Alaris GP User manual

Page 1
Alaris
®
GP Volumetric Pump
Technical Service Manual
Page 2
This manual has been prepared for use by qualified service personnel only.
performance of parts or equipment resulting from unauthorised repair or modification.
t Cardinal Health, 1180 Rolle, Switzerland
Alaris®, IVAC®, Guardrails® and Asena® are registered trademarks of Cardinal Health, Inc. or one of its subsidiaries.
All other trademarks belong to their respective owners.
© 2006-2008. Cardinal Health, Inc. or one of its subsidiaries All rights reserved.
This product uses zlib (http://www.zlib.net/) © 1995-2005 Jean-loup Gailly and Mark Adler Technician Mode uses the lwIP communication stack (http://www.sics.se/~adam/lwip/) © 2001, 2002 Swedish Institute of Computer Science. All rights reserved. Redistribution and use in source and binary forms, with or without modi cation, are permitted provided that the following conditions are met:
1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.
2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
3. The name of the author may not be used to endorse or promote products derived from this software without speci c prior written permission.
THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
Alaris® GP Volumetric Pump 2/84 1000SM00013 Issue 3
Page 3
Contents
Chapter
1. General Information 4
2. Configuration & Calibration 8
3. Preventative Maintenance 19
4. Troubleshooting 39
5. Circuit Descriptions 44
6. Corrective Maintenance 48
Appendix
A. Electromagnetic Compatibility 68
B. Disposal 73
C. Spare Parts Listings 75
D. Service Contacts 81
E. Document History 83
Alaris® GP Volumetric Pump 3/84 1000SM00013 Issue 3
Page 4
General Information
In this chapter
Introduction 5
Features of the Pump 6
Chapter 1
General Precautions 7
Page 5
General Information
Introduction
The Alaris® GP Volumetric Pump and the Alaris® GP Guardrails® Volumetric Pump (hereinafter referred to as 'Pump') is a small lightweight volumetric infusion pump that provides accurate and reliable infusions over a range of rates.
The pump is designed to meet the infusion requirements as specified in the Directions For Use (DFU) for all hospital departments including general wards, critical and intensive care, operating rooms and accident and emergency rooms.
This pump is suitable for use by appropriately trained clinicians or nurses. This pump can be used for intravenous infusion modes. Supporting fluid & drug therapy, blood transfusions and parenteral feeding.
Product Familiarity
Ensure that you are fully familiar with the pump by carefully studying the Directions for Use (DFU) prior to operation and prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are introduced from time to time.
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the Alaris® GP Volumetric Pump and the Alaris® GP Guardrails® Volumetric Pump.
This manual is intended for use by personnel experienced in medical equipment testing and maintenance procedures .
Conventions Used in this Manual
BOLD Used for Display names, self-test codes, controls and indicators referenced in this manual, for
example, Battery Indicator, access code 212, ON/OFF button.
'Single quotes' Used to indicate cross-references made to another section of this manual. For example, see
Chapter 2, 'Configuration & Calibration'.
underline Used to indicate a link to another section within this manual.
Italics Used to refer to other documents or manuals. For example, refer to the relevant Directions for Use
(DFU) for further information. Also used for emphasis, for example, ...if the gap still measures less
than...
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test or maintenance that is important to know about. A typical example is drawing attention to a software upgrade that you should check has been installed.
Alaris® GP Volumetric Pump 5/84 1000SM00013 Issue 3
Page 6
Features of the Pump
Alarm indicator
Display
General Information
Alaris®
GP
Door
Softkeys
Chevrons
Run
Bolus
Hold
Options
AC power indicator
Release lever for rotating cam
Mute
Pressure Battery indicator
On/Off
Door Lever
Handle
Flow sensor connector (cover removed for clarity)
RS232/Nursecall connector (cover removed for clarity)
Rotating cam to lock onto
horizontal rectangular bars.
Mains fuses
Mains inlet
IR communications port
Potential Equalisation (PE)
Connector
Alaris® GP Volumetric Pump 6/84 1000SM00013 Issue 3
Folded pole clamp
Medical device interface (MDI)
Page 7
General Precautions
Attention consult accompanying documents: Prior to using this pump, carefully read the Operating
w V
B
M
Alaris®
GP
Precautions described in the Directions for Use (DFU).
This pump contains static-sensitive components. Observe strict precautions for the protection of static sensitive components when attempting to repair and service the pump.
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care to locate the pump away from any such hazardous sources.
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed. Refer all servicing to qualified service personnel.
This pump is protected against the effects of high energy radio frequency emissions and is designed to be fail safe if extremely high levels of interference are encountered. Should false alarm conditions be encountered, either remove the source of the interference or regulate the infusion by another appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise suspected to have been damaged, remove it from service for inspection by qualified service personnel.
General Information
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If the integrity of the external protective conductor in the installation or its arrangement is in doubt, the pump should be operated from the battery.
Alaris® GP Volumetric Pump 7/84 1000SM00013 Issue 3
Page 8
Chapter 2
Configuration and Calibration
In this Chapter
Entering Service Mode 9
Service Mode - Factory Defaults 10
Service Mode - Configuration 11
Service Mode - Data Set Transfer 14
Service Mode - Calibration 15
Page 9
Configuration and Calibration
Entering Service Mode
Warning -
At no time should Service Mode be entered while the pump is connected to a patient. Service Mode should only be accessed by qualified and trained personnel.
Service Mode can be accessed via a three-digit access code that is entered using the following procedure:
1. Hold down b and turn the pump ON a.
2. Enter the access code 212 using the f keys and the NEXT softkey.
3. When the code shows on screen, press OK to confirm.
Select the required option using the f keys and the OK softkey.
For the Alaris® GP Volumetric Pump the options will be as follows:
SERVICE
Factory Defaults Load a default data set. Confirm to perform a
cold start.
Configuration This menu comprises a list of options which are
configurable by the user.
Data Set Transfer Upload a data set to pump.
Calibration This menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVP Performance Verification Procedure Tests.
For the Alaris® GP Guardrails® Volumetric Pump the options will be as follows:
CQI Events Download For future implementation
Data Set Transfer Upload a data set to pump.
Configuration This menu comprises a list of options which are
configurable by the user.
Calibration This menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVP Performance Verification Procedure Tests.
Factory Defaults Load a default data set. Confirm to perform a
cold start.
Factory Defaults
Configuration Data Set Transfer Calibration Test Verification/PVP
SELECT WITH
QUIT
SERVICE
CQI Events Download
Data Set Transfer Configuration Calibration Test Verification/PVP Factory Defaults
OK
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 9/84 1000SM00013 Issue 3
OK
Page 10
Configuration and Calibration
Service Mode - Factory Defaults
Factory Defaults
Select the required option using the f keys and the OK softkey.
Default Data Set Replace the current data set with a default data
set.
Cold Start Confirm Confirm clearing and resetting the data set and calibration data to the factory defaults.
Clear CQI Log File* Confirm clearing all pump history and resetting the data set to the factory default.
* Alaris® GP Guardrails® Volumetric Pump only.
FACTORY DEFAULTS
Default Data Set
Cold Start Confirm Clear CQI Log File
SELECT WITH
Default Data Set
1. Press OK to confirm loading the default data set.
QUIT
OK
DEFAULT DATA SET
******
Restoring factory default data set will delete the clinically approved data set installed. Delete approved and restore factory default?
QUIT
WARNING!
******
OK
Cold Start Confirm
1. Press OK to confirm performing a cold start.
Alaris® GP Volumetric Pump 10/84 1000SM00013 Issue 3
COLD START CONFIRM
******
All data set and cal data will be reset to default! The pump will require a full calibration before returning to clinical use.
QUIT
WARNING!
******
OK
Page 11
Configuration and Calibration
Service Mode - Factory Defaults (continued)
Clear CQI Log File
1. Press OK to confirm clearing the CQI Log File.
CLEAR CQI LOG FILE
Service Mode - Configuration
Select the required option using the f keys and the OK softkey.
Date/Time Sets the current date and time used for event
logging.
Software Versions Displays the pump software versions.
Serial Number Configure the displayed serial number.
Pump Reference Pump specific text to be displayed in user
mode at start up. (20 characters max.)
Language Configure the Language used for display
messages.
Backlight & Contrast Adjust the Backlight and Contrast values
Current Data Set File Displays the current data set file details.
******
This will remove all instrument history and will restore the factory default data set by deleting the clinically approved installed data set.
Clear current log?
QUIT
******
OK
SERVICE CONFIGURATION
Date/Time
Software Versions Serial Number Pump Reference Language Backlight & Contrast Current Data Set File
SELECT WITH
WARNING!
Date/Time
1. Set the correct date and time using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
QUIT
DATE / TIME
AUG–04–2006
09:48
ADJUST WITH
QUIT NEXT
OK
OK
Alaris® GP Volumetric Pump 11/84 1000SM00013 Issue 3
Page 12
Configuration and Calibration
Service Mode - Configuration (continued)
Software Versions
Press 1. OK to exit after verifying Software Version fitted, display will vary depending on software version fitted.
SOFTWARE VERSIONS
SRP 001.005.004
MP 001.005.004 FDP 001.005.004 SP 001.005.004
QUIT
Serial Number
1. Set the serial number required (maximum 9 characters) using the
f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
OK
SOFTWARE VERSIONS
SWV 001.009.000
PKG 001.009.000 MPU 001.009.000 MPT 001.009.000 FDP 001.007.016 SP 001.005.000 LANGUAGE 001.004.005
QUIT
OK
SERIAL NUMBER
8026 - 00000
Pump Reference
1. Set the Pump Reference required using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
Write Success: .
ADJUST WITH
QUIT NEXT
OK
PUMP REFERENCE
---------------------
ADJUST WITH
QUIT NEXT
OK
Alaris® GP Volumetric Pump 12/84 1000SM00013 Issue 3
Page 13
Configuration and Calibration
Service Mode - Configuration (continued)
Language
1. Select the required Language using the f keys.
2. Press OK to confirm.
Languages available will be dependant on the pump software version.
LANGUAGE
ENGLISH
NORSK - NORWEGIAN DEUTSCH - GERMAN FRANCAIS - FRENCH NEDERLANDS - DUTCH SVENSKA - SWEDISH ESPAñOL - SPANISH ITALIANO - ITALIAN NONE - VANILLA
SELECT WITH
QUIT
OK
Backlight & Contrast
1. Use the f keys to adjust Backlight and Contrast to required
setting.
2. Press PAR AM to change between Backlight and Contrast.
3. Press OK to confirm.
Select the Dimming parameter to see what the display would look like when dimmed.
Current Data Set File
1. Press OK to exit after verifying current data set information.
BACKLIGHT & CONTRAST
Backlight = 30%
Contrast = 70%
Dimming = 25%
ADJUST WITH
QUIT PARAM
OK
CURRENT DATA SET
Data Set ID: ABCD Name: Data_set_EXAMPLE Version: 2 Last update: 01-Jun-2006 12:34 Hospital name: Basingstoke General
QUIT
Alaris® GP Volumetric Pump 13/84 1000SM00013 Issue 3
OK
Page 14
Configuration and Calibration
Service Mode - Data Set Transfer
Upload data set to an Alaris® GP Volumetric Pump
Equipment required:
Alaris® GP Editor Software Kit (1000SP01310) - includes the Alaris® GP Transfer Tool
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979) - optional
USB to RS232 converter 4 way hub (1000EL00980) - optional
PC - for requirements see Upgrading Firmware PC requirements
Using the Alaris® GP Transfer Tool allows a released data set to be uploaded to an Alaris® GP Volumetric Pump.
Warning ­At no time should the Alaris® GP Transfer Tool be used to upload to an Alaris® GP Volumetric Pump while the pump is connected to a patient.
In Service Mode select Data Set Transfer using the f keys and the OK softkey.
Using the Alaris® GP Transfer Tool select data set to be uploaded. 1. Press the 2. RS232 or IrDA softkey to select the Comms mode being used.
3.
Connect the pump to PC. Press the 4. START softkey to begin transfer. Please ensure the data set ID shown on the pump is identical to the one transferred.5. Press 6. PASS softkey to confirm correct transfer and exit. To transfer the data set to another pump repeat steps 2 to 6.7.
