allations requiring connection to external exhaust systemError! Bookmark not defined.
6.1 Canopy/thimble connection 25
6.2 Hard connection 25
6.3 IFlow alarm monitoring system instructions 26
VII 、Label instructions 27
APPENDIX 1 Error! Book
DOWNFLOW/HAND INTAKE SETTINGS Error! Book
BIOHAZARD CABINETS-Class Error! Book
mark not defined.
mark not defined.
mark not defined.
FIGURE1 BASE FRAME ASSEMBLY 27
FIGURE 2 BOTH SIDE PANELS
_Toc265052624
FIGURE 4 MAIN PARTS Error! Book
mark not defined.
FIGURE 5 CONTROL PANEL 30
FIGURE 6 MEASUREMENT OF BIOHAZARD CABINET 28
FIGURE 7: AIR FLOW PATTERN AND PROTECTED AREA 30
FIGURE 8: FRONT ACCESS PANEL REMOVAL 30
FIGURE 9: GLASS CLIP 32
FIGURE 10: PRESSURE MONITOR CALIBRATION SETUP Error! Book
FIGURE 11: EXHAUST SYSTEM Error! Book
mark not defined.
mark not defined.
FIGURE 12: ELECTRICAL SCHEMATIC 33
VIII 、W
arranty
3
I、Introduction
Investigators and technicians concerned with microbiological safety have, for many years,
utilized specialized containment enclosures to limit their exposure to harmful pathogens. In
addition, it is recognized that exacting research, as well as routine pathology, is often
compromised by environmental contamination. The ENVIRCO Class II Biohazard Cabinet is
designed to limit the contamination exposure of both the worker and the work when handling
biohazard material. The Class II enclosure is an open-face cabinet with a high volume of air
circulated through an internal HEPA filter. Air is drawn into the open front of the cabinet and
exhausted through a HEPA filter. The controlled airflow direction and volume balance provide
protection for the worker, the work (sample), and against cross-contamination within the
cabinet.
This product has been tested to meet the requirements of CAN/CSA-C22.2 No.61010-1,
secondedition, including Amendment 1,or a later version of the same standard incorporating
the same level of testing requirements.
Important
Volatile, friable or poisonousradioactive substancesarenot allowed in the
non-exhaust-cover cabinet.
Sterilize the cabinet before turning off PAO test switch or wash water valve
Do not hinder airflow flowing out of HEPA filter.
Disconnect power before accessingthe operational board.
Maximum power of this equipment is 1000W.
Connect Only To Properly Grounded Outlet. Do Not Remove Ground Pin.
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
If needed, please use the emergency power cut-off, located on the front panel.
4
Technical Specifications
Model: 11228 -001
Nominal Size: 4Ft.(1.2 meters)
Internal Dimensions: (WxDxH): 55.9″x 33.4″x 85
Internal Size: (W x D x H): 48″x 26.1″x25.6″ (1220 x 665 x 65
Voltage Resistance: 2 seconds no breakdown with DC1200V
Cabinet Construction: Work area: 1.5mmstainless steel
Frame: cold-roll steel sheets with electrostatic coating
Working Environment: Use indoors only
Environmental temperature: 59°F– 95°F (15Ԩ-35Ԩ)
Relative humidity: ≤75
%;
Range of atmospheric pressure: 70kPa~106kPa
5
Power supply: AC 120V±10%, 60Hz ±1Hz
Serial number is on cabinet; see operational board mark.
II、Installation
2.1Position
The Biological safety cabinet should be positioned in an airflow-protected zone in order to
prevent the airflow from being adversely affected. This can be caused by ventilation
systems,air-conditioning or personnel movement.If the test shows that the interference of
other airflow overpasses the inlet speed into cabinet air flow, indoor infectious gas can enter
the biological safety cabinet work area. Therefore, it is necessary that the cabinet be
positioned correctly.The relationship between the safety cabinet exhaust air and indoor air
exhaust ventilation pipes should also be notedso that the air emission exhaustsfrom the top of
the cabinetand the cabinet itselfshould be positioned to avoid restrictions from its
exhaust.Biological safety cabinets should be in the downstream airflow direction;if possible,
the best way to install this cabinet is to reserve a 12” clearance on all four sides of the cabinet.
