Drager Drager_Evita_4,2Dura_Ventilator_-_Error_code_manual Scales (Analog) Repair and Troubleshooting

HR.00005
Evita 4 / 2dura / XLError code list for
Repeated warm start
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Replace both cartridges (Air and O2)
MODDAT: 08.11.2002 14:55:39
HR.00006
HR.00013
There is probably a problem with the +15V voltage supply. This may be caused by stickingmixer cartridges in the pneumatic system or defects in the power supply unit.Short­term mains power failure to units without DC modules may also result in this error combination.
Communication problems
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Power failure alarm
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Power failure alarm
CONSEQ_SERVICE:
User is not informed of voltage failure
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:28
MODDAT: 08.11.2002 14:55:38
FAILURE:
The alarm generator is not activated on power failure
SOLUTION_SERVICE:
Replace CO2 Carrier PCB
HR.00016
HR.00020
Capacitance loss of Gold Cap capacitors
No items of special note found in log
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
None
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Replace mixer
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:39
No items of special note found in log
SOLUTION_SERVICE:
Not applicable
MODDAT: 12.12.2002 09:18:52
With this fault pattern, a mixer problem generates various error codes. As it cannot be clearly stated which side is affected, the complete mixer must be replaced
Replace complete mixer
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 1 of 171
02.00.001
Evita 4 / 2dura / XLError code list for
Restart, reset
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:20
02.00.002
FUNCTION_SERVICE:
Cold/warm start saving in RAM on CPU 68332 PCB is monitored.
CONSEQ_SERVICE:
Restart, reset
- Software error on CPU 68332 PCB. - CPU 68332 PCB is faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
Data packets between CPU PCB and Pneumatic Controller PCB cannot be fully transferred.
- CPU 68332 PCB is faulty. - Pneumatic Controller PCB is faulty.
1 LAST_SOFTWARE: 99,99
FAILURE:
SOLUTION_SERVICE:
Install relevant software on CPU 68332 PCB. If the software installation was not successful, replace the CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. Replace Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, download the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:20
02.00.003
02.00.004
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
After unit start it was not possible to establish synchronisation between CPU 68332 PCB and Pneumatic Controller PCB within 10 seconds.
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
Data packets between CPU PCB and Pneumatic Controller PCB cannot be fully transferred.
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. Replace Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
MODDAT: 11.04.2003 12:01:20
MODDAT: 11.04.2003 12:01:20
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 2 of 171
02.00.005
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software.
MODDAT: 11.04.2003 12:01:20
02.00.020
02.01.001
- RAM or software error on CPU PCB. - CPU 68332 PCB is faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
- RAM or software error on CPU PCB. - RAM or software error on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:20
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:20
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). After replacing the PCB, install the relevant software.
Hardware or SW error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB. If readout plausible, software error, otherwise most probably hardware error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Table for X: 1 PAW high adjuster selected 2 PAW high arrow selected 3 MV min adjuster selected 4 MV min high arrow selected 5 O2 adjuster selected 6 Last adjuster plus one 7 PAW high adjuster not selected 8 PAW high arrow not selected 9 MV min adjuster not selected 10MV min arrow not selected 11 O2 adjuster not selected 12 PAW high line 30 top 13 Pressure scale values top 30 14 PAW high line 60 top 15 Pressure scale values top 60 16 PAW high line 100 top 17 Pressure scale values top 100 18 PAW high line 30 bottom 19 Pressure scale values bottom 30 20 PAW high line 60 bottom 21 Pressure scale values bottom 60 22 PAW high line 100 bottom 23 Pressure scale values bottom 100 24 PAW high line 30 bargraph 25 Pressure scale values bargraph 30 26 PAW high line 60 bargraph 27 Pressure scale values bargraph 60 28 PAW high line 100 bargraph 29 Pressure scale values bargraph 10030 30 PAW high readout 1 31 PAW high readout 2 32 MV min readout 1 33 MV min readout 2 34 Cold start display
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 3 of 171
02.01.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:21
02.01.003
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:21
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.01.004
02.01.005
O2 adjuster, Si copy consistency
FIRST_SOFTWARE:
FUNCTION_SERVICE:
si1_Err, O2 adjuster, Si copy consistency X alarm, reset
CONSEQ_SERVICE: SOLUTION_SERVICE:
Safety-relevant alarm values are checked for plausibility. Defective CPU 68332 PCB Defective Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:14
MODDAT: 11.04.2003 12:01:21
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 4 of 171
02.01.006
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:21
02.01.007
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:21
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.01.020
02.01.021
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 1,04
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
MODDAT: 11.04.2003 12:01:21
MODDAT: 11.04.2003 12:01:21
Only up to SW 1.04 (not EvitaXL). RAM or software error.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 5 of 171
02.01.022
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB, install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB. After replacing Pneumatic Controller PCB, install the relevant software.
MODDAT: 11.04.2003 12:01:21
02.02.001
02.02.002
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB. If readout plausible, software error, otherwise most probably hardware error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Table for X see error 02.01.001.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:21
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:21
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
Hardware or SW error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB. If readout plausible, software error, otherwise most probably hardware error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Table for X see 02.01.001.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 6 of 171
02.02.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:21
02.02.020
FUNCTION_SERVICE:
The screen content (on Evita 2 dura, 7-segment displays) is read and compared with the requirement.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or software error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB. If readout plausible, software error, otherwise most probably hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura). Table for X see error 02.01.001.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB faulty.
1 LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:21
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.03.001
02.03.002
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or transmission error from Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) via CPU 8332 PCB to Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
1 LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
MODDAT: 11.04.2003 12:01:22
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 7 of 171
02.03.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:22
02.03.004
02.03.005
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
02.03.006
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
1 LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 8 of 171
02.03.007
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:22
02.03.020
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.04.001
02.04.002
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
MODDAT: 11.04.2003 12:01:22
MODDAT: 11.04.2003 12:01:22
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 9 of 171
02.04.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:22
02.04.004
FUNCTION_SERVICE:
In neo mode according to safety software wrong flow sensor being used
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In adult mode according to safety software wrong flow sensor being used.
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
REASON / SHORTREASON_CUSTOMER:
02.04.005
02.04.020
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety software: RESET - RAM error
CONSEQ_SERVICE:
An inconsistency in the values for the filtered air pressure on the CPU 68332 PCB and the Pneumatic Controller PCB has occurred.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
MODDAT: 11.04.2003 12:01:40
MODDAT: 11.04.2003 12:01:22
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 10 of 171
02.05.001
Evita 4 / 2dura / XLError code list for
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Install the relevant software on the CPU 68332 PCB and the Pneumatic Controller PCB.
MODDAT: 11.04.2003 12:01:22
02.05.002
02.05.003
Software error or impermissible "BTPS shut-off" activated at ambient pressure below 961 mbar.
Flow insp, current counter
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
The HPSV activation monitor signals error.
HPSV cartridge (AIR or O2) or one of the two HPSV Activation PCBs or the Pneumatic Controller PCB is faulty. In diagnosis mode “ Valves ” if error is current status of error can be read.
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
The HPSV cartridge or the HPSV Activation PCB or the Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace HPSV cartridge (AIR or O2). If replacement of the HPSV cartridge was not successful, replace the HPSV Activation PCB (AIR or O2). If replacement of the HPSV Activation PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
MODDAT: 11.04.2003 12:01:22
02.05.004
FUNCTION_SERVICE: FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE:
Software error or impermissible BTPS shut-off activated at ambient pressure below 961 mbar.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
Install the relevant software on the CPU 68332 PCB and on the Pneumatic Controller PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 11 of 171
02.05.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software
MODDAT: 11.04.2003 12:01:22
02.06.001
02.06.002
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Software/hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
- Probably software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) - CAN communication with CPU 68332 PCB and Pneumatic Controller PCB disturbed.
