1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.
2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the
documentation and/or other materials provided with the distribution.
3. The name of the author may not be used to endorse or promote products derived from this software without speci c prior written
permission.
THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO,
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE
AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT
LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER
CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE)
ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
Alaris® GP Volumetric Pump 2/84 1000SM00013 Issue 3
Page 3
Contents
Chapter
1. General Information 4
2. Configuration & Calibration 8
3. Preventative Maintenance 19
4. Troubleshooting 39
5. Circuit Descriptions 44
6. Corrective Maintenance 48
Appendix
A. Electromagnetic Compatibility 68
B. Disposal 73
C. Spare Parts Listings 75
D. Service Contacts 81
E. Document History 83
Alaris® GP Volumetric Pump 3/84 1000SM00013 Issue 3
Page 4
General Information
In this chapter
Introduction 5
Features of the Pump 6
Chapter 1
General Precautions 7
Page 5
General Information
Introduction
The Alaris® GP Volumetric Pump and the Alaris® GP Guardrails® Volumetric Pump (hereinafter referred to as 'Pump') is a
small lightweight volumetric infusion pump that provides accurate and reliable infusions over a range of rates.
The pump is designed to meet the infusion requirements as specified in the Directions For Use (DFU) for all hospital
departments including general wards, critical and intensive care, operating rooms and accident and emergency rooms.
This pump is suitable for use by appropriately trained clinicians or nurses. This pump can be used for intravenous infusion
modes. Supporting fluid & drug therapy, blood transfusions and parenteral feeding.
Product Familiarity
Ensure that you are fully familiar with the pump by carefully studying the Directions for Use (DFU) prior to operation and
prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are
introduced from time to time.
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the Alaris® GP Volumetric Pump and the Alaris®
GP Guardrails® Volumetric Pump.
This manual is intended for use by personnel experienced in medical equipment testing and maintenance procedures .
Conventions Used in this Manual
BOLDUsed for Display names, self-test codes, controls and indicators referenced in this manual, for
example, Battery Indicator, access code 212, ON/OFF button.
'Single quotes'Used to indicate cross-references made to another section of this manual. For example, see
Chapter 2, 'Configuration & Calibration'.
underlineUsed to indicate a link to another section within this manual.
ItalicsUsed to refer to other documents or manuals. For example, refer to the relevant Directions for Use
(DFU) for further information. Also used for emphasis, for example, ...if the gap still measures less
than...
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or
information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test
or maintenance that is important to know about. A typical example is drawing attention to a
software upgrade that you should check has been installed.
Alaris® GP Volumetric Pump 5/84 1000SM00013 Issue 3
Page 6
Features of the Pump
Alarm indicator
Display
General Information
Alaris®
GP
Door
Softkeys
Chevrons
Run
Bolus
Hold
Options
AC power indicator
Release lever for rotating cam
Mute
Pressure
Battery indicator
On/Off
Door Lever
Handle
Flow sensor
connector (cover
removed for clarity)
RS232/Nursecall
connector (cover
removed for clarity)
Rotating cam to lock onto
horizontal rectangular bars.
Mains fuses
Mains inlet
IR communications port
Potential Equalisation (PE)
Connector
Alaris® GP Volumetric Pump 6/84 1000SM00013 Issue 3
Folded pole clamp
Medical device
interface (MDI)
Page 7
General Precautions
Attention consult accompanying documents: Prior to using this pump, carefully read the Operating
w
V
B
M
Alaris®
GP
Precautions described in the Directions for Use (DFU).
This pump contains static-sensitive components. Observe strict precautions for the protection of static
sensitive components when attempting to repair and service the pump.
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care
to locate the pump away from any such hazardous sources.
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed.
Refer all servicing to qualified service personnel.
This pump is protected against the effects of high energy radio frequency emissions and is designed
to be fail safe if extremely high levels of interference are encountered. Should false alarm conditions
be encountered, either remove the source of the interference or regulate the infusion by another
appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise
suspected to have been damaged, remove it from service for inspection by qualified service personnel.
General Information
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If
the integrity of the external protective conductor in the installation or its arrangement is in doubt, the
pump should be operated from the battery.
Alaris® GP Volumetric Pump 7/84 1000SM00013 Issue 3
Page 8
Chapter 2
Configuration and Calibration
In this Chapter
Entering Service Mode 9
Service Mode - Factory Defaults 10
Service Mode - Configuration 11
Service Mode - Data Set Transfer 14
Service Mode - Calibration 15
Page 9
Configuration and Calibration
Entering Service Mode
Warning -
At no time should Service Mode be entered while the pump is connected to a patient.
Service Mode should only be accessed by qualified and trained personnel.
Service Mode can be accessed via a three-digit access code that is entered using the following procedure:
1. Hold down b and turn the pump ON a.
2. Enter the access code 212 using the f keys and the NEXT softkey.
3. When the code shows on screen, press OK to confirm.
Select the required option using the f keys and the OK softkey.
For the Alaris® GP Volumetric Pump the options will be as follows:
SERVICE
Factory DefaultsLoad a default data set. Confirm to perform a
cold start.
ConfigurationThis menu comprises a list of options which are
configurable by the user.
Data Set TransferUpload a data set to pump.
CalibrationThis menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVPPerformance Verification Procedure Tests.
For the Alaris® GP Guardrails® Volumetric Pump the options will be as follows:
CQI Events DownloadFor future implementation
Data Set TransferUpload a data set to pump.
ConfigurationThis menu comprises a list of options which are
configurable by the user.
CalibrationThis menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVPPerformance Verification Procedure Tests.
Factory DefaultsLoad a default data set. Confirm to perform a
cold start.
Factory Defaults
Configuration
Data Set Transfer
Calibration
Test Verification/PVP
SELECT WITH
QUIT
SERVICE
CQI Events Download
Data Set Transfer
Configuration
Calibration
Test Verification/PVP
Factory Defaults
OK
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 9/84 1000SM00013 Issue 3
OK
Page 10
Configuration and Calibration
Service Mode - Factory Defaults
Factory Defaults
Select the required option using the f keys and the OK softkey.
Default Data Set Replace the current data set with a default data
set.
Cold Start Confirm Confirm clearing and resetting the data set and
calibration data to the factory defaults.
Clear CQI Log File* Confirm clearing all pump history and resetting
the data set to the factory default.
* Alaris® GP Guardrails® Volumetric Pump only.
FACTORY DEFAULTS
Default Data Set
Cold Start Confirm
Clear CQI Log File
SELECT WITH
Default Data Set
1. Press OK to confirm loading the default data set.
QUIT
OK
DEFAULT DATA SET
******
Restoring factory
default data set will
delete the clinically
approved data set
installed.
Delete approved and
restore factory default?
QUIT
WARNING!
******
OK
Cold Start Confirm
1. Press OK to confirm performing a cold start.
Alaris® GP Volumetric Pump 10/84 1000SM00013 Issue 3
COLD START CONFIRM
******
All data set and cal
data will be reset to
default!
The pump will require a
full calibration before
returning to clinical use.
QUIT
WARNING!
******
OK
Page 11
Configuration and Calibration
Service Mode - Factory Defaults (continued)
Clear CQI Log File
1. Press OK to confirm clearing the CQI Log File.
CLEAR CQI LOG FILE
Service Mode - Configuration
Select the required option using the f keys and the OK softkey.
Date/TimeSets the current date and time used for event
logging.
Software VersionsDisplays the pump software versions.
Serial NumberConfigure the displayed serial number.
Pump ReferencePump specific text to be displayed in user
mode at start up. (20 characters max.)
LanguageConfigure the Language used for display
messages.
Backlight & ContrastAdjust the Backlight and Contrast values
Current Data Set FileDisplays the current data set file details.
******
This will remove all
instrument history and will
restore the factory default
data set by deleting the
clinically approved
installed data set.
Clear current log?
QUIT
******
OK
SERVICE CONFIGURATION
Date/Time
Software Versions
Serial Number
Pump Reference
Language
Backlight & Contrast
Current Data Set File
SELECT WITH
WARNING!
Date/Time
1. Set the correct date and time using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
QUIT
DATE / TIME
AUG–04–2006
09:48
ADJUST WITH
QUITNEXT
OK
OK
Alaris® GP Volumetric Pump 11/84 1000SM00013 Issue 3
Page 12
Configuration and Calibration
Service Mode - Configuration (continued)
Software Versions
Press 1. OK to exit after verifying Software Version fitted, display will vary depending on software version fitted.
SOFTWARE VERSIONS
SRP 001.005.004
MP 001.005.004
FDP 001.005.004
SP 001.005.004
QUIT
Serial Number
1. Set the serial number required (maximum 9 characters) using the
f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
OK
SOFTWARE VERSIONS
SWV 001.009.000
PKG 001.009.000
MPU 001.009.000
MPT 001.009.000
FDP 001.007.016
SP 001.005.000
LANGUAGE 001.004.005
QUIT
OK
SERIAL NUMBER
8026 - 00000
Pump Reference
1. Set the Pump Reference required using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
Write Success: .
ADJUST WITH
QUITNEXT
OK
PUMP REFERENCE
---------------------
ADJUST WITH
QUITNEXT
OK
Alaris® GP Volumetric Pump 12/84 1000SM00013 Issue 3
Page 13
Configuration and Calibration
Service Mode - Configuration (continued)
Language
1. Select the required Language using the f keys.
2. Press OK to confirm.
Languages available will be dependant on the pump
software version.
LANGUAGE
ENGLISH
NORSK - NORWEGIAN
DEUTSCH - GERMAN
FRANCAIS - FRENCH
NEDERLANDS - DUTCH
SVENSKA - SWEDISH
ESPAñOL - SPANISH
ITALIANO - ITALIAN
NONE - VANILLA
SELECT WITH
QUIT
OK
Backlight & Contrast
1. Use the f keys to adjust Backlight and Contrast to required
setting.
2. Press PAR AM to change between Backlight and Contrast.
3. Press OK to confirm.
Select the Dimming parameter to see what the display
would look like when dimmed.
