Agilent Technologies 6610XA User Manual

SERVICE GUIDE
for Agilent 6610xA
Power Modules
Serial Numbers:
Agilent Model 66101A: US34440101 and up Agilent Model 66102A: US34210101 and up Agilent Model 66103A: US34210101 and up Agilent Model 66104A: US34450101 and up Agilent Model 66105A: US34210101 and up Agilent Model 66106A: US34210101 and up
For instruments with higher serial numbers, a change page may be included.
For instruments with lower serial numbers, refer to Appendix A.
Agilent Part No. 5959-3364 Printed in USA Microfiche Part No. 5959-3365 Update: June, 2000
CERTIFICATION
Agilent Technologies certifies that this product met its published specifications at time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Bureau of Standards, to the extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other International Standards Organization members.
WARRANTY
This Agilent Technologies hardware product is warranted against defects in material and workmanship for a period of three years from date of delivery. Agilent software and firmware products, which are designated by Agilent Technologies for use with a hardware product and when properly installed on that hardware product, are warranted not to fail to execute their programming instructions due to defects in material and workmanship for a period of 90 days from date of delivery. During the warranty period Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Agilent Technologies does not warrant that the op eration of the software, firmware, or hardware shall be uninterrupt ed or error free.
For warranty service, with the exception of warranty options, this product must be returned to a service facility designated by Agilent Technologies. Customer shall prepay shipping charges (and shall pay all duty and taxes) for products returned to Agilent Technologies for warranty service. Except for products returned to Customer from another country, Agilent Technologies shall pay for return of products to Customer.
Warranty services outside the country of initial purchase are included in Agilent Technologies ’ product price, only if Customer pays Agilent Technologies international prices (defined as destination local currency price, or U.S. or Geneva Export price).
If Agilent Technologies is unable, within a reasonable time to repair or replace any product to condition as warranted, the Customer shall be entitled to a refund of the purchase price upon return of the product to Agilent Technologies.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Customer, Customer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation and maintenance. NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE THE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
ASSISTANCE
The above statements apply only to the standard product warranty. Wa rranty options, extended support contracts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest Agilent Technologies Sales and Service office for further information on Agilent Technologies’ full line of Support Programs.
2
SAFETY SUMMARY
The following general safety precautions must be obser ved during all phas es of oper ation of this power module. Failure to comply with these precautions or with specific warnings elsewhere in this guide violates safety standards of design, manufacture, and intended use of the power module. Agilent Technologies assumes no liability for the user’s failure to comply with these requirements.
GROUND THE POWER MODULE.
When installed in a properly wired mainframe, this product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the mainframe must be connected to an electrical ground. The mainframe must be connected to the a c power supply mains through a three-conductor power cable , with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury.
DO NOT EXCEED INPUT RATINGS.
The mainframe is equipped with a power line filter to reduce electromagnetic interference and must be connected to a properly grounded receptacle to minimize electric shock hazard. Operation at line voltages or frequencies in excess of those stated on the line rating label may cause leakage currents in excess of 5.0 mA peak.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE.
Do not operate the power module in the presence of flammable gases or fumes.
BEFORE APPLYING POWER.
Verify that the power module line voltage switch is set to match the available line voltage.
SAFETY SYMBOLS.
Instruction manual symbol: the product will be marked with this symbol when it is necessary for the user to refer to the instruction manual (refer to Table of Contents) .
Indicates hazardous voltages. Indicate earth (ground) terminal.
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met.
Power modules that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
3
Printing History
The current edition of this guide is indicated belo w. Reprints of this guide containing minor corrections and updates may have the same printing date. New editions are identified by a new printing date and, in some cases, by a new part number. A new edition incorporates all new or corrected material since the previous edition. Changes to the guide occurring between editions are covered by change sheets shipped with this guide. Also, if the serial number prefix of your power module is higher than those listed on the title page of this guide, then it may or may not include a change sheet. That is because even though the higher serial prefix indicates a design change, that change may not affect the content of the guide.
Edition 1 . . . . October, 1993 (Agilent Model 6610xA)
Update . . . . . June, 2000
Copyright 1993, 2000 Agilent Technologies, Inc.
This document contains proprietary information protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced, or translated into another language without pr ior consent of Agilent Techno logies The information contained in this document is subject to change without notice.
4

