Anritsu MA24330A, MA24340A, MA24350A Data Sheet

Technical Data Sheet
Microwave CW USB Power Sensors
Low Cost, Compact, and Highly Accurate Power Sensors for RF and Microwave Applications
MA24330A
10 MHz to 33 GHz
MA24340A
10 MHz to 40 GHz
MA24350A
10 MHz to 50 GHz
MA243x0A Specifications
Introduction
The MA243x0A series Microwave CW USB Power Sensors employ a single-path diode architecture to provide fast, accurate average power measurements from 10 MHz up to 50 GHz with 90 dB of dynamic range.
Features and Benefits
Broad Frequency Range (10 MHz up to 50 GHz): Ideal for general purpose, aerospace and defense, satellite and wireless
communications applications
Accurate Power Measurements with over 90 dB Dynamic Range
Best-in-Class Damage Protection (+26 dBm CW, +32 dBm peak < 10 µs): Protects instrumentation investment
No Zeroing Required (for signals > –50 dBm) and Elimination of 1 mW Reference Calibration: Reduces test time and handling in
production while maintaining absolute accuracy
Advanced Trigger Capabilities: Facilitates time dependent power measurements
NIST Traceable Calibration: Provides high-accuracy measurements and ensures absolute accuracy
Calibration Traceable to SI Units via National Metrology Institutes
Easy to Use with PC or Select Anritsu Handheld Instruments: No benchtop power meter unit needed
Silicone Protective Covering (removable): Provides additional field durability
External Trigger Latching: For pulses as narrow as 20 ns
MA243x0A Series Microwave CW USB Power Se nsors
Specifications MA243x0A
Table of Contents
Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Sensor Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measurement Uncertainty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
PowerXpert™ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operational Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Regulatory Compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Optional Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Definitions
Warm-Up Time 60 minutes
Operating Temperature Range 0 °C to 50 °C
Characteristic Performance Characteristic specifications are not tested and are not warranted.
ISO GUM Measurement Uncertainty Zero and Noise uncertainty expressed with coverage factor of k=3.
Calibration Cycle Anritsu recommended calibration interval is 12 months.
All specifications and characteristics apply under the following conditions, unless otherwise stated:
Average and Relative Power uncertainty expressed with coverage factor of k=2.
All specifications subject to change without notice. For the most current data sheet, please visit the Anritsu
www.anritsu.com
web site:
Notes Sensors may have degraded performance when dropped without the removable protective covering. This
cover is required for warranted operation.
MA243x0A TDS PN: 11410-00906 Rev. E 3 of 9
MA243x0A Specifications
Sensor Specifications
Frequency
MA24330A 10 MHz to 33 GHz, K(m) Connector MA24340A 10 MHz to 40 GHz, K(m) Connector MA24350A 10 MHz to 50 GHz, V(m) Connector
Power Measurement
<50 MHz
Dynamic Range –70 dBm to +20 dBm
Damage Levels at RF Port +26 dBm, ±20 V DC (+32 dBm peak <10 μs pulse and 10 % duty cycle), minimum
