The measuring device described in this Operating Manual is to be used only for measuring the flow
rate of conductive fluids in closed pipes.
A minimum conductivity of 20 μS/cm is required for measuring demineralized water. Most liquids
can be measured as of a minimum conductivity of 5 μS/cm.
Examples:
• Acids, alkalis,
• Drinking water, wastewater, sewage sludge,
• Milk, beer, wine, mineral water, etc.
Resulting from incorrect use or from use other than that designated the operational safety of the
measuring devices can be suspended. The manufacturer accepts no liability for damages being
produced from this.
1.2Installation, commissioning and operation
Please note the following:
• Installation, connection to the electricity supply, commissioning and maintenance of the device
must be carried out by trained, qualified specialists authorized to perform such work by the
facility's owner-operator. The specialist must have read and understood this Operating Manual
and must follow the instructions it contains.
• The device must be operated by persons authorized and trained by the facility's owner-operator.
Strict compliance with the instructions in the Operating Manual is mandatory.
• With regard to special fluids, including fluids used for cleaning, Endress+Hauser will be happy to
assist in clarifying the corrosion-resistant properties of wetted materials.
However, minor changes in temperature, concentration or in the degree of contamination in the
process may result in variations in corrosion resistance. For this reason, Endress+Hauser does not
accept any responsibility with regard to the corrosion resistance of wetted materials in a specific
application.
The user is responsible for the choice of suitable wetted materials in the process.
• If welding work is performed on the piping system, do not ground the welding appliance through
the Promag flowmeter.
• The installer must ensure that the measuring system is correctly wired in accordance with the
wiring diagrams. The transmitter must be grounded apart from when special protective measures
are taken (e.g. galvanically isolated SELV or PELV power supply)
• Invariably, local regulations governing the opening and repair of electrical devices apply.
1.3Operational safety
Please note the following:
• Measuring systems for use in hazardous environments are accompanied by separate Ex
documentation, which is an integral part of this Operating Manual. Strict compliance with the
installation instructions and ratings as stated in this supplementary
documentation is mandatory. The symbol on the front of this Ex documentation indicates the
approval and the certification body (e.g.
• The measuring device complies with the general safety requirements in accordance with
EN 61010-1, the EMC requirements of IEC/EN 61326 and NAMUR Recommendations NE 21
and NE 43.
• Depending on the application, the seals of the process connections of the Promag H sensor require
periodic replacement.
0 Europe, 2 USA, 1 Canada).
Endress+Hauser5
Safety instructionsPromag 50
• When hot fluid passes through the measuring tube, the surface temperature of the housing
increases. In the case of the sensor, in particular, users should expect temperatures that can be
close to the fluid temperature. If the temperature of the fluid is high, implement sufficient
measures to prevent burning or scalding.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser distributor will supply you with current information and updates to these
Operating Instructions.
1.4Return
• Do not return a measuring device if you are not absolutely certain that all traces of hazardous
substances have been removed, e.g. substances which have penetrated crevices or diffused
through plastic.
• Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be
charged to the owner-operator.
1.5Notes on safety conventions and icons
The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the
factory in a condition in which they are safe to operate. The devices comply with the applicable
standards and regulations in accordance with EN 61010-1 "Safety requirements for electrical
equipment for measurement, control and laboratory use".
The devices can, however, be a source of danger if used incorrectly or for anything other than the
designated use. Consequently, always pay particular attention to the safety instructions indicated in
this Operating Manual by the following icons:
#
Warning!
"Warning" indicates an action or procedure which, if not performed correctly, can result in injury
or a safety hazard. Comply strictly with the instructions and proceed with care.
Caution!
"
"Caution" indicates an action or procedure which, if not performed correctly, can result in incorrect
operation or destruction of the device. Comply strictly with the instructions.
!
Note!
"Note" indicates an action or procedure which, if not performed correctly, can have an indirect
effect on operation or trigger an unexpected response on the part of the device.
