wipe ‘O’ ring with diluted washing up liquid and fit this onto the screw. Slide this assembly into the bore until the bush is flush with the door. Fit the other bush on the reverse side, a plain washer and then the Nyloc nut.
ST-SM33l P33/51
Tighten this nut until the ‘O’ ring starts to become compressed then tighten a further half turn (note small gap illustrated, do not tighten nut to eliminate). Fit the remaining washers, nut and ring terminal as shown below. Finally for vacuum units perform a leak test (special function F4).
Note: The sensor assembly on early models may be different to that shown below which is now the preferred arrangement. If replacing an assembly the version shown below should be fitted as detailed above, see Parts list 1 (item 34) for component part numbers.
8.26 Thermal fuse (non-vac. units)
IMPORTANT If the thermal fuse trips, or needs replacing,
the condition of the chamber MUST be inspected for any signs of damage or overheating. To replace the thermal fuse (not illustrated, strapped on the outside of the chamber behind neck ring) simply release the spring clip and disconnect its spade connections. Replace with a new one ensuring the fuse is in close contact with the chamber wall and correctly oriented.
8.27 Display board
T o replace the display board (fitted behind the display panel) remove all connections and unclip it from the front moulding allowing the ribbon cable from the membrane key pad to pass through the slot in the board. Replacement is the reverse but ensure the correct procedures are used when handling the PCA to protect it from static discharge and the membrane key pad ribbon cable is passed through the slot in the display board and reconnected to it. The PCA will require configuring and calibration using the Autoclave Service Interface Software.
8.24 Reset thermostat (overheat)
8.24.1 T o remove the Reset thermostat (item 25, Fig.3) and attached bulb, remove two spade connectors from the switch noting the terminals each was connected to. Release switch from the base by removing two retaining screws and shakeproof washers. Free the bulb inside chamber by releasing the clamp around bulb and heater element. Remove capillary tube and bulb through chamber wall by undoing the tubes captive nut from the chambers neck ring. Release capillary tubing from all constraints noting the path taken.
8.24.2 Replace new thermostat and bulb by reversing the above using a new 'O' ring seal (part number 391 189) and ensure it is correctly in place around the capillary tube plug before fitting. When repositioning the clamp and bulb they should be approximately 50mm in from the front of the elements left hand segment (viewed from the door end of the chamber). On non-vacuum units lay the bulb into the clamp on top of the cycling (auto reset) thermostat bulb. Fold the clamp legs over to trap the bulb (or bulbs) with a large pair of pliers. Ensure the bulb (or bulbs) are held tight in the clamp and that the tube (or tubes) are routed away from above the element so that they are not damaged by the chamber furniture. Finally for vacuum units perform a leak test (special function F4).
8.25 Cycling thermostat (non-vac. units)
To remove and replace the cycling thermostat (item 3, Fig.4) follow the procedure for the reset thermostat (see section 8.24). Both items are removed and replaced using the same procedure.
8.28 Mains fuse holders
To replace a mains fuse holder (item 8, Fig.2) it will be necessary to remove the water discharge solenoid valve completely, see section 8.4. Fuse holder replacement entails releasing the spade connections, unclipping the old one from the base of the unit and clipping back a new one. Replace the spade connections to the correct terminals then replace the water discharge solenoid valve. Ensure the cover and cable tie are replaced.
8.29 Membrane key pad
To replace the membrane key pad (item 3, Fig.2) entails removing the ribbon cable from the key pad to the display board and then prising the key pad away from the front moulding (it is held in place by adhesive). Clean the moulding face before re-attaching a new key pad by removing the release paper on the adhesive and pressing in place. Reattach the ribbon cable to the display board.
8.30 Reservoir level sensor
Do not remove the sensor connections before draining the reservoir. The reservoir level sensor is fitted in the
base of the reservoir and consists of two parts each connected electrically. One connection is to the water fill pipe spigot. The second is above the water fill pipe spigot and it must be connected correctly to avoid leaks and make good electrical connection. This is fully detailed in section
8.5.3. The only parts that would normally require
replacement are the sealing washers and PTFE tape around the screw. Do not forget to refill the reservoir.
