Alaris GW User manual

Alaris
®
GW Volumetric Pump
Technical Service Manual
Cardinal Health cannot accept any liability for any breakdown or deterioration in performance of parts or equipment
resulting from unauthorised repair or modification.
t Cardinal Health, 1180 Rolle, Switzerland
Alaris® and Asena® are registered trademarks of Cardinal Health, Inc. or one of its subsidiaries.
All other trademarks belong to their respective owners.
© 2003-2006 Cardinal Health, Inc. or one of its subsidiaries. All rights reserved.
Alaris® GW Volumetric Pump 2/77 1000SM00006 Issue 4
Contents
Chapter
1.
2.
3.
4.
5.
6.
Appendix
Introduction & Start up 4
Configuration & Calibration 10
Routine Maintenance 15
Troubleshooting 24
Circuit Descriptions 28
Spare Parts Replacement Procedures 32
Specifications 5
A.
Disposal 6
B.
Spare Parts Listing 6
C.
Fitting and Replacement Guidelines 6
D.
Compatible IV Infusion Sets 6
E.
Configuration Records 7
F.
Service Contacts 7
G.
Document History 7
H.
2
0
2
6
9
2
4
6
Alaris® GW Volumetric Pump 3/77 1000SM00006 Issue 4

Introduction & Start up

In this chapter
Introduction 5
General precautions 5
Chapter 1
Views of the Alaris® GW Volumetric Pump 6
Controls and indicators 7
Operating the Alaris® GW Volumetric Pump 8
Starting the Alaris® GW Volumetric Pump 9
Features of the Alaris® GW Volumetric Pump 9
Introduction & Start up

Introduction

The Alaris® GW Volumetric Pump is designed to deliver a continuous and accurate infusion. High performance, comprehensive alarm protection and sophisticated monitoring systems, combined with simple operation, make this pump ideal for general care and critical care in a variety of areas within a hospital.
The Asena® brand name has been recently changed to the Alaris® brand name. This change in brand name has no effect on the intended use or functionality of the product. Recommended disposable products for use with this product may refer to either the Asena® brand name or Alaris® brand name and both types are suitable for use with this infusion pump.
Familiarity
Before operation, ensure that you are fully familiar with this pump by carefully studying the Directions for Use (DFU) prior to attempting any repairs or servicing.
As part of a policy of continuous improvement, product enhancements and changes are introduced from time to time.
Purpose
This Technical Service Manual describes how to set up, test and maintain the Alaris® GW Volumetric Pump. It is intended for use by personnel experienced in medical equipment testing and maintenance procedures.
Symbols
Wherever you see this symbol in the manual you will find a Hints & Tips note that we hope you will find useful. These notes provide useful advice or information that may help you perform the task more effectively.
Wherever you see this symbol in the manual you will find a Toolbox note that highlights an aspect of test or maintenance that is important for you to know about. A typical example is a software upgrade that you should
check has been installed.

General precautions

Please read the general Operating Precautions described in the Directions for Use carefully prior to using this pump.
w
This pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive components when attempting to repair and service the pump.
V
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to locate the pump away from any such hazardous sources.
B
An electrical shock hazard exists if the casing of the pump is opened or removed. Refer all servicing to qualified service personnel.
A
This pump is protected against the effects of high energy radio frequency emissions and is designed to fail safe if extremely high levels of interference are encountered. Should false alarm conditions be encountered, either remove
M
the source of the interference or regulate the infusion by another appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by a qualified service engineer.
*
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the integrity of the external protective conductor in the installation or its arrangement is in doubt, the pump should be operated
Alaris® GW Volumetric Pump 5/77 1000SM00006 Issue 4
from the battery.
Introduction & Start up

Views of the Alaris® GW Volumetric Pump

Front view
Display Indicators
Flow Stop Mechanism When activated the mechanism will stop fluid flow.
Air Sensor
Bevel Releases the flow stop mechanism lever arm when the door is closed.
Main Display - Displays the infusion rate, VTBI, VI and time remaining for VTBI / Time infusions. Display will flash when pump is operating on battery.
Door Latch Press outer latch to open the door.
Tubing Guide Guide to assist in the loading of the IV infusion set.
Keypad
Flow Direction Label
Pressure Sensor
Rear view
Release Lever Rotates cam to release the pump from horizontal rectangular bars.
Rotating Cam Attaches the pump to rectangular bars or Alaris® DS Docking Station.
AC Fuses
AC Power Connector
Potential Equalisation (PE) Connector
RS232 Connector / Nursecall
Flow Sensor Interface
Folded Pole Clamp
Infrared Communications port (IrDA)
Alaris® GW Volumetric Pump 6/77 1000SM00006 Issue 4