Upload data set to an Alaris® GP Guardrails® Volumetric Pump
Equipment required:
Guardrails® Editor V3.1 Software Kit (1000SP01389) or Guardrails® Editor V3.1 Transfer Tool Software Kit (1000SP01390)
RS232 cable (1000SP01183)
USB to RS232 Converter cable (1000EL00979)
USB to RS232 converter 4 way hub (1000EL00980) - optional
PC - for requirements see Upgrading Firmware PC requirements
Using the Guardrails® Editor V3.1 Transfer Tool allows an approved data set to be uploaded to an Alaris® GP Guardrails® Volumetric Pump.
Warning ­At no time should the Guradrails® Editor V3.1 Transfer Tool be used to upload to an Alaris® GP Guardrails® Volumetric Pump while the pump is connected to a patient.
In Service Mode select
Using the 1. Guardrails® Editor V3.1 Transfer Tool Transfer Tool select data set to be uploaded.
2.
Connect the pump to PC. Press the 3. START softkey to begin transfer.
Please ensure the data set ID shown on the pump is identical to the one transferred.
4.
Press 5. ACCEPT softkey to confirm correct transfer and exit. To transfer the data set to another pump repeat steps 2 to 5.6.
Data Set Transfer using the f keys and the OK softkey.
Caution: Loading the Data Set Transfer Tool software is considered a non-clinical service activity. Interconnecting the pump with a PC may cause the safety or electromagnetic environment to change while the connection exists. The threat of higher leakage currents or EMI disturbances may be present. Disconnect the IrDA or RS232 cable connection at both ends following software upload activities.
For more information relating to the Alaris® GP Editor Software or the Guardrails® Editor V3.1 Software refer to the relevant Directions For Use supplied with the
software.
Alaris® GP Volumetric Pump 14/84 1000SM00013 Issue 3
Page 15
Configuration and Calibration
Service Mode - Calibration
Zero Point Calibration
Select the required option using the f keys and the OK softkey.
Wait for the pressure sensors to park.1.
Ensure that an infusion set is not installed.2.
Press the 3. START softkey.
Pump will countdown for 15 seconds.4.
Press the 5. ACCEPT softkey.
Press the 6. PASS softkey.
CALIBRATION
Zero Point Cal
Pressure Cal Volumetric Cal Battery Cal
SELECT WITH
QUIT
OK
Pressure Calibration
Equipment required:
Calibrated Pressure Gauge, minimum specification of
Accuracy = 0.10% of full scale
Full scale = 1500mmHg
Pressure Calibration Set (1000SP01422) Use to calibrate 10 pumps • and then change
Set up equipment as per figure 2-1 and allow 30 seconds before proceeding.
Select the required option using the f keys and the OK softkey.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal Battery Cal
SELECT WITH
QUIT
OK
Alaris® GP Volumetric Pump 15/84 1000SM00013 Issue 3
Page 16
Configuration and Calibration
Service Mode - Calibration (continued)
Pressure Calibration continued
Pressure source (50ml/100ml syringe or similar device)
Pressure Gauge
Figure 2 - 1 Pressure Calibration Equipment Set Up
Vent to atmosphere
Turn 3-way tap to close from atmosphere.1.
Press the 2. START softkey.
Apply a pressure of 200mmHg and pump will countdown for 15 seconds.3.
Press the 4. ACCEPT softkey.
Apply a pressure of 800mmHg and pump will countdown for 15 seconds.5.
Press the 6. ACCEPT softkey.
Turn 3-way tap to vent to atmosphere.7.
Turn 3-way tap to close from atmosphere. 8.
Press the 9. NEXT softkey to proceed to Verification Procedure.
Alaris®
GP
Verification Procedure
Software version v1.7.x and below Software version v1.9.0 and above
Apply a pressure of 200mmHg and wait for 5 10. seconds.
Press the 11. NEXT softkey.
Apply a pressure of 400mmHg and wait for 5 12. seconds.
Apply a pressure of 500mmHg and wait for 5 10. seconds.
Press the 11. NEXT softkey.
Turn 3-way tap to vent to atmosphere.12.
Press the 13. PASS softkey.
Press the 13. NEXT softkey.
Apply a pressure of 600mmHg and wait for 5 14. seconds.
Press the 15. NEXT softkey.
Apply a pressure of 800mmHg and wait for 5 16. seconds.
Press the 17. NEXT softkey.
Turn 3-way tap to vent to atmosphere.18.
Press the 19. PASS softkey.
Alaris® GP Volumetric Pump 16/84 1000SM00013 Issue 3
Page 17
Configuration and Calibration
Service Mode - Calibration (continued)
Volumetric Calibration
Select the required option using the f keys and the OK softkey.
Load the primed Infusion Set (60793) into the Pump and set-up 1. as shown in Figure 2-2 below and adjust the fluid level so that the meniscus is level with the zero mark.
Press 2. START to begin. Test will run and fluid will be delivered into the burette.
When 3. Volume delivered! is displayed, check accumulated air in line value is less than 100µl then enter the volume delivered into the burette using the f keys and the OK softkey. If accumulated air in line value is greater than 100µl then repeat test.
If measured value is 19.3ml or less then enter
19.2ml and if value is 20.6ml or higher then enter
20.7ml.
CALIBRATION
Zero Point Cal Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
If no calibration is required (Volume delivered within limits of 4.
19.4ml to 20.5ml) then press PASS to confirm and exit.
If the Calibration value is changed automatically then press 5. VERIFY and repeat steps 2 to 4.
If the pump still fails replace the Platen and Fingers then repeat 6. the calibration procedure.
450mm
± 20mm
Alaris®
GP
VOLUMETRIC CALIBRATION
Rate 125ml/h VTBI 20.0ml Cal Value 920 Acc’d Air 0μl
START to begin...
QUIT
STARTDETAILS
Figure 2 - 2 Volumetric Calibration Equipment Set Up
Alaris® GP Volumetric Pump 17/84 1000SM00013 Issue 3
Page 18
Configuration and Calibration
Service Mode - Calibration (continued)
Battery Calibration
Select the required option using the f keys and the OK softkey.
Connect AC Mains to the Pump and press 1. CAL to begin. When calibration is complete it will display 2. CALIBRATION
SUCCESS or CALIBRATION FAILURE. Press PASS to confirm successful calibration or REPEAT to perform calibration again.
CALIBRATION
Zero Point Cal Pressure Cal Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
BATTERY CALIBRATION
Serial No 12345 Current 0mA FCC 2502mAh Chrg Remain 2453mAh Rel Chrg 98% Temperature 22°C Calibration
CAL to calibrate!
CAL can exceed 10 hrs!
QUIT
DETAILS
CAL
BATTERY CALIBRATION
Serial No 12345 Current 0mA FCC 2335mAh Chrg Remain 2332mAh Rel Chrg 100% Temperature 26°C Calibration
BATTERY CALIBRATION
Serial No 12345 Current -340mA FCC 2502mAh Chrg Remain 2453mAh Rel Chrg 98% Temperature 22°C Calibration
CALIBRATING . . .
Discharge
QUIT
DETAILS
89%
CAL
BATTERY CALIBRATION
Serial No 12345 Current 0mA FCC 2502mAh Chrg Remain 2453mAh Rel Chrg 100% Temperature 22°C Calibration X
CALIBRATION SUCCESS
QUIT
Alaris® GP Volumetric Pump 18/84 1000SM00013 Issue 3
REPEAT
PASS
CALIBRATION FAILURE
QUIT
REPEAT
Page 19
Chapter 3
Preventative Maintenance
In this chapter
Preventative Maintenance 20
Visual Inspection 20
Recommended Cleaning and Storage 21
Updates 23
Battery Test and Replacement 25
Service Mode - Test Verification/PVP 25
Performance Verification Procedure 38
Page 20
Preventative Maintenance
Preventative Maintenance
To ensure the pump remains in good operating condition, routine and preventative maintenance inspections are required. Routine maintenance inspections should be performed by hospital/facility before each use, see Directions For Use for details.
Preventative maintenance inspections should be performed at least every year.
For the preventative maintenance inspection the following should be performed:
Full visual inspection of the pump, internal and external
Fitting of all updates required
Battery test and/or replacement
Clean the pump
Performance Verification Procedures
Following all spare part replacement and repair activities, testing must be performed in accordance with the Performance Verification Procedure (PVP). Additional testing and calibration may be required after certain repairs are completed, see table in Chapter 6 ‘Corrective Maintenance’ for more information.
Visual Inspection
Open the pump, as per Chapter 6 ‘Corrective Maintenance’ and visually inspect the interior of the pump.
Visually inspect the exterior of the pump checking the following:
Labels should be replaced as required if not flat, legible or fully adhered.
Check Keypad for any sign of wear and replace as required.
Case components must be checked for damage and replaced if necessary.
Check the pole clamp is not damaged and that it functions correctly.
Inspect the AC power supply plug and cable for damage.
The case should be clean and free from IV solution residue, especially near moving parts.
Check for dried solution deposits on accessible areas of pumping mechanism.
Alaris® GP Volumetric Pump 20/84 1000SM00013 Issue 3
Page 21
Preventative Maintenance
Recommended Cleaning and Storage
Cleaning the Pump: -
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution.
Recommended cleaners are:
Brand Concentration
Hibiscrub 20% (v/v) Virkon 1% (w/v)
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
- Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Before cleaning always switch OFF and disconnect from the AC power supply. Never allow fluid to enter the casing and avoid excess fluid build up on the Pump. Do not use aggressive cleaning agents as these may damage the exterior surface of the Pump. Do not steam autoclave, ethylene oxide sterilise or immerse this Pump in any fluid.
Storing the Pump: -
If the Pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection. Once every 3 months during storage, carry out functional tests as described in this technical service manual and ensure that the internal battery is fully charged.
Cleaning and storing the Infusion Set: -
The Infusion Set is a disposable single use item and should be discarded after use according to hospital protocol.
Alaris® GP Volumetric Pump 21/84 1000SM00013 Issue 3
Page 22
Preventative Maintenance
Recommended Cleaning and Storage (continued)
Cleaning the door: -
Periodically during use (as per hospital policy), clean the door by wiping over with a lint-free cloth, lightly dampened with warm water and a standard disinfectant / detergent solution. Dry door before use. To aid cleaning of a door which has been heavily soiled, contaminated or if the door operation is not free moving, then the door may be removed (see procedure below) then immersed and soaked in warm water with a standard disinfectant / detergent. The door should be allowed to dry fully prior to use.
Door Removal
1. Remove the screw securing the lower hinge lock.
2. Open the lower hinge lock.
3. Pull the door away from lower hinge pin and lift up to remove the door.
4. Clean the door.
5. Refit door in reverse order. Ensure screw is refitted with a torque of 70cNm.
Cleaning the Flow Sensor: -
Before the transfer of the flow sensor to a new infusion set and periodically during use, clean the flow sensor by wiping over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution. Ensure the connector does not get wet. Dry flow sensor before use. To aid cleaning of flow sensors which have been heavily soiled, contaminated or if the handle operation is not free moving, then the flow sensor may be immersed and soaked in clean soapy water (see note below). The inside of the spring mechanism can be cleaned by activating it whilst submerged in the water. After cleaning, the sensor should be allowed to dry fully prior to use.
Warning -
The plug of the flow sensor must not be immersed in water as damage will occur.
Alaris® GP Volumetric Pump 22/84 1000SM00013 Issue 3
Page 23
Updates
Upgrading firmware
Requirements
PC • Minimum hardware system requirements: Software requirements:
1GHz Pentium processor• 512MB RAM• Microsoft Windows 2000 • 1GB of free space on the system hard
drive Available configurable RS232 9 pin serial
or USB communications ports
Preventative Maintenance
CD ROM drive
K
eyboard
(service pack 4), or XP (service pack 2)
Video resolution of 1024 X 768 pixels
Mouse
and 16 bit colour depth The Software Maintenance Utility (SMU) (1000CD00028)• RS232 cable (1000SP01183) • USB to RS232 Converter cable (1000EL00979) - optional• USB to RS232 converter 4 way hub (1000EL00980) - optional• Firmware CD
Preparation
Load the SMU software onto the PC• Connect RS232 connector (using USB to RS232 converter if required) to each pump being updated• Disconnect the Battery Remove the Battery Compartment Plug
Battery Compartment Plug
Software Upload
1. Through the Rear Case, there is a set of three dip switches on the bottom of the Control PCB. Switch 1 should be ON and switches 2 and 3 should be OFF.
2. Using a small flat blade screwdriver or round ended tweezers configure the dip switch settings to:
Switch 1 = OFFa. Switch 2 = ONb. Switch 3 = ON c.
