The 3500/60 & 61 modules provide six channels of
temperature monitoring and accept both Resistance
Temperature Detector (RTD) and Thermocouple (TC)
temperature inputs. The modules condition these inputs and
compare them against user-programmable alarm setpoints.
The 3500/60 and 3500/61 provide identical functionality
except that the 3500/61 provides recorder outputs for each of
its six channels while the 3500/60 does not.
The user programs the modules to perform either RTD or TC
temperature measurements using the 3500 Rack
Configuration Software. Different I/O modules are available in
RTD/TC non-isolated or TC isolated versions. The user can
configure the RTD/TC non-isolated version to accept either TC
or RTD, or a mixture of TC and RTD inputs. The TC isolated
version provides 250 Vdc of channel-to-channel isolation to
protect against external interference.
When used in a Triple Modular Redundant (TMR)
configuration, temperature monitors must be installed
adjacent to each other in groups of three. When used in this
configuration, the system employs two types of voting to
ensure accurate operation and to avoid single-point failures.
141540 Rev. L
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Specifications
Inputs
SignalAccepts from 1 to 6 RTD or TC
transducer signals.
Input
Impedance
Power Consumption
3500/60Nominal consumption of 7
3500/61 Nominal consumption of 9
Tranducers
TCs
Type E -100 °C to +1000 °C, (-148 °F to
When using any of the isolated
thermocouple I/O modules (-03 or -04
I/O ordering option) with Type E
thermocouples, the available full-scale
range will be reduced if the 3500
system is operated in an ambient
temperature above +35C. The reduced
range will be -60C to +1000C (-76F to
+1832F). Rack configuration software
will allow the user to configure a
channel down to -100C but the system
will not function properly in this
scenario and therefore should not be
configured to operate with these
settings.
0.00385 alphas are the
worldwide industrial
standard and are
recommended for all
applications
System Isolation: 500Vdc
Channel to Channel Isolation:
250 Vdc
.Type T-160 °C to +400 °C, (-256 °F to
+752 °F).
RTDs
Isolation is only required for
fault scenarios and these
voltages will not be present on
2/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
the I/O module inputs when a
wiring or field fault occurs.
Outputs
Front Panel LEDs
OK LEDIndicates when the
Temperature Monitor is
operating properly.
TX/RX LEDIndicates then the Temperature
Monitor is communicating with
other modules in the 3500 rack.
Bypass LEDIndicates when the
Temperature Monitor is in
Bypass Mode.
RTD Current Source Value
925 ±15 µA @ 25° C per transducer (single
supply for the 4-wire RTD and two supplies for
the 3-wire).
Recorder
+4 to +20 mA. Values are proportional to
monitor full-scale. Individual recorder values
are provided for each channel. Monitor
operation is unaffected by short circuits on
recorder outputs.
RTDs and TCs (except for 10Ω Copper RTDs)
Resolution1 °C or 1 °F
Accuracy
Internal
Termination
NonIsolated
Bulkhead
Rack
Standard
Rack
±3 °C at 25 °C
(±5.4 °F at 77 °F).
±3 °C at 25 °C
(±5.4 °F at 77 °F).
External Termination Non-Isolated
Bulkhead
Rack
Standard
Rack
±3 °C at 25 °C
(±5.4 °F at 77 °F).
±1 °C at 25 °C
(±1.8 °F at 77 °F).
Internal Termination Isolated
Bulkhead
Rack
±2 °C at 25 °C
(±3.6 °F at 77 °F).
Voltage Compliance (current output)
0 to +12 Vdc range across load. Load
resistance is 0 to 600 Ω.
Resolution
0.3662 µA per bit ±0.15% error at room
temperature ±0.4% error over temperature
range.
Signal Conditioning
Specified at +25 °C (+77 °F) unless
otherwise noted.
Full-scale range for each
channel is set in the field via
3500 Configuration Software. No
calibration is required.
±3 °C at 25 °C
±5.4 °F at 77 °F).
Standard
Rack
±3 °C at 25 °C
±5.4 °F at 77 °F).
External Termination Isolated
Bulkhead
Rack
Standard
Rack
±1 °C at 25 °C
(±1.8 °F at 77 °F).
°C at 25 °C
(±1.8 °F at 77 °F)
10 Ω Copper RTDs
Resolution1°C or 1 °F
Accuracy±3 °C at 25 °C
3/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
(±5.4 °F at 77 °F).
