any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G7100-90000
Edition
12/2009
Printed in Germany
Agilent Technologies, Deutschland GmbH
Hewlett-Packard-Strasse 8
76337 Waldbronn
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
Research Use Only
Agilent 7100 Capillary Electrophoresis System
In This Guide…
This guide contains information to install and operate the
Agilent 7100 Capillary Electrophoresis System.
1Introduction
This chapter gives an introduction to the instrument,
instrument overview and internal connectors.
2Site Requirements and Specifications
This chapter provides information on environmental
requirements, physical and performance specifications.
3Installing the Instrument
In this chapter you will find information about how to unpack
and to set up the Agilent CE and information about the
configuration of Instrument.
4LAN Configuration
This chapter provides information on connecting the CE
instrument to the Agilent ChemStation PC.
5Using the Agilent 7100 Capillary Electrophoresis System
This chapter provides information on how to set up the CE
instrument for an analysis and explains the basic settings.
6Troubleshooting and Diagnostics
This chapter gives an overview about the troubleshooting and
diagnostic features and the different user interfaces.
7Hardware Information
This chapter describes the instrument in more detail on
hardware and electronics.
8Maintenance
This chapter describes the maintenance of the CE instrument.
Agilent 7100 Capillary Electrophoresis System User Manual3
9Parts and Materials for Maintenance
This chapter provides information on parts for maintenance.
10Appendix
This chapter provides additional information on safety, legal
and web content associated with your 7100 CE instrument.
4Agilent 7100 Capillary Electrophoresis System User Manual
Contents
1Introduction13
Introduction to the Agilent 7100 Capillary Electrophoresis System14
Getting Help14
Where to Find Information15
What Learning Products are Available?16
2Site Requirements and Specifications17
Site Requirements18
Power Consideration18
Power Cords19
Bench Space20
Environment20
Physical Specifications21
Performance Specifications22
Anforderungen an den Aufstellungsort23
Stromversorgung23
Netzkabel24
Platzbedarf25
Umgebungsbedingungen25
Abmessungen und Gewicht26
Leistungsspezifikationen27
Exigences d’installation28
Remarques à propos de l’alimentation28
Câbles d’alimentation29
Encombrement30
Environnement30
Caractéristiques physiques31
Caractéristiques de performance32
Agilent 7100 Capillary Electrophoresis System User Manual5
Contents
Requisiti ambientali33
Alimentazione33
Cavi di alimentazione34
Spazio necessario35
Ambiente35
Specifiche fisiche36
Specifiche delle prestazioni37
Requisitos de las instalaciones38
Consideraciones sobre la corriente38
Cables de alimentación39
Espacio en el banco40
Entorno40
Especificaciones físicas41
Especificaciones de rendimiento42
設置について43
電源について43
電源コード44
作業台スペース45
環境条件45
物理的仕様46
性能仕様47
场地要求48
电源要求48
电源线49
工作台位置50
环境50
物理规格51
性能规格52
3Installing the Instrument53
Unpacking the Agilent CE Instrument54
Damaged Packaging54
Unpacking the CE Instrument54
6Agilent 7100 Capillary Electrophoresis System User Manual
Delivery Checklist55
Installing the Instrument57
Physical Connection57
4LAN Configuration61
What You Have to Do First62
TCP/IP Parameter Configuration63
Configuration Switch64
Initialization Mode Selection65
Link Configuration Selection69
Storing the Settings Permanently with Bootp70
Manual Configuration71
With Telnet72
Agilent Bootp Service75
Location of the Media Access Control (MAC) Address76
5Using the Agilent 7100 Capillary Electrophoresis System77
System Definition78
Controlling the Agilent CE Instrument79
Before You Start the Agilent CE Instrument79
Starting Up the Agilent CE Instrument79
Starting Your Computer and the Agilent ChemStation Software81
Starting the Agilent ChemStation Software the First Time82
Working with the CE Diagram84
Common Software Toolbar84
Software Toolbar85
Graphical User Interface Explanation86
Menus in the CE and DAD Diagram88
System Vialtable90
How to Prepare your Agilent CE Instrument for Analysis91
Capillary Electrophoresis, CE92
How to Use Your Agilent CE Instrument for Capillary Electrophoresis
