Det-Tronics PIRVOLLT User Manual

Instructions 95-8631
Volumetric Infrared Hydrocarbon Gas Process Monitor for Light Hydrocarbons PIRVOLLT
3.1 Rev: 1/12 95-8631
Table Of Contents
APP LICATION .............................................1
FE ATURES ...............................................1
SPECIFICATIONS ..........................................1
DESCRIPTION ............................................4
Detection Method .......................................4
Current Loop Output .....................................4
Signal Display Device ....................................4
Operating Modes .......................................4
INSTALLATION ............................................5
Device Mounting Orientation ..............................5
Termination Boxes (PIRTB) ................................5
General Wiring Requirements .............................5
Wiring Procedure .......................................6
Detector Separation (Optional) .............................7
STARTUP PROCEDURE ....................................8
TROUBLESHOOTING ......................................8
MAINTENANCE ...........................................8
Disassembly and Cleaning Procedure .......................9
REPLACEMENT PARTS .................................... 11
DEVICE REPAIR AND RETURN .............................11
ORDERING INFORMATION .................................12
APPENDIX A – CSA APPROVAL .............................13
APPENDIX B – ATEX / CE APPROVAL ........................14
APPENDIX C – IECEx APPROVAL ............................15
APPLICATION
PIRVOLLT is a point-type infrared hydrocarbon gas process monitor. It provides a continuous 4-20 milliampere (mA) output signal that is proportional to methane vapor concentrations from 0 to 100% by volume.
PIRVOLLT is designed for use in extractive vapor sampling systems where the vapor concentration is likely to be above the lower flammability limit (LFL). Typical applications for the PIRVOLLT include hydrocarbon well logging and petroleum exploration where a volumetric measurement of the target gas is required.
INSTRUCTIONS
Volumetric Infrared Hydrocarbon
Gas Process Monitor
for Light Hydrocarbons
PIRVOLLT
PIRVOLLT is furnished with a sealed aluminum or stainless steel vapor sampling enclosure. Brass compression­type tubing fittings on the enclosure (for use with 0.25 inch o.d. tubing) allow easy connection to the inlet and exhaust lines of the vapor sample extraction system.
PIRVOLLT is CSA certified for use in Class I, Division 1, Groups B, C and D; and Class I, Division 2, Groups A, B, C and D areas.
FEATURES
• Routine calibration not required.
• Continuous self-test automatically indicates a fault or
fouled optics condition.
• Performs well in the presence of high concentrations or constant background levels of hydrocarbons and in oxygen depleted atmospheres.
• There are no known poisons, e.g. silicones or hydrides, that compromise the integrity of the measurement.
SPECIFICATIONS
INPUT VOLTAGE—
24 Vdc nominal. Operating range is 18 to 32 Vdc including ripple.
POWER CONSUMPTION (Watts)—
Input Voltage: 18 Vdc 24 Vdc 32 Vdc
Nominal 3.5 4.6 6.2 Maximum 4.0 5.5 7.0
DETECTION RANGE—
0 to 100% by volume methane.
DETECTABLE GASES—
PIRVOLLT responds to most hydrocarbon gases, however, the signal processing program is only accurate for methane.
ACCURACY—
(Det-Tronics verified) ±2% v/v from 0 to 10% v/v ±3.5% v/v from 10 to 50% v/v ±6% v/v from 50 to 90% v/v ±7% v/v from 90 to 100% v/v.
3.1 ©Detector Electronics Corporation 2012 Rev:
1/12
95-8631
Table 1—Current Loop Output Levels
and Corresponding Status Indications
Current Level Status
23.0 mA Over 90% by volume
20.0 mA 90% by volume
4.0 mA Zero gas level (0% by volume)
3.9-2.4 mA Negative zero drift
2.2 mA Zero calibration in progress
2.0 mA Span calibration in progress
1.8 mA Calibration complete - remove gas
1.6 mA Calibration fault
1.0 mA Fouled optics
0.8 mA 24 Vdc line low (less than 17.5 Vdc)
0.6 mA
Calibrate input active at power-up (probable wiring fault)
0.4 mA Active channel fault
0.2 mA Reference channel fault
0.0 mA CPU system fault, warmup
900
800
700
600
500
MAXIMUM LOOP RESISTANCE (OHMS)
400
18
20 22 24
POWER SUPPLY VOLTAGE (VDC)
C1964
Figure 1—4 to 20 mA Current Loop Resistance
26 28 30 32
LOOP RESISTANCE (OHMS)
HUMIDITY (non-condensing)—
0 to 99% relative humidity (Det-Tronics verified) 5 to 95% relative humidity (CSA verified).
