Agilent Technologies 83480A User Manual

Agilent 83480A Digital Communications Analyzer Agilent 54750A Digitizing Oscilloscope Quick Start Guide
© Copyright Agilent Technologies 2000 All Rights Reserved. Repro­duction, adaptation, or trans­lation without prior written permission is prohibited, except as allowed under copy­right laws.
Agilent Part No. 83480-90049 Printed in USA June 2000
Agilent Technologies Lightwave Division 1400 Fountaingrove Parkway Santa Rosa, CA 95403-1799, USA (707) 577-1400
Notice.
The information contained in this document is subject to change without notice. Com­panies, names, and data used in examples herein are ficti­tious unless otherwise noted. Agilent Technologies makes no warranty of any kind with regard to this material, includ­ing but not limited to, the implied warranties of mer­chantability and fitness for a particular purpose. Agilent Technologies shall not be lia­ble for errors contained herein or for incidental or conse­quential damages in connec­tion with the furnishing, performance, or use of this material.
Restricted Rights Legend.
Use, duplication, or disclo­sure by the U.S. Government is subject to restrictions as set forth in subparagraph (c) (1) (ii) of the Rights in Technical Data and Computer Software clause at DFARS 252.227-7013 for DOD agencies, and sub­paragraphs (c) (1) and (c) (2) of the Commercial Computer Software Restricted Rights clause at FAR 52.227-19 for other agencies.
Warranty.
This Agilent Technologies instrument product is war­ranted against defects in
material and workmanship for a period of one year from date of shipment. During the war­ranty period, Agilent Technol­ogies will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility desig­nated by Agilent Technolo­gies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay ship­ping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country.
Agilent Technologies war­rants that its software and firmware designated by Agi­lent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Tech­nologies does not warrant that the operation of the instru­ment, or software, or firmware will be uninterrupted or error­free.
Limitation of Warranty.
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer­supplied software or interfac­ing, unauthorized modifica­tion or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Agilent Technologies specifically dis­claims the implied warranties of merchantability and fitness for a particular purpose.
Exclusive Remedies.
The remedies provided herein are buyer's sole and exclusive remedies. Agilent Technolo-
gies shall not be liable for any direct, indirect, special, inci­dental, or consequential dam­ages, whether based on contract, tort, or any other legal theory.
Safety Symbols.
CAUTION
The
caution
sign denotes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in damage to or destruction of the product. Do not proceed beyond a cau­tion sign until the indicated conditions are fully under­stood and met.
WAR NING
The
warning
sign denotes a hazard. It calls attention to a procedure which, if not cor­rectly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning sign until the indicated conditions are fully understood and met.
The instruction man­ual symbol. The prod­uct is marked with this warning symbol when it is necessary for the user to refer to the instructions in the manual.
The laser radiation symbol. This warning symbol is marked on products which have a laser output.
The AC symbol is used to indicate the required nature of the line module input power.
The ON symbols are
|
used to mark the posi­tions of the instrument power line switch.
The OFF symbols
are used to mark the positions of the instru­ment power line switch.
The CE mark is a reg­istered trademark of the European Commu­nity.
The CSA mark is a reg­istered trademark of the Canadian Stan­dards Association.
The C-Tick mark is a registered trademark of the Australian Spec­trum Management Agency.
This text denotes the
ISM1-A
instrument is an Industrial Scientific and Medical Group 1 Class A product.
Typographical Conven­tions.
The following conventions are used in this book:
Key type
for keys or text located on the keyboard or instrument.
Softkey type
for key names that are displayed on the instru­ment’s screen.
Display type
for words or characters displayed on the computer’s screen or instru­ment’s display.
User type
for words or charac-
ters that you type or enter.
Emphasis
type for words or characters that emphasize some point or that are used as place holders for text that you type.
ii
What You’ll Find in This Book
What You’ll Find in This Book
This book gives you a quick introduction to the instrument. It is not intended as a comprehensive guide, but rather as a starting point to start using the instrument quickly. This book is divided into three sections:
Chapter 1 provides a description of the instrument and a quick tour of
the front panel in a descriptive format, with examples that help to illustrate the text.
Chapter 2 provides several tutorials that show how to get up and
running, how to scale a waveform vertically or horizontally, and how to use the Help menu.
Chapter 3 provides a few simple hints in case you are experiencing
difficulty in either getting the mainframe or plug-in module to perform properly.
To best meet your learning needs, you can use this book in one of two ways:
1
Read “A Quick Tour of the Front Panel” on page 1-4 before performing the exercises.
2
Go straight to the exercises. This method is for users who prefer to learn how to use the front panel by experimenting with the front-panel controls.
If you are already familiar with the information in this book, you can refer to the
Agilent 83480A, Agilent 54750A User’s Guide
use the instrument to make many common measurements. It assumes you have already learned how to use the front panel of the instrument, by first reading this quick start guide.
All calibration and repair information is contained in the optional
Agilent 83480A, Agilent 54750A Service Guide
information is contained in the
mer’s Guide
.
Agilent 83480A, Agilent 54750A Program-
, which shows you how to
, while all programming
iii
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC Publica­tion 61010-1, Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, and has been supplied in a safe condition. The instruction documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
WARNING
WARNING
WARNING
WARNING
WARNING
There are many points in the instrument which can, if contacted, cause personal injury. Be extremely careful. Any adjustments or service procedures that require operation of the instrument with protective covers removed should be performed only by trained service personnel.
If this instrument is not used as specified, the protection provided by the equipment could be impaired. This instrument must be used in a normal condition (in which all means for protection are intact) only.
To prevent electrical shock, disconnect the Agilent (Product variable) from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
This is a Safety Class 1 product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
WARNING
For continued protection against fire hazard, replace line fuse only with same type and ratings, (type T 0.315A/250V for 100/120V operation and 0.16A/250V for 220/240V operation). The use of other fuses or materials is prohibited. Verify that the value of the line­voltage fuse is correct.
iv
General Safety Considerations
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
Before this instrument is switched on,
make sure its primary power circuitry has been adapted to the voltage of the ac power source. Failure to set the ac power input to the correct voltage could cause damage to the instrument when the ac power cable is plugged in.
This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and 664 respectively.
VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4°C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.
Always use the three-prong ac power cord supplied with this instrument. Failure to ensure adequate earth grounding by not using this cord may cause instrument damage.
connect ac power until you have verified the line voltage is correct.
Do not
Damage to the equipment could result.
This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range.
CAUTION
Electrostatic discharge (ESD) on or near input connectors can damage circuits inside the instrument. Repair of damage due to misuse is
covered under
not
warranty. Before connecting any cable to the electrical input, momentarily short the center and outer conductors of the cable together. Personnel should be properly grounded, and should touch the frame of the instrument before touching any connector.
v

