Keep these ‘Instructions for use’ in a safe convenient place for future reference.
Eschmann After Sales Service Department
The Eschmann After Sales Service Department is staffed and equipped to provide advice and
assistance during normal office hours. To avoid delays when making enquiries, please quote the
Model and Serial Number of your Operation Table which is shown on the Serial Number Plate
located on the table base (or the trunk section for the T20-m). Please ensure you include all
alpha and numeric digits of the Serial Number.
For further information visit www.eschmann.co.uk
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
Tel: +44 (0) 1903 765040. Fax: +44 (0) 1903 875711.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Patents and Trade marks
The ESCHMANN name and logo are trade marks of Eschmann Holdings Limited.
“T20-a”, “T20-m” and “T20-s” are trade marks of Eschmann Holdings Limited.
“Eschmann Equipment” is a trading name of Eschmann Holdings Limited.
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.
The CE marking affixed to the product certifies that it complies with the
European Medical Devices Directive 93/42/EEC and related legislation.
Instructions for use
T-IM114b April 2006
T20
Series
OPERATION TABLE
The T20 Series of operation table
T20-a
The T20-a table is a
powered operating
table suitable for a full range of general and specialist surgical
procedures. This operating table has an X-ray translucent traversing
tabletop with a built-in X-ray cassette tunnel and is capable of
supporting a maximum patient weight of 300kg (mobile 135kg). The
slimline column is offset relative to top and base for ease of C-arm
access. The T20-a table features a corded handset
controlling Trendelenburg, reverse Trendelenburg,
lateral tilt, flexion/extension (including 90º chair
position) traverse and height. The handset also
provides a battery level indication. Ergonomically
designed foot pedals are located at the head and leg end of the table base to provide braked, castor and 360°
mobility. The batteries in the table base are mains rechargeable with a standby battery in case of emergency.
Covers to the top and base are purple and made of a special scratch resistant, hard-wearing and easy to clean
seamless acrylic capped ABS. The mattress is moulded and antistatic with no seams.
T20-s
The T20-s table is a
powered operating
table suitable for a full range of general surgical procedures. This operating
table has an X-ray translucent traversing tabletop with a built-in X-ray
cassette tunnel and is capable of supporting a maximum patient weight of
300kg (mobile 135kg). The slimline column is offset relative to top and base
for ease of C-arm access. The T20-s table features a
corded handset controlling Trendelenburg, reverse
Trendelenburg, lateral tilt, flexion/extension (including
90° chair position) traverse and height. The handset
also provides a battery level indication. Ergonomically
designed foot pedals are located at the head and leg end
of the table to accommodate different operating theatre layouts providing braked, castor and 360° mobility. The
batteries in the table base are mains rechargeable with a standby battery in case of emergency. Covers to the top
and base are ocean blue and made of a special scratch resistant, hard-wearing and easy to clean seamless
acrylic capped ABS. The mattress is moulded and antistatic with no seams.
T20-m
The T20-m table is a
powered operating table
(maximum patient weight of 300kg or 200kg mobile) featuring all the
functions of the T20-a table with a mobile base that is supported by
four 125mm castors to enhance manoeuvrability and usability of the
operating table. The T20-m table is suitable for theatre environments
that require the functionality of a trolley system whilst maintaining the
performance of a table for a full range of surgical
procedures. The ergonomically designed single foot
pedal is located at the head end of the table and
has three positions to enable the table to be static
(braked), rotated 360° or moved in a straight line.
Covers to the top and base are purple and made of
a special scratch resistant, hard wearing and easy to clean seamless
acrylic capped ABS. The mattress is moulded and antistatic with no seams.
T-IM114bP3/62
CONTENTS
1.0 PRELIMINARY INFORMATION ..............6
1.1General 6
1.2About this manual.......................................... 6
1.1.5Ensure the table has been correctly installed
before starting to use it, see section 3.0.
WARNING
Read this preliminary information carefully and
note ALL of the warnings, cautions and safety
notes contained within these ‘Instructions for
use’ before using this Operation Table. Keep
these ‘Instructions for use’ close-to-hand at all
times for reference.
1.1General
1.1.1T20 Series operation tables are classified as
battery powered, mobile, general purpose, four section
operation tables. Their intended function is to support and
position a patient, in conjunction with their associated
accessories, during general surgical operations and
procedures in an operating theatre. Their intended
application is for use by medically qualified personnel,
trained in the use of the T20 Series powered operating
tables, during surgical operations and procedures in
accordance with these instructions.
1.1.2These ‘Instructions for Use’ should be referred to
for details of the following T20 Series Powered Operation
Tables, see the table’s Serial Number Plate for the actual
table Serial Number and REF No.:
T20-aCatalogue (REF) number prefixed T20
Serial Numbers (SN) prefixed T2AC*...... or above.
T20-sCatalogue (REF) number prefixed T20
Serial Numbers (SN) prefixed T2SB*...... or above.
T20-mCatalogue (REF) number prefixed T2M
Serial Numbers (SN) prefixed T2MA*...... or above.
* The last digit of the SN prefix is the design modification
state, this may increase during the build life of the tables.
NOTE: Some models have ‘Denyer’, ‘USA’, or ‘Euro (Kifa)’
accessory side rails, standard Eschmann accessories
that locate onto these may not fit, please check with
Eschmann Equipment before purchasing accessories.
1.1.3The T20 Series operation tables and accessories
must only be used as detailed within these ‘Instructions
for use’. Failure to do so could result in injury to patients or
users, or damage to the operation table and accessories.
Always ensure that all warnings and cautions detailed
within these ‘Instructions for use’, are strictly complied with.
Appendix 1 is provided for Hospitals that wish to keep a
log of those people trained in the safe use of this table.
Eschmann Equipment offers full training in the safe use of
these tables, please contact our After Sales Service
Department for details.
1.1.4‘Instructions for use’ and ‘Service manuals’ (see
section 3.3.9) as applicable should be readily accessible
for reference prior to and when operating, cleaning and
servicing the operation table. All manuals are available from
Eschmann Equipment, see inside front cover for address.
1.1.6All servicing and maintenance procedures should
be carried out by engineers who have been trained by
Eschmann. Training, a Service, or a Service Contract can
be arranged through our After Sales Service Department
(see inside front cover for contact details). Do not remove
the table covers at any time (danger of accessible mains
voltage), this should only be carried out during maintenance
procedures. Ensure that the operation table is regularly
serviced and checked for safety at least every six months.
1.2 About this manual
1.2.1Within the text of these ‘Instructions for use’ the
following terminology is used:
i)
Left and right
the side of the table when viewed from the long trunk
end (i.e. the head end with the table in its normal
configuration, head section fitted to the long trunk
as shown in Fig. 2.1). This is also the patient’s ‘left’
and ‘right’ when lying on the tabletop in a supine
position, with their head on the head section and
with the table in its normal configuration.
ii)
Auto-level
the automatic sequence of movements to return the
table to a preprogrammed position by pressing and
holding a single handset button. This does not affect
the head or leg sections, or any other manually
operated, or powered accessory, fitted to the table.
iii)
LEDs.
This abbreviation is used when referring to
the various indicator lights on the table or handset
which are light emitting diodes (LEDs).
iv)
Cranially and Caudally.
section moves away from column) and ‘caudally’ (leg
section moves away from column) are used to
describe tabletop movement (table in the normal
orientation). Normal table orientation is with the head
section in the long trunk, normal patient orientation is
with the patient’s head on the head section.
1.2.2These ‘Instructions for use’ have been split into
specific sections for ease of finding information (see the
main headings in ‘Contents’). Where applicable and within
each section, adequate cross references to other sections
are made to eliminate the need to duplicate information.
1.2.3Section 5 (Operation) details how to use the T20
Series of tables and operate their controls correctly. It is
strongly recommended that the user has read and is
familiar with sections 1, 2 and 4 before passing onto
section 5 and starting to use this operation table.
1.2.4Within this manual the sections in bold type
headed ‘WARNING’ give guidance on possible actions that
could lead to injury of the patient, or theatre staff, and
potential damage to the operation table. Sections headed
‘CAUTION’ give guidance on possible actions that could
lead to damage of this operation table which could then
lead to injury of the patient, or theatre staff.
. The terms ‘left’ and ‘right’ refer to
. The term ‘auto-level’ is used to describe
The term ‘cranially’ (head
P6/62T-IM114b
T20
Series
OPERATION TABLE
1.3Table description
1.3.1The T20 Series of four section operation tables
have been designed to provide facilities for General
Surgery, including Minimal Access procedures, Urology
and Gynaecology, Thoracic, Ophthalmic and ENT,
Neurosurgery, Plastic and Maxillo-Facial surgery, and nontraction Orthopaedic Surgery, they allow for intra-operative
radiography using a C-arm image intensifier. They are not
suitable for Orthopaedic procedures requiring traction.
1.3.2Careful design has minimised traps for potential
contamination stopping fluid entering the table during
normal use, cleaning and disinfection procedures. They
are stable, rigid in use and the robust construction provides
protection from patient trolley or C-arm knocks whilst still
providing easy access for servicing.
1.3.3Tabletop movements (i.e. Trendelenburg, height,
tilt, break and traverse) are electrically powered and
incorporate ‘soft start and stop’ motions for patient comfort
and safety. The pedestal base is foot operated and the
standard head and leg sections are hand operated (see
sections 5.1 and 5.2 respectively).
1.3.4The T20 Series of table are easy to operate,
theatre staff can quickly learn how to use them correctly
and safely. Tabletop control is provided by a corded
handset, or an optional footswitch (for Trendelenburg and
height control only) or an optional infrared handset. Hand
controls override footswitch control at all times and if a
corded handset is plugged into the table this will take
precedence over the infrared handset.
1.3.5Power for all powered table movements is
provided by internal rechargeable batteries. There are two
battery sets, main batteries and standby batteries. These
are recharged by an internal battery charger which requires
connection to mains voltage using the mains cord supplied.
Note: If the main table batteries fail or become critically
low (i.e. warning indicators were ignored and batteries have
not been recharged) the table cannot be powered directly
from the mains. However by depressing and holding the
standby battery switch, powered table movements can
again be achieved (using power from standby batteries).
1.3.6In emergencies (e.g. handset failure or a critically
low battery) a standby control panel on the column can be
used to control the table (however this MUST be limited to
emergency use ONLY, certain safety features are overridden when this panel is used, see section 4.6.5). It may
be required to press the standby battery switch if the main
battery charge level is too low.
1.3.7The bases on T20-a and T20-s tables are fitted
with enclosed multidirectional castors, that can be engaged
at any table height, making it easy to move on most theatre
floors. They have two foot pedals providing either castor,
wheel or braked orientations. The T20-m table is provided
with four large castors that can be adjusted by a single
foot pedal to provide wheel and castor modes and a braked
position (see section 5.1).
1.3.8Visual indicators have been restricted to essential
functions and information only (e.g. table switched ‘on’ see
section 5.3.2.2, battery level indication and battery charging
state see section 5.3.1). Audible signals are only used to
signal that the table has been switched ‘on’ for a long time
without operation and that it should be switched ‘off’ see
section 8.1, a ‘beep’ is also provided when switching ‘on’
(the audible inactivity ‘beep’ can be configured ‘off’ during
a service if required).
1.3.9Tabletops have a lightweight X-ray translucent
surface (designed to reduce shadows on images) and an
X-ray cassette tunnel with the facility for an X-ray cassette
(430mm x 340mm) to be placed at any point beneath the
full length of the patient’s body, see section 6.3. The
standard side rails allow placement of clamps and most
standard accessories. Simple button operated catches
release the head and leg sections when required.
1.3.10The tabletop can be adjusted into the following
patient positions:
♦Supine with C-arm access to patient from nipple
region to feet
♦Supine with C-arm access to patient from groin to
head
♦Supine Extension with C-arm access to break area
(Cholecystectomy-type procedure)
♦Supine Flexion (‘Lawn chair’ position)
♦Lateral
♦Lateral Extension with C-arm access to break
(Nephrectomy position)
♦Supine Lithotomy with C-arm access to whole of
the Urinary tract
♦Supine Lithotomy with or without Trendelenburg
♦Prone with or without extension at waist or hips (with
C-arm access)
♦90° Chair position with patient’s knees at the same
level as the heart, offset to the head end.
1.3.11For additional information see section 6.0 for
patient positioning notes, section 6.2 for the obese patient,
6.3 for radiographic procedures and 6.4 for illustrations
and details of safe loading.
1.3.12Castor covers (spats) are supplied as an optional
item with the T20-m table. These can be fitted by an
Eschmann Engineer at any time, please contact the After
Sales Service Department.
T-IM114bP7/62
2.0 TABLE PARTS AND SYMBOLS
2.1Part identification
2.1.1Fig. 2.1 shows the table top in its normal
configuration and identifies the major parts of the table
top.
2.1.2Fig. 2.2 identifies the various parts of the T20-a
and T20-s table base and column.
SNThis symbol indicates the unit serial number is
as indicated adjacent to the symbol.
REFThis symbol indicates the catalogue number is
as indicated adjacent to the symbol.
This symbol indicates that the date of manufacture is as indicated adjacent to the symbol.
This symbol indicates the connection point for
a footswitch.
2.1.3Fig. 2.3 identifies the various parts of the T20-m
base.
2.2Symbols and graphics
To enable an easy reference to all the symbols and graphics
used on the T20 Series of tables (and within these
‘Instructions for use’) the following grouped sections show
all the symbols and graphics used.
2.2.1 Symbols general
The following symbols are shown on various parts of the
table, handset or Serial Number Plate.
IPX 4 This symbol (splash proof) denotes that the
equipment (the table) meets the requirements of IEC529
for protection from splashing water.
IPX 6 This symbol (protection against heavy seas) denotes
that the equipment (the handset) meets the requirements
of IEC529 in that water from heavy seas or water projected
from powerful jets shall not enter in harmful quantities.
This symbol indicates that the equipment is for
use on alternating current.
This symbol indicates that fuses adjacent to the
symbol have a rating and type as detailed.
This symbol warns the user to read the accompanying documents, these ‘Instructions for use’.
Symbols and near the main table ‘on/off’
switch, indicate ‘OFF’ and ‘ON’ respectively.
With the mains cord attached the equipment has
‘Class II’ protection against electric shock.
The patient leakage current, with mains voltage
on the applied parts, meets the requirement for
type BF medical electrical equipment and are
defibrillator proof.
This symbol indicates the connection point for
the corded handset.
This symbol indicates that the table section
to which it is applied (e.g. head section, under
the mattress) should not be used as a seat.
This symbol indicates the ‘Safe working load’
of the section to which it is applied can safely
support an evenly distributed load to the value
indicated, in this example 25kg.
This symbol indicates the ‘Minimum breaking
load’ of the section to which it is applied. An
evenly distributed load (in this example 100kg
or greater) may break the section.
This symbol on the table base indicates the
table should not be loaded above 300kg,
nor moved with a load above 135kg for the
T20-a and T20-s tables, or 200kg for the
T20-m table.
This symbol is used to identify the standby
113150-01
battery switch.
This graphic (T20-m table only) adjacent to the
mains socket, identifies the relationship between
the colour of the mains ‘on’ LED and the battery
charge state, see section 5.3.1.
This graphic (T20-a and T20-s tables
only) adjacent to the mains socket,
identifies the relationship between the
colour of the mains ‘on’ LED and the
battery charge state for the main and
standby batteries, see section 5.3.1.
This symbol is used to indicate the table’s duty
cycle which is the ratio of the operating time to
the sum of the operating time and the ensuing
interval.
P8/62T-IM114b
T20
Series
OPERATION TABLE
2.2.2 Handset button symbols
The following symbols are shown on the handset buttons
to indicate their function. Use of the handset is fully detailed
in section 5.3.3 of this manual.
Trendelenburg - Press to rotate tabletop in the
Trendelenburg (head down) direction.
Reverse Trendelenburg - Press to rotate
tabletop in the Reverse Trendelenburg (head
up) direction.
Height down - Press to move tabletop down.