VSWR 1.9:1 1.17:1 1.08:1 1.16:11.21:11.29:11.44:11.5:1
Response
Signal Channel Rise Time 8 μs characteristic
Sampling Rate 140 kS/s
Trigger
Source1Bus, Continuous, Internal, External
Arm Type (for Internal/External) Auto, Single, Multiple, Standby
50 to 150 MHz
>0.15 to 2 GHz
>2 to 6 GHz
>6 to 18 GHz
>18 to 33 GHz
>33 to 40 GHz
>40 to 50 GHz
Internal Trigger
External Trigger
High Level Input Voltage 2.3 V min, 3.0 V max
Low Level Input Voltage 1.3 V min, 1.6 V max
Trigger Repetition Period 7.1 μs min
Dynamic Range –35 dBm to +20 dBm
Level Accuracy ±0.5 dB characteristic
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
Hysteresis 0 dB to 10 dB, with 0.1 dB resolution
Trigger Hold Off 0 s to 10 s, with 0.01 ms resolution
External Trigger Input MCX (female), 5.5 V maximum
Input Impedance 4 kΩ nominal
Type TTL/CMOS
Slope Positive or Negative
Delay Range –5 ms to +10 s
Delay Resolution 10 μs
2
Latency
Trigger Pulse Width 20 ns min
Trigger Holdoff 0 s to 10 s with 0.01 ms resolution
7.1 μs max
1. Bus trigger is not available in PowerXpert application.
2. Latency is defined as the time delay between the defined edge of the applied trigger and the sensor switching into the triggered state.
Specifications MA243x0A
Measurement Uncertainty
Average Power (dB)325 °C to 35 °C:
Range (dBm) 18 GHz >18 GHz to 40 GHz >40 GHz to 50 GHz –70 to <+15 0.11 0.13 0.19 +15 to +20 0.14 0.17 0.23
0°C to 50°C: Range (dBm) 18 GHz >18 GHz to 40 GHz >40 GHz to 50 GHz
–70 to <+15 0.14 0.17 0.25 +15 to +20 0.18 0.21 0.29
4
Zero
Range (dBm) Watt dBm Watt dBm –70 to –20 9.68E-11 –70.14 8.90E-11 –70.50 >–20 to 0 4.95E-09 –53.05 4.14E-09 –53.83 >0 to +20 1.56E-08 –48.08 1.72E-08 –47.65
5
Noise
Range (dBm) Watt dBm –70 to –20 3.53E-11 –74.52 >–20 to 0 6.51E-11 –71.86 >0 to +20 6.30E-10 –62.01
Set Drift
3. Power uncertainty expressed with coverage factor of k=2 for CW measurement after zero operation. Includes calibration factor and linearity over temperature uncertainties, but not the effects of mismatch, zero set and drift, or noise.
4. Zero uncertainty expressed with coverage factor of k=3. One hour warm-up followed by a Zero operation. Measured with 256 averages and 40 ms aperture and with the temperature kept within ±1 °C. Zero Set: Average of the reported power over one hour. Zero Drift: Two sigma value of the reported power over one hour. Specified Zero Set, Drift and Noise are valid at 30 °C ± 5 °C.
5. Two sigma noise at 10.2 seconds of integration time (integration time = aperture time x averaging number). Effect of noise can be reduced by increasing the number of averages and/or increasing the aperture time. Noise is inversely proportional to the square root of number of ADC samples used per measurement; the number of ADC samples per measurement is the product of the sample rate, aperture time, and number of averages used. Noise uncertainty is expressed with coverage factor of k=3.
MA243x0A TDS PN: 11410-00906 Rev. E 5 of 9
MA243x0A Specifications
PowerXpert™
PC Requirements
(version 3.0 or greater)
Processor and RAM Minimum: Equivalent to Intel® Pentium® III with 1 GB RAM or Intel® Pentium® IV with 512 MB RAM
Operating System Microsoft® Windows® 8, Windows® 7, and Windows® XP
Hard-Disk Free Space 100 MB minimum
Display Resolution 1024 × 768 minimum
System
Measurand Average power
Measurement Resolution 0.01 dB max via PowerXpert™, 0.001 dB max via remote command
Offset Correction
Averaging Auto, Manual
Number of Averages (Manual)
Auto Average Resolution
Auto Average Source Scope Data Point Number: 1 to 16,384
Continuous Average Mode
Duty Cycle Correction 0.01 % to 100 %
Aperture Time 0.01 ms to 1 s
Measurement Time
Buffer Size 8192
Scope Mode
Capture Time 0.01 ms to 1 s
Data Points 1 to 16,384
Resolution 0.01 ms max
Measurement Time
List Mode
Number of Measurements 1 to 1000
Input Parameters Frequency (GHz), aperture time (ms), averages
Recommended: Equivalent to Intel® Pentium® IV with 1 GB RAM
Interface USB 2.0 high speed
6
–100 dB to +150 dB
Type Moving, Repeat
7
1 to 65,536
8
1 dB, 0.1 dB, 0.01 dB
9
N x (aperture time x Ct) + T Continuous: >2,100 readings/s (minimum aperture, one average) Buffered: >5,600 readings/s (minimum aperture, one average)
10
N x (capture time x Ct) + (Pn x 0.038 ms) + T
com
com
6. Offset correction feature is available only through the PowerXpert application. There is no remote command for it in the sensor firmware.
