Agilent Technologies 7200 Troubleshooting And Maintenance Manual

Agilent 7200 Accurate-Mass/Q-TOF GC/MS System
Troubleshooting and Maintenance Manual
Agilent Technologies
Notices
CAUTION
WARNING
No p art o f this manu al may be re produce d in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G3850-90009
Edition
First Edition, August 2014
Printed in USA
Agilent Technologies, Inc. 5301 Stevens Creek Boulevard Santa Clara, CA 95051
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Troubleshooting and Maintenance Manual

Contents

1 Introduction
Abbreviations Used 8
The 7200 Accurate-Mass Quadrupole Time-of-Flight GC/MS
System 10
7200 Accurate-Mass Q-TOF GC/MS Description 12
Side Panel AC Power Connectors 13
Back Panel Connectors 14
Interfacing Start Events to External Devices 15
Remote control processor 15 Remote start signals 15 System ready 15 Start run input 16
Important Safety Warnings 17
Safety and Regulatory Certifications 20
Intended Use 23
Cleaning/Recycling the Product 23
Moving or Storing the MS 23
2 General Troubleshooting
Troubleshooting Tips and Tricks 26
General Symptoms 27
Chromatographic Symptoms 29
Mass Spectra General Symptoms 34
Pressure Symptoms 36
Temperature Symptoms 38
Troubleshooting and Maintenance Manual 3
Common Types of Errors 40
Air Leaks 45
Contamination 46
3 CI Troubleshooting
Common CI-Specific Problems 50
Troubleshooting Tips and Tricks 51
Air Leaks 52
Pressure-Related Symptoms 55
Signal-Related Symptoms 58
Tuning-Related Symptoms 64
The CI ion source is dirty 65 Air leak 65
4 General Maintenance
Before Starting 68
Scheduled maintenance 68 Tools, spare parts, and supplies 69 High voltage precautions 69 Dangerous temperatures 69 Chemical residue 70 Electrostatic discharge 71
To Refill the EI Calibration Vial 72
Materials needed 72 Refill 72
To Refill the IRM Vial 74
Materials needed 74 Procedure 74
To Connect the GC Nitrogen Gas Source to the Collision Cell 76
4 Troubleshooting and Maintenance Manual
5 CI Maintenance
Materials needed 76 Procedure 76
To Replace the Seals in the RIS Probe 77
Materials needed 77 Procedure 77
To Separate the GC from the MS 79
Materials needed 79 Procedure 79
To Position the GC Next to the MS 82
Procedure 82
To Move or Store the MS 83
Materials needed 83 Procedure 83
To Access the Left Side Lifting Handle 86
Materials needed 86 Procedure 87
To Minimize Foreline Pump Damage from Ammonia 90
To Replace the Methane/Isobutane Gas Purifier 91
To Clean the Reagent Gas Supply Lines 92
To Refill the CI Calibration Vial 93
Materials needed 93 Refill 93
6Vacuum System
Overview 96
Maintaining the Vacuum System 97
Periodic maintenance 97 Other procedures 97
Troubleshooting and Maintenance Manual 5
Vacuum System Components 97
Common Vacuum System Problems 98
Foreline Pump 99
To check the oil mist filter 99 To check the foreline pump fluid level 100 To add foreline pump fluid 100 To replace the foreline pump fluid 101
Side Plate 104
Vacuum Seals 104
Calibration Valves 105
EI Calibration Valve 105 CI Calibration Valve 106 IRM Calibration Valves 107
7Replacement Parts
To Order Parts 110
Electronics 111
Vacuum System 116
Analyzer 121
RIS Manifold 130
GC/MS Interface 132
Consumables and Maintenance Supplies 133
6 Troubleshooting and Maintenance Manual
Agilent 7200 Accurate-Mass Q-TOF GC/MS System Troubleshooting and Maintenance Manual
1 Introduction
Abbreviations Used 8 The 7200 Accurate-Mass Quadrupole Time-of-Flight GC/MS System 10 7200 Accurate-Mass Q-TOF GC/MS Description 12 Side Panel AC Power Connectors 13 Interfacing Start Events to External Devices 15 Important Safety Warnings 17 Safety and Regulatory Certifications 20 Intended Use 23 Cleaning/Recycling the Product 23 Moving or Storing the MS 23
This section provides general information about the 7200 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) GC/MS System, including a hardware description and general safety warnings.
Agilent Technologies
7
1 Introduction