Height up - Press to move tabletop up.
Tilt - Press to tilt tabletop down on the left (when
viewed from the long trunk end)
Tilt - Press to tilt tabletop down on the right
(when viewed from the long trunk end)
2.2.3 Handset graphics
This graphic is shown on the corded
handset to indicate the battery charge
level for both the main and standby
batteries, see section 5.3.1.
2.2.4 Standby control panel button symbols
The following symbols are shown on the standby control
panel buttons, indicating the function they select. The
arrows (upper or lower) indicate the direction the selected
function will move, if the corresponding direction button
(i.e. upper or lower) is pressed, see section 5.3.4.
Button selects Trendelenburg function.
Button selects Break function.
Button selects Height function.
Break down - Press to move the break down
(i.e. short trunk moves down w.r.t. long trunk)
Break up - Press to move the break up
(i.e. short trunk moves up w.r.t. long trunk)
Traverse Cranially (towards the head) - Press
to move the tabletop Cranially (i.e. tabletop
moves in the long trunk direction).
Traverse Caudally (towards the feet) - Press
to move the tabletop Caudally (i.e. tabletop
moves in the short trunk direction).
Flexion - Press to move tabletop into Flexion.
Extension - Press to move table into Extension.
Return to Level - Press to return tabletop
to a preset level position.
Button selects Tilt function.
Button selects Traverse function.
Direction button - Press button to obtain movement
indicated by upper arrow of function button.
Direction button - Press button to obtain movement
indicated by lower arrow of function button.
T-IM114bP9/62
View from Head end
2
1
9
3
4
8
7
5
6
10
9
1
2
3
4
9
9
6
5
12
1Head section
2Long trunk section
3Short trunk section
4Leg section
5Leg section gas spring
6Short trunk section release button
7Accessory side rail (can also be used to stow handset*)
8Long trunk section release button
9Head or leg section end block (use to stow UK/EURO handset)
10Head section release bar
11Head section gas spring
12Leg section release bar
7
View from Leg end
11
8
* Clip on handset must match side rail type, see accessory section 4.8
Fig. 2.1 Identification of the main parts of the T20 Series table top.
P10/62T-IM114b
5
T20
OPERATION TABLE
7
8
Series
T20
T20-a
T20-s
3
1
2
4
Detail ADetail CDetail B
3
12
A
10
3
11
12
B
11
B
C
6
9
13
10
10
11
11
A
C
10
1Standby battery switch
2Battery fuses
3Connection socket for Eschmann handset (ONLY)
4Table ‘On’ LED (green). Bright or Dim see section 5.3.1)
5Main table ‘On/Off’ switch (
6Connection socket for Eschmann footswitch (ONLY)
7Mains ‘on’ and Main Battery charging state LED. Red, Amber or Green.
8Mains ‘on’ and Standby Battery charging state LED. Red, Amber or Green.
9Connection socket for mains (ONLY use Eschmann mains cord supplied)
10Castor foot pedal
11Wheel foot pedal
12Standby control panel
13Mains fuses
= Off, = On)
Fig. 2.2 Identification of the main parts of the T20-a and T20-s table base and column.
T-IM114bP11/62
3
1
4
2
1Standby battery switch
2Mains ‘on’ and Battery charging state LED. Red, Amber or Green.
3Connection socket for mains (ONLY use Eschmann mains cord supplied)
4Foot pedal
5Mains fuses
Note: Castor covers (spats) are supplied as an optional item (see section 1.3.12)
The following items on the column are identical to the T20-a table:
Connection socket for Eschmann handset (ONLY) - see item 3, Fig. 2.2
Connection socket for Eschmann footswitch (ONLY) - see item 6, Fig. 2.2
Standby control panel - see item 12, Fig. 2.2
Main table ‘On/Off’ switch - see item 5, Fig. 2.2
Table ‘On’ LED (green) - see item 4, Fig. 2.2
Fig. 2.3 Identification of the main parts of the T20-m table base and column.
5
Use webbing straps of minimum width 100mm
and a minimum breaking load of 1000kg.
Position adequate padding in the areas indicated
(on both sides) to avoid damaging the table covers.
T20-a & T20-s Position straps inside,
but as close as possible to the castors.
T20-m Position straps
inside castors as indicated.
Fig. 3.1 Lifting the T20 Series operation table
P12/62T-IM114b
T20
Series
OPERATION TABLE
3.0 INSTALLATION
3.1General
3.1.1In the U.K. the table is delivered un-crated with
the head and leg section fitted. The mattress set is boxed
individually and placed on the trunk sections. The handset,
mains cord and literature are supplied loose. Any
accessories ordered with the table will be packed
individually. Some of the following unpacking and assembly
information is not therefore applicable to U.K. customers.
For overseas markets the table is usually packed in a
container with the head and leg section fitted, together with
a boxed mattress set a mains cord and the literature. The
handset is packed within the container in an antistatic bag.
Other accessories are usually packed separately, but some
may be included in the main case and should be unpacked
and stored separately during the table installation.
3.1.2When delivered packaged, carefully remove the
T20 operation table (having first removed any accessories
and packing restraints from within the container) from the
packing case as follows:-
iRemove the walls of the case leaving the table on
the base still in the braked position as packed.
Remove any chocks from the pallet base to enable
later table movement (iv below).
iiPosition the ramp provided (in the packing case)
adjacent to the base of the case.
iiiFollow the instruction provided in section 5.1 of this
manual and place the table base into the ‘Wheel’
orientation or mode.
ivWith at least two people to support the table’s weight,
push it (do not pull it) off of the pallet base and down
the ramp.
vNote that the table should not be wheeled over rough
ground, always use a trolley until a smooth floor area
has been reached.
viShould it be necessary to lift the operation table refer
to section 3.2 where suitable lifting points and
methods are detailed.
3.1.3Unpack the mattress set and fit a mattress to
each tabletop section as detailed in section 5.2.6.
3.1.4Any packaging materials should be recycled or
disposed of in accordance with current legislation.
3.1.5The T20 Series of table are powered by internal
rechargeable batteries which are connected and charged
before delivery (see disposal note in section 8.6.3). The
tables also have standby batteries. An internal mains
powered battery charger is incorporated in the table’s base,
to charge both the main batteries and the standby batteries.
CAUTION
It is most important that fuses of the correct
type, size and rating are installed (see Technical
Data, section 9.5.4).
3.1.6The T20 Series of table require a mains electrical
supply corresponding to the voltage shown on the Serial
Number Plate located on the table base. Only use the
Eschmann mains supply cord provided with the table. If
the plug supplied fitted to the cord is not suitable it should
be replaced with a suitable plug wired as below. If the plug
is a fused type, a 10A fuse must be fitted. The mains supply
cord must always be wired as follows:
Brown internal cord to LIVE
Blue internal cord to NEUTRAL
Green internal cord to Earth
(Note: The T20-a, T20-s and T20-m tables are
Class II, Type BF, there is no EARTH connection
through to the table)
3.1.7The table batteries should be recharged (see
section 5.3.1) and the table operated through the cycle of
movements detailed below to check and ensure correct
function, before the tables are first used.
Full Trendelenburg / reverse Trendelenburg
Maximum to minimum height
Maximum tilt, left and right
Maximum traverse, caudally and cranially
Maximum to minimum break
CAUTION
To complete and maintain the antistatic
pathway the table must be used on an
electrically conductive or antistatic floor and
with mattresses supplied by Eschmann
Equipment.
3.1.8The table has an antistatic pathway from the
tabletop, through an internal resistor, to the castors. To
complete the antistatic pathway, the table must be used
on an electrically conductive, or on an antistatic floor (also
see the warning in section 6.1).
3.1.9As with all medical electrical equipment care
should be taken with regard to electromagnetic
compatibility (EMC) during installation. These instructions
are written in line with the latest international standards
(EN 60601-1-2:2001) and are designed to minimise the
risk of electromagnetic compatibility issues. The T20 Series
of table should be installed and put into service in
accordance with the EMC information provided in the
Technical Data section of these ‘Instructions for Use’
(Section 9.11).
3.1.10The table should be cleaned and disinfected prior
to its first use as detailed in section 8.2 and 8.3 and then
commissioned in accordance with any local procedures
applicable to new equipment, this should include staff
training. Eschmann supply a range of wall charts with the
table, additional training aids and on-site training can be
arranged, contact Eschmann for more information.
T-IM114bP13/62
3.2Lifting the operation table
WARNING
The table is heavy and at least four strong
people are required to lift it. Ensure that
adequate precautions are taken (e.g. wear
protective shoes, use the correct straps).
3.2.1The T20 Series of table should only be lifted as
a last resort. Ideally it should be placed on a trolley directly
from the delivery vehicle, or moved on the base of the
delivery packing case (overseas only) and then rolled down
the ramp provided.
3.2.2If required the T20 table should only be lifted by
placing suitable webbing straps underneath the table base
in the positions indicated in Fig. 3.1 (having placed the
T20-a and T20-s tables into their ‘castor’ orientation to
increase ground clearance) and observing the notes that
follow. The table should only be lifted the minimum amount
required and not carried. Lift the table sufficiently high to
allow a fully decked pallet to be slid underneath. The table
should be placed into its ‘braked’ orientation whilst on the
pallet. Transport the table on the pallet using a forklift truck
or similar equipment ensuring the table is strapped securely
to the pallet.
3.2.3Extreme care should be taken to pad the straps
where they pass the base covers and the table top sections
to avoid damage. Take special care not to cause damage
to the lower edge of the base covers.
3.2.4Before lifting remove all tabletop sections,
accessories and mattresses, to minimise the weight to
approx. 248kg (T20-a and T20-s tables) or 196kg (T20-m
table). Place the table top into a level plane in both
directions (i.e. tilt and Trendelenburg) and traverse the top
so the long trunk is central to the column as shown in
Fig. 3.1. Lower the short trunk and then the table top to
their maximum limits.
3.2.5When lowering the table after the lift take care
not trap feet under the tables base. Inspect the table for
any signs of damage and check all functions prior to placing
the table back into service.
3.3Technical
3.3.1The following sections are provided for the user
to note prior to using a T20 Series of operation table.
3.3.2The T20 Series of table meets the requirements
of international standards (see section 9.8) and conforms
dimensionally to meet most requirements, for the full table
technical specification details refer to the Technical Data,
section 9.0.
3.3.3The table should only be used on an antistatic
floor and is classified as type ‘BF’ (i.e. the table has isolation
from earth equivalent to that of type ‘BF’ equipment when
the mains cord is attached).
3.3.4The antistatic properties of the table depend upon
the use of the recommended mattresses (i.e. Eschmann
antistatic mattresses ONLY) also see section 3.1.8.
3.3.5Only Eschmann accessories listed in this manual
should be used on the table and in accordance with the
‘User Handbook’ supplied with the accessory. Accessories
available from Eschmann are listed in section 7.0. Other
accessories, especially those that could compromise table
stability, must not be used. Use of other equipment with
T20 tables should only be considered after evaluating the
safety of the patient and personnel. For accessories that
fit onto the side rail ensure that they are compatible with
the side rail fitted to the table. These tables can be supplied
with standard UK, Euro USA or Denyer style side rails,
see section 1.1.2
3.3.6Provision of a diagnostic port within the table
enables access for reprogramming the tables software,
down-loading of fault information and service functions.
Use of this port MUST be limited to trained service
personnel only, and should only be used in accordance
with the correct Eschmann manuals (see 3.3.9 below).
3.3.7The table has four fuses (two on the T20-m) that
the user has access to in the event of failure. The position
of these are shown in Fig. 2.2 and 2.3 and are replaced as
detailed in section 5.3.5.
3.3.8Eschmann can provide customers with manuals
(see 3.3.9 below), for use by them in maintaining their own
equipment. These manuals contain schematic diagrams,
component part lists, descriptions and calibration
instructions which will assist the customer’s Eschmann
trained personnel to service the equipment or replace parts
(which should only be obtained from Eschmann).
3.3.9The following manuals are applicable to the T20
Series of operation table and their accessories (the part
number is in brackets following the manual reference) they
are available to order, see inside front cover for contact
details:-
T-SM52 (113175) - Service manual
T-IPL40 (113177) - Illustrated parts list
T-IM94 (111012) - Application software manual
T-IM56 (698907) - General accessory leaflet
Note: Some accessories are provided with their
own ‘User/Service Handbooks’.
3.3.10Appendix 1 provides a log that can be used to
record those people trained in the safe use of this operation
table. It is suggested that this is used to ensure that ALL
personnel using this table, are aware of all the warnings
and cautions contained within these ‘Instructions for use’.
3.3.11The T20 Series of table and their accessories,
as listed in these ‘Instructions for use’, do not contain
‘Latex’.
P14/62T-IM114b
T20
Series
OPERATION TABLE
4.0 SAFETY NOTES & CAUTIONS
4.1Warnings
The warnings and cautions that follow must be followed,
they are repeated as applicable within the text of these
‘Instructions for use’ to emphasise their importance. Ensure
you are familiar with them before using the table.
WARNINGS
The T20 Series of table have been designed to
minimise the possibility of accidental
electrosurgery burns. Contact with any metal
surfaces (e.g. table side rail, or other equipment
etc.) can cause burns during electrosurgery
and must be avoided.
The T20 tables are not rated as AP or APG and
should not therefore be used in the presence
of explosive gases.
The T20 tables have been designed for patients
weighing up to 300kg (47 stone) with their centre
of gravity (normally the umbilicus) positioned
close to the column on the trunk sections.
However patient positioning and additional
loads from accessories can compromise table
stability and strength. Ensure that loading does
not compromise table stability or damage the
table (see graphs in section 6.4).
To comply with BS EN 60601-1:1990 some
accessories have been designed for a
maximum evenly distributed load, see the ‘User
Handbook’ supplied with each accessory.
With the table in (or during transition into) the
‘castor’ or ‘wheel’ orientation, the centre of
gravity of the patient (normally the umbilicus)
should lie no more than 200mm away from the
centre of the column. Whenever this is not
practical the table should be adequately
supported (e.g. by at least two able people).
The head and leg sections are designed to
support and position the corresponding part
of the patient’s weight only. Damage leading to
failure of the section may be caused if
excessive weight is applied. Take care when
handling these sections to avoid strain and
ensure no body parts or objects are trapped
when replacing or adjusting them.
Only use Eschmann accessories and sections
that are compatible with this table. When parts
are replaced during maintenance procedures,
ensure that ONLY parts supplied by, or from,
Eschmann Equipment are used. Alternatives,
although similar, may affect the safety of the
table. Eschmann cannot be held responsible
for service, modification or adjustments to the
equipment, when performed by other than
Eschmann accredited personnel.
Where given, descriptions and illustrations of
patient operating positions, and positioning
procedures, are for guidance only. It is the
responsibility of the operating surgeon to make
sure that the positions, and the positioning
procedures are appropriate for the operation
to be performed, and the safety of the patient.
(See section 6.0).
During any table positioning procedure care
should be taken to ensure the patient’s safety.
In particular during Trendelenburg and tilt
movements the patient should be supported
to ensure they remain secure on the tabletop.
The patient’s weight should be supported
whenever the sections are adjusted or removed
from the table during repositioning.
During ANY movement of the table or tabletop,
ensure that no part of either patient or hospital
staff, or object (e.g. drapes, infusion tubing,
diathermy connections, ECG cords etc.) can
become trapped between any moving and/or
stationary equipment, or in a pinch point,
causing injury or damage to equipment.
Particular table movements that should be
operated with care are reducing the height and
Trendelenburg, which can cause trapping
situations. Always ensure adequate slack is
available in drapes and tubing for the movement
required (e.g. maximum Trendelenburg).
When moving the table (e.g. in or out of theatre)
with a patient (maximum weight 135kg for T20-a
and T20-s, or 200kg for T20-m, see section 5.1)
always ensure that the patient‘s limbs are
secure on the tabletop to prevent crushing or
trapping them against another object, always
use cot sides (available as an accessory).
