Detcon DM User Manual

Detcon Model DM Transmitter Test Fixture
Operators Instruction Manual
October 22, 2007 * Document 3391 * Revision 1.0
DETCON, Inc.
3200 Research Forest Dr., Building A-1
The Woodlands, Texas 77387
Ph. 713-559-9200 / Fax 281-298-2868
www.detcon.com
Table of Contents
1.0 Description
2.0 Features
3.0 Specifications
4.0 Operation
5.0 Test Fixture Calibration
6.0 Warranty
Page 2 of 8
1.0 Description
The Detcon Model DM Transmitter Test Fixture is designed to test the operation of Detcon Model DM Series H2S Transmitters. Models covered are series DM-5xx, DM­550x, and DM-6xx.
NOTE: Model series DM-4xx loop powered transmitters are not compatible with this test fixture.
2.0 Features
Reference figure 1 below for feature locations.
1 - Internal power supply and power switch 2 - Plug-in connector for DM-5xx transmitters 3 - mA output digital LED meter 4 - Plug-in connector for DM-550 and DM-6xx transmitters 5 - RS-485 output (DM-550 and DM-6xx) 6 - Fault and alarm relay output LEDs (DM-550 and DM-6xx) 7 - % Range selector switch and zero fault switch
Page 3 of 8
3.0 Specifications
Compatible withthe Following Models Model Series DM-524x, DM-550x, DM624x Input Power 100-240VAC, 50 / 60Hz Power Consumption < 17 Watts Digital Current Meter Accuracy: ± 0.3% FS Warranty Two year Outputs Serial RS-485 Modbus™
4.0 Operation
Before using the test fixture, the technician should first be familiar with the operation of Detcon DM series toxic gas sensors. Review the applicable sensor instruction manual(s) for detailed operating instructions.
The test fixture is designed to verify performance of the Detcon DM series transmitters only. The test fixture circuitry mimics the sensor portion of the complete sensor assembly.
NOTE 1: The test fixture is designed to be used with only one transmitter at a time. Plugging more than one transmitter in the test fixture at the same time (P1 and P2) will cause erroneous readings.
NOTE 2: Reference Figure 1 for location of test fixture components. Before starting verify that SW1 is in the “0” position.
NOTE 3: Model series DM-4xx loop powered transmitters are not compatible with this test fixture.
4.1 DM-5xx Transmitters
Require tools:
Detcon PN 327-000000-000 MicroSafe™ Programming Magnet
NOTE: These instructions cover testing of the main features of the DM-5xx transmitter. See the applicable DM-5xx instruction manual for a complete list of program features and detailed operating procedures.
Page 4 of 8
1 - Turn the power switch on. The mA LED meter should illuminate.
2 - Plug the DM-5xx transmitter into P1.
3 - Perform a zero cal.
a) Enter the calibration menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “1-ZERO 2-SPAN”, then withdraw the magnet. Note that the “CAL” LED is on. Also note that the CAL LED is illuminated and that the mA Output LED meter on the test fixture indicates
2.0 mA.
b) Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display reads: “ZERO 0%”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is complete the display will read “ZERO COMPLETE” for 5 seconds and then return to the normal operations menu, “0 % LEL”. Note that the mA Output LED meter on the test fixture indicates 4.0 mA.
4 - Perform a span cal.
NOTE: DM series transmitters are universal and are programmed to operate at a specific range as well as calibration level (see the DM-5xx instruction manual for further information). To do a proper span cal, you must know the range and calibration level of your transmitter. In most instances the calibration level is set to half of the range. In this case, when performing a span cal, you will set SW1 to 50% of scale regardless of what ppm range is requested by the transmitter during the calibration procedure.
a) Enter the calibration menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “1-ZERO 2-EXIT”, then withdraw the magnet.
b) From the calibration menu “1-ZERO 2-EXIT” proceed into the span adjust function by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “APPLY ## PPM” then withdraw the programming magnet. Note that the CAL LED is illuminated and that the mA Output LED meter on the test fixture indicates 2.0 mA.
c) Turn SW1 to the “50%” position.The display will change to “AUTOSPAN ## PPM”. When the sensor signal is stable it will auto span to the requested concentration and the display will change to “SPAN COMPLETE” for two seconds and then “REMOVE GAS”. Turn SW1 back to the “0” position. The display will return to the normal operation menu, “0 PPM ###”.
5 - Check different ranges. Turn SW1 to each position (0, 10, 25, 50, 100) and verify a matching transmitter display reading (±10%). These position settings are the % of the full
Page 5 of 8
scale range of the transmitter so calculate the proper reading accordingly. For example, If the range of the transmitter is 0-100 ppm, the reading will equal the number indicated by the SW1 position. If the range of the transmitter is 0-50 PPM, the reading will equal half of the number indicated by the SW1 position, and so on. The 100 position will cause an over-range condition resulting in the display flashing “### PPM ###”. Also verify an LED mA output reading of 4.0 mA (0), 5.6mA (10), 8.0mA (25), 12.0mA (50), and
20.0mA (100). Return SW1 to the “0” position.
