Bently Nevada 3500/64M Data sheet

3500/64M Dynamic Pressure Monitor
Datasheet
Bently Nevada Machinery Condition Monitoring

Description

The 3500/64M Dynamic Pressure Monitor is a single slot, four­channel monitor that accepts input from high temperature pressure transducers and uses this input to drive alarms. The monitor’s one measured variable per channel is bandpass dynamic pressure.
You can use the 3500 Rack Configuration Software to configure the bandpass corner frequencies along with an additional notch filter. The monitor provides a recorder output for control system applications.
The primary purpose of the 3500/64M Dynamic Pressure Monitor is to provide the following:
l Machinery protection by continuously comparing
monitored parameters against configured alarm setpoints to drive alarms
l Essential machine information for operations and
maintenance personnel
Each channel, depending on configuration, conditions its input signal to generate various parameters called measured variables. You can configure alert and danger setpoints for each active measured variable.
Document: 141536
Rev. L
3500/64M Dynamic Pressure Monitor Datasheet
Update rate approximately 100 ms or less

Specifications

Inputs
Signal
Input Impedance
3 wire transducer 10 kΩ
2 wire 1.5 MΩ to 3.5 MΩ typical
Sensitivity
Dynamic pressure 100 mV/psi (1.45 mV/mBar)
Outputs
Front Panel LEDs
OK LED
TX/RX LED
Bypass LED
Buffered Transducer
Outputs
Output Impedance
Transducer Supplies
3 wire -24 Vdc
2 wire 3.3 mA current source @ 22 Vdc (nominal)
Recorder outputs
Voltage Compliance (current output)
Resolution
Indicates when the 3500/64M Dynamic Pressure Monitor is operating properly.
Indicates when the 3500/64M is communicating with other modules in the 3500 rack.
Indicates when the 3500/64M is in Bypass Mode.
The front of each monitor has one coaxial connector for each channel.
Each connector is short-circuit protected.
The Cascade Mode does not cascade the buffered transducer outputs.
550 Ω
+4 to +20 mA Values are proportional to monitor full-scale.
The monitor provides individual recorder values for each channel.
Monitor operation is unaffected by short circuits on recorder outputs.
0 to +12 Vdc range across load
Load resistance is 0 to 600 Ω.
0.3662 µA per bit ±0.25% error at room temperature
-0.66 to +0.70% error over temperature range
Accepts from 1 to 4 pressure transducer signals
Signal Conditioning
Dynamic Pressure - Direct Filter
5 Hz to 4 KHz
Low mode
High mode
If no LP filter is chosen, the range extends to approximately 5.285 KHz
10 Hz to 14.75 KHz Fixed low pass
Low and high filtering modes are options for a channel pair. Channels 1 and 2 form a pair, and channels 3 and 4 are the other pair. You may select different band pass options on each channel of a channel pair. However, the channels within the pair must operate in the same filtering mode.
You can set up the signal processing so that the monitor feeds only the channel 1 input to all four channels. This feature is called Cascade Mode and is denoted as 1 >ALL in the 3500 Rack Configuration Software.
In Cascade Mode, you can select filter mode options for a channel pair only. One transducer provides input to four channels for different filtering requirements. As a result, you can configure four separate bandpass filter options and four separate full-scale ranges with one transducer input.
The two modes of filtering provide different qualities of filtering.
Low Mode
Filter Quality
High pass
Low pass (LP)
Fixed low pass
LP = none
10-pole (200dB per decade, 60 dB per octave)
10-pole (200dB per decade, 60 dB per octave)
-78 dB minimum attenuation in the stop band.
2/9 141536 Rev. L
3500/64M Dynamic Pressure Monitor Datasheet
High Mode
Filter Quality
6-pole (120 dB per decade, 36 dB per High pass Low pass
Line rejection (notch) filter
Filter quality response
octave)
-65 dB minimum attenuation in the
stop band
The line rejection filter has two
settings, 50 or 60 Hz. Filter response
and center frequency selections are
valid for both settings.
-0.175 dB (98%) of Full Scale at
Center Frequency of +2 Hz and
above
-0.175 dB (98%) of Full Scale at
Center Frequency of –2 Hz and below
-35 dB (1.8%) of Full Scale from –0.5
Hz of Center Frequency to +0.5 Hz of
Center Frequency
Measured Variables
Dynamic pressure direct
The primary value for each channel
Physical

Alarms

Use the 3500 Rack Configuration Software to set alert and danger levels for the direct values measured by the monitor.
Alarm Setpoints
Accuracy of alarm setpoints
Alarm Time Delays
You can program alarm delays using the 3500 Rack Configuration Software from one to 60 seconds in one second intervals.
Alert
Danger
Alarms are adjustable from 0 to 100% of full-scale for each measured value. However, when the full-scale range exceeds the range of the transducer, the range of the transducer will limit the setpoint.
Within 0.13% of the desired value
From one to 60 seconds in one second intervals
0.1 seconds or from one to 60 seconds in 0.1 second intervals
Monitor Module (Main Board)
Dimensions (Height x Width x Depth)
Weight 0.82 kg (1.8 lb)
I/O Modules (non-barrier)
Dimensions (Height x Width x Depth)
Weight 0.20 kg (0.44 lb)
I/O Modules (with barrier)
Dimensions (Height x Width x Depth)
Weight 0.46 kg (1.01 lb)
241.3 mm x 24.4 mm x 241.8 mm (9.50 in x 0.96 in x 9.52 in)
241.3 mm x 24.4 mm x 99.1 mm (9.50 in x 0.96 in x 3.90 in)
241.3 mm x 24.4 mm x 163.1 mm (9.50 in x 0.96 in x 6.42 in)
3/9 141536 Rev. L
Loading...
+ 6 hidden pages