Table of Contents

Overview.................................................................................................................................................................................. 7
About This Manual............................................................................................................................................................... 7
Related Documents........................................................................................................................................................... 8
Revisions and Updates.......................................................................................................................................................... 8
Manual.............................................................................................................................................................................. 8
Firmware........................................................................................................................................................................... 8
Safety Considerations........................................................................................................................................................... 9
Electrostatic Discharge......................................................................................................................................................... 9
Service Tools and Equipment............................................................................................................................................... 9
Verification and Performance Tests....................................................................................................................................11
Operation Verification Tests...............................................................................................................................................11
Performance Tests..............................................................................................................................................................11
Measurement Techniques............................................................................................................................................... 12
Setup for Most Tests................................................................................................................................................... 12
Current-Monitoring Resistor....................................................................................................................................... 12
Programming.................................................................................................................... .............................................. 12
Constant Voltage (CV) Tests.......................................................................................................................................... 13
CV Setup.....................................................................................................................................................................13
Voltage Programming/Readback Accuracy................................................................................................................13
CV Load Effect........................................................................................................................................................... 13
CV Source Effect........................................................................................................................................................ 14
CV Noise (PARD)...................................................................................................................................................... 14
Transient Recovery Time............................................................................................................................................14
Constant Current (CC) Tests........................................................................................................................................... 15
CC Setup..................................................................................................................................................................... 15
Current Programming/Readback Accuracy................................................................................................................. 15
CC Load and Line Regulation .................................................................................................................................... 15
CC Load Effect........................................................................................................................................................... 16
CC Source Effect........................................................................................................................................................ 16
CC Noise (PARD)......................................................................................................................................................17
Troubleshooting.................................................................................................................................................................... 25
Troubleshooting Sequence..................................................................................................................................................25
Accessing the Flowcharts.................................................................................................................................................... 25
If You Experienced a Selftest Failure or RunTime Error Message................................................................................. 26
If You Experienced Other Problems With the Module................................................................................................... 26
Test Points..........................................................................................................................................................................48
Built-In Test Functions....................................................................................................................................................... 48
EEPROM Troubleshooting and Initialization.....................................................................................................................51
Troubleshooting.............................................................................................................................................................. 51
Initialization.................................................................................................................................................................... 51
Disassembly Procedures..................................................................................................................................................... 56
Removing the Module and Module Connector............................................................................................................... 56
Removing the Cover.......................................................................................................................................................56
Removing the Fan........................................................................................................................................................... 56
Removing the Front Panel Assembly..............................................................................................................................57
Removing the Power Board............................................................................................................................................ 57
Removing the Bias Board............................................................................................................................................... 57
Removing the Front Panel Board.................................................................................................................................... 57
5
Installing the Power Board.............................................................................................................................................58
Installing the Bias Board ................................................................................................................................................58
Installing the Front Panel Board and Front Panel Assembly.......................................................................................... 59
Installing the Fan............................................................................................................................................................59
Installing the Cover.........................................................................................................................................................59
Theory of Operation............................................................................................................................................................. 61
AC Input and Bias Supplies................................................................................................................................................ 61
Microprocessor Circuits......................................................................................................................................................61
CV/CC DACs.....................................................................................................................................................................61
CV Amplifier...................................................................................................................................................................... 63
CC Amplifier...................................................................................................................................................................... 63
OR Gates ............................................................................................................................................................................ 63
Pulse-Width Modulator, FETS, and Isolation Transformer................................................................................................ 64
Downprogrammer Circuit................................................................................................................................................... 64
Readback Multiplexer......................................................................................................................................................... 64
OV Circuit..........................................................................................................................................................................65
Replacement Parts................................................................................................................................................................67
Diagrams ............................................................................................................................................................................... 83
Schematic Sheets................................................................................................................................................................ 83
Component Location Diagrams.......................................................................................................................................... 83
Schematic Notes.................................................................................................................................................................84
Manual Backdating............................................................................................................................................................ 97
Index.................................................................................................................................................................................... 99
6
1

Overview

About This Manual

This manual is a complete guide for servicing the Agilent 6610xA Power Modules. Troubleshooting information for the Agilent 66000A MPS Mainframe is provided in the mainframe service manual (see "Related Documents"), which should be used with this manual.
Note Troubleshooting the Agilent 6610xA Power Modules requires the use of one working Agilent 66000A
MPS Mainframe.
This manual is organized into the following major sections:
Overview discusses safety and equipment considerations to be aware of when troubleshooting. It also lists the tools and
equipment required for troubleshooting.
Verification and Performance consists of a series of tests that are used to verify....
Troubleshooting procedures systematically verify selected circuits to isolate the cause of a problem. Suggestions fo r
remedying the problem are also provided. The following troubleshooting procedures are provided:
OverallMicroprocessorError MessagesOvervoltage CircuitOutput CircuitCV and CC CircuitsDownprogrammer CircuitsDAC CircuitReadback Circuit
Disassembly procedures are provided at the end of the chapter.
Theory of Operation provides a brief description of the overall operation of the power module.
Replacement Parts lists part numbers for electrical components and miscellaneous chassis parts.
Schematics and component location diagrams are provided as an aid in troubleshooting and repair.
Overview
7
Related Documents
The following documents are related to this manual:
Manual Part Number
Agilent 66000A MPS Mainframe
Installation Guide Agilent Series 6610xA MPS Power
Modules User’s Guide Agilent Series 6610xA MPS Power
Modules Programming Guide Agilent Series 66000A MPS
Mainframe Service Manual