P34/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
8.31 Water pump
8.31.1 T o remove the water pump (item 2, Fig.4) after draining the reservoir or clamping the supply pipe,
remove it from the spring retaining clip for ease of access. Disconnect the electrical connection by releasing the spring retaining clip on the plug attached to the wiring loom. Cut off the cable ties from the silicone tubes (taking care not to damage the tubes) where they connect to the pump and release the tubes noting the orientation to ensure correct reconnection.
8.31.2 Replace the pump by the reverse of the removal procedure ensuring the pipes are connected to the correct pump ports (using Hellerine oil part number 670170 sparingly) with new cable ties. Refill the reservoir if drained or remove the pipe clamp.
8.32 T est probe port & pressure safety gauge
Replacement of a test probe port or pressure safety gauge is self explanatory , but ensure the sealing face is clean and use PTFE tape when refitting. It is not essential that the gauge needle points up vertically at ambient pressure. Finally for vacuum units perform a leak test (special function F4).
8.33 Door
To replace an autoclave door in the unlikely event that it has been damaged first remove the water level sensors (see 8.23) the three test probe ports and pressure safety gauge (see 8.32). The door is removed by releasing the two screws marked ‘A’ in Fig.3 (but not removing them from the hinge pin). Replacement is the reverse but refer back to sections 8.32 and 8.23. Note that a complete door assembly is available, Part No. 391212.
8.34 Band heater resettable thermostat
T o replace the band heater resettable thermostat (item 29, Fig.3) remove the electrical spade connections and the nuts and washers that clamp it to the band heater. Replacement is the reverse of removal but ensure the thermostat is in contact with the band heater and the contact area is clean and not corroded.
8.35 Vacuum pump maintenance
8.35.1 If the vacuum performance of the pump is poor, but it is operating correctly, it can be refurbished rather than replaced, by the following maintenance procedure. Refurbishment of the vacuum pump entails the replacement of the diaphragm with hold down plate, valve plate seal and ‘O’ ring in both heads. During dismantling carefully note part orientation for reassembly, failure to assemble in the correct orientation could result in a pump working in reverse, or, with pump heads working against each other. Obtain pump refurbishment kit Part No. 391 185 and proceed as follows:
Fig. 6 Pump maintenance part identification
8.35.2 Disconnect the two pipes (one on each pump head) by cutting off the cable tie around the silicone pipe (taking care not to damage the pipe). Remove the electrical connection and the pumps retaining 'O' ring and remove the pump from the autoclave.
8.35.3 Remove the four head screws (1, Fig.6) in each head and lift off the two head assemblies (2, Fig.6) together .
8.35.4Remove the four diaphragm retainer plate screws (3, Fig.6) in each head, lift off both diaphragm retainer plates (4, Fig.6) remove both diaphragms (5, Fig.6) and discard the old diaphragm retainer plates and diaphragm.
ST-SM33l P35/51
8.35.5 Place new diaphragms on con-rod tops (6, Fig.6) with the new diaphragm retainer plates and replace the screws in a cross pattern using Loctite 242 (Part No.306033) on threads and torque to 2.5Nm, re-torque the first screw to each head.
8.35.6 Remove the valve plate (7, Fig.6) from both head assemblies by removing the four retaining screws (8, Fig.6) and discard the valve plate seals (9, Fig.6) beneath the valve plates.
8.35.7 Remove the two flapper retaining screws (10, Fig.6) in each valve plate and clean the flappers (11, Fig.6) and valve plates. Replace the flappers to their original positions and secure with screws.
8.35.8 Fit new valve plate seals and replace the valve plate assemblies and secure by tightening screws in a cross pattern and torque to 2Nm, re-torque the first screw to each head.
8.35.9 Separate the heads, replace the 'O' ring seals (12, Fig.6) around the connecting tube (13, Fig.6) and reassemble the heads together.
8.35.10 Position one of the diaphragms at the mid stroke position and ensure that the diaphragm is seated correctly. Replace the head and head screws and secure leaving screws loose, position the second head, secure it and finally torque screws in both heads in a cross pattern to 3Nm, re­torque the first screw to each head.
