The Alaris® GW Volumetric Pump is designed to deliver a continuous and accurate infusion. High performance, comprehensive alarm
protection and sophisticated monitoring systems, combined with simple operation, make this pump ideal for general care and critical
care in a variety of areas within a hospital.
The Asena® brand name has been recently changed to the Alaris® brand name. This change in brand name has no effect on the
intended use or functionality of the product. Recommended disposable products for use with this product may refer to either the
Asena® brand name or Alaris® brand name and both types are suitable for use with this infusion pump.
Familiarity
Before operation, ensure that you are fully familiar with this pump by carefully studying the Directions for Use (DFU) prior to
attempting any repairs or servicing.
As part of a policy of continuous improvement, product enhancements and changes are introduced from time to time.
Purpose
This Technical Service Manual describes how to set up, test and maintain the Alaris® GW Volumetric Pump. It is intended for use by
personnel experienced in medical equipment testing and maintenance procedures.
Symbols
Wherever you see this symbol in the manual you will find a Hints & Tips note that we hope you will find useful.
These notes provide useful advice or information that may help you perform the task more effectively.
Wherever you see this symbol in the manual you will find a Toolbox note that highlights an aspect of test or
maintenance that is important for you to know about. A typical example is a software upgrade that you should
check has been installed.
General precautions
Please read the general Operating Precautions described in the Directions for Use carefully prior to using this pump.
w
This pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive
components when attempting to repair and service the pump.
V
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to locate the
pump away from any such hazardous sources.
B
An electrical shock hazard exists if the casing of the pump is opened or removed. Refer all servicing to qualified
service personnel.
A
This pump is protected against the effects of high energy radio frequency emissions and is designed to fail safe if
extremely high levels of interference are encountered. Should false alarm conditions be encountered, either remove
M
the source of the interference or regulate the infusion by another appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to
have been damaged, remove it from service for inspection by a qualified service engineer.
*
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the integrity
of the external protective conductor in the installation or its arrangement is in doubt, the pump should be operated
Flow Stop Mechanism
When activated the
mechanism will stop fluid
flow.
Air Sensor
Bevel
Releases the flow stop
mechanism lever arm when
the door is closed.
Main Display - Displays the infusion rate, VTBI, VI and time
remaining for VTBI / Time infusions. Display will flash when
pump is operating on battery.
Door Latch
Press outer latch to open
the door.
Tubing Guide
Guide to assist in the loading of
the IV infusion set.
Keypad
Flow Direction Label
Pressure Sensor
Rear view
Release Lever
Rotates cam to release
the pump from horizontal
rectangular bars.
Rotating Cam
Attaches the pump to
rectangular bars or Alaris®
DS Docking Station.
Switches the pump between Primary and Secondary infusion modes (if enabled).
ON / OFF
Switches the pump on and off.
PRIME / BOLUS
Primes the IV infusion set.
Administers bolus during the infusion.
CLEAR / SILENCE ALARM
Resets numeric values to zero.
Silences alarms and warnings for 1 minute.
ENTER
Scrolls between rate, time, VTBI and total volume infused (VI).
Enters values for selected infusion/configuration parameters.
Confirms the rate during an infusion titration.
RUN / HOLD
Starts and stops the infusion.
Cancels alarm.
CHEVRONS
Increases or decreases the infusion rate, TIME limit and VTBI.
Press and hold to increase the selection speed.
Used to adjust user selectable options.
Introduction & Start up
When any of the following are illuminated:
AC POWER INDICATOR - The pump is connected to an AC power supply.
RATE The pump is displaying the infusion rate in millilitres per hour (ml/h).
VTBI The pump is displaying the Volume To Be Infused (VTBI) in millilitres (ml).
VI The pump is displaying the Volume Infused (VI) in millilitres (ml).
TIME The pump is displaying the infusion time in hours : mins.
MICRO The pump is operating in the MICRO mode. When not illuminated, the pump is in the STANDARD
mode.