7. Maximum number of averages allowed in Continuous Average mode is 65,536. In Scope mode, the maximum number of averages is equal to 16,777,216 divided by the number of data points.
8. Averaging resolution of 0.001 dB is not available with the PowerXpert application. It is defined as the place after the decimal to which the reading becomes stable. For example, if 0.01 is selected, then the reading will typically be stable within ±0.01 dB. Please refer to the remote operation chapter in the user guide for information regarding access to this feature.
9. Speed is defined as the data throughput at the “A” end of the USB A to Micro-B Cable (p/n 2000-2010-R), where: Number of Repeat Averages = N (N = 1 for moving average mode) Capture Time Coefficient = Ct = 8.238 Command Processing Time = T Speed may vary depending on the speed of and load on the CPU controlling the sensor. Specified results obtained with Intel® Core™ i5-3550 CPU running at 3.30 GHz
10. Speed is defined as the data throughput at the “A” end of the USB A to Micro-B Cable (p/n 2000-2010-R), where: Number of Repeat Averages = N (N = 1 for moving average mode) Capture Time Coefficient = Ct = 8.238 Number of Points = Pn Command Processing Time = T
= 0.347 ms
com
= 0.289 ms
com
Specifications MA243x0A
External
Trigger Input
(TTL/CMOS)
K and V Type connectors designed for use with a torque wrench ensuring repeatable connections
Trigger Output
USB Micro-B port for connectivity to host
(PC or Anritsu handheld instrument)
Color LED
reports functional
status of the sensor
General
RF Connector K male (MA24330A, MA24340A)
Interface to Host USB 2.0 high speed
Current Consumption 410 mA to 450 mA characteristic (20 ºC to 30 ºC)
Weight 397 g (0.88 lb)
Warranty 1 year
Operational Requirements
Operating Temperature Range 0 ºC to 50 ºC
Storage Temperature Range –40 ºC to +71 ºC
Humidity 45 % relative humidity at 50 ºC (non-condensing)
Altitude 4600 m operational max
Shock 30 g
Vibration Sinusoidal: 5 Hz to 55 Hz, 3 g max
V male (MA24350A)
Size 110 mm x 46 mm x 25.6 mm, excluding K or V connector and silicone protective covering
Tests were performed per MIL-PRF-28800F (Class 3).
75 % relative humidity at 40 ºC (non-condensing) 95 % relative humidity at 30 ºC (non-condensing)
half-sine, 11 ms duration
n
Random: 10 Hz to 500 Hz Power Spectral Density: 0.03 g
2
/ Hz
Regulatory Compliance
European Union EMC 2014/30/EU, EN 61326:2013, CISPR 11/EN 55011, IEC/EN 61000-4-2/3/4/5/6/8/11
Australia and New Zealand RCM AS/NZS 4417:2012
South Korea KCC-REM-A21-0004
MA243x0A TDS PN: 11410-00906 Rev. E 7 of 9
Low Voltage Directive 2014/35/EU Safety EN 61010-1:2010 RoHS Directive 2011/65/EU
MA243x0A Specifications
Ordering Information
Available Models
MA24330A 33 GHz USB Power Sensor MA24340A 40 GHz USB Power Sensor MA24350A 50 GHz USB Power Sensor
Included Accessories
Available Options
MA24330A-097 MA24340A-097
MA24330A-098 MA24340A-098 MA24350A-098
