Agilent Technologies 7100 User Manual

User Manual

Agilent 7100 Capillary Electrophoresis System

Agilent Technologies
Notices
CAUTION
WARNING
© Agilent Technologies, Inc. 2009 No p art o f this manu al may be re produce d in
any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G7100-90000
Edition
12/2009 Printed in Germany Agilent Technologies, Deutschland GmbH
Hewlett-Packard-Strasse 8 76337 Waldbronn
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connec­tion with the furnishing, use, or per­formance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a U.S. Government prime contract or subcon­tract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer soft­ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or contract clause. Use, duplication or disclo­sure of Software is subject to Agilent Tech­nologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
Research Use Only
Agilent 7100 Capillary Electrophoresis System
In This Guide…
This guide contains information to install and operate the Agilent 7100 Capillary Electrophoresis System.
1 Introduction
This chapter gives an introduction to the instrument, instrument overview and internal connectors.
2 Site Requirements and Specifications
This chapter provides information on environmental requirements, physical and performance specifications.
3 Installing the Instrument
In this chapter you will find information about how to unpack and to set up the Agilent CE and information about the configuration of Instrument.
4 LAN Configuration
This chapter provides information on connecting the CE instrument to the Agilent ChemStation PC.
5 Using the Agilent 7100 Capillary Electrophoresis System
This chapter provides information on how to set up the CE instrument for an analysis and explains the basic settings.
6 Troubleshooting and Diagnostics
This chapter gives an overview about the troubleshooting and diagnostic features and the different user interfaces.
7 Hardware Information
This chapter describes the instrument in more detail on hardware and electronics.
8 Maintenance
This chapter describes the maintenance of the CE instrument.
Agilent 7100 Capillary Electrophoresis System User Manual 3
9 Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.
10 Appendix
This chapter provides additional information on safety, legal and web content associated with your 7100 CE instrument.
4 Agilent 7100 Capillary Electrophoresis System User Manual

