There is probably a problem with the +15V voltage supply.
This may be caused by stickingmixer cartridges in the
pneumatic system or defects in the power supply unit.Shortterm mains power failure to units without DC modules may
also result in this error combination.
Communication problems
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Power failure alarm
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Power failure alarm
CONSEQ_SERVICE:
User is not informed of voltage failure
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:08.11.2002 14:55:28
MODDAT:08.11.2002 14:55:38
FAILURE:
The alarm generator is not activated on power failure
SOLUTION_SERVICE:
Replace CO2 Carrier PCB
HR.00016
HR.00020
REASON_SERVICE:
Capacitance loss of Gold Cap capacitors
No items of special note found in log
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
None
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Replace mixer
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:08.11.2002 14:55:39
No items of special note found in log
SOLUTION_SERVICE:
Not applicable
MODDAT:12.12.2002 09:18:52
With this fault pattern, a mixer problem generates various
error codes. As it cannot be clearly stated which side is
affected, the complete mixer must be replaced
Replace complete mixer
Copyright Dräger Medical - For internal use only!
2003-04-16Page 1 of 171
02.00.001
Evita 4 / 2dura / XLError code list for
Restart, reset
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:20
02.00.002
FUNCTION_SERVICE:
Cold/warm start saving in RAM on CPU 68332 PCB is
monitored.
CONSEQ_SERVICE:
Restart, reset
REASON_SERVICE:
- Software error on CPU 68332 PCB. - CPU 68332 PCB is
faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
Data packets between CPU PCB and Pneumatic Controller
PCB cannot be fully transferred.
REASON_SERVICE:
- CPU 68332 PCB is faulty. - Pneumatic Controller PCB is
faulty.
1LAST_SOFTWARE: 99,99
FAILURE:
SOLUTION_SERVICE:
Install relevant software on CPU 68332 PCB. If the software
installation was not successful, replace the CPU 68332 PCB
and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. Replace Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, download the
relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:20
02.00.003
02.00.004
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
After unit start it was not possible to establish synchronisation
between CPU 68332 PCB and Pneumatic Controller PCB
within 10 seconds.
REASON_SERVICE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
Data packets between CPU PCB and Pneumatic Controller
PCB cannot be fully transferred.
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. Replace Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
MODDAT:11.04.2003 12:01:20
MODDAT:11.04.2003 12:01:20
REASON_SERVICE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 2 of 171
02.00.005
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software.
MODDAT:11.04.2003 12:01:20
02.00.020
02.01.001
REASON_SERVICE:
- RAM or software error on CPU PCB. - CPU 68332 PCB is
faulty.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
- RAM or software error on CPU PCB. - RAM or software error
on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:20
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:20
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura). After replacing the PCB, install the
relevant software.
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332
PCB or Pneumatic Controller PCB. If readout plausible,
software error, otherwise most probably hardware error on
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura). Table for X: 1 PAW high adjuster selected
2 PAW high arrow selected 3 MV min adjuster selected 4 MV
min high arrow selected 5 O2 adjuster selected 6 Last
adjuster plus one 7 PAW high adjuster not selected 8 PAW
high arrow not selected 9 MV min adjuster not selected 10MV
min arrow not selected 11 O2 adjuster not selected 12 PAW
high line 30 top 13 Pressure scale values top 30 14 PAW high
line 60 top 15 Pressure scale values top 60 16 PAW high line
100 top 17 Pressure scale values top 100 18 PAW high line
30 bottom 19 Pressure scale values bottom 30 20 PAW high
line 60 bottom 21 Pressure scale values bottom 60 22 PAW
high line 100 bottom 23 Pressure scale values bottom 100 24
PAW high line 30 bargraph 25 Pressure scale values
bargraph 30 26 PAW high line 60 bargraph 27 Pressure scale
values bargraph 60 28 PAW high line 100 bargraph 29
Pressure scale values bargraph 10030 30 PAW high readout
1 31 PAW high readout 2 32 MV min readout 1 33 MV min
readout 2 34 Cold start display
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 3 of 171
02.01.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:21
02.01.003
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or SW error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:21
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
02.01.004
02.01.005
O2 adjuster, Si copy consistency
FIRST_SOFTWARE:
FUNCTION_SERVICE:
si1_Err, O2 adjuster, Si copy consistency X alarm, reset
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Safety-relevant alarm values are checked for plausibility.
Defective CPU 68332 PCB Defective Pneumatic Controller
PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:08.11.2002 14:55:14
MODDAT:11.04.2003 12:01:21
Copyright Dräger Medical - For internal use only!
2003-04-16Page 4 of 171
02.01.006
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:21
02.01.007
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety-relevant alarm values are checked for plausibility.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:21
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
02.01.020
02.01.021
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 1,04
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
MODDAT:11.04.2003 12:01:21
MODDAT:11.04.2003 12:01:21
REASON_SERVICE:
Only up to SW 1.04 (not EvitaXL). RAM or software error.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 5 of 171
02.01.022
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB. After replacing CPU 68332 PCB,
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB. After replacing Pneumatic Controller PCB, install the
relevant software.
MODDAT:11.04.2003 12:01:21
02.02.001
02.02.002
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332
PCB or Pneumatic Controller PCB. If readout plausible,
software error, otherwise most probably hardware error on
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura). Table for X see error 02.01.001.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:21
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:21
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332
PCB or Pneumatic Controller PCB. If readout plausible,
software error, otherwise most probably hardware error on
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura). Table for X see 02.01.001.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 6 of 171
02.02.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:21
02.02.020
FUNCTION_SERVICE:
The screen content (on Evita 2 dura, 7-segment displays) is
read and compared with the requirement.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or software error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332
PCB or Pneumatic Controller PCB. If readout plausible,
software error, otherwise most probably hardware error on
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura). Table for X see error 02.01.001.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB faulty.
1LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:21
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
02.03.001
02.03.002
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or transmission error from Graphics Controller PCB
(Evita 4, EvitaXL) or Front Panel PCB (Evita 2 dura) via CPU
8332 PCB to Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
1LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
MODDAT:11.04.2003 12:01:22
Copyright Dräger Medical - For internal use only!
2003-04-16Page 7 of 171
02.03.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:22
02.03.004
02.03.005
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
02.03.006
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
1LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4, EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 8 of 171
02.03.007
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:22
02.03.020
FUNCTION_SERVICE:
Key readout on Graphics Controller PCB (Evita 4, EvitaXL) or
Front Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software or hardware error on Graphics Controller PCB (Evita
4, EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
Graphics Controller PCB (Evita 4, EvitaXL) or Front Panel
PCB (Evita 2 dura) is faulty.
Replace Graphics Controller PCB (Evita 4, EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
02.04.001
02.04.002
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
MODDAT:11.04.2003 12:01:22
MODDAT:11.04.2003 12:01:22
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 9 of 171
02.04.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:22
02.04.004
FUNCTION_SERVICE:
In neo mode according to safety software wrong flow sensor
being used
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In adult mode according to safety software wrong flow sensor
being used.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
REASON / SHORTREASON_CUSTOMER:
02.04.005
02.04.020
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Safety software: RESET - RAM error
CONSEQ_SERVICE:
An inconsistency in the values for the filtered air pressure on
the CPU 68332 PCB and the Pneumatic Controller PCB has
occurred.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
MODDAT:11.04.2003 12:01:40
MODDAT:11.04.2003 12:01:22
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 10 of 171
02.05.001
Evita 4 / 2dura / XLError code list for
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Install the relevant software on the CPU 68332 PCB and the
Pneumatic Controller PCB.
MODDAT:11.04.2003 12:01:22
02.05.002
02.05.003
REASON_SERVICE:
Software error or impermissible "BTPS shut-off" activated at
ambient pressure below 961 mbar.
Flow insp, current counter
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
The HPSV activation monitor signals error.
REASON_SERVICE:
HPSV cartridge (AIR or O2) or one of the two HPSV
Activation PCBs or the Pneumatic Controller PCB is faulty. In
diagnosis mode “ Valves ” if error is current
status of error can be read.
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
The HPSV cartridge or the HPSV Activation PCB or the
Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace HPSV cartridge (AIR or O2). If replacement of the
HPSV cartridge was not successful, replace the HPSV
Activation PCB (AIR or O2). If replacement of the HPSV
Activation PCB was not successful, replace the Pneumatic
Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
MODDAT:11.04.2003 12:01:22
02.05.004
FUNCTION_SERVICE:FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error or impermissible BTPS shut-off activated at
ambient pressure below 961 mbar.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
Install the relevant software on the CPU 68332 PCB and on
the Pneumatic Controller PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 11 of 171
02.05.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software
MODDAT:11.04.2003 12:01:22
02.06.001
02.06.002
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Software/hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
- Probably software error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) - CAN
communication with CPU 68332 PCB and Pneumatic
Controller PCB disturbed.
Software error or hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
Software error or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura) is faulty or CPU 68332 PCB.
Install software on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). If the software installation
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and
download the relevant software. If replacement of the
Graphics Controller PCB or Front Panel PCB was not
successful, replace the CPU 68332 PCB and install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:22
FUNCTION_SERVICE:FAILURE:
Software error or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura) is faulty or CPU 68332 PCB
is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Install software on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). If the software installation
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and
download the relevant software. If replacement of the
Graphics Controller PCB or Front Panel PCB was not
successful, replace the CPU 68332 PCB and install the
relevant software.
