Keep thIs ‘Service Manual’ in a safe convenient place for future reference.
Operating controls
& display
Autoclave
components
Operation cycle
Maintenance
Part removal
& replacement
Technical data
Eschmann After Sales Service Department
The Eschmann After Sales Service Department is staffed and equipped to provide advice and
assistance during normal office hours. To avoid delays when making enquires, please quote the
Model and Serial Number of your Autoclave which is shown on the Serial Number plate, the
location of which is shown below. Please ensure you include all alpha and numeric digits of the
Serial Number.
The Serial Number
F
4
L
0
S
0
4
m
VA
A
C
6
U
9
U
6
M
plate is located here
1
8
1
1
S
3
N
A
L
2
S
x
4
3
5
8
B
1
0
5
9
0
0
R
9
0
3
9
E
2
4
-
F
0
5
6
8
3
4
5
6
7
6
2
3
0
V
4
.
4
A
E
s
c
5
h
0
m
-
a
6
n
n
0
P
E
H
e
q
t
u
e
z
i
r
p
R
m
o
e
a
n
W
d
t
,
,
e
L
s
a
t
n
S
c
u
i
s
n
s
g
e
,
x
,
E
n
B
g
N
la
1
n
5
M
d
8
A
T
D
J
.
E
I
N
E
N
G
L
A
N
D
789789-1
For further information visit www.eschmann.co.uk
All correspondence relating to the after sales service of Eschmann Equipment to be addressed to :
UK Customers
Eschmann Equipment, Peter Road, Lancing, West Sussex BN15 8TJ, England.
T el: +44 (0) 1903 765040. Fax: +44 (0) 1903 762006.
Overseas Customers
Contact your local distributor. In case of doubt contact Eschmann Equipment.
Patents and Trade marks
The ESCHMANN name and logo are registered trade marks of Eschmann Holdings Limited .
“Eschmann Equipment” is a trading name of Eschmann Holdings Limit ed.
“LS5”, “Little Sister”, “QuickV ac” and “SES2555” are t rade marks of Eschmann Holdings Limit ed.
Patents : Patents Pending plus - Pat. US5090033 and Pat. GB2238407
All rights reserved. This booklet is protected by copyright. No part of it may be reproduced, stored in a
retrieval system or transmitted in any form or by any means, electronic, mechanical, photocopying,
recording or otherwise without written permission from Eschmann Holdings Limited.
The information in this publication was correct at the time of going to print. The Company, however,
reserves the right to modify or improve the equipment referred to.
If the CE mark is affixed to the product, it indicates compliance with Council Directive
93/42/EEC of 14 June 1993 concerning medical devices.
Read this service manual before servicing the
autoclave and keep it in a safe convenient place
for future reference. Always have at hand the
‘Instructions for Use’ as supplied with the
autoclave and be familiar with the cautions and
warnings contained within it. The servicing of
this autoclave should be under the control of a
competent person with sterilization training.
The autoclave should only be serviced as
specified by the manufacturer and detailed
within these service instructions.
1.1This service manual applies to the following
autoclaves, fitted with software version 1.51 or later
(software version is displayed briefly when the autoclave
is switched ‘on’ or B1 is pressed):-
SES Little Sister 5 Autoclave (from Serial Number
L5NA3C1001) REF 87-060-01
SES Little Sister 5 Vacuum Autoclave (from Serial
Number L5SA3C1001) REF 87-060-25
SES QuickVac V acuum Autoclave (from Serial Number
LQSA3D1001) REF 87-061-47
SES 2555 Autoclave (from Serial Number S5NA3C1001)
REF 87-060-61
SES 2555 Vacuum Autoclave (from Serial Number
S5SA3C1001) REF 87-060-73
SES 2555Vacuum Autoclave (from Serial Number
SQSA3D1001) REF 87-060-73
SES 2555 Autoclave - Australian (from Serial Number
S5NA3C1001) REF 87-061-09
SES 2555 Vacuum Autoclave - Australian (from Serial
Number S5SA3C1001) REF 87-061-21
SES 2555 Autoclave - Singapore and New Zealand (from
Serial Number S5NA3C1001) REF 87-060-85
SES 2555 Vacuum Autoclave - Singapore and New
Zealand (from Serial Number S5SA3C1001) R EF 87- 060-97
1.2This Service Manual contains a technical
description, complete maintenance procedures and an
illustrated list of spare parts that are available from the
Eschmann After Sales Service Department.
Note: When sterilizing lubricated dental handpieces, the
reservoir water should be changed every week to prevent
contamination of the door seal, and other elastomer
components, used in the pressure system.
1.5It is recommended that the reservoir is drained,
allowed to dry and refilled with water* every week. At every
service interval the reservoir must be removed,thoroughly cleaned and dried (see section 8.5), and then
refilled with water*.
* DO NOT USE T A P WATER Eschmann recommend the
use of ‘Sterile Water for Irrigation’ alternatively clean
distilled or deionized water, or water treated by reverse
osmosis may be used.
2Description
2.1The autoclave is a portable steam unit heated by
an electric element (or elements depending on the model).
For sterilization of wrapped loads the vacuum unit creates
a vacuum in the chamber. The unit is supplied to suit the
mains electrical supply shown in the technical data section.
2.2The autoclave is electronically controlled and has
various sterilizing programmes as detailed in the
‘Instructions for Use’. For typical sterilization cycle times,
refer to the technical data section.
2.3The required sterilizing cycle is selected by pressing
various buttons on the control panel. LEDs illuminate to
show which temperature and cycle have been selected.
