Alaris Imed Gemini PC-2 User manual

imeci
5.1
INTRODUCTION
This
section
instructions,
procedures,
and
reassembly
operational
GEMINI
Potentially
within
performance
PC-2º
the instrument AC
power.
for
maintenance recommended operated
Printed
damaged
removed
applied pads
can
metal
foil
that
type
therefore,
irons
and
should
replacing
on
integrated
be
cut
CMOS electrical damaged
activity
protected
ESD
protective
personnel
maintenance
preventive
mode
troubleshooting
procedures
test
Volumetric
Pump/Controller.
WARNING
lethal
PC-2
case
is
operated
When
using
the
action,
the
instrument
the
CAUTION
circuit
be
before
boards
when
and
replaced.
to
the
circuit cause and
base
is
essentially
only
vacuum
used
components
circuit
attempting
(PCBs)
integrated
board
de-lamination
material.
low-temperature
solder
when
removing
on
components
NOTE
devices
is
are
charges
during
not
environment
and
repair
performed
sensitive
procedures
grounding.
SECTION
maintenance
operating
routines,
and a comprehensive
for
when
using
case
it
Excessive
irreparable;
removal
PCBs.
unsoldering
may
if
using
disassembly
the
IMEDS
are
present
the
external
is
opened
is
be
are
traces
of
the
Damage
soldering
and
Leads
should
to
be
the
repair
in
an
including
easily
are
heat
and
of
tools
and
static
ESD
5 - MAINTENANCE
5.2
PREVENTIVE
The
GEMINI
with
the requirements. incorporates a diagnostic instrument's parameters. irregularities functional Alarms
notification. in
use
troubleshooting
Maintenance-free scheduled can on an interval once a year Verification responsibility such
the
graphs
forming
5.2.1
using
solutions. permit
or
the
non-volatile
by
biotechnical
be
‘as
tests
factory
Cleaning
Always
before
sterilize/
Do
solution.
any
cleanup
PC-2
goal
of
The
subsystems Detection
or
operation
Malfunction
Problems
preventive
required’
for
preventive
based
of
proper
of
and
at
nominal
general
unplug
cleaning.
autoclave
not
immerse
of
the
This
MAINTENANCE
is
minimizing
routine
and
of
operating
that
activates
alerts
of
RAM
error
this
personnel
and
repair
operation
maintenance
by
performing
basis.
The
maintenance
on
normal
operation
the
user,
verification
cost.
in
detail
the
maintenance
Instructions
CAUTION
the
AC
Do
the
the
of
the
PC-2
following
list
is
considered
of
all
and
assembled
maintenance
microprocessor
that
operating
system
affect
the
instrument's
audio
and
for
nature
log
between
At
the
may
The
procedures
on
power
not
instrument.
PC-2
are
for
subsequent
in
performing
regularly
inspections
routine
recommended
inspections
use
and
operation.
is
the
user’s
be
performed
following
the
PC-2.
cord
steam-
in
any
may
be
cleaned
recommended
contaminates.
the
visual
recorded
option,
at
para-
for
per-
to
is
5-1
UPDG3
тес]
Isopropyl
Warm
Household
household
These
free swab. moved, a cloth be and
Then
completely moistened
water dried
solutions
cloth; a soft
Once
used
the
rinses
with a soft,
Prior the
of
the
electrolyte female
contamination;
the
alcohol
soapy
water
Bleach
bleach
may
bristle
the
contamination
soaked
to
rinse
the
all
of
the
entire
instrument
rinsed
with
using
fresh
the
instrument
lint
to
reattaching
instrument,
power
and
power
cord.
(10%
to 9 parts
be
applied
brush
with
entire
another
water.
free
cloth.
WARNING
the
ensure
input
module
dry
thoroughly.
on
the
if
contaminated,
solution,
and/or a cotton
has
fresh
instrument
surface
Following
must
AC
that
power
i.e. 1 part
water)
using a soft,
been
water
should
cloth
thoroughly
the
be
thoroughly
power
the
is
clean
Check
cord
lint
re-
should
solution.
be
fresh
cord
to
male base
of
any
the
for
replace
utilizing (sys (reading
the in the read instruments
should operating to check.
should replaced.
5.2.3.2
1.
2.
the
batt)
test
will
instrument
sestion
5.5
battery
6.5
volts
be
charged
condition,
stabilize
Batteries
be
subjected
Lithium
Turn
instrument
cord
from
Open
the
A/D
terminals.
primarily
for 4 hours
5.5.1).
3.
Connect a digital (TP3)
and
Board.
Voltage
to
check
be 0.5 true
disassembly
and
display’s
the
battery
voltage),
procedures
connecting a voltmeter
Battery
+0.3
volts.