6Endress+Hauser
Promag 50Identification
Promag 50
-20°C (-4°F) <Tamb<+60°C (+140°F)
IP67 / NEMA/Type 4XOrder Code:
Ser.No.:
TAG No.:
50PXX-XXXXXXXXXXXX
12345678901
ABCDEFGHJKLMNPQRST
20-55VAC/16-62VDC
50-60Hz
I-OUT (HART), f-OUT
15VA/W
i
EPD / MSÜ
STATUS-IN
2
3
4
5
9
8
1
N12895
ECC
6
7
2Identification
2.1Device designation
The flow measuring system consists of the following components:
• Promag 50 transmitter
• Promag D, Promag L, Promag W, Promag P or Promag H sensor
In the compact version, the transmitter and sensor form a single mechanical unit; in the remote
version they are installed separately.
2.1.1Nameplate of the transmitter
A0005412
Fig. 1: Nameplate specifications for the "Promag 50" transmitter (example)
1Ordering code/serial number: See the specifications on the order confirmation for the meanings of the individual
letters and digits.
2Power supply, frequency, power consumption
3Additional information:
EPD/MSÜ: with Empty Pipe Detection
ECC: with electrode cleaning
4Outputs available:
I-OUT (HART): with current output (HART)
f-OUT (HART): with frequency output
STATUS-IN: with status input (power supply)
5Reserved for information on special products
6Observe device documentation
7Reserved for additional information on device version (approvals, certificates)
8Permitted ambient temperature range
9Degree of protection
Endress+Hauser7
IdentificationPromag 50
-20°C (-4°F)<Tamb<+60°C (+140°F)NEMA/Type4X
50PXX-XXXXXXXXXXXX
1.0000/0000
–10 ...150°C/+14 ...300°F°C°F
PFA
12345678901 RY
ABCDEFGHJKLMNPQRST
DN100 DIN EN PN40/pnom =PS= 40bar
EPD/MSÜ, R/B
TM:
Order Code:
Materials:
K-factor:
Ser.No.:
TAG No.:
PROMAG P
1
2
3
7
13
12
1.4435/316L
Electrodes:
0.2% CAL
4
5
6
10
11
i
9
8
IP67
2007
N12895
2.1.2Nameplate of the sensor
A0004374
Fig. 2: Nameplate specifications for the "Promag" sensor (example)
1Ordering code/serial number: See the specifications on the order confirmation for the meanings of the individual
letters and digits.
2Calibration factor with zero point
3Nominal diameter / Pressure rating
4Fluid temperature range
5Materials: lining/measuring electrodes
6Reserved for information on special products
7Permitted ambient temperature range
8Observe device documentation
9Reserved for additional information on device version (approvals, certificates)
10Calibration tolerance
11Additional information (examples):
– EPD/MSÜ: with Empty Pipe Detection electrode
– R/B: with reference electrode
12Degree of protection
13Flow direction
Fig. 3: Nameplate specifications for transmitter (example)
1Serial number
2Possible configuration of current output
3Possible configuration of relay contacts
4Terminal assignment, cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L– for DC
5Signals present at inputs and outputs, possible configuration and terminal assignment (20 to 27),
see also "Electrical values of inputs/outputs"
6Version of device software currently installed
7Installed communication type, e.g.: HART, PROFIBUS PA, etc.
8Information on current communication software (Device Revision and Device Description), e.g.:
Dev. 01 / DD 01 for HART
9Date of installation
Endress+Hauser9
10Current updates to data specified in points 6 to 9
2.2Certificates and approvals
The devices are designed to meet state-of-the-art safety requirements in accordance with sound
engineering practice. They have been tested and left the factory in a condition in which they are
safe to operate.
The devices comply with the applicable standards and regulations in accordance with EN 61010-1
"Safety requirements for electrical equipment for measurement, control and laboratory use" and
with the EMC requirements of IEC/EN 61326/A1.
The measuring system described in this Operating Manual is therefore in conformity with the
statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the
device by affixing to it the CE mark.
The measuring system meets the EMC requirements of the Australian Communications and Media
Authority (ACMA)
IdentificationPromag 50
2.3Registered trademarks
KALREZ® and VITON
Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP
®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
®
HART
Registered trademark of the HART Communication Foundation, Austin, USA
HistoROM™, S-DAT®, Field Xpert™, FieldCare
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
®
®
, Fieldcheck®, Applicator
®
10Endress+Hauser
Promag 50Installation
3Installation
3.1Incoming acceptance, transport and storage
3.1.1Incoming acceptance
On receipt of the goods, check the following:
• Check the packaging and the contents for damage.
