Endress+Hauser NMS81 Operating Manual

BA01459G/00/EN/04.18 71411069 2018-07-13
Valid as of version
Products Solutions Services
Operating Instructions
Proservo NMS81
Tank Gauging
Proservo NMS81
Order code:
Ext. ord. cd.:
Ser. no.:
www.endress.com/deviceviewer
Endress+Hauser Operations App
XXXXXXXXXXXX
XXXXX-XXXXXX
XXX.XXXX.XX
Serial number
1.
3.
2.
A0023555
2 Endress+Hauser
Proservo NMS81 Table of contents

Table of contents

1 About this document ................ 4
1.1 Document function ..................... 4
1.2 Symbols .............................. 4
1.3 Documentation ........................ 7
1.4 Registered trademarks ................... 9
2 Basic safety instructions ........... 10
2.1 Requirements for the personnel ........... 10
2.2 Designated use ....................... 10
2.3 Workplace safety ...................... 11
2.4 Operational safety ..................... 11
2.5 Product safety ........................ 11
3 Product description ................ 12
3.1 Product design ........................ 12
4 Incoming acceptance and product
identification ..................... 13
4.1 Incoming acceptance ................... 13
4.2 Product identification .................. 13
4.3 Storage and transport .................. 15
5 Installation ....................... 17
5.1 Requirements ........................ 17
5.2 Mounting of the device ................. 35
5.3 Post-installation check .................. 45
6 Electrical connection .............. 46
6.1 Terminal assignment ................... 46
6.2 Connecting requirements ................ 61
6.3 Ensuring the degree of protection .......... 62
6.4 Post-connection check .................. 62
7 Operability ........................ 63
7.1 Overview of the operation options ......... 63
7.2 Structure and function of the operating
menu .............................. 64
7.3 Access to the operating menu via the local or
remote display and operating module. ...... 66
7.4 Access to the operating menu via the service
interface and FieldCare ................. 79
7.5 Access to the operating menu via Tankvision
Tank Scanner NXA820 and FieldCare ....... 80
8 System integration ................ 83
8.1 Overview of the Device Description files
(DTM) .............................. 83
9 Commissioning .................... 84
9.1 Terms related to tank measurement ........ 84
9.2 Initial settings ........................ 85
9.3 Calibration .......................... 87
9.4 Configuring the measuring device .......... 95
9.5 Configuring the tank gauging application ... 107
9.6 Advanced settings .................... 129
9.7 Simulation .......................... 129
9.8 Protecting settings from unauthorized
access ............................. 129
10 Operation ....................... 130
10.1 Reading off the device locking status ...... 130
10.2 Reading off measured values ............ 130
10.3 Gauge commands .................... 131
11 Diagnostics and troubleshooting .. 137
11.1 General trouble shooting ............... 137
11.2 Diagnostic information on local display ..... 138
11.3 Diagnostic information in FieldCare ....... 141
11.4 Overview of the diagnostic messages ...... 143
11.5 Diagnostic list ....................... 149
11.6 Reset measuring device ................ 150
11.7 Device information ................... 150
11.8 Firmware history ..................... 150
12 Maintenance .................... 151
12.1 Maintenance tasks .................... 151
12.2 Endress+Hauser services ............... 151
13 Repair ........................... 152
13.1 General information on repairs ........... 152
13.2 Spare parts ......................... 152
13.3 Endress+Hauser services ............... 153
13.4 Return ............................. 153
13.5 Disposal ........................... 153
14 Accessories ...................... 154
14.1 Device-specific accessories .............. 154
14.2 Communication-specific accessories ....... 157
14.3 Service-specific accessories .............. 157
14.4 System components ................... 157
15 Operating menu .................. 158
15.1 Overview of the operating menu .......... 158
15.2 "Operation" menu ..................... 169
15.3 "Setup" menu ........................ 185
15.4 "Diagnostics" menu .................... 316
Index ................................. 328
Endress+Hauser 3
About this document Proservo NMS81
DANGER
WARNING
CAUTION
NOTICE

1 About this document

1.1 Document function

These Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.

1.2 Symbols

1.2.1 Safety symbols

Symbol Meaning
DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury.
NOTE!
This symbol contains information on procedures and other facts which do not result in personal injury.

1.2.2 Electrical symbols

Symbol Meaning
Direct current
Alternating current
Direct current and alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system.
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.
Equipotential connection
A connection that has to be connected to the plant grounding system: This may be a potential equalization line or a star grounding system depending on national or company codes of practice.
4 Endress+Hauser
Proservo NMS81 About this document
A
1.
1.

1.2.3 Tool symbols

Symbol Meaning
Torx screwdriver
A0013442
Flat blade screwdriver
A0011220
Cross-head screwdriver
A0011219
Allen key
A0011221
Hexagon wrench
A0011222

1.2.4 Symbols for certain types of information

Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
Notice or individual step to be observed
, 2., 3.… Series of steps
Result of a step
Help in the event of a problem
Visual inspection

1.2.5 Symbols in graphics

Symbol Meaning
1, 2, 3 ... Item numbers
, 2., 3.… Series of steps
A, B, C, ... Views
A-A, B-B, C-C, ... Sections
Endress+Hauser 5
About this document Proservo NMS81
-
.
Symbol Meaning
Hazardous area
Indicates a hazardous area.
Safe area (non-hazardous area)
Indicates the non-hazardous area.

1.2.6 Symbols at the device

Symbol Meaning
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
Temperature resistance of the connection cables
Specifies the minimum value of the temperature resistance of the connection cables.
6 Endress+Hauser
Proservo NMS81 About this document

1.3 Documentation

For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer : Enter the serial number from the nameplate (www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate.

1.3.1 Technical Information (TI)

The Technical Information contains all the technical data on the device and provides an overview of the accessories and other products that can be ordered for the device.
Device Technical Information
Proservo NMS81 TI01249G

1.3.2 Brief Operating Instructions (KA)

The Brief Operating Instructions contain all the essential information from incoming acceptance to initial commissioning.
Device Brief Operating Instructions
Proservo NMS81 KA01203G

1.3.3 Operating Instructions (BA)

The Operating Instructions contain all the information that is required in various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
It also contains a detailed explanation of each individual parameter in the operating menu (except the Expert menu). The description is aimed at those who work with the device over the entire life cycle and perform specific configurations.
Device Operating Instructions
Proservo NMS81 BA01459G

1.3.4 Description of Device Parameters (GP)

The Description of Device Parameters provides a detailed explanation of each individual parameter in the 2nd part of the operating menu: the Expert menu. It contains all the device parameters and allows direct access to the parameters by entering a specific code. The description is aimed at those who work with the device over the entire life cycle and perform specific configurations.
Device Description of Device Parameters
Proservo NMS81 GP01077G
Endress+Hauser 7
About this document Proservo NMS81

1.3.5 Safety instructions (XA)

Ordering feature 010 "Approval" Meaning XA
BC ATEX II 1/2G Ex db [ia Ga] IIC T6 Ga/Gb XA01495G
FD FM C/US XP-AIS Cl.I Div.1 Gr.BCD T6
AEx db [ia Ga] IIC T6 Ga/Gb
GC EAC Ga/Gb Ex db [ia Ga] IIC T6...T1 X XA01711G
IC IEC Ex db [ia Ga] IIC T6 Ga/Gb XA01495G
1)
KC
MC INMETRO Ex d[ia] IIC T6 Ga/Gb XA01705G
NC NEPSI Ex d[ia] IIC T6 Ga/Gb XA01704G
TC TIIS Ex d[ia] IIC T4 Ga/Gb XA01600G
1) KC approval is covered with IEC Ex approval.
KC Ex d[ia] IIC T6 Ga/Gb XA01495G
XA01496G
8 Endress+Hauser
Proservo NMS81 About this document

1.4 Registered trademarks

FieldCare
Registered trademark of the Endress+Hauser Process Solutions AG, Reinach, Switzerland
MODBUS
Registered trademark of the MODBUS-IDA, Hopkinton, MA, USA
®
®
Endress+Hauser 9
Basic safety instructions Proservo NMS81

2 Basic safety instructions

2.1 Requirements for the personnel

The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements:
Trained, qualified specialists must have a relevant qualification for this specific function
and task. Are authorized by the plant owner/operator.
Are familiar with federal/national regulations.
Before starting work, read and understand the instructions in the manual and
supplementary documentation as well as the certificates (depending on the application). Follow instructions and comply with basic conditions.
The operating personnel must fulfill the following requirements:
Are instructed and authorized according to the requirements of the task by the facility's
owner-operator. Follow the instructions in this manual.

2.2 Designated use

Application and measured materials
Depending on the version ordered, the measuring device can also measure potentially explosive, flammable, poisonous and oxidizing media.
Measuring devices for use in hazardous areas, in hygienic applications or in applications where there is an increased risk due to process pressure, are labeled accordingly on the nameplate.
To ensure that the measuring device remains in proper condition for the operation time:
Only use the measuring device in full compliance with the data on the nameplate and
the general conditions listed in the Operating Instructions and supplementary documentation. Check the nameplate to verify if the device ordered can be put to its intended use in the
approval-related area (e.g. explosion protection, pressure vessel safety). Use the measuring device only for media against which the process-wetted materials
are adequately resistant. If the measuring device is not operated at atmospheric temperature, compliance with
the relevant basic conditions specified in the associated device documentation is absolutely essential. Protect the measuring device permanently against corrosion from environmental
influences. Observe the limit values in the "Technical Information".
The manufacturer is not liable for damage caused by improper or non-designated use.
Residual risk
During operation the sensor may assume a temperature near the temperature of the measured material.
Danger of burns due to heated surfaces!
For high process temperatures: Install protection against contact in order to prevent
burns.
10 Endress+Hauser
Proservo NMS81 Basic safety instructions

2.3 Workplace safety

For work on and with the device:
Wear the required personal protective equipment according to federal/national
regulations.

2.4 Operational safety

Risk of injury.
Operate the device in proper technical condition and fail-safe condition only.
The operator is responsible for interference-free operation of the device.
Conversions to the device
Unauthorized modifications to the device are not permitted and can lead to unforeseeable dangers.
If, despite this, modifications are required, consult with the manufacturer.
Repair
To ensure continued operational safety and reliability,
Carry out repairs on the device only if they are expressly permitted.
Observe federal/national regulations pertaining to repair of an electrical device.
Use original spare parts and accessories from the manufacturer only.
Hazardous area
To eliminate a danger for persons or for the facility when the device is used in the hazardous area (e.g. explosion protection, pressure vessel safety):
Based on the nameplate, check whether the ordered device is permitted for the
intended use in the hazardous area. Observe the specifications in the separate supplementary documentation that is an
integral part of these Instructions.

2.5 Product safety

This measuring device is designed in accordance with good engineering practice to meet state-of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate. It meets general safety standards and legal requirements.

2.5.1 CE mark

The measuring system meets the legal requirements of the applicable EC guidelines. These are listed in the corresponding EC Declaration of Conformity together with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
Endress+Hauser 11
Product description Proservo NMS81
2
1
3
4
5
6
7
8
2
3
4
9

3 Product description

3.1 Product design

A0028698
 1 Design of Proservo NMS81
1 Housing 2 Display and operating module (can be operated without opening the cover) 3 Process connection (Flange) 4 Displacer
 2 Configuration of NMS81
1 Front cover 2 Display 3 Modules 4 Sensor unit 5 Housing 6 Wire drum 7 Bracket 8 Housing cover 9 Displacer
A0028866
12 Endress+Hauser

Proservo NMS81 Incoming acceptance and product identification

4 Incoming acceptance and product
identification

4.1 Incoming acceptance

Upon receipt of the goods check the following:
• Are the order codes on the delivery note and the product sticker identical?
• Are the goods undamaged?
• Do the nameplate data match the ordering information on the delivery note?
• If required (see nameplate): Are the Safety Instructions (XA) enclosed?
If one of these conditions is not satisfied, contact your Endress+Hauser Sales Center.