Batteries
operated
for
12
hours
then
unplugged
prior
to
testing
Battery
to
further
off
less
Check
and
unplug
outlet.
instrument
Vaay
case
voltmeter
(U8,
pin
system
voltage
or
by
voltage
installed
on
battery
in a non-
and
performing
than
62
testing
AC
(see
section
between
16)
on
the
battery
following
described
across
should
in
power
allowed
voltage
volts
or
be
power
ground
Analog
5.2.2
Perform section
checks:
+
Mechanical
the
mechanical
2.2
of
this
Inspect
excessive or
damaged
Perform a pumping spring roller the functioning
Replace
5.2.3
Perform
described
5.2.3.1
Perform a battery entering for
Maintenance
the
wear
1000
hours
or
integrity
clamp
Pump
mode.
the
damaged
Electrical
the
in
section
Battery
the
maintenance
Inspection
manual
and/or
of
instrument
worn
replace
test
while
the
If
channel
or
Inspection
electrical
2.3.2
Voltage
voltage
Mode
inspection
plus
the
pumping
holes
the
mechanism
by
closing
pump
only
one
will
missing
inspections
of
this
Check
check
mode
operating
described
following
seal
for
every
90
operating
seal.
extension
the
IV
is
operating
spring
not
by
(see
procedures)
is
spring.
manual.
either
section
in
days time.
set
in
5.3 and
4.
Disconnect Power
5.
After
display
the
meter
Volts.
If
6.
Reconnect POWER
7.
If
the lithium should
5.3
MAINTENANCE/DIAGNOSTIC
The
PC-2
Diagnostic
access
the
the
Maintenance/Diagnostic
biotechnician to
verify
operation
features, independent check access
keypad,
system
the
and
communications
Supply
the
audio
to
stabilize
ON
voltage
battery
be
replaced.
is
configured
Mode
software
can use
to
operate
of
Analog
error
to
the
battery
Board.
initiates,
(~30
Reading
the
battery
control
reading
that
to
is
(B1)
on
with a Maintenance/
allows
diagnostic
the
diagnostic
of
the
operator
the
pumping
the
alarm/malfunction
to
Digital
log,
to
test
check
port.
the
input
harness
allow
at
the
seconds},
should
harness
silence
<2.25
the
and press
the
Volts,
Logic
Board
MODE
biotechnicians
subsystem.
Mode
is
test
interface
mechanisms
interrupts,
voltages,
the
lamps
port
and
J6
on
the
voltmeter
then
note
be
>2.25
audio.
the
to
Once
the
routines
to
and
serial
the
to
UPDD3
5-2
Integer
The
selection
parameters.
Keypad/Series
Maintenance
and
to
2.xx
Mode
is
set serial
Software
communications
for
1.
Unplug recommended
be possible).
the
performed
AC
power
that
maintenance
using
battery
imed
cord
(it
is
power
operations
Fractional
The
Setup
and
to
5.3.1
Keypad/Series Mode
set
communications
Maintenance/Diagnostic
Sequence
Integer
The
the initialization:
Keypad/Series
Maintenance
following
order
TIMEBASE
LAMP
TEST KEYPAD ERROR MOTOR PUMP
TEST SERIAL A/D
VOLTAGE INPUT COMMUNICATIONS
LANGUAGE POWERDOWN
Fractional
The
following
Keypad/Series
Diagnostic
initialization:
ERROR PUMP MOTOR A/D
VOLTAGE
INPUT LAMP
TEST
KEYPAD
TIMEBASE
SERIAL
ROM
CRC DISPLAY
POWERDOWN
5.3.2
Maintenance/Diagnostic
is
Mode
TEST
LOG
HOMING
PORT
PORT
Mode
LOG
TEST
HOMING
PORT
TEST
PORT
3.xx
utilized
2.xx
test
after
Maintenance
CHECK
DISPLAY
TEST
TESTS
DISPLAY
DISPLAY
SELECTION
TEST
test
routines
after
diagnostic
DISPLAY
TEST
DISPLAY
DISPLAY
CHECK
TESTS
TEST
Software
for
language
parameters.
Mode
Software
SETUP
3.xx
appear
mode
Mode
selection
Test
appears
Mode
in
Operation
in
the
2.
Install
Call/Maintenance
Press MONITOR
3.
Press
Integer
+
Release
the
Maintenance
Release
and
hold switch.
the
POWER
All
LEDs
and
seconds
then
Keypad/Series
"maintenance
on
the
Channel B Operator
display
(2.xx
version).
Fractional +
"diagnostics
Keypad/Series
<stop>" - scrolls channel B Operator (v3.xx = installed
4.
The
PC-2
is
now
Maintenance/Diagnostic
Software
The
to next
5.3.3
The
both
Where
Fractional
Maintenance
following
the
Release
maintenance
change
v2.xx
differences
both
and
during
procedural
and
keypad
procedures
from
V2.13
Plug
in
Plug
connector
V2.33
the
and
COMPUTER
ON
control
Subsequent
displays - illuminate
extinguish.
2.xx
Software
v2.xx" - scrolls
is
the
installed
3.xx
PC2
v3.xx
to
continuously
Information
software
initialized
version).
in
the
Mode.