• Check the shipment, make sure nothing is missing and that the scope of supply matches your
order.
3.1.2Transport
The following instructions apply to unpacking and to transporting the device to its final location:
• Transport the devices in the containers in which they are delivered.
• Do not remove the protective plates or caps on the process connections until you are ready to
install the device. This is particularly important in the case of sensors with PTFE linings.
Special notes on flanged devices
Caution!
"
• The wooden covers mounted on the flanges from the factory protect the linings on the flanges
during storage and transportation. In case of Promag L they are additionally used to hold the lap
joint flanges in place. Do not remove these covers until immediately before the device in the
pipe.
• Do not lift flanged devices by the transmitter housing, or the connection housing in the case of
the remote version.
#
Transporting flanged devices DN 300 (12")
Use webbing slings slung round the two process connections. Do not use chains, as they could
damage the housing.
Warning!
Risk of injury if the measuring device slips. The center of gravity of the assembled measuring device
might be higher than the points around which the slings are slung.
At all times, therefore, make sure that the device does not unexpectedly turn around its axis or slip.
a0004294
Fig. 4: Transporting sensors with DN 300 ( 12")
Endress+Hauser11
InstallationPromag 50
Transporting flangeddevices DN > 300 (> 12")
Use only the metal eyes on the flanges for transporting the device, lifting it and positioning the
sensor in the piping.
Caution!
"
Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing. This
would buckle the casing and damage the internal magnetic coils.
a0004295
Fig. 5: Transporting sensors with DN > 300 (> 12")
3.1.3Storage
Please note the following:
• Pack the measuring device in such a way as to protect it reliably against impact for storage
(and transportation). The original packaging provides optimum protection.
• The storage temperature corresponds to the operating temperature range of the measuring
transmitter and the appropriate measuring sensors ä 99.
• Do not remove the protective plates or caps on the process connections until you are ready to
install the device. This is particularly important in the case of sensors with PTFE linings.
• The measuring device must be protected against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
• Choose a storage location where moisture does not collect in the measuring device. This will help
prevent fungus and bacteria infestation which can damage the liner.
12Endress+Hauser
Promag 50Installation
h 2 x DN³
3.2Installation conditions
3.2.1Dimensions
The dimensions and installation lengths of the sensor and transmitter can be found in the "Technical
Information" for the device in question. This document can be downloaded as a PDF file from
www.endress.com. A list of the "Technical Information" documents available is provided in the
"Documentation" section on ä 116.
3.2.2Mounting location
Entrained air or gas bubble formation in the measuring tube can result in an increase in measuring
errors.
Avoid the following locations:
• Highest point of a pipeline. Risk of air accumulating!
• Directly upstream from a free pipe outlet in a vertical pipeline.
A0008154
Fig. 6: Mounting location
Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and
the consequent risk of damage to the lining of the measuring tube. Information on the lining's
resistance to partial vacuum can be found on ä 103.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or
peristaltic pumps. Information on the measuring system's resistance to vibration and shock can be
found on ä 99.
A0003203
Fig. 7: Installation of pumps
Endress+Hauser13
InstallationPromag 50
5 x DN
2 x DN
³
³
h
2
1
Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration.
The Empty Pipe Detection function (EPD ä 73) offers additional protection by detecting empty
or partially filled pipes.
Caution!
"
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable
to install a cleaning valve.
A0008155
Fig. 8: Installation in a partially filled pipe
Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes whose length h 5 m
(16.4 ft). This precaution is to avoid low pressure and the consequent risk of damage to the lining
of the measuring tube.
This measure also prevents the system losing prime, which could cause air pockets. Information on
the lining's resistance to partial vacuum can be found on ä 103.