Controls and indicators

PRIMARY / SECONDARY
Switches the pump between Primary and Secondary infusion modes (if enabled).
ON / OFF
Switches the pump on and off.
PRIME / BOLUS
Primes the IV infusion set.
Administers bolus during the infusion.
CLEAR / SILENCE ALARM
Resets numeric values to zero.
Silences alarms and warnings for 1 minute.
ENTER
Scrolls between rate, time, VTBI and total volume infused (VI).
Enters values for selected infusion/configuration parameters.
Confirms the rate during an infusion titration.
RUN / HOLD
Starts and stops the infusion.
Cancels alarm.
CHEVRONS
Increases or decreases the infusion rate, TIME limit and VTBI.
Press and hold to increase the selection speed.
Used to adjust user selectable options.
Introduction & Start up
When any of the following are illuminated:
AC POWER INDICATOR - The pump is connected to an AC power supply.
RATE The pump is displaying the infusion rate in millilitres per hour (ml/h).
VTBI The pump is displaying the Volume To Be Infused (VTBI) in millilitres (ml).
VI The pump is displaying the Volume Infused (VI) in millilitres (ml).
TIME The pump is displaying the infusion time in hours : mins.
MICRO The pump is operating in the MICRO mode. When not illuminated, the pump is in the STANDARD
mode.
SEC The pump is operating in the SECONDARY mode. When not illuminated, the pump is in the PRIMARY mode. ml/hr (Millilitres / hour) When ml is illuminated the pump displays the VTBI or VI. When the hr is illuminated the pump displays the rate or infusion time.
Infusion indicator - Infusing in STANDARD mode.
Infusion indicator - Infusing in MICRO mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in STANDARD mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in MICRO mode.
Alaris® GW Volumetric Pump 7/77 1000SM00006 Issue 4
Introduction & Start up

Operating the Alaris® GW Volumetric Pump

Follow the instructions supplied with the individual IV infusion set.
Use of non-specified IV infusion sets may impair the operation of the pump and the
accuracy of the infusion.
Refer to the "Compatible IV Infusion Sets" section in Appendix D, for a list of
recommended IV infusion sets.
Ensure that the tubing is inserted completely into the pumping channel, avoiding any
slack.
Position the IV fluid container to avoid spillage onto the pump.
When using 273-003 IV infusion sets, ensure a separation of at least 30cm is
maintained between the pump and the upper Y-site.
1.
Close the in-line clamp on the IV infusion set. Press the door latch to open the door.
Release the flow stop mechanism by pushing the lever arm up
2. and to the right.
3.
Insert the IV infusion set from left to right into the slot provided, following the flow direction label, avoiding any slack.
Re-engage the flow stop mechanism by pushing the lever left
4. and down.
Close the door. Use the recesses in the door to press firmly to
5. ensure the latch is correctly applied. Open in-line clamp on the IV infusion set.
Observe fluid chamber and check for no fluid flow.
6.
Alaris® GW Volumetric Pump 8/77 1000SM00006 Issue 4
Introduction & Start up

Starting the Alaris® GW Volumetric Pump

Close in-line clamp.
1.
Load IV infusion set.
2.
Half fill the drip chamber.
3.
Open in-line clamp.
4.
Press to switch pump on.
5.
Prime () set: Press once. Whilst is displayed press again and hold to clear all visible air from the line.
6.
Connect to test equipment as required.
7.
Set-up pump as described below:
8.
RATE / VOLUME
Enter rate using .
1.
Press once to scroll to VTBI.
2.
Enter VTBI using .
3.
Press to scroll to VI.
4.
If necessary, press to clear VI.
5.
Press to start infusion.
6. On completion, infusion will continue at KVO rate until is
pressed.
If a flow sensor is used, VTBI is set to OFF; by pressing the a VTBI may be set.

Features of the Alaris® GW Volumetric Pump

VTBI / TIME
If necessary, press to put pump on hold.
1.
Press for 2 seconds, press twice and will
2.
display, use Keys to scroll to On.
Press twice to return to hold mode.
3.
Enter VTBI using .
4.
Press to scroll to TIME .
5.
Enter TIME using .
6.
Press to scroll to VI
7.
If necessary, press to clear VI.
8.
Press to start infusion.
9.
Panel Lock - prevents certain key operations
Press together to turn Loc On or to turn Loc Off.
1.
Rate Titration - while infusing
Press to adjust rate.
1.
Press to confirm change of rate.
2.
Bolus Infusions - while infusing
Press twice and hold, release after administering the desired bolus volume.
1.
Micro Mode* - pump must be on hold
Press and hold for 2 seconds.
1.
Press three times - will be displayed.
2.
Use to turn micro mode on or off.
3.
Press once to return to or set-up.
4.
* configurable options, refer to Directions For Use.
Alaris® GW Volumetric Pump 9/77 1000SM00006 Issue 4
Chapter 2

Configuration & Calibration

In this chapter
Access codes 11
Configurable options 11
Calibration procedures 13
Configuration & Calibration
Configuration & Calibration

Access codes

The pump software contains a number of configuration and test routines that can be accessed by the user. The majority of tests are driven from a technical access code (see below).
Entering Access Codes
With the pump OFF.
1.
Press and hold and press and release .
2.
The pump will alarm and briefly show the software version
3. installed in the pump. It will then display “CodE”.
Release the key and the pump will display "0".
4.
Use the keys to select the code required from the list.
5.
Press the key to confirm your choice.
6.
If an invalid code is entered, the pump will display “CodE”
7. followed by “0”.

