Fe-Probe Measurements on iron and steel substrate
NFe-Probe Measurements on non-m agnetic metal substr ates, such as
Aluminum, Zinc, Copper, Brass, stainless steel
Range see specification of probes
Resolution 0.1 !m 0 to 99.9 !m 0.01 mil 0 to 9.99
mil
(standard set ti ng) 1 !m 100 to 999 !m 0.1 mil 10 to 100
mil
0.01 mm " 1.00 mm 1 mil > 100 mil
Accuracy (factor y cali bration on zero reference plates supplied with the gauge)
± (1 !m + 2%*) 0 to 2000 !m ± (0.04 mil + 2%*) 0 to 80 mil
± 3.5%* > 2000 !m (80 mil)
(* of reading)
With the one-poi nt or two-point calibration the accuracy can still be
improved using the optionally avail able test shims.
Minimum Fe-Probe 10 x 10 mm# (0.4" x 0.4")
Area NFe-Probe 6 x 6 mm# (0.24" x 0.24")
Minimum convex 5 mm (0.2")
Curvature concave 30 mm (1.2")
Minimum Fe-Probe 0.2 mm (8 mil)
Substrate Thi ckness NFe-Probe 0.05 mm (2 mil)
Memory Capacity Basic
Readings 100 up to 13,000 (max. 2000 r eadings per batch)
Batches 1 999
Calibrations 1 100
Interf ace wireless 2. 4 GHz, range max. 10 m (30ft) in undisturbed environment
Temperature Range
Storage -10°C to 60°C (14°F to 140°F)
Operation 0°C to 50°C (32°F to 122°F)
Power Supply 2 x batteries (AA) 1.5V Alkali, or 2 x rechargeable batt eries (AA) 1.2V
Dimensions 124 mm x 67 mm x 33 mm ( 4.9" x 2.6" x 1.3")
Weight ca. 120 g (4.3 oz) gauge with batteries and probe
Premium
byko-test 8500
Instruction Manua l
15
MAIN MENUMeasuring Mode
MAIN MENU
conti nued
Calibr ationCustom Calibr ation
Setup
Select SubstrateFe / NFe
Measurem ent T ypeSingle Measur ement
Uppe r Li mit
Lower Limit
AveragingOFF
Delete All
Memory O FF
Average Z ero
Display
Wireless Interface
Display Sy s t em Info
Language
Buzzer
Date / Time
New Batch
Select Batch
Data Memory
StatisticsSelect Batch for DisplaySelec t Stat istic Par ameter s
Fe
NFe
Com bined Measurem ent
Continuous M ode
Set
Delete
Set
Delete
Input No. of Readings
Input N a me of Ba tch
Select
Display R eadingsReadings
Inf oDate / Time
Uppe r Li mit
Lower Limit
Renam e BatchMod ify Batc h Na me
Delete BatchOK
Delete All
Reset to Factory Calibration
Define New Calibration
Select Calibr ationSelect
OFF
Input N o. of Z er o R eadings
Unit
Resolution
Backlight
Orient ation
ON
OFF
Select Language
ON
OFF
Set Time
Set Date
Input Limit Value
OK
Canc el
Input Limit Value
OK
Canc el
Set
Delete
Set
Delete
Canc el
Inf oDate / Time
Delete ReadingOK
Input Limit Value
OK
Canc el
Input Limit Value
OK
Canc el
Input Name
Calibr ate
Renam e Cali b r ation
Delet e C alibrationOK
!m
mil
Coarse
Normal
Fine
Auto
OFF
Normal
Flip
= selection by scr olling t hrough list
= basic model
One Point C al
Two Poi n t Ca l
Modify Name
Canc el
= selection by scrolling through list
= basic model
Canc el
Perform Calibra tion
2
5 Setting of Zero Ref erence
When using the probe for the first time, after a batter y exchange, with varying appli cati ons,
or after a long peri od the probe was not used, it is necessary to take a zero reference.
Press the !- key. Then pl ace the probe on one of the reference plates in the case or on an
appropriat e uncoated Fe- or NFe-substrate. A cont rol number is shown in the display. Then
lift the probe of f by minimum 5 cm (2"). After a second control number appears shortly in the
display, i t switches to 0.0!m (mil). The control numbers are not of importance f or the user.