Abbreviations Used

The abbreviations in Table 1 are used in discussing this product. They are collected here for convenience.
Tab l e 1 Abbreviations
Abbreviation Definition
AC Alternating current
ALS Automatic liquid sampler
BFB Bromofluorobenzene (calibrant)
CC Collision cell
CI Chemical ionization
DC Direct current
DFTPP Decafluorotriphenylphosphine (calibrant)
DIP Direct insertion probe
EI Electron impact
EPC Electronic pneumatic control
eV Electron volt
GC Gas chromatograph
id Inside diameter
IRM Internal Reference Mass
LAN Local Area Network
m/z Mass to charge ratio
MFC Mass flow controller
MS Mass spectrometer
MS1 Front quadrupole
NCI Negative chemical ionization
OFN Octafluoronaphthalene (sample)
8 Troubleshooting and Maintenance Manual
Tab l e 1 Abbreviations (continued)
Abbreviation Definition
PCI Positive chemical ionization
PFDTD Perfluoro-5,8-dimethyl-3,6,9-trioxydodecane (calibrant)
PFET 2,4,6-tris (Pentafluoroethyl)-1,3,5-triazine
PFTBA Perfluorotributylamine (calibrant)
Q-TOF Quadrupole time-of-flight
Quad Quadrupole mass filter
RF Radio frequency
RFPA Radio frequency power amplifier
TOF Time-of-flight
Torr Unit of pressure, 1 mm Hg
Turbo Turbomolecular vacuum pump
Introduction 1
Troubleshooting and Maintenance Manual 9
1 Introduction

The 7200 Accurate-Mass Quadrupole Time-of-Flight GC/MS System

The 7200 Accurate-Mass Quadrupole Time-of-Flight (Q-TOF) GC/MS System is a standalone capillary GC detector for use with the Agilent 7890 Series gas chromatograph. The 7200 Q-TOF features:
Three turbomolecular vacuum pumps
Rotary vane foreline pump
Independently MS-heated EI or CI ion source
Removable ion source (RIS) probe with bayonet and cooling chamber, which
allows quick change from EI to CI source with minimal loss of vacuum in the instrument
Independently MS-heated hyperbolic quadrupole mass filter, which can be heated to high temperatures, minimizing the contamination typical with low temperature analyses
Single hexapole collision cell
Vacuum-insulated f light tube with dual-stage ion mirror
Fast electronics, allowing fast sampling rates
Analog to digital detector
Independently GC-heated GC/MS interface with automatic retraction
during source removal
Physical description
The 7200 Q-TOF GC/MS is approximately 48 cm high, 71 cm wide, and 89 cm deep. The flight tube extends 84 cm up over the top of the instrument. The RIS probe handle, when attached, extends 48 cm from the front of the instrument.
The weight of the instrument is 152 kg for the turbo pump mainframe. The attached foreline (roughing) pump weighs an additional 22.2 kg.
The basic components of the instrument are: the frame/cover assemblies, the vacuum system, the GC/MS interface, the removable ion source, the f light tube electronics, the collision cell, the detector, and the analyzer.
10 Troubleshooting and Maintenance Manual
Vacuum gauge
The 7200 Q-TOF GC/MS is equipped with four ion vacuum gauges:
RIS vacuum chamber
Vacuum manifold chamber
TOF vacuum manifold chamber
Turbomolecular vacuum pumps exhaust
The MassHunter Workstation can be used to read the pressure (high vacuum) in the vacuum manifold, at the turbomolecular vacuum pump discharge, and the flight tube.
Ionization modes
The G3851BA 7200 Accurate-Mass Q-TOF GC/MS comes standard with both an EI and CI removable ion source (RIS).
A methane/isobutane gas purifier is provided and is required. It removes oxygen, water, hydrocarbons, and sulfur compounds.
The MS CI system has been optimized to achieve the relatively high source pressure required for CI while still maintaining high vacuum in the collision cell, quadrupole, and TOF tube. Special seals along the flow path of the reagent gas and very small openings in the ion source keep the source gases in the ionization volume long enough for the appropriate reactions to occur.
Introduction 1
The interface has special plumbing for reagent gas. A retractable insulating seal fits onto the tip of the interface and is used for both EI and CI.
Switching back and forth between CI and EI sources takes less than 30 minutes with the new removable ion source. The RIS allows the instrument to remain close to pressure, and provides a cooling chamber with N2 purge for rapid source cooling without venting the machine. This saves hours in cycle time over the traditional unit.
Troubleshooting and Maintenance Manual 11
1 Introduction
7200 Series MS
MS power switch
7890 GC
GC power switch