Current Data Set File
1. Press OK to exit after verifying current data set information.
BACKLIGHT & CONTRAST
Backlight = 30%
Contrast = 70%
Dimming = 25%
ADJUST WITH
QUITPARAM
OK
CURRENT DATA SET
Data Set ID: ABCD
Name:
Data_set_EXAMPLE
Version: 2
Last update:
01-Jun-2006 12:34
Hospital name:
Basingstoke General
QUIT
Alaris® GP Volumetric Pump 13/84 1000SM00013 Issue 3
OK
Page 14
Configuration and Calibration
Service Mode - Data Set Transfer
Upload data set to an Alaris® GP Volumetric Pump
Equipment required:
Alaris® GP Editor Software Kit (1000SP01310) - includes the Alaris® GP Transfer Tool•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) - optional•
USB to RS232 converter 4 way hub (1000EL00980) - optional•
PC - for requirements see Upgrading Firmware PC requirements•
Using the Alaris® GP Transfer Tool allows a released data set to be uploaded to an Alaris® GP Volumetric Pump.
Warning At no time should the Alaris® GP Transfer Tool be used to upload to an Alaris®
GP Volumetric Pump while the pump is connected to a patient.
In Service Mode select Data Set Transfer using the f keys and the OK softkey.
Using the Alaris® GP Transfer Tool select data set to be uploaded. 1.
Press the 2. RS232 or IrDA softkey to select the Comms mode being used.
3.
Connect the pump to PC.
Press the 4. START softkey to begin transfer.
Please ensure the data set ID shown on the pump is identical to the one transferred.5.
Press 6. PASS softkey to confirm correct transfer and exit.
To transfer the data set to another pump repeat steps 2 to 6.7.
Upload data set to an Alaris® GP Guardrails® Volumetric Pump
Equipment required:
Guardrails® Editor V3.1 Software Kit (1000SP01389) or Guardrails® Editor V3.1 Transfer Tool Software Kit (1000SP01390)•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) •
USB to RS232 converter 4 way hub (1000EL00980) - optional•
PC - for requirements see Upgrading Firmware PC requirements•
Using the Guardrails® Editor V3.1 Transfer Tool allows an approved data set to be uploaded to an Alaris® GP Guardrails®
Volumetric Pump.
Warning At no time should the Guradrails® Editor V3.1 Transfer Tool be used to upload
to an Alaris® GP Guardrails® Volumetric Pump while the pump is connected to
a patient.
In Service Mode select
Using the 1. Guardrails® Editor V3.1 Transfer Tool Transfer Tool select data set to be uploaded.
2.
Connect the pump to PC.
Press the 3. START softkey to begin transfer.
Please ensure the data set ID shown on the pump is identical to the one transferred.
4.
Press 5. ACCEPT softkey to confirm correct transfer and exit.
To transfer the data set to another pump repeat steps 2 to 5.6.
Data Set Transfer using the f keys and the OK softkey.
Caution: Loading the Data Set Transfer Tool software is considered a non-clinical service activity.
Interconnecting the pump with a PC may cause the safety or electromagnetic environment to
change while the connection exists. The threat of higher leakage currents or EMI disturbances may
be present. Disconnect the IrDA or RS232 cable connection at both ends following software upload
activities.
For more information relating to the Alaris® GP Editor Software or the Guardrails®
Editor V3.1 Software refer to the relevant Directions For Use supplied with the
software.
Alaris® GP Volumetric Pump 14/84 1000SM00013 Issue 3
Page 15
Configuration and Calibration
Service Mode - Calibration
Zero Point Calibration
Select the required option using the f keys and the OK softkey.
Wait for the pressure sensors to park.1.
Ensure that an infusion set is not installed.2.
Press the 3. START softkey.
Pump will countdown for 15 seconds.4.
Press the 5. ACCEPT softkey.
Press the 6. PASS softkey.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
Pressure Calibration
Equipment required:
Calibrated Pressure Gauge, minimum specification of •
Accuracy = 0.10% of full scale•
Full scale = 1500mmHg•
Pressure Calibration Set (1000SP01422) Use to calibrate 10 pumps •
and then change
Set up equipment as per figure 2-1 and allow 30 seconds before
proceeding.
Select the required option using the f keys and the OK softkey.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
Alaris® GP Volumetric Pump 15/84 1000SM00013 Issue 3
Page 16
Configuration and Calibration
Service Mode - Calibration (continued)
Pressure Calibration continued
Pressure source
(50ml/100ml syringe
or similar device)
Pressure
Gauge
Figure 2 - 1 Pressure Calibration Equipment Set Up
Vent to
atmosphere
Turn 3-way tap to close from atmosphere.1.
Press the 2. START softkey.
Apply a pressure of 200mmHg and pump will countdown for 15 seconds.3.
Press the 4. ACCEPT softkey.
Apply a pressure of 800mmHg and pump will countdown for 15 seconds.5.
Press the 6. ACCEPT softkey.
Turn 3-way tap to vent to atmosphere.7.
Turn 3-way tap to close from atmosphere. 8.
Press the 9. NEXT softkey to proceed to Verification Procedure.
Alaris®
GP
Verification Procedure
Software version v1.7.x and belowSoftware version v1.9.0 and above
Apply a pressure of 200mmHg and wait for 5 10.
seconds.
Press the 11. NEXT softkey.
Apply a pressure of 400mmHg and wait for 5 12.
seconds.
Apply a pressure of 500mmHg and wait for 5 10.
seconds.
Press the 11. NEXT softkey.
Turn 3-way tap to vent to atmosphere.12.
Press the 13. PASS softkey.
Press the 13. NEXT softkey.
Apply a pressure of 600mmHg and wait for 5 14.
seconds.
Press the 15. NEXT softkey.
Apply a pressure of 800mmHg and wait for 5 16.
seconds.
Press the 17. NEXT softkey.
Turn 3-way tap to vent to atmosphere.18.
Press the 19. PASS softkey.
Alaris® GP Volumetric Pump 16/84 1000SM00013 Issue 3
Page 17
Configuration and Calibration
Service Mode - Calibration (continued)
Volumetric Calibration
Select the required option using the f keys and the OK softkey.
Load the primed Infusion Set (60793) into the Pump and set-up 1.
as shown in Figure 2-2 below and adjust the fluid level so that the
meniscus is level with the zero mark.
Press 2. START to begin. Test will run and fluid will be delivered into
the burette.
When 3. Volume delivered! is displayed, check accumulated air
in line value is less than 100µl then enter the volume delivered
into the burette using the f keys and the OK softkey. If
accumulated air in line value is greater than 100µl then repeat test.
If measured value is 19.3ml or less then enter
19.2ml and if value is 20.6ml or higher then enter
20.7ml.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
If no calibration is required (Volume delivered within limits of 4.
19.4ml to 20.5ml) then press PASS to confirm and exit.
If the Calibration value is changed automatically then press 5.
VERIFY and repeat steps 2 to 4.
If the pump still fails replace the Platen and Fingers then repeat 6.
the calibration procedure.
450mm
± 20mm
Alaris®
GP
VOLUMETRIC CALIBRATION
Rate 125ml/h
VTBI 20.0ml
Cal Value 920
Acc’d Air 0μl
START to begin...
QUIT
STARTDETAILS
Figure 2 - 2 Volumetric Calibration Equipment Set Up
Alaris® GP Volumetric Pump 17/84 1000SM00013 Issue 3
Page 18
Configuration and Calibration
Service Mode - Calibration (continued)
Battery Calibration
Select the required option using the f keys and the OK softkey.
Connect AC Mains to the Pump and press 1. CAL to begin.
When calibration is complete it will display 2. CALIBRATION
SUCCESS or CALIBRATION FAILURE. Press PASS to confirm
successful calibration or REPEAT to perform calibration again.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 98%
Temperature 22°C
Calibration
CAL to calibrate!
CAL can exceed 10 hrs!
QUIT
DETAILS
CAL
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2335mAh
Chrg Remain 2332mAh
Rel Chrg 100%
Temperature 26°C
Calibration
BATTERY CALIBRATION
Serial No 12345
Current -340mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 98%
Temperature 22°C
Calibration
CALIBRATING . . .
Discharge
QUIT
DETAILS
89%
CAL
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 100%
Temperature 22°C
Calibration X
CALIBRATION SUCCESS
QUIT
Alaris® GP Volumetric Pump 18/84 1000SM00013 Issue 3
REPEAT
PASS
CALIBRATION FAILURE
QUIT
REPEAT
Page 19
Chapter 3
Preventative Maintenance
In this chapter
Preventative Maintenance 20
Visual Inspection 20
Recommended Cleaning and Storage 21
Updates 23
Battery Test and Replacement 25
Service Mode - Test Verification/PVP 25
Performance Verification Procedure 38
Page 20
Preventative Maintenance
Preventative Maintenance
To ensure the pump remains in good operating condition, routine and preventative maintenance inspections are
required. Routine maintenance inspections should be performed by hospital/facility before each use, see Directions For Use for details.
Preventative maintenance inspections should be performed at least every year.
For the preventative maintenance inspection the following should be performed:
Full visual inspection of the pump, internal and external•
Fitting of all updates required•
Battery test and/or replacement•
Clean the pump•
Performance Verification Procedures
•
Following all spare part replacement and repair activities, testing must be performed in
accordance with the Performance Verification Procedure (PVP). Additional testing and
calibration may be required after certain repairs are completed, see table in Chapter 6
‘Corrective Maintenance’ for more information.
Visual Inspection
Open the pump, as per Chapter 6 ‘Corrective Maintenance’ and visually inspect the interior of the pump.
Visually inspect the exterior of the pump checking the following:
Labels should be replaced as required if not flat, legible or fully adhered.
•
Check Keypad for any sign of wear and replace as required. •
Case components must be checked for damage and replaced if necessary. •
Check the pole clamp is not damaged and that it functions correctly.•
Inspect the AC power supply plug and cable for damage.•
The case should be clean and free from IV solution residue, especially near moving parts. •
Check for dried solution deposits on accessible areas of pumping mechanism.•
Alaris® GP Volumetric Pump 20/84 1000SM00013 Issue 3
Page 21
Preventative Maintenance
Recommended Cleaning and Storage
Cleaning the Pump: -
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a
lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution.
Recommended cleaners are:
Brand Concentration
Hibiscrub 20% (v/v)
Virkon 1% (w/v)
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
- Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Before cleaning always switch OFF and disconnect from the AC power supply.
Never allow fluid to enter the casing and avoid excess fluid build up on the
Pump.
Do not use aggressive cleaning agents as these may damage the exterior
surface of the Pump.
Do not steam autoclave, ethylene oxide sterilise or immerse this Pump in any
fluid.
Storing the Pump: -
If the Pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in
a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in this technical service manual and ensure
that the internal battery is fully charged.
Cleaning and storing the Infusion Set: -
The Infusion Set is a disposable single use item and should be discarded after use according to hospital protocol.