Configurable options

The default settings are configurable as displayed in the table below.
Each of the configurable options has a code which must only be altered by a qualified service engineer with reference to the technical service manual.
Any changes made that are not confirmed by pressing will not be saved on power up.
Use the keys to adjust the selected option. Press the key to confirm your choice. The following access codes can be used to
configure the pump:
Code Description Default Summary
Select one of the two available infusion modes:
21 Enable Volume/Time Infusions OFF
22 Maximum Priming Volume 40ml
23
24 Maximum VTBI in MICRO Mode 999ml
25 Bolus Rate 400ml/h
26 Maximum Bolus Volume 5ml
27 Keep Vein Open (KVO) Rate 5ml/h
28 Single Bubble Alarm Volume* 100µL
30 Enable Secondary Infusions OFF
31 Default Occlusion Pressure Hi Set the default occlusion alarm value (Lo, Normal or Hi) at power-on. 32 Alarm Volume Level 4 Set alarm volume level between 1 (low) and 7 (high).
35 Enable MICRO Mode OFF
36 Maximum Infusion Rate 999ml/h Set the maximum infusion rate between 1 and 999ml/h.
38 ASCII Mode Comms OFF
Clear Infusion Parameters to Zero on Power On
OFF
(CLoC) OFF: Input a Flow Rate and Volume to be Infused (VTBI)
(CLoC) ON: Input a Volume to be Infused (VTBI) and Time for infusion. The Maximum volume (OFF, 1 - 40ml) to be infused during priming
sequence. Before starting an infusion, pressing the b key will initiate set priming sequence.
OFF: Previous infusion parameters: last rate, VTBI (and time of
infusion if applicable) and volume infused are displayed on power on.
ON: Previous infusion parameters are reset to zero on power on. Set the maximum allowable VTBI between 0.1ml and 999ml, in micro
mode only. Set the Bolus rate between 1 and 999ml/h (providing the default bolus
volume is greater than zero). Pump will infuse at this rate when b key is pressed twice and held.
Set the maximum bolus volume between Off and 99ml. Maximum volume that will be delivered whilst the b key is held down during an infusion.
Set the Keep Vein Open (KVO) rate (OFF, 1.0 - 5.0ml/h). At the end of the infusion, the pump can either stop pumping or continue infusing at a Keep Vein Open (KVO) rate.
Set the maximum size of air bubble (50µL, 100µL, 250µL, 500µL) that can be passed through the pump without causing an alarm.
OFF: Disable automatic secondary infusions.
ON: Enable automatic secondary infusions.
OFF: Standard infusion mode.
ON: Enable micro mode.
OFF: Disable ASCII communications mode.
ON: Enable ASCII communications mode.
Alaris® GW Volumetric Pump 11/77 1000SM00006 Issue 4
Configuration & Calibration
Configuration & Calibration
Configurable options (continued)
Code Description Default Summary
39 Odd Parity Comms OFF
40 Pump Address Comms 1 Set pump address used for communications (1 to 250).
41 Flow Sensor Connection Mode AUTO
42 Set-up of Current Time and Date
44 Language Selection EnGL Set language used (EnGL, dEut, FrAn, ItAL, ESPA, nEd, SE).
45 IrDA Communications Selection ON
46 Nurse Call Activation ON
47 Drops per ml of Fluid 20
48 Silent Mode OFF
User Select Mode Configuration:
• Pressure Limit Enabled
49
50 Flow Sensor Sensitivity Level nor
200
201 Reset EEPROM data -
202 Repair EEPROM data -
• Alarm Volume Enabled
• Timed Infusions Enabled
• Micro Infusions Enabled
Reset all configurable options to default
OFF
- Resets all configurable options to factory default.
OFF: Disable odd communications parity bit generation.
ON: Enable odd communications parity bit generation.
AUTO: Pump automatically detects flow sensor if connected.
ON: Pump will only operate with a flow sensor connected. Set current time (00:00 to 23:59), and date (01/01/00-31/12/99) for
event logging. Does not automatically adjust for Summer time.
OFF: RS232 comms enabled.
ON: IrDA comms enabled. OFF: Disable activation state of the nurse call (active low output from
pump).
ON: Enable activation state of the nurse call (active high output
from pump).
Select number of drops per ml of fluid (1 to 200). Defined by the type of set. Reference the packaging of IV infusion set.
OFF: Audible response to a key press is given.
ON: No audible response to a key press is given.
OFF: Disable mode.
ON: Enable mode.
nor: Normal sensitivity.
Hi: High sensitivity.
Reset EEPROM data code to a defined state if EEPROM checksum error.
NOTE: The pump will need to be returned to factory for reconfiguration if this option is used.
Detects and repairs any corrupted memory segments, resetting any repaired areas back to the factory defaults.
NOTE: The pump will need to be returned to factory for reconfiguration if this option is used.
*Single Bubble Alarm Volume
Although an individual bubble may not exceed the pre-programmed threshold, the accumulative volume of bubbles, in a 15 minute window, may be sufficient to initiate an air-in­line alarm, indicated by an “Air OCCL” message.
Before making any amendments to configuration settings:
Care should be taken to document existing configuration settings to enable changes
to be reverted if required. Configuration requirements may vary from ward to ward therefore care should be taken to ensure any configuration settings are appropriate for the ward concerned and users are aware of any changes to configuration settings prior to use.
Subsequently, sharing of Alaris® GW Volumetric Pumps between wards may be inappropriate.
Alaris® GW Volumetric Pump 12/77 1000SM00006 Issue 4
Configuration & Calibration
Configuration & Calibration
This section outlines the procedures for calibration of the Alaris® GW Volumetric Pump.
All of these calibrations should only be carried out by qualified biomedical engineers. If in any doubt about how to perform the tests, in particular the pressure sensor calibration checks, contact your local Cardinal Health Service Centre who will be able to assist.