Contents

1 Introduction 13
Introduction to the Agilent 7100 Capillary Electrophoresis System 14
Getting Help 14 Where to Find Information 15 What Learning Products are Available? 16
2 Site Requirements and Specifications 17
Site Requirements 18
Power Consideration 18
Power Cords 19
Bench Space 20
Environment 20 Physical Specifications 21 Performance Specifications 22 Anforderungen an den Aufstellungsort 23
Stromversorgung 23
Netzkabel 24
Platzbedarf 25
Umgebungsbedingungen 25 Abmessungen und Gewicht 26 Leistungsspezifikationen 27 Exigences d’installation 28
Remarques à propos de l’alimentation 28
Câbles d’alimentation 29
Encombrement 30
Environnement 30 Caractéristiques physiques 31 Caractéristiques de performance 32
Agilent 7100 Capillary Electrophoresis System User Manual 5
Contents
Requisiti ambientali 33
Alimentazione 33
Cavi di alimentazione 34
Spazio necessario 35
Ambiente 35 Specifiche fisiche 36 Specifiche delle prestazioni 37 Requisitos de las instalaciones 38
Consideraciones sobre la corriente 38
Cables de alimentación 39
Espacio en el banco 40
Entorno 40 Especificaciones físicas 41 Especificaciones de rendimiento 42
設置について 43
電源について 43
電源コード 44
作業台スペース 45
環境条件 45 物理的仕様 46 性能仕様 47
场地要求 48
电源要求 48
电源线 49
工作台位置 50
环境 50 物理规格 51 性能规格 52
3 Installing the Instrument 53
Unpacking the Agilent CE Instrument 54
Damaged Packaging 54
Unpacking the CE Instrument 54
6 Agilent 7100 Capillary Electrophoresis System User Manual
Delivery Checklist 55 Installing the Instrument 57
Physical Connection 57
4 LAN Configuration 61
What You Have to Do First 62 TCP/IP Parameter Configuration 63 Configuration Switch 64 Initialization Mode Selection 65 Link Configuration Selection 69 Storing the Settings Permanently with Bootp 70 Manual Configuration 71
With Telnet 72 Agilent Bootp Service 75 Location of the Media Access Control (MAC) Address 76
5 Using the Agilent 7100 Capillary Electrophoresis System 77
System Definition 78 Controlling the Agilent CE Instrument 79
Before You Start the Agilent CE Instrument 79
Starting Up the Agilent CE Instrument 79
Starting Your Computer and the Agilent ChemStation Software 81
Starting the Agilent ChemStation Software the First Time 82 Working with the CE Diagram 84
Common Software Toolbar 84
Software Toolbar 85
Graphical User Interface Explanation 86
Menus in the CE and DAD Diagram 88
System Vialtable 90 How to Prepare your Agilent CE Instrument for Analysis 91 Capillary Electrophoresis, CE 92
How to Use Your Agilent CE Instrument for Capillary Electrophoresis
Separations 92
Contents
Agilent 7100 Capillary Electrophoresis System User Manual 7
Contents
Agilent CE Method Parameters 92
Vials 96
Cassette Temperature 97
High Voltage System 98
Posttime and Stoptime 100
Replenishment 100
Preconditioning 101
Injection 103
Timetable 105
Detector Parameters 107
Switching the Lamp On 110
Preparing the Replenishment System if Needed 111
Preparing Vials for Buffer and Sample 115
Installing a Capillary in the Alignment Interface 124
Installing a Capillary in an Empty Capillary Cassette 128
Inserting the Cassette 132
Fraction Collection 135 Capillary Electrophoresis, Plus High Pressure CE+p 136
Requirements 136
Outlet Lift Connection to the Tubing System 138
Preparing the Agilent CE Instrument for High Pressure Use 140
Operation and Function in CE+p Mode 143
Injection Options with CE+p 146
Timetable Options with CE+p 146
CE+p Control Using the CE Diagram 147 Capillary Electrochromatography, CEC 148
How to Use Your Agilent CE Instrument for Capillary
Electrochromatography 148 Preparing the Agilent CE Instrument for High Pressure Use 150 CEC-Specific GUI Explanation 153 Injection Options with CEC 156 Timetable Options with CEC 157 CEC Control Using the CE Diagram 158
8 Agilent 7100 Capillary Electrophoresis System User Manual
Running the CEC Analysis 159
6 Troubleshooting and Diagnostics 161
CE Problem Solving 162
Power On Fail 162 Leaks 163 Blocked Replenishment Needle 165 Leakage Current 165 Broken Capillary 166 Contaminated Insulation Plate 166 Empty Injection Vial 167 Dirt in the Vials 167 Air Bubble in Injection Vial or Run Buffer Vials 167 Air Bubble in Capillary 167 Problems with the Replenishment System and Buffers
Containing SDS 168
Agilent Lab Advisor Software 169
Configuration 169
Contents
7 Hardware Information 171
Identifying Instrument Components 172 7100 Capillary Electrophoresis Instrument Electronics 173 Firmware 176
Resident System 176 Main System 176 Firmware Updates 177
Diode Array Detector 179
DAD Control 180 Digital to Analog Converter 181
Changing the Fuses for the Power Supplies 182 Installing Drainage Tubing 183 External Water Bath for Tray Cooling (Optional) 184 Analog to Digital Converter 186
Agilent 7100 Capillary Electrophoresis System User Manual 9
Contents
8 Maintenance 189
Standard Operating Procedures for Maintenance of Your
Agilent CE Instrument 190
Overview of Maintenance 191 Early Maintenance Feedback (EMF) 194
EMF Counters 194 Using the EMF Counters 194 Setting the EMF Limits 195
Cleaning the Electrodes, Pre-punchers and Insulation Plate 196
Prepare the Agilent CE Instrument 197 Remove the Covers 198 Accessing the Electrodes 199 Accessing the Liquid Handling Module 201 Accessing and Removing the Pre-punchers 203 Cleaning the Electrodes 205 Cleaning the Pre-punchers 206 Cleaning the Insulation Plate 207 Reinstalling the Pre-punchers 208 Reinstalling the Electrodes and the Insulation Plate 210 Reinstall the Capillary Cassette 215 Reinstall the Front Cover 215
Maintenance of the Replenishment System 216
Changing the Buffer Composition in the Replenishment System 217 The Replenishment System Will Not Be Used for Some Time 218 Further Action 219 Operations for Replenishment Maintenance 219
Changing the Air Inlet Filter 222
Accessing the Air Filter 222
Changing the Lamp 224
Removing the Existing Lamp 225 Inserting a New Lamp 227 Further Action 227
Cleaning the Instrument 228
10 Agilent 7100 Capillary Electrophoresis System User Manual
9 Parts and Materials for Maintenance 229
Overview of Maintenance Parts 230
10 Appendix 231
General Safety Information 232
Operation 232
Allgemeine Sicherheitsinformation 235 Informations générales de sécurité 238 Informazioni generali sulla sicurezza 241 Información de seguridad 244
安全に関する一般的な情報 247
一般安全信息 250
Setting Up a Test Sample Run 252
Method Parameter CE (Default Settings) 253 Method Parameter DAD 255
Chemical and Biological Safety 256 The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) 257
Radio Interference 258 Sound Emission 259 Legal Notice 260 Agilent Technologies on Internet 261
Contents
Index 263
Agilent 7100 Capillary Electrophoresis System User Manual 11
Contents
12 Agilent 7100 Capillary Electrophoresis System User Manual
Agilent 7100 Capillary Electrophoresis System User Manual
1 Introduction
Introduction to the Agilent 7100 Capillary Electrophoresis System 14 Where to Find Information 15 What Learning Products are Available? 16
This chapter gives an introduction to the Agilent 7100 Capillary Electrophoresis System and where to find specific information or help.
Agilent Technologies
13
1 Introduction