NOTE
V2.13
plug
must
be
one
test
routine
the
Mode
steps
v3.xx
software
occur
between
controls
are
motor
used
homing
Test
Suite
are
versions.
the
to
described.
select a test
the
Nurse
(rear
panel).
CONTROL/
and check:
for
continuously
Information
software
exit
press
on
the
display
installed
to
the
test.
applicable
Integer
and
3
to
Performing
will
diagnostic
Prior
to
operating
Maintenance
instrument
the
provide
access
mode
WARNING
the
Mode
is
not
connected
following
to
the
tests:
PC-2
in
ensure
that
procedures
PC-2's
maintenance/
the
the
to a patient.
as
described
Software The
used
maintenance
Software The
used
maintenance
5-3
Release
press
function
to
Release
COMPUTER
to
sequence
V2.13
of
the
through
mode
tests.
V2.33 and
CONTROUMONITOR
through
mode
tests.
Audio
the
individual
Subsequent
the
individual
Control
switch
switch
UPDO3
is
is
im
Each
confirmation
pressed
scrolled displayed
select
test
deselect
in
test
Integer
Actuation identifier
Fractional
Diagnostic Diagnostic
scrolling
TIMEBASE
1.
Press
and »
2.
Integer
Fractional
test
routine
is
identified
message.
to
initiate a specific
test
identifier
test
phase
control
the
the
Maintenance
may
current
test
Keypad/Series
of
the
is
or
PAUSE/STOP
scrolling
and
Keypad/Series
mode
mode
via a powerdown
will
powerdown
can or
powerdown
CHECK
the
appropriate
check:
"timebase Information
Press
Keypad/Series
*
"running" - displays
«
Channel B VTBI
check" - scrolls
display.
START
and
between 0 and "failed"
Information
clear
allow
*
"swing" - displays
+
numerical
channel B VTBI windows.
-
-
-
will
display
display.
the
Operator
the
test
Keypad/Series
values
Press
PAUSE/STOP
display
+
*
If expected and
Rate
for
data
"(halted)
channel A Operator press
the
displays
PAUSE/STOP
‘swing’
deviation, a failure
"FAIL"
CLEAR
resume.
by a scrolled
Once
the
START
test
is
replaced
identifier.
be
used
and
Mode
2.xx
interrupt
the
only
be
test
by a statically
The
appropriate
at
any
proceed
test
sequence.
Software
key
while a test
the
instrument.
3.xx
exited
test
in
the
Pump
select
on
check:
2.xx
statically
display - alternates
1.
If
displayed
on
the
Operator
Pressing
CLEAR
Information
to
resume.
3.xx
statically
sequence
display's
recording.
tenths
once
rapidly
to
statically
Information
again
value
exceeds
appears
in
will
the
allow
channel
key
is
the
time
to
to
the
next
test
when
the
identifier
Test
control
the
value
once
Operator
is
will
display
and
in
and
the
units
freeze
in
the
display
to
unfreeze.
the
is
invoked
B
the
test
to
are
22,
the
Press the between user between user
difference readings
modes.
LAMP
1.
Press
and check: «
"lamp
Information
Press
2.
+
Rate
sequentially 9 9
(The
display a decimal position;
Indicator
+
"ου,
*
Controller
Secondary
Traffic
illuminate test
+
Audio
KEYPAD
1.
Press and *
Press
+
Press
or
and
active
corresponding
information
either
display
to
old
keypress),
old
keypress)
between
updated
TEST
(with
the
appropriate
test" - scrolls
START
and
VTBI
followed
channel B VTBI
instead
will
Information
in
"*,
"and
and
cycle
alert - sounds
TEST
the
appropriate
check:
“keypad
Information
"start”
Information
test" - scrolls
START
each
no
COMPUTER
Subsequent)
test)
on
the © or © controls
the
"max"
and
new
timer
reading
"min"
and
new
timer
or
"dynamic"
old
every
AUDIO)
test
and
check:
displays - Flash
numbers
by
1.1.1.1.
point
the
illuminate)
upper
case,
and
10
on
1 1
Display
Battery
displays - scroll
difference
reading
new
msec)
seleot control
1 1
through
will
in
the
Operation
the
numbers
"?"
and
Pump
(Piggyback),
Battery
display
and
display
key
response,
indicator
during
check:
statically
(except
delivery
once
test
on
POWER
Audio
mode,
Communication
LEDs
one
half
per
select control
Operator
on
Control
CONTROUMONITOR which
one
nomenclature
the
channel B Operator
always
at a time
terminates
and
verify
for
to
taggle
difference
since
last
since
last
timer
data
once
through 9 9
9.9.9.9.
not
Units
the
"0"
of
the
second.
once
ON
which
(V2.13)
(V2.34
the
that the
the
key
-
UPDO3
5-4
ERROR
1.
Press
and
*
2.
Press +
«
3.
Press log:
+
4.
Press
+
LOG
DISPLAY
the
appropriate
check:
“error
log
display" - scrolls
information
START
“empty”
(no
display
and
check:
errors channel B Operator this
case
proceed
mode
test.