3. Start the SMU facility to upload the software.
4. Select the Firmware zip file.
5. Select the Comm settings - Comm Port and Baud rate of 460800 (USB to RS232 converter) or 115200 (RS232).
6. Connect the Battery to the Battery Cable.
7. Press Start.
Switch 1 turns the safety battery circuit on but is not required for programming. Switch 2 forces the pump to turn on. Switch 3 turns the pump into ‘Boot Mode’, this is only used for programming.
Alaris® GP Volumetric Pump 23/84 1000SM00013 Issue 3
Page 24
Preventative Maintenance
Updates (continued)
Upgrading firmware continued
8. Once the green bar has reached the far right hand side and the time has reached 0:00 and the flashing green light is a steady green light, the RS232 connector can be removed from the pump.
9. Disconnect the Battery and turn OFF all dip switches.
10. Wait 5 seconds then reconnect the Battery.
11. Configure the Dip Switches to:
Switch 1 = ONa. Switch 2 = ON then OFFb. Switch 3 = OFFc.
12. Refit the Battery Compartment Plug, this prevents fluid ingress.
13. Power up the pump in Service Mode, enter access code 212, then select Configuration > Date/Time and set the current
date and time.
If the Control, Interface or RS232 PCB is replaced, the pump must be re-programmed.
Alaris® GP Volumetric Pump 24/84 1000SM00013 Issue 3
Page 25
Preventative Maintenance
Battery Test and Replacement
To test the battery perform the battery calibration, as outlined in the procedure in Chapter 2 ‘Configuration and Calibration, and verify that all pass criteria are met. If pass criteria are not met then replace the battery.
Battery charge retention will eventually degrade. So where retention is critical the internal battery should be replaced every three years.
Replace the Main Battery
Remove the two case screws in battery cover, remove 1. cover and battery.
Fit new battery.2. Replace battery cover and secure with 2 screws.3.
Battery Cover
It is essential that the battery pack is calibrated after fitting as failure to do so will degrade the quoted auxiliary battery power on this product.
Battery pack
Screw/Washer (x2) 40cNm
Service Mode - Test Verification/PVP
Warning -
At no time should Service Mode be entered while the pump is connected to a patient. Service Mode should only be accessed by qualified and trained personnel.
Enter access code 212 to view the Service Mode menu (see 'Entering Service Mode' in Chapter 2 for instructions).
Select the Test Verification/PVP option using the f keys and the OK softkey. Press the Quit softkey to return the pump to the access code screen.
Select the required option using the f keys and the OK softkey.
PVP Work Flow Performance Verification Procedure (PVP) tests. Alarms Functionality Test the Alarms function correctly. User Interface Display, Keypad, LEDs and Audio Tests. Power Supplies Test AC Mains and Battery operation. Sensor Tests Perform tests on the pump sensors. Comms Tests RS232, IrDA and Nurse Call Tests.
TEST/VERIFY
PVP Work Flow
Alarms Functionality User Interface Power Supplies Sensor Tests Comms Tests
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 25/84 1000SM00013 Issue 3
OK
Page 26
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
PVP Work Flow
This test is used to con rm that the Pump is functioning correctly.
Press the START softkey to begin.
The PVP Work Flow will iterate through the following tests:
Software Versions
Date/Time (read only)
Serial Number (read only)
Door Frame
Safety Clamp
Sear
Audio Test
Chequerboard
LED’s
Keypad
Nurse Call
RS232 Loop Back
Pumping E ciency Test
Downstream Occlusion Pressure Test
PVP WORK FLOW
PVP Work Flow Test
QUIT
START
Alarms Functionality
Volumetric (Accuracy Test) Calibration
See individual test details in this chapter or in Chapter 2 'Con guration and Calibration' for instructions.
Alaris® GP Volumetric Pump 26/84 1000SM00013 Issue 3
Page 27
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Alarms Functionality
1. Press the START softkey to begin.
ALARMS FUNCTIONALITY
2. Pump starts an infusion.
3. Check all alarms sound and display correctly.
4. Alarms to test are:
AC Mains Disconnect
Door Open
Upstream occlusion
Air In Line
****
Ensure that a fluid filled
*****
set is in use.
Subsequent alarm screens should be cleared using the CANCEL softkey.
INFORMATION!
QUIT
START
ALARMS FUNCTIONALITY
AC Mains . Door Open . Upstream Occ . Air In Line .
5. When all the alarms have been tested press the PASS softkey if all the
alarms worked correctly or REPEAT softkey if alarms still need to be checked.
Waiting for alarm . . .
QUIT
STOP
ALARMS FUNCTIONALITY
AC Mains Door Open Upstream Occ Air In Line
All alarms detected
QUIT
REPEAT
PASS
Alaris® GP Volumetric Pump 27/84 1000SM00013 Issue 3
Page 28
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
User Interface
Select the required option using the f keys and the OK softkey.
Chequerboard Display Chequerboard Test. Keypad Tests Perform Keypad Test to check all keys work when
pressed.
LED Tests Check all LEDs display in correct order.
Audio Tests Check Primary and Secondary Audio sounders.
USER INTERFACE
Chequerboard
Keypad Tests LED Tests Audio Tests
SELECT WITH
Chequerboard Pattern
1. Check pattern is displayed correctly and to pass. If pattern is incorrect
then press
2. Press PASS softkey to confirm pass and exit.
to fail.
QUIT
QUIT
OK
Keypad Tests
1. Press the START softkey to begin.
2. Press all the keys and a
3. Press PASS softkey to confirm pass and exit.
will indicate each key to pass.
KEYPAD TESTS
[ . ] [ . ] [ . ]
[ . ] [ . ] [ . ] [ . ]
[ . ] [ . ] [ . ] [ . ] [ . ] [ . ]
START to begin...
START
Alaris® GP Volumetric Pump 28/84 1000SM00013 Issue 3
Page 29
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
LED Tests
1. Press the START softkey to begin.
2. Check LEDs are displayed correctly and
displayed then press
3. Press PASS softkey to confirm pass and exit.
to fail.
to pass. If an LED is not
LED TESTS
Safety Processor Primary Red Secondary Red Primary Amber Secondary Amber Start Stop Battery
START to begin...
QUIT
START
Audio Tests
1. Press the START softkey to begin.
2. Check Audio sounds are correct and
correct then press to fail.
3. Press PASS softkey to confirm pass and exit.
to pass. If Audio sounds are not
AUDIO TESTS
Primary Audio Working Test
Secondary Audio Test
START to begin...
QUIT
STARTDETAILS
Alaris® GP Volumetric Pump 29/84 1000SM00013 Issue 3
Page 30
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Power Supplies
Select the required option using the f keys and the OK softkey.
AC Mains Test Tests the AC mains removal detection.
Battery Test the battery. To perform calibration see Chapter
2 'Configuration & Calibration'.
POWER SUPPLIES
AC Mains Test
Battery
SELECT WITH
AC Mains Test
1. Press the START softkey to begin.
2. Check AC Mains connected/disconnected is correctly indicated and press
softkey or press softkey if not correctly detected.
3. Press PASS softkey to confirm pass and exit.
QUIT
OK
AC MAINS TEST
AC Connection
START to begin...
QUIT
STARTDETAILS
Battery
1. Review the battery information.
2. Press the DETAILS softkey to see further battery details.
3. Press QUIT softkey to exit.
Alaris® GP Volumetric Pump 30/84 1000SM00013 Issue 3
BATTERY CALIBRATION
Serial No 1001 Current 0mA FCC 2714mAh Chrg Remain 2477mAh Rel Chrg 91% Temperature 25°C
QUIT
DETAILS
Page 31
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Sensor Tests
In Sensor Tests menu select required test using the f keys and the OK softkey.
Flow Sensor Test Check Flow Sensor is connected and drops
count.
Door Frame Test Check door registers as open and closed. Safety Clamp Test Check Safety clamp detection registers clamp
enabled or disabled.
Sear Test Check Sear detection registers sear enabled or
disabled.
Air In Line Test Check Air In Line sensor detects fluid and air
correctly.
Run-In Mode Performs a continuous infusion for burn in
testing.
Pumping Efficiency Test
Pump Finger Height
or Parking Test*
* Test name has changed to Parking Test for latest software versions however the procedure is the same.
This test is used to con rm that the Pump is able to generate su cient pressure.
Allows the mechanism to be parked.
SENSOR TESTS
Flow Sensor Test
Door Frame Test Safety Clamp Test Sear Test Air In Line Test Run-In Mode Pumping Effic’y Test Pump Finger Height
SELECT WITH
QUIT
OK
Flow Sensor Test
1. Plug flow sensor into connector on rear of the pump.
2. Press the START softkey to begin.
3. Check Flow sensor operation is correct and press
displayed correctly or press softkey if drops are not detected.
4. Press PASS softkey to confirm pass and exit.
softkey if drops are
FLOW SENSOR TEST
Working Connected Rate 125ml/h VTBI 20.0ml Drop count Test
START to begin...
QUIT
DETAILS
START
Alaris® GP Volumetric Pump 31/84 1000SM00013 Issue 3
Page 32
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Door Frame Test
1. Press the START softkey to begin.
2. Check Door open/closed is correctly indicated and press
press
softkey if not correctly detected.
3. Press PASS softkey to confirm pass and exit.
softkey or
DOOR FRAME TEST
Door Frame Test
START to begin...
QUIT
START
Safety Clamp Test
1. Press the START softkey to begin.
2. Check Clamp enabled/disabled is correctly indicated and press
or press softkey if not correctly detected.
3. Press PASS softkey to confirm pass and exit.
Sear Test
1. Press the START softkey to begin.
2. Check Sear in/out is correctly indicated and press
softkey if not correctly detected.
3. Press PASS softkey to confirm pass and exit.
softkey or press
softkey
SAFETY CLAMP TEST
Working Clamp enabled Test
Opto Tx En 0mV Opto Tx Dis 0mV
START to begin...
QUIT
START
SEAR TEST
Working Sear in place Test
START to begin...
QUIT
Alaris® GP Volumetric Pump 32/84 1000SM00013 Issue 3
DETAILS
START
Page 33
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Air In Line Test
1. Press the START softkey to begin.
2. Insert a fluid filled Infusion Set and an air filled Infusion Set.
3. Confirm pump detects fluid and air correctly and press
press softkey if not correctly detected.
4. Press PASS softkey to confirm pass and exit.
softkey or
AIR IN LINE TEST
Upstream: INVALID Detecting Downstream: VALID Detecting
START to begin...
QUIT
DETAILS
START
Run-In Mode
1. Load an Infusion Set.
2. Set the rate required using the f keys and press the START
softkey to begin test.
3. Press STOP softkey when test is completed.
4. Press PASS softkey to confirm pass and exit.
RUN-IN MODE
Rate 125ml/h
CHANGE RATE
START to begin...
QUIT
START
Alaris® GP Volumetric Pump 33/84 1000SM00013 Issue 3
Page 34
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Pumping E ciency Test
This test is used to con rm that the Pump is able to generate su cient pressure. This is done by infusing into a calibrated pressure gauge and checking that the correct line pressure is achieved. The test set-up is as per  gure 3 - 1.
1. Insert the Infusion Set (60793) and the in-line roller clamp closed to
prevent  uid  ow.
2. Close the door and open the roller clamp on the set. Ensure that the
3-way tap to the transducer is closed to the atmosphere.
3. From the Pressure System Test menu screen, highlight Pumping E c'y
Test and press OK softkey.
4. Press the RATE softkey to select a rate of 50ml/h.
5. Press the START softkey and start the timer.
6. When 1000mmHg is reached stop the timer and then press the STOP
softkey and open the 3-way tap to atmosphere.
7. Record that the time taken to reach 1000mmHg was 2 minutes or less.
8. Press the PASS softkey to con rm pass and exit.
Note: If a DRV2 fault code is encountered during the pumping e ciency
test and the pressure has exceeded 1000mmHg, the fault should be ignored, and the pump power should be cycled to reset the condition.
PUMPING EFFICIENCY TEST
Rate 50ml/h
RATE for 50/125 ml/h START to begin...
QUIT
STARTRATE
Pressure Gauge
150 mm ± 5 mm
Figure 3 - 1 Pressure Tests Equipment Set Up
3-way tap
Pump Finger Height (Parking Test)
1. Press the START softkey to begin.
2. Mechanism will run and park.
3. Press DONE softkey to confirm and exit.
450 mm
± 20 mm
Alaris®
GP
PUMP FINGER HEIGHT
Parked
START to begin...
QUIT
START
Alaris® GP Volumetric Pump 34/84 1000SM00013 Issue 3
Page 35
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
COMMS Test
Select the required option using the f keys and the OK softkey.