Revisions and Updates

66000-90001 Provides product specifications, installation, and
5959-3386
5959-3362
66000-90003 Provides replaceable-parts lists, circuit diagrams, and
Description
connection information.
Describes the operation of the MPS Power Modules.
Describes SCPI programming commands, status
reporting.
verification procedures for the Agilent Series 66000A MPS Mainframe.
Manual
Agilent Technologies instruments are identified by a two-part, ten-character serial number, such as 3246A-00517. The first five characters (e.g. 3246A) are the serial prefix, which is the same for all identically made instruments. The last five digits (e.g. 00517) are a unique serial number assigned to each instrument. If a significant design change is made, the prefix changes but the last five numbers co ntinue in sequence.
This manual applies to Agilent 6610xA Power Modules with the same serial prefixes and serial numbers equal to or higher than the ones shown on the title page. If the prefix number on your power module is higher than the one on the title page, then the module was made after publication of the manual and may have hardware and/or firmware differences not covered in the manual. If there are such differences, they are documented in one or more "Manual Changes" sheets sent with the manual.
If the prefix number on your power module is lower than the one on the title page, then the module was made before the publication of the manual and may have differences not covered in the manual. These differences are documented in appendix A.
Firmware
This manual applies to p ower modules that run on software revision A.00.01 through A.00.04. Firmware revision A.00.04 is backward compatible with all previous revisions. The ROM chip (U2) has a label that specifies the firmware revision.
8
Overview
Safety Considerations
The Agilent 6610xA Power Modules are Safety Class 1 instruments that connect to a protective earth terminal when properly installed in an Agilent 66000A MPS Mainframe. Refer to the Safety Summary page at the beginning of this manual for general safety procedures and the meaning of safety symbols appearing in the manual and on the module.

Electrostatic Discharge

The Modular Power System has components that can be damaged by ESD (electrostatic discharge).
when complete failure does not occur.
When working on the Modular Power System, observe all anti-static work practices. This includes, but is not limited to:
Working at a static-free station, such as a table covered with static-dissipative laminate or with a conductive table mat
(Agilent p/n 9300-0797).
Using a conductive wrist strap (Agilent p/n 9300-0969 or 9300-0970).
Grounding all metal equipment at the station to a single, common ground.
Connecting low-impedance test equipment to static sensitive components only when those components have power
applied to them.
Removing power from the Modular Power System before removing or installing printed circuit boards.
Failure to observe standard anti-static practices can result in serious degradation of performance even

Service Tools and Equipment

The following tools are required to repair the Agilent 6610xA MPS Modules:
T10 TORX driver
7mm HEX NUT drivers
#2 POZI driver
Slot-head screwdriver - medium
DIP component removal tool
Soldering iron and solder
De-solder removal tool
Needle-nose pliers
Anti-static work mat and wrist strap
Overview
9
The following equipment is required to verify and troubleshoot the Agilent 6610xA Power Modules:
Table 1-1. Service Test Equipment
Equipment Required Critical Specifications Recommended Model Use
GPIB Controller Full GPIB capabilities HP Series 200/300 computer T
Oscilloscope Sensitivity: 1mV
Agilent 54504A T Bandwidth Limit: 20MHz Probe: 1:1 with RF tip
Rms Voltmeter True RMS Band width: 20MHz
Agilent 3400B T Sensitivity: 100µV
Digital Voltmeter Resolution: 10nV @ 1V (8 1/2 digit) Agilent 3458A Accuracy: 20 ppm V,T
Variable Voltage Transformer
CC RMS Noise Resistor (R
)
M
Electronic Load Voltage and current range must exceed range
Adjustable from- 13% to + 6% of input range. 1KVA min.
Agilent 66101A,102A - 100A (.01 Ω)
2
0.04% Agilent 66103A,106A - 15A (.01 Ω)
3
0.04%
of module under test. Power Range: 600W minimum
T
±
Guildline 9230/100
V,T
Guildline 9230/15
±
Agilent 6050A Load mainframe
V,T and Agilent 60503A module (240V) or Agilent 60507A module (120V).
Load Resistor (RL)
Agilent 66101A- 0.1 Ω ± 5% 300W Agilent 66102A-106A- 1.0 Ω ± 5% 300W
Ohmite C300KR10 Ohmite C300KIR0 T
1
MPS Mainframe Agilent 66000A V,T
MPS Keyboard
4
Agilent 66001A V,T
1 V = Verification; T = Troubleshooting. 2 Guildline Shunt Accuracy determined by Power Coefficient ( ± 0.01% + (0.0004%/watt)) 3 Guildline Shunt Accuracy determined by Power Coefficient ( ± 0.01% + (0.002%/watt)) 4 Use of the keyboard is recommended for displaying selftest error messages.
10
Overview
2