Alaris® GP Volumetric Pump 21/84 1000SM00013 Issue 3
Page 22
Preventative Maintenance
Recommended Cleaning and Storage (continued)
Cleaning the door: -
Periodically during use (as per hospital policy), clean the door by wiping over with a lint-free cloth, lightly dampened with
warm water and a standard disinfectant / detergent solution. Dry door before use.
To aid cleaning of a door which has been heavily soiled, contaminated or if the door operation is not free moving, then
the door may be removed (see procedure below) then immersed and soaked in warm water with a standard disinfectant
/ detergent.
The door should be allowed to dry fully prior to use.
Door Removal
1. Remove the screw securing the lower hinge lock.
2. Open the lower hinge lock.
3. Pull the door away from lower hinge pin and lift up to remove the door.
4. Clean the door.
5. Refit door in reverse order. Ensure screw is refitted with a torque of 70cNm.
Cleaning the Flow Sensor: -
Before the transfer of the flow sensor to a new infusion set and periodically during use, clean the flow sensor by wiping
over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution. Ensure the
connector does not get wet. Dry flow sensor before use.
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if the handle operation is not free
moving, then the flow sensor may be immersed and soaked in clean soapy water (see note below). The inside of the spring
mechanism can be cleaned by activating it whilst submerged in the water.
After cleaning, the sensor should be allowed to dry fully prior to use.
Warning -
The plug of the flow sensor must not be immersed in water as damage will
occur.
Alaris® GP Volumetric Pump 22/84 1000SM00013 Issue 3
Page 23
Updates
Upgrading firmware
Requirements
PC •
Minimum hardware system requirements: Software requirements:•
1GHz Pentium processor• 512MB RAM• Microsoft Windows 2000 •
1GB of free space on the system hard •
drive
Available configurable RS232 9 pin serial •
or USB communications ports
Preventative Maintenance
CD ROM drive•
K•
eyboard
(service pack 4), or XP
(service pack 2)
Video resolution of 1024 X 768 pixels •
Mouse•
and 16 bit colour depth
The Software Maintenance Utility (SMU) (1000CD00028)•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) - optional•
USB to RS232 converter 4 way hub (1000EL00980) - optional•
Firmware CD•
Preparation
Load the SMU software onto the PC•
Connect RS232 connector (using USB to RS232 converter if required) to each pump being updated•
Disconnect• the Battery
Remove the Battery Compartment Plug•
Battery
Compartment
Plug
Software Upload
1. Through the Rear Case, there is a set of three dip switches on the bottom of the Control PCB. Switch 1 should be ON and
switches 2 and 3 should be OFF.
2. Using a small flat blade screwdriver or round ended tweezers configure the dip switch settings
to:
Switch 1 = OFFa.
Switch 2 = ONb.
Switch 3 = ON c.
3. Start the SMU facility to upload the software.
4. Select the Firmware zip file.
5. Select the Comm settings - Comm Port and Baud rate of 460800 (USB to RS232 converter) or 115200 (RS232).
6. Connect the Battery to the Battery Cable.
7. Press Start.
Switch 1 turns the safety battery circuit on but is not required for programming.
Switch 2 forces the pump to turn on.
Switch 3 turns the pump into ‘Boot Mode’, this is only used for programming.
Alaris® GP Volumetric Pump 23/84 1000SM00013 Issue 3
Page 24
Preventative Maintenance
Updates (continued)
Upgrading firmware continued
8. Once the green bar has reached the far right hand side and the time has reached 0:00 and the flashing green light is a
steady green light, the RS232 connector can be removed from the pump.
9. Disconnect the Battery and turn OFF all dip switches.
10. Wait 5 seconds then reconnect the Battery.
11. Configure the Dip Switches to:
Switch 1 = ONa.
Switch 2 = ON then OFFb.
Switch 3 = OFFc.
12. Refit the Battery Compartment Plug, this prevents fluid ingress.
13. Power up the pump in Service Mode, enter access code 212, then select Configuration > Date/Time and set the current
date and time.
If the Control, Interface or RS232 PCB is replaced, the pump must be re-programmed.
Alaris® GP Volumetric Pump 24/84 1000SM00013 Issue 3
Page 25
Preventative Maintenance
Battery Test and Replacement
To test the battery perform the battery calibration, as outlined in the procedure in Chapter 2 ‘Configuration and
Calibration’, and verify that all pass criteria are met. If pass criteria are not met then replace the battery.
Battery charge retention will eventually degrade. So where retention is critical the internal battery should be replaced
every three years.
Replace the Main Battery
Remove the two case screws in battery cover, remove 1.
cover and battery.
Fit new battery.2.
Replace battery cover and secure with 2 screws.3.
Battery Cover
It is essential that the battery pack is
calibrated after fitting as failure to do
so will degrade the quoted auxiliary
battery power on this product.
Battery pack
Screw/Washer (x2) 40cNm
Service Mode - Test Verification/PVP
Warning -
At no time should Service Mode be entered while the pump is connected to a patient.
Service Mode should only be accessed by qualified and trained personnel.
Enter access code 212 to view the Service Mode menu (see 'Entering Service Mode' in Chapter 2 for instructions).
Select the Test Verification/PVP option using the f keys and the OK softkey. Press the Quit softkey to return the
pump to the access code screen.
Select the required option using the f keys and the OK softkey.
PVP Work FlowPerformance Verification Procedure (PVP) tests.
Alarms FunctionalityTest the Alarms function correctly.
User InterfaceDisplay, Keypad, LEDs and Audio Tests.
Power SuppliesTest AC Mains and Battery operation.
Sensor TestsPerform tests on the pump sensors.
Comms TestsRS232, IrDA and Nurse Call Tests.
TEST/VERIFY
PVP Work Flow
Alarms Functionality
User Interface
Power Supplies
Sensor Tests
Comms Tests
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 25/84 1000SM00013 Issue 3
OK
Page 26
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
PVP Work Flow
This test is used to con rm that the Pump is functioning correctly.
Press the START softkey to begin.
The PVP Work Flow will iterate through the following tests:
Software Versions•
Date/Time (read only)•
Serial Number (read only)•
Door Frame•
Safety Clamp•
Sear•
Audio Test•
Chequerboard•
LED’s•
Keypad•
Nurse Call•
RS232 Loop Back•
Pumping E ciency Test•
Downstream Occlusion Pressure Test•
PVP WORK FLOW
PVP Work Flow Test
QUIT
START
Alarms Functionality•
Volumetric (Accuracy Test) Calibration•
See individual test details in this chapter or in Chapter 2 'Con guration and
Calibration' for instructions.
Alaris® GP Volumetric Pump 26/84 1000SM00013 Issue 3
Page 27
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Alarms Functionality
1. Press the START softkey to begin.
ALARMS FUNCTIONALITY
2. Pump starts an infusion.
3. Check all alarms sound and display correctly.
4. Alarms to test are:
AC Mains Disconnect
•
Door Open•
Upstream occlusion•
Air In Line•
****
Ensure that a fluid filled
*****
set is in use.
Subsequent alarm
screens should be
cleared using the
CANCEL softkey.
INFORMATION!
QUIT
START
ALARMS FUNCTIONALITY
AC Mains .
Door Open .
Upstream Occ .
Air In Line .
5. When all the alarms have been tested press the PASS softkey if all the
alarms worked correctly or REPEAT softkey if alarms still need to be
checked.
Waiting for alarm . . .
QUIT
STOP
ALARMS FUNCTIONALITY
AC Mains
Door Open
Upstream Occ
Air In Line
All alarms detected
QUIT
REPEAT
PASS
Alaris® GP Volumetric Pump 27/84 1000SM00013 Issue 3
Page 28
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
User Interface
Select the required option using the f keys and the OK softkey.
ChequerboardDisplay Chequerboard Test.
Keypad TestsPerform Keypad Test to check all keys work when
pressed.
LED TestsCheck all LEDs display in correct order.
Audio TestsCheck Primary and Secondary Audio sounders.
USER INTERFACE
Chequerboard
Keypad Tests
LED Tests
Audio Tests
SELECT WITH
Chequerboard Pattern
1. Check pattern is displayed correctly and to pass. If pattern is incorrect
then press
2. Press PASS softkey to confirm pass and exit.
to fail.
QUIT
QUIT
OK
Keypad Tests
1. Press the START softkey to begin.
2. Press all the keys and a
3. Press PASS softkey to confirm pass and exit.
will indicate each key to pass.
KEYPAD TESTS
[ . ] [ . ] [ . ]
[ . ] [ . ] [ . ] [ . ]
[ . ] [ . ] [ . ]
[ . ] [ . ] [ . ]
START to begin...
START
Alaris® GP Volumetric Pump 28/84 1000SM00013 Issue 3
2. Set the rate required using the f keys and press the START
softkey to begin test.
3. Press STOP softkey when test is completed.
4. Press PASS softkey to confirm pass and exit.
RUN-IN MODE
Rate 125ml/h
CHANGE RATE
START to begin...
QUIT
START
Alaris® GP Volumetric Pump 33/84 1000SM00013 Issue 3
Page 34
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Pumping E ciency Test
This test is used to con rm that the Pump is able to generate su cient
pressure. This is done by infusing into a calibrated pressure gauge and
checking that the correct line pressure is achieved. The test set-up is as per
gure 3 - 1.
1. Insert the Infusion Set (60793) and the in-line roller clamp closed to
prevent uid ow.
2. Close the door and open the roller clamp on the set. Ensure that the
3-way tap to the transducer is closed to the atmosphere.
3. From the Pressure System Test menu screen, highlight Pumping E c'y
Test and press OK softkey.
4. Press the RATE softkey to select a rate of 50ml/h.
5. Press the START softkey and start the timer.
6. When 1000mmHg is reached stop the timer and then press the STOP
softkey and open the 3-way tap to atmosphere.
7. Record that the time taken to reach 1000mmHg was 2 minutes or less.
8. Press the PASS softkey to con rm pass and exit.
Note: If a DRV2 fault code is encountered during the pumping e ciency
test and the pressure has exceeded 1000mmHg, the fault should
be ignored, and the pump power should be cycled to reset the
condition.
PUMPING EFFICIENCY TEST
Rate 50ml/h
RATE for 50/125 ml/h
START to begin...
QUIT
STARTRATE
Pressure
Gauge
150 mm
± 5 mm
Figure 3 - 1 Pressure Tests Equipment Set Up
3-way tap
Pump Finger Height (Parking Test)
1. Press the START softkey to begin.
2. Mechanism will run and park.
3. Press DONE softkey to confirm and exit.
450 mm
± 20 mm
Alaris®
GP
PUMP FINGER HEIGHT
Parked
START to begin...