Software error or hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
Software error or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty or CPU 68332 PCB.
Install software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and download the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:22
FUNCTION_SERVICE: FAILURE:
Software error or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty or CPU 68332 PCB is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Install software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and download the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
Software error on Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura) or CAN communication with CPU 68332 PCB and Pneumatic Controller PCB disturbed.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 12 of 171
02.06.003
Evita 4 / 2dura / XLError code list for
Software error or hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Software error or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB is faulty (Evita 2 dura) or CPU 68332 PCB is faulty.
Install software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and download the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:41
02.06.004
- Software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). - Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB.
Software/hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Probably software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN communication with CPU 68332 PCB and Pneumatic Controller PCB disturbed.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
Software error or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB is faulty (Evita 2 dura) or CPU 68332 PCB is faulty.
Install software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and download the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.06.020
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
2003-04-16 Page 13 of 171
02.06.021
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:40
02.07.020
02.08.001
RAM or software error on CPU 68332 PCB, Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:40
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
Software error or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Install relevant software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB. If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and install the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
Probably software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN communication with CPU 68332 PCB and Pneumatic Controller PCB disturbed.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 14 of 171
02.08.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB is faulty.
Install relevant software on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB. If the software installation was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and install the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
02.08.020
02.08.021
Probably software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN communication with CPU 68332 PCB and Pneumatic Controller PCB disturbed.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 15 of 171
02.09.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:23
02.09.002
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
02.09.003
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 16 of 171
02.09.004
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or Pneumatic Controller PCB is faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
02.09.020
02.09.021
Software error on CPU 68332 PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 17 of 171
02.11.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:23
02.11.002
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
02.11.020
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 18 of 171
02.12.001
Evita 4 / 2dura / XLError code list for
The inspiratory pressure is too high for longer than ... second
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:18:52
FUNCTION_SERVICE:
si12_Err, Paw high too long, reset
CONSEQ_SERVICE: SOLUTION_SERVICE:
The inspiratory pressure is too high for longer than ... second and cannot be relieved by way of the emergency venting valve Y1.3. The inspiratory pressure is measured on the Pneumatic Controller PCB and on the CO2 Carrier PCB. Possible causes: Inspiratory pressure measurement on Pneumatic Controller PCB (1st channel) or CO2 Carrier PCB (2nd channel) defective. Read values with service mode. If the error is on the CO2 Carrier PCB, the cause may be contact problems. The CO2 Carrier PCB must be secured by both fixing screws; Inspiratory pressure sensor faulty; Pneumatic fault; Excessive measurement from expiratory pressure sensor (Note: If ground (GND) is missing the Paw sensors measure approx. 115 mbar).
FAILURE:
The inspiratory pressure is too high for longer than ... second and cannot be relieved by way of the emergency venting valve Y1.3. The inspiratory pressure is measured on the Pneumatic Controller PCB and on the CO2 Carrier PCB. Possible causes: Inspiratory pressure measurement on Pneumatic Controller PCB (1st channel) or CO2 Carrier PCB (2nd channel) defective. Read values with service mode. If the error is on the CO2 Carrier PCB, the cause may be contact problems. The CO2 Carrier PCB must be secured by both fixing screws; Inspiratory pressure sensor faulty; Pneumatic fault; Excessive measurement from expiratory pressure sensor (Note: If ground (GND) is missing the Paw sensors measure approx. 115 mbar).
Replace valve Y1.3; Replace inspiratory pressure sensor; Replace Pneumatic Controller PCB; Replace CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
02.12.020
02.13.020
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
MODDAT: 11.04.2003 12:01:23
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 19 of 171
02.14.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.14.002
02.14.020
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
02.15.001
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 20 of 171
02.15.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:23
02.16.001
02.16.002
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Hardware boot, test data not reset, Reset
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB is faulty.
Replace CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
02.16.020
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 21 of 171
02.17.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.17.002
02.17.020
RAM or software error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or SW error on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic Controller PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software. If replacement of the Graphics Controller PCB or Front Panel PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
FUNCTION_SERVICE: FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 22 of 171
02.17.021
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB is faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.18.001
02.18.002
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
FUNCTION_SERVICE: FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 23 of 171
02.18.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.18.020
02.19.001
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 24 of 171
02.19.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.19.003
02.21.000
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
02.21.001
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:41
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 25 of 171
02.22.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:24
02.40.001
02.40.002
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Cold/warm start save monitoring
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Cold/warm start saving in RAM on CPU 68332 PCB is monitored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Data packets between CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:41
FAILURE:
The Pneumatic Controller PCB detects Service codes 02.40... to 02.62. They correspond to 02.00... to 02.22... of the CPU 68332 PCB.
See explanation under Service codes 02.00... to 02.22... of CPU 68332 PCB.
MODDAT: 12.12.2002 09:19:02
Data packets between CPU 68332 PCB and Pneumatic Controller PCB cannot be fully transferred.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.40.003
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Synchronisation not possible
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
After unit start it was not possible to establish synchronisation between CPU 68332 PCB and Pneumatic Controller PCB within 10 seconds.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 26 of 171
02.40.004
Evita 4 / 2dura / XLError code list for
Data packets between CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Data packets between CPU 68332 PCB and Pneumatic Controller PCB cannot be fully transferred.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:02
02.40.005
02.40.020
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332
Replace CPU PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.41.001
02.41.002
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
MODDAT: 12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 27 of 171
02.41.003
Evita 4 / 2dura / XLError code list for
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:02
02.41.004
02.41.005
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.41.006
02.41.007
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
MODDAT: 12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 28 of 171
02.41.020
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:02
02.41.021
02.41.022
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
Only up to SW 1.04. RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.42.001
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
1 LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 29 of 171
02.42.002
Evita 4 / 2dura / XLError code list for
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:02
02.42.003
02.42.020
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
The screen content (on Evita 2 dura also 7-segment displays) is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:02
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.43.001
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW or transmission error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW or transmission error from Graphics Controller PCB or Front Panel PCB via CPU 68332 PCB and Pneumatic Controller PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 30 of 171
02.43.002
Evita 4 / 2dura / XLError code list for
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:03
02.43.003
02.43.004
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
02.43.005
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 31 of 171
02.43.006
Evita 4 / 2dura / XLError code list for
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:03
02.43.007
02.43.020
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front Panel PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.44.001
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 32 of 171
02.44.002
Evita 4 / 2dura / XLError code list for
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:03
02.44.003
02.44.004
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
In neoflow mode the wrong flow sensor is used.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
In adult mode the wrong flow sensor is used.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
The wrong flow sensor is used in neoflow mode according to the safety software.
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
The wrong flow sensor is used in adult mode according to the safety software. Hardware or SW error on CPU 68332 PCB or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.44.020
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Inconsistency of values for filtered air pressure
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
An inconsistency in the values for the filtered air pressure on the CPU PCB and Pneumatic Controller PCB has occurred
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 33 of 171
02.45.001
Evita 4 / 2dura / XLError code list for
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Hardware or SW error on CPU PCB 68332 or Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:03
02.45.002
02.45.003
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty HPSV Controller Faulty HPSV cartridge
HPSV activation monitor signals error.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty HPSV cartridge Faulty LP HPSV controller Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
SW error or impermissible BTPS shut-off activated at ambient pressure below 961 mbar.
Replace HPSV Controller PCB; Replace HPSV cartridge
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
HPSV activation monitor signals error.