REASON_SERVICE:
Software error on Graphics Controller PCB (Evita 4) or Front
Panel PCB (Evita 2 dura) or CAN communication with CPU
68332 PCB and Pneumatic Controller PCB disturbed.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 12 of 171
02.06.003
Evita 4 / 2dura / XLError code list for
Software error or hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Software error or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB is faulty (Evita 2 dura) or CPU 68332 PCB
is faulty.
Install software on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). If the software installation
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and
download the relevant software. If replacement of the
Graphics Controller PCB or Front Panel PCB was not
successful, replace the CPU 68332 PCB and install the
relevant software.
MODDAT:11.04.2003 12:01:41
02.06.004
REASON_SERVICE:
- Software error on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). - Faulty CAN
communication to CPU 68332 PCB and Pneumatic Controller
PCB.
Software/hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Probably software error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN
communication with CPU 68332 PCB and Pneumatic
Controller PCB disturbed.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
Software error or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB is faulty (Evita 2 dura) or CPU 68332 PCB
is faulty.
Install software on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). If the software installation
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or the Front Panel PCB (Evita 2 dura) and
download the relevant software. If replacement of the
Graphics Controller PCB or Front Panel PCB was not
successful, replace the CPU 68332 PCB and install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
02.06.020
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
2003-04-16Page 13 of 171
02.06.021
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:40
02.07.020
02.08.001
REASON_SERVICE:
RAM or software error on CPU 68332 PCB, Pneumatic
Controller PCB or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:40
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
Software error or Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura) or CPU 68332 PCB is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Install relevant software on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB. If the software installation was
not successful, replace the Graphics Controller PCB (Evita 4;
EvitaXL) or the Front Panel PCB (Evita 2 dura) and install the
relevant software. If replacement of the Graphics Controller
PCB or Front Panel PCB was not successful, replace the CPU
68332 PCB and install the relevant software.
REASON_SERVICE:
Probably software error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN
communication with CPU 68332 PCB and Pneumatic
Controller PCB disturbed.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 14 of 171
02.08.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) or CPU 68332 PCB is faulty.
Install relevant software on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB. If the software installation was
not successful, replace the Graphics Controller PCB (Evita 4;
EvitaXL) or the Front Panel PCB (Evita 2 dura) and install the
relevant software. If replacement of the Graphics Controller
PCB or Front Panel PCB was not successful, replace the CPU
68332 PCB and install the relevant software.
MODDAT:11.04.2003 12:01:23
02.08.020
02.08.021
REASON_SERVICE:
Probably software error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CAN
communication with CPU 68332 PCB and Pneumatic
Controller PCB disturbed.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 15 of 171
02.09.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:23
02.09.002
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic
Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic
Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
02.09.003
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic
Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:23
Copyright Dräger Medical - For internal use only!
2003-04-16Page 16 of 171
02.09.004
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or Pneumatic
Controller PCB is faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Graphics Controller PCB (Evita 4; EvitaXL) or
Front Panel PCB (Evita 2 dura) and install the relevant
software. If replacement of the Graphics Controller PCB or
Front Panel PCB was not successful, replace the Pneumatic
Controller PCB and install the relevant software.
MODDAT:11.04.2003 12:01:23
02.09.020
02.09.021
REASON_SERVICE:
Software error on CPU 68332 PCB or Graphics Controller
PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) or
Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 17 of 171
02.11.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:23
02.11.002
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic
Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The CPU 68332 PCB monitors the timing of the Pneumatic
Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
02.11.020
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
Copyright Dräger Medical - For internal use only!
2003-04-16Page 18 of 171
02.12.001
Evita 4 / 2dura / XLError code list for
The inspiratory pressure is too high for longer than ... second
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:18:52
FUNCTION_SERVICE:
si12_Err, Paw high too long, reset
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
The inspiratory pressure is too high for longer than ... second
and cannot be relieved by way of the emergency venting valve
Y1.3. The inspiratory pressure is measured on the Pneumatic
Controller PCB and on the CO2 Carrier PCB. Possible
causes: Inspiratory pressure measurement on Pneumatic
Controller PCB (1st channel) or CO2 Carrier PCB (2nd
channel) defective. Read values with service mode. If the
error is on the CO2 Carrier PCB, the cause may be contact
problems. The CO2 Carrier PCB must be secured by both
fixing screws; Inspiratory pressure sensor faulty; Pneumatic
fault; Excessive measurement from expiratory pressure
sensor (Note: If ground (GND) is missing the Paw sensors
measure approx. 115 mbar).
FAILURE:
The inspiratory pressure is too high for longer than ... second
and cannot be relieved by way of the emergency venting valve
Y1.3. The inspiratory pressure is measured on the Pneumatic
Controller PCB and on the CO2 Carrier PCB. Possible
causes: Inspiratory pressure measurement on Pneumatic
Controller PCB (1st channel) or CO2 Carrier PCB (2nd
channel) defective. Read values with service mode. If the
error is on the CO2 Carrier PCB, the cause may be contact
problems. The CO2 Carrier PCB must be secured by both
fixing screws; Inspiratory pressure sensor faulty; Pneumatic
fault; Excessive measurement from expiratory pressure
sensor (Note: If ground (GND) is missing the Paw sensors
measure approx. 115 mbar).
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:23
MODDAT:11.04.2003 12:01:23
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 19 of 171
02.14.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:24
02.14.002
02.14.020
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
02.15.001
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 20 of 171
02.15.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:23
02.16.001
02.16.002
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Hardware boot, test data not reset, Reset
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
FAILURE:
CPU 68332 PCB is faulty.
Replace CPU 68332 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
02.16.020
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:23
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 21 of 171
02.17.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic
Controller PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software. If
replacement of the Graphics Controller PCB or Front Panel
PCB was not successful, replace the CPU 68332 PCB and
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
MODDAT:11.04.2003 12:01:24
02.17.002
02.17.020
REASON_SERVICE:
RAM or software error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or on CPU 68332
PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or SW error on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura) or CPU 68332
PCB or Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
Graphics Controller PCB (Evita 4; EvitaXL) or Front Panel
PCB (Evita 2 dura) or CPU 68332 PCB or Pneumatic
Controller PCB is faulty.
Replace Graphics Controller PCB (Evita 4; EvitaXL) or Front
Panel PCB (Evita 2 dura) and install the relevant software. If
replacement of the Graphics Controller PCB or Front Panel
PCB was not successful, replace the CPU 68332 PCB and
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
FUNCTION_SERVICE:FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 22 of 171
02.17.021
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB is faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:24
02.18.001
02.18.002
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
FUNCTION_SERVICE:FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 23 of 171
02.18.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:24
02.18.020
02.19.001
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 24 of 171
02.19.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:24
02.19.003
02.21.000
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:24
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
02.21.001
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:41
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 25 of 171
02.22.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:24
02.40.001
02.40.002
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Cold/warm start save monitoring
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Cold/warm start saving in RAM on CPU 68332 PCB is
monitored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Data packets between CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:41
FAILURE:
The Pneumatic Controller PCB detects Service codes 02.40...
to 02.62. They correspond to 02.00... to 02.22... of the CPU
68332 PCB.
See explanation under Service codes 02.00... to 02.22... of
CPU 68332 PCB.
MODDAT:12.12.2002 09:19:02
Data packets between CPU 68332 PCB and Pneumatic
Controller PCB cannot be fully transferred.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.40.003
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Synchronisation not possible
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
After unit start it was not possible to establish synchronisation
between CPU 68332 PCB and Pneumatic Controller PCB
within 10 seconds.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 26 of 171
02.40.004
Evita 4 / 2dura / XLError code list for
Data packets between CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Data packets between CPU 68332 PCB and Pneumatic
Controller PCB cannot be fully transferred.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:02
02.40.005
02.40.020
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332
Replace CPU PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.41.001
02.41.002
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
MODDAT:12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16Page 27 of 171
02.41.003
Evita 4 / 2dura / XLError code list for
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:02
02.41.004
02.41.005
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.41.006
02.41.007
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Safety-relevant alarm values are checked for plausibility
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
Safety-relevant alarm values are checked for plausibility. RAM
or SW error on CPU 68332 PCB or Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
MODDAT:12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16Page 28 of 171
02.41.020
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:02
02.41.021
02.41.022
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
Only up to SW 1.04. RAM or SW error on CPU PCB 68332 or
Pneumatic Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.42.001
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
1LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
Copyright Dräger Medical - For internal use only!
2003-04-16Page 29 of 171
02.42.002
Evita 4 / 2dura / XLError code list for
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:02
02.42.003
02.42.020
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
Screen content (on Evita 2 dura 7-segment displays) is read
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
The screen content (on Evita 2 dura also 7-segment displays)
is read and compared with the requirement.
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:02
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.43.001
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW or transmission error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
SW or transmission error from Graphics Controller PCB or
Front Panel PCB via CPU 68332 PCB and Pneumatic
Controller PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 30 of 171
02.43.002
Evita 4 / 2dura / XLError code list for
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:03
02.43.003
02.43.004
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
02.43.005
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 31 of 171
02.43.006
Evita 4 / 2dura / XLError code list for
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:03
02.43.007
02.43.020
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
SW or hardware error reading buttons
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on Graphics Controller PCB or Front
Panel PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
SW or hardware error reading buttons on Graphics Controller
PCB or Front Panel PCB
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.44.001
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 32 of 171
02.44.002
Evita 4 / 2dura / XLError code list for
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:03
02.44.003
02.44.004
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
In neoflow mode the wrong flow sensor is used.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
In adult mode the wrong flow sensor is used.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
The wrong flow sensor is used in neoflow mode according to
the safety software.