CURRENT OUTPUT—
4 to 20 mA current source (non-isolated). For output levels below 4.0 mA and above 20 mA, refer to Table 1.
To convert PIRVOLLT current output to % by volume gas concentration, use the following formula:
(milliamp signal – 4.0) x 6.25 = % by volume methane
Example: (4.0 – 4.0) x 6.25 = 0% (12.0 – 4.0) x 6.25 = 50% by volume (20.0 – 4.0) x 6.25 = 100% by volume.
Maximum loop resistance: 580 ohms at +24 Vdc. See Figure 1 for further information.
WIRING—
The PIRVOLLT has five 22 AWG wires, 20 inches (50 cm) long. Red = + 24 volts dc Black = – (common) White = 4 to 20 mA signal output Yellow = Calibration input Green = Chassis ground
Field Wiring: 18 AWG minimum is recommended for field wiring. Larger diameter wire may be required to maintain a minimum of 18 Vdc (including ripple) at the sensor for all operating conditions (see Figure 2). For maximum EMI/ RFI protection, shielded cable is recommended.
TEMPERATURE RANGE—
Operating: –40°F to +167°F (–40°C to +75°C). Storage: –40°F to +185°F (–40°C to +85°C).
ENCLOSURE MATERIALS—
Copper-free aluminum (clear anodized) or 316 stainless steel.
SAMPLE DRAW CUP—
Material: Aluminum or stainless steel. Fittings: Nickel-plated Brass. Tubing Size: 1/4 inch O.D.
DIMENSIONS—
See Figure 3 for dimensions of the PIRVOLLT and Figure 4 for dimensions of the PointWatch Termination Box Model PIRTB.
SHIPPING WEIGHT (Approximate)—
Aluminum: 1.9 pounds (0.9 kilogram). Stainless Steel: 5.0 pounds (2.3 kilograms).
2500
2000
1500
1000
500
0
18 20 22 24
MAXIMUM DISTANCE FROM POWER SUPPLY TO PIRVOLLT IN FEET
Figure 2—PIRVOL Wiring Requirements
POWER SUPPLY VOLTAGE (VDC)
26 28 30 32
A2432
16 AWG14 AWG12 AWG
18 AWG
3.1 95-8631
2
3.2 (78)
7
5.8 (147)
6.1 (154)
4.7
(11.9)
2.75 (70)
3.46 (8.8)
5.86
(14.9)
2.7
(6.9)
Figure 3—PIRVOL Dimensions in Inches (mm)
5.2
(13.2)
4.6
(116)
5.86
(14.9)
2.7
(6.9)
5.2
(13.2)
B2029
6.57
(16.7)
1.28 (3.3)
4.7
(11.9)
3.46
(8.8)
3.77 (9.6)
1.28 (3.3)
A230
Figure 4—PIRTB Dimensions in Inches (cm)
C2281
3.1 95-8631
3
CERTIFICATION—
®
For complete approval details for the PIRVOLLT Process Gas Monitor and the PointWatch Termination Box Model PIRTB, refer to the appropriate Appendix:
Appendix A - CSA Appendix B - ATEX/CE Appendix C - IECEx
OPERATING MODES
Warmup
When power is applied to the PIRVOLLT, it enters a Warmup mode (for approximately one minute) in which it performs diagnostic checks and allows the sensors to stabilize before beginning normal operation. The current output during this period is 0 mA. At the end of the warmup period with no faults present, the PIRVOLLT automatically enters the Normal operating mode. If a fault is present after the warmup, the current output will indicate a fault.
WARNING
Always ensure that the detector/termination box hazardous (classified) location ratings are applicable for the intended use.