QUIT
START
Alaris® GP Volumetric Pump 34/84 1000SM00013 Issue 3
Page 35
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
COMMS Test
Select the required option using the f keys and the OK softkey.
For further details please contact Cardinal Health.
RS232 Loop Back
1. Link pins 2 & 3 of the RS232 connector on rear of the pump.
2. Press the START softkey to begin.
3. Check RS232 operation is correct and
item. If RS232 Test fails a is displayed to indicate the failure.
4. Press PASS softkey to confirm pass and exit.
for pass are shown after each
SELECT WITH
QUIT
OK
RS232 LOOP BACK
Message sent 0
Message received 0
Transmit count 0
Received count 0
Apply loopback by
connecting pins 2 & 3
QUIT
START
NURSE CALL TEST
1. Press the START softkey to begin.
2. Check Nurse Call operation is correct and
each item. If Nurse Call fails a is displayed to indicate the failure.
3. Press PASS softkey to confirm pass and exit.
for pass are shown after
NURSE CALL TEST
Output
Monitor
Test
START to begin...
QUIT
Alaris® GP Volumetric Pump 35/84 1000SM00013 Issue 3
DETAILS
START
Page 36
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Occlusion Test
This test can be only done as part of the PVP Work Flow.
Use the Infusion Set ten times only and then replace. Record how many times the Infusion Set has been used.
Note: The Occlusion Pressure Test is carried out with uid in the Infusion Set.
This test is used to con rm that the pressure sensor is correctly calibrated and able to detect an occlusion at the correct line
pressure. This is done by pumping into a calibrated pressure gauge and checking that an alarm occurs at the correct line
pressure. The test set-up is as per gure 3 - 1.
Put the uid lled Infusion Set (60793) into the Pump.1.
Enter the 2. PVP Work Flow and proceed to the Occlusion Pressure Test.
Open the 3-way tap to atmosphere then press the 3. LEVEL softkey to adjust the alarm level to L5.
Con gure the Calibrated Pressure Gauge to hold the Peak/MAX Pressure reading, in preparation for the test.4.
Press the 5. START softkey to begin running the Pump at a rate of 125ml/h. Allow the Pump to run for 1 minute, so that the
pressure reading stabilises.
Turn the tap to occlude the Infusion Set into the pressure gauge. 6.
The Pump will continue to infuse and it will be observed that the pressure reading increases. Eventually a high-pressure 7.
alarm will occur and the Pump will stop infusing. Note the reading on the pressure gauge and con rm that it is 500mmHg
±100mmHg.
7. Press PASS softkey if Pump passes test at all levels.
If the pressure is outside of tolerance pressure calibration is required. Calibration should be performed
as per procedure in Chapter 2 ‘Configuration and Calibration’. If the pump continues to fail the occlusion
test then the pressure sensors should be replaced and perform the calibration procedure again.
Occlusion Test (Optional)
This test can be done in normal operating mode to check the occlusion without having to perform the full PVP Work Flow.
Use the Infusion Set ten times only and then replace. Record how many times the Infusion Set has been used.
Note: The Occlusion Pressure Test is carried out with uid in the Infusion Set.
This test is used to con rm that the pressure sensor is correctly calibrated and able to detect an occlusion at the correct line
pressure. This is done by pumping into a calibrated pressure gauge and checking that an alarm occurs at the correct line
pressure. The test set-up is as per gure 3 - 1.
Put the fluid filled Infusion Set (60793) into the Pump.1.
Press the2. a button to turn the pump on.
Open the 3-way tap to atmosphere. 3.
Set the Rate to 125ml/h.4.
Press the 5. VTBI softkey and set VTBI to 10ml.
Press the 6. e button and set the Pressure Alarm Limit to L5.
Con gure the Calibrated Pressure Gauge to hold the Peak/MAX Pressure reading, in preparation for the test.7.
Press the 8. b button to begin running the Pump at a rate of 125ml/h. Allow the Pump to run for 15 seconds, so that the
pressure reading stabilises.
Turn the tap to occlude the Infusion Set into the pressure gauge. 9.
The Pump will continue to infuse and it will be observed that the pressure reading increases. Eventually a high-pressure 10.
alarm will occur and the Pump will stop infusing. Note the reading on the pressure gauge and confirm that it is
500mmHg ±100mmHg.
Open the 3-way tap to atmosphere. 11.
Hold the12. a button down for approximately three seconds to turn the pump off.
If the pressure is outside of tolerance pressure calibration is required. Calibration should be performed
as per procedure in Chapter 2 ‘Configuration and Calibration’. If the pump continues to fail the occlusion
test then the pressure sensors should be replaced and perform the calibration procedure again.
Alaris® GP Volumetric Pump 36/84 1000SM00013 Issue 3
Page 37
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
Volumetric Accuracy
This test can be done as part of the PVP Work Flow or in the calibration
menu.
Load the Infusion Set (60793) into the Pump and set-up as shown in 1.
Figure 3-2 below and adjust the fluid level so that the meniscus is
level with the zero mark.
Press 2. START to begin. Test will run and fluid will be delivered into the
burette.
When 3. Volume delivered! is displayed, check accumulated air in line
value is less than 100μl then enter the volume delivered into the
burette using the f keys and the OK softkey. If accumulated
air in line value is greater than 100μl then repeat test.
If measured value is 19.3ml or less then enter 19.2ml
and if values is 20.6ml or higher then enter 20.7ml.
VOLUMETRIC CALIBRATION
Rate 125ml/h
VTBI 20.0ml
Cal Value 920
Acc’d Air 0μl
START to begin...
QUIT
STARTDETAILS
If no calibration is required then press 4. PASS to confirm and exit.
If the Calibration value is changed automatically then press 5. VERIFY
and repeat steps 2 to 4.
If the pump still fails replace the Platen and Fingers then repeat the 6.
calibration procedure.
450 mm
± 20 mm
Alaris®
GP
Figure 3 - 2 Volumetric Accuracy Equipment Set Up
Alaris® GP Volumetric Pump 37/84 1000SM00013 Issue 3
Page 38
Preventative Maintenance
Performance Verification Procedure
Model / Serial Number:Service Order / Inventory Number:
Hospital Name / Reference:Software Version:
INSPECTIONPhysical inspection and clean
Check all functions in PVP Work Flow
Enter access code 212 and go to PVP Work Flow
Software Versions•
Date/Time•
Serial Number•
Door Frame•
Safety Clamp•
Sear•
Audio Test•
Chequerboard•
LED’s•
SELF TEST
CH3
Keypad•
Nurse Call•
RS232 Loop Back•
Pumping E ciency Test•
• Time taken = 2 minutes or less
CH3
_____ minutes _____ seconds
Downstream Occlusion Pressure Test•
• Occlusion alarm = 500 ± 100 mmHg
Alarms Functionality•
Volumetric (Accuracy Test) Calibration•
• Delivery = 20 ml ± 0.6 ml (3%)
SETUPSet rate to zero (or lowest value possible), clear Volume Infused and VTBI
Class I Type CF
Alternatively attach printed test results
Earth Resistance Test <= 0.2 Ω
ELECTRICAL
SAFETY TESTS
Earth Leakage Current <= 500 μA
Enclosure Leakage Current <= 100 μA
Verification
Performed
By____________________________
Sign
CHX
indicates the chapter number in the Technical Service Manual (TSM) - 1000SM00013.
CH2
E.G.
= Refer to TSM Chapter 2.
____________________________
Print
L5 ___________ mmHg
_____________ ml
_____________ Ω
_____________ μA
_____________ μA
_____________________
Date
Alaris® GP Volumetric Pump 38/84 1000SM00013 Issue 3
Page 39
Troubleshooting
In this Chapter
Chapter 4
Log Downloads 40
Introduction 41
Software Fault Codes 41
General Fault Diagnosis 43
Exception Error Handling 43
Page 40
Troubleshooting
Log Downloads
PC Setup (first time only)
1. Navigate through the Start menu, select Settings, then Network Connections.
2. Select New Connection Wizard.
3. Click Next.
4. Select Set up an advanced connection option and click Next.
5. Select Connect directly to another computer option and click Next.
6. Select Guest option and click Next.
7. Enter AlarisGP as the Computer Name and click Next.
8. Select the appropriate COM port and click Next.
9. Select the Connection Availability required and click Next.
10. Tick check box if a shortcut is required on the desktop and click Finish.
11. Connect AlarisGP Dialog box is displayed and click Properties.
12. On General tab click Configure.
13. Set Maximum speed (bps): to 115200, uncheck Enable Hardware flow control and click OK.
14. On Options tab check Display progress while connecting and uncheck Prompt for name and password,
certificate, etc..
15. On Security tab click Settings.
16. Check Unencrypted password (PAP) only and click OK. Click Yes on the confirmation dialog that is displayed
17. On Networking tab click Settings.
18. Check Enable LCP extensions and Enable software compression then click OK.
19. Check Internet Protocol (TCP/IP) and QoS Packet Scheduler, highlight Internet Protocol (TCP/IP) then click
Properties.
20. Check Use the following IP address and enter an IP address of 192.168.3.2 then click Advanced.
21. Uncheck U
22. Click OK.
23. Click OK.
24. The PC will dial the pump, refer to download procedure.
se default gateway on remote network and click OK.
PC Setup (second time)
1. Navigate through the Start menu, select Settings, then Network ConnectionsAlarisGP.
2. The PC will dial the pump.
3. Refer to download procedure.
Event Log Download
1. Switch the pump on in Service Mode.
2. Once communication is established open a web browser and enter http://192.168.3.1 into the address bar.
3. Download log.
Warning At no time should the Event Log be downloaded while the pump is connected
to a patient.
For pumps with software version v1.9.x and above also download the
Presentation Style Sheet to enable the logs to be viewed (this file only
needs to be downloaded once). Also the downloaded event log needs to
be stored in the same directory as the Presentation Style Sheet. To view the
downloaded event log open file with Microsoft Excel and select style sheet.
Alaris® GP Volumetric Pump 40/84 1000SM00013 Issue 3
Page 41
Troubleshooting
Troubleshooting
Introduction
Use this troubleshooting guide to help identify the cause of errors and faults which may occur as a result of damage to
the pump or failure of an internal component. The following table lists the error messages and describes what action to
take to resolve the problem. A general fault diagnosis checklist is also provided. For information on alarm procedures and
messages, refer to the DFU.