IrDA Test Check IrDA operates correctly.
RS232 Loop Back Check RS232 operates correctly.
Nurse Call Test Check Nurse Call operates correctly.
COMMS TESTS
IrDA Test
RS232 Loop Back Nurse Call Test
IrDA Test requires specialist equipment.
For further details please contact Cardinal Health.
RS232 Loop Back
1. Link pins 2 & 3 of the RS232 connector on rear of the pump.
2. Press the START softkey to begin.
3. Check RS232 operation is correct and
item. If RS232 Test fails a is displayed to indicate the failure.
4. Press PASS softkey to confirm pass and exit.
for pass are shown after each
SELECT WITH
QUIT
OK
RS232 LOOP BACK
Message sent 0 Message received 0 Transmit count 0 Received count 0
Apply loopback by
connecting pins 2 & 3
QUIT
START
NURSE CALL TEST
1. Press the START softkey to begin.
2. Check Nurse Call operation is correct and
each item. If Nurse Call fails a is displayed to indicate the failure.
3. Press PASS softkey to confirm pass and exit.
for pass are shown after
NURSE CALL TEST
Output Monitor Test
START to begin...
QUIT
Alaris® GP Volumetric Pump 35/84 1000SM00013 Issue 3
DETAILS
START
Page 36
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Occlusion Test
This test can be only done as part of the PVP Work Flow. Use the Infusion Set ten times only and then replace. Record how many times the Infusion Set has been used.
Note: The Occlusion Pressure Test is carried out with  uid in the Infusion Set. This test is used to con rm that the pressure sensor is correctly calibrated and able to detect an occlusion at the correct line pressure. This is done by pumping into a calibrated pressure gauge and checking that an alarm occurs at the correct line pressure. The test set-up is as per  gure 3 - 1.
Put the  uid  lled Infusion Set (60793) into the Pump.1. Enter the 2. PVP Work Flow and proceed to the Occlusion Pressure Test. Open the 3-way tap to atmosphere then press the 3. LEVEL softkey to adjust the alarm level to L5. Con gure the Calibrated Pressure Gauge to hold the Peak/MAX Pressure reading, in preparation for the test.4. Press the 5. START softkey to begin running the Pump at a rate of 125ml/h. Allow the Pump to run for 1 minute, so that the pressure reading stabilises. Turn the tap to occlude the Infusion Set into the pressure gauge. 6. The Pump will continue to infuse and it will be observed that the pressure reading increases. Eventually a high-pressure 7. alarm will occur and the Pump will stop infusing. Note the reading on the pressure gauge and con rm that it is 500mmHg
±100mmHg.
7. Press PASS softkey if Pump passes test at all levels.
If the pressure is outside of tolerance pressure calibration is required. Calibration should be performed as per procedure in Chapter 2 ‘Configuration and Calibration’. If the pump continues to fail the occlusion
test then the pressure sensors should be replaced and perform the calibration procedure again.
Occlusion Test (Optional)
This test can be done in normal operating mode to check the occlusion without having to perform the full PVP Work Flow. Use the Infusion Set ten times only and then replace. Record how many times the Infusion Set has been used.
Note: The Occlusion Pressure Test is carried out with  uid in the Infusion Set. This test is used to con rm that the pressure sensor is correctly calibrated and able to detect an occlusion at the correct line pressure. This is done by pumping into a calibrated pressure gauge and checking that an alarm occurs at the correct line pressure. The test set-up is as per  gure 3 - 1.
Put the fluid filled Infusion Set (60793) into the Pump.1.
Press the2. a button to turn the pump on. Open the 3-way tap to atmosphere. 3. Set the Rate to 125ml/h.4. Press the 5. VTBI softkey and set VTBI to 10ml.
Press the 6. e button and set the Pressure Alarm Limit to L5.
Con gure the Calibrated Pressure Gauge to hold the Peak/MAX Pressure reading, in preparation for the test.7.
Press the 8. b button to begin running the Pump at a rate of 125ml/h. Allow the Pump to run for 15 seconds, so that the pressure reading stabilises.
Turn the tap to occlude the Infusion Set into the pressure gauge. 9. The Pump will continue to infuse and it will be observed that the pressure reading increases. Eventually a high-pressure 10.
alarm will occur and the Pump will stop infusing. Note the reading on the pressure gauge and confirm that it is 500mmHg ±100mmHg.
Open the 3-way tap to atmosphere. 11.
Hold the12. a button down for approximately three seconds to turn the pump off.
If the pressure is outside of tolerance pressure calibration is required. Calibration should be performed as per procedure in Chapter 2 ‘Configuration and Calibration’. If the pump continues to fail the occlusion
test then the pressure sensors should be replaced and perform the calibration procedure again.
Alaris® GP Volumetric Pump 36/84 1000SM00013 Issue 3
Page 37
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Volumetric Accuracy
This test can be done as part of the PVP Work Flow or in the calibration menu.
Load the Infusion Set (60793) into the Pump and set-up as shown in 1. Figure 3-2 below and adjust the fluid level so that the meniscus is level with the zero mark.
Press 2. START to begin. Test will run and fluid will be delivered into the burette.
When 3. Volume delivered! is displayed, check accumulated air in line value is less than 100μl then enter the volume delivered into the burette using the f keys and the OK softkey. If accumulated air in line value is greater than 100μl then repeat test.
If measured value is 19.3ml or less then enter 19.2ml and if values is 20.6ml or higher then enter 20.7ml.
VOLUMETRIC CALIBRATION
Rate 125ml/h VTBI 20.0ml Cal Value 920 Acc’d Air 0μl
START to begin...
QUIT
STARTDETAILS
If no calibration is required then press 4. PASS to confirm and exit. If the Calibration value is changed automatically then press 5. VERIFY
and repeat steps 2 to 4. If the pump still fails replace the Platen and Fingers then repeat the 6. calibration procedure.
450 mm ± 20 mm
Alaris®
GP
Figure 3 - 2 Volumetric Accuracy Equipment Set Up
Alaris® GP Volumetric Pump 37/84 1000SM00013 Issue 3
Page 38
Preventative Maintenance
Performance Verification Procedure
Model / Serial Number: Service Order / Inventory Number:
Hospital Name / Reference: Software Version:
INSPECTION Physical inspection and clean
Check all functions in PVP Work Flow
Enter access code 212 and go to PVP Work Flow
Software Versions
Date/Time
Serial Number
Door Frame
Safety Clamp
Sear
Audio Test
Chequerboard
LED’s
SELF TEST
CH3
Keypad
Nurse Call
RS232 Loop Back
Pumping E ciency Test
Time taken = 2 minutes or less
CH3
_____ minutes _____ seconds
Downstream Occlusion Pressure Test
Occlusion alarm = 500 ± 100 mmHg
Alarms Functionality
Volumetric (Accuracy Test) Calibration
Delivery = 20 ml ± 0.6 ml (3%)
SETUP Set rate to zero (or lowest value possible), clear Volume Infused and VTBI
Class I Type CF
Alternatively attach printed test results
Earth Resistance Test <= 0.2 Ω
ELECTRICAL
SAFETY TESTS
Earth Leakage Current <= 500 μA
Enclosure Leakage Current <= 100 μA
Verification Performed By ____________________________
Sign
CHX
indicates the chapter number in the Technical Service Manual (TSM) - 1000SM00013.
CH2
E.G.
= Refer to TSM Chapter 2.
____________________________
Print
L5 ___________ mmHg
_____________ ml
_____________ Ω
_____________ μA
_____________ μA
_____________________
Date
Alaris® GP Volumetric Pump 38/84 1000SM00013 Issue 3
Page 39
Troubleshooting
In this Chapter
Chapter 4
Log Downloads 40
Introduction 41
Software Fault Codes 41
General Fault Diagnosis 43
Exception Error Handling 43
Page 40
Troubleshooting
Log Downloads
PC Setup (first time only)
1. Navigate through the Start menu, select Settings, then Network Connections.
2. Select New Connection Wizard.
3. Click Next.
4. Select Set up an advanced connection option and click Next.
5. Select Connect directly to another computer option and click Next.
6. Select Guest option and click Next.
7. Enter AlarisGP as the Computer Name and click Next.
8. Select the appropriate COM port and click Next.
9. Select the Connection Availability required and click Next.
10. Tick check box if a shortcut is required on the desktop and click Finish.
11. Connect AlarisGP Dialog box is displayed and click Properties.
12. On General tab click Configure.
13. Set Maximum speed (bps): to 115200, uncheck Enable Hardware flow control and click OK.
14. On Options tab check Display progress while connecting and uncheck Prompt for name and password,
certificate, etc..
15. On Security tab click Settings.
16. Check Unencrypted password (PAP) only and click OK. Click Yes on the confirmation dialog that is displayed
17. On Networking tab click Settings.
18. Check Enable LCP extensions and Enable software compression then click OK.
19. Check Internet Protocol (TCP/IP) and QoS Packet Scheduler, highlight Internet Protocol (TCP/IP) then click
Properties.
20. Check Use the following IP address and enter an IP address of 192.168.3.2 then click Advanced.
21. Uncheck U
22. Click OK.
23. Click OK.
24. The PC will dial the pump, refer to download procedure.
se default gateway on remote network and click OK.
PC Setup (second time)
1. Navigate through the Start menu, select Settings, then Network Connections AlarisGP.
2. The PC will dial the pump.
3. Refer to download procedure.
Event Log Download
1. Switch the pump on in Service Mode.
2. Once communication is established open a web browser and enter http://192.168.3.1 into the address bar.
3. Download log.
Warning ­At no time should the Event Log be downloaded while the pump is connected to a patient.
For pumps with software version v1.9.x and above also download the Presentation Style Sheet to enable the logs to be viewed (this file only needs to be downloaded once). Also the downloaded event log needs to be stored in the same directory as the Presentation Style Sheet. To view the downloaded event log open file with Microsoft Excel and select style sheet.
Alaris® GP Volumetric Pump 40/84 1000SM00013 Issue 3
Page 41
Troubleshooting
Troubleshooting
Introduction
Use this troubleshooting guide to help identify the cause of errors and faults which may occur as a result of damage to the pump or failure of an internal component. The following table lists the error messages and describes what action to take to resolve the problem. A general fault diagnosis checklist is also provided. For information on alarm procedures and messages, refer to the DFU.
Software Fault Codes
Code
DFS1 Door Flow Stop Flow Stop Sensor Fault Clean AIL/Safety Clamp Housing. DFS2 Sear Sensor Fault
DFS3 Platen Fault Check door sensor and door are correctly DFS4 Hall Fault
DFS5 Pressure System Fault Pressure Sensors or Interface PCB. DRV1 Drive Park Fault
DRV3 Linearisation Fault DRV4 Inhibit Fault Check Connections between Control and
DRV5 Rate Control Fault Chassis or Interface PCB. DRV6 Calibration Fault Calibrate the pump.
DSP1 Downstream Pressure Sensor Fault Downstream pressure sensor, Interface PCB
DSP2 Calibration Fault Calibrate pressure.
FLD1 Fluid Channel Stale Fault Check Connections between Control and FLD2 Volume Display Fault
FLW1 Drip Chamber Measurement Fault Check Flow Sensor. Try another Flow
HDW1 Hardware Excess Interrupts HDW2 Stale HDW3 Platform Fault HDW4 Serial Number Corrupt HDW5 ADC Reference Failure IFS1 File System Persistent Storage Fault SD Card or Control PCB. IFS2 Policies Cfg Fault Con gure and calibrate pump.
Module Failure Action/Replace
Check connections and cables. AIL/Safety Clamp Housing.
positioned and not damaged. Door sensor, door or pressure sensors.
Chassis or Interface PCB.DRV2 Motor Control Fault
Interface PCBs. Chassis, Interface PCB or Control PCB.
SD Card, Chassis or Interface PCB.
or cable.
SD Card, Downstream pressure sensor, Interface PCB or cable.
Interface PCBs. Interface PCB.
Sensor. Check cable connections to Interface PCB. Comms PCB or Interface PCB.
Control PCB or Interface PCB.
SD Card or Control PCB.