"newest"
"nn: the
(nn = number
=
error
[v2.xx]
Next format
followed
ccc"
[v3.xx]
channel B Operator
identification
or
00
Units
(1)
oldest
or
"no
CLEAR
Error
log
clears - "wait"
on
the
a
one
second
either
“okay”
to
or
by
will
of
entry
to
129
digit
key
entry
will
older“
key
and
Information
verification
or
"fail".
test
select
on
logged)
appears
Information
the
next
"nn:
cc”
display
statically
Information
from
00
code - 00
[v3.xx]).
to
read
display
will
appear.
check:
displays
display
stage
control
on
once
the
display.
maintenance
[v2.xx]
the
in
to
to
"nn:
or
display
09,
99
error
cc"
on
οσ(ο)
statically
during
In
by
5.
Press
«
Pumping
‘sync’ and Operator
»
Pumping
“failed”
Information
Software
2.
Press +
Channel A VTBI the
"homestep" channel A Operator
3.
Use (between 0 and
>199
4.
Press +
Pumping selected display
the
*
Pumping
channel A VTBI
numerical step
START
“homing”
and
mechanism
position
displays
check:
(number 7 finger
Information
mechanism
displays
Release
START
statically
V2.33
and
check:
and
displays
last
selected
home
displays
Information
the
Data
Entry
199)
for
will
revert
display
START
and
check:
mechanism
step,
"-
- -
-"
and
"homing"
channel A Operator
mechanism
display
value
of
the
imed
statically
display.
stops and
step
statically
homing. A selection
to
operates
appears
displays
stops
to
seek
extended)
in
the
"homed"
in
the
Operator
Subsequent
either a "0"
and
in
the
display.
to
select a step
"0".
to
seek
in
the
statically
Information
and
the
shaws
the
homing
display.
the
or
or
the
VTBI
in
MOTOR
Software
1.
2.
3.
4.
HOMING
Press
the
and
check:
«
“motor
Operator
Release
Press
START
+
“chan
A*
information
Press
START
+
Pumping ‘sync’
and
"homing"
Operator
+
Pumping
“failed”
Informatian
Press
START
*
"chan
B
Operator
TEST
appropriate
homing
test" - scrolls
Information
V2.13
and
displays
and
mechanism
(number 7 finger
displays Information mechanism
displays
display.
and
B"
displays
Information
test
select
display.
check:
statically
check:
statically
display.
stops
statically
in
check:
statically
display.
control
on
in
the
to
and
the
in
the
once
channel
Operator
seek
the
extended)
in
the
“homed” Operator
Channel
B
or
+
"homed"
channel A Operator
«
Press channel
#4.
PUMP
1.
TEST
Press
and
check:
*
"pump
the
information
Software
2.
Press +
The
Release
START
upper CONTROLLER Rate
indicator
*
"PG-2
channels
or
"failed"
PAUSE/STOP
B,
then
repeat
appropriate
test"
scrolls
display.
V2.13
and
position
Delivery
illuminates
Vx.xx"
scrolls
Information
displays
Information
once
steps
test
select
on
the
check:
of
either
Mode/Retative
once
display
statically
display
to
select #2
through
control
Operator
the
PUMP
for
each
across
of
both
in
once
channel
the
UPDO3
the
or
imeci
*
"ACCESS
the
+
“MAINTENANCE” the
channels.
Use
the
section
abbreviated Operators
operate
The
instrument
independent
2.
Press +
"normal"
A
Use
the
select test
mode.
step
between
mode
channel B Operator
Press
operation.
+
"MAINTENANCE"
both
*
Channel A Rate and "о"
*
Last
mode/Operating
Use
the
3.3.1 instrument.
When
ACCESS
operation.
When
+
Calculated threshold
is
displayed
CHANNEL"
Operator
Operator
3.3.1.1
Information
Information
procedures
of
procedures
Reference
the
instrument.
will
of
any
Release
START
V2.33
and
displays
Information
Units
(1)
either
the
‘normal’
Each
normal
selected
START
selected
procedures
to
set
channel A is
will
to
Information
Pump
up
channel A and
CHANNEL
in
the
Pressure
START
occlusion
for
selected
in
scrolls
scrolls
this
manual
described
Guide
NOTE
continue alarm
and
Subsequent
check:
statically
digit
Data
or
actuation
and
pressure.
display
Information
enable
the
scrolls
VTBI
or
Controller
indicator
described
operating,
B
then
mode
in
step
45
pressure
infusion
the
RATE
once
across
display
of
both
continuously display
of
described
or
the
in
to
set
up
both
the
and
operating
conditions.
on
the
channel
Entry
Control
‘pressure’
will
alternately
pump
The
statically
on
display.
PC-1
for
continuously
displays
show
delivery
will
illuminate.
in
Section
operate
press
setup
for
and
after
above,
check
voltage
parameters
display
on
in
the
on
the
to
the
»
Voltage
Strain
Any
time
while
in
and
VTBI
VTBI
values.