4.2 Product identification

The following options are available for identification of the measuring device:
• Nameplate specifications
• Extended order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer ( www.endress.com/deviceviewer ): All information about the measuring device is displayed.
• Enter the serial number from the nameplates into the Endress+Hauser Operations App or scan the 2-D matrix code (QR code) on the nameplate with the Endress+Hauser Operations App: all the information for the measuring device is displayed.
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer: Enter the serial number from the nameplate (www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the 2-D matrix code (QR code) on the nameplate.
Endress+Hauser 13
Incoming acceptance and product identification Proservo NMS81
1
3
21
22
23
6
7
8
9
14
16
15
19
20
11
18
12
13
24
25
26
28
27
4
5
Ext. ord. cd.:
Order code: Ser. no.:
Tp max.:
Mat.:
Date:
FW:
Dev.Rev.:
ex works
DeviceID:
Ta
Ta:
if modification see sep. label
X =
MWP:
Tank ID: Tank ref.height:
2
Density range:
Kg/m³
17
10

4.2.1 Nameplate

A0027791
 3 Nameplate
1 Manufacturer address 2 Device name 3 Order code 4 Serial number 5 Extended order code 6 Supply voltage 7 Maximum process pressure 8 Maximum process temperature 9 Permitted ambient temperature (Ta)
10 Temperature resistance of cable 11 Thread for cable entry 12 Material in contact with process 13 Device ID 14 Firmware version 15 Device revision 16 Metrology certification numbers 17 Customized parametrization data 18 Ambient temperature range 19 CE mark / C-tick mark 20 Additional information on the device version 21 Ingress protection 22 Certificate symbol 23 Data concerning the Ex approval 24 General certificate of approval 25 Associated Safety Instructions (XA) 26 Manufacturing date 27 RoHS mark 28 QR code for the Endress+Hauser Operations App
14 Endress+Hauser
Proservo NMS81 Incoming acceptance and product identification
Proservo NMS
&'./0
1234:
-20°C + °C60
67:
  89:;;<=び?@のBC、E/0をGわないでください。
  PQRはT9をUVしないでください。
  XY34 Z[のケーブルを`aしてください。85
  &'67cdefg
hi
+,-$%(2)
"#$%
j"#$% Qk
+,-$%(3)
lm$%(4) lm$%(5)
lm$%(6) no,-$%(1)(2) no+-$%(1)(2)
lm$%(1)
lm$%(2)
lm$%(3)
,-$%(1)
+,-$%( )1
+,-$%(4)
'p qr 2s がu vし ない こと をy zし てか らT 9を Uけ てく ださ い。
&'():NMS
XA01600G
Ex d[ia] IIC T4 Ga/Gb
エンドレスハウザー)
い た で8 9の  を らな いで くだ さい 。
:
1
2
3 4 5 6 7 8
9 10 11 12 13 14 15 16
17
Endress+Hauser 15
A0032435
 4 Nameplate Proservo NMS8x for TIIS
1 Product type 2 Ex type 3 Input/Output circuit (1) 4 Input/Output circuit (2) 5 Signal circuit (1) 6 Signal circuit (2) 7 Signal circuit (3) 8 Output circuit (1) 9 Power supply 10 Input/output circuit (3) 11 Input/output circuit (4) 12 Signal circuit (4) 13 Signal circuit (5) 14 Signal circuit (6) 15 Contact output circuit (1) (2) 16 Contact input circuit (1) (2) 17 Drawing number

4.2.2 Manufacturer address

Endress+Hauser SE+Co. KG Hauptstraße 1 79689 Maulburg, Germany
Address of the manufacturing plant: See nameplate.

4.3 Storage and transport

4.3.1 Storage conditions

• Storage temperature: –50 to +80 °C (–58 to +176 °F)
• Store the device in its original packaging.
Incoming acceptance and product identification Proservo NMS81

4.3.2 Transport

NOTICE
Risk of injury
Transport the measuring device to the measuring point in its original packaging.
Take into account the mass center of the device in order to avoid unintended tilting.
Comply with the safety instructions, transport conditions for devices over 18kg
(39.6lbs) (IEC61010).
16 Endress+Hauser
Proservo NMS81 Installation

5 Installation

5.1 Requirements

5.1.1 Type of tanks

Depending on the type of tank and application, different installation procedures are recommended for NMS8x.
Type of tanks Without guide system With stilling well With guide wires
Fixed roof tank
Floating roof tank
Covered floating roof tank
Pressurized or bullet tank
Tank with agitator or heavy turbulence
• A stilling well is required in a floating roof tank and a covered floating roof tank.
Endress+Hauser 17
Installation Proservo NMS81
• Guide wires cannot be installed in a floating roof tank. When the measuring wire is exposed to free space, it may break due to an external shock.
• Installing guide wires is not allowed in pressurized tanks because the wires would prevent closing the valve for replacing the wire, wire drum, or displacer. NMS8x installation position is important for applications without the guide wire system in order to prevent the measuring wire from being broken (refer to Operating Instructions for details ).
18 Endress+Hauser
Proservo NMS81 Installation
B
2
3
4
5
6
1
4
5
6
A
1
3
3
4
5
C
6
1
Typical tank installation
 5 Typical tank installation
A Fixed roof tank B High pressure tank C Floating roof tank with stilling well 1 NMS8x 2 Ball valve 3 Measuring wire 4 Displacer 5 Stilling well 6 Prothermo NMT53x
A0026725
Endress+Hauser 19
Installation Proservo NMS81

5.1.2 Displacer selection guide

A wide variety of displacers are available to suit different application. Proper displacer selection ensures optimal performance and longevity. The following guidelines will assist you in selecting the most suitable displacer for your application.
Displacer types
The following NMS8x displacers are available.
30 mm (1.18 in) 50 mm (1.97 in) 70 mm (2.76 in) 110 mm (4.33 in)
316L/PTFE 316L/Alloy C/PTFE 316L 316L
A0026731 A0026732
A0026729
A0026730
20 Endress+Hauser
Proservo NMS81 Installation
A
Ø30(1.18)
a
B
86 (3.39)
Ø30(1.18)
3.6 (0.14)
121 (4.76)
165.5 (6.52)
64 (2.52)
153.6 (6.05)
185.6 (7.31)
9.5 (0.37)
a
Displacer dimensions
A0029579
A ⌀30 mm (1.18 in) 316L cylindrical displacer B ⌀30 mm (1.18 in) PTFE cylindrical displacer a Immersion point
Item
Weight (g) 261 250
Volume (ml) 84.3 118
Balance volume (ml) 41.7 59
⌀30 mm (1.18 in) 316L
cylindrical displacer
⌀30 mm (1.18 in) PTFE
cylindrical displacer
The weight, volume, and balance volume are individually determined by each displacer and also might vary depending on the values stated above.
Endress+Hauser 21
Installation Proservo NMS81
A
B
C
Ø50.8 (2)
a
45 (1.77)
85 (3.35)
129 (5.08)
2.5 (0.09)
Ø50.8 (2)
a
45 (1.77)
2.5 (0.09)
85 (3.35)
Ø50 (1.97)
a
35 (1.38)
82.7 (3.26)
46.7 (1.84)
11.5 (0.45)
11.5 (0.45)
129 (5.08)
A ⌀50 mm (1.97 in) 316L cylindrical displacer B ⌀50 mm (1.97 in) AlloyC cylindrical displacer C ⌀50 mm (1.97 in) PTFE cylindrical displacer a Immersion point
Item
Weight (g) 253 253 250
Volume (ml) 143 143 118
Balance volume (ml) 70.7 70.7 59
⌀50 mm (1.97 in) 316L
cylindrical displacer
The weight, volume, and balance volume are individually determined by each
⌀50 mm (1.97 in) AlloyC
cylindrical displacer
⌀50 mm (1.97 in) PTFE
displacer and also might vary depending on the values stated above.
A0029580
cylindrical displacer
22 Endress+Hauser
Proservo NMS81 Installation
A
B
46.6 (1.83)
1 (0.03)
90 (3.54)
Ø70(2.76)
a
a
67 (2.64)
26 (1.02)
2 (0.07)
10 (0.39)
Ø110 (4.33)
A0029582
A ⌀70 mm (2.76 in) 316L conical displacer B ⌀110 mm (4.33 in) 316L conical displacer a Immersion point
Item ⌀70 mm (2.76 in) 316L conical displacer ⌀110 mm (4.33 in) 316L conical displacer
Weight (g) 245 223
Volume (ml) 124 108
Balance volume (ml) 52.8 36.3
The weight, volume, and balance volume are individually determined by each displacer and also might vary depending on the values stated above.
Recommended displacer by application
Application Product level Interface level Density
Viscous liquid 50 mm (1.97 in) PTFE Not Recommended Not Recommended
Crude oil 50 mm (1.97 in) 316L
50 mm (1.97 in) PTFE
Black oil 50 mm (1.97 in) 316L 50 mm (1.97 in) 316L 50 mm (1.97 in) 316L
White oil 70 mm (2.76 in) 316L 70 mm (2.76 in) 316L 70 mm (2.76 in) 316L
Liquefied gas, LPG/LNG 70 mm (2.76 in) 316L 70 mm (2.76 in) 316L 70 mm (2.76 in) 316L
Corrosive liquid
50 mm (1.97 in) Alloy C 50 mm (1.97 in) PTFE
50 mm (1.97 in) 316L 50 mm (1.97 in) PTFE
50 mm (1.97 in) Alloy C 50 mm (1.97 in) PTFE
50 mm (1.97 in) 316L 50 mm (1.97 in) PTFE
50 mm (1.97 in) Alloy C 50 mm (1.97 in) PTFE
Endress+Hauser 23
Installation Proservo NMS81
D
1
1
d

5.1.3 Mounting without a guide system

NMS8x is mounted on a nozzle of the tank roof without a guide system. Sufficient clearance inside the nozzle is necessary to allow the displacer to move without hitting the inner walls (for details of D, →  25).
A0026734
 6 No guide system
D1Inner diameter of the tank nozzle d Diameter of the displacer
1 Displacer
24 Endress+Hauser
Proservo NMS81 Installation
L
1
L
2
L
1
L
2
A
B
L
3
L
3
D
2
D
1
d
r
D
1
d
D
2
p (L )x