Alaris® GP Volumetric Pump 41/84 1000SM00013 Issue 3
Page 42
Software Fault Codes (continued)
Troubleshooting
Troubleshooting
Code
MMI1 MMI Primary Audio Fault Speaker, Control PCB or Interface PCB. MMI2 Stuck Key Fault_Stop MMI3 Stuck Key Fault_Start MMI4 Stuck Key Fault_OnO MMI5 Stuck Key Fault_IncInc MMI6 Stuck Key Fault_Inc MMI7 Stuck Key Fault_Dec MMI8 Stuck Key Fault_DecDec MMI9 Stuck Key Fault_Menu MMI10 Stuck Key Fault_Bolus MMI11 Stuck Key Fault_Soft1 MMI12 Stuck Key Fault_Soft2 MMI13 Stuck Key Fault_Soft3 MMI14 Stuck Key Fault_Mute MMI15 Stuck Key Fault_Pressure POW1 Power Monitor Battery Fault POW2 Charge Fault PRG1 Program Flow Control Fault PRG2 Abort Fault (Prg) PRG3 Abort Fault (data) PRG4 Critical Data Corruption Fault PRG5 Image Corruption PRG6 Assertion Fault REM1 Remote Comms Nurse Call Failure Check Comms connections.
RTC1 Instrument RTC Init Failure Con gure clock. RTC2 RTC Over ow Imminent
SCM1 Pump Crisis Switch pump o and then back on.
USP1 UpstreamPressure Sensor Fault Upstream pressure sensor, Interface PCB or
USP2 Calibration Fault Calibrate pressure.
Module Failure Action/Replace
Key has been registered as stuck for 2 minutes. Check keypad operation in Service Mode. Keypad or Control PCB.
Battery or Control PCB.
SD Card, Control PCB or Interface PCB.
Comms PCB or Control PCB.
Perform cold start. Control PCB.
Contol PCB or Interface PCB.
cable.
SD Card, Upstream pressure sensor, Interface PCB or cable.
Alaris® GP Volumetric Pump 42/84 1000SM00013 Issue 3
Page 43
Troubleshooting
Troubleshooting
General Fault Diagnosis
Failure Action/Replace
Display missing vertical lines
If Control PCB is issue 9 and below, then replace with latest issue Control PCB.No response from keypad or LED Safety alarm is activated Failure of RS232 communications Replace Comms PCB.
Parts to Check/Test
Dropped or damaged
Exposed to fluids
No battery power
General Fault
No AC mains power
Delivery rates out of tolerance
Front Case
 


Rear Case
Labels & Keypads
Mechanism
Control PCB
Interface PCB
Power PCB
Display PCB
Door
Battery
Mains Lead


Fuses
Exception Error Handling
Exception errors include Assertion Errors and are used to trap logical errors in the software execution.
The pump will display the error type, the title of the software module in which the error occurred and the line number. The user should make a note of these for use in diagnosis. This information is stored in the event log.
After an error, the pump will not store information when powered down. When the pump is switched on again, the user should always confirm clear setup.
Alaris® GP Volumetric Pump 43/84 1000SM00013 Issue 3
Page 44
Circuit Descriptions
In this chapter
Functional Module Block Diagram 45
Chapter 5
Module Overview Functional Description 46
Module Overview Functional Description (continued) 47
Page 45
Circuit Descriptions
R
Y
J4
Circuit Descriptions
Functional Module Block Diagram
GP PRESSURE
PCB UPSTREAM
GP PRESSURE
PCB DOWN
STREAM
MOTOR
ENCODER PCB
AIL/SAFETY
CLAMP HOUSING
THREE PHASE
STEPPER MOTO
DOOR DETECT FLEXIBLE PCB
CHASSIS
J15
J16
J6
J7
J3
J13
PRESSURE
MEASUREMENT
MODULE
ENCODER
INTERFACE
MODULE
AIR IN LINE
MODULE
MOTOR DRIVE
MODULE
J2
J8
SAFETY
PROCESSOR
MODULE
FLUID DELIVERY
PROCESSOR
MODULE
INTERFACE PCB
DROP SENSOR
MODULE
J5
SPEAKER
SD MEMORY
CARD
BATTERY PACK
GAS GAUGE
PCB
BATTERY
CELLS
J3
J2
AUDIBLE ALARM
MODULE
BOOT FLASH
SD RAM
MAIN
PROCESSOR
MODULE
LED DRIVER
POWER
REGULATION
MODULE
POWER
MANAGEMENT
MODULE
J13
J11
J1
J14
J9
CONTROL PCB
POWER SUPPL
J2 J1
PCB
IRDA FLE XIBLE
PCB
J9 J3
COMMS PCB
J6
(RS232 & NURSE CALL)
DISPLAY PCB
MEMBRANE
KEYPAD
IEC MAINS
INLET
Alaris® GP Volumetric Pump 45/84 1000SM00013 Issue 3
Page 46
Circuit Descriptions
Circuit Descriptions
Module Overview Functional Description
The Pumps are designed to be serviced generally to major assembly level. The PCBs are designed as non-serviceable items and as such, can only be replaced as complete parts.
The major assemblies are:
Control PCB Interface PCB Display PCB Comms PCB IrDA Flexible PCB GP Pressure PCB x 2 Motor Encoder PCB IEC Mains Inlet Door Detect Flexible PCB Battery Pack Power Supply PCB Membrane Keypad AIL/Safety Clamp Housing Motor
Cardinal Health will make available, on request, circuit diagrams which will assist appropriately qualified technical personnel to repair those parts of the device which are designated by the manufacturer as repairable.
Control PCB
The Control PCB is broken into a number of functional blocks. A description of each block follows:
Main Processor Module At the heart of the system, the main processor provides
all the high level control functionality. It processes data provided by the safety processor and fluid delivery processor and provides the interface to the user via the display, keypad, audible alarm and LED driver. It also provides external communications via the RS232 Nurse Call and IrDA interfaces. The main processors memory consists of a secure digital (SD) memory card (Industrial Grade) SDRAM and Boot Flash. The firmware is stored on the SD card and copied into the SDRAM at power-up and executed. Event logs, calibration information and other systems configuration information are also stored on the SD card.
Safety Processor Module Running from an independent power supply provided
by a rechargeable lithium coin cell battery the safety processor monitors the operation of the main processor and fluid delivery processor. In the event of a fault it is able to sound a secondary alarm, illuminate the alarm beacon and stop the motor. It also provides real time clock and power on / off functionality.
Power Management and Power Regulation Modules The power management module consists of a multi-
chemistry smart battery charger controlled by the gas gauge within the battery pack. Power from the battery and the mains power supply unit is routed through a number of switches which provide a smooth transition between mains and battery operation. The power regulation module provides regulated supply rails for the display back light and the digital and analogue systems.
Audible Alarm Module The primary audible alarm controlled by the main
processor module and independently monitored by the safety processor. Alarm tones are derived from a PWM signal generated by the main processor. The signal is passed through a limiter and active filter before being amplified and output via the speaker. The safety processor measures the amount of current passing through the speaker to determine correct operation.
Interface PCB
The Interface PCB provides all the low level control and monitoring functionality of the system. It is broken into the following functional blocks:
Fluid Delivery Processor Module This module provides the interface between the motor
drive and sensor systems and the main processor.
Drop Sensor Module This module provides the interface between the fluid
delivery processor and an IVAC® 180 Flow Sensor. It incorporates automatic gain control to minimise the effects of fogging and changes in ambient light levels. Connection of the drop sensor is automatically detected.
Pressure Measurement Module The fluid delivery processor connects to the upstream
and downstream pressure transducers via its internal analogue to digital converter. The fluid delivery processor is able to determine if the sensors are working correctly by monitoring the voltage across the force transducer and by switching in a known off set to the amplifier.
Encoder Interface Module Using optical encoders the fluid delivery processor is able
to determine the direction and speed of the motor, the position of the cam, the status of the door seers and the flow stop device.
Air In Line Module This is an ultrasonic system used to detect air bubbles in
the line. A swept frequency signal is used to excite the piezo crystals in the AIL/Safety Clamp Housing. When fluid is present in the tube the signal is coupled across the gap and received by another piezo crystal. The received signal is amplified and passed through a detector to indicate whether air or fluid is present in the line.
Motor Drive Module The fluid delivery processor generates three PWM control
signals which are used to determine the amount of current flowing through each phase of the stepper motor. The motor is driven using micro steps. The Safety Processor is able to prevent operation of the motor if it believes that a system fault has occurred.
Alaris® GP Volumetric Pump 46/84 1000SM00013 Issue 3
Page 47
Circuit Descriptions
Module Overview Functional Description (continued)
Display PCB
This is an ISTN negative mode graphics display with built in temperature compensation.
Comms PCB and IrDA Flexible PCB
Data from the main processor is routed via this board to either the isolated RS232 interface or the IrDA interface. An isolated nurse call interface is also provided via the RS232 connector. The status of the nurse call relay is monitored
and fed back to the safety processor.
GP Pressure PCB
Two pressure boards are used, one above the pumping mechanism to measure upstream pressure and one below to measure down stream pressure. The tubing is compressed against a force transducer. As pressure builds up in the line the tubing expands and hence the force measured increases. Similarly as the pressure falls the tubing contracts and the force decreases. The software converts the force into a relative pressure measurement. The pressure board contains a silicon bridge force sensor, an instrumentation amplifier and diagnostic systems to check gain and the voltage across the force sensor.
Motor Encoder PCB
The Motor Encoder PCB sits above the encoder wheel. The wheel consists of two discs one with multiple teeth the other with a single slot. The wheel with the multiple teeth runs through a dual channel slotted optical switch which produces two digital encoder signals. The fluid delivery processor is able to interpret the phase and frequency of these signals to determine the speed and direction of the cam shaft. The disc with the single slot runs through a single channel slotted optical switch that produces a single digital signal from which the mechanism can be set into the park position.
Door Detect Flexible PCB
The status of the door is monitored using a magnet embedded in the door frame and a digital Hall Effect device mounted on the end of the Door Detect Flexible PCB.
Battery Pack
The battery pack contains a smart gas gauge device that provides charge information to the charger and the status of the battery (capacity, voltage, current and temperature) to the main processor. The pack also contains a thermal fuse and thermal cut out. The battery will be charged when ever the unit is connected to the main supply.
Power Supply
A universal input switched mode power supply used to regulate the mains input voltage.
Membrane Keypad
The membrane keypad consists of fourteen keys and LED’s to indicate battery, mains, start and stop. The on / off key connects to the main processor and the safety processor. The safety processor manages the power up sequence and the main processor power down.
IEC Mains Inlet
A medical grade filtered mains inlet with fuses in the live and neutral lines. The fuses can be accessed by removing the external splash cover and opening the fuse draw.
Motor
A three phase stepper motor coupled to the cam shaft by a toothed drive belt. The motor does 5,689 microsteps per ml.
Air In Line(AIL)/Safety Clamp Housing
The AIL/Safety Clamp Housing contains the ultra sonic piezo transducers used by the air in line system.
These transducers connect to the Interface PCB via the Air In Line Flexible PCB. A reflective optical sensor in the AIL/Safety Clamp Housing allows the fluid delivery processor to determine the status of the seers used to retract the flow stop. A photo transistor and a photo diode are used to determine if the Safety Clamp slide is open or closed. All the drives to the optical sensors are modulated to prevent cross talk and determine correct operation. The optical sensors connect to the Interface PCB via the Safety Clamp Detect Flexible PCB.
Alaris® GP Volumetric Pump 47/84 1000SM00013 Issue 3
Page 48
Corrective Maintenance
In this chapter
Corrective Maintenance 49
Chapter 6
Torque Guide 50
Access To Pump 51
Rear Case and Subassemblies 53
Front Case and Subassemblies 59
Keypads and Labels 66
Page 49
Corrective Maintenance
Corrective Maintenance
Ensure the pump is disconnected from the AC power supply and switched off before attempting to service.
VThe pump contains static-sensitive components and therefore strict ESD precautions should be
observed at all times. Batteries should be disposed of as outlined by the local country regulations. Do not send batteries back to the manufacturer. Only use Cardinal Health recommended spare parts.
This chapter contains procedures required to properly disassemble, repair and replace parts and then to reassemble the pump.
Following all spare part replacement and repair activities, testing must be performed in accordance with the Performance Verification Procedure (PVP), see Chapter 3, ‘Preventative Maintenance’. Additional testing and calibration may be required after certain repairs are completed, see table below for more information.
Repair/Replacement of
Performance Verification Procedure
Battery Calibration
Volumetric Verification / Calibration
Zero Point Calibration
Pressure Calibration
Test/calibration to perform
= Required
Blank = Optional
Front Case
Rear Case
Labels & Keypads
Chassis / Pump
Mechanism
Control PCB
Power PCB
Display PCB
Interface PCB
Battery
Pressure Sensors
Door
AIL/Safety Clamp
Housing
  




Alaris® GP Volumetric Pump 49/84 1000SM00013 Issue 3
Page 50
Corrective Maintenance
Torque Guide
The torque levels established during the manufacturing process are outlined in this chapter, for example 40cNm. Torque levels selected apply throughout product life
Use the information as a guide to the 'do not exceed' torque levels when servicing the pump. When servicing, it is recommended that torque is applied gradually until the component is secure. In any process do not exceed the stated levels.