The cycle is started by pressing the start button. The
sterilizing/drying cycle proceeds automatically until
complete and the printer (if fitted) starts automatically when
the start button is pressed.
2.4Indication of cycle status or error codes during a
cycle are provided by the digital display and printer (if fitted).
3Operating controls & display
3.1The following operating controls and displays are
all positioned on the front of the unit. Their use is explained
in greater detail within the ‘Instruction for Use’.
Mains switch The symbol ‘I‘ indicates ‘on’ and the symbol
‘0’ indicates ‘off’ (see 9, Fig. 2).
1.3Ensure that routine servicing is carried out at regular
intervals by either Eschmann trained personnel or suitably
trained engineers only, otherwise the warranty could be
infringed.
1.4Keep the ‘Instructions for Use’ and this ‘Service
Manual’ readily accessible for reference purposes prior to
and during operation, cleaning and servicing of the autoclave.
CAUTION
In common with other systems containing static
water reservoirs, water used in this unit can become
contaminated over a period of time, or following an
aborted cycle, and should be treated as a potential
risk of infection.
Push buttons (see Fig. 1a and 1b)
B1 Display change button. Press B1 to display the
cycle counter in display window D1.
B2 Cycle temperature button. Each time B2 is
pressed the cycle temperature alternates between
134°C and 121°C.
B3 Unwrapped load (with drying) cycle button.
B4 Unwrapped load (without drying) cycle button.
B5 Programme ‘P’ button. For details see Section 7.2.
B6 Cycle start button. (Always press and hold)
B7 Open door button. (Always press and hold)
B8 Single wrapped load (with drying) cycle button.
(V acuum autoclaves only)
P4/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
B9 Hollow instruments (without drying) cycle
button. (QuickV ac and later SES2555 models only)
Note: During some maintenance procedures you can
press and hold buttons to scroll display data quickly .
Digital displays (see Fig. 1a and 1b)
D1 Time, cycle counter, error number display or
minutes left during sterilization or drying phase.
For more details on the error codes see Table 4.
D2 Temperature display.
D3 Pressure display.
NOTE: Displays may show other information in special
functions mode, see the ‘Special functions table’.
Indicators (see Fig. 1a & 1b)
A1 Low water indicator.
A2 Process display indicator. (See Table 1 in the
instruction for Use).
A3 Door unlocked indicator.
A4 Door locked indicator.
NOTE: FOR EASY REFERENCE THESE FIGURES (1a & 1b) ARE REPEATED
ON A FOLD OUT AT THE BACK OF THIS MANUAL
CSelection indicators. These indicators illuminate
to show which function has been selected (cycle or
temperature).
4Autoclave components
4.1The following components and equipment are
incorporated in the autoclave. They are illustrated in Fig.2,
3 and 4. Fig.2 shows the autoclave with covers on Fig. 3
and 4 with covers off. In Fig. 2 and 3 the vacuum unit is
shown, some components are particular to the vacuum
unit only (vacuum and QuickV ac models are similar apart
from the control panel). Fig.4 shows the position of
components peculiar to the non-vacuum unit and some
common components.
Door close solenoid, Fig.3 item 8. This is operated when the
door closed microswitch (Fig.3 item 19) is actuated by
closing the door. The lock (Fig.3 item 12) holds the door
closed until the sterilizing cycle is complete. It will also
keep the door closed under all fault conditions. As
absence of power is also a ‘fault’ the unit power switch
must be set to ‘on’ in order to open the door. (See errors
and error clearing section).
Fig. 1a Control panel for SES2555 and LS5Fig. 1b Control panel for QuickVac
and later SES2555 Models
ST-SM33lP5/51
Fig. 2 Main component identification (covers on)
Door open solenoid, Fig.3 item 6. This is operated when the
door open button B7 is pressed to release the primary
door lock, or the controller signals it to open the door
onto the secondary catch during drying on nonvacuum autoclaves.
Secondary catch, Fig.2 item 12. To ensure that the door
does not fly open should there be residual pressure in
the chamber when the open door button is operated.
Primary door lock, Fig.3 item 12. Used to lock the door shut
during all cycles, cannot be released until Pressure
door lock (Fig.3 item 34) opens when pressure falls
below 0.2Bar(Gauge).
Door closed microswitch, Fig.3, item 19. This is used to
signal the controller that the door is properly closed.
It should operate just as the door is fully closed.
Door locked microswitch, Fig. 3, item 18. This is used to
signal the controller that the door is properly locked.
It should operate just as the door is fully locked.
Chamber pressure gauge, Fig.3 item 21. Fitted behind the
door cover, it indicates the approximate chamber
pressure (without electrical power).
Water reservoir, Fig.3 item 5. This is used to hold water
which is admitted into the chamber via the water fill
valve (Fig.3, item 11). The water reservoir also receives
hot water and steam vapour discharged from the
chamber towards the end of the cycle, via the water
discharge valve (Fig.3, item 16). The vacuum pump
(Fig.3, item 9) on vacuum units also discharges into
the water reservoir.
Reservoir water level sensor (fitted in base of reservoir
position indicated in Fig. 4 item 1). The water level
sensor in the reservoir will stop the cycle being started
if there is insufficient water in the reservoir to complete
a chamber fill. Indicator A1 will be illuminated in a low
water condition.
Gauze filter, Fig. 2 item 6. This filters the water leaving the
chamber (easily removed for cleaning).
Mains switch, Fig. 2 item 9. This switch controls mains
power to the autoclave.