8.35.11 Reassemble the pump in the autoclave and test function and flow before reconnecting the pipes. Finally test the autoclave by running at least one complete cycle. If an abnormal noise is heard, obtain assistance from the Eschmann After Sales Service Department (details inside front cover).
8.36 Pressure door lock
8.36.1 If the pressure door lock does not work correctly it should be replaced as a complete unit. There are no user serviceable parts within. However before replacement check that all pipework to the pressure door lock are not blocked or damaged and that the M6 screw on the plunger has not come loose. If either of the later are applicable replace as below.
8.36.2 To replace the pressure door lock release the pipe from the top connection taking care not to damage the pipe. If required also replace the pipe set (part number
111805). Release the four M4 retaining screws and remove the pressure door lock. Remove from the old lock the M6 screw and the spacer for fitting to the new pressure door lock (part number 427035).
8.36.3 Fit the old M6 screw and the spacer to the new pressure door lock using threadlock (part number 306033) on the screw thread. Replace the pressure door lock assembly and connect the tubing to the top connection using a tie wrap (part number 695777). Also use a tie wrap to hold the pipe in place. T est function after replacement to ensure correct operation.
P36/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
APPENDIX 1
Before replacing a Power PCA check to see if the autoclave is fitted with a Mains Filter (Part number 1 11703) as shown in the illustration below (item number 1). Then proceed as applicable:
Mains Filter fitted. If a Mains Filter is fitted the Power PCA (Part Number 111653) can be
replaced as detailed in section 8.7. In this case the autoclave will already be fitted with the new Power Loom (Part Number 111856) and Power PCA (Part Number 111653).
Mains Filter not fitted. If a Mains Filter is not fitted then before replacing the Power PCA (which
should be replaced with the latest version, Part Number 111653 now common to both vacuum and non-vacuum autoclaves) ensure that both a Mains Filter* is fitted (see illustration below) and the existing Power Loom** (Part Number 391288) is upgraded. Then replace the Power PCA as detailed in section 8.7
* The Mains Filter can be supplied as part of a kit (Part number 1 1 1700) which includes all the items required to fix it to the autoclave as shown in the illustration below. Ensure the new Power Loom (Part Number 111856) is fitted or the existing Power Loom (Part Number 391288) is upgraded by using loom upgrade kit (Part number 1 11727) see **.
** The existing Power Loom (Part number 391288) can be modified to eliminate replacing the whole loom by using loom upgrade kit (Part number 1 1 1727). This kit comes complete with a leaflet providing full fitting instructions. The loom upgrade kit also provides the parts required to upgrade the crimp connections to the Immersion heater, Band heater, Fuse holder and Mains switch.
Parts and fixing detail for Mains Filter.
ST-SM33l P37/51
NOTE: The fixings (screws, nuts, washers etc.) for each part are detailed in the parts list following each item.
Note: Illustration shows 3 twinlocks fitted, product now supplied with only 1 and two blank plugs. Also pressure door lock not illustrated.