A0008157
Fig. 9: Measures for installation in a down pipe
1Vent valve
2Pipe siphon
hLength of down pipe
14Endress+Hauser
Promag 50Installation
A
1
22
A
3
3.2.3Orientation
An optimum orientation position helps avoid gas and air accumulations and deposits in the
measuring tube. However, Promag offers the additional Empty Pipe Detection (EPD) function to
ensure the detection of partially filled measuring tubes, e.g. in the case of degassing fluids or varying
process pressure:
• Electrode Cleaning Circuit (ECC) for applications with accretive fluids, e.g. electrically
conductive deposits ( "Description of Device Functions" manual).
• Empty Pipe Detection (EPD) ensures the detection of partially filled measuring tubes, e.g. in the
case of degassing fluids ( ä 73)
• Exchangeable Measuring Electrodes for abrasive fluids ( ä 92)
Vertical orientation
This is the ideal orientation for self-emptying piping systems and for use in conjunction with Empty
Pipe Detection.
Fig. 10: Vertical orientation
Horizontal orientation
The measuring electrode plane should be horizontal. This prevents brief insulation of the two
measuring electrodes by entrained air bubbles.
Caution!
"
Empty Pipe Detection functions correctly only when the measuring device is installed horizontally
and the transmitter housing is facing upward ( å 10). Otherwise there is no guarantee that
Empty Pipe Detection will respond if the measuring tube is only partially filled or empty.
A0008158
Fig. 11: Horizontal orientation
1EPD electrode for the detection of empty pipes (not with Promag D and Promag H (DN 2 to 15; 1/12" to ½"))
2Measuring electrodes for signal detection
3Reference electrode for the potential equalization (not with Promag D and H)
Endress+Hauser15
A0003207
InstallationPromag 50
5 x DN³
³ 2 x DN
L
Inlet and outlet run
If possible, install the sensor upstream from fittings such as valves, T-pieces, elbows, etc. The
following inlet and outlet runs must be observed in order to meet accuracy specifications:
• Inlet run: 5 × DN
•Outlet run: 2 × DN
A0003210
Fig. 12: Inlet and outlet runs
3.2.4Vibrations
Secure the piping and the sensor if vibration is severe.
Caution!
"
If vibrations are too severe, we recommend the sensor and transmitter be mounted separately.
Information on resistance to vibration and shock can be found on ä 99.
A0003208
Fig. 13: Measures to prevent vibration of the device (L > 10 m (32.8 ft))
16Endress+Hauser
Promag 50Installation
100
10
0.5
d / D
[mbar]
0.60.70.80.9
1m/s
2m/s
3m/s
4 m/s
5m/s
6m/s
7m/s
8m/s
1
D
d
max. 8°
3.2.5Foundations, supports
If the nominal diameter is DN 350, mount the sensor on a foundation of adequate load-bearing
strength.
Caution!
"
Risk of damage.
Do not support the weight of the sensor on the metal casing: the casing would buckle and damage
the internal magnetic coils.
A0003209
Fig. 14: Correct support for large nominal diameters (DN 350)
!
3.2.6Adapters
Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in largerdiameter pipes.
The resultant increase in the rate of flow improves measuring accuracy with very slow-moving
fluids. The nomogram shown here can be used to calculate the pressure loss caused by reducers and
expanders.
Note!
The nomogram only applies to liquids of viscosity similar to water.
1.Calculate the ratio of the diameters d/D.
2.From the nomogram read off the pressure loss as a function of flow velocity (downstream from
the reduction) and the d/D ratio.
Fig. 15: Pressure loss due to adapters
A0011907
Endress+Hauser17
InstallationPromag 50
3.2.7Nominal diameter and flow rate
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum velocity of flow is between 2 and 3 m/s (6.5 to 9.8 ft/s)
The velocity of flow (v), moreover, has to be matched to the physical properties of the fluid:
• v < 2 m/s (v < 6.5 ft/s): for abrasive fluids
• v > 2 m/s (v > 6.5 ft/s): for fluids producing buildup
!
Note!
Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor
( ä 17).