MA24330A-099 MA24340A-099 MA24350A-099
Optional Accessories
Calibrated Torque Wrenches
Precision Fixed Attenuators
10585-00021 Quick Start Guide
2000-1605-R 1.5 m BNC(m) to MCX(m) cable 2000-2010-R 1.83 m USB A to Micro-B cable
Option 97: ISO/IEC 17025 and ANSI/NCSL Z540-1 or ANSI/NCSLI Z540.3 (includes test report, uncertainty data, and accreditation symbol)
Option 98: Standard calibration ISO/IEC 17025 and ANSI/NCSL Z540-1
Option 99: Premium calibration ISO/IEC 17025 and ANSI/NCSL Z540-1 (includes test report and uncertainty data)
01-201 Calibrated torque wrench for K and V connectors
41KB-3 DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
41KB-6 DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f) 41KB-10 DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f) 41KB-20 DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
41KC-3 DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f)
41KC-6 DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f) 41KC-10 DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f) 41KC-20 DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f)
41V-3 DC to 60 GHz, 3 dB, 50 Ω, V(m) to V(f)
41V-6 DC to 60 GHz, 6 dB, 50 Ω, V(m) to V(f) 41V-10 DC to 60 GHz, 10 dB, 50 Ω, V(m) to V(f) 41V-20 DC to 60 GHz, 20 dB, 50 Ω, V(m) to V(f)
43KB-3 DC to 26.5 GHz, 3 dB, 50 Ω, K(m) to K(f)
43KB-6 DC to 26.5 GHz, 6 dB, 50 Ω, K(m) to K(f) 43KB-10 DC to 26.5 GHz, 10 dB, 50 Ω, K(m) to K(f) 43KB-20 DC to 26.5 GHz, 20 dB, 50 Ω, K(m) to K(f)
43KC-3 DC to 40 GHz, 3 dB, 50 Ω, K(m) to K(f)
43KC-6 DC to 40 GHz, 6 dB, 50 Ω, K(m) to K(f) 43KC-10 DC to 40 GHz, 10 dB, 50 Ω, K(m) to K(f) 43KC-20 DC to 40 GHz, 20 dB, 50 Ω, K(m) to K(f)
Precision Coaxial Adapters
33KFKF50B DC to 40 GHz, 50 Ω, K(f) to K(f)
33KKF50B DC to 40 GHz, 50 Ω, K(m) to K(f)
33VFVF50C DC to 70 GHz, 50 Ω, V(f) to V(f)
33VVF50C DC to 70 GHz, 50 Ω, V(m) to V(f)
34NKF50 DC to 18 GHz, 50 Ω, N(m) to K(f)
34NFKF50 DC to 18 GHz, 50 Ω, N(f) to K(f)
Training at Anritsu
Anritsu has designed courses to help you stay up to date with technologies important to your job. For available training courses, visit:
www.anritsu.com/training
• United States Anritsu Americas Sales Company
450 Century Parkway, Suite 190 Allen, TX 75013, U.S.A. Phone: +1-800-Anritsu (1-800-267-4878)
• Canada Anritsu Electronics Ltd.
700 Silver Seven Road, Suite 120 Kanata, Ontario K2V 1C3, Canada Phone: +1-613-591-2003 Fax: +1-613-591-1006
• Brazil Anritsu Eletronica Ltda.
Praça Amadeu Amaral, 27 - 1 Andar 01327-010 - Bela Vista - Sao Paulo - SP Brazil Phone: +55-11-3283-2511 Fax: +55-11-3288-6940
• Mexico Anritsu Company, S.A. de C.V.
Blvd Miguel de Cervantes Saavedra #169 Piso 1, Col. Granada Mexico, Ciudad de Mexico, 11520, MEXICO Phone: +52-55-4169-7104
• United Kingdom Anritsu EMEA L td.
200 Capability Green Luton, Bedfordshire, LU1 3LU, U.K. Phone: +44-1582-433200 Fax: +44-1582-731303
• France Anritsu S.A.