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
The wrong flow sensor is used in adult mode according to the
safety software. Hardware or SW error on CPU 68332 PCB or
Pneumatic Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
02.44.020
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Inconsistency of values for filtered air pressure
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
An inconsistency in the values for the filtered air pressure on
the CPU PCB and Pneumatic Controller PCB has occurred
Copyright Dräger Medical - For internal use only!
2003-04-16Page 33 of 171
02.45.001
Evita 4 / 2dura / XLError code list for
Hardware or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Hardware or SW error on CPU PCB 68332 or Pneumatic
Controller PCB
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:03
02.45.002
02.45.003
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error Faulty CAN communication to CPU 68332 PCB and Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
02.46.003
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 35 of 171
02.46.004
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:03
02.46.020
02.46.021
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:03
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
02.47.020
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Faulty Graphics Controller PCB Faulty Front Panel PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU 68332 PCB or Pneumatic
Controller PCB, error on Graphics Controller PCB or Front
Panel PCB
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 36 of 171
02.48.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:04
02.48.002
02.48.020
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
02.48.021
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Pneumatic Controller
PCB Faulty CPU 68332 PCB Faulty Front Panel PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 37 of 171
02.49.001
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB or Front Panel PCB
Faulty CAN communication to CPU 68332 PCB and
Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:04
02.49.002
02.49.003
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Faulty Graphics Controller PCB Faulty Front Panel PCB.
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by
the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by
the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.49.004
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by
the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 38 of 171
02.49.020
Evita 4 / 2dura / XLError code list for
SW error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
"SW error on Graphics Controller PCB; Front Panel PCB;
CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:04
02.49.021
02.51.001
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
Faulty Graphics Controller PCB Faulty Front Panel PCB.
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.51.002
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
The timing of the Pneumatic Controller PCB is monitored by
the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 39 of 171
02.51.020
Evita 4 / 2dura / XLError code list for
Timing of Pneumatic Controller PCB
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The timing of the Pneumatic Controller PCB is monitored by
the CPU 68332 PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:04
02.52.001
02.52.020
REASON_SERVICE:
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
Inspiratory pressure measurement
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
The inspiratory pressure is too high for longer than 1 second
and cannot be relieved by way of the emergency venting valve
Y1.3. The inspiratory pressure is measured on the Pneumatic
Controller PCB and on the CO2 Carrier PCB
Inspiratory pressure measurement on Pneumatic Controller
PCB or CO2 Carrier PCB faulty. Inspiratory pressure sensor
faulty Fault in pneumatic unit
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 40 of 171
02.54.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:04
02.54.002
02.54.020
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.55.001
02.55.020
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:04
MODDAT:12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16Page 41 of 171
02.56.001
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:05
02.56.002
02.56.020
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332
"Replace CPU PCB; Replace Pneumatic Controller PCB"
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
02.57.001
02.57.002
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16Page 42 of 171
02.57.020
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:05
02.57.021
02.58.001
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Pneumatic Controller PCB Faulty CPU 68332 PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
02.58.002
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 43 of 171
02.58.003
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
MODDAT:12.12.2002 09:19:05
02.58.020
02.59.001
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty Graphics Controller PCB Faulty Front Panel PCB
Faulty CPU 68332 PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
"RAM or SW error on Graphics Controller PCB or Front Panel
PCB; CPU 68332 PCB; Pneumatic Controller PCB"
Replace CPU PCB; Replace Pneumatic Controller PCB;
Replace Graphics Controller PCB or Front Panel PCB
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
02.59.002
REASON_SERVICE:
Faulty CPU PCB Faulty Pneumatic Controller PCB
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty CPU PCB Faulty Pneumatic Controller PCB
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:19:05
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 44 of 171
02.59.003
Evita 4 / 2dura / XLError code list for
RAM or software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
RAM or SW error on CPU PCB 68332 or Pneumatic
Controller PCB.
Replace CPU PCB; Replace Pneumatic Controller PCB
MODDAT:12.12.2002 09:19:05
02.61.000
02.61.001
REASON_SERVICE:
Faulty CPU PCB Faulty Pneumatic Controller PCB
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:41
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:42
CPU 68332 PCB is faulty.
Replace CPU 68332 PCB and install relevant software.
02.62.001
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
The Pneumatic Controller PCB detects Service codes 02.40...
to 02.62.... They correspond to 02.00... to 02.22... of the CPU
68332 PCB.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Replace CPU 68332 PCB and install the
relevant software. If replacement of the CPU 68332 PCB was
not successful, replace the Pneumatic Controller PCB and
install the relevant software. EvitaXL: - See explanation under
Service codes 02.00... to 02.22... of CPU 68332 PCB.
MODDAT:11.04.2003 12:01:42
Copyright Dräger Medical - For internal use only!
2003-04-16Page 45 of 171
02.71.001
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:24
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the
loudspeaker (approx. 50 Ohms) is tapped via a series resistor
(5 Ohms) on the Graphics Controller PCB/Front Panel PCB
and made available to the system as digital information.
Based on this signal a target/actual comparison of the
loudspeaker activation is made. The loudspeaker is located in
the control panel/front panel. Note: See also errors 02.71.012
and 02.71.013.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Evita 4; Evita 2 dura: Software error, if the error occurs
between multiple operations of suction mode without
ventilation occurring in the meantime. This error is eliminated
on Evita 4 as from SW 2.21 (Japan from SW 1.09) and on
Evita 2 dura in versions later than 3.00. Cable error between
CPU 68332 PCB and loudspeaker in control panel.
Loudspeaker is faulty. Trouble-shooting: Measure resistance
of loudspeaker via connecting cable of control panel incl.
cable between pins 9 and 10 of SUB-D cable connector. Test
value = 55 plus/minus 8 Ohms. This value incl. shunt
resistance must be 5 Ohms above the value of the
loudspeaker which can be measured after opening the control
panel. - With diagnosis mode at lowest volume setting, see
"Repair Instructions - Service no./modes, error list" section
3.2.1, Diagnosis page “Microprocessor ”,
front”, page 60. - With external DS mode via PC with
lowest volume setting, see "Repair Instructions - Service
no./modes, error list" section 9.4.7, Testing loudspeaker, page
143, or see " Repair Instructions - Service no./modes, error
list" section 9.5.7, Testing loudspeaker, page 153. EvitaXL:
Loudspeaker detection is faulty.
FAILURE:
Audible alarm absent
Evita 4/Evita 2 dura: - Measure resistance of loudspeaker via
connecting cable of control panel incl. cable between pins 9
and 10 of SUB-D cable connector. Test value = 55 plus/minus
8 Ohms. This value incl. shunt resistance must be 5 Ohms
above the value of the loudspeaker which can be measured
after opening the control panel. - With diagnosis mode at
lowest volume setting, see "Repair Instructions - Service
no./modes, error list" section 3.2.1, Diagnosis page
“Microprocessor ” of “;Front”,
page 60. - With external DS mode via PC with lowest volume
setting, see "Repair Instructions - Service no./modes, error
list" section 9.4.7, Testing loudspeaker, page 143, or see "
Repair Instructions - Service no./modes, error list" section
9.5.7, Testing loudspeaker, page 153. EvitaXL: Replace
Graphics Controller 8 PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 46 of 171
02.71.002
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:24
02.71.003
FUNCTION_SERVICE:
Flow metering on Pneumatic Controller PCB faulty
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Evita 4/Evita 2 dura: - As of SW 1.05: Replace Pneumatic
Controller PCB. To SW 1.04: Error is reset if ambient
pressure on switch-on is below 961 mbar, but is not displayed
on-screen. Only replace Pneumatic Controller PCB if error
message is displayed on-screen in operation. To SW 1.04:
Error is reset if ambient pressure on switch-on is below 961
mbar, but is not displayed on-screen. Only replace Pneumatic
Controller PCB if error message is displayed on-screen in
operation. EvitaXL: - A/D conversion on Pneumatic Controller
PCB is faulty.
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
FAILURE:
Evita 4; Evita 2 dura: As of SW 1.05: Replace Pneumatic
Controller PCB. To SW 1.04: Error is reset if ambient
pressure on switch-on is below 961 mbar, but is not displayed
on-screen. Only replace Pneumatic Controller PCB if error
message is displayed on-screen in operation. EvitaXL: The
Pneumatic Controller PCB is faulty.