SEC The pump is operating in the SECONDARY mode. When not illuminated, the pump is in the PRIMARY
mode.
ml/hr (Millilitres / hour) When ml is illuminated the pump displays the VTBI or VI. When the hr is illuminated
the pump displays the rate or infusion time.
Infusion indicator - Infusing in STANDARD mode.
Infusion indicator - Infusing in MICRO mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in STANDARD mode.
Infusion indicator - Displays fluid drops detected by the flow sensor when infusing in MICRO mode.
The pump software contains a number of configuration and test routines that can be accessed by the user. The majority of tests are
driven from a technical access code (see below).
Entering Access Codes
With the pump OFF.
1.
Press and hold and press and release .
2.
The pump will alarm and briefly show the software version
3.
installed in the pump. It will then display “CodE”.
Release the key and the pump will display "0".
4.
Use the keys to select the code required from the list.
5.
Press the key to confirm your choice.
6.
If an invalid code is entered, the pump will display “CodE”
7.
followed by “0”.
Configurable options
The default settings are configurable as displayed in the table below.
Each of the configurable options has a code which must only be altered by a qualified service
engineer with reference to the technical service manual.
Any changes made that are not confirmed by pressing will not be saved on power up.
Use the keys to adjust the selected option. Press the key to confirm your choice. The following access codes can be used to
configure the pump:
CodeDescriptionDefaultSummary
Select one of the two available infusion modes:
21Enable Volume/Time InfusionsOFF
22Maximum Priming Volume40ml
23
24Maximum VTBI in MICRO Mode 999ml
25Bolus Rate 400ml/h
26Maximum Bolus Volume 5ml
27Keep Vein Open (KVO) Rate 5ml/h
28Single Bubble Alarm Volume*100µL
30Enable Secondary Infusions OFF
31Default Occlusion PressureHiSet the default occlusion alarm value (Lo, Normal or Hi) at power-on.
32Alarm Volume Level 4Set alarm volume level between 1 (low) and 7 (high).
35Enable MICRO Mode OFF
36Maximum Infusion Rate 999ml/hSet the maximum infusion rate between 1 and 999ml/h.
38ASCII Mode CommsOFF
Clear Infusion Parameters to Zero
on Power On
OFF
(CLoC) OFF: Input a Flow Rate and Volume to be Infused (VTBI)
(CLoC) ON: Input a Volume to be Infused (VTBI) and Time for infusion.
The Maximum volume (OFF, 1 - 40ml) to be infused during priming
sequence. Before starting an infusion, pressing the b key will initiate set
priming sequence.
OFF: Previous infusion parameters: last rate, VTBI (and time of
infusion if applicable) and volume infused are displayed on
power on.
ON: Previous infusion parameters are reset to zero on power on.
Set the maximum allowable VTBI between 0.1ml and 999ml, in micro
mode only.
Set the Bolus rate between 1 and 999ml/h (providing the default bolus
volume is greater than zero). Pump will infuse at this rate when b key is
pressed twice and held.
Set the maximum bolus volume between Off and 99ml. Maximum
volume that will be delivered whilst the b key is held down during an
infusion.
Set the Keep Vein Open (KVO) rate (OFF, 1.0 - 5.0ml/h). At the end of the
infusion, the pump can either stop pumping or continue infusing at a
Keep Vein Open (KVO) rate.
Set the maximum size of air bubble (50µL, 100µL, 250µL, 500µL) that
can be passed through the pump without causing an alarm.
40Pump Address Comms 1Set pump address used for communications (1 to 250).
41Flow Sensor Connection Mode AUTO
42Set-up of Current Time and Date
44Language Selection EnGLSet language used (EnGL, dEut, FrAn, ItAL, ESPA, nEd, SE).