Introduction to the Agilent 7100 Capillary Electrophoresis System

Introduction to the Agilent 7100 Capillary Electrophoresis System
This handbook gives an overview of your Agilent 7100 Capillary Electrophoresis System. It is designed to get you started working with the system. We strongly advise you to make extensive use of the online help when working with the system. There you find detailed reference and task information which complements the overview given in this guide. Refer to
“Getting Help” later in this section for information on using the online help of
the ChemStation. Refer to “Where to Find Information” for help in finding information on certain tasks.
Getting Help
You can get help by choosing the Help item from the top menu. This puts you into the help index. From there you can browse through the information by selecting the appropriate jumps (underlined words). To access specific information in the help:
When you are in the Agilent ChemStation or Lab Advisor software:
select a context-sensitive element and press F1.
When you are in the online help:
search for a specific keyword using the search command,
select a jump - a jump is an underlined word you can select to go to a topic
related to the word, or
select a word or phrase with a dotted underline to see its definition. The online help also contains online information on how to use help.
14 Agilent 7100 Capillary Electrophoresis System User Manual

Where to Find Information

Tabl e 1 Where to find information
Task Refer to
Introduction 1
Where to Find Information
Installation of the Agilent CE System,
installing the external water bath, installing the high sensitivity cell
Setting up the Agilent CE System for
CE analysis, CE+p
How to use the Agilent CE System in
the CE+p mode
How to use the Agilent CE System for
capillary electrochromatography
How to maintain the Agilent CE
System
Setting up a sequence Online help "Understanding Your ChemStation"
Data analysis Online help "Understanding Your ChemStation"
Setting up a report Online help "Understanding Your ChemStation"
Spectral library search “ChemStation Applications” handbook and online help
Setting up customized reports “ChemStation Applications” handbook
Safety information Safety information in “Appendix” on page 231
Learning ChemStation concepts “Understanding Your ChemStation” handbook
Learning the theory of the technique High Performance Capillary Electrophoresis: An
"The Core Agilent CE Instrument" “Installing the
Instrument” on page 53
Chapter “Using the Agilent 7100 Capillary
Electrophoresis System” on page 77"
“Capillary Electrophoresis, Plus High Pressure
CE+p” on page 136
“Capillary Electrochromatography, CEC” on page 148
Chapter “Maintenance” on page 189
Introduction” The CE Partner CD-ROM: “The comprehensive, interactive tool for beginners and advanced users of CE”
Application Agilent web page (http://www.agilent.com), search for
Application notes
Agilent 7100 Capillary Electrophoresis System User Manual 15
1 Introduction

What Learning Products are Available?