Pay special attention to the electrical line: every biological safety cabinet must adopt its
own exclusive line directly into the socket. Do not, under any circumstances, allow the use of
an extension cord without permission from the manufacturer to modify the powercord.
This biological safety cabinet can operate 24 hours a daysuccessively. All circuits and motors
are adequate and have protection measures.
2.2 Disassembly and handling
When the equipment is received,please check the box for outside damage conditions. Broken
glass or other damage should be notedupon receipt of equipment andcarrier should be
notified immediately.
1) If you do not receive return authorization, the equipment cannot be returned to the factory.
2) If the equipment is in cold weather transportation, it should be placed near the desired
location and left at room temperature for 24 hours prior toassembly.
3) First, remove all surface protection material. Read Item #7 below. Do not try to lift the
packageunder frame before the device is completely exposed.
4) When removing the packaging materials, carefully put assembly parts to one side. Excess
packaging tape residue can be wiped clean with alcohol.
5) Check all parts in the transport making sure there are no damaged parts. If damaged parts
are found, immediately report them to the transportation company. If there is damage, do not
open the packaging;the consigner may be asked to inspect the box.
6) Lift the Equipment off the packageunder frame and release the four stop blocks.
7) The security cabinet is very heavy and is equipped with highly efficient HEPA filters that
should be moved carefully. The safety cabinet should be transferred by a vehiclesuch as a
fork lift.
8) The cabinet should not be upside down during shipment under any circumstances.If so
contact the carrier
6
2.3 Assembly and install
1) Assemble the under-frame, see installation Figure 1.
FIGURE 1: BASE FRAME ASSEMBLY
1. Seven M10×50 bolts
2. Horizontal support without hole is fastened with two M10×50 bolts
3. Horizontal support with hole is fastened with two M10×50 bolts
4. Left-side support
5. T-style support is fastened with three M10×50 bolts
6. Right-side support
7
2) Unload the two side panels see Figure 2.
3) Place cabinet onto under-frame and secure with bolts. Bolts are in the cabinet work zone
bag.Make the connection between the under frame and cabinet body (sees Figure 3), and
then mount side panels;
FIGURE 3: CONNECTION OF BASE AND CABINET BODY
1. RIGHT PANEL
2. CABINET
BODY
8
3. LEFT PANEL
4. M10 BOLT
5. BASE FRAME
4) Adjust under-frame and level the working surface to ensure the maximum stability on floor.
5) Remove the internal packing materials, empty all parts and clean all debris from inside of
the cabinet
Note: Any residual debris may lead to damage to the HEPA filters and fan.
6) Check packing list to ensure that all materials and accessories are included.
7) Check the internal and external safety cabinet carefully and inspect all components that
may have loosened during transport. If necessary, check for filter leakage. Check the front
panel fixed bolts; the bolts should not be too tight. If you need to check the air filter surface,
you can release the fixed wing nut bolt, remove from the front of the wing nut and then
carefully inspect.
8) Check to make sure the work surface is firmly installed.
9) If the safety cabinet is equipped with water and air, connection to building or other services
must be in accordance with local codes.
10) Install the exhaust HEPA protection device on the exhaust filter: The protection device will
be placed around the filter opening, fixed with a nut.
NOTE: Do not, underany circumstances, put hand into exhaust filter.
11) Remove the four screws and lift out work surface panel. Remove the plug on the
down-board;secure the valve connecting the pieces to the down-board, connecting the drain
valve and making sure valve is in a closed drain position, (parallel to the ground).