Recommended flow (SI units)
Nominal
diameter
[mm]Min./max. full scale value (v 0.3 or 10 m/s) in [dm³/min]
2––––0.06 to 1.8
4––––0.25 to 7
8––––1 to 30
15–––4 to 1004 to 100
259 to 300–9 to 3009 to 3009 to 300
32––15 to 50015 to 500–
4025 to 700–25 to 70025 to 70025 to 700
5035 to 110035 to 110035 to 110035 to 110035 to 1100
6560 to 200060 to 200060 to 200060 to 200060 to 2000
8090 to 300090 to 300090 to 300090 to 300090 to 3000
100145 to 4700145 to 4700145 to 4700145 to 4700145 to 4700
125–220 to 7500220 to 7500220 to 7500–
[mm]Min./max. full scale value (v 0.3 or 10 m/s) in [m³/h]
150–20 to 60020 to 60020 to 600–
200–35 to 110035 to 110035 to 1100–
250–55 to 170055 to 170055 to 1700–
300–80 to 240080 to 240080 to 2400–
350–110 to 3300110 to 3300110 to 3300–
375–140 to 4200140 to 4200––
400–140 to 4200140 to 4200140 to 4200–
450–180 to 5400180 to 5400180 to 5400–
500–220 to 6600220 to 6600220 to 6600–
600–310 to 9600310 to 9600310 to 9600–
700–420 to 13500420 to 13500––
750–480 to 15200480 to 15200––
800–550 to 18000550 to 18000––
900–690 to 22500690 to 22500––
1000–850 to 28000850 to 28000––
1050–950 to 40000950 to 40000––
1200–1250 to 400001250 to 40000––
1400––1700 to 55000––
1600––2200 to 70000––
1800––2800 to 90000––
2000––3400 to 110000––
Promag DPromag LPromag WPromag PPromag H
18Endress+Hauser
Promag 50Installation
Recommended flow (US units)
Nominal diameterPromag DPromag LPromag WPromag PPromag H
[inch]Min./max. full scale value (v 0.3 or 10 m/s) in [gal/min]
/"––––0.015 to 0.5
1
/"–– ––0.07 to 2
/"–– ––0.25 to 8
/"–––1.0 to 271.0 to 27
1"2.5 to 80–2.5 to 802.5 to 802.5 to 80
1
/"––4 to 1304 to 130–
/"7 to 1907 to 1907 to 1907 to 1907 to 190
1
2"10 to 30010 to 30010 to 30010 to 30010 to 300
2
/"16 to 50016 to 50016 to 50016 to 50016 to 500
3"24 to 80024 to 80024 to 80024 to 80024 to 800
4"40 to 125040 to 125040 to 125040 to 125040 to 1250
5"–60 to 195060 to 195060 to 1950–
6"–90 to 265090 to 265090 to 2650–
8"–155 to 4850155 to 4850155 to 4850–
10"–250 to 7500250 to 7500250 to 7500–
12"–350 to 10600350 to 10600350 to 10600–
14"–500 to 15000500 to 15000500 to 15000–
15"–600 to 19000600 to 19000––
16"–600 to 19000600 to 19000600 to 19000–
18"–800 to 24000800 to 24000800 to 24000–
20"–1000 to 300001000 to 300001000 to 30000–
24"–1400 to 440001400 to 440001400 to 44000–
28"–1900 to 600001900 to 60000––
30"–2150 to 670002150 to 67000––
32"–2450 to 800002450 to 80000––
36"–3100 to 1000003100 to 100000––
40"–3800 to 1250003800 to 125000––
42"–4200 to 1350004200 to 135000––
48"–5500 to 1750005500 to 175000––
[inch]Min./max. full scale value (v 0.3 or 10 m/s) in [Mgal/d]
54"––9 to 300––
60"––12 to 380––
66"––14 to 500––
72"––16 to 570––
78"––18 to 650––
Endress+Hauser19
InstallationPromag 50
200
100
5
10100200
L
max
[m][m]
[µS/cm]
L
max
[ft]
2006000
400
3.2.8Length of connecting cable
In order to ensure measuring accuracy, comply with the following instructions when installing the
remote version:
• Fix cable run or lay in armored conduit. Cable movements can falsify the measuring signal
especially in the case of low fluid conductivities.
• Route the cable well clear of electrical machines and switching elements.
• Ensure potential equalization between sensor and transmitter, if necessary.