DESCRIPTION
DETECTION METHOD
PIRVOLLT operates on the infrared absorption principle. A beam of modulated light is projected from an internal infrared source to a reflector, which sends it back to a pair of infrared sensors. One of the sensors is designated reference and the other active, with different optical filters in front of the two sensors to make them selective to different infrared wavelengths. The reference wavelength is unaffected by combustible gases, while the active wavelength is absorbed by combustible gases. The ratio of the active to the reference wavelength is computed within the PIRVOLLT to determine the concentration of gas present. This value is then converted into a 4 to 20 mA current output for connection to external display and control systems.
CURRENT LOOP OUTPUT
During normal operation, the PIRVOLLT has a 4 to 20 mA current output that is proportional to gas concentrations from 0 to 100% by volume methane. A current output less than 4 mA indicates a fault condition (see Table 1).
SIGNAL DISPLAY DEVICE
If a local or remote digital display for indication of detected gas concentration is required, it is recommended to utilize a scalable display device with a clearly marked unit of measurement in % by volume. Under no circumstances should a % LFL unit of measure display module be used to display the PIRVOLLT signal output.
Normal
In the normal operating mode, the 4 to 20 mA signal level corresponds to the detected gas concentration. The PIRVOLLT continuously checks for system faults or initiation of calibration, and automatically changes to the appropriate mode.
Fault
Faults detected during warmup, normal operation, or calibration are indicated by the current loop output as shown in Table 1.
Calibration
The PIRVOLLT is calibrated at the factory for detection of methane. No field calibration is required under normal circumstances.
In the event the PIRVOLLT optics are fouled and complete optics disassembly for cleaning is required, device re­zeroing may be required after re-assembly. Refer to the “Calibration” section for details.
NOTE A momentary connection of the calibration lead wire to DC negative (common) of the power supply initiates the zero and span calibration sequence. When wiring the PIRVOLLT, use care to isolate it from accidental contact with the PIRTB and/or other conductors.
3.1 95-8631
4
INSTALLATION
CORRECT
INCORRECT
DEVICE MOUNTING ORIENTATION
It is highly recommended that the PIRVOLLT be installed in the horizontal position. See Figure 5.
TERMINATION BOXES (PIRTB)
The PIRVOLLT is designed to be threaded into a termination box, which is then mounted to a solid, vibration-free wall or post.
The use of shielded cable in conduit or shielded armored cable is recommended for optimum RFI/EMI protection. In applications where the wiring cable is installed in conduit, the conduit must not be used for wiring to other electrical equipment. To assure proper operation, the resistance of the connecting wire must be within the specified limits. The maximum distance between the PIRVOLLT and power source is determined by the power supply capability and wire size. See Figure 2 to determine the proper wire size and maximum wiring distance allowed.
Det-Tronics offers the Model PIRTB termination box in two styles for use specifically with the PIRVOLLT:
• Tall Cover Model includes a window. See Figure 6.
• Short Cover Model uses solid cover. See Figure 7. This
model can also be used for detector separation.
The PIRTB terminals are UL/CSA rated for 14 to 22 AWG wire, and DIN/VDE rated for 2.5 mm2 wire.
GENERAL WIRING REQUIREMENTS
NOTE The wiring procedures in this manual are intended to ensure proper functioning of the device under normal conditions. However, because of the many variations in wiring codes and regulations, total compliance to these ordinances cannot be guaranteed. Be certain that all wiring complies with applicable regulations relating to the installation of electrical equipment in a hazardous area. If in doubt, consult the authority having jurisdiction before wiring the system.
It is important that moisture not be allowed to come in contact with the electrical connections of the system.
The use of proper piping techniques, breathers, glands, and seals are required to prevent water ingress and/or maintain the explosion-proof rating.
Figure 5—Recommended Orientation of PIRVOL
3.1 95-8631
5
Figure 6—Tall Cover PIRTB with Window
Figure 7—Short Cover PIRTB
WIRING PROCEDURE
IMPORTA N T
Do not apply power until the wiring procedure is complete and has been veried.
1. Locate the sensor where it will be accessible for maintenance. Excessive heat or vibration can result in premature failure of any electronic device and should be avoided if possible.