The process of zero adj ustment is completed now.
: The zero reference can also be set by placing the probe first and then pressing the
Note
!- key.
Regarding zero adj ustment on rough surfaces see § 7.1.2.
6 Measuring Mode
The gauge offers different options to select the measuring mode.
Setting of the measuring mode can be done following the normal menu sequence (see §
7.2.1). The setting can also be accessed directly using the key " when the menu is not activated.
Fe or NFe-Mode
Fe / NFe-Mode
Combined Mode
IMPORTANT
thickness of 50 #m (2 mil) in order to give correct readings.
: For those appl ications with cl early defined type of substrat e the
measuring mode can be fixed to Fe or NFe respectively.
: For applications with varrying substrates (Fe- and NFe-
substrates) the Fe/NFe
For example if the substrate changes from steel to al uminum and
the probe is placed on the new surf ace for the first time, the
gauge displays the message " No Fe-substrate – repeat". When
the probe is placed again t he new m easurement is taken with the
NFe-mode.
The procedure is accor dingly when changing fr om a non-fer rous
to a ferrous substrate.
When being switched on the gauge starts with the measuri ng
mode, which was used befor e switch off.
: This mode can be used when for example an object made from
steel is galvanized with zinc and then painted. The gauge
displays the t hickness of the zinc coating and the paint
separately.
: In the combined mode the NFE-substrate must have a minimum
-mode should be set.
Content
1 General 5
2 Change of Probes 5
3 Power Supply and Change of Batte ries 5
4 Handling 5
5 Setting of Zero Reference 6
6 Measuring Mode 6
7 Menu 7
7.1 Calibration 7
7.1.1 User-Specific Calibrati on 7
7.1.1.1 Delete All User-Specific Calibrations 7
7.1.1.2 Factory Calibration 7
7.1.1.3 Creation of a New User-Specific Calibration 7
7.1.1.4 Selection of User-Specific Calibration 8
7.1.1.4.1 Activation of User-Specific Calibration 8
7.1.1.4.2 Calibration Method 8
7.1.1.4.2.1 One-Point-Calibration 8
7.1.1.4.2.2 Two-Point-Calibration 8
7.1.1.4.3 Rename User-Specific Calibration 8
7.1.1.4.4 Delete User-Specific Calibration 8
7.1.2 Average Ze ro Reference on Rough Sur faces (BMR) 8
7.2 Measuring M ode Settings 9
7.2.1 Selection of Substrate 9
7.2.2 Single Measurement / Conti nuous Measureme nt 9
7.2.3 Upper Limit 9
7.2.3.1 Setting of Upper Limit 9
7.2.3.2 Delete Upp e r Limit 9
7.2.4 Lower Limit 9
7.2.5 Averaging of Readings 9
7.3 Memory Mode 10
7.3.1 Delete Entire Memory 10
7.3.2 ON / OFF of Memory Mode 10
7.3.3 Creating a New Batch 10
7.3.4 Selection / Configurati on of a Batch 10
7.3.4.1
7.3.4.2 Display of Readings of a Batch 10
7.3.4.2.1 Information About a Readi ng 10
Activation of a Batch 10
6 3
7.3.4.2.2 Delete a Reading 10
7.3.4.3 Information About a Batch 11
7.3.4.4 Limits 11
7.3.4.5 Rename a Batch 11
7.3.4.6 Delete a Batch 11
7.4 Statistics 11
7.4.1 Selection of Statisti cal Parameters of the Active Batc h to be Displayed 11
7.4.2 Display of Statistics of not Activated Batches 11
7.5 Gauge-Setup 11
7.5.1 LC-Display 11
7.5.1.1 Unit of Readings 11
7.5.1.2 Resolution 12
7.5.1.3 Backlight 12
7.5.1.4 Orientation 12
7.5.2 Wireless Interface 12
7.5.3 System-Information 12
7.5.4 Language 12
7.5.5 Buzzer 12
7.5.6 Date / Time 12
7.5.6.1 Setting of Time 13
7.5.6.2 Setting of Date 13
8 Installation of the QuaNix
®
-Software 13
9 Installation of the Drivers for the USB-RF-Dongle 13
10 Setting of the Gauge Opt ions via QuaNix®-Software 13
11 Download of a New Language into the Gauge 13
12 Update of Software in Gauge and Probes 13
13 Symbols Used in the Display 14
14 Scope of Supply 14
15 Technical Data 15
1 General
The gauges of byko-test 8500 family, together with t he corr esponding probes, constitute a
modular system. The handhel d gauges and the probes can be arbitrarily combined. The
probes can be used integr ated in the gauge as well as attached to it via adapter cable. Thi s
enables the user t o choose the optimum measuring configuration for his appli cation.