What Learning Products are Available?
The Agilent CE System is supplied with the following:
User's Guide
High Performance Capillary Electrophoresis: An Introduction
The Agilent ChemStation and Lab Advisor/Instrument Utilities software DVD is supplied with the following electronic documents:
Installing Your Agilent ChemStation for LC and CE Systems
Installing and Understanding Your Spectra Module
Understanding Your Agilent ChemStation
Online help
Understanding Your Agilent ChemStation for CE Systems
16 Agilent 7100 Capillary Electrophoresis System User Manual
Agilent 7100 Capillary Electrophoresis System User Manual
2 Site Requirements and Specifications
Site Requirements 18 Physical Specifications 21 Performance Specifications 22 Anforderungen an den Aufstellungsort 23 Abmessungen und Gewicht 26 Leistungsspezifikationen 27 Exigences d’installation 28 Caractéristiques physiques 31 Caractéristiques de performance 32 Requisiti ambientali 33 Specifiche fisiche 36 Specifiche delle prestazioni 37 Requisitos de las instalaciones 38 Especificaciones físicas 41 Especificaciones de rendimiento 42
設置について 43 物理的仕様 46 性能仕様 47
场地要求 48 物理规格 51 性能规格 52
This chapter provides information on the environmental requirements, as well as physical and performance specifications.
Agilent Technologies
17
2 Site Requirements and Specifications
WARNING
WARNING

Site Requirements

Site Requirements
A suitable environment is important to ensure optimal performance of the 7100 Capillary Electrophoresis instrument.
Before you begin installation, check carefully that the place you have chosen meets the requirements below.
Power Consideration
The instrument power supply has wide ranging capabilities and accepts any standard line voltage in the range mentioned in Table 2 on page 21. Consequently, there is no voltage selector on the back of the detector.
The module is partially energized when switched off, as long as the power cord is plugged in.
18 Agilent 7100 Capillary Electrophoresis System User Manual
Repair or maintenance work on the module can lead to personal injuries, for example a shock hazard, when the cover is opened and the module is connected to the power.
Make sure that it is always possible to access the power plug.
Remove the power cable from the instrument before opening the cover.
Do not connect the power cable to the instrument while the electronic box in the
back of the instrument is opened or side panels are removed.
Incorrect line voltage on the instrument Shock hazard or damage to your instrument can result if the device is connected to a
line voltage that is higher than specified.
Connect your instrument to the specified line voltage.
CAUTION
Inaccessible power plug.
WARNING
WARNING
In the case of an emergency it must be possible to disconnect the instrument from the power line at any time.
Make sure the power connector of the instrument can be easily reached and unplugged.
Provide sufficient space behind the power socket of the instrument to unplug the cable.
Power Cords
Different power cords are delivered with the instrument. The female end of all power cords is identical. It plugs into the power-input socket at the rear. The male end of each power cord is different and designed to match the wall socket of a particular country or region.
Absence of a ground connection or use of an unspecified power cord
Site Requirements and Specifications 2
Site Requirements
The absence of a ground connection or the use of an unspecified power cord can lead to an electric shock or short circuit.
Never operate your instrument from a power outlet without a ground connection.
Never use a power cord other than the Agilent Technologies power cord designed
for your region.
Use of unsupplied cables Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Agilent 7100 Capillary Electrophoresis System User Manual 19
2 Site Requirements and Specifications
CAUTION
Site Requirements
Bench Space
The instrument dimensions and weight mean that the instrument can be placed on almost any desk or laboratory bench. It needs an additional 2.5 cm (1.0 inch) of space on either side and approximately 8 cm (3.1 inches) in the rear for air circulation and electric connections.
If the Agilent 7100 Capillary Electrophoresis instrument is to be placed on a bench, make sure that the bench is designed to bear the weight of the system.
The instrument must be operated in a upright position.
Environment
Your instrument will work within the specifications for ambient temperatures and relative humidity described in Table 2 on page 21.
Condensation within the module Condensation will damage the system electronics.
20 Agilent 7100 Capillary Electrophoresis System User Manual
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
If your module was shipped in cold weather, leave it in its box and allow it to warm slowly to room temperature to avoid condensation.