Three
the tarily then a
mode
the
To
press
switch
routine
motor
pumping
to
resumes
one
time
presentation
infusion
change
COMPUTER
once.
back
communication
the
communications
Fractional
2.
Press
-
3.
Use
the
4.
Press
Keypad/Series
START
A
Operator
the
"normal"
START operation. +
"MAINTENANCE" both
Operator
e
Channel A Rate and
+
Operating selected
5.
Follow
3.3.1
operate
6.
When
*
the
to
the
in
Calculated
threshold is
displayed
*
Voltage
Strain
relative
Beam
NOTES
the
pumping
the
pressure
display
revolutions
mechanism
check
occlusion
operation.
interruption
parameters.
pump
This
to
step
cable
and
Information
or © control
and
"pressure"
to
enable
Information
indicator
delivery
procedures
set
up
channel(s) A and/or B and
instrument.
the
Pressure
START
Beam
in
occlusion
for
selected
in
the
displays
to
pressure
mode,
revert
Rate
sensed
in
the
VTBI
mechanism
the
to
displaying
and
decremented
following
stops
pressure
This
of
the
with a flash
test
mode
selection,
CONTROL/MONITOR
will
loop
the
#3
providing
is
NOT
data
port
3.xx
Software
check:
staticaily
on
display.
to
toggle
options.
the
PC-2
scrolls
continuously
displays
VTBI
displays
illuminates
mode.
described
mode
and
step
#5
above,
infusion
RATE
to
in
the
display
sensed
VTBI
at
the
stops
RATE
the
START,
momen-
and
will
result
pressure
display
test
a
connected
(RS-232C).
the
channel
between
for
on
show
for
last
in
Section
after
check
voltage
parameters
at
display.
in
of
to
the
the
UPDO3
5-6
Any
time while and
VTBI selected VTBI
Three
the
pumping tarily then
a
one
mode the
infusion
To
change press switch
routine
communication
the
communications
SERIAL
1.
PORT
Press check: +
“serial Information
Install
may
Press
«
"echo"
Information
Press
*
Each
communications
a ? (e.g. A
*
"okay" display
3
seconds;
the
To
rerun
loop
Press
+
"send"
Information
NOTES
the
pumping
in
the
pressure
display
infusion
revert Rate
values.
motor
revolutions
mechanism
to
check
resumes
time
presentation
occlusion
operation.
interruption
with a flash
parameters.
pump
COMPUTER
once.
back
test
CONTROL/MONITOR
This
will
to
step
cable
TESTS
the
appropriate
port
tests”
scrolls
Communications
connector
also
be
performed
vice a communications
START
displays
on
and
check:
statically
display.
START
alpha,
appears
if
each
and
check:
numeric
character
?)
in
in
byte
sequence.
the
otherwise
channel B VTB!
the
test,
press
the
test
PAUSE/STOP
and
statically
mechanism
mode,
the
to
displaying
and
decremented
following
stops
momen-
pressure
This
will
of
the
display
mode
#3
is
data
test
selection,
loop
the
providing
NOT
port
select
on
the
the
rear
test
connected
(RS-232-C).
control
plug
panel.
using a remote
plug.
on the
and
symbol
set
is
sent
is
echoed
“FAIL”
appears
display.
CLEAR.
back
to
step 3 above.
check:
on
the
stops
RATE
the
START,
and
result
in
of
a
to
and
in
the
Test
Operator
in
the
sent
with
Information
within
in
This
will
ni
Press
+
Byte selected “okay”
display
display.
8.
To
loop
9.
Press +
"receive" Information
+
display.
10.
(To
be C2
PC-2).
A/D
VOLTAGE
1.
Press
and
+
"A/D
Operator
2.
Press +
"A
B
*
Channel À strain
the
Integer
3.
Press «
"B Channel B Operator
+
Channel B strain the
R
Press
+
"sys
Channel B Operator
+
System
Channel B VTBI
m
Press +
"Vímains)" Channel B Operator
+ A numerical
Channel B VTBI
display.
START
stream
rerun
the
and
is
baud
appears
or
"FAIL"
the
test,
test
PAUSE/STOP
displays
display.
START
conduct
connected
the
to a computer
Communication
DISPLAY
the
appropriate
check:
voltage
Information
START
strain"
channel B VTBI
Keypad/Series
and
displays
Information
PAUSE/STOP
strain“
displays
Channel B VTBI
PAUSE/STOP
batt"
displays
battery
PAUSE/STOP
displays
value
otherwise
check:
sent
out
rate.
Upon
on the
appears
press
CLEAR.
back
and
check:
statically
causes
"receive"
protocol
test
scrolls
display.
check:
statically
beam
2.xx
and
statically
Information
beam
display.
and
statically
Information
voltage + 2
and
statically
Information
(=2.50)
display
approximately
и
imecl
at
the
currently
completion
Information
in
the
VTBI
This
will
to
step 6 above.
on
the
Operator
"input
?" to
test
the
PC-2
must
terminal
or
another
select
on
on
display
voltage
check"
on
voltage
check:
on
displays
check:
appears
if
AC
control
the
the
displays
the
display
displays
the
display
on
the
display
power
00.01
using
once
channel
on
on
the
is
will
in
in
the
UPD03
тес]
AC
a
quantitative
for
6.
