Agilent Model 66101A: US34440101 and up
Agilent Model 66102A: US34210101 and up
Agilent Model 66103A: US34210101 and up
Agilent Model 66104A: US34450101 and up
Agilent Model 66105A: US34210101 and up
Agilent Model 66106A: US34210101 and up
For instruments with higher serial numbers,
a change page may be included.
For instruments with lower serial numbers,
refer to Appendix A.
Agilent Part No. 5959-3364 Printed in USA
Microfiche Part No. 5959-3365 Update: June, 2000
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent
Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of
Standards, to the extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other International
Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three
years from date of delivery. Agilent software and firmware products, which are designated by Agilent Technologies for use
with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their
programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During
the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective.
Agilent Technologies does not warrant that the op eration of the software, firmware, or hardware shall be uninterrupt ed or
error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated
by Agilent Technologies. Customer shall prepay shipping charges (and shall pay all duty and taxes) for products returned to
Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent
Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies ’ product price, only if
Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva
Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the
Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer,
Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental
specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED
OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT
TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Wa rranty options, extended support contracts, product
maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent
Technologies Sales and Service office for further information on Agilent Technologies’ full line of Support Programs.
2
SAFETY SUMMARY
The following general safety precautions must be obser ved during all phas es of oper ation of this power module.
Failure to comply with these precautions or with specific warnings elsewhere in this guide violates safety
standards of design, manufacture, and intended use of the power module. Agilent Technologies assumes no
liability for the user’s failure to comply with these requirements.
GROUND THE POWER MODULE.
When installed in a properly wired mainframe, this product is a Safety Class 1 instrument (provided with a protective earth
terminal). To minimize shock hazard, the mainframe must be connected to an electrical ground. The mainframe must be
connected to the a c power supply mains through a three-conductor power cable , with the third wire firmly connected to an
electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or
disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
DO NOT EXCEED INPUT RATINGS.
The mainframe is equipped with a power line filter to reduce electromagnetic interference and must be connected to a
properly grounded receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those
stated on the line rating label may cause leakage currents in excess of 5.0 mA peak.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the power module in the presence of flammable gases or fumes.
BEFORE APPLYING POWER.
Verify that the power module line voltage switch is set to match the available line voltage.
SAFETY SYMBOLS.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to
refer to the instruction manual (refer to Table of Contents) .
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not
correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING
sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if
not correctly performed or adhered to, could result in damage to or destruction of part or all of the
product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and
met.
Power modules that appear damaged or defective should be made inoperative and secured against unintended operation
until they can be repaired by qualified service personnel.
3
Printing History
The current edition of this guide is indicated belo w. Reprints of this guide containing minor corrections and updates may
have the same printing date. New editions are identified by a new printing date and, in some cases, by a new part number.
A new edition incorporates all new or corrected material since the previous edition. Changes to the guide occurring
between editions are covered by change sheets shipped with this guide. Also, if the serial number prefix of your power
module is higher than those listed on the title page of this guide, then it may or may not include a change sheet. That is
because even though the higher serial prefix indicates a design change, that change may not affect the content of the guide.
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document
may be photocopied, reproduced, or translated into another language without pr ior consent of Agilent Techno logies The
information contained in this document is subject to change without notice.