Physical Specifications

NOTE
Tabl e 2 Physical specifications
Type Specification Comments
Weight 35 kg (77.2 lbs)
Site Requirements and Specifications 2
Physical Specifications
Dimensions (width × depth × height)
Line voltage 100–240 VAC ± 10% Wide-ranging capability
Line frequency 50 or 60 Hz ± 5%
Power consumption 350 VA / 300 W / 1024 BTU/h Maximum
Ambient operating temperature
Ambient non-operating temperature
Humidity below 80% at 31 °C (87.8 °F) Non-condensing
Operating altitude Up to 2000 m (6500 ft)
External cooling max. 0.5 bar (7.2 psi),
External pressure 2-12 bar (29-174 psi) Oil-free air or nitrogen
Safety standards: IEC, CSA, UL
Housing
350 x 510 x 590 mm (13.8 x 20.1 x 23.2 inches)
5–40 °C (41–104 °F)
-40 to 70°C (-40 to 158 °F)
max. 50 °C (122 °F)
Installation category II, Pollution degree 2
All materials are recyclable
Waterbath
For indoor use only
For applications run in CEC mode (Capillary Electrochromatography) or CE+p mode (Capillary Electrophoresis mode with optional usage of higher pressure), an external pressure supply of oil-free air or nitrogen with a maximum pressure of 15 bar (218 psi) can be used. A respective male adapter that fits into the instrument’s female adapter and PTFE tubing are part of the shipment.
The Agilent 7100 Capillary Electrophoresis System is designed to operate in a typical electromagnetic environment where RF transmitters, such as mobile phones, should not be used in close proximity.
Agilent 7100 Capillary Electrophoresis System User Manual 21
2 Site Requirements and Specifications

Performance Specifications

Performance Specifications
Tabl e 3 Performance specifications for the Agilent 7100 Capillary Electrophoresis System
Type Specification Comments
Safety features Extensive diagnostics, error detection and display,
current leak detection; low current limit Liquid leak sensor; safety sensors on door and cover disabling high voltage; vial sensor;
Control and data evaluation Agilent ChemStation under Windows® XP (SP3)
Communication LAN, controller-area network (CAN), RS-232C,
APG remote: ready, start, stop and shut-down signals. USB (future use). Analog out and in.
GLP features Early maintenance feedback (EMF), electronic
records of maintenance and errors
0 to1 V
22 Agilent 7100 Capillary Electrophoresis System User Manual