45IrDA Communications Selection ON
46Nurse Call ActivationON
47Drops per ml of Fluid 20
48Silent Mode OFF
User Select Mode Configuration:
• Pressure Limit Enabled
49
50Flow Sensor Sensitivity Levelnor
200
201Reset EEPROM data -
202Repair EEPROM data -
• Alarm Volume Enabled
• Timed Infusions Enabled
• Micro Infusions Enabled
Reset all configurable options to
default
OFF
-Resets all configurable options to factory default.
OFF: Disable odd communications parity bit generation.
ON: Enable odd communications parity bit generation.
AUTO: Pump automatically detects flow sensor if connected.
ON: Pump will only operate with a flow sensor connected.
Set current time (00:00 to 23:59), and date (01/01/00-31/12/99) for
event logging. Does not automatically adjust for Summer time.
OFF: RS232 comms enabled.
ON: IrDA comms enabled.
OFF: Disable activation state of the nurse call (active low output from
pump).
ON: Enable activation state of the nurse call (active high output
from pump).
Select number of drops per ml of fluid (1 to 200). Defined by the type of
set. Reference the packaging of IV infusion set.
OFF: Audible response to a key press is given.
ON: No audible response to a key press is given.
OFF: Disable mode.
ON: Enable mode.
nor: Normal sensitivity.
Hi: High sensitivity.
Reset EEPROM data code to a defined state if EEPROM checksum error.
NOTE: The pump will need to be returned to factory for reconfiguration
if this option is used.
Detects and repairs any corrupted memory segments, resetting any
repaired areas back to the factory defaults.
NOTE: The pump will need to be returned to factory for reconfiguration
if this option is used.
*Single Bubble Alarm Volume
Although an individual bubble may not exceed the pre-programmed threshold, the
accumulative volume of bubbles, in a 15 minute window, may be sufficient to initiate an air-inline alarm, indicated by an “Air OCCL” message.
Before making any amendments to configuration settings:
Care should be taken to document existing configuration settings to enable changes
to be reverted if required. Configuration requirements may vary from ward to ward
therefore care should be taken to ensure any configuration settings are appropriate
for the ward concerned and users are aware of any changes to configuration settings
prior to use.
Subsequently, sharing of Alaris® GW Volumetric Pumps between wards may be
inappropriate.
This section outlines the procedures for calibration of the Alaris® GW Volumetric Pump.
All of these calibrations should only be carried out by qualified biomedical engineers. If in any doubt about how to perform the tests,
in particular the pressure sensor calibration checks, contact your local Cardinal Health Service Centre who will be able to assist.
Calibration procedures
Recommended Calibration Equipment
Specialised test equipment is not required for the majority of the functional tests to be carried out on the pump. In order to calibrate
or verify the occlusion alarm point or volumetric accuracy, the following equipment will be necessary:
IV infusion set, suitable for the Alaris® GW Volumetric Pump with standard LUER lock taps for connecting to other test equipment.
If the standard sets are not available, it is possible to order a basic "test set" that is available from your local ALARIS® Service Centre
- part number 0000TG00074. Note that all sets should only be used for a single calibration operation.
Pressure gauge for measuring liquid pressure, with a full scale of 0-2 bar (0-1500mmHg) ±20mmHg.
either:
Class A 50ml glass burette with graduations down to 0.1ml and calibrated down to ± 0.05ml.
or:
Calibrated scales accurate to at least ± 0.01g.
Volumetric Accuracy Calibration (CODE 18)
Volumetric Accuracy Calibration (CODE 18)
The volumetric accuracy calibration routine is used for manufacture of the pump only.
The volumetric accuracy calibration routine is used for manufacture of the pump only.
Displaying the Volumetric and Pressure Calibration Values (CODE 2)
Use of this access code simply displays the calibration values stored in the software.
Enter the access code 2.
1.
Press to step through all Cal values.
2.
Switch the pump "OFF" if there are no further tests to be done at that time.
3.
Battery Calibration (CODE 4)
The pump must be connected to the AC power source throughout the duration of the test. The test can be aborted at any time by
switching the pump off using the key; no change is made to the battery low point calibration value stored previously in the
pump. This calibration should only be performed on a fully charged battery. Ensure the pump is plugged into the mains for at least
24 hours before starting this procedure.