Press
+
"V
B
+
A/D
(=2.50) VTBI
voltage
PAUSE/STOP
(ref)"
Operator
converter
is
display.
МОТЕ
measurement
evaluation,
or
absence
and
statically
Information
reference
displayed
on
(V
mains)
but
rather a test
of
AC
check:
on
the
display
voltage + 2
the
Channel
is
not
Channel
B
7.
Press
(O)
+
"V(audio)” displays
Operator
»
8.
Press
+
"V(NiCad)" Operator
+
NiCad
channel B VTBI
acceptable
and
check:
Information
00.00
(Random
if
keypad
and
check:
displays
Information
battery
voltage + 2
reading
statically
display
will
display
numerical
statically
display
display
2.71).
on
as no display
are
pressed).
of
displays
the
audio
the
may
on
is
the
7.
Press
+
«
I
Fractional
3.
Press
»
+
4.
Press
«
*
5.
Press and
»
> A numerical
AC a
for
6.
Press
«
+
PAUSE/STOP
“V(audio)"
Operator
occur,
"B
strain"
Information
0000
will
(Random
if
keypad
Keypad/Series
the
[0 control
displays
numerical
Channel B Operator
Channel B strain
the
Channel 8 VTBI
(O)
and
"sys
batt"
displays
B
Operator
System
channel B VTBI
battery
"V(mains)"
channel B Operator
channel B VTBI present;
otherwise
beam
check:
Information
voltage
check:
displays
value
display
display.
NOTE
voltage
quantitative
presence
"V
(ref)"
B
A/D
(=2.50)
VTBI
measurement
(0)
and
converter
is
displayed
display
evaluation,
or
absence
check:
statically
Information
reference
and
check:
statically
on
display
display
3.xx
as no
display
are
pressed).
and check:
statically
Information
display.
statically
statically
Information
(=2.50)
approximately
on
on
voltage
on display + 2
displays
on
appears
if
AC
power
(V
mains)
but
rather a test
of
AC
on
the
display
voltage + 2
the
channel
the
audio
may
the
display
the
channel
on
the
display
on
is
00.01
is
voltage.
channel
B
is
in
the
the
will
not
INPUT
1.
2.
RATE 1000
γι
B
integer
3.
PORT
Press
the
and
check:
»
“input
Information
The
RATE
during
for
Press
»
+
BI
this
both
START
Information Each
units
digit
either a "0"
as
described
100
10
1
- git
3
Keypad/Series
Press
+
+
PAUSE/STOP
"selftest"
Information Each units either a "0"
as
described
DISPLAY
appropriate
port
test
scrolls
NOTE
and
VTBI
display
test
have
the
channels.
and
check:
digit
in
of
Digit = 0 |
=}
Digit = G
=1
Digit = 0
-3
Ogit
= O
=1
=
Digit = 0 | ga | Door
=
digit
in
digit
of
statically
display
the
RATE
the
VTBI
or
"1"
to
in
the
sco
Eo
|MOTOR/ | Sensor
SYNC | Sensor
2.xx
statically
display
the
RATE
the
VTBI
or
"1"
to
in
the
display
display
following
[Fluid
Air
Sensor
MALFUNCTION
Air Fluid
Daor
Software
and
check:
display
following
select control
on the
presentations
same
on
meaning
the
and
will
present
sensor
the
table:
in
sensor
in
sensor
in
sensor
.
in
Open Closed
on
display and
sensor
sees
Opoque
sces
tight
the
will
sensor
table:
for
the
present
once
output
lest
status
UPDO3
5-8
Fractional
3.
Press © and *
*
RATE - 1000
VTBI Y
E
Integer COMMUNICATIONS
1.
2.
3.
4.
Keypad/Series
"test"
displays information Each
digit units
digit
either a "0" as
described
Digit
100
Digit - 0
10
Digit
Y
Digit = O
digit - 0 | ooog | Sensor
Keypad/Series
Press
appropriate check: *
"communications Operator
Press
START
-
“serial Operator
+
"0000" will display.
Use number.
Press
*
"baudrate"
+
Currently 600, display
#"
or
display
Data
STOP
1200,
Entry
on
3.xx
check:
statically
on
the
Operator
display
in
the
RATE
of
the
VTBI
or
"1"
in
the
= 0
=
Information
Information
дк
seo. | Slide
-1 = 0
-4
=!
currently
eco | Sensor
SYNC | Sensor
1
SETUP
test
and
check:
on
the
and
check:
displays
Information
selected
2400,
the
Channel B VTBI
display
display
to
following
[Sensor
Sensor
Slide
Sensor
Sensor
Sansor
2.xx
select
setup"
statically
display
installed
Channel B VTBI
to
statically
display
baudrate
4800
sensor
table:
МоНипейол
Operating
Clamp Clamp
Connected
nol sees
sees
Opercting Malfunction
control
scrolls
on
serial
change
on
(110, 300,
or
9600)
will
and
present
In Out
Connected
Opaque
Light
ONLY
and
on
the
the
number
the
the
will
display.
the
status
serial
7.