7200 Accurate-Mass Q-TOF GC/MS Description

Figure 1 is an overview of a typical 7200 Accurate-Mass Q-TOF GC/MS system.
Figure 1 7200 Q-TOF GC/MS System
12 Troubleshooting and Maintenance Manual

Side Panel AC Power Connectors

Introduction 1
Figure 2 Side panel power supply (left) and back panel connections (right)
Foreline pump power receptacle (top)
The foreline pump power cord receptacle located on the left side of the MS provides AC power for the foreline pump. If the power switch is off, no power is supplied to the foreline pump.
Main power cord receptacle (bottom)
The AC power cord located on the left side of the MS brings in all electrical power for the MS. The power cord can be detached from the MS.
Troubleshooting and Maintenance Manual 13
1 Introduction

Back Panel Connectors

Figure 3 Side panel power supply (left) and back panel connections (right)
Remote start connector
The remote start connector is the external connector for the remote start circuitry on the LAN/MS control card. It receives remote start signals from the GC.
LAN (I/O) connector
The LAN cable from the data system is connected to the LAN communications connector. It carries all data communication between the PC and the MS.
14 Troubleshooting and Maintenance Manual

Interfacing Start Events to External Devices

Start Ground
Ready

Remote control processor

The remote control processor on the LAN/MS control card synchronizes start-run signals with GCs and other devices. The functions of the remote control processor are extended to the remote start (Remote) connector (Figure 4) on the back panel of the MS. The remote start cable connects the GC and the MS. An optional cable can extend these events to another instrument.

Remote start signals

It is often necessary to communicate with external devices (for example, a purge-and-trap) during a run. Typically, these communications are requests to send a system-ready signal. They also include:
Receive a start run signal from an external device
Program the timing of events during a run
Introduction 1
Figure 4 Remote start connector

System ready

When interfacing to an external device, it is often desirable to send a system-ready signal to the device. In the case of a multi-sample Tekmar purge-and-trap, each sample is purged onto a trap where it waits for a ready signal. On receipt of the ready signal, the desorption cycle begins. When a specific temperature is reached, the purge-and-trap closes a contact to indicate the run has started.
Troubleshooting and Maintenance Manual 15
1 Introduction

Start run input

The ready pin on the remote start connector on the GC is held low at all times except when the GC, MS, and data system are all ready. On system ready, a logic high of 5 VDC is present between that pin and any ground. This same high can be detected between the ready and ground pins on the remote start connector on the MS.
The best way to generate a start run signal is to use the remote start connector on the GC. Since remote start cables are made for most common devices, this is often the simplest way. A general-purpose remote start cable (Y-Remote Start/Stop, NON APG p/n G1530-61200), is also available that terminates in spade lugs. Care must be taken to ensure that the system is actually ready before the start run signal is sent.
If necessary, the remote start connector on the back of the MS can be used to send the start run signal. A contact closure between the start and ground pins will start the run if the system is ready.
16 Troubleshooting and Maintenance Manual