Plug a mains lead into the pump.
1.
Prepare a fluid-filled looped tubing, load it into the pump and close the door.
2.
Enter the access code 4.
3.
The test will start automatically. The pump infuses at 999 ml/h internally switching to the battery power source to discharge the
4.
battery; the air-in-line alarm is disabled. Throughout this test the time indicator will increase in minute intervals.
As soon as the pump detects that the battery is discharged, the time shown on the main display will stop increasing and begin to
5.
flash.
If the time is greater than two hours and the low point voltage value is within the allowable range of 6.5 to 7.8 volts, then the
6.
display shows "PASS" / "xx:xx" / "bx.x", the pass indication, elapsed time and battery low point value in volts. Otherwise the
display shows "FAIL" / "xx:xx" / "bx.x".
Press the key. When the key is pressed, the low point calibration value will be stored.
7.
If the pump is switched off before the key is pressed, the calibration value will be lost and the
If the pump is switched off before the key is pressed, the calibration value will be lost and the
test will have to be repeated. When is pressed, the pump will revert to the technical service
test will have to be repeated. When is pressed, the pump will revert to the technical service
entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump
entry mode and flash CodE, followed by 0, this allows you to begin other tests. Switch the pump
OFF if there are no further tests to be done at that time.
OFF if there are no further tests to be done at that time.
If the pump fails the calibration, try to charge the battery, and repeat the test. If this fails again then replace the battery and / or
An internal pressure sensor is used to detect downstream occlusions. This sensor requires calibration whenever a new sensor, Main
PCB is fitted or if door is changed or removed. It is necessary when servicing a pump, to carry out an occlusion pressure test to verify
that the sensor is calibrated correctly (see self test routine in Chapter 3 Routine Maintenance). A calibrated pressure gauge will be
needed in order to perform this calibration.
PRI
SEC
Pressure gauge
± 20mmHg
When the pressure sensor is replaced the null pressure value must be checked and adjusted, if required, prior to
calibration. Check the null pressure value as follows:
1. Enter the access code 12. Go to test 8 and press e.
2. With no set loaded and door open check displayed value is 11 ±4.
3. If the reading is outside of tolerance then adjust R2 on the Pressure Sensor PCB until displayed value is within
tolerance.
Load a set into the pump to be calibrated and prime the set. Connect to pressure gauge as shown in diagram above.
Enter the access code 17.
Apply pressure required for each step and when pressure required is displayed on pressure gauge for 10 seconds (allows pressure to
settle) press e. Calibration values will be returned. Press e to go to next step.
PrES step - 150mmHg ± 40mmHg
HI step - 650mmHg ± 40mmHg.
CAP step - 1000mmHg ± 40mmHg.
Difference step
XX or XXX indicates calibration values, that have
no tolerance values applicable.
YYY indicates calibration values that should be
between 115 and 214.
ZZ indicates a calibration value that is the
difference between XXX and XX and should be
between 37 and 54.
For routine maintenance, the following self-test and performance verification procedures should be performed in addition to the tasks
described in the section on General Cleaning and Inspection for Damage.
Refer to the relevant “Directions for Use” for the recommended routine maintenance period.
Test procedures
Important Service Information:
Important Service Information:
Testing and Calibration of Volumetric pumps is very dependent on the tubing set used.
Testing and Calibration of Volumetric pumps is very dependent on the tubing set used.
For this reason, a new set of tubing should be used for each pump tested, and the tubing
For this reason, a new set of tubing should be used for each pump tested, and the tubing
should be thrown away once all tests are completed. Recommended test set is part number
should be thrown away once all tests are completed. Recommended test set is part number
0000TG00074.
0000TG00074.
Entering Access Codes (Technician Mode)
Note: See Chapter 2 Configuration & Calibration for information on how to enter access codes
CodeTestDescription
1
2
3Main Self TestSee Self Test Routine Table.