Heating element, Fig.3 item 13, and Fig.4 item 4. The
heating element consists of a single immersion element
inside the chamber. The heating element is controlled
by a solid state relay on the Power PCA (Fig.3,
item 26) and protected from overheating by a manual
reset thermostat (Fig.3, item 25). Refer to
TECHNICAL DATA for heater element loading.
Power board, Fig.3 item 26. There are a number of key
functions provided by the power board:
Solid state relays (SSR) control the mains supply to
the water heater element (10A), band heater (10A)
and vacuum pump (2A).
Mechanical relay provides additional safety for heaters
and pump circuits.
Connections for display board (40-way ribbon cable,
J1), 24 volts from transformer (J2), temperature
sensors (J3), solenoid valve drives, door
microswitches and water level sensors (J4), mains to
heaters and vacuum pump (J5) and mains in (J6).
P6/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
Note: Item 34 the
‘Pressure door
lock’ is only fitted
on later units.
Note: Items 15 and
20 are shown with
‘twinlocks’ fitted, the
autoclave is supplied
with a blank plug only
Fig. 3 Main component identification (covers off) vacuum and QuickVac autoclave
cables and tubing removed for clarity.
ST-SM33lP7/51
4
1
2
3
Fig. 4 Main component identification (covers off) non-vacuum autoclave with cables,
pressure door lock and tubing removed for clarity (some parts common to vacuum autoclave).
Fuse for protecting the 24V d.c. supply (T3.15A).
Two voltage regulators.
Autoclave bleeper.
Pressure transducer with connection via pipe to
item 3). The display board interfaces with the power
board to control every aspect of autoclave
management. The main features are:
7 segment displays (D1, D2 and D3 of Fig. 1) to
provide time, temperature and pressure indication
and other information for special and protected
functions and error detail.
LEDs to indicate low water, cycle status and door
status (A1, A2, A3 and A4 of Fig. 1)
LEDs to provide visual indication of cycle selection (C
of Fig. 1).
Connections to Power PCA (40 way ribbon cable, J4),
key pad (11 way ribbon cable, J3), printer (16 way ribbon
cable, J5) and serial port (10 way ribbon cable, J2).
Battery to maintain real time clock.
Microprocessor to control the autoclave. This receives
information from the front panel and all the sensors
(temperature, pressure, door microswitches, chamber
water level, reservoir level etc.). If any errors are
detected they are shown on the display, and printed
(if a printer is fitted) as error codes.
Reset thermostat (overheat). Fig.3 item 25. The reset
thermostat is fitted underneath the unit and is
connected in series with the chamber heater element.
The reset thermostat is operated by a fluid-filled
capsule clamped to the heating element, providing
protection if the temperature of the heater surface
exceeds 250°C. It will remake electrically if the reset
button underneath the cabinet is pressed, after giving
the heater element time to cool. (Position is indicated
by a label on the left-hand side of the unit).
Cycling thermostat, Fig.4 item 3. (Non-vacuum only) This
is used to control the temperature of the chamber
heating element, in conjunction with the
microprocessor, during drying.
Band heater overheat cutout, Fig.3 item 29. (Vacuum units
only). The band heater cutout is fitted on the band
heater and is connected in series with the band heater.
It contains a bimetallic disc thermostat which operates
if the temperature of the heater surface exceeds 250°C.
The cutout will remake electrically if the reset button is
pressed when the heater has cooled.
Mains fuses, Fig. 2 item 8. Two fuses are on the front panel
of the unit rated as shown in TECHNICAL DATA,
connected in the ‘mains supply’ to the unit.
Transformer, Fig.3 item 27. The transformer converts the
mains voltage to 20V a.c. It is rated at 75VA.
Water fill solenoid valve, Fig.3 item 11. The water fill solenoid
valve controls the entry of water into the chamber. It is
electrically operated from the 24V d.c. supply generated
and signalled from the power board.
Water discharge solenoid valve, Fig.3 item 16. The water
discharge solenoid valve is used at the end of the
sterilizing cycle to allow water and steam vapour from
the chamber to pass back into the reservoir. The valve
is electrically operated from a 24V d.c. supply
generated and signalled from the power board.
Steam bleed solenoid valve, Fig.3 item 2. The steam bleed
solenoid valve is used at the beginning of the heating
phase to allow air and steam vapour to pass back into
the reservoir, and at the start of discharge to remove
P8/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
pressure before the discharge valve opens. The valve
is electrically operated from a 24V d.c. supply
generated and signalled from the power board.
Safety valve, Fig.3 item 24. The safety valve is factory set
[pressure rating 2.85Bar (Gauge)] to release excess
elevated pressure from within the chamber. It is a
primary safety device and must not be readjusted. It
should operate between 2.72 and 3.13Bar (Gauge).
Air admit solenoid valve, Fig.3 item 1. (Vacuum units only)
The air admit solenoid valve controls the admission of
air (filtered by the Bacterial Filter, Fig.2 item 1) into the
chamber during the drying phase.
Vacuum solenoid valve, Fig.3 item 3. (Vacuum units only)
When open, this valve allows the vacuum pump to
suck air and steam from the chamber. The valve is
electrically operated from a 24V d.c. supply generated
and signalled from the power board.
Vacuum pump, Fig.3 item 9. (Vacuum units only) The
vacuum pump is a two stage diaphragm pump and
used to suck air and steam from the chamber. The
pump is electrically operated from the mains supply
and signalled from the power board.
Bacterial Filter, Fig.2 item 1. (Vacuum units only) The
bacterial filter filters the air entering the chamber
during the drying phase.