Calibration procedures

Recommended Calibration Equipment
Specialised test equipment is not required for the majority of the functional tests to be carried out on the pump. In order to calibrate or verify the occlusion alarm point or volumetric accuracy, the following equipment will be necessary:
IV infusion set, suitable for the Alaris® GW Volumetric Pump with standard LUER lock taps for connecting to other test equipment. If the standard sets are not available, it is possible to order a basic "test set" that is available from your local ALARIS® Service Centre
- part number 0000TG00074. Note that all sets should only be used for a single calibration operation. Pressure gauge for measuring liquid pressure, with a full scale of 0-2 bar (0-1500mmHg) ±20mmHg.
either:
Class A 50ml glass burette with graduations down to 0.1ml and calibrated down to ± 0.05ml.
or:
Calibrated scales accurate to at least ± 0.01g.
Volumetric Accuracy Calibration (CODE 18)
Volumetric Accuracy Calibration (CODE 18)
The volumetric accuracy calibration routine is used for manufacture of the pump only.
The volumetric accuracy calibration routine is used for manufacture of the pump only.
Displaying the Volumetric and Pressure Calibration Values (CODE 2)
Use of this access code simply displays the calibration values stored in the software.
Enter the access code 2.
1. Press to step through all Cal values.
2. Switch the pump "OFF" if there are no further tests to be done at that time.
3.
Battery Calibration (CODE 4)
The pump must be connected to the AC power source throughout the duration of the test. The test can be aborted at any time by switching the pump off using the key; no change is made to the battery low point calibration value stored previously in the pump. This calibration should only be performed on a fully charged battery. Ensure the pump is plugged into the mains for at least
24 hours before starting this procedure.
Plug a mains lead into the pump.
1.
Prepare a fluid-filled looped tubing, load it into the pump and close the door.
2.
Enter the access code 4.
3.
The test will start automatically. The pump infuses at 999 ml/h internally switching to the battery power source to discharge the
4. battery; the air-in-line alarm is disabled. Throughout this test the time indicator will increase in minute intervals.
As soon as the pump detects that the battery is discharged, the time shown on the main display will stop increasing and begin to
5. flash.
If the time is greater than two hours and the low point voltage value is within the allowable range of 6.5 to 7.8 volts, then the
6. display shows "PASS" / "xx:xx" / "bx.x", the pass indication, elapsed time and battery low point value in volts. Otherwise the display shows "FAIL" / "xx:xx" / "bx.x".
Press the key. When the key is pressed, the low point calibration value will be stored.
7.
If the pump is switched off before the key is pressed, the calibration value will be lost and the
If the pump is switched off before the key is pressed, the calibration value will be lost and the test will have to be repeated. When is pressed, the pump will revert to the technical service
test will have to be repeated. When is pressed, the pump will revert to the technical service entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump
entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump OFF if there are no further tests to be done at that time.
OFF if there are no further tests to be done at that time.
If the pump fails the calibration, try to charge the battery, and repeat the test. If this fails again then replace the battery and / or
8. Power Supply Unit.
Alaris® GW Volumetric Pump 13/77 1000SM00006 Issue 4
Configuration & Calibration
Configuration & Calibration
Calibration procedures (continued)
Pressure Sensor Calibration (CODE 17)
An internal pressure sensor is used to detect downstream occlusions. This sensor requires calibration whenever a new sensor, Main PCB is fitted or if door is changed or removed. It is necessary when servicing a pump, to carry out an occlusion pressure test to verify that the sensor is calibrated correctly (see self test routine in Chapter 3 Routine Maintenance). A calibrated pressure gauge will be needed in order to perform this calibration.
PRI
SEC
Pressure gauge ± 20mmHg
When the pressure sensor is replaced the null pressure value must be checked and adjusted, if required, prior to calibration. Check the null pressure value as follows:
1. Enter the access code 12. Go to test 8 and press e.
2. With no set loaded and door open check displayed value is 11 ±4.
3. If the reading is outside of tolerance then adjust R2 on the Pressure Sensor PCB until displayed value is within tolerance.
Load a set into the pump to be calibrated and prime the set. Connect to pressure gauge as shown in diagram above. Enter the access code 17. Apply pressure required for each step and when pressure required is displayed on pressure gauge for 10 seconds (allows pressure to settle) press e. Calibration values will be returned. Press e to go to next step.
PrES step - 150mmHg ± 40mmHg
HI step - 650mmHg ± 40mmHg.
CAP step - 1000mmHg ± 40mmHg.
Difference step
XX or XXX indicates calibration values, that have no tolerance values applicable.
YYY indicates calibration values that should be between 115 and 214.
ZZ indicates a calibration value that is the difference between XXX and XX and should be between 37 and 54.
Alaris® GW Volumetric Pump 14/77 1000SM00006 Issue 4

