ALTANA Haze-Gard Plus Short Instructions

Additives & Instruments
A member of
Measure what you see.
delete all
delete last
statistic result
on/off on/off
operate set up reference
trans-
mittance
hazeclarity
haze-gard plus Operation
On the left Side:
How to Calibrate?
clarity-port
Main Switch on/off
Voltage Selector
Fuse
Connection for:
Power Cable
Foot Switch
PC
haze-port
operate button:
Measure
set up button:
SET-UP MENU
reference button:
Measure reference in Continuous mode
SET UP MENUE
¢
Calibrate
£
Print Statistic
£
Print Online: off
£
Cont. Mode: off
£
Language USA
Escape: SET UP
Cursor keys:
Used in SET-UP mode to select individual menu items and set values. In any measuring mode, these keys are used to switch statistics functions on/off, and to delete values from statistics memory
Calibration
Zero Calibration
Calibration
100% Calibration
Calibration
Place clarity stand
75.5
at clarity-port
BYK-Gardner GmbH · Lausitzer Strasse 8 · 82538 Geretsried · Germany · Tel. +49-8171-3493-0 · Fax +49-8171-3493-140 · www.byk.com/instruments
Calibration is necessary at the first startup and at a regular interval (two months recommended):
Press set up button to call up
the calibration routine
Cover haze-port with dark
standard and press operate
Remove cover from haze-port
and start the 100% measuremen
Place the clarity standard at the
clarity port and press operate
269 020 584 E 0806
Please note the detailed operating instructions in English on CD.
Measure what you see.
Additives & Instruments
A member of
delete all
delete last
statistic result
on/off on/off
operate set up reference
trans-
mittance
hazeclarity
How to measure How to measure continuously?
3
1
4
Display example:
STATISTIC n = 5
H 30.6 m = 30.1 s = 0.5 C 20.8 m = 20.3 s = 0.40
2
Select Parameter
Press a button to add a parameter. Press the button again to display this parameter only.
Place sample at:
haze-port for Haze and Transm. clarity-port for Clarity
Measure: Press operate
button or foot switch
Statistics:
1 Switches statistics on and off 2 Switches all statistical data on
and off
3 Last saved measurement is
deleted
4 All statistics in the memory
are deleted
The Statistics mode can be selected for any combination of parameters. In Continuous mode, statistics can be run for only one parameter
SET UP MENUE
£
Calibrate
£
Print Statistik
£
Print Online: off
¢
Con. Mode: on
£
Language USA
Escape: SET UP
How to change settings?
SET UP MENUE
£
Calibrate
£
Print Statistik
£
Print Online: off
£
Cont. Mode: on
¢
Language USA
Escape: SET UP
Continuous Mode:
Select Cont. Mode “on” in the
SET-UP Menu
Select the desired parameter
to be measured (only one is possible in this mode).
Haze measurement needs the transmittance value as a reference. For a new sample press the reference button first. To add readings to statistics, press operate button.
Press set up button to
display the SET UP menu
Use the cursor keys to select
the desired item: Print function, Language, Beeper, Date/Time, Contrast, Calibration values