Chamber temperature sensors, Fig.3 item 23. These are
used to sense the chamber temperature. Two are
provided within the single sensor unit to provide
controller accuracy.
Band heater temperature sensor, Fig.3 item 28. (Vacuum
units only) The band heater temperature sensor is
used to control the band temperature during the
drying phase.
Thermocouple entry port, Fig.3 item 22. Marked TT on the
chamber door. This is used to insert a thermocouple
into the chamber to allow the operating temperature
to be measured and, if necessary, adjusted.
Pressure test port, Fig.3 item 15#. Marked PT on the
chamber door. The pressure test port is for inserting
a pressure probe to monitor chamber pressure.
Spare test port, Fig.3 item 20#. Spare temperature or
pressure test port for use as required during test
procedures.
programme selector buttons, digital display windows
and programme indicator LEDs. It is connected to the
display board via an 11-way ribbon cable.
Pressure transducer (on the power board). The pressure
transducer monitors the pressure in the chamber and
generates chamber pressure signals for cycle
monitoring, control, and display.
Printer, Fig.2 item 5. The printer, if fitted, starts automatically
when a cycle is selected and will print out hardcopy of
the sterilization cycle. Details of printer output is given
in the ‘Instruction for Use’.
Water pump, Fig.4 item 2. (Non-vacuum units only) Used
to pump water into the chamber from the reservoir.
The pump is operated from a 24V d.c. supply generated
and signalled from the power board.
Fan, Fig.3 item 10. The fan draws air over the chamber for
rapid cooling between cycles. The fan is electrically
operated from the 24V supply and signalled from the
power board.
Band heater, Fig.3 item 4. (Vacuum units only) The band
heater heats the chamber during the drying cycle. The
heating element is controlled by a solid state relay on
the Power PCA and protected from overheating by a
manual reset thermostat (Fig.3, item 25). Refer to
TECHNICAL DATA for heater element loading.
Chamber water level sensors, Fig. 2 items 10 and 11 and
Fig.3 items 14 and 17. These signal the power board
when the minimum and maximum water levels have
been achieved in the chamber during filling. They are
connected to the power board.
Thermal fuse, not illustrated, strapped on the outside of the
chamber behind neck ring (Non vacuum units only).
The thermal fuse is connected in series with the
chamber heater element, providing protection if the
temperature of the thermal fuse exceeds 184°C
(chamber will be above 250°C). It cannot be reset and
must be replaced if it trips.
RS232 serial connection port (9-way D type), Fig.2 item 4.
This connection port can be used to link the autoclave
to a P.C. for calibration, customization, software
upgrades, fault diagnosis etc. providing the correct
equipment is available (e.g. software).
Magnetic door latch, Fig.3 item 7. (Vacuum units only)
Used to hold the primary door lock (Fig.3 item 12) in
place during the vacuum cycle.
Pressure door lock, Fig. 3 item 34. This lock stops the door
opening when the pressure inside the chamber is
above 0.2Bar (Gauge). Note: Not fitted on early units.
5Operation cycle
5.1A detailed knowledge of the operation of the
autoclave is not necessary to be able to repair it effectively;
however, a basic understanding of the various processes
of autoclave operation which occur during a cycle is given
in the following paragraphs.
CAUTION
Ensure that the Autoclave is switched off
before filling the reservoir.
# Note: Items 15 and 20 are shown with ‘twinlocks’ fitted, the autoclave is supplied with a blanking plug only
ST-SM33lP9/51
TABLE 1 - KEY TO DISPLAY A2
Phase 1 LED 1Cycle started.
LED 2Chamber minimum fill achieved.
LED 3Chamber filled with water.
Phase 2 LED 1Heater on.
LED 2Approximately 90°C reached.
LED 3Final heating phase started.
Phase 3 LED 1Sterilization phase started.
LED 2-7 One sixth of the sterilization time
completed for each LED illuminated.
Phase 4 LED 1Condensing stage started
LED 2Water removed from chamber.
LED 3Pressure approaching ambient.
Note: Eschmann recommend the use of ‘Sterile Water
for Irrigation’ alternatively clean distilled or deionized water ,
or water treated by reverse osmosis may be used.
5.2Power is switched on by selecting the power switch
(O-I) to I (Fig.2 item 9).
NOTE: If the door was locked closed when switching
‘on’ indicator A4 will flash and button B7 should be
pressed to unlock the door, the secondary catch can
then be released to open the door fully.
5.3After loading the chamber close the door and hold it
firmly shut until the automatic door lock operates. This is
shown by indicator A3 switching off and indicator A4
switching on.
NOTE: A cycle cannot be selected (as detailed in the
‘Instruction for Use’) with the door open.
5.4 When the door is closed, with the power switched
on, this is sensed by the power board via the door closed
microswitch. Indicator A4 will not illuminate until the
automatic lock has operated and the door locked
microswitch has been actuated by the primary lock. After
cycle selection (as detailed in the ‘Instruction for Use’) the
cycle is started by pressing B6. If the autoclave has a printer
it will start printing automatically .
5.5As the cycle progresses indicator A2 will show the
cycle stage reached (see Table 1) and the digital displays
D2 and D3 will display the actual chamber temperature
and pressure respectively .
5.6Control of the cycle is fully automatic with
temperature information being monitored by temperature
sensors and displayed on the front panel. Timing is
controlled by the power board and cycle times cannot be
adjusted. By comparing measured values with known time/
temperature relationships, the power board is able to detect
faults such as lack of water at the fill stage, or loss of water
and steam during the process, and it will indicate such
problems by displaying errors codes (e.g. 'Err03') which
will be accompanied by an audible warning signal (see
Table 4 for details on clearing errors and error codes).