About This Manual............................................................................................................................................................... 7
Related Documents........................................................................................................................................................... 8
Revisions and Updates.......................................................................................................................................................... 8
Service Tools and Equipment............................................................................................................................................... 9
Verification and Performance Tests....................................................................................................................................11
Setup for Most Tests................................................................................................................................................... 12
Constant Voltage (CV) Tests.......................................................................................................................................... 13
Voltage Programming/Readback Accuracy................................................................................................................13
Constant Current (CC) Tests........................................................................................................................................... 15
CC Setup..................................................................................................................................................................... 15
Current Programming/Readback Accuracy................................................................................................................. 15
CC Load and Line Regulation .................................................................................................................................... 15
CC Load Effect........................................................................................................................................................... 16
CC Source Effect........................................................................................................................................................ 16
CC Noise (PARD)......................................................................................................................................................17
Accessing the Flowcharts.................................................................................................................................................... 25
If You Experienced a Selftest Failure or RunTime Error Message................................................................................. 26
If You Experienced Other Problems With the Module................................................................................................... 26
Test Points..........................................................................................................................................................................48
Built-In Test Functions....................................................................................................................................................... 48
EEPROM Troubleshooting and Initialization.....................................................................................................................51
Removing the Module and Module Connector............................................................................................................... 56
Removing the Cover.......................................................................................................................................................56
Removing the Fan........................................................................................................................................................... 56
Removing the Front Panel Assembly..............................................................................................................................57
Removing the Power Board............................................................................................................................................ 57
Removing the Bias Board............................................................................................................................................... 57
Removing the Front Panel Board.................................................................................................................................... 57
5
Installing the Power Board.............................................................................................................................................58
Installing the Bias Board ................................................................................................................................................58
Installing the Front Panel Board and Front Panel Assembly.......................................................................................... 59
Installing the Fan............................................................................................................................................................59
Installing the Cover.........................................................................................................................................................59
Theory of Operation............................................................................................................................................................. 61
AC Input and Bias Supplies................................................................................................................................................ 61
CC Amplifier...................................................................................................................................................................... 63
OR Gates ............................................................................................................................................................................ 63
Pulse-Width Modulator, FETS, and Isolation Transformer................................................................................................ 64
OV Circuit..........................................................................................................................................................................65
This manual is a complete guide for servicing the Agilent 6610xA Power Modules. Troubleshooting information for the
Agilent 66000A MPS Mainframe is provided in the mainframe service manual (see "Related Documents"), which should be
used with this manual.
NoteTroubleshooting the Agilent 6610xA Power Modules requires the use of one working Agilent 66000A
MPS Mainframe.
This manual is organized into the following major sections:
■Overview discusses safety and equipment considerations to be aware of when troubleshooting. It also lists the tools and
equipment required for troubleshooting.
■ Verification and Performance consists of a series of tests that are used to verify....
■ Troubleshooting procedures systematically verify selected circuits to isolate the cause of a problem. Suggestions fo r
remedying the problem are also provided. The following troubleshooting procedures are provided:
Disassembly procedures are provided at the end of the chapter.
■ Theory of Operation provides a brief description of the overall operation of the power module.
■ Replacement Parts lists part numbers for electrical components and miscellaneous chassis parts.
■ Schematics and component location diagrams are provided as an aid in troubleshooting and repair.
Overview
7
Related Documents
The following documents are related to this manual:
ManualPart Number
Agilent 66000A MPS Mainframe
Installation Guide
Agilent Series 6610xA MPS Power
Modules User’s Guide
Agilent Series 6610xA MPS Power
Modules Programming Guide
Agilent Series 66000A MPS
Mainframe Service Manual
Revisions and Updates
66000-90001Provides product specifications, installation, and
5959-3386
5959-3362
66000-90003Provides replaceable-parts lists, circuit diagrams, and
Description
connection information.
Describes the operation of the MPS Power Modules.
Describes SCPI programming commands, status
reporting.
verification procedures for the Agilent Series 66000A
MPS Mainframe.
Manual
Agilent Technologies instruments are identified by a two-part, ten-character serial number, such as 3246A-00517. The first
five characters (e.g. 3246A) are the serial prefix, which is the same for all identically made instruments. The last five digits
(e.g. 00517) are a unique serial number assigned to each instrument. If a significant design change is made, the prefix
changes but the last five numbers co ntinue in sequence.
This manual applies to Agilent 6610xA Power Modules with the same serial prefixes and serial numbers equal to or higher
than the ones shown on the title page. If the prefix number on your power module is higher than the one on the title page,
then the module was made after publication of the manual and may have hardware and/or firmware differences not covered
in the manual. If there are such differences, they are documented in one or more "Manual Changes" sheets sent with the
manual.