Part view covers on
Fig. 7 Parts list 1 - Electrical components, shown covers off
P38/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
PARTS LIST 1 - ELECTRICAL COMPONENTS
Item Part No. Description Qty. Spares
1 391318 Fuse holder 2 1 2 (NI) 301817 Fuse (10 Amp) 2 2 3 391319 Switch, mains 1 ­3a (NI) 391188 . Fuse / Switch cover 1 ­4 (NI) 39114 5 Socket, IEC 1 ­5 391157 Fan 1 ­5a(NI) 111691 . Finger guard 1 -
710219 . Screw M4 x 16mm cap head (was 10mm) 4 -
710731 . Washer M4 shakeproof 4 ­6 (NI) 391299 Ribbon cable (40-way), power to display PCA 1 ­6a (NI) 71 1348 . Ribbon cable clamp 1 ­7 (NI) 391324 Loom, signal 2 (valves, fan, w .l. sensors, microswitches) 1 ­7a(NI) 391323 Loom, signal 1 (mains transformer, water pump,
water reservoir, door solenoids) 1 ­8 (NI) 391288 Loom, power (see Appendix 1) 1­9 (NI) 391388 Loom, water level 1 -
10 (NI) 301842 Grommet 1 -
11 391298 Transformer 1 -
307249 . Screw M4 x 8mm Pozi head 2 ­710731 . Washer M4 shakeproof 2 -
12 391036 Sensor , temperature 1 1
391050 . Washer, Dowty 10mm dia. 1 1
13 391067 Element, heating 1 1
416149 . Clamp, element temperature sensor 1 ­425181 . Washer 1 1
420024 . Nut, element 1 ­14* (NI) 391069 Thermal fuse assembly 10 amp 1 1 15# 391070 Band heater 1 ­16* 391061 Water pump 1 -
391054 . Spring clip 1 -
710734 . Washer M3 shakeproof 1 -
391086 . Screw M3 x 6mm Pozi head 1 ­17# 391084 Cycling Thermostat 1 1
391086 . Screw M3 x 6mm Pozi head 2 -
710734 . Washer M3 shakeproof 2 -
391 1 89 . Spare ‘O’ ring seal As Reqd. 1 18 391040 Reset thermostat (overheat) 1 1
391086 . Screws M3 x 6mm Pozi head 2 -
710734 . Washers M3 shakeproof 2 -
391 1 89 . Spare ‘O’ ring seal As Reqd. 1 19 391325 Solenoid, door lock/open 2 1
307249 . Screws M4 x 8mm Pozi head 4 -
710731 . Washer M4 shakeproof 4 ­20# 391060 Sensor, band heater temperature 1 1
709861 . Washer M4 plain 4 -
710731 . Washer M4 shakeproof 1 -
307046 . Nut 4BA 1 ­21 391066 Microswitch (door locked), lever, 1 1
709861 . Washer M4 plain 2 -
391048 . Truarc retaining disc 2 2 22 391065 Microswitch (door closed), roller, 1 1
709861 . Washer M4 plain 2 -
391048 . Truarc retaining disc 2 2 * = non-vacuum units only # = vacuum units only (NI) = Not illustrated (Continued on page 37)
ST-SM33l P39/51
NOTE: The fixings (screws, nuts, washers etc.) for each part are detailed in the parts list following each item.
Note: Illustration shows 3 twinlocks fitted, product now supplied with only 1 and two blank plugs. Also pressure door lock not illustrated.
Part view covers on
Fig. 7 (repeated) Parts list 1 - Electrical components, shown covers off
P40/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
PARTS LIST 1 (Continued) - ELECTRICAL COMPONENTS
Item Part No. Description Qty. Spares
23 391057 Solenoid valve (only 3 fitted on non-vacuum) 5 1
710211 . Screw M4 x 10mm cap head 2 per valve -
710731 . Washer M4 shakeproof 2 per valve -
391075 . Spare ‘O’ ring for solenoid valve As reqd. 5 24# 391053 Vacuum pump (for pump kit see 8.35.1) 1 -
391072 . ‘O’ ring 1 1 25# 391314 Power board PCA, vacuum autoclave (see Appendix 1) 1-
391097 . PCA insulation sheet 1 -
696643 . PCA slide 1 -
301796 . Spacer 2 -
424171 . Nut, thumb 2 2 26* 391308* Power board PCA, non-vacuum autoclave (see Appendix 1) 1-
391097 . PCA insulation sheet 1 -
696643 . PCA slide 1 -
301796 . Spacer 2 -
424171 . Nut, thumb 2 ­27 (NI) 391312 Display board PCA 1 ­28# 391016 Membrane key pad for vacuum autoclave (yellow buttons) 1 ­28a## 111772 Membrane key pad for QuickVac & SES2555 (blue buttons) 1 ­29* 391015 Membrane key pad for non-vac. autoclave (yellow buttons) 1 ­30 391010 Printer assembly complete 1 ­30a 391 1 87 Printer ribbon retaining clip (if printer fitted) 1 ­31 391301 Cable assembly, printer (16-way) 1 ­32 391302 Cable assembly , RS232 (9-way) 1 -