The gauge options can not only be set using the simple menu. With the QuaNix
all options can be comfortably preset on the PC and finally downloaded into the gauge.
A wireless interface is used for the data exchange between gauge and PC.
2 Change of Probes
The probes or the housi ng of t he adapter cable lock in when inserted in the gauge. To
remove them from the gauge press the button on the rear side of the gauge and pull the
probe or the adapter out.
After inserti ng a probe the gauge displays inf ormation about type of probe, serial number
and number of softwar e version.
These data can be called anytime by pressing key ! when the gauge is in the normal
display mode.
3 Power Supply and Change of Batteries
The byko-test 8500 gauge is powered by 2 batteries AA (1.5 V) alkal ine, or 2 rechargeable
batteri es AA (1.2 V). Please note that the capacity of the rechar geable batteries is
significantly lower.
Step 1: A change of batteries is advisable. However quite a number of
measurements can still be taken. The EL-backlight is still active.
Step 2: (small symbol
deactivated.
Step 3: (large symbol in center of displ ay
taken.
During change of batteries menu settings and all readings in the memory remain st ored.
: Only setting of dat e and time can get lost, when the change of batt eri es takes longer
Note
than about 2 minutes.
IMPORTANT
Discharged batt eries might leak and damage the gauge.
: Empty batteries should be disposed of properly. If possible, please use t he special
Note
services of your council.
4 Handling
The gauge switches on automat ically when the probe is placed on the surface of the object
under test. It can al so be switched on by one of the keys of the keypad. If the probe is not
used for longer than about 30 seconds the gauge switches off again.
Place the probe evenl y on the object under test. On bent surfaces use the V-grooves
aligned parallel to the axis of curvatur e to insure that the probe is vertical ly placed to the
surface.
An error message wil l be di splayed when the probe is placed incorr ectly. For measurements
beyond the measuring range the gauge will display "Infi".
Do not take measurements on magnetized parts. Magnetic fi elds can influence the results.
Also electro-magnetic fields can influence the NFe-measurements.
: Please take batteri es out of the gauge when storing it over a longer period!
®
-Software
) Measurements can sti ll be taken. The EL-backlight is
) No more measurements can be
4 5
7.3 Memory Mode
With this option t he memor y m ode is switched on or off. Also configurat ion and handling of
the batches is controlled.
After pressing key ! the cursor moves to the active item of the following menu level:
- "OFF", if the memory mode is switched off or
- the active bat ch in the memory mode
7.3.1 Delete Entire Memory
The entire memory i s delet ed when confirming this menu item by pressing key !.
7.3.2 ON / OFF of Me mo ry M ode
To set the gauge to the memory mode either a new batch must be created or an already
existing one must be selected and activated with !.
By selecting "O FF" t he mem ory mode is swi tched off .
7.3.3 Creating a New Batch
After pressing key ! the number of the new batch is displayed, e. g.
> B 0005 <
The cursor highlights the first digit.
The number can be accepted by holdi ng the key ! pressed until the cursor moves to "OK" .
The new batch is confirmed by pressing key !. It is added to the list and mar ked as acti ve
batch. Then the gauge r etunrs to the normal display m ode. Global limits acti vated before the
new batch was created, wil l al so be used in the new batch.
When the number appear s the new bat ch can also be given an individual nam e with max. 8
digits. Each of the possible digits can be set to an alphanumeric sign using the keys "and
#. The sign is confirmed by pr essing key !.