If the prefix number on your power module is lower than the one on the title page, then the module was made before the
publication of the manual and may have differences not covered in the manual. These differences are documented in
appendix A.
Firmware
This manual applies to p ower modules that run on software revision A.00.01 through A.00.04. Firmware revision A.00.04 is
backward compatible with all previous revisions. The ROM chip (U2) has a label that specifies the firmware revision.
8
Overview
Safety Considerations
The Agilent 6610xA Power Modules are Safety Class 1 instruments that connect to a protective earth terminal when
properly installed in an Agilent 66000A MPS Mainframe. Refer to the Safety Summary page at the beginning of this manual
for general safety procedures and the meaning of safety symbols appearing in the manual and on the module.
Electrostatic Discharge
The Modular Power System has components that can be damaged by ESD (electrostatic discharge).
when complete failure does not occur.
When working on the Modular Power System, observe all anti-static work practices. This includes, but is not limited to:
■ Working at a static-free station, such as a table covered with static-dissipative laminate or with a conductive table mat
(Agilent p/n 9300-0797).
■ Using a conductive wrist strap (Agilent p/n 9300-0969 or 9300-0970).
■ Grounding all metal equipment at the station to a single, common ground.
■ Connecting low-impedance test equipment to static sensitive components only when those components have power
applied to them.
■ Removing power from the Modular Power System before removing or installing printed circuit boards.
Failure to observe standard anti-static practices can result in serious degradation of performance even
Service Tools and Equipment
The following tools are required to repair the Agilent 6610xA MPS Modules:
■ T10 TORX driver
■ 7mm HEX NUT drivers
■ #2 POZI driver
■ Slot-head screwdriver - medium
■ DIP component removal tool
■ Soldering iron and solder
■ De-solder removal tool
■ Needle-nose pliers
■ Anti-static work mat and wrist strap
Overview
9
The following equipment is required to verify and troubleshoot the Agilent 6610xA Power Modules:
1 V = Verification; T = Troubleshooting.
2 Guildline Shunt Accuracy determined by Power Coefficient ( ± 0.01% + (0.0004%/watt))
3 Guildline Shunt Accuracy determined by Power Coefficient ( ± 0.01% + (0.002%/watt))
4 Use of the keyboard is recommended for displaying selftest error messages.
10
Overview
2
Verification and Performance Tests
This chapter contains test procedures to check the operation of the Agilent 6610xA Power Modules. The required test
equipment is specified in Chapter 1. Sample performance test record sheets are included at the end of the chapter.
Instructions are given for performing the tests either from a GPIB controller or the MPS keyboard. Two types of procedures
are provided: Operation Verification tests and Performance tests.
Operation Verification
Performance
If you encounter failures or out-of-specification test results, see "Troubleshooting Pr ocedures" in Chapter 3 of this manual.
Those procedures will determine if repair and/or calibration is required.
These tests do not check all parameters, but comprise a short procedure to verify that the
power module is performing properly.
These test all the Performance Specifications (not Supplementary Characteristics) listed in
Table 1-1 of the Power Module User’s Guide.
NoteThe power module must pass the selftest at power-on before the following tests can be performed. If the
module fails selftest, refer to the overall troubleshooting procedures in Chapter 3.
SHOCK HAZARD The tests should only be performed by qualified personnel. During the performance of
these tests, hazardous voltages may be present at the output of the module.
Operation Verification Tests
To assure that the module is operating properly, without testing all specified parameters, perform the following test
procedures:
1. Perform the turn-on and checkout procedures given in Chapter 3 of the Power Module User’s Guide.
2. Perform the Voltage Programming/Readback Accuracy and the Current Programming/Readback Accuracy Performance
Tests in this chapter.
Performance Tests
The following paragraphs provide test procedures for verifying the module’s compliance with the specifications listed in
Table 1-1 of the Power Module User’s Guide. All of the performance test specifications are listed in the Performance Test
Records at the end of this chapter. You can record the actual measured values in the column provided.