Use
Units
through
8.
Press
+
"stopbits”
Operator
»
Currently
‘1.5’ VTBI
9.
Use
Units
through
Release
LANGUAGE
1.
Press
check:
and
*
"language
Operator
2.
Press
*
"currently statically display.
3.
Press
+
Operator
Subsequent
the
language
and
back
instruments
back
to
"english"
instruments.
Fractional
ROM
CRC
1.
Press and
check:
+
"ROM
(1)
digit
and
select
STOP
and
check:
Information
selected
ог
'2")
(1)
digit
and
select
V2.3x
SELECTION
the
appropriate
selection"
Information
START
the
and
on
the
one
(1)
Information
presses
of
selected
to
"english"
and
on
"italian",
on
the
Keypad/Series
DISPLAY
the
appropriate
CRC
display”
Information
Data
Entry
desired
statically
display
stop
bits
on
the
Data
Entry
desired
test
scrolis
display.
check:
language"
control
statically
display.
the
(1)
to:
"french",
the
"a"
then
"b"
version
3.xx
test
scrolls
imed
Control
parity
on
option
Channel
Control
select control
on
Information
once
on
control
then
version
"swedish",
Software
select
on the
to
option.
the
(‘1',
B
to
option.
the
and
check:
the
will
change
"spanish",
EPROM
then
EPROM
ONLY
control
step
step
once
once
2.
Press
5.
Use
Units
(1)
digit
Data
Entry
Control
through
6.
Press
*
*
and
STOP
information Currently 'E'(even) B
VTBI
or
select
and
check:
display
‘OFF’
desired
statically
baudrate.
parity
option
on
the
on
to
'O'(odd),
the
Channel
step
START
«
"m
xxxxx" information identifier reference
Engineers
and
display
and
xxxxx
for
use
ONLY).
check:
in
the
(n = ROM
= a
by
IMED
block
hexadecimal
40003
imeci
POWERDOWN
1.
Press
the
and
check:
*
"powerdown
Information
2.
Press
START
+
"POWER"
across
Information
seconds
+
One
shuts
The
maintenance
prior
to
service
TEST
appropriate
test”
test
select control
scrolls
display.
and
check:
"OFF
n"
the
channel A and B Operator
displays
('n'
counts
from 5 to 1 [v2.xx]
second
down.
aíter
'n' = 1
WARNING
plug
must
returning
for
patient
the
instrument
care.
on
the
statically
or 3 to
the
instrument
be
once
Operator
down
in
1
removed
to
5.4
TROUBLESHOOTING
The
troubleshooting
Table
5-1
are
maintenance section procedure test
that
if a control
Keypad Lamp
test; check The order
the
corrective
of
corrective
should
the
instrument
If
the
test
repair a microprocessor
your
facility,
returned
mode
5.3.
The
is
to
replicates
key
test;
if a LED
if
the
failure
actions
reduce
equipment
it
to
the
routines
correlated
test
recommended
perform
the
is
not
instrument
causes
actions
probability.
in
repair
to
patient
is
recommended
factory
presented
the
functioning - run the
segment
are
the
sequence
time
care
system
for
to
described
troubleshooting
Maintenance
discrepancy;
is
out - run
fails
to
under
Initialization.
listed
in a descending
Performing
and
expedite
to
troubleshoot
is
the
repair.
the
provided
not
instrument
in
the
in
Mode
e.g.
the
-
the
returning
and
available
at
be
UPDO3
5-10
imed
Table
5-1.
Troubleshooting/Fault
Isolation
Guide
LED
Alarm
Test/Fault
display
tone
INITIALIZATION
No
No
Battery
Battery
F1
POWER
Transistor
Transistor
Lithium/NiCad
Digital
Audio
Audio
Analog
<5.58
harness
on
Power
ON
Logic
Harness
Oscillator
Board
Harness
Lead
Cause
Volts
fuse
blown
Supply
switch
Battery
Failure
Failure
PCB
inoperative
Unplugged
Disconnected
Failure
Disconnected
Failure
(v2.xx
blown
ONLY)
Corrective
Connect
Replace
Replace
Check
Test/Replace
Connect
Repair
Replace
Replace
Connect
Replace
Replace
AC
fuse
fuse
Keypad
Transistor
Broken
Lithium/NiCad
Digital
Audio
Audio
Analog
Action
Power
Cable
Keypad
Harness
Wire
Logic
Harness
Oscillator
Board
Connector
Battery
Board
LEDs
VERSION
‘maintenance
vice
'PC-2 ‘maintenance
Model/Version
Model/Version corrupted
stay
'PC-2
V2.13'
ON
DISPLAY
V2.13'
V2.13'
vice
V2.13°
fails
display
to
scroll
Digital
Maintenance
Maintenance
Maintenance
Digital
Display
Power
Digital
Display
Logic
Logic
Board
Supply
Logic
Board
Failure
Plug
Plug
Piug
Failure
Failure
Installed
not
Faulty
Failure
Board
Failure
5-11
Installed
Failure
Replace
Remove
install
Replace
Replace
Replace
Replace
Replace
Replace
Digital
Maintenance
Maintenance
Maintenance
Digital
Display
Power
Digital
Display
Logic
Logic
Supply
Logic
Board
Plug
Plug
Plug
Board
Board
Board
Board
Board
imed
TIMEBASE
Test
CHECK
fails
to
step
Comm.