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace HPSV cartridge
REASON / SHORTREASON_CUSTOMER:
02.45.004
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 34 of 171
02.45.020
Evita 4 / 2dura / XLError code list for
SW error Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
SW error or impermissible BTPS shut-off at ambient pressure below 961 mbar.
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace HPSV cartridge
MODDAT: 12.12.2002 09:19:03
02.46.001
02.46.002
Faulty HPSV cartridge Faulty LP HPSV controller Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
02.46.003
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 35 of 171
02.46.004
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:03
02.46.020
02.46.021
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
02.47.020
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU PCB; Replace Pneumatic Controller PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB Faulty Graphics Controller PCB Faulty Front Panel PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU 68332 PCB or Pneumatic Controller PCB, error on Graphics Controller PCB or Front Panel PCB
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 36 of 171
02.48.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:04
02.48.002
02.48.020
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
02.48.021
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
Faulty Graphics Controller PCB Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 37 of 171
02.49.001
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB or Front Panel PCB Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:04
02.49.002
02.49.003
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB Faulty Graphics Controller PCB Faulty Front Panel PCB.
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.49.004
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 38 of 171
02.49.020
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB; Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:04
02.49.021
02.51.001
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB Faulty Graphics Controller PCB Faulty Front Panel PCB.
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.51.002
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 39 of 171
02.51.020
Evita 4 / 2dura / XLError code list for
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The timing of the Pneumatic Controller PCB is monitored by the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:04
02.52.001
02.52.020
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
Inspiratory pressure measurement
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
The inspiratory pressure is too high for longer than 1 second and cannot be relieved by way of the emergency venting valve Y1.3. The inspiratory pressure is measured on the Pneumatic Controller PCB and on the CO2 Carrier PCB
Replace valve Y1.3; Replace inspiratory pressure sensor; Replace Pneumatic Controller PCB; Replace CO2 Carrier PCB
02.53.020
REASON_SERVICE:
Inspiratory pressure measurement on Pneumatic Controller PCB or CO2 Carrier PCB faulty. Inspiratory pressure sensor faulty Fault in pneumatic unit
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 40 of 171
02.54.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:04
02.54.002
02.54.020
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.55.001
02.55.020
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:04
MODDAT: 12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 41 of 171
02.56.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:05
02.56.002
02.56.020
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.57.001
02.57.002
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 42 of 171
02.57.020
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:05
02.57.021
02.58.001
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
02.58.002
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 43 of 171
02.58.003
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
MODDAT: 12.12.2002 09:19:05
02.58.020
02.59.001
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB; Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.59.002
Faulty CPU PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty CPU PCB Faulty Pneumatic Controller PCB
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 44 of 171
02.59.003
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT: 12.12.2002 09:19:05
02.61.000
02.61.001
Faulty CPU PCB Faulty Pneumatic Controller PCB
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:41
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:42
CPU 68332 PCB is faulty.
Replace CPU 68332 PCB and install relevant software.
02.62.001
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
The Pneumatic Controller PCB detects Service codes 02.40...
to 02.62.... They correspond to 02.00... to 02.22... of the CPU
68332 PCB.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. EvitaXL: - See explanation under Service codes 02.00... to 02.22... of CPU 68332 PCB.
MODDAT: 11.04.2003 12:01:42
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 45 of 171
02.71.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:24
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the loudspeaker (approx. 50 Ohms) is tapped via a series resistor (5 Ohms) on the Graphics Controller PCB/Front Panel PCB and made available to the system as digital information. Based on this signal a target/actual comparison of the loudspeaker activation is made. The loudspeaker is located in the control panel/front panel. Note: See also errors 02.71.012 and 02.71.013.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Evita 4; Evita 2 dura: Software error, if the error occurs between multiple operations of suction mode without ventilation occurring in the meantime. This error is eliminated on Evita 4 as from SW 2.21 (Japan from SW 1.09) and on Evita 2 dura in versions later than 3.00. Cable error between CPU 68332 PCB and loudspeaker in control panel. Loudspeaker is faulty. Trouble-shooting: Measure resistance of loudspeaker via connecting cable of control panel incl. cable between pins 9 and 10 of SUB-D cable connector. Test value = 55 plus/minus 8 Ohms. This value incl. shunt resistance must be 5 Ohms above the value of the loudspeaker which can be measured after opening the control panel. - With diagnosis mode at lowest volume setting, see "Repair Instructions - Service no./modes, error list" section
3.2.1, Diagnosis page “Microprocessor ”, front”, page 60. - With external DS mode via PC with lowest volume setting, see "Repair Instructions - Service no./modes, error list" section 9.4.7, Testing loudspeaker, page 143, or see " Repair Instructions - Service no./modes, error list" section 9.5.7, Testing loudspeaker, page 153. EvitaXL: Loudspeaker detection is faulty.
FAILURE:
Audible alarm absent
Evita 4/Evita 2 dura: - Measure resistance of loudspeaker via connecting cable of control panel incl. cable between pins 9 and 10 of SUB-D cable connector. Test value = 55 plus/minus 8 Ohms. This value incl. shunt resistance must be 5 Ohms above the value of the loudspeaker which can be measured after opening the control panel. - With diagnosis mode at lowest volume setting, see "Repair Instructions - Service no./modes, error list" section 3.2.1, Diagnosis page “Microprocessor ” of “;Front”, page 60. - With external DS mode via PC with lowest volume setting, see "Repair Instructions - Service no./modes, error list" section 9.4.7, Testing loudspeaker, page 143, or see " Repair Instructions - Service no./modes, error list" section
9.5.7, Testing loudspeaker, page 153. EvitaXL: Replace Graphics Controller 8 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 46 of 171
02.71.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:24
02.71.003
FUNCTION_SERVICE:
Flow metering on Pneumatic Controller PCB faulty
CONSEQ_SERVICE: SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - As of SW 1.05: Replace Pneumatic Controller PCB. To SW 1.04: Error is reset if ambient pressure on switch-on is below 961 mbar, but is not displayed on-screen. Only replace Pneumatic Controller PCB if error message is displayed on-screen in operation. To SW 1.04: Error is reset if ambient pressure on switch-on is below 961 mbar, but is not displayed on-screen. Only replace Pneumatic Controller PCB if error message is displayed on-screen in operation. EvitaXL: - A/D conversion on Pneumatic Controller PCB is faulty.
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
FAILURE:
Evita 4; Evita 2 dura: As of SW 1.05: Replace Pneumatic Controller PCB. To SW 1.04: Error is reset if ambient pressure on switch-on is below 961 mbar, but is not displayed on-screen. Only replace Pneumatic Controller PCB if error message is displayed on-screen in operation. EvitaXL: The Pneumatic Controller PCB is faulty.