Separations92
Contents
Agilent 7100 Capillary Electrophoresis System User Manual7
Contents
Agilent CE Method Parameters92
Vials96
Cassette Temperature97
High Voltage System98
Posttime and Stoptime100
Replenishment100
Preconditioning 101
Injection103
Timetable105
Detector Parameters 107
Switching the Lamp On110
Preparing the Replenishment System if Needed111
Preparing Vials for Buffer and Sample115
Installing a Capillary in the Alignment Interface124
Installing a Capillary in an Empty Capillary Cassette 128
Inserting the Cassette 132
Fraction Collection135
Capillary Electrophoresis, Plus High Pressure CE+p136
Requirements136
Outlet Lift Connection to the Tubing System138
Preparing the Agilent CE Instrument for High Pressure Use140
Operation and Function in CE+p Mode143
Injection Options with CE+p146
Timetable Options with CE+p146
CE+p Control Using the CE Diagram147
Capillary Electrochromatography, CEC148
How to Use Your Agilent CE Instrument for Capillary
Electrochromatography148
Preparing the Agilent CE Instrument for High Pressure Use150
CEC-Specific GUI Explanation153
Injection Options with CEC156
Timetable Options with CEC157
CEC Control Using the CE Diagram158
8Agilent 7100 Capillary Electrophoresis System User Manual
Running the CEC Analysis159
6Troubleshooting and Diagnostics161
CE Problem Solving162
Power On Fail162
Leaks163
Blocked Replenishment Needle165
Leakage Current165
Broken Capillary166
Contaminated Insulation Plate166
Empty Injection Vial167
Dirt in the Vials167
Air Bubble in Injection Vial or Run Buffer Vials167
Air Bubble in Capillary167
Problems with the Replenishment System and Buffers
Resident System176
Main System176
Firmware Updates 177
Diode Array Detector179
DAD Control180
Digital to Analog Converter181
Changing the Fuses for the Power Supplies182
Installing Drainage Tubing183
External Water Bath for Tray Cooling (Optional)184
Analog to Digital Converter186
Agilent 7100 Capillary Electrophoresis System User Manual9
Contents
8Maintenance189
Standard Operating Procedures for Maintenance of Your
Agilent CE Instrument190
Overview of Maintenance191
Early Maintenance Feedback (EMF)194
EMF Counters194
Using the EMF Counters194
Setting the EMF Limits195
Cleaning the Electrodes, Pre-punchers and Insulation Plate196
Prepare the Agilent CE Instrument197
Remove the Covers198
Accessing the Electrodes199
Accessing the Liquid Handling Module201
Accessing and Removing the Pre-punchers203
Cleaning the Electrodes205
Cleaning the Pre-punchers206
Cleaning the Insulation Plate207
Reinstalling the Pre-punchers208
Reinstalling the Electrodes and the Insulation Plate210
Reinstall the Capillary Cassette215
Reinstall the Front Cover215
Maintenance of the Replenishment System216
Changing the Buffer Composition in the Replenishment System217
The Replenishment System Will Not Be Used for Some Time218
Further Action219
Operations for Replenishment Maintenance219
Changing the Air Inlet Filter222
Accessing the Air Filter222
Changing the Lamp224
Removing the Existing Lamp225
Inserting a New Lamp227
Further Action227
Cleaning the Instrument228
10Agilent 7100 Capillary Electrophoresis System User Manual
9Parts and Materials for Maintenance229
Overview of Maintenance Parts230
10Appendix231
General Safety Information232
Operation232
Allgemeine Sicherheitsinformation235
Informations générales de sécurité238
Informazioni generali sulla sicurezza241
Información de seguridad244
安全に関する一般的な情報247
一般安全信息250
Setting Up a Test Sample Run252
Method Parameter CE (Default Settings)253
Method Parameter DAD255
Chemical and Biological Safety256
The Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC)257
Radio Interference258
Sound Emission259
Legal Notice260
Agilent Technologies on Internet261
Contents
Index263
Agilent 7100 Capillary Electrophoresis System User Manual11
Contents
12Agilent 7100 Capillary Electrophoresis System User Manual
Agilent 7100 Capillary Electrophoresis System
User Manual
1
Introduction
Introduction to the Agilent 7100 Capillary Electrophoresis System 14
Where to Find Information 15
What Learning Products are Available? 16
This chapter gives an introduction to the Agilent 7100 Capillary
Electrophoresis System and where to find specific information or help.