Anforderungen an den Aufstellungsort

WARNUNG
WARNUNG
Eine geeignete Umgebung ist wichtig für die optimale Leistungsfähigkeit des 7100 Kapillarelektrophorese-Geräts.
Prüfen Sie sorgfältig, ob der von Ihnen gewählte Aufstellort die nachfolgenden Anforderungen erfüllt, bevor Sie mit der Installation beginnen.
Stromversorgung
Das Weitbereichsnetzteil des Geräts arbeitet mit Standard-Netzspannungen gemäß den Angaben in Tabelle 4 auf Seite 26. Daher ist an der Rückseite des Detektors kein Spannungswahlschalter vorhanden.
Auch im ausgeschalteten Zustand fließt im Modul Strom, solange das Netzkabel eingesteckt ist.
Site Requirements and Specifications 2
Anforderungen an den Aufstellungsort
Agilent 7100 Capillary Electrophoresis System User Manual 23
Die Durchführung von Reparatur- oder Wartungsarbeiten am Modul kann zu Personenschäden wie z. B. durch Stromschlag führen, wenn das Modulgehäuse geöffnet wird, während das Modul an die Netzspannung angeschlossen ist.
Stellen Sie den freien Zugang zu den Netzkabeln sicher.
Trennen Sie das Netzkabel vom Gerät, bevor Sie das Gehäuse öffnen.
Schließen Sie das Netzkabel nicht an das Gerät an, während der Elektronikkasten
an der Geräterückseite geöffnet oder Seitenteile entfernt sind.
Falsche Netzspannung am Gerät Wenn das Gerät an eine höhere Spannung als spezifiziert angeschlossen wird, kann
dies zu Überspannungsschäden oder einer völligen Zerstörung des Geräts führen.
Schließen Sie das Gerät an die angegebene Netzspannung an.
2 Site Requirements and Specifications
VORSICHT
WARNUNG
WARNUNG
Anforderungen an den Aufstellungsort
Unzugänglicher Netzstecker. In einem Notfall muss es jederzeit möglich sein, das Gerät vom Stromnetz zu trennen.
Stellen Sie sicher, dass der Netzstecker des Geräts leicht zugänglich ist.
Lassen Sie hinter der Netzbuchse des Geräts genügend Platz zum Herausziehen
des Kabels.
Netzkabel
Das Gerät wird mit unterschiedlichen Netzkabeln ausgeliefert. Der weibliche Stecker ist bei jedem Netzkabel identisch. Er wird an die Netzanschlussbuchse an der Rückseite angeschlossen. Die Stecker am anderen Ende der Netzkabel sind unterschiedlich und erfüllen die Normen unterschiedlicher Länder oder Regionen.
Nicht vorhandene Erdung oder Verwendung eines nicht spezifizierten Netzkabels Bei der Verwendung des Geräts ohne Erdung oder mit einem nicht spezifizierten
Netzkabel können Stromschläge und Kurzschlüsse auftreten.
24 Agilent 7100 Capillary Electrophoresis System User Manual
Betreiben Sie Ihr Gerät niemals an einer Spannungsquelle ohne Erdung.
Verwenden Sie niemals ein anderes als das von Agilent zum Einsatz im jeweiligen
Land bereitgestellte Kabel.
Verwendung nicht im Lieferumfang enthaltener Kabel Die Verwendung von Kabeln, die nicht von Agilent Technologies geliefert
wurden, kann zu einer Beschädigung der elektronischen Komponenten oder zu Personenschäden führen.
Verwenden Sie ausschließlich Originalkabel von Agilent Technologies, um eine
einwandfreie Funktion und die Einhaltung der Sicherheits- und EML-Bestimmungen zu gewährleisten.
Platzbedarf
VORSICHT
Die Abmessungen und das Gewicht des Geräts ermöglichen seine Aufstellung auf praktisch jedem Schreibtisch oder Laborarbeitstisch. Das Gerät benötigt an jeder Seite zusätzlich 2,5 cm Platz und ungefähr 8 cm an der Rückseite für die elektrischen Anschlüsse und für eine ausreichende Luftzirkulation.
Wenn das Agilent 7100 Kapillarelektrophorese System auf einem Labortisch aufgestellt werden soll, stellen Sie sicher, dass der Tisch ausreichende Tragfähigkeit für das System aufweist.
Das Gerät ist stehend zu betreiben!
Umgebungsbedingungen
Ihr Gerät arbeitet bei der Umgebungstemperatur und der relativen Luftfeuchtigkeit, die in den Spezifikationen in Tabelle 4 auf Seite 26 angegeben sind.
Kondensation im Inneren des Moduls
Site Requirements and Specifications 2
Anforderungen an den Aufstellungsort
Agilent 7100 Capillary Electrophoresis System User Manual 25
Kondensation führt zur Beschädigung der Systemelektronik.
Vermeiden Sie die Lagerung, den Versand oder den Betrieb der Pumpe unter Bedingungen, die zu einer Kondensation in der Pumpe führen können.
Nach einem Transport bei niedrigen Temperaturen muss das Gerät zur Vermeidung von Kondensation in der Verpackung verbleiben, bis es sich auf Raumtemperatur erwärmt hat.
2 Site Requirements and Specifications
HINWEIS