5.1.4 Mounting with a stilling well

The stilling well diameter that is required to protect the measuring wire without disturbing its operation varies depending on the tank height. The stilling well could either be of constant diameter, or narrower at its upper part and wider at its lower part. The following figure shows two examples of the latter case, namely a concentric stilling well and an asymmetric stilling well.
A0029573
 7 Mounting with concentric stilling well
A Front view B Side view L1Length from the center of the calibration window to the upper part of the stilling well
L2Length from the center of the calibration window to the bottom of the stilling well L3Length from the center of the calibration window to the bottom of the flange D1Diameter of upper part of stilling well D2Diameter of stilling well d Diameter of displacer
p
Longitudinal wire position from the center of the flange (Lx) r Radial direction offset
Endress+Hauser 25
Installation Proservo NMS81
L
1
L
2
L
1
L
2
A
B
L
3
L
3
D
2
D
1
d
r
D
1
d
D
2
p (L )x
A0026733
 8 Mounting with asymmetric stilling well
A Front view B Side view L1Length from the center of the calibration window to the upper part of the stilling well
L2Length from the center of the calibration window to the bottom of the stilling well L3Length from the center of the calibration window to the bottom of the flange D1Diameter of upper part of stilling well D2Diameter of stilling well d Diameter of displacer
p
Longitudinal wire position from the center of the flange (Lx) r Radial direction offset
• L3: length from center of the calibration window to the bottom of the flange (77 mm (3.03 in) + flange thickness). For JIS 10K 150A RF, the flange thickness is 22 mm (0.87 in).
• When using an asymmetric stilling well, take into account the lateral shift of the displacer and follow the NMS8x mounting direction as shown in the figure.
• To calculate the required stilling well diameters, the formula below should be used. The following tables contain the necessary parameters in order to calculate the dimensions of the stilling well. Be sure to have appropriate dimensions of the stilling well according to each dimension in the table.
• The radial direction offset (r) is required for only the 47 m (154.20 ft) and 55 m (180.45 ft) wire drum. For all other drums, the offset is 0 mm/in.
26 Endress+Hauser
Proservo NMS81 Installation
Feature: 110 Description
(Measuring range; Wire;
Diameter)
G1
H1
Feature: 120 Description
1AA 316L; 30 mm (1.18 in) cylindrical 30 mm (1.18 in)
1AC 316L; 50 mm (1.97 in) cylindrical 50 mm (1.97 in)
1BE 316L; 70 mm (2.76 in) conical 70 mm (2.76 in)
1BJ 316L;110 mm (4.33 in) conical 110 mm (4.33 in)
2AA PTFE; 30 mm (1.18 in) cylindrical 30 mm (1.18 in)
2AC PTFE; 50 mm (1.97 in) cylindrical 50 mm (1.97 in)
3AC AlloyC276; 50 mm (1.97 in)
4AC 316L polished; 50 mm (1.97 in)
4AE 316L polished; 70 mm (2.76 in)
5AC PTFE; 50 mm (1.97 in) cylindrical,
47 m (154.20 ft); 316L;
0.15 mm (0.00591 in)
55 m (180.45 ft); 316L
0.15 mm (0.00591 in)
(Displacer material; Type)
cylindrical
cylindrical
conical
hygienic white
NMS80 NMS81 NMS83 r
6 mm (0.24 in)
6 mm (0.24 in)
NMS80 NMS81 NMS83 d
50 mm (1.97 in)
50 mm (1.97 in)
70 mm (2.76 in)
50 mm (1.97 in)
Parameter Description
d Diameter of displacer
p(Lx) Longitudinal wire position from the center of the flange
The value can be determined by using following graph.
r Radial direction offset
s Safety factor recommended: 5 mm (0.197 in)
The following graph shows the lateral shift of the displacer depending on the measured distance for the different wire drums.
Endress+Hauser 27
Installation Proservo NMS81
L , L (mm) ( )
1 2
in
p (Lx) (mm) ( )in
-1.4 -1.3 -1.2 -1.1 -1.0 -0.9 -0.8 -0.7 -0.6 -0.5 -0.4 -0.3 -0.2 -0.1 0.0 0.1 0.2 0.3
0
100
200
300
400
500
600
700
800
900
1000
1100
1200
1300
1400
1500
1600
1700
1800
1900
2000
2100
2200
0 2000 4000 6000 8000
10000 12000 14000 16000 18000 20000 22000 24000 26000 28000 30000 32000 34000 36000 38000 40000 42000 44000 46000 48000 50000 52000 54000 56000
-36 -32 -28 -24 -20 -16 -12 -8 -4 0 4 8
a
d
e
c
b
f
 9 Lateral shift of displacer according to measurement range
a 16 m (A3) (NMS80/NMS81/NMS83) b 22 m (C2) (NMS80/NMS81/NMS83) c 28 m (D1) (NMS80/NMS81) d 36 m (F1) (NMS80/NMS81) e 47 m (G1) (NMS81) f 55 m(H1) (NMS81)
Upper diameter of stilling well
The dimension of D1 has to be the largest value of the dimensions D1a, D1b , D1c, and D according to the following formula.
28 Endress+Hauser
D1 Dimension
(Example)
>68.1 mm
(2.68 in)
D
Dimension
1x
Example Parameter
68.1 mm (2.68 in)
65.6 mm (2.58 in)
D
1a
D
1b
Description Formula
D1 dimension when the displacer is at the center of the calibration window
D1 dimension when the displacer is at the upper part of the stilling well
= 2 x ( |p (0)|+ d/2 + s)
= 2 x (|p (L1)|+ d/2 + s)
A0027997
1d
Proservo NMS81 Installation
D1 Dimension
(Example)
D
Dimension
1x
Example Parameter
50.9 mm (2.00 in)
D
1c
D
1d
Description Formula
D1 dimension when the displacer is at the bottom of the stilling well
D1 dimension when the radial direction offset is considered. This calculation is used only with the 47 m (154.20 ft) wire drum (G1 in Feature110) and 55 m (180.45 ft) (H1 in feature
110)
= 2 x ( |p (L2)|+ s)
= 2 x (d/2 + r + s)
Example: L1 = 1 000 mm, L2 = 20 000 mm, d = 50 mm, s = 5.0, 28 m drum
Lower diameter of stilling well
The dimension of D2 has to be the larger value of the dimensions D1 and D2b .
See the table below.
Concentric pipe
D2 Dimension
(Example)
>100.9 mm
(3.97 in)
D2x Dimension
Example Parameter
68.1 mm (2.68 in)
100.9 mm (3.97 in)
D
1
D
2b
Description Formula
Calculated D1 value
D2 dimension when the displacer is in L2 length
= 2 x (|p (L2)| + d/2 + s)
Example: L2 = 20 000 mm, d = 50 mm, s = 5.0, 28 m drum
Asymmetric pipe
D2 Dimension
(Example)
>84.5 mm
(3.33 in)
D2x Dimension
Example Parameter
68.1 mm (2.68 in)
84.5 mm (3.33 in)
D
1
D
2b
Description Formula
Calculated D1 value
D2 dimension that the displacer can pass through (nth groove)
Example: L2 = 20 000 mm, d = 50 mm, s = 5.0, 28 m drum
= |p (L2)| + d/2 + s + D1/2
Endress+Hauser 29
Installation Proservo NMS81
Recommendations for NMS8x mounting with a stilling well
Follow the recommendations for mounting NMS8x with a stilling well.
• Keep the pipe connection welds smooth.
• When drilling holes into the pipe, keep the interior surface of the holes clear of metal chips and burrs.
• Coat or paint the interior surface of the pipe to prevent corrosion.
• Keep the pipe as vertical as possible. Check using a plumb bob.
• Install the asymmetric pipe under the valve and align the centers of the NMS8x and the valve.
• Set the center of the lower part of the asymmetric pipe in the direction of the lateral motion.
• Observe the recommendations as per API MPMS chapter 3.1B.
• Confirm grounding between NMS8x and the tank nozzle.
30 Endress+Hauser
Proservo NMS81 Installation
2
1
3
7
4
8
12
13
9
10 11
5
6
100 (3.94)
B
D
A
C

5.1.5 Mounting with guide wires

It is also possible to guide the displacer with guide wires to prevent swinging.
A0026819
 10 Guide wire; dimensions mm (in)
No. Description
A Calibration chamber
B Spring and sleeve
C Guide wire sleeve
D Crimp tool
1 NMS8x
2 Reducer plate
3 Spring, SUS304
4 Sleeve, SUS316
5 Calibration chamber for maintenance
6 Tank
7 Measuring wire
8 Guide wire, SUS316
9 Displacer
10 Anchor hook plate, SUS304
11 Welding point
12 Wire ring, SUS316L
13 Flange
Endress+Hauser 31
Installation Proservo NMS81
6
2
3
1
4
5
1. 2. 3. 4. 5.
Guide wire installation
Guide wire installation procedure
1. Install NMS8x [1] on the reducer plate.
2. Perform calibration steps (→  87) before the displacer [3] is attached to the guide wires.
 Make sure that the displacer does not touch the guide wires during calibration.
This could be done by mounting the NMS8x to the reducer plate [2] prior to fitting the guide wires [4].
Perform calibration steps so that displacer does not touch the guide wires if the guide wires are already installed to the reducer plate.
3. Secure the guide wires to the hooks of the springs [5].
4. Secure the springs to the reducer plate.
5. Put the guide wires through the displacer guide ring [6] and set the displacer.
This completes the guide wire installation procedure.
 11 Guide wire installation
1 NMS8x 2 Reducer plate 3 Displacer 4 Guide wires 5 Springs 6 Displacer guide ring
32 Endress+Hauser
A0026887
Proservo NMS81 Installation
+1°
-1°
1

5.1.6 Alignment of NMS8x

Flange
Confirm that the size of the nozzle and the flange is matched prior to mounting NMS8x on the tank. The flange size and the rating of NMS8x vary depending on the customer’s specifications.
• Check the flange size of NMS8x.
• Mount the flange on the top of the tank. The deviation of the flange from the horizontal plane should not exceed +/- 1 degree.
• When mounting NMS8x on a long nozzle, make sure that the displacer does not touch the inner wall of the nozzle.
 12 Allowable inclination of mounting flange
1 Nozzle
When NMS8x is installed without a guide system, follow the recommendations below:
• Confirm the mounting nozzle is in the sector between 45 and 90 degrees (or -45 and -90 degrees)away from the inlet pipe of the tank. This prevents heavy swinging of the displacer caused by waves or turbulence from the inlet liquid.
• Confirm the mounting nozzle is 500 mm (19.69 in) or more away from the tank wall.
• Confirm the minimum measuring level is at 500 mm (19.69 in) or more above the top of the inlet pipe by setting the low stop (for details of low stop setting, →  97). This protects the displacer from direct flow of the inlet liquid.
• If a stilling well cannot be mounted in the tank due to the shape or condition of the tank, attaching a guide system is recommended. Consult E+H services for further information.
A0026889
Endress+Hauser 33
Installation Proservo NMS81
500 (19.69)
90°-90°
0°
45°-45°
1
1
500 (19.69)
2
2
A0026890
 13 Recommended position for mounting NMS8x and minimum measuring level; dimensions mm (in)
1 Inlet pipe 2 Tank nozzle
• Before pouring liquid into the tank, confirm that liquid flowing through the inlet of the pipe will not contact the displacer directly.
• When discharging liquid out of the tank, ensure that the displacer will not get caught in the liquid current and sucked into the outlet pipe.

5.1.7 Electrostatic charge

When liquid measured by NMS8x has a conductivity of 1 uS/m or less, it is quasi­nonconductive. In this case, using a stilling well or guide wire is recommended. This releases the electrostatic charge on the liquid surface.
34 Endress+Hauser
Proservo NMS81 Installation

5.2 Mounting of the device

The NMS8x is delivered in two different packing styles depending on the mounting method of the displacer.
• For the all-in-one method, the displacer is mounted on the measuring wire of NMS8x.
• For the displacer shipped separately method, it is necessary to install the displacer on the measuring wire inside NMS8x.

5.2.1 Available installations

The following installation procedures are available for NMS8x.
• Mounting without guide system
• Mounting with stilling well
• Mounting with guide wire
Mounting
options
Type of tanks
Type of installations
Without guide system
(Free-space mounting)
• All-in one
• Displacer shipped separately
• Displacer installation through calibration window
With stilling well With guide wire
• All-in one
• Displacer shipped separately
• Displacer installation through calibration window
Displacer shipped separately
Endress+Hauser 35
Installation Proservo NMS81
316L
50
ml
70.8
ml
141.6
g
256.6
Ø
B
V
D
V
D
m
NMS8
Displacer
12345678901
Ser.no.:
712xxxxx
Spare part-no.:
HNBR
O-Ring
28m
Range
NMS8
98765432109
Ser. no.:
3
1.
2.
Spare part-no.: 712xxxxx
12345678901
712xxxxx
m 256.6g
D
V 141.6mL
D
V 70.8mL
B
!50 316L
mm
302.xxx
Cir.
NMS8
Wiredrum
98765432109
Ser.no.:
1.35g/10m
HNBR
28 m
WireWgt.
O-Ring
Range
Spare part-no.: 712xxxxx

5.2.2 Verification of displacer and wire drum

Prior to installation of NMS8x, confirm that the serial numbers of displacer and the wire drum match with those printed to the label attached on the housing.
A0028025
 14 Verification of displacer and wire drum
36 Endress+Hauser
Proservo NMS81 Installation
A

5.2.3 Tools to be required for installation

The following tools are required when installing NMS8x.
Tools Figures Notes
Box end wrench Use the following size
• 24 mm (0.94 in)
• 26 mm (1 in)
• 30 mm (1.2 in)
• 32 mm (1.3 in)
Crescent wrench Use the size of 350 mm (13.78 in)
Allen key Use the size of 3 mm (0.12 in)or 5 mm (0.17 in)
Screw driver
• Cross-head screwdriver
• Flat-blade screwdriver
Wire cutters or terminal pliers
Crimp terminal
Water pump pliers
Density calibration test weight
A:
Signal and power supply:
0.2 to 2.5 mm2 (24 to 13 AWG)
• Ground terminal in the terminal compartment: max. 2.5 mm2 (13 AWG)
• Ground terminal at the housing: max. 4 mm2 (11 AWG)
This tool is used especially for density measurement application (optional).
Endress+Hauser 37
Installation Proservo NMS81
5
6
3
4
2
1
1.
2.
3.
4.
5.
6.