Agilent Technologies
13
1Introduction
Introduction to the Agilent 7100 Capillary Electrophoresis System
Introduction to the Agilent 7100 Capillary Electrophoresis
System
This handbook gives an overview of your Agilent 7100 Capillary
Electrophoresis System. It is designed to get you started working with the
system. We strongly advise you to make extensive use of the online help when
working with the system. There you find detailed reference and task
information which complements the overview given in this guide. Refer to
“Getting Help” later in this section for information on using the online help of
the ChemStation. Refer to “Where to Find Information” for help in finding
information on certain tasks.
Getting Help
You can get help by choosing the Help item from the top menu. This puts you
into the help index. From there you can browse through the information by
selecting the appropriate jumps (underlined words). To access specific
information in the help:
When you are in the Agilent ChemStation or Lab Advisor software:
• select a context-sensitive element and press F1.
When you are in the online help:
• search for a specific keyword using the search command,
• select a jump - a jump is an underlined word you can select to go to a topic
related to the word, or
• select a word or phrase with a dotted underline to see its definition. The
online help also contains online information on how to use help.
14Agilent 7100 Capillary Electrophoresis System User Manual
Where to Find Information
Tabl e 1Where to find information
TaskRefer to
Introduction1
Where to Find Information
• Installation of the Agilent CE System,
installing the external water bath,
installing the high sensitivity cell
• Setting up the Agilent CE System for
CE analysis, CE+p
• How to use the Agilent CE System in
the CE+p mode
• How to use the Agilent CE System for
capillary electrochromatography
• How to maintain the Agilent CE
System
• Setting up a sequenceOnline help "Understanding Your ChemStation"
• Data analysisOnline help "Understanding Your ChemStation"
• Setting up a reportOnline help "Understanding Your ChemStation"
• Spectral library search“ChemStation Applications” handbook and online help
• Setting up customized reports“ChemStation Applications” handbook
• Safety informationSafety information in “Appendix” on page 231
• Learning ChemStation concepts“Understanding Your ChemStation” handbook
• Learning the theory of the techniqueHigh Performance Capillary Electrophoresis: An
"The Core Agilent CE Instrument" “Installing the
Instrument” on page 53
Chapter “Using the Agilent 7100 Capillary
Electrophoresis System” on page 77"
“Capillary Electrophoresis, Plus High Pressure
CE+p” on page 136
“Capillary Electrochromatography, CEC” on page 148
Chapter “Maintenance” on page 189
Introduction”
The CE Partner CD-ROM: “The comprehensive,
interactive tool for beginners and advanced users of
CE”
• ApplicationAgilent web page (http://www.agilent.com), search for
Application notes
Agilent 7100 Capillary Electrophoresis System User Manual15
1Introduction
What Learning Products are Available?
What Learning Products are Available?