Verification and Performance Tests

This chapter contains test procedures to check the operation of the Agilent 6610xA Power Modules. The required test equipment is specified in Chapter 1. Sample performance test record sheets are included at the end of the chapter. Instructions are given for performing the tests either from a GPIB controller or the MPS keyboard. Two types of procedures are provided: Operation Verification tests and Performance tests.
Operation Verification
Performance
If you encounter failures or out-of-specification test results, see "Troubleshooting Pr ocedures" in Chapter 3 of this manual. Those procedures will determine if repair and/or calibration is required.
These tests do not check all parameters, but comprise a short procedure to verify that the power module is performing properly.
These test all the Performance Specifications (not Supplementary Characteristics) listed in Table 1-1 of the Power Module User’s Guide.
Note The power module must pass the selftest at power-on before the following tests can be performed. If the
module fails selftest, refer to the overall troubleshooting procedures in Chapter 3.
SHOCK HAZARD The tests should only be performed by qualified personnel. During the performance of these tests, hazardous voltages may be present at the output of the module.

Operation Verification Tests

To assure that the module is operating properly, without testing all specified parameters, perform the following test procedures:
1. Perform the turn-on and checkout procedures given in Chapter 3 of the Power Module User’s Guide.
2. Perform the Voltage Programming/Readback Accuracy and the Current Programming/Readback Accuracy Performance
Tests in this chapter.