If a torque driver is available for servicing this will help control the applied torque; otherwise, be aware that excess force may cause the
Where a torque level is not stated then fixing should be hand-tight.
component to fail.
Always use the correct torque level when performing an assembly stage.
Take care with the torque applied when re-assembling parts.
The head patterns of the fasteners are of the following types:
- Torx T8
- Torx T10
- Allen key 2mm
- Small flat blade
- Hex 4.5mm
- Hex 10mm
Always select the correct tool and bit pattern for the fastener.
.
Alaris® GP Volumetric Pump 50/84 1000SM00013 Issue 3
Page 51
Corrective Maintenance
Access To Pump
Replacement Procedure
1. Remove the two case screws with integral flat washer in battery cover, remove cover and battery.
2. Remove the five case screws with integral flat washer .
3. Carefully separate case halves.
4. Remove screw holding earth cable to mechanism and disconnect four other cables.
5. Where necessary, remove the feet and/or seal.
6. Reassemble in reverse order.
(A) Battery pack
(B) Battery Cover
(C) Screw/Washer (x2) 40cNm
(D) Feet
Spare parts
Item Description Part Number
A Asena LVP Battery Pack 1000SP00487 B Cover Battery Asena LVP 1000ME00589 C Alaris GP Fastener Spares Kit 1000SP01252 D Foot Battery Cover Asena LVP 1000ME00590 E Foot Front Asena LVP 1000ME00649 F Seal Case Nickel/graphite 1000ME01611 G Alaris GP Rear Case Kit 1000SP01250
H Alaris GP Front Case Kit 1000SP01248
Alaris® GP Volumetric Pump 51/84 1000SM00013 Issue 3
Page 52
Access To Pump (continued)
(C) Screw/Washer (x5) 40cNm
Corrective Maintenance
(H) Front Case kit
(G) Rear Case kit
(F) Case seal
(C) Screw 70cNm
(E) Feet (x2)
Alaris® GP Volumetric Pump 52/84 1000SM00013 Issue 3
Page 53
Rear Case and Subassemblies
Power Supply Unit (PSU) & Speaker
Replacement Procedure
1. Disconnect the Mains Inlet cable.
2. Remove the three PSU screws.
3. Remove earth wire screw and washer.
4. Remove PSU and insulator.
5. Pull the speaker up and out.
6. Reassemble in reverse order.
(C) Screw/Washer (x4) 40cNm
(A) PSU
Corrective Maintenance
(B) Speaker & (D)* Self adhesive pad
Spare parts
Item Description Part Number
A Alaris GP PSU PCB Kit 1000SP01305 B Alaris GP Speaker KIt 1000SP01306 C Alaris GP Fastener Spares Kit 1000SP01252 D* Pad Self Adhesive Double Sided 12x12mm 0000ME00423 * Item not shown
Alaris® GP Volumetric Pump 53/84 1000SM00013 Issue 3
Page 54
Corrective Maintenance
Rear Case and Subassemblies (continued)
Mains inlet, IrDA PCB, PE stud and magnet
Replacement Procedure
1. Remove nut and washer to remove PE stud.
2. Remove the two screws on Mains inlet retaining plate.
3. Remove mains inlet retaining plate.
4. Remove magnet by lifting one end.
5. Unclip IrDA PCB and remove.
6. Unclip Mains Inlet and remove.
7. Reassemble in reverse order.
(A) AC Mains Inlet
(A) AC Mains Inlet Gasket
(B) IrDA PCB
(C) Screw (x2) 40cNm
(C) Washer
(D) Magnet
(B) IrDA PCB bracket
(C) Nut 150cNm
(E) PE Stud
Spare parts
Item Description Part Number
A Alaris GP Mains Inlet Kit 1000SP01251 B Alaris GP IrDA PCB Flexi Kit 1000SP01308 C Alaris GP Fastener Spares Kit 1000SP01252 D Magnet IR Detect 1000ME01303 E Stud PE Connector M6 Thread X 15 0000ME00141 * Bussmann Fuse Gmd-1.25A 0000ME00770
Alaris® GP Volumetric Pump 54/84 1000SM00013 Issue 3
Page 55
Corrective Maintenance
Rear Case and Subassemblies (continued)
Pole clamp
Replacement Procedure
1. Remove three pole clamp screws.
2. Reassemble in reverse order.
(B) Bung
(A) Pole Clamp
(B) Screw (x3) 70cNm
Spare parts
Item Description Part Number
A Asena SP, Assy, Pole Clamp 1000SP00115 B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 55/84 1000SM00013 Issue 3
Page 56
Corrective Maintenance
Rear Case and Subassemblies (continued)
Rail cam
Replacement Procedure
1. Remove screw from lever release.
2. Remove screw from lever rail cam.
3. Remove spring from the lever rail cam.
4. Reassemble in reverse order.
(B) Screw 70cNm
(B) Screw 70cNm
(A) Rail Cam Kit
(C) Rail Cam only Kit
Spare parts
Item Description Part Number
A Alaris GP Docking Station Kit 1000SP01307 B Alaris GP Fastener Spares Kit 1000SP01252 C Alaris SP Cam Rail Clamp Only Kit 1000SP01323
Alaris® GP Volumetric Pump 56/84 1000SM00013 Issue 3
(A) Rail Cam Kit
Page 57
Corrective Maintenance
Rear Case and Subassemblies (continued)
RS232 Connector & Comms PCB
Replacement Procedure
1. Remove two retaining screws and washers from assembly.
2. Remove RS232 connector cover and two RS232 socket screws.
3. Remove RS232 Connector & Comms PCB assembly.
4. Reassemble in reverse order.
(C) Screw/washer 40cNm
(A) Comms PCB
(C) Screw 40cNm
(B) RS232 Connector cover
(D) Flow Sensor Connector cover
(C) Screw (x2) 40cNm
(C) Screw (x2) 40cNm
(A) RS232 Socket Screw (x2) 40cNm
(E) Fuse cover
Spare parts
Item Description Part Number
A Alaris GP Comms PCB Kit 1000SP01256 B Cover RS232 1000ME01745 C Alaris GP Fastener Spares Kit 1000SP01252 D Asena GW, Assy, Cover Dust Drop Sensor 1000ME00291 E Fuse Cover 1000ME00655
Alaris® GP Volumetric Pump 57/84 1000SM00013 Issue 3
Page 58
Corrective Maintenance
Rear Case and Subassemblies (continued)
Handle
Replacement Procedure
1. Remove two screw from handle block.
2. Remove handle block, handle spring and handle.
3. Reassemble in reverse order.
Refitting notes:
1) Make sure that the handle spring is in front of the handle and not behind it.
(D) Screw (x2) 40cNm
(C) Handle block
(B) Handle spring
(A) Handle
Spare parts
Item Description Part Number
A Asena LVP Overmould Handle 1000ME01845 B Handle Spring Asena LVP 1000ME00630 C Asena LVP Handle Retaining Block 1000ME00632 D Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 58/84 1000SM00013 Issue 3
Page 59
Corrective Maintenance
Front Case and Subassemblies
Door
Replacement Procedure
1. Remove the two screws securing the hinge locks.
2. Open the two hinge locks.
3. Remove the door.
4. Unscrew the two hinge pins.
5. Reassemble in reverse order.
(B) Screw (x2) 70cNm
(A) Membrane
Spare parts
Item Description Part Number
A Asena LVP Assembly Membrane 1000ME00667 B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 59/84 1000SM00013 Issue 3
Page 60
Corrective Maintenance
Front Case and Subassemblies (continued)
Door continued
(A) Hinge pin (x2) 60cNm
(B) Door
(A) Hinge pin seal (x2)
Spare parts
Item Description Part Number
A Alaris GP Hinge Pin Kit 1000SP01246 B Alaris GP Door Kit 1000SP01244
Alaris® GP Volumetric Pump 60/84 1000SM00013 Issue 3
Page 61
Corrective Maintenance
Front Case and Subassemblies (continued)
Chassis assembly
Replacement Procedure
1. Disconnect four cables from the Interface PCB and one cable from the Control PCB.
2. Remove the two screws securing the roller mounting.
3. Remove the roller mounting and gasket.
4. Remove the snap rivet securing the door detector flexible circuit.
5. Remove the two screws securing the chassis.
6. Carefully withdraw the chassis.
7. Reassemble in reverse order.
(A) Screw (x2) 90cNm
(B) Screw (x2) 65cNm
(B) Snap rivet
(A) Gasket
(A) Roller mounting bracket
Spare parts
Item Description Part Number
A Alaris GP Roller Mounting Bracket Kit 1000SP01303 B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 61/84 1000SM00013 Issue 3
Page 62
Corrective Maintenance
Front Case and Subassemblies (continued)
Chassis assembly breakdown
Replacement Procedure
1 Carefully withdraw the chassis.
2. Remove four screws securing the motor.
3. Remove the motor and gasket.
4. Unclip and remove the pressure sensors.
5. Reassemble in reverse order.
(A) Motor
(A) Motor Insulator
(B) Screw (x4) 70cNm
Spare parts
(C) Chassis
(D) Pressure Sensor (x2)
(E) Seal (x4)
Item Description Part Number
A Alaris GP Motor Kit 1000SP01296 B Alaris GP Fastener Spares Kit 1000SP01252 C Alaris GP Pumping Mech (Minus Motor) Kit 1000SP01247 D Alaris GP Pressure Sensor Kit 1000SP01254 E V Seals Hinge Pins V5a-NBR 0000ME00767
Alaris® GP Volumetric Pump 62/84 1000SM00013 Issue 3
Page 63
Corrective Maintenance
Front Case and Subassemblies (continued)
Control PCB and Interface PCB
Replacement Procedure
1. Disconnect display and keypad cables from Control PCB.
2. Remove the three retaining screws and washers.
3. Remove the three pillar supports.
4. When fitting Control PCB ensure all flexi and cables are routed clear of PCB.
5. Reassemble in reverse order.
(A) Interface PCB
(C) Screw (x3) 65cNm
(B) Control PCB
Spare parts
Item Description Part Number
A Alaris GP Interface PCB Kit 1000SP01300 B Alaris GP Control PCB Kit 1000SP01298 C Alaris GP Fastener Spares Kit 1000SP01252 Note: Three Pillar Supports are supplied with both PCB kits.
Pillar supports (x3)
Alaris® GP Volumetric Pump 63/84 1000SM00013 Issue 3
Page 64
Corrective Maintenance
Front Case and Subassemblies (continued)
Display PCB
Replacement Procedure
1. Remove the two fixing screws from display frame.
2. Remove Display Frame and Gasket as required.
3. Reassemble in reverse order.
(A) Display
(B) Display Frame
(C) Screw (x2) 65cNm
Spare parts
Item Description Part Number
A Alaris GP Display Kit 1000SP01255 B Alaris GP Display Accessories Kit 1000SP01297 C Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 64/84 1000SM00013 Issue 3
Page 65
Corrective Maintenance
Front Case and Subassemblies (continued)
AIL/Safety Clamp Housing, Top Retainer & Door Sensor Flexible Circuit
Replacement Procedure
1. Remove the four fixing screws from AIL/Safety Clamp Housing.
2. Remove AIL/Safety Clamp Housing and remove seal as required.
3. Remove one screw securing coloured insert as required.
4. Remove one screw from Top Retainer.
5. Remove Top Retainer and remove O ring as required.
6. Remove Door Sensor Flexible Circuit as required.
7. Reassemble in reverse order.
Refitting notes:
1) Door Sensor Flexible Circuit is retained using hot melt glue.
2) The Screw (F) for the insert torque is 25cNm instead of the original 10cNm due to the new insert plastic which is more tolerant and will not crack. Any cracked Inserts should be replaced by the kit part number 1000SP01417.