Evita 4/Evita 2 dura: - Replace both supply pressure
sensors. - Replace Pneumatic Controller PCB and download
the relevant software. EvitaXL: - Replace Pneumatic
Controller PCB and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to the power failure. The duration of this power failure
alarm is dependent on the capacitance of the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. A unit malfunction is signalled if one of the following
criteria is met: a) Final charging voltage too high (above 11 V),
voltage regulator faulty b) Final charging voltage too low
(below 8 V), charging circuit faulty c) Reference voltage too
high (above 6.3 V), discharging circuit faulty d) Reference
voltage too low (below 4.5 V), no Gold Cap e) Capacitance
less than 30mF, Gold Cap capacitance too low
CONSEQ_SERVICE:
Gold Cap capacitor, capacitance too low
REASON_SERVICE:
Gold Cap capacitance is too low
FAILURE:
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - see Repair Instructions - Electronic
components, section 5.5, Conversions, page 35 CO2 Carrier
PCB. EvitaXL: - Replace CPU 68332 PCB and install the
relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 47 of 171
02.71.004
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, voltage too high
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to the power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high (above 11
V), voltage regulator defectiveb) Final charging voltage too
low (below 8 V), charging circuit defectivec) Reference
voltage too high (above 6.3 V), discharging circuit
defectived) Reference voltage too low (below 4.5 V), no
Gold Cape) Capacitance less than 30mF, Gold Cap
capacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to the power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high (above 11
V), voltage regulator defectiveb) Final charging voltage too
low (below 8 V), charging circuit defectivec) Reference
voltage too high (above 6.3 V), discharging circuit
defectived) Reference voltage too low (below 4.5 V), no
Gold Cape) Capacitance less than 30mF, Gold Cap
capacitance too low
see Repair Instructions - Electroniccomponents, section 5.5,
Conversions,page 35 CO2 Carrier PCB. Replace CO2 carrier
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 48 of 171
02.71.005
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, voltage too low
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to a power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high(above 11
V), voltage regulatordefectiveb) Final charging voltage too
low(below 8 V), charging circuit defectivec) Reference
voltage too high(above 6.3 V), discharging
circuitdefectived) Reference voltage too low(below 4.5
V), no Gold Cape) Capacitance less than 30mF, Gold
Capcapacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to a power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high(above 11
V), voltage regulatordefectiveb) Final charging voltage too
low(below 8 V), charging circuit defectivec) Reference
voltage too high(above 6.3 V), discharging
circuitdefectived) Reference voltage too low(below 4.5
V), no Gold Cape) Capacitance less than 30mF, Gold
Capcapacitance too low
see Repair Instructions - Electroniccomponents, section 5.5,
Conversions,page 35 CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 49 of 171
02.71.006
Evita 4 / 2dura / XLError code list for
Gold Cap capacitor, tester
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:08.11.2002 14:55:37
FUNCTION_SERVICE:
Gold Cap capacitor, tester defective
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to the power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high (above 11
V), voltage regulator defectiveb) Final charging voltage too
low (below 8 V), charging circuit defectivec) Reference
voltage too high (above 6.3 V), discharging circuit
defectived) Reference voltage too low (below 4.5 V), no
Gold Cape) Capacitance less than 30mF, Gold Cap
capacitance too low
FAILURE:
In case of a voltage failure the Gold Cap on the CO2 Carrier
PCB supplies the piezo alarm generator with voltage and so
alerts to the power failure. The duration of this power failure
alarm is dependent on the capacitanceof the Gold Cap and
so both the function and the capacitance are subjected to an
hourly test, first performed 5 minutes after the unit is switched
on. Aunit malfunction is signalled if one of the following
criteria is met:a) Final charging voltage too high (above 11
V), voltage regulator defectiveb) Final charging voltage too
low (below 8 V), charging circuit defectivec) Reference
voltage too high (above 6.3 V), discharging circuit
defectived) Reference voltage too low (below 4.5 V), no
Gold Cape) Capacitance less than 30mF, Gold Cap
capacitance too low
see Repair Instructions - Electronic components, section 5.5,
Conversions, page 35 CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
02.71.007
Boot test faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Boot test faulty
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
CPU 68332 PCB (more probably) or Pneumatic Controller
PCB defective. Note: Displayed in addition to other error
messages.
1LAST_SOFTWARE: 99,99
FAILURE:
CPU 68332 PCB (more probably) or Pneumatic Controller
PCB defective. Note: Displayed in addition to other error
messages.
Replace CPU PCB; Replace Pneumatic Controller PCB.
a) CO2 Carrier PCB defectiveb) CPU 68332 PCB defective
Piezo alarm generator: Current too high
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Piezo alarm generator, current too high
1LAST_SOFTWARE: 99,99
FAILURE:
As well as signalling a power failure alarm, the piezo also
functions as a replacement alarm for the loudspeaker in the
event of a defect in the CPU 68332 (standstill). To ensure
this, the piezo is suppressed on the CO2 Carrier PCB by the
CPU 68332 PCBwith a square signal. If this cycle misses,
the resultant muting of the piezo is cancelled and a
continuous beep sounds. To test the piezo the muting is
purposely cancelled. The current through the piezo, which
now risesbecause of the cancelled muting, can be
measured a a voltage drop via a series resistor by the A/D
converter on the CO2 Carrier PCB.
- With diagnosis mode, see "Repair Instructions- Service
no./modes, error list" section 3.3.2, Diagnosis page
Microprocessor of electronics, page 66.- With external DS
mode via PC, see "Repair Instructions - Service no./modes,
error list" section 9.4.6, Testing horn, page 142.
REASON / SHORTREASON_CUSTOMER:
MODDAT:08.11.2002 14:55:18
FAILURE:
As well as signalling a power failure alarm, the piezo also
functions as a replacement alarm for the loudspeaker in the
event of a defect in the CPU 68332 (standstill). To ensure
this, the piezo is suppressed on the CO2 Carrier PCB by the
CPU 68332 PCBwith a square signal. If this cycle misses,
the resultant muting of the piezo is cancelled and a
continuous beep sounds. To test the piezo the muting is
purposely cancelled. The current through the piezo, which
now risesbecause of the cancelled muting, can be
measured a a voltage drop via a series resistor by the A/D
converter on the CO2 Carrier PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
- With diagnosis mode, see "Repair Instructions- Service
no./modes, error list" section 3.3.2, Diagnosis page
Microprocessor of electronics, page 66.- With external DS
mode via PC, see "Repair Instructions - Service no./modes,
error list" section 9.4.6, Testing horn, page 142.
REASON_SERVICE:
a) CO2 Carrier PCB defectiveb) CPU 68332 PCB defective
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 51 of 171
02.71.010
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:24
02.71.011
FUNCTION_SERVICE:
As well as signalling a power failure alarm, the piezo also
functions as a replacement alarm for the loudspeaker in the
event of a fault in the CPU 68332 (standstill). To ensure this,
the piezo is suppressed on the CO2 Carrier PCB by the CPU
68332 PCB with a square signal. If this cycle misses, the
resultant muting of the piezo is cancelled and a continuous
beep sounds. To test the piezo the muting is purposely
cancelled. The current through the piezo, which now rises
because of the cancelled muting, can be measured as a
voltage drop via a series resistor by the A/D converter on the
CO2 Carrier PCB.
CONSEQ_SERVICE:
Piezo alarm generator, current too low
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
FAILURE:
Evita 4/Evita 2 dura: - CO2 Carrier PCB or CPU 68332 PCB is
faulty. EvitaXL: - The piezo cable is unplugged or the piezo is
faulty or the CO2 Carrier PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
Diagnosis page Microprocessor; of electronics, page 66. With external DS mode via PC, see "Repair Instructions Service no./modes, error list" section 9.4.6, Testing horn,
page 142. EvitaXL: Plug in the piezo cable. If plugging in the
piezo cable was not successful, replace the piezo. If
replacing the piezo was not successful, replace the CO2
Carrier PCB.
MODDAT:11.04.2003 12:01:25
FUNCTION_SERVICE:
The specified setting of the nebuliser valve is continuously
checked against the returned signal from the activation
electronics.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Evita 4/Evita 2 dura: - Drug nebuliser valve monitoring is
faulty. EvitaXL: - The drug nebuliser valve comparator is
faulty.
FAILURE:
Drug nebuliser valve is not connected or Pneumatic Controller
PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the drug nebuliser valve is
connected is not detected.)
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 52 of 171
02.71.012
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:25
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the
loudspeaker (approx. 50 Ohms) is tapped via a series resistor
(5 Ohms) on the Graphics Controller PCB and made available
to the system as digital information. Based on this signal a
target/actual comparison of the loudspeaker activation is
made. If this target/actual comparison detects an error, a unit
malfunction is generated.
CONSEQ_SERVICE:
Loudspeaker is detected too infrequently.
FAILURE:
Evita 4/Evita 2 dura: a) Cable fault between CO2 Carrier PCB
and loudspeaker in control panel or b) Loudspeaker faulty or
c) Graphics Controller PCB faulty or d) CPU 68332 PCB
faulty EvitaXL: -Graphics Controller 8 PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Trouble-shooting: Measure resistance of
loudspeaker via connecting cable of control panel incl. cable
between pins 9 and 10 of SUB-D cable connector. Test value
= 55 plus/minus 8 Ohms. This value incl. shunt resistance
must be 5 Ohms above the value of the loudspeaker which
can be measured after opening the control panel. - With
diagnosis mode at lowest volume setting, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
Diagnosis page “Microprocessor” of
“Front”, page 60. - With external DS mode via
PC with lowest volume setting, see "Repair Instructions Service no./modes, error list" section 9.4.7, Testing
loudspeaker, page 143, or see " Repair Instructions - Service
no./modes, error list" section 9.5.7, Testing loudspeaker, page
153. Note: See also error 02.71.001 Repair cable error
between CO2 Carrier PCB and loudspeaker in control panel.
Replace loudspeaker. Replace Graphics Controller PCB and
install the relevant software. Replace CPU 68332 PCB and
install the relevant software. EvitaXL: - Replace Graphics
Controller 8 PCB and install the relevant software.