Contents

General Safety Considerations iiv
1 Introducing the Instrument
The Agilent 83480A, Agilent 54750A—At a Glance 1-2 A Quick Tour of the Front Panel 1-4
2 Operating the Instrument
Operating the Instrument 2-2 Equipment Used in These Exercises 2-3 Exercise 1. Installing a Plug-in Module 2-4 Exercise 2. Turning On the Instrument 2-5 Exercise 3. Connecting a Signal 2-6 Exercise 4. Finding the Signal 2-7 Exercise 5. Modifying the Setup 2-8 Exercise 6. Making Measurements on the Signal 2-13 Exercise 7. Using the Help Menu 2-17
3 In Case of Difficulty
In Case of Difficulty 3-2 If the Mainframe Does Not Operate 3-3 If the Plug-in Does Not Operate 3-4
Contents-1
1
Entry devices 1-4 Disk drive 1-4 CAL signal 1-4 Hardkeys 1-5 Display 1-5 Softkeys 1-6

Introducing the Instrument

Introducing the Instrument
The Agilent 83480A, Agilent 54750A—At a Glance
The Agilent 83480A, Agilent 54750A—At a Glance
The Agilent 83480A-series digital communications analyzer or Agilent 54750A-series digitizing oscilloscope are modular, high-performance mainframes that contain digitizers, timebases, and display circuitry behind vertical channel plug-in modules containing samplers, IF amplifiers, power monitoring circuitry and O/E converter (Agilent 83485A).
The Agilent 83480A-series digital communications analyzer include these unique capabilities:
• A full range of custom and standard telecom and datacom masks and tem­plates with automated conformance and margin testing.
• Automated extinction-ratio and q-factor measurements.
• The ability to set the timebase display to bit period, select the appropriate telecom bit rate, and enter the number of bits to be displayed.
If you wish to add these capabilities to an Agilent 54750A-series digitizing oscilloscope, or use the Agilent 83485A optical plug-in module, a firmware upgrade must first be installed. Order the Agilent 83480K communications firmware kit and follow the installation instructions.
NOTE
The Agilent 83480A and Agilent 54750A mainframes do designed for use with the Agilent 54710A and Agilent 54720A.
The instrument also has software modularity by having a 3-1/2 inch disk drive and flash ROMs to allow for simple firmware upgrades of the firmware fea­tures.
The plug-in modules provide optical and electrical signal conditioning for the A/D converters that are inside the mainframe.
1-2
accept plug-in modules
not
Introducing the Instrument
The Agilent 83480A, Agilent 54750A—At a Glance
Certification
Agilent Technologies certifies that this product met its published specifica­tions at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute’s calibration facility, and to the calibration facilities of other Interna­tional Standards Organization members.
1-3
Introducing the Instrument