The Agilent CE System is supplied with the following:
• User's Guide
• High Performance Capillary Electrophoresis: An Introduction
The Agilent ChemStation and Lab Advisor/Instrument Utilities software DVD
is supplied with the following electronic documents:
• Installing Your Agilent ChemStation for LC and CE Systems
• Installing and Understanding Your Spectra Module
• Understanding Your Agilent ChemStation
• Online help
• Understanding Your Agilent ChemStation for CE Systems
16Agilent 7100 Capillary Electrophoresis System User Manual
Agilent 7100 Capillary Electrophoresis System
User Manual
2
Site Requirements and Specifications
Site Requirements 18
Physical Specifications 21
Performance Specifications 22
Anforderungen an den Aufstellungsort 23
Abmessungen und Gewicht 26
Leistungsspezifikationen 27
Exigences d’installation 28
Caractéristiques physiques 31
Caractéristiques de performance 32
Requisiti ambientali 33
Specifiche fisiche 36
Specifiche delle prestazioni 37
Requisitos de las instalaciones 38
Especificaciones físicas 41
Especificaciones de rendimiento 42
設置について 43
物理的仕様 46
性能仕様 47
场地要求 48
物理规格 51
性能规格 52
This chapter provides information on the environmental requirements, as well
as physical and performance specifications.
Agilent Technologies
17
2Site Requirements and Specifications
WARNING
WARNING
Site Requirements
Site Requirements
A suitable environment is important to ensure optimal performance of the
7100 Capillary Electrophoresis instrument.
Before you begin installation, check carefully that the place you have chosen
meets the requirements below.
Power Consideration
The instrument power supply has wide ranging capabilities and accepts any
standard line voltage in the range mentioned in Table 2 on page 21.
Consequently, there is no voltage selector on the back of the detector.
The module is partially energized when switched off, as long as the power cord is
plugged in.
18Agilent 7100 Capillary Electrophoresis System User Manual
Repair or maintenance work on the module can lead to personal injuries, for
example a shock hazard, when the cover is opened and the module is connected to
the power.
• Make sure that it is always possible to access the power plug.
• Remove the power cable from the instrument before opening the cover.
• Do not connect the power cable to the instrument while the electronic box in the
back of the instrument is opened or side panels are removed.
Incorrect line voltage on the instrument
Shock hazard or damage to your instrument can result if the device is connected to a
line voltage that is higher than specified.
• Connect your instrument to the specified line voltage.
CAUTION
Inaccessible power plug.
WARNING
WARNING
In the case of an emergency it must be possible to disconnect the instrument from the
power line at any time.
• Make sure the power connector of the instrument can be easily reached and
unplugged.
• Provide sufficient space behind the power socket of the instrument to unplug the
cable.
Power Cords
Different power cords are delivered with the instrument. The female end of all
power cords is identical. It plugs into the power-input socket at the rear. The
male end of each power cord is different and designed to match the wall socket
of a particular country or region.
Absence of a ground connection or use of an unspecified power cord
Site Requirements and Specifications2
Site Requirements
The absence of a ground connection or the use of an unspecified power cord can
lead to an electric shock or short circuit.
• Never operate your instrument from a power outlet without a ground connection.
• Never use a power cord other than the Agilent Technologies power cord designed
for your region.
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
• Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Agilent 7100 Capillary Electrophoresis System User Manual19
2Site Requirements and Specifications
CAUTION
Site Requirements
Bench Space
The instrument dimensions and weight mean that the instrument can be
placed on almost any desk or laboratory bench. It needs an additional 2.5 cm
(1.0 inch) of space on either side and approximately 8 cm (3.1 inches) in the
rear for air circulation and electric connections.
If the Agilent 7100 Capillary Electrophoresis instrument is to be placed on a
bench, make sure that the bench is designed to bear the weight of the system.
The instrument must be operated in a upright position.
Environment
Your instrument will work within the specifications for ambient temperatures
and relative humidity described in Table 2 on page 21.
Condensation within the module
Condensation will damage the system electronics.
