1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.
2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the
documentation and/or other materials provided with the distribution.
3. The name of the author may not be used to endorse or promote products derived from this software without speci c prior written
permission.
THIS SOFTWARE IS PROVIDED BY THE AUTHOR “AS IS” AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO,
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE
AUTHOR BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT
LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER
CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE)
ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
Alaris® GP Volumetric Pump 2/84 1000SM00013 Issue 3
Contents
Chapter
1. General Information 4
2. Configuration & Calibration 8
3. Preventative Maintenance 19
4. Troubleshooting 39
5. Circuit Descriptions 44
6. Corrective Maintenance 48
Appendix
A. Electromagnetic Compatibility 68
B. Disposal 73
C. Spare Parts Listings 75
D. Service Contacts 81
E. Document History 83
Alaris® GP Volumetric Pump 3/84 1000SM00013 Issue 3
General Information
In this chapter
Introduction 5
Features of the Pump 6
Chapter 1
General Precautions 7
General Information
Introduction
The Alaris® GP Volumetric Pump and the Alaris® GP Guardrails® Volumetric Pump (hereinafter referred to as 'Pump') is a
small lightweight volumetric infusion pump that provides accurate and reliable infusions over a range of rates.
The pump is designed to meet the infusion requirements as specified in the Directions For Use (DFU) for all hospital
departments including general wards, critical and intensive care, operating rooms and accident and emergency rooms.
This pump is suitable for use by appropriately trained clinicians or nurses. This pump can be used for intravenous infusion
modes. Supporting fluid & drug therapy, blood transfusions and parenteral feeding.
Product Familiarity
Ensure that you are fully familiar with the pump by carefully studying the Directions for Use (DFU) prior to operation and
prior to attempting any repairs or servicing. As part of continuous improvement, product enhancements and changes are
introduced from time to time.
Purpose of this Manual
This Technical Service Manual describes how to set up, test and maintain the Alaris® GP Volumetric Pump and the Alaris®
GP Guardrails® Volumetric Pump.
This manual is intended for use by personnel experienced in medical equipment testing and maintenance procedures .
Conventions Used in this Manual
BOLDUsed for Display names, self-test codes, controls and indicators referenced in this manual, for
example, Battery Indicator, access code 212, ON/OFF button.
'Single quotes'Used to indicate cross-references made to another section of this manual. For example, see
Chapter 2, 'Configuration & Calibration'.
underlineUsed to indicate a link to another section within this manual.
ItalicsUsed to refer to other documents or manuals. For example, refer to the relevant Directions for Use
(DFU) for further information. Also used for emphasis, for example, ...if the gap still measures less
than...
Wherever this symbol is shown a Hints & Tips note is found. These notes provide useful advice or
information that may help to perform the task more effectively.
Wherever this symbol is shown a Toolbox note is found. These notes highlight an aspect of test
or maintenance that is important to know about. A typical example is drawing attention to a
software upgrade that you should check has been installed.
Alaris® GP Volumetric Pump 5/84 1000SM00013 Issue 3
Features of the Pump
Alarm indicator
Display
General Information
Alaris®
GP
Door
Softkeys
Chevrons
Run
Bolus
Hold
Options
AC power indicator
Release lever for rotating cam
Mute
Pressure
Battery indicator
On/Off
Door Lever
Handle
Flow sensor
connector (cover
removed for clarity)
RS232/Nursecall
connector (cover
removed for clarity)
Rotating cam to lock onto
horizontal rectangular bars.
Mains fuses
Mains inlet
IR communications port
Potential Equalisation (PE)
Connector
Alaris® GP Volumetric Pump 6/84 1000SM00013 Issue 3
Folded pole clamp
Medical device
interface (MDI)
General Precautions
Attention consult accompanying documents: Prior to using this pump, carefully read the Operating
w
V
B
M
Alaris®
GP
Precautions described in the Directions for Use (DFU).
This pump contains static-sensitive components. Observe strict precautions for the protection of static
sensitive components when attempting to repair and service the pump.
An explosion hazard exists if the pump is used in the presence of flammable anaesthetics. Exercise care
to locate the pump away from any such hazardous sources.
Dangerous Voltage. An electrical shock hazard exists if the casing of the pump is opened or removed.