6 - Check zero fault. Turn SW1 to the zero fault position and note the “FLT” LED on the transmitter illuminates and the display reads “Zero Fault”. Also note that the mA output LED display indicates a reading of 000. Return SW2 to the NORMAL position.
4.2 DM-550 and DM-6xx Transmitters
Require tools:
Detcon PN 327-000000-000 MicroSafe™ Programming Magnet
NOTE: These instructions cover testing of the main features of the DM-550 and DM-6xx transmitters. See the applicable DM-550 and DM-6xx instruction manual for a complete list of program features and detailed operating procedures.
1 - Turn the power switch on. Both digital LED meters should illuminate.
2 - Plug the DM-550/DM-6xx transmitter into P2.
3 - Perform a zero cal.
a) Enter the calibration menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “1-ZERO 2-SPAN”, then withdraw the magnet. Note that the “CAL” LED is on. Also note that the CAL LED is illuminated and that the mA Output LED meter on the test fixture indicates
2.0 mA.
b) Next, enter the zero menu by holding the magnet stationary over “PGM 1” for 3 seconds until the display reads: “ZERO 0%”, then withdraw the magnet. The sensor has now entered the auto zero mode. When it is complete the display will read “ZERO COMPLETE” for 5 seconds and then return to the normal operations menu, “0 % LEL”. Note that the mA Output LED meter on the test fixture indicates 4.0 mA.
4 - Perform a span cal.
NOTE: DM series transmitters are universal and are programmed to operate at a specific range as well as calibration level (see the DM-550/DM-6xx instruction manual for further information). To do a proper span cal, you must know the range and calibration level of your transmitter. In most instances the calibration level is set to half of the range. In this
Page 6 of 8
case, when performing a span cal, you will set SW1 to 50% of scale regardless of what ppm range is requested by the transmitter during the calibration procedure.
a) Enter the calibration menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “1-ZERO 2-EXIT”, then withdraw the magnet.
b) From the calibration menu “1-ZERO 2-EXIT” proceed into the span adjust function by holding the programming magnet stationary over “PGM 1” for 3 seconds until the display reads “APPLY ## PPM” then withdraw the programming magnet. Note that the CAL LED is illuminated and that the mA Output LED meter on the test fixture indicates 2.0 mA.
c) Turn SW1 to the “50%” position.The display will change to “AUTOSPAN ## PPM”. When the sensor signal is stable it will auto span to the requested concentration and the display will change to “SPAN COMPLETE” for two seconds and then “REMOVE GAS”. Turn SW1 back to the “0” position. The display will return to the normal operation menu, “0 PPM ###”.
5 - Check different ranges. Turn SW1 to each position (0, 10, 25, 50, 100) and verify a matching transmitter display reading (±10%). These position settings are the % of the full scale range of the transmitter so calculate the proper reading accordingly. For example, If the range of the transmitter is 0-100 ppm, the reading will equal the number indicated by the SW1 position. If the range of the transmitter is 0-50 PPM, the reading will equal half of the number indicated by the SW1 position, and so on. The 100 position will cause an over-range condition resulting in the display flashing “### PPM ###”. Also verify an LED mA output reading of 4.0 mA (0), 5.6mA (10), 8.0mA (25), 12.0mA (50), and
20.0mA (100). Return SW1 to the “0” position.
6 - Check zero fault. Turn SW1 to the zero fault position and note the “FLT” LED on the transmitter illuminates and the display reads “Zero Fault”. Also note that the test fixture fault LED illuminates and that the mA output LED display indicates a reading of 000. Return SW2 to the NORMAL position.
7 - Check alarm relay outputs. Turn SW1 clockwise until both alarm LEDs on the transmitter illuminate. Verify that both alarm LEDs on the test fixture also illuminate. Return SW1 to the “0” position. If alarms on the transmitter are programmed for latching, swipe the programming magnet over PGM1 or PGM2 to reset.
8 - RS-485 Out (P3). The DM-550x and DM-624x operate as Modbus slave devices. RS­485 operation is checked by connecting to an RS-485 Modbus Master device. Reference the applicable instruction manual for further information.
Page 7 of 8
5.0 Test Fixture Calibration
The DM Transmitter Test Fixture is calibrated at the Detcon factory. SW1 range adjustment potentiometers are factory set and should not be adjusted by the user. If calibration is required, return the test fixture to Detcon for calibration.
6.0 Warranty
Detcon, Inc., as manufacturer, warrants each DM Transmitter Test Fixture for a two year period under the conditions described as follows: The warranty period begins on the date of shipment to the original purchaser and ends two years thereafter. Should the test fixture fail to perform in accordance with published specifications within the warranty period, return it to Detcon, Inc., 3200 A-1 Research Forest Dr., The Woodlands, Texas 77381, for necessary repairs or replacement.
Page 8 of 8
Loading...