Cold Junction Compensation Sensor (used for TC
measurements)
Accuracy
±1° C at 25 °C
(±1.8 °F at 77 °F).
Alarms
Alarm
Setpoints
Alarm Time Delays
The user can program alarm delays using
software as follows:
AlertFrom 1 to 60 seconds in 1 second
The user can set Alert and
Danger setpoints for the value
measured by the monitor using
software configuration. Alarms
are adjustable from 0 to 100% of
full-scale for each measured
value. The exception is when the
full-scale range exceeds the
range of the sensor. In this case,
the range of the sensor will limit
the setpoint. Accuracy of alarms
are to within 0.13% of the desired
value. The Temperature
Monitors have both under and
over alarm setpoints.
intervals.
DangerFrom 1 to 60 seconds in 0.5
second intervals
Measured Values
Measured values are temperature
measurements used to monitor the machine.
The Temperature Monitors return
temperature measured values.
4/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Environmental Limits
Operating
Temprature
Storage
Temperature
Physical
Monitor Module
Dimensions
(Height x Width x
Depth)
Weight
I/O Modules
-30°C to +65°C (-22°F to +150
°F) when used with
Internal/External Termination
I/O Modules
0°C to +65 °C (32 °F to +150 °F)
when used with Internal Barrier
I/O Modules (Internal
Termination)
-40°C to +85°C (-40°F to
+185°F)
241.3 mm x 24.4 mm x
241.8 mm
(9.50 in x 0.96 in x 9.52 in)
0.91 kg (2.0 lb)
Compliance and
Certifications (Approvals
pending)
FCC
This device complies with part 15 of the
FCC Rules. Operation is subject to the
following two conditions:
l This device may not cause harmful
interference.
l This device must accept any
interference received, including
interference that may cause
undesired operation.
EMC
European Community Directive:
EMC Directive 2014/30/EU
Standards:
Dimensions
(Height x Width x
Depth)
Weight0.45 kg (1.0 lb)
I/O Modules (internal barrier)
Dimensions
(Height x Width x
Depth)
Weight0.46 kg (1.01 lb)
241.3 mm x 24.4 mm x 99.1
mm
(9.50 in x 0.96 in x 3.90 in)
241.3 mm x 24.4 mm x 163.1
mm
(9.50 in x 0.96 in x 6.42 in)
Rack Space Requirements
Monitor
Module
I/O Modules1 full-height rear slot
1 full-height front slot
EN 61000-6-2; Immunity for
Industrial Environments
EN 61000-6-4; Emissions for
Industrial Environments
Electrical Safety
European Community Directive:
LV Directive 2014/35/EU
Standards:
EN 61010-1
RoHS
European Community Directive:
RoHS Directive 2011/65/EU
Cyber Security
Designed to meet IEC 62443
5/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Maritime
DNV GL rules for classification – Ships,
offshore units, and high speed and light
craft
ABS Rules for Condition of Classification,
Part 1
l Steel Vessels Rules
l Offshore Units and Structures
Functional Safety
SIL 2
Hazardous Area Approvals
For the detailed listing of country and
product specific approvals, refer to the
Approvals Quick Reference Guide
(108M1756) available from Bently.com.
CSA/NRTL/C
When used with
I/O module
ordering options
without internal
barriers
When used with
I/O module
ordering options
with internal
barriers
Class I, Zone 2: AEx/Ex nA nC
ic IIC T4 Gc;
Class I, Zone 2: AEx/Ex ec nC
ic IIC T4 Gc;
Class I, Division 2, Groups A,
B, C, and D;
T4 @ Ta= -20˚C to +65˚C (4˚F to +149˚F)
When installed per drawing
149243 or 149244.
Class I, Zone 2: AEx/Ex nA nC
ic [ia Ga] IIC T4 Gc;
Class I, Zone 2: AEx/Ex ec nC
ic [ia Ga] IIC T4 Gc;
Class I, Division 2, Groups A,
B, C, and D (W/ IS Output for
Division 1)
T4 @ Ta= -20˚C to +65˚C (4˚F to +149˚F)
When installed per drawing
138547.