5.2.4 Installation for all-in-one

In the case of a 50 mm (1.97 in) or 70 mm (2.76 in) diameter displacer, the device can be delivered by all-in-one method.
Displacer is shipped separately according to the following specifications.
• 47 m (154.2 ft) measuring range
• 55 m (180.5 ft) measuring range
• 110 mm (4.33 in) measuring range
• NPS8 in flange
• Cleaned from oil+grease option
A0027013
 15 Removing packing materials
1 Tape 2 Fixing band 3 Displacer holder 4 Wire drum stopper 5 Drum housing cover 6 Screws and bolts
Steps Procedures Notes
Hold the gauge so that it stays horizontal against the flange.
1
2
3
Cut the fixing bands [2].
Remove the displacer holder [3] and packing material of the displacer.
Mount NMS8x on the nozzle . • Make sure that the measuring wire hangs
Remove screws and M6 bolts [6] (M10 bolts for stainless steel housing) to remove the drum housing cover [5].
Loosen two screws and remove the wire drum stopper [4].
• Perform these steps before mounting NMS8x on the nozzle.
• Do not tilt NMS8x after removing the displacer holder.
vertically.
• Confirm that there are no kinks or other defects in the measuring wire.
Be sure not to lose the O-ring and the fixing bolts for the cover of the drum housing.
38 Endress+Hauser
Proservo NMS81 Installation
7.
8.
9.
Steps Procedures Notes
Remove the tape [1] from the wire drum
4
5
6
carefully.
Mount the drum housing cover. Confirm that the O-ring is in the drum housing
Turn on the power of NMS8x.
• Remove the tape by hands to avoid damaging the wire drum.
• Make sure that the measuring wire is wound so that it fits correctly in the grooves.
cover.
Sensor, reference, and drum calibration steps
are not required because they are all performed prior to delivery.
Endress+Hauser 39
Installation Proservo NMS81
1.
200 (7.87)
200 (7.87)
130 (5.12) 130 (5.12)
2.
3.
6
1
2
4
3
4.
5.
6.
7.
8.
9.
10.
1
3
2

5.2.5 Installation for displacer shipped separately method

It is necessary to remove the wire drum from NMS8x, remove the tape on the wire drum, mount the wire drum in the drum housing, and install the displacer on the measuring wire.
Use blocks or a pedestal to secure NMS8x and provide an environment where electrical power can be supplied to NMS8x.
Procedures Figures
Secure NMS8x on the blocks or pedestal.
Confirm that there is enough space under NMS8x.
Be careful not to drop NMS8x.
Dimensions mm (in)
Remove screws and M6 bolts [6] (M10 bolts for stainless steel housing).
Remove the wire drum cover [5], wire drum stopper [4], and the bracket [2].
Remove the wire drum [1] from the drum housing.
Remove the tape [3] on the wire drum.
Unwind the measuring wire approximately 250 mm (9.84 in) so that the wire ring is positioned under the flange.
Mount the wire drum on NMS8x.
Mount the bracket.
• Take special care to not hit the wire drum against the
housing due to strong magnet force.
• Handle the measuring wire with care. It may kink.
• Be sure that the wire is wound correctly in the grooves.
Hook the displacer [3] on the ring [2].
• Be sure that the wire is wound correctly in the grooves.
• If not, remove the displacer and the wire drum, and repeat step 7.
A0027015
40 Endress+Hauser
A0029115
Proservo NMS81 Installation
11.
1
2
3
4
12.
13.
14.
15.
16.
17.
18.
1
19.
20.
21.
Procedures Figures
Turn on the power of NMS8x.
Perform sensor calibration
Secure the displacer [2] to the measuring wire [1] using the securing wire [4].
Install the ground wire [3] of the displacer (for details of displacer ground wire installation →  44).
Perform reference calibration.
Turn off the power.
Mount the wire drum cover [5].
• For sensor calibration, →  90
• For reference calibration, →  92.
Mount NMS8x on the tank nozzle [1].
Confirm that the displacer does not touch the inner wall of the nozzle.
Turn on the power.
A0027016
Perform drum calibration.
For drum calibration, →  93
A0027018
Endress+Hauser 41
Installation Proservo NMS81
1.
1
2.
2
1
3
4
6
3.
4.
5.
6.
3
4
2
1
7.
8.
9.
10.
1
2
3
4
11.
12.
13.

5.2.6 Installation through the calibration window

In the case of a 50 mm (1.97 in) diameter displacer, the displacer can be installed through the calibration window.
It is only possible to install the following displacers through the calibration window: 50 mm SUS, 50 mm alloy C, 50 mm PTFE
Procedures Figures
Remove the calibration window cover [1].
A0027019
Remove M6 bolts and screws [6] (M10 bolts for stainless steel housing).
Remove the cover [5], wire drum stopper [4], and the bracket [3].
Remove the wire drum [1] from the drum housing.
Remove the tape [2] that is securing the wire.
Handle the measuring wire with care. It may kink.
Holding the wire drum [1] with one hand, unwind the measuring wire [3] approximately 500 mm (19.69 in).
Secure the wire [3] temporarily with the tape [2].
Insert the wire ring [4] into the drum housing.
Pull the wire ring out through the calibration window.
Handle the measuring wire with care.
Insert the wire drum [4] temporarily into the drum housing.
Hook the displacer [3] on the wire ring.
Secure the displacer to the measuring wire using the securing wire [2].
Install the ground wire [1] for the displacer (for details of displacer ground wire installation →  44).
• Take special care to not hit the wire drum against the
housing due to strong magnet force.
• Handle the measuring wire with care. It may kink.
A0029117
A0027020
A0027983
42 Endress+Hauser
Proservo NMS81 Installation
14.
2
1
15.
16.
17.
18.
2
3
4
5
1
19.
20.
21.
22.
23.
24.
25.
26.
Procedures Figures
Remove the wire drum from the drum housing and unwind the measuring wire approximately 500 mm (19.69 in).
Hold the wire drum [1] up and place the displacer [2] into the calibration window.
Hold the displacer at the center of the calibration window.
Hold the other hand (wire drum) up to add tension to the measuring wire in order not to drop the displacer rapidly.
A0027985
Let go of the displacer [2].
Remove the tape from the wire drum [5].
Insert the wire drum into the drum housing.
Mount the bracket [4].
Be sure that the wire is wound correctly in the grooves.
Turn on the power of NMS8x and move the displacer up using the Move displacer wizard→  88 until the wire ring can be seen in the calibration window.
• Confirm that there are no kinks or other defects in the
measuring wire.
• Confirm that the displacer does not touch the inner wall of the nozzle.
Perform sensor calibration.
For sensor calibration, →  90
Perform reference calibration.
For reference calibration, →  92.
Mount the drum housing cover [3] and the calibration window cover [1].
Perform drum calibration.
For drum calibration, →  93
A0027987
Endress+Hauser 43
Installation Proservo NMS81
1
2
3
4
2
5
6
7
1. 2. 3. 4.

5.2.7 Displacer ground wire installation

Depending on the application and Ex requirements, electrical grounding of the displacer is required. There are different procedures depending on the displacer type, which are described below.
For details of displacer installation →  35
Standard displacer installation
1. Mount the displacer [3] on the wire ring [1].
2. Wind the securing wire [4] on the wire hook.
3. Wind the ground wire [2] between the washers [5] twice.  If grounding is not required for non-explosion-proof applications, skip this step.
4. Secure the nut [6] with a wrench [7].
This completes the displacer installation procedure.
 16 Displacer installation
1 Wire ring 2 Ground wire 3 Displacer 4 Securing wire 5 Washer 6 Nut 7 Wrench
PTFE displacer installation
1. Remove the screw [1] using a flathead screwdriver.
2. Mount the displacer [4] on the PFA covered ring [2].
3. Remove the PFA cover approximately 10 mm (0.39 in) for conductivity.
4. Install the ground wire [6] onto the displacer from the wire insertion slot [3] until the ground wire contacts to the wall of the screw hole [5].
5. Tighten the screw [1].  Hold the ground wire with finger tips so that the wire does not come out from the
slot.
6. Lift the displacer using a screwdriver and confirm that the ground wire does not come out from the slot.
This completes the PTFE displacer installation.
A0028694
44 Endress+Hauser
Proservo NMS81 Installation
1. 2.
3. 4.
5. 6.
1
2
3
4
5
10 (0.39)
 17 PTFE displacer installation; dimensions mm (in)
1 Screw 2 PFA covered ring 3 Wire insertion slot 4 Displacer 5 Screw hole 6 Ground wire

5.3 Post-installation check

Is the device undamaged (visual inspection)?
m
Does the device conform to the measuring point specifications?
For example:
• Process temperature
m
• Process pressure (refer to the chapter on "Material load curves" of the "Technical Information" document)
• Ambient temperature range
• Measuring range
Are the measuring point identification and labeling correct (visual inspection)?
m
Is the device adequately protected from precipitation and direct sunlight?
m
A0028696
Endress+Hauser 45
Electrical connection Proservo NMS81
D
E
G
F
C
B
A
1
1
1
1
1      3
2
2
2      4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
POWER
i
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
G1 N G3 L
AC 85...264 V

6 Electrical connection

6.1 Terminal assignment

 18 Terminal compartment (typical example) and ground terminals
Terminal area Module
Up to four I/O modules, depending on the order code
A/B/C/D (slots for I/O modules)
• Modules with four terminals can be in any of these slots.
• Modules with eight terminals can be in slot B or C.
The exact assignment of the modules to the slots is dependent on the device version
→  49.
E HART Ex i/IS interface
• E1: H+
• E2: H-
F Remote display
• F1: VCC (connect to terminal 81 of the remote display)
• F2: Signal B (connect to terminal 84 of the remote display)
• F3: Signal A (connec t to terminal 83 of the remote display)
• F4: Gnd (connect to terminal 82 of the remote display)
Power consumption: 28.8 VA
Power supply: 85 to 264 V
G
• G1: N
1)
AC
• G2: not connected
• G3: L
Protective ground connection (M4 screw)
A0018339
A0027012
1) Maximum power varies depending on the configuration of the modules. As the value of 28.8 VA shows
12 w.
maximum apparent power, select the applicable cables accordingly. The actual consumed effective power is
46 Endress+Hauser
Proservo NMS81 Electrical connection
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
4
F
1
1 3
2
2 4
HR
CDI
WP SIM
G
1
3
2
POWER
1
3
Vcc
Gnd
A B
2
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
F
1 3 2 4
HR
WP
Vcc
Gnd
A
B
81 82 83 84
F1 F2 F3 F4

6.1.1 Power supply

A0033413
G1 N G2 not connected G3 L 4 Green LED: indicates power supply
Supply voltage
85 to 264 VAC, 50/60 Hz, 28.8 VA
1)
The supply voltage is also indicated on the nameplate.

6.1.2 Remote display and operating module DKX001

 19 Connection of the remote display and operating module DKX001 to the Tank Gauging device (NMR8x,
NMS8x or NRF8x)
1 Remote display and operating module 2 Connecting cable 3 Tank Gauging device (NMR8x, NMS8x or NRF8x)
A0037025
The remote display and operating module DKX001 is available as an accessory. For details refer to SD01763D.
• The measured value is indicated on the DKX001 and on the local display and operating module simulataneously.
• The operating menu cannot be accessed on both modules at the same time. If the
1) maximum value; actual value depending on modules installed. 28.8 VA includes the nominal power, and the cabling specification has to meet
Endress+Hauser 47
this value. On the other hand, the effective power consumption is 12 W.
operating menu is entered in one of these modules, the other module is automatically locked. This locking remains active until the menu is closed in the first module (back to measured value display).
Electrical connection Proservo NMS81
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
E
1
1
2
CDI
i
G
1
3
2
POWER
3

6.1.3 HART Ex i/IS interface

A0033414
E1 H+ E2 H­3 Orange LED: indicates data communication
This interface always operates as the main HART master for connected HART slave transmitters. The Analog I/O modules, on the other hand, can be configured as a HART master or slave →  56 →  58.
48 Endress+Hauser
Proservo NMS81 Electrical connection
Spare parts for: Proservo NMS81 Ser.-no.:
8A21AC098AF4
Spare part
Spare no./structure
Displacer
Wire Drum
Additional information:
XPF0002-AABICR+ AAEAEBEFLALC76 XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ XPF0002-AABEFEG+ 71273689 7122xxxx XPF0002-AABEFEG+ AAAAACDEFEG+ XPF0002-AAACABADJ+ AAHAHCHRIJJAJBKP
71023451
71023451
71023451
Cover
IOM-V1
IOM-A/RTD
IOM-D
SlotA
Slot B
Slot C
Slot D
IO Mod FF
Display Display asm.
Detector
Main electr.
SMS electr.
6
7
8
9 10
11
12
5
4
3
2
1
www.endress.com/deviceviewer
D
C
B
A