(F) Screw 25cNm
(F) Insert
(C) Screw (x4) 65cNm
(A) AIL/Safety Clamp Housing
(A) Seal
(B) Door Sensor
(C) Screw 65cNm
(E) Seal
(D) Top Retainer
Spare parts
Item Description Part Number
A Alaris GP AIL/Safety Clamp Housing Kit 1000SP01249 B Door Detect Flexible Circuit 1000EL00643 C Alaris GP Fastener Spares Kit 1000SP01252 D Asena LVP GP Top Retainer 1000ME00701 E Seal O Ring 6ID 1CSDIA Silicon 0000ME00691 F Spares Kit (Orange Clip & Screw) 1000SP01417
Alaris® GP Volumetric Pump 65/84 1000SM00013 Issue 3
Page 66
Corrective Maintenance
Keypads and Labels
Replacement Procedure
1. Discard keypad when removed as it cannot be reused.
2. Fit replacement keypad after removing backing paper from underside. Handle replacement keypad carefully to avoid damage.
3. Remove label(s) from case as required.
4. Clean case where replacement label(s) are to be fitted.
5. Fit replacement label(s) taken from label sheet as required.
6. Ensure keypad membrane flexi tail is routed correctly.
(A) Keypad
Spare parts
Item Description Part Number
A Alaris GP Keypad BOM 1000LB00623 B Alaris GP Door Label Set 1000LB01040 B Alaris GP Guardrails Door Label 1000LB01475 C Alaris GP Label Set BOM 1000LB00614
Alaris® GP Volumetric Pump 66/84 1000SM00013 Issue 3
Page 67
Corrective Maintenance
Keypads and Labels (continued)
C.1
B
C.3
C.1
Cardinal Health
1180 Rolle
Switzerland
#ü³Ôüóú #ü³Ôü³ðú #ü³Ôü³óú ð³Ă³üóú #ü³Ôü³ú ð³Ă³üüÉú #ü³Ôü³Ôú ․⸵ъ ъ ъ Ъ½½Ъяяхъ Ъ½½Ъяэхъ Ъ½½Ъя»хъ цяэ´´½хъ цэцЪ½ЦхъъЦĈЦĈчхъЦЪ»´½хъъ яЦЦяĈэц´бчхъ½ЦЦЪĈ´´эбэхъъ ЦЦцЪэцхъ ъ яĈ´э»чĈхъ яĈ´э»чЪхъ яĈ´э»чэхъ ъ ЪэĈĈЪхъ ЪэĈĈэхъ ЪэĈĈцхъ ъ ЪЪЪцЦЦчхъ ЪЪЪцЦЦцхъ ЪЪЪцЦЦЦхъ ъ ч½Ĉцээ»б
2 x T 1.25A 20 x 5 mm 250V
~ 50 - 60Hz 60VA (Max.)
100 - 230V
1000LB00614 Iss 41000LB00614 Iss 4
Alaris® GP Volumetric Pump Label Set 1000LB00614 Iss 4 1000AW01013 Iss 4
C.3
C.2
C.2
B
B
Serial Information box
Alaris® GP Volumetric Pump 67/84 1000SM00013 Issue 3
Page 68
Appendix A
Electromagnetic Compatibility
Page 69
Electromagnetic Compatibility
Electromagnetic Compatibility
Warning:
The use of any accessory, transducer, or cable with the Pump other than those specified may result in increased emissions or decreased immunity of the pump.
The Pump should not be used adjacent to or stacked with other equipment, however if adjacent or stacked use is necessary, the Pump should be observed to verify normal operation in the configuration in which it will be used.
Caution:
The Pump is a CISPR 11 Group 1 Class B Medical Equipment System and intended for use by healthcare professionals only.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed, put into service and used
according to the EMC information provided in the accompanying documents.
Portable and Mobile RF communications can affect Medical Electrical Equipment.
Operating the Pump near equipment which radiates high energy radio frequencies (electro surgical or cauterizing equipment,
portable radios, cellular telephones, etc.) may cause false alarm conditions. If this happens, reposition the Pump away from the source of interference or turn off the Pump and manually regulate the flow.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pump should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
CISPR 11
RF Emissions
CISPR 11
RF Emissions
EN 61000-3-2
Harmonic Emissions
EN 61000-3-3
Voltage Fluctuations, Flicker Emissions
Group 1
Class B
Class A
Complies
The pump uses RF energy only for its internal function in the normal product offering. Therefore, its RF emissions are very low and are not likely to cause any interface in nearby electronic equipment.
The pump is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Alaris® GP Volumetric Pump 69/84 1000SM00013 Issue 3
Page 70
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of Pump should assure that it is used in such an environment.
Immunity Test
EN 61000-4-2 Electro-Static
Discharge (ESD)
EN 61000-4-4
Electrical Fast Transient, Burst (EFT) (Note 3)
EN 61000-4-5
Power Line Surge (Note 3)
EN 61000-4-8 Power Frequency Magnetic Field (50/60 Hz)
EN 61000-4-11
Voltage Dips, Short Interruptions, and Voltage Variations (Note 3)
Compliance Level
EN 60601-2-24
EN 60601-1-2
±8 kV contact (Note
2) ±15 kV air (Note 2)
±2 kV for power supply lines N/A (Note 4)
±1 kV Line(s) to Line(s) ±2 kV Line(s) to Earth
400 A/m 50 Hz (Note 2)Power frequency magnetic fields should be at levels characteristic of a
<5 % UT (>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment.
typical location in a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment. If the user of the pump requires continued operation during power mains interruptions, it is recommended that the pump be powered from an uninterruptible power supply or a battery.
The pump does employ an internal short duration battery.
Electromagnetic Environment – Guidance
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Note 1—UT is the AC mains voltage prior to application of the test level. Note 2—Compliance levels raised by EN 60601-2-24. Note 3—Performed at the Minimum and Maximum Rated Input Voltage. Note 4—Cardinal Health recommends using signal cables of less than 3 meters in length and this requirement is applicable only if signal cables are 3 meters or more in length. (EN 60601-1-2:2002, Clause 36.202.4)
Alaris® GP Volumetric Pump 70/84 1000SM00013 Issue 3
Page 71
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
LIFE SUPPORT Equipment
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pump should ensure that it is used in such an environment.
Immunity Test Compliance Level
EN 60601-2-24
EN 60601-1-2
EN 61000-4-6
Conducted RF
EN 61000-4-3
Radiated RF
10 V rms
(Note 3)
10 V/m
(Note 3)
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the pump, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
3.5 d = [-----] √P
V
1
12 d = [-----] √P 80 MHz to 800 MHz
V
2
12 d = [-----] √P 80 MHz to 2.5 GHz E
1
23 d = [-----] √P 800 MHz to 2.5 GHz E
1
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
a
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, b should be less than the compliance level in each frequency range.
c
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1—At 80 MHz and 800 MHz, the higher frequency range applies. Note 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Note 3—Compliance levels raised by EN 60601-2-24.
a The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease
the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the pump is used exceeds the applicable RF compliance level above, the pump should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the pump.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Alaris® GP Volumetric Pump 71/84 1000SM00013 Issue 3
Page 72
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Recommended Separation Distances for LIFE SUPPORT Equipment between
portable and mobile RF communications equipment and the Pump
The Pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled.
The user of the Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Pump as recommended below, according to the maximum output power
of the communications equipment.
Separation Distance According to Frequency of Transmitter
Rated Maximum Output
Power of Transmitter
W
150 kHz to 80 MHz
Outside ISM bands 3.5
d = [------] √P
V
1
150 kHz to 80 MHz
In ISM bands
12
d = [------] √P
V2
m
80 MHz to 800 MHz
12
d = [ ------] √P
E1
800 MHz to 2.5 GHz
23
d = [------] √P
E1
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1—At 80 MHz and 800 MHz, the separation distance for the higher frequency range apply. Note 2—The ISM (Industrial, Scientific, and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. Note 3—An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. Note 4—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
0.03 0.12 0.12 0.23
0.11 0.38 0.38 0.73
0.35 1.20 1.20 2.30
1.11 3.80 3.80 7.28
3.50 12.00 12.00 23.00
Alaris® GP Volumetric Pump 72/84 1000SM00013 Issue 3
Page 73
Disposal
Appendix B
Page 74
Disposal
Ensure the Pump is disconnected from the AC power supply and switched off before attempting to service.
VThe Pump contains static-sensitive components and therefore strict ESD precautions should be
observed at all times. Only use Cardinal Health recommended spare parts. Following all spare part replacement and repair activities, testing must be performed in accordance with the Performance Verification Procedure (PVP), see Chapter 3 ‘Preventative Maintenance’.
Disposal
Information on Disposal for Users of Waste Electrical & Electronic Equipment
This U symbol on the product and/or accompanying documents means that used electrical and electronic products should
not be mixed with municipal waste. If you wish to discard electrical and electronic equipment, please contact your Cardinal Health affiliate office or distributor for further information. Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human health and the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
This symbol is only valid in the European Union. The product should be disposed of taking environmental factors into consideration. To ensure no risk or hazard, remove the internal rechargeable battery and the Nickel Metal Hydride battery from the control board and dispose of as outlined by the local country regulations. All other components can be safely disposed of as per local regulations.
Battery Removal
Removal Procedure
Remove the two case screws with integral flat washer in battery cover, remove cover and battery.