Each digit can be modified using the backspace key $.
: When creating a new batch after other batches (e.g. B 0001, B 0002) have been
NOTE
deleted, the new number wil l be e. g. B 0003. Only after deleting the entire m em ory t he
number will restar t with B0001.
7.3.4 Selection / Configuration of a Batch
When selecting this menu item with the gauge set to memory mode the cursor marks the
active batch.
A new batch can be selected by scr olling with the keys " and #.
7.3.4.1 Activation of a Batch
When selecting menu it em "SELECT" and then pressing the key ! the selected bat ch is
activated. The gauge returns to the normal display mode.
7.3.4.2 Display of Readings of a Batch
When confirming t he item "Show Readings" the list of readings stored in the batch is
displayed. Scr olling through the li st i s done with the keys "and #.
7.3.4.2.1 Information About a Reading
Information about date and time when a highlighted reading was taken can be viewed via
the following menu "I NFO".
7.3.4.2.2 Delete a Reading
With this menu item a single reading in the batch can be deleted.
7 Menu
The Structur e of the menu i s shown on the first two pages of the instruction manual. It
contains all possible options. Depending on the version of the gauge only part of these
options is available. However the procedures t o handle the available opt ions remain the
same.
Function of t he keys in the menu mode:
! Activation of the menu, call of the next menu level , selection of the menu item
$ Backspace, cancellation of a function
! , " scrolling through the list within a menu level
The current menu it em is indicated by a dark background.
#mark of activated menu item
The menu mode is left after a shor t actuation of the probe tip. The gauge ret urns to the
normal displ ay m ode.
7.1 Calibration
The probe is supplied wit h a factory calibration, which provides correct readings for most of
the applications. This calibration is stored in the probe and can not be changed by the user.
Special calibrations can be set for those applications where the readings are not within
specified tolerance or a higher accuracy within a certain range is required.
Depending on the gauge version, up to 100 special cali brations can be set and stored. Each
calibrat ion can be individually named.
7.1.1 User-Specific Calibration
7.1.1.1 Delete All User-Specific Calibrations
When confirming this option by pressing key ! all specific calibrations ar e delet ed.
The probe is reset to the f a ctory calibration.
7.1.1.2 Factory Calibration
When confirming this option by pressing key ! the probe is reset to the factory calibration.
The gauge returns t o the normal measuring mode.
7.1.1.3 Creation of a New User-Specific Calibration
When activati ng the option "New Calibration" t he number of the new calibration is displayed,
e.g.
>C 004 <
The cursor highlights the first sign. This number can be selected by holding key ! until the
cursor moves to"OK" . After confirmation by pressi ng key ! the new calibration can be
defined as descri bed in § 7.1.1.5.2.
The number can be modified t o an indi vidual name with up to 8 alphanumeric signs. This
modificat ion can be done when the new number is displayed. Each position can be adjusted
with the keys " and # until the desired sign appears. The sign is confirmed by pressing
key !. The cursor moves on to the next position.
A correcti on of previously set signs can be done anyti m e by pressing the backspace key $.
10
7
7.1.1.4 Selection of User-Specific Calibration
On this menu level all speci fic calibrations are listet below the option " N ew Calibration". To
scroll t hrough the list use the keys ! and ". By pressing key # the corresponding m enu is
opened. The activat ed calibration is marked by a dot right to the name.
7.1.1.4.1 Activation of User-Specific Calibration
The selected calibration is activated by pressing #.
A warning is displayed, when this calibration has not yet been performed. In thi s case the
menu moves on to the list from which one of the two calibration procedures can be selected
(see following §§).
7.1.1.4.2 Calibration
Prior to performing a calibration it is recommended t o set the zero reference on the
corresponding uncoat ed substrate.
Two types of calibration are available:
7.1.1.4.2.1 One-Point-Calibration
With a one-point cali bration the accuracy of the probe can be optimized in the vicinity of an
expected coating t hickness
To do so a test shim with a thickness of about the same as the expected coating thickness
is placed on the uncoated substrate. Then the probe is placed on the shim. The displayed
reading can be adjust ed up or down t o the thickness of the shim by pressing ! or ".