4
5Volumetric Accuracy Verification TestSee Volumetric Accuracy Verification Test (code 5) detailed description in this document.
10Alarm history log
11Display current time and date-
12
17Pressure CalibrationRefer to Chapter 2 Configuration & Calibration.
18Volumetric CalibrationThis is a manufacturing code, volumetric calibration should not be carried out.
19Reduced Volumetric CalibrationThis is a manufacturing code, volumetric calibration should not be carried out.
Display volumetric, pressure
calibration and battery Cal values
Automatic Battery Maintenance/
Discharge Test
Access to individual tests within the
Main Self Test
This enables the user to put in their own 4 digit reference number or asset number, together with
the date the pump was last serviced in the format Wk:Yr (15:01) week 15 of 2001. Defaults to 0 &
0:00.
Reference only, 4 values shown:
CAL - 20.00 if pump has not been calibrated (range 16-24)
Pres - DXXX delta value, (range 36 to 55, default to 46)
CAP - CXXX Maximum value, (range 110 to 219, default 163)
bAt - bx.xx (range 6.44 to 7.86, default 7.15)
Takes approximately 2-5 hours, if it takes less than 2 hours it will display ‘fail’ The pump will then
switch to charging, it is recommended this be done for 24 hours.
This will show the last 10 error/alarm codes Use the key to step through.
Note: Not in Sequential Test (code 3), available only through code 12.
Test Nr. Description
9 EEPROM Checksum Test. Display two 16-bit EEPROM check sums. During power
up the processor calculates EPROM checksums values and checks them against
those stored, if a difference is detected a corruption in data has occurred.
11 Pumping Mechanism Test. This test disables the pressure, door and AIL sensors.
The pump will pump into a pressure gauge and display the pressure reached, this is
exited by pressing the enter key.
12 Bubble Measurement Test. This displays the size of the air bubble detected in
Enter access code 3. Press the key to advance to next test.
LevelTestDescription
Confirm display indicates correct button pressed.
When level is entered the display will show “b-1”, press buttons 1 to 8, after pressing button 8 the test
will automatically proceed to level 2.
1Keypad Test
b-1 b-2 b-3 b-4 b-5 b-6 b-7 b-8
Check all LEDs.
The pump will run through a count-up series to illuminate each segment of the 7 segment LEDs, and
2Display Test
3Alarm Test
4Door Test
5Air Sensor Test
7Motor Opto Test
8Occlusion Pressure Test
cycle through all of the green LEDs.
Confirm all LEDs are working. At the end of this test all LEDs will illuminate.
Press the key to advance to level 3.
Confirm the alarm is working and a distinctive change is heard between volume levels. Pump displays
‘ALAR’ and alarms for 0.5 seconds at each volume (1 to 7).
Press the key to advance to level 4.
Confirm the change of state between door open (d-0) and door closed (d-1).
Press the key to advance to level 5.
Confirm the change of state between an air filled set (a-0) and a fluid filled set (a-1).
Press the key to advance to level 7.
Displays PASS or FAIL. Confirm displays PASS. The pump runs the motor forwards a turn, then backwards
a turn. The processor checks that it sees both motor optos come on at the correct time.
Press the key to advance to level 8.
Test requires a calibrated pressure gauge.
Connect the pump IV Infusion set to the pressure gauge via a 3 way tap.
Press the key for 10 seconds. Press the key, the pump will run at 125ml/h and the display will
show the current pressure sensor reading ‘xxx’.
After 10 seconds close off the 3 way tap so that the pump delivers into the pressure gauge.
Confirm that an alarm occurs and a ‘’ appears on the pump display. The pressure displayed on the
pressure gauge should be 500 mmHg +/- 150 mmHg.
Carry out the next test or press the key to advance to level 10.
10Drop Sensor Test
Check for correct drop count.
This test counts the number of drops detected, if the sensor is not present then ‘Off’ will be displayed.