Harness
Disconnected
Reconnect
Comm.
Harness
Test
Test
displays
‘failed’
No
test
LAMP
TEST
LED illuminate
No
Audio
KEYPAD
fails
to
display
segment
adjust
TEST
start
=2',
w/AUDIO
fails
&
to
ACCESS
Digital
Keypad
Keypad
Digital
Crystal
Digital
Display
Display
Digital
Audio
Switch
Logic
Failure
Failure
Interface
Logic
Failure
and
Oscillator
Logic
Failure
Board
Board
Logic
Failure
Control
Failure
Failure
Failure
Failure
Switch
out
of
Failure
sync.
Replace
Replace
Test/Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
Replace
ACCESS
Digital
Display
Digital
Digital
Digital
Display
Display
Digital
Audio
switch
Logic
Keypad
Board
Logic
Logic
Logic
Board
Board
Logic
Control
Board
Board
Board
Board
Board
Switch
Key/Display
Invalid
ERROR
MOTOR
Motor
Homes
PUMP
Pump
OCCLUDED-FLUID
Key
LOG
HOMING
Fails
to
TEST
Stops:
DISPLAY
TEST
to
home
other
than
or
(See
SIDE
Decoder
Table
Motor
Digital
Analog
Motion
Disconnected
Motion
Digital
Transducer
Transducer
Failure
5-2
for a listing
Harness
Logic
Board
Sensor
Sensor
Logic
and
Disconnected
Failure
Failure
Harness
Failure
Failure
Harness
Failure
description
(v2.xx
Disconnected
ONLY)
Replace
of
Error
Reconnect
Replace
Replace
Reconnect
Replace
Replace
Replace
Transducer
Digital
Log
Codes)
Digital
Analog
Motion
Digital
Transducer
Logic
Motor
Logic
Motion
Logic
Harness
Board
Harness
Board
Board
Sensor
Sensor
Board
Harness
Reconnect
u
imeci
HELP
HELP
SERIAL
RATE
ACCURACY
INTERNAL
PORT
TEST
ERROR
Digital
Analog
Motion
Motion
Motor
Stepper
Analog
Logic
Board
Sensor
Disconnected
Sensor
Harness
Motor
Digital
AIL/SCD-Door
AIL/SCD
Door
Digital
Logic
Board
Disconnected
Sensor
Sensor
Logic
Board
Failure
Failure
Harness
Failure
Disconnected
Failure
Board
Board
Failure
Failure
Sensor
Failure
Failure
Failure
(v2.xx
(v2.xx
Harness
ONLY)
ONLY)
Replace
Replace
Reconnect
Replace
Reconnect
Replace
Replace
Replace
Reconnect
Replace
Replace
Replace
Digital
Analog
Logic
Motion
Motion
Stepper
Digital
Analog
AIL/SCD
Door
Digital
Sensor
Motor
Logic
AIVSCD-Door
Sensor
Logic
Board
Board
Sensor
Harness
Motor
Board
Sensor
Harness
Board
Harness
Board
"echo"
AID
|
i
>'0’,
"A
<1.0
pumping
test
fail
VOLTAGE
or B strain"
set
or B strain"
or
DISPLAY
not
installed
>2.0
segment
reading
reading
with
dy
installed
Faulty
Digital
Power
Analog
Strain
Strain
Digital
Analog
Strain
Strain
Analog
Communication
Logic
Supply
Beam
Beam
Logic
Beam
Beam
Digital
Logic
Board
Board
Board
Board
Board
Failure
Out
of
Failure
Board
Failure
Out
of
Failure
Board
Failure
Plug
Failure
Failure
(v2.xx
Calibration
Failure
(v2.xx
Calibration
Failure
(v2.xx
ONLY)
ONLY)
ONLY)
Replace
Replace
Replace
Replace
Recalibrate Section
Replace
Replace
Replace
Replace Recalibrate
Replace
Replace
Comm
Digital
Power
Analog
Strain
5.7)
Strain
Digital
Analog
Strain
Strain
Digital
Analog
Emulator
Logic
Board
Supply
Board
Beam
Beam
Logic
Board
Board
Beam
Beam
Logic
Board
Board
Plug
Board
(see
Loading...
+ 29 hidden pages