Always keep the patient under observation (e.g.
check respiratory and circulatory system and
for the possibility of pressure sores etc.) and
correctly positioned whilst on the table.
Ensure that electrical equipment connected to
the communication port (available during
maintenance procedures only) complies with
appropriate electrical safety standards. Note
that standards compliance of this product may
be affected if noncompliant equipment is
attached to the communication port.
T-IM114bP15/62
CAUTIONS
Do not place either heavy accessories, or, long
accessories that could impose high torques,
to the side rails, as this may lead to damage of
the rails.
♦Ensure that only the Eschmann mains cord supplied
with the table is used to connect the table to the mains.
♦Ensure that only Eschmann supplied parts are used
during part replacement.
DO NOT:
Do not exceed the duty cycle for any table
motor drive as detailed in the technical data
section 9.5.5.
Moving the table over soft floors (e.g. carpet
or ‘cushion’ flooring) will increase resistance
to movement compared to normal hospital
‘hard’ flooring and care should be taken.
4.2Do’s and Don’ts
Attention to the following points will prolong the life and
efficiency of the T20 Series of table and will help to avoid
the risk of accidents, or damage. Other safety notes and
warnings are also given within the text of this manual and
these should be noted during use of the table.
DO:
♦Keep these ‘Instructions for use’ close-to-hand.
♦Read these ‘Instructions for use’ carefully before
adjusting, moving or using the table.
♦Use the table on an antistatic floor to prevent
inadvertent static buildup.
♦Use only the correct Eschmann mattresses and
accessories that are compatible with the T20 Series
of table as detailed in these instructions.
♦Check that handset cables and standby controls are
not damaged before use.
♦Check that the table and its accessories are not worn
or damaged, or are in any way not suitable for the
intended purpose, before use.
♦Check that all the sections (e.g. head and leg) and
accessories are secure, and put the table base in the
‘braked’ position before use.
♦Ensure that all cables are not stretched leading to
disconnection or damage during movement or
readjustment of the operation table or patient.
♦Remove table accessories and their clamps (in
particular rotary clamps) from side rails, when they
are not being used.
♦Read and follow the instructions for cleaning, and for
the care of the table and mattresses.
♦Switch ‘off’ and disconnect from the mains electrical
supply prior to cleaning and/or disinfecting the table
and when it is not being recharged.
♦Ensure that the table and accessories are serviced at
regular intervals (every six months is the
recommended frequency) only by Eschmann trained
personnel, or by accredited agents.
♦Do not lift the table by its tabletop.
♦Do not move the table with a patient without cot sides
in place on either side of the tabletop.
♦Do not push the table over rough surfaces, use a
trolley.
♦Do not drop the table (or individual sections).
♦Do not put heavy weights on the table sections,
observe the maximum advised loading.
♦Do not put sharp objects on, or against, mattresses,
pads, or the radiographic tabletop.
♦Do not place any objects on the base covers
♦Do not drop heavy objects onto the radiographic
tabletop or base covers.
♦Do not spill oil, ether, or other fluids onto the mattresses
or the pads.
♦Do not pull the table by any of the tabletop sections,
or accessories, always push it.
♦Do not service this equipment unless you have been
trained by Eschmann.
4.3Daily ‘Before use’ test
It is recommended that a ‘Daily test’ is carried out before
using the table every day. First check that the batteries
are charged if they are not charge them (see section 5.3.1.1
and 5.3.1.2) and ensure that a suitable practice is put in
place so they are charged at the end of every shift in future.
Then check the table responds to the following movements
from the handset (each movement only needs to be brief,
one or two seconds each):
Trendelenburg / reverse Trendelenburg
Height Up / Down
Tilt Left / Right
Traverse Caudally / Cranially
Break Up /Down
Then check at least two of the above plus Trendelenburg
using the Standby control panel.
4.4Accessories
The accessories available from Eschmann for the T20
Series of table are listed in section 7.0. Use of other
equipment with the T20 tables should only be considered
after evaluating the safety of the patient and personnel.
Inadvertent use of incorrect accessories could damage the
table and lead to injury. Always ensure that the information
in the ‘Instruction’ or ‘User Handbook’ supplied with the
accessory are complied with and follow all the safety notes
P16/62T-IM114b
contained within them during use. For accessories that fit
onto the side rails ensure that they are compatible with the
side rails fitted to these tables. These tables can be supplied
with standard UK, USA or Denyer style side rails.
4.5Manual handling
WARNING
Ensure care is taken when moving adjusting
or lifting any part of the table or patient. Note
the guidelines provided.
4.5.1During adjustment or changing the configuration
of the T20 series of operation tables, there are occasions
when the user should be aware of the safe practises to be
employed during manual handling or adjustment of parts
of the table. For information, the weights of the heaviest
sections and accessories commonly used are listed in
Appendix 4. Appendix 4 also contains other information
on manual handling. When lifting, carrying or fitting these
components it is recommended that care is taken and that
two or more people are employed when required.
T20
Series
OPERATION TABLE
4.5.2These ‘Instructions for use’ advise supporting the
weight of the patient during adjustment of any section,
obviously this requires the intervention of several
personnel, some supporting the patient’s limbs and others
adjusting the table sections.
4.5.3When moving the table note that extra effort is
required to start the table moving, take care not to strain
limbs or back. The table should not be moved when heavily
loaded note the warnings and cautions provided. It is good
practice to use two or more people when moving a table.
T-IM114bP17/62
5.0 OPERATING THE TABLE
manual operations achieved by using one or both of the
foot pedals on the table base.
This section has been split into sections as follows:
5.1Moving/operating the table base.
5.2Using the removable sections.
5.3Using the table’s powered/electrical functions.
5.1Operating the table base
WARNING
Always push the table (do not pull it) at a
suitable height ensuring that it is stable at all
times, take care not to collide with personnel
or equipment. Maximum stability will be at
minimum height. Ensure that the patient is
adequately supported and restrained
(especially limbs) using cot sides as
appropriate (also see section 5.1.3). Ensure that
all connections via cord or tube, to the patient
or table, have either been disconnected, or are
only attached to equipment that will move with
the table to avoid inadvertent disconnection.
When changing from ‘castor’ or ‘wheel’ orientation, to ‘braked’, ensure that no objects (e.g.
cords, tubing etc.) can become trapped beneath
the table base and floor (T20-a and T20-s tables
only). Do not move the table with a patient
weighing more than 135kg (T20-a and T20-s
tables) or 200kg (T20-m table) and ensure
stability is maintained.
If table is difficult to move check for objects
under castors and that castors are maintained
and kept clean and free from foreign objects
(also see section 1.3.12).
The table should always be left in the ‘braked’ orientation
with both pedals in their raised positions. Do not leave the
table with the wheel pedal up and the castor pedal down.
When familiar with sections 5.1.1.1 - 5.1.1.2, use Fig. 5.3
as a quick reference guide for pedal operation.
5.1.1.1‘Wheel’ orientation (from ‘braked’)
T20-a wheel pedal operation is easily achieved if the pedal
is pressed down with the right foot whilst steadying yourself
with hands on the tabletop. Stand on the side of the table
on which the footpad is located. T20-s wheel pedal operation
is easily achieved if you steady yourself with your hands
on the tabletop, note that you can stand on either side of
the table to operate the pedal. For either type of pedal use
a steady ‘press’ rather than a ‘rapid depression’ of the pedal,
this provides easier identification of the ‘click’ and avoids
pressing the pedal passed its locking position.
To place the table into the ‘Wheel’ orientation from the
‘braked’ orientation, press the wheel pedal (item 11, Fig. 2.2)
down steadily until you hear an audible ‘click’ (action
indicated in Fig. 5.1a or 5.1b). This ‘click’ indicates that the
pedal has locked in the down position. Do not continue to
press the pedal after the ‘click’ as this will release the internal
catch and the pedal will not lock down. If this does happen
the pedal must be allowed to rise fully (this resets the internal
catch) before pressing it down again.
When placed into the ‘Wheel’ orientation the table base is
supported on two wheels at the short trunk end and two
castors at the long trunk end. To move the table in ‘wheel’
orientation always push it (do not pull it) from the long trunk
end, moving the end nearest to you, left or right, to steer the
table in the required direction. The table is in ‘wheel’
orientation when the wheel pedal is in the lowered position
and castor pedal is in the raised position.
5.1.1.2‘Castor’ orientation (from ‘wheel’)
The table base has three formats, the T20-a and T20-s
tables (basically similar, differentiated by the design of the
foot pedal) see section 5.1.1 and the T20-m table which
has a single foot pedal, see section 5.1.2.
5.1.1 T20-a and T20-s table base
The T20-a and T20-s tables can be moved easily on builtin castors and wheels. Normally the table rests on brake
pads at the long trunk end and wheels at the short trunk
end, these provide a secure and static location on the
operating theatre floor. Lower the table to a suitable height
to achieve a stable position before moving the table.
To move the T20-a and T20-s tables they are placed onto
their wheels and castors in two ways, providing both a
‘castor’ orientation and a ‘wheel’ orientation. The latter
enables easy movement of the table in a straight line (down
a corridor for example). Moving the table into either of these
orientations does not require battery power, they are
T20-a castor pedal operation is easily achieved if the pedal
is pressed with the left foot whilst steadying yourself with
hands on the tabletop. Stand on the side of the table on
which the footpad is located. T20-s castor pedal operation
is easily achieved if you steady yourself with your hands
on the tabletop, note that you can stand on either side of
the table to operate the pedal. For either type of pedal use
a steady ‘press’ rather than a ‘rapid depression’ of the pedal,
this provides easier identification of the ‘click’ and avoids
pressing the pedal passed its locking position.
To place the table into the ‘castor’ orientation from the ‘wheel’
orientation, press the castor pedal (item 10, Fig. 2.2) down
steadily until you hear an audible ‘click’ (action indicated in
Fig. 5.1a and 5.1b). This ‘click’ indicates that the pedal has
locked in the down position. Do not continue to press the
pedal after the ‘click’ as this will release the internal catch
and the pedal will not lock down. If this does happen the
pedal must be allowed to rise fully (this resets the internal
catch) before pressing it down again.
P18/62T-IM114b
T20
Series
OPERATION TABLE
When placed into the ‘castor’ orientation the table base is
supported on four castors, two at each end of the table.
This orientation enables the table to be moved in any
direction including sideways and swivelling within its own
length. To move the table always push it in the required
direction, never pull it. The table is in ‘castor’ orientation when
both pedals are in their lowered positions.
5.1.1.3‘Castor’ orientation (from ‘braked’)
To place the table into ‘castor’ orientation from the ‘braked’
orientation follow 5.1.1.1 to place the table into ‘wheel’
orientation and then follow 5.1.1.2 to complete the move
into the ‘castor’ orientation.
Note: It is not critical that sections 5.1.1.1 and 5.1.1.2 are
carried out in this sequence the reverse is equally suitable
and correct.
5.1.1.4‘Braked’ orientation
To place the table into the ‘braked’ orientation move the pedals
(items 10 and 11, Fig. 2.2) into their raised position. If either
pedal is in the lowered position raise them, in any sequence.
For the T20-a table stand on the standby control panel side of
the table and use the foot advised in 5.1.1.1 or 5.1.1.2 as
appropriate. For the T20-s table you can stand on either
side of the table. Steady yourself with your hands on the
tabletop and press the pedal ‘firmly down’ see Fig. 5.2a
and 5.2b, this disengages the internal locking catch.
Release pressure on the pedal and allow it to rise. The
table will gently lower onto its brake pads and wheels, the
motion is softened and braked by an internal damper.
The table is in ‘braked’ orientation when both pedals are in
their raised positions.
Fig. 5.1a Lowering a T20-a foot pedal
(T20-a wheel foot pedal illustrated)
Fig. 5.2a Raising a T20-a foot pedal
(T20-a castor foot pedal illustrated)
Fig. 5.1b Lowering a T20-s foot pedal
(T20-s castor foot pedal illustrated)
Fig. 5.2b Raising a T20-s foot pedal
(T20-s castor foot pedal illustrated)
T-IM114bP19/62
=
BRAKED
TABLE WILL NOT MOVE
Fig. 5.3 T20-a and T20-s foot pedal quick reference guide
5.1.2 T20-m table base
WARNING
Before using the table, position the castors
under the base covers to help protect them
from any falling contamination.
The T20-m table base has been provided with four large
castors. The table rests on these castors at all times
providing either a secure and static location on the
operating theatre floor (‘braked’), or two modes for easy
movement (‘castor’ and ‘wheel’ modes). Adjust the table
to a suitable height to achieve a stable position before
moving the table.
The T20-m table should always be left ‘braked’ with the
foot pedal in its lowest position. When familiar with sections
5.1.2.1 to 5.1.2.4, refer to the label on the table base as a
quick reference guide to pedal operation (label is illustrated
in Fig. 5.4).
To move the table it is placed into its ‘castor’ mode or ‘wheel’
mode. The latter enables easy movement of the table in a
straight line (down a corridor for example). The ‘castor’
mode provides full free wheeling mobility with 360° rotation
and movement sideways.
Adjusting the table into either of these modes does not
require battery power, they are manual operations achieved
by using the single foot pedal on the table base (see item 4,
Fig. 2.3).
5.1.2.1Foot pedal operation
Foot pedal operation is easily achieved when the pedal is
operated with either foot whilst steadying yourself with your
hands on the tabletop. Stand on either side of the table or
the pedal end, which ever is the most suitable and easy.
=
PUSH IN A STRAIGHT LINE
CASTOR
WHEEL
=
Do not operate the pedal from the end of the table when a
long table section (e.g. a leg section) has been fitted to the
long trunk end, this may require unnecessary stretching
by the operator to reach the pedal. Operate the pedal from
the side of the table.
Use a steady motion rather than a ‘rapid’ movement of the
foot pedal, this provides easy identification of the ‘snap’
into any of its three positions.
5.1.2.2‘Braked’
To place the table into the ‘braked’ orientation press the
foot pedal (item 4, Fig. 2.3) down to its lowest position.
Operate the pedal as detailed in section 5.1.2.1. and press
the foot pedal down until it snaps into its lowest position.
The pedal can be moved from its highest ‘wheel’ position
through its central ‘castor’ position and into the ‘braked’
position in one easy movement.
The table is ‘braked’ when the foot pedal is in its lowest
position.
5.1.2.3‘Castor’ mode
To place the table into the ‘castor’ mode, move the foot
pedal (item 4, Fig. 2.3) into its central position. Operate
the pedal as detailed in section 5.1.2.1 and move the foot
pedal until it snaps into its central position.
The pedal is moved up from its lowest ‘braked’ position by
lifting the pedal up with the top of the foot, or down from its
raised ‘wheel’ position by pressing the pedal down with the
ball of the foot, until the pedal snaps into the central ‘castor’
position.
The table is in ‘castor’ mode when the foot pedal is in its
central position.
Note: When the table has been moved to the required
location always leave the table ‘braked’.
PUSH IN ANY DIRECTION
P20/62T-IM114b
Press pedal down fully with the castors
in any position to brake all four castors
(see WARNING in section 5.1.2).
Lift pedal up, or press pedal down,
into the central position to enable the
table to be 'castored'. All four castors
will swivel allowing full table mobility.
Select 'castor' mode' (pedal in the central position) and
push the table until the short trunk end castors are correctly
aligned as indicated. Lift the pedal up fully into the 'wheel'
position (to stop the short trunk end castors swivelling).
This will enable the table to be 'wheeled' in a straight line.
Push and steer the table from the pedal (head) end.
A
B
Detail of the required short trunk end castor postion, which should be obtained, before
lifting the foot pedal into its top position to place the table base into the 'wheel' mode.
Note:
In position A the castor leads its mount, in position B the castor trails its mount.
WHEEL
CASTOR
BRAKE
See A above
See B above
113136-01
T20
Series
OPERATION TABLE
Illustration of the table’s base ‘quick
reference guide’ label. Note the
booklet symbol that indicates
reference to these instructions
should be observed.