Evita 4/Evita 2 dura: - Replace both supply pressure sensors. - Replace Pneumatic Controller PCB and download the relevant software. EvitaXL: - Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to the power failure. The duration of this power failure alarm is dependent on the capacitance of the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. A unit malfunction is signalled if one of the following criteria is met: a) Final charging voltage too high (above 11 V), voltage regulator faulty b) Final charging voltage too low (below 8 V), charging circuit faulty c) Reference voltage too high (above 6.3 V), discharging circuit faulty d) Reference voltage too low (below 4.5 V), no Gold Cap e) Capacitance less than 30mF, Gold Cap capacitance too low
CONSEQ_SERVICE:
Gold Cap capacitor, capacitance too low
Gold Cap capacitance is too low
FAILURE:
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - see Repair Instructions - Electronic components, section 5.5, Conversions, page 35 CO2 Carrier PCB. EvitaXL: - Replace CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 47 of 171
02.71.004
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, voltage too high
CONSEQ_SERVICE: SOLUTION_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to the power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high (above 11 V), voltage regulator defectiveb) Final charging voltage too low (below 8 V), charging circuit defectivec) Reference voltage too high (above 6.3 V), discharging circuit defectived) Reference voltage too low (below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Cap capacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to the power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high (above 11 V), voltage regulator defectiveb) Final charging voltage too low (below 8 V), charging circuit defectivec) Reference voltage too high (above 6.3 V), discharging circuit defectived) Reference voltage too low (below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Cap capacitance too low
see Repair Instructions - Electroniccomponents, section 5.5, Conversions,page 35 CO2 Carrier PCB. Replace CO2 carrier
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 48 of 171
02.71.005
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, voltage too low
CONSEQ_SERVICE: SOLUTION_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to a power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high(above 11 V), voltage regulatordefectiveb) Final charging voltage too low(below 8 V), charging circuit defectivec) Reference voltage too high(above 6.3 V), discharging circuitdefectived) Reference voltage too low(below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Capcapacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to a power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high(above 11 V), voltage regulatordefectiveb) Final charging voltage too low(below 8 V), charging circuit defectivec) Reference voltage too high(above 6.3 V), discharging circuitdefectived) Reference voltage too low(below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Capcapacitance too low
see Repair Instructions - Electroniccomponents, section 5.5, Conversions,page 35 CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 49 of 171
02.71.006
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, tester
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, tester defective
CONSEQ_SERVICE: SOLUTION_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to the power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high (above 11 V), voltage regulator defectiveb) Final charging voltage too low (below 8 V), charging circuit defectivec) Reference voltage too high (above 6.3 V), discharging circuit defectived) Reference voltage too low (below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Cap capacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier PCB supplies the piezo alarm generator with voltage and so alerts to the power failure. The duration of this power failure alarm is dependent on the capacitanceof the Gold Cap and so both the function and the capacitance are subjected to an hourly test, first performed 5 minutes after the unit is switched on. Aunit malfunction is signalled if one of the following criteria is met:a) Final charging voltage too high (above 11 V), voltage regulator defectiveb) Final charging voltage too low (below 8 V), charging circuit defectivec) Reference voltage too high (above 6.3 V), discharging circuit defectived) Reference voltage too low (below 4.5 V), no Gold Cape) Capacitance less than 30mF, Gold Cap capacitance too low
see Repair Instructions - Electronic components, section 5.5, Conversions, page 35 CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
02.71.007
Boot test faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Boot test faulty
CONSEQ_SERVICE: SOLUTION_SERVICE:
CPU 68332 PCB (more probably) or Pneumatic Controller PCB defective. Note: Displayed in addition to other error messages.
1 LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB (more probably) or Pneumatic Controller PCB defective. Note: Displayed in addition to other error messages.
Replace CPU PCB; Replace Pneumatic Controller PCB.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:37
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 50 of 171
02.71.008
Evita 4 / 2dura / XLError code list for
Piezo alarm generator, trigger aux
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 08.11.2002 14:55:18
02.71.009
FUNCTION_SERVICE:
Piezo alarm generator, auxiliary alarm trigger defective
CONSEQ_SERVICE: SOLUTION_SERVICE:
a) CO2 Carrier PCB defectiveb) CPU 68332 PCB defective
Piezo alarm generator: Current too high
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Piezo alarm generator, current too high
1 LAST_SOFTWARE: 99,99
FAILURE:
As well as signalling a power failure alarm, the piezo also functions as a replacement alarm for the loudspeaker in the event of a defect in the CPU 68332 (standstill). To ensure this, the piezo is suppressed on the CO2 Carrier PCB by the CPU 68332 PCBwith a square signal. If this cycle misses, the resultant muting of the piezo is cancelled and a continuous beep sounds. To test the piezo the muting is purposely cancelled. The current through the piezo, which now risesbecause of the cancelled muting, can be measured a a voltage drop via a series resistor by the A/D converter on the CO2 Carrier PCB.
- With diagnosis mode, see "Repair Instructions- Service no./modes, error list" section 3.3.2, Diagnosis page Microprocessor of electronics, page 66.- With external DS mode via PC, see "Repair Instructions - Service no./modes, error list" section 9.4.6, Testing horn, page 142.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:18
FAILURE:
As well as signalling a power failure alarm, the piezo also functions as a replacement alarm for the loudspeaker in the event of a defect in the CPU 68332 (standstill). To ensure this, the piezo is suppressed on the CO2 Carrier PCB by the CPU 68332 PCBwith a square signal. If this cycle misses, the resultant muting of the piezo is cancelled and a continuous beep sounds. To test the piezo the muting is purposely cancelled. The current through the piezo, which now risesbecause of the cancelled muting, can be measured a a voltage drop via a series resistor by the A/D converter on the CO2 Carrier PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
- With diagnosis mode, see "Repair Instructions- Service no./modes, error list" section 3.3.2, Diagnosis page Microprocessor of electronics, page 66.- With external DS mode via PC, see "Repair Instructions - Service no./modes, error list" section 9.4.6, Testing horn, page 142.
a) CO2 Carrier PCB defectiveb) CPU 68332 PCB defective
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 51 of 171
02.71.010
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:24
02.71.011
FUNCTION_SERVICE:
As well as signalling a power failure alarm, the piezo also functions as a replacement alarm for the loudspeaker in the event of a fault in the CPU 68332 (standstill). To ensure this, the piezo is suppressed on the CO2 Carrier PCB by the CPU 68332 PCB with a square signal. If this cycle misses, the resultant muting of the piezo is cancelled and a continuous beep sounds. To test the piezo the muting is purposely cancelled. The current through the piezo, which now rises because of the cancelled muting, can be measured as a voltage drop via a series resistor by the A/D converter on the CO2 Carrier PCB.
CONSEQ_SERVICE:
Piezo alarm generator, current too low
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
FAILURE:
Evita 4/Evita 2 dura: - CO2 Carrier PCB or CPU 68332 PCB is faulty. EvitaXL: - The piezo cable is unplugged or the piezo is faulty or the CO2 Carrier PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, Diagnosis page Microprocessor; of electronics, page 66. ­With external DS mode via PC, see "Repair Instructions ­Service no./modes, error list" section 9.4.6, Testing horn, page 142. EvitaXL: Plug in the piezo cable. If plugging in the piezo cable was not successful, replace the piezo. If replacing the piezo was not successful, replace the CO2 Carrier PCB.
MODDAT: 11.04.2003 12:01:25
FUNCTION_SERVICE:
The specified setting of the nebuliser valve is continuously checked against the returned signal from the activation electronics.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Drug nebuliser valve monitoring is faulty. EvitaXL: - The drug nebuliser valve comparator is faulty.
FAILURE:
Drug nebuliser valve is not connected or Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the drug nebuliser valve is connected is not detected.)
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 52 of 171
02.71.012
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:25
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the loudspeaker (approx. 50 Ohms) is tapped via a series resistor (5 Ohms) on the Graphics Controller PCB and made available to the system as digital information. Based on this signal a target/actual comparison of the loudspeaker activation is made. If this target/actual comparison detects an error, a unit malfunction is generated.
CONSEQ_SERVICE:
Loudspeaker is detected too infrequently.