Refer all servicing to qualified service personnel.
This pump is protected against the effects of high energy radio frequency emissions and is designed
to be fail safe if extremely high levels of interference are encountered. Should false alarm conditions
be encountered, either remove the source of the interference or regulate the infusion by another
appropriate means.
If the pump is dropped, subjected to excessive moisture, humidity or high temperature, or otherwise
suspected to have been damaged, remove it from service for inspection by qualified service personnel.
General Information
When connected to an external power source, a three-wire (Live, Neutral, Earth) supply must be used. If
the integrity of the external protective conductor in the installation or its arrangement is in doubt, the
pump should be operated from the battery.
Alaris® GP Volumetric Pump 7/84 1000SM00013 Issue 3
Chapter 2
Configuration and Calibration
In this Chapter
Entering Service Mode 9
Service Mode - Factory Defaults 10
Service Mode - Configuration 11
Service Mode - Data Set Transfer 14
Service Mode - Calibration 15
Configuration and Calibration
Entering Service Mode
Warning -
At no time should Service Mode be entered while the pump is connected to a patient.
Service Mode should only be accessed by qualified and trained personnel.
Service Mode can be accessed via a three-digit access code that is entered using the following procedure:
1. Hold down b and turn the pump ON a.
2. Enter the access code 212 using the f keys and the NEXT softkey.
3. When the code shows on screen, press OK to confirm.
Select the required option using the f keys and the OK softkey.
For the Alaris® GP Volumetric Pump the options will be as follows:
SERVICE
Factory DefaultsLoad a default data set. Confirm to perform a
cold start.
ConfigurationThis menu comprises a list of options which are
configurable by the user.
Data Set TransferUpload a data set to pump.
CalibrationThis menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVPPerformance Verification Procedure Tests.
For the Alaris® GP Guardrails® Volumetric Pump the options will be as follows:
CQI Events DownloadFor future implementation
Data Set TransferUpload a data set to pump.
ConfigurationThis menu comprises a list of options which are
configurable by the user.
CalibrationThis menu comprises a list of calibrations which
can be performed by the user.
Test Verification/PVPPerformance Verification Procedure Tests.
Factory DefaultsLoad a default data set. Confirm to perform a
cold start.
Factory Defaults
Configuration
Data Set Transfer
Calibration
Test Verification/PVP
SELECT WITH
QUIT
SERVICE
CQI Events Download
Data Set Transfer
Configuration
Calibration
Test Verification/PVP
Factory Defaults
OK
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 9/84 1000SM00013 Issue 3
OK
Configuration and Calibration
Service Mode - Factory Defaults
Factory Defaults
Select the required option using the f keys and the OK softkey.
Default Data Set Replace the current data set with a default data
set.
Cold Start Confirm Confirm clearing and resetting the data set and
calibration data to the factory defaults.
Clear CQI Log File* Confirm clearing all pump history and resetting
the data set to the factory default.
* Alaris® GP Guardrails® Volumetric Pump only.
FACTORY DEFAULTS
Default Data Set
Cold Start Confirm
Clear CQI Log File
SELECT WITH
Default Data Set
1. Press OK to confirm loading the default data set.
QUIT
OK
DEFAULT DATA SET
******
Restoring factory
default data set will
delete the clinically
approved data set
installed.
Delete approved and
restore factory default?
QUIT
WARNING!
******
OK
Cold Start Confirm
1. Press OK to confirm performing a cold start.
Alaris® GP Volumetric Pump 10/84 1000SM00013 Issue 3
COLD START CONFIRM
******
All data set and cal
data will be reset to
default!
The pump will require a
full calibration before
returning to clinical use.
QUIT
WARNING!
******
OK
Configuration and Calibration
Service Mode - Factory Defaults (continued)
Clear CQI Log File
1. Press OK to confirm clearing the CQI Log File.
CLEAR CQI LOG FILE
Service Mode - Configuration
Select the required option using the f keys and the OK softkey.
Date/TimeSets the current date and time used for event
logging.
Software VersionsDisplays the pump software versions.
Serial NumberConfigure the displayed serial number.
Pump ReferencePump specific text to be displayed in user
mode at start up. (20 characters max.)
LanguageConfigure the Language used for display
messages.