A Quick Tour of the Front Panel

A Quick Tour of the Front Panel
This section describes how to use the front panel. This information will help you gain the full use of your instrument in the shortest possible time. This sec­tion divides the front panel into the following areas:
• Entry devices
• Disk drive
• CAL signal
• Hardkeys
•Display
•Softkeys
Entry devices
Disk drive
CAL signal
The entry devices include the knob, arrow keys, and keypad. The knob pro­vides continuous control, while the arrow keys are step controls. The keypad allows you to enter specific values.
The disk drive is a 3-1/2 inch, high-density, MS-DOS
®
compatible drive. You can use the disk drive for loading applications, storing and loading waveforms, storing and loading instrument setups, printing waveforms to a file, or loading new system firmware.
MS-DOS
is a registered trademark of Microsoft Corporation.
The CAL signal is a dc voltage level used during the calibration routines. It is internally generated and only activated during calibration routines. It can be controlled by the user by pressing:
Utility
Calibrate
,
1-4
Introducing the Instrument
A Quick Tour of the Front Panel
Hardkeys
The two types of hardkeys are instant action and menu.
Clear display
and
Run
are examples of instant action keys because as soon as you press them, the instrument changes operating states.
Time base
and
Trigger
are examples of menu keys. Pressing them causes softkey menus to come up on the right side of the display.
Display
Figure 1-1. The front panel of the instrument
Figure 1-2 shows the main areas of the display, and how you can interpret
some of the instrument’s settings from the display.
1-5
Introducing the Instrument
A Quick Tour of the Front Panel
Softkeys
Figure 1-2. The instrument display
The softkeys are to the right of the display, and instead of placing a label on the softkey itself, the label is actually located on the display next to the soft­key. These labels on the display are called menus, and the menu being dis­played depends on which key has been pressed. For example, pressing
base
brings up the Time Base menu, while pressing
Trigger
brings up the Trigger
Time
menu.
There are five types of softkeys, and there are some differences on how each type of softkey operates. The five types of softkeys are toggle, menu, continu­ous, list, and single choice. Because the Trigger menu contains all five types of softkeys, it is used over the next few pages to demonstrate the differences among the softkeys.
When you press the
Trigger
key, the Trigger softkey menu comes up on the
right side of the display as shown in Figure 1-3.
Toggle softkeys
A typical example of the toggle softkeys is
Sweep
. A toggle softkey allows you to choose between two or more selections, and you change selections by pressing that same softkey.
1-6
Sweep
has choices of either triggered or freerun.
Introducing the Instrument
A Quick Tour of the Front Panel
The highlighted choice is the selection that is currently active. In Figure 1-3, triggered is highlighted, which indicates that the instrument is set to triggered sweep.
Figure 1-3. Toggle softkey selection
Menu softkeys
A typical example of the menu softkeys is
External Scale...
(refer to Figure 1-3
on page 1-7). The ellipses after the word "scale" indicate that when you press
this softkey, a new softkey menu comes up on the display that allows you to make several additional choices (refer to Figure 1-4). To exit this menu, sim­ply press the
Done
softkey. The display returns to the previous menu.
1-7
Introducing the Instrument
A Quick Tour of the Front Panel
Figure 1-4. Menu softkey selection
Continuous softkeys
A typical example of the continuous softkeys is
Level
. It has a range of values
that can be changed with the keypad, arrow keys, or knob.
Figure 1-5. Continuous softkey selection
1-8
Introducing the Instrument
A Quick Tour of the Front Panel
Because the Trigger menu only has one of this type of softkey, it is active when the menu comes up on the display. You do not need to press the
Level
softkey when you want to change the trigger level. However, the Time base menu has at least two of this type of softkey,
and Position
. In this case, only one of the softkeys is active at a time. To make
Bit Rate
(
Agilent 83480A only
) and
Scale
the other softkey active, you must press the corresponding softkey. You can tell which softkey is currently active because it has a different colored back­ground from the rest of the softkeys.
When using time base scale or channel scale, the knob and arrow keys operate in two modes. One mode is a 1-2-5 sequence, while the other mode is a fine mode that allows you to make smaller incremental changes. You activate the fine mode by pressing with the word
Fine
active, you will see the word You get out of the Fine mode by simply pressing by pressing the arrow key with the word
Shift
on the keypad, followed by pressing the arrow key
written in blue above that key. When the fine mode is
Fine
at the top-right corner of the graticule area.
Shift
on the keypad, followed
Fine
written above it.
List softkeys
A typical example of the list softkeys is
Hysteresis
. It has several stacked rect­angles that represents a list of choices you can choose from. The top rectangle shows the selection that is currently active. In the example below, normal is the current selection.
Figure 1-6. List softkey selection
1-9
Introducing the Instrument
A Quick Tour of the Front Panel
You change selections by pressing the
Hysteresis
softkey, which brings up a sec­ond level softkey menu on the display. Then, you use the softkeys, arrow keys, or knob to change selections. To activate that selection, press the then the display returns to the previous softkey menu. If you press the
Enter
softkey,
Cancel
softkey, the instrument returns to the previous selection.
Single choice softkeys
A typical example of the single choice softkeys is
Enter
. There are several varia­tions of this type of softkey throughout the many softkey menus. When you press the softkey, the instrument executes that softkey command.
Figure 1-7. Single choice softkey selection
1-10
2
Equipment Used in These Exercises 2-3 Exercise 1. Installing a Plug-in Module 2-4 Exercise 2. Turning On the Instrument 2-5 Exercise 3. Connecting a Signal 2-6 Exercise 4. Finding the Signal 2-7 Exercise 5. Modifying the Setup 2-8 Exercise 6. Making Measurements on the Signal 2-13 Exercise 7. Using the Help Menu 2-17