FAILURE:
Evita 4/Evita 2 dura: a) Cable fault between CO2 Carrier PCB and loudspeaker in control panel or b) Loudspeaker faulty or c) Graphics Controller PCB faulty or d) CPU 68332 PCB faulty EvitaXL: -Graphics Controller 8 PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Trouble-shooting: Measure resistance of loudspeaker via connecting cable of control panel incl. cable between pins 9 and 10 of SUB-D cable connector. Test value = 55 plus/minus 8 Ohms. This value incl. shunt resistance must be 5 Ohms above the value of the loudspeaker which can be measured after opening the control panel. - With diagnosis mode at lowest volume setting, see "Repair Instructions - Service no./modes, error list" section 3.2.1, Diagnosis page “Microprocessor” of “Front”, page 60. - With external DS mode via PC with lowest volume setting, see "Repair Instructions ­Service no./modes, error list" section 9.4.7, Testing loudspeaker, page 143, or see " Repair Instructions - Service no./modes, error list" section 9.5.7, Testing loudspeaker, page
153. Note: See also error 02.71.001 Repair cable error between CO2 Carrier PCB and loudspeaker in control panel. Replace loudspeaker. Replace Graphics Controller PCB and install the relevant software. Replace CPU 68332 PCB and install the relevant software. EvitaXL: - Replace Graphics Controller 8 PCB and install the relevant software.
EvitaXL: - The sound detection is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 53 of 171
02.71.013
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:25
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the loudspeaker (approx. 50 Ohms) is tapped with a series resistor (5 Ohms) on the Graphics Controller PCB and made available to the system as digital information. Based on this signal a target/actual comparison of the loudspeaker activation is made. If this target/actual comparison detects an error, a unit malfunction is generated.
CONSEQ_SERVICE:
Loudspeaker detected too frequently.
FAILURE:
Evita 4/Evita 2 dura: a) Cable error between CO2 Carrier PCB and loudspeaker in control panel or b) Loudspeaker faulty or c) Graphics Controller PCB faulty or d) CPU 68332 PCB faulty EvitaXL: - The Graphics Controller 8 PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Trouble-shooting: Measure resistance of loudspeaker via connecting cable of control panel incl. cable between pins 9 and 10 of SUB-D cable connector. Test value = 55 plus/minus 8 Ohms. This value incl. shunt resistance must be 5 Ohms above the value of the loudspeaker which can be measured after opening the control panel. - With diagnosis mode at lowest volume setting, see "Repair Instructions - Service no./modes, error list" section 3.2.1, Diagnosis page “Microprocessor” of “Front”, page 60. - With external DS mode via PC with lowest volume setting, see "Repair Instructions ­Service no./modes, error list" section 9.4.7, Testing loudspeaker, page 143, or see " Repair Instructions - Service no./modes, error list" section 9.5.7, Testing loudspeaker, page
153. Note: See also error 02.71.001 Repair cable error between CO2 Carrier PCB and loudspeaker in control panel. Replace loudspeaker. Replace Graphics Controller PCB or Front Panel PCB and install the relevant software. Replace CPU 68332 PCB and install the relevant software. EvitaXL: ­Replace Graphics Controller 8 PCB and install the relevant software.
02.71.014
Evita 4/Evita 2 dura: a) Cable error between CO2 Carrier PCB and loudspeaker in control panel or b) Loudspeaker faulty or c) Graphics Controller PCB faulty or d) CPU 68332 PCB faulty EvitaXL: - The sound detection is faulty.
+ 15 V is too low
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances. A shortfall of + 14 V results in a unit malfunction signal.
CONSEQ_SERVICE:
+ 15 V is too low or not present.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:25
FAILURE:
Power supply unit is faulty or CO2 Carrier PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Check of power supply unit voltages (several possibilities): Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.2.1, diagnosis page “Microprocessor” of “Front”, page 60. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page 133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit EvitaXL: Check power supply unit voltages and replace power supply unit as necessary. If the power supply unit is OK, replace the CO2 Carrier PCB.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 54 of 171
02.71.015
Evita 4 / 2dura / XLError code list for
+10V reference voltage too low
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:25
02.71.016
FUNCTION_SERVICE:
The voltage is reduced on the CO2 Carrier PCB from the + 15 V and also measured there. A shortfall of +9 V results in a unit malfunction signal.
CONSEQ_SERVICE: SOLUTION_SERVICE:
The reference voltage +10V falls short of the tolerance.
AD conversion for O2 metering is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The pre-amplified voltage of the O2 sensor is read-in on the Pneumatic Controller PCB by two independent A/D converters. Trouble-shooting: With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section
3.4.2, diagnosis page “ Sensors” Pneumatic”, page 73. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.4, Sensors, page 125.
CONSEQ_SERVICE:
The voltages are different.
1 LAST_SOFTWARE: 99,99
FAILURE:
Power supply unit or CO2 Carrier PCB is faulty.
Measure power supply unit voltages, replace power supply unit as necessary. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:25
FAILURE:
The Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software.
02.71.017
The raw voltage and the amplified voltage of the O2 sensor are not within the tolerance.
Activation of O2/Air switch-over valve is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/Air switch-over valve is continuously checked against the returned signal of the activation electronics on the Pneumatic Controller PCB.
CONSEQ_SERVICE:
The returned activation signal for the O2/AIR switch-over valve produces incorrect values.
The switch-over valve is not connected or is faulty.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:25
FAILURE:
Trouble-shooting: With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page “ Valves” of Pneumatic”, page 70. With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the O2/AIR switch-over valve is connected is not detected.)
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 55 of 171
02.71.018
Evita 4 / 2dura / XLError code list for
Wrong drug nebuliser gas
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
02.71.019
FUNCTION_SERVICE:
During a drug nebulisation the specified setting of the O2/AIR switch-over valve is continuously checked against the returned signal from the activation electronics. Up to SW 3.21 lower tolerance range. As from SW 3.22 tolerance range extended and so error occurs less frequently.
CONSEQ_SERVICE:
Returned activation signal of O2/AIR switch-over valve producing wrong values.
Wrong drug nebuliser gas.
Cold/warm start detection is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The cold/warm start detection is effected with the power switch auxiliary contacts of the On switch in the power supply unit. The CO2 Carrier PCB reads in the power switch auxiliary contacts.
1 LAST_SOFTWARE: 99,99
FAILURE:
The Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page "Pneumatic valves". - With external service mode, see Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves. - Replace Pneumatic Controller PCB and install the relevant software. EvitaXL: ­Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:25
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page “ Microprocessor” of “ Electronic”, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.2, Power supply unit status, page 134. Possible cause of error: - Power switch auxiliary contacts in power switch are faulty or wrongly connected. - CO2 Carrier PCB faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Replace power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB. EvitaXL: Replace power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB. If replacement of the CO2 Carrier PCB was not successful, replace the CPU 68332 PCB and install the relevant software.
Evita 4/Evita 2 dura: - Power switch contacts not functioning. - CO2 Carrier PCB faulty. EvitaXL: - Cold start detection is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 56 of 171
02.71.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
MODDAT: 11.04.2003 12:01:25
02.71.021
CONSEQ_SERVICE:
The data to initialise the interrupt masks are incorrectly stored.
RAM or software error on CPU 68332 PCB or Pneumatic Controller PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Time basis is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The quartz times of the CPU 68332 PCB and the Pneumatic Controller PCB are compared. In the event of a difference an error message is delivered.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:25
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
02.72.001
02.72.002
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 57 of 171
02.72.004
02.72.005
02.72.006
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
02.72.007
02.72.008
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 58 of 171
02.72.009
02.72.010
02.72.020
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
MODDAT: 12.12.2002 09:19:05
02.72.040
02.72.041
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 59 of 171
02.72.042
02.72.050
02.72.051
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.060
02.72.061
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 60 of 171
02.72.070
02.72.071
02.72.080
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.090
02.72.100
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 61 of 171
02.72.201
02.72.202
02.72.203
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.204
02.72.205
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 62 of 171
02.72.206
02.72.207
02.72.208
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.209
02.72.211
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 63 of 171
02.72.212
02.72.213
02.72.214
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.215
02.72.216
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 64 of 171
02.72.217
02.72.218
02.72.219
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.221
02.72.222
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 65 of 171
02.72.223
02.72.224
02.72.225
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.226
02.72.227
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 66 of 171
02.72.228
02.72.229
02.72.231
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.232
02.72.233
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 67 of 171
02.72.234
02.72.235
02.72.236
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
02.72.237
02.72.238
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
MODDAT: 12.12.2002 09:19:06
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 68 of 171
02.72.239
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
MODDAT: 12.12.2002 09:19:06
03.01.001
03.01.002
Touchscreen not detected on booting
FIRST_SOFTWARE:
FUNCTION_SERVICE:
After Evita 4 power-on the touchscreen is not detected by the Graphics Controller PCB. The touchscreen communicates via the RS232 interface with the Graphics Controller PCB.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
After unit power-on the touchscreen is not detected by the Graphics Controller PCB. The touchscreen communicates via the RS232 interface with the Graphics Controller PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting, Evita 4: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.2.1, diagnosis page "Microprocessor" of "Front", page 60. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.5.5, Testing touchscreen, page
150.