Performance Tests

The following paragraphs provide test procedures for verifying the module’s compliance with the specifications listed in Table 1-1 of the Power Module User’s Guide. All of the performance test specifications are listed in the Performance Test Records at the end of this chapter. You can record the actual measured values in the column provided.
Verification and Performance Tests 11
Measurement Techniques
Setup for Most Tests
Most tests are performed at the rear terminals as shown in the following figure. Measure the dc voltage directly at the + S and - S terminals. Set the connector sense switch for local sensing and use adequate wire gauge for load leads as described in Chapter 2 of the Power Module User’s Guide.
Figure 2-1. Performance Test Setup
Many of the test procedures require the use of a variable load capable of dissipating the required power (see Table 1-1). If a variable resistor is used, switches must be used to connect, disconnect, and short the load resistor. For most tests, an electronic load can be used. The electronic load is considerably easier to use than load resistors but some may not be fast enough to test transient recove ry time and may be too noisy for the noise (PARD) tests. Fixed load resistors may be used in place of a variable load, with minor changes to the test procedures in this chapter. Also, if computer controlled test setups are used, the relatively slow (compared to computers and system voltmeters) settling time and slew rates of the power module may have to be taken into account. WAIT statements can be used in the test program if the test system is faster than the module.
Current-Monitoring Resistor
To eliminate output current measurement error caused by voltage drops in the leads and connections, connect the current monitoring resistor between the output and the load as a four-terminal device (see R monitoring leads inside the load lead connections directly at the monitoring points on the resistor element.
in Figure 2-1). Connect the current
M
Programming
Table 2-1 lists the programming and current values for each module. You may program the module from the MPS Keyboard or from a GPIB controller when performing the tests. The test procedures are written assuming that you know how to do either or both. Complete instructions for remote and local programming are given in the module Power Module Programming Guide and Power Module User’s Guide.
12 Verification and Performance Tests
Table 2-1. Power Module Voltage and Current Values
Agilent
Model
66101A 8V 8.190V 16A 16.380A 10V 66102A 20V 20.475V 7.5A 7.678A 24V 66103A 35V 35.831V 4.5A 4.607A 42V 66104A 60V 61.425V 2.5A 2.559A 72V 66105A 120V 122.85V 1.25A 1.280A 144V 66106A 200V 204.75V 0.75A 0.768A 240V
Full-Scale
Voltage
Max. Prog.
Voltage
Full-Scale
Current
Max. Prog.
Current
Max. Prog.
Overvoltage
Constant Voltage (CV) Tests
CV Setup
If more than one meter or a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to avoid mutual coupling effects. For constant voltage dc tests, connect only to + S and - S because the module regulates the output voltage that appears between those terminals, not between the + and - output terminals. Use coaxial cable or shielded 2-wire cable to avoid noise pickup on the test leads.
Voltage Programming/Readback Accuracy
This test verifies that the voltage programming, readback, and front panel display functions are within specifications. Note that the values read back over the GPIB should be identical to those displayed on the keyboard display.
1. Turn off the power module and connect a digital voltmeter between the + S and--S terminals (see Figure 2-1).
2. Turn on the module and program it for zero volts and the maximum programmable current (see Table 2-1) with the load
off.
3. Record the output voltage readings on the DVM and the keyboard display. The readings should be within the limits
specified in the Performance Test Record Tables under CV PROGRAMMING @ 0 VOLTS, for the particular model being tested. The CV annunciator should be on and the output current reading should be approximately zero.
4. Program the output voltage to full scale (see Table 2-1).
5. Record the output voltage readings on the DVM and the keyboard display. The readings should be within the limits
specified in the Performance Test Record Tables under CV PROGRAMMING @ FULL SCALE, for the particular model being tested.
CV Load Effect
This test measures the change in output voltage resulting from a change in output current from full-load to no-load.
1. Turn off the module and connect the output as shown in Figure 2-1 with the DVM connected between the + S and - S
terminals.
2. Turn on the module and program the current to the maximum programmable value and the voltage to the full-scale
value (see Table 2-1).
3. Adjust the load for the full-scale current (see Table 2-1) as indicated on the keyboard display. The front panel CV
annunciator must be on. If it is not, ad just the load so that the output c urrent drops slightly until the annunciator comes on.
4. Record the output voltage reading on the DVM connected to + S and - S.
5. Open the load and again record the DVM voltage reading.
6. The difference between the DVM readings in steps (4) and (5) is the load effect voltage and should not exceed the
value listed in the Performance Test Record Tables under CV LOAD EFFECT, for the model being tested .
Verification and Performance Tests 13
CV Source Effect
This test measures the change in output voltage that results from a change in ac line voltage from the minimum to maximum value within the line voltage specifications.
1. Turn off the module and c onnect the ac power line through a variable-voltage tr ansformer.
2. Connect the output as shown in Figure 2-1 with the DVM connected between the + S and - S terminals. Set the
transformer to nominal line voltage (either 115Vac or 230Vac).
3. Turn on the module and program the current to the maximum programmable value and the output voltage to the
full-scale value (see Table 2-1).
4. Adjust the load for the full-scale current value (see Table 2-1) as indicated on the keyboard display. The front panel CV
annunciator must be on. If it is not, ad just the load so that the output c urrent drops slightly until the annunciator comes on.
5. Adjust the transformer to the LOW line voltage (e.g., 87Vac for a 115Vac nominal input, or 174Vac for a 230Vac
nominal input).
6. Record the output voltage reading on the DVM.
7. Adjust the transformer to the HIGH line voltage (e.g., 132Vac for a 115Vac nominal input, or 250Vac for a 230Vac
nominal input).
8. Record the output voltage reading on the DVM.
9. The difference between the DVM readings in steps (6) and (8) is the source effect voltage and should not exceed the
value listed in the Performance Test Record Tables under CV SOURCE EFFECT, for the model being tested.
CV Noise (PARD)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac voltage superimposed on the dc output voltage. This test measures CV PARD, specified as the rms or peak-to-peak output voltage over the frequency range of 20Hz to 20MHz.
1. Turn off the module and connect the output as shown in Figure 2-1 to an oscilloscope (ac coupled) between the + and--
terminals. Set the oscilloscope’s bandwidth limit to 20MHz (30MHz on the Agilent 54504A) and use an RF tip on the oscilloscope probe.
2. Turn on the module and program the current to the maximum programmable value and the output voltage to the
full-scale value (see Table 2-1).
3. Adjust the load for the full-scale current value (see Table 2-1) as indicated on the keyboard display.
4. The waveform on the oscilloscope should not exceed the peak-to-peak limits in the Performance Test Record Tables
under CV NOISE (PARD), for the model being tested.
5. Disconnect the oscilloscope and connect an ac rms voltmeter in its place. The rms voltage reading should not exceed
the rms limits in the Performance Test Record Tables under CV NOISE (PARD) for the model being tested.
Transient Recovery Time
This test measures the time for the output voltage to recover to within the specified value following a 10% change in the load current.
1. Turn off the module and connect the output as shown in Figure 2-1 with the oscilloscope across the + S and -S
terminals.
2. Turn on the module and program the output voltage to the full-scale value and the current to the maximum
programmable value (see Table 2-1).
3. Set the load to the Constant Current mode and program the load current to 90% of the power module full-scale rated
current.
4. Set the electronic load’s transient generator frequency to 100Hz and its duty cycle to 50%.
5. Program the load’s transient level to the module’s full-scale current value and turn the transient on.
6. Adjust the oscilloscope for a waveform similar to that in Figure 2-2.
7. The output voltage should return to within 100mV of the nominal value in less than 1ms. Check both loading and
14 Verification and Performance Tests
unloading transients by triggering on the positive and negative slope.
Figure 2-2. Transient Response Waveform