RTC1InstrumentRTC Init FailureCon gure clock.
RTC2RTC Over ow Imminent
SCM1Pump CrisisSwitch pump o and then back on.
USP1UpstreamPressureSensor FaultUpstream pressure sensor, Interface PCB or
USP2Calibration FaultCalibrate pressure.
ModuleFailureAction/Replace
Key has been registered as stuck for 2
minutes.
Check keypad operation in Service Mode.
Keypad or Control PCB.
Battery or Control PCB.
SD Card, Control PCB or Interface PCB.
Comms PCB or Control PCB.
Perform cold start.
Control PCB.
Contol PCB or Interface PCB.
cable.
SD Card, Upstream pressure sensor,
Interface PCB or cable.
Alaris® GP Volumetric Pump 42/84 1000SM00013 Issue 3
Page 43
Troubleshooting
Troubleshooting
General Fault Diagnosis
FailureAction/Replace
Display missing vertical lines
If Control PCB is issue 9 and below, then replace with latest issue Control PCB.No response from keypad or LED
Safety alarm is activated
Failure of RS232 communicationsReplace Comms PCB.
Parts to Check/Test
Dropped or damaged
Exposed to fluids
No battery power
General Fault
No AC mains power
Delivery rates out of
tolerance
Front Case
Rear Case
Labels & Keypads
Mechanism
Control PCB
Interface PCB
Power PCB
Display PCB
Door
Battery
Mains Lead
Fuses
Exception Error Handling
Exception errors include Assertion Errors and are used to trap logical errors in the software execution.
The pump will display the error type, the title of the software module in which the error occurred and the line number.
The user should make a note of these for use in diagnosis. This information is stored in the event log.
After an error, the pump will not store information when powered down. When the pump is switched on again, the user
should always confirm clear setup.
Alaris® GP Volumetric Pump 43/84 1000SM00013 Issue 3
Alaris® GP Volumetric Pump 45/84 1000SM00013 Issue 3
Page 46
Circuit Descriptions
Circuit Descriptions
Module Overview Functional Description
The Pumps are designed to be serviced generally to major
assembly level. The PCBs are designed as non-serviceable
items and as such, can only be replaced as complete
parts.
The major assemblies are:
Control PCB Interface PCB
Display PCB Comms PCB
IrDA Flexible PCB GP Pressure PCB x 2
Motor Encoder PCB IEC Mains Inlet
Door Detect Flexible PCB Battery Pack
Power Supply PCB Membrane Keypad
AIL/Safety Clamp Housing Motor
Cardinal Health will make available, on request, circuit
diagrams which will assist appropriately qualified technical
personnel to repair those parts of the device which are
designated by the manufacturer as repairable.
Control PCB
The Control PCB is broken into a number of functional
blocks. A description of each block follows:
Main Processor Module
At the heart of the system, the main processor provides
all the high level control functionality. It processes data
provided by the safety processor and fluid delivery
processor and provides the interface to the user via the
display, keypad, audible alarm and LED driver. It also
provides external communications via the RS232 Nurse
Call and IrDA interfaces. The main processors memory
consists of a secure digital (SD) memory card (Industrial
Grade) SDRAM and Boot Flash. The firmware is stored on
the SD card and copied into the SDRAM at power-up and
executed. Event logs, calibration information and other
systems configuration information are also stored on the
SD card.
Safety Processor Module
Running from an independent power supply provided
by a rechargeable lithium coin cell battery the safety
processor monitors the operation of the main processor
and fluid delivery processor. In the event of a fault it is able
to sound a secondary alarm, illuminate the alarm beacon
and stop the motor. It also provides real time clock and
power on / off functionality.
Power Management and Power Regulation Modules
The power management module consists of a multi-
chemistry smart battery charger controlled by the gas
gauge within the battery pack. Power from the battery and
the mains power supply unit is routed through a number
of switches which provide a smooth transition between
mains and battery operation. The power regulation
module provides regulated supply rails for the display
back light and the digital and analogue systems.
Audible Alarm Module
The primary audible alarm controlled by the main
processor module and independently monitored by the
safety processor. Alarm tones are derived from a PWM
signal generated by the main processor. The signal is
passed through a limiter and active filter before being
amplified and output via the speaker. The safety processor
measures the amount of current passing through the
speaker to determine correct operation.
Interface PCB
The Interface PCB provides all the low level control and
monitoring functionality of the system. It is broken into
the following functional blocks:
Fluid Delivery Processor Module
This module provides the interface between the motor
drive and sensor systems and the main processor.
Drop Sensor Module
This module provides the interface between the fluid
delivery processor and an IVAC® 180 Flow Sensor. It
incorporates automatic gain control to minimise the
effects of fogging and changes in ambient light levels.
Connection of the drop sensor is automatically detected.
Pressure Measurement Module
The fluid delivery processor connects to the upstream
and downstream pressure transducers via its internal
analogue to digital converter. The fluid delivery processor
is able to determine if the sensors are working correctly by
monitoring the voltage across the force transducer and by
switching in a known off set to the amplifier.
Encoder Interface Module
Using optical encoders the fluid delivery processor is able
to determine the direction and speed of the motor, the
position of the cam, the status of the door seers and the
flow stop device.
Air In Line Module
This is an ultrasonic system used to detect air bubbles in
the line. A swept frequency signal is used to excite the
piezo crystals in the AIL/Safety Clamp Housing. When fluid
is present in the tube the signal is coupled across the gap
and received by another piezo crystal. The received signal
is amplified and passed through a detector to indicate
whether air or fluid is present in the line.
Motor Drive Module
The fluid delivery processor generates three PWM control
signals which are used to determine the amount of current
flowing through each phase of the stepper motor. The
motor is driven using micro steps. The Safety Processor is
able to prevent operation of the motor if it believes that a
system fault has occurred.
Alaris® GP Volumetric Pump 46/84 1000SM00013 Issue 3
This is an ISTN negative mode graphics display with built
in temperature compensation.
Comms PCB and IrDA Flexible PCB
Data from the main processor is routed via this board to
either the isolated RS232 interface or the IrDA interface. An
isolated nurse call interface is also provided via the RS232
connector. The status of the nurse call relay is monitored
and fed back to the safety processor.
GP Pressure PCB
Two pressure boards are used, one above the pumping
mechanism to measure upstream pressure and one
below to measure down stream pressure. The tubing is
compressed against a force transducer. As pressure builds
up in the line the tubing expands and hence the force
measured increases. Similarly as the pressure falls the
tubing contracts and the force decreases. The software
converts the force into a relative pressure measurement.
The pressure board contains a silicon bridge force sensor,
an instrumentation amplifier and diagnostic systems to
check gain and the voltage across the force sensor.
Motor Encoder PCB
The Motor Encoder PCB sits above the encoder wheel. The
wheel consists of two discs one with multiple teeth the
other with a single slot. The wheel with the multiple teeth
runs through a dual channel slotted optical switch which
produces two digital encoder signals. The fluid delivery
processor is able to interpret the phase and frequency
of these signals to determine the speed and direction of
the cam shaft. The disc with the single slot runs through
a single channel slotted optical switch that produces a
single digital signal from which the mechanism can be set
into the park position.
Door Detect Flexible PCB
The status of the door is monitored using a magnet
embedded in the door frame and a digital Hall Effect
device mounted on the end of the Door Detect Flexible
PCB.
Battery Pack
The battery pack contains a smart gas gauge device that
provides charge information to the charger and the status
of the battery (capacity, voltage, current and temperature)
to the main processor. The pack also contains a thermal
fuse and thermal cut out. The battery will be charged
when ever the unit is connected to the main supply.
Power Supply
A universal input switched mode power supply used to
regulate the mains input voltage.
Membrane Keypad
The membrane keypad consists of fourteen keys and LED’s
to indicate battery, mains, start and stop. The on / off key
connects to the main processor and the safety processor.
The safety processor manages the power up sequence and
the main processor power down.
IEC Mains Inlet
A medical grade filtered mains inlet with fuses in the live
and neutral lines. The fuses can be accessed by removing
the external splash cover and opening the fuse draw.
Motor
A three phase stepper motor coupled to the cam shaft by
a toothed drive belt. The motor does 5,689 microsteps per
ml.
Air In Line(AIL)/Safety Clamp Housing
The AIL/Safety Clamp Housing contains the ultra sonic
piezo transducers used by the air in line system.
These transducers connect to the Interface PCB via the
Air In Line Flexible PCB. A reflective optical sensor in
the AIL/Safety Clamp Housing allows the fluid delivery
processor to determine the status of the seers used to
retract the flow stop. A photo transistor and a photo diode
are used to determine if the Safety Clamp slide is open or
closed. All the drives to the optical sensors are modulated
to prevent cross talk and determine correct operation. The
optical sensors connect to the Interface PCB via the Safety
Clamp Detect Flexible PCB.
Alaris® GP Volumetric Pump 47/84 1000SM00013 Issue 3
Page 48
Corrective Maintenance
In this chapter
Corrective Maintenance 49
Chapter 6
Torque Guide 50
Access To Pump 51
Rear Case and Subassemblies 53
Front Case and Subassemblies 59
Keypads and Labels 66
Page 49
Corrective Maintenance
Corrective Maintenance
Ensure the pump is disconnected from the AC power supply and switched off before attempting to
service.
VThe pump contains static-sensitive components and therefore strict ESD precautions should be
observed at all times.
Batteries should be disposed of as outlined by the local country regulations. Do not send batteries
back to the manufacturer.
Only use Cardinal Health recommended spare parts.
This chapter contains procedures required to properly disassemble, repair and replace parts and then to reassemble the
pump.
Following all spare part replacement and repair activities, testing must be performed in accordance with the Performance
Verification Procedure (PVP), see Chapter 3, ‘Preventative Maintenance’. Additional testing and calibration may be
required after certain repairs are completed, see table below for more information.
Repair/Replacement of
Performance Verification
Procedure
Battery Calibration
Volumetric Verification /
Calibration
Zero Point Calibration
Pressure Calibration
Test/calibration to perform
= Required
Blank = Optional
Front Case
Rear Case
Labels & Keypads
Chassis / Pump
Mechanism
Control PCB
Power PCB
Display PCB
Interface PCB
Battery
Pressure Sensors
Door
AIL/Safety Clamp
Housing
Alaris® GP Volumetric Pump 49/84 1000SM00013 Issue 3
Page 50
Corrective Maintenance
Torque Guide
The torque levels established during the manufacturing process are outlined in this chapter, for example 40cNm. Torque levels
selected apply throughout product life
Use the information as a guide to the 'do not exceed' torque levels when servicing the pump. When servicing, it is recommended that
torque is applied gradually until the component is secure. In any process do not exceed the stated levels.