Abmessungen und Gewicht

Abmessungen und Gewicht
Tabelle 4 Abmessungen und Gewicht
Typ Spezifikationen Kommentare
Gewicht 35 kg
Abmessungen (Breite × Tiefe × Höhe)
Netzspannung 100–240 V Wechselstrom ± 10 % Weitbereichsnetzteil
Netzfrequenz 50 oder 60 Hz ± 5 %
Leistungsaufnahme 350 VA / 300 W / 1024 BTU/h Maximal
zulässige Umgebungstemperatur bei Lagerung
Umgebungstemperatur bei Lagerung
Luftfeuchtigkeit unter 80 % bei 31 °C Nicht kondensierend
Betriebshöhe bis zu 2000 m
350 x 510 x 590 mm (23,2 x 13,8 x 20,1 Zoll)
5–40 °C
-40 bis 70 °C
26 Agilent 7100 Capillary Electrophoresis System User Manual
Externe Kühlung max. 0,5 bar (7,2 psi), max. 50 °C Wasserbad
Externer Druck 2-12 bar (29-174 psi) Ölfreie Luft oder Stickstoff
Sicherheitsstandards: IEC, CSA, UL
Gehäuse
Installationskategorie II, Verschmutzungsgrad 2.
Alle Materialien sind recyclebar.
Nur für den Einsatz im Innenbereich geeignet
Für Applikationen im CEC-Modus (Capillary Electrochromatography) oder CE+p-Modus (Kapillarelektrophorese-Modus mit optionaler Verwendung höheren Druckes) kann eine externe Druckversorgung mit ölfreier Luft oder Stickstoff und einem Druckmaximum von 15 bar (218 psi) verwendet werden. Ein entsprechender männlicher Adapter für den weiblichen Adapter des Geräts und PTFE-Schläuche werden mitgeliefert.
Das Agilent 7100 Kapillarelektrophorese-System ist auf den Betrieb in einer typischen elektromagnetischen Umgebung ausgelegt, in deren unmittelbarer Nähe keine HF-Sender wie z. B. Mobiltelefone verwendet werden dürfen.
Site Requirements and Specifications 2

Leistungsspezifikationen

Leistungsspezifikationen
Tabelle 5 Leistungsspezifikationen für das Agilent 7100 Kapillarelektrophorese-System
Typ Spezifikationen Kommentare
Sicherheitsvorkehrungen Umfangreiche Diagnose, Fehlererkennung und
-anzeige, Kriechstromerkennung; Grenzwert für zu niedrigen Strom Flüssigkeitslecksensor; Sicherheitssensoren an Tür und Abdeckung zur Trennung der Hochspannung; Probenflaschensensor;
Steuerung und Datenauswertung
Kommunikation LAN, CAN (Controller-Area Network), RS-232C,
GLP-Eigenschaften Frühwarnsystem für fällige Wartungen (EMF, Early
Agilent ChemStation unter Windows® XP (SP3)
APG Remote: Ready-, Start-, Stop- und Shut-down-Signale. USB (zukünftige Ver wen dung) . Analoger Ein-/Ausgang.
Maintenance Feedback), elektronische Aufzeichnung von Wartungsarbeiten und Fehlermeldungen
0 bis 1 V
Agilent 7100 Capillary Electrophoresis System User Manual 27
2 Site Requirements and Specifications
AVERTISSEMENT
AVERTISSEMENT