REASON_SERVICE:
EvitaXL: - The sound detection is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 53 of 171
02.71.013
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:25
FUNCTION_SERVICE:
The activation signal from the CPU 68332 PCB to the
loudspeaker (approx. 50 Ohms) is tapped with a series
resistor (5 Ohms) on the Graphics Controller PCB and made
available to the system as digital information. Based on this
signal a target/actual comparison of the loudspeaker
activation is made. If this target/actual comparison detects an
error, a unit malfunction is generated.
CONSEQ_SERVICE:
Loudspeaker detected too frequently.
FAILURE:
Evita 4/Evita 2 dura: a) Cable error between CO2 Carrier PCB
and loudspeaker in control panel or b) Loudspeaker faulty or
c) Graphics Controller PCB faulty or d) CPU 68332 PCB
faulty EvitaXL: - The Graphics Controller 8 PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Trouble-shooting: Measure resistance of
loudspeaker via connecting cable of control panel incl. cable
between pins 9 and 10 of SUB-D cable connector. Test value
= 55 plus/minus 8 Ohms. This value incl. shunt resistance
must be 5 Ohms above the value of the loudspeaker which
can be measured after opening the control panel. - With
diagnosis mode at lowest volume setting, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
Diagnosis page “Microprocessor” of
“Front”, page 60. - With external DS mode via
PC with lowest volume setting, see "Repair Instructions Service no./modes, error list" section 9.4.7, Testing
loudspeaker, page 143, or see " Repair Instructions - Service
no./modes, error list" section 9.5.7, Testing loudspeaker, page
153. Note: See also error 02.71.001 Repair cable error
between CO2 Carrier PCB and loudspeaker in control panel.
Replace loudspeaker. Replace Graphics Controller PCB or
Front Panel PCB and install the relevant software. Replace
CPU 68332 PCB and install the relevant software. EvitaXL: Replace Graphics Controller 8 PCB and install the relevant
software.
02.71.014
REASON_SERVICE:
Evita 4/Evita 2 dura: a) Cable error between CO2 Carrier PCB
and loudspeaker in control panel or b) Loudspeaker faulty or
c) Graphics Controller PCB faulty or d) CPU 68332 PCB
faulty EvitaXL: - The sound detection is faulty.
+ 15 V is too low
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances. A shortfall of + 14 V results in a unit
malfunction signal.
CONSEQ_SERVICE:
+ 15 V is too low or not present.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:25
FAILURE:
Power supply unit is faulty or CO2 Carrier PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Check of power supply unit voltages
(several possibilities): Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
diagnosis page “Microprocessor” of
“Front”, page 60. - With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.4.1, Testing voltages, page 133. If OK = Replace
CO2 Carrier PCB If not OK = Replace power supply unit
EvitaXL: Check power supply unit voltages and replace power
supply unit as necessary. If the power supply unit is OK,
replace the CO2 Carrier PCB.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 54 of 171
02.71.015
Evita 4 / 2dura / XLError code list for
+10V reference voltage too low
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:25
02.71.016
FUNCTION_SERVICE:
The voltage is reduced on the CO2 Carrier PCB from the + 15
V and also measured there. A shortfall of +9 V results in a unit
malfunction signal.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
The reference voltage +10V falls short of the tolerance.
AD conversion for O2 metering is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The pre-amplified voltage of the O2 sensor is read-in on the
Pneumatic Controller PCB by two independent A/D
converters. Trouble-shooting: With diagnosis mode, see
"Repair Instructions - Service no./modes, error list" section
3.4.2, diagnosis page “ Sensors”
Pneumatic”, page 73. - With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.3.4, Sensors, page 125.
CONSEQ_SERVICE:
The voltages are different.
1LAST_SOFTWARE: 99,99
FAILURE:
Power supply unit or CO2 Carrier PCB is faulty.
Measure power supply unit voltages, replace power supply
unit as necessary. If replacing the power supply unit was not
successful, replace the CO2 Carrier PCB.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:25
FAILURE:
The Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software.
02.71.017
REASON_SERVICE:
The raw voltage and the amplified voltage of the O2 sensor
are not within the tolerance.
Activation of O2/Air switch-over valve is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/Air switch-over valve is
continuously checked against the returned signal of the
activation electronics on the Pneumatic Controller PCB.
CONSEQ_SERVICE:
The returned activation signal for the O2/AIR switch-over
valve produces incorrect values.
REASON_SERVICE:
The switch-over valve is not connected or is faulty.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:25
FAILURE:
Trouble-shooting: With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page “ Valves” of
Pneumatic”, page 70. With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the O2/AIR switch-over valve is
connected is not detected.)
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 55 of 171
02.71.018
Evita 4 / 2dura / XLError code list for
Wrong drug nebuliser gas
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
02.71.019
FUNCTION_SERVICE:
During a drug nebulisation the specified setting of the O2/AIR
switch-over valve is continuously checked against the
returned signal from the activation electronics. Up to SW 3.21
lower tolerance range. As from SW 3.22 tolerance range
extended and so error occurs less frequently.
CONSEQ_SERVICE:
Returned activation signal of O2/AIR switch-over valve
producing wrong values.
REASON_SERVICE:
Wrong drug nebuliser gas.
Cold/warm start detection is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The cold/warm start detection is effected with the power
switch auxiliary contacts of the On switch in the power supply
unit. The CO2 Carrier PCB reads in the power switch auxiliary
contacts.
1LAST_SOFTWARE: 99,99
FAILURE:
The Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page "Pneumatic valves". - With external service
mode, see Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves. - Replace Pneumatic
Controller PCB and install the relevant software. EvitaXL: Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:25
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page “ Microprocessor” of “
Electronic”, page 66. - With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.4.2, Power supply unit status, page 134. Possible
cause of error: - Power switch auxiliary contacts in power
switch are faulty or wrongly connected. - CO2 Carrier PCB
faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Evita 4/Evita 2 dura: Replace power supply unit. If replacing
the power supply unit was not successful, replace the CO2
Carrier PCB. EvitaXL: Replace power supply unit. If
replacing the power supply unit was not successful, replace
the CO2 Carrier PCB. If replacement of the CO2 Carrier PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software.
REASON_SERVICE:
Evita 4/Evita 2 dura: - Power switch contacts not
functioning. - CO2 Carrier PCB faulty. EvitaXL: - Cold start
detection is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 56 of 171
02.71.020
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
CPU 68332 PCB or Pneumatic Controller PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
MODDAT:11.04.2003 12:01:25
02.71.021
CONSEQ_SERVICE:
The data to initialise the interrupt masks are incorrectly stored.
REASON_SERVICE:
RAM or software error on CPU 68332 PCB or Pneumatic
Controller PCB or on Graphics Controller PCB (Evita 4;
EvitaXL) or Front Panel PCB (Evita 2 dura).
Time basis is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The quartz times of the CPU 68332 PCB and the Pneumatic
Controller PCB are compared. In the event of a difference an
error message is delivered.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:25
FAILURE:
CPU 68332 PCB or Pneumatic Controller PCB faulty.
Replace CPU 68332 PCB and install the relevant software. If
replacement of the CPU 68332 PCB was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
02.72.001
02.72.002
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
Copyright Dräger Medical - For internal use only!
2003-04-16Page 57 of 171
02.72.004
02.72.005
02.72.006
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
02.72.007
02.72.008
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 58 of 171
02.72.009
02.72.010
02.72.020
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
MODDAT:12.12.2002 09:19:05
02.72.040
02.72.041
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 59 of 171
02.72.042
02.72.050
02.72.051
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.060
02.72.061
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 60 of 171
02.72.070
02.72.071
02.72.080
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.090
02.72.100
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 61 of 171
02.72.201
02.72.202
02.72.203
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.204
02.72.205
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 62 of 171
02.72.206
02.72.207
02.72.208
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.209
02.72.211
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 63 of 171
02.72.212
02.72.213
02.72.214
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.215
02.72.216
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 64 of 171
02.72.217
02.72.218
02.72.219
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.221
02.72.222
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 65 of 171
02.72.223
02.72.224
02.72.225
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.226
02.72.227
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 66 of 171
02.72.228
02.72.229
02.72.231
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.232
02.72.233
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 67 of 171
02.72.234
02.72.235
02.72.236
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
02.72.237
02.72.238
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
MODDAT:12.12.2002 09:19:06
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 68 of 171
02.72.239
Evita 4 / 2dura / XLError code list for
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
MODDAT:12.12.2002 09:19:06
03.01.001
03.01.002
Touchscreen not detected on booting
FIRST_SOFTWARE:
FUNCTION_SERVICE:
After Evita 4 power-on the touchscreen is not detected by the
Graphics Controller PCB. The touchscreen communicates via
the RS232 interface with the Graphics Controller PCB.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
After unit power-on the touchscreen is not detected by the
Graphics Controller PCB. The touchscreen communicates via
the RS232 interface with the Graphics Controller PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting, Evita 4: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
diagnosis page "Microprocessor" of "Front", page 60. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.5.5, Testing touchscreen, page
150.
Replace touchscreen. If replacement of the touchscreen was
not successful, replace the Graphics Controller PCB (Evita 4;
EvitaXL) and install the relevant software.
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
diagnosis page “Microprocessor” of
“Front”, page 60. - With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.5.5, Testing touchscreen, page 150.