The balances need to be switched on for 30 minutes prior to use to enable the electronics to
The balances need to be switched on for 30 minutes prior to use to enable the electronics to
warm up and settle.
warm up and settle.
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing
Always use new test tubing for each test. If the test ever has to be repeated, a new set of tubing
must be used.
must be used.
Do not move the desk during testing, it will upset the balance readings.
Do not move the desk during testing, it will upset the balance readings.
This test is used to confirm that the pumping accuracy of the system as a whole, including the tubing, is within the specified limits.
In the most controlled conditions, a needle should be used to pump liquid into the weighing beaker to prevent liquid touching the
sides of the beaker and to provide some back-pressure so that leaks/overflows do not affect the readings. As a result of these and
other errors, if the system fails just marginally, it is worth performing the test a second time. If it still fails, return the pump to your local
ALARIS® Service Centre for further analysis. The head height on IV infusion set should be approximately 30cm.
Setup ASetup Bor...
Setup ASetup Bor...
Class A Burette (50ml)
Class A Burette (50ml)
0.1ml graduations
0.1ml graduations
Burette clamp
Burette clamp
Equipment stand
Equipment stand
Table or bench
Table or bench
Needle
Needle
Beaker
Beaker
Scales (±0.01g)
Scales (±0.01g)
Enter access code 5. The pump will initially display “”.
1.
If it is necessary to prime the set, press and hold the button. The pump will display "" and allow the set to be primed,
2.
ignoring any air-in-line alarms.
Release once the set is primed and the pump will again display “”.
Zero scales or burette. Press to begin the verification check.
4.
The pump will automatically pump at 125ml/h for a total of 20ml, which will take approximately 9½ minutes. During the run, it
5.
will display the volume infused up to that point and intermittently flash "r" with the run indicator, to show that this is a
verification run. At the end of this run, the pump should display "20.00" and “wait”.
Allow the scales to settle and then note the reading on the scales. If using a burette, take the final reading of volume infused.
6.
The volume infused should be 20.00ml ± 5%.
Press the key and the pump will revert to the technical service entry mode and flash "CodE", followed by "0", enabling you to
7.
begin other tests if required. Switch the pump "OFF" if there are no further tests to be done at that time.
Flow Stop Mechanism Test
It is necessary to check that the flow stop device can hold sufficient pressure and thus prevent free flow.
Load set into the pump. Use same setup as for pressure calibration (see Chapter 2 Configuration & Calibration).
1.
Open the door and check that the arm stays in the UP position by lifting the arm onto the ledge.
2.
Close the door fully.
3.
Reopen the door and note that the flow stop arm has activated into the down position.
4.
Pull the test tubing out of the pump and place the tube in front of the flow stop mechanism arm in the down position.
5.
Close the door and reopen the door. Check the tubing locator has loaded the tube fully under the clamp and is flat. This checks
6.
whether the tubing locator is the correct way around and works.
Ensure the door is open and the flow stop is closed. Apply 650mmHg to the distal end of the IV infusion set for 10 seconds. Then
7.
reduce pressure to 500mmHg to the distal end and verify pressure gauge reading is 500mmHg. Ensure the pressure does not
drop by more than 15mmHg in 30 seconds.
To ensure that this pump remains in good operating condition, it is important to keep it clean and carry out the routine procedures
described below. All servicing should only be performed by a qualified service engineer, with reference to this manual.
Thoroughly clean external surfaces of the pump by wiping over with a lint-free cloth, lightly dampened with warm water and a
standard disinfectant/detergent solution.
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
- Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Recommended cleaners are:
Brand Concentration
Hibiscrub 20% (v/v)
Virkon 1% (w/v)
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow liquid
to enter the casing and avoid excess fluid build up on the pump. Do not use aggressive cleaning
to enter the casing and avoid excess fluid build up on the pump. Do not use aggressive cleaning
agents as these may damage the exterior surface of the pump. Do not steam autoclave, ethylene
agents as these may damage the exterior surface of the pump. Do not steam autoclave, ethylene
oxide sterilise or immerse this pump in any fluid.
oxide sterilise or immerse this pump in any fluid.