6.1.4 Slots for I/O modules

The terminal compartment contains four slots (A, B, C and D) for I/O modules. Depending on the device version (ordering features 040, 050 and 060) these slots contain different I/O modules. The table below shows which module is located in which slot for a specific device version.
The slot assignment for the device is also indicated on a label attached to the back cover of the display module.
A0030120
1 Label showing (among other things) the modules in the slots A to D. A Cable entry for slot A B Cable entry for slot B C Cable entry for slot C D Cable entry for slot D
Endress+Hauser 49
Electrical connection Proservo NMS81
NMx8x - XX XX XXxxxx ...
060050040
D
C
B
A
1
1
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
"Primary Output" (040) = "Modbus" (A1)
Ordering feature
040
Primary
Output
A1 X0 X0 Modbus - - -
A1 X0 A1 Modbus - - Digital
A1 X0 A2 Modbus - Digital Digital
A1 X0 A3 Modbus Digital Digital Digital
A1 X0 B1 Modbus Modbus - -
A1 X0 B2 Modbus Modbus - Digital
A1 X0 B3 Modbus Modbus Digital Digital
A1 A1 X0 Modbus Analog Ex d/XP - -
A1 A1 A1 Modbus Analog Ex d/XP - Digital
A1 A1 A2 Modbus Analog Ex d/XP Digital Digital
A1 A1 B1 Modbus Modbus Analog Ex d/XP -
A1 A1 B2 Modbus Modbus Analog Ex d/XP Digital
A1 A2 X0 Modbus Analog Ex d/XP Analog Ex d/XP -
A1 A2 A1 Modbus Analog Ex d/XP Analog Ex d/XP Digital
A1 A2 B1 Modbus Analog Ex d/XP Analog Ex d/XP Modbus
A1 B1 X0 Modbus Analog Ex i/IS - -
A1 B1 A1 Modbus Analog Ex i/IS - Digital
A1 B1 A2 Modbus Analog Ex i/IS Digital Digital
A1 B1 B1 Modbus Modbus Analog Ex i/IS -
A1 B1 B2 Modbus Modbus Analog Ex i/IS Digital
A1 B2 X0 Modbus Analog Ex i/IS Analog Ex i/IS -
A1 B2 A1 Modbus Analog Ex i/IS Analog Ex i/IS Digital
A1 B2 B1 Modbus Analog Ex i/IS Analog Ex i/IS Modbus
A1 C2 X0 Modbus Analog Ex i/IS Analog Ex d/XP -
A1 C2 A1 Modbus Analog Ex i/IS Analog Ex d/XP Digital
A1 C2 B1 Modbus Analog Ex i/IS Analog Ex d/XP Modbus
050 Secondary IO Analog
060
Secondary
IO Digital
Ex d/XP
Terminal area
A0023888
50 Endress+Hauser
Proservo NMS81 Electrical connection
NMx8x - XX XX XXxxxx ...
060050040
D
C
B
A
1
1
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
"Primary Output" (040) = "V1" (B1)
Ordering feature
040
Primary
Output
B1 X0 X0 V1 - - -
B1 X0 A1 V1 - - Digital
B1 X0 A2 V1 - Digital Digital
B1 X0 A3 V1 Digital Digital Digital
B1 X0 B1 V1 Modbus - -
B1 X0 B2 V1 Modbus - Digital
B1 X0 B3 V1 Modbus Digital Digital
B1 A1 X0 V1 Analog Ex d/XP - -
B1 A1 A1 V1 Analog Ex d/XP - Digital
B1 A1 A2 V1 Analog Ex d/XP Digital Digital
B1 A1 B1 V1 Modbus Analog Ex d/XP -
B1 A1 B2 V1 Modbus Analog Ex d/XP Digital
B1 A2 X0 V1 Analog Ex d/XP Analog Ex d/XP -
B1 A2 A1 V1 Analog Ex d/XP Analog Ex d/XP Digital
B1 A2 B1 V1 Analog Ex d/XP Analog Ex d/XP Modbus
B1 B1 X0 V1 Analog Ex i/IS - -
B1 B1 A1 V1 Analog Ex i/IS - Digital
B1 B1 A2 V1 Analog Ex i/IS Digital Digital
B1 B1 B1 V1 Modbus Analog Ex i/IS -
B1 B1 B2 V1 Modbus Analog Ex i/IS Digital
B1 B2 X0 V1 Analog Ex i/IS Analog Ex i/IS -
B1 B2 A1 V1 Analog Ex i/IS Analog Ex i/IS Digital
B1 B2 B1 V1 Analog Ex i/IS Analog Ex i/IS Modbus
B1 C2 X0 V1 Analog Ex i/IS Analog Ex d/XP -
B1 C2 A1 V1 Analog Ex i/IS Analog Ex d/XP Digital
B1 C2 B1 V1 Analog Ex i/IS Analog Ex d/XP Modbus
050
Secondary
IO Analog
060
Secondary
IO Digital
Ex d/XP
Terminal area
A0023888
Endress+Hauser 51
Electrical connection Proservo NMS81
NMx8x - XX XX XXxxxx ...
060050040
D
C
B
A
1
1
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
"Primary Output" (040) = "4-20mA HART Ex d" (E1)
Ordering feature
040
Primary
Output
E1 X0 X0 - Analog Ex d/XP - -
E1 X0 A1 - Analog Ex d/XP - Digital
E1 X0 A2 - Analog Ex d/XP Digital Digital
E1 X0 A3 Digital Analog Ex d/XP Digital Digital
E1 X0 B1 Modbus Analog Ex d/XP - -
E1 X0 B2 Modbus Analog Ex d/XP - Digital
E1 X0 B3 Modbus Analog Ex d/XP Digital Digital
E1 A1 X0 - Analog Ex d/XP Analog Ex d/XP -
E1 A1 A1 - Analog Ex d/XP Analog Ex d/XP Digital
E1 A1 A2 Digital Analog Ex d/XP Analog Ex d/XP Digital
E1 A1 B1 Modbus Analog Ex d/XP Analog Ex d/XP -
E1 A1 B2 Modbus Analog Ex d/XP Analog Ex d/XP Digital
E1 B1 X0 - Analog Ex d/XP Analog Ex i/IS -
E1 B1 A1 - Analog Ex d/XP Analog Ex i/IS Digital
E1 B1 A2 Digital Analog Ex d/XP Analog Ex i/IS Digital
E1 B1 B1 Modbus Analog Ex d/XP Analog Ex i/IS -
E1 B1 B2 Modbus Analog Ex d/XP Analog Ex i/IS Digital
050 Secondary IO Analog
060
Secondary
IO Digital
Ex d/XP
Terminal area
A0023888
52 Endress+Hauser
Proservo NMS81 Electrical connection
NMx8x - XX XX XXxxxx ...
060050040
D
C
B
A
1
1
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
"Primary Output" (040) = "4-20mA HART Ex i" (H1)
Ordering feature
040
Primary
Output
H1 X0 X0 - Analog Ex i/IS - -
H1 X0 A1 - Analog Ex i/IS - Digital
H1 X0 A2 - Analog Ex i/IS Digital Digital
H1 X0 A3 Digital Analog Ex i/IS Digital Digital
H1 X0 B1 Modbus Analog Ex i/IS - -
H1 X0 B2 Modbus Analog Ex i/IS - Digital
H1 X0 B3 Modbus Analog Ex i/IS Digital Digital
H1 A1 X0 - Analog Ex i/IS Analog Ex d/XP -
H1 A1 A1 - Analog Ex i/IS Analog Ex d/XP Digital
H1 A1 A2 Digital Analog Ex i/IS Analog Ex d/XP Digital
H1 A1 B1 Modbus Analog Ex i/IS Analog Ex d/XP -
H1 A1 B2 Modbus Analog Ex i/IS Analog Ex d/XP Digital
H1 B1 X0 - Analog Ex i/IS Analog Ex i/IS -
H1 B1 A1 - Analog Ex i/IS Analog Ex i/IS Digital
H1 B1 A2 Digital Analog Ex i/IS Analog Ex i/IS Digital
H1 B1 B1 Modbus Analog Ex i/IS Analog Ex i/IS -
H1 B1 B2 Modbus Analog Ex i/IS Analog Ex i/IS Digital
050
Secondary
IO Analog
060
Secondary
IO Digital
Ex d/XP
Terminal area
A0023888
Endress+Hauser 53
Electrical connection Proservo NMS81
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
WP
on
SIM
2
23344
112233445566778
8
A
1 2 3 4
A1-4
i
D
1
2 3 4
D1-4

6.1.5 Terminals of the "Modbus" or "V1" module

A0031200
 20 Designation of the "Modbus" or "V1" modules (examples); depending on the device version these modules
may also be in slot B or C.
Depending on the device version, the "Modbus" and/or "V1" module may be in different slots of the terminal compartment. In the operating menu the "Modbus" and "V1" interfaces are designated by the respective slot and the terminals within this slot: A1-4, B1-4, C1-4, D1-4.
Terminals of the "Modbus" module
Terminal
X1 S Cable shielding connected via a capacitor to EARTH
X2 0V Common reference
X3 B- Non-inverting signal line
X4 A+ Inverting signal line
Designation of the module in the operating menu: Modbus X1-4; (X = A, B, C or D)
1) In this column, "X" stands for one of the slots "A", "B", "C", or "D".
1)
Name Description
Terminals of the "V1" module
Terminal
X1 S Cable shielding connected via capacitor to EARTH
X2 not connected
X3 B- Protocol loop signal -
X4 A+ Protocol loop signal +
Designation of the module in the operating menu: V1 X1-4; (X = A, B, C or D)
1)
Name Description
54 Endress+Hauser
1) In this column, "X" stands for one of the slots "A", "B", "C", or "D".
Proservo NMS81 Electrical connection
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
WP
on
SIM
22334
4
112233445566778
8
i
B
1 2 3 4 5 6 7 8
B1-3 B4-8
C
1 2 3 4 5 6 7 8
C1-3 C4-8

6.1.6 Terminals of the "Analog I/O" module (Ex d /XP or Ex i/IS)

A0031168
Terminals Function Connection diagrams Designation in the operating menu
B1-3 Analog input or output
C1-3 Analog I/O C1-3 (→  213)
B4-8 Analog input RTD: →  59 Analog IP B4-8 (→  207)
C4-8 Analog IP C4-8 (→  207)
(configurable)
• Passive usage: →  56
• Active usage: →  58
Analog I/O B1-3 (→  213)
Endress+Hauser 55
Electrical connection Proservo NMS81
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
ca
b
-
+
!
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
a
+
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
b

6.1.7 Connection of the "Analog I/O" module for passive usage

• In the passive usage the supply voltage for the communication line must be supplied by an external source.
• The wiring must be in accordance with the intended operating mode of the Analog I/O module; see the drawings below.
"Operating mode" = "4..20mA output" or "HART slave +4..20mA output"
 21 Passive usage of the Analog I/O module in the output mode
a Power supply b HART signal output c Analog signal evaluation
"Operating mode" = "4..20mA input" or "HART master+4..20mA input"
A0027931
56 Endress+Hauser
 22 Passive usage of the Analog I/O module in the input mode
a Power supply b External device with 4...20mA and/or HART signal output
A0027933
Proservo NMS81 Electrical connection
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
a
+
!
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
b
"Operating mode" = "HART master"
A0027934
 23 Passive usage of the Analog I/O module in the HART master mode
a Power supply b Up to 6 external devices with HART signal output
Endress+Hauser 57
Electrical connection Proservo NMS81
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
b
a
-
+
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
-
+
a

6.1.8 Connection of the "Analog I/O" module for active usage

• In the active usage the supply voltage for the communication line is supplied by the device itself. There is no need of an external power supply.
• The wiring must be in accordance with the intended operating mode of the Analog I/O module; see the drawings below.
• Maximum current consumption of the connected HART devices: 24 mA (i.e. 4 mA per device if 6 devices are connected).
• Output voltage of the Ex-d module: 17.0 V@4 mA to 10.5 V@22 mA
• Output voltage of the Ex-ia module: 18.5 V@4 mA to 12.5 V@22 mA
"Operating mode" = "4..20mA output" or "HART slave +4..20mA output"
A0027932
 24 Active usage of the Analog I/O module in the output mode
a HART signal output b Analog signal evaluation
"Operating mode" = "4..20mA input" or "HART master+4..20mA input"
58 Endress+Hauser
 25 Active usage of the Analog I/O module in the input mode
a External device with 4...20mA and/or HART signal output
A0027935
Proservo NMS81 Electrical connection
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
22334
4
112233445566778
8
i
G
1
3
2
POWER
E
G
F
1
1
1 3
2
2 4
HR
CDI
WP
on
SIM
i
D
C
B
A
1122334
4
112233445566778
8
G
1
3
2
POWER
a
-
+
A
C
B
1 112 22
3 33
4 445 556 667 778 88
"Operating mode" = "HART master"
A0027936
 26 Active usage of the Analog I/O module in the HART master mode
a Up to 6 external devices with HART signal output
The maximum current consumption for the connected HART devices is 24 mA (i.e. 4 mA per device if 6 devices are connected).