Phase 5 LED 1Start of drying phase.
LED 2Drying phase 50% completed.
LED 3Cycle completed (audible bleep).
5.7The autoclave operates at temperatures slightly
above the usual recommended minimums. The operating
temperature for the 121°C cycle is set to 122.5°C, and the
134°C cycle is set for 135.5°C.
5.8Operating information relating to the printer is given
in the ‘Instructions for Use’.
5.9The overall time for the cycle is not fixed and
depends on many factors such as the supply voltage, the
load, and the ambient temperature. However, the power
board will ensure a satisfactory sterilization cycle even
when these factors vary over wide ranges. T ypical cycles
details are explained in section 6.
6Cycle details
6.1A typical vacuum cycle (e.g. 134° Wrapped extended
cycle) progresses through several stages. The following
information provides details of the controls actuated and
the resultant functions performed by the autoclave during
selection and running of a typical vacuum sterilization cycle.
Note: The water reservoir must be above minimum and
the door must be closed.
Select temperature 134° with B2 button. (121° light
extinguished, 134° light illuminated or vice versa for
each press of button B2).
Push B8 button. (Wrapped LED illuminated, all other
cycle LED’s extinguished).
Push B5 button. (Extend LED illuminates, gives
sterilization time of 18 minutes at 134°C).
Push B6 button. (First LED on progress indicator
illuminates. If fitted, the printer records time and cycle
data. Mains relay is on. V acuum pump is on.
V acuum inlet valve open. All other valves closed).
Wait for pressure to drop to 32kPa. (Vacuum inlet valve
closed. V acuum pump off. If fitted, the printer outputs
‘TPV’ followed by the time and temperature and
pressure values).
P10/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
Programmed pause. (If chamber pressure does not rise
by more than 0.9kPa in the last minute, water fill
valve opens).
Wait for water to reach minimum level. (Second LED on
progress indicator illuminates).
Wait for water to reach maximum level. (W ater fill valve
closes. Third LED on progress indicator illuminates).
Immersion heater on. (Fourth LED on progress indicator
illuminates. T emperature rises).
Wait for the pressure to rise to 50kPa above atmospheric.
(Fifth LED on progress indicator illuminates. Steam
bleed valve opens. If fitted, the printer outputs ‘TPP’
followed by the time and temperature and pressure
values).
Programmed pause.(Steam bleed valve closes.
Sixth LED on progress indicator illuminates. Heater
off. If pressure above 115kPa, fan on until 115kPa
achieved. Fan off, vacuum inlet valve opens and
pump starts).
Wait for pressure to fall 20kPa below atmospheric. (Pump
off, vacuum valve closes, heater comes on).
Wait for temperature to reach 135.0°. (Seventh LED on
progress indicator illuminates. If fitted, the printer
outputs ‘TSS’ followed by current time, temperature
and pressure values, D1 indicates remaining
sterilization time in minutes, initially 18 mins).
Eighth to thirteenth LED’s on progress indicator illuminate
at regular intervals. (T emperature should not deviate
from 135.5° by more than ±1.5° during sterilisation
plateau. If fitted, the printer outputs the temperature
and pressure values every 20 seconds. Immersion
heater duty drops to approximately 20%).
sounds. If fitted, the printer outputs ‘TSE’ followed
by the temperature and pressure values. Fan on.
Fourteenth LED on progress indicator illuminates.
Steam bleed valve pulses to exhaust steam).
Wait for 20 seconds. (Air admit valve opens).
(Note: Above two steps repeat until end of drying time)
Wait for half of specified drying time. (Eighteenth LED
on progress indicator illuminates).
Wait until five minutes before the end of the specified
drying time. (Band heater disabled. Fan on).
Wait until the end of specified drying time. (Nineteenth
LED on progress indicator illuminates. Mains relay
off. Three beeps sound. If fitted, the printer outputs
‘TDE’ followed by the temperature and pressure
values, ‘Cycle Complete’ followed by the time. D1
displays time).
6.2A typical non-vacuum cycle (e.g. 134° unwrapped
with drying extended cycle) progresses through several
stages. The following information provides details of the
controls actuated and the resultant functions performed
by the autoclave during selection and running of a typical
non-vacuum sterilization cycle. Note: The water reservoir
must be above minimum and the door must be closed.
Select temperature 134° with B2 button. (121° light
extinguished, 134° light illuminated or vice versa for
each press of button B2).
Push B3 button. (With drying LED illuminated, all other
cycle LED’s extinguished).
Push the button B5. (Extend LED illuminated, gives
sterilization time of 18 minutes at 134°C).
Push the button B6. (First LED on progress indicator
illuminates. If fitted, the printer records time and cycle
data. Water pump is on. Water fill valve opens.
Steam bleed valve opens. All other valves are
closed).
Wait for water to reach minimum level. (Second LED on
progress indicator illuminates).
Wait for water to reach maximum level. (W ater fill valve
closes. Water pump is off. Third LED on progress
indicator illuminates).
Wait for temperature to drop to 120°C. (W ater discharge
valve opens. Band heater enabled.
Wait for the water in the chamber to drain. (Fifteenth
LED on progress indicator illuminates).
Wait for rate of pressure drop below 40kPa/minute.
(Sixteenth LED on progress indicator illuminates.
Seventeenth LED on progress indicator illuminates.
If fitted, the printer outputs ‘TDS’ followed by the
temperature and pressure values. Vacuum pump
on. Vacuum inlet valve opens at 115kPa. Beep
sounds. D1 displays drying time left).