Exigences d’installation

Exigences d’installation
Un environnement adéquat est indispensable pour obtenir des performances optimales de l’instrument d’électrophorèse capillaire 7100.
Avant de commencer l’installation, vérifiez soigneusement que l’emplacement choisi est conforme aux exigences ci-dessous.
Remarques à propos de l’alimentation
Le module d’alimentation de l’instrument présente une tolérance importante. Il accepte ainsi n’importe quelle tension de secteur se situant dans la plage de tolérance précisée dans le Tableau 6, page 31. Ceci explique l’absence de sélecteur de tension à l’arrière du détecteur.
Le module reste partiellement activé lorsqu’il est éteint, tant que le cordon d’alimentation est branché.
Certaines opérations de réparation ou d’entretien sur le module peuvent occasionner des blessures, par exemple l’électrocution, si le capot est ouvert et que le module est branché.
Veillez à ce que la prise d’alimentation électrique reste à tout moment accessible.
Débranchez le câble d’alimentation de l’instrument avant d’ouvrir le capot.
Ne branchez pas le câble d’alimentation sur l’instrument si le boîtier
électronique à l’arrière est ouvert ou si les panneaux latéraux sont enlevés.
Tension de secteur incorrecte au niveau de l’instrument Il existe un risque d’électrocution ou de dommages à l’instrument en cas de
raccordement à une ligne d’alimentation de tension supérieure à celle spécifiée.
Connectez l’instrument à la tension indiquée.
28 Agilent 7100 Capillary Electrophoresis System User Manual
ATTENTION
Prise d’alimentation inaccessible.
AVERTISSEMENT
AVERTISSEMENT
En cas d’urgence, il doit être possible de débrancher à tout moment l’instrument du secteur.
Veillez à faciliter l’accès au connecteur d’alimentation de l’instrument et à permettre de le débrancher facilement.
Maintenez un espace suffisant derrière la prise d’alimentation de l’instrument pour pouvoir débrancher le câble.
Câbles d’alimentation
Différents câbles d’alimentation sont livrés avec l’instrument. L’extrémité femelle est la même pour tous les câbles. Elle s’enfiche dans l’embase d’alimentation située à l’arrière. L’extrémité mâle, qui se branche sur la prise de courant murale, varie selon le pays ou la région.
Absence de raccordement à la terre et utilisation d’un câble d’alimentation non adapté
Site Requirements and Specifications 2
Exigences d’installation
Agilent 7100 Capillary Electrophoresis System User Manual 29
L’absence de raccordement à la terre et l’utilisation d’un câble d’alimentation non adapté peuvent entraîner une électrocution ou un court-circuit.
N’utilisez jamais une prise de courant sans mise à la terre.
N’utilisez jamais de câble d’alimentation autre que le modèle Agilent
Technologies destiné à votre pays.
Utilisation de câbles non fournis L’utilisation de câbles non fournis par Agilent Technologies pourrait
endommager les composants électroniques ou provoquer des blessures.
Pour garantir un bon fonctionnement et le respect des règles de sécurité ou de
compatibilité électromagnétique, n’utilisez jamais d’autres câbles que ceux fournis par Agilent Technologies.
2 Site Requirements and Specifications
ATTENTION
Exigences d’installation
Encombrement
Les dimensions et le poids de l’instrument permettent un placement sur quasiment n’importe quel bureau ou paillasse de laboratoire. Veillez à prévoir un dégagement de 2,5 cm de part et d’autre et d’environ 8 cm à l’arrière pour assurer la circulation de l’air et le passage des raccordements électriques
Si le système d’électrophorèse capillaire Agilent 7100 doit être placé sur une paillasse, assurez-vous que celle-ci est conçue pour supporter le poids du système.
En fonctionnement, l’instrument doit être en position verticale.
Environnement
Votre instrument fonctionnera conformément aux spécifications dans les conditions de température ambiante et d’humidité relative précisées dans le
Tableau 6, page 31.
Condensation à l’intérieur du module
30 Agilent 7100 Capillary Electrophoresis System User Manual
La condensation endommage les circuits électroniques du système.
N’entreposez pas, ne transportez pas et n’utilisez pas votre module dans des conditions où les fluctuations de température risqueraient de provoquer de la condensation à l’intérieur du module.
Si le module a été transporté par temps froid, ne le sortez directement pas de son emballage : laissez-le atteindre progressivement la température ambiante pour éviter toute condensation.
Loading...
+ 236 hidden pages