2. The PIRTB should be electrically connected to earth ground.
3. Figure 8 shows typical wiring for stand alone operation. Figure 9 shows typical wiring for PIRVOLLT with Det-Tronics’ PIRTB. Figure 10 shows the wiring terminals located inside the PIRTB. The PIRVOLLT wiring color code is:
Red lead = +24 volts dc Black lead = – (common) White lead = 4 to 20 mA signal output Yellow lead* = Calibration input Green lead = Chassis ground
* If the calibration wire (yellow lead) is not being
used, do not connect this wire to ground. Trim excess length and insulate wire so no shorting can occur.
+
24 VDC
POWER SUPPLY
+
PIRVOL
NOTES: CALIBRATION PUSHBUTTON, CURRENT METER AND POWER SUPPLY ARE NOT SUPPLIED.
CALIBRATE SWITCH SHOWN INSTALLED, HOWEVER, FIELD CALIBRATION IS NOT REQUIRED UNDER NORMAL CIRCUMSTANCES.
RED
BLACK
WHITE
YELLOW
GREEN
4 TO 20 mA
Figure 8—Typical PIRVOL Wiring, Stand Alone Conguration
PIRTB TERMINATION BOX
4-20 mA IN
24 Vdc
SPARE
CAL
4 – 20
– +
RET
+24
CHASSIS
CAL
4 – 20
RET
+24
GREEN YELLOW WHITE BLACK RED
Figure 9—Typical PIRVOL Wiring, with PIRTB
CALIBRATE
PIRVOL
PROCESS MONITOR
D1756
B2031
3.1 95-8631
6
CALIBRATE SWITCH
HOLD CALIBRATION MAGNET AT OUTSIDE BASE OF TERMINATION BOX AT THIS LOCATION TO ACTIVATE CALIBRATION SWITCH
REMOTE LED
CONTROL DEVICE, SUCH AS TRANSMITTER
PIRTB
TERMINATION
BOX
SUITABLE WIRING. MUST MEET ALL LOCAL CODES.
Figure 11—Detector Separation
PIRVOL
PROCESS MONITOR
D2030
C2056
Figure 10—PIRTB Terminals and Calibration Switch
DETECTOR SEPARATION (OPTIONAL)
In applications where the PIRVOLLT must be installed in a different location from a control device (such as a transmitter), a PIRTB termination box must be installed at the PIRVOLLT location to make the electrical connection. Refer to Figure 11 for a typical separation diagram.
Wiring Requirements for Detector Separation
Shielded four wire cable is recommended for connecting the PIRTB to the transmitter. Cable with a foil shield is recommended. The shield of the cable should be open at the PIRTB and connected to earth ground at the transmitter termination box. Ensure that the shield wire is clipped short and insulated with electrical tape to prevent accidental grounding at the open end.
The maximum distance between the PIRTB and the transmitter is limited by the resistance of the connecting wiring, which is a function of the gauge of the wire being used.
Mounting and Connecting Procedure for Detector Separation
The PIRTB should be mounted to a solid, vibration-free wall or post. The termination box should be electrically connected to earth ground.
1. Lubricate the sensor threads with low vapor pressure silicone grease, then install the sensor in the conduit entry of the PIRTB. It should be tight to ensure an explosion-proof installation, however, do not overtighten.
2. Connect the PIRVOLLT wires to the terminal strip in the PIRTB.
3. Connect the leadwires from the transmitter to the appropriate terminals inside the PIRTB. Do not ground the shield at the PIRTB. Ground the shield at the transmitter end only.
4. Check the connections inside the PIRTB and place the cover on the termination box.
5. Mount and wire the transmitter as described in the instruction manual provided by the manufacturer.
NOTE It is important to maintain a minimum of +18 volts dc (including ripple) at the PIRVOLLT. When determining the appropriate wire size for the installation, refer to Figure 2. Be sure to take into account the distance from the power supply to the PIRVOLLT or to the transmitter and then to the PIRVOLLT to ensure that the power requirements are met.
3.1 95-8631
7
STARTUP PROCEDURE
TROUBLESHOOTING
NOTE Caution should be exercised when applying this and any electronic device to explosive vapor extraction and measurement applications. Typical hazards in these applications include small tting vapor leaks, resulting in dangerous gas concentrations near or within the system enclosure. Any spark or arc may result in a re or explosion in these conditions. Always inspect all ttings for leakage before commissioning, and always remove power from all electronic devices before performing maintenance on the vapor extraction system tubing, ttings, and hardware.