Wait for chamber pressure to fall 5kPa below
atmospheric. (Air admit valve opens).
Wait for pressure to return to atmospheric. (Air admit
valve closes).
Mains relay on. (Immersion heater on. Fourth LED on
progress indicator illuminates. Temperature rises).
Wait for the temperature to be above 90° and to rise at
over 1.8°/minute. (Fifth LED on progress indicator
illuminates).
Programmed pause and condition check. (Steam bleed
valve closed. Sixth LED on progress indicator
illuminates).
Wait for temperature to reach 135.0°. (Seventh LED on
progress indicator illuminates. If fitted, the printer
outputs ‘TSS’ followed by the current time and the
temperature and pressure values. Immersion heater
duty drops. Temperature should not deviate from
135.5° by more than ±1.5° during sterilisation
plateau. Eighth to thirteenth LED’s on progress
ST-SM33lP11/51
indicator illuminate at regular intervals. If fitted, the
printer outputs the temperature and pressure values
every 20 seconds. D1 indicates remaining
sterilization time, initially 18 mins).
sounds. If fitted, the printer outputs ‘TSE’ followed
by the temperature and pressure values, and
maximum and minimum sterilization temperatures).
Fan on. (Fourteenth LED on progress indicator
illuminates. Steam bleed valve pulses to exhaust
steam).
Wait for temperature to drop to 120°C. (W ater discharge
valve opens).
Wait for the water in the chamber to drain. (Chamber
minimum indicator off. Fifteenth LED on progress
indicator illuminates).
Wait for rate of pressure drop to go below 60kPa/minute.
(Sixteenth LED on progress indicator illuminates.
Seventeenth LED on progress indicator illuminates.
Fan off at 1 15kPa. Door catch releases and door is
opened by residual pressure. Beep sounds. D1
displays remaining drying time).
display D1 will show F0, each time the ‘P’ button B5 is
pressed, this will increase from F0 up to F8* (F4,F5,F6
vacuum autoclave only), and then back to F0 again, until
either the start button B6 is pressed to select that function,
or, the autoclave is switched ‘off’ or B7 is pressed (with
the door open) to cancel special function selection. To
access F0 the autoclave PIN number (password) is
required. If this is known the protected functions are
available and are detailed in T able 3 - protected functions.
* Door must be shut before access to F4, F5, F6 and F7 is
permitted.
7.2Protected functions.
Protected functions are accessed through special function
F0 (see above) using the autoclave PIN number (password).
If this is known the protected functions are available as
detailed in Table 3 - protected functions. When access to
the protected functions has been made all displays and
indicators are blank except D1 which shows P0. For each
press of B5 the “P” number will increase by 1 and then back
to P0. Cycle to the number required and press button B6 to
perform that function as detailed in Table 3.
7.3Errors and faults.
Immersion heater duty to 4%. (Steam bleed valve opens.
If fitted, the printer outputs ‘TDS’ followed by the
temperature and pressure values. D1 displays
drying time left).
Wait for half of specified drying time. (Eighteenth LED
on progress indicator illuminates).
Wait for second half of specified drying time. (Nineteenth
LED on progress indicator illuminates. Immersion
heater disabled. Mains relay off. Fan on. Three
beeps sound. If fitted, the printer outputs ‘TDE’
followed by the temperature and pressure values,
‘Cycle Complete’ followed by the time. D1 displays
time).
7Maintenance
For easy reference to the buttons, display and indicator
numbers (B1, D1, A1, etc.) used within the following
sections note that Figs. 1a and 1b are repeated on a fold
out page at the back of this manual. Open out this page to
provide a quick reference guide.
7.1Special functions.
Switch ‘off’ the mains switch (item 9, Fig.2). Press and
hold in the ‘P’ button B5 whilst switching the mains switch
back ‘on’ (display D1 will show F0) to make available
several special functions as detailed in Table 2 - Special
functions. When the autoclave is switched on in this way
display D1 only will be illuminated (with other displays as
required depending on which function is selected). Initially
7.3.1 Power failure
If, after switching ‘on’ power there is no visual display , first
check power supply connections, both fuses at the front of
the unit (see Fuse Renewal) and the fuse in the mains
plug (if it is a fused plug). If power fails during a cycle,
check supply conditions and fuses at the front of the unit
and in the mains plug if fused, also check the mains lead
is correctly connected at both ends. Once power is restored
the display D1 will show the error number (D2 and D3 will
be blank) until cleared by switching the autoclave ‘off’ and
then ‘on’ whilst depressing the start button B6.
7.3.2 Aborted cycles and errors
If an error occurs during a cycle (i.e. any time after pressing
the start button B6), the cycle will abort and provided power
supply to the unit is maintained, the error number will be
indicated by the visual display D1 (displays D2 and D3 will
continue to show actual temperature and pressure). To
display the last error number in display D1 press B1 and
B5 together at any time. Errors are detailed in Table 4 Fault diagnosis with error number details.
7.3.3 Reset thermostat (overheat)
Some solutions to an error number will indicate that the
manual reset thermostat may have tripped and caused
the error. Should the heating element manual reset
thermostat operate this can be reset by pressing the ‘reset
button’ on the underside of the autoclave (approximately
11cm in on the left hand side underneath the unit as
indicated on the left hand side of the unit).
P12/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
7.4Calibration.