If a torque driver is available for servicing this will help control the applied torque; otherwise, be aware that excess force may cause the
Where a torque level is not stated then fixing should be hand-tight.
component to fail.
Always use the correct torque level when performing an assembly stage.
Take care with the torque applied when re-assembling parts.
The head patterns of the fasteners are of the following types:
- Torx T8
- Torx T10
- Allen key 2mm
- Small flat blade
- Hex 4.5mm
- Hex 10mm
Always select the correct tool and bit pattern for the fastener.
.
Alaris® GP Volumetric Pump 50/84 1000SM00013 Issue 3
Page 51
Corrective Maintenance
Access To Pump
Replacement Procedure
1. Remove the two case screws with integral flat washer in battery
cover, remove cover and battery.
2. Remove the five case screws with integral flat washer .
3. Carefully separate case halves.
4. Remove screw holding earth cable to mechanism and
disconnect four other cables.
5. Where necessary, remove the feet and/or seal.
6. Reassemble in reverse order.
(A) Battery pack
(B) Battery Cover
(C) Screw/Washer (x2) 40cNm
(D) Feet
Spare parts
Item Description Part Number
A Asena LVP Battery Pack 1000SP00487
B Cover Battery Asena LVP 1000ME00589
C Alaris GP Fastener Spares Kit 1000SP01252
D Foot Battery Cover Asena LVP 1000ME00590
E Foot Front Asena LVP 1000ME00649
F Seal Case Nickel/graphite 1000ME01611
G Alaris GP Rear Case Kit 1000SP01250
H Alaris GP Front Case Kit 1000SP01248
Alaris® GP Volumetric Pump 51/84 1000SM00013 Issue 3
Page 52
Access To Pump (continued)
(C) Screw/Washer (x5) 40cNm
Corrective Maintenance
(H) Front Case kit
(G) Rear Case kit
(F) Case seal
(C) Screw 70cNm
(E) Feet (x2)
Alaris® GP Volumetric Pump 52/84 1000SM00013 Issue 3
Page 53
Rear Case and Subassemblies
Power Supply Unit (PSU) & Speaker
Replacement Procedure
1. Disconnect the Mains Inlet cable.
2. Remove the three PSU screws.
3. Remove earth wire screw and washer.
4. Remove PSU and insulator.
5. Pull the speaker up and out.
6. Reassemble in reverse order.
(C) Screw/Washer (x4) 40cNm
(A) PSU
Corrective Maintenance
(B) Speaker & (D)* Self adhesive pad
Spare parts
Item Description Part Number
A Alaris GP PSU PCB Kit 1000SP01305
B Alaris GP Speaker KIt 1000SP01306
C Alaris GP Fastener Spares Kit 1000SP01252
D* Pad Self Adhesive Double Sided 12x12mm 0000ME00423
* Item not shown
Alaris® GP Volumetric Pump 53/84 1000SM00013 Issue 3
Page 54
Corrective Maintenance
Rear Case and Subassemblies (continued)
Mains inlet, IrDA PCB, PE stud and magnet
Replacement Procedure
1. Remove nut and washer to remove PE stud.
2. Remove the two screws on Mains inlet retaining plate.
3. Remove mains inlet retaining plate.
4. Remove magnet by lifting one end.
5. Unclip IrDA PCB and remove.
6. Unclip Mains Inlet and remove.
7. Reassemble in reverse order.
(A) AC Mains Inlet
(A) AC Mains
Inlet Gasket
(B) IrDA PCB
(C) Screw (x2) 40cNm
(C) Washer
(D) Magnet
(B) IrDA PCB
bracket
(C) Nut 150cNm
(E) PE Stud
Spare parts
Item Description Part Number
A Alaris GP Mains Inlet Kit 1000SP01251
B Alaris GP IrDA PCB Flexi Kit 1000SP01308
C Alaris GP Fastener Spares Kit 1000SP01252
D Magnet IR Detect 1000ME01303
E Stud PE Connector M6 Thread X 15 0000ME00141
* Bussmann Fuse Gmd-1.25A 0000ME00770
Alaris® GP Volumetric Pump 54/84 1000SM00013 Issue 3
Page 55
Corrective Maintenance
Rear Case and Subassemblies (continued)
Pole clamp
Replacement Procedure
1. Remove three pole clamp screws.
2. Reassemble in reverse order.
(B) Bung
(A) Pole Clamp
(B) Screw (x3) 70cNm
Spare parts
Item Description Part Number
A Asena SP, Assy, Pole Clamp 1000SP00115
B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 55/84 1000SM00013 Issue 3
Page 56
Corrective Maintenance
Rear Case and Subassemblies (continued)
Rail cam
Replacement Procedure
1. Remove screw from lever release.
2. Remove screw from lever rail cam.
3. Remove spring from the lever rail cam.
4. Reassemble in reverse order.
(B) Screw 70cNm
(B) Screw 70cNm
(A) Rail Cam Kit
(C) Rail Cam only Kit
Spare parts
Item Description Part Number
A Alaris GP Docking Station Kit 1000SP01307
B Alaris GP Fastener Spares Kit 1000SP01252
C Alaris SP Cam Rail Clamp Only Kit 1000SP01323
Alaris® GP Volumetric Pump 56/84 1000SM00013 Issue 3
(A) Rail Cam Kit
Page 57
Corrective Maintenance
Rear Case and Subassemblies (continued)
RS232 Connector & Comms PCB
Replacement Procedure
1. Remove two retaining screws and washers from assembly.
2. Remove RS232 connector cover and two RS232 socket screws.
3. Remove RS232 Connector & Comms PCB assembly.
4. Reassemble in reverse order.
(C) Screw/washer 40cNm
(A) Comms PCB
(C) Screw 40cNm
(B) RS232
Connector cover
(D) Flow Sensor
Connector cover
(C) Screw (x2) 40cNm
(C) Screw (x2) 40cNm
(A) RS232 Socket
Screw (x2) 40cNm
(E) Fuse cover
Spare parts
Item Description Part Number
A Alaris GP Comms PCB Kit 1000SP01256
B Cover RS232 1000ME01745
C Alaris GP Fastener Spares Kit 1000SP01252
D Asena GW, Assy, Cover Dust Drop Sensor 1000ME00291
E Fuse Cover 1000ME00655
Alaris® GP Volumetric Pump 57/84 1000SM00013 Issue 3
Page 58
Corrective Maintenance
Rear Case and Subassemblies (continued)
Handle
Replacement Procedure
1. Remove two screw from handle block.
2. Remove handle block, handle spring and handle.
3. Reassemble in reverse order.
Refitting notes:
1) Make sure that the handle spring is in front of the handle and not behind it.
(D) Screw (x2) 40cNm
(C) Handle block
(B) Handle spring
(A) Handle
Spare parts
Item Description Part Number
A Asena LVP Overmould Handle 1000ME01845
B Handle Spring Asena LVP 1000ME00630
C Asena LVP Handle Retaining Block 1000ME00632
D Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 58/84 1000SM00013 Issue 3
Page 59
Corrective Maintenance
Front Case and Subassemblies
Door
Replacement Procedure
1. Remove the two screws securing the hinge locks.
2. Open the two hinge locks.
3. Remove the door.
4. Unscrew the two hinge pins.
5. Reassemble in reverse order.
(B) Screw (x2) 70cNm
(A) Membrane
Spare parts
Item Description Part Number
A Asena LVP Assembly Membrane 1000ME00667
B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 59/84 1000SM00013 Issue 3
Page 60
Corrective Maintenance
Front Case and Subassemblies (continued)
Door continued
(A) Hinge pin (x2) 60cNm
(B) Door
(A) Hinge pin seal (x2)
Spare parts
Item Description Part Number
A Alaris GP Hinge Pin Kit 1000SP01246
B Alaris GP Door Kit 1000SP01244
Alaris® GP Volumetric Pump 60/84 1000SM00013 Issue 3
Page 61
Corrective Maintenance
Front Case and Subassemblies (continued)
Chassis assembly
Replacement Procedure
1. Disconnect four cables from the Interface PCB and one cable from the Control PCB.
2. Remove the two screws securing the roller mounting.
3. Remove the roller mounting and gasket.
4. Remove the snap rivet securing the door detector flexible circuit.
5. Remove the two screws securing the chassis.
6. Carefully withdraw the chassis.
7. Reassemble in reverse order.
(A) Screw (x2) 90cNm
(B) Screw (x2) 65cNm
(B) Snap rivet
(A) Gasket
(A) Roller mounting
bracket
Spare parts
Item Description Part Number
A Alaris GP Roller Mounting Bracket Kit 1000SP01303
B Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 61/84 1000SM00013 Issue 3
Page 62
Corrective Maintenance
Front Case and Subassemblies (continued)
Chassis assembly breakdown
Replacement Procedure
1 Carefully withdraw the chassis.
2. Remove four screws securing the motor.
3. Remove the motor and gasket.
4. Unclip and remove the pressure sensors.
5. Reassemble in reverse order.
(A) Motor
(A) Motor Insulator
(B) Screw (x4) 70cNm
Spare parts
(C) Chassis
(D) Pressure Sensor (x2)
(E) Seal (x4)
Item Description Part Number
A Alaris GP Motor Kit 1000SP01296
B Alaris GP Fastener Spares Kit 1000SP01252
C Alaris GP Pumping Mech (Minus Motor) Kit 1000SP01247
D Alaris GP Pressure Sensor Kit 1000SP01254
E V Seals Hinge Pins V5a-NBR 0000ME00767
Alaris® GP Volumetric Pump 62/84 1000SM00013 Issue 3
Page 63
Corrective Maintenance
Front Case and Subassemblies (continued)
Control PCB and Interface PCB
Replacement Procedure
1. Disconnect display and keypad cables from Control PCB.
2. Remove the three retaining screws and washers.
3. Remove the three pillar supports.
4. When fitting Control PCB ensure all flexi and cables are routed clear of PCB.
5. Reassemble in reverse order.
(A) Interface PCB
(C) Screw (x3) 65cNm
(B) Control PCB
Spare parts
Item Description Part Number
A Alaris GP Interface PCB Kit 1000SP01300
B Alaris GP Control PCB Kit 1000SP01298
C Alaris GP Fastener Spares Kit 1000SP01252
Note: Three Pillar Supports are supplied with both PCB kits.