• The permitted connecting cable length L
is determined by the fluid conductivity ( å 16).
max
A minimum conductivity of 20 μS/cm is required for measuring demineralized water. Most
liquids can be measured as of a minimum conductivity of 5 μS/cm.
• The maximum connecting cable length is 10 m (32.8 ft) when empty pipe detection
(EPD ä 73) is switched on.
Fig. 16: Permissible cable length for the remote version
Area shaded gray = permitted range
Lmax = connecting cable length in [m]
Fluid conductivity in [μS/cm]
A0010734
20Endress+Hauser
Promag 50Installation
5
1
2
3
4
3.3Installation instructions
3.3.1Installing the Promag D sensor
The sensor is installed between the pipe flanges with a mounting kit. The device is centered using
recesses on the sensor ( ä 22).
!
"
Note!
A mounting kit consisting of mounting bolts, seals, nuts and washers can be ordered separately
( ä 76). Centering sleeves are provided with the device if they are required for the installation.
Caution!
When installing the transmitter in the pipe, observe the necessary torques ( ä 23).
When installing the sensor, make sure that the seals used do not project into the pipe cross-section.
Caution!
"
Risk of short circuit! Do not use electrically conductive sealing compounds such as graphite! An
electrically conductive layer could form on the inside of the measuring tube and short-circuit the
measuring signal.
!
Endress+Hauser21
Note!
Use seals with a hardness rating of 70° Shore.
InstallationPromag 50
1
1
1
1
1
1
1
1
1
1
1
1
2
2
22
3
3
3
3
3
3
3
3
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
Arrangement of the mounting bolts and centering sleeves
The device is centered using recesses on the sensor. The arrangement of the mounting bolts and the
use of the centering sleeves supplied depend on the nominal diameter, the flange standard und the
pitch circle diameter.
Process connection
EN (DIN)ANSIJIS
DN 25 to 40
(DN 1" to 1 ½")
A0010896A0010824A0010896
DN 50
(DN 2")
DN 65
DN 80
(DN 3")
DN 100
(DN 4")
A0010897A0010825A0010825
–––––––––––––––––
A0012170
A0010898A0010827A0010826
A0012171
A0012168A0012168A0012169
1 = Mounting bolts with centering sleeves
2 = EN (DIN) flanges: 4-hole with centering sleeves
3 = EN (DIN) flanges: 8-hole without centering sleeves
22Endress+Hauser
Promag 50Installation
Screw tightening torques (Promag D)
Please note the following:
• The tightening torques listed below are for lubricated threads only.
• Always tighten the screws uniformly and in diagonally opposite sequence.
• Overtightening the screws will deform the sealing faces or damage the seals.
• The tightening torques listed below apply only to pipes not subjected to tensile stress.
The tightening torques apply to situations where an EPDM soft material flat seal (e.g. 70 Shore) is
used.
Tightening torques, mounting bolts and centering sleeves for EN (DIN) PN 16
* A centering sleeve is not required. The device is centered directly via the sensor housing.
Mounting bolts
Centering sleeve
length
Tightening torque [lbf · ft]
with a process flange with a
Endress+Hauser23
InstallationPromag 50
3.3.2Installing the Promag L sensor
Caution!
"
• The protective covers mounted on the two sensor flanges (DN 50…300) are used to hold the lap
joint flanges in place and to protect the PTFE liner during transportation. Consequently, do not
remove these covers until immediately before the sensor is installed in the pipe.
• The covers must remain in place while the device is in storage.
• Make sure that the lining is not damaged or removed from the flanges.
!
Note!
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the customer.
The sensor is designed for installation between the two piping flanges.
• Observe in any case the necessary screw tightening torques on ä 25
• If grounding disks are used, follow the mounting instructions which will be enclosed with the
shipment
• To comply with the device specification, a concentrical installation in the measuring section is
required
Fig. 18: Installing the Promag L sensor
Seals
Comply with the following instructions when installing seals:
• Hard rubber lining additional seals are always necessary.
• Polyurethane lining no seals are required.
• PTFE lining no seals are required.
• For DIN flanges, use only seals according to EN 1514-1.
• Make sure that the seals do not protrude into the piping cross-section.
Caution!
"
Risk of short circuit!