Constant Current (CC) Tests

CC Setup
Follow the general setup instructions in the Measurement Techniques paragraph and the specific instructions given in the following paragrap hs.
Current Programming/Readback Accuracy
This test verifies that the current programming and readback are within specification. The accuracy of the current monitoring resistor is determined by the power coefficient (see Table 1-1).
1. Turn off the module and connect the current monitoring resistor directly across the output and a DVM across the
resistor. Refer to the "Current Monitoring Resistor" paragraph.
2. Turn on the module and program the output voltage to 5V and the current to zero.
3. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and
record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits specified in the Performance Test Record Tables under CC PROGRAMMING @ 0 AMPS, for the particular model being tested.
4. Program the output voltage to 5V and the current to full-scale (see Table 2-1).
5. Divide the voltage drop (DVM reading) across the current monitoring resistor by its resistance to convert to amps and
record this value as (Io). The actual current (Io) and the reading on the keyboard display should be within the limits specified in the Performance Test Record Tables under CC PROGRAMMING @ FULL SCALE, for the particular model being tested.
CC Load and Line Regulation
These tests (CC Load Effect and CC Source Effect given below) are tests of the dc regulation of the power module’s output current. To insure that the values read are not the instantaneous measurement of the ac peaks of the output current ripple, several dc measurements should be made and the average of these readings calculated.
Verification and Performance Tests 15
The following steps show how to set up an Agilent 3458A System Voltmeter from its front panel to take a statistical average of 100 readings.
represents the unlabeled shift key in the FUNCTION/RANGE group.
1. Program 10 power line cycles per samp le by pressing
2. Program 100 samples per trigger by pressing
3. Set up the voltmeter to take measurements in the statistical mode as follows:
4. a. Press
b. Press c. Press
5. Now set up the voltmeter to read the average value of the measurements as follows:
6. a. Press
b. Press c. Press
7. Execute the average reading program by pressing
8. Wait for 100 readings and then read the average measurement by pressing
To repeat the measurement, perform steps (7) and (8).
CC Load Effect
This test measures the change in output current for a change in the load from full-scale output voltage to short circuit.
1. Turn off the module and connect the output to be tested as shown in Figure 2-1 with the voltmeter connected across the
2. Turn on the module and program the current to the full-scale current value and the output voltage to the maximum
3. Adjust the load in the CV mode for full-scale voltage as indicated on the keyboard display. Check that the CC
4. Record the output current reading (DVM reading ÷ current monitor resistance value in ohms).
until the MATH function is selected; then press . until the STAT function is selected; then press .
until the RMATH function is selected; then press . until the MEAN function is selected; then press .
current monitoring resistor.
programmable voltage value (see Table 2-1).
annunciator is on. If it is not, adjust the load to drop the output voltage slightly until the annunciator c omes on.
.
.
.
.
.
. Record this as your result.
Note You may want to use the average reading program described previously.
5. Short the load switch and record the output current reading.
6. The difference in the current readings in steps (4) and (5) is the load effect and should not exceed the limit specified in
the Performance Test Record Tables under CC LOAD EFFECT, for the particular model being tested.
CC Source Effect
This test measures the change in output current that results when the ac line voltage changes from the minimum to the maximum value within the specifications.
1. Turn off the module and c onnect the ac power line through a variable-voltage tr ansformer.
2. Connect the output terminals as shown in Figure 2-1 with the voltmeter connected across the current monitoring
resistor. Set the transformer to the nominal line voltage.
3. Turn on the module and program the current to the full-scale value and the output voltage to the maximum
programmable value (see Table 2-1).
4. Adjust the load in the CV mode for full-scale voltage as indicated on the front panel display. Check that the CC
annunciator is on. If it is not, adjust the load to drop the output voltage slightly until the annunciator c omes on.
5. Adjust the transformer to the LOW line voltage (e.g., 87Vac for a 115Vac nominal input, or 174Vac for a 230Vac
nominal input).
6. Record the output current reading (DVM reading ÷ current monitoring resistor in ohms).
16 Verification and Performance Tests
Note You may want to use the average reading program described previously.
7. Adjust the transformer to the HIGH line voltage (e.g., 132Vac for a 115Vac nominal input, or 250Vac for a 230Vac
nominal input), and record the output current reading.
8. The difference in the current readings in steps (6) and (7) is the CC source effect and should not exceed the values
listed in the Performance Test Record Tables under CC SOURCE EFFECT, for the particular model being tested.
CC Noise (PARD)
Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac current as well as an ac voltage superimposed on the dc output. Constant current (CC) PARD is specified as the rms output current in a frequency range 20Hz to 20MHz with the module in CC operation.
