TROUBLESHOOTING
The following is a partial listing of error messages. For a complete listing, refer to the Instruction Manual.
E 1 No Load Detected Check all electrical connections to the electrophoresis
• The electrophoresis cell(s) are cell and whether the cell contains the appropriate buffer
not plugged in. volume. Then,
• The cell was unplugged during a run • Press the key to resume the run or,
• The current load dropped below • To clear the error code, press any key (other than ).
the accepted rating of 5 mA.
E 2 Short Circuit Check for and correct any short circuit or excessive load
The current load exceeded problem. Excessive load due to a high buffer concentration
2.5 A. will require the buffer be remade. Then,
• Press the key to resume the run or,
• To clear the error code, press any key (other than ).
E 5 Power Failure During a Timed Run To clear the error code, press any key (other than ).
Power Failure detect (PFd) was The key is disabled until the error code is cleared.
activated; the run was completed
after the power failure. Reminder:
Time is reset to zero after each timed run. Enter a time
Caution: value prior to each timed run. PFd is disabled after each
If a timed run with PFd activation is run. If PFd is required, activate it prior to starting the run.
terminated by turning the main
power switch off, the power supply
will resume operation under the
previous run parameters when the main
power switch is turned back on.
E 6 Power Failure During a Timed Run To clear the error code, press any key (other than ).
Power Failure detect (PFd) was The key is disabled until the error code is cleared.
not
activated; run was terminated
either due to power failure or Reminder:
because the previous run was Time is reset to zero after each timed run. Enter a time
stopped by turning off the power value prior to each timed run.
supply or unplugging the unit. If PFd is required, activate it prior to starting the run.
E 7 Power Failure During an Untimed Run To clear the error code, press any key (other than ).
Run was terminated either due to The key is disabled until the error code is cleared
power failure, or because the from the display.
previous run was stopped by
turning off the power supply or Reminder:
unplugging the unit. PFd can only be activated prior to a TIMED run.
E 9 Change in Load Resistance Check and correct any potential resistance problems. Then,
• Electrophoresis cells were added • Press the key to resume the run or,
or removed during a run • To clear the error code, press any key (other than ).
• Buffer leaking in a connected cell
• Loose connection in a
connected cell.
E10 Unacceptable Value(s) Entered • To clear the error code, press any key (other than )
• No value entered and enter acceptable values.
• Voltage below 5 V • When acceptable values are entered, press the
• Current below 0.01 A. key to resume the run.
E11 Ground Leak Detected • Check and correct the cell for an improper connection to
earth ground.
• Restart the power supply by turning the power switch off
and then on.