Important Safety Warnings

WARNING
WARNING
There are several important safety notices to always keep in mind when using the MS.
Many internal parts of the MS carry dangerous voltages
If the MS is connected to a power source, even if the power switch is off, potentially dangerous voltages exist on:
The wiring between the MS power cord and the AC power supply
The AC power supply itself
The wiring from the AC power supply to the power switch
With the power switch on, potentially dangerous voltages also exist on:
All electronics boards in the instrument
The internal wires and cables connected to these boards
The wires for any heater (oven, detector, inlet, or valve box)
Introduction 1
Troubleshooting and Maintenance Manual 17
All these parts are shielded by covers. With the covers in place, it should be difficult to accidentally make contact with dangerous voltages. Unless specifically instructed to, never remove a cover unless the detector, inlet, and oven are turned off.
If the power cord insulation is frayed or worn, the cord must be replaced. Contact your Agilent service representative.
Electrostatic discharge is a threat to MS electronics
The printed circuit boards in the MS can be damaged by electrostatic discharge. Do not touch any of the boards unless it is absolutely necessary. If you must handle them, wear a grounded wrist strap and take other antistatic precautions.
1 Introduction
WARNING
WARNING
WARNING
Precautions to take to prevent an explosion
The use of hydrogen gas is specifically prohibited with this product.
You MUST make sure the top thumbscrew on the front analyzer side plate and the top thumbscrew on the rear analyzer side plate are both fastened finger-tight. Do not overtighten the thumbscrews; this can cause air leaks.
You MUST leave the collision cell chamber top plate shipping brackets fastened. Do not remove the shipping brackets from the top plate for normal operation; they secure the top plate in the event of an explosion.
Failure to secure your MS as described above greatly increases the chance of personal injury in the event of an explosion.
18 Troubleshooting and Maintenance Manual
Many parts are dangerously hot
Many parts of the GC/MS operate at temperatures high enough to cause serious burns. These parts include, but are not limited to the:
Inlet
Oven and its contents
Val ve box
Column nuts attaching the column to an inlet or detector
Foreline pump
GC/MS transfer line
Always cool these areas of the system to room temperature before working on them. They will cool faster if you first set the temperature of the heated zone to room temperature. Turn the zone off after it has reached the setpoint. If you must perform maintenance on hot parts, use a wrench and wear gloves. Whenever possible, cool the part of the instrument that you will be maintaining before you begin working on it.
WARNING
Be careful when working behind the instrument. During cool-down cycles, the GC
WARNING
WARNING
emits hot exhaust that can cause burns.
The insulation around the inlets, detectors, valve box, and the insulation cups is made of refractory ceramic fibers. To avoid inhaling fiber particles, we recommend the following safety procedures: ventilate your work area; wear long sleeves, gloves, safety glasses, and a disposable dust/mist respirator; dispose of insulation in a sealed plastic bag in accordance with local regulations; wash your hands with mild soap and cold water after handling the insulation.
The oil pan under the standard foreline pump can be a fire hazard
Oily rags, paper towels, and similar absorbents in the oil pan could ignite and damage the pump and other parts of the MS.
Introduction 1
Combustible materials (or flammable/nonflammable wicking material) placed under, over, or around the foreline (roughing) pump constitutes a fire hazard. Keep the pan clean, but do not leave absorbent material such as paper towels in it.
Troubleshooting and Maintenance Manual 19
1 Introduction