1. Inhibit the output loads that are actuated by the system to prevent activation of these devices.
2. Check that the PIRVOLLT has been wired properly.
3. Apply power to the system and allow the PIRVOLLT to operate for a minimum of 2 hours, then check zero and verify gas response.
4. Place the system in normal operation by reactivating the output loads.
Refer to Table 2 to isolate and correct malfunctions with the PIRVOLLT.
MAINTENANCE
It is recommended to have spare IR Modules on hand (see “Spare Parts” section). Use Table 2 to isolate and correct malfunctions.
IMPORTANT MAINTENANCE NOTES
• Hydrocarbon-based grease emits hydrocarbon
vapors, which will be measured by PIRVOLLT, resulting in elevated gas concentration readings. Use only silicone-based grease (not
hydrocarbon-based grease) when lubricating threads on the PIRVOLLT and associated termination box. A suitable grease is listed under
“Spare Parts” in the Ordering Information section of this manual.
• It is recommended to keep spare IR modules
(See “Spare Parts” section) for eld replacement in the event of a malfunction.
Table 2—Troubleshooting Table
Current Level Status Corrective Action
2.4 to 3.9 mA Negative Zero Indication This fault can be caused by the presence of background gas during calibration. Ensure that background gas is not present and recalibrate the unit. If fault does not clear, perform disassembly and cleaning procedure, then recalibrate. If fault still does not clear, replace electronics assembly.
1.6 mA Calibration fault Repeat zero calibration. If the condition persists, replace electronics assembly.
1.0 mA Fouled optics Perform disassembly and cleaning procedure, then recalibrate.
0.8 mA +24 Vdc line low (less than +17.5 Vdc) Ensure that input voltage is correct and that power connections are good. If fault does not clear, replace the electronics assembly.
0.6 mA Calibrate input active at power-up Ensure that calibration line is not shorted and that the calibration switch is open. If fault does not clear, replace the unit.
0.4 mA Active channel fault Replace electronics assembly.
0.2 mA Reference channel fault Replace electronics assembly.
0.0 mA CPU system fault, warmup Ensure that power is applied and that the warmup period is complete (1 minute). If fault does not clear, replace the unit.
3.1 95-8631
8
DISASSEMBLY AND CLEANING PROCEDURE
The PIRVOLLT should be inspected periodically to ensure that its performance is not impaired by fouled optics or by clogging of the sample draw cup or hydrophobic filter. Inspection and/or periodic maintenance involves three different areas.
IMPORTANT
Remove power before disconnecting and removing the PIRVOLLT for maintenance.
Sample Draw Cup. Perform a visual inspection of the sample draw cup and associated plumbing, and clean as required. External filtration of sample stream is recommended.
Hydrophobic Filter. The flow of gas through the filter can be inhibited by an accumulation of mud or other contaminants. To inspect the hydrophobic filter, disassemble the PIRVOLLT as described below. If the filter appears to be fouled, replace it.
Optics. Cleaning of the optical surfaces is required only if an optical fault is indicated (1.0 mA current output signal from the PIRVOLLT). This procedure is most easily accomplished on a bench.
Required materials: Clean, flat work surface, foam tipped swabs (no cotton), isopropyl alcohol, screwdriver.
CAUTION
The PIRVOLLT contains semiconductor devices that are susceptible to damage by electrostatic discharge. An electrostatic charge can build up on the skin and discharge when an object is touched. Therefore, use caution when handling the device, taking care not to touch electronic components or terminals. If the electronics assembly is removed, it should be placed in an
anti-static bag or box while stored or transported.
A static safeguarded work area is highly recommended (if available) for disassembly and cleaning of the PIRVOLLT.
1. Remove the sample draw cup from the front of the PIRVOLLT by pulling it straight off the electronics assembly (no tools needed). See Figure 12.
2. Perform a visual inspection of the sample draw cup and associated plumbing, and clean as required. Also inspect the two O-rings inside the sample draw cup and clean, lubricate or replace as required.