Routine Maintenance

In this chapter
Routine maintenance 16
Test procedures 16
Chapter 3
General cleaning and inspection for damage 20
Storage 20
Data transfer 21
Performance verification procedure 23
Routine Maintenance

Routine Maintenance

Routine maintenance
For routine maintenance, the following self-test and performance verification procedures should be performed in addition to the tasks described in the section on General Cleaning and Inspection for Damage.
Refer to the relevant “Directions for Use” for the recommended routine maintenance period.

Test procedures

Important Service Information:
Important Service Information:
Testing and Calibration of Volumetric pumps is very dependent on the tubing set used.
Testing and Calibration of Volumetric pumps is very dependent on the tubing set used. For this reason, a new set of tubing should be used for each pump tested, and the tubing
For this reason, a new set of tubing should be used for each pump tested, and the tubing should be thrown away once all tests are completed. Recommended test set is part number
should be thrown away once all tests are completed. Recommended test set is part number 0000TG00074.
0000TG00074.
Entering Access Codes (Technician Mode)
Note: See Chapter 2 Configuration & Calibration for information on how to enter access codes
Code Test Description
1
2
3 Main Self Test See Self Test Routine Table.
4
5 Volumetric Accuracy Verification Test See Volumetric Accuracy Verification Test (code 5) detailed description in this document.
10 Alarm history log
11 Display current time and date -
12
17 Pressure Calibration Refer to Chapter 2 Configuration & Calibration.
18 Volumetric Calibration This is a manufacturing code, volumetric calibration should not be carried out.
19 Reduced Volumetric Calibration This is a manufacturing code, volumetric calibration should not be carried out.
67 Learn configuration settings See Teach Learn procedure.
Input a pump reference number, and service date
Display volumetric, pressure calibration and battery Cal values
Automatic Battery Maintenance/ Discharge Test
Access to individual tests within the Main Self Test
This enables the user to put in their own 4 digit reference number or asset number, together with the date the pump was last serviced in the format Wk:Yr (15:01) week 15 of 2001. Defaults to 0 & 0:00.
Reference only, 4 values shown: CAL - 20.00 if pump has not been calibrated (range 16-24) Pres - DXXX delta value, (range 36 to 55, default to 46) CAP - CXXX Maximum value, (range 110 to 219, default 163) bAt - bx.xx (range 6.44 to 7.86, default 7.15)
Takes approximately 2-5 hours, if it takes less than 2 hours it will display ‘fail’ The pump will then switch to charging, it is recommended this be done for 24 hours.
This will show the last 10 error/alarm codes Use the key to step through.
Note: Not in Sequential Test (code 3), available only through code 12.
Test Nr. Description
9 EEPROM Checksum Test. Display two 16-bit EEPROM check sums. During power
up the processor calculates EPROM checksums values and checks them against those stored, if a difference is detected a corruption in data has occurred.
11 Pumping Mechanism Test. This test disables the pressure, door and AIL sensors.
The pump will pump into a pressure gauge and display the pressure reached, this is exited by pressing the enter key.
12 Bubble Measurement Test. This displays the size of the air bubble detected in
microlitres.
Alaris® GW Volumetric Pump 16/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance
Test procedures (continued)
Self test routine
Enter access code 3. Press the key to advance to next test.
Level Test Description
Confirm display indicates correct button pressed. When level is entered the display will show “b-1”, press buttons 1 to 8, after pressing button 8 the test will automatically proceed to level 2.
1 Keypad Test