The adjustment of t he calibration is confirmed and activated by pressing key #. The gauge
returns to the normal measuring mode.
.
7.1.1.4.2.2 Two-Point-Calibration
A two-point calibration is used to optimize the probe for a certain range of coating thickness.
As a first step place a test shim with a thickness close to the lower
thickness range on t he uncoated substrate. Then place the probe on the shim. Adjust the
reading to the thickness of the shim using the keys ! and " until the reading shows the
thickness of t he shim. The adjustment is terminated by pressing the key #.
In a second step place a t est shi m with a thickness close to the upper
thickness range on t he uncoated substrate. Then place the probe on the shim. Adjust the
reading to the thickness of the shim using the keys ! and " until the reading shows the
thickness of t he shim. The adjustment is terminated by pressing the key #, and the
calibration is activated.
After the second step the gauge returns to the normal measuring mode.
limit of the coating
limit of the coating
7.1.1.4.3 Rename User-Specific Calibration
Modificat ion of the name of a calibration is done as described in § 7.1.1.2.
After changing the nam e the gauge automatically moves back to menu item of § 7.1.1.4.
7.1.1.4.4 Delete User-Specific Calibration
After pressing key # the display shows "OK" and "Cancel" respecti vely. When confirmi ng
"OK" the user-specific calibration is deleted.
7.1.2 Average Zero Reference on Rough Surfaces (BMR)
On rough substrate surfaces the zero readings wi ll vary from one spot to the next. This
variation increases with increasing roughness.
This option allows the user to find an average zero reference. Select a number of zero
readings that need to be taken before the gauge calculat es the average zero (Base Met al
Reading = BMR). This values is stored and used to determine the coating thi ckness on the
corresponding coat ed substrate.
When pressing the key # in the menu "Average Zero" the options "OFF" and "Val ue" show
up.
The option "OFF" resets the gauge to a single zero reading.
When selecting the option "Value" the number of zero r eadings for the BMR calculation can
be set using the keys ! and ". The number is confir m ed by pressi ng key #.
7.2 Measuring Mode Settings
7.2.1 Selection of Substrate
In addition to the nor m al Fe- and NFe-mode the gauge provides a combination mode. In this
mode both measurement s, Fe and
This combined measurement can be helpful when a metal coating (e.g. zinc, chromium) on
a ferrous substrate is coated e.g. by lacquer.
Important
: The non-ferrous metal coating must have a minimum thickness of 50 $m (2 mil).
NFe, are taken at the same time.
7.2.2 Single Measurement / Continuous Measurement
Beside the single m ode the gauge can al so be set to a continuous mode. In the continuous
mode the gauge performs repeated measurements in short intervals as long the probe i s
placed on the surf ace.
The continuous mode is i ndicated by the symbol .
7.2.3 Upper Limit
7.2.3.1 Setting of Upper Limit
After pressing key # the last reading is displayed, if no limit was set before. Ot herwi se the
current val ue is di splayed.
The upper limit is set by pressing ! or " and confirmed with #.
A reading exceeding the upper limit is indicated by the symbol ! and an acoustical signal.
7.2.3.2 Delete Upper Limit
After pressing key # the options "OK" and "CANCEL" appear on the display. With the
confirmation of the selection "OK" the upper li mit is deleted. When selecting "CANCEL" t he
limit remains unchanged.
7.2.4 Lower Limit
Selection and deleting of the lower limit is perf ormed in the same way as described in §
7.2.3.
A reading small er than t he lower limit is indicated by the symbol " and an acoustical signal.
7.2.5 Averaging of Readings
With large variations in coating thickness or to meet the requirement s of a test standard the
gauge can be programmed to calculate and display / store an average coating thi ckness
from a predefi ned num ber of measur em ents.
To activate t he averaging the number of measurements m ust be set in the next menu level
by pressing !or ". The selected number must be confirmed with #.
This mode is indicated by the sign x.
When measuring in the averaging mode the actual reading i s displayed together with a
counter for the remaining readings unti l t he average is displayed at nor m al si ze.