MODDAT:11.04.2003 12:01:25
MODDAT:11.04.2003 12:01:25
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace touchscreen (Evita 4; EvitaXL). If replacement of the
touchscreen was not successful, replace the Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) and install the relevant software.
REASON_SERVICE:
- Touchscreen (Evita 4; EvitaXL) is faulty. - Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 69 of 171
03.01.003
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:25
03.02.001
FUNCTION_SERVICE:
The touchscreen communicates via the RS232 interface with
the Graphics Controller PCB (Evita 4; EvitaXL) and the Front
Panel PCB (Evita 2 dura).
CONSEQ_SERVICE:
Similar to error 03.01.001, but occurred in operation. After unit
power-on the screen is not detected by the Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura).
REASON_SERVICE:
Touchscreen (Evita 4; EvitaXL) is faulty. Graphics Controller
PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) is
faulty.
Mode switch detected
FIRST_SOFTWARE:
FUNCTION_SERVICE:
1. During operation as unit fault (software error on Graphics
Controller PCB) Please notify the responsible DrägerService
product support executive immediately 2. The error message
can be ignored if: The unit ventilates with no volume control.
Then choose "Configuration -> Ventilation->3032->Modes" to
alter the AutoFlow setting. Detection: Before unit fault
03.02.001 the log reports "Autoflow on" or "Autoflow off" as
appropriate
4,1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.2.1,
diagnosis page “Microprocessor” of
“Front”, page 60. - With external service mode,
see "Repair Instructions - Service no./modes, error list"
section 9.5.5, Testing touchscreen, page 150.
SOLUTION_SERVICE:
Replace touchscreen (Evita 4; EvitaXL). If replacement of the
touchscreen was not successful, replace the Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:12.12.2002 09:18:52
FAILURE:
1. Unit fault 2. No effect
CONSEQ_SERVICE:
Software error on Graphics Controller PCB
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
SOLUTION_SERVICE:
Please notify the responsible DrägerService product support
executive immediately
Copyright Dräger Medical - For internal use only!
2003-04-16Page 70 of 171
03.07.001
Evita 4 / 2dura / XLError code list for
"Reset wire" error
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:26
04.01.001
FUNCTION_SERVICE:
Error reset wire
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
There is an open circuit in the reset wire from the CPU 68332
PCB to the Graphics Controller PCB (Evita 4; EvitaXL) or
Front Panel PCB (Evita 2 dura).
Power supply unit error, 5V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
FAILURE:
Control panel or Front Panel PCB is not connected. Cable to
control panel or to Front Panel PCB is faulty. CPU 68332
PCB is faulty. Graphics Controller PCB or Front Panel PCB is
faulty. Connector between PCBs is faulty.
Evita 4/Evita 2 dura: 1) Check front panel cable, replace front
panel cable as necessary. 2) Remove housing cover so
service LEDs on CPU 68332 PCB are visible. If necessary,
plug the CPU 68332 PCB into one of the upper slots. 3)
Connect control panel ready for use. 4) Switch on unit. 5) LED
6 must light up briefly after power-on: If it does not = CPU
68332 PCB faulty If it does = CPU 68332 PCB probably OK,
replace Graphics Controller PCB (Evita 4; EvitaXL) Front
Panel PCB (Evita 2 dura) and install the relevant software.
EvitaXL: Check control panel connector and replace
connector as necessary. Check cable to control panel and
replace cable as necessary. Replace CPU 68332 PCB and
install the relevant software. If replacement of the CPU 68332
PCB was not successful, replace the Graphics Controller 8
PCB and install the relevant software. Check connector
between PCBs and replace connector as necessary.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:40
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 5V supply voltage is above 6 V
REASON_SERVICE:
Power supply unit (source) or CO2 Carrier PCB (monitor) is
faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 71 of 171
04.01.002
Evita 4 / 2dura / XLError code list for
Power supply unit error, 5V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 5V supply voltage is below 4 V.
REASON_SERVICE:
Power supply unit is faulty (source) or CO2 Carrier PCB
(source monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.003
Power supply unit error, 12V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 12V supply voltage is above 14.4 V.
REASON_SERVICE:
Power supply unit is faulty (source) or CO2 Carrier PCB
(monitor) is faulty.
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FAILURE:
Trouble-shooting: - Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 72 of 171
04.01.004
Evita 4 / 2dura / XLError code list for
Power supply unit error, 12V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 12V supply voltage is below 9.6 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.005
Power supply unit error, 15V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20 %.
CONSEQ_SERVICE:
Power supply unit error: 15V supply voltage is above 18 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 73 of 171
04.01.006
Evita 4 / 2dura / XLError code list for
Power supply unit error, 15V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 15V supply voltage is below 12 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.007
Power supply unit error, -15V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: -15V supply voltage is above -18 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 74 of 171
04.01.008
Evita 4 / 2dura / XLError code list for
Power supply unit error, -15V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: -15V supply voltage is below -12 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.009
Power supply unit error, 24V supply voltage
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20%.
CONSEQ_SERVICE:
Power supply unit error: 24V supply voltage is above 28.8 V
REASON_SERVICE:
Evita 4/Evita 2 dura: a) Power supply unit faulty (source) b)
CO2 Carrier PCB faulty (monitor) EvitaXL: - Power supply unit
error; 24V supply voltage is above 28.8 V. The operating
voltages accessible by way of an A/D converter on the CO2
Carrier PCB are compared cyclically against the defined
tolerances of plus/minus 20%.
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor of Electronic, page 66. - With
external service mode, see "Repair Instructions - Service
no./modes, error list" section 9.4.1, Testing voltages, page
133. If OK = Replace CO2 Carrier PCB If not OK = Replace
power supply unit Important: 001 Power supply unit +5V too
high (for digital components) 002 Power supply unit +5 V too
low 003 Power supply unit +12V too high (for HPSVs, fan in
power supply unit) 004 Power supply unit +12V too low 005
Power supply unit +15V too high (for analog components) 006
Power supply unit +15V too low 007 Power supply unit -15V
too high (for analog components) 008 Power supply unit -15V
too low 009 Power supply unit +24V too high (for valves and
control panel) 010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 75 of 171
04.01.010
Evita 4 / 2dura / XLError code list for
Power supply unit error, 24V supply voltage
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:41
FUNCTION_SERVICE:
The operating voltages accessible on the CO2 Carrier PCB
with an A/D converter are compared cyclically against the
defined tolerances of plus/minus 20 %.
CONSEQ_SERVICE:
Power supply unit error: 24V supply voltage is below 19.2 V
REASON_SERVICE:
Power supply unit (source) is faulty or CO2 Carrier PCB
(monitor) is faulty.
FAILURE:
Trouble-shooting: Check of power supply unit voltages
(several possibilities): - Measure voltage with multimeter, see
Repair Instructions - Electronic components, section 1, Power
supply unit, page 1 - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.3.2,
diagnosis page Microprocessor ”of
“Electronic”, page 66. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.4.1, Testing voltages, page 133. If OK = Replace
CO2 Carrier PCB If not OK = Replace power supply unit
Important: 001 Power supply unit +5V too high (for digital
components) 002 Power supply unit +5 V too low 003 Power
supply unit +12V too high (for HPSVs, fan in power supply
unit) 004 Power supply unit +12V too low 005 Power supply
unit +15V too high (for analog components) 006 Power supply
unit +15V too low 007 Power supply unit -15V too high (for
analog components) 008 Power supply unit -15V too low 009
Power supply unit +24V too high (for valves and control panel)
010 Power supply unit +24V too low
SOLUTION_SERVICE:
Check power supply unit voltages. If the voltages are
incorrect, replace the power supply unit. If replacing the
power supply unit was not successful, replace the CO2 Carrier
PCB.
REASON / SHORTREASON_CUSTOMER:
04.01.011
05.01.001
Battery voltage is too high
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
Electronics: Battery voltage is above 30 V.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Short in pneumatics fan
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D
converter with a measuring resistor. In normal operation the
fan current must fluctuate by more than 50 mV. The
minimums and maximums of this fan voltage are measured
for 60 seconds each. If the difference between the maximum
and minimum is less than 50 mV there is a fault.
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit) The measured
fan voltage is greater than or equal to 2.9 V (short)
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Replace battery. EvitaXL: - Check
power supply unit voltages, replace the power supply unit as
necessary. - Replace CO2 Carrier PCB.
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.3,
diagnosis page Microprocessor”
of“Pneumatic”, page 77. - With external
service mode, see "Repair Instructions - Service no./modes,
error list" section 9.3.5, Voltages, page 129. Fan connected?
Fan rotating? No = Replace fan Yes = Replace Pneumatic
Controller PCB
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as
necessary. If replacement of the fan was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:41
MODDAT:11.04.2003 12:01:26
REASON_SERVICE:
PCB cooling fan in pneumatic unit is faulty or the fan is not
connected.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 76 of 171
05.01.002
Evita 4 / 2dura / XLError code list for
Fan (pneumatic unit)
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:26
05.01.003
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D
converter. In normal operation this fan current must fluctuate
by more than 50 mV. The minimums and maximums of this
fan voltage are measured for 60 seconds each. If the
difference between the maximum and minimum is less than
50 mV there is a fault.
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit). The
measured fan voltage is greater than or equal to 2.9 V (short).