20Agilent 7100 Capillary Electrophoresis System User Manual
• Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
• If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Physical Specifications
NOTE
Tabl e 2Physical specifications
TypeSpecificationComments
Weight35 kg (77.2 lbs)
Site Requirements and Specifications2
Physical Specifications
Dimensions
(width × depth × height)
Line voltage100–240 VAC ± 10%Wide-ranging capability
Line frequency50 or 60 Hz ± 5%
Power consumption350 VA / 300 W / 1024 BTU/h Maximum
Ambient operating
temperature
Ambient non-operating
temperature
Humiditybelow 80% at 31 °C (87.8 °F)Non-condensing
Operating altitudeUp to 2000 m (6500 ft)
External coolingmax. 0.5 bar (7.2 psi),
External pressure2-12 bar (29-174 psi)Oil-free air or nitrogen
Safety standards: IEC,
CSA, UL
Housing
350 x 510 x 590 mm
(13.8 x 20.1 x 23.2 inches)
5–40 °C (41–104 °F)
-40 to 70°C (-40 to 158 °F)
max. 50 °C (122 °F)
Installation category II,
Pollution degree 2
All materials are recyclable
Waterbath
For indoor use only
For applications run in CEC mode (Capillary Electrochromatography) or CE+p
mode (Capillary Electrophoresis mode with optional usage of higher pressure), an
external pressure supply of oil-free air or nitrogen with a maximum pressure of 15
bar (218 psi) can be used. A respective male adapter that fits into the instrument’s
female adapter and PTFE tubing are part of the shipment.
The Agilent 7100 Capillary Electrophoresis System is designed to operate in a typical
electromagnetic environment where RF transmitters, such as mobile phones, should not be
used in close proximity.
Agilent 7100 Capillary Electrophoresis System User Manual21
2Site Requirements and Specifications
Performance Specifications
Performance Specifications
Tabl e 3Performance specifications for the Agilent 7100 Capillary Electrophoresis System
TypeSpecificationComments
Safety featuresExtensive diagnostics, error detection and display,
current leak detection; low current limit
Liquid leak sensor; safety sensors on door and
cover disabling high voltage; vial sensor;
Control and data evaluationAgilent ChemStation under Windows® XP (SP3)
22Agilent 7100 Capillary Electrophoresis System User Manual
Anforderungen an den Aufstellungsort
WARNUNG
WARNUNG
Eine geeignete Umgebung ist wichtig für die optimale Leistungsfähigkeit
des 7100 Kapillarelektrophorese-Geräts.
Prüfen Sie sorgfältig, ob der von Ihnen gewählte Aufstellort die nachfolgenden
Anforderungen erfüllt, bevor Sie mit der Installation beginnen.
Stromversorgung
Das Weitbereichsnetzteil des Geräts arbeitet mit Standard-Netzspannungen
gemäß den Angaben in Tabelle 4 auf Seite 26. Daher ist an der Rückseite
des Detektors kein Spannungswahlschalter vorhanden.
Auch im ausgeschalteten Zustand fließt im Modul Strom, solange das Netzkabel
eingesteckt ist.
Site Requirements and Specifications2
Anforderungen an den Aufstellungsort
Agilent 7100 Capillary Electrophoresis System User Manual23
Die Durchführung von Reparatur- oder Wartungsarbeiten am Modul kann zu
Personenschäden wie z. B. durch Stromschlag führen, wenn das Modulgehäuse
geöffnet wird, während das Modul an die Netzspannung angeschlossen ist.
• Stellen Sie den freien Zugang zu den Netzkabeln sicher.
• Trennen Sie das Netzkabel vom Gerät, bevor Sie das Gehäuse öffnen.
• Schließen Sie das Netzkabel nicht an das Gerät an, während der Elektronikkasten
an der Geräterückseite geöffnet oder Seitenteile entfernt sind.