6/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
ATEX/IECEx
When used with
I/O module
ordering options
without internal
barriers
When used with
I/O module
ordering options
with internal
barriers
II 3 G
Ex nA nC ic IIC T4 Gc;
Ex ec nC ic IIC T4 Gc;
T4 @ Ta= -20˚C to +65˚C (4˚F to +149˚F)
When installed per drawing
149243 or 149244.
II 3(1) G
Ex nA nC ic [ia Ga] IIC T4 Gc;
Ex ec nC ic [ia Ga] IIC T4 Gc;
T4 @ Ta= -20˚C to +65˚C (4˚F to +149˚F)
When installed per drawing
138547.
7/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Terminations
Ordering Information
For the detailed listing of country and
product specific approvals, refer to the
Approvals Quick Reference Guide
(108M1756) available from Bently.com.
Without Recorder Outputs
3500/60-AA-BB
A:I/O Module Type
0 1RTD/TC Non-isolated I/O Module
Internal Terminations
0 2RTD/TC Non-isolated I/O Module
External Terminations
0 3TC Isolated I/O Module Internal
Terminations
0 4TC Isolated I/O Module External
Terminations
0 5RTD/TC Non-isolated I/O Module
with Internal Barriers
0 5RTD/TC Non-isolated I/O Module
with Internal Barriers
B:Agency Approval Option
0 0None
0 1CSA/NRTL/C (Class 1, Div 2)
0 2ATEX/ IECEx/ CSA (Class 1, Zone
2)
B:Agency Approval Option
0 0None
0 1CSA/NRTL/C (Class 1, Div 2)
0 2ATEX/ IECEx/ CSA (Class 1, Zone
2)
Recorder Outputs
3500/61-AA-BB
A:I/O Module Type
0 1RTD/TC Non-isolated I/O Module
Internal Terminations
0 2RTD/TC Non-isolated I/O Module
External Terminations
0 3TC Isolated I/O Module Internal
Terminations
0 4TC Isolated I/O Module External
8/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Internal Barriers and Internal
Terminations. (Not-Isolated)
10/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Graphs and Figures
1. Status LEDs
2. 3500/60 Main Module Front View
3. ISO TC Temp I/O Module (Internal Terminations)
4. RTD/TC Temp I/O Module (Internal Terminations)
5. ISO TC Temp I/O Module (External Terminations)
6. RTD/TC Temp I/O Module (External Terminations)
Figure 1: Front and Rear Views of the 3500/60 Temperature Monitor (Without Recorder Outputs)
11/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
1. Status LEDs
2. 3500/61 Main Module Front View
3. ISO TC Temp I/O Module (Internal Terminations)
4. RTD/TC Temp I/O Module (Internal Terminations)
5. ISO TC Temp I/O Module (External Terminations)
6. RTD/TC Temp I/O Module (External Terminations)
Figure 2: Front and Rear Views of the 3500/61 Temperature Monitor (Recorder Outputs)
12/13
3500/60 and 3500/61 Temperature Modules
Datasheet141540 Rev. L
Copyright 2020 Baker Hughes Company. All rights reserved.
Bently Nevada and Orbit Logo are registered trademarks of Bently Nevada, a Baker Hughes Business, in the United
States and other countries. The Baker Hughes logo is a trademark of Baker Hughes Company. All other product and
company names are trademarks of their respective holders. Use of the trademarks does not imply any affiliation
with or endorsement by the respective holders.
Baker Hughes provides this information on an “as is” basis for general information purposes. Baker Hughes does not
make any representation as to the accuracy or completeness of the information and makes no warranties of any
kind, specific, implied or oral, to the fullest extent permissible by law, including those of merchantability and fitness
for a particular purpose or use. Baker Hughes hereby disclaims any and all liability for any direct, indirect,
consequential or special damages, claims for lost profits, or third party claims arising from the use of the
information, whether a claim is asserted in contract, tort, or otherwise. Baker Hughes reserves the right to make
changes in specifications and features shown herein, or discontinue the product described at any time without
notice or obligation. Contact your Baker Hughes representative for the most current information.
The information contained in this document is the property of BakerHughes and its affiliates; and is subject to
change without prior notice. It is being supplied as a service to our customers and may not be altered or its content
repackaged without the express written consent of Baker Hughes. This product or associated products may be
covered by one or more patents. See Bently.com/legal.
1631 Bently Parkway South, Minden, Nevada USA 89423
Phone: 1.775.782.3611 or 1.800.227.5514 (US only)
Bently.com
13/13
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.