1. Turn off the module and connect the load resistor (R
) and rms voltmeter as shown in Figure 2-1. Use only a resistive
L
load for this test. Keep leads as short as possible to reduce noise pickup.
2. Check the test setup for noise with the module turned off. Other equipment (e.g. computers, DMM, etc.) may affect the
reading.
3. Turn on the module and program the current to full-scale and the output voltage to the maximum programmable value
(see Table 2-1).
4. The output current should be at the full-scale rating with the CC Mode on.
5. Divide the reading on the rms voltmeter by the shunt resistance to obtain rms current. It should not exceed the values
listed in the Performance Test Record Tables under CC NOISE (Rms) for the particular unit being tested.
Verification and Performance Tests 17
Table 2-2. Performance Test Record for Model Agilent 66101A (8V, 16A)
Pre Cal_____________Post Cal______________
Model Agilent 66101A Temperature___________________________Test Performed By_______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 1mV
CV SOURCE EFFECT 0.5mV
CV NOISE (PARD) Peak to Peak Rms
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 3mV
DVM ± 2mV
8V ± 5.4mV
DVM ± 3.6mV
5mV 2mV
0 ± 6mA
Io ± 6mA
1.5µV
1.5µV
88µV 88µV
1µV
1µV
0.87mV 150µV
150µA 150µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT 0.5mA
CC SOURCE EFFECT 0.75mA
CC NOISE (RMS) 8mA 0.9mA ________
18 Verification and Performance Tests
16A ± 10.8mA
Io ± 9.2mA
2.1mA
2.1mA
15µA 15µA
________ ________
________
________
Table 2-3. Performance Test Record for Model Agilent 66102A (20V, 7.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66102A Temperature___________________________Test Performed By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 1mV
CV SOURCE EFFECT 0.5mV
CV NOISE (PARD) Peak to Peak Rms
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 8mV
DVM ± 3mV
20V ± 14mV
DVM ± 7mV
7mV 3mV
0 ± 3mA
Io ± 3mA
1.6µV
1.6µV
335µV 335µV
20µV
20µV
0.872mV 500µV
150µA 150µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT 0.2mA
CC SOURCE EFFECT 0.5mA
CC NOISE (RMS) 4mA 0.7mA ________
7.5A ± 5.3mA Io ± 4.5mA
Verification and Performance Tests 19
1mA 1mA
9µA 9µA
________ ________
________
________
Table 2-4. Performance Test Record for Model Agilent 66103A (35V, 4.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66103A Temperature___________________________Test Performed By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 1mV
CV SOURCE EFFECT 1mV
CV NOISE (PARD) Peak to Peak Rms
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 13mV
DVM ± 8mV
35V ± 23.5mV
DVM ± 15mV
10mV
5mV
0 ± 2mA
Io ± 2mA
1.7µV
1.7µV
526µV 526µV
28µV
28µV
0.90mV 500µV
15µA 15µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT 0.2mA
CC SOURCE EFFECT 0.3mA
CC NOISE (RMS) 2mA
20 Verification and Performance Tests
4.5A ± 3.4mA Io ± 2.9mA
0.7mA
0.7mA
3µA 3µA
250µA
________ ________
________
________
________
Table 2-5. Performance Test Record for Model Agilent 66104A (60V, 2.5A)
Pre Cal_____________Post Cal______________
Model Agilent 66104A Temperature___________________________Test Performed By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 2mV
CV SOURCE EFFECT 2mV
CV NOISE (PARD) Peak to Peak Rms
1
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 27mV
DVM ± 16mV
60V ± 45mV
DVM ± 28mV
15mV
9mV
0 ± 0.6mA
Io ± 0.6mA
1.8µV
1.8µV
845µV 845µV
40µV
40µV
1.0mV
1.5mV
15µA 15µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT 0.1mA
CC SOURCE EFFECT 0.1mA
CC NOISE (RMS) 1mA 0.2mA ________
1 18mV Rms when the ac input is between 87 and 104Vac.
2.5A ± 1.4mA Io ± 1.1mA
Verification and Performance Tests 21
323µA 323µA
2µA 2µA
________ ________
________
________
Table 2-6. Performance Test Record for Model Agilent 66105A (120V, 1.25 A)
Pre Cal_____________Post Cal______________
Model Agilent 66105A Temperature___________________________Test Performed By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 4mV
CV SOURCE EFFECT 3mV
CV NOISE (PARD) Peak to Peak Rms
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 54mV
DVM ± 32mV
120V ± 90mV DVM ± 56mV
25mV 18mV
0 ± 0.6mA
Io ± 0.6mA
2.1µV
2.1µV
1.7mV
1.7mV
230µV
230µV
1.0mV
1.5mV
15µA 15µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT
CC SOURCE EFFECT
CC NOISE (RMS) 1mA 0.2mA ________
22 Verification and Performance Tests
1.25A ± 975µA Io ± 850µA
50µA1.4 50µA1.4µA
157µA 157µA
µ
A
________ ________
________
________
Table 2-7. Performance Test Record for Model Agilent 66106A (200V, 0.75A)
Pre Cal_____________Post Cal______________
Model Agilent 66106A Temperature___________________________Test Performed By______________________
Serial Number_______________________Humidity________________________Date__________________________
Test Description Specification Measurement
Uncertainty
CV PROGRAMMING @ 0 volts Output Accuracy (DVM Reading) Display and Readback Accuracy
CV PROGRAMMING @ full-scale Output Accuracy (DVM Reading) Display and Readback Accuracy
CV LOAD EFFECT 7mV
CV SOURCE EFFECT 5mV
CV NOISE (PARD) Peak to Peak Rms
TRANSIENT RECOVERY in 1 ms 100mV 15mV ________
CC PROGRAMMING @ 0 amps Output Accuracy (Io) Display and Readback Accuracy
0 ± 90mV
DVM ± 54mV
200V ± 150mV
DVM ± 94mV
50mV 30mV
0 ± 0.4mA
Io ± 0.3mA
2.6µV
2.6µV
2.7mV
2.7mV
350µV
350µV
1.5mV
5.0mV
15µA 15µA
Actual
________ ________
________ ________
________
________
________ ________
________ ________
CC PROGRAMMING @ full-scale Output Accuracy (Io) Display and Readback Accuracy
CC LOAD EFFECT
CC SOURCE EFFECT
CC NOISE (RMS) 1mA 0.2mA ________
0.75A ± 625µA Io ± 450µA
10µA1 30µA1µA
Verification and Performance Tests 23
100µA 100µA
µ
A
________ ________
________
________
3

