Druck PTX7901 Data sheet

The PTX 7900 series combines modular design with the latest advances in ASIC technology and surface mounted electronics. This provides a lightweight and cost effective alternative to pressure gauges and switches in process industry applications.
At the heart of the transmitter is GE Druck’s own piezo­resistive silicon sensor. This technology is extensively qualified and proven, for example in aerospace and subsea programmes which demand the highest levels of performance and long term reliability.
A Hastelloy isolation diaphragm and electron beam welded stainless steel body, ensures compatibility with a wide range of hostile media. A 316 stainless steel electronics housing with aluminium/bronze end cap, provides a maintenance-free assembly suitable for the most arduous environments.
The PTX 7900 pressure transmitter also offers excellent RFI immunity, meeting the highest level of CE marking requirements for heavy industrial use.
GE Druck
g
PTX 7900 Series
Pressure Transmitters
Standard ranges or custom scaled
Gauge and absolute configurations
Hastelloy and stainless steel construction
User accessible zero and span adjustment
RFI protected to CE Heavy Industrial
ATEX Intrinsically Safe and Flameproof
versions
OPTIONS
(O) Standard - CE Category 1 Pressure
Accessory to Pressure Equipment directive (PED) 97/23/EC. Note: ‘Operating Pressure Range’ is equivalent to maximum working pressure (Ps) as referred to in the PED
(I) Intrinsically Safe Approval
118 0
II 1G EExia IIC T4 (Ta = 80⬚C)
to ATEX directive 94/9/EC
(D) Flameproof Approval
1180
II 2G EExd IIC T6 (Ta = 70⬚C) to ATEX directive 94/9/EC Other T class and ambient temperature available.
All options are compliant with EMC Directive 89/336/EEC
EMC Emissions: EN50081-1, EN55022 EMC Immunity: EN61000-6-2: 1999
(10V/m Heavy Industrial).
ORDERING INFORMATION
(1) Select model number: Code Model PTX 7900 Base Model - M20 female conduit entry PTX 7901 Base Model - 1/2” NPT female conduit entry PTX 7902 Base Model - Pg 13.5 female conduit entry
Code Pressure Connection
1 G1/2Female 2 G1/2Male to BS EN387-1 3 1/2NPT Female 4 1/2NPT Male
Code Approvals
0 None - Safe Area use I ATEX Intrinsically Safe
D ATEX Flameproof
PTX 7900 - 1 - D Typical Model No.
(2) State pressure range/units
Continuing development can necessitate specification changes without notice.
STANDARD SPECIFICATIONS
Pressure Measurement
Operating Pressure Ranges
0 to 100, 160, 250 mbar, 500 mbar, 1, 2, 3.5, 7, 10, 20, 35, 70 bar gauge and absolute. 0 to 140, 200, 350, 700 bar sealed gauge and absolute. Compound ranges available on request.
Note: Any pressure unit and span can be specified between 100 mbar and 700 bar F.S.
Overpressure
The operating range can be exceeded by the following without degrading performance: 12 x for 100 mbar 8 x for 160mbar range 6 x for ranges 250 and 500 mbar 4 x for ranges 1 and 2 bar 3 x for ranges 3.5 to 140 bar (200 bar max) 2 x for ranges 200 to 700 bar (1000 bar max)
Pressure Containment
Gauge ranges: 16 x for 100 mbar 12 x for 160 mbar range 8 x for ranges 250 and 500 mbar 6 x for ranges 1 and 2 bar 4 x for ranges 3.5 to 70 bar (250 bar max) Absolute/sealed gauge ranges: 250 bar for absolute ranges 100 mbar to 140 bar. 1000 bar for 200 bar to 700 bar ranges
Pressure Media
Fluids compatible with 316L stainless steel and Hastelloy C276. (NACE compatible grades).
Supply Voltage
9 to 30 V at PTX terminals. (9 to 28V for IS units). Maximum load () = 50 x (Supply Voltage - 9).
Supply Sensitivity
0.005% F.S./Volt.
Insulation Resistance
>10 M@ 500 Vd.c. (@ 20C).
Surge Protection
Withstands 2kV spike. Spike test conforms to EN61000-4-5.
Output Current
4 - 20 mA (2-wire) for zero to FS pressure.
Performance
Accuracy
0.25% F.S. Combined Non-linearity, Hysteresis and Repeatability.
Zero Offset and Span Setting
Factory set ±0.05 mA. Further nominal ±5% site adjustment via non-interacting potentiometers.
Long Term Stability
At standard reference conditions the calibration will not change by more than 0.1% F.S. per year.
Operating Temperature Range
Ambient: -40 to 100⬚C Process: -40 to 120⬚C
Temperature Effects
For ranges 500mbar and above, output will not deviate from room temp calibration by more than:­1% F.S. over - 10to 50C (0.7%F.S. typical) 2% F.S. over - 20to 80C (1.5% F.S. typical)
(< 500mbar, values increase pro-rata with span.)
Physical
Pressure Connection
G
1
/
2 female, G
1
/
2 male to BS EN387-1
(DIN 16288),
1
/2 NPT female or 1/2 NPT male.
Electrical Connection
M20
1
/
2 female conduit.
1/2 NPT or PG 13.5 female conduit (via adaptor).
Ingress Protection
Designed to meet IP67 when properly installed with conduit fitting connection.
Weight
1kg nominal.
INSTALLATION DRAWINGS - Dimensions in mm
Ø60
134
78
36
M20 x 1.5
44
19
38
1/2 NPT Male
G 1/2 Male
G 1/2 or 1/2 NPT
(Female)
PTX 7900 Series
Pressure Transmitters
Druck Limited Fir Tree Lane, Groby, Leicester, LE6 0FH, UK +44 (0)116 2317100 Fax:+44 (0)116 2317103 E-Mail: sales@druck.com www.druck.com
Agent:
05/04
GE Druck
g
Group