Backlight & ContrastAdjust the Backlight and Contrast values
Current Data Set FileDisplays the current data set file details.
******
This will remove all
instrument history and will
restore the factory default
data set by deleting the
clinically approved
installed data set.
Clear current log?
QUIT
******
OK
SERVICE CONFIGURATION
Date/Time
Software Versions
Serial Number
Pump Reference
Language
Backlight & Contrast
Current Data Set File
SELECT WITH
WARNING!
Date/Time
1. Set the correct date and time using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
QUIT
DATE / TIME
AUG–04–2006
09:48
ADJUST WITH
QUITNEXT
OK
OK
Alaris® GP Volumetric Pump 11/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Configuration (continued)
Software Versions
Press 1. OK to exit after verifying Software Version fitted, display will vary depending on software version fitted.
SOFTWARE VERSIONS
SRP 001.005.004
MP 001.005.004
FDP 001.005.004
SP 001.005.004
QUIT
Serial Number
1. Set the serial number required (maximum 9 characters) using the
f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
OK
SOFTWARE VERSIONS
SWV 001.009.000
PKG 001.009.000
MPU 001.009.000
MPT 001.009.000
FDP 001.007.016
SP 001.005.000
LANGUAGE 001.004.005
QUIT
OK
SERIAL NUMBER
8026 - 00000
Pump Reference
1. Set the Pump Reference required using the f keys.
2. Press NEXT to continue to next item to change.
3. Press OK to confirm.
Write Success: .
ADJUST WITH
QUITNEXT
OK
PUMP REFERENCE
---------------------
ADJUST WITH
QUITNEXT
OK
Alaris® GP Volumetric Pump 12/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Configuration (continued)
Language
1. Select the required Language using the f keys.
2. Press OK to confirm.
Languages available will be dependant on the pump
software version.
LANGUAGE
ENGLISH
NORSK - NORWEGIAN
DEUTSCH - GERMAN
FRANCAIS - FRENCH
NEDERLANDS - DUTCH
SVENSKA - SWEDISH
ESPAñOL - SPANISH
ITALIANO - ITALIAN
NONE - VANILLA
SELECT WITH
QUIT
OK
Backlight & Contrast
1. Use the f keys to adjust Backlight and Contrast to required
setting.
2. Press PAR AM to change between Backlight and Contrast.
3. Press OK to confirm.
Select the Dimming parameter to see what the display
would look like when dimmed.
Current Data Set File
1. Press OK to exit after verifying current data set information.
BACKLIGHT & CONTRAST
Backlight = 30%
Contrast = 70%
Dimming = 25%
ADJUST WITH
QUITPARAM
OK
CURRENT DATA SET
Data Set ID: ABCD
Name:
Data_set_EXAMPLE
Version: 2
Last update:
01-Jun-2006 12:34
Hospital name:
Basingstoke General
QUIT
Alaris® GP Volumetric Pump 13/84 1000SM00013 Issue 3
OK
Configuration and Calibration
Service Mode - Data Set Transfer
Upload data set to an Alaris® GP Volumetric Pump
Equipment required:
Alaris® GP Editor Software Kit (1000SP01310) - includes the Alaris® GP Transfer Tool•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) - optional•
USB to RS232 converter 4 way hub (1000EL00980) - optional•
PC - for requirements see Upgrading Firmware PC requirements•
Using the Alaris® GP Transfer Tool allows a released data set to be uploaded to an Alaris® GP Volumetric Pump.
Warning At no time should the Alaris® GP Transfer Tool be used to upload to an Alaris®
GP Volumetric Pump while the pump is connected to a patient.
In Service Mode select Data Set Transfer using the f keys and the OK softkey.
Using the Alaris® GP Transfer Tool select data set to be uploaded. 1.
Press the 2. RS232 or IrDA softkey to select the Comms mode being used.
3.
Connect the pump to PC.
Press the 4. START softkey to begin transfer.
Please ensure the data set ID shown on the pump is identical to the one transferred.5.
Press 6. PASS softkey to confirm correct transfer and exit.
To transfer the data set to another pump repeat steps 2 to 6.7.