6.1.9 Connection of a RTD

A0026371
A 4-wire RTD connection B 3-wire RTD connection C 2-wire RTD connection
Endress+Hauser 59
Electrical connection Proservo NMS81
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
WP
on
SIM
22334
4
1
122334455667788
A
1 2 3 4
A1-2 A3-4
i
C
1
2 3 4 5 6 7
C1-2 C3-4

6.1.10 Terminals of the "Digital I/O" module

A0026424
 27 Designation of the digital inputs or outputs (examples)
• Each Digital IO Module provides two digital inputs or outputs.
• In the operating menu each input or output is designated by the respective slot and two terminals within this slot. A1-2, for example, denotes terminals 1 and 2 of slot A. The same is valid for slots B, C and D if they contain a Digital IO module.
• For each of these pairs of terminals, one of the following operating modes can be selected in the operating menu: – Disable – Passive Output – Passive Input – Active Input
60 Endress+Hauser
Proservo NMS81 Electrical connection

6.2 Connecting requirements

6.2.1 Cable specification

Terminals
Terminal Wire cross section
Signal and power supply
• Spring terminals (NMx8x-xx1...)
• Screw terminals (NMx8x-xx2...)
Ground terminal in the terminal compartment max. 2.5 mm2 (13 AWG)
Ground terminal at the housing max. 4 mm2 (11 AWG)
Power supply line
Standard device cable is sufficient for the power line.
HART communication line
• Standard device cable is sufficient if only the analog signal is used.
• Shielded cable is recommended if using the HART protocol. Observe the grounding concept of the plant.
0.2 to 2.5 mm2 (24 to 13 AWG)
Modbus communication line
• Observe the cable conditions from the TIA-485-A, Telecommunications Industry Association.
• Additional conditions: Use shielded cable.
V1 communication line
• Two wire (twisted pair) screened or un-screened cable
• Resistance in one cable: ≤ 120 Ω
• Capacitance between lines: ≤ 0.3 µF
Endress+Hauser 61
Electrical connection Proservo NMS81

6.3 Ensuring the degree of protection

To guarantee the specified degree of protection, carry out the following steps after the electrical connection:
1. Check that the housing seals are clean and fitted correctly. Dry, clean or replace the seals if necessary.
2. Tighten all housing screws and screw covers.
3. Firmly tighten the cable glands.
4. To ensure that moisture does not enter the cable entry, route the cable so that it loops down before the cable entry ("water trap").
A0013960
5. Insert blind plugs appropriate for the safety rating of the device (e.g. Ex d/XP).

6.4 Post-connection check

Are cables or the device undamaged (visual inspection)?
m
Do the cables comply with the requirements?
m
Do the cables have adequate strain relief?
m
Are all cable glands installed, firmly tightened and correctly sealed?
m
Does the supply voltage match the specifications on the transmitter nameplate?
m
Is the terminal assignment correct →  46?
m
If required: Is the protective earth connected correctly ?
m
If supply voltage is present: Is the device ready for operation and do values appear on the display
m
module?
Are all housing covers installed and firmly tightened?
m
Is the securing clamp tightened correctly?
m
62 Endress+Hauser
Proservo NMS81 Operability

7 Operability

7.1 Overview of the operation options

The device is operated via an operating menu →  64. This menu can be accessed by the following interfaces:
• The display and operating module at the device or the remote display and operating module DKX001 (→  66).
• FieldCare connected through the service interface in the terminal compartment of the device (→  79).
• FieldCare connected through Tankvision Tank Scanner NXA820 (remote operation; →  80).
• FieldCare connected through Commubox FXA195 (→  157) to a HART interface of the device.
Confirm that the servo motor stops before changing parameters for safety use.
Endress+Hauser 63
Operability Proservo NMS81

7.2 Structure and function of the operating menu

Menu Submenu /
parameter
Operation Proservo
parameters
Level Shows the measured and calculated
Temperature Shows the measured and calculated
Density Shows the measured and calculated
Pressure Shows the measured and calculated
GP values Shows the general purpose values.
Setup Standard
parameters
Calibration Calibration of the measurement
Advanced setup Contains further parameters and
Diagnostics Diagnostic
parameters
Diagnostic list Contains up to 5 currently active error
Device information Contains information needed to
Simulation Used to simulate measured values or
Device check Contains all parameters needed to
1)
Expert
Contains all parameters of the device (including those which are already contained in one of the other menus). This menu is organized according to the function blocks of the device.
The parameter of the Expert menu are described in: GP01077G (NMS81)
System Contains all general device parameters
Sensor Contains all parameters needed to
Input/output Contains submenus to configure the
Communication Contains all parameters needed to
Application Contains submenus to configure
Meaning
Contains parameters to operate Proservo (e.g. Gauge command).
level values.
temperature values.
density values.
pressure values.
Standard commissioning parameters
submenus:
• to adapt the device to special measuring conditions.
• to process the measured value.
• to configure the signal output.
Indicates:
• The latest diagnostic messages and their timestamps.
• The operating time (overall time and time since last restart).
• The time according to the real-time clock.
messages.
identify the device.
output values.
check the measurement capability of the device.
which do not affect the measurement or the communication interface.
configure the measurement.
analog and discrete I/O modules and connected HART devices.
configure the digital communication interface.
• the tank gauging application
• the tank calculations
• the alarms.
64 Endress+Hauser
Proservo NMS81 Operability
Menu Submenu /
parameter
Tank values Shows measured and calculated tank
Diagnostics Contains all parameters needed to
1) On entering the "Expert" menu, an access code is always requested. If a customer specific access code has not been defined, "0000" has to be entered.
Meaning
values
detect and analyze operational errors.
Endress+Hauser 65
Operability Proservo NMS81
XX XX XX XX X
mm
1
2

7.3 Access to the operating menu via the local or remote display and operating module.

• Operating via the remote display and operating module DKX001 (→  47) or the local display and operating module at the device are equivalent.
• The measured value is indicated on the DKX001 and on the local display and operating module simulataneously.
• The operating menu cannot be accessed on both modules at the same time. If the operating menu is entered in one of these modules, the other module is automatically locked. This locking remains active until the menu is closed in the first module (back to measured value display).

7.3.1 Display and operating elements

The device has an illuminated liquid crystal display (LCD) that shows measured and calculated values as well as the device status in the standard view. Other views are used to navigate through the operating menu and to set parameter values.
The device is operated by three optical keys, namely "-", "+" and "E". They are actuated when the appropriate field on the protective glass of the front is touched with the finger ("touch control").
A0028345
 28 Display and operating elements
1 Liquid crystal display (LCD) 2 Optical keys; can be operated through the cover glass.
66 Endress+Hauser
Proservo NMS81 Operability
X X X X X X XX X
5
8
1
3
6
mm
4841.00
Levelbal.
F
4
7
#

7.3.2 Standard view (measured value display)

A0028702
 29 Typical appearance of the standard view (measured value display)
1 Display module 2 Device tag 3 Status area 4 Display area for measured values 5 Display area for measured value and status symbols 6 Gauge status indication 7 Gauge status symbol 8 Measured value status symbol
Status symbols
Symbol Meaning
"Failure"
A device error is present. The measured value is no longer valid.
A0013956
"Function check"
The device is in service mode (e.g. during a simulation).
A0013959
"Out of specification"
The device is operated:
• Outside of its technical specifications (e.g. during startup or a cleaning)
A0013958
• Outside of the configuration carried out by the user (e.g. level outside configured span)
"Maintenance required"
Maintenance is required. The measured value is still valid.
A0013957
Measured value symbols
Symbol 1 Symbol 2 Measured value
• Tank level
A0028148
A0028149
A0028528
A0028528
A0028528
Endress+Hauser 67
A0027993
A0028150
• Measured level
• Tank level %
Water level
Liquid temperature
Vapor temperature
A0027990
Air temperature
A0027991
• Tank ullage
• Tank ullage %
Observed density value
Operability Proservo NMS81
1
2
3
1
2
3
4
Symbol 1 Symbol 2 Measured value
Average profile density
A0028150
A0027991
P1 (bottom)
A0028151
A0028151
A0028151
A0027992
A0027992
A0027992
A0027992
A0028149 A0028529
A0028149
A0028150 A0028529
A0028150
A0028150
A0028145
A0027994
A0028141
A0028142
A0028146
A0028141
A0028142
A0028146
A0028147
A0027989
A0013957
A0027989
P2 (middle)
P3 (top)
GP 1 value
This is used for an external device.
GP 2 value
This is used for an external device.
GP 3 value
This is used for an external device.
GP 4 value
This is used for an external device.
Upper I/F level
Lower I/F level
Upper density
Middle density
Lower density
Bottom level
Displacer position
Gauge command and gauge status symbols
Symbol 1 Symbol 2 Meaning
Gauge command
A0028139
This shows current command.
Gauge status
: Displacer is unbalanced (Level/Interface not found yet).
: Displacer is balanced (Level/Interface measurement valid).
A0028143 A0028144
A0027995 A0028138
A0028140
: Displacer is moving up. : Displacer is moving down.
: Displacer stopped.
68 Endress+Hauser
Proservo NMS81 Operability
Measured value status symbols
Symbol Meaning
Status "Alarm"
The measurment is interrupted. The output assumes the defined alarm value. A diagnostic message is generated.
A0012102
Status "Warning"
The device continues measuring. A diagnostic message is generated.
A0012103
Calibration to regulatory standards disturbed
Is displayed in the following situations:
A0031169
• The write protection switch is OFF. →  77
• The write protection switch is ON but the level value can currently not be guaranteed because the displacer is not balanced.
Locking state symbols
Symbol Meaning
Display parameter
Marks display-only parameters which cannot be edited.
A0011978
Device locked
• In front of a parameter name: The device is locked via software and/or hardware.
• In the header of the measured value screen: The device is locked via hardware.
A0011979
Meaning of the keys in the standard view
Key Meaning
Enter key
• Pressing the key briefly opens the operating menu.
• Pressing the key for 2 s opens the context menu: – Level (visible if the keylock is inactive):
A0028326
Shows the measured levels.
Keylock on (visible if the keylock is inactive):
Activates the keylock.
Keylock off (visible if the keylock is active):
Deactivates the keylock.
Endress+Hauser 69
Operability Proservo NMS81
3
1
2
/../Setup
00215-1
10,000.00 mm
Empty
Set level Calibration

7.3.3 Navigation view

A0028347-EN
 30 Navigation view
1 Current submenu or wizard 2 Quick access code 3 Display area for navigation
Navigation symbols
Symbol Meaning
Operation
Is displayed:
• in the main menu next to the selection Operation
A0011975
• in the header, if you are in the Operation menu.
Setup
Is displayed:
• in the main menu next to the selection Setup
A0011974
• in the header, if you are in the Setup menu
Expert
Is displayed:
• in the main menu next to the selection Expert
A0011976
• in the header, if you are in the Expert menu
Diagnostics
Is displayed:
• in the main menu next to the selection Diagnostics
A0011977
• in the header, if you are in the Diagnostics menu
Submenu
A0013967
Wizard
A0013968
Parameter locked
When displayed in front of a parameter name, indicates that the parameter is locked.
A0013963
70 Endress+Hauser
Proservo NMS81 Operability
Meaning of the keys in the navigation view
Key Meaning
Minus key
Moves the selection bar upwards in a picklist.
A0028324
Plus key
Moves the selection bar downwards in a picklist.
A0028325
Enter key
• Pressing the key briefly opens the selected menu, submenu or
A0028326
A0028327
parameter.
• For parameters: Pressing the key for 2 s opens the help text for the function of the parameter (if present).
Escape key combination (press keys simultaneously)
• Pressing the keys briefly – Exits the current menu level and takes you to the next higher level. – If help text is open, closes the help text of the parameter.
• Pressing the keys for 2 s returns you to the measured value display ("standard view").
Endress+Hauser 71
Operability Proservo NMS81
2
1
0 mm
5000.0 mm
/../Set level
Level
Set level

7.3.4 Wizard view

A0028350-EN
 31 Wizard view on the display module
1 Current wizard 2 Display area for navigation
Wizard navigation symbols
Symbol Meaning
Parameters within a wizard
A0013972
Switches to the previous parameter.
A0013978
Confirms the parameter value and switches to the next parameter.
A0013976
Opens the editing view of the parameter.
A0013977
In the wizard view the meaning of the keys is indicated by the navigation symbol directly above the respective key (softkey functionality).
72 Endress+Hauser
Proservo NMS81 Operability
2
1
3 40 1 2
9
5
6 87
20
0
9
.