Operating the Instrument

Operating the Instrument

Operating the Instrument
Operating the Instrument
This chapter contains several exercises that will familiarize you with the oper­ation of the instrument.
Exercises 1–3 show how to get the instrument up and running.
Exercises 4 and 5 demonstrate the vertical, horizontal, and trigger menus by
having you display a signal and then use the channel and time base menus to expand the signal.
Exercise 6 shows you how to use the markers and automatic
measurement features to make a pulse width and voltage amplitude measurement.
Exercise 7 introduces you to the Help menu.
You will find it easier to perform these exercises if you are at least a little familiar with how to use the front panel. The best way to learn how to use the front panel is to read Chapter 1, “Introducing the Instrument”.
Note
Clean the cabinet using a damp cloth only.
2-2
Operating the Instrument

Equipment Used in These Exercises

Equipment Used in These Exercises
All of the pictures of the display in this book were generated using an Agilent 83480A with an Agilent 83483A, 54751A electrical plug-in module, and the signal from the Agilent 8133A pulse generator. If you are using a different plug-in module or signal source, your display and instrument settings may vary from the pictures in this book.
With the Agilent 83480A, the Agilent 83485A optical/electrical plug-in module could also have been used. In this case, we would connect the signal to the electrical input of the Agilent 83485A.
2-3
Operating the Instrument

Exercise 1. Installing a Plug-in Module

Exercise 1. Installing a Plug-in Module
The purpose of a plug-in module is to provide measurement channels, includ­ing sampling for the mainframe. A plug-in module scales the input signal, sets the bandwidth of the system, and allows the offset to be adjusted so the signal can be viewed. The output of the plug-in module is an analog signal that is applied to the A/D converters inside the mainframe.
1
Slide the plug-in module into either slots 1 and 2 or slots 3 and 4.
Note
The plug-in module will
If the mainframe is already turned on, you do not need to turn it off before you install or remove a plug-in module. After a plug-in module is installed, the mainframe identifies the plug-in module and the slot in which it is installed.
Note
In the examples to follow, it is assumed the Agilent 83483A or Agilent 54751A plug-in module is resident in slots 1 and 2 of the mainframe.
2
Finger-tighten the knurled screw at the bottom of the plug-in module.
The RF connectors on the rear panel of the plug-in module are spring-loaded. Finger-tighten the knurled screw to ensure there is a good ground connection between the plug-in module and the mainframe.
3
To remove the module loosen the knurled screws and then pull on the screws to take out the module. Do
operate if plugged into slots 2 and 3.
not
pull on the cables or connectors.
not
2-4
Operating the Instrument