Verification and Performance Tests 11
Measurement Techniques
Setup for Most Tests
Most tests are performed at the rear terminals as shown in the following figure. Measure the dc voltage directly at the + S
and - S terminals. Set the connector sense switch for local sensing and use adequate wire gauge for load leads as described
in Chapter 2 of the Power Module User’s Guide.
Figure 2-1. Performance Test Setup
Many of the test procedures require the use of a variable load capable of dissipating the required power (see Table 1-1). If a
variable resistor is used, switches must be used to connect, disconnect, and short the load resistor. For most tests, an
electronic load can be used. The electronic load is considerably easier to use than load resistors but some may not be fast
enough to test transient recove ry time and may be too noisy for the noise (PARD) tests. Fixed load resistors may be used in
place of a variable load, with minor changes to the test procedures in this chapter. Also, if computer controlled test setups
are used, the relatively slow (compared to computers and system voltmeters) settling time and slew rates of the power
module may have to be taken into account. WAIT statements can be used in the test program if the test system is faster than
the module.
Current-Monitoring Resistor
To eliminate output current measurement error caused by voltage drops in the leads and connections, connect the current
monitoring resistor between the output and the load as a four-terminal device (see R
monitoring leads inside the load lead connections directly at the monitoring points on the resistor element.
in Figure 2-1). Connect the current
M
Programming
Table 2-1 lists the programming and current values for each module. You may program the module from the MPS Keyboard
or from a GPIB controller when performing the tests. The test procedures are written assuming that you know how to do
either or both. Complete instructions for remote and local programming are given in the module Power ModuleProgramming Guide and Power Module User’s Guide.
12 Verification and Performance Tests
Table 2-1. Power Module Voltage and Current Values
If more than one meter or a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to
avoid mutual coupling effects. For constant voltage dc tests, connect only to + S and - S because the module regulates the
output voltage that appears between those terminals, not between the + and - output terminals. Use coaxial cable or shielded
2-wire cable to avoid noise pickup on the test leads.
Voltage Programming/Readback Accuracy
This test verifies that the voltage programming, readback, and front panel display functions are within specifications. Note
that the values read back over the GPIB should be identical to those displayed on the keyboard display.
1. Turn off the power module and connect a digital voltmeter between the + S and--S terminals (see Figure 2-1).
2. Turn on the module and program it for zero volts and the maximum programmable current (see Table 2-1) with the load
off.
3. Record the output voltage readings on the DVM and the keyboard display. The readings should be within the limits
specified in the Performance Test Record Tables under CV PROGRAMMING @ 0 VOLTS, for the particular model
being tested. The CV annunciator should be on and the output current reading should be approximately zero.
4. Program the output voltage to full scale (see Table 2-1).
5. Record the output voltage readings on the DVM and the keyboard display. The readings should be within the limits
specified in the Performance Test Record Tables under CV PROGRAMMING @ FULL SCALE, for the particular
model being tested.
CV Load Effect
This test measures the change in output voltage resulting from a change in output current from full-load to no-load.
1. Turn off the module and connect the output as shown in Figure 2-1 with the DVM connected between the + S and - S
terminals.
2. Turn on the module and program the current to the maximum programmable value and the voltage to the full-scale
value (see Table 2-1).
3. Adjust the load for the full-scale current (see Table 2-1) as indicated on the keyboard display. The front panel CV
annunciator must be on. If it is not, ad just the load so that the output c urrent drops slightly until the annunciator comes
on.
4. Record the output voltage reading on the DVM connected to + S and - S.
5. Open the load and again record the DVM voltage reading.
6. The difference between the DVM readings in steps (4) and (5) is the load effect voltage and should not exceed the
value listed in the Performance Test Record Tables under CV LOAD EFFECT, for the model being tested .
Verification and Performance Tests 13
CV Source Effect
This test measures the change in output voltage that results from a change in ac line voltage from the minimum to maximum
value within the line voltage specifications.
1. Turn off the module and c onnect the ac power line through a variable-voltage tr ansformer.
2. Connect the output as shown in Figure 2-1 with the DVM connected between the + S and - S terminals. Set the
transformer to nominal line voltage (either 115Vac or 230Vac).