3. Loosen the two captive screws on top of the mirror assembly (Figure 12) and remove the mirror assembly and hydrophobic filter from the electronics assembly.
4. Thoroughly douse the interior of the mirror assembly as well as the foam tipped swab with isopropyl alcohol. Use the swab to gently cleanse the surfaces of the reflecting mirrors inside the mirror assembly. After cleaning with the swab, flush out the mirror assembly using a liberal amount of isopropyl alcohol. Tip the mirror assembly with mirror openings downward to remove accumulated isopropyl alcohol and particle contaminants. Repeat the alcohol flush to remove any remaining contaminants. Allow the mirror assembly to air dry in a dust-free location.
IMPORTA N T
Do not insert any sharp object into the mirror assembly. Scratching of the mirrors will void the PIRVOLLT warranty. Do not use cotton tipped swabs or buds as they are likely to leave fiber residue.
5. Clean the two windows on the electronics assembly using the procedure described above. When the mirror assembly and electronics assembly are thoroughly dry, proceed with re-assembly.
6. Slide a new hydrophobic filter over the end of the mirror assembly, then carefully seat the mirror assembly and hydrophobic filter against the base of the electronics assembly. Be careful not to crumple or fold the hydrophobic filter.
7. Tighten the two captive screws on the top of the mirror assembly. Tighten the screws evenly. Do not over-tighten (6 inch-pounds recommended).
8. Slide the sample draw cup back onto the PIRVO LLT.
9. Check for leaks around the PIRVOLLT sample draw cup.
3.1 95-8631
9
UL APPROVED O-RINGS
SAMPLE DRAW CUP
WINDOWS
ELECTRONICS ASSEMBLY
MIRROR ASSEMBLY
Figure 12—PIRVOLLT Disassembly for Cleaning
CALIBRATION
In the event the PIRVOLLT optics are fouled and complete dis-assembly is required for cleaning, a zero calibration of the device may be required. This procedure ensures the signal output level is at 4 mA, and does not require application of span gas.
NOTE Zero drift will be indicated by a constant offset in one direction either above or below 4 mA. The presence of background gas would be indicated by a small but constantly changing output.
IMPORTANT CALIBRATION NOTES
Ensure that the detector has been operating for at least two hours before calibrating.
Do not open the explosion-proof enclosure
when power is applied to the system unless the appropriate permits have been procured.
HYDROPHOBIC FILTER
MIRRORS (2)
C2032
Calibration Procedure
To perform a zero calibration of the PIRVOLLT, follow the sequence described in Table 3.
NOTE The calibration sequence is initiated by momentarily connecting the calibration lead to the negative lead (common) of the power supply using either the Cal Magnet or an external switch.
If the Det-Tronics PIRTB with magnetic Cal Switch
is being used, this is accomplished by holding the Cal Magnet near the side of the termination box for one second. The location of the Cal Switch is shown in Figure 10. An alternate way of accomplishing this is to install a pushbutton switch between the yellow lead and the power supply common.
Factory calibration is recommended for complete device calibration. Contact Det-Tronics for additional information.
3.1 95-8631
10
Table 3—Calibration Sequence
Description Current
Normal operation/no gas present 4.0 mA Off
Initiate calibration
Zero calibration complete
Calibration fault indication
NOTE: Failure to re-apply magnet to set zero level will result in Fault status if span gas is not applied. Span gas is typically
applied at the factory only.
2.2 mA On steady Initiate the calibration sequence by holding the
2.0 mA Flashing Re-apply magnet (or activate switch) for one second.
1.6 mA Off See Troubleshooting Table.
REPLACEMENT PARTS
LED
Operator Action
If the possibility of background gases exists, purge the sensor with clean air to ensure accurate calibration.
calibration magnet against the side of the termination box or activating the external calibrate pushbutton for one second.
Span calibration is not required. Device will revert to live operation using new zero level and original span level from factory calibration.
Pack the unit properly. Always use sufficient packing material. Where applicable, use an antistatic bag as
The PIRVOLLT is not designed to be repaired in the field.
protection from electrostatic discharge. If a problem should develop, refer to the Troubleshooting section. If it is determined that the problem is caused by an electronic defect, the device must be returned to the factory for repair.