391087 . Screw, M3 x 8mm, Pozi head 2 -
710734 . Washer, M3 shakeproof 2 ­33 759647 . Band heater reset thermostat 1 -
709861 . Washer M4 plain 2 -
710731 . Washer M4 shakeproof 2 -
307046 . Nut 4BA 2 ­34 111523 DOOR SENSOR ASSEMBL Y KIT (two sensors as below) 1 per unit -
111520 . Insulation Bush 2 ­1 11522 . Screw M4 x 35mm 1 ­1 11519 . ‘O’ ring (VITON) 1 -
1 11521 . Nut M4 (NYLOC) 1 ­709861 . Plain Washers M4 3 ­710731 . Shakeproof Washers M4 1 ­710371 . Nut M4 (plain) 1 -
35 - - - - - - RESERVOIR WATER LEVEL SENSOR COMPRISING
700237 . Screw M5 x 20mm cap head 1 ­709862 . Washer M5 plain. 2 ­391074 . Washer, PTFE 1 1
- - - - - - EARTH FIXINGS 693640 . Metway clamps (primary earth only) 2 ­709861 . Washer M4 plain 3 ­710731 . Washer M4 shakeproof 3 ­710371 . Nut M4 3 -
- - - - - MISCELLANEOUS 695777 Cable tie As Reqd. 6 301600 Tape, PTFE As Reqd. 1 714188 IEC Mains lead 1 -
* = non-vacuum units only # = vacuum units only ## = QuickV ac and later SES2555 vacuum units only (NI) = Not illustrated
Note: For information on the Mains Filter see Appendix 1.
ST-SM33l P41/51
NOTE: The fixings (screws, nuts, washers etc.) for each part are detailed in the parts list following each item.
Fig. 8 Parts list 2 - Pipes and plumbing components
P42/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
PARTS LIST 2 - PIPES AND PLUMBING COMPONENTS
Item Part No. Description Qty. Spares
1 301349 Twin lock test plug 1 1 2# (NI) 391140 Screw, 10mm x 12mm Skt cap head 1 ­# 391050 . Washer , Dowty 10mm dia. 2 2 3 391018 Reservoir spout cup 1 -
(NOTE: The new preferred cup fixings shown below may differ from those initially supplied)
112012 . Washer, M5 x 16mm (top fixing only) 1 -
1 12011 . W asher, spring M5 2 -
710271 . Screw, M5 x 16mm soc. cap head 2 -
4 391019 Reservoir 1 -
710119 . Screw, M5 x 10mm Pozi head 2 ­710733 . Washer, M5 shakeproof 2 -
1 11312 . Reinforced hose (inside reservoir) 1 ­5 391035 Condensing coil 1 ­6 391056 Headed bushes (older units only) 2 -
710119 . Screw, M5 x 10mm pozi head (older units only) 2 -
7 391197 Pipe spigot and filter assembly 1 -
307252 . Washer (large) 1 1 8 391059 Reservoir cap 1 ­9 391064 Manifold 1 ­10 391079 Cable grommet 1 1 11* 391095 Domed plug 1 ­12# 391080 Elbow 1 -
301357 . Nut, ¼” BSP 1 ­13 391055 Pipe spigot 1 ­14 3911 66 1.25” Bore hose 1 1
391062 . Jubilee clip 2 ­15# 391033 Air filter mount 1 ­16* 391063 Domed plug 1 ­17# 713580 Bacterial filter and straight connector 1 4 18 3911 68 Discharge / Air bleed pipe assembly 1 1 19 3911 69 Fill / Empty pipe assembly for Vacuum autoclave 1 1
(inc. spiral wrap 391 190) OR
1 11011 Fill / Empty pipe assembly for Non-Vacuum autoclave 1 1
(inc. spiral wrap 391 190)
20 3911 70 Silicone tube, 400mm x 3mm Ø x 2mm wall 1 1
OR If Pressure door lock fitted:
1 11805 Pipe set, pressure door lock 1 1 21* 391171 Silicone tube, 280mm x 5mm Ø x 2mm wall 1 1 22# 391172 Silicone tube, 450mm x 5mm Ø x 2mm wall 3 ­23 391156 Restrictor 1 ­24 380010 Drain tap 1 ­25 3911 75 Gauge, pressure/vacuum safety 1 ­26 380020 Plug, 1/4” BSP 884/2 2 1 27 3911 32 Washer, Dowty 1/4” BSP 2 1
- - - - - MISCELLANEOUS 695777 Cable tie As reqd. 10 301600 PTFE tape As reqd. 1 670170 Hellerine oil (pipe fitting lubricant) As reqd. 1 391077 Silicone tube (per metre) (3mmØ x 2mm wall) As reqd. ­391078 Silicone tube (per metre) (5mmØ x 2mm wall) As reqd. -
* = non-vacuum units only # = vacuum units only (NI) = Not illustrated
ST-SM33l P43/51
NOTE: The fixings (screws, nuts, washers etc.) for each part are detailed in the parts list following each item.