Fig. 5.4 Operating the T20-m table base
T-IM114bP21/62
5.1.2.4‘Wheel’ mode
WARNING
Do not push the table in the ‘wheel’ mode until
you are sure the castors have moved into their
correct orientation for the ‘wheel’ mode, as
detailed in the second stage below and Fig. 5.4.
If the table is pushed with the castors locked out
of position this will cause undue wear leading to
failure of the short trunk end castors.
Placing the table into the ‘wheel’ mode is a four stage
procedure which will ensure that the short trunk end
castors are locked in position correctly.
First, if the table is ‘braked’ move the foot pedal (item 4,
Fig. 2.3) into its ‘castor’ position as detailed in section
5.1.2.3.
Second, push the table forwards from the long trunk end
until both the short trunk end castors have swivelled into
the position ‘B’ shown in Fig. 5.4. They do not need to be
exactly in line, but they should not be as shown in ‘A’ Fig. 5.4
(i.e. leading their mounts).
Third, lift the pedal up with the top of the foot, as detailed
in section 5.1.2.1, until the pedal snaps into its highest
‘wheel’ position.
Fourth and finally, continue to push the table in a straight
line, the short trunk end castors will lock automatically in
line with the table base. This enables the table to be moved
easily down a corridor, being steered from the long trunk
end.
The table is in ‘wheel’ mode when the foot pedal is in its
highest position and the short trunk end castors have locked
in-line with the table base.
Note: When the table has been moved to the required
location adjust the table into ‘castor’ mode to enable full
mobility and positioning and then leave the table ‘braked’
(press the pedal fully down).
5.1.3 Moving table with a patient
The normal table position when moving it with a patient is
level in both directions (i.e. tilt and Trendelenburg) and
with the patient in the supine or lateral recovery position.
Fig. 5.5 shows an alternative position that can be used,
the notes in Fig. 5.5 also apply when moving the table with
the patient in the supine or lateral recovery position.
Notes:
Always use cot sides.
Lower table to a suitable height.
Ensure patient limbs are secure.
Check patient weight and position complies with Fig. 6.3 and Fig. 6.4.
Always push table from the head end (do not pull).
Position also applies to T20-a and T20-s tables (T20-m table illustrated).
Fig. 5.5 Alternative table position for moving table with a patient
P22/62T-IM114b
T20
Series
OPERATION TABLE
5.2Using the removable sections.
WARNING
Ensure that nothing becomes trapped
(e.g. fingers, tubing, cords) when attaching
tabletop sections. When removing a section be
prepared to support the full weight of the
section when the guide pins disengage. Always
carry the section holding the side rails. Never
hold or pick up the section using the black
gas support struts and take care not to operate
the release handle accidentally. Do not operate
the release handle when the section is removed
from the table. For users of small stature, when
handling larger sections (e.g. the leg section)
two people should work together to avoid strain
injuries. Also to minimise weight for all users
remove mattresses before fitting or removing
a section. Users of earlier Eschmann tables
(e.g. MR and RX Series) should note that T20
guide pins are shorter and disengage earlier.
Always ensure that the sections have been
correctly and securely fitted before use and
only use the correct Eschmann sections.
To attach the section to a table after re-aligning the pins
insert the pins into the tabletop for 90% of their length,
actuate the release handle and adjust the section until
horizontal and then push the section fully home until the
locking catches engage (also see section 5.2.2).
During long term storage of a section it should be positioned
with the pins up, this ensures continued lubrication of the
gas spring seals (e.g. attached to the table and fully
lowered). If stored separately from the table take care not
to actuate the release handle during storage since this will
alter alignment of the pins. Should this happen inadvertently
see section 5.2.1. For short term storage and to aid manual
handling (e.g. during section reversal) use of the ‘T’ series
Head/Leg accessory trolley (REF TA-040013) is
recommended.
5.2.1 Re-aligning a section’s pins
Realignment of a section’s pins is only required if they have
become mis-aligned whilst the section has been removed
from the table (see Fig. 5.6 for illustration of aligned and
mis-aligned pins).
WARNING
Take care when realigning a sections pins. Ensure
that fingers are clear of the gas springs during
actuation of the release handle.
To realign the pins of a table section place the section on
its side (see Fig. 5.6) and support it by its upper side rail
only keeping fingers etc. well clear of the gas springs
(see 1, Fig. 5.6). Then actuate the release bar (see 2,
Fig. 5.6) until the gas springs have moved both pins to the
ends of their travel as shown (see 3, Fig. 5.6).
Fig. 5.6 Section pin alignment
5.2.2 Attaching a removable section
When attaching a section first check to see if the pins are
aligned with each other. It is possible that during storage
the release handle may have been actuated and the pins
are no longer aligned as shown in Fig. 5.6. If they are aligned
attach the section as detailed below, if they are not, align
them and attach the section as detailed in section 5.2.1.
To attach the removable tabletop sections (i.e. head or
leg) hold the section firmly with two hands aligning the pins
of the section (item 1, Fig. 5.7) with the location holes
(item 2, Fig. 5.7) in the fixed tabletop section as shown in
Fig. 5.7.
Fig. 5.7 Aligning section pre-attachment
T-IM114bP23/62
Gently insert the section pins into the tabletop and slide
the section into the tabletop evenly until the locking
catches engage with an audible click. The section will
slide in easily if the weight of the section is gently
supported with both hands. Take care not to trap anything
(especially fingers) between the section and the tabletop.
Check that the section is fully inserted and the safety catches
have engaged by gently pulling on the section. Both release
buttons (items 6 and 8, Fig. 2.1) should be out when the
section has been correctly located and locked.
5.2.3 Removing a tabletop section
When removing a tabletop section (e.g. a head or leg
section) ensure that it is horizontal before removal as
this will aid later replacement (pins will be aligned with
the section). If the section is not horizontal adjust it to
horizontal before removal (see section 5.2.4).
Remove the section from the tabletop by supporting the
section’s weight and pressing in the right-hand trunk
section button (viewed from end of table) as indicated by
1 in Fig. 5.8. The button will stay in when pressed correctly.
Press in and hold in the left-hand trunk section button
(viewed from end of table) as indicated by 2 in Fig. 5.8
whilst pulling the section evenly away from the fixed
section about 2 - 3cm (see 3, Fig. 5.8). When employing
two people one should press the button whilst the other
pulls the section out the 2 - 3cm. Gently supporting the
weight of the section (i.e. slightly lifting whilst pulling)
will make this action easier. Then release the left-hand
trunk section button.
Now using both hands (one on each side of the section
as shown in 4 of Fig. 5.8) continue to gently pull the section
out evenly from the trunk section until the guide pins are
free. Again gently lifting and supporting the weight of the
section will make this action easier. When employing two
people, one should hold the second button in whilst the
other pulls the section out evenly, both can then support
the weight when the pins are free.
Fig. 5.8 Removing a table section
When raising the section the gas springs will assist the
movement, when lowering the section gentle pressure may
be required to overcome slight resistance of the gas
springs. Adjust the section by rotating it up or down to the
required position (see 2, Fig. 5.9). When the correct position
has been obtained release the locking bar. The section is
now locked in place automatically by the gas springs (see 3,
Fig. 5.9).
5.2.4 Removable section adjustment
The head and leg sections are operated in the same way.
The section’s weight is supported during adjustment and
held locked in position by two gas springs (items 5 and
11, Fig. 2.1). Lowering the section’s release handle during
adjustment automatically locks the section in place.
To adjust either section (having noted the warnings at
the beginning of section 5.2) grasp the end of the section
on both sides with thumbs up and the finger tips resting
on the release handle (items 10 or 12, Fig. 2.1) underneath
the section. Whilst supporting the section in place (it may
move when the release handle is raised), gently lift the
release handle with the finger tips until the section is free
to move up or down (see 1, Fig. 5.9).
Fig. 5.9 Adjusting a head or leg section
P24/62T-IM114b
T20
Series
OPERATION TABLE
5.2.5 Changing sections during a procedure
The table sections can be changed during a surgical
procedure but extreme care must be taken to adequately
support the patient throughout such adjustments to the
table configuration. Also note the warning at the beginning
of section 5.2.
When replacing a section during a procedure take extreme
care not to trap any part of the patient, surgical drapes,
surgical equipment, cords or tubing. Also note the warnings
in section 4.1.
It is suggested that the mattress is located onto the section
before fitting the section onto the tabletop when
repositioning it during a procedure (allow for the extra
weight).
5.2.6 Attaching/Removing mattresses
WARNING
Larger mattress sections can be heavier and
more bulky than normally envisaged (see
section 7.0 and the note in section 9.1). Take
care when handling to avoid stain injuries, use
two people if required.
Only use Eschmann mattresses and ensure
they are free of cuts and tears, replacing as
required. Do not use without a mattress fitted
to the table.
Mattresses are attached to the tabletop by two mushroom
shaped pins per section. These pins ensure that the
mattress stays in place on the tabletop in all table
configurations. To ensure that the mattress is correctly
located onto the pins proceed as follows:-
iThe larger mattress sections can be bulky to handle
individually if of small stature, use two people to avoid
strain.
iiEnsure the mattress is the correct one for the section
you are fitting it to and orientate it relative to the
tabletop so the mushroom pins are lined up with the
mating recesses in the underside of the mattress.
iiiLower the mattress onto the tabletop and gently ease
one location recess at a time onto and over the
respective mushroom pin. The mattress should lie
flat on the tabletop when this has been achieved
without evidence that the pin is holding the mattress
off the tabletop surface.
To remove mattresses reverse the above procedure gently
easing the mattress off the pins, one pin at a time. Ensure
minimal sideways force is applied to the mattress during
this procedure to prolong mattress life.
Always ensure that mattresses are free from any cuts and
abrasions, such damage can compromise patient safety
by promoting cross contamination. Replace worn
mattresses with new ones as soon as possible.
5.3Powered and electrical functions.
This section is split into four sections as follows:
5.3.1 Batteries and mains
5.3.2 Powered motions
5.3.3 Using the handset
5.3.4 Using the standby control panel
5.3.5 Changing fuses
Note: Whilst it is safe to use the table when connected to
the mains it is good practice to use the tables batteries to
power the table, rather than connect the table to the mains.
5.3.1 Batteries and mains
5.3.1.1Battery charging introduction
WARNING
Ensure that the mains supply used is suitable
(i.e. 100-240V, 50-60Hz) before attaching the
table to the mains supply. Only use the mains
cord supplied with the table and charge the
batteries daily.
CAUTION
Continued use of the table batteries when
‘critically low’ can damage the batteries. Charge
batteries regularly to maintain peak performance.
Do not remove the table from charge until both
charging LED’s are ‘green’ to avoid false battery
level indication on the handset.
Batteries within the table base are mains rechargeable and
should power the table continuously for 3 days normal use
(from a full charge). However to maintain peak battery
performance the table batteries should be placed on charge
at the end of each day or shift. Daily charging should be
encouraged as good practice, do not wait for either ‘low’battery indication before charging. Battery charge level
is indicated by LEDs on the corded handset, see Fig. 5.10.
The table should be placed on charge as detailed in section
5.3.1.2. Also see battery management sheet Appendix 3.
Note: Over the first few charge cycles (from new) battery
capacity increases until they reach a stable maximum level.
Batteries should be checked for adequate charge before
using the table for any surgical procedure. The corded
handset has two battery level indicators, one for the main
battery and one for the standby battery, see Fig. 5.10.
These indicators will only operate when the corded handset
is plugged into the table and the table is switched ‘on’.
Each indicator contains three coloured LEDs, these
indicate the information shown over the page.
T-IM114bP25/62
Main battery level indicator:
Green LED ‘on’ = Battery level satisfactory.
Amber LED ‘on’ = Low battery. An indication that
the remaining battery charge is only adequate for
one day’s average use. Batteries MUST be
recharged at the end of the current shift.
Red LED ‘on’ = Critically low battery. An indication
that the remaining battery charge is critically low.
Batteries MUST be recharged as soon as possible
and before the next procedure. Charging during the
current procedure is not recommended.
Standby battery level indicator:
Green LED ‘on’ = Battery level satisfactory.
Amber LED ‘on’ = Low battery. Battery level is low
and only adequate for two or three Trendelenburg
movements, they MUST be recharged at the end of
the current shift.
Red LED ‘on’ = Critically low battery. Battery level is
critically low and only adequate for one Trendelenburg
movement. They MUST be recharged as soon as
possible and before the next procedure. Charging
during the current procedure is not recommended.
Handset (battery charge level)
5.3.1.2Battery charging
To recharge the table batteries (normal and standby)
connect the mains cord supplied with the table (do notuse any other mains cord) into the mains inlet (item 18,
Fig. 2.2 for T20-a and T20-s tables, or item 3 of 2.3 for
the T20-m table).
Connect the mains cord to a suitable mains supply (check
mains voltage is 100-240V, 50-60Hz) and switch the supply
‘on’ if controlled by a switch.
The battery charging LEDs (items 7 and 8 of Fig. 2.2 for
T20-a and T20-s tables, or item 2.3 for the T20-m table)
will illuminate ‘red’ to indicate ‘bulk’ recharge, ‘amber’ to
indicate that the batteries are in a 2 hour ‘top-up’ charge
and will illuminate ‘green’ to indicate a fully charged battery
on ‘trickle’ charge (Note. The T20-m table only has one
LED indicating both main and standby charging state).
The batteries should be fully charged within six hours (if
the table is not in use) but it is safe to leave the mains
supply ‘on’ when the batteries are fully recharged and both
LEDs are ‘green’ (or in the case of the T20-m table the
single LED is ‘green’). See CAUTION in section 5.3.1.1.
Note: Always ensure the table is switched ‘off’ to minimise
the battery charging time.
It is recommended that the mains cord is stored on the
‘T’ Series accessory trolley (an accessory) to avoid loss
and inadvertent use of an incorrect mains cord. It should
be wrapped around the cable cleats.
When the main battery ‘critically low’ warning is indicated
the table cannot be operated by either handset or the
footswitch. To continue to operate the table either connect
it to a suitable mains supply and recharge the batteries
(see 5.3.1.2) or use the standby batteries (see 5.3.1.4).
Note: The standby control panel can be used even when
the ‘critically low battery’ indication is given but note the
warning in section 5.3.4.
5.3.1.3Battery changing
Batteries in the table base should only be changed by service
personnel. They should last for many years if recharged
regularly and should not require user intervention (also see
section 8.6.3 for the caution concerning disposal).
5.3.1.4Standby batteries
WARNING
The standby batteries are for emergency use
only and will only provide power for a few
movements.
The table is provided with standby batteries to cover the
unlikely event of a main battery failure or fault. To use the
standby batteries, the standby battery switch (see item 1,
Fig. 2.2 for T20-a and T20-s tables, or item 1 Fig. 2.3 fot
T20-m table) must be pressed and held depressed whilst
the required motions are conducted using the normal hand
and foot controls.
The standby batteries are only used when the standby
battery switch is depressed and they are automatically
recharged during the normal battery charging process.
P26/62T-IM114b
T20
Series
OPERATION TABLE
5.3.2 Powered motions
WARNING
Always ensure that the patient is secure or
adequately supported during all tabletop
adjustments and that such adjustments do not
compromise table stability.
If a patient is on the table and the handset is
clipped onto the tabletop, take care that the
handset buttons are not inadvertently activated
(e.g. by a patient Transporter or person) causing
unwanted table movements.
Similarly take care at all times to ensure that the
handset is not inadvertently activated.
During all tabletop adjustments be aware of
pinch points between the moving and static
parts (e.g. the break hinge and head and leg
hinges).
Ensure that all theatre drapes, electrical cords
and any medical tubing etc. are clear from
entrapment in pinch points and that adequate
excess drape, cord, or tube, are available to cater
for the adjustment required.
Ensure that any motion and adjustment
selected will not cause any part of the table to
hit or collide with any person or object.
Examples being:-
i)Leg section hitting floor when table is
low and reverse Trendelenburg selected.
ii) A fully lowered leg section hitting table
base when tabletop is lowered.
iii) Head section hitting anaesthetist’s seat
when Trendelenburg selected.