Replace touchscreen. If replacement of the touchscreen was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) and install the relevant software.
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.2.1, diagnosis page “Microprocessor” of “Front”, page 60. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.5.5, Testing touchscreen, page 150.
MODDAT: 11.04.2003 12:01:25
MODDAT: 11.04.2003 12:01:25
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace touchscreen (Evita 4; EvitaXL). If replacement of the touchscreen was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
- Touchscreen (Evita 4; EvitaXL) is faulty. - Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 69 of 171
03.01.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:25
03.02.001
FUNCTION_SERVICE:
The touchscreen communicates via the RS232 interface with the Graphics Controller PCB (Evita 4; EvitaXL) and the Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:
Similar to error 03.01.001, but occurred in operation. After unit power-on the screen is not detected by the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Touchscreen (Evita 4; EvitaXL) is faulty. Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Mode switch detected
FIRST_SOFTWARE:
FUNCTION_SERVICE:
1. During operation as unit fault (software error on Graphics Controller PCB) Please notify the responsible DrägerService product support executive immediately 2. The error message can be ignored if: The unit ventilates with no volume control. Then choose "Configuration -> Ventilation->3032->Modes" to alter the AutoFlow setting. Detection: Before unit fault
03.02.001 the log reports "Autoflow on" or "Autoflow off" as appropriate
4,1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.2.1, diagnosis page “Microprocessor” of “Front”, page 60. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.5.5, Testing touchscreen, page 150.
SOLUTION_SERVICE:
Replace touchscreen (Evita 4; EvitaXL). If replacement of the touchscreen was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 12.12.2002 09:18:52
FAILURE:
1. Unit fault 2. No effect
CONSEQ_SERVICE:
Software error on Graphics Controller PCB
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
SOLUTION_SERVICE:
Please notify the responsible DrägerService product support executive immediately
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 70 of 171
03.07.001
Evita 4 / 2dura / XLError code list for
"Reset wire" error
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:26
04.01.001
FUNCTION_SERVICE:
Error reset wire
CONSEQ_SERVICE: SOLUTION_SERVICE:
There is an open circuit in the reset wire from the CPU 68332 PCB to the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Power supply unit error, 5V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
FAILURE:
Control panel or Front Panel PCB is not connected. Cable to control panel or to Front Panel PCB is faulty. CPU 68332 PCB is faulty. Graphics Controller PCB or Front Panel PCB is faulty. Connector between PCBs is faulty.
Evita 4/Evita 2 dura: 1) Check front panel cable, replace front panel cable as necessary. 2) Remove housing cover so service LEDs on CPU 68332 PCB are visible. If necessary, plug the CPU 68332 PCB into one of the upper slots. 3) Connect control panel ready for use. 4) Switch on unit. 5) LED 6 must light up briefly after power-on: If it does not = CPU 68332 PCB faulty If it does = CPU 68332 PCB probably OK, replace Graphics Controller PCB (Evita 4; EvitaXL) Front Panel PCB (Evita 2 dura) and install the relevant software. EvitaXL: Check control panel connector and replace connector as necessary. Check cable to control panel and replace cable as necessary. Replace CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller 8 PCB and install the relevant software. Check connector between PCBs and replace connector as necessary.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:40
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 5V supply voltage is above 6 V
Power supply unit (source) or CO2 Carrier PCB (monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 71 of 171
04.01.002
Evita 4 / 2dura / XLError code list for
Power supply unit error, 5V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 5V supply voltage is below 4 V.
Power supply unit is faulty (source) or CO2 Carrier PCB (source monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.003
Power supply unit error, 12V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 12V supply voltage is above 14.4 V.
Power supply unit is faulty (source) or CO2 Carrier PCB (monitor) is faulty.
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FAILURE:
Trouble-shooting: - Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 72 of 171
04.01.004
Evita 4 / 2dura / XLError code list for
Power supply unit error, 12V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 12V supply voltage is below 9.6 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.005
Power supply unit error, 15V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20 %.
CONSEQ_SERVICE:
Power supply unit error: 15V supply voltage is above 18 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 73 of 171
04.01.006
Evita 4 / 2dura / XLError code list for
Power supply unit error, 15V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 15V supply voltage is below 12 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.007
Power supply unit error, -15V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: -15V supply voltage is above -18 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 74 of 171
04.01.008
Evita 4 / 2dura / XLError code list for
Power supply unit error, -15V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: -15V supply voltage is below -12 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.009
Power supply unit error, 24V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 24V supply voltage is above 28.8 V
Evita 4/Evita 2 dura: a) Power supply unit faulty (source) b) CO2 Carrier PCB faulty (monitor) EvitaXL: - Power supply unit error; 24V supply voltage is above 28.8 V. The operating voltages accessible by way of an A/D converter on the CO2 Carrier PCB are compared cyclically against the defined tolerances of plus/minus 20%.
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor of Electronic, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 75 of 171
04.01.010
Evita 4 / 2dura / XLError code list for
Power supply unit error, 24V supply voltage
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB with an A/D converter are compared cyclically against the defined tolerances of plus/minus 20 %.
CONSEQ_SERVICE:
Power supply unit error: 24V supply voltage is below 19.2 V
Power supply unit (source) is faulty or CO2 Carrier PCB (monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages (several possibilities): - Measure voltage with multimeter, see Repair Instructions - Electronic components, section 1, Power supply unit, page 1 - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.3.2, diagnosis page Microprocessor ”of “Electronic”, page 66. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.4.1, Testing voltages, page 133. If OK = Replace CO2 Carrier PCB If not OK = Replace power supply unit Important: 001 Power supply unit +5V too high (for digital components) 002 Power supply unit +5 V too low 003 Power supply unit +12V too high (for HPSVs, fan in power supply unit) 004 Power supply unit +12V too low 005 Power supply unit +15V too high (for analog components) 006 Power supply unit +15V too low 007 Power supply unit -15V too high (for analog components) 008 Power supply unit -15V too low 009 Power supply unit +24V too high (for valves and control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are incorrect, replace the power supply unit. If replacing the power supply unit was not successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.011
05.01.001
Battery voltage is too high
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
Electronics: Battery voltage is above 30 V.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Short in pneumatics fan
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D converter with a measuring resistor. In normal operation the fan current must fluctuate by more than 50 mV. The minimums and maximums of this fan voltage are measured for 60 seconds each. If the difference between the maximum and minimum is less than 50 mV there is a fault.
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit) The measured fan voltage is greater than or equal to 2.9 V (short)
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Replace battery. EvitaXL: - Check power supply unit voltages, replace the power supply unit as necessary. - Replace CO2 Carrier PCB.