Do not use electrically conductive sealing compounds such as graphite! An electrically conductive
layer could form on the inside of the measuring tube and short-circuit the measuring signal.
Ground cable
• If necessary, special ground cables for potential equalization can be ordered as an accessory
( ä 76).
• Information on potential equalization and detailed mounting instructions for the use of ground
cables can be found on ä 55.
a0004296
24Endress+Hauser
Promag 50Installation
Screw tightening torques (Promag L)
Please note the following:
• The tightening torques listed below are for lubricated threads only.
• Always tighten the screws uniformly and in diagonally opposite sequence.
• Overtightening the screws will deform the sealing faces or damage the seals.
• The tightening torques listed below apply only to pipes not subjected to tensile stress.
Promag L tightening torques for EN (DIN)
Nominal diameterEN (DIN)Max. tightening torque
Hard rubberPolyurethanePTFE
[mm]Pressure rating
[bar]
50PN 10/164 × M 16-1540
65*PN 10/168 × M 16-1022
80PN 10/168 × M 16-1530
100PN 10/168 × M 16-2042
125PN 10/168 × M 16-3055
150PN 10/168 × M 20-5090
200PN 108 × M 20-65130
250PN 1012 × M 20-5090
300PN 1012 × M 20-55100
350PN 612 × M 20111120350PN 1016 × M 20112118400PN 616 × M 209098400PN 1016 × M 24151167450PN 616 × M 20112126450PN 1020 × M 24153133500PN 620 × M 20119123500PN 1020 × M 24155171600PN 620 × M 24139147600PN 1020 × M 27206219700PN 624 × M 24148139700PN 1024 × M 27246246800PN 624 × M 27206182800PN 1024 × M 30331316900PN 624 × M 27230637900PN 1028 × M 30316307-
1000PN 628 × M 272182081000PN 1028 × M 334024051200PN 632 × M 303192991200PN 1032 × M 36564568-
350Table E12 × M 24203-400Table E12 × M 24226-450Table E16 × M 24226-500Table E16 × M 24271-600Table E16 × M 30439-700Table E20 × M 30355-750Table E20 × M 30559-800Table E20 × M 30631--
900Table E24 × M 30627-1000Table E24 × M 30634-1200Table E32 × M 30727--
AS 2129Threaded
fasteners
Pressure ratingHard rubberPolyurethanePTFE
Max. tightening torque
Promag L tightening torques for AS 4087
Nominal dia-
meter
[mm][Nm][Nm][Nm]
350PN 1612 × M 24203--
375PN 1612 × M 24137--
400PN 1612 × M 24226--
450PN 1612 × M 24301--
500PN 1616 × M 24271--
600PN 1616 × M 27393--
700PN 1620 × M 27330--
750PN 1620 × M 30529--
800PN 1620 × M 33631--
900PN 1624 × M 33627--
1000PN 1624 × M 33595-1200PN 1632 × M 33703--
AS 4087Threaded
fasteners
Pressure ratingHard rubberPolyurethanePTFE
Max. tightening torque
26Endress+Hauser
Promag 50Installation
3.3.3Installing the Promag W sensor
!
"
Note!
Bolts, nuts, seals, etc. are not included in the scope of supply and must be supplied by the customer.
The sensor is designed for installation between the two piping flanges.
• Observe in any case the necessary screw tightening torques on ä 27
• If grounding disks are used, follow the mounting instructions which will be enclosed with the
shipment
a0004296
Fig. 19: Installing the Promag W sensor
Seals
Comply with the following instructions when installing seals:
• Hard rubber lining additional seals are always necessary.
• Polyurethane lining no seals are required.
• For DIN flanges, use only seals according to EN 1514-1.
• Make sure that the seals do not protrude into the piping cross-section.
Caution!
Risk of short circuit!
Do not use electrically conductive sealing compounds such as graphite! An electrically conductive
layer could form on the inside of the measuring tube and short-circuit the measuring signal.
Ground cable
• If necessary, special ground cables for potential equalization can be ordered as an accessory
( ä 76).
• Information on potential equalization and detailed mounting instructions for the use of ground
cables can be found on ä 55
Screw tightening torques (Promag W)
Please note the following:
• The tightening torques listed below are for lubricated threads only.