Det-Tronics reserves the right to apply a service
charge for repairing returned product damaged as
NOTE
a result of improper packaging.
DEVICE REPAIR AND RETURN
Return all equipment transportation prepaid to the factory Prior to returning devices, contact the nearest local
Detector Electronics office so that a Return Material Identification (RMI) number can be assigned. A
written statement describing the malfunction must accompany the returned device or component to assist and expedite nding the root cause of the failure.
in Minneapolis.
NOTE It is highly recommended that a complete spare be kept on hand for eld replacement to ensure continuous protection.
3.1 95-8631
11
ORDERING INFORMATION
PIRVOLLT PROCESS GAS MONITOR
Refer to the PIRVOLLT Model Matrix.
REPLACEMENT IR MODULE
Without cup, for field replacement only.
Specify:
– Aluminum or stainless steel
– 3/4 inch or M25
PIRTB TERMINATION BOXES
Short Cover Termination box (solid cover, with 5 ports)
Part Number Description
006414-016 3/4 inch entries, aluminum 006414-017 3/4 inch entries, stainless steel 006414-018 M25 entries, aluminum 006414-019 M25 entries, stainless steel
STOP PLUGS AND REDUCERS
Part Number Description
101197-001 Stop plug, 3/4 inch, aluminum 101197-004 Stop plug, 3/4 inch, stainless steel 103517-001 Stop plug, M25, aluminum, IP66 101197-003 Stop plug, M25, stainless steel, IP66 102804-001 Reducer, M25 to M20, aluminum 102804-003 Reducer, M25 to M20, stainless steel
SPARE PARTS
Part Number Description
007128-001 Aluminum Sample Draw Cup with O-Rings and Fittings
007128-003 Stainless Steel Sample Draw Cup with O-Rings and Fittings
007128-002 O-Ring Kit (UL)
006297-003 Hydrophobic Filter
006680-001 Silicone Grease
Tall Cover Termination box (with window and 5 ports)
Part Number Description
006414-020 3/4 inch entries, aluminum 006414-021 3/4 inch entries, stainless steel 006414-022 M25 entries, aluminum 006414-023 M25 entries, stainless steel
PIRVOLLT MODEL MATRIX
MODEL DESCRIPTION PIRVOL Point Infrared Volumetric Process Gas Monitor
TYPE APPLICATION
LT Light Hydrocarbons
TYPE MATERIAL
A Aluminum S Stainless Steel
TYPE THREAD TYPE
3 3/4” NPT 2 M20
TYPE SAMPLING CUP
A With Sampling Cup B Without Sampling Cup (replacement part only)
ASSISTANCE
For assistance in ordering a system to meet the needs of a specific application, contact:
Detector Electronics Corporation 6901 West 110th Street Minneapolis, Minnesota 55438 USA Operator: (952) 941-5665 or (800) 765-FIRE Customer Service: (952) 946-6491 Fax: (952) 829-8750 Web site: www.det-tronics.com E-mail: det-tronics@det-tronics.com
TYPE APPROVALS
C CSA E ATEX/CE/IECEx
3.1 95-8631
12
APPENDIX A
CSA APPROVAL
PIRVOLLT Process Gas Monitor
CLASS 2258 02 - PROCESS CONTROL EQUIPMENT - For Hazardous Locations
Class I, Div. 1, Groups B, C & D (T5); Class I, Div. 2, Groups A, B, C & D (T3C).
PointWatch Termination Box Model PIRTB
CLASS 4828 01 - SIGNAL APPLIANCES - Combustible Gas Detection Instruments - For Hazardous Locations
Class I, Div. 1, Groups B, C & D (T6). Class I, Div. 2, Groups A, B, C & D (T6).
3.1 95-8631
13
ATEX CERTIFICATION
PIRVOLLT Process Gas Monitor
0539 Ex d IIB+H2 T4-T6 Gb DEMKO 09 ATEX 148373X T6 (T T5 (T T4 (T IP66.
II 2 G
= –55°C to +50°C)
amb
= –55°C to +60°C)
amb
= –55°C to +75°C)
amb
APPENDIX B
ATEX / CE APPROVAL
EN Standards:
Read and understand instruction manual before operating.