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.3, diagnosis page Microprocessor” of“Pneumatic”, page 77. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.5, Voltages, page 129. Fan connected? Fan rotating? No = Replace fan Yes = Replace Pneumatic Controller PCB
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as necessary. If replacement of the fan was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:41
MODDAT: 11.04.2003 12:01:26
PCB cooling fan in pneumatic unit is faulty or the fan is not connected.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 76 of 171
05.01.002
Evita 4 / 2dura / XLError code list for
Fan (pneumatic unit)
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:26
05.01.003
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D converter. In normal operation this fan current must fluctuate by more than 50 mV. The minimums and maximums of this fan voltage are measured for 60 seconds each. If the difference between the maximum and minimum is less than 50 mV there is a fault.
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit). The measured fan voltage is greater than or equal to 2.9 V (short).
PCB cooling fan in pneumatic unit is faulty or not connected.
Fan (pneumatic unit)
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D converter. In normal operation this fan current must fluctuate by more than 50 mV. The minimums and maximums of this fan voltage are measured for 60 seconds each. If the difference between the maximum and minimum is less than 50 mV there is a fault.
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.3, diagnosis page Microprocessor” of“Pneumatic”, page 77. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.5, Voltages, page 129. Fan connected? Fan rotating? No = Replace fan. Yes = Replace Pneumatic Controller PCB
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as necessary. If replacement of the fan was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.3, diagnosis page Microprocessor” of“Pneumatic”, page 77. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.5, Voltages, page 129. Fan connected? Fan rotating? No = Replace fan Yes = Replace Pneumatic Controller PCB
05.02.001
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit). The measured fan voltage is greater than or equal to 2.9 V (short).
PCB cooling fan in pneumatic unit is faulty or not connected, or the Pneumatic Controller PCB is faulty.
"Reset-up wire" error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as necessary. If replacement of the fan was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:26
Trouble-shooting: 1. Remove housing cover so service LEDs on CPU 68332 PCB are visible. If necessary, plug the PCB into one of the upper slots. 2) Switch on unit. 3) The following LEDs must light up briefly after power-on: LED 4 (together with LED 3) on reset-up and LED 5 on reset-down. If no = CPU 68332 PCB faulty If yes = CPU 68332 PCB probably OK, replace Pneumatic Controller PCB
Check connector between electronic and pneumatic units, replace connector as necessary. If the connector was OK, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
Connector between PCBs, CPU 68332 PCB or Pneumatic Controller PCB is faulty.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 77 of 171
05.02.002
Evita 4 / 2dura / XLError code list for
"Reset down wire" error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Trouble-shooting: 1. Remove housing cover so service LEDs on CPU 68332 PCB are visible. If necessary, plug the PCB into one of the upper slots. 2) Switch on unit. 3) The following LEDs must light up briefly after power-on: LED 4 (together with LED 3) on reset-up and LED 5 on reset-down. If no = CPU 68332 PCB faulty If yes = CPU 68332 PCB probably OK, replace Pneumatic Controller PCB
Check connector between electronic and pneumatic units, replace connector as necessary. If the connector was OK, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:26
05.02.003
Connector between PCBs, CPU 68332 PCB or Pneumatic Controller PCB is faulty.
Disable wire
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Open circuit in disable wire from CPU 68332 PCB to Pneumatic Controller PCB. Connector between PCBs, CPU 68332 PCB or Pneumatic Controller PCB is faulty.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:26
Trouble-shooting: 1) Remove housing cover so service LEDs on CPU 68332 PCB are visible. If necessary, plug the PCB into one of the upper slots. 2) Switch on unit. 3) LED 7 must flash after power-on and remain lit until the pneumatics are activated (approx. 10 s): If no = CPU 68332 PCB faulty If yes = CPU 68332 PCB probably OK, replace Pneumatic Controller PCB Note: Only possible with software version 1.00 to 1.05, 2.00 and 2.10
Check connector between electronic and pneumatic units, replace connector as necessary. If the connector was OK, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
05.04.001
Copyright Dräger Medical - For internal use only!
O2/AIR switch-over valve cannot be switched on
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is continuously checked against the returned signal from the activation electronics. If the specified setting and the returned signal do not match, a unit error message is produced.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the O2/AIR switch-over valve is connected is not detected.)
MODDAT: 11.04.2003 12:01:26
2003-04-16 Page 78 of 171
05.04.002
Evita 4 / 2dura / XLError code list for
O2/AIR switch-over valve is not switched off
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:26
05.04.003
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is continuously checked against the returned signal from the activation electronics. If the specified setting and the returned signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
O2/AIR switch-over valve is not switched off.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
O2/AIR switch-over valve has not switched
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is continuously checked against the returned signal from the activation electronics. If the specified setting and the returned signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
O2/AIR switch-over valve has not switched
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the O2/AIR switch-over valve is connected is not detected.)
MODDAT: 11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the O2/AIR switch-over valve is connected is not detected.)
05.04.004
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Setting of O2/AIR switch-over valve is incorrect
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is continuously checked against the returned signal from the activation electronics. If the specified setting and the returned signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
Setting of O2/AIR switch-over valve is incorrect
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software. (Note: Whether the O2/AIR switch-over valve is connected is not detected.)
MODDAT: 11.04.2003 12:01:26
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 79 of 171
05.04.005
Evita 4 / 2dura / XLError code list for
O2/AIR switch-over valve not switching to AIR
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:26
05.04.006
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar in the tubing system is built up only with one connected pressure supply (O2 or AIR). Then the O2/AIR switch-over valve "Y1.1" switches to the absent compression gas. This bleeds the tubing system to below 10 mbar.
CONSEQ_SERVICE:
O2/AIR switch-over valve not switching to AIR
Evita 4/Evita 2 dura: The system cannot build up the test pressure of 30 mbar in the unit check or the built-up pressure is not relieved fast enough. EvitaXL: - The O2/AIR switch­over valve is not switching to AIR.
O2/AIR switch-over valve not switching to O2
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar in the tubing system is built up only with one connected pressure supply (O2 or AIR). Then the O2/AIR switch-over valve Y1.1 switches to the absent compression gas. This bleeds the tubing system to below 10 mbar.
1 LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Eliminate leaks in the system - Replace switch-over valve Y1.1 - Replace supply pressure sensors ­Replace the airway pressure sensors - Calibrate PEEP valve. EvitaXL: - Check that the test lung is rigid enough. - Check whether an airway pressure of 30 mbar is built up. Replace valve "Y1.1" in the gas port. - Replace expiratory airway pressure sensor.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page “ Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
05.05.001
CONSEQ_SERVICE:
The system cannot build up the test pressure of 30 mbar in the unit check or the built-up pressure is not relieved fast enough.
O2/Air switch-over valve not switching to O2.
Safety valve cannot be switched on
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
Safety valve cannot be switched on.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 1,04
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Eliminate leaks in the system - Replace switch-over valve Y1.1 - Replace supply pressure sensors ­Replace the airway pressure sensors - Calibrate PEEP valve. EvitaXL: - Check that the test lung is rigid enough. - Check whether an airway pressure of 30 mbar is built up. Replace valve "Y1.1" in the gas port. - Replace expiratory airway pressure sensor.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety valve was not successful, replace the Pneumatic Controller PCB and install the relevant software.
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 80 of 171
05.05.002
Evita 4 / 2dura / XLError code list for
Safety valve cannot be switched off
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 1,04
Only up to SW 1.04. Safety valve or safety valve actuator on Pneumatic Controller PCB is faulty.
MODDAT: 11.04.2003 12:01:26
05.05.003
05.05.004
CONSEQ_SERVICE:
Safety valve cannot be switched off.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Safety valve cannot be switched
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE:
Safety valve cannot be switched.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Setting of safety valve is incorrect
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
1 LAST_SOFTWARE: 1,04
1 LAST_SOFTWARE: 1,04
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety valve was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety valve was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on Pneumatic Controller PCB is faulty.