3. Turn on the module and program the current to the maximum programmable value and the output voltage to the
full-scale value (see Table 2-1).
4. Adjust the load for the full-scale current value (see Table 2-1) as indicated on the keyboard display. The front panel CV
annunciator must be on. If it is not, ad just the load so that the output c urrent drops slightly until the annunciator comes
on.
5. Adjust the transformer to the LOW line voltage (e.g., 87Vac for a 115Vac nominal input, or 174Vac for a 230Vac
nominal input).
6. Record the output voltage reading on the DVM.
7. Adjust the transformer to the HIGH line voltage (e.g., 132Vac for a 115Vac nominal input, or 250Vac for a 230Vac
nominal input).
8. Record the output voltage reading on the DVM.
9. The difference between the DVM readings in steps (6) and (8) is the source effect voltage and should not exceed the
value listed in the Performance Test Record Tables under CV SOURCE EFFECT, for the model being tested.
CV Noise (PARD)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac voltage
superimposed on the dc output voltage. This test measures CV PARD, specified as the rms or peak-to-peak output voltage
over the frequency range of 20Hz to 20MHz.
1. Turn off the module and connect the output as shown in Figure 2-1 to an oscilloscope (ac coupled) between the + and--
terminals. Set the oscilloscope’s bandwidth limit to 20MHz (30MHz on the Agilent 54504A) and use an RF tip on the
oscilloscope probe.
2. Turn on the module and program the current to the maximum programmable value and the output voltage to the
full-scale value (see Table 2-1).
3. Adjust the load for the full-scale current value (see Table 2-1) as indicated on the keyboard display.
4. The waveform on the oscilloscope should not exceed the peak-to-peak limits in the Performance Test Record Tables
under CV NOISE (PARD), for the model being tested.
5. Disconnect the oscilloscope and connect an ac rms voltmeter in its place. The rms voltage reading should not exceed
the rms limits in the Performance Test Record Tables under CV NOISE (PARD) for the model being tested.
Transient Recovery Time
This test measures the time for the output voltage to recover to within the specified value following a 10% change in the
load current.
1. Turn off the module and connect the output as shown in Figure 2-1 with the oscilloscope across the + S and -S
terminals.
2. Turn on the module and program the output voltage to the full-scale value and the current to the maximum
programmable value (see Table 2-1).
3. Set the load to the Constant Current mode and program the load current to 90% of the power module full-scale rated
current.
4. Set the electronic load’s transient generator frequency to 100Hz and its duty cycle to 50%.
5. Program the load’s transient level to the module’s full-scale current value and turn the transient on.
6. Adjust the oscilloscope for a waveform similar to that in Figure 2-2.
7. The output voltage should return to within 100mV of the nominal value in less than 1ms. Check both loading and
14 Verification and Performance Tests
unloading transients by triggering on the positive and negative slope.
Figure 2-2. Transient Response Waveform
Constant Current (CC) Tests
CC Setup
Follow the general setup instructions in the Measurement Techniques paragraph and the specific instructions given in the
following paragrap hs.
Current Programming/Readback Accuracy
This test verifies that the current programming and readback are within specification. The accuracy of the current
monitoring resistor is determined by the power coefficient (see Table 1-1).
1. Turn off the module and connect the current monitoring resistor directly across the output and a DVM across the
resistor. Refer to the "Current Monitoring Resistor" paragraph.
2. Turn on the module and program the output voltage to 5V and the current to zero.
3. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and
record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits
specified in the Performance Test Record Tables under CC PROGRAMMING @ 0 AMPS, for the particular model
being tested.
4. Program the output voltage to 5V and the current to full-scale (see Table 2-1).
5. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and
record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits
specified in the Performance Test Record Tables under CC PROGRAMMING @ FULL SCALE, for the particular
model being tested.