Safety and Regulatory Certifications

The 7200 Q-TOF GC/MS conforms to the following safety standards:
Canadian Standards Association (CSA): CAN/CSA-C22.2 No. 61010-1-04
CSA/Nationally Recognized Test Laboratory (NRTL): UL 61010–1
International Electrotechnical Commission (IEC): 61010–1
EuroNorm (EN): 61010–1
The 7200 Q-TOF GC/MS conforms to the following regulations on Electromagnetic Compatibility (EMC) and Radio Frequency Interference (RFI):
CISPR 11/EN 55011: Group 1, Class A
IEC/EN 61326-1
AUS/NZ
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB—001 du Canada.
The 7200 Q-TOF GC/MS is designed and manufactured under a quality system registered to ISO 9001.
Information
The Agilent Technologies 7200 Accurate-Mass Q-TOF GC/MS meets the following IEC (International Electrotechnical Commission) classifications: Equipment Class I, Laboratory Equipment, Installation Category II, and Pollution Degree 2.
This unit has been designed and tested in accordance with recognized safety standards and is designed for use indoors. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the MS has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation.
20 Troubleshooting and Maintenance Manual
Symbols
Introduction 1
Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard.
Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
See accompanying instructions for more information.
Indicates a hot surface.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
Indicates potential explosion hazard.
Indicates radioactivity hazard.
Indicates electrostatic discharge hazard.
Indicates that you must not discard this electrical/electronic product in domestic household waste.
Troubleshooting and Maintenance Manual 21
1 Introduction
Electromagnetic compatibility
This device complies with the requirements of CISPR 11. Operation is subject to the following two conditions:
This device may not cause harmful interference.
This device must accept any interference received, including interference
that may cause undesired operation.
If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try one or more of the following measures:
1 Relocate the radio or antenna.
2 Move the device away from the radio or television.
3 Plug the device into a different electrical outlet, so that the device and the
radio or television are on separate electrical circuits.
4 Make sure that all peripheral devices are also certified.
5 Make sure that appropriate cables are used to connect the device to
peripheral equipment.
6 Consult your equipment dealer, Agilent Technologies, or an experienced
technician for assistance.
Changes or modifications not expressly approved by Agilent Technologies could void the user’s authority to operate the equipment.
Sound emission declaration
Sound pressure
Sound pressure Lp < 70 dB according to EN 27779:1991 and EN ISO 3744:1995.
Schalldruckpegel
Schalldruckpegel LP < 70 dB nach EN 27779:1991 und EN ISO 3744:1995.
22 Troubleshooting and Maintenance Manual

Intended Use

Agilent products must only be used in the manner described in the Agilent product user guides. Any other use may result in damage to the product or personal injury. Agilent is not responsible for any damages caused, in whole or in part, by improper use of the products, unauthorized alterations, adjustments or modifications to the products, failure to comply with procedures in Agilent product user guides, or use of the products in violation of applicable laws, rules or regulations.

Cleaning/Recycling the Product

To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth. For recycling, contact your local Agilent sales office.

Moving or Storing the MS

Introduction 1
The best way to keep your MS functioning properly is to keep it pumped down and hot, with carrier gas flow. If you plan to move or store your MS, a few additional precautions are required. The MS must remain upright at all times; this requires special caution when moving. The MS should not be left vented to atmosphere for long periods. For more information, see “To Move or Store the
MS” on page 83.
Troubleshooting and Maintenance Manual 23
1 Introduction
24 Troubleshooting and Maintenance Manual
Agilent 7200 Accurate-Mass Q-TOF GC/MS System Troubleshooting and Maintenance Manual
2 General Troubleshooting
Troubleshooting Tips and Tricks 26 General Symptoms 27 Chromatographic Symptoms 29 Mass Spectra General Symptoms 34 Pressure Symptoms 36 Temperature Symptoms 38 Common Types of Errors 40 Air Leaks 45 Contamination 46
This is a quick reference to symptoms and possible causes of the most common problems experienced by users. For each symptom, one or more possible causes are listed. In general, the causes listed first are the most likely causes or the easiest to check and correct.
This chapter does not include corrective actions for the possible causes listed. Some of the corrective actions required may be dangerous if performed incorrectly. Do not attempt any corrective actions unless you are sure you know the correct procedure and the dangers involved. See the other chapters in this manual for more information.
If the material in this chapter and in the online help proves insufficient to help you diagnose a problem, contact your Agilent Technologies service representative.
Agilent Technologies
25
2 General Troubleshooting