7.3.5 Numeric editor

A0028341
 32 Numeric editor on the display module
1 Display area of the entered value 2 Input mask
Symbol Meaning
Selection of numbers from 0 to 9.
A0013998
Inserts decimal separator at the input position.
A0016619
Inserts minus sign at the input position.
A0016620
Confirms selection.
A0013985
Moves the input position one position to the left.
A0016621
Exits the input without applying the changes.
A0013986
Clears all entered characters.
A0014040
Meaning of the keys in the numeric editor
Key Meaning
Minus key
In the input mask, moves the selection bar to the left (backwards).
A0028324
Plus key
In the input mask, moves the selection bar to the right (forwards).
A0028325
Enter key
• Pressing the key briefly adds the selected number to the current
A0028326
A0028327
decimal place or carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.
Escape key combination (press keys simultaneously)
Closes the text or numeric editor without applying changes.
Endress+Hauser 73
Operability Proservo NMS81
2
1
XYZ
ABC
_
Aa1

7.3.6 Text editor

A0028342
 33 Text editor on the display module
1 Display area of the entered text 2 Input mask
Text editor symbols
Symbol Meaning
Selection of letters from A to Z
A0013997
Toggle
• Between upper-case and lower-case letters
• For entering numbers
A0013981
• For entering special characters
Confirms selection.
A0013985
Switches to the selection of the correction tools.
A0013987
Exits the input without applying the changes.
A0013986
Clears all entered characters.
A0014040
Correction symbols under
Clears all entered characters.
A0013989
Moves the input position one position to the right.
A0013991
Moves the input position one position to the left.
A0013990
Deletes one character immediately to the left of the input position.
A0013988
74 Endress+Hauser
Proservo NMS81 Operability
Meaning of the keys in the text editor
Key Meaning
Minus key
In the input mask, moves the selection bar to the left (backwards).
A0028324
Plus key
In the input mask, moves the selection bar to the right (forwards).
A0028325
Enter key
• Pressing the key briefly
A0028326
A0028327
– Opens the selected group. – Carries out the selected action.
• Pressing the key for 2 s confirms the edited parameter value.
Escape key combination (press keys simultaneously)
Closes the text or numeric editor without applying changes.

7.3.7 Keypad lock

Automatic keypad lock
Operation via the local display is automatically locked:
• after a start-up or restart of the device.
• if the device has not been operated via the display for > 1 minute.
When attempting to access the operating menu while the keylock is enabled, the
Keylock on message appears.
Disabling the keypad lock
1. The keylock is enabled. Press  for at least 2 seconds.  A context menu appears.
2. Select Keylock off from the context menu.  The keylock is disabled.
Manual activation of the keypad lock
After commissioning of the device the keypad lock can be activated manually.
1. The device is in the measured value display. Press  for at least 2 seconds.  A context menu appears.
2. Select Keylock on from the context menu.  The keylock is enabled.
Endress+Hauser 75
Operability Proservo NMS81

7.3.8 Access code and user roles

Meaning of the access code
An access code can be defined in order to distinguish between the following user roles:
User role Definition
Maintenance • Knows the access code.
• Has write access to all parameters (except service parameters).
Operator • Doesn't know the access code.
• Has write access to only a few parameters.
• The description of parameters states which role is needed at least for read and write access to each parameter.
• The current user role is indicated by the Access status display parameter.
• If the access code is "0000", every user is in the Maintenance role. This is the default setting on delivery of the device.
Defining an access code
1. Navigate to: Setup → Advanced setup → Administration → Define access code → Define access code
2. Enter the intended access code (max. 4 digits).
3. Repeat the same code in the Confirm access code parameter.  The user is in the Operator role. The -symbol appears in front of all write-
protected parameters.
Switching to the "Maintenance" role
If the -symbol appears on the local display in front of a parameter, the parameter is write-protected because the user is in the Operator role. To switch to the Maintenance role, proceed as follows:
1. Press .  The input prompt for the access code appears.
2. Enter the access code.  The user is in the Maintenance role. The -symbol in front of the parameters
disappears; all previously write-protected parameters are now re-enabled.
Switching back to the "Operator" role automatically
The user automatically switches back to the Operator role:
• if no key is pressed for 10 minutes in the navigation and editing mode.
• 60 s after going back from the navigation and editing mode to the standard view (measured value display).
76 Endress+Hauser
Proservo NMS81 Operability
HR
WP
on
SIM
A
1 2
3 mm
2.5 Nm (1.84 lbf ft)

7.3.9 Write protection switch

The operating menu can be locked by a hardware switch in the connection compartment. In this locking state W&M related parameters are read only.
A0028693
1. Loosen the securing clamp.
2. Unscrew the housing cover.
3. Pull out the display module with a gentle rotation movement.
4. Using a flat blade screwdriver or a similar tool, set the write protection switch (WP) into the desired position. ON: operating menu is locked; OFF: operating menu is unlocked.
5. Put the display module onto the connection compartment, screw the cover closed and tighten the securing clamp.
To avoid acces to the write protection switch, the cover of the connection compartment can be secured by a lead seal.
The display module can be attached to the edge of the electronics compartment. This makes it easier to access the lock switch.
Endress+Hauser 77
A0033285
Operability Proservo NMS81
X X X X X X XX X
20.50
XX
XX
A0033452
 34 NMS81: Display module attached to the edge of the terminal compartment
Indication of the locking state
 35 Write protection symbol in the header of the display
Write protection via locking switch is indicated as follows:
Locking status (→  199) = Hardware locked
•  appears in the header of the display.
78 Endress+Hauser
A0028695
A0015870
Proservo NMS81 Operability
2
CDI
1

7.4 Access to the operating menu via the service interface and FieldCare

A0026993
 36 Operation via service interface
1 Service interface (CDI = Endress+Hauser Common Data Interface) 2 Commubox FXA291 3 Computer with "FieldCare" operating tool and "CDI Communication FXA291" COM DTM
The "Save/Restore" function
After a device configuration has been saved to a computer and restored to the device using the Save/Restore function of FieldCare, the device must be restarted by the following setting: SetupAdvanced setupAdministrationDevice reset = Restart device. This ensures correct operation of the device after the restore.
Endress+Hauser 79
Operability Proservo NMS81
NXA820
5
6
4
1 2 3
7

7.5 Access to the operating menu via Tankvision Tank Scanner NXA820 and FieldCare

7.5.1 Wiring scheme

A0025621
 37 Connection of Tank Gauging devices to FieldCare via the Tankvision Tank Scanner NXA820
1 Proservo NMS8x 2 Tankside Monitor NRF81 3 Micropilot NMR8x 4 Field protocol (e.g. Modbus, V1) 5 Tankvision Tank Scanner NXA820 6 Ethernet 7 Computer with FieldCare installed
80 Endress+Hauser
Proservo NMS81 Operability

7.5.2 Establishing the connection between FieldCare and the device

1. Make sure the HART CommDTM NXA is installed and update the DTM catalogue if required.
2. Create a new project in FieldCare.
3.
Add a new device: NXA HART Communication
4.
Open the configuration of the DTM and enter the required data (IP address of the NXA820; "Password" = "hart"; "Tank identification" only with NXA V1.05 or higher)
A0028515
A0028516
Endress+Hauser 81
Operability Proservo NMS81
5.
A0028517
Select Create network from the context menu.  The device is detected and the DTM is assigned.
6.
A0032427
 The device can be configured.
The "Save/Restore" function
After a device configuration has been saved to a computer and restored to the device using the Save/Restore function of FieldCare, the device must be restarted by the following setting: SetupAdvanced setupAdministrationDevice reset = Restart device. This ensures correct operation of the device after the restore.
82 Endress+Hauser
Proservo NMS81 System integration

8 System integration

8.1 Overview of the Device Description files (DTM)

To integrate the device via HART into FieldCare, a Device Description file (DTM) according to the following specification is required:
Manufacturer ID 0x11
Device type (NMS8x) 0x112D
HART specification 7.0
DD files For information and files see:
www.endress.com
Endress+Hauser 83
Commissioning Proservo NMS81
4
13
A
B
C
D
E
F
G
5
H
1
2
14
15
16
17
10
11
12
6
7
8
9
3

9 Commissioning

9.1 Terms related to tank measurement

 38 Terms concerning NMS8x installation (e.g. NMS81)
A Liquid level B Upper interface C Lower interface D Gas phase E Upper phase F Middle phase G Lower phase H Tank bottom 1 Gauge reference height 2 Empty 3 Datum plate 4 Tank ullage 5 Tank level 6 Tank reference height 7 High stop level 8 Displacer position 9 Standby level 10 Upper interface level 11 Lower interface level 12 Low stop level 13 Dipping reference
A0026916
84 Endress+Hauser
Proservo NMS81 Commissioning
14 Mechanical stop 15 Slow hoist zone 16 Distance 17 Reference position

9.2 Initial settings

Depending on NMS8x specification, some of the initial settings described below may not be required.

9.2.1 Setting the display language

Setting the display language via the display module
1. While in the standard view (→  67), press "E". If required, select Keylock off from the context menu and press "E" again.
 The Language parameter appears.
2. Open the Language parameter and select the display language.
Setting the display language via an operating tool (e.g. FieldCare)
1. Navigate to: Setup → Advanced setup → Display → Language
2. Select the display language.
This setting only affects the language on the display module. To set the language in the operating tool use the language setting functionality of FieldCare or DeviceCare, respectively.

9.2.2 Setting the real-time clock

Setting the real-time clock via the display module
1. Navigate to: Setup → Advanced setup → Date / time → Set date
2. Use the following parameters to set the the real-time clock to the current date and time: Year, Month, Day, Hour, Minutes.
Setting the real-time clock via an operating tool (e.g. FieldCare)
1. Navigate to: Setup → Advanced setup → Date / time
2.
Go to the Set date parameter and select the Start option.
Endress+Hauser 85
Commissioning Proservo NMS81
3.
Use the following parameters to set the date and time: Year, Month, Day, Hour, Minutes.
4.
Go to the Set date parameter and select the Confirm time option.  The real-time clock is set to the current date and time.
86 Endress+Hauser
Proservo NMS81 Commissioning

9.3 Calibration

After installing or replacing NMS8x or its parts (sensor module, detector unit, wire drum, or measuring wire), several calibration steps are required. All calibration steps may not be required, depending on whether the device is being installed, adjusted, or replaced (see table below).
Type of installation/replacement Calibration step
Sensor calibration Reference calibration Drum calibration
All-in one Not required Not required Not required
Displacer shipped separately Required Required Required
Displacer installation through calibration window
Replacement/ maintenance
Drum Required Required Required
Displacer Not required Required Required
Sensor module Not required Required Required
Detector unit Required Required Required
Required Required Required