CAUTION
Do not exceed the duty cycle for any motor
drive as detailed in section 9.5.5.
5.3.2.1General
The major tabletop adjustments and motions are powered.
These powered motions are controlled by a corded handset
supplied with the table (see 5.3.3) and the standby control
panel (item 12, Fig. 2.2 see section 5.3.4). An optional
infrared handset or an optional footswitch (see accessory
list section 7.0) can also be used to control the tabletop.
The use of these accessories is explained in the leaflet
provided with them.
Powered tabletop motions are, Trendelenburg, height, tilt,
brake, traverse, flexion and return to level. Each motion in
either direction is controlled by pressing and holding the
appropriate handset, footswitch or standby control panel
button. The table must be switched ‘on’ for powered
motions to operate. Section 5.3.3.2 details handset button
functions, section 5.3.4 provides standby control panel
information.
All signals from controllers will be treated in a strict priority
order to avoid multiple activation errors from different
controllers. Inputs will be responded to in the sequence
below:
1Standby control panel
2Corded handset
3Infrared handset
4Footswitch control.
Button presses from a controller lower on this list will be
ignored if any button is pressed on a controller higher on
the list and the input from the higher controller will be
actioned.
The table cannot be powered by footswitch, or either
handset, when the main battery ‘critically low’ indication is
displayed (see Fig. 10, main battery LED is ‘red’) unless
the standby battery switch (item 1, Fig. 2.2 for T20-a and
T20-s tables, or item 1, Fig. 2.3 for the T20-m table) is
held depressed. However the tabletop can be controlled
by the standby control panel (item 12, Fig. 2.2) situated on
the side of the column. This standby control panel has
limited functionality in comparison to either handset and
should ONLY be used in an emergency (e.g. handset or
footswitch failure, control system failure, or ‘critically low’
battery).
Use of this standby control panel in any situation other
than an emergency SHOULD be avoided as certain safety
devices are overridden. The controls and buttons of the
standby control panel are detailed in section 5.3.4. In the
unlikely event of insufficient battery power when using the
standby control panel, press and hold the standby battery
switch (item 1, Fig. 2.2 for T20-a and T20-s tables, or item
1, Fig. 2.3 for the T20-m table) whilst conducting powered
motions.
In the unlikely event of failure of the main table batteries,
‘press and hold’ the standby battery switch (item 1, Fig. 2.2
or 2.3) in the ‘on’ position and use the corded or infrared
handset or the standby control panel to control the table.
This enables the standby batteries to power the table,
however the main battery fault should be rectified as soon
as possible.
T-IM114bP27/62
5.3.2.2Tabletop motions
5.3.3 Using the handset
WARNING
Ensure you have read and understood the
safety warnings listed in section 4.1 and 5.3.2
before using any of the powered motions.
All powered tabletop movements are programmed to
provide a gentle transition from stationary into the required
motion and back again to stationary (i.e. initially the motion
starts slowly and then speeds up to maximum speed and
then slows again before coming to a stop). This also
enables accurate positioning at slow speed by using
repeated short button presses. Alternatively long duration
button presses allow large changes in tabletop position at
maximum speed.
To operate the tabletop powered motions the table must
be switched ‘on’ ( ) at the table on/off switch (item 5,
Fig. 2.2). The table will emit a single ‘beep’ and the green
LED (item 4, Fig. 2.2) will illuminate to show the table is
‘on’. The LED will be bright during use but only dim if the
operating system has gone into ‘sleep’ mode (saving
battery power). The table will instantly respond to any
control input even when in ‘sleep’ mode. Also see section
8.1.
If two buttons are pressed on a handset or footswitch at
the same time neither will have any effect. The function of
the second button pressed is ignored and automatically
cancels the function of the first button pressed and any
table motion stops. Releasing either of the buttons will
enable the function of the remaining button (if still pressed)
to be actioned. This is to eliminate multiple button activation
errors.
Most powered tabletop motions will pause briefly when
they pass through the level position (Note: not applicable
to height and traverse). This is to enable each motion to
be returned to a level position individually.
Be aware that a lowered leg section could hit the table base
when the top is lowered, or the head section could collide
with an anaesthetist’s chair during a Trendelenburg
movement.
5.3.3.1General
WARNING
Do not plug two handsets into the table at the
same time. When two handsets are connected,
neither handset will operate the table.
The corded handset will operate any T20 operation table
(note that the infrared handset is configured to match a
specific table). Input from the corded handset will override
any signals from an infrared handset. The corded handset
simply plugs into either of the table handset sockets (item 3
of Fig. 2.2 and item 2, Fig. 5.11). Never plug handsets into
both sockets at the same time. See section 5.3.3.2 for
details of button function and section 5.3.1.1 for an
explanation of the battery state LEDs.
When using the corded handset always ensure that the cord
is well clear of any moving parts, pinch points and possible
entrapment from table movements. Also ensure that the cord
will cater for any table movement and that such table
movements do not stretch the cord excessively.
When not in use it is suggested the handset is clipped onto
the accessory side rail (item 7, Fig. 2.1). Note that when
clipped onto the table in this way it can be operated with
one finger, without the need to hold it in the other hand. The
handset must be compatible with the side rail fitted to the
table, see sections 1.1.2. The UK handset will also clip onto
head/leg end blocks (item 9, Fig. 2.1).
To remove the handset plug from its socket the release
button (item 1, Fig. 5.11) must be pressed and held in during
removal. When removing the corded handset plug from its
socket always grip and pull the plug, do not pull the cord
only as this may damage the cord or internal connections.
Handset buttons provide a tactile feedback to enable the
user to detect when a button has been pressed or released,
this is in the form of a ‘pop’ or ‘click’. Also, all buttons are
‘de-bounced’ to ensure that only deliberate button presses
are responded to (i.e. an accidental quick activation is
ignored).
Always ensure that there is adequate space around the
table for the movement selected and that the movement
required will not cause injury to patient or medical
personnel. Look for possible trap and or pinch points
between parts of the table and stationary objects.
Fig. 5.11 Controller sockets and
release buttons
P28/62T-IM114b
T20
Series
OPERATION TABLE
5.3.3.2Handset button functions
Handset buttons have their function indicated by graphics
printed on the handset, see Fig. 5.12.
To activate the required tabletop motion press and hold
the appropriate button until the required amount of change
in tabletop position has been achieved, then release the
button. The actual motion for each button is fully detailed
below, refer to Fig. 5.12 for button reference.
Note: Buttons B1 and B2 provide the described movement
for a table top in the normal orientation (head section in
long trunk) with the patient’s head on the head section. If
the patient orientation is reversed button functions B1 and
B2 are also reversed.
B1 Trendelenburg (orange button). Press to
rotate the tabletop in the Trendelenburg
(head down) direction.
B2 Reverse Trendelenburg (orange button).
Press to rotate the tabletop in the
Reverse Trendelenburg (head up)
direction.
B3 Height down. Press to lower the whole
tabletop.
B11 Extension - Press to move table into
Extension.
B12 Flexion - Press to move table into
Flexion.
B13 Return to Level - Press to return the
tabletop to a ‘preset’ level position.
This may invoke all powered
movements to achieve the ‘preset’
level position. The sequence of movements is tilt,
Trendelenburg and Break together, a short pause and
then maximum traverse cranially. (Note: Height is not
affected).
Tilt, Trendelenburg and Break motions will pause
momentarily when passing through a ‘level’ position.
Note that ‘Break’ is level when the trunk sections are inline with each other, they will not be horizontal unless
Trendelenburg is horizontal.
B4 Height up. Press to raise the whole
tabletop.
B5 Tilt left - Press to tilt the whole tabletop
down on the left-hand side when viewed
from the long trunk end.
B6 Tilt right - Press to tilt the whole tabletop
down on the right-hand side when viewed
from long trunk end.
B7 Break down - Press to move the tabletop
in the break down (Extension) direction.
B8 Break up - Press to move the tabletop in
the break up (Flexion) direction.
B9 Traverse Cranially - Press to move the
tabletop Cranially (i.e. the tabletop will
traverse and increase the distance
between the end of the long trunk section
and the column).
B10 Traverse Caudally - Press to move the
tabletop Caudally (i.e. the tabletop will
traverse and decrease the distance
between the end of the long trunk section
and the column).
T-IM114bP29/62
T20-a / T20-s Serial label (end of table base)
T20-m Serial label (on trunk section)
SN T20A1JI090
REF T20-1-3-1101
2005-08
Fig. 5.12 Handset button functions and serial labels
(labels are examples only and text on actual table will be different to that shown above)
P30/62T-IM114b
T20
Series
OPERATION TABLE
5.3.4 Using the standby control panel
WARNING
The standby control panel MUST be used with
extreme care for Trendelenburg movements. All
programmed safety features are overridden in
this mode (e.g. should the tabletop hit an
object, motor protection is inhibited and
damage to them could occur). Also do not
exceed 30° of Trendelenburg (or reverse) from
the standby panel.
The standby control panel is located on the side of the
column (item 12, Fig. 2.2). The panel has five function
buttons (OB1 to OB5, see Fig. 5.13) and two direction
buttons (OB6 and OB7, see Fig. 5.13). The motion of the
function button is described by its graphical symbol
(symbols are shown and detailed in section 2.2.4).
When controlling the tabletop from this standby control
panel certain inbuilt safety features are overridden. Also
there is no ‘soft start’ to the powered Trendelenburg
movements and the tabletop will not pause as it passes
through the level (Trendelenburg) position. Only five
functions can be controlled from this panel Trendelenburg,
Height, Tilt, Break and Traverse.
OB1Trendelenburg - Press and hold this button, then
press required direction button to rotate the tabletop in
the Trendelenburg (button OB6, Fig. 5.13) or reverse
Trendelenburg direction (button OB7, Fig. 5.13)
OB2Height - Press and hold this button, then press
required direction button to change the height of the
tabletop (button OB6, Fig. 5.13 is for Height up;
button OB7, Fig. 5.13 is for Height down).
OB3Tilt - Press and hold this button, then press
the required direction button to tilt the tabletop
(button OB6, Fig. 5.13 is for Tilt down on the left;
button OB7, Fig. 5.13 is for Tilt down on the right,
when viewed from the long trunk end of the table).
OB4Break - Press and hold this button, then press
required direction button to move the tabletop in the
Break up or Flexion direction (button OB6, Fig. 5.13)
or Break down or Extension direction (button OB7,
Fig. 5.13).
OB5Traverse - Press and hold this button, then
press required direction button to traverse the
tabletop (button OB6, Fig. 5.13 is for cranially; button
OB7, Fig. 5.13 is for caudally).
5.3.5 Fuse replacement
The fuses are located as shown in Fig. 2.2 items 2 and 13
for T20-a and T20-s tables and Fig. 2.3 item 5 for the T20-m
table. Fuses adjacent to the mains inlet socket are only
applicable to the mains supply for the battery charger.
Fuses in the side of the table base adjacent to standby
battery switch (T20-a and T20-s tables ONLY) are
applicable to the main batteries and standby batteries.
Fig. 5.13 Standby control panel
To operate the tabletop from this panel select a required
function by ‘pressing and holding’ a function button
(i.e. OB1 to OB5, Fig. 5.13) and then select a direction for
this function by ‘pressing and holding’ a direction button
(i.e. OB6 or OB7, Fig. 5.13).
Pressing the upper or lower direction button will select the
direction indicated by the corresponding arrow on the
function button. The table will move whilst both the function
button and the direction button are pressed, releasing either
will stop the motion. Motions for each button are fully
detailed in the following sections.
These fuses are replaced as follows:-
iRemove the mains cord from the table before
replacing any of the fuses and switch the table ‘off’
with switch item 5, Fig. 2.2.
iiTurn the fuse cover cap anticlockwise to remove
the cap with the fuse inside.
iiiEnsure you have the correct fuse (consult the
markings adjacent to the fuse or the Technical Data,
section 9.0).
ivPlace the new fuse into the cap and replace the cap
by screwing it in clockwise.
T-IM114bP31/62
6.0 PATIENT POSITIONING
in various positions is shown in Figs. 6.3 and 6.4. For
guidance on the treatment of the obese patient, see
section 6.2.
6.1General
WARNING
Do not use any table sections or accessories
without the correct Eschmann mattress or pad in
place (except for accessories such as the
lightweight leg section which can be used for short
periods of use, without a mattress, during patient
positioning). Mattresses are an important part of
the antistatic pathway and help prevent pressure
sores developing.
Ensure the patient’s weight is evenly distributed and that
the mattress or pad is correctly positioned on each section
of the table. Check at frequent intervals during long
procedures to ensure that pressure sores are not being
promoted. See section 5.2.6 for details on attaching and
removing mattresses.
Note the graphics in section 6.4 which provide maximum
loading details for various table positions. The normal
maximum loading, for the patient’s centre of gravity (C of G)
55
81
5
5
7
7
11
15
15
22
22
10
10
15
15
135kg
10
200kg
15
5
5
8
8
11
300kg
22
22
22
11
33
124
33
11
Fig. 6.1 Approximate patient weight
distributions
The Figures in section 6.4 provide details of various patient
positions that can be arranged with the T20 Series of table.
Note that these illustrations are examples only and care
should be taken to ensure patient safety and table stability.
The terms caudally and cranially are explained in section
1.2.1 (iv).
6.2Treatment of the obese patient
Always ensure that the table is correctly loaded to maintain
stability within the guidelines shown in the graphs of section
6.4 noting the maximum patient weight of 300kg. Note that
the centre of gravity for the maximum patient weight of
300kg must not be outside of the top plateaux section of
the graphs.
Several width extenders can be placed on either side of the
table to help support the patient but note the maximum load
for this accessory and that width extenders should not be
attached to each other (also see Warnings and Cautions in
the width extenders ‘User Handbook’ ).
The table should not be moved across the floor with a
patient weight over 135kg (T20-a and T20-s tables) or
200kg (T20-m table). Although it is possible to place the
T20-a and T20-s table into ‘castor’ or ‘wheel’ with a patient
over 135kg and up to 300kg this is not advised.
6.3Radiographic procedures
All ‘C-arm’ image intensifiers are suitable for use with this
table.
Fig. 6.2 shows typical patient positions for imaging the
upper and lower torso with the patient in the Supine
position. The graphics would equally apply to patients in
the Prone position.
6.4Table positions and safe loading
In Figures 6.5 to 6.10, common table and patient positions
and safe loading graphs are provided, to give guidance on
patient positioning and table loading. Because it would not
be possible to cover every situation these should be
considered as examples only. Refer to Figs. 6.5 to 6.10
and then use Figs. 6.3 and 6.4 to establish the maximum
patient weight allowed for that configuration. The weight of
additional accessories fitted to the table should be included
in any estimation of correctly applied loads.
Figs. 6.3 and 6.4 show graphs of the position of the
maximum patient weights (centre of gravity) that can be
positioned on the table in four table configurations. Because
all variations can not be covered, a degree of interpolation
is required by the user, to ensure that table stability is
maintained. Fig. 6.1 provides an approximate patient
weight distribution for guidance.
P32/62T-IM114b
T20
OPERATION TABLE
Patient positioning for maximum ‘C’ arm imaging of the required area.
Additionally there is an X-ray cassette tunnel, accessible from either end.
Lower torso imaging
Patient in normal orientation with table
head and leg sections in the normal
position. Patient positioned down the
table to maximize the amount of lower
torso in the imaging area.
Series
NOTE: All illustrations are
examples only. Care should be
taken in patient positioning to
ensure stability, see section 6.3.
For the obese patient see section
6.1.
Upper torso imaging
Patient in normal orientation with table
head and leg sections reversed from their
normal position (i.e. leg section in the
long trunk). Patient positioned up the
table to maximize the amount of upper
torso in the imaging area.
Fig. 6.2 Patient positioning for radiographic procedures
T-IM114bP33/62
When moving the table with
a patient, ensure their centre
of gravity lies within the bold
vertical lines in all cases.