Falsche Netzspannung am Gerät
Wenn das Gerät an eine höhere Spannung als spezifiziert angeschlossen wird, kann
dies zu Überspannungsschäden oder einer völligen Zerstörung des Geräts führen.
• Schließen Sie das Gerät an die angegebene Netzspannung an.
2Site Requirements and Specifications
VORSICHT
WARNUNG
WARNUNG
Anforderungen an den Aufstellungsort
Unzugänglicher Netzstecker.
In einem Notfall muss es jederzeit möglich sein, das Gerät vom Stromnetz zu trennen.
• Stellen Sie sicher, dass der Netzstecker des Geräts leicht zugänglich ist.
• Lassen Sie hinter der Netzbuchse des Geräts genügend Platz zum Herausziehen
des Kabels.
Netzkabel
Das Gerät wird mit unterschiedlichen Netzkabeln ausgeliefert. Der weibliche
Stecker ist bei jedem Netzkabel identisch. Er wird an die Netzanschlussbuchse
an der Rückseite angeschlossen. Die Stecker am anderen Ende der Netzkabel
sind unterschiedlich und erfüllen die Normen unterschiedlicher Länder oder
Regionen.
Nicht vorhandene Erdung oder Verwendung eines nicht spezifizierten Netzkabels
Bei der Verwendung des Geräts ohne Erdung oder mit einem nicht spezifizierten
Netzkabel können Stromschläge und Kurzschlüsse auftreten.
24Agilent 7100 Capillary Electrophoresis System User Manual
• Betreiben Sie Ihr Gerät niemals an einer Spannungsquelle ohne Erdung.
• Verwenden Sie niemals ein anderes als das von Agilent zum Einsatz im jeweiligen
Land bereitgestellte Kabel.
Verwendung nicht im Lieferumfang enthaltener Kabel
Die Verwendung von Kabeln, die nicht von Agilent Technologies geliefert
wurden, kann zu einer Beschädigung der elektronischen Komponenten oder
zu Personenschäden führen.
• Verwenden Sie ausschließlich Originalkabel von Agilent Technologies, um eine
einwandfreie Funktion und die Einhaltung der Sicherheits- und EML-Bestimmungen
zu gewährleisten.
Platzbedarf
VORSICHT
Die Abmessungen und das Gewicht des Geräts ermöglichen seine Aufstellung
auf praktisch jedem Schreibtisch oder Laborarbeitstisch. Das Gerät benötigt
an jeder Seite zusätzlich 2,5 cm Platz und ungefähr 8 cm an der Rückseite
für die elektrischen Anschlüsse und für eine ausreichende Luftzirkulation.
Wenn das Agilent 7100 Kapillarelektrophorese System auf einem Labortisch
aufgestellt werden soll, stellen Sie sicher, dass der Tisch ausreichende
Tragfähigkeit für das System aufweist.
Das Gerät ist stehend zu betreiben!
Umgebungsbedingungen
Ihr Gerät arbeitet bei der Umgebungstemperatur und der relativen
Luftfeuchtigkeit, die in den Spezifikationen in Tabelle 4 auf Seite 26
angegeben sind.
Kondensation im Inneren des Moduls
Site Requirements and Specifications2
Anforderungen an den Aufstellungsort
Agilent 7100 Capillary Electrophoresis System User Manual25
Kondensation führt zur Beschädigung der Systemelektronik.
• Vermeiden Sie die Lagerung, den Versand oder den Betrieb der Pumpe unter
Bedingungen, die zu einer Kondensation in der Pumpe führen können.
• Nach einem Transport bei niedrigen Temperaturen muss das Gerät zur Vermeidung
von Kondensation in der Verpackung verbleiben, bis es sich auf Raumtemperatur
erwärmt hat.