Upload data set to an Alaris® GP Guardrails® Volumetric Pump
Equipment required:
Guardrails® Editor V3.1 Software Kit (1000SP01389) or Guardrails® Editor V3.1 Transfer Tool Software Kit (1000SP01390)•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) •
USB to RS232 converter 4 way hub (1000EL00980) - optional•
PC - for requirements see Upgrading Firmware PC requirements•
Using the Guardrails® Editor V3.1 Transfer Tool allows an approved data set to be uploaded to an Alaris® GP Guardrails®
Volumetric Pump.
Warning At no time should the Guradrails® Editor V3.1 Transfer Tool be used to upload
to an Alaris® GP Guardrails® Volumetric Pump while the pump is connected to
a patient.
In Service Mode select
Using the 1. Guardrails® Editor V3.1 Transfer Tool Transfer Tool select data set to be uploaded.
2.
Connect the pump to PC.
Press the 3. START softkey to begin transfer.
Please ensure the data set ID shown on the pump is identical to the one transferred.
4.
Press 5. ACCEPT softkey to confirm correct transfer and exit.
To transfer the data set to another pump repeat steps 2 to 5.6.
Data Set Transfer using the f keys and the OK softkey.
Caution: Loading the Data Set Transfer Tool software is considered a non-clinical service activity.
Interconnecting the pump with a PC may cause the safety or electromagnetic environment to
change while the connection exists. The threat of higher leakage currents or EMI disturbances may
be present. Disconnect the IrDA or RS232 cable connection at both ends following software upload
activities.
For more information relating to the Alaris® GP Editor Software or the Guardrails®
Editor V3.1 Software refer to the relevant Directions For Use supplied with the
software.
Alaris® GP Volumetric Pump 14/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Calibration
Zero Point Calibration
Select the required option using the f keys and the OK softkey.
Wait for the pressure sensors to park.1.
Ensure that an infusion set is not installed.2.
Press the 3. START softkey.
Pump will countdown for 15 seconds.4.
Press the 5. ACCEPT softkey.
Press the 6. PASS softkey.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
Pressure Calibration
Equipment required:
Calibrated Pressure Gauge, minimum specification of •
Accuracy = 0.10% of full scale•
Full scale = 1500mmHg•
Pressure Calibration Set (1000SP01422) Use to calibrate 10 pumps •
and then change
Set up equipment as per figure 2-1 and allow 30 seconds before
proceeding.
Select the required option using the f keys and the OK softkey.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
Alaris® GP Volumetric Pump 15/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Calibration (continued)
Pressure Calibration continued
Pressure source
(50ml/100ml syringe
or similar device)
Pressure
Gauge
Figure 2 - 1 Pressure Calibration Equipment Set Up
Vent to
atmosphere
Turn 3-way tap to close from atmosphere.1.
Press the 2. START softkey.
Apply a pressure of 200mmHg and pump will countdown for 15 seconds.3.
Press the 4. ACCEPT softkey.
Apply a pressure of 800mmHg and pump will countdown for 15 seconds.5.
Press the 6. ACCEPT softkey.
Turn 3-way tap to vent to atmosphere.7.
Turn 3-way tap to close from atmosphere. 8.
Press the 9. NEXT softkey to proceed to Verification Procedure.
Alaris®
GP
Verification Procedure
Software version v1.7.x and belowSoftware version v1.9.0 and above
Apply a pressure of 200mmHg and wait for 5 10.
seconds.
Press the 11. NEXT softkey.
Apply a pressure of 400mmHg and wait for 5 12.
seconds.
Apply a pressure of 500mmHg and wait for 5 10.
seconds.
Press the 11. NEXT softkey.
Turn 3-way tap to vent to atmosphere.12.
Press the 13. PASS softkey.
Press the 13. NEXT softkey.
Apply a pressure of 600mmHg and wait for 5 14.
seconds.
Press the 15. NEXT softkey.
Apply a pressure of 800mmHg and wait for 5 16.
seconds.
Press the 17. NEXT softkey.
Turn 3-way tap to vent to atmosphere.18.
Press the 19. PASS softkey.
Alaris® GP Volumetric Pump 16/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Calibration (continued)
Volumetric Calibration
Select the required option using the f keys and the OK softkey.
Load the primed Infusion Set (60793) into the Pump and set-up 1.
as shown in Figure 2-2 below and adjust the fluid level so that the
meniscus is level with the zero mark.