B
A
B
A
Fig. 6.3 Maximum patient weight v table position graphs
(head and leg section in normal positions).
P34/62T-IM114b
When moving the table with
a patient, ensure their centre
of gravity lies within the bold
vertical lines in all cases.
A
T20
OPERATION TABLE
Series
B
B
A
Fig. 6.4 Maximum patient weight v table position graphs
(head and leg section reversed).
T-IM114bP35/62
Fig. 6.5 Patient in various ‘Supine’ positions
(Note: For maximum patient weights see Fig. 6.3 and 6.4)
P36/62T-IM114b
T20
OPERATION TABLE
Series
Fig. 6.6 Patient in various ‘Prone’ positions
(Note: For maximum patient weights see Fig. 6.3 and 6.4)
T-IM114bP37/62
Note warnings in section 2.1
regarding patient support.
Fig. 6.7 Patient in various ‘Lawn chair’ & ‘Trendelenburg’ positions
(Note: For maximum patient weights see Fig. 6.3 and 6.4)
P38/62T-IM114b
T20
OPERATION TABLE
Series
Fig. 6.8 Patient in various ‘Chair’ positions
(Note: For maximum patient weights see Fig. 6.3 and 6.4), but note that with
the table and patient positioned as above these are basically stable positions)
T-IM114bP39/62
Note: Maximum patient weight
for these two positions is 85kg.
Note: Iliac crest support may be
useful to reduce knee damage
for all three positions.
Fig. 6.9 Patient in various ‘Proctology’ positions
(Note: For maximum patient weights see above and Fig. 6.3 and 6.4), note that table
stability in the top two graphics is the lowest and can only be used for the lighter patient)
P40/62T-IM114b
T20
OPERATION TABLE
Series
Fig. 6.10 Patient in ‘Lithotomy’, ‘Prone Laminectomy’ and ‘Lateral Nephrectomy’
(Note: For maximum patient weights see Fig. 6.3 and 6.4)
T-IM114bP41/62
7.0 ACCESSORIES
WARNING
When fitting any accessory ensure that it has
been securely attached and the maximum
loading has been noted before using it to support
part of the patient’s weight. Also see guidance
notes in section 4.4.
The catalogue (REF) numbers for the standard T20 Series
accessories are shown in the following list, where
applicable these are supplied with a ‘User/Service
Handbook’. This list may not be exhaustive as new
accessories are added to the range . Please check with
Eschmann Equipment for the latest additions.
Some of the accessories listed below are also available
with alternative side rails, please contact Eschmann for
more information.
IMPORTANT NOTE: Please check with Eschmann
Equipment or their local agent before using
accessories that fit onto the side rail of tables supplied
with non-standard (e.g. US, Denyer or Euro) side rails.
REFDescription
T00-3110001Head section², purple, UK side rail
T00-3310001Head section², blue, UK side rail
T00-3210001Leg section², purple, UK side rail
T00-3410001Leg section², blue, UK side rail
T00-5300001Corded handset, blue, (UK side rail)
T00-5400001Corded handset, purple, (UK side rail)
TA-0303026Divided leg², pair, purple, UK side rail
TA-0303028Divided leg², pair, blue, UK side rail
TA-0202001Width extender², long, UK side rail
TA-0202007Width extender², short, UK side rail
TA-0203001Foot rest/extension², UK side rail
TA-0204089Back elevator
TA-0301001Lightweight leg
TA-0302001Ophthalmic head section
TA-0303004Accessory attachment bar
TA-0401013Accessory trolley
TA-0601011Leg abduction support
TA-0801090X-ray cassette tray
81-342-00Perineal instrument tray
81-367-85Douche tray
Mattresses (original)
T20-411-0001Mattress, 50mm, complete set
T20-412-0001Mattress, 75mm, complete set
T20-441-0001Mattress, 50mm, leg section
T20-442-0001Mattress, 75mm, leg section
TA-0501002Mattress, 50mm, long width extender
TA-0501008Mattress, 50mm, short width extender
TA-0501009Mattress, 75mm, long width extender
TA-0501010Mattress, 75mm, short width extender
TA-0501006Mattress, Ophthalmic head section
TA-0501015Mattress, 25mm, lightweight leg
TA-0501016Mattress, 50mm, lightweight leg
Mattresses (new Mk.II*)
T20-414-0000Mattress, 50mm, complete set
T20-421-2093Mattress, 50mm, trunk section
T20-431-2094Mattress, 50mm, head/footrest
T20-441-2095Mattress, 50mm, leg section
TA-0501096Mattress, 50mm, long width extender
TA-0501097Mattress, 50mm, short width extender
TA-0502118Mattress, 50mm, divided leg, pair
TA-0501100Mattress, 25mm, lightweight leg
* Introduced in November 2005. Weights for these new
mattreses are approximately one third of the original
mattresses and can can be identified by white breather
ports on their base.
Side rail clamps
TA-020-1082Anti drift rotary clamp (for UK side rail)
TA-020-1086Anti drift rotary clamp (for Euro side rail)
TA-020-1087Anti drift rotary clamp (for US side rail)
TA-020-1084Drop handle direct on clamp (UK/Euro)
TA-020-1085Drop handle direct on clamp (UK/US)
²
These accessories come without a mattress which
should be ordered separately, see list above.
Eschmann Equipment also supply:
A ‘Modular Patient Positioning System (M.P.P.S.)’ which
utilises a range of modular components that can be
easily configured to suit both the surgical requirement
and the patient’s anatomy and a range of support and
Gel pads, suitable for use on the T20 Series of tables.
Direct Placement Leg Holders (DPLH) which can be
used as an alternative to knee crutches and Lloyd
Davies supports for positioning patients into the
lithotomy position. In addition they have been cited in
orthopaedics for assisting both positioning of the patient
and facilitating C-arm access in orthopaedic trauma.
A range of support and Gel pads, suitable for use on
these tables.
For information on any of the above please contact
Eschmann Equipment, at the address on the back cover,
to request advice or product information leaflets.
Alternatively visit the Eschmann website at
‘www.eschmann.co.uk’
P42/62T-IM114b
The following accessories can be used with the T20 Series
of tables. Their use is self explanatory but always ensure
all clamps are fully tightened before using to support the
applicable part of the patient’s weight. Some of these
require additional clamps to secure them to the table this
is indicated in the following list. Again this list may not be
exhaustive as new accessories are added to the range.
Please check with Eschmann Equipment for the latest
additions.
REFDescription
81-250-15Lithotomy supports **
81-264-10Shoulder rests
81-286-18Anaesthetic screen, single *
81-290-10Arm support, Carter Braine #
81-300-19Narrow arm table with pad #
81-314-14Arm table, square without support ##
81-344-13Instrument table *
81-368-23Perspex arm support, right-angle
81-378-38Foot stocks
81-404-13Pelvis support #
81-412-15Buttock support #
81-420-17Lloyd Davies lithotomy supports
81-428-15Knee crutches
81-436-17Pelvis / chest support #
81-446-13Wristlet #
81-462-17Patient restraint strap
81-462-33Cot side, tall
81-463-06Infusion pole
81-466-40Ophthalmic head flap
81-490-03Kidney position supports ##
81-492-16Lateral support, adult
81-504-19Lateral support, child
81-569-13Laminectomy support pad
81-569-80Laminectomy support **
81-812-17X-ray cassette holder, lateral
81-820-19Arm table, square with support
81-827-52X-ray cassette tray, end loading
81-828-41X-ray cassette tray, side loading
81-931-44Orthopaedic knee crutch
* Requires one, Anti drift rotary clamp
** Requires two, Anti drift rotary clamps
# Requires one, Drop handle direct on clamp
## Requires two, Drop handle direct on clamps
†
†
†
†
T20
Series
OPERATION TABLE
† Ensure these clamps match the tables side rail
(e.g. UK, Euro or US) see ‘Side rail clamps’ on
the previous page.
Spare mains leads can be ordered under the following
Spare Part Numbers:
391177 - T20-a/T20-m, Australian mains lead, not fused.
714188 - T20-a/T20-m, U.K. mains lead (10A fuse).
715254 - T20-a/T20-m, Euro mains lead, not fused.
111589 - T20-a/T20-m, U.S.A. mains lead, not fused.
111466 - T20-s, U.K. angled mains lead (10A fuse).
T-IM114bP43/62
8.0 AFTER USE, CLEANING & CARE
8.1 After use procedures
CAUTION
Do not immerse the handsets or footswitch in
liquids when cleaning or disinfecting.
After each procedure the table should be cleaned and
disinfected in accordance with local procedures which take
into account the guidelines provided in section 8.2 and 8.3
of this manual.
Accessories should be removed from the table and after
cleaning and disinfection the head and leg section should
be fully lowered as good practise.
The table battery LEDs should be checked to see if the
batteries require recharging, recharge if applicable. If the
preceding procedure was the last one for the day or shift,
then the batteries should be placed on charge as a routine
procedure (see section 5.3.1.2)
The table must be switched ‘off’ ( ) at the table on/off switch
(item 5 of Fig. 2.2) when the table is not in use. The green
LED item 4 Fig. 2.2 will not be illuminated to show the
table is ‘off’.
If the table has been left switched ‘on’ and no movement
command has been given via any controller (hand, foot or
standby) for a period of six hours, an audible warning sounds
(a slow repeating double ‘beep’). This is provided to warn
the user to switch the table ‘off’ ( ) at the table on/off switch
(item 5 of Fig. 2.2). If the table is still in use when this alarm
sounds (e.g. table being used for a long procedure) and the
user does not want the table switched ‘off’
(e.g. Trendelenburg may be required urgently) the alarm
can be stopped by pressing any control input briefly, or,
switching the table ‘off’, waiting 10 seconds and then
switching the table back ‘on’. The alarm will sound again in
six hours if no movement command has been given and
the table has not been switched ‘off’. This feature can be
disabled or the time period of six hours can be altered,
contact Eschmann Equipment for more information.
8.2 Cleaning
NOTE
Eschmann Equipment cannot accept liability
for the efficiency or the effects of any cleaning
or disinfection techniques. Only use the
materials detailed in the following sections.
WARNING
Disconnect from the mains electrical supply
before cleaning or disinfecting the operation
table, mattresses, or accessories. When
cleaning with a brush wear suitable eye
protection (brushes are prone to ‘flick’
particles) and at all times wear suitable
personal protection (e.g. gloves and overalls).
If you have any particular cleaning problems, contact the
Eschmann Equipment After Sales Service Department at
the address given inside the front cover of this Manual.
8.2.1 Operation table and accessories
Fully raise the tabletop, remove all accessories and all the
mattresses, then wash the table with hot (55°C) neutral
(pH7) detergent solution (diluted in accordance with the
manufacturers instructions) and rinse with clean water. Use
a small brush to clean areas of difficult access.
Clean non-electrical accessories as detailed above for the
table but only use a damp cloth to clean the handsets and
footswitch.
8.2.2 Mattresses and pads
WARNING
Mattresses and pads should be checked for any
cuts or tears in the outer covering. If any of
these defects are found, the mattress must be
replaced to eliminate possible biological
hazards.
CAUTION
Remove spilt ether and anaesthetic liquid from
mattresses and pads immediately.
When cleaning mattresses and pads:
vDO NOT use phenolic disinfectants.
vDO NOT use abrasive cleaning agents.
vDO NOT use organic solvents (e.g. petroleum spirit,
carbon tetrachloride, or tetra-chloromethane).
vDO NOT allow petroleum-based oil, vegetable oil,
wax, or grease to remain on the mattress.
vDO NOT dry mattresses with direct heat (e.g.
radiators, electric fires, or hot-air blowers) allow them
to dry at normal room temperatures.
Wash mattresses and pads thoroughly, with hot (55°C)
neutral (pH7) detergent solution (diluted in accordance with
the manufacturers instructions) and then rinse with clean
water. Use a small brush to clean areas of limited access.
Remove stubborn stains and deposits with good quality
vinyl cleaner using a soft-bristle brush. Dry all surfaces
with absorbent paper, to avoid damage do not leave them
wet in contact with another surface (e.g. the table top).
P44/62T-IM114b
T20
Series
OPERATION TABLE
8.3 Disinfection
NOTE
All equipment and accessories returned to
Eschmann Equipment must be accompanied
by a Decontamination Certificate, signed by an
authorized person of managerial status.
Appendix 2 shows an example of a suitable
Decontamination Certificate that can be
photocopied for use.
CAUTION
Hypochlorite solutions can damage metal
parts, after disinfection, ensure it is rinsed of
thoroughly with water. Prolonged exposure to
hypochlorite solutions may degrade mattress
material. Avoid hypochlorite solutions and
other liquids coming into contact with any
internal parts of the table.
8.3.1 Disinfection procedure
A well ventilated area should be designated and used as the
disinfection area. Access to the area should be restricted to
those people involved in the disinfection process.
The following disinfection procedure is used by Eschmann
Equipment, and its use is recommended if no other local
approved procedures are available.
8.3.2 Table, accessories mattresses and pads
Disinfect the operation table, accessories, mattresses and
pads as follows:
iDisassemble the table and/or accessories as far
as possible without the use of tools, remove
mattresses and pads.
iiScrub all surfaces and crevices with hot (55°C)
neutral (pH7) detergent solution (diluted in
accordance with the manufacturers instructions) to
remove all visible contamination. Use a small brush
to clean areas of limited access repositioning the
table as required to gain access to all surfaces. For
stubborn mattress stains see section 8.2.2.
iiiWash down with hot (55°C) water.
ivDry all surfaces with absorbent paper.
vWash down all surfaces and crevices with one of
the solutions below:
viDry all surfaces with absorbent paper, to avoid
damage do not leave mattresses wet in contact with
another surface.
vii Dispose of all cleaning material and solutions in
accordance with authorized disposal procedures.
8.4Care
WARNING
The head and leg section gas-springs are filled
with gas at high pressure, do not try to open
them. The gas-springs should be replaced
immediately any signs of leaking or
deterioration in performance are noted (e.g.
movement of section when locked).
CAUTION
Do not lubricate the head or the
leg section, gas-springs.
Once a week:
iRemove fluff and debris from the head and leg
section guide pin location sockets and if required
spray the pins and into the sockets with a suitable
light lubricant such as WD40.
iiRemove all mattresses and check them for any cuts,
scuffs or other damage and replace as required.
iiiCheck that the table covers are not cracked,
chipped, or otherwise damaged and arrange
replacement as required.
ivCheck the table for any signs of wear or damage
that requires attention and arrange remedial action
if required.
vEnsure batteries have been regularly charged and
charge them if required.
viOn the T20-m table clean the castors and remove
any debris caught between them.
Once every six months the table should receive a safety
check and service as detailed in the service manual (which
provides full details of part replacement, safety checks and
routine maintenance). During this service the tables
calibration should be checked and if required it should be
recalibrated. Also at least once every year the electrical
conductivity of the table should be checked. To arrange
these contact the Eschmann After Sales Service
Department, see inside front cover for contact details.
va 70% solution of industrial methylated spirit in
water.
va 1000 to 5000mg/litre solution of hypochlorite
in water (see Caution note above).
viWash down thoroughly with clean water.
T-IM114bP45/62
8.5Storing the operation table (long term)
8.6.3 Environmental considerations
The Operation table should be covered, and stored in a
clean environment, with no extremes of temperature, see
the Technical Data section for details.
The table should be stored:
vWith the base in the braked position.
vWith the head and leg sections fully lowered.
vWith the tabletop level in both planes.
vWith ‘on/off’ switch (item 5, Fig. 2.2) ‘off’.
Every two weeks:
vCharge the table batteries (see section 5.3.1.2)
until both charging LEDs are green.
The storage maintenance inspection must be conducted
by a trained engineer. When the inspection is finished, the
‘Storage Maintenance Record’ (see page 47) should be
completed and signed.
CAUTION
Do not store mattresses and pads with objects
(especially objects having sharp edges and
protrusions) resting on their padded surface
as this could damage them. Always store them
flat.