12 avenue du Québec, Bâtiment Iris 1- Silic 612, 91140 Villebon-sur-Yvette, France Phone: +33-1-60-92-15-50 Fax: +33-1-64-46-10-65
• Germany Anritsu GmbH
Nemetschek Haus, Konrad-Zuse-Platz 1 81829 München, Germany Phone: +49-89-442308-0 Fax: +49-89-442308-55
• Italy Anritsu S.r.l.
Via Elio Vittorini 129, 00144 Roma, Italy Phone: +39-6-509-9711 Fax: +39-6-502-2425
List Revision Date: 20200602
• Sweden Anritsu AB
Isafjordsgatan 32C 164 40 Kista, Sweden Phone: +46-8-534-707-00
• Finland Anritsu AB
Teknobulevardi 3-5 FI-01530 Vantaa, Finland Phone: +358-20-741-8100 Fax: +358-20-741-8111
• Denmark Anritsu A/S
c/o Regus Winghouse, Ørestads Boulevard 73, 4th floor, 2300 Copenhagen S, Denmark Phone: +45-7211-2200
• Russia Anritsu EMEA Ltd.
Representation Office in Russia
Tverskaya str. 16/2, bld. 1, 7th floor Moscow 125009, Russia Phone: +7-495-363-1694 Fax: +7-495-935-8962
• Spain Anritsu EMEA Ltd.
Representation Office in Spain
Paseo de la Castellana, 141. Planta 5, Edificio Cuzco IV 28046, Madrid, Spain Phone: +34-91-572-6761
• United Arab Emirates Anritsu EMEA Ltd.
Dubai Liaison Office
902 Aurora Tower P O Box: 500311- Dubai Internet City Dubai, United Arab Emirates Phone: +971-4-3758479 Fax: +971-4-4249036
• India Anritsu India Private Limited
6th Floor, Indiqube ETA, No.38/4 Adjacent to EMC2, Doddanekundi, Outer Ring Road Bengaluru 560048, India Phone: +91-80-6728-1300 Fax: +91-80-6728-1301
• Singapore Anritsu Pte. Ltd.
11 Chang Charn Road, #04-01, Shriro House Singapore 159640 Phone: +65-6282-2400 Fax: +65-6282-2533
• P.R. China (Shanghai) Anritsu (China) Co., Ltd.
Room 2701-2705, Tower A New Caohejing International Business Center No. 391 Gui Ping Road Shanghai 200233, P.R. China Phone: +86-21-6237-0898 Fax: +86-21-6237-0899
• P.R. China (Hong Kong) Anritsu Company Ltd.
Unit 1006-7, 10/F. Greenfield Tower, Concordia Plaza No. 1 Science Museum Road Tsim Sha Tsui East, Kowloon Hong Kong, P.R. China Phone: +852-2301-4980 Fax: +852-2301-3545
• Japan Anritsu Corporation
8-5, Tamura-cho, Atsugi-shi, Kanagawa, 243-0016 Japan Phone: +81-46-296-6509 Fax: +81-46-225-8352
• South Korea Anritsu Corporation, Ltd.
5FL, 235 Pangyoyeok-ro Bundang-gu, Seongnam-si Gyeonggi-do 13494, South Korea Phone: +82-31-696-7750 Fax: +82-31-696-7751
• Australia Anritsu Pty. Ltd.
Unit 20, 21-35 Ricketts Road Mount Waverley, Victoria 3149, Australia Phone: +61-3-9558-8177 Fax: +61-3-9558-8255
• Taiwan Anritsu Company Inc.
7F, No. 316, Sec. 1, NeiHu Rd. Taipei 114, Taiwan Phone: +886-2-8751-1816 Fax: +886-2-8751-1817
Data subject to change without notice. For the most recent specifications, visit: www.anritsu.com.
9
9 of 9
® Anritsu All trademarks are registered trademarks of their respective companies.
Copyright January 2021, Anritsu Company, USA. All Rights Reserved.
MA243x0A TDS, PN: 11410-00906, Rev. E
Anritsu utilizes recycled paper and environmentally conscious inks and toner.