Battery pack
Battery Cover
Alaris® GP Volumetric Pump 74/84 1000SM00013 Issue 3
Page 75
Spare Parts Listing
In this chapter
Appendix C
Spare Parts Kits 76
Individual Components 77
Keypad & Labels 77
Software 77
Test Equipment 77
Kit Bill Of Materials (BOM) 78
Page 76
Spare Parts Listing
Spare Parts Kits
Part Number Description
1000SP01244 Alaris GP Door Kit 1000SP01246 Alaris GP Hinge Pin Kit 1000SP01247 Alaris GP Pumping Mech (Minus Motor) Kit 1000SP01248 Alaris GP Front Case Kit 1000SP01249 Alaris GP AIL/Safety Clamp Housing Kit 1000SP01250 Alaris GP Rear Case Kit 1000SP01251 Alaris GP Mains Inlet Kit 1000SP01252 Alaris GP Fastener Spares Kit 1000SP01253 Alaris GP Battery Compartment Kit 1000SP01254 Alaris GP Pressure Sensor Kit 1000SP01255 Alaris GP Display Kit 1000SP01256 Alaris GP Comms PCB Kit 1000SP01296 Alaris GP Motor Kit 1000SP01297 Alaris GP Display Accessories Kit 1000SP01298 Alaris GP Control PCB Kit 1000SP01299 Alaris GP SD Card Kit 1000SP01300 Alaris GP Interface PCB Kit 1000SP01301 Alaris GP Encoder PCB Kit 1000SP01302 Alaris GP Battery Pack Kit 1000SP01303 Alaris GP Roller Mounting Bracket Kit 1000SP01304 Alaris GP Feet Kit 1000SP01305 Alaris GP PSU PCB Kit 1000SP01306 Alaris GP Speaker Kit 1000SP01307 Alaris GP Docking Station Kit 1000SP01308 Alaris GP IrDA PCB Flexi Kit 1000SP00115 Asena SP, Assy, Pole Clamp 1000SP00487 Asena LVP Battery Pack 1000SP01417 Spares Kit (Orange Clip & Screw) 1000SP01323 Alaris SP Cam Rail Clamp Only Kit
Alaris® GP Volumetric Pump 76/84 1000SM00013 Issue 3
Page 77
Spare Parts Listing
Individual Components
Part Number Description
0000ME00141 Stud PE Connector M6 Thread X 15 0000ME00768 Cables Ties Hayco 3623793 0000ME00770 Bussmann Fuse Gmd-1.25a 1000EL00605 Drop Sensor Cable 1000EL00606 RS232 Nurse Call Cable 1000EL00607 PSU Cable 1000ME00291 Asena GW, Assy, Cover Dust Drop Sensor 1000ME00630 Handle Spring Asena LVP 1000ME00632 Asena LVP Handle Retaining Block 1000ME01845 Asena LVP Overmould Handle 1000ME00636 Asena LVP Mains Inlet Bracket 1000ME00655 Fuse Cover 1000ME01303 Magnet IR Detect 1000ME01745 Cover RS-232 1000ME00589 Cover Battery Asena LVP 1000ME00590 Foot Battery Cover Asena LVP 1000ME00649 Foot Front Asena LVP 1000ME01611 Seal Case Nickel/graphite 1000ME00667 Asena LVP Assembly Membrane 0000ME00767 V Seals Hinge Pins V5a-nbr 1000EL00643 Door Detect Flexible Circuit 1000ME00701 Asena LVP GP Top Retainer 0000ME00691 Seal O Ring 6id 1csdia silicon
0000ME00423 Pad Self Adhesive Double Sided 12x12mm
Keypad & Labels
Part Number Description
1000LB00623 Alaris GP Keypad BOM 1000LB01040 Alaris GP Door Label Set 1000LB00614 Alaris GP Label Set BOM 1000LB01475 Alaris GP Guardrails Door Label
Software
Part Number Description
1000SP01412 Alaris LVP GP F/ware Upgrade V1.7.18 Kit 1000CD00028 Alaris SMU 1000SP01310 Alaris GP Editor Software Kit
Test Equipment
Part Number Description
60793 GP Series Infusion Set, 2 Y 1000EL00979 Converter Cable -USB To Serial 1000EL00980 Converter Cable -USB To 4x Serial 1000SP01183 Cable RS232 Filtered
Alaris® GP Volumetric Pump 77/84 1000SM00013 Issue 3
Page 78
Spare Parts Listing
Kit Bill Of Materials (BOM)
Kit Description Description Qty
1000SP01244 Alaris GP Door Kit Alaris GP Door 1
Assy Screw & Washer M2, 5 X 6 With Patch 2 Alaris GP Door Label 1
1000SP01246 Alaris GP Hinge Pin Kit Assembly Insulated Door Hinge 2
V Seals Hinge Pins V5a-NBR 2
1000SP01247 Alaris GP Pumping Mech (Minus Motor)
Kit
1000SP01248 Alaris GP Front Case Kit Case Front Moulded Asena LVP 1
Alaris GP Pumping Mech 1
Rivet Snap 03 6.5/7.5 Panel Thickness 1 V Seals Hinge Pins V5a-NBR 6 Screw M3x8 Button Head With Patch 4
Alaris GP Keypad Bom 1 Asena LVP Assembly Membrane 1 Seal O Ring 6ID 1CSDIA Silicon 1 Asena LVP GP Top Retainer 1 Screw M3 X 8 Pan Washer Head With Patch 1 Foam Gasket Display Asena LVP 1 Door Detect Flexible Circuit 1 Rivet Snap 03 6.5/7.5 Panel Thickness 1 Foot Front Asena LVP 2 Seal Case Nickel/graphite 1 V Seals Hinge Pins V5a-NBR 2 Adhesive Hot Melt Stick 12mm White 1
1000SP01250 Alaris GP Rear Case Kit Case Rear Moulded Asena LVP 1
Seal Case Nickel/Graphite 1 Alaris GP Label Set BOM 1 Spacer Speaker Label 1
1000SP01251 Alaris GP Mains Inlet Kit Mains Inlet Scha ner Fn9260b Inc Cable 1
Asena LVP Gasket Mains Inlet 1
Alaris® GP Volumetric Pump 78/84 1000SM00013 Issue 3
Page 79
Spare Parts Listing
Kit Bill Of Materials (BOM)
Kit Description Description Qty
1000SP01252 Alaris GP Fastener Spares Kit Plug Pull M12 10
Screw M3 X 8 Pan Washer Head With Patch 10 Screw M3x8 Button Head With Patch 30 Screw M3x8 Pan Hd T10 Skt & Flat Washer 50 Screw M3x10 Button Hd T10 20 Screw M3x8 Button Hd T10 Shkpf Washer 30 Nut M6 Zp+P 10 Washer M6 Internal Shakeproof St Z & C 10 Screw Pt Kc30x10 Csk (T8)-Rogard 10 Screw Pt K25x12 Pan Hd T8 Hexalobular Sk 10 Screw M3x12 Pan Hd Torx (T10) 20 Washer M3 Shakeproof External 10 Screw Lock RS232 4-40 UNC -4-40 UNC 10 Screw M3x6 Flat Head Hexalobular Skt-8 10 Screw, Pan Hd Slotted 4-40 UNC Nylon 10 Screw M2, 5 X 8 Csk Head, With Patch 10 Assy Screw & Washer M2, 5 X 6 With Patch 10 Assembly Screw & Washer, M3x8 With Patch 10 Screw M3 X 8 Extra Low Head With Patch 10 Plastic Snap Rivet 10 Rivet Snap 03 6.5/7.5 Panel Thickness 10 Alaris GP LVP Fixings Kit Instructions 1
1000SP01253 Alaris GP Battery Compartment Kit Cover Battery Asena LVP 1
Foot Battery Cover Asena LVP 2
1000SP01256 Alaris GP Comms PCB Kit Asena LVP RS232 Nurse Call 1
Screw Lock RS232 4-40 UNC -4-40 UNC 2 Gasket,PVC Foam,597/8/9 Fso 1 Insulator RS232 PCB Asena LVP 1 Plastic Snap Rivet 1
1000SP01297 Alaris GP Display Accessories Kit LCD Light Mask 1
Foam Gasket Display Asena LVP 1 Asena LVP LCD Retainer 1
1000SP01298 Alaris GP Control PCB Kit Asena LVP Control Card PCB 1
Rivet Memory CRD 1 Spacer PCB Richco PST- 6-01 2 Asena LVP PCB Clip 3
1000SP01299 Alaris GP SD Card Kit SD Card 1
Rivet Memory CRD 1
Alaris® GP Volumetric Pump 79/84 1000SM00013 Issue 3
Page 80
Spare Parts Listing
Kit Bill Of Materials (BOM)
Kit Description Description Qty
1000SP01300 Alaris GP Interface PCB Kit Asena LVP Interface PCB 1
Asena LVP PCB Clip 3
1000SP01301 Alaris GP Encoder PCB Kit Asena GP Motor Encoder PCB 1
Flexi Encoder To Interface PCB 1 Nylon Spacer 2
1000SP01303 Alaris GP Roller Mounting Bracket Kit Assembly Roller Mounting Bracket 1
Gasket Assembly Roller LVP 1 Screw M3x4 Cap Head Shoulder With Patch 2
1000SP01304 Alaris GP Feet Kit Foot Front Asena LVP 10
Foot Battery Cover Asena LVP 10
1000SP01305 Alaris GP PSU PCB Kit Assembly PSU Asena LVP 1
Insulator PSU Asena LVP 1
1000SP01307 Alaris GP Docking Station Kit Cam Rail Clamp 1
Seal Ring V 10mm Dia 1 Linkage MDI 1 Lever Intermediate MDI Asena LVP 1 Spring Extension OD 4.76 WD 0.66 L 19.05 1 Lever Rail Cam Asena LVP 1 Lever Release Medical Device Interface 1 Seal Ring V 6mm Dia 1
1000SP01308 Alaris GP IrDA PCB Flexi Kit IrDA Flexible Circuit 1
Bracket IrDA GP/VP 1
1000SP01254 Alaris GP Pressure Sensor Kit Housing Upper Pressure Sensor Asena LVP 1
Actuator Pressure Sensor Asena LVP 1 Asena GP Pressure Sensor PCB 1 Housing Lower Pressure Sensor Asena LVP 1 Flexi Pressure Board to Interface Board 1
1000SP01417 Spares Kit (Orange Clip & Screw) Insert Coloured Housing Flowstop AIL 1
Screw M2, 5 X 8 Csk Head, With Patch 1
Alaris® GP Volumetric Pump 80/84 1000SM00013 Issue 3
Page 81
Service Contacts
Appendix D
Page 82
Service Contacts
Service Contacts
For service, contact your local Affiliate Office or Distributor.
AE
Cardinal Health, PO Box 5527, Dubai, United Arab Emirates. Tel: (971) 4 28 22 842 Fax: (971) 4 28 22 914 www.cardinalhealth.com/ international/distributors/ alaris
AU
Cardinal Health, 3/167 Prospect Highway, PO Box 355 Seven Hills, NSW 2147, Australia. Tel: (61) 2 9838 0255 Fax: (61) 2 9674 4444 www.cardinalhealth.com/au
techservice-au@
cardinal.com
BE
Cardinal Health, Leuvensesteenweg 248 D, 1800 Vilvoorde, Belgium. Tel: (32) 2 267 38 99 Fax: (32) 2 267 99 21 www.cardinalhealth.com/be
gmb-ctsi-tech.belux@
cardinal.com
CA
Cardinal Health, 235 Shields Court, Markham, Ontario L3R 8V2, Canada. Tel: (1) 905-752-3333 Fax: (1) 905-752-3343 www.cardinalhealth.com/ca
CN
Cardinal Health, Shanghai Representative Office, Suite 9B, Century Ba-Shi Building, 398 Huai Hai Rd(M.), Shanghai 200020, China. Tel: (56) 8621-63844603 Tel: (56) 8621-63844493 Fax: (56) 8621-6384-4025
DE
Cardinal Health, Pascalstr. 2, 52499 Baesweiler, Deutschland. Tel: (49) 2401 604 0 Fax: (49) 2401 604 121 www.cardinalhealth.com/de
DK
Cardinal Health, Postboks 29, 2820 Gentofte, Danmark. Tlf. (45)70 20 30 74 Fax. (45)70 20 30 98
ES
Cardinal Health, Edificio Veganova, Avenida de La Vega, nº1, Bloque 1 - Planta 1, 28108 Alcobendas, Madrid, España. Tel: (34) 902 555 660 Fax: (34) 902 555 661 www.cardinalhealth.com/es
servicio.tecnico@cardinal.
com
FR
Cardinal Health, Immeuble Antares ­Technoparc, 2, rue Charles-Edouard Jeanneret. 78300 POISSY, France. Tél: (33) 1 30 06 74 60 Fax: (33) 1 39 11 48 34 www.cardinalhealth.com/fr
FR-Assistance-
Technique@cardinal.com
GB
Cardinal Health, The Crescent, Jays Close, Basingstoke, Hampshire, RG22 4BS, United Kingdom. Tel: (44) 0800 917 8776 Fax: (44) 1256 330860 www.cardinalhealth.com/alaris
UK-Technical-Support@
cardinal.com
HU
Cardinal Health, Döbrentei tér 1, H-1013 Budapest, Magyarország. Tel: (36) 14 88 0232 Tel: (36) 14 88 0233 Fax: (36) 12 01 5987
AlarisCE@cardinalhealth.
com
IT
Cardinal Health, Via Ticino 4, 50019 Sesto Fiorentino, Firenze, Italia. Tél: (39) 055 30 33 93 00 Fax: (39) 055 34 00 24 www.cardinalhealth.com/it
assistenza.tecnica@
cardinal.com
NL
Cardinal Health, De Molen 8-10, 3994 DB Houten, Nederland. Tel: (31) 30 228 97 11 Fax: (31) 30 225 86 58 www.cardinalhealth.com/nl
tech.benelux@cardinal.
com
NO
Cardinal Health Solbråveien 10 A, 1383 ASKER, Norge. Tel: (47) 66 98 76 00 Fax: (47) 66 98 76 01 www.cardinalhealth.com/no
technical.supportNO@
cardinal.com
NZ
Cardinal Health, 14B George Bourke Drive, Mt Wellington 1060, PO Box 14-518, Panmure 1741, Auckland, New Zealand Tel: 09 270 2420 Freephone: 0508 422734 Fax: 09 270 6285 www.cardinalhealth.com/nz
techservice-nz@cardinal.
com
SE
Cardinal Health, Hammarbacken 4B, 191 46 Sollentuna, Sverige. Tel: (46) 8 544 43 200 Fax: (46) 8 544 43 225 www.cardinalhealth.com/se
technical.supportSE@
cardinal.com
US
Cardinal Health, 10020 Pacific Mesa Blvd., San Diego, CA 92121, USA. Tel: (1) 858 617 2000 Fax: (1) 858 617 2900 www.cardinalhealth.com/alaris
ZA
Cardinal Health, Unit 2 Oude Molen Business Park, Oude Molen Road, Ndabeni, Cape Town 7405, South Africa. Tel: (27) (0) 860 597 572 Tel: (27) 21 510 7562 Fax: (27) 21 5107567 www.cardinalhealth.com/za
SA-Technical-Support@
cardinal.com
Alaris® GP Volumetric Pump 82/84 1000SM00013 Issue 3
Page 83
Document History
Appendix E
Page 84
Document History
Document History
Issue Date CO No. Author Update Description
1 20/01/06 5999 Ian Tyler Initial release
New Tech mode section.
Spare Part Replacement Procedures chapter added.
2 December 06 7111 Ian Tyler
3 September 08 8587 Ian Tyler
Troubleshooting chapter added.
New Software features added.
Spare Parts Listing appendix added. Introduce Calibration information.
Update Preventative Maintenance information.
Add information on the Alaris® GP Guardrails® Volumetric Pump.
Update Corrective Maintenance information.
Alaris® GP Volumetric Pump 84/84 1000SM00013 Issue 3
Loading...