Pillar supports (x3)
Alaris® GP Volumetric Pump 63/84 1000SM00013 Issue 3
Page 64
Corrective Maintenance
Front Case and Subassemblies (continued)
Display PCB
Replacement Procedure
1. Remove the two fixing screws from display frame.
2. Remove Display Frame and Gasket as required.
3. Reassemble in reverse order.
(A) Display
(B) Display Frame
(C) Screw (x2) 65cNm
Spare parts
Item Description Part Number
A Alaris GP Display Kit 1000SP01255
B Alaris GP Display Accessories Kit 1000SP01297
C Alaris GP Fastener Spares Kit 1000SP01252
Alaris® GP Volumetric Pump 64/84 1000SM00013 Issue 3
Page 65
Corrective Maintenance
Front Case and Subassemblies (continued)
AIL/Safety Clamp Housing, Top Retainer & Door Sensor Flexible Circuit
Replacement Procedure
1. Remove the four fixing screws from AIL/Safety Clamp Housing.
2. Remove AIL/Safety Clamp Housing and remove seal as required.
3. Remove one screw securing coloured insert as required.
4. Remove one screw from Top Retainer.
5. Remove Top Retainer and remove O ring as required.
6. Remove Door Sensor Flexible Circuit as required.
7. Reassemble in reverse order.
Refitting notes:
1) Door Sensor Flexible Circuit is retained using hot melt glue.
2) The Screw (F) for the insert torque is 25cNm instead of the
original 10cNm due to the new insert plastic which is more
tolerant and will not crack. Any cracked Inserts should be
replaced by the kit part number 1000SP01417.
(F) Screw 25cNm
(F) Insert
(C) Screw (x4) 65cNm
(A) AIL/Safety
Clamp Housing
(A) Seal
(B) Door Sensor
(C) Screw 65cNm
(E) Seal
(D) Top Retainer
Spare parts
Item Description Part Number
A Alaris GP AIL/Safety Clamp Housing Kit 1000SP01249
B Door Detect Flexible Circuit 1000EL00643
C Alaris GP Fastener Spares Kit 1000SP01252
D Asena LVP GP Top Retainer 1000ME00701
E Seal O Ring 6ID 1CSDIA Silicon 0000ME00691
F Spares Kit (Orange Clip & Screw) 1000SP01417
Alaris® GP Volumetric Pump 65/84 1000SM00013 Issue 3
Page 66
Corrective Maintenance
Keypads and Labels
Replacement Procedure
1. Discard keypad when removed as it cannot be reused.
2. Fit replacement keypad after removing backing paper from underside. Handle replacement keypad carefully to avoid
damage.
3. Remove label(s) from case as required.
4. Clean case where replacement label(s) are to be fitted.
5. Fit replacement label(s) taken from label sheet as required.
6. Ensure keypad membrane flexi tail is routed correctly.
(A) Keypad
Spare parts
Item Description Part Number
A Alaris GP Keypad BOM 1000LB00623
B Alaris GP Door Label Set 1000LB01040
B Alaris GP Guardrails Door Label 1000LB01475
C Alaris GP Label Set BOM 1000LB00614
Alaris® GP Volumetric Pump 66/84 1000SM00013 Issue 3
Alaris® GP Volumetric Pump Label Set1000LB00614 Iss 41000AW01013 Iss 4
C.3
C.2
C.2
B
B
Serial Information box
Alaris® GP Volumetric Pump 67/84 1000SM00013 Issue 3
Page 68
Appendix A
Electromagnetic Compatibility
Page 69
Electromagnetic Compatibility
Electromagnetic Compatibility
Warning:
• The use of any accessory, transducer, or cable with the Pump other than those specified may result in increased emissions or
decreased immunity of the pump.
• The Pump should not be used adjacent to or stacked with other equipment, however if adjacent or stacked use is necessary, the
Pump should be observed to verify normal operation in the configuration in which it will be used.
Caution:
• The Pump is a CISPR 11 Group 1 Class B Medical Equipment System and intended for use by healthcare professionals only.
• Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed, put into service and used
according to the EMC information provided in the accompanying documents.
• Portable and Mobile RF communications can affect Medical Electrical Equipment.
• Operating the Pump near equipment which radiates high energy radio frequencies (electro surgical or cauterizing equipment,
portable radios, cellular telephones, etc.) may cause false alarm conditions. If this happens, reposition the Pump away from the
source of interference or turn off the Pump and manually regulate the flow.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pump should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment – Guidance
CISPR 11
RF Emissions
CISPR 11
RF Emissions
EN 61000-3-2
Harmonic Emissions
EN 61000-3-3
Voltage Fluctuations,
Flicker Emissions
Group 1
Class B
Class A
Complies
The pump uses RF energy only for its internal function in the normal product
offering. Therefore, its RF emissions are very low and are not likely to cause any
interface in nearby electronic equipment.
The pump is suitable for use in all establishments, including domestic, and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Alaris® GP Volumetric Pump 69/84 1000SM00013 Issue 3
Page 70
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of Pump should assure that it is used in such an environment.
Immunity Test
EN 61000-4-2 Electro-Static
Discharge (ESD)
EN 61000-4-4
Electrical Fast Transient, Burst
(EFT) (Note 3)
EN 61000-4-5
Power Line Surge
(Note 3)
EN 61000-4-8 Power
Frequency Magnetic Field
(50/60 Hz)
EN 61000-4-11
Voltage Dips, Short
Interruptions, and Voltage
Variations
(Note 3)
Compliance Level
EN 60601-2-24
EN 60601-1-2
±8 kV contact (Note
2)
±15 kV air (Note 2)
±2 kV for power
supply lines
N/A (Note 4)
±1 kV Line(s) to
Line(s)
±2 kV Line(s) to Earth
400 A/m 50 Hz (Note 2)Power frequency magnetic fields should be at levels characteristic of a
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the relative humidity should be
at least 30 %.
Mains power quality should be that of a typical commercial or hospital
environment.
Mains power quality should be that of a typical commercial or hospital
environment.
typical location in a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital
environment.
If the user of the pump requires continued operation during power mains
interruptions, it is recommended that the pump be powered from an
uninterruptible power supply or a battery.
The pump does employ an internal short duration battery.
Electromagnetic Environment – Guidance
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Note 1—UT is the AC mains voltage prior to application of the test level.
Note 2—Compliance levels raised by EN 60601-2-24.
Note 3—Performed at the Minimum and Maximum Rated Input Voltage.
Note 4—Cardinal Health recommends using signal cables of less than 3 meters in length and this requirement is applicable only if signal cables are 3
meters or more in length. (EN 60601-1-2:2002, Clause 36.202.4)
Alaris® GP Volumetric Pump 70/84 1000SM00013 Issue 3
Page 71
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
LIFE SUPPORT Equipment
The Pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pump should ensure that it is used in such an environment.
Immunity TestCompliance Level
EN 60601-2-24
EN 60601-1-2
EN 61000-4-6
Conducted RF
EN 61000-4-3
Radiated RF
10 V rms
(Note 3)
10 V/m
(Note 3)
Electromagnetic Environment – Guidance
Portable and mobile RF communications equipment should be used no closer to any part of
the pump, including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended Separation Distance
3.5d = [-----] √P
V
1
12d = [-----] √P 80 MHz to 800 MHz
V
2
12d = [-----] √P 80 MHz to 2.5 GHzE
1
23d = [-----] √P 800 MHz to 2.5 GHzE
1
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
a
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, b
should be less than the compliance level in each frequency range.
c
Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1—At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
Note 3—Compliance levels raised by EN 60601-2-24.
a The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz are intended to decrease
the likelihood that mobile/portable communications equipment could cause interference if it is inadvertently brought into patient areas. For this reason,
an additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in these frequency ranges.
b Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the pump is used exceeds
the applicable RF compliance level above, the pump should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the pump.
c Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Alaris® GP Volumetric Pump 71/84 1000SM00013 Issue 3
Page 72
Electromagnetic Compatibility
Electromagnetic Compatibility (continued)
Recommended Separation Distances for LIFE SUPPORT Equipment between
portable and mobile RF communications equipment and the Pump
The Pump is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled.
The user of the Pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Pump as recommended below, according to the maximum output power
of the communications equipment.
Separation Distance According to Frequency of Transmitter
Rated Maximum Output
Power of Transmitter
W
150 kHz to 80 MHz
Outside ISM bands 3.5
d = [------] √P
V
1
150 kHz to 80 MHz
In ISM bands
12
d = [------] √P
V2
m
80 MHz to 800 MHz
12
d = [ ------] √P
E1
800 MHz to 2.5 GHz
23
d = [------] √P
E1
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1—At 80 MHz and 800 MHz, the separation distance for the higher frequency range apply.
Note 2—The ISM (Industrial, Scientific, and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957
MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
Note 3—An additional factor of 10/3 is used in calculating the recommended separation distance for transmitters in the ISM frequency bands between
150 kHz and 80 MHz and in the frequency range 80 MHz to 2.5 GHz to decrease the likelihood that mobile/portable communications equipment could
cause interference if it is inadvertently brought into patient areas.
Note 4—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
0.030.120.120.23
0.110.380.380.73
0.351.201.202.30
1.113.803.807.28
3.5012.0012.0023.00
Alaris® GP Volumetric Pump 72/84 1000SM00013 Issue 3
Page 73
Disposal
Appendix B
Page 74
Disposal
Ensure the Pump is disconnected from the AC power supply and switched off before attempting to
service.
VThe Pump contains static-sensitive components and therefore strict ESD precautions should be
observed at all times.
Only use Cardinal Health recommended spare parts.
Following all spare part replacement and repair activities, testing must be performed in accordance
with the Performance Verification Procedure (PVP), see Chapter 3 ‘Preventative Maintenance’.
Disposal
Information on Disposal for Users of Waste Electrical & Electronic Equipment
This U symbol on the product and/or accompanying documents means that used electrical and electronic products should
not be mixed with municipal waste.
If you wish to discard electrical and electronic equipment, please contact your Cardinal Health affiliate office or distributor
for further information.
Disposing of this product correctly will help to save valuable resources and prevent any potential negative effects on human
health and the environment which could otherwise arise from inappropriate waste handling.
Information on Disposal in Countries outside the European Union
This symbol is only valid in the European Union. The product should be disposed of taking environmental factors into
consideration. To ensure no risk or hazard, remove the internal rechargeable battery and the Nickel Metal Hydride battery
from the control board and dispose of as outlined by the local country regulations. All other components can be safely
disposed of as per local regulations.