8 9
13 Symbols used in the Display
∞ Fe measurement on steel or iron substrate
∞ NFe measurement on non-m agnetic substrate
∞ Error handling error
∞ Infi measurement beyond range, wrong substrate
∞ ! upper limit exceeded
∞ " lower limit exceeded
∞ x average
∞ # standard deviation
∞ $wireless interface activated
∞ continuous measuri ng m ode
∞ Buzzer switched of f
∞ batteries week, measurement s can still be taken
∞ batteries empty
14 Scope of Supply
∞ Gauge with opti onal probe
∞ Adapter cable for probe
∞ 2 batteries AA 1,5V (alkaline)
∞ Instruction manual
∞ Test certificate for optional probe
∞ Soft pouch with clip
∞ Case for tr ansport and storage
Optionall y available
∞ Fe-Probe 2 m m (80 mil)
∞ Fe-Probe 5 m m (200 mil)
∞ NFe-Probe 2 mm (80 mil)
∞ Dual-Probe Fe/NFe 2 / 2 mm (80 / 80 mil)
∞ Dual-Probe Fe/NFe 5 / 2 mm (200 / 80 mil)
∞ Calibration shims, Zero refer ence plates, also available as set
∞ QuaNix
configuration (*)
∞ Wireless interface for data tr ansfer between gauge and PC (*)
(*) Already included i n the scope of supply of the byko-test 8500Premium version
®
-Software for transfer to and evaluation of the data in EXCEL and gauge
7.3.4.3 Information About a Batch
The item "INFO" retur ns information about date and time when the batch was created, and
how many readings it contains.
7.3.4.4 Limits
Upon starti ng thi s funct ion the gauge displays the limits stored with the batch. If no limits are
set in the batch, the gauge di splays the last reading.
Upper and lower limits can be set for each batch indivi dually using the same procedure as
described in §§ 7.2.3 and 7.2.4.
The modified lim it s are used in the active batch instead of the global limits.
7.3.4.5 Rename a Batch
Renaming a batch is done i n the same way as described in § 7.3.3.
7.3.4.5 Delete a Batch
The entire bat ch can be deleted when calling the delete function and finally confirming the
procedure by pressi ng the key %.
7.4 Statistics
The gauge calculates the following parameters:
- Average
- Standard deviation
- Maximum
- Minimum
Within the menu for the statistics two di splay modes are used:
a) In the memory mode two statistical param eters of the active batch can be
displayed. These param eters are updated with each new reading added to the
batch.
b) When calling the statistics of a non-acti ve batch all parameters are displ ayed.
When entering thi s menu, the list of all batches is displayed.
7.4.1 Selection of Statistical Parameters of the Active Batch to be
Displayed
After selecting the active batch and pressing key % the list of statistical parameters is
displayed. With the keys ! and " the individual parameters can be select ed and be
activated by pr essing key %.
: As a maximum 2 parameters can be activated to be displayed !
NOTE
To leave this menu item press the (backspace-) key &. For an immediate return to the
normal displ a y mode just press the probe tip shortl y.
7.4.2 Display of Statistics of not activated Batches
When selecting a non-active batch using keys !and " and confirming the selection with
key % all statistical parameters of this batch are displayed in a popup-window.
7.5 Gauge-Setup
7.5.1 LC-Display
7.5.1.1 Unit of Readings
The readings can be displayed either in 'm or in mil (standard setting: 'm).
The unit can be select ed by using keys ! or ". The selection is confirmed by pressing key
%.
14
11
7.5.1.2 Resolution
The resolution of the readings can be adapted to the application. E.g. on surfaces wit h high
roughness the resol ution should be reduced because of the higher variance of t he readings.
The resoluti on can be "coarse", "normal" or "fine".
7.5.1.3 Backlight
In the "AUTO"-mode the backlight is turned on for about 7 seconds when a measurem ent is
taken.
The backlight can be switched off permanently selecting "OFF".
Switching off the backlight extends battery life time.
7.5.1.4 Orientation
Orientation of the display can be adapted to the handl ing of the gauge.