Labels should be flat and legible. Any label, if no longer fully adhered, must be replaced if it represents a path for fluid ingress.
Case components must be checked for damage that may affect function, fluid ingress route or present a user hazard and be re-
placed if necessary.
Check the pole clamp screws are not loose and that the threads are not damaged. Check that it folds away and ensure arm is not
bent.
Inspect the AC power supply plug and cable for damage.
Clean the flow sensor by wiping over with a cloth, lightly dampened with warm water and a
Clean the flow sensor by wiping over with a cloth, lightly dampened with warm water and a
standard disinfectant/detergent solution. Ensure the connector does not get wet. Dry flow sensor
standard disinfectant/detergent solution. Ensure the connector does not get wet. Dry flow sensor
before use.
before use.
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if handles
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if handles
operation is not free, the flow sensor may be immersed and soaked in clean soapy water.
operation is not free, the flow sensor may be immersed and soaked in clean soapy water.
Activating the spring mechanism of the sensor whilst immersed will assist in cleaning the inside
Activating the spring mechanism of the sensor whilst immersed will assist in cleaning the inside
of the mechanism. After cleaning the sensor should be allowed to dry fully prior to use.
of the mechanism. After cleaning the sensor should be allowed to dry fully prior to use.
Caution: the plug of the sensor should not be immersed as damage will occur.
Caution: the plug of the sensor should not be immersed as damage will occur.
Storage
If the pump is to be stored for an extended period it should be cleaned and the internal battery fully charged. Store in a clean, dry
atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in this chapter and ensure that the internal battery is fully
charged.
Please note during long term storage of the pump the Real Time Clock circuitry is being
Please note during long term storage of the pump the Real Time Clock circuitry is being
maintained by BT1 on the control PCB. Under long term storage conditions it is recommended
maintained by BT1 on the control PCB. Under long term storage conditions it is recommended
that the pump is powered in Technician Mode for a period of 24 hours so as to keep the BT1
that the pump is powered in Technician Mode for a period of 24 hours so as to keep the BT1
charged, and eliminate the possibility of depleting BT1 and inducing Err9 faults at power up.
charged, and eliminate the possibility of depleting BT1 and inducing Err9 faults at power up.
The optional upgrade of the Alaris® GW Volumetric Pump software to V5R1F should be considered at the next product service for all
Alaris® GW Volumetric Pumps fitted with software version V4R2C. Perform upgrades by acquiring the software upgrade kits specified
in the spares parts listings. Note: when upgrading Alaris® GW Volumetric Pumps from software version V4R1B, first install the V4R2C
software upgrade kit to enable the flash upload capabilty.
The major features of the V5R1F software include:
Teach / Learn Capability;
Additional configuration options:
Silent Mode;
SELECT Mode Options;
Drop Sensor Connection Mode;
Drop Sensor Light Sensitivity Level;
Alarm volume level factory default now 4 (was 7);
Clear Infusion Parameters to Zero default now OFF (was ON).
Additional Technician Mode Configuration options:
New Volumetric Calibration Mode to reduce calibration time (Code 19);
New EEPROM Memory Management to improve work with Teach / Learn and to eliminate need to recalibrate pumps
following firmware upgrade (Codes 200, 201, 202).
Automatic setting of VTBI to OFF when used with drop sensor;
Elimination of '' error in Mode resulting in nuisance alarms;
VI now cleared in Mode and retained upon power down.
Recalibration is not required when upgrading from software version V4R2C, although all configuration parameters will be returned to
factory defaults.
PC Requirements
Microsoft Windows 95, 98, 2000 or NT operating system
9pin D-type PC serial port or IrDA port.