Press 2. START to begin. Test will run and fluid will be delivered into
the burette.
When 3. Volume delivered! is displayed, check accumulated air
in line value is less than 100µl then enter the volume delivered
into the burette using the f keys and the OK softkey. If
accumulated air in line value is greater than 100µl then repeat test.
If measured value is 19.3ml or less then enter
19.2ml and if value is 20.6ml or higher then enter
20.7ml.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
If no calibration is required (Volume delivered within limits of 4.
19.4ml to 20.5ml) then press PASS to confirm and exit.
If the Calibration value is changed automatically then press 5.
VERIFY and repeat steps 2 to 4.
If the pump still fails replace the Platen and Fingers then repeat 6.
the calibration procedure.
450mm
± 20mm
Alaris®
GP
VOLUMETRIC CALIBRATION
Rate 125ml/h
VTBI 20.0ml
Cal Value 920
Acc’d Air 0μl
START to begin...
QUIT
STARTDETAILS
Figure 2 - 2 Volumetric Calibration Equipment Set Up
Alaris® GP Volumetric Pump 17/84 1000SM00013 Issue 3
Configuration and Calibration
Service Mode - Calibration (continued)
Battery Calibration
Select the required option using the f keys and the OK softkey.
Connect AC Mains to the Pump and press 1. CAL to begin.
When calibration is complete it will display 2. CALIBRATION
SUCCESS or CALIBRATION FAILURE. Press PASS to confirm
successful calibration or REPEAT to perform calibration again.
CALIBRATION
Zero Point Cal
Pressure Cal
Volumetric Cal
Battery Cal
SELECT WITH
QUIT
OK
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 98%
Temperature 22°C
Calibration
CAL to calibrate!
CAL can exceed 10 hrs!
QUIT
DETAILS
CAL
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2335mAh
Chrg Remain 2332mAh
Rel Chrg 100%
Temperature 26°C
Calibration
BATTERY CALIBRATION
Serial No 12345
Current -340mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 98%
Temperature 22°C
Calibration
CALIBRATING . . .
Discharge
QUIT
DETAILS
89%
CAL
BATTERY CALIBRATION
Serial No 12345
Current 0mA
FCC 2502mAh
Chrg Remain 2453mAh
Rel Chrg 100%
Temperature 22°C
Calibration X
CALIBRATION SUCCESS
QUIT
Alaris® GP Volumetric Pump 18/84 1000SM00013 Issue 3
REPEAT
PASS
CALIBRATION FAILURE
QUIT
REPEAT
Chapter 3
Preventative Maintenance
In this chapter
Preventative Maintenance 20
Visual Inspection 20
Recommended Cleaning and Storage 21
Updates 23
Battery Test and Replacement 25
Service Mode - Test Verification/PVP 25
Performance Verification Procedure 38
Preventative Maintenance
Preventative Maintenance
To ensure the pump remains in good operating condition, routine and preventative maintenance inspections are
required. Routine maintenance inspections should be performed by hospital/facility before each use, see Directions For Use for details.
Preventative maintenance inspections should be performed at least every year.
For the preventative maintenance inspection the following should be performed:
Full visual inspection of the pump, internal and external•
Fitting of all updates required•
Battery test and/or replacement•
Clean the pump•
Performance Verification Procedures
•
Following all spare part replacement and repair activities, testing must be performed in
accordance with the Performance Verification Procedure (PVP). Additional testing and
calibration may be required after certain repairs are completed, see table in Chapter 6
‘Corrective Maintenance’ for more information.
Visual Inspection
Open the pump, as per Chapter 6 ‘Corrective Maintenance’ and visually inspect the interior of the pump.
Visually inspect the exterior of the pump checking the following:
Labels should be replaced as required if not flat, legible or fully adhered.
•
Check Keypad for any sign of wear and replace as required. •
Case components must be checked for damage and replaced if necessary. •
Check the pole clamp is not damaged and that it functions correctly.•
Inspect the AC power supply plug and cable for damage.•
The case should be clean and free from IV solution residue, especially near moving parts. •
Check for dried solution deposits on accessible areas of pumping mechanism.•
Alaris® GP Volumetric Pump 20/84 1000SM00013 Issue 3
Preventative Maintenance
Recommended Cleaning and Storage
Cleaning the Pump: -
Before the transfer of the Pump to a new patient and periodically during the use, clean the Pump by wiping over with a
lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution.