b-1 b-2 b-3 b-4 b-5 b-6 b-7 b-8
Check all LEDs. The pump will run through a count-up series to illuminate each segment of the 7 segment LEDs, and
2 Display Test
3 Alarm Test
4 Door Test
5 Air Sensor Test
7 Motor Opto Test
8 Occlusion Pressure Test
cycle through all of the green LEDs. Confirm all LEDs are working. At the end of this test all LEDs will illuminate.
Press the key to advance to level 3.
Confirm the alarm is working and a distinctive change is heard between volume levels. Pump displays ‘ALAR’ and alarms for 0.5 seconds at each volume (1 to 7).
Press the key to advance to level 4.
Confirm the change of state between door open (d-0) and door closed (d-1). Press the key to advance to level 5.
Confirm the change of state between an air filled set (a-0) and a fluid filled set (a-1). Press the key to advance to level 7.
Displays PASS or FAIL. Confirm displays PASS. The pump runs the motor forwards a turn, then backwards a turn. The processor checks that it sees both motor optos come on at the correct time.
Press the key to advance to level 8.
Test requires a calibrated pressure gauge. Connect the pump IV Infusion set to the pressure gauge via a 3 way tap.
Press the key for 10 seconds. Press the key, the pump will run at 125ml/h and the display will show the current pressure sensor reading ‘xxx’. After 10 seconds close off the 3 way tap so that the pump delivers into the pressure gauge.
Confirm that an alarm occurs and a ‘’ appears on the pump display. The pressure displayed on the pressure gauge should be 500 mmHg +/- 150 mmHg.
Carry out the next test or press the key to advance to level 10.
10 Drop Sensor Test
Check for correct drop count. This test counts the number of drops detected, if the sensor is not present then ‘Off’ will be displayed.
Alaris® GW Volumetric Pump 17/77 1000SM00006 Issue 4
Routine Maintenance
30cm
30cm
30cm
30cm
Routine Maintenance
Test procedures (continued)
Volumetric Accuracy Verification Test (CODE 5)
Important Notes:
Important Notes:
The balances need to be switched on for 30 minutes prior to use to enable the electronics to
The balances need to be switched on for 30 minutes prior to use to enable the electronics to warm up and settle.
warm up and settle.
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing must be used.
must be used.
Do not move the desk during testing, it will upset the balance readings.
Do not move the desk during testing, it will upset the balance readings.
This test is used to confirm that the pumping accuracy of the system as a whole, including the tubing, is within the specified limits.
In the most controlled conditions, a needle should be used to pump liquid into the weighing beaker to prevent liquid touching the sides of the beaker and to provide some back-pressure so that leaks/overflows do not affect the readings. As a result of these and other errors, if the system fails just marginally, it is worth performing the test a second time. If it still fails, return the pump to your local ALARIS® Service Centre for further analysis. The head height on IV infusion set should be approximately 30cm.
Setup A Setup Bor...
Setup A Setup Bor...
Class A Burette (50ml)
Class A Burette (50ml)
0.1ml graduations
0.1ml graduations
Burette clamp
Burette clamp
Equipment stand
Equipment stand
Table or bench
Table or bench
Needle
Needle
Beaker
Beaker
Scales (±0.01g)
Scales (±0.01g)
Enter access code 5. The pump will initially display “”.
1.
If it is necessary to prime the set, press and hold the button. The pump will display "" and allow the set to be primed,
2. ignoring any air-in-line alarms.
Release once the set is primed and the pump will again display “”.
3.
Alaris® GW Volumetric Pump 18/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance
Test procedures (continued)
Zero scales or burette. Press to begin the verification check.
4.
The pump will automatically pump at 125ml/h for a total of 20ml, which will take approximately 9½ minutes. During the run, it
5. will display the volume infused up to that point and intermittently flash "r" with the run indicator, to show that this is a verification run. At the end of this run, the pump should display "20.00" and “wait”.
Allow the scales to settle and then note the reading on the scales. If using a burette, take the final reading of volume infused.
6. The volume infused should be 20.00ml ± 5%.
Press the key and the pump will revert to the technical service entry mode and flash "CodE", followed by "0", enabling you to
7. begin other tests if required. Switch the pump "OFF" if there are no further tests to be done at that time.
Flow Stop Mechanism Test
It is necessary to check that the flow stop device can hold sufficient pressure and thus prevent free flow.
Load set into the pump. Use same setup as for pressure calibration (see Chapter 2 Configuration & Calibration).
1.
Open the door and check that the arm stays in the UP position by lifting the arm onto the ledge.
2.
Close the door fully.
3.
Reopen the door and note that the flow stop arm has activated into the down position.
4.
Pull the test tubing out of the pump and place the tube in front of the flow stop mechanism arm in the down position.
5.
Close the door and reopen the door. Check the tubing locator has loaded the tube fully under the clamp and is flat. This checks
6. whether the tubing locator is the correct way around and works.
Ensure the door is open and the flow stop is closed. Apply 650mmHg to the distal end of the IV infusion set for 10 seconds. Then
7. reduce pressure to 500mmHg to the distal end and verify pressure gauge reading is 500mmHg. Ensure the pressure does not drop by more than 15mmHg in 30 seconds.
Alaris® GW Volumetric Pump 19/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance

General cleaning and inspection for damage

To ensure that this pump remains in good operating condition, it is important to keep it clean and carry out the routine procedures described below. All servicing should only be performed by a qualified service engineer, with reference to this manual.
Thoroughly clean external surfaces of the pump by wiping over with a lint-free cloth, lightly dampened with warm water and a
standard disinfectant/detergent solution.
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
- Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Recommended cleaners are:
Brand Concentration
Hibiscrub 20% (v/v)
Virkon 1% (w/v)
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid to enter the casing and avoid excess fluid build up on the pump. Do not use aggressive cleaning
to enter the casing and avoid excess fluid build up on the pump. Do not use aggressive cleaning agents as these may damage the exterior surface of the pump. Do not steam autoclave, ethylene
agents as these may damage the exterior surface of the pump. Do not steam autoclave, ethylene oxide sterilise or immerse this pump in any fluid.
oxide sterilise or immerse this pump in any fluid.
Labels should be flat and legible. Any label, if no longer fully adhered, must be replaced if it represents a path for fluid ingress.
Case components must be checked for damage that may affect function, fluid ingress route or present a user hazard and be re-
placed if necessary. Check the pole clamp screws are not loose and that the threads are not damaged. Check that it folds away and ensure arm is not
bent. Inspect the AC power supply plug and cable for damage.
Clean the flow sensor by wiping over with a cloth, lightly dampened with warm water and a
Clean the flow sensor by wiping over with a cloth, lightly dampened with warm water and a standard disinfectant/detergent solution. Ensure the connector does not get wet. Dry flow sensor
standard disinfectant/detergent solution. Ensure the connector does not get wet. Dry flow sensor before use.
before use.
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if handles
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if handles operation is not free, the flow sensor may be immersed and soaked in clean soapy water.
operation is not free, the flow sensor may be immersed and soaked in clean soapy water. Activating the spring mechanism of the sensor whilst immersed will assist in cleaning the inside
Activating the spring mechanism of the sensor whilst immersed will assist in cleaning the inside of the mechanism. After cleaning the sensor should be allowed to dry fully prior to use.
of the mechanism. After cleaning the sensor should be allowed to dry fully prior to use. Caution: the plug of the sensor should not be immersed as damage will occur.
Caution: the plug of the sensor should not be immersed as damage will occur.

Storage

If the pump is to be stored for an extended period it should be cleaned and the internal battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in this chapter and ensure that the internal battery is fully charged.
Please note during long term storage of the pump the Real Time Clock circuitry is being
Please note during long term storage of the pump the Real Time Clock circuitry is being maintained by BT1 on the control PCB. Under long term storage conditions it is recommended
maintained by BT1 on the control PCB. Under long term storage conditions it is recommended that the pump is powered in Technician Mode for a period of 24 hours so as to keep the BT1
that the pump is powered in Technician Mode for a period of 24 hours so as to keep the BT1 charged, and eliminate the possibility of depleting BT1 and inducing Err9 faults at power up.
charged, and eliminate the possibility of depleting BT1 and inducing Err9 faults at power up.
Alaris® GW Volumetric Pump 20/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance

Data transfer

Upgrading Software
The optional upgrade of the Alaris® GW Volumetric Pump software to V5R1F should be considered at the next product service for all Alaris® GW Volumetric Pumps fitted with software version V4R2C. Perform upgrades by acquiring the software upgrade kits specified in the spares parts listings. Note: when upgrading Alaris® GW Volumetric Pumps from software version V4R1B, first install the V4R2C software upgrade kit to enable the flash upload capabilty.
The major features of the V5R1F software include:
Teach / Learn Capability;
Additional configuration options:
Silent Mode;
SELECT Mode Options;
Drop Sensor Connection Mode;
Drop Sensor Light Sensitivity Level;
Alarm volume level factory default now 4 (was 7);
Clear Infusion Parameters to Zero default now OFF (was ON).
Additional Technician Mode Configuration options:
New Volumetric Calibration Mode to reduce calibration time (Code 19);
New EEPROM Memory Management to improve work with Teach / Learn and to eliminate need to recalibrate pumps following firmware upgrade (Codes 200, 201, 202).
Automatic setting of VTBI to OFF when used with drop sensor;
Elimination of '' error in Mode resulting in nuisance alarms;
VI now cleared in Mode and retained upon power down.
Recalibration is not required when upgrading from software version V4R2C, although all configuration parameters will be returned to factory defaults.
PC Requirements
Microsoft Windows 95, 98, 2000 or NT operating system
9pin D-type PC serial port or IrDA port.
Tools required
CD-ROM 1000SP00493 - Alaris® GW Volumetric Pump Software Distribution Disk V5R1F
Programming Kit 1000SP00172 (Suitable for all ASENA® Infusion Pumps) or RS232 Cable 1000SP00336
Follow the programming instructions given in publication 1000PB01365 (supplied with 1000SP00493).
1.
Load the software program onto your PC from the Distribution Disk.
2.
Select the Alaris LVP SMU icon (WinSmug)
3.
Place the infrared programming device approximately 50mm directly behind the IrDA window on the rear case or connect
4. the RS232 cable to the 9 pin D type serial port connector situated on the side of the pump.
Select the appropriate port (e.g. COM1) on the Alaris LVP SMU program and Press 'Upload'.
5. Briefly depress the key on the pump to be upgraded; the pump will now display "PROG".
6. Switch the pump on for normal operation when the upgrade is complete. If required, the pump will display whilst
7. automatically completing a test sequence during which the EEPROM memory will be re-partitioned.
When the upgrade is complete, enter the Technician Mode and verify the correct software version has been installed;
8. initiate a factory Reset (Code 200).
Perform the Self Test checks (Code 3).
9.
Power failure. Power failures may occur when using laptops when communicating with the Alaris®
Power failure. Power failures may occur when using laptops when communicating with the Alaris® GW Volumetric Pump due to power requirements. External power may be used in conjunction
GW Volumetric Pump due to power requirements. External power may be used in conjunction with IrDA or RS232 to compensate for lack of power from the laptop.
with IrDA or RS232 to compensate for lack of power from the laptop.
Bright sunlight and strong fluorescent lighting affect the Infrared programming system. If any
Bright sunlight and strong fluorescent lighting affect the Infrared programming system. If any errors are reported then the RS232 method of upgrading the software should be used.
errors are reported then the RS232 method of upgrading the software should be used.
Alaris® GW Volumetric Pump 21/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance
Data transfer (continued)
Teach Learn (Software Versions V5R1F and above)
For both the teach and learn pumps in Technician Mode enable IrDA communications (Code 45), and ensure that ASCII / Binary
1. mode and parity bit options (Codes 38 and 39 respectively) are the same.
Turn the teaching pump on in normal operation. Note: For multiple teach-learn operations, to avoid call-back alarm every
2. 2 minutes, turn the teaching pump on in Technician Mode.
Enter Technician Code 67 on the learning pump.
3.
Align the two IrDA ports on the pumps (optimum distance 50 mm).
4. Depress the key to initiate learning.
5.
A progress bar will travel across the learn pump.
6.
When successful, the learn pump will display 'PASS'.
7.
If the learning pump is unable to learn all configuration parameters then the display will show 'ConF' followed by a list of the
8. configuration parameters that could not be learnt; these will instead contain the factory default settings. This could occur if, for example, the software version of the learning pump is newer than that of the teaching pump.
Possible Reasons for failure
Possible Reasons for failure
IrDA not enabled on both pumps;
IrDA not enabled on both pumps;
ASCII / Binary and parity bit options are not the same;
ASCII / Binary and parity bit options are not the same;
If the software versions are not compatible;
If the software versions are not compatible;
If the pump models are different;
If the pump models are different;
The line of sight between the IrDA windows was obstructed during data
The line of sight between the IrDA windows was obstructed during data
transfer.
transfer.
Event Log Download
A PC application known as the Event Log Download Utility (ELDU) is available to download the event log from the Alaris® GW Volumetric Pump.
ELDU Operation
Click on ELDU icon on PC.
1.
Click Accept to agree with Restrictions of Use and continue;
2.
Select Configure from drop-down menu;
3.
Select Setup Pump and choose Alaris® GW as pump type;
4.
Select Settings to select log to be downloaded;
5.
Check communications are set up as follows:
6. Required PC Comm port selected Character type and parity match pump configuration
Click OK to confirm
7.
Align the IrDA converter with the IrDA window (optimum distance 50 mm), or connect an RS 232 cable.
8. Power up the pump by pressing the key.
9.
Click Download log from the main PC screen.
10.
Press Close when finished.
11.
Select File from drop-down menu and save file. Log may be printed here as required.
12.
Alaris® GW Volumetric Pump 22/77 1000SM00006 Issue 4
Routine Maintenance
Routine Maintenance

Performance verification procedure

Model / Serial Number: Service Order / Inventory Number:
Hospital Name / Reference: Software Version:
Physical Inspection and Clean
Recommended When Serviced Updates
Grease the pumping mechanism with Molykote PG54. TSM
Upgrade to software version V5R1F or above. TSM
Set/Confirm time and date - access code 42
Set service date - access code 1
Check all functions in Self Test - access code 3
During standard infusion check the following:
KVO Operation Flow-stop mechanism test
Alarms Functionality Check
Door , AIL (OCC/AIR), Upstream Occlusion (OCC/AIR), Power fail, Time Out, Downstream Occlusion (HI PRESS). Ensure pump works on battery and AC mains
UPDATE
REF.
Fitted
Not fitted / Not
Applicable
Rate Accuracy Verification Test
(Automatic test in Code 5)
Rate set to 125ml/h, VTBI set to 20ml. Volume infused = 19 to 21ml. _____________ml
Pressure Tests
(Automatic test in Code 12 test 8)
Pump set to alarm at 500mmHg. Pressure = 350 to 650mmHg. _____________mmHg
Set Rate to Zero (or lowest value possible), Clear Volume infused and VTBI Clear Error/Alarm/Battery logs (As required)
Electrical Safety Test Class I Type CF
Alternatively attach printed test results
Earth Resistance Test <=0.2 Ω _____________Ω
Earth Leakage Current <=500 µA _____________µA
Enclosure Leakage Current <=100 µA _____________µA
Verification Performed By ________________________
Sign
For additional information, refer to: 1000SM00006 - Alaris® GW Volumetric Pump Technical Service Manual (TSM)
_______________________
Print
_________________
Date
Alaris® GW Volumetric Pump 23/77 1000SM00006 Issue 4

Troubleshooting

In this chapter
Software alarm codes & displayed messages 25
Alarm types 27
Chapter 4
General fault diagnosis 27
Loading...
+ 53 hidden pages