Battery Removal
Removal Procedure
Remove the two case screws with integral
flat washer in battery cover, remove cover
and battery.
Battery pack
Battery Cover
Alaris® GP Volumetric Pump 74/84 1000SM00013 Issue 3
Page 75
Spare Parts Listing
In this chapter
Appendix C
Spare Parts Kits 76
Individual Components 77
Keypad & Labels 77
Software 77
Test Equipment 77
Kit Bill Of Materials (BOM) 78
Page 76
Spare Parts Listing
Spare Parts Kits
Part NumberDescription
1000SP01244Alaris GP Door Kit
1000SP01246Alaris GP Hinge Pin Kit
1000SP01247Alaris GP Pumping Mech (Minus Motor) Kit
1000SP01248Alaris GP Front Case Kit
1000SP01249Alaris GP AIL/Safety Clamp Housing Kit
1000SP01250Alaris GP Rear Case Kit
1000SP01251Alaris GP Mains Inlet Kit
1000SP01252Alaris GP Fastener Spares Kit
1000SP01253Alaris GP Battery Compartment Kit
1000SP01254Alaris GP Pressure Sensor Kit
1000SP01255Alaris GP Display Kit
1000SP01256Alaris GP Comms PCB Kit
1000SP01296Alaris GP Motor Kit
1000SP01297Alaris GP Display Accessories Kit
1000SP01298Alaris GP Control PCB Kit
1000SP01299Alaris GP SD Card Kit
1000SP01300Alaris GP Interface PCB Kit
1000SP01301Alaris GP Encoder PCB Kit
1000SP01302Alaris GP Battery Pack Kit
1000SP01303Alaris GP Roller Mounting Bracket Kit
1000SP01304Alaris GP Feet Kit
1000SP01305Alaris GP PSU PCB Kit
1000SP01306Alaris GP Speaker Kit
1000SP01307Alaris GP Docking Station Kit
1000SP01308Alaris GP IrDA PCB Flexi Kit
1000SP00115Asena SP, Assy, Pole Clamp
1000SP00487Asena LVP Battery Pack
1000SP01417Spares Kit (Orange Clip & Screw)
1000SP01323Alaris SP Cam Rail Clamp Only Kit
Alaris® GP Volumetric Pump 76/84 1000SM00013 Issue 3
Page 77
Spare Parts Listing
Individual Components
Part NumberDescription
0000ME00141Stud PE Connector M6 Thread X 15
0000ME00768Cables Ties Hayco 3623793
0000ME00770Bussmann Fuse Gmd-1.25a
1000EL00605Drop Sensor Cable
1000EL00606RS232 Nurse Call Cable
1000EL00607PSU Cable
1000ME00291Asena GW, Assy, Cover Dust Drop Sensor
1000ME00630Handle Spring Asena LVP
1000ME00632Asena LVP Handle Retaining Block
1000ME01845Asena LVP Overmould Handle
1000ME00636Asena LVP Mains Inlet Bracket
1000ME00655Fuse Cover
1000ME01303Magnet IR Detect
1000ME01745Cover RS-232
1000ME00589Cover Battery Asena LVP
1000ME00590Foot Battery Cover Asena LVP
1000ME00649Foot Front Asena LVP
1000ME01611Seal Case Nickel/graphite
1000ME00667Asena LVP Assembly Membrane
0000ME00767V Seals Hinge Pins V5a-nbr
1000EL00643Door Detect Flexible Circuit
1000ME00701Asena LVP GP Top Retainer
0000ME00691Seal O Ring 6id 1csdia silicon
0000ME00423Pad Self Adhesive Double Sided 12x12mm
Keypad & Labels
Part NumberDescription
1000LB00623Alaris GP Keypad BOM
1000LB01040Alaris GP Door Label Set
1000LB00614Alaris GP Label Set BOM
1000LB01475Alaris GP Guardrails Door Label
Software
Part NumberDescription
1000SP01412Alaris LVP GP F/ware Upgrade V1.7.18 Kit
1000CD00028Alaris SMU
1000SP01310Alaris GP Editor Software Kit
Test Equipment
Part NumberDescription
60793GP Series Infusion Set, 2 Y
1000EL00979Converter Cable -USB To Serial
1000EL00980Converter Cable -USB To 4x Serial
1000SP01183Cable RS232 Filtered
Alaris® GP Volumetric Pump 77/84 1000SM00013 Issue 3
Page 78
Spare Parts Listing
Kit Bill Of Materials (BOM)
KitDescriptionDescriptionQty
1000SP01244 Alaris GP Door KitAlaris GP Door1
Assy Screw & Washer M2, 5 X 6 With Patch2
Alaris GP Door Label1
1000SP01246 Alaris GP Hinge Pin KitAssembly Insulated Door Hinge2
V Seals Hinge Pins V5a-NBR2
1000SP01247 Alaris GP Pumping Mech (Minus Motor)
Kit
1000SP01248 Alaris GP Front Case KitCase Front Moulded Asena LVP1
Alaris GP Pumping Mech1
Rivet Snap 03 6.5/7.5 Panel Thickness1
V Seals Hinge Pins V5a-NBR6
Screw M3x8 Button Head With Patch4
Alaris GP Keypad Bom1
Asena LVP Assembly Membrane1
Seal O Ring 6ID 1CSDIA Silicon1
Asena LVP GP Top Retainer1
Screw M3 X 8 Pan Washer Head With Patch1
Foam Gasket Display Asena LVP1
Door Detect Flexible Circuit1
Rivet Snap 03 6.5/7.5 Panel Thickness1
Foot Front Asena LVP2
Seal Case Nickel/graphite1
V Seals Hinge Pins V5a-NBR2
Adhesive Hot Melt Stick 12mm White1
1000SP01250 Alaris GP Rear Case KitCase Rear Moulded Asena LVP1
Seal Case Nickel/Graphite1
Alaris GP Label Set BOM1
Spacer Speaker Label1
1000SP01251 Alaris GP Mains Inlet KitMains Inlet Scha ner Fn9260b Inc Cable1
Asena LVP Gasket Mains Inlet1
Alaris® GP Volumetric Pump 78/84 1000SM00013 Issue 3
Page 79
Spare Parts Listing
Kit Bill Of Materials (BOM)
KitDescriptionDescriptionQty
1000SP01252 Alaris GP Fastener Spares KitPlug Pull M1210
Screw M3 X 8 Pan Washer Head With Patch10
Screw M3x8 Button Head With Patch30
Screw M3x8 Pan Hd T10 Skt & Flat Washer50
Screw M3x10 Button Hd T1020
Screw M3x8 Button Hd T10 Shkpf Washer30
Nut M6 Zp+P10
Washer M6 Internal Shakeproof St Z & C10
Screw Pt Kc30x10 Csk (T8)-Rogard10
Screw Pt K25x12 Pan Hd T8 Hexalobular Sk10
Screw M3x12 Pan Hd Torx (T10)20
Washer M3 Shakeproof External10
Screw Lock RS232 4-40 UNC -4-40 UNC10
Screw M3x6 Flat Head Hexalobular Skt-810
Screw, Pan Hd Slotted 4-40 UNC Nylon10
Screw M2, 5 X 8 Csk Head, With Patch10
Assy Screw & Washer M2, 5 X 6 With Patch10
Assembly Screw & Washer, M3x8 With Patch10
Screw M3 X 8 Extra Low Head With Patch10
Plastic Snap Rivet10
Rivet Snap 03 6.5/7.5 Panel Thickness10
Alaris GP LVP Fixings Kit Instructions1
1000SP01253 Alaris GP Battery Compartment KitCover Battery Asena LVP1
Foot Battery Cover Asena LVP2
1000SP01256 Alaris GP Comms PCB KitAsena LVP RS232 Nurse Call1
Alaris® GP Volumetric Pump 79/84 1000SM00013 Issue 3
Page 80
Spare Parts Listing
Kit Bill Of Materials (BOM)
KitDescriptionDescriptionQty
1000SP01300 Alaris GP Interface PCB KitAsena LVP Interface PCB1
Asena LVP PCB Clip3
1000SP01301 Alaris GP Encoder PCB KitAsena GP Motor Encoder PCB1
Flexi Encoder To Interface PCB1
Nylon Spacer2
1000SP01303 Alaris GP Roller Mounting Bracket KitAssembly Roller Mounting Bracket1
Gasket Assembly Roller LVP1
Screw M3x4 Cap Head Shoulder With Patch2
1000SP01304 Alaris GP Feet KitFoot Front Asena LVP10
Foot Battery Cover Asena LVP10
1000SP01305 Alaris GP PSU PCB KitAssembly PSU Asena LVP1
Insulator PSU Asena LVP1
1000SP01307 Alaris GP Docking Station KitCam Rail Clamp1
Seal Ring V 10mm Dia1
Linkage MDI1
Lever Intermediate MDI Asena LVP1
Spring Extension OD 4.76 WD 0.66 L 19.051
Lever Rail Cam Asena LVP1
Lever Release Medical Device Interface1
Seal Ring V 6mm Dia1
1000SP01308 Alaris GP IrDA PCB Flexi KitIrDA Flexible Circuit1
Cardinal Health,
10020 Pacific Mesa Blvd.,
San Diego, CA 92121,
USA.
Tel: (1) 858 617 2000
Fax: (1) 858 617 2900
www.cardinalhealth.com/alaris
ZA
Cardinal Health,
Unit 2 Oude Molen Business
Park,
Oude Molen Road, Ndabeni,
Cape Town 7405, South Africa.
Tel: (27) (0) 860 597 572
Tel: (27) 21 510 7562
Fax: (27) 21 5107567
www.cardinalhealth.com/za
SA-Technical-Support@
cardinal.com
Alaris® GP Volumetric Pump 82/84 1000SM00013 Issue 3
Page 83
Document History
Appendix E
Page 84
Document History
Document History
IssueDateCO No.AuthorUpdate Description
120/01/065999Ian TylerInitial release
New Tech mode section.
Spare Part Replacement Procedures chapter added.
2December 067111Ian Tyler
3September 088587Ian Tyler
Troubleshooting chapter added.
New Software features added.
Spare Parts Listing appendix added.
Introduce Calibration information.
Update Preventative Maintenance information.
Add information on the Alaris® GP Guardrails® Volumetric Pump.
Update Corrective Maintenance information.
Alaris® GP Volumetric Pump 84/84 1000SM00013 Issue 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.