Fig. 9 sheet 1 Parts list 3 - General assembly components
P44/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
PARTS LIST 3 - GENERAL ASSEMBLY COMPONENTS (Fig. 9 Sheet 1)
Item Part No. Description Qty. Spares
1 391002 Door cover 1 -
391009 . Pin, spring, printer location 1 ­391006 . Hinge pin, top 1 -
2 391012 Pin, door linkage 1 -
391013 . Spring, door linkage pin 1 -
3 391004 Hinge plate, door bottom 1 -
710091 . Screw, M5 x 10mm Pan head 2 -
4 391003 Printer door 1 ­5 391146 Fastener, ¼ turn 1 ­6 391143 Front panel moulding 1 -
391005 . Bush, door hinge 2 ­710119 . Screw, M5 x 10mm Pozi head (was Pan head) 4 -
1 11193 . Shouldered washer (top fixing screws only if fitted) 2 ­7 391102 Overlay label 1 ­8 391022 Door, chamber 1 ­9 391028 Seal, door 1 2 10 391037 Filter , gauze chamber 1 2 11 391043 Top cover (older models) 1 ­ or 11 1648 Top cover (with side fixings) 1 -
1 11767 . M5 x 10mmTorx pin screw 2 ­12 391046 Foot, small rear 2 -
391090 . Screw, M5 x 16mm Pozi head 2 -
13 391038 Foot, large front 2 -
391090 . Screw, M5 x 16mm Pozi head 2 ­30 3911 87 Printer ribbon retaining clip (if printer fitted) 1 ­31 111409 Blanking plate 1 ­32 110953 . Screw, M4 x 12mm Torx pin c.sk. 2 ­33 111692 Printer roll holder 1 ­34 3911 98 Door spacer 1 -
- - - - - - MISCELLANEOUS
391141 Label, LS5 Door cover 1 -
391142 Label, SES2555 Door cover (green) 1 -
391138 Label, reset press here symbol 1 -
1 11782 Label, QuickVac Door cover 1 ­1 11783 Label, SES2555 Door cover (blue) 1 -
* = non-vacuum units only # = vacuum units only (NI) = Not illustrated
(continued on page 46)
ST-SM33l P45/51
NOTE: The fixings (screws, nuts, washers etc.) for each part are detailed in the parts list following each item.
For pipe set see Parts list 2
35
38,39
36,37
Note: Illustration shows 3 twinlocks fitted, product now supplied with only 1 and two blank plugs.