Exercise 2. Turning On the Instrument

Exercise 2. Turning On the Instrument
1
Set the line-voltage switch on the rear panel to the correct voltage selection for your power source.
2
Connect a power cord from a power source to the rear-panel connector of the mainframe.
3
Set the rear-panel line switch to the ON position.
CAUTION
When the mainframe is set to the 115 VAC mode, use the supplied power cord. If you use an alternate power cord that of the instrument, the power cord may become warm or the instrument may not operate.
4
Set the front-panel power switch to ON.
After a few seconds, you will see a graticule on the display and a softkey menu on the right side of the display. If the display is blank or if the message,
Power-On Test Failed
Chapter 3, “In Case of Difficulty” or to the optional
Service Guide
.
, appears at the top of the display, refer to
cannot
handle the power requirements
Agilent 83480A, 54750A
2-5
Operating the Instrument

Exercise 3. Connecting a Signal

Exercise 3. Connecting a Signal
CAUTION
CAUTION
Electrostatic discharge (ESD) on or near input connectors can damage circuits inside the instrument. Repair of damage due to misuse is warranty. Before connecting any cable to the electrical input, momentarily short the center and outer conductors of the cable together. Personnel should be properly grounded, and should touch the frame of the instrument before touching any connector. For more information regarding ESD, refer to the
Agilent 83480A, 54750A User’s Guide
1
Using the Agilent 8133A pulse generator (or equivalent), set a trigger and square wave signal to the following:
Trigger level . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV (do not exceed 2 Vp-p)
Pulse level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mV (do not exceed 2 Vp-p)
Pulse frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 MHz
Cables can build up high voltage levels that can discharge into the channel or trigger inputs and cause damage. Cables should be discharged using a short or a 50Ω terminator prior to connection.
2
Make sure the trigger and pulse signals are enabled.
3
Connect the trigger to the trigger input (channel 2) on the Agilent 83483A, 54751A electrical plug-in module.
4
Connect the pulse signal to the electrical input (channel 1) on the Agilent 83483A, 54751A electrical plug-in module.
.
covered under
not
5
Display the Setup menu (on the Agilent 83480A, 54750A) and select the default setup by pressing:
Setup
Default setup
,
The default setup puts the instrument in a predefined operating condition. This gives you a known starting point for the following exercises. The default condi­tions are listed in the setup section of the
.
Guide
2-6
Agilent 83480A, 54750A User’s
Operating the Instrument

Exercise 4. Finding the Signal

Exercise 4. Finding the Signal
The purpose of this exercise is to demonstrate the Autoscale function. The Autoscale function can be used to automatically set up the display for most input signals.
To activate the Autoscale function, press
Figure 2-1 will be displayed.
When the Autoscale function is enabled, the instrument quickly analyzes the signal and sets up the vertical, horizontal, and trigger to optimize the display of the signal. Autoscale can find repetitive signals with:
• a frequency greater than or equal to 50 Hz
• a duty cycle greater than 1%
• an amplitude greater than 50 mV p-p
Autoscale
. A screen similar to
Figure 2-1. Signal displayed using the Autoscale function
If you unintentionally press the function, press existed just prior to pressing the
Shift, Undo Autoscale
Autoscale
Autoscale
key or wish to disable the Autoscale
. The instrument is reset to the settings that
key.
2-7
Operating the Instrument