Mattresses and pads should not be left in direct or
excessive heat and they must be stored flat. Do not leave
them with other accessories resting on them or leaning
against another object.
8.6Maintenance
8.6.1 General
The Service Manual, which can be ordered from the
Eschmann After Sales Service Department, contains the
routine service schedule and details of how to replace parts
when required. Some parts of the table are not user
serviceable and this is detailed in the Service Manual, also
refer to section 1.1.6, 3.3.9 and 8.4.
8.6.2 Fault diagnosis
Table 1 lists possible causes for faults and conditions that
may arise. Some of these may require further investigation
by trained engineers in conjunction with the Service Manual
and circuit diagrams. Where the remedy will require an
engineer to rectify the fault or condition this is indicated by
the phrase “Engineer to...” in the remedy column. If any
fault persists (e.g. blown fuse) this should be investigated
by an engineer.
WARNING
Gas springs contain nitrogen gas and a small
quantity of hydraulic oil at very high pressure,
these must be vented before disposal, consult
the Service Manual for the safe procedure.
DISPOSAL NOTE
This equipment contains environmentally
hazardous lead-acid batteries. If the batteries
fail, or if the equipment is to be disposed of, it
is recommended that the batteries are taken to
a disposal site designated for the disposal of
lead-acid batteries, or that the batteries are
collected by an agent who specialises in the
collection of lead-acid batteries.
During normal use there are no environmental
considerations that need to be considered. During the
design stage several materials were considered to be
unacceptable and materials such as Cadmium, CFC filled
capacitors and devices containing mercury, have not been
used within this equipment.
During cleaning and disinfection procedures, the potentially
contaminated waste materials produced during these
actions, should be handled in accordance with local
procedures and National legislation for the disposal of
potentially contaminated waste.
At the end of the working life of the table it should be
dismantled and recycled as much as possible in line with
the recommended procedure available from Eschmann
Equipment and in accordance with local procedures and
National legislation.
8.6.4 Technical Lifetime
This product has a technical lifetime, which by Eschmann
Holdings Limited is considered to be 10 years. At the time
of delivery the product fulfils the existing regulations and
standards but as with all other electro-mechanical products,
the Eschmann T20 Series of tables is subject to ageing
and wear, and even though the product may have
undergone regular service in accordance with the
recommended service schedule, Eschmann Holdings
Limited can not guarantee the product’s safety after the
expiry of the technical lifetime.
Eschmann Holdings Limited recommends that a T20 Series
table is taken out of service 10 years after the date of
manufacture as shown on the serial label fixed on the
table’s base. Provision of spare parts and service by
Eschmann Holdings Limited after the expiry of the specified
technical lifetime does not mean an extension of Eschmann
Holdings Limited liabilities.
P46/62T-IM114b
T20
OPERATION TABLE
STORAGE MAINTENANCE RECORD (for Table Serial No : ................................................)
Series
Date ofDate of
Battery ChargeInspection
SignatureComment
T-IM114bP47/62
TABLE 1 - FAULT DIAGNOSIS
FAULTPOSSIBLE CAUSEREMEDY
Table will not move inTable not switched ‘on’.Switch table ‘on’, switch 5, Fig. 2.2.
response to handset
or footswitch.Main batteries critically low,Recharge batteries see section 5.3.1.2
see 5.3.1.1 and Fig. 5.9.OR, in an emergency only control
the table using the standby control
panel as detailed in section 5.3.4,
OR, switch in the standby batteries as
detailed in section 5.3.1.4.
Corded handset or footswitchConnect handset (see 5.3.3) or
not connected to table.footswitch to table.
Handset or footswitchReplace handset or footswitch with
faulty, or possible flat batteryone known to work, or engineer to
in infrared handset.arrange replacement of infrared hand
control batteries.
Infrared handset has a differentUse a compatible handset, see
identifier code to the table.infrared handset ‘User Handbook’..
Infrared handset has infraredRemove labels/fingers from window.
windows blocked by labels/fingers.
Table not configured for infraredArrange to have table configured for an
handset.infrared handset by a Service Engineer.
Battery fuse blown.Replace fuse as detailed in section 5.3.5,
for T20-a and T20-s only, or, switch to
standby battery as detailed
in section 5.3.1.4.
Table has reached the limitPress alternative button.
for that button function.
Two buttons on one controllerRelease both buttons and press
being pressed at the same time.only one button at a time.
An internal error has beenRelease button, wait 2 seconds for
detected in the control system.error to clear, press button again.
Signal from infrared handsetMove handset around table or
blocked by object or personnel.closer to the pedestal.
Signal from one controllerOnly use one type of handset
being overridden by anotherwith a table at any time.
controller (see section 4.6.1.2).
Main table batteries have failed.Switch in the standby batteries as
detailed in section 5.3.1.4.
Maximum load of function selectedAssist table motion or reposition patient
has been exceeded.to reduce the offset load.
Table motion physically blockedMove object or reposition table.
(e.g. leg section hitting floor)
Two handsets connected to tableDisconnect one (either) handset.
P48/62T-IM114b
T20
Series
OPERATION TABLE
TABLE 1 - FAULT DIAGNOSIS (continued)
FAULTPOSSIBLE CAUSEREMEDY
Table will not moveTable base in ‘braked’ position.Place table into the ‘castor’ or ‘wheel’
when pushed.orientation as detailed in section 5.1.
Foreign object trapped underCheck for foreign object and remove
a wheel or castor.it if found.
Table difficult to manoeuvre.Table in ‘wheel’ orientation.Place table into ‘castor’ orientation.
Table difficult to moveTable in ‘castor’ orientation.Place table in ‘wheel’ orientation.
in a straight line.
Table not stable on floorTable is not in ‘braked’ position.Place table into the ‘braked’ orientation
and moves when pushedas detailed in section 5.1.
Table not stable on floorFloor uneven or object underMove table to a flatter area
and wobbles when ‘braked’.a wheel or castor.or remove object.
No charge LED(s) ‘on’Mains supply faulty or notCheck mains supply, or switch ‘on’
when table connectedswitched ‘on’.at mains supply.
to mains.
Mains cable not connectedReconnect mains cable correctly at
correctly at both ends.each end.
Mains cable faulty.Replace mains cable.
Mains fuse in supplyReplace fuse in mains plug (10A).
cable blown.
Mains supply fuse inReplace fuse as detailed in section 5.3.5.
table blown.
Charge LED colour doesTable batteries at fault.Engineer to change batteries.
not change to green (even
after twelve hours).
Table section does notWeight of section not supportedSupport section weight when attaching
attach or releaseduring attaching or removal.and removing section (i.e. gently lift end
easily from table.furthest from table).
Unexpected table movement. Another infrared controllerCheck for another infrared controller
being used locally.and discontinue its use, or engineer to
change table infrared code.
Excessive motor noise.Faulty drive.Engineer to remedy fault.
Table audible alarmTable left switched ‘on’ and noSee section 8.1 and either stop the
sounding (‘beeping’)command input given for 6 hours.alarm (if table still in use) or switch the
table ‘off’ (if table not in use). If the table
is not in use, place it on charge.
T-IM114bP49/62
9.0 TECHNICAL DATA
9.1Weights
The nominal weights of the standard table components
are listed below (50mm mattress*):
Leg section (without mattress).... 11.50kg
Leg section mattress ........ 1.00kg
Head section (without mattress).... 7.75kg
Head section mattress...... 0.75kg
T20-a and T20-s (base and column only)192kg
T20-m (base and column only).... 140kg
Long and short trunk assembly..55.75kg
Long and short trunk mattress.... 2.50kg
Total weight T20-a and T20-s.... 271kg
Total weight T20-m........219kg
Lightweight leg section...... 4.50kg
Footrest (incl. mattress)...... 7.00kg
Width extender (incl. mattress).... 6.00kg
Perineal instrument tray...... 3.00kg
* NOTE:50mm mattress weights above are for the new style of
50mm mattress introduced in November 2005. Original mattress
weights are approximately three times those shown above. New
mattresses can be identified by white breather ports on their
base (see section 7.0).
Max. extension (4)........230°
Max. flexion (4) ..........90°
Head section (5)........±45°
Leg section (3) ........-100° +55°
Max. lateral tilt (7)........±18°
9.4Table loading
See graphs in section 6.0 for maximum patient weight
distribution charts for various table positions and
orientations.
Maximum section loading is stated on each item and is
detailed below for reference:
Leg section....44kg
Head section....22kg
For accessories the maximum loading is stated on each
item and detailed in the User Handbook supplied with them.
For specific notes on the obese patient see section 6.2.
9.5Electrical
9.5.1 Antistatic requirements
The table has an antistatic pathway from the table top,
through an internal resistor, to a castor.
9.2Dimensions
The following are the nominal major dimensions of the
standard table (with head and leg sections fitted) and
include the standard 50mm mattress and side rail if
applicable (see Fig. 9.1 for illustration, the number in
brackets after the item details which part):
Overall length (2)......2102mm
Maximum height (4)......1120mm*
Minimum height (5)......720mm*
Maximum top traverse (2)....250mm
Overall width T20-a and T20-s (6) ..600mm
Overall width T20-m (6)....725mm
Side rail (UK)...... 31.5 x 6.2mm
* 5mm less for tables built prior to June 2005
9.3Movements
The following table movements are maximums and cannot
be assumed to be available in all combinations of table
positioning (see Fig. 9.1 for illustration, the number in
brackets after the item details which part). For example,
with the table in its normal configuration and at minimum
height the maximum Trendelenburg and leg section
movements cannot be achieved due to the proximity of
the floor. Similarly with a large amount of tilt set, maximum
Trendelenburg cannot be achieved.
Max. Trendelenburg (1)......35°
Max. reverse Trendelenburg (1)....35°
CAUTION
To complete and maintain the antistatic
pathway the table must be used on an
electrically conductive or antistatic floor and
with mattresses supplied by Eschmann
Equipment.
9.5.2 Batteries
Table base:
Main batteries:
Two, 12V 10Ah, sealed lead acid
Standby batteries:
Two, 12V 1.2Ah, sealed lead acid
(see battery disposal caution in section 8.6.3)
Infrared handset:
Two, 1.5V size AA Alkaline (Note: Must only be
changed in accordance with the notes in the ‘Service
Manual’ to ensure the IP rating is not compromised).
9.5.3 Internal battery charger
Input
100-240V a.c. 50-60Hz 2.5A max
Output
29.2V d.c. 2A (max.) when charge state LED
red or orange, 27.6V d.c. when charge state
LED green
P50/62T-IM114b
T20
Series
OPERATION TABLE
1
4
5
2
6
725mm*
610mm*
T20-a and T20-sT20-m
3
7
These are maximum movements for each aspect, they may not be available in certain combinations (e.g. maximum
tilt and maximum Trendelenburg). Movements that could cause damage cannot be catered for (e.g. position of the
leg section). Also see sections 9.2 and 9.3.
* The maximum width of the T20-m table is across the base at the foot pedal end, this is 610mm with
the castors in line with the base and 725mm with both castors at 90° to the base as indicated.
Battery fuses (item 2, Fig. 2.2). Note: On the T20-m
these fuses are fitted internally.
2 x T20A (6.35 x 32mm) 500V
Mains plug fuse (if fitted) 10A.
Internal fuses (only accessible by engineer)
Battery fuses (joining each battery pair)
2 x 30A blade type (1 per battery pair)
2 x T20A (6.35 x 32mm) 500V (these fuses
are fitted externally on the T20-a and T20-s)
9.5.5 Duty cycle
This symbol is used to indicate the table’s duty cycle
which, in the worst case, is ‘60s : 600s’, the ratio of the
operating time to the sum of the operating time and the
ensuing interval (see note below). Each motor drive has
its own duty cycle and this is dependent on loading and
table position as detailed below.
Trendelenburg1 : 4 (60s : 240s) at a maximum torque
of 417Nm (e.g. 135kg load offset 31cm from the
fulcrum, or a 300kg load offset 14cm from the
fulcrum. Where this load is the patient and
accessories weight and the offset is how far the loads
centre of gravity is, from the centre of the column).
Traverse1 : 4 (60s : 240s) when the table is horizontal
and at the maximum load of 300kg, or at 17.5° with
a patient weight of 225kg, or at 35° with a patient
weight of 150kg.
9.7.1 Electrosurgical equipment (h.f.)
The T20 Series of operation table have been designed to
minimise the possibility of accidental electrosurgery burns
and can be used in conjunction with electrosurgical
equipment. However contact with any metal surfaces (e.g.
table side rail, or other equipment etc.) can cause burns
during electrosurgery and must be avoided.
9.7.2 Defibrillation equipment
With the mains cord attached the equipment has a
defibrillator proof applied part with type BF protection
against electric shock.
9.7.3 RF communications equipment
See section 9.11 ‘Electromagnetic compatibility’.
9.8Standards compliance
The table has been designed and built to comply with the
following international standards:
BS EN 60601-1: 1990 and all amendments to date
BS EN 60601-1-2: 2001
BS EN 60601-2-46: 1998
BS EN 60601-1-4: 1997
9.9Environmental conditions
9.9.1 Operating environment
The table has been designed to operate in the following
environment:
Temperature range....10°C to +40°C
Pressure range ......69KPa to 106KPa.
Humidity range ......30% to 75% RH
non-condensing.
Break1 : 10 (60s : 600s) at maximum patient weight
of 300kg.
Height1 : 4 (60s : 240s) at maximum patient weight
of 300kg (not offset).
Tilt1 : 4 (60s : 240s) at maximum patient weight
of 300kg.
Note: The duty cycles above are all for the worst case (i.e.
maximum loads). For reduced loading the above duty
cycles can be increased.
9.6Classification and symbology
All the symbols used on this table are shown and explained
in section 2.2. These detail the safety category and class
of this table as marked on the table or section by the use
of these symbols.
9.9.2 Transport & storage environment
The table can be transported and stored safely, in the
following environment:
Temperature range....-30°C to +50°C
Pressure range ......69KPa to 106KPa
Humidity range ......30% to 90% RH
non-condensing.
9.10 Alarms
9.10.1Visual
Visual alarm functions of this table are LED indicators for
mains ‘on’ and battery charge state (see Figs. 2.3 and 2.3)
situated on the table base and battery charge level
indication (see Fig. 5.9) provided on the corded handset.
Ensure you are familiar with these before using this
operation table (see sections 5.3.1). There is also an LED
indicator to show that the table is switched ‘on’ (see item 4,
Fig. 2.2 and 2.3).
P52/62T-IM114b
T20
Series
OPERATION TABLE
9.10.2Audible
Audible alarm indication is provided (a double ‘beep’
repeating every eight seconds) to warn that the table has
been left ‘on’ for an extended period and should be switched
‘off’ (see section 8.1 for more information). A single ‘beep’
also sounds when the table is switched ‘on’.
9.11 Electromagnetic compatibility (EMC)
9.11.1Interference considerations
The Eschmann T20 Series of tables have been designed
and manufactured in such a way as to remove, or minimise
as far as possible, risks connected with reasonably
foreseeable environmental conditions such as magnetic
fields and external electrical influences (i.e. electrical
interference). The risks of electrical interference from or to
other devices normally used with these tables have been
taken into account. The steps taken to achieve this are to
ensure compliance with relevant international
electromedical standards {i.e. EN 60601-1-2:2001)
regarding electromagnetic compatibility. This compliance
has been confirmed by independent testing. However it is
not possible to simulate all the conditions that may be
encountered, the compliance testing therefore provides
only a very good indication as to the susceptibility or
suppression of emissions to or from a device.
It is highly unlikely that any electrical interference problems
will be encountered with these tables. However, should
such interference be suspected, the following the following
guidance is provided.
9.11.2Interference from other equipment
The Eschmann T20 Series of tables have been designed
to ensure that when using them in close proximity with ‘other
correctly designed’ electrical equipment, interference with
the tables control systems does not occur. Eschmann cannot
guarantee that other equipment used in an operating theatre
is properly constructed so as to avoid electrical interference
with them. This could be a problem particularly with very old
equipment (i.e. it does not conform to the latest standards).