Tools required
CD-ROM 1000SP00493 - Alaris® GW Volumetric Pump Software Distribution Disk V5R1F
Programming Kit 1000SP00172 (Suitable for all ASENA® Infusion Pumps) or RS232 Cable 1000SP00336
Follow the programming instructions given in publication 1000PB01365 (supplied with 1000SP00493).
1.
Load the software program onto your PC from the Distribution Disk.
2.
Select the Alaris LVP SMU icon (WinSmug)
3.
Place the infrared programming device approximately 50mm directly behind the IrDA window on the rear case or connect
4.
the RS232 cable to the 9 pin D type serial port connector situated on the side of the pump.
Select the appropriate port (e.g. COM1) on the Alaris LVP SMU program and Press 'Upload'.
5.
Briefly depress the key on the pump to be upgraded; the pump will now display "PROG".
6.
Switch the pump on for normal operation when the upgrade is complete. If required, the pump will display whilst
7.
automatically completing a test sequence during which the EEPROM memory will be re-partitioned.
When the upgrade is complete, enter the Technician Mode and verify the correct software version has been installed;
8.
initiate a factory Reset (Code 200).
Perform the Self Test checks (Code 3).
9.
Power failure. Power failures may occur when using laptops when communicating with the Alaris®
Power failure. Power failures may occur when using laptops when communicating with the Alaris®
GW Volumetric Pump due to power requirements. External power may be used in conjunction
GW Volumetric Pump due to power requirements. External power may be used in conjunction
with IrDA or RS232 to compensate for lack of power from the laptop.
with IrDA or RS232 to compensate for lack of power from the laptop.
Bright sunlight and strong fluorescent lighting affect the Infrared programming system. If any
Bright sunlight and strong fluorescent lighting affect the Infrared programming system. If any
errors are reported then the RS232 method of upgrading the software should be used.
errors are reported then the RS232 method of upgrading the software should be used.
For both the teach and learn pumps in Technician Mode enable IrDA communications (Code 45), and ensure that ASCII / Binary
1.
mode and parity bit options (Codes 38 and 39 respectively) are the same.
Turn the teaching pump on in normal operation. Note: For multiple teach-learn operations, to avoid call-back alarm every
2.
2 minutes, turn the teaching pump on in Technician Mode.
Enter Technician Code 67 on the learning pump.
3.
Align the two IrDA ports on the pumps (optimum distance 50 mm).
4.
Depress the key to initiate learning.
5.
A progress bar will travel across the learn pump.
6.
When successful, the learn pump will display 'PASS'.
7.
If the learning pump is unable to learn all configuration parameters then the display will show 'ConF' followed by a list of the
8.
configuration parameters that could not be learnt; these will instead contain the factory default settings. This could occur if, for
example, the software version of the learning pump is newer than that of the teaching pump.
Possible Reasons for failure
Possible Reasons for failure
IrDA not enabled on both pumps;
IrDA not enabled on both pumps;
ASCII / Binary and parity bit options are not the same;
ASCII / Binary and parity bit options are not the same;
If the software versions are not compatible;
If the software versions are not compatible;
If the pump models are different;
If the pump models are different;
The line of sight between the IrDA windows was obstructed during data
The line of sight between the IrDA windows was obstructed during data
transfer.
transfer.
Event Log Download
A PC application known as the Event Log Download Utility (ELDU) is available to download the event log from the Alaris® GW
Volumetric Pump.
ELDU Operation
Click on ELDU icon on PC.
1.
Click Accept to agree with Restrictions of Use and continue;
2.
Select Configure from drop-down menu;
3.
Select Setup Pump and choose Alaris® GW as pump type;
4.
Select Settings to select log to be downloaded;
5.
Check communications are set up as follows:
6.
Required PC Comm port selected
Character type and parity match pump configuration
Click OK to confirm
7.
Align the IrDA converter with the IrDA window (optimum distance 50 mm), or connect an RS 232 cable.
8.
Power up the pump by pressing the key.
9.
Click Download log from the main PC screen.
10.
Press Close when finished.
11.
Select File from drop-down menu and save file. Log may be printed here as required.