2Site Requirements and Specifications
HINWEIS
Abmessungen und Gewicht
Abmessungen und Gewicht
Tabelle 4 Abmessungen und Gewicht
TypSpezifikationenKommentare
Gewicht35 kg
Abmessungen
(Breite × Tiefe × Höhe)
Netzspannung100–240 V Wechselstrom ± 10 %Weitbereichsnetzteil
Netzfrequenz50 oder 60 Hz ± 5 %
Leistungsaufnahme350 VA / 300 W / 1024 BTU/hMaximal
zulässige
Umgebungstemperatur
bei Lagerung
Umgebungstemperatur
bei Lagerung
Luftfeuchtigkeitunter 80 % bei 31 °CNicht kondensierend
Betriebshöhebis zu 2000 m
350 x 510 x 590 mm
(23,2 x 13,8 x 20,1 Zoll)
5–40 °C
-40 bis 70 °C
26Agilent 7100 Capillary Electrophoresis System User Manual
Externe Kühlungmax. 0,5 bar (7,2 psi), max. 50 °CWasserbad
Externer Druck2-12 bar (29-174 psi)Ölfreie Luft oder Stickstoff
Sicherheitsstandards: IEC,
CSA, UL
Gehäuse
Installationskategorie II,
Verschmutzungsgrad 2.
Alle Materialien sind recyclebar.
Nur für den Einsatz im
Innenbereich geeignet
Für Applikationen im CEC-Modus (Capillary Electrochromatography) oder
CE+p-Modus (Kapillarelektrophorese-Modus mit optionaler Verwendung höheren
Druckes) kann eine externe Druckversorgung mit ölfreier Luft oder Stickstoff und
einem Druckmaximum von 15 bar (218 psi) verwendet werden. Ein
entsprechender männlicher Adapter für den weiblichen Adapter des Geräts und
PTFE-Schläuche werden mitgeliefert.
Das Agilent 7100 Kapillarelektrophorese-System ist auf den Betrieb in einer typischen
elektromagnetischen Umgebung ausgelegt, in deren unmittelbarer Nähe keine HF-Sender
wie z. B. Mobiltelefone verwendet werden dürfen.
Site Requirements and Specifications2
Leistungsspezifikationen
Leistungsspezifikationen
Tabelle 5 Leistungsspezifikationen für das Agilent 7100 Kapillarelektrophorese-System
TypSpezifikationenKommentare
SicherheitsvorkehrungenUmfangreiche Diagnose, Fehlererkennung und
-anzeige, Kriechstromerkennung; Grenzwert für
zu niedrigen Strom
Flüssigkeitslecksensor; Sicherheitssensoren
an Tür und Abdeckung zur Trennung der
Hochspannung; Probenflaschensensor;
Steuerung und
Datenauswertung
KommunikationLAN, CAN (Controller-Area Network), RS-232C,
GLP-EigenschaftenFrühwarnsystem für fällige Wartungen (EMF, Early
Agilent ChemStation unter Windows® XP (SP3)
APG Remote: Ready-, Start-, Stop- und
Shut-down-Signale. USB (zukünftige
Ver wen dung) .
Analoger Ein-/Ausgang.
Maintenance Feedback), elektronische
Aufzeichnung von Wartungsarbeiten und
Fehlermeldungen
0 bis 1 V
Agilent 7100 Capillary Electrophoresis System User Manual27
2Site Requirements and Specifications
AVERTISSEMENT
AVERTISSEMENT
Exigences d’installation
Exigences d’installation
Un environnement adéquat est indispensable pour obtenir des performances
optimales de l’instrument d’électrophorèse capillaire 7100.
Avant de commencer l’installation, vérifiez soigneusement que l’emplacement
choisi est conforme aux exigences ci-dessous.
Remarques à propos de l’alimentation
Le module d’alimentation de l’instrument présente une tolérance importante.
Il accepte ainsi n’importe quelle tension de secteur se situant dans la plage
de tolérance précisée dans le Tableau 6, page 31. Ceci explique l’absence
de sélecteur de tension à l’arrière du détecteur.
