Bio-Rad Plasmid Mini Purification Module User Manual

Bio-Rad Laboratories, Inc.
2000 Alfred Nobel Dr. Hercules, CA 94547 USA (510) 741-1000
1-800-424-6723
4110111 Rev C
Aurum
Plasmid Mini Kit
Instruction Manual
For technical service, call your local Bio-Rad office, or in the US, call 1-800-424-6723
Table of Contents
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Section 2 Kit Components . . . . . . . . . . . . . . . . . . . . . . . . .1
Section 3 Storage Conditions . . . . . . . . . . . . . . . . . . . . . . .2
Section 4 Necessary Supplies . . . . . . . . . . . . . . . . . . . . . .2
Section 5 Guidelines for Using the
Aurum Plasmid Mini Kit . . . . . . . . . . . . . . . . . . . . .3
Section 6 Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Section 7 Troubleshooting Guide . . . . . . . . . . . . . . . . . . .11
Section 8 Ordering Information . . . . . . . . . . . . . . . . . . . .13
Section 1 Introduction
The new Aurum plasmid mini kit is optimized for the purification of up to 20 µg of plasmid DNA, rapidly and inexpensively, without the use of toxic reagents or alcohol precipitations. The use of membranes to bind and purify plasmid DNA, in combination with an optimized column design, minimizes handling and allows plasmid purification to be carried out in either a vacuum or spin format. In either format, the final elution is carried out in a microcentrifuge. Plasmid DNA purified with the Aurum plasmid mini kit is eluted into a small volume of aqueous buffer and is free of salts, bacterial chromosomal DNA, and RNA. The exceptional purity of the plasmid produced by this system makes it ideal for use in automated fluorescent sequencing and in any other molecular biology application.
Section 2 Kit Components
The Aurum plasmid mini kit contains the following components:
resuspension solution 25 ml
lysis solution 25 ml
neutralization solution 40 ml
wash solution 25 ml (5x concentrate)
elution solution 16 ml
plasmid mini columns 100
2 ml capless wash tubes 100
The solutions are specifically formulated for the Aurum plasmid mini kit. They are NOT interchangeable with solutions used in other kits or protocols.
1
Section 3 Storage Conditions
Solutions and columns should be stored at room temperature. If precipitation is observed in any solution, warm solution to 37ºC to redissolve, and allow to return to room temperature before use. Do not expose any of the solutions to temperatures above 37ºC. If the kit is used infrequently, storage of the resuspension solution at 4ºC is recommended to preserve the RNase.
Section 4 Necessary Supplies
Equipment and supplies to be provided by the customer:
1.5–2.0 ml capped microcentrifuge tubes 200
95–100% ethanol, or reagent-grade (denatured) ethanol, 100 ml
Microcentrifuge (12,000 x g)
Additional equipment required for vacuum format:
Bio-Rad Aurum vacuum manifold with vacuum regulator and column adaptor plate (Cat. #732-6470), or other vacuum manifold with luer fittings (Figure 1).
Note: Please read Section 7, Instrument Setup and Use for the Column Adaptor Plate in the Aurum vacuum manifold instruction
manual for proper vacuum setup conditions.
Vacuum source (capability of -20 to -23" Hg required)