The autoclave can be calibrated by running the calibration
cycles (Protected functions P3, P4 and P5 consecutively)
This will require the positioning of probes (from calibrated
instruments or chart recorders) into the autoclave chamber
through the test ports. The three available test ports are
positioned in the autoclave door, one marked TT (a twinlock)
for the chamber test probe, one marked PT (a blank plug)
for the chamber pressure probe and a spare blank plug for
probes that could be positioned in the load should this be
required. To insert a probe remove the port cover and
carefully insert ROUNDED tipped probes into the chamber
through the seal built into the port, take care not to damage
the port seals and replace the cover after use. During
calibration of the band heater a temperature probe should
be placed under the spring clip provided on the heater, this
will require the top cover to be removed to gain access. It is
important that the top cover is replaced for accurate
calibration.
Running protected function P5 will enable chamber
temperature and pressure calibration, band heater
calibration and a pressure relief valve test. The latter can
be avoided if this test is not required during calibration.
7.5LED Function test.
Annually
iCheck that the pressure relief valve operates
correctly at the set pressure.
iiInspect the pressure system for integrity.
iiiCheck door microswitches and interlocks.
ivCheck door locking mechanism for integrity
vCheck pressure indicators for correct operation.
NOTE:
Annual inspections should only be undertaken by a
‘Competent Person’. Eschmann can provide
comprehensive service contracts which cover preventive
maintenance to ensure trouble free operation of your
autoclave as well as six monthly and annual inspections
of the pressure system to satisfy the requirements of the
Pressure Systems Safety Regulations 2000.
To test the function of all LEDs (i.e. to check that they are
capable of illumination) press and hold buttons B6 and B7
together with the door open. All LEDs will be illuminated
while the buttons are held depressed together.
7.6Connections, air, water and electrical.
During maintenance refer to Figs. 5, 10 and 1 1 for details
on all autoclave connections. Fig. 5 shows all the air and
water connections whilst Figs. 10 and 1 1 show the electrical
connections for the non-vacuum and vacuum units
respectively. (Note: Figs. 10 and 11 open out at the back
of this manual). When refitting tie wraps to pipes, use ‘Tie
wrap tool 5401’ set at tension 2.
7.7Safety Checks
The following safety checks, detailed in the ‘Instructions
for Use’, are repeated here as they are important.
Operators should ensure that the following safety checks
are carried out.
Weekly
iCheck that the door opens and closes easily .
iiCheck the door seal for any signs of damage.
iiiCheck the secondary door catch latches effectively.
ivCheck for any obvious escape of steam or water
during a cycle (apart from the normal escape via
reservoir vents).
Any of the above defects should be attended to by a
‘Competent Person’ immediately and the autoclave should
not be used until repair has been effected (see note below).
ST-SM33lP13/51
Fig. 5 Pipe connections and identification
P14/51ST-SM33l
LS5/SES2555 & QuickVac
VACUUM & NON- VACUUM AUTOCLAVES
TABLE 2 - SPECIAL FUNCTIONS (See section 7.1 which provides instructions for
accessing these special functions and Fig. 1a & 1b foldout, at the back of this manual)
DISPLAYFUNCTIONMETHOD OF USE
F0Protected functionsTo use protected P functions requires knowledge of the PIN code (code is in HEX
format). With display D1 showing F0 press the start button B6, display D1 shows 0.
Use B2 and B3 to adjust D1 to show the first digit of the PIN number, then press B4
to confirm it. D1 will show a 0 in the next position of the PIN number, use B2, B3
and B4 to adjust and confirm the second digit. Continue until all digits have been
entered, D1 will show P0. Press button B5 to cycle through the protected functions
P0-P9. Use the information in T able 3 - Protected Functions to carry out the selected
P function and return the autoclave to normal operation mode.
F1Printer test(Note: The printer should be loaded with paper before entering this special
functions mode). With display D1 showing F1 press the start button B6 to start
the printer test, a character set printout, display D1 shows F11. To stop the test
press button B7, the printer will stop at the end of the test cycle, display D1
shows F0. T o resume normal use switch ‘off’ the mains switch (item 9, Fig. 2) and
then switch back ‘on’, or press B7 with the door open.
F2Cycle historyWith display D1 showing F2 press the start button B6 to start printing a summary
of the last five cycles completed by the unit (cycle temperature, cycle type, cycle
count, date and time at start of cycle, date and time at end of cycle and the cycle
result). Display D1 returns to F0. To resume normal use switch ‘off’ the mains
switch (item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open.
F3Set delayed startWith display D1 showing F3 press the start button B6 to enter the delayed start
time setting mode and delayed start initiation, display D1 will show .
If the start hour displayed in D2 is correct (as set last time) and the minutes can
be assumed correct (as set last time) press the start button B6 to programme the
autoclave to start automatically at this set time. The display D1 will change to F0
when B6 is pressed. Press B7 with the door open to exit the setting mode.
If the start time is not correct when entering function F3 (display D1 showing
, display D2 showing the hour set in 24 hour clock mode) press B2 to
increase the hour (or B3 to reduce it). Press B4 to confirm the hour when it is
correct. Display D1 will show , to increase the minutes (shown in D2)
press B2, or B3 to reduce them. Press B6 to confirm the minutes. The display D1
will show F0. Press B7 with the door open to exit the setting mode.
The autoclave should be left switched ‘on’ with the required load in place, the
door shut and the required cycle selected as follows. When the door is shut a ‘ ’
flashes ‘on and off’ after the time in display D1. This continues while the program
selection and any adjustments to it are made (see ‘Instructions for Use’ if ‘’
appears in display D2). When cycle selection has been completed and B6 is
pressed to confirm them, the ‘ ’ stops flashing. The cycle selected will be run
automatically , starting at the time set, a ‘ ’ will be displayed in D1 after the time to
show that a delayed start time has been set.