REASON_SERVICE:
PCB cooling fan in pneumatic unit is faulty or not connected.
Fan (pneumatic unit)
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The activation current of the fan is measured with an A/D
converter. In normal operation this fan current must fluctuate
by more than 50 mV. The minimums and maximums of this
fan voltage are measured for 60 seconds each. If the
difference between the maximum and minimum is less than
50 mV there is a fault.
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.3,
diagnosis page Microprocessor”
of“Pneumatic”, page 77. - With external
service mode, see "Repair Instructions - Service no./modes,
error list" section 9.3.5, Voltages, page 129. Fan connected?
Fan rotating? No = Replace fan. Yes = Replace Pneumatic
Controller PCB
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as
necessary. If replacement of the fan was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.3,
diagnosis page Microprocessor”
of“Pneumatic”, page 77. - With external
service mode, see "Repair Instructions - Service no./modes,
error list" section 9.3.5, Voltages, page 129. Fan connected?
Fan rotating? No = Replace fan Yes = Replace Pneumatic
Controller PCB
05.02.001
CONSEQ_SERVICE:
The measured fan voltage is 0 V (open circuit). The
measured fan voltage is greater than or equal to 2.9 V (short).
REASON_SERVICE:
PCB cooling fan in pneumatic unit is faulty or not connected,
or the Pneumatic Controller PCB is faulty.
"Reset-up wire" error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Check fan in pneumatic unit is rotating, replace fan as
necessary. If replacement of the fan was not successful,
replace the Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:26
Trouble-shooting: 1. Remove housing cover so service LEDs
on CPU 68332 PCB are visible. If necessary, plug the PCB
into one of the upper slots. 2) Switch on unit. 3) The following
LEDs must light up briefly after power-on: LED 4 (together
with LED 3) on reset-up and LED 5 on reset-down. If no =
CPU 68332 PCB faulty If yes = CPU 68332 PCB probably OK,
replace Pneumatic Controller PCB
Check connector between electronic and pneumatic units,
replace connector as necessary. If the connector was OK,
replace the CPU 68332 PCB and install the relevant
software. If replacement of the CPU 68332 PCB was not
successful, replace the Pneumatic Controller PCB and install
the relevant software.
REASON_SERVICE:
Connector between PCBs, CPU 68332 PCB or Pneumatic
Controller PCB is faulty.
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 77 of 171
05.02.002
Evita 4 / 2dura / XLError code list for
"Reset down wire" error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Trouble-shooting: 1. Remove housing cover so service LEDs
on CPU 68332 PCB are visible. If necessary, plug the PCB
into one of the upper slots. 2) Switch on unit. 3) The following
LEDs must light up briefly after power-on: LED 4 (together
with LED 3) on reset-up and LED 5 on reset-down. If no =
CPU 68332 PCB faulty If yes = CPU 68332 PCB probably OK,
replace Pneumatic Controller PCB
Check connector between electronic and pneumatic units,
replace connector as necessary. If the connector was OK,
replace the CPU 68332 PCB and install the relevant
software. If replacement of the CPU 68332 PCB was not
successful, replace the Pneumatic Controller PCB and install
the relevant software.
MODDAT:11.04.2003 12:01:26
05.02.003
REASON_SERVICE:
Connector between PCBs, CPU 68332 PCB or Pneumatic
Controller PCB is faulty.
Disable wire
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Open circuit in disable wire from CPU 68332 PCB to
Pneumatic Controller PCB. Connector between PCBs, CPU
68332 PCB or Pneumatic Controller PCB is faulty.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:26
Trouble-shooting: 1) Remove housing cover so service LEDs
on CPU 68332 PCB are visible. If necessary, plug the PCB
into one of the upper slots. 2) Switch on unit. 3) LED 7 must
flash after power-on and remain lit until the pneumatics are
activated (approx. 10 s): If no = CPU 68332 PCB faulty If yes
= CPU 68332 PCB probably OK, replace Pneumatic
Controller PCB Note: Only possible with software version 1.00
to 1.05, 2.00 and 2.10
Check connector between electronic and pneumatic units,
replace connector as necessary. If the connector was OK,
replace the CPU 68332 PCB and install the relevant
software. If replacement of the CPU 68332 PCB was not
successful, replace the Pneumatic Controller PCB and install
the relevant software.
REASON / SHORTREASON_CUSTOMER:
05.04.001
Copyright Dräger Medical - For internal use only!
O2/AIR switch-over valve cannot be switched on
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is
continuously checked against the returned signal from the
activation electronics. If the specified setting and the returned
signal do not match, a unit error message is produced.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the O2/AIR switch-over valve is
connected is not detected.)
MODDAT:11.04.2003 12:01:26
2003-04-16Page 78 of 171
05.04.002
Evita 4 / 2dura / XLError code list for
O2/AIR switch-over valve is not switched off
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:26
05.04.003
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is
continuously checked against the returned signal from the
activation electronics. If the specified setting and the returned
signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
O2/AIR switch-over valve is not switched off.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
O2/AIR switch-over valve has not switched
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is
continuously checked against the returned signal from the
activation electronics. If the specified setting and the returned
signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
O2/AIR switch-over valve has not switched
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the O2/AIR switch-over valve is
connected is not detected.)
MODDAT:11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the O2/AIR switch-over valve is
connected is not detected.)
05.04.004
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Setting of O2/AIR switch-over valve is incorrect
FIRST_SOFTWARE:
FUNCTION_SERVICE:
The specified setting of the O2/AIR switch-over valve is
continuously checked against the returned signal from the
activation electronics. If the specified setting and the returned
signal do not match, a unit error message is produced.
CONSEQ_SERVICE:
Setting of O2/AIR switch-over valve is incorrect
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software. (Note: Whether the O2/AIR switch-over valve is
connected is not detected.)
MODDAT:11.04.2003 12:01:26
Copyright Dräger Medical - For internal use only!
2003-04-16Page 79 of 171
05.04.005
Evita 4 / 2dura / XLError code list for
O2/AIR switch-over valve not switching to AIR
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:26
05.04.006
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar in the tubing
system is built up only with one connected pressure supply
(O2 or AIR). Then the O2/AIR switch-over valve "Y1.1"
switches to the absent compression gas. This bleeds the
tubing system to below 10 mbar.
CONSEQ_SERVICE:
O2/AIR switch-over valve not switching to AIR
REASON_SERVICE:
Evita 4/Evita 2 dura: The system cannot build up the test
pressure of 30 mbar in the unit check or the built-up pressure
is not relieved fast enough. EvitaXL: - The O2/AIR switchover valve is not switching to AIR.
O2/AIR switch-over valve not switching to O2
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar in the tubing
system is built up only with one connected pressure supply
(O2 or AIR). Then the O2/AIR switch-over valve Y1.1 switches
to the absent compression gas. This bleeds the tubing system
to below 10 mbar.
1LAST_SOFTWARE: 99,99
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Eliminate leaks in the system - Replace
switch-over valve Y1.1 - Replace supply pressure sensors Replace the airway pressure sensors - Calibrate PEEP valve.
EvitaXL: - Check that the test lung is rigid enough. - Check
whether an airway pressure of 30 mbar is built up. Replace
valve "Y1.1" in the gas port. - Replace expiratory airway
pressure sensor.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page “ Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
05.05.001
CONSEQ_SERVICE:
The system cannot build up the test pressure of 30 mbar in
the unit check or the built-up pressure is not relieved fast
enough.
REASON_SERVICE:
O2/Air switch-over valve not switching to O2.
Safety valve cannot be switched on
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
Safety valve cannot be switched on.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 1,04
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Eliminate leaks in the system - Replace
switch-over valve Y1.1 - Replace supply pressure sensors Replace the airway pressure sensors - Calibrate PEEP valve.
EvitaXL: - Check that the test lung is rigid enough. - Check
whether an airway pressure of 30 mbar is built up. Replace
valve "Y1.1" in the gas port. - Replace expiratory airway
pressure sensor.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on
Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety
valve was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
Copyright Dräger Medical - For internal use only!
2003-04-16Page 80 of 171
05.05.002
Evita 4 / 2dura / XLError code list for
Safety valve cannot be switched off
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 1,04
Only up to SW 1.04. Safety valve or safety valve actuator on
Pneumatic Controller PCB is faulty.
MODDAT:11.04.2003 12:01:26
05.05.003
05.05.004
CONSEQ_SERVICE:
Safety valve cannot be switched off.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Safety valve cannot be switched
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:
Safety valve cannot be switched.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Setting of safety valve is incorrect
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
1LAST_SOFTWARE: 1,04
1LAST_SOFTWARE: 1,04
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety
valve was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
MODDAT:11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on
Pneumatic Controller PCB is faulty.
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety
valve was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
MODDAT:11.04.2003 12:01:26
Only up to SW 1.04. Safety valve or safety valve actuator on
Pneumatic Controller PCB is faulty.
05.05.005
CONSEQ_SERVICE:
Setting of safety valve is incorrect
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Pressure bleed via safety valve is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
In the unit check a test pressure of 30 mbar is built up in the
patient system. Then it is bled to below 10 mbar via safety
valve "Y1.3".
CONSEQ_SERVICE:
Pressure bleed with safety valve "Y1.3" is faulty.
REASON_SERVICE:
The system cannot build up the test pressure of 30 mbar in
the unit check or the built-up pressure is not relieved fast
enough.