Exercise 5. Modifying the Setup

Exercise 5. Modifying the Setup
The purpose of this exercise is to show you the Channel and Time Base menus.
Changing the vertical scale
1
To display the Channel menu and change the vertical scaling, press:
Channel
(located on the plug-in module),
Change the scaling so the vertical height of the signal is about eight divisions, using the knob, arrow keys or keypad.
Scale
Selecting Fine mode
When the knob and arrow keys are ing in a 1-2-5 sequence. You can change the scaling in 1 mV increments by using the keypad, or by setting the knob and arrow keys to the Fine mode.
2
To set the instrument to the Fine mode, press:
Shift, Fine
The word fine mode is activated.
FINE
appears at the top-right corner of the display area when the
in the Fine mode, they change the scal-
not
2-8
Operating the Instrument
Exercise 5. Modifying the Setup
Using the keypad
When using the Agilent 83483A, 54751A electrical plug-in module, the knob or arrow key changes the scaling from 50 mV/div to 100 mV/div, in one step. At 50 mV/div the signal is clipped, and at 100 mV/div the signal is to take advantage of the full vertical size of the display. In this case, you can use the keypad to set the scaling to 75 mV/div.
3
To set the scaling to 75 mV with the keypad, type the number 75 on the keypad. Then press
milli
, for millivolts, to select the unit of measure.
expanded
not
When you select a key with a unit of measure written on it, you do press the that do ing on the value you enter with the keypad, the instrument may round the value up or down. For example, if you try to enter 75.75 mV/div, the instru­ment will automatically round the entry to 75.8 mV/div.
Enter
key. You only press the Enter key if you are entering numbers
require a unit of measure or that are in volts or seconds. Depend-
not
need to
not
Setting the offset
Offset is similar to the vertical position on analog instruments, except offset is calibrated. If offset is set to –10 mV, then you know the vertical center of the display is –10 mV. (For optical channels, offset is referenced two divisions above the bottom of the display.)
4
To move the waveform vertically on the display, press:
Offset
Readjust the scaling and offset until the signal is expanded vertically and is cen­tered on the display as shown in Figure 2-2.
2-9
Operating the Instrument
Exercise 5. Modifying the Setup
Figure 2-2. Signal vertical display
Note — Agilent 83480A only
You can automatically adjust the vertical scale, instead of following the preceding pro­cedure by pressing:
Channel,
Channel autoscale
The amplitude scale is optimized for the input signal while the timebase settings are left alone.
Changing the time base
5
To display the Time Base menu and change the horizontal scale, press:
Time base
Use the knob, arrow keys or keypad to change the scaling so the horizontal width of the positive portion of the signal is about eight divisions.
2-10
Scale
,
Operating the Instrument
Exercise 5. Modifying the Setup
Using the keypad
6
When the knob and arrow keys are not in the Fine mode, they operate in a 1-2-5 sequence, changing the scale from 5 ns/div to 2 ns/div in one large step. You can use Fine mode so the knob and arrow keys adjust in smaller increments, or you can use the keypad to enter a smaller value, like 1.4 ns/div.
Using the Position function
7
To horizontally center the signal on the display, press:
Position
The Position function moves the waveform horizontally on the display. (Posi­tion is sometimes referred to as delay.) The reference point is indicated by an arrow at the bottom of the display as shown in Figure 2-3.
Figure 2-3. Signal horizontal display
2-11
Operating the Instrument
Exercise 5. Modifying the Setup
Setting the reference
8
To set the reference to the left or center of the graticule area, press:
Reference
2-12
Operating the Instrument

Exercise 6. Making Measurements on the Signal

Exercise 6. Making Measurements on the Signal
The purpose of this exercise is to familiarize you with the Marker menu and the automatic measurement features of the instrument. You will begin by using the manual markers to make a pulse width and an amplitude measure­ment on channel 1. After that, you will make the same measurements using the automatic measurement features. There are many additional features built into the instrument. Refer to the more information on making automatic measurements.
Selecting the manual marker mode
1
To select the manual marker mode, press:
Agilent 83480A, 54750A User’s Guide
for
Marker
Change the mode selection by using the knob, arrow keys, or softkeys.
mode, manual, Enter
,
Setting the X, Y source
In the manual marker mode, there are two X-markers and two Y-markers. You can position the X-markers horizontally and the Y-markers vertically on the display. Use the X-markers to make timing measurements and the Y-markers to make amplitude measurements.
2
To set the X1, Y1 Source to channel 1, press:
X1, Y1 Source, channel 1, Enter
3
To set the X2, Y2 Source to channel 1, press:
X2, Y2 Source, channel 1, Enter
2-13
Operating the Instrument
Exercise 6. Making Measurements on the Signal
Positioning the markers
4
To position the X1 marker, press:
X1 Position
Set the X1 marker to the left edge of the pulse.
5
To position the X2 marker, press:
X2 Position
Set the X2 marker to the right edge of the pulse.
6
To position the Y1 marker, press:
Y1 Position
Set the Y1 marker to the bottom of the pulse.
7
To position the Y2 marker, press:
Y2 Position
Set the Y2 marker to the top of the pulse.
The display will be similar to Figure 2-4.
Figure 2-4. Signal custom measurements
2-14
Operating the Instrument
Exercise 6. Making Measurements on the Signal
Notice the ∆X and ∆Y at the bottom of the display. The ∆X value is the width measurement, and the ∆Y value is the amplitude measurement. The markers allow you to make custom measurements on signals. Remember, they are tied directly to the waveform data stored in memory, they are just positions on the display.
Making an automatic +width measurement
8
To make an automatic +width measurement on channel 1, press:
not
Shift, +width
The automatic measurement results are displayed at the bottom left of the screen.
channel 1, Enter
,
Making an automatic amplitude measurement
9
To make an automatic amplitude measurement on channel 1, press:
Shift, V amptd
The IEEE pulse parameter standards are used for all of the automatic measure­ments, unless changes are made in the define measurement menu.
Automatic measurements are preferred because they are faster and more repeatable than manual marker measurements. Automatic measurements use the waveform data stored in memory for all calculations, and the waveform data in memory has a higher resolution than the resolution of the display. The manual markers are positions on the display.
The automatic measurement results are placed on the display where the verti­cal scale and offset values are normally displayed as shown in Figure 2-5.
channel 1
,
Enter
,
tied to the waveform data in memory, they are just
not
2-15
Operating the Instrument
Exercise 6. Making Measurements on the Signal
Figure 2-5. Signal automatic measurements
Displaying the scale and offset values
10
To redisplay the vertical scale and offset values, press:
Shift, Clr meas
The automatic measurement results are replaced by the vertical scale and off­set values.
2-16