Fig. 9 sheet 2 Parts list 3 - General assembly components
P46/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
PARTS LIST 3 - GENERAL ASSEMBLY COMPONENTS (Fig. 9 Sheet 2)
Item Part No. Description Qty. Spares
14 391024 Secondary catch 1 -
391049 . Spring 1 1 15 391032 Pivot pin (secondary catch) 1 ­16 391083 Door beam 1 -
391047 . Pin, Sellok 2mm Ø x 16mm (microswitch supt) 4 4 17 391025 Chamber neck ring assy 1 -
710441 . Bung, M10 grub screw 2 -
710280 . Bung, M6 grub screw 2 -
710019 . Screw, M5 x 10 slotted head (was Pozi hd) 2 -
710733 . Washer, M5 shakeproof 2 ­18 391051 Jubilee clip 1 ­19 391029 Pin, door pivot 1 -
307239 . Screw, M5 x 16 cap head 2 -
710733 . Washer, M5 shakeproof 2 ­20 391030 Pin, door beam pivot 1 -
699158 . Pin, Sellok 4mm dia x 25mm 1 1 21 391082 Catch, primary lock 1 ­22 391031 Pin, primary catch pivot 1 -
699158 . Pin, Sellok 4mm dia x 25mm 1 1 23 3911 05 Pin, Microswitch actuation 1 ­24 424444 Calibration clip 1 -
709861 . Washer, M4 plain 1 -
710731 . Washer, M4 shakeproof 1 -
307046 . Nut, 4BA 1 ­25 391042 Link, primary lock solenoid 1 -
699188 . Pin, 4mm x 16mm Sellok (for solenoid spindle) 2 -
699158 . Pin, Sellok 4mm x 25mm 1 -
391074 . Washer, PTFE 3 ­26 3911 39 Catch, magnetic (vacuum autoclave only) 1 -
710371 . Nut M4 1 -
391086 . Screw, M3 x 6 Pozi head 2 -
710734 . Washer, M3 shakeproof 2 ­27 391071 Pressure relief valve 1 1 27a 391 1 91 Safety valve shield 1 ­28 391041 Chassis and base plate assembly 1 ­29 3911 99 Plastic spacing washer (1.19mm) 1 ­ or 391201 Plastic spacing washer (0.84mm) 1 ­35 111804 Bracket pressure door lock 1 ­36 427035 Pressure door lock assembly 1 ­37 110535 Screw M4 x 12mm PV Pan head Pozi 2 ­38 710154 Screw M6 x 25mm Soc cap head (use retainer 306033) 1 ­39 111762 Spacer 1 -
- - - - - - MISCELLANEOUS
670070 (or 306055) Grease, door mechanism As reqd. 1
306234 Pipe adhesive, chamber plugs As reqd. -
301600 PTFE Tape As reqd. 1
670650 Screw retainer As reqd. 1
695777 Cable tie As reqd. 6
391173 Universal rack - -
391 1 74 Vertical cassette holder - -
REF 87-041-45 Dental handpiece furniture set - -
* = non-vacuum units only # = vacuum units only (NI) = Not illustrated
ST-SM33l P47/51
9 TECHNICAL DATA
9.1 Electrical
Supply 230V 50/60 Hz a.c.
For use with alternating current Loading at 230V 2.0kW 10A (max.) Fuses T10A (Part.No.301817) (x2)
T3.15A (Part.No.696207) (x1)
9.2 Sterilizing data
Standard:
Sterilizing time at 134/137°C is 3min. 5sec., at 121/124°C is 15min. and at 134/137°C (extended) is 18min. (but all can be extended
All:
Drying time (if selected) is from 17 minutes (but can be extended).
Operating pressure at 121°C is 1.03 bar, at 124°C is 1.23 bar , at 134°C is 2.03 bar and at 137°C is 2.32 bar.
9.3 Maximum Load
Tray - - - 1.5kg maximum per tray Cassette - - 2.0kg maximum per cassette Chamber - - 8.0kg maximum total chamber load
9.4 Water reservoir capacity
8.0 litre
9.5 Weight (approx.)
Non-V acuum Net = 42.5kg
Shipping = 55.5kg.
For V acuum add 6.5kg.
9.6 Safety standards
EN 61010-1 (1993) inc. Amendment 2 (1995) EN 61010-2-041 (1996)
9.7 Design pressure
2.85bar (gauge)
P48/51 ST-SM33l
Eschmann Equipment, Peter Road, Lancing, West Sussex, BN15 8TJ, England.
Tel: +44 (0) 1903 753322. Fax: +44 (0) 1903 766793. www.eschmann.co.uk
LS5/SES2555 & QuickV ac
VACUUM & NON- VACUUM AUTOCLAVES
Fig. 10 Schematic electrical diagramme, non-vacuum autoclave
ST-SM33l P49/51
LS5/SES2555 & QuickV ac
VACUUM & NON- VACUUM AUTOCLAVES
Fig. 11 Schematic electrical diagramme, vacuum autoclave
P50/51 ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
Fig. 1a Control panel for SES2555 and LS5
Fig. 1b Control panel for QuickVac
and 2555QV models
ST-SM33l P51/51
Loading...