9.3.1 Verification of displacer and wire drum

Prior to installation of NMS8x, confirm that all of the following data of the displacer and the wire drum on the nameplate match with those programmed into the device.
Parameters to be confirmed
Parameters Navigate to:
Displacer diameter Setup → Advanced setup → Sensor config → Displacer → Displacer diameter
Displacer weight Setup → Advanced setup → Sensor config → Displacer → Displacer weight
Displacer volume Setup → Advanced setup → Sensor config → Displacer → Displacer volume
Displacer balance volume Setup → Advanced setup → Sensor config → Displacer → Displacer balance
volume
Drum circumference Setup → Advanced setup → Sensor config → Wiredrum
Wire weight Expert → Sensor → Sensor config → Wiredrum → Wire weight
Endress+Hauser 87
Commissioning Proservo NMS81
/../Displacer
08010-1
256.6g
Displacer diamet
Displacer weight
Displacer volume
316L
50
ml
70.8
ml
141.6
g
256.6
Ø
B
V
D
V
D
m
NMS8
Displacer
mm
302.xxxCir.
NMS8
Wiredrum
1.35g/10m
HNBR
28 m
WireWgt.
O-Ring
Range
Spare part-no.: 712xxxxx
Ser.no.: 12345678901
Ser.no.: 12345678901
Spare part-no.: 712xxxxx
Data verification
Data verification procedure
1. Check the displacer diameter, weight, volume, and balance volume for the Displacer diameter parameter, the Displacer weight parameter, the Displacer volume
parameter, and the Displacer balance volume parameter.
2. Check the drum circumference and wire weight for the Drum circumference parameter and Wire weight parameter.
This completes the data verification procedure.
A0028697
 39 Data verification

9.3.2 Move displacer

The move displacer operation is optional and can be used to change the current position of the displacer in order to perform the calibration steps more easily.
1. Make sure that the wire drum stopper has been removed.
2. Navigate to: Setup → Calibration → Move displacer → Move distance
3. Input the relative moving distance for the Move distance parameter.
4. Select the Move down option or the Move up option
5. Select the Yes.
This completes move displacer commands procedure.
88 Endress+Hauser
Proservo NMS81 Commissioning
A0029119
 40 Move displacer
Endress+Hauser 89
Commissioning Proservo NMS81
m
S
m
O2
m
z
m
B
m
O1

9.3.3 Sensor calibration

Sensor calibration adjusts the weight measurement of the detector unit. The calibration consists of three steps as follows.
• ADC zero calibration
• ADC offset calibration
• ADC span calibration
For the ADC offset weight calibration, either 0 g or an offset weight (0 to 100 g) can be used.
Using an offset weight other than 0 g is recommended for density measurement.
A0029472
 41 Concept of sensor calibration
m Weight of displacer B Binary value of AD-Converter mSSpan weight
mo1Offset weight in case of 0 to 100 g (50 g is recommended.) mo2Offset weight in case of 0 g mzZero weight
90 Endress+Hauser
Proservo NMS81 Commissioning
1.
2.
3.
4.
Calibration procedure
Step Using displacer Using offset weight Description
• Navigate to: Setup → Calibration → Sensor calibration → Sensor calibration
• Input the offset weight for the Offset weight parameter used in step 3 (0.0 g in case of using the displacer only).
• Input the value for the Span weight parameter used in step 4 (weight of displacer indicated on nameplate).
A0028000 A0028000
• Hold up or remove the displacer.
• Select for next parameter.
Measuring zero weight option is shown on the display.
• Wait until the Zero calibration parameter shows the Finished option and calibration status shows Idle.
When the displacer is being held up, do not release
A0027999
A0027999
A0028000 A0028000
A0028001
A0028002
it until this step is completed.
• Confirm that the Offset calibration parameter shows the Place offset weight option.
• Hold up the displacer or attach the offset weight.
• Select for next parameter.
Measuring offset weight option is shown on the display.
• Wait until the Offset calibration parameter shows the Finished option and Calibration status shows Idle.
When the displacer is being held up, do not release
it until this step is completed.
• Release the displacer or mount it on the measuring ring if an offset weight was used in the previous step.
• Select for next parameter.
Measuring span weight option is shown on the display.
• Confirm that the Span calibration parameter shows the Finished option and Calibration status shows Idle.
• Select the Next option.
• Confirm that the Sensor calibration parameter shows the Calibration finished option and Calibration status shows Idle.
This completes sensor calibration procedure.
Do not swing the displacer and keep it in as stable
a position as possible.
Endress+Hauser 91
Commissioning Proservo NMS81
1
R
1.1. 2. 3.

9.3.4 Reference calibration

The reference calibration defines the zero distance position of the displacer from the mechanical stop.
1. Navigate to: Setup → Calibration → Reference calibration → Reference calibration
2. Select the Start option
3. Check the reference position (e.g. 70 mm (2.76 in)).  The reference position is preset prior to delivery.
4. Confirm that the displacer is correctly attached to the measuring wire.
5. The reference calibration starts automatically.
This completes the reference calibration.
 42 Reference calibration sequence
1 Mechanical stop R Reference position
A0029121
92 Endress+Hauser
Proservo NMS81 Commissioning
1
2
3
50

9.3.5 Drum calibration

1. Navigate to: Setup → Calibration → Drum calibration → Drum calibration
2. Ensure a distance of 500 mm (19.69 in) or more from the bottom of the displacer to the liquid level.
3. Confirm that the displacer weight is correct for the Set high weight parameter.
4. Select the Start option.  The drum calibration starts automatically.
The drum calibration records fifty points which will take approximately eleven minutes.
5. Select the No option as usual for the Make low table parameter.  To make a low table for special applications, select the Yes and use 50 g weight.
This completes drum calibration procedure.
To cancel any calibration, press  +  simultaneously. If the drum calibration is canceled while making the new table, the old table remains effective. If making a new table fails due to an obstruction, NMS8x will not accept the new table and shows an error message.
A0029122
 43 Making drum table
Endress+Hauser 93
Commissioning Proservo NMS81

9.3.6 Commissioning check

This procedure is to confirm that all calibration steps have been completed appropriately.
1. Navigate to: Diagnostics → Device check → Commissioning check → Commissioning check
2. Select the Start option.  Executing option is shown on the verify drum table.
3. Select the Next option.
4. Confirm that the Commissioning check wizard shows the Finished option.
5. Confirm that the Result drum check parameter is passed.
This completes the commissioning check procedure.
94 Endress+Hauser
Proservo NMS81 Commissioning

9.4 Configuring the measuring device

Configuration task Description
Configuring the level and interface measurement
Level calibration Setting for open tank with liquid →  98
Configuring the density measurement Setting spot density →  101

9.4.1 Configuring the level and interface measurement

Setting density →  95
Setting tank height →  96
Setting high and low stop →  97
Setting for open tank without liquid →  99
Setting for closed tank →  100
Setting process condition →  101
Setting tank profile →  104
Setting interface profile →  105
Setting manual profile →  106
The level measurement is to measure the position where the displacer is balanced (immersion point) in the liquid. When the liquid surface level changes, the displacer continuously follows the position to measure the liquid level. To define the appropriate level measurement, the following settings are required prior to operation.
The interface measurement can determine the interface between different liquids in a tank (e.g. water and oil).Up to two different interfaces can be determined within a maximum of three phases in a tank.
Setting the density of application
Density values for three liquid phases are set as follows prior to delivery.
• Upper density: 800 kg/m
• Middle density: 1 000 kg/m
• Lower density: 1 200 kg/m
3
3
3
Change the data to reflect the actual density values. For tanks with only one liquid phase, set the upper density. For tanks with two or three phases, set middle and bottom densities as well.
Number of phases Parameters to be set
1 phase Upper density
2 phases Upper/middle density
3 phases Upper/middle/lower density
When performing an interface measurement, the minimum density difference between phases should be at least 100 kg/m3.
Setting the density
1. Navigate to: Setup → Upper density , Setup → Middle density and Setup → Lower density
2. Input the value to Upper, Middle, and Lower densities accordingly.
Endress+Hauser 95
Commissioning Proservo NMS81
A
B
C
D
E
F
A0026983
 44 Tank configuration
A Liquid level B Upper interface C Lower interface D Upper phase (density) E Middle phase (density) F Lower phase (density)
Setting the tank height
To measure the tank level correctly, the tank reference height and empty (distance from reference point to datum plate) must be set in advance.
• Tank reference height: Set by the customer to represent the height of the tank. Distance between the dipping reference and the datum plate. Used for percentage calculation and as reference for the ullage level.
• Empty: Distance between the zero point of device and datum plate. Empty is automatically adjusted by the Set level parameter.
• Refer to Level calibration for details how to determine the empty parameter accurately. →  98
Setting the tank reference height and empty
1. Navigate to: Setup → Empty
2. Input the empty value.
3. Navigate to: Setup → Tank reference height
4. Input the value of tank reference height.
96 Endress+Hauser
Proservo NMS81 Commissioning
1
3
2
5
4
A0028032
 45 Tank height
1 High stop 2 Low stop 3 Datum plate 4 Tank reference height 5 Empty
Setting the high stop and low stop
The high stop and low stop determine the highest and lowest points of displacer movement. Set these data to the desired actual upper and lower limit values.
If the displacer should be able to determine a tank bottom that is below the datum plate, set the low stop to a negative value. To make sure that the displacer travels up to the reference position, set the high stop to a value greater than or equal to empty.
High stop and low stop setting procedure
1. Navigate to: Setup → High stop level
2. Input the actual value for high stop.
3. Navigate to: Setup → Low stop level
4. Input the actual value for low stop.
This completes upper and lower stop setting procedure.
Endress+Hauser 97
Commissioning Proservo NMS81
a
L
1

9.4.2 Level calibration

The following table shows the most likely options for setting the level calibration.
Open tank with liquid Open tank without liquid Closed tank
A0028865
A0029120
Setting for an open tank with liquid
Level setting procedure
1. Navigate to: Setup → Gauge command
2. Select the Level option for the Gauge command parameter.  The displacer automatically searches for the point where it balances.
3. Wait until the displacer is balanced on the liquid.
4. Perform dipping to determine the liquid level (L) in the tank.
5. Navigate to: Setup → Set level
6. Input the determined level value for the Set level parameter.
The Set level parameter adjusts the Empty parameter to reflect the new level value.
This completes setting for open tank with liquid procedure.
A0029126
A0028033
 46 Set level for opened tank
1 Displacer L Measured value
98 Endress+Hauser
Proservo NMS81 Commissioning
Example: Ea = 28m, Bv = 10.5m, Z0 = 0.5m
En = 28m - 10.5m = 17m       
Z0
2
1
En
Ea
Bv
Setting for an open tank without liquid
If there is no liquid in the tank, the following procedure can be used to set the tank bottom or datum plate to 0 mm for the tank level.
Level setting procedure
1. Navigate to: Operation → Gauge command → Gauge command
2. Select the Bottom level option to measure the tank bottom.
3. Navigate to: Operation → One-time command status
4. Wait until the Finished option is shown.
5. Navigate to: Operation → Level → Bottom level
6. Read the Bottom level parameter (Bv).
7. Navigate to: Setup → Empty
8. Read the actual empty value (Ea).
9. Calculate the new empty value using following formula.  En = Ea - Bv - Z0
10. Input the calculated value for the Empty parameter. 
• The parameter Z0 defines the distance between the desired 0mm level value and the physical tank bottom (if displacer measures the datum plate, Z0 = 0 mm (0 in)).
• Bottom level operation considers the immersion depth of the displacer in the measurement.
This completes the level setting for open tank without liquid procedure.
 47 Open tank without liquid
1 Tank bottom 2 Datum plate Ea Initial empty setting Bv Initial bottom level En New empty Z0 Distance from tank bottom to datum plate
A0029473
A0028133
Endress+Hauser 99
It is recommended to repeating the level calibration when there is liquid in the tank (→  98).
Commissioning Proservo NMS81
a
Bv
Ea
L
1
2
Setting for a closed tank
For tanks that cannot be hand-dipped, follow the procedure shown below.
Level setting procedure
1. Navigate to: Operation → Gauge command → Gauge command
2. Select the Bottom level option to measure the tank bottom.  NMS8x measures the tank bottom and returns to level if the post gauge
command is set to level (default).
3. Navigate to: Operation → One-time command status
4. Wait until the Finished option is shown.
5. Navigate to: Operation → Level → Bottom level
6. Read the bottom value (Bv).
7. Navigate to: Operation → Level → Tank level (a)
8. Calculate the level value (L) by using following formula.  L = a - Bv
9. Navigate to: Setup → Set level
10. Input the value L for the Set level parameter.
This completes the level setting procedure.
If the datum plate is not zero (e.g. Z mm), adjust the set level value (L) by subtracting Z from the value L (L= a-Bv-Z).
 48 Closed tank
1 Initial zero level position 2 Datum plate Ea Initial setting of Empty Bv Bottom level a Tank level L Set level value
A0028137
100 Endress+Hauser
Loading...