Exercise 7. Using the Help Menu

Exercise 7. Using the Help Menu
Operating the Instrument
To display the Help menu, press played.
Figure 2-6. The instrument Help menu
A three-column index appears on the display. The left column lists the fea­tures of the instrument, the middle column lists the front-panel key, and (if needed) the right column lists the softkey you press to find that feature. You can use either the knob or arrow keys to scroll through the Help menu.
Help
. A screen similar to Figure 2-6 is dis-
2-17
3
If the Mainframe Does Not Operate 3-3 If the Plug-in Does Not Operate 3-4

In Case of Difficulty

In Case of Difficulty

In Case of Difficulty
In Case of Difficulty
This chapter provides suggestions for you to follow if the mainframe or plug-in module fails to operate. For a complete listing of display messages, refer to the
Agilent 83480A, 54750A User’s Guide
.
CAUTION
For complete service information, refer to the optional
54750A Service Guide
Review the procedure being performed when the problem occurred. A few minutes spent performing some simple checks, before calling Agilent Technol­ogies or returning the unit for service, may save waiting for your instrument to be repaired.
Electrostatic discharge (ESD) on or near input connectors can damage circuits inside the instrument. Repair of damage due to misuse is warranty. Before connecting any cable to the electrical input, momentarily short the center and outer conductors of the cable together. Personnel should be properly grounded, and should touch the frame of the instrument before touching any connector. Refer to the for further information regarding ESD.
.
Agilent 83480A, 54750A User’s Guide
Agilent 83480A,
covered under
not
3-2

If the Mainframe Does Not Operate

If the Mainframe Does Not Operate
If the mainframe does not operate, check the following:
1
Is the line fuse good?
2
Does the line socket have power?
3
Is the unit plugged in to the proper ac power source?
4
Is the mainframe turned on?
5
Is the rear-panel line switch set to on?
6
Will the mainframe power up
If the mainframe still does not power up, refer to the optional
Agilent 83480A, 54750A Service Guide
fied service department. Refer to the for a complete listing of Agilent Sales and Service Offices.
without
the plug-in module installed?
or return the mainframe to a quali-
Agilent 83480A, 54750A User’s Guide
In Case of Difficulty
3-3
In Case of Difficulty

If the Plug-in Does Not Operate

If the Plug-in Does Not Operate
If the plug-in does not operate:
1
Make the following checks:
• Is the plug-in module firmly seated in the mainframe slot?
• Are the knurled screws at the bottom of the plug-in module finger-tight?
• If other equipment, cables, and connectors are being used with the plug-in
module, are they connected properly and operating correctly?
• Review the procedure for the test being performed when the problem ap-
peared. Are all the settings correct? Can the problem be reproduced?
• Are the connectors clean? Refer to the
for more information about cleaning the connectors.
Guide
• Does the plug-in module work in the other two slots (1,2 or 3,4)?
2
Perform the following procedures:
• Make sure the instrument is ready to acquire data by pressing
• Verify the pulse output is connected to channel 1 on the Agilent 83483A,
54751A electrical plug-in module.
• Verify the trigger output is connected to the trigger input.
• Find any signals on the channel inputs by pressing
• Verify the Agilent 8133A pulse generator is set to:
Waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . square wave
Amplitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 500 mV
Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50 MHz
Trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200 mV
Pulse/trigger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . enabled
• See if any signals are present at the channel inputs by pressing:
Trigger
Sweep, freerun
,
Agilent 83480A, 54750A User’s
Run
.
Autoscale
.
3-4
In Case of Difficulty
If the Plug-in Does Not Operate
After viewing the signal, press
triggered
.
• Make sure Channel Display is on by pressing:
Channel
Display on off
,
• Make sure the channel offset is adjusted so the waveform is not clipped off the display.
• If you are using the plug-in module only as a trigger source, make sure at least one other channel is turned on.
If all of the channels are turned off, the mainframe will
not
trigger.
• Make sure the mainframe identifies the plug-in module by pressing:
Utility
System config...
,
The calibration status of the plug-in modules is listed near the bottom of the display, in the box labeled "
Plug-ins
". If the model number of the plug-in module is listed next to the appropriate slot number, then the mainframe has identified the plug-in.
~known"
If "
is displayed, instead of the model number of the plug-in
module, remove and reinsert the plug-in module in the same slot. If
~known
"
" is still displayed, then the memory contents of the plug-in module are corrupt. Refer to the service guide for the plug-in module or contact a qualified service department. If all of the above steps check out okay, and the plug-in module still does not operate properly, then the problem is beyond the scope of this book. Refer to the plug-in module service guide or return the plug-in to a qualified service department.
3-5