In the unlikely event that electromagnetic interference is
caused by ‘other’ noncompliant equipment, such ‘other’
equipment should not be activated at the same time these
tables are switched ‘on’ and their control systems are active.
these tables, the other equipment should not be activated
at the same time as the tables are switched ‘on’ and their
control systems are active.
9.11.4RF communications equipment
As with all medical electrical equipment the user should
be aware that portable and mobile RF communications
equipment can affect medical electrical equipment such
as these tables.
9.11.5Accessories
As with all medical electrical equipment, and in line with
the latest European standard (EN 60606-1-2:2001) the user
should be warned that the use of accessories, transducers
and cables other than those specified below (with the
exception of those sold by Eschmann Equipment as
replacement parts for internal components) may result in
increased emissions or decreased immunity of the these
tables and their associated accessories.
The following Eschmann Equipment accessories can be
used with these tables without affecting the table’s
electromagnetic compatibility with subclauses 36.201 and
36.202 of EN60601-1-2:2001. Other similar accessories
or equivalents that could affect compliance with the
requirements of subclauses 36.201 and 36.202 of
EN60601-1-2:2001 should not be used.
Mains cord (for part number see section 7.0).
T00-5300001Handset, blue, (UK side rail)
T00-5400001Handset, purple, (UK side rail)
(also, other handsets for non UK side rail)
All the other accessories listed in section 7.0 are designed
for use with the Eschmann T20 Series of tables and have
no EMC implications.
9.11.6Installation
WARNING
The T20 Series tables should not be stacked
with other equipment. If used adjacent to other
equipment these tables should be observed to
verify normal operation in the configuration in
which it will be used.
9.11.3Interference with other equipment
The Eschmann T20 Series of tables have been designed
to ensure that when used in close proximity with other
correctly designed equipment they do not cause
interference with such other correctly designed equipment.
Eschmann cannot guarantee that other equipment used
in an operating theatre is properly constructed to withstand
electrical interference with adjacent equipment. This could
be a problem particularly with very old equipment (i.e. it
does not conform to the latest standards). In the unlikely
event that electromagnetic interference is experienced with
The Eschmann T20 Series of tables have been tested for
use in close proximity with other equipment and should be
installed in accordance with the EMC tables that follow in
section 9.11.7.
9.11.7EMC tables
The following guidance (in table format) is provided in line
with EN60601-1-2:2001. The table references (i.e. 201,
202, etc.) are those used within the standard. All other
tables referenced within the standard do not apply to the
T20 Series of table.
T-IM114bP53/62
Table 201 - Guidance and manufacturer’s declaration - electromagnetic emissions
Guidance and manufacturer’s declaration - electromagnetic emissions
The T20 Series of operation table are intended for use in the electromagnetic environment specified below. The
customer or the user of these operation tables should assure that they are used in such an environment
RF emissionsGroup 1The T20 Series of operation table use RF energy only for
CISPR 11its internal function. Therefore, RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissionsClass A
CISPR 11
Harmonic emissionsClass A
IEC 61000-3-2
Voltage fluctuations/Complies
flicker emissions
IEC 61000-3-3
The T20 Series of operation table are suitable for use in
all establishments other than domestic premises or those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
P54/62T-IM114b
T20
Series
OPERATION TABLE
Table 202 - Guidance and manufacturer’s declaration - electromagnetic immunity
Guidance and manufacturer’s declaration - electromagnetic immunity
The T20 Series of operation table are intended for use in the electromagnetic environment specified below. The
customer or the user of these operation tables should assure that it is used in such an environment
Electrostatic±6 kV contact±6 kV contactFloors should be wood, concrete or
discharge (ESD)ceramic tile. If floors are covered with
±8 kV air±8 kV airsynthetic material, the relative humidity
IEC 61000-4-2should be at least 30%.
Electrical fast±2 kV for power±2 kV for powerMains power quality should be that of a
transient/burstsupply linessupply linestypical commercial or hospital
±1 kV for input/±1 kV for input/environment.
IEC 61000-4-4output linesoutput lines
Surge±1 kV differential±1 kV differentialMains power quality should be that of a
modemodetypical commercial or hospital
IEC 61000-4-5environment.
±2 kV
common mode
Voltage dips, short<5%
interruptions and(>95% dip in
U
T
U
voltage variationsfor 0,5 cyclefor 0,5 cycleThe T20 Series of operation table have
on power supplydual classification (Class 2 and Internally
input lines40%
U
T
(>60% dip in
U
IEC 61000-4-11for 5 cyclesfor 5 cycles
±2 kV
common mode
<5%
U
)(>95% dip in
T
40%
)(>60% dip in
T
T
U
T
Mains power quality should be that of a
U
)typical commercial or hospital environment.
T
powered).
U
)
T
70%
U
T
(>30% dip in
U
70%
U
)(>30% dip in
T
T
U
)
T
for 25 cyclesfor 25 cycles
<5%
U
T
(>95% dip in
U
<5%
U
)(>95% dip in
T
T
U
)
T
for 5 secfor 5 sec
Power frequency3A/m3A/mPower frequency magnetic fields should
(50/60Hz)be at levels characteristic of a typical
magnetic fieldlocation in a typical commercial or
IEC 61000-4-8hospital environment.
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
T-IM114bP55/62
Table 204 - Guidance and manufacturer’s declaration - electromagnetic immunity - for EQUIPMENT
and SYSTEMS that are not LIFE-SUPPORTING.
Guidance and manufacturer’s declaration - electromagnetic immunity
The T20 Series of operation table are intended for use in the electromagnetic environment specified below. The
customer or the user of these operation tables should assure that it is used in such an environment.
Portable and mobile RF communications
equipment should be used no closer to any
part of the T20 Series of operation table,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distances
Conducted RF 3Vrms 3V d = 1.2 √
P
IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 3V/m 3V/m d = 1.2 √ P 80 MHz to 800 MHz
IEC 61000-4-3 80 MHz to 2,5 GHz
d = 2.3 √ P 800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the T20 Series of operation
table are used exceeds the applicable RF compliance level above, the T20 table should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the T20 table.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m
a
P56/62T-IM114b
T20
Series
OPERATION TABLE
Table 206 - Recommended separation distances between portable and mobile RF communications
equipment and the T20 Series of operation table - for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING.
Recommended separation distances between portable and mobile RF
communications equipment and the T20 Series of operation table
The T20 Series of operation table are intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the T20 operation table can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the T20 operation table as recommended below, according to the maximum output power of
the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum output m
power of transmitter 150 kHz to 80 Mhz 80 MHz to 800 MHz 800 MHz to 2.5 Ghz
W
3,5
d
= [——]
V
P
√
1
3,5
d
= [——]
7
d
P
√
E
1
= [——]
E
1
√
P
0.010.120.120.23
0.10.370.370.74
11.21.22.3
103.73.77.4
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
T-IM114bP57/62
APPENDIX 1
TRAINING LOG FOR THE ESCHMANN T20 SERIES OPERATION TABLE
NAME POSITION DATE INSTRUCTOR AUTHORIZED BY REVIEW DATE
Ensure that ALL the personnel listed above have been fully trained in the safe use of this T20 Operation
table. This should include a thorough understanding of all the safety notes and cautions contained within
this ‘Instructions for use’ manual.
Photocopy this page as required and ensure that ALL personnel have not passed the review date without
a refresher course. Append additional documents applicable to your unit/hospital (e.g. cross contamination
procedures) and list them below in the space provided, if they should be included in the training.
ADDITIONAL DOCUMENTS
P58/62T-IM114b
T20
Series
OPERATION TABLE
APPENDIX 2
DECONTAMINATION CERTIFICATE
Make and description of equipment/item:
Catalogue number (REF)*:
Serial Number (SN)*:
This is to certify that the above equipment/item has been decontaminated in accordance with the
attached procedure and that this certificate has been issued by the authorized person detailed
below. The equipment/item has also been suitably marked to show it has been decontaminated in
accordance with the attached procedure and that this certificate has been issued.
Certificate issued by:
Title:
Position:
Hospital/unit address:
Date:
Additional information (optional)
Reason for return:
For the attention of:
* If applicable and known
T-IM114bP59/62
APPENDIX 3
BATTERY MANAGEMENT SHEET FOR THE ESCHMANN T20 OPERATION TABLE
CAUTION
Continued use of the table batteries in the ‘critically low’ state can damage the batteries.
Batteries within the table base are mains rechargeable, to maintain peak battery performance the table batteries should be placed
‘on charge’ at the end of each day or shift, do not wait a ‘low’ battery indication, it is good practice to charge daily. Batteries must
be charged at the end of the shift when a ‘low’ indication is given and as soon as possible when a ‘critically low’ indication is given.
Batteries should be checked for adequate charge before using the table for any surgical procedure. The orange battery LEDs on the
corded handset indicate a ‘low’ battery and the red LEDs indicate a ‘critically low battery’. When the ‘critically low’ battery warning is
indicated the table cannot be operated by either of the handsets or the footswitch. To continue to operate the table either connect it to
a suitable mains supply and recharge the batteries, or use the standby batteries or control panel. To use the standby batteries the
standby battery switch must be pressed and held depressed whilst the required motions are conducted using the normal controls.
To recharge the table batteries (normal and standby) connect the mains cord supplied (do not use any other cord) into the
mains inlet. Connect the mains cord to a suitable mains supply (check voltage) and switch the supply ‘on’ if controlled by a
switch. The battery charging LEDs will illuminate ‘red’ to indicate ‘bulk’ recharge, ‘amber’ to indicate ‘top-up’ charge and ‘green’
to indicate a fully charged battery on ‘trickle’ charge. Note: The standby control panel can be used even when the ‘critically
low’ battery indication is given but note the warning given in section 4.6.5 of the ‘Instructions for Use’.
DATE
TIMETIME
‘ON’‘OFF’
BAT’Y IND’N*
MAIN STANDBY
DATE
TIMETIME
‘ON’‘OFF’
BAT’Y IND’N*
MAIN STANDBY
* BAT’Y IND’N - Indicate battery state after charge. ‘G = Green or OK’, ‘A = Amber or Low’, or ‘R = Red or Critically Low’ as
indicated on the corded handset. Please photocopy this page and use to record battery charging data.
P60/62T-IM114b
APPENDIX 4
T20
OPERATION TABLE
MANUAL HANDLING SAFETY NOTES AND ADVICE
Series
1Manual handling
1.1During configuration or adjustment of the T20 series
of table there are occasions where the user should be
aware of the safe practises to be employed during manual
handling or adjustment of parts of the table. For information
the weights of the heaviest accessories commonly used
are listed below. When lifting, carrying or fitting these
accessories it is recommended that care is taken and two
people are employed when required.
Leg section, without mattress, 11.50kg (25.3lb)
Leg section, with 50mm mattress, 14.25kg (31.4lb)
Leg section, with 75mm mattress, 15.63kg (34.4lb)
Head section, without mattress, 7.75kg (17.1lb)
Head section, with 50mm mattress, 9.75kg (21.5lb)
Head section, with 75mm mattress, 10.75kg (23.7lb)
Long and short trunk 50mm mattress, 7.25kg (16.0lb)
Long and short trunk 75mm mattress 10.88kg (23.9lb)
Footrest, without mattress, 5.00kg (11.0lb)
Footrest, including 50mm mattress, 7.00kg (15.4lb)
A divided leg section, 7.70kg (16.9lb)
A divided leg section, with 50mm mattress, 9.90kg (21.8lb)
A divided leg section, with 75mm mattress, 11.0kg (24.2lb)
Long width extender, with 75mm mattress, 6.00kg (13.2lb)
* NOTE: 50mm mattress weights included in the values
above are for the original style of 50mm mattress. The
latest 50mm mattress weights are approximately one third
of those indicated above. New mattresses can be identified
by the white breather ports on their base (see section 7.0).
1.2When adjusting a section such as the head and leg
section the weight is partially supported* by the gas springs,
but together with the weight of the patient’s limbs, the total
weight required to be lifted could be much greater than the
weight of the section and mattress*. It is not possible to
quantify this weight exactly as each situation will vary, but
Fig. 4.9 should assist in an approximation, taking into
account the patient’s actual total weight.
These ‘Instructions for use’ advise supporting the weight
of the patient during adjustment of any section, obviously
this requires the intervention of several personnel, some
supporting the patient’s limbs and others adjusting the table
sections.
* Because of the complexities involved in both the way the
patient loading acts on a section, and the assisting force
provided by the gas spring varies, as the section is raised
or lowered about its hinge, the load at a particular point
cannot be exactly stated.
1.3Manual handling is covered by legislation and the
Manual Handling Operations Regulations 1992 require
employers to avoid the need for employees to carry out
manual handling work that could cause injury by
mechanising processes, or reorganising work, so far as is
reasonably practicable. It states that it is the Duty of
Employees while at work to make full and proper use of
any system of work provided for his use by his employer
to comply with these regulations. To this end the following
publication could be used to create such a system of
working, “Manual Handling in the health services”, Health
and Safety Commission, 1998.
1.4There are no recommended maximum weights
specified in the “Manual Handling Operations Regulations
1992” or the “Health and Safety (miscellaneous
amendment) Regulations 2002” but the notes following may
be considered useful in compiling a safe working practice.
The reason no specific maximum weights are stipulated is
that factors such as an individuals capabilities (height,
gender, strength, etc.) the load (size, availability of handholds, weight, etc.) the task (motion required, duration,
distance, etc.) the environment and other factors all need
to be assessed when considering the task.
2How to lift safely
Where the ‘load’ is a patient the person or team must
communicate with the patient throughout the lifting
procedure if they are conscious. Here are some important
points, using a basic lifting operation as an example, these
points can be remembered as the four ‘Ps’.
iPLAN - Plan the lift.
Where is the load to be placed? Use appropriate
handling aids if possible. Do you need help with the
load? Remove obstructions such as discarded
wrapping materials. For a long lift, such as floor-toshoulder height, consider resting the load midway
on a table or bench in order to change grip.
iiPREPARE - Prepare load and position feet.
Make the load secure, easy to hold, safe and
compact. Clear obstructions and any floor hazards
giving good visibility and space. Position the feet
apart, giving a balanced and stable base for lifting
with the leading leg as far forward as is comfortable.
iiiPOSTURE - Adopt a good posture.
When lifting from a low level, bend the knees. But
do not kneel or overflex the knees. Keep the back
straight (tucking in the chin helps). Lean forward a
little over the load if necessary to get a good grip.
Get as close to the load as possible. Keep the
shoulders level and facing in the same direction as
the hips. Try to keep the arms within the boundary
formed by the legs, elbows tucked in. The best
T-IM114bP61/62
position and type of grip depends on the
circumstances and individual preference; but it must
be secure. A hook grip is less tiring than keeping
the fingers straight. If you need to vary the grip as
the lift progresses, do it as smoothly as possible.
ivPERFORM - Keep close to the load.
Keep the load close to your trunk for as long as
possible. Keep the heaviest side of the load next to
your trunk. If a close approach to the load is not
possible, slide it towards you before trying to lift. Lift
smoothly, keeping control of the load. Don’t twist
the trunk when turning to the side.
3Team lifting
Never be pushed into lifting with others if you are unsure
about their knowledge or ability. Say NO. The key factors
involved in team lifting are the three 'Cs':
iCommunication
There must be a team leader to direct the procedure.
The leader should give a command to lift or move,
first making sure that every member of the team
understands the command to be used. The leader
must be in control of the procedure from start to
finish.
iiCooperation
The team must lift together. It helps if the team
members are similar in height and build (but this is
not essential). The team needs to have trust and
confidence in each other. If one member of a team
fails to lift or move properly it is probable that
someone will be injured.
iiiCoordination
The team members must communicate, if only by a
nod of the head. They must indicate when they are
ready. If they are tiring or need a break, telling other
members in sufficient time to allow the load to be
lowered safely, before they drop it, is important.
P62/62T-IM114b
Eschmann Equipment, Peter Road, Lancing, West Sussex, BN15 8TJ, England.