12) The device is equipped with a HEPA exhaust protection device on the top. If the
equipment request is to connect to the external exhaust system (used in toxic, poisonous gas
or steam), be sure to refer to Chapter 6 of the external exhaust systems.
NOTE: The minimum distance between exhaust HEPA filter and the ceiling is 11.8”
(300mm). If the exhaust HEPA filter is blocked, sufficient air will not flow into the outlet
operating area. The inlet operating the open airflow speed should be checked on
location, when the exhaust HEPA filter is at the top of the clean room, less than 15.9”
(406mm).
13) Install ultraviolet and fluorescent lamp.
14) Installation of the main components, see Figure 4.
9
FIGURE 4: MAIN PARTS
1. Exhaust filter protector hood
2. Operational board
3. UV light
4. Control panel
5. Light
6. Power lock
7. Base frame
8. Emergency stop
10
2.4 Check
1) Connect the biological safety cabinet’s dedicated power line to power outlet on a 120V
power source, with a total power of 20 amps. The biological safety cabinet must be the only
equipment on the circuit; see the electrical equipment requirements data on the cover.
2) Turn the key on the front panel and press the power button located on the control panel.
There will be a delay on the fan in order to achieve the required speed. In order to prevent too
much current, the fan needs a delay of about 10-15 seconds in order to achieve full speed
before motor is running. Similarly, the fan also needsa 10-15 second delay to stop. Stable
voltage is provided by a built-in voltage regulator circuit for the motor/blower which helps to
eliminate air fluctuations. Please remember to turn fan of before shutting down.
3) Press the LIGHT button on control panel to turn on the lights. If the lamp is not bright, check
the lamptube or socket. It may have come loose between the lamp and lamp holder during
shipping.
4) Press the UV button on the control panel, turn on the UV lamp and check to make sure the
UV light is working properly.
NOTE: For safety purpose the UV lights will only operate when the glass dooris fully
closed.
5) Press key SOCKET button on the control panel and check the safety cabinet's internal
power socket. If the safety cabinet is equipped with two sockets, this button will control both.
11
III. Adjustments and Testing
3.1 Adjust on site
Before packing and transport, this safety cabinet has been tested and confirmed in our
factory; however, another test after installation must be performed. Only qualified personnel
can adjust this device and in order to ensure the accuracy of the on-site testing, this test
equipment must be calibrated. If you cannot implement this operation, please contact
ENVIRCO to provide a local on-site testing services company.
On site adjustment should include the following
- Check the safety cabinet inflow
- Check the safety cabinet internal decline airflow rate
- Check airflow inside the cabinet by using smoke simulator
- When the cabinet is connected to the building’s exhaust system, balance the air emission.
Toensure that your new equipment is accurate, perform an on-site inspection. The equipment
cannot be used without this pre-test, performed by qualified personnel.
Do not adjust the fan speed without fully testing decline of the wind speed and the inflow air.
The manufacturer has already set the correct air flow; no further adjustments should be
required. Do not change the valve set of the ratio between the exhaust and air supply unless a
qualified person thinks it is necessary to do.
3.2 Tests and standards
3.2.1 Filter leakage test
The standard aerosol testing for the integrity of the HEPA filtersis generally performed
annually. Qualified personnel must use a calibrated device when aerosol testing. A
referenced list will be provided by the factory.
:
NOTE: Place connection between supplyair and exhaust filters hose test for pressure
(the middle hose) photometer upper (PAO test point connection).
3.2.2 Air adjustment at the opening of the work area
The exhaust HEPA filter should be installed on the top of the biosafety cabinet. The air volume
exhausted is equal to the air volume entering from the work area opening. The average
velocity entering the opening of the work area has been identified during the design of the bio
safety cabinet. The test should conform to the rules of NSF49. The opening air velocity testing
data is in APPENDIX 1. To adjust the fan speed, adjustments should be according to the data
specified in APPENDIX 1.
12
Loading...
+ 26 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.