Index

Symbols
+width key, 2-15
A
amplitude, 2-7
measurement, 2-14
scale, 2-10 analog signal, 2-4 automatic measurement, 2-13, 2-15 Autoscale key, 2-7, 3-4
B
bandwidth, 2-4 Bit Rate softkey, 1-8
C
cabinet, cleaning, iiv CAL Signal overview, 1-4 Calibrate softkey, 1-4 Cancel softkey, 1-9 care
of cabinet, iiv Channel autoscale softkey, 2-10 Channel key, 2-8, 3-5 channel scale, 1-9 classification
product, iiv cleaning
cabinet, iiv Clear display key, 1-5 Clr meas key, 2-16 connecting a signal, 2-6 continuous softkeys, 1-8 custom measurement, 2-14
D
Default setup softkey, 2-6 description of the instrument, 1-1 disk drive overview, 1-4 display
messages, 3-2
overview, 1-5 Display softkey, 3-5 Done softkey, 1-7
duty cycle, 2-7
E
electrical plug-in module, 2-3, 2-8 Enter softkey, 1-91-10, 2-13 entry device overview, 1-4 equipment used for display, 2-3 expanding the signal, 2-8 External Scale softkey, 1-7
F
finding input signals, 2-7 Fine softkey, 1-9, 2-8 freerun softkey, 3-4 front panel, 1-4, 1-10 fuse
values, iiv
H
hardkey overview, 1-5 Help key, 2-17 help menu, 2-17 horizontal
display, 2-7 scale, 2-10
Hysteresis softkey, 1-9
I
IEC Publication 61010-1, iiv input signal, 2-4, 2-7 installing a plug-in module, 2-4 instant action key, 1-5 instrument
description, 1-1 settings, 1-5
L
Level softkey, 1-8 line voltage switch, 2-5 list softkeys, 1-9
Index-1
Index
M
manual marker measurements, 2-15 manual softkey, 2-13 Marker key, 2-13 measuring the signal, 2-13 menu
display, 1-6 key, 1-5
softkeys, 1-7 messages on display, 3-2 milli key, 2-9 mode softkey, 2-13 modifying the setup, 2-8
O
offset, 2-16
adjustment, 2-4 Offset softkey, 2-9 optical/electrical plug-in module, 2-3
P
plug-in module, 2-8
installing, 2-4 Position softkey, 1-8, 2-11 power cord, 2-5 pulse
frequency, 2-6
generator, 2-3, 2-6
level, 2-6
setup modification, 2-8 Shift key, 1-9 signal
connection, 2-6 markers, 2-8
measurement, 2-13 single choice softkeys, 1-10 softkey overview, 1-6 square wave signal, 2-6 Sweep softkey, 1-6, 3-4 System configuration softkey, 3-5
T
Time base key, 1-5 time base scale, 1-9 toggle softkeys, 1-6 trigger
display, 2-7
level, 2-6
signal, 2-6 Trigger key, 1-51-6, 3-4 triggered softkey, 3-4 troubleshooting, 3-23-5 turning on the instrument, 2-5
U
Undo Autoscale key, 2-7 unit of measure, 2-9 using the front panel, 1-4, 1-10 Utility key, 1-4, 3-5
R
rear panel
connector, 2-5
line switch, 2-5 Reference softkey, 2-12 repetitive signals, 2-7 Run key, 1-5, 3-4
S
safety, iiv
laser classification, iiv scale increments, 2-8 Scale softkey, 1-8, 2-8, 2-10 Setup key, 2-6
Index-2
V
V amptd key, 2-15 vertical
display, 2-7, 2-9 scale, 2-16
voltage selection, 2-5
W
width measurement, 2-14
X
X1, Y1 Source softkey, 2-13
X1/X2 Position softkeys, 2-14 X2, Y2 Source softkey, 2-13
Y
Y1/Y2 Position softkeys, 2-14
Index
Index-3
Loading...