Troubleshooting Tips and Tricks

Rule 1: “Look for what has been changed.”
Many problems are introduced accidentally by human actions. Every time any system is disturbed, there is a chance of introducing a new problem.
If the MS was just pumped down after maintenance, suspect air leaks or incorrect assembly.
If carrier gas or helium gas purifier was just changed, suspect leaks or contaminated or incorrect gas.
If the GC column was just replaced, suspect air leaks or a contaminated or bleeding column.
Rule 2: “If complex isn’t working, go back to simple.”
A complex task is not only more difficult to perform but also more difficult to troubleshoot. If you’re having trouble detecting your sample, verify that autotune is successful.
Rule 3: “Divide and conquer.”
This technique is known as “half-split” troubleshooting. If you can isolate the problem to only part of the system, it is much easier to locate.
To determine whether an air leak is in the GC or the MS, you can vent the MS, remove the column, and install the blank interface ferrule. If the leak goes away, it was in the GC.
26 Troubleshooting and Maintenance Manual

General Symptoms

This section describes symptoms you might observe when first turning on the GC/MS system. All of these symptoms would prevent operation of the system.
GC does not turn on
Nothing happens when the GC is switched on. The GC fans do not turn on and the keypad display does not light.
Disconnected GC power cord
No voltage or incorrect voltage at the electrical outlet
Failed fuse in the GC
GC power supply is not working correctly
MS does not turn on
Nothing happens when the MS is switched on. The foreline pump does not start. The cooling fan for the high-vacuum pump does not turn on.
Disconnected MS power cord
No voltage or incorrect voltage at the electrical outlet
Failed primary fuses - Not user replaceable
MS electronics are not working correctly
General Troubleshooting 2
Foreline pump is not operating
The MS is receiving power (the fan is operating) but the foreline pump is not operating.
A large air leak (usually the analyzer door open) has caused pumpdown failure. You must power cycle the MS to recover from this state.
Disconnected foreline pump power cord
Malfunctioning foreline pump
Check power switch on foreline pump
Troubleshooting and Maintenance Manual 27
2 General Troubleshooting
MS turns on but then the foreline pump shuts off
The MS will shut down both the foreline pump and the turbo pumps if the system fails to pump down correctly. This is usually because of a large air leak or the side plate has not sealed correctly. This feature helps prevent the foreline pump from sucking air through the system, which can damage the analyzer and the turbo pumps.
You must power cycle the MS to recover from this state.
28 Troubleshooting and Maintenance Manual

Chromatographic Symptoms

These are symptoms you may observe in the chromatograms generated by data acquisition. In general, these symptoms do not prevent you from operating your GC/MS system. They indicate, however, that the data you are acquiring may not be the best data obtainable. These symptoms can be caused by instrument malfunctions but are more likely caused by incorrect chromatographic technique.
Two of the symptoms, Low sensitivity and Poor repeatability also apply to mass spectral data.
No peaks
If an analysis shows no chromatographic peaks, only a flat baseline or minor noise, run the automated tune program. If the MS passes tune, the problem is most likely related to the GC. If the MS does not pass tune, the problem is most likely in the MS.
Passes tune
General Troubleshooting 2
Incorrect sample concentration
No analytes present
Syringe missing from the ALS or not installed correctly
Injection accidentally made in split mode instead of splitless mode
Empty or almost empty sample vial
Dirty GC inlet
Leaking GC inlet*
Loose column nut at the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Does not pass tune
Calibration vial is empty
Excessive foreline or analyzer chamber pressure
Very d i r t y i o n s o ur c e
Troubleshooting and Maintenance Manual 29
2 General Troubleshooting
Calibration valve is not working correctly
Bad signal cable connection
Filament has failed or is not connected correctly
Bad ion source wiring connection
Bad detector wiring connection
Failed MS detector
Peaks are tailing
Active sites in the sample path
Injection is too large
Incorrect GC inlet temperature
Insufficient column flow
GC/MS interface temperature is too low
Ion source temperature is too low
Peaks are fronting
Column film thickness mismatched with analyte concentration (column overload)
Initial oven temperature is too low
Active sites in the sample path
Injection is too large
GC inlet pressure too high
Insufficient column flow
Peaks have flat tops
Insufficient solvent delay
Incorrect scale on the display
Injection is too large
30 Troubleshooting and Maintenance Manual
Loading...
+ 108 hidden pages