CC Load and Line Regulation
These tests (CC Load Effect and CC Source Effect given below) are tests of the dc regulation of the power module’s output
current. To insure that the values read are not the instantaneous measurement of the ac peaks of the output current ripple,
several dc measurements should be made and the average of these readings calculated.
Verification and Performance Tests 15
The following steps show how to set up an Agilent 3458A System Voltmeter from its front panel to take a statistical average
of 100 readings.
represents the unlabeled shift key in the FUNCTION/RANGE group.
1. Program 10 power line cycles per samp le by pressing
2. Program 100 samples per trigger by pressing
3. Set up the voltmeter to take measurements in the statistical mode as follows:
4. a. Press
b. Press c. Press
5. Now set up the voltmeter to read the average value of the measurements as follows:
6. a. Press
b. Press c. Press
7. Execute the average reading program by pressing
8. Wait for 100 readings and then read the average measurement by pressing
To repeat the measurement, perform steps (7) and (8).
CC Load Effect
This test measures the change in output current for a change in the load from full-scale output voltage to short circuit.
1. Turn off the module and connect the output to be tested as shown in Figure 2-1 with the voltmeter connected across the
2. Turn on the module and program the current to the full-scale current value and the output voltage to the maximum
3. Adjust the load in the CV mode for full-scale voltage as indicated on the keyboard display. Check that the CC
4. Record the output current reading (DVM reading ÷ current monitor resistance value in ohms).
until the MATH function is selected; then press .
until the STAT function is selected; then press .
until the RMATH function is selected; then press .
until the MEAN function is selected; then press .
current monitoring resistor.
programmable voltage value (see Table 2-1).
annunciator is on. If it is not, adjust the load to drop the output voltage slightly until the annunciator c omes on.
.
.
.
.
.
. Record this as your result.
Note You may want to use the average reading program described previously.
5. Short the load switch and record the output current reading.
6. The difference in the current readings in steps (4) and (5) is the load effect and should not exceed the limit specified in
the Performance Test Record Tables under CC LOAD EFFECT, for the particular model being tested.
CC Source Effect
This test measures the change in output current that results when the ac line voltage changes from the minimum to the
maximum value within the specifications.
1. Turn off the module and c onnect the ac power line through a variable-voltage tr ansformer.
2. Connect the output terminals as shown in Figure 2-1 with the voltmeter connected across the current monitoring
resistor. Set the transformer to the nominal line voltage.
3. Turn on the module and program the current to the full-scale value and the output voltage to the maximum
programmable value (see Table 2-1).
4. Adjust the load in the CV mode for full-scale voltage as indicated on the front panel display. Check that the CC
annunciator is on. If it is not, adjust the load to drop the output voltage slightly until the annunciator c omes on.
5. Adjust the transformer to the LOW line voltage (e.g., 87Vac for a 115Vac nominal input, or 174Vac for a 230Vac
nominal input).
6. Record the output current reading (DVM reading ÷ current monitoring resistor in ohms).
16 Verification and Performance Tests
Note You may want to use the average reading program described previously.
7. Adjust the transformer to the HIGH line voltage (e.g., 132Vac for a 115Vac nominal input, or 250Vac for a 230Vac
nominal input), and record the output current reading.
8. The difference in the current readings in steps (6) and (7) is the CC source effect and should not exceed the values
listed in the Performance Test Record Tables under CC SOURCE EFFECT, for the particular model being tested.
CC Noise (PARD)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac current as well as
an ac voltage superimposed on the dc output. Constant current (CC) PARD is specified as the rms output current in a
frequency range 20Hz to 20MHz with the module in CC operation.
1. Turn off the module and connect the load resistor (R
) and rms voltmeter as shown in Figure 2-1. Use only a resistive
L
load for this test. Keep leads as short as possible to reduce noise pickup.
2. Check the test setup for noise with the module turned off. Other equipment (e.g. computers, DMM, etc.) may affect the
reading.
3. Turn on the module and program the current to full-scale and the output voltage to the maximum programmable value
(see Table 2-1).