E.g. when the probe is att ached to the gauge via adapter cable and the gauge is put down
on a table the display can be flipped by 180° so that reading the display is the same as if the
gauge were handheld.
To set the displ ay ori entation press key ! to enter the sel ection. W ith keys "or # select
"Normal" or "Flip" and confirm with key !.
Note: Functi ons of t he keys " and #respectively $and ! are exchanged. This allows the
user to operat e the keys with the gauge lying on the table as he is used to.
This menu item is left by pr essing the (backspace-)key $.
7.5.2 Wireless Interface
When the wireless i nterface is active the gauge can r eceive commands and inputs from the
QuaNix
interf ace (see § 8). Also updates of the software of gauge and probes can be loaded
through the i nter face (see § 10).
The active int erf ace is indicated by the symbol % in the upper right corner of the display and
a blinking red li ght on the keypad (also when the gauge is switched off).
When data are sent via t he wir eless i nterface the red LED of the keypad is fl ashing.
The interf ace can be swit ched off for power saving when it is not used.
®
-Software. E.g. t he gauge can be programmed for different applications via the
7.5.3 System-Information
The following information is displayed:
- Serial number of the gauge
- Software/Hardware-version of the gauge
- Memory in use (in %)
7.5.4 Language
The gauge can store 3 different languages. As a standard Engli sh and German are already
stored in the gauge. To download additional languages see § 9.
To change the language, call the list of languages and scrol l to the desired language with
the keys "or #, then press ! to confirm.
7.5.5 Buzzer
The buzzer can be switched on and off.
With the buzzer i n the off m ode the di splay shows the symbol .
7.5.6 Date / Time
The gauge has an internal clock so that with each reading and batch time and date can be
stored.
Setting of date and ti me can be required
- after a change of batteries, lasting longer than about 2 minutes
- after change between winter and summer ti m e
- change of time zone
7.5.6.1 Setting of Time
Setting of time is menu guided in the display. Values are m odif ied by using keys " and # .
Setting of hour and minutes are confirmed by pressing key !.
Correcti ons can be done by stepping back with key $.
7.5.6.2 Setting of Date
Setting of the date is done in the same way as that of the time.
8 Installation of the QuaNix®-Software
The installation of the QuaNix
into the PC.
9 Installation of the Drivers for the USB-RF-Dongle
The driver for the USB-RF-D ongle is stored on the same CD as the files for the QuaNix
Software.
When the dongle is connect e to the PC for the first time, a window pops up with the
message that a new hardware has been identified and the necessary driver needs to be
installed. The correct path for t he CD needs to be typed in or can be searched by cl icking on
the "Search" button. Then the installation can be started.
The installation procedure is onl y necessary when t he dongle is connected to the PC for the
fir s t time.
10 Setting of the Gauge Options via QuaNix
The settings descr ibed in § 6 can also be programmed and downloaded into the gauge
using the QuaNix
®
-Software. Especially the input of names for speci fic calibrations and
batches is done in a much easi er way with this tool.
The wireless interface dongle must be connected t o the PC and the QuaNix
needs to be start ed. In the gauge the wireless interface must be activated (see § 7.5. 2).
When the software has i dentified the gauge, all available menu options and their settings
are displayed in the software menu "Gauge Setti ngs" or can be downloaded from the gauge
by clicking on "Download Settings from the Gauge".
Now all settings of the gauge can be performed on the monitor and be rechecked. When all
selections have been done the m odifed settings will be transferred to the gauge by cl icking
on "Transfer Settings to Gauge".
11 Download of a New Language into the Gauge
A new language can be downloaded into the gauge using the "LanguageEditor" of the
QuaNix
®
-software.
The LanguageEditor pr ovides a table containing the complete set of commands and notes
that can be displayed.
12 Update of Software in Gauge and Probes
The software i n the gauge and the probes can be updated by the user. As a result of our
continuous product improvement updates of the software of gauge and probe are made
available to our cust omers on our homepage www.automation.de
After downloading the updates to the PC they can be loaded into the gauge or t he probe via
the wireless interface.
®
-software is automatically started after insertion of the CD
®
-Software
®
-software
.
®
-
12 13
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.