Recommended cleaners are:
Brand Concentration
Hibiscrub 20% (v/v)
Virkon 1% (w/v)
Do not use the following disinfectant types:
- NaDcc (such as PRESEPT)
- Hypochlorites (such as CHLORASOL)
- Aldehydes (such as CIDEX)
- Cationic Surfactants (such as Benzalkonium Chloride)
- Iodine (such as Betadine)
- Concentrated Isopropyl alcohol based cleaners will degrade plastic parts.
Before cleaning always switch OFF and disconnect from the AC power supply.
Never allow fluid to enter the casing and avoid excess fluid build up on the
Pump.
Do not use aggressive cleaning agents as these may damage the exterior
surface of the Pump.
Do not steam autoclave, ethylene oxide sterilise or immerse this Pump in any
fluid.
Storing the Pump: -
If the Pump is to be stored for an extended period it should be first cleaned and the internal battery fully charged. Store in
a clean, dry atmosphere at room temperature and, if available, employ the original packaging for protection.
Once every 3 months during storage, carry out functional tests as described in this technical service manual and ensure
that the internal battery is fully charged.
Cleaning and storing the Infusion Set: -
The Infusion Set is a disposable single use item and should be discarded after use according to hospital protocol.
Alaris® GP Volumetric Pump 21/84 1000SM00013 Issue 3
Preventative Maintenance
Recommended Cleaning and Storage (continued)
Cleaning the door: -
Periodically during use (as per hospital policy), clean the door by wiping over with a lint-free cloth, lightly dampened with
warm water and a standard disinfectant / detergent solution. Dry door before use.
To aid cleaning of a door which has been heavily soiled, contaminated or if the door operation is not free moving, then
the door may be removed (see procedure below) then immersed and soaked in warm water with a standard disinfectant
/ detergent.
The door should be allowed to dry fully prior to use.
Door Removal
1. Remove the screw securing the lower hinge lock.
2. Open the lower hinge lock.
3. Pull the door away from lower hinge pin and lift up to remove the door.
4. Clean the door.
5. Refit door in reverse order. Ensure screw is refitted with a torque of 70cNm.
Cleaning the Flow Sensor: -
Before the transfer of the flow sensor to a new infusion set and periodically during use, clean the flow sensor by wiping
over with a lint-free cloth lightly dampened with warm water and a standard disinfectant / detergent solution. Ensure the
connector does not get wet. Dry flow sensor before use.
To aid cleaning of flow sensors which have been heavily soiled, contaminated or if the handle operation is not free
moving, then the flow sensor may be immersed and soaked in clean soapy water (see note below). The inside of the spring
mechanism can be cleaned by activating it whilst submerged in the water.
After cleaning, the sensor should be allowed to dry fully prior to use.
Warning -
The plug of the flow sensor must not be immersed in water as damage will
occur.
Alaris® GP Volumetric Pump 22/84 1000SM00013 Issue 3
Updates
Upgrading firmware
Requirements
PC •
Minimum hardware system requirements: Software requirements:•
1GHz Pentium processor• 512MB RAM• Microsoft Windows 2000 •
1GB of free space on the system hard •
drive
Available configurable RS232 9 pin serial •
or USB communications ports
Preventative Maintenance
CD ROM drive•
K•
eyboard
(service pack 4), or XP
(service pack 2)
Video resolution of 1024 X 768 pixels •
Mouse•
and 16 bit colour depth
The Software Maintenance Utility (SMU) (1000CD00028)•
RS232 cable (1000SP01183) •
USB to RS232 Converter cable (1000EL00979) - optional•
USB to RS232 converter 4 way hub (1000EL00980) - optional•
Firmware CD•
Preparation
Load the SMU software onto the PC•
Connect RS232 connector (using USB to RS232 converter if required) to each pump being updated•
Disconnect• the Battery
Remove the Battery Compartment Plug•
Battery
Compartment
Plug
Software Upload
1. Through the Rear Case, there is a set of three dip switches on the bottom of the Control PCB. Switch 1 should be ON and
switches 2 and 3 should be OFF.