4. The output current should be at the full-scale rating with the CC Mode on.
5. Divide the reading on the rms voltmeter by the shunt resistance to obtain rms current. It should not exceed the values
listed in the Performance Test Record Tables under CC NOISE (Rms) for the particular unit being tested.
Verification and Performance Tests 17
Table 2-2. Performance Test Record for Model Agilent 66101A (8V, 16A)
Pre Cal_____________Post Cal______________
Model Agilent 66101A Temperature___________________________Test Performed
By_______________________
Serial Number_______________________Humidity________________________Date__________________________
CC PROGRAMMING @ 0 amps
Output Accuracy (Io)
Display and Readback Accuracy
0 ± 90mV
DVM ± 54mV
200V ± 150mV
DVM ± 94mV
50mV
30mV
0 ± 0.4mA
Io ± 0.3mA
2.6µV
2.6µV
2.7mV
2.7mV
350µV
350µV
1.5mV
5.0mV
15µA
15µA
Actual
________
________
________
________
________
________
________
________
________
________
CC PROGRAMMING @ full-scale
Output Accuracy (Io)
Display and Readback Accuracy
CC LOAD EFFECT
CC SOURCE EFFECT
CC NOISE (RMS)1mA0.2mA________
0.75A ± 625µA
Io ± 450µA
10µA1
30µA1µA
Verification and Performance Tests 23
100µA
100µA
µ
A
________
________
________
________
3
Troubleshooting
This chapter provides troubleshooting and repair information for the Agilent 6610xA Power Modules. Before attempting to
troubleshoot the modules, ensure that the problem is not with an external circuit or application, the mainframe, the GPIB
controller, the application program, or the input power line.
SHOCK HAZARD Most of the troubleshooting procedures given in this chapter are performed with power
applied and protective covers removed. Such maintenance should be performed only by trained
service personnel who are aware of the hazards (for example, tire and electrical shock).
Hazardous voltages are present throughout the P ower and Bias board assemblies. Ac line voltage is present
on some connectors even when the module is not turned on. The dc rail voltage to the module is 310 volts
when the unit is turned on.
Troubleshooting Sequence
1. Verify that the module is at fault.
2. Use the flowcharts to locate the functional block at fault.
3. Troubleshoot the functional block per flowchart instructions.
Selftest routines built into firmware are provided to help isolate a problem to a particular circuit on the board. Once a
problem has been isolate d to a circuit, suggestions are given in the appropriate flowchart as to what component may be at
fault.
This instrument uses components that can be damaged or suffer serious performance degradation as a
result of ESD (electrostatic discharge). Observe the standard anti-static precautions discussed in
Chapter 1 to avoid damage to components.
Accessing the Flowcharts
The following troubleshooting flowcharts are provided to help you identify and isolate a problem to a specific circuit. Once
a problem has been narrowed down, suggestions are provided as to the components that may be responsible fo r the problem.
Refer to the schematic diagrams for detailed information on circuit wiring and component function.
Oscilloscope waveforms are provided on some flowcharts along with the settings required to generate the waveforms.
Connect the module to ac mains through an isolation transformer when making measurements with an
oscilloscope or any equipment that has the measuring input grounded.
If You Experienced a Selftest Failure or RunTime Error Message
If you experienced a selftest failure or runtime error message as described in Chapter 3 of the Power Module User’s Guide,
disassemble the module and proceed directly to the flowchart in Figure 3-3 "Troubleshooting Error Messages".
If You Experienced Other Problems With the Module
If you experienced problems other than a selftest failure or a runtime error message, disassemble the module and proceed
directly to Figure 3-1 "Overall Troubleshooting".
26 Troubleshooting
Figure 3-1. Overall Troubleshooting (Sheet 1 of 3)
Troubleshooting 27
28 Troubleshooting
Figure 3-1. Overall Troubleshooting (Sheet 2 of 3)
Figure 3-1. Overall Troubleshooting (Sheet 3 of 3)
Troubleshooting 29
Figure 3-2. Troubleshooting Microprocessor Circuits (Sheet 1 of 2)
30 Troubleshooting
Loading...
+ 71 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.