Special Conditions for Safe Use (‘X’):
The Process Gas Monitor, Model PIRVOL, has a threaded wire feed through incorporating flying leads. This feed through is to be screwed into the cable entry opening of a suitable certified (Ex ‘d’ or Ex ‘e’) enclosure in which the flying leads are to be terminated.
In order to maintain the Ingress Protection rating IP66, the back end of the Process Gas Monitor, Model PIRVOL, with the flying leads is to be screwed into a suitable certified flameproof enclosure with an Ingress Protection rating of at least IP66.
EN 50270: 2006 EN 60079-0: 2009 EN 60079-1: 2007
EN 60529: 1991+A1: 2000.
CE MARK
The PIRVOLLT Process Gas Monitor was tested and found to be compliant with EN50270 when wired in conduit or with shielded cable. All screen drains shall be terminated to the chassis.
PointWatch Termination Box Model PIRTB
0539
Ex d IIC T5-T6 Gb EN 60079-29-1 DEMKO 02 ATEX 131326 T6 (Tamb = –55°C to +60°C) T5 (Tamb = –55°C to +75°C) IP66.
II 2 G
EN Standards: EN 60079-0: 2009 EN 60079-1: 2007 EN 60529: 1991+A1: 2000
EN 60079-29-1: 2007.
All cable entry devices and blanking elements shall be certified in type of explosion protection flameproof enclosure ‘d’, suitable for the conditions of use and correctly installed. Unused apertures shall be closed with suitable certified blanking elements.
For ambient temperatures below –10°C and above +60°C, use field wiring suitable for both minimum and maximum ambient temperatures.
3.1 95-8631
14
PIRVOLLT Process Gas Monitor
IECEx ULD 11.0007X Ex d IIB+H2 T4-T6 Gb T6 (Tamb = –55°C to +50°C) T5 (Tamb = –55°C to +60°C) T4 (Tamb = –55°C to +75°C) IP66.
APPENDIX C
IECEx APPROVAL
IEC Standards:
CONDITIONS OF CERTIFICATION:
The Process Gas Monitor, Model PIRVOL, has a threaded wire feed through incorporating flying leads. This feed through is to be screwed into the cable entry opening of a suitable certified (Ex ‘d’ or Ex ‘e’) enclosure in which the flying leads are to be terminated.
In order to maintain the Ingress Protection rating IP66, the back end of the Process Gas Monitor, Model PIRVOL, with the flying leads is to be screwed into a suitable certified flameproof enclosure with an Ingress Protection rating of at least IP66.
PointWatch Termination Box Model PIRTB
IECEx ULD 10.0002 Ex d IIC T5-T6 Gb T6 (Tamb = –55°C to +60°C) T5 (Tamb = –55°C to +75°C) IP66.
IEC Standards: IEC 60079-0: 2007 IEC 60079-1: 2007
IEC 60079-0: 2007 IEC 60079-1: 2007 IEC 60529, 2.1.ed.+Corr. 1:2003+2:2007.
IEC 60529, 2.1.ed.+Corr. 1:2003+2:2007.
All cable entry devices and blanking elements shall be certified in type of explosion protection flameproof enclosure ‘d’, suitable for the conditions of use and correctly installed. Unused apertures shall be closed with suitable certified blanking elements.
For ambient temperatures below –10°C and above +60°C, use field wiring suitable for both minimum and maximum ambient temperatures.
3.1 95-8631
15
95-8631
X3301 Multispectrum
IR Flame Detector
PointWatch Eclipse®
IR Combustible Gas Detector
®
Universal Display
FlexVu
w/ GT3000 Toxic Gas Detector
Eagle Quantum Premier
Safety System
Detector Electronics Corporation
6901 West 110th Street
Minneapolis, MN 55438 USA
T: 952.941.5665 or 800.765.3473
F: 952.829.8750
®
W: http://www.det-tronics.com
E: det-tronics@det-tronics.com
Det-Tronics and the DET-TRONICS logo are registered trademarks or trademarks of Detector Electronics Corporation in the United States,
other countries, or both. Other company, product, or service names may be trademarks or service marks of others.
© Copyright Detector Electronics Corporation 2012. All rights reserved.
Loading...