Troubleshooting

This chapter provides troubleshooting and repair information for the Agilent 6610xA Power Modules. Before attempting to troubleshoot the modules, ensure that the problem is not with an external circuit or application, the mainframe, the GPIB controller, the application program, or the input power line.
SHOCK HAZARD Most of the troubleshooting procedures given in this chapter are performed with power applied and protective covers removed. Such maintenance should be performed only by trained
service personnel who are aware of the hazards (for example, tire and electrical shock).
Hazardous voltages are present throughout the P ower and Bias board assemblies. Ac line voltage is present on some connectors even when the module is not turned on. The dc rail voltage to the module is 310 volts when the unit is turned on.

Troubleshooting Sequence

1. Verify that the module is at fault.
2. Use the flowcharts to locate the functional block at fault.
3. Troubleshoot the functional block per flowchart instructions.
Selftest routines built into firmware are provided to help isolate a problem to a particular circuit on the board. Once a problem has been isolate d to a circuit, suggestions are given in the appropriate flowchart as to what component may be at fault.
This instrument uses components that can be damaged or suffer serious performance degradation as a result of ESD (electrostatic discharge). Observe the standard anti-static precautions discussed in
Chapter 1 to avoid damage to components.

Accessing the Flowcharts

The following troubleshooting flowcharts are provided to help you identify and isolate a problem to a specific circuit. Once a problem has been narrowed down, suggestions are provided as to the components that may be responsible fo r the problem. Refer to the schematic diagrams for detailed information on circuit wiring and component function.
þ Figure 3-1 - Overall Troubleshooting þ Figure 3-2 - Troubleshooting Microprocessor Circuits þ Figure 3-3 - Troubleshooting Error Messages þ Figure 3-4 - Troubleshooting Overvoltage at Turn-on
Troubleshooting 25
þ Figure 3-5 - Troubleshooting No Output þ Figure 3-6 - Troubleshooting High Output þ Figure 3-7 - Troubleshooting W ill Not Overvoltage þ Figure 3-8 - Troubleshooting CV Accuracy þ Figure 3-9 - Troubleshooting CC Accuracy þ Figure 3-10 - Troubleshooting Downprogrammer þ Figure 3-11 - Troubleshooting DAC Circuit þ Figure 3-12 - Troubleshooting Readback Circuits
Oscilloscope waveforms are provided on some flowcharts along with the settings required to generate the waveforms.
Connect the module to ac mains through an isolation transformer when making measurements with an oscilloscope or any equipment that has the measuring input grounded.
If You Experienced a Selftest Failure or RunTime Error Message
If you experienced a selftest failure or runtime error message as described in Chapter 3 of the Power Module User’s Guide, disassemble the module and proceed directly to the flowchart in Figure 3-3 "Troubleshooting Error Messages".
If You Experienced Other Problems With the Module
If you experienced problems other than a selftest failure or a runtime error message, disassemble the module and proceed directly to Figure 3-1 "Overall Troubleshooting".
26 Troubleshooting
Figure 3-1. Overall Troubleshooting (Sheet 1 of 3)
Troubleshooting 27
28 Troubleshooting
Figure 3-1. Overall Troubleshooting (Sheet 2 of 3)
Figure 3-1. Overall Troubleshooting (Sheet 3 of 3)
Troubleshooting 29
Figure 3-2. Troubleshooting Microprocessor Circuits (Sheet 1 of 2)
30 Troubleshooting
Loading...
+ 71 hidden pages