• Always tighten the screws uniformly and in diagonally opposite sequence.
• Overtightening the screws will deform the sealing faces or damage the seals.
• The tightening torques listed below apply only to pipes not subjected to tensile stress.
Endress+Hauser27
InstallationPromag 50
Tightening torques for:
•EN (DIN) ä 28
• JIS ä 30
•ANSI ä 29
• AWWA ä 30
• AS 2129 ä 31
• AS 4087 ä 31
Promag W tightening torques for EN (DIN)
Nominal diameterEN (DIN)
[mm]Pressure rating [bar]Hard rubberPolyurethane
25PN 404 × M 12-15
32PN 404 × M 16-24
40PN 404 × M 16-31
50PN 404 × M 164840
65*PN 168 × M 163227
65PN 408 × M 163227
80PN 168 × M 164034
80PN 408 × M 164034
100PN 168 × M 164336
100PN 408 × M 205950
125PN 168 × M 165648
125PN 408 × M 248371
150PN 168 × M 207463
150PN 408 × M 2410488
200PN 108 × M 2010691
200PN 1612 × M 207061
200PN 2512 × M 2410492
250PN 1012 × M 208271
250PN 1612 × M 249885
250PN 2512 × M 27150134
300PN 1012 × M 209481
300PN 1612 × M 24134118
300PN 2516 × M 27153138
350PN 612 × M 20111120
350PN 1016 × M 20112118
350PN 1616 × M 24152165
350PN 2516 × M 30227252
400PN 616 × M 209098
400PN 1016 × M 24151167
400PN 1616 × M 27193215
400PN 2516 × M 33289326
450PN 616 × M 20112126
450PN 1020 × M 24153133
450PN 1620 × M 27198196
450PN 2520 × M 33256253
500PN 620 × M 20119123
500PN 1020 × M 24155171
500PN 1620 × M 30275300
500PN 2520 × M 33317360
600PN 620 × M 24139147
600PN 1020 × M 27206219
600 *PN 1620 × M 33415443
600PN 2520 × M 36431516
700PN 624 × M 24148139
700PN 1024 × M 27246246
700PN 1624 × M 33278318
Threaded
fasteners
Max. tightening torque [Nm]
28Endress+Hauser
Promag 50Installation
Nominal diameterEN (DIN)
[mm]Pressure rating [bar]Hard rubberPolyurethane
700PN 2524 × M 39449507
800PN 624 × M 27206182
800PN 1024 × M 30331316
800PN 1624 × M 36369385
800PN 2524 × M 45664721
900PN 624 × M 27230637
900PN 1028 × M 30316307
900PN 1628 × M 36353398
900PN 2528 × M 45690716
1000PN 628 × M 27218208
1000PN 1028 × M 33402405
1000PN 1628 × M 39502518
1000PN 2528 × M 52970971
1200PN 632 × M 30319299
1200PN 1032 × M 36564568
1200PN 1632 × M 45701753
1400PN 636 × M 33430398
1400PN 1036 × M 39654618
1400PN 1636 × M 45729762
1600PN 640 × M 33440417
1600PN 1040 × M 45946893
1600PN 1640 × M 5210071100
1800PN 644 × M 36547521
1800PN 1044 × M 45961895
1800PN 1644 × M 5211081003
2000PN 648 × M 39629605
2000PN 1048 × M 4510471092
2000PN 1648 × M 5613241261
2510K4 × M 16-19
2520K4 × M 16-19
3210K4 × M 16-22
3220K4 × M 16-22
4010K4 × M 16-24
4020K4 × M 16-24
5010K4 × M 164033
5020K8 × M 162017
6510K4 × M 165545
6520K8 × M 162823
8010K8 × M 162923
8020K8 × M 204235
10010K8 × M 163529
10020K8 × M 205648
12510K8 × M 206051
12520K8 × M 229179
15010K8 × M 207563
15020K12 × M 228172
20010K12 × M 206152
20020K12 × M 229180
25010K12 × M 2210087
25020K12 × M 24159144
30010K16 × M 227463
30020K16 × M 24138124