2. Using a small flat blade screwdriver or round ended tweezers configure the dip switch settings
to:
Switch 1 = OFFa.
Switch 2 = ONb.
Switch 3 = ON c.
3. Start the SMU facility to upload the software.
4. Select the Firmware zip file.
5. Select the Comm settings - Comm Port and Baud rate of 460800 (USB to RS232 converter) or 115200 (RS232).
6. Connect the Battery to the Battery Cable.
7. Press Start.
Switch 1 turns the safety battery circuit on but is not required for programming.
Switch 2 forces the pump to turn on.
Switch 3 turns the pump into ‘Boot Mode’, this is only used for programming.
Alaris® GP Volumetric Pump 23/84 1000SM00013 Issue 3
Preventative Maintenance
Updates (continued)
Upgrading firmware continued
8. Once the green bar has reached the far right hand side and the time has reached 0:00 and the flashing green light is a
steady green light, the RS232 connector can be removed from the pump.
9. Disconnect the Battery and turn OFF all dip switches.
10. Wait 5 seconds then reconnect the Battery.
11. Configure the Dip Switches to:
Switch 1 = ONa.
Switch 2 = ON then OFFb.
Switch 3 = OFFc.
12. Refit the Battery Compartment Plug, this prevents fluid ingress.
13. Power up the pump in Service Mode, enter access code 212, then select Configuration > Date/Time and set the current
date and time.
If the Control, Interface or RS232 PCB is replaced, the pump must be re-programmed.
Alaris® GP Volumetric Pump 24/84 1000SM00013 Issue 3
Preventative Maintenance
Battery Test and Replacement
To test the battery perform the battery calibration, as outlined in the procedure in Chapter 2 ‘Configuration and
Calibration’, and verify that all pass criteria are met. If pass criteria are not met then replace the battery.
Battery charge retention will eventually degrade. So where retention is critical the internal battery should be replaced
every three years.
Replace the Main Battery
Remove the two case screws in battery cover, remove 1.
cover and battery.
Fit new battery.2.
Replace battery cover and secure with 2 screws.3.
Battery Cover
It is essential that the battery pack is
calibrated after fitting as failure to do
so will degrade the quoted auxiliary
battery power on this product.
Battery pack
Screw/Washer (x2) 40cNm
Service Mode - Test Verification/PVP
Warning -
At no time should Service Mode be entered while the pump is connected to a patient.
Service Mode should only be accessed by qualified and trained personnel.
Enter access code 212 to view the Service Mode menu (see 'Entering Service Mode' in Chapter 2 for instructions).
Select the Test Verification/PVP option using the f keys and the OK softkey. Press the Quit softkey to return the
pump to the access code screen.
Select the required option using the f keys and the OK softkey.
PVP Work FlowPerformance Verification Procedure (PVP) tests.
Alarms FunctionalityTest the Alarms function correctly.
User InterfaceDisplay, Keypad, LEDs and Audio Tests.
Power SuppliesTest AC Mains and Battery operation.
Sensor TestsPerform tests on the pump sensors.
Comms TestsRS232, IrDA and Nurse Call Tests.
TEST/VERIFY
PVP Work Flow
Alarms Functionality
User Interface
Power Supplies
Sensor Tests
Comms Tests
SELECT WITH
QUIT
Alaris® GP Volumetric Pump 25/84 1000SM00013 Issue 3
OK
Preventative Maintenance
Service Mode - Test Verification/PVP (continued)
PVP Work Flow
This test is used to con rm that the Pump is functioning correctly.
Press the START softkey to begin.
The PVP Work Flow will iterate through the following tests:
Software Versions•
Date/Time (read only)•
Serial Number (read only)•
Door Frame•
Safety Clamp•
Sear•
Audio Test•
Chequerboard•
LED’s•
Keypad•
Nurse Call•
RS232 Loop Back•
Pumping E ciency Test•
Downstream Occlusion Pressure Test•
PVP WORK FLOW
PVP Work Flow Test
QUIT
START
Alarms Functionality•
Volumetric (Accuracy Test) Calibration•
See individual test details in this chapter or in Chapter 2 'Con guration and
Calibration' for instructions.
Alaris® GP Volumetric Pump 26/84 1000SM00013 Issue 3
Loading...
+ 58 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.