Le module reste partiellement activé lorsqu’il est éteint, tant que le cordon
d’alimentation est branché.
Certaines opérations de réparation ou d’entretien sur le module peuvent
occasionner des blessures, par exemple l’électrocution, si le capot est
ouvert et que le module est branché.
•
Veillez à ce que la prise d’alimentation électrique reste à tout moment accessible.
• Débranchez le câble d’alimentation de l’instrument avant d’ouvrir le capot.
• Ne branchez pas le câble d’alimentation sur l’instrument si le boîtier
électronique à l’arrière est ouvert ou si les panneaux latéraux sont enlevés.
Tension de secteur incorrecte au niveau de l’instrument
Il existe un risque d’électrocution ou de dommages à l’instrument en cas de
raccordement à une ligne d’alimentation de tension supérieure à celle
spécifiée.
• Connectez l’instrument à la tension indiquée.
28Agilent 7100 Capillary Electrophoresis System User Manual
ATTENTION
Prise d’alimentation inaccessible.
AVERTISSEMENT
AVERTISSEMENT
En cas d’urgence, il doit être possible de débrancher à tout moment l’instrument
du secteur.
• Veillez à faciliter l’accès au connecteur d’alimentation de l’instrument
et à permettre de le débrancher facilement.
• Maintenez un espace suffisant derrière la prise d’alimentation de l’instrument
pour pouvoir débrancher le câble.
Câbles d’alimentation
Différents câbles d’alimentation sont livrés avec l’instrument. L’extrémité
femelle est la même pour tous les câbles. Elle s’enfiche dans l’embase
d’alimentation située à l’arrière. L’extrémité mâle, qui se branche sur la prise
de courant murale, varie selon le pays ou la région.
Absence de raccordement à la terre et utilisation d’un câble d’alimentation
non adapté
Site Requirements and Specifications2
Exigences d’installation
Agilent 7100 Capillary Electrophoresis System User Manual29
L’absence de raccordement à la terre et l’utilisation d’un câble d’alimentation
non adapté peuvent entraîner une électrocution ou un court-circuit.
• N’utilisez jamais une prise de courant sans mise à la terre.
• N’utilisez jamais de câble d’alimentation autre que le modèle Agilent
Technologies destiné à votre pays.
Utilisation de câbles non fournis
L’utilisation de câbles non fournis par Agilent Technologies pourrait
endommager les composants électroniques ou provoquer des blessures.
• Pour garantir un bon fonctionnement et le respect des règles de sécurité ou de
compatibilité électromagnétique, n’utilisez jamais d’autres câbles que ceux
fournis par Agilent Technologies.
2Site Requirements and Specifications
ATTENTION
Exigences d’installation
Encombrement
Les dimensions et le poids de l’instrument permettent un placement sur
quasiment n’importe quel bureau ou paillasse de laboratoire. Veillez à prévoir
un dégagement de 2,5 cm de part et d’autre et d’environ 8 cm à l’arrière pour
assurer la circulation de l’air et le passage des raccordements électriques
Si le système d’électrophorèse capillaire Agilent 7100 doit être placé sur une
paillasse, assurez-vous que celle-ci est conçue pour supporter le poids du
système.
En fonctionnement, l’instrument doit être en position verticale.
Environnement
Votre instrument fonctionnera conformément aux spécifications dans les
conditions de température ambiante et d’humidité relative précisées dans le
Tableau 6, page 31.
Condensation à l’intérieur du module
30Agilent 7100 Capillary Electrophoresis System User Manual
La condensation endommage les circuits électroniques du système.
• N’entreposez pas, ne transportez pas et n’utilisez pas votre module dans des
conditions où les fluctuations de température risqueraient de provoquer de
la condensation à l’intérieur du module.
• Si le module a été transporté par temps froid, ne le sortez directement pas de
son emballage : laissez-le atteindre progressivement la température ambiante
pour éviter toute condensation.
Loading...
+ 236 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.