1LAST_SOFTWARE: 99,99
SOLUTION_SERVICE:
Replace safety valve "Y1.3". If replacement of the safety
valve was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
MODDAT:11.04.2003 12:01:26
FAILURE:
Trouble-shooting: - With diagnosis mode, see "Repair
Instructions - Service no./modes, error list" section 3.4.1,
diagnosis page ” Valves” of
“Pneumatic”, page 70. - With external service
mode, see "Repair Instructions - Service no./modes, error list"
section 9.3.1, Switching valves, page 123.
SOLUTION_SERVICE:
Replace safety valve "Y1.3" in the gas port.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 81 of 171
05.05.006
Evita 4 / 2dura / XLError code list for
Test pressure cannot be built up
FIRST_SOFTWARE:
1LAST_SOFTWARE: 1,04
MODDAT:11.04.2003 12:01:26
10.01.020
10.01.040
FUNCTION_SERVICE:
In the unit check a pressure of 30 mbar is to be built up in the
tubing system with the test specimen connected.
CONSEQ_SERVICE:
Test pressure of 30 mbar cannot be built up in unit check.
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
Software error (timeout)
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Timeout error on Pneumatic Controller PCB.
Software error (timeout)
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
FAILURE:
Cause: - Tubing system leaking - Test lung not connected
Note: As from SW 1.04 only a “ F “ is set for
this error in the unit check.
SOLUTION_SERVICE:
Evita 4/Evita 2 dura: - Tubing system is leaking, eliminate
leakage. - PEEP valve is not calibrated, calibrate PEEP
valve. EvitaXL: - The test lung is not connected. - Check the
tubing system for leakage.
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
10.02.001
FUNCTION_SERVICE:FAILURE:
The Pneumatic Controller PCB is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
Timeout error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
RAM error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 82 of 171
10.03.001
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:27
10.97.???
10.99.001
REASON_SERVICE:
Flash-EPROM error on Pneumatic Controller PCB.
ROSY SHUTDOWN
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Pneumatic controller
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
ROSY SHUTDOWN software problems of operating system.
The operating system is running on the CPU 68332 PCB, the
Pneumatic Controller PCB, the Graphics Controller PCB
(Evita 4) or Front Panel PCB (Evita 2 dura) and the
Communication PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:08.11.2002 14:55:37
FAILURE:
In case of problems please consult Lübeck
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
10.99.020
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 83 of 171
10.99.030
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:27
10.99.040
10.99.050
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.060
10.99.061
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
MODDAT:11.04.2003 12:01:27
Copyright Dräger Medical - For internal use only!
2003-04-16Page 84 of 171
10.99.062
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:27
10.99.063
10.99.064
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.065
10.99.070
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
MODDAT:11.04.2003 12:01:27
Copyright Dräger Medical - For internal use only!
2003-04-16Page 85 of 171
10.99.071
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:27
10.99.080
10.99.081
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:27
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.082
10.99.083
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
The Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
MODDAT:11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16Page 86 of 171
10.99.084
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:28
10.99.085
10.99.086
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.087
10.99.088
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
MODDAT:11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16Page 87 of 171
10.99.101
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:28
10.99.110
10.99.111
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
Power supply unit or Pneumatic Controller PCB is faulty.
Other cause: If the mixer generates abnormal noises on
power-on, an HPS valve is impermissibly placing load on the
minus 15 V.
Check power supply unit supply voltages, replace power
supply unit as necessary. If replacement of the power supply
unit was not successful, replace the Pneumatic Controller
PCB and install the relevant software.
10.99.112
REASON_SERVICE:
Faulty A/D conversion on Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Faulty A/D conversion on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
Power supply unit or Pneumatic Controller PCB is faulty.
Other cause: If the mixer generates abnormal noises on
power-on, an HPS valve is impermissibly placing load on the
minus 15 V.
Check power supply unit voltages, replace power supply unit
as necessary. If replacement of the power supply unit was
not successful, replace the Pneumatic Controller PCB and
install the relevant software.
REASON / SHORTREASON_CUSTOMER:
Copyright Dräger Medical - For internal use only!
2003-04-16Page 88 of 171
10.99.801
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:28
10.99.901
10.99.911
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.912
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
Copyright Dräger Medical - For internal use only!
2003-04-16Page 89 of 171
10.99.913
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty. Most probable cause: Pneumatic Controller
PCB.
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:28
10.99.914
10.99.915
REASON_SERVICE:
Error in CAN controller on Pneumatic Controller PCB or CPU
68332 PCB or on Graphics Controller PCB (Evita 4; EvitaXL)
or Front Panel PCB (Evita 2 dura). Most probable cause:
Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:
Pneumatic Controller PCB is faulty.
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
FAILURE:
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
10.99.916
Copyright Dräger Medical - For internal use only!
Software error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Wrong software version on Pneumatic Controller PCB. Note:
The same software version must be installed on all
downloadable PCBs!
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
Wrong software version on Pneumatic Controller PCB.
Install software on Pneumatic Controller PCB. Note: The
same software version must be installed on all downloadable
PCBs!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 90 of 171
10.99.917
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:28
10.99.918
10.99.919
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.920
10.99.921
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
The Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:28
MODDAT:11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16Page 91 of 171
10.99.922
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:29
10.99.923
10.99.924
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.925
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16Page 92 of 171
10.99.926
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics
Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura) is
faulty. Most probable cause: Pneumatic Controller PCB
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:29
10.99.927
10.99.928
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most
probable cause: Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most
probable cause: Pneumatic Controller PCB.
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:
CAN ERROR_INT2
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the Graphics Controller 8 PCB
and install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most
probable cause: Pneumatic Controller PCB
Copyright Dräger Medical - For internal use only!
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 93 of 171
10.99.929
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:29
10.99.930
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura). Most
probable cause: Pneumatic Controller PCB
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB, on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
Pneumatic Controller PCB or CPU 68332 PCB Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty. Most probable cause: Pneumatic Controller
PCB.
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON / SHORTREASON_CUSTOMER:
10.99.935
Copyright Dräger Medical - For internal use only!
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or on CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
1LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura).
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:42
2003-04-16Page 94 of 171
10.99.936
Evita 4 / 2dura / XLError code list for
Hardware error
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
Pneumatic Controller PCB or CPU 68332 PCB or Graphics
Controller PCB (Evita 4; EvitaXL) or Front Panel PCB (Evita 2
dura) is faulty.
Replace Pneumatic Controller PCB and install the relevant
software. If replacement of the Pneumatic Controller PCB
was not successful, replace the CPU 68332 PCB and install
the relevant software. If replacement of the CPU 68332 PCB
was not successful, replace the Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura) and
install the relevant software.
MODDAT:11.04.2003 12:01:42
10.99.940
10.99.941
REASON_SERVICE:
Hardware error in CAN controller on Pneumatic Controller
PCB or CPU 68332 PCB or on Graphics Controller PCB
(Evita 4; EvitaXL) or Front Panel PCB (Evita 2 dura).
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
10.99.942
Copyright Dräger Medical - For internal use only!
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
2003-04-16Page 95 of 171
10.99.950
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:29
10.99.951
10.99.952
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.953
10.99.960
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
MODDAT:11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16Page 96 of 171
10.99.970
Evita 4 / 2dura / XLError code list for
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
MODDAT:11.04.2003 12:01:29
10.99.971
10.99.972
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
1LAST_SOFTWARE: 99,99
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
10.99.973
Pneumatic Controller PCB is faulty.
FIRST_SOFTWARE:
FUNCTION_SERVICE:FAILURE:
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:
Software error on Pneumatic Controller PCB.
1LAST_SOFTWARE: 99,99
The Pneumatic Controller PCB is faulty.
Replace Pneumatic Controller PCB and install the relevant
software.
REASON / SHORTREASON_CUSTOMER:
MODDAT:11.04.2003 12:01:29
Copyright Dräger Medical - For internal use only!
2003-04-16Page 97 of 171
11.01.001
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:29
11.01.003
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
11.01.004
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!
2003-04-16Page 98 of 171
11.01.005
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:30
11.01.006
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
11.01.007
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!
2003-04-16Page 99 of 171
11.01.008
Evita 4 / 2dura / XLError code list for
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
1LAST_SOFTWARE: 99,99
MODDAT:11.04.2003 12:01:30
11.01.009
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
11.01.010
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
HPS "AIR" valve or HPSV Controller PCB is faulty
FIRST_SOFTWARE:
FUNCTION_SERVICE:
This error message occurs on Evita 4 as from software
version 2.21 (Japan as from software version 1.09) and on
Evita 2 dura with software versions later than 3.00. With older
software versions only a mixer fault is displayed and no unit
error message is stored.
CONSEQ_SERVICE:SOLUTION_SERVICE:
REASON_SERVICE:REASON / SHORTREASON_CUSTOMER:
1LAST_SOFTWARE: 99,99
FAILURE:
HPS "AIR" valve or HPSV Controller PCB for "AIR" (bottom
slot on pneumatic unit) is faulty.
Trouble-shooting and remedy: Swap the two HPSV Controller
PCBs. If the error then occurs in the HPSV O2 valve, the error
is in the HPSV Controller PCB now located in the middle slot.
If the error remains in the HPS "AIR" valve, the HPS "AIR"
valve is faulty.
MODDAT:11.04.2003 12:01:30
Copyright Dräger Medical - For internal use only!