05.05.005
CONSEQ_SERVICE:
Setting of safety valve is incorrect
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Pressure bleed via safety valve is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar is built up in the patient system. Then it is bled to below 10 mbar via safety valve "Y1.3".
CONSEQ_SERVICE:
Pressure bleed with safety valve "Y1.3" is faulty.
The system cannot build up the test pressure of 30 mbar in the unit check or the built-up pressure is not relieved fast enough.
1 LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety valve was not successful, replace the Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair Instructions - Service no./modes, error list" section 3.4.1, diagnosis page ” Valves” of “Pneumatic”, page 70. - With external service mode, see "Repair Instructions - Service no./modes, error list" section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace safety valve "Y1.3" in the gas port.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 81 of 171
05.05.006
Evita 4 / 2dura / XLError code list for
Test pressure cannot be built up
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 1,04
MODDAT: 11.04.2003 12:01:26
10.01.020
10.01.040
FUNCTION_SERVICE:
In the unit check a pressure of 30 mbar is to be built up in the tubing system with the test specimen connected.
CONSEQ_SERVICE:
Test pressure of 30 mbar cannot be built up in unit check.
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
Software error (timeout)
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Timeout error on Pneumatic Controller PCB.
Software error (timeout)
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
FAILURE:
Cause: - Tubing system leaking - Test lung not connected Note: As from SW 1.04 only a “ F “ is set for this error in the unit check.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Tubing system is leaking, eliminate leakage. - PEEP valve is not calibrated, calibrate PEEP valve. EvitaXL: - The test lung is not connected. - Check the tubing system for leakage.
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
10.02.001
FUNCTION_SERVICE: FAILURE:
The Pneumatic Controller PCB is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software.
Timeout error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
RAM error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 82 of 171
10.03.001
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:27
10.97.???
10.99.001
Flash-EPROM error on Pneumatic Controller PCB.
ROSY SHUTDOWN
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Pneumatic controller
CONSEQ_SERVICE: SOLUTION_SERVICE:
ROSY SHUTDOWN software problems of operating system. The operating system is running on the CPU 68332 PCB, the Pneumatic Controller PCB, the Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura) and the Communication PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 08.11.2002 14:55:37
FAILURE:
In case of problems please consult Lübeck
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
10.99.020
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 83 of 171
10.99.030
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:27
10.99.040
10.99.050
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.060
10.99.061
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
MODDAT: 11.04.2003 12:01:27
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 84 of 171
10.99.062
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:27
10.99.063
10.99.064
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.065
10.99.070
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
MODDAT: 11.04.2003 12:01:27
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 85 of 171
10.99.071
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:27
10.99.080
10.99.081
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.082
10.99.083
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
The Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
MODDAT: 11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 86 of 171
10.99.084
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:28
10.99.085
10.99.086
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.087
10.99.088
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
MODDAT: 11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 87 of 171
10.99.101
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:28
10.99.110
10.99.111
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
Power supply unit or Pneumatic Controller PCB is faulty. Other cause: If the mixer generates abnormal noises on power-on, an HPS valve is impermissibly placing load on the minus 15 V.
Check power supply unit supply voltages, replace power supply unit as necessary. If replacement of the power supply unit was not successful, replace the Pneumatic Controller PCB and install the relevant software.
10.99.112
Faulty A/D conversion on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Faulty A/D conversion on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
Power supply unit or Pneumatic Controller PCB is faulty. Other cause: If the mixer generates abnormal noises on power-on, an HPS valve is impermissibly placing load on the minus 15 V.
Check power supply unit voltages, replace power supply unit as necessary. If replacement of the power supply unit was not successful, replace the Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 88 of 171
10.99.801
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:28
10.99.901
10.99.911
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.912
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 89 of 171
10.99.913
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty. Most probable cause: Pneumatic Controller PCB.
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:28
10.99.914
10.99.915
Error in CAN controller on Pneumatic Controller PCB or CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most probable cause: Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Pneumatic Controller PCB is faulty.
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
FAILURE:
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
Software error on Pneumatic Controller PCB.
10.99.916
Copyright Dräger Medical - For internal use only!
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Wrong software version on Pneumatic Controller PCB. Note: The same software version must be installed on all downloadable PCBs!
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
Wrong software version on Pneumatic Controller PCB.
Install software on Pneumatic Controller PCB. Note: The same software version must be installed on all downloadable PCBs!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 90 of 171
10.99.917
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:28
10.99.918
10.99.919
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.920
10.99.921
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
The Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:28
MODDAT: 11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 91 of 171
10.99.922
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:29
10.99.923
10.99.924
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.925
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 92 of 171
10.99.926
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura) is faulty. Most probable cause: Pneumatic Controller PCB
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:29
10.99.927
10.99.928
Hardware error in CAN controller on Pneumatic Controller PCB or on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most probable cause: Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware error in CAN controller on Pneumatic Controller PCB or on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most probable cause: Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
CAN ERROR_INT2
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the Graphics Controller 8 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
Hardware error in CAN controller on Pneumatic Controller PCB or on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most probable cause: Pneumatic Controller PCB
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 93 of 171
10.99.929
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:29
10.99.930
Hardware error in CAN controller on Pneumatic Controller PCB or on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most probable cause: Pneumatic Controller PCB
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware error in CAN controller on Pneumatic Controller PCB, on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
Pneumatic Controller PCB or CPU 68332 PCB Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty. Most probable cause: Pneumatic Controller PCB.
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.935
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Hardware error in CAN controller on Pneumatic Controller PCB or on CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
1 LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:42
2003-04-16 Page 94 of 171
10.99.936
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is faulty.
Replace Pneumatic Controller PCB and install the relevant software. If replacement of the Pneumatic Controller PCB was not successful, replace the CPU 68332 PCB and install the relevant software. If replacement of the CPU 68332 PCB was not successful, replace the Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and install the relevant software.
MODDAT: 11.04.2003 12:01:42
10.99.940
10.99.941
Hardware error in CAN controller on Pneumatic Controller PCB or CPU 68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
Software error on Pneumatic Controller PCB.
10.99.942
Copyright Dräger Medical - For internal use only!
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16 Page 95 of 171
10.99.950
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:29
10.99.951
10.99.952
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.953
10.99.960
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
MODDAT: 11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 96 of 171
10.99.970
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
MODDAT: 11.04.2003 12:01:29
10.99.971
10.99.972
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
1 LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.973
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE: FAILURE:
CONSEQ_SERVICE: SOLUTION_SERVICE:
Software error on Pneumatic Controller PCB.
1 LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT: 11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 97 of 171
11.01.001
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:29
11.01.003
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
11.01.004
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 98 of 171
11.01.005
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:30
11.01.006
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
11.01.007
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 99 of 171
11.01.008
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1 LAST_SOFTWARE: 99,99
MODDAT: 11.04.2003 12:01:30
11.01.009
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
11.01.010
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software version 2.21 (Japan as from software version 1.09) and on Evita 2 dura with software versions later than 3.00. With older software versions only a mixer fault is displayed and no unit error message is stored.
CONSEQ_SERVICE: SOLUTION_SERVICE:
REASON_SERVICE: REASON / SHORTREASON_CUSTOMER:
1 LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller PCBs. If the error then occurs in the HPSV O2 valve, the error is in the HPSV Controller PCB now located in the middle slot. If the error remains in the HPS "AIR" valve, the HPS "AIR" valve is faulty.
MODDAT: 11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!
2003-04-16 Page 100 of 171