On completion of the cycle the autoclave is ready for the load to be removed and
normal use resumed without the need to exit from the special functions mode.
Note: If the same delayed start time is required each day the setting can be
entered quickly once display D1 shows F3, by pressing (B6 B6 B7).
F4Leak TestWith display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONLY)
Note. To conduct this test the
chamber temperature must be
below 50°C to avoid an Error 38,
and the door closed before
selecting F4.
display D1 shows F4, press B6. Indicator A4 will flash, press B7 to open the door,
indicator A3 will illuminate. Close the door, A4 will illuminate, press B6 this starts
an automatic leak test which checks to see if the autoclave system is still pressure
tight. Display D3 will show the chamber pressure, after a few minutes display D2
will show the pressure rise to the nearest 0.1kPa. After a further 10 minutes if the
pressure has not risen by more than 1.3kPa display D2 will show indicating
the autoclave has passed the test (or to indicate the autoclave has failed the
test) and an audible sound will be heard. (continues on page 16)
ST-SM33lP15/51
TABLE 2 (continued) - SPECIAL FUNCTIONS
DISPLAYFUNCTIONMETHOD OF USE
F4 (continued from page 15)Press button B5 and wait until display D2 and D3 go blank. To resume normal
use switch ‘off’ the mains switch (item 10, Fig. 2) and then switch back ‘on’, or
press B7 with the door open.
F5Air Detection TestWith display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONL Y)
Note. To conduct this test the
chamber temperature must be
below 50°C to avoid an Error 38,
and the door closed before
selecting F5. Connect a needle
valve (calibrated to provide a
leak of between 1.0 and 1.3kPa
per minute) via a short length of
6mm hose to the antibacterial
filter before pressing B6.
F6Steam penetration test. With display D1 showing F0 close the autoclave door (if open), press B5 until
(Vacuum Autoclave ONLY)
Note. To conduct this test the
door must be closed before
selecting F6.
display D1 shows F5, press B6. Indicator A4 will flash, press B7 to open the door ,
indicator A3 will illuminate. Close the door , A4 will illuminate, press start button B6.
This starts an automatic air detection test which checks if the autoclave system
detects a leak if present (it performs a leak test similar to F4 first, to ensure no leaks
are present, then admits air through the needle valve). Display D3 shows chamber
pressure, after a few minutes display D2 will show pressure rise to the nearest
0.1kPa. After a further 10 minutes if pressure has not risen by more than 1.3kPa
display . Display D2 will then show or according to result (test aborts on
). If first test is the air valve will open to conduct detection test. After one
minute if pressure <34kPa the leak rate is measured for one minute (if =>34kPa
test aborts), display D3 will show pressure. If the rate of pressure rise is more than
0.9kPa display D2 will show indicating the autoclave has passed the test, or
to indicate the autoclave has failed the test (i.e. pressure rise equal or less
than 0.9kPa). To resume normal use switch ‘off’ the mains switch (item 9, Fig. 2)
and then switch back ‘on’, or press B7 with the door open.
display D1 shows F6, press B6. Indicator A4 will flash, press B7 to open the
door, indicator A3 will illuminate. Place the Eschmann LS5 Vacuum T ype-S and
QuickV ac Steam Penetration T est Device in it’ s pouch (see test device instructions
for use) and into the chamber. Close the door, A4 will illuminate, press the start
button B6, this will start a special B8 cycle (no drying), D2 and D3 will show
temperature and pressure as normal during the cycle. At the end of the cycle the
normal audible sound will be heard. Press B7 to open the door, display D1 will
be showing F0. To resume normal use switch ‘off’ the mains switch (item 9, Fig.
2) and then switch back ‘on’, or press B7 with the door open. (NOTE: This
function can also be used to carry out HTM2010 Performance Qualification (PQ)
Tests in conjunction with the ‘Instructions for Testing’ manual ST -IM55).
F7Disinfect functionWith display D1 showing F7 load the autoclave with the load to be disinfected
Note. To conduct this test the
door must be closed before
selecting F7.
F8Adjust date and timeWith display D1 showing F8 press the start button B6 to enter the adjust date and
and close the door. Press the start button B6 to start the disinfect function (a 2.5
minute disinfect time at 1 10°C), D1 will continue to show F7 during the cycle and
D2 and D3 will show temperature and pressure. At the end of the cycle the normal
audible sound will be heard. Press B7 to open the door and remove the load,
display D1 will be showing F0 . T o resume normal use switch ‘off’ the mains switch
(item 9, Fig. 2) and then switch back ‘on’, or press B7 with the door open. (NOTE:
The ex-factory condition is for this function to be disabled.)
time function. Display D1 will show and D2 will show the year, to increase
the year press B2 (or B3 to reduce it) when correct press button B4 to confirm
the year. The display D1 will show and D2 will show the month in numerical
format, to increase the month press B2 (or B3 to reduce it) when correct press
button B4 to confirm. The display D1 will show and D2 will show the day in
numerical format, to increase the day press B2 (or B3 to reduce it) when correct
press button B4 to confirm. The display D1 will show and D2 will show the
hour , to increase the hour press B2 (or B3 to reduce it) when correct press button
B4 to confirm. The display D1 will show and D2 will show the minute, to
increase the minute press B2 (or B3 to reduce it) when correct press button B6,
display D1 shows F0 display D2 and D3 will be blank. The time and date are now
set (seconds were set to zero when B6 was pressed). Press B